General Electric BT03 - VOLUSON 730 EXPERT SERVICE MANUAL_SM_105899_3 Enclosed Oil-Less Scroll Air Compressors with Dryers

Page 1
PRELIMINARY
GE Healthcare
Voluson® 730Expert
Service Manual
Voluson® 730Expert systems with Serial number A09000 onwards
Voluson® 730Expert systems with Software version SW 3.x.x (BT03)
(APRIL 29, 2008) E_BT03_COVER.FM
Part Number: 105899
Revision: 3
0123
Page 2
Page 3
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
Important Precautions
• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE TRANSLATION SERVICES.
WARNING
WARNING
(EN)
(EN)
AVERTISSEMENT
(FR)
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL OR OTHER HAZARDS.
• CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.
• SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE TRADUIRE.
• NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.
WARNUNG
(DE)
• DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT, IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG ZU SORGEN.
• WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN ANWWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND NACHVOLLZIEHEN KÖNNEN.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
- i
Page 4
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
• ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.
• SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, LA TRADUCCIÓN ES RESPONSABILIDAD DEL CLIENTE.
AVISO
(ES)
• NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE OTRA NATURALEZA.
• ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL EM INGLÊS.
• SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
ATENÇÃO
(PT)
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
• NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
• O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS, MECÂNICOS OU OUTROS.
AVVERTENZA
(IT)
HOIATUS
(ET)
• IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.
• KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES.
• KUI KLIENDITEENINDUSE OSUTAJA NÕUAB JUHENDIT INGLISE KEELEST ERINEVAS KEELES, VASTUTAB KLIENT TÕLKETEENUSE OSUTAMISE EEST.
• ÄRGE ÜRITAGE SEADMEID TEENINDADA ENNE EELNEVALT KÄESOLEVA TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST.
• KÄESOLEVA HOIATUSE EIRAMINE VÕIB PÕHJUSTADA TEENUSEOSUTAJA, OPERAATORI VÕI PATSIENDI VIGASTAMIST ELEKTRILÖÖGI, MEHAANILISE VÕI MUU OHU TAGAJÄRJEL.
ii -
Page 5
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
VAROITUS
(FI)
ΠΡΟΕΙ∆ΟΠΟΙΗΣΗ
(EL)
FIGYELMEZTETÉS
(HU)
VIÐVÖRUN
(IS)
• ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU.
• EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU.
• REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG SKILIÐ ÞESSA ÞJÓNUSTUHANDBÓK.
• EF EKKI ER FARIÐ AÐ ÞESSARI VIÐVÖRUN GETUR ÞAÐ VALDIÐ MEIÐSLUM ÞJÓNUSTUVEITANDA, STJÓRNANDA EÐA SJÚKLINGS VEGNA RAFLOSTS, VÉLRÆNNAR EÐA ANNARRAR HÆTTU.
- iii
Page 6
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
VÝSTRAHA
(CS)
• DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK.
• HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
ADVARSEL
(DA)
• FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE DENNE SERVICEMANUAL ER BLEVET LÆST OG FORSTÅET.
• MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFØRE SKADE PÅ GRUND AF ELEKTRISK, MEKANISK ELLER ANDEN FARE FOR TEKNIKEREN, OPERATØREN ELLER PATIENTEN.
WAARSCHUWING
(NL)
BRÎDINÂJUMS
(LV)
• DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS VERKRIJGBAAR.
• ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST, DAN IS DE KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN.
• PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS.
• INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, ZOU HET ONDERHOUDSPERSONEEL, DE OPERATOR OF EEN PATIËNT GEWOND KUNNEN RAKEN ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK, MECHANISCHE OF ANDERE GEVAREN.
iv -
Page 7
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
ĮSPĖJIMAS
(LT)
• DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK.
• HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE.
ADVARSEL
(NO)
• IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT.
• MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ GRUNN AV ELEKTRISK STØT, MEKANISKE ELLER ANDRE FARER.
OSTRZEŻENIE
(PL)
ATENŢIE
(RO)
- v
Page 8
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
ОСТОРОЖНО!
(RU)
UPOZORNENIE
(SK)
VARNING
(SV)
• DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA.
• OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA ÖVERSÄTTNINGSTJÄNSTER.
• FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN.
• OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA I SKADOR PÅ SERVICETEKNIKERN, OPERATÖREN ELLER PATIENTEN TILL FÖLJD AV ELEKTRISKA STÖTAR, MEKANISKA FAROR ELLER ANDRA FAROR.
DİKKAT
(TR)
vi -
Page 9
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
(JA)
(ZH-CN)
(KO)
- vii
Page 10
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
DAMAGE IN TRANSPORTATION - FOR USA ONLY
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested within this 14 day period.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY
All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be performed by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall be performed by qualified GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
OMISSIONS & ERRORS
If there are any omissions, errors or suggestions for improving this documentation, please contact the GE Healthcare Global Documentation Group with specific information listing the system type, manual title, part number, revision number, page number and suggestion details. Mail the information to: Service Documentation, 9900 Innovation Drive (RP-2123), Wauwatosa, WI 53226, USA.
GE Healthcare employees should use the iTrak System to report all documentation errors or omissions.
SERVICE SAFETY CONSIDERATIONS
DANGER
!! DANGER:
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
WARNINGWARNING
!! WARNING:
Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety
glasses, and kneeling pad, to reduce the risk of injury.
viii -
Page 11
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
LEGAL NOTES
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare.
GE Healthcare - Kretztechnik may revise this publication from time to time without written notice.
TRADEMARKS
All products and their name brands are trademarks of their respective holders.
COPYRIGHTS
All Material Copyright© 2003 - 2008 by General Electric Inc. All Rights Reserved.
- ix
Page 12
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
Revision History
Revision Date Reason for change
0 September 12, 2003 Initial Release (from Software Version 3.0.x onwards)
1 December 11, 2003 from Software Version 3.0.3 onwards
2 March 08, 2004 from Software Version 3.1.x onwards
3 May 2008 general update, implementaion of new parts, release of Software version 3.1.6
List of Effected Pages
Pages Revision Pages Revision Pages Revision
Title Page
Important Precautions (incl. Legal Notes) pages i to ix
Rev History/LOEP page x
Table of Contents pages xi to xxvi
Chapter 1 - Introduction
pages 1-1 to 1-12
Chapter 2 - Site Preparation
3
pages 2-1 to 2-10
Chapter 3 - Setup Instructions
3
pages 3-1 to 3-52
Chapter 4 - Functional Checks
3
pages 4-1 to 4-44
Chapter 5 - Components and Functions (Theory)
3
pages 5-1 to 5-56
Chapter 6 - Service Adjustments
3
pages 6-1 to 6-8
3
3
3
3
3
Chapter 7 - Diagnostics/Troubleshooting
pages 7-1 to 7-30
Chapter 8 - Replacement Procedures
pages 8-1 to 8-24
Chapter 9 - Renewal Parts
pages 9-1 to 9-36
Chapter 10 - Care & Maintenance
pages 10-1 to 10-26
Index pages I to IV
Back Cover
3
3
3
3
3
3
x -
Page 13
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
Table of Contents
CHAPTER 1
Introduction
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Purpose of Chapter 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Purpose of Service Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Typical Users of the Basic Service Manual . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Voluson® 730Expert Models Covered by this Manual . . . . . . . . . . . . . . . . . 1 - 2
System History - Hardware and Software Versions . . . . . . . . . . . . . . . . . . . 1 - 2
Purpose of Operator Manual(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Important Conventions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Conventions Used in Book . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Standard Hazard Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4
Product Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Human Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Mechanical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 8
Safe Practices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 8
Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 8
Labels Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 9
Dangerous Procedure Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 9
Lockout/Tagout Requirements (For USA Only) . . . . . . . . . . . . . . . . . . . . . . 1 - 10
Returning/Shipping Probes and Repair Parts . . . . . . . . . . . . . . . . . . . . . . . 1 - 10
Electromagnetic Compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 10
What is EMC? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 10
Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 11
Electrostatic Discharge (ESD) Prevention . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 11
Customer Assistance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 11
Contact Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 11
System Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 12
xi Table of Contents
Page 14
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
CHAPTER 2
Site Preparation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Purpose of Chapter 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
General Console Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Console Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Lighting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Electrical Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Voluson® 730Expert Power Requirements . . . . . . . . . . . . . . . . . . . 2 - 3
Inrush Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Site Circuit Breaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Site Power Outlets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Main Power Plug . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
EMI Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Probe Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Time and Manpower Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
System Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Physical Dimensions of Voluson® 730Expert . . . . . . . . . . . . . . . . . . 2 - 5
Weight without Monitor and Peripherals . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Acoustic Noise Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Electrical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Facility Needs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 6
Purchaser Responsibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 6
Mandatory Site Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 7
Site Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 7
Recommended Ultrasound Room Layout . . . . . . . . . . . . . . . . . . . . . 2 - 8
Networking Pre-installation Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 9
Purpose of the DICOM Network Function . . . . . . . . . . . . . . . . . . . . . 2 - 9
DICOM Option Pre-installation Requirements . . . . . . . . . . . . . . . . . . 2 - 9
xii -
Page 15
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
CHAPTER 3
Setup Instructions
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
The Purpose of Chapter 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Set Up Reminders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Average Installation Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Installation Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Moving/Lifting the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
System Acclimation Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Control Panel Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Brake Pedal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3
Safety Reminders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3
Receiving and Unpacking the Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 4
Preparing for Set Up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 6
Verify Customer Order . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 6
System Voltage Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 7
EMI Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 7
Connection of Auxiliary Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 8
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 8
Monitor Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 10
Foot Switch Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 10
ECG-preamplifier Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 11
Global Modem Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 12
S-VHS Video Recorder Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 13
Mitsubishi HS-MD3000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 13
Sony SVO-9500MD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 14
B/W Video Printer Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 15
Line Printer Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 16
Digital Color Printer Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 17
Bluetooth Printer Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 18
HP 5600/5900 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 18
Canon Pixma MP600 / MP610 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 19
External USB-Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 20
External USB-Devices - Connection . . . . . . . . . . . . . . . . . . . . . . . . 3 - 20
External USB-Devices - Disconnection . . . . . . . . . . . . . . . . . . . . . . 3 - 20
Completing the Set Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 22
Connecting the Unit to a Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 22
Power On / Boot Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 22
- xiii
Page 16
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
Scanner Power On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 22
Back-end Processor Boot Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 23
Power Off / Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 24
Back-end Processor Power Down . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 24
Scanner Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 24
Transducer Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 25
Printer Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 26
Installing Line Printer HP 990cxi or HP 995c . . . . . . . . . . . . . . . . . . . . . . . . 3 - 27
Install the HP printer software/driver . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 27
Installing Digital Color Printer Sony UP-D21MD or UP-D23MD . . . . . . . . . . 3 - 29
Install the UP-D21MD / UP-D23MD printer software/driver . . . . . . . 3 - 29
Printer Installation manually . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 31
Adjustment of Printer Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 36
HP 990cxi / HP 995c - Printer Settings . . . . . . . . . . . . . . . . . . . . . . . 3 - 37
UP-D21MD / UP-D23MD - Printer Settings . . . . . . . . . . . . . . . . . . . . 3 - 38
Remote Control Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 40
System Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 41
System Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 41
To invoke the Setup procedure: . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 41
How to enter Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 42
How to enter Hospital Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 42
How to change Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 42
How to change Touch Panel Scheme . . . . . . . . . . . . . . . . . . . . . . . . 3 - 42
On-Board Optional Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 43
External I/O Connection Panel (GES) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 44
External I/O Pin Outs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 45
Video Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 46
Electrical Specifications on VGA Connector . . . . . . . . . . . . . . . . . . . 3 - 47
Available Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 47
Software/Option Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 47
Network IP Address Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 48
Map Network Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 49
Connectivity Setup Worksheet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 50
Paperwork . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 51
Product Locator Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 51
User Manual(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 51
xiv -
Page 17
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
CHAPTER 4
Functional Checks
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Purpose of Chapter 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Required Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
General Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Power On / Boot Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Scanner Power On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Power Off / Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
Scanner Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
System Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 4
Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 4
Touch Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 5
Monitor Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 6
Functional Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 7
2D Mode Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 8
Additional (optional) Operating Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 10
B-Flow Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 10
XTD-View Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 10
M Mode Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 11
Spectral Doppler Mode Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 12
Color Doppler Mode Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 13
Volume Mode Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 15
Pre-Volume Mode Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 15
Functions after the 3D Acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 17
Sub Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 18
Using Cine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 19
Activating Cine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 19
Cine-Split Function (Multiple Format) . . . . . . . . . . . . . . . . . . . . . . . 4 - 19
Activating 2D Auto Cine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 19
Spectral Doppler- or M Cine Loop . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 19
Activating 3D Rotation Cine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 19
Activating Volume Cine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 19
Activating Auto Cine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 19
Basic Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 20
Distance and Tissue Depth Measurements (2D and M Mode) . . . . 4 - 20
Circumference/Area (using Ellipse or Trace) Measurements . . . . . 4 - 21
Volume Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 21
3D MultiPlane Measurements . . . . . . . . . . . . . . . . . . . . . . . . 4 - 21
Velocity Measurements (Spectral Doppler Mode) . . . . . . . . . . . . . . 4 - 22
Acceleration Velocity and Velocity Ratio . . . . . . . . . . . . . . . . 4 - 22
- xv
Page 18
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
Average Velocity (Manual Trace) . . . . . . . . . . . . . . . . . . . . . . 4 - 22
Average Velocity (Auto Trace) . . . . . . . . . . . . . . . . . . . . . . . . 4 - 22
Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 23
OB Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 23
GYN Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 23
Cardiac Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 23
Vascular Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 23
Report Pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 23
Probe/Connectors Usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 24
Connecting a probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 24
Activating the probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 24
Deactivating the probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 24
Disconnecting the probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 24
Image Management (Sonoview) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 25
Using the optional MOD (Magneto-Optical Drive) . . . . . . . . . . . . . . . . . . . . 4 - 26
Formatting Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 27
Adjusting the Write Speed of the DVD/CD Recorder. . . . . . . 4 - 28
Backup and Restore Database, Preset Configurations and Images . . . . . . . . . . . . 4 - 29
Save User Settings Only (Application Settings) . . . . . . . . . . . . . . . . . . . . . . 4 - 30
Load User Settings Only (Application Settings) . . . . . . . . . . . . . . . . . . . . . . 4 - 31
Save Full Backup (Presets, Configurations & Application Settings) . . . . . . 4 - 33
Load Full Backup (Presets, Configurations & Application Settings) . . . . . . . 4 - 35
Delete Full Backup (Presets, Configurations & Application Settings) . . . . . 4 - 37
Archiving Images . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 38
Software Configuration Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 39
Peripheral Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 40
ECG Check Out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 40
Power Supply Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 40
Mechanical Function Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 41
Rotation of the Control Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 41
Brakes and Direction Locks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 41
Site Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 42
Site Log - System (Service Database) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 42
