• The InstaTrak® 3500 PlusSurgical Navigation System–Advanced ENT
Application is intended as an aid to the surgeon to precisely locate anatomical
structures in the human body, and to precisely locate the tip and trajectory of
surgical instrumentation in reference to those anatomical structures.
Indications for Use
•The InstaTrak® 3500 Plus–Advanced ENT Application is indicated for use in
surgical procedures that require precise location of surgical instrumentation with
reference to rigid anatomical structures visible on medical images.
Contraindications
• The InstaTrak® 3500 PlusAdvanced ENT Application should not be used for
any non-ENT or non-skull-base surgical procedures.
• The InstaTrak® 3500 PlusAdvanced ENT Application is contraindicated for
skeletally immature patients.
• The InstaTrak® 3500 PlusAdvanced ENT Application should not be used with
patients suspected of having Creutzfeld-Jacob disease. Sterilization of cables
cannot be guaranteed.
•For instruments used with the InstaTrak® 3500 Plus Advanced ENT
Application, follow the specific contraindications for that device.
Warnings and Precautions
• The InstaTrak®3500 Plus Advanced ENT Application provides information from
medical imaging that enhances the information provided to the surgeon. The
InstaTrak® 3500 Advanced ENT Application should be used only as an adjunct
for surgical guidance. It is not a replacement for surgical judgment, expertise, or
knowledge of the anatomy. The InstaTrak® 3500 Plus Advanced ENT Application is for use by qualified surgeons only and should not be used by any
person not qualified to use this device.
Page 1-2
• Clinical tracking accuracy is subject to many factors. Navigation should be used
only by personnel trained in the use of navigation and experienced with the
Page 13
INTRODUCTION
anatomy on which navigation will be used. It is the responsibility of the user to
verify tracking accuracy during the procedure to ensure the navigation system is
performing adequately. The user should plan the procedure so that it can be
safely completed without the use of navigation.
Caution: U.S. Federal law restricts this device to sale by or on the order of a
physician.
• The safety and effectiveness of the InstaTrak® 3500 Advanced ENT Application
has not been validated for use on patients who have electronic devices that make
an internal, direct connection to the cardiovascular system, brain or central
nervous system.
• Only persons who have attended the in-service training courses conducted by GE
Healthcare Technologies Clinical Training Specialists, should attempt to operate
this equipment.
• GE Healthcare Technologies also strongly recommends that radiologists and
scanner technologists acquaint themselves with the intra-operative use of the
InstaTrak® 3500 Plus by reading this manual and attending surgery. Radiology
and OR personnel are both critical components of the InstaTrak® 3500 Plus
Advanced ENT Application team.
•This manual covers the use of the InstaTrak® 3500 Plus Advanced ENT
Application and should be used in conjunction with the InstaTrak® 3500 Plus
Operator’s Manual. The InstaTrak® 3500 Plus Operator’s Manual contains
instructions and information for the basic system platform. Information about
any of the features and functions that are common to all of the applications can
be found in the InstaTrak® 3500 Plus Operator’s Manual.
• Do not connect any equipment to the system, which is not specifically described
in this manual. Whenever AC powered equipment is connected to the system, the
user must confirm that applicable Standards for leakage current are met by the
interconnected system any any applied parts.
• Please follow the warnings and cautions embedded in individual sections of this
manual as appropriate. For instruments used with the InstaTrak
®
3500 Plus,
follow the specific warnings and precautions for that device.
Note: The screen images in this manual are for representation purposes only. The actual
images on your display might vary.
Note: No natural latex is used in the manufacture of any of GE Healthcare Technologies
products or packaging used for the InstaTrak
®
3500 Plus system.
Note: This device has been tested for electromagnetic emission and susceptibility in
accordance with IEC 60601-1-2.
Page 1-3
Page 14
INTRODUCTION
The InstaTrak 3500® Plus
System Components
The components of the InstaTrak® 3500 Plus include
• A high-powered computer,
• A high-resolution touch-screen display,
• A keyboard/mouse, and
• A tracking system.
The patient’s CT or MRI images are acquired and sent to the InstaTrak® 3500 Plus prior
to surgery. The scan is used during the surgical navigation procedure.
The keyboard is used to enter patient identifiers and text. The touch-screen or mouse is
used to interact with the display.
The tracking system is used to track the position of the surgeon’s instruments.
Note The InstaTrak® 3500 Plus uses a specific tracking system. If your System is not
equipped with the correct tracking system, the following message appears after
attempting to enter the Visualization Display for surgical navigation.
Figure 1-1. Incompatible Tracker Message
If this message is seen, the InstaTrak 3500 Plus cannot be used for surgical navigation.
Contact your GE Healthcare Technologies Representative.
Page 1-4
Page 15
INTRODUCTION
HIPAA Privacy Rule Compliance
The InstaTrak® Plus Surgical Navigation System is compliant with the Privacy Rule, a
Federal regulation under the Health Insurance Portability and Accountability Act
(HIPAA) of 1996 protecting the privacy of individually identifiable health information.
Users of the InstaTrak® Plus Surgical Navigation System have the option of assigning a
password before system operation and to de-identify patient data before copying.
Note: To protect sensitive healthcare information, GE Healthcare Technologies strongly
recommends requiring a password to operate the system and de-identifying patient data
before transferring or copying to a CD or floppy disk.
Page 1-5
Page 16
Page 17
2
P
RE-OPERATIVE
Page 18
PRE-OPERATIVE
Scanner Compatibility
The Advanced ENT Application of the InstaTrak® 3500 Plus is designed to interface
with any CT or MRI scanner that can output data in the DICOM 3.0 storage class format.
Data Transfer
Before the InstaTrak® 3500 Plus may be used in surgery, a CT or MRI scan must be
transferred to the system.
If the scanner has the capability to archive DICOM data onto a CD, then the scans can be
downloaded onto a CD directly from the scanner or Radiology workstation and then
loaded onto the InstaTrak® 3500 Plus.
The scans also can be transferred to the system via an Ethernet network. Connectivity
must be established between the scanner or Radiology workstation and the InstaTrak® 3500 Plus or ConneCTstat™ Plus in order to transfer scans in this manner.
There are some basic prerequisites for successful connectivity:
The system uses the DICOM (Digital Imaging and Communications in Medicine) 3.0
standard for network based image transfers. This standard was developed jointly by the
American College of Radiology (ACR) and National Electrical Manufacturers
Association (NEMA) to facilitate data transfer between medical equipment. Many
scanners support this standard by providing a storage or transfer facility to a DICOMcapable device, either directly from the scanner or by means of a workstation. In some
cases, older scanners can be made to output DICOM data through the use of third party
equipment.
Data may be transferred to either the InstaTrak® 3500 Plus or the ConneCTstat™ Plus
through an Ethernet network connection. The Ethernet port is a twisted pair RJ-45
(10/100 Base-T) connector (no ground connection) located on the back panel of the
InstaTrak 3500 Plus. The Ethernet capability conforms to the ISO 802.3 standard.
Radiology
Surgical Navigation Systems use information from medical images to aid doctors during
surgery. Depending on the surgical procedure, a CT or MRI is obtained using the
scanning protocols provided in this Operator’s Manual. The axial scan is transferred to
the system, where the images are automatically reformatted into coronal, sagittal and
three-dimensional views.
Connectivity is the process of establishing communication between the scanner and the
InstaTrak® 3500 Plus. A GE Healthcare Field Service Engineer works with your
network administrator to establish a method for transferring the medical images to the
system. Refer to the Service section in the InstaTrak® 3500 Plus Operator’s Manual for
more information about connectivity.
Page 2-2
Page 19
PRE-OPERATIVE
Transferring Scans To the InstaTrak
3500 Plus
There are a number of options for transferring scans to the system:
A ConneCTstat™ Plus can be placed in a Radiology Department and connected to
your network. Images are transferred to the ConneCTstat™ Plus and sent to a CD,
which can then be inserted into the InstaTrak® 3500 Plus. This enables an outside
Radiology group to perform the medical imaging that will be used during surgery.
The InstaTrak® 3500 Plus can be directly connected to the hospital network
enabling the scans to be transferred from the scanner to the system, which is
connected to the network in the OR. It is also possible to transfer scans between the
InstaTrak® Plus and the ConneCTstat™ Plus via a network connection.
The InstaTrak® 3500 Plus can be transported to the Radiology Department and
hooked up to a network connection. Scans are transferred directly from the scanner
to the system.
If the scanner has the capability to archive DICOM data onto a CD, then the scans
may be downloaded onto a CD directly from the scanner or Radiology workstation
and then loaded onto the InstaTrak® 3500 Plus.
The medical images used for surgical navigation must meet specific criteria. GE
Healthcare has developed protocols that must be followed in order to ensure high quality
images for use with the InstaTrak® 3500 Plus. The GE Healthcare Clinical Specialist
will work with your CT and/or MRI technologists to ensure that the protocols are
understood and followed.
Scanning Protocols
The scanning protocols are designed to meet the following goals:
• Maximize system accuracy during surgery
• Obtain Automatic Registration, when applicable
• Minimize scan time
• Minimize patient radiation exposure
• Optimize image quality
Page 2-3
Page 20
PRE-OPERATIVE
Registration
Registration is the process of aligning the patient scan with the actual physical features of
the patient. For a more detailed description of registration, refer to System Operation in
the InstaTrak 3500 Plus Operator’s Manual.
A valid registration is necessary in order to use the system for surgical navigation.
Methods
Several different registration options are available with the Advanced ENT Application
of the InstaTrak® 3500 Plus:
• Automatic Headset registration - The headset is worn during the scan.
• Auto Plus registration – The headset is worn during the scan.