Site Log - Paper Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 43
xvi -
Page 19
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
CHAPTER 5
Components and Functions (Theory)
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Purpose of Chapter 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 2
Description of Voluson® 730Expert Operating Modes . . . . . . . . . . . . . . . . 5 - 6
B-Mode or 2D-Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6
Tissue Harmonic Imaging (THI) incl. Coded Harmonics . . . . 5 - 6
M-Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6
Color Doppler Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6
Color Flow Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6
Power Doppler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 7
Tissue Doppler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 7
Pulsed (PW) Doppler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 7
3D Imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 7
3D Data Collection and Reconstruction . . . . . . . . . . . . . . . . . . . . . . 5 - 7
3D Image Presentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 8
3D Rendering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 8
Block diagram Voluson® 730Expert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 9
Data Flow Control Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 11
B-Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 11
Special B-Mode Techniques . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 12
M-Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 12
D-Mode (Pulsed Wave- and Continuous Wave Doppler) . . . . . . . . 5 - 13
D-Mode Autotrace (draws PC-calculated envelope to D-Spectrum) 5 - 13
CFM-Mode (Color Flow Mode) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 14
3D-Mode (Freezes after 1 volume sweep) . . . . . . . . . . . . . . . . . . . . 5 - 14
Real Time 4D-Mode (nonstop volume rendering) . . . . . . . . . . . . . . 5 - 14
CRI-Mode (Compound Resolution Imaging) . . . . . . . . . . . . . . . . . . 5 - 14
VCI-Mode (Volume Contrast Imaging) . . . . . . . . . . . . . . . . . . . . . . . 5 - 14
Extern-Video-Mode (display Video from Video-Recorder) . . . . . . . . 5 - 14
Sonoview write mode (store Image to Sonoview) . . . . . . . . . . . . . . 5 - 15
Description of Software Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 16
Real Time 4D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 16
DICOM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 16
Real Time 4D-Biopsy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 16
VOCAL - Virtual Organ Computer-aided Analysis . . . . . . . . . . . . . . 5 - 17
B-Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 17
XTD-View (Extended View) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 17
DiagnoSTIC (Spatio-Temporal Image Correlation) . . . . . . . . . . . . . 5 - 17
CRI - Compound Resolution Imaging . . . . . . . . . . . . . . . . . . . . . . . 5 - 17
VCI - Volume Contrast Imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 17
Description of Hardware Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 18
CW - Continuous Wave Doppler . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 18
- xvii
Page 20
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
ECG Preamplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 18
MOD (Magneto-Optical Drive) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 18
Scan/Freeze Foot switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 18
Global Modem (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 18
Location in the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 19
LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 19
Main board Chassis GEF Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 20
FrontEnd Processor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 21
FrontEnd - Board Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 22
CPU - Probe Connector Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 22
CPR - Beamformer-Motherboard . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 22
CPD - Beamformer Sub-board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 23
CRW - CW-Doppler Board (optional) . . . . . . . . . . . . . . . . . . . . . . . . 5 - 24
CPZ - Cover Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 24
CPK - Motherboard of GEF-Module . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 24
CRS - Signal Processor Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 25
CPP - Power Supply Secondary Board + Motor Power stage . . . . . . 5 - 26
BackEnd Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 27
Block diagram CKV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 28
BackEnd - Board Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 29
SBC - Single Board Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 29
PCI VGA Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 29
CKV - DMA-Controller / Video-Card . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 29
Hard Disk Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 29
CPE - Back Panel I/O-Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 29
CPM - PC-Motherboard Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 29
CPP - Power Supply Secondary Board + Motor Power stage . . . . . . 5 - 30
Internal I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 31
Top Console. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 34
Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 38
External I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 39
Peripherals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 40
General Information - GEM (Disk Drive Module) . . . . . . . . . . . . . . . . . . . . . 5 - 40
ECG-preamplifier (MAN6 - optional) . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 40
DVD/CD+(R)W Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 40
Magneto-Optical Drive (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 40
Power Distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 41
CPN6 - Primary Power Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 41
xviii -
Page 21
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
Mechanical Concept and Overview . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 41
Major Functions of CPN6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 41
Fuses on Rear Panel (CPN6) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 42
Fuses inside CPN6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 42
CPN80-81 - Primary Power Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 43
Mechanical Concept and Overview . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 43
Major Functions of CPN80-81 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 43
Fuses on Rear Panel (CPN80-81) . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 44
Fuses inside CPN80-81 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 44
Disk Drive Module (GEM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 44
Fuses on CPE-Board for Disk Drive Module (GEM) . . . . . . . . . . . . 5 - 44
Mechanical Descriptions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 45
Physical Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 45
Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 46
Top Console Positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 46
Rotation of the Control Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 46
Assembly Drawing GW & GEU & Monitor . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 47
Air Flow Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 48
Air Flow Distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 48
Air Flow Distribution Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 48
Service Platform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 49
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 49
Access / Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 49
iLinq Interactive Platform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 50
Service Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 51
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 51
Access / Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 51
Service Login . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 51
Auto Tester . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 52
Service Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 53
Delete all Patients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 53
Export System Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 53
Maintenance Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 53
Monitor Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 55
Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 56
Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 56
NLS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 56
EUM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 56
UIS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 56
SSW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 56
- xix
Page 22
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
CHAPTER 6
Service Adjustments
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Purpose of Chapter 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Regulatory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Monitor Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Monitor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Degauss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3
Color Temp - Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3
Contrast and Brightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 4
Geometry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 4
Convergence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 4
Color temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 4
B-Mode Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 4
Control Console, Transport Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 5
Control Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 5
Transport Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 5
Trackball Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 6
Daylight Saving Time (DST) - New Dates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 7
xx -
Page 23
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
CHAPTER 7
Diagnostics/Troubleshooting
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Purpose of Chapter 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Collect Vital System Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 2
Needed Data - in case of System crash or Errors . . . . . . . . . . . . . . . . . . . 7 - 3
Check Points Voltages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 4
How to check power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 4
Screen Captures and Logs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 5
Capturing a screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 5
Export Log’s and System Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 5
Export System Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 6
How to use the Auto Tester program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 7
Minimum Configuration to Boot/Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 10
Minimum Configuration to Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 10
Troubleshooting Trees and Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 11
System does not Power On / Boot Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 11
Noise disturbs the Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 12
Trackball . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 13
System Does Not Power Off / Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 14
Monitor Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 15
Unable to Record to VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 16
Printer Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 17
DVD/CD+(R)W Troubleshooting (DVD/CD Drive) . . . . . . . . . . . . . . . . . . . 7 - 19
MOD Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 19
Network Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 20
No Connection to the Network at All . . . . . . . . . . . . . . . . . . . . . . . . 7 - 20
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 21
- xxi
Page 24
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
CHAPTER 8
Replacement Procedures
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
Purpose of Chapter 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
Returning/Shipping Probes and Repair Parts . . . . . . . . . . . . . . . . . . . . . . . 8 - 2
Ultrasound Application Software (UIS) Installation Procedure. . . . . . . . . . . . . . . . . 8 - 2
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 2
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 2
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 2
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 2
Software - Installation Procedure (via Service Page) . . . . . . . . . . . . . . . . . . 8 - 5
User Settings Only (Application Settings) Loading Procedure. . . . . . . . . . . . . . . . . 8 - 8
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 8
Loading Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 8
Full Backup (Presets, Configurations & Appl. Settings) Loading Procedure . . . . . . 8 - 8
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 8
Loading Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 8
Service Platform (SSW) Upgrade Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 9
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 9
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 9
Upgrade Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 9
Electronic User Manual (EUM) Upgrade Procedure . . . . . . . . . . . . . . . . . . . . . . . . 8 - 11
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 11
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 11
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 11
EUM - Upgrade Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 12
Software and Functional Checks after the Upgrade . . . . . . . . . . . . . . . . . . . . . . . . 8 - 14
Replacement or Activation of Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 15
Operation for activating Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 15
Operation for installing an Option: . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 15
Transfer of Patient Database and Images from System-to-System. . . . . . . . . . . . . 8 - 16
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 16
Transfer of Patient Database and Images via Sonoview . . . . . . . . . . . . . . . 8 - 17
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 17
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 17
xxii -
Page 25
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 17
Backup all Exams of the “old” system . . . . . . . . . . . . . . . . . . . . . . . 8 - 17
Restore all Exams (of the “old” system) to the “new” system . . . . . 8 - 18
Replacement of the Monitor Task Lamp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 21
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 21
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 21
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 21
Task Lamp - Removal Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 21
Task Lamp - Installation Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 21
Replacement of the Probe Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 22
Replacement of the Probe Holder (at GW140) . . . . . . . . . . . . . . . . . . . . . . 8 - 22
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 22
Probe Holder - Replacement Procedure . . . . . . . . . . . . . . . . . . . . . 8 - 22
Replacement of the Trackball top fixation ring . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 23
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 23
Trackball top fixation ring - Replacement Procedure . . . . . . . . . . . . . . . . . 8 - 23
Replacement of Digipots and TGC Slider controls . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 23
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 23
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 23
Cap and/or Spring - Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . 8 - 23
- xxiii
Page 26
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
CHAPTER 9
Renewal Parts
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
Purpose of Chapter 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
List of Abbreviations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 2
Parts List Groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 3
Housing (GW) and additional Console Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 4
User Interface (GEU Top Console) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 7
Monitor + Monitor Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 9
Disk Drives (GEM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 10
Main Power Module (CPN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 11
Main Board Module (GEF). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 13
FrontEnd (US-Part) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 14
BackEnd Processor (PC-Part) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 16
Options and Upgrades . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 18
Miscellaneous Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 19
Optional Peripherals and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 26
Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 30
2D-Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 30
Real-Time 4D Volume Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 32
CW-Pencil Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 33
Biopsy Needle Guides . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 34
xxiv -
Page 27
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
CHAPTER 10
Care & Maintenance
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Periodic Maintenance Inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Purpose of Chapter 10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Why do Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 2
Keeping Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 2
Quality Assurance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 2
Maintenance Task Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 2
How often should care & maintenance tasks be performed? . . . . . . . . . . . 10 - 2
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 5
Special Tools, Supplies and Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 5
Specific Requirements for Care & Maintenance . . . . . . . . . . . . . . . 10 - 5
System Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 6
Preliminary Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 6
Functional Checks (see also Chapter 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 7
System Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 7
Peripheral/Option Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 8
Input Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 8
Mains Cable Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 8
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 8
General Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 8
Physical Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 9
Optional Diagnostic Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 9
Probe Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 10
Probe Related Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 10
Basic Probe Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 10
Basic Probe Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 10
Disinfection by means of the RIC-Holder . . . . . . . . . . . . . . . . . . . . . 10 - 11
Using a Phantom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 12
Electrical Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 12
Safety Test Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 12
GEMS Leakage Current Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 13
Outlet Test - Wiring Arrangement - USA & Canada . . . . . . . . . . . . . . . . . . 10 - 14
Grounding Continuity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 15
Meter Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 15
Chassis Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 16
- xxv
Page 28
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 SERVICE MANUAL
Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 16
Generic Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 16
Data Sheet for Chassis Source Leakage Current . . . . . . . . . . . . . . . 10 - 17
Isolated Patient Lead (Source) Leakage–Lead to Ground . . . . . . . . . . . . . . 10 - 17
Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 17
Generic Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 18
Isolated Patient Lead (Source) Leakage–Lead to Lead . . . . . . . . . . . . . . . . 10 - 18
Isolated Patient Lead (Sink) Leakage-Isolation Test . . . . . . . . . . . . . . . . . . 10 - 19
Data Sheet for ECG Leakage Current . . . . . . . . . . . . . . . . . . . . . . . . 10 - 20
Probe Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 21
Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 21
Generic Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 21
No Meter Probe Adapter Procedure . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 22
Data Sheet for Transducer Source Leakage Current . . . . . . . . . . . . 10 - 23
When There's Too Much Leakage Current.... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 24
Chassis Fails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 24
Probe Fails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 24
Peripheral Fails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 24
Still Fails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 24
New Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 24
ECG Fails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 24
xxvi -
Page 29
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 1 Introduction
Section 1-1 Overview
1-1-1 Purpose of Chapter 1
This chapter describes important issues related to safely servicing the Voluson® 730Expert scanner. The service provider must read and understand all the information presented in this manual before installing or servicing a unit.
Table 1-1 Contents in Chapter 1
Section Description Page Number
1-1
1-2
1-3
1-4
1-5
Overview
Important Conventions
Safety Considerations
Electromagnetic Compatibility (EMC)
Customer Assistance
1-1-2 Purpose of Service Manual
This Service Manual provides installation and service information for the Voluson® 730Expert Ultrasound Scanning System and contains the following chapters:
1.) Chapter 1 - Introduction: Contains a content summary and warnings.
2.) Chapter 2 - Site Preparation: Contains pre-installation requirements for the Voluson® 730Expert.
3.) Chapter 3 - Setup Instructions: Contains installation procedures.
4.) Chapter 4 - Functional Checks: Contains functional checks that are recommended as part of the
installation, or as required during servicing and periodic maintenance.
5.) Chapter 5 - Components and Functions (Theory): Contains block diagrams and functional
explanations of the electronics.
6.) Chapter 6 - Service Adjustments: Contains instructions on how to make available adjustments to
the Voluson® 730Expert.
7.) Chapter 7 - Diagnostics/Troubleshooting: Provides procedures for running diagnostic or related
routines for the Voluson® 730Expert.
8.) Chapter 8 - Replacement Procedures: Provides disassembly procedures and reassembly
procedures for all changeable Field Replaceable Units (FRU).
9.) Chapter 9 - Renewal Parts: Contains a complete list of field replaceable parts for the
Voluson® 730Expert.
10.)Chapter 10 - Care & Maintenance: Provides periodic maintenance procedures for the
Voluson® 730Expert.
1-1
1-3
1-7
1-10
1-11
Chapter 1 - Introduction 1-1
Page 30
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
1-1-3 Typical Users of the Basic Service Manual
Service Personnel (installation, maintenance, etc.).
Hospital’s Service Personnel
Contractors (Some parts of Chapter 2 - Pre-Installation)
1-1-4 Voluson® 730Expert Models Covered by this Manual
Table 1-2 Voluson® 730Expert Model Designations
GE Part Number Kretz # Description
H48621N ---------- Voluson® 730Expert Main Body (BT03)
1-1-5 System History - Hardware and Software Versions
This manual applies to Voluson® 730Expert systems:
- with Serial number A09000 - (onwards)
- with Software version 3.0.x (BT03) installed
- with Software version 3.1.x (BT03) installed
- that were upgraded to BT03 (SW 3.0.x or higher installed)
1-1-6 Purpose of Operator Manual(s)
The Operator Manual(s) should be fully read and understood before operating the Voluson® 730Expert and also kept near the unit for quick reference.
1-2 Section 1-1 - Overview
Page 31
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Section 1-2 Important Conventions
1-2-1 Conventions Used in Book
Icons
Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter.
Safety Precaution Messages
Various levels of safety precaution messages may be found on the equipment and in the service information. The different levels of concern are identified by a flag word that precedes the precautionary message. Known or potential hazards are labeled in one of following ways:
DANGER
!! DANGER:
WARNINGWARNING
!! WARNING:
CAUTION
!! CAUTION:
NOTICE
!! NOTICE:
INDICATES THE PRESENCE OF A HAZARD THAT WILL CAUSE SEVERE PERSONAL INJURY OR DEATH IF THE INSTRUCTIONS ARE IGNORED.
INDICATES THE PRESENCE OF A HAZARD THAT CAN CAUSE SEVERE PERSONAL INJURY AND PROPERTY DAMAGE IF INSTRUCTIONS ARE IGNORED.
Indicates the presence of a hazard that will or can cause minor personal injury and property damage if instructions are ignored.
Equipment Damage Possible Notice is used when a hazard is present that can cause property damage but has absolutely no
personal injury risk.
Example: Disk drive will crash.
NOTE: Notes provide important information about an item or a procedure.
Information contained in a NOTE can often save you time or effort.
Chapter 1 - Introduction 1-3
Page 32
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
1-2-2 Standard Hazard Icons
Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
Table 1-3 Standard Hazard Icons
ELECTRICAL MECHANICAL RADIATION
LASER HEAT PINCH
LASER
LIGHT
Other hazard icons make you aware of specific procedures that should be followed.
Table 1-4 Standard Icons Indicating a Special Procedure be Used
AVOID STATIC ELECTRICITY TAG AND LOCK OUT WEAR EYE PROTECTION
TAG
&
LOCKOUT
Date
Signed
EYE
PROTECTION
1-4 Section 1-2 - Important Conventions
Page 33
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
1-2-3 Product Icons
The following table describes the purpose and location of safety labels and other important information provided on the equipment.
Table 1-5 Product Icons
LABEL/SYMBOL PURPOSE/MEANING LOCATION
Identification and Rating Plate
Device Listing/Certification
Labels
0123
Type/Class Label
IP Code (IPX 1) IP Code (IPX 7)
“CAUTION
This unit weighs...
Special care must be used to
avoid..."
Manufacturer's name and address
Model and serial numbers
Electrical ratings
Laboratory logo or labels denoting conformance with industry safety standards such as UL or IEC.
Council Directive 93/42/EEC concerning medical devices: The CE mark affixed to the equipment testifies compliance to the directive.
Used to indicate the degree of safety or protection.
Indicates the degree of protection provided by the enclosure per IEC 529. IPX 1 and IPX 7 indicates drip proof.
Equipment Type BF (man in the box symbol) IEC 878-02-03 indicates B Type equipment having even more electrical isolation than standard Type B equipment because it is intended for intimate patient contact.
This precaution is intended to prevent injury that may result if one person attempt to move the unit considerable distances or on an incline due to the weight of the unit.
Rear side of the unit
Monitor rear side
On each probe
Rear side of the unit
Rear side of the monitor
Rear side of the unit on the plug of each probe
Footswitch Probes
Probe connectors
Front side of the ECG-preamplifier (MAN)
Rear of Power Supply
Used in the Service and User Manual which should be adjacent to equipment at all times for quick reference.
"CAUTION" The equilateral triangle is usually used in combination with other symbols to advise or warn the user.
ATTENTION - Consult accompanying documents " is intended to alert the user to refer to the operator manual or other instructions when complete information cannot be provided on the label.
"CAUTION - Dangerous voltage" (the lightning flash with arrowhead in equilateral triangle) is used to indicate electric shock hazards.
Chapter 1 - Introduction 1-5
Various
Rear side of Power Supply
Rear side of Monitor
Page 34
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Table 1-5 Product Icons (Continued)
LABEL/SYMBOL PURPOSE/MEANING LOCATION
"Mains OFF" Indicates the power off position of the mains power switch.
"OFF/Standby" Indicates the power off/standby position of the power switch.
CAUTION This Power Switch DOES NOT ISOLATE Mains Supply
"Mains ON" Indicates the power on position of the mains power switch.
ON switch of the isolation transformer for auxiliary devices.
OFF switch of the isolation transformer for auxiliary devices.
"Protective Earth" Indicates the protective earth (grounding) terminal.
"Equipotentiality" Indicates the terminal to be used for connecting equipotential conductors when interconnecting (grounding) with other equipment.
Rear of system at mains switch (F1)
Adjacent to On-Off/Standby switch left below the Control panel.
Rear of system at mains switch (F1)
Rear of system at the switch for auxiliary devices (F2)
Rear of system at the switch for auxiliary devices (F2)
Internal, Rear of Power Supply
Rear of Power Supply
DANGEROUS ELECTRIC VOLTAGE. Pull the mains plug before opening the unit!
Waste Electrical and Electronic Equipment (WEEE) Disposal. This symbol indicates that waste electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment.
These symbols indicate that at least one of the six hazardous substances of the China RoHS Labelling Standard is above the RoHS limitation. The number inside the circle is referred to as the Environmental Friendly Use Period (EFUP). It indicates the number of years that the product, under normal use, will remain harmless to health of humans or the environment.
EFUP = 10 for Short Use Products
EFUP = 20 for Medium Use Products
LAMP CONTAINS MERCURY, DISPOSE ACCORDING TO
+J
STATE/LOCAL LAW
KTD100272_2
This product consists of devices that may contain mercury, which must be recycled or disposed of in accordance with local, state, or country laws. (Within this system, the backlight lamps in the Touch Panel contain mercury.)
1-6 Section 1-2 - Important Conventions
Rear of Power Supply (if “new” version of CPN80-81 is installed)
Rear side of the unit
on the plug of each probe
(availability depends on delivery status)
Rear side of the unit
on the plug of each probe
(availability depends on delivery status)
Rear side of the unit
on rear side of the Touch Panel
(availability depends on delivery status)
Page 35
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Section 1-3 Safety Considerations
1-3-1 Introduction
The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
1-3-2 Human Safety
Operating personnel must not remove the system covers. Servicing should be performed by authorized personnel only. Only personnel who have participated in a Voluson® 730Expert Training are authorized to service the equipment.
1-3-3 Mechanical Safety
!! WARNING:
!! WARNING:
!! WARNING:
!! CAUTION:
!! CAUTION:
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
CAUTION
CAUTION
USE EXTREME CAUTION WHEN ELEVATING THE UNIT, OR IF IT IS RAISED FOR A REPAIR OR MOVED ALONG ANY INCLINE. IT MAY BECOME UNSTABLE WHICH COULD CAUSE THE UNIT TO TIP OVER.
ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN HANDLING AND PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A DAMAGED OR DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN SERIOUS INJURY AND EQUIPMENT DAMAGE.
NEVER USE A PROBE THAT HAS FALLEN TO THE FLOOR. EVEN IF IT LOOKS OK, IT MAY BE DAMAGED.
Always lower and center the Operator I/O Panel before moving the scanner.
The Voluson® 730Expert weighs 136 kg or more, depending on installed peripherals, (300 lbs., or more) when ready for use.
Care must be used when moving it or replacing its parts. Failure to follow the precautions listed could result in injury, uncontrolled motion and costly damage.
ALWAYS:
• Use the handle to move the system. • Be sure the pathway is clear.
• Use slow, careful motions. • Do not let the system strike walls or door frames. Two people are required when moving on inclines or lifting more than 16 kg (35 lbs).
CAUTION
!! CAUTION:
Keep heat venting holes on the monitor unobstructed to avoid overheating of the monitor.
NOTE: Special care should be taken when transporting the unit in a vehicle:
Secure the unit in an upright position and lock the caster wheels (brake).
DO NOT use the Control Panel as an anchor point.
Place the probes in their carrying case.
Eject any DVD, CD or Magneto Optical disk (MOD) from their drive.
Ensure that the Voluson® 730Expert system is firmly secured while inside the vehicle.
Prevent vibration damage by driving cautiously. Avoid unpaved roads, excessive speeds, and erratic stops or starts.
Chapter 1 - Introduction 1-7
Page 36
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
1-3-4 Electrical Safety
1-3-4-1 Safe Practices
To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The system is equipped with a three-conductor AC power cable. This must be plugged into an approved electrical outlet with safety ground. If an extension cord is used with the system, make sure that the total current rating of the system does not exceed the extension cord rating.
The power outlet used for this equipment should not be shared with other types of equipment.
Both the system power cable and the power connector meet international electrical standards.
WARNINGWARNING
!! WARNING:
CONNECTING A VOLUSON® 730EXPERT SCANNER TO THE WRONG VOLTAGE LEVEL WILL MOST LIKELY DESTROY THE SCANNER.
1-3-4-2 Probes
All the probes for the Voluson® 730Expert are designed and manufactured to provide trouble-free, reliable service. To ensure this, correct handling of probes is important and the following points should be noted:
Do not drop a probe or strike it against a hard surface, as this may damage the transducer elements, acoustic lens, or housing.
Inspect the probe prior to each use for damage or degradation to the Housing, Cable strain relief, Lens and Seal.
Do not use a cracked or damaged probe. In this event, call your field service representative immediately to obtain a replacement.
Avoid pulling, pinching or kinking the probe cable, since a damaged cable may compromise the electrical safety of the probe.
To avoid the risk of a probe accidentally falling, do not allow the probe cables to become entangled, or to be caught in the machine’s wheels.