• Fiducial registration - Fiducial markers are placed on the patient before the
scan.
Fast Fiducial ™ is a computer algorithm to determine the position
and center of the fiducial marker
•AccuMatch™ Surface Registration (if system is equipped) - No markers are
placed on the patient.
Figure 2-1. Scanning/Surgery
Registration Methods and Scan Compatibility
•CT: All registration methods may be used
Page 2-4
Page 21
PRE-OPERATIVE
• MRI: Fiducial or AccuMatch™ registration only.
Note: The headset contains embedded metal markers and may not be worn during an MRI
scan.
Page 2-5
Page 22
PRE-OPERATIVE
Automatic Registration Headset
Headset Selection Guidelines
• For patients 14 years or older, use the Adult Headset
• For pediatric patients under the age of 14, head circumference must be
measured to determine the appropriate Pediatric Headset size.
• Measure the head circumference directly above the eyebrows used the tape
measure provided in each headset box.
Headset Description Reorder # Patient Head Circumference
Adult 100497 Adults and Children age 14 or over Greater than 57 cm
Pediatric Large 100948 Large Children Greater than 52 cm
Pediatric Small 100949 Small Children Less than 52 cm
Table 2-1. Headset Selection Guidelines
Note: Patients, 14 years or older with a head circumference less than 57 cm, may be better
suited for the Pediatric Large Headset. Pediatric patients, less than 14 years with a head
circumference greater than 57 cm, may be better suited for using the Adult Headset. We
recommend that the surgeon fit the Headset to the patient to ensure proper fit.
WARNING
Some patients may experience discomfort when the Headset is worn for an
extended period of time. Such discomfort should be temporary. Patients with
TMJ, impaired circulation, or other conditions that may be aggravated by use of
the Headset, should be given special consideration. Temporary pain or
temporary loss of sensation may be experienced with use of the Headset.
CAUTION
The use of the Headset is not recommended on patients less than three (3) years
of age unless a proper fit can be determined.
WARNING
The Headset is a single-use patient item. Using the same Headset on another
patient can result in system inaccuracy due to changes in Headset geometry
leading to a patient safety hazard.
CAUTION
Prior to scanning, ensure the small metal markers embedded in front and side of
the Headset are present. If any markers are missing, do not use the Headset.
Automatic registration is not possible if the markers are missing or are not
present on the CT scan.
Page 2-6
Page 23
PRE-OPERATIVE
CAUTION
To maximize system and tracking accuracy, the Headset used during the CT
scan should be used in surgery. If a surgical fixation device, such as a Mayfield,
will be used, ensure that the Headset is positioned exactly as in the CT scanner.
Page 2-7
Page 24
PRE-OPERATIVE
Verification Pad Placement
2 adult verification pads are provided in each Adult Headset box. 2 pediatric verification
pads are provided in each Pediatric Small Headset box. 2 adult and 2 pediatric
verification pads are provided in each Pediatric Large Headset box. The clinician must
select the appropriate Verification Pad for each patient.
The verification pad must be worn during the CT scan for patient comfort and to ensure
proper fit of the Headset. Place the verification pad on the patient’s nose by aligning the
upper center curve of the verification pad with the bridge of the patient’s nose.
Figure 2-2. Verification Pad Placement
Verification Pad
WARNING
The Verification Pad provided with the headset must be used to increase patient
comfort and to reduce the possibility of bruising of the skin under the nosepiece
of the Headset.
Page 2-8
Page 25
Headset Placement
Select the appropriate Adult, Large Pediatric or Pediatric Headset. Gently set the
nosepiece on the bridge of nose and place each earpiece into the external ear canals.
Figure 2-3. Headset Placement
PRE-OPERATIVE
Tragus
WARNING
Store Headset only in the supplied box. Ensure that the Headset is positioned
properly in the box to maintain its shape. Failure to do so can result in system
inaccuracy due to changes in Headset geometry.
WARNING
The Headset is a single-use patient item. Using the same Headset on another
patient can result in system inaccuracy due to changes in Headset geometry
leading to a patient safety hazard.
Fiducial markers are used to correlate the scanned images with the patient. The fiducial
markers must be placed on the patient before the scan and must not be moved until after
the registration process in surgery is complete.
For best results, wipe the areas on which the fiducials are to be placed with alcohol to
remove oil. Place the fiducial markers on the patient according to the placement below.
Note: To ensure registration accuracy, once the fiducial markers are placed they cannot be
moved or replaced until the registration process is complete.
CAUTION
If a fiducial marker falls off, it must not be replaced. Replacing a fiducial marker
may result in registration inaccuracy.
Page 2-9
Page 26
PRE-OPERATIVE
Fiducial Marker Placement Guidelines
• Place more fiducial markers than necessary in case one or more is displaced.
A minimum of four (4) fiducial markers is required for a relevant
registration.
• Surround the operative area and do not place too close together
• Apply to areas where the skin is relatively immobile.
• Apply to bony areas. Avoid areas where skin is loose.
• Apply within the medical images scanning range only.
• Do not apply to areas of trauma as swelling may displace their position.
• Do not apply fiducials in a straight line
For best results, GE recommends fiducial marker placement on the following areas:
• Above the eyebrow at the hairline
• Lateral forehead
• Vertex, slightly off the midline
• Asterion (junction of parietal, occipital and temporal bones)
• Above and behind the ear
Figure 2-4. Fiducial Marker Placement
CT Scanning Protocol
The following table provides CT scanner guidelines to obtain the best results for use with
the InstaTrak® 3500 Plus.
Keep all parameters consistent throughout the scan.
Page 2-10
Scanning Plane
Patient Position
Axial
Supine
Page 27
PRE-OPERATIVE
Gantry Angle
Matrix Size 512
Slice Thickness 1-3mm
Scanning mode Helical
Table increment 1mm
Algorithm Standard brain
Table 2-2. Recommended CT Scanner Protocol
Zero
CAUTION
Scanner settings listed in bold print indicate that no changes are allowed. All
others may be set at the discretion of the user.
Patient Preparation
Remove all metal from the patient’s head
Place the fiducial markers or the Automatic Headset on the patient, as applicable
Be certain to place the Verification Pad on the patient as directed before placing the
Headset
Fiducial markers or the Headset are not required if AccuMatch™ Surface
Registration will be used
Position the patient on the scanner table for a supine axial scan.
MRI Scanning Protocol
The following table provides MRI scanner setting-guidelines to obtain the best results for
use with the InstaTrak® 3500 Plus.
Scanning Plane
Patient Position
Data Acquisition
Matrix Size 256
Slice Thickness 1.5mm
Axial
Supine
3D Volumetric
Table 2-3. Recommended MRI Protocol
Page 2-11
Page 28
PRE-OPERATIVE
CAUTION
Scanner settings in bold print indicate that no changes are allowed. All others
may be set at the discretion of the user.
Patient Preparation
• Remove all metal from the patient
• Place fiducial markers on the patient, if applicable
• Position the patient on the scanner table in the supine position
Scanning
• Use a protocol for T1 3D volumetric data or 3D reconstructions
• Scan using 1.5mm slice thickness with contiguous, non-overlapping slices
Images must be acquired in the axial plane.
− If the abnormal area is not clearly visualized, then the protocol should be
adjusted to use T2.
• Make certain that all of the requested anatomy appears on the scan
• If fiducial markers are used, all markers must appear on the scan
• If no markers were placed on the patient, the nose, ears, top of the head and
the back of the head must appear on the scan.
Note: The GE Healthcare Automatic Headset cannot be used during an MRI scan. If using a
MRI scan, fiducial or AccuMatch™ Surface Registration must be performed.
Page 2-12
Page 29
GETTING STARTED
3
Page 30
GETTING STARTED
HIPAA “Privacy Rule” Compliance
The InstaTrak® Plus Surgical Navigation System is compliant with the Privacy Rule, a
Federal regulation under the Health Insurance Portability and Accountability Act
(HIPAA) of 1996 protecting the privacy of individually identifiable health information.
Users of the InstaTrak® Plus Surgical Navigation System have the option of assigning a
password before system operation and to de-identify patient data before copying.
Note: To protect sensitive healthcare information, GE Healthcare Technologies strongly
recommends requiring a password to operate the system and de-identifying patient data
before transferring or copying to a CD or floppy disk
Page 3-2
Page 31
GETTING STARTED
Transferring Data
Scanner Transfer
The following instructions apply to transferring image data from either a CT or an MRI
scanner to either the InstaTrak® 3500 Plus or the ConneCTstat™ Plus. Refer to Service
in the InstaTrak® 3500 PlusOperator’s Manual for more details about connectivity.
Note: The System should be disconnected from the network connection before using in
surgery.
• Plug the system into a hospital-grade electrical outlet.
• Plug the network connector into the receptacle marked “Ethernet Port” on the
back of the system.
• Turn on the system.
• Wait for the MAIN MENU to appear.
CAUTION
Each scan requires approximately one hundred twenty-five (125) images of free
disk space. If disk space is unavailable, older scans must be deleted. See
System Operation in the InstaTrak® 3500 Plus Operator’s Manual for instructions
on how to delete images.
Network Transfer
After scanning is done and all reconstructions are complete, initiate the transfer between
the scanner and the InstaTrak® 3500 Plus. A menu appears showing the transfer status.
To end the transfer of a scan, select an entry while it is being transferred and then select
the Cancel button. Any scan that has already been transferred must be deleted manually
from the PATIENT DATABASE MENU.
Figure 3-1. Network Transfer Menu
Page 3-3
Page 32
GETTING STARTED
Transferring Between Systems
A scan may be selected and transferred between InstaTrak systems via an Ethernet
network. Before beginning the transfer, ensure that both the InstaTrak® 3500 Plus or the
ConneCTstat™ Plus are connected to the hospital network and that both systems have
the same AE (Application Entity) Title.