Never immerse the probe connector or adapter into any liquid.
NOTE: For detailed information on handling probes, refer to the Voluson® 730Expert Basic User Manual and
the care card supplied with the probe.
1-8 Section 1-3 - Safety Considerations
Page 37
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
1-3-5 Labels Locations
12
8* 6
3
45
7*
Figure 1-1 Labeling
1.) Caution Label
2.) CE-Label
3.) UL-Label
4.) Identification “Main” Label
5.) CW-Doppler (only if the CW-Doppler option is installed)
6.) Quality Control Label
7.) *place for additional label (e.g., “Hg Vermont label”, “Homologation label” for Japan or China only)
8.) *place for Disposal Label (WEEE)
* availability depends on delivery status of the Voluson® 730Expert system
1-3-6 Dangerous Procedure Warnings
Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed.
DANGER
!! DANGER:
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
WARNINGWARNING
!! WARNING:
WARNINGWARNING
!! WARNING:
EXPLOSION WARNING
DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE. OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT CONSTITUTES A DEFINITE SAFETY HAZARD.
DO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT
BECAUSE OF THE DANGER OF INTRODUCING ADDITIONAL HAZARDS, DO NOT INSTALL SUBSTITUTE PARTS OR PERFORM ANY UNAUTHORIZED MODIFICATION OF THE EQUIPMENT.
Chapter 1 - Introduction 1-9
Page 38
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
1-3-7 Lockout/Tagout Requirements (For USA Only)
Follow OSHA Lockout/Tagout requirements to protect service personnel from injuries caused by unexpected energizing or start-up of equipment during service, repair, or maintenance.
NOTICE
!! NOTICE:
Energy Control and Power Lockout for Voluson® 730Expert.
When servicing parts of the system where there is exposure to voltage greater than 30 Volts: Unplug the system Maintain control of the system power plug
TAG
&
LOCKOUT
Signed
Date
There are no test points to verify isolation, you must wait for at least 20 seconds for capacitors to discharge.
Beware that the Power Supply (CPN), Front End Processor and Back End Processor may be energized even if the power is turned off when the cord is still plugged into the AC Outlet.
UT
1-3-8 Returning/Shipping Probes and Repair Parts
Equipment being returned must be clean and free of blood and other infectious substances.
GEMS policy states that body fluids must be properly removed from any part or equipment prior to shipment. GEMS employees, as well as customers, are responsible for ensuring that parts/equipment have been properly decontaminated prior to shipment. Under no circumstance should a part or equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or and ultrasound probe).
The purpose of the regulation is to protect employees in the transportation industry, as well as the people who will receive or open this package.
NOTE: The US Department of Transportation (DOT) has ruled that “items what were saturated and/or dripping
with human blood that are now caked with dried blood; or which were used or intended for use in patient care” are “regulated medical waste” for transportation purpose and must be transported as a hazardous material.
Section 1-4 Electromagnetic Compatibility (EMC)
1-4-1 What is EMC?
Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings including other equipment, power sources and persons with which the device must interface. Inadequate compatibility results when a susceptible device fails to perform as intended due interference from its environment or when the device produces unacceptable levels of emission to its environment. This interference is often referred to as radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through space or conducted over interconnecting power of signal cables. In addition to electromagnetic energy, EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and disturbances in the electrical power supply.
For applicable standards refer to Chapter 2 in the Basic User Manual of the Voluson® 730Expert.
1-10 Section 1-4 - Electromagnetic Compatibility (EMC)
Page 39
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
1-4-2 Compliance
The Voluson® 730Expert unit conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient requirements.
NOTE: For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh, clamps, are in
good condition, installed tightly without skew or stress. Proper installation following all comments noted in this service manual is required in order to achieve full EMC performance.
1-4-3 Electrostatic Discharge (ESD) Prevention
WARNINGWARNING
!! WARNING:
DO NOT touch any boards with integrated circuits prior to taking the necessary ESD precautions:
1.) When installing boards, ESD may cause damage to a board. ALWAYS connect yourself, via an arm-wrist strap, to the advised ESD connection point located on the rear of the system (to the right of the power connector).
2.) Follow general guidelines for handling of electrostatic sensitive equipment.
WARNINGWARNING
!! WARNING:
Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts, conductors and components. Always use non-conductive handles designed for the removal and replacement of ESD sensitive parts. All parts that have the potential for storing energy must be discharged or isolated before making contact.
Section 1-5 Customer Assistance
1-5-1 Contact Information
If this equipment does not work as indicated in this service manual or in the Basic User Manual, or if you require additional assistance, please contact the local distributor or appropriate support resource, as listed below.
NOTE: Prepare vital system information (see: Section 7-2 on page 7-2) before you call:
System Type
System Serial number (also visible on label on back of the system)
Application Software version
Backup version
additional information about installed software
Chapter 1 - Introduction 1-11
Page 40
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
1-5-1 Contact Information (cont’d)
Table 1-6 Phone Numbers for Customer Assistance
Location Phone Number
USA
GE Medical Systems
Ultrasound Service Engineering
9900 Innovation Drive (RP-2123)
Wauwatosa, WI 53226, USA
Canada 1-800-668-0732
Latin America
Europe
GE Medical Systems Kretztechnik GmbH & Co OHG
Service Department - Ultrasound
Tiefenbach 15
A-4871 Zipf (Austria)
Asia Support Center - Singapore
GE Ultrasound Asia
Service Department - Ultrasound
298 Tiong Bahru Road #15-01/06
Central Plaza
Singapore 169730
Japan Support Center
Service On-site
Service: Parts
Applications support
Fax:
Service
Applications support
Phone: +43 (0)7682-3800-26
Fax: +43 (0)7682-3800-47
Phone: +65-6291-8528
Fax: +65-6291-7006
Phone: 81-426-48-2944
Fax: 81-426-48-2905
1-800–437–1171
1-800-558-2040
1-800-682-5327 or 1-262-524-5698
1-414-721-4387
1-800-321-7937
1-262-524-5698
1-5-2 System Manufacturer
Table 1-7 System Manufacturer
Manufacturer Telephone FAX
GE Medical Systems - Kretztechnik GmbH & Co OHG
Tiefenbach 15
A-4871 Zipf
Austria
1-12 Section 1-5 - Customer Assistance
+43-7682-3800-0 +43-7682-3800-47
Page 41
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 2 Site Preparation
Section 2-1 Overview
2-1-1 Purpose of Chapter 2
This chapter provides information required to plan and prepare for installation of a Voluson® 730Expert ultrasound unit. Included are descriptions of the facility and electrical needs to be met by the purchaser.
Table 2-1 Contents in Chapter 2
Section Description Page Number
2-1
2-2
2-3
Overview
General Console Requirements
Facility Needs
2-1
2-2
2-6
Chapter 2 - Site Preparation 2-1
Page 42
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Section 2-2 General Console Requirements
2-2-1 Console Environmental Requirements
Table 2-2 Environmental Requirements
!! CAUTION:
CAUTION
Operating
Temperature
10 to 40oC
(50 to 104
o
F)
If the system has been in storage or has been transported, please see the acclimation requirements before powering
ON and/or using the system (see: Section 3-2-2 "Installation Warnings" on page 3-2).
Operating
Humidity
30 to 80% rH
non-condensing
Heat Dissipation
3446 BTU pr hour
Storage
Temperature
-10 to 40
(14 to 104
o
C
o
F)
Storage Humidity
< 90% rH non-
2-2-1-1 Cooling
The cooling requirement for the Voluson® 730Expert is 3446 BTU/hr. This figure does not include cooling needed for lights, people, or other equipment in the room.
NOTE: Each person in the room places an additional 300 BTU/hr. demand on the cooling system.
2-2-1-2 Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source of EMI which could degrade image quality. These controls should be selected to minimize possible interference.
2-2-2 Electrical Requirements
NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound
equipment. This dedicated power shall originate at the last distribution panel before the system.
condensing
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size Ground wire from the distribution panel to the Ultrasound outlet.
NOTE: Please note that image artifacts can occur, if at any time within the facility, the Ground from the main
facility's incoming power source to the Ultrasound unit is only a conduit.
2-2 Section 2-2 - General Console Requirements
Page 43
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
2-2-2-1 Voluson® 730Expert Power Requirements
Table 2-3 Electrical Specifications for Voluson® 730Expert
Voltage Tolerances Current Frequency
100 VAC ±10% 10.10 A 50, 60 Hz (±2%)
110 VAC ±10% 9.20 A 50, 60 Hz (±2%)
115 VAC ±10% 8.80 A 50, 60 Hz (±2%)
130 VAC ±10% 7.80 A 50, 60 Hz (±2%)
230 VAC ±10% 4.40 A 50, 60 Hz (±2%)
240 VAC ±10% 4.20 A 50, 60 Hz (±2%)
Power Consumption nominal 1010 VA including all options.
Mains outlets: Mains socket ST1, ST2, ST3, ST4, ST5 for accessories. All mains outlets are co-switched by the unit’s mains switch via built-in isolation transformer. Output voltage for: ST1 - ST5: 115V or 230V.
CAUTION
!! CAUTION:
Modification of voltage setting only by an authorized service person! The maximum power consumption of equipment (inclusive color video monitor) connected to these outlets must not exceed 350VA!
2-2-2-2 Inrush Current
Inrush current is not a factor to consider due to the inrush current limiting properties of the power supplies.
2-2-2-3 Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be readily accessible.
CAUTION
!! CAUTION:
POWER OUTAGE MAY OCCUR. The Voluson® 730Expert requires a dedicated single branch circuit. To avoid circuit overload
and possible loss of critical care equipment, make sure you DO NOT have any other equipment operating on the same circuit.
2-2-2-4 Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without extension cords. Other adequate outlets for the external peripherals, medical and test equipment needed to support this unit must also be present within 1 m (3.2 ft.) of the unit. Electrical installation must meet all current local, state, and national electrical codes.
2-2-2-5 Main Power Plug
The Voluson® 730Expert is supplied with a main power plug, as standard. In the event that the unit arrives without a power plug, or with the wrong plug, contact your GE dealer. When necessary, the installation engineer will supply the appropriate power plug to meet the applicable local regulations.
Chapter 2 - Site Preparation 2-3
Page 44
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
2-2-3 EMI Limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transients in the air or wiring. Ultrasound machines also generate EMI. The Voluson® 730Expert complies with limits as stated on the EMC label. However, there is no guarantee that interference will not occur in a particular installation.
NOTICE
Possible EMI sources should be identified before the unit is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. Sources of EMI include the following:
medical lasers
scanners
cauterizing guns
•computers
•monitors
•fans
gel warmers
microwave oven
light dimmers
portable phones
broadcast stations and mobile broadcasting machines
Table 2-4 EMI Prevention/Abatement
EMI Rule Details
Be aware of RF sources.
Keep the unit at least 5 meters (16.4 feet) away from other EMI sources. Special shielding may be required to eliminate interference problems caused by high frequency, high powered radio or video broadcast signals.
Ground the unit.
Replace and/or reassemble all screws, RF gaskets, covers and cores.
Replace broken RF gaskets.
Do not place labels where RF gaskets touch metal.
Use GE- specified harnesses and peripherals.
Take care with cellular phones.
Properly dress peripheral cables.
Poor grounding is the most likely reason a unit will have noisy images. Check grounding of the power cord and power outlet.
After you finish repairing or updating the system, replace all covers and tighten all screws. Any cable with an external connection requires a magnet wrap at each end. Install the shield over the front of card cage. Loose or missing covers or RF gaskets allow radio frequencies to interfere with the ultrasound signals.
If more than 20% or a pair of the fingers on an RF gasket are broken, replace the gasket. Do not turn ON the unit until any loose metallic part is removed and replaced, if required.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit RF leakage. In case a label has been found in such a location, move the label to a different, appropriate location.
The interconnect cables are grounded and require ferrite beads and other shielding. Cable length, material, and routing are all important; do not make any changes that do not meet all specifications.
Cellular phones may transmit a 5 V/m signal that causes image artifacts.
Do not allow cables to lie across the top of the card cage or hang out of the peripheral bays. Loop the excess length for peripheral cables inside the peripheral bays. Attach the monitor cables to the frame.
2-4 Section 2-2 - General Console Requirements
Page 45
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
2-2-4 Probe Environmental Requirements
Operation: Ambient temperature 18° to 30° C
Storage: -10° to 50° C
NOTE: Temperature in degrees C. Conversion to degrees F = °C (9/5) + 32).
NOTICE
!! NOTICE:
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
-10 TO + 50 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
2-2-5 Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six weeks before delivery, for enough time to make necessary changes.
CAUTION
!! CAUTION:
Have two people available to deliver and unpack the Voluson® 730Expert. Attempts to move the unit considerable distances or on an incline by one person could result in injury or damage or both.
2-2-6 System Specifications
2-2-6-1 Physical Dimensions of Voluson® 730Expert
The physical dimensions of the Voluson® 730Expert unit are summarized in Ta bl e 2 -5 .
Table 2-6 lists the size of Voluson® 730Expert, with monitor and without on-board peripherals.
Table 2-5 Physical Dimensions of Voluson® 730Expert
Height Width Depth
142 cm / 55.9 inches 68 cm / 26.8 inches 100 cm / 39.4 inches
2-2-6-2 Weight without Monitor and Peripherals
Table 2-6 Weight of Voluson® 730Expert with Monitor and without other Peripherals
Model Weight [kg] Weight [lbs.]
Voluson® 730Expert 136 300
2-2-6-3 Acoustic Noise Output
max. 57dB(A)
2-2-6-4 Electrical Specifications
Please refer to Section 2-2-2-1 "Voluson® 730Expert Power Requirements" on page 2-3.
Chapter 2 - Site Preparation 2-5
Page 46
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Section 2-3 Facility Needs
2-3-1 Purchaser Responsibilities
The work and materials needed to prepare the site is the responsibility of the purchaser. Delay, confusion, and waste of manpower can be avoided by completing pre installation work before delivery.
Use the Pre-installation checklist (provided in Ta ble 2- 7) to verify that all needed steps have been taken.
Table 2-7 Voluson® 730Expert Pre-Installation Check List
Action Yes No
Schedule at least 3 hours for installation of the system.
Notify installation team of the existence of any variances from the basic installation.
Make sure system and probes have been subject to acclimation period.
Environmental cooling is sufficient.
Lighting is adjustable to adapt to varying operational conditions of the scanner.
Electrical facilities meet system requirements.
EMI precautions have been taken and all possible sources of interference have been removed.
Mandatory site requirements have been met.
If a network is used, IP address has been set for the system and a dedicated network outlet is available.
Purchaser responsibility includes:
Procuring the materials required.
Completing the preparations before delivery of the ultrasound system.
Paying the costs for any alterations and modifications not specifically provided in the sales contract.
NOTE: All electrical installations that are preliminary to the positioning of the equipment at the site prepared for
the equipment must be performed by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations, and testing must also be performed by qualified personnel. The products involved (and the accompanying electrical installations) are highly sophisticated and special engineering competence is required. All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
The desire to use a non–listed or customer provided product or to place an approved product further from the system than the interface kit allows presents challenges to the installation team. To avoid delays during installation, such variances should be made known to the individuals or group performing the installation at the earliest possible date (preferably prior to the purchase).
The ultrasound suite must be clean prior to delivery of the machine. Carpet is not recommended because it collects dust and creates static. Potential sources of EMI (electromagnetic interference) should also be investigated before delivery. Dirt, static, and EMI can negatively impact system reliability.
2-6 Section 2-3 - Facility Needs
Page 47
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
2-3-2 Mandatory Site Requirements
NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound
equipment. This dedicated power shall originate at the last distribution panel before the system. The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet. Please note that image artifacts can occur, if at any time within the facility, the ground from the main facility's incoming power source to the Ultrasound unit is only a conduit.
The following are mandatory site requirements. Additional (optional) recommendations, as well as a recommended ultrasound room layout, are provided in Section 2-3-3 "Site Recommendations" .
Door opening is at least 76 cm (30 in) wide.
Proposed location for unit is at least 0.3 m (1 ft.) from the wall for cooling
Clean and protected space for storage of probes (either in their case or on a rack).
Material to safely clean probes (performed using a plastic container, never metal).
Power outlet and place for any external peripheral are within 2 m (6.5 ft) of each other with
peripheral within 1 m of the unit to connect cables.
NOTE: The Voluson® 730Expert has four outlets inside the unit. One is for the monitor and three for on board
peripherals.
In case of network option:
An active network outlet in the vicinity of the ultrasound unit.
A network cable of appropriate length (regular Pin-to-Pin network cable).
An IT administrator who will assist in configuring the unit to work with your local network.
A fixed IP address is required. Refer to the form provided at page 3-50 for network details that are required.
NOTE: All relevant preliminary network port installations at the prepared site must be performed by authorized
contractors. The purchaser of GE equipment must utilize only qualified personnel to perform servicing of the equipment.
2-3-3 Site Recommendations
The following are (optional) site recommendations. Mandatory site requirements are provided in the
Mandatory Site Requirements section, above.
Door is at least 90 cm (3 ft.) wide
Circuit breaker for dedicated power outlet is easily accessible
Sink with hot and cold water
Receptacle for bio–hazardous waste, like used probe sheaths
Emergency oxygen supply
Storage for linens and equipment
Nearby waiting room, lavatory, and dressing room
Dual level lighting (bright and dim)
Lockable cabinet ordered by GE for its software and proprietary manuals
Chapter 2 - Site Preparation 2-7
Page 48
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
2-3-3-1 Recommended Ultrasound Room Layout
Figure 2-1 below shows a floor plan illustrating the recommended layout of the Ultrasound Room and
depicting the minimal room layout requirements.
VOLUSON
35.5 IN.
(92 CM)
Figure 2-1 Recommended Floor Plan 4.3m x 5.2m (14ft x 17ft)
Dedicated Power Outlets
Hospital Network
Cabinet for Software and Manuals (optional)
2-8 Section 2-3 - Facility Needs
Page 49
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
2-3-4 Networking Pre-installation Requirements
2-3-4-1 Purpose of the DICOM Network Function
DICOM (Digital Imaging and Communications in Medicine) services provide the operator with clinically useful features for moving images and patient information over a hospital network. Examples of DICOM services include the transfer of images to workstations for viewing or transferring images to remote printers. As an added benefit, transferring images in this manner frees up the on­board monitor and peripherals, enabling viewing to be done while scanning continues. With DICOM, images can be archived, stored, and retrieved faster, easier, and at a lower cost.
2-3-4-2 DICOM Option Pre-installation Requirements
To configure the Voluson® 730Expert to work with other network connections, the site’s network administrator must provide some necessary information.
To configure the Voluson® 730Expert ultrasound unit to work with other network connections, the network administrator must provide some necessary information.
Use the Connectivity Setup Worksheet on page 3-50 to record required information that must include:
• Voluson® 730Expert Details: DICOM network details for the Voluson® 730Expert unit,
incl. the host name, local port, IP address, AE title and net mask.
• Routing Information: IP addresses for the default gateway and other routers in use at
the site.
• DICOM Application Information: Details of DICOM devices in use at the site, including the DICOM
host name, AE title, DICOM port number and IP addresses.
Installation see: Section 3-11 "Network IP Address Configuration" on page 3-48.
Chapter 2 - Site Preparation 2-9
Page 50
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
This page was intentionally left blank.
2-10 Section 2-3 - Facility Needs
Page 51
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 3 Setup Instructions
Section 3-1 Overview
3-1-1 The Purpose of Chapter 3
This chapter contains information needed to setup the Voluson® 730Expert unit. Included are procedures to receive, unpack and configure the equipment. A worksheet is provided (see: page 3-50) to help ensure that all the required information is available, prior to setup the system.
Table 3-1 Contents in Chapter 3
Section Description Page Number
3-1
3-2
3-3
3-4
3-5
3-6
3-7
3-8
3-9
3-10
3-12
3-11
3-13
Section 3-2 Set Up Reminders
Overview
Set Up Reminders
Receiving and Unpacking the Equipment
Preparing for Set Up
Connection of Auxiliary Devices
Completing the Set Up
Printer Installation
System Configuration
Available Probes
Software/Option Configuration
Connectivity Setup Worksheet
Network IP Address Configuration
Paperwork
3-1
3-1
3-4
3-6
3-8
3-22
3-26
3-41
3-47
3-47
3-50
3-48
3-51
3-2-1 Average Installation Time
Table 3-2 Average Installation Time
Description Average Installation Time Comments
Unpacking the scanner 0.5 hours
Scanner /options / printers 0.5 to 1.5 hours Dependant on the required configuration
DICOM Option 0.5 - 1.5 hours Dependant on the configuration amount
Chapter 3 - Setup Instructions 3-1
Page 52
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-2-2 Installation Warnings
1.) Since the Voluson® 730Expert weighs approximately 136 kg (300 lbs.) without options, two people are required to unpack it. Two people are also required for installing any additional items in excess of 16 kg / 35 pounds.(e.g., Monitor).
2.) There are no operator serviceable components. To prevent shock, do not remove any covers or
panels. Should problems or malfunctions occur, unplug the power cord.
Only qualified service personnel should carry out servicing and troubleshooting.
3-2-2-1 Moving/Lifting the System
When pulling, moving or lifting the system, grasp it only at the rear handle of the trolley and the handle underneath the foot rest.
WARNING
WARNING
Do NOT pull or lift the system with the front handle of the user interface (operator panel).
Figure 3-1 moving or lifting the system
3-2-2-2 System Acclimation Time
After being transported, the Voluson® 730Expert system may be very cold or hot. It requires one hour for each 2.5°C increment it's temperature is below 10°C or above 40°C.
CAUTION
!! CAUTION:
Equipment damage possibility. Turning the system on without acclimation after arriving at site may cause the system to be damaged.
Table 3-3 Acclimation Time
60 55 50 45 40 35 30 25 20 15 10 5 0 -5 -10 -15 -20 -25 -30 -35 -40
°C
140 131 122 113 104 96 86 77 68 59 50 41 32 23 14 5 -4 -13 -22 -31 -40
°F
864200000002468101214161820
hrs
3-2-2-3 Control Panel Position
If weight is placed on the Control Panel (UI) in it’s extended position the console could tip over.
WARNINGWARNING
!! WARNING:
The system should NOT be moved with the Control Panel (UI) extended. Move the Control Panel to it’s centered and locked position. Refer to Section 6-4 on page 6-5.
WARNINGWARNING
!! WARNING:
Monitor mounting mechanism may break if not properly supported (e.g., with packing foam) during transportation.
3-2 Section 3-2 - Set Up Reminders
Page 53
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-2-2-4 Brake Pedal Operation
WARNINGWARNING
!! WARNING:
REMEMBER: If the front wheels are engaged for transportation, pressing the release brake pedals (brakes on front wheels under the foot rest) once disengages the lock.
3-2-3 Safety Reminders
DANGER
!! DANGER:
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH THE UNIT!
CAUTION
!! CAUTION:
CAUTION
!! CAUTION:
CAUTION
!! CAUTION:
CAUTION
!! CAUTION:
CAUTION
!! CAUTION:
Two people should unpack the unit because of its weight. Two people are required whenever a part weighing 16kg (35 lb.) or more must be lifted.
If the unit is very cold or hot, do NOT turn on its power until it has had sufficient time to acclimate to its operating environment.
To prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use a three to two prong adapter. This defeats safety grounding.
Do NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is present.
Do not use a 20 Amp to 15 Amp adapter on the 120 Vac unit’s power cord. This unit requires a dedicated 16 A circuit.
!! CAUTION:
!! CAUTION:
!! CAUTION:
CAUTION
CAUTION
CAUTION
DO NOT operate this unit unless all board covers and frame panels are securely in place, to ensure optimal system performance and cooling. (When covers are removed, EMI may be present).
OPERATOR MANUAL(S) The User Manual(s) should be fully read and understood before operating the Voluson® 730Expert and kept near the unit for quick reference.
ACOUSTIC OUTPUT HAZARD Although the ultrasound energy transmitted from the Voluson® 730Expert probe is within FDA limits, avoid unnecessary exposure. Ultrasound energy can produce heat and mechanical damage.
Figure 3-2 Environmental Labels
Chapter 3 - Setup Instructions 3-3
Page 54
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Section 3-3 Receiving and Unpacking the Equipment
!! CAUTION:
!! NOTICE:
CAUTION
NOTICE
Transport only with forklift or stracker truck. During transport pay attention to the point of gravity (“tilt and drop” indicator)!
Have two people available to unpack the Voluson® 730Expert. Attempts to move the unit considerable distances or on an incline by one person could result in injury or damage or both.
The envelope with delivery address, packing list and invoice is located on the front panel of the crate.
Check whether delivery is complete (according to packing list) and check visual damage!
Figure 3-3 envelope at front panel of the crate
The device must only be transported in the original packaging!
Unpack the devices such a way that packaging can be reused.
A drill with size 20 torx bit and/or a Philips 2 screwdriver will be needed to open the crate.
Table 3-4 Unpacking Procedure
Step Task
1.
Loosen the screws and remove top panel from crate; top panel is used for the ramp.
tilt and drop indicator
center
of
gravity
3-4 Section 3-3 - Receiving and Unpacking the Equipment
Page 55
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Table 3-4 Unpacking Procedure
Step Task
2.
Open front panel by removing the screws from the side panels. Remove probes and accessories from the wooden box.
probes and accessories
WARNINGWARNING
!! WARNING:
Do not open side panels prior to
screws
screws
removing the front panel! Accessories could drop down and DAMAGE the user interface!
packing
3.
4.
Caution: Two people are needed in the next step due to the weight of the equipment.
5.
Loosen all screws to remove the wooden rack for accessories as well as the left, right and back panel.
wooden rack for accessories
Carefully remove foam packing material and plastic bag from the ultrasound unit and monitor.