Note: Only one scan may be selected and transferred at a time.
To transfer a scan between systems
• Plug in both systems and turn the power on.
• Select the PATIENT DATABASE button. The PATIENT DATABASE
MENU appears.
• Highlight the correct scan.
• Select the Send Data To…. button from the PATIENT DATABASE
MENU. The TRANSFER DATA MENU appears.
• Select the Network button from the TRANSFER DATA MENU. All the
DICOM Data Providers that are listed on the DICOM CONFIGURATION
MENU appear.
• Select the Host Name assigned to thereceivingsystem.
• Select the Send button.
• A message appears to confirm the transfer. Select Yes to continue or No to
cancel the process.
Page 3-4
Page 33
Selected
Patient
Scans
De-Identify
Button
GETTING STARTED
Transfer Data
Task Window
Send Data To…
Button
Eject Button
Figure 3-2. Transfer Data Between Systems
Page 3-5
Page 34
GETTING STARTED
Transferring Using a CD
If a ConneCTstat™ Plus is in the Radiology Department, the CT or MRI scans can be
placed on a blank CD-R for transfer to the InstaTrak® 3500 Plus.
To load the CT or MRI scan for viewing or navigation purposes, the scan must first be
transferred to the system Hard Drive. Scans cannot be directly loaded from the CD.
Scans may be downloaded onto a CD and viewed on a PC. Refer to the InstaTrak® 3500 Plus Operator’s manual for more information on the DICOM CD Viewer.
Note: The InstaTrak® 3500 Plus or ConneCTstat™ Plus uses a CD-R that can be written to
one time only.
Note: Before copying patient data to a CD, the patient’s name can be de-identified. GE
Healthcare strongly recommends the de-identification of patient data before copying.
• There are three ways to open the CD drawer:
− The drawer will open on command when certain message windows are
selected.
− The button is pressed on the CD Drive.
− The Eject button is s elected from the PATIENT DATABASE MENU.
• There are two ways to close the CD drawer:
Hard Drive
Button
CD Drive
Button
− The drawer will close on command.
− The CD drawer is gently pushed to activate closing. Do not force the CD
drawer.
Page 3-6
Figure 3-3. Transferring Data Using A CD
Page 35
GETTING STARTED
Transferring Scans: CD to Hard Drive
• Open the CD drawer and place the CD containing the scaninto the drive.
• Close the CD drawer.
• Select the CD Drive button from the PATIENT DATABASE MENU.
• Highlight the desired scan (s). Multiple scans may be copied at the same time
as long as all scans are listed consecutively in the PATIENT DATABASE
MENU.
• Select the Send Data To… button. The TRANSFER DATA MENU appears.
Hard Drive is selected automatically.
• Select Send.
• A message window appears. Select Yes to confirm or No to cancel.
After the transfer process is complete, select the Hard Drive button to confirm that the
scan was successfully transferred.
Hard Drive to CD
From the PATIENT DATABASE MENU, select the desired scan (s). More than one scan
can be copied at one time. Snapshots appearing in the database can be included in a
series.
• If the scans are listed consecutively, highlight one and drag the cursor to the
next until all scans or snapshots to be copied are highlighted.
• If the scans are not listed consecutively, and a keyboard is attached to the
system, highlight a scan, then select and hold down the Control button on
the keyboard. Highlight the other scans or snapshots until all items to be
copied are highlighted.
• Select the Send Data To…. button from the PATIENT DATABASE
MENU. The TRANSFER DATA task window appears. CD Drive is selected
automatically.
Note: To remove the patient name from the copied data, select the De-Identify Data button.
The De-Identify task window appears allowing you to select a new identifier for the
patient.
• Select the Send button.
• Select Yes to confirm or No to cancel the copying process.
• If Yes is selected, the CD drawer opens automatically.
• If a blank disk is inserted in the CD drive, select Continue. The CD drawer
will automatically close
• Select Cancel to cancel the copy process
• Note the amount of data to be copied and verify that there is sufficient space
on the CD. You must use a blank CD-R.
Page 3-7
Page 36
GETTING STARTED
.
Figure 3-4. Sufficient CD Space Message
If the highlighted series of scans and snapshots are correct, select Yes. The CD
copying process begins automatically
Figure 3-5. Confirm CD Copy Message
When the transfer process is complete, the CD Complete message appears. Select
OK.
Figure 3-6. CD Complete Message
When the CD copying process is complete and the OK button is selected, the CD drawer
opens automatically.
To confirm that the scan has been copied onto the CD, close the drawer and select the CD Drive button on the PATIENT DATABASE MENU. The copied scans and snapshots
appear on the PATIENT DATABASE MENU. To remove the CD, select the Eject
button.
Page 3-8
Page 37
GETTING STARTED
Auto Registration Verification
Figure 3-7. Auto Registration Column
If the patient’s CT scan was performed using a Headset, an automatic registration may be
used. To verify that automatic registration may be used:
• From the MAIN MENU, select the Patient Database button. The PATIENT
DATABASE MENU appears.
− The Auto Reg column on the far right of the PATIENT DATABASE
MENU provides information on the status of Automatic HeadsetRegistration.
− If Yes appears in the Auto Reg column, then Automatic Headset
Registration can be used.
−If No appears in the Auto Reg column, either the Automatic Headset
Registration was not successful, the Headset was not used during the
patient scan, or the embedded fiducial markers were not present in the
scan volume.
Note: If an Automatic Headset Registration is not possible as indicated by a No in the Auto
Reg column, an AccuMatch™ Surface Registration, automatic fiducial registration (if at
least 5 marker balls are present on the patient scan), or fiducial registration (if patient
was scanned with fiducial markers) may be performed.
Page 3-9
Page 38
GETTING STARTED
CAUTION
If dashes appear or anything other than Yes or No is listed in the Auto Reg
column, the scan was not done to protocol and cannot be used in surgery with
the InstaTrak 3500 Plus.
Note: The automatic registration status will not appear on the CD. The Auto Reg column for
the patient scan will display “- - -“. To verify automatic registration, check the status of
the scan on the hard drive. When the scan is transferred to the Hard Drive, postprocessing will reinitiate and the correct auto registration will reappear.
Page 3-10
Page 39
4
OPERATING ROOM
Page 40
OPERATING ROOM
Operating Room Setup
HIPAA “Privacy Rule” Compliance
The InstaTrak® Plus Surgical Navigation System is compliant with the Privacy Rule, a
Federal regulation under the Health Insurance Portability and Accountability Act
(HIPAA) of 1996 protecting the privacy of individually identifiable health information.
Users of the InstaTrak® Plus Surgical Navigation System have the option of assigning a
password before system operation and to de-identify patient data before copying.
Note: To protect sensitive healthcare information, GE Medical System strongly recommends
requiring a password to operate the system and de-identifying patient data before
transferring or copying to a CD or floppy disk
Logon and Logout
Since the system contains sensitive patient data, you have the option of requiring users to
Logon to the system and to Logout when system use is complete. The Logon process
requires a password (default or selected by you) when the system is powered on. When
finished using the System, you have the option to logout or shutdown. The System
shutdown and logout process is covered in the Post-Operative section of this Operator’s Manual and in the InstaTrak 3500 Plus Operator’s Manual.
Password Protection
Upon installation, the InstaTrak® Plus Surgical Navigation System is protected with a
default password. The password is required to operate the system. If desired, you can
change or eliminate the password. Since the system contains sensitive patient
information, GE Healthcare strongly recommends that the password requirement not be
eliminated.
Powering and Logging On To The System
The system should be positioned in the Operating Room so that the surgeon can easily
see and reach the display. If the operative area is crowded, the articulating arm can be
extended. If you want to interact with the display during the surgical procedure, a
sterilizable stylus is supplied or the display can be covered with a sterile drape.
• Plug the system into a grounded hospital grade outlet. Disconnect the system
from the network connection, if applicable.
• Turn on the power switch (bottom center of unit). After a short period of
time, the Logon menu will appear:
Page 4-2
Page 41
OPERATING ROOM
Figure 4-1. Log-on Menu
• The Logon Name is genav and appears automatically.
• Enter the password. The default password is tmp123 but may have been
changed by your System Administrator or Biomedical Department.
Note: If you do not know the password, contact your hospital’s System Administrator or
Biomedical Department.
•When the correct password is entered, the MAIN MENU appears.
Page 4-3
Page 42
OPERATING ROOM
Changing The Password
If you want to change the password:
• Select the Change Password button. The CHANGE PASSWORD TASK
WINDOW appears:
Figure 4-2. Change Password Task Window
Note:To change the password, the current password must be entered. The default password is
tmp123. If you do not know the password, contact your hospital’s System
Administrator or Biomedical Department.
• Enter the Logon Name: genav, and enter the current and new password.
When completed, select OK.
• Enter the new password into the LOGIN TASK WINDOW
• Select the Login button and the MAIN MENU will appear.
Page 4-4
Page 43
OPERATING ROOM
Disabling the Logon Requirement
You have the option to eliminate the Logon-with-password requirement and may choose
to retain the patient’s name on copied scans and snapshots. To choose these options:
•Select System Administration from the MAIN MENU
System Administration
Button
Figure 4-3. System Administration Button
•Select User Options. The USER OPTIONS TASK WINDOW appears.
Page 4-5
Page 44
OPERATING ROOM
User Options
Task Window
User Options
Button
Figure 4-4. User Options
• Select Remove User Login/Logout.
Note: GE Healthcare strongly recommends requiring users to log in and out of the system with
a password and not automatically displaying the patient’s name on snapshots
Page 4-6
Page 45
OPERATING ROOM
Copying Sensitive Patient Data
Patient snapshots can be copied to a floppy disk or CD. Unless the patient identifiers are
removed, the snapshots contain the patient name. The patient identifier can be replaced.
Copying snapshots to a Floppy Disk or CD
• Select the snapshot images to copy from the PATIENT DATABASE MENU
• Select the View Snapshots button or double-click on the patient’s name. The
IMAGE PREVIEW MENU appears:
De-Identify
Button
Copy Images to
Floppy or CD
Buttons
Border Around Snapshot
Figure 4-5. Image Preview Menu
• Select anywhere on the image by touching the screen or clicking with the left
mouse button. The border of the image turns green meaning that the image is
selected.
• Select Next or Previous to view and choose individual images or Select All
to copy the entire series.
Before copying the images, you have the option to remove the patient identifier
on the copied images. To remove the patient identifier, select De-Identify Data.
Page 4-7
Page 46
OPERATING ROOM
De-Identify Data
Task Window
Figure 4-6. De-Identify Data Window
The DE-IDENTIFY DATA WINDOW displays the Original Name and a New
Name consisting of pre-selected letters and numbers. If desired, input a New
Name.
• When completed, select Continue.
• To continue copying, select Copy Images to Floppy.
Page 4-8
Page 47
OPERATING ROOM
ElectroMagnetic Tracking Guidelines
The InstaTrak™ 3500 Plus utilizes an electromagnetic tracking system. For best results,
you must be aware of the amount of metal in the operative area and try to eliminate
ferrous metal placed between the transmitter and receiver.
Standard operating room equipment and techniques should be used along with the
following recommended guidelines for reducing the amount of ferrous metal in the
operative area.
• Select a radiolucent OR table whenever possible. If using a metal OR table,
place an extra mattress pad on the table.
• Use plastic clips for securing surgical drapes
• Place intubation tube holder and Mayo stand at least 12 inches away from the
operative area
• Position the InstaTrak® 3500 Plus so that it is at least 3 feet away from the
transmitter
• When metal instruments such as a speculum or retractors are required, ensure
that they are non-ferrous. Non-ferrous instruments are supplied by GE
Healthcare. For more information, see your GE Healthcare Technologies
Representative.
• The system displays a message if the field is distorted or excessive metal is
detected.
WARNING
Do not ignore any field distortion messages. When the field is distorted or
excessive metal is detected, accuracy can be affected that can lead to a patient
hazard.
WARNING
When instruments, such as mouth gag, nasal speculum or retractors, are used
ensure they are made from low content metal or non-magnetic materials. These
instruments may be obtained from GE Healthcare. Failure to use non-magnetic
instruments may decrease accuracy or result in an inoperable system.
CAUTION
Notify GE Healthcare if any red warning message occurs repeatedly without an
apparent cause or cannot be eliminated by the user.
Page 4-9
Page 48
OPERATING ROOM
Standard ENT Equipment and Instruments
The following lists equipment and instruments generally used for ENT and skull-base
surgical navigation procedures. The list should be used as a guide only.
Figure 4-7. Aspirators
• Headset, Adult or Pediatric
• Aspirator(s):
− 7 F or 10 F
− Extended Straight
− 15 Degree, 45 Degree, 90 Degree
• Debrider Attachment
• Button Probe, non-sterile, if AccuMatch™ surface registration is desired
• Transmitter
• One or Two Receivers
Note: GE Healthcare Technologies strongly recommends sterilizing the receivers before use.
Page 4-10
Page 49
Transsphenoidal Procedures
• Automatic Headset, Adult or Pediatric
• Verification Pad, Adult or Pediatric
• Extended Straight Aspirator
• 3M 1010 Steri-Drapes or other clear plastic drape for the Headset
• Button Probe, non-sterile, if AccuMatch™ surface registration is desired
• Nasal Speculum, non-ferrous metal
Skull-base Procedures
• Axcess™ System for unrestricted surgical approach. For procedures when
the use of the headset is not possible. Refer to the Axcess System Operator’s Manual for complete details.
• Non-sterile Pointer or Button Probe for registration process
• Nasal Speculum (optional)
OPERATING ROOM
Figure 4-8. Button Probe
Figure 4-9. Axcess System
General and Optional Equipment and Instruments
• Button Probe if using optional AccuMatch™ surface registration
• Calibration Post for calibrating the Gyrus Diego or Stryker Debriders
• Flat Panel Sterile Drape for display
• Fiducial markers if using MRI and fiducial registration is desired
Page 4-11
Page 50
OPERATING ROOM
• Camera drape bags for the receiver, if sterilization is not possible
• Composite or S-video cable to connect to endoscope or microscope
• Video Scan Converter/Recorder
Note: No natural rubber latex is used in any GE Healthcare Technologies product or
packaging.
WARNING
Only surgical instruments or attachments to instruments supplied or
recommended by GE Healthcare will provide the level of accuracy specified by
this System. We cannot guarantee the accuracy with instruments that are not
recommended by GE Healthcare and using such instruments might compromise
the accuracy leading to a potential patient hazard.
CAUTION
Do not connect any equipment that is not specifically described in this manual to
the System. Whenever AC powered equipment is connected to the System, the
user must confirm that applicable standards for leakage current are met by the
interconnected system.
CAUTION
Due to the magnets in the Instruments, do not put magnetic media, such as a
floppy disk, near any of the Instruments as the data could be erased.
Page 4-12
Page 51
OPERATING ROOM
Setting Up The InstaTrak 3500 Plus
Refer to System Operation in the InstaTrak® 3500 PlusOperator’s Manual for more
detailed instructions.
Note:Select refers either to touching the touch screen with a finger or a pointer or clicking
with the computer mouse.
WARNING
The Network Connection must be unplugged before the InstaTrak® 3500 Plus is
placed in the vicinity of a patient. Not unplugging the network connection could
cause a potential patient hazard.
After turning on the InstaTrak® 3500 Plus and entering the Logon information, the
MAIN MENU appears.
Figure 4-10. Main Menu
•Attach the transmitter and a receiver to the front of the InstaTrak® 3500
Plus.
If a video display of the microscope or endoscope camera is desired on the system
during the surgical procedure, a connection must be made between the endoscopic
camera and the InstaTrak®3500 Plus using composite or S-video cable.
Page 4-13
Page 52
OPERATING ROOM
To establish the link between the system and the microscope or endoscope
camera:
− Plug one end of either a composite cable or an S-Video cable into the
appropriate Video In connector on the back of the system.
− Plug the other end of the cable into the output on the microscope or
endoscopic system.
Loading Images
After selecting the Patient Database button from the MAIN MENU, the PATIENT
DATABASE TASK WINDOW is displayed.
Patient
Scans
Application
Selection
Load
Button
Figure 4-11. Patient Database Menu with Application Selections
Note: If using a CD as the method for transferring the scans, the scan must be transferred to the
Hard Drive before loading.
• Select the patient’s scan from the PATIENT DATABASE.
• Select the Load button.
• If you are using the mouse, you can also double-click on the scan to load it.
Page 4-14
Page 53
OPERATING ROOM
Post processing is the process of formatting the scan and may be required before the scan
can be loaded. If so, the Post processing Message appears.
Figure 4-12 Post Processing Required Dialog
• Select OK to post process the images. Once post processing is complete, you
must select the Load button to load the scan.
• Select Cancel to stop the post processing procedure.
• Once the Load button is selected, a confirmation message will appear. This
enables you to verify that the correct scan has been selected.
• Select Yes to confirm
Note: Unless Image Fusion is available on your InstaTrak 3500 Plus, only one scan can be
loaded at a time.
• If a scan is already loaded, a message appears to confirm that you will be
replacing the existing scan with the new scan.
• Select Yes to confirm
Note: Replacing one scan with another will not delete the previously loaded scan.
Page 4-15
Page 54
OPERATING ROOM
• Select the correct Application, such as Advanced ENT
The following message appears indicating that the scan is being loaded.
Visualization Display
After loading the patient scan, the VISUALIZATION MENU appears.
Message
Center
Figure 4-13. Visualization Menu
The Message Center displays important or informative messages. The messages
indicate tasks that need to be completed in order to initiate tracking or they indicate the
status of the system. The message background will be red or yellow depending on the
status or severity. Red messages must be responded to immediately and always indicate
a severe problem or urgent condition.
The icons on the bottom right of the Display enable quick access to display adjustments.
Page 4-16
Page 55
OPERATING ROOM
Figure 4-14. Display Adjustment
The icons are:
1. Snapshot
2. Brightness and Contrast
3. Crosshair Adjustment
4. Zoom
1 2 3 4
Page 4-17
Page 56
OPERATING ROOM
Snapshot
The Camera icon enables a snapshot to be taken at any time during the
procedure. When selected, a bright light indicates that a snapshot is being
taken. The snapshot is available in the PATIENT DATABASE and can be
stored, viewed, or copied to a floppy disk or CD.
Contrast/Brightness Adjustment
The Contrast/Brightness icon enables adjustment to the appearance of the
display. To make adjustments, select the contrast/brightness icon. When
selected, the IMAGE CONTROLS TASK WINDOW appears on the left side
of the display.
Contrast and
Brightness Sliders
Scan Opacity Slider
Figure 4-15. Image Picture
• To adjust the contrast and/or brightness of the scan, use the sliders.
• Select Close when adjustments are complete.
Note: The Scan Opacity slider and Color Map selector are useful with the Image Fusion
option. See the Image Fusion Operator’s Manual for more information.