Disengage the brakes and slowly move unit down the ramp (top panel).
brakes
Brakes
top panel
is used for ramp
Note: Packing crate and material should be stored for future use.
Chapter 3 - Setup Instructions 3-5
Page 56
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
b
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Section 3-4 Preparing for Set Up
3-4-1 Verify Customer Order
1.) After unpacking the equipment, it is important to verify that all items ordered by the customer have been received. Compare all items listed on the packing slip (delivery note) with those received.
!! NOTICE:
NOTICE
It is recommended to keep and store the shipping carton and all other packing materials (including the support foams, anti-static plastic cover, etc.), in case the unit has to be moved to a different location. Unpack the devices such a way that packaging can be reused. For warranty purposes, storage of the above is required for one year from date of purchase.
2.) Visually inspect the system components using the following checklist.
Table 3-5 Damage Inspection Checklist - Voluson® 730Expert System
Step Item Recommended Procedure
1
Main label Enter Serial Number: __________________________ (printed on main label on back of the system)
2
Console Verify that the system is switched OFF and unplugged. Clean the console and control panel.
3
Control Console
4
Probes
5
LCD Display
6
Fans Verify that the system’s cooling fans and peripheral fans are operating.
7
Rear Panel
Physically inspect the control console for missing or damaged items. After switching on the system, verify the proper illumination of all the control panel buttons.
Check all probes for wear and tear on the lens, cable, and connector. Look for bent or damaged pins on the connector and in the connector socket on the unit. Verify that the EMI fingers around the probe connector socket housing are intact. Check the probe locking mechanism and probe switch.
Clean the LCD display by gently wiping with a dry, soft, lint-free non-abrasive folded cloth. Inspect the monitor for scratches and raster burn.
Check the rear panel connectors for bent pins, loose connections and loose or missing hardware. Screw all the cable connectors tightly to the connector sockets on the panel. Verify that the labeling is in good condition.
8
9
10
11
Covers
Peripherals
Power Cord
System Voltage setting
Check that all screws are tightly secured in place, that there are no dents or scratches and that no internal parts are exposed.
Check and clean the peripherals in accordance with the manufacturer’s directions. To prevent EMI or system overheating, dress the peripheral cables inside the peripheral cover.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or any deterioration. Verify continuity. Replace the power cord, as required.
Verify that the Voluson® 730Expert is set to the correct voltage. see: Section 3-4-2 "System Voltage Settings" on page 3-7
NOTE: Report any items that are missing, back-ordered, or damaged, to your GE Healthcare - Kretztechnik
sales representative. The contact address is shown in Contact Information on page 1-11.
3-6 Section 3-4 - Preparing for Set Up
Page 57
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-4-2 System Voltage Settings
Verify that the scanner is set to the correct voltage. The Voltage settings for the Voluson® 730Expert Scanner is found on the identification plate, on the rear of the system.
Figure 3-4 Identification Plate
WARNINGWARNING
!! WARNING:
CONNECTING A Voluson® 730Expert SCANNER TO THE WRONG VOLTAGE LEVEL WILL MOST LIKELY DESTROY THE SCANNER.
3-4-3 EMI Protection
This unit has been designed to minimize the effects of Electo-Magnetic Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system from image artifacts caused by this interference. For this reason, it is imperative that all covers and hardware are installed and secured before the unit is put into operation.
Ensure that the system is protected from electromagnetic interference (EMI), as follows:
Operate the system at least 15 feet away from equipment that emits strong electromagnetic
radiation.
Operate the system in an area enclosed by walls, floors and ceilings comprised of wood, plaster or
concrete, which help prevent EMI.
Shield the system when operating it in the vicinity of radio broadcast equipment, if necessary.
Do not operate mobile phones or other EMI emitting devices in the ultrasound room.
Verify that all EMI rules listed in the following table are followed:
The Voluson® 730Expert environmentally qualified facilities, in terms of the prevention of radio wave interference. Operation of the ultrasound unit television sets situated near the medical equipment. For further details and EMI Prevention/Abatement refer to Section 2-2-3 "EMI Limitations" on page 2-4.
ultrasound unit is approved for use in hospitals, clinics and other
in an inappropriate environment can cause electronic interference to radios and
Chapter 3 - Setup Instructions 3-7
Page 58
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Section 3-5 Connection of Auxiliary Devices
3-5-1 Preparations
NOTE: Normally the auxiliary devices and peripherals come already installed with the system.
1.) Carefully remove plastic caps using a knife and loosen screws and washers (see: Figure 3-5.)
2.) When the cover is loose on top, pull the rear cover out and move upwards.
plastic caps; screws and washers
Figure 3-5 Remove rear cover plate
3.) Disconnect the GND ground-cable from the back of the rear cover plate.
GND ground cable
Figure 3-6 GND ground-cable
4.) Connect Peripherals according to correct connection scheme described in the following sub-
sections (refer to Table 3-6 on page 3-9).
5.) Connect the GND ground-cable at the rear cover plate.
6.) Mount rear cover plate, reattach screw and washers and plug-in the caps.
3-8 Section 3-5 - Connection of Auxiliary Devices
Page 59
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-5-1 Preparations (cont’d)
Table 3-6 below outlines Voluson® 730Expert hardware installation procedures described in the sub-
sections.
Table 3-6 Connection Procedures
Sub-section Description Page Number
!! WARNING:
!! CAUTION:
WARNINGWARNING
CAUTION
3-5-2
3-5-3
3-5-4
3-5-5
3-5-6
3-5-7
3-5-8
3-5-9
3-5-10
3-5-11
Monitor Connection
Foot Switch Connection
ECG-preamplifier Connection
Global Modem Connection
S-VHS Video Recorder Connection
B/W Video Printer Connection
Line Printer Connection
Digital Color Printer Connection
Bluetooth Printer Connection
External USB-Devices
3-10
3-10
3-11
3-12
3-13
3-15
3-16
3-17
3-18
3-20
After each installation, the leakage currents have to be measured according to IEC 60601-1 respectively UL 60601-1.
Please observe that some printers may not be medical devices! If the Bluetooth Printer and/or Line Printers are not medical devices, they have to be located outside of the patient environment (according to IEC 60601-1 / UL 60601-1).
2,5 m
1,5 m1,5 m
1,5 m
Chapter 3 - Setup Instructions 3-9
Page 60
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-5-2 Monitor Connection
NOTE: The monitor comes already installed with the system.
3-5-3 Foot Switch Connection
Figure 3-7 Foot Switch Connection Scheme
NOTE: After physical connection, adjust the Footswitch (Left/Right) as described in
Section 3-7-5 "Remote Control Selection" on page 3-40.
3-10 Section 3-5 - Connection of Auxiliary Devices
Page 61
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-5-4 ECG-preamplifier Connection
Figure 3-8 ECG Connection Scheme
Chapter 3 - Setup Instructions 3-11
Page 62
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-5-5 Global Modem Connection
Figure 3-9 Global Modem Connection Scheme
WARNINGWARNING
!! WARNING:
CAUTION
!! CAUTION:
Never install the telephone wiring, or use the modem during an electric storm; there may be a remote risk of electrical shock from lightning.
Use only the power adapter supplied with the modem and connect it as shown. Use of any other power adapter will void the warranty and could damage the modem.
3-12 Section 3-5 - Connection of Auxiliary Devices
Page 63
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-5-6 S-VHS Video Recorder Connection
NOTE: Please refer to connection scheme belonging to the used VCR type:
Section 3-5-6-1 "Mitsubishi HS-MD3000" on page 3-13
Section 3-5-6-2 "Sony SVO-9500MD" on page 3-14
3-5-6-1 Mitsubishi HS-MD3000
NOTE: There are two types of Mitsubishi HS-MD3000 VCR (PAL and NTSC) approved by GE Healthcare -
Kretztechnik. Use the standard VCR type (PAL or NTSC) that is suited to your region.
!! CAUTION:
CAUTION
ONLY the specific GE - Kretztechnik Version of this video recorder type may be used in connection with the system!
!! NOTICE:
NOTICE
Figure 3-10 Mitsubishi - Video Recorder Connection Scheme
Please use the proper connection set and remote cable. see: Section 9-12 "Optional Peripherals and Accessories" on page 9-26.
Chapter 3 - Setup Instructions 3-13
Page 64
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-5-6-2 Sony SVO-9500MD
Figure 3-11 Sony - Video Recorder Connection Scheme
Set DIP Switches on back of the VCR as shown:
Switch 1: OFF-Audio is muted during fast playback
Switch 2: OFF-FF/REW mode is used during searches.
Switch 3: ON-VCR counter is reset upon cassette ejection.
Switch 4: OFF-The VCR can be controlled by the system.
Switch 5 + 6: ON-The Baud rate is set to 19200 bit/sec.
(must be the same as Voluson® 730Expert)
Figure 3-12 DIP Switches
NOTICE
!! NOTICE:
Please use the proper connection set and remote cable. see: "Section 9-12 Optional Peripherals and Accessories (cont’d)" on page 9-28.
3-14 Section 3-5 - Connection of Auxiliary Devices
Page 65
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-5-7 B/W Video Printer Connection
!! NOTICE:
NOTICE
Figure 3-13 B/W Video Printer Connection Scheme
Set DIP Switches on back of the Black/White printer as shown in the image (all DOWN).
Please use the proper connection set. see: Section 9-12 "Optional Peripherals and Accessories" on page 9-26.
Chapter 3 - Setup Instructions 3-15
Page 66
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-5-8 Line Printer Connection
Figure 3-14 Line Printer Connection Scheme
CAUTION
!! CAUTION:
NOTICE
!! NOTICE:
NOTICE
!! NOTICE:
Please observe that the Line Printer has to be located outside of the patient environment (acc. IEC 60601-1 / UL 60601-1).
The switch of the printer has to be in ON position before starting the system. Leave printer switch always in the ON position.
Please use the proper connection set. see: Section 9-12 "Optional Peripherals and Accessories" on page 9-26.
NOTE: For further installation instructions see:
Section 3-7-1 "Installing Line Printer HP 990cxi or HP 995c" on page 3-27.
3-16 Section 3-5 - Connection of Auxiliary Devices
Page 67
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-5-9 Digital Color Printer Connection
Figure 3-15 Digital Color Printer Connection Scheme
CAUTION
!! CAUTION:
CAUTION
!! CAUTION:
NOTICE
!! NOTICE:
NOTICE
!! NOTICE:
CAUTION
!! CAUTION:
Pay attention to lateral distances. See Instruction Manual of the printer!
The Printer Supply Voltage must be the same as the Output Voltage of the Voluson® 730Expert Power Out Connectors (Power Supply)!
The switch of the printer has to be in ON position before starting the system. Leave printer switch always in the ON position.
Please use the proper connection set. see: Section 9-12 "Optional Peripherals and Accessories" on page 9-26.
The Sony UP-D23MD printer must never be connected to USB-ports on the SBC backplane. Please use one of the two outlets on the back of the GEM (Disk Drive Module) only.
NOTE: For further installation instructions see:
Section 3-7-2 "Installing Digital Color Printer Sony UP-D21MD or UP-D23MD" on page 3-29.
Chapter 3 - Setup Instructions 3-17
Page 68
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-5-10 Bluetooth Printer Connection
NOTE: Please refer to connection scheme belonging to the used B/W Printer type:
Section 3-5-10-1 "HP 5600/5900 Series" on page 3-18
Section 3-5-10-2 "Canon Pixma MP600 / MP610" on page 3-19
3-5-10-1 HP 5600/5900 Series
Figure 3-16 HP Bluetooth Printer - Connection Scheme
CAUTION
!! CAUTION:
CAUTION
!! CAUTION:
NOTICE
!! NOTICE:
NOTICE
!! NOTICE:
Please observe that the complete Bluetooth Printer Assembly has to be located outside of the patient environment (acc. IEC 60601-1 / UL 60601-1).
The printer being used may not be a medical device. The Bluetooth Printer Set and the Power Supply of the Bluetooth Printer Adapter is also not a medical device. The equipment meets the requirements of the EN60950 Standard.
The switch of the printer has to be in ON position before starting the system. Leave printer switch always in the ON position.
Please use the proper Bluetooth Printer Connection set. see: Table 9-15 "Optional Peripherals and Accessories" on page 9-27.
3-18 Section 3-5 - Connection of Auxiliary Devices
Page 69
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-5-10-2 Canon Pixma MP600 / MP610
Canon Bluetooth
Printer Adapter
!! CAUTION:
!! CAUTION:
!! NOTICE:
CAUTION
CAUTION
NOTICE
Figure 3-17 Canon Bluetooth Printer - Connection Scheme
Please observe that the complete Bluetooth Printer Assembly has to be located outside of the patient environment (acc. IEC 60601-1 / UL 60601-1).
The Printer and the Bluetooth Printer Adapter used may not be a medical device. The equipment meets the requirements of the EN60950 Standard.
The switch of the printer has to be in ON position before starting the system. Leave printer switch always in the ON position.
Chapter 3 - Setup Instructions 3-19
Page 70
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-5-11 External USB-Devices
WARNINGWARNING
!! WARNING:
Do not connect or disconnect any external USB-devices to or from the system while scanning a patient! The appearing dialogs could distract you from the scan!
3-5-11-1 External USB-Devices - Connection
When an external USB-storage device (such as a USB-memory stick or external hard disk) is connected to the Voluson® 730Expert system, Windows detects the device and automatically installs a driver. During this process, several dialogs may pop up, starting with the „Found New Hardware“ dialog.
Figure 3-18 Found New Hardware - USB Device
The device is then accessible using the drive letter the system assigned to it.
NOTICE
!! NOTICE:
When connecting external USB devices, be sure to execute Safety Directions found in Chapter 2 of the Voluson® 730Expert Basic User Manual.
3-5-11-2 External USB-Devices - Disconnection
Before an external USB-device (e.g., USB-memory stick) can be disconnected, the system has to be informed about the removal of the device! For this purpose the System Setup -
Figure 4-22 on page 4-29) has a STOP USB DEVICES
BACKUP page (see:
button.
!! CAUTION:
CAUTION
Unplugging or ejecting USB devices without first stopping them can often cause the system to crash and possibly result in loss of valuable data.
By clicking the STOP USB DEVICES
button, the „Unplug or Eject Hardware“ dialog is started.
Using this dialog, the USB-devices can be stopped before they are physically disconnected.
Figure 3-19 Unplug or Eject Hardware
3-20 Section 3-5 - Connection of Auxiliary Devices
Page 71
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-5-11-2 External USB-Devices - Disconnection (cont’d)
The „Unplug or Eject Hardware“ dialog shows all USB-devices that are connected to the system. On every system is an USB mass storage device, the DVD/CD writer, which has the drive letter (F:). If the system has an optional MO-drive installed as well, it is listed too, and has the drive letter (E:).
To stop the external device, select it and click the STOP
button.
A dialog shows which components will be stopped. To finish the process, click [OK].
Figure 3-20 Stop a Hardware device
Finally, a dialog shows that the device was stopped successfully. The device can now be safely disconnected from the system.
!! CAUTION:
CAUTION
Figure 3-21 Safe to Remove Hardware
By clicking OK
Close this dialog by clicking CLOSE
Afterwards select OK
, the „Unplug or Eject Hardware“ dialog is active again.
.
to reboot the system.
If the system’s DVD/CD writer or (optional) MO-drive was stopped by accident, simply stop the external device as well and reboot the system. During reboot, the DVD/CD writer and the MO-drive will be installed again.
Chapter 3 - Setup Instructions 3-21
Page 72
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Section 3-6 Completing the Set Up
3-6-1 Connecting the Unit to a Power Source
The connection of the Voluson® 730Expert ultrasound unit to a power source should be performed by a qualified person who has completed basic Voluson® 730Expert System User Training. Use only the power cords, cables and plugs provided by or designated by GE Healthcare - Kretztechnik to connect the unit to the power source.
CAUTION
!! CAUTIO N: Prior to
Prior to connect the Voluson® 730Expert unit to a power source, verify compliance with all electrical and safety requirements. Check the power cord to verify that it is intact and of hospital-grade. Products equipped with a power source (wall outlet) plug should be connected to the fixed power socket that has a protective grounding conductor. Never use an adapter or converter to connect with a power source plug (e.g., a three-prong to two-prong converter).
WARNINGWARNING
!! WARNING:
The unit’s power must be supplied from a separate, properly rated outlet to avoid risk of fire. See Section 2-2-2-1 "Voluson® 730Expert Power Requirements" on page 2-3 for rating information. The power cord should not, under any circumstances, be altered to a configuration rated less than that specified for the current.
CAUTION
!! CAUTION:
CAUTION
!! CAUTION:
Whenever disconnecting the Voluson® 730Expert system from the electrical outlet, always observe the safety precautions. First unplug the main power cable from the wall outlet socket, then from the unit itself. Remove by pulling on the cable connector - DO NOT pull on the cable.
The Voluson® 730Expert requires all covers! Operate this system only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
3-6-2 Power On / Boot Up
3-6-2-1 Scanner Power On
1.) Connect the Power Cable to the back of the system.
2.) Connect the Main Power Cable to a hospital grade power receptacle with the proper rated voltage.
Never use an adapter that would defeat the safety ground.
3.) CPN6 only - Switch ON the Circuit Breaker (F1) and the Power Switch of peripherals (F2).
4.) CPN80-81 only - Switch ON the Circuit Breaker at the rear of the system.
Circuit Breaker
Connector for Mains Power Cable
Figure 3-22 Circuit Breaker at rear of system with CPN6 installed
NOTICE
!! NOTICE:
When AC power is applied to the scanner, the ON/OFF switch on the Control panel is illuminated, indicating that the System (including the Back-end Processor) is in standby mode.
3-22 Section 3-6 - Completing the Set Up
Power Switch of Peripherals
Page 73
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-6-2-2 Back-end Processor Boot Up
Press the ON/OFF Standby switch left below the control panel.
ON/OFF Standby switch
Figure 3-23 ON/OFF Standby Switch
NOTE: The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the
Standby switch. The switch of printers has to be in ON position before starting the system. However, be aware some auxiliary equipment may switch itself to standby mode when Standby power is on (e.g., Color video printer) and must therefore be switched on separately. (CPN6 only - Auxiliary equipment need not to be switched ON/OFF separately if the F2 power switch on the rear of the system is always ON.)
When the ON/OFF Standby switch left below the Control Panel is pressed, the System (including the Back-end Processor) starts and the operating system is loaded which then leads the application software to activate the scanner.
The system automatically performs an initialization sequence which includes the following:
- Loading the operating system.
- Running a quick diagnostic check of the system.
- Detecting connected probes
As soon as the software has been loaded, the system enters 2D-Mode with the probe and application that were used before the system was shut down.
NOTE: Total time used for start-up is about 2 minutes.
Chapter 3 - Setup Instructions 3-23
Page 74
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-6-3 Power Off / Shutdown
NOTICE
!! NOTICE:
After turning off a system, wait at least 10 seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
3-6-3-1 Back-end Processor Power Down
1.) Press the ON/OFF Standby switch left below the Control Panel. see: Figure 3-23.
3-6-3-2 Scanner Shutdown
1.) Press the ON/OFF Standby switch left below the Control Panel.
2.) Switch OFF the Circuit Breaker at the rear of the system.
NOTE: The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the
Standby switch. So the auxiliary equipment need not to be switched ON/OFF separately.
WARNINGWARNING
!! WARNING:
Disconnection of the Main Power Cable is necessary! For Example: When repairing the system.
3.) After complete power down, unscrew the 2 screws and remove the pull-out protection to disconnect
the main power cable from the system or unplug it from the AC wall outlet socket.
Circuit Breaker
Mains Power Cable
Figure 3-24 Circuit Breaker, Protection and Power Cable on back of Voluson® 730Expert
CAUTION
!! CAUTION:
2 screws
pull-out Protection
4.) Press once on the brakes to block the front wheels (brakes on front wheels under the foot rest).
5.) Prior to disconnect a probe freeze the image.
6.) Disconnect probes. (Turn the probe locking handle counterclockwise and then pull the connector
straight out of the probe port.)
7.) If required, open the right-hand side door, remove the probe cable from the cable holder and then
close the door.
DO NOT disconnect a probe while running (Live Scan “Write” mode)! A software error may occur. In this case switch the unit OFF (perform a reset).
3-24 Section 3-6 - Completing the Set Up
Page 75
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-6-4 Transducer Connection
Connect a transducer to one of the three rightmost transducer receptacle as follows:
1.) Inspect the probe and probe socket to verify that it is free of debris.
2.) Ensure that the transducer twist lock lever is at horizontal position.
3.) Insert the connector on the receptacle guide pin until it touches the receptacle mating surface.
4.) Twist the transducer twist lock lever to vertical position to lock it in place.
Twist the lever to the horizontal position to disconnect the transducer.
5.) Open the right-hand side door, lay the cable into the intended cable holder and close the door.
Figure 3-25 Transducer Connection
CAUTION
!! CAUTION:
CAUTION
!! CAUTION:
If the cable spout on the right-hand door is missing, don’t pull the probe cable. Please insert the spout in the designated place to avoid damage of the probe cable.
Do not bend the probe cable acutely. Fault conditions can result in electric shock hazard. Do not touch the surface of probe connectors which are exposed when the probe is removed. Do not touch the patient when connecting or disconnecting a probe.
NOTE: Prior to connecting or disconnecting a probe, freeze the image.
It is not necessary to turn OFF power to connect or disconnect a transducer.
Chapter 3 - Setup Instructions 3-25
Page 76
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Section 3-7 Printer Installation
NOTE: For Connection schemes refer to Section 3-5 "Connection of Auxiliary Devices" on page 3-8.
For further installation instructions see:
Section 3-7-1 "Installing Line Printer HP 990cxi or HP 995c" on page 3-27
Section 3-7-2 "Installing Digital Color Printer Sony UP-D21MD or UP-D23MD" on page 3-29
Section 3-7-3 "Printer Installation manually" on page 3-31
Section 3-7-4 "Adjustment of Printer Settings" on page 3-36
!! CAUTION:
CAUTION
The Bluetooth Printer Connection set as well as the Color Deskjet printer MUST NOT be installed by the user!
For installation please contact your local distributor or GE service representative.
3-26 Section 3-7 - Printer Installation
Page 77
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-1 Installing Line Printer HP 990cxi or HP 995c
1.) Power off/Shutdown the system as described in: Section 3-6-3 on page 3-24.
2.) Connect the printer as described on page 3-8 and reinstall the rear cover plate.
NOTE: For connection schemes refer to Section 3-5-8 on page 3-16.
NOTICE
!! NOTICE:
Do not connect the USB-cable to the printer!
NOTE: “Mouse functions” can be performed by using the trackball for moving the cursor.
“Normal select” (Click) = left/right trackball key; “Opening a context menu” = upper trackball key
3.) Turn ON the printer, then switch ON the power of the system and wait till the system has booted.
NOTE: The power switch of the printer has to be in ON position before starting the system!
3-7-1-1 Install the HP printer software/driver
Perform the following steps if this printer was never installed on the Voluson® 730Expert!
1.) Connect the USB cable to the printer and the system. The windows ‘Searching for Drivers …’,
‘Found new Hardware …’ and finally the following windows appear.
select “Search for a suitable driver for my device (recommended)”
Click on “Specify a location”
Figure 3-26 Found New Hardware Wizard
Chapter 3 - Setup Instructions 3-27
Page 78
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-1-1 Install the HP printer software/driver (cont’d)
2.) Click the NEXT
3.) Use the BROWSE
buttons to start the Hardware Wizard and to locate the driver files.
button to search the following path on the hard disk (see: Figure 3-27)
- C:\Utilities\PrinterDriver\HPDeskjet990c for HP 990cxi
- C:\Utilities\PrinterDriver\HPDeskjet995c for HP 995c, and then click OK
Figure 3-27 Search for Network path on hard disk
4.) Confirm the correct path and click NEXT
5.) Confirm the installation by clicking FINISH
to install the driver. All necessary files are copied.
to close the Hardware Wizard.
.
Figure 3-28 Confirm correct path and finish the Installation
6.) Close all open windows and restart the system (turn off and on the system).
NOTICE
!! NOTICE:Ver
After boot up of the system, verify the correct settings in the printer “Properties”, see: Section 3-7-4
"Adjustment of Printer Settings" on page 3-36.
3-28 Section 3-7 - Printer Installation
Page 79
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-2 Installing Digital Color Printer Sony UP-D21MD or UP-D23MD
1.) Power off/Shutdown the system as described in: Section 3-6-3 on page 3-24.
2.) Connect the printer as described on page 3-8 and reinstall the rear cover plate.
NOTE: For connection schemes refer to Section 3-5-9 on page 3-17.
NOTICE
!! NOTICE:
Do not connect the USB-cable to the printer!
NOTE: “Mouse functions” can be performed by using the trackball for moving the cursor.
“Normal select” (Click) = left/right trackball key; “Opening a context menu” = upper trackball key
3.) Turn ON the printer, then switch ON the power of the system and wait till the system has booted.
NOTE: The power switch of the printer has to be in ON position before starting the system!
3-7-2-1 Install the UP-D21MD / UP-D23MD printer software/driver
Perform the following steps if this printer was never installed on the Voluson® 730Expert!
1.) Connect the USB cable to the printer and the system. The windows ‘Searching for Drivers …’,
‘Found new Hardware …’ and finally the following windows appear.
2.) Click the NEXT
select “Search for a suitable driver for my device (recommended)”
Click on “Specify a location”
Figure 3-29 Found New Hardware Wizard
buttons to start the Hardware Wizard and to locate the driver files.