Page 4-18
Page 57
Crosshair Adjustment
The tip of the surgeon’s instrument will appear on the Visualization display’s
orthogonal views as a crosshair. By selecting the Crosshair icon, the
appearance can be changed.
OPERATING ROOM
Figure 4-16. Crosshair Adjustment
The Crosshair Adjustment choices are:
• 5 mm scale - crosshairs with tick marks every 5 mm
• Plain – crosshairs with no tick marks
• Target – Places a small, yellow cross at the target location
• None - no crosshairs appear on the scan
Crosshair Adjustment
Page 4-19
Page 58
OPERATING ROOM
Zoom Control Adjustment
The displayed scan size can be made larger or smaller by selecting the Zoom
Controls icon. When the size is selected, all orthogonal displayed views are
adjusted.
Figure 4-17. Zoom Control
The Zoom Control choices are:
• 1:1 magnification
• 2:1 magnification
• 4:1 magnification
Zoom
Zoom
Control
Control
Page 4-20
Page 59
Display Layouts
Orthogonal Views
Medical scans can be viewed directly on the InstaTrak® 3500 Plus. The axial scan is
loaded into the system and reconstructed into coronal and sagittal views. The 2D axial,
sagittal, and coronal views are termed orthogonal.
You can scroll through the scan using the Arrow buttons to view the scan sequentially or
click within the two-dimensional (2D) images or the three-dimensional (3D) model to
locate a point of interest.
OPERATING ROOM
Up and Down
Up and DownArrows
Arrows
}
Figure 4-18. Up and Down Arrows
To view a point of interest, position the cursor on an orthogonal image and click with the
left mouse button. The other views automatically display the same image point.
If you want to view the scan without the crosshairs, the crosshairs can be removed by
selecting None from the Crosshair icon on the display.
Page 4-21
Page 60
OPERATING ROOM
3D Surface Model
The 3D model may be manipulated and adjusted by using the icons on the right of the 3D
model.
Expand 3D
Model Button
Figure 4-19. Surface Model Adjustment
To access these capabilities, select the icon.
1.
2.
3.
4.
5.
3D Surface and
Adjustment
Icons
Page 4-22
6.
1. Opacity of 3D model surface. This button is used most frequently with the
Auto-Segmentation. Available with the Cranial Application only.
Page 61
OPERATING ROOM
2. Point to quickly view a point of interest on the model.
3. Rotate to turn the 3D surface model.
4. Zoom to enlarge or shrink the size of the model.
5. Position to move the model side-to-side and up and down.
6. Reset to quickly adjust the model to default, face-front position.
Expand Button
The 3D model can be expanded to full screen by selecting the Expand button at the top
right of the 3D model. When the 3D Model is expanded to full screen, the button changes
to Shrink to exit the full screen mode.
Page 4-23
Page 62
OPERATING ROOM
Tracking and Surgical Review Views
Different surgical review or tracking views can be selected for each quadrant. A different
view may be selected and changed at any time. The current selection can be seen in the
top left of every quadrant. The views selected at the end of the procedure are stored
when the InstaTrak® 3500 Plus is shut down and displayed during the next use of the
system.
To change the view, select the button at the top left of the quadrant. A pull-down menu
appears enabling you to change the layout for surgical review or tracking. Each quadrant
layout must be changed individually.
Figure 4-20. View Selection Pull Down Menu
Page 4-24
Page 63
Surgical Review Views
Selecting Surgical Review enables you to select a view when not tracking. The view can
be changed at any time. You can select a view for each quadrant.
OPERATING ROOM
Figure 4-21. Surgical Review Pull Down Menu
The Surgical Review view choices are:
• Coronal
• Axial
• Sagittal
• 3D Surface Model
Page 4-25
Page 64
OPERATING ROOM
Tracking View Views
Selecting Tracking View enables you to select a view for tracking. The view may be
changed at any time. You must select a view for each quadrant.
The available Tracking Views are determined by the instrument to be used during
tracking. You must select either Probes or Instruments before selecting a Tracking View.
Figure 4-22. Probes or Instruments Pull Down Menu
Probes are any instrument where a trajectory may be displayed. Examples are the
pointer. Oblique Views are available when Probes is selected.
Instruments are any instrument without a trajectory. An example is an Aspirator.
Page 4-26
Page 65
OPERATING ROOM
If Probes are selected, you can choose a Tracking View from the pull-down list.
Figure 4-23. Tracking View Pull Down List - Probes
The available Tracking Views for Probes are:
• Coronal
• Axial
• Sagittal
• 3D Surface
• Axial Oblique
• Sagittal Oblique
• Perpendicular Oblique
• 3D Surface with Axial Oblique
• 3D Surface with Sagittal Oblique
• 3D Surface with Excavate Brick
• Video
Page 4-27
Page 66
OPERATING ROOM
If Instruments are selected, you can choose a TrackingView from the pull-down list.
Figure 4-24. Tracking View Pull Down List - Instruments
When Instruments is selected, the layout choices are:
• Coronal
• Axial
• Sagittal
• 3D Surface
• 3D Surface with Excavate Brick
• Video
Note: Once the tracking views for either Probes or Instruments are set, they will automatically
appear whenever the applicable instrument is tracked.
Page 4-28
Page 67
OPERATING ROOM
Using the Automatic Registration
Headset
Attaching Transmitter to the Headset
The Automatic Registration Headset may be used for procedures where access to patient
anatomy will not be obstructed by use of the Headset. When AccuMatch™ surface
registration is preferred, the Headset can also function simply as a Transmitter
attachment device.
Figure 4-25. Attaching Transmitter to the Headset
• Align the rotating latch on the transmitterto the slot on the Headset.
• Insert the transmitterinto the slotsin the Headset.
• Secure the transmitterto the Headsetby rotating the latch to lock into place
of the transmitter.
WARNING
These instructions must be followed for use of the Automatic Headset. Failure to
follow these instructions may impact System accuracy leading to a potential
safety hazard.
Attaching the Verification Pad
The Verification Pad must be used with the Automatic Headset. Two pads are provided
in each headset box. The Verification Pad improves patient comfort and system
accuracy.
Page 4-29
Page 68
OPERATING ROOM
• Place a Verification Pad on the patient’s nose after wiping the area to remove
dirt and oils.
• Peel the backing from the Verification Pad to expose the adhesive.
• Align the upper center curve of the Verification Pad with the bridge of the
patient’s nose. Place the verification point on the bony part of the nose. The
verification point should be accessible after the headset is in place.
Upper Center Curve
Verification Point
WARNING
To ensure system accuracy the Verification Pad must be used during the CT
scan and the surgical procedure.
WARNING
Use the cushioned Verification Pad provided with the Headset to increase patient
comfort and to reduce the possibility of bruising to the skin.
Placing the Automatic Headset
Tragus
Page 4-30
Page 69
OPERATING ROOM
• Place the Headset on the patient by expanding the earpieces and positioning
the nosepiece into the deepest notch of the bridge of the nose. Do not pull or
displace the skin
• Align the earpieces over the patient’s ear canals and behind the tragus.
• Slowly allow the earpieces to work into the deepest portion of the ear canal.
Be sure that any accuracy markers used do not change position as the
Headset is placed on the patient.
WARNING
If head stabilization is required, ensure that the head frame does not displace,
press, or pull on the Headset. Displacement of the Headset may cause system
inaccuracy leading to a potential safety hazard.
WARNING
Some patients may experience temporary discomfort, pain, or loss of sensation
when the headset is worn for an extended period of time. Patients with TMJ,
impaired circulation, or other conditions that may be aggravated by use of the
headset, should be given special consideration.
WARNING
A headset must only be used with one patient. Using the same headset with
different patients can result in system inaccuracy due to changes in headset
geometry resulting in a potential safety hazard.
WARNING
To maximi ze System and tracking accuracy, the same headset used during the
CT scan should be used in surgery. A Verification Pad must be used if one was
used during the CT scan.
Draping
When draping the patient, use only lightweight plastic draping to avoid displacing the
Headset. Ensure that the Lateral Markers on the side of the Headset are easily accessible
for registration.
WARNING
When draping the headset prior to surgery, do not allow the draping to distort or
pull on the headset. This can result in system inaccuracy leading to a potential
safety hazard.
If using AccuMatch™ surface registration, an unsterile receiver and Button Probe will be
used for the registration process. The receiver used for the surgical procedure, should be
sterilized before use. If sterilization is not possible, the receiver must be completely
covered with a sterile camera or laser drape bag to maintain sterility. Place the camera
Page 4-31
Page 70
OPERATING ROOM
drape bag over the Receiver, smooth out any wrinkles in the bag at the attachment point
of the Instrument.
CAUTION
If tape or other restraining devices are used to help hold the camera drape bag in
place, care must be taken not to impede or prevent the exchange of the
instruments.
If desired, the Display Touch Screen may be fully draped. Draping the Display Touch
Screen enables the surgeon to interact with the System during the surgical procedure.
Using Transmitter Arm Attachment
If using cranial stabilization, the transmitter may be attached to the cranial stabilization
system. The transmitter must be securely attached in the surgical field before the
registration process can occur. The transmitter can be attached by using one of the three
devices available with the InstaTrak 3500 Plus:
Transmitter Attachment Arm for use with the Mayfield
System
®
Cranial Stabilization
Universal Transmitter Attachment Arm for headframes other than the Mayfield®.
Automatic Registration Headset
If the surgery is performed using a head frame system, such as the Mayfield, the patient
should be positioned before attaching the transmitter.
If the Headset will be used, immobilization of the head is not required since the
transmitter is attached to the Headset.
The patient should not be draped until after the registration process is complete.
Transmitter attachment, instrument calibration and registration are performed in a nonsterile environment.
Attaching the Transmitter Arm to the Mayfield® Cranial
Stabilization System
GE Healthcare Technologies has designed a transmitter attachment arm that connects
specifically to the Mayfield® Cranial Stabilization System.
• Position the patient’s head securely in the Mayfield headframe.
• Insert the starburst connecter of the transmitterattachment arm into the
starburst connector of the Mayfield.
• Tighten using the handle until the teeth of the connector are locked and the
transmitter attachment arm is securely attached.
Page 4-32
Page 71
OPERATING ROOM
Figure 4-26. Starburst Connector
• Position the transmitterattachment armso that it will not be intrusive.
• Position the transmitterattachment arm so that the transmitterwill be located
no more than ten (10) inches away from the surgical site.
• While holding the transmitter Arm in the desired location, tighten the two (2)
additional clamps. The clamps must be very tight to ensure there is no
movement and that the transmitter arm is as rigid as possible.
Figure 4-27. Positioning of Transmitter Arm
with Transmitter less than 10” from Surgical Site
WARNING
Any movement of the transmitter arm after the registration process will cause
inaccuracy that might lead to a potential safety hazard. Ensure that the
transmitter arm is securely attached and will not be bumped during the surgical
procedure.
Page 4-33
Page 72
OPERATING ROOM
Attaching the Transmitter to the Transmitter Arm
The transmitter is attached to the transmitter arm by aligning the transmitter’s rotating
latch to the slot in the transmitter arm.
Figure 4-28. Align Transmitter Latch with Slot
• Insert the transmitter into the corresponding slot in the transmitter attachment
arm
Figure 4-29. Rotate Transmitter Latch to Secure Transmitter
• Rotate the latch on the transmitterto lock it and to prevent it from slipping
CAUTION
If the Transmitter Arm is not completely rigid, registration accuracy might be
compromised.
Page 4-34
Page 73
OPERATING ROOM
Attaching the Universal Transmitter Arm
The Universal Transmitter Arm may be used if a surgical headframe other than the
Mayfield® Cranial Stabilization System is used. The Universal Transmitter Arm is
designed to attach to any headframe or stabilization system.
• Attach the clamp on the Universal Transmitter Arm to the head frame at a
point of adequate grip. The transmitter must be positioned within ten (10)
inches of the surgical site
• Tighten all of the clamps on the Universal Transmitter Arm.
• Open the latch on the Universal Transmitter Arm and insert the transmitter
CAUTION
Headframes constructed of radiolucent materials may be damaged by over
tightening of the universal clamp.
Draping
If using either the transmitter attachment arm for the Mayfield® or the universal
transmitter arm, drape the patient in the usual manner, with the following adjustments:
Figure 4-30. Open Latch -- Insert Transmitter -- Close Latch
• Close the latch. If the transmitter does not fit securely, turn the knob on the
transmitter arm in small increments to tighten the attachment.
• Place a clear cassette or camera drape bag over the transmitter attachment
arm.
• Cut a small hole in the drape to allow access to the transmitter for calibration
purposes.
Page 4-35
Page 74
OPERATING ROOM
Figure 4-31. Transmitter and Attachment Arm Draping
Page 4-36
Page 75
OPERATING ROOM
Attaching an Instrument
Note: All instruments should be sterilized before use.
Note: Multiple sterilization cycles may cause discoloration of the plastic components. This is
normal and will not affect the performance of the device.
The InstaTrak® 3500 Plus automatically will detect which instrument is attached to the
receiver. Confirm that the correct instrument type is indicated before using the
instrument. The Button Probe will be identified as a Pointer.
CAUTION
Prior to each use, visually inspect all Instruments and docking stations. Discard if
they are visually distorted or broken.
Attaching a Pointer
The pointer is supplied in a sterile package and may be used one time only during
surgery. The exception is the Pointer used only for registration purposes. This pointer
may be reused but must be cleaned and sterilized between uses as recommended.
These instructions apply to the pointer and to the button probe and may be followed in
any order, that is, the instrument may be attached to the docking station before attaching
to the receiver.
• Insert the back of the docking station into the slot on the receiver and gently
press until the receiver is locked into the docking station
Page 4-37
Page 76
OPERATING ROOM
• Align the notch on the instrument with the channel on the docking station.
Be certain that the non-functioning suction hole is facing up and magnetic
detectors are facing the receiver.
Snap the instrument into the docking station
CAUTION
Do not use the instrument if the package has been opened or damaged.
Using the Button Probe, P/N 1003632
Note: The following assembly instructions apply only to the Button Probe, P/N 1003632. The
newer version of the Button Probe, P/N 1006909, does not require assembly.
The button probe, P/N 1003632 must be assembled before using.
• Insert the button lever through the button hole.
Figure 4-32. Insertion of button lever
• Squeeze the tabs of the button lever then insert the pivot pins into the pivot
holes.
Page 4-38
• Attach the button probe to the receiver using the instructions for attaching an
instrument shown above.
Page 77
OPERATING ROOM
CAUTION
Sterilization of the Button Probe should be limited to ten cycles.
CAUTION
Prior to each use, visually inspect the Button Probe and the docking station.
Discard any part that is visually distorted, broken and/or in which the probe is no
longer seated securely within the molded handle.
Note: Multiple sterilization cycles may cause discoloration of the plastic components. This is
normal and will not affect the performance of the device.
CAUTION
Both versions of the Button Probe and docking station must be cleaned and
sterilized prior to each use, if using during the surgical procedure.
Nasal Speculum
When using the System in conjunction with a surgical microscope, it is recommended that
you use GE Healthcare’s Nasal Speculum or other titanium or non-magnetic speculum
CAUTION
For optimal system accuracy, the hinge of the speculum should be positioned
away from the patient’s nose.
Page 4-39
Page 78
OPERATING ROOM
Debrider Attachments
The Debrider Attachment is a device that provides a means of attaching the receiver to
almost any of the standard microdebriders being used in sinus surgery. This enables the
InstaTrak 3500 Plus to track the location of the tip of the microdebrider.
Compatible Microdebriders:
• Gyrus Diego™ Powered Dissector
• Gyrus Turbo 7000
• Smith and Nephew ESSential
• XOMED
• XOMED
• Stryker
• Stryker
• Linvatec E9000
®
STRAIGHTSHOT®
®
MAGNUM™
®
Hummer 2
®
Hummer TPS debriders
™
High Speed Shaver Handpiece.
®
Shaver Handpiece
Note: The Debrider Attachment, along with debrider, must be cleaned and sterilized before
use. The Receiver may be sterilized, if desired. Sterilization guidelines may be found in
the InstaTrak 3500 Plus Operator’s Manual. If not sterile, cover the Receiver with a
sterile camera drape bag.
WARNING
When using the debrider, the system should be used for localizing only. Do not
attempt to track with the system while debriding. Tracking while debriding may
result in unintended tissue removal.
Page 4-40
Page 79
OPERATING ROOM
Gyrus™ Diego™ Powered Dissector
Components
The Front Ring Assembly consists of the Receiver Attachment, Position Adjustment
Arm, Position Adjustment Screw and the Front Ring.
Back Ring, with attached screw
Calibration Post
Allen wrench, with small and large ends
Figure 4-33. Gyrus Diego Powered Dissector, Unassembled
Page 4-41
Page 80
OPERATING ROOM
Assembly
• Slide the Front Ring Assembly onto the Gyrus Diego dissector.
• Slide the Back Ring onto the dissector.
• Line up the screw on the Back Ring with the Front Ring Assembly. Use the
large end of the Allen wrench to tighten the attached screw until there is no
movement between the dissector and the attachment. Verify that the
attachment is securely fitted to the dissector by applying a side-to-side force
to the attachment. If the Attachment is loose, tighten the screw.
• The Gyrus Diego Powered Dissector Attachment may be attached to the
Gyrus Diego Powered Dissector in one of three positions: right, left or
center. Figure 4-34 shows the attachment in the center position.
Figure 4-34. Gyrus Diego Powered Dissector, Assembled
To change the position of the Receiver Attachment:
• Remove the Position Adjustment Screw from the Front Ring Assembly by
using the small end of the Allen wrench.
Page 4-42
Page 81
OPERATING ROOM
• Move the Receiver Attachment into one of the three ratcheted positions.
• Replace the screw into the hole and tighten with the Allen wrench.
After the attachment is firmly connected to the Diego dissector, the receiver may be
attached to the Gyrus™ Diego™ PoweredDissector Attachment.
• Snap the attachment onto the receiver foot-first in the same manner as
attaching the Docking Station to the receiver. The system will automatically
detect that a debrider/ dissector is attached.
Disassembly
The Diego Powered Dissector Attachment must be removed from the Diego dissector
before cleaning and sterilization. The Front Ring Assembly must be removed from the
Back Ring, but the Front Ring Assembly may remain as one piece and does not have to
be dismantled prior to sterilization.
• Use the large end of the Allen wrench to loosen the screw on the Back Ring.
The screw is permanently attached to the Back Ring and may not be
removed.
• Slide the Back Ringoff of the dissector.
• Remove the Front Ring Assembly.
Page 4-43
Page 82
OPERATING ROOM
Gyrus Turbo 7000 Debrider
Figure 4-35. Debrider Attachment to the Gyrus TURBO 7000 Debrider
• Unscrew the knob on the Debrider Attachment to open the clamp. It is not
necessary to completely remove the knob from the clamp.
• Attach the Debrider Attachment to the Turbo 7000 Debrider by placing the
clamp around the midsection of the debrider, close to the front of the
debrider. The word Blade on the Debrider Attachment must face the cutter
blade. Position the Debrider Attachment directly above the suction tube.
• Align the knob with the mating threads on the clamp. Turn the knob
clockwise to tighten.
• Verify that the Debrider Attachment is securely fitted to the debrider by
applying a side-to-side force to the Debrider Attachment. If the Debrider
Attachment is loose, tighten the thumbscrew. The Debrider Attachment must
not be loose or be able to move.
• Snap the Snap-on Receiver onto the top platform of the Debrider
Attachment, similar to attaching the receiver to the Docking Station of an
aspirator. The InstaTrak 3500 Plus will automatically detect that the debrider
is attached to a receiver.
Page 4-44
Page 83
OPERATING ROOM
Gyrus (Smith and Nephew) ESSential® Shaver
Handpiece
Figure 4-36. Debrider Attachment Attached to the Gyrus ESSential Shaver Handpiece
• Unscrew the knob on the Debrider Attachment to open the clamp. It is not
necessary to completely remove the knob from the clamp.
• Attach the Debrider Attachment to the ESSential Shaver Handpiece by
placing the clamp around the midsection of the debrider. The word Blade on
the Debrider Attachment must face the cutter blade. Position the Debrider
Attachment above the suction tube.
Page 4-45
Page 84
OPERATING ROOM
XOMED® STRAIGHTSHOT® and MAGNUM™
Debriders
Figure 4-37. Debrider Attachment to the XOMED STRAIGHTSHOT Debrider
• Attach the Debrider Attachment to the STRAIGHTSHOT or MAGNUM
Debrider by placing the vise around the midsection of the debrider. The
word, Blade, on the Debrider Attachment must face the cutter blade. Position
the Debrider Attachment at a comfortable angle on the debrider. The
recommended position is forty-five degrees (45°) to the right of the handle
with the top indicated by the suction tubing. This position generally allows
the least interference with an endoscope during the surgical procedure.
• Press and turn the thumbscrew on the Debrider Attachment clockwise to
secure the Debrider Attachment into place. Tighten until the end of the screw
fits securely into one of the grooves on the debrider.
Note: To prevent accidental disengagement and movement during use, the thumbscrew is
equipped with a positive engagement feature. To turn the thumbscrew, apply downward
pressure to the top of the thumbscrew while turning.
• Verify that the Debrider Attachment is securely fitted to the debrider by
applying a side-to-side force to the Debrider Attachment. If the Debrider
Attachment is loose, tighten the thumbscrew. The Debrider Attachment must
not be loose or be able to move.
• Snap the receiver onto the top platform of the Debrider Attachment, similar
to attaching the Receiver to the Docking Station of an aspirator. The
InstaTrak 3500 Plus will automatically detect that the debrider is attached to
a receiver.
Page 4-46
Page 85
OPERATING ROOM
Stryker® Hummer 2 and Hummer TPS debriders:
Figure 4-38. Debrider Attachment to the Stryker Hummer 2 Debrider
Note: In order to attach the debrider attachment to the Hummer TPS debrider the debrider
must be modified. The front collar of the Hummer TPS must be replaced with a new
front collar similar to another Stryker debrider. If the debrider has not been modified,
contact GE Medical Systems Navigation and Visualization for additional information.
• Unscrew the knob on the Debrider Attachment to open the clamp. It is not
necessary to completely remove the knob from the clamp.
• Attach the Debrider Attachment to the Hummer 2 or Hummer TPS Debrider
by placing the clamp around the midsection of the debrider. The word, Blade,
on the Debrider Attachment must face the cutter blade. Position the Debrider
Attachment at a comfortable angle on the debrider. The recommended
position is thirty degrees (30º) off center of the suction tubing to ensure that
the on/off switch can be easily accessed during the procedure.
• Align the knob with the mating threads on the clamp. Turn the knob
clockwise to tighten.
• Verify that the Debrider Attachment is securely fitted to the debrider by
applying a side-to-side force to the Debrider Attachment. If the Debrider
Attachment is loose, tighten the thumbscrew. The Debrider Attachment must
not be loose or be able to move.
• Snap the Receiver onto the top platform of the Debrider Attachment, similar
to attaching the receiver to the Docking Station of an aspirator. The
InstaTrak 3500 Plus will automatically detect that the debrider is attached to
a receiver.
Page 4-47
Page 86
OPERATING ROOM
Linvatec E9000™ High Speed Shaver Handpiece
Figure 4-39. Debrider Attachment to Linvatec High Speed Shaver Handpiece
• Unscrew the knob on the Debrider Attachment to open the clamp. It is not
necessary to completely remove the knob from the clamp.
• Attach the Debrider Attachment to the E9000 High Speed Shaver Handpiece
by placing the clamp around the midsection of the debrider. The word, Blade,
on the Debrider Attachment must face the cutter blade. Position the Debrider Attachment at a comfortable angle on the debrider. The recommended
position is 30º off center of the suction tube to ensure the on/off switch can
be used during the procedure.
• Align the knob with the mating threads on the clamp. Turn the knob
clockwise to tighten.
• Verify that the Debrider Attachment is securely fitted to the debrider by
applying a side-to-side force to the Debrider Attachment. If the Debrider
Attachment is loose, tighten the thumbscrew. The Debrider Attachment must
not be loose or be able to move.
• Snap the InstaTrak Receiver onto the top platform of the Debrider
Attachment, similar to attaching the receiver to the Docking Station of an
aspirator. The InstaTrak 3500 Plus will automatically detect that the
debrider is attached to a receiver.
Page 4-48
Page 87
Calibration
Calibrating Instruments
Calibration is the process that identifies the position of the tip of the instrument. Every
instrument must be calibrated before using it for surgical navigation. The Calibrate
button on the VISUALIZATION MENU remains orange until calibration is completed
and then changes to green.
CAUTION
If a non-sterile pointer was calibrated for registration, the sterile pointer should
also be calibrated even if the Calibrate button remains green. This action
ensures the best accuracy.
To calibrate an instrument:
Place the tip of the instrument in the dimple on the transmitter.
OPERATING ROOM
3
2
1
Figure 4-40. Calibration
• Select the Calibrate button from the VISUALIZATION MENU.
• If calibration is successful on one point, the Calibrate button turns green.
• If the following message appears, keep the instrument in the dimple
4
Page 4-49
Page 88
OPERATING ROOM
• Slightly rotate or move the instrument through three (3) additional points. A bell
sounds and the cursor moves down the dialog box as each point is accepted.
• When all points are accepted, the Calibrate button turns green.
Figure 4-41. 4 Point Calibration Dialog
Note: In order to pass calibration, the instrument tip must be in the transmitter dimple when a
point is collected. If desired, place a finger on the instrument tip to ensure that it is held
in the dimple.
CAUTION
If a non-sterile pointer was calibrated for registration, the sterile pointer should
also be calibrated even if the Calibrate button remains green. This action
ensures the greatest accuracy.
Calibrating on Transmitter Dimple
• Turn the dissector to its normal “right-side-up” position
• Place the tip of the dissector into the dimple on the Transmitter, with the
dissector held “right-side-up” and at a slight angle, as shown in Figure 4-42.
• Collect 4 points by rotating the Receiver to 9:00, 11:00, 1:00, and 3:00 positions.
• The Receiver cable must not touch the Transmitter during Calibration.
Page 4-50
Page 89
OPERATING ROOM
Figure 4-42. Calibrating the Gyrus Diego Powered Dissector
Note: The receiver cable must not touch the transmitter during Calibration.
Note: Calibration for Diego Powered Dissector is similar to the calibration procedure for the
Aspirator, but the four points should be collected in 60° intervals in two (upper)
quadrants (between 9:00 to 3:00).
.
Figure 4-43. Suggested Calibration Points
Page 4-51
Page 90
OPERATING ROOM
Stryker Hummer and TPS Debriders
A sterile Calibration Post must be used to calibrate the Stryker debriders when attached to
the debrider attachment. Cut a small hole in the Headset draping and insert the sterile
Calibration Post into the hole in the transmitter. Turn the sterile Calibration Post slightly
to lock into place.
Using the dimple on the sterile Calibration Post, follow the instructions for calibration.
Note: In order to calibrate the debrider attachment on the Stryker debrider, the sterile
Calibration Post must be attached to the transmitter.
Page 4-52
Page 91
Registration
Registration is the process of aligning the scan to the patient’s physical space. This
process enables surgical navigation to be performed accurately.
For more details about registration, refer to Getting Started in this manual and to System Operation in the InstaTrak 3500 PlusOperator’s Manual.
There are three methods of registration that may be used with the Advanced ENT
Application:
• AccuMatch™ Surface Registration
• Fiducial Registration
• Automatic Headset Registration
You must access the REGISTRATION MENU to perform either an AccuMatch™
surface registration or a fiducial registration.
You don’t have to use the REGISTRATION MENU if you are using the Headset with
automatic registration.
OPERATING ROOM
Automatic Registration Headset
The Automatic Registration Headset can be used for procedures where access to patient
anatomy will not be obstructed by use of the device. Registration, the alignment of the
images and the patient’s anatomy, is done automatically with the Headset. Automatic
registration and Auto Plus registration may only be obtained through use of the Headset
at the time of the scan and in the OR. When AccuMatch™ registration is preferred, the
Headset may function simply as a transmitter attachment device.
Auto Plus Registration
Auto Plus registration requires that the patient wear the Headset with Lateral Markers
during the CT scan and during surgery. Both of the Lateral Markers must be visible on
the CT scan and automatic registration must be listed in the Auto Reg column as Yes.
Lateral Markers Lateral Markers
Figure 4-44 Lateral Marker Position
Page 4-53
Page 92
OPERATING ROOM
After calibrating and verifying the instrument, an orange message box may appear,
indicating that the lateral markers have been identified by the System.
Figure 4-45 Lateral Marker Message
Collecting Lateral Markers
To collect the lateral markers on the Headset:
• Touch the tip of the instrument to one of the lateral markers on the side of the
Headset. A bell sounds when the point is collected.
• Touch the other marker. A bell sounds when the point is collected.
If the RMS is less than three (3) mm, the VISUALIZATION MENU is displayed and
tracking is initiated. If the RMS is greater than three (3) mm, registration is unacceptable
and a yellow dialog appears with the option to recollect the markers.
To continue without using the lateral markers, select the Cancel button and return to
Automatic Registration.
Page 4-54
Page 93
OPERATING ROOM
Performing a Fiducial or AccuMatch™
Registration
To perform a fiducial or AccuMatch™ registration you must first calibrate the non-sterile
registration pointer. After calibrating, select the Registration button. The
REGISTRATION MENU appears with Automatic, Fiducial, or AccuMatch selected as
the method of registration.
Figure 4-46. Use the Registration Button to Select Method
To change the method of registration, select the Registration Method button. A pulldown menu appears.
Figure 4-47. Registration Method Pull Down Menu
Page 4-55
Page 94
OPERATING ROOM
Select AccuMatch™ or Fiducial, as applicable.
Fiducial Registration
A fiducial registration requires that fiducial markers be placed on the patient prior to the
scan. The fiducial markers contain a radio-opaque material that shows up on the medical
image. These fiducial markers must remain on the patient until the registration process
has been completed.
Fast Fiducial Registration is an exclusive GE Healthcare feature where the system
software automatically locates the markers on the scan, eliminating the process of
identifying the markers manually as in a standard fiducial registration.
The InstaTrak® 3500 Plus stores the location of each fiducial marker that is
automatically identified on the medical image. The actual physical marker on the patient
must be touched with the instrument tip in order to “collect” the point for a valid
registration. The system then calculates the registration RMS value.
CAUTION
Fast Fiducial Registration is not compatible with T2 weighted MR scans because
fiducial markers do not show up clearly on the images.
Page 4-56
Page 95
OPERATING ROOM
Fiducial Registration Instructions
Use the following procedure for both the standard fiducial registration and fast fiducial
registration methods. The term fiducial applies to the donut type of fiducial marker
placed on the patient before the scan.
Note: Before proceeding, rotate the 3-D model so that the majority of the fiducials are visible.
1. After the instrument has been calibrated, select the Registration button. If
necessary, select Fiducial as the registration method.
2. If the system has identified the fiducial markers on the scan, the Collect
Fiducials message appears automatically, and a Fast Fiducial Registration may
be performed.
Fiducial
Fiducial
Registration
Registration
Selected
Selected
Registration
Registration
Points
Points
List
List
Figure 4-48. Fast Fiducial Registration Display
3. A red arrow points to the first fiducial marker on the 3-D model. The crosshairs
are centered on the fiducial and the Registration Points List shows all fiducials
that were located as Uncollected. In this case, skip to Step 13 to continue the
registration process.
4. If the Collect Fiducials message does not appear, and Number One (1) on the
Registration Points List is <Insertion Point>, the system was unable to perform
a valid registration. The Fast Fiducial Registration is not active and a standard
fiducial registration must be performed. In this case, follow all instructions to
complete the fiducial registration process or to add fiducials to the Registration
Points List.
Page 4-57
Page 96
OPERATING ROOM
5. Scroll through the patient’s images and 3D model to identify a fiducial marker.
Crosshairs
Centered
on Fiducial
Markers
Crosshairs
Centered
On Fiducial
Markers
Crosshairs Centered
On Fiducial Markers
on 3D Model
Figure 4-49. Identifying Fiducial Markers with 4:1 Magnification
6. Place the crosshairs in the center of the first fiducial location on the images.
Alignment can be checked in all three planes axial, coronal and sagittal or on the
3D model. Adjust the crosshair position until the center of the crosshairs appears
in the center of the fiducial marker on each one of the orthogonal images.
Note: Select 4:1 magnification for a more precise selection during this step.
Note: As points are collected, the System automatically highlights the applicable “insertion
point” on the list. If an insertion point is reselected manually, the automatic process is
inactive and the cursor must be moved manually from that point on.
7. Select the Image Point button. The Insertion Point List now states:
a. Uncollected
b. <insertion point>
8. Locate the second fiducial marker and center the crosshairs.
9. Select the Image Point button. The Insertion PointList now states:
a. Uncollected
b. Uncollected
c. <insertion point>
10. Continue this process until all fiducials seen on the scan are listed in the
Insertion Point List as Uncollected.
Page 4-58
Page 97
OPERATING ROOM
11. Select the Begin Rec. Acquis. (Begin Receiver Acquisition) button. The Collect
Fiducials message appears and a red arrow points to the fiducial marker on the
3D Model.
12. Fast Fiducial Registration instructions resume at this point.
CAUTION
If more than seven (7) fiducial markers were placed and are identified by the
System, you must collect them in the order indicated by the position of the Red
Arrow on the Display. If seven (7) or less fiducial markers were identified by the
System, and no markers were skipped, you may collect the fiducial markers in
any order.
13. Gently place the pointer in the center of the first fiducial on the patient. Be
careful not to depress the surface of the skin.
a. A high-pitched bell sound indicates that the point was successfully collected.
b. A lower-pitch tone indicates that the system was unable to collect the point.
If there is motion of the instrument or excessive metal interference during a point
collection, the following messages appear and the point is not collected. Select OK to
remove the message.
The point may be recollected. Place the instrument tip on the center of the fiducial
marker, but reposition the receiver in order to reduce metal interference. Select the
Collect Receiver Pts. button.
14. Continue this process until all the fiducials are listed in the Insertion Point List
as collected. Ensure that points are collected only once.
WARNING
It is important that a point is collected only once. Collecting the same point twice
may cause system inaccuracies that could lead to a potential patient safety
hazard.
15. The system calculates the best fit of the image point positions and the actual
patient locations. This value can be seen in the Registration RMS. If the
Registration RMS is acceptable, RMS < 3 mm, you will automatically proceed to
the Visualization Menu to continue with the procedure.
16. If the registration is unacceptable, RMS > 3mm, the following dialog appears:
Figure 4-50. Field Interference Messages
Page 4-59
Page 98
OPERATING ROOM
Figure 4-51. Registration Unacceptable Message
a. Select Yes to recollect the receiver points by touching the center of the
fiducial marker with the tip of the instrument.
b. To repeat the entire process, select the Reset Receiver Pts button to return all
of the image points to Uncollected.
c. If No is selected, the Registration RMS appears on the menu and a residual
distance in millimeters is listed for each insertion point. If one or more of the
values is significantly higher than the other points, and there are enough
fiducials listed (four [4] are required for a valid registration), entries may be
deleted. To delete points, select the entry, and then select Delete Single Entry. The system recalculates the Registration RMS.
17. In rare instances the selected points cannot determine a registration because the
information is mathematically indeterminate. This occurs if the points are too
close together or are positioned in a straight line. If at least four (4) registration
points cannot be collected, a valid registration may not be possible.
18. When the registration process is complete, select the Visualization button to
return to the VISUALIZATION MENU and continue with the procedure.
AccuMatch™ Surface Registration
AccuMatch™ Surface Registration is an optional feature. If your system is equipped, the
REGISTRATION MENU will list AccuMatch as one of the available methods of
registration. If your system is not equipped, the REGISTRATION MENU will list
AccuMatch, but the button will be disabled.
If the patient was not scanned with fiducial markers or the headset, the AccuMatch
registration may be used as an alternate registration process.
Page 4-60
Page 99
OPERATING ROOM
A Button Probe is used to perform an AccuMatch registration. Before beginning the
registration process, the Button Probe must be calibrated and the transmitter must be
securely attached to the patient using either the headset or one of the transmitter
attachment arms.
Guidelines for AccuMatch Registration
• Scan must include anatomical landmarks (nose and ears).
• Collect points in areas where bone is in close proximity to the skin surface
and where the skin is relatively immobile.
• Do not collect points on surface areas of trauma, loose skin (cheeks, previous
craniotomy sites) or movable parts (mandible).
CAUTION
AccuMatch™ Surface Registration is highly dependent on user technique, which
may result in System accuracy variations.
To perform an AccuMatch Registration:
1. After the button probe has been calibrated, select the Registration button. The
REGISTRATION MENU appears.
2. If the Collect/Cancel message appears, select Cancel
3. Select AccuMatch from the pull-down Registration menu.
4. The AccuMatchSURFACE REGISTRATION MENU appears with the 3-D
model displayed on the full screen.
Page 4-61
Page 100
OPERATING ROOM
AccuMatch
Registration
Selected
Find, Touch,
Paint Buttons
Figure 4-53 . AccuMatch Registration Menu
AccuMatch Registration process consists of three (3) steps:
1. Find - Three (3) points are placed on the 3D model using the mouse.
2. Touch - The corresponding three (3) points are touched on the patient.
Note: An RMS number appears after the Touch process is complete. The RMS reflects the
correlation of the three points as set on the 3D model and touched on the patient. If this
RMS is greater than fifteen (15), a message appears indicating that the registration
process must be repeated.
3. Paint - A series of lines are “painted” on the surface of the patient’s skin.
Note: A new RMS replaces the previous RMS in the upper left corner of the screen. The new
RMS reflects the correlation of the patient’s scan with the AccuMatch Registration point
collection.
Step #1 - Find
•Select the Find button.
Page 4-62
• Using the mouse, find three (3) points on the 3D Model. These three (3)
points should surround the area of surgical interest and be as widely
dispersed as possible while still on a single rigid anatomical structure.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.