Chapter 3 - Setup Instructions 3-29
Page 80
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-2-1 Install the UP-D21MD / UP-D23MD printer software/driver (cont’d)
3.) Use the BROWSE
button to search the following path (see: Figure 3-27)
• C:\Utilities\PrinterDriver\SonyUP-D21MD or
• C:\Utilities\PrinterDriver\SonyUP-D23MD and then click OK
.
Figure 3-30 Search for Network path
4.) Confirm the correct path and click NEXT
to install the driver. All necessary files are copied.
5.) The Message: Digital Signature Not Found appears. Click YES
6.) Confirm the installation by clicking FINISH
to close the Hardware Wizard.
.
Figure 3-31 Confirm correct path and finish the Installation
7.) Close all open windows and restart the system (turn off and on the system).
NOTICE
!! NOTICE:Ver
After boot up of the system, verify the correct settings in the printer “Properties”, see: Section 3-7-4
"Adjustment of Printer Settings" on page 3-36.
3-30 Section 3-7 - Printer Installation
Page 81
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-3 Printer Installation manually
1.) On the Touch Panel, press UTILITIES.
2.) In the Utilities menu, touch SYSTEM SETUP
3.) Select the
SERVICE page. The “password window” appears automatically.
to invoke the setup desktop on the screen.
Figure 3-32 System Setup Service page
4.) Enter the password SHE and click the ACCEPT
button to display the Service Tools window.
“Printer” button
Figure 3-33 Service Tools window
5.) Click on the PRINTER
button.
Chapter 3 - Setup Instructions 3-31
Page 82
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-3 Printer Installation manually (cont’d)
6.) Click the ADD PRINTER
button.
A warning message appears: Please read this message carefully and click YES
if you have skills to do this.
Figure 3-34 Add Printer and Printer Installation/Properties
7.) Click the NEXT
button to start the Add Printer Wizard.
8.) Select the ‘Local Printer‘, deselect “Automatically install Plug and Play printer” and then click NEXT
.
Figure 3-35 Add Printer Wizard
3-32 Section 3-7 - Printer Installation
Page 83
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-3 Printer Installation manually (cont’d)
9.) Select the corresponding Printer Port (e.g., Figure 3-36 = USB001) and click NEXT
Figure 3-36 Select Printer Port
10.)In the following window select the HAVE DISK
button.
.
11.)Use the BROWSE
Figure 3-37 Have Disk...
button to search the Printer Driver path (C:\Utilities\PrinterDriver\xxxx).
Chapter 3 - Setup Instructions 3-33
Page 84
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-3 Printer Installation manually (cont’d)
Figure 3-38 Select Printer Driver path (C:\Utilities\PrinterDriver\....)
12.)Click OPEN
13.)Verify the selected Printer Driver path and confirm with OK
, select the “xxx.inf” file and click OPEN again.
.
Figure 3-39 verify selected Printer Driver path
14.)Select the manufacturer and model of your printer and confirm with the NEXT
button.
3-34 Section 3-7 - Printer Installation
Page 85
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-3 Printer Installation manually (cont’d)
15.)Assign a name, decide if the printer should be used as default printer and confirm with NEXT
.
see: Figure 3-40.
Figure 3-40 Assign name and select Printer Sharing - no
16.)Select “Do not share this printer” and confirm the “Printer Sharing” window (Figure 3-40) by clicking
NEXT
.
17.)The “Complete the Add Printer Wizard” window appears on the screen.
!! NOTICE:Ver
NOTICE
Figure 3-41 Complete manual Printer Installation
18.)Complete the manual Printer Installation with the FINISH
19.)Close all open windows, close the “System Setup” with SAVE & EXIT
button.
and restart the system
(turn off and on the system).
After boot up of the system, verify the correct settings in the printer “Properties”, see: Section 3-7-4
"Adjustment of Printer Settings" on page 3-36.
Chapter 3 - Setup Instructions 3-35
Page 86
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-4 Adjustment of Printer Settings
1.) After system restart, touch the UTILITIES key, and then SYSTEM SETUP on the Touch Panel.
2.) Select the
3.) Enter the password SHE and click the ACCEPT
4.) Click on the PRINTER
5.) Select the desired printer from the pull-down menu and click the EDIT SETTINGS
6.) Confirm the warning message with the YES
SERVICE page. The “password window” appears automatically.
button.
button.
button.
button. The “Printer Properties” appear.
!! WARNING:
WARNINGWARNING
hp deskjet 990c
select the desired printer from the pull-down menu
Figure 3-42 Select the desired printer
To adjust the Line printer see: Section 3-7-4-1 "HP 990cxi / HP 995c - Printer Settings" .
To adjust the Color printer see: Section 3-7-4-2 "UP-D21MD / UP-D23MD - Printer Settings" .
After each printer installation, the leakage currents have to be measured acc. IEC 60601-1 resp. UL60601-1.
3-36 Section 3-7 - Printer Installation
Page 87
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-4-1 HP 990cxi / HP 995c - Printer Settings
1.) Call up the ‘Printer Properties’; operation see: Section 3-7-4 "Adjustment of Printer Settings" .
2.) Select the
3.) Select the
GENERAL page and click the PRINTING PREFERENCES... button.
SETUP page for adjusting print quality and paper size.
“Paper type” should be set to ‘Automatic’. See left Figure 3-43.
4.) Select the
“Orientation” must
FEATURES page for adjusting ‘Two-Side Printing’ if desired.
be set to ‘Portrait’. See right Figure 3-43.
Figure 3-43 HP Printer - Settings
5.) For saving the adjusted printer settings click APPLY
6.) Select the
PORTS page and select/verify the correct USB printer port.
7.) For saving the adjusted printer settings click APPLY
and then OK.
and then OK.
Finally close the ‘Printers’ -window with the close button and exit System Setup with SAVE&EXIT
8.) Assign the HP 990cxi / HP 995c printer as Report Printer;
see: Section 3-7-5 "Remote Control Selection" on page 3-40.
9.) Print report page(s) containing measurements.
For operation see Basic User Manual of Voluson® 730Expert.
10.)Turn off the system!
.
Chapter 3 - Setup Instructions 3-37
Page 88
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-4-2 UP-D21MD / UP-D23MD - Printer Settings
1.) Call up the ‘Printer Properties’; operation see: Section 3-7-4 "Adjustment of Printer Settings" .
!! NOTICE:
NOTICE
Settings for Paper Size MUST match with the used Paper (large/small) and also the right color ink cartridge has to be used. Otherwise you will get an error message at printing.
2.) Select the
PAPER page and select:
- Paper Size: UPC-21L (large) / UPC-21S (small)
- Orientation: Landscape (recommended when using large paper size)
- High Speed (check mark on)
Figure 3-44 Paper page
3-38 Section 3-7 - Printer Installation
Page 89
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-4-2 UP-D21MD / UP-D23MD - Printer Settings (cont’d)
3.) Select the
GRAPHICS page. From the “Color Adjust” pop-up menu select:
a.) Color Balance: Cyan = 0; Magenta = 0; Yellow = 0 b.) Gamma Select: Gamma 1
Figure 3-45 Graphics page (Color Balance + Gamma Select)
c.) Color Correction: set Printer Hardware Color Correction d.) Lightness: Sharpness = 7 or 8; Dark = 0; Gamma = -12; Light = 8
Figure 3-46 Graphics page (Color Correction + Lightness)
4.) For saving the adjusted printer settings click APPLY
and then OK.
Finally close the ‘Printers’-window with the close button and exit System Setup with SAVE&EXIT
5.) Assign the Printer to remote keys PRINT A
and/or PRINT B; see: Section 3-7-5 on page 3-40.
Chapter 3 - Setup Instructions 3-39
.
Page 90
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-7-5 Remote Control Selection
To assign an auxiliary device (e.g., printer) to the remote keys PRINT A and/or PRINT B, or to adjust Foot switch, etc.:
1.) On the Touch Panel, touch UTILITIES
2.) In the Utilities menu, touch SYSTEM SETUP
3.) Select the
VCR Source
Print A
.
to invoke the setup desktop on the screen.
PERIPHERALS page.
Foot Switch Remote Control
Report Printer
Print B
Figure 3-47 Peripherals page
Remote Print A: Select the desired Printer for the remote control PRINT A
Remote Print B: Select the desired Printer for the remote control PRINT B
NOTE: Optionally the Remote Control can be done by Foot switches.
Therefore select “Print A” or “Print B” in “Foot Switch Left” or “Foot Switch Right” - section.
Report Printer: Select the desired Report Printer from the drop-down menu.
NOTICE
!! NOTICE:
The selected Report Printer is usually used for printing reports and images from Sonoview.
Foot Switch Left/Right: Select desired function of the Foot switch Left and Right.
After adjustment, click SAVE&EXIT
to save Settings and exit System Setup.
key.
key.
3-40 Section 3-7 - Printer Installation
Page 91
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Section 3-8 System Configuration
3-8-1 System Setup
Modifications of system parameters are supported by diverse dialog pages and windows on the system setup desktop:
General - Date, Time, Clinic Name, Language, Screen saver, etc.
User Settings - to save User programs, 3D/4D programs, Auto Text, Doppler 2D Refresh, etc.
Peripherals - to adjust assignment of PRINT
•Option - shows which options are installed in the system
Service - enter the password to get access to the Service Tools functions
Backup - Save/Load User Settings Only, Save/Load/Delete Full Backup
•Network - to set up all DICOM, Sonoview and Network configuration nodes
System Info - shows which Software/Hardware version is installed in the system
keys, Foot Switch, selection of Save Destination, etc.
NOTICE
!! NOTICE:
More detailed information pertaining System Setup adjustments is found in the Voluson® 730Expert Basic User Manual; see: Table 9-17, “System Manuals,” on page 9-29.
3-8-1-1 To invoke the Setup procedure:
1.) Touch the UTILITIES
key on the Touch Panel.
2.) Touch the SYSTEM SETUP
In general operations are done with the trackball and the trackball keys (mouse emulation).
Trackball (mouse position):
positions the pointing device (arrow) on the desktop
left trackball key (left mouse button):
sets, fixates markers and activates pages/buttons etc. marked by the pointing device
upper trackball key (right mouse button):
no function in system desktop
key in the “Utilities” menu to activate the setup desktop screen.
right trackball key (left mouse button):
sets, fixates markers and activates pages/buttons etc. marked by the pointing device
Chapter 3 - Setup Instructions 3-41
Page 92
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-8-1-2 How to enter Date and Time
Select the
1.) Select the “Date Format” (only one can be active).
2.) Click the DATE/TIME
3.) Click the TIME FORMAT
4.) Click SAVE&EXIT
select Time Format
select Date/Time
and Time Zone
enter Clinic Name
GENERAL page in the System Setup see: Figure 3-48.
button to activate a sub dialog window to enter date, time and time zone.
button to activate a sub dialog window to choose the preferred time
format.
to save Settings and exit System Setup.
select Date Format
Display options
change Language
Exit without saving
Figure 3-48 System Setup - General page
3-8-1-3 How to enter Hospital Name
Select the
GENERAL page in the System Setup see: Figure 3-48.
1.) Select the text box to enter a new “Clinic Name” with the keyboard.
2.) Click SAVE&EXIT
to save Settings and exit System Setup.
The clinic name will be copied into the Hospital ID in the information header.
3-8-1-4 How to change Language
Select the
GENERAL page in the System Setup see: Figure 3-48.
1.) Select the desired language from the pop-up menu.
2.) Click SAVE&EXIT
to save Settings and exit System Setup.
NOTE: After changing the language the system has to reboot.
3-8-1-5 How to change Touch Panel Scheme
Select the
USER SETTING page in the System Setup.
1.) Select the desired color scheme display of the Touch Panel from the pop-up menu.
2.) Click SAVE&EXIT
to save Settings and exit System Setup.
Save Settings & Exit
3-42 Section 3-8 - System Configuration
Page 93
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-8-2 On-Board Optional Peripherals
Mains outlets: Mains socket ST1, ST2, ST3, ST4, ST5 for accessories. All mains outlets are co-switched by the unit’s mains switch via built-in isolation transformer. Output voltage for: ST1 - ST5: 115V or 230V.
!! CAUTION:
CAUTION
Modification of voltage setting only by an authorized service person! The maximum power consumption of equipment (inclusive color video monitor) connected to these outlets must not exceed 350VA!
Table 3-7 Approved Peripherals
Device Manufacturer Model Video Signal
B/W Video Printer SONY UP-895MD NTSC/PAL
Digital Color Video Printer
Video Cassette Recorder
Line Printer
Color Deskjet Printer (Bluetooth)
Magneto-Optical Drive (MOD) Fujitsu MCM3130AP
Global Modem Multi-Tech MT5634ZBA
ECG Preamplifier MAN6
Footswitch MFT7
SONY UP-D21MD USB-Port
SONY UP-D23MD USB-Port
SONY
Mitsubishi
Hewlett Packard hp deskjet 990cxi USB-Port
Hewlett Packard hp deskjet 995c USB-Port
Hewlett Packard HP 5600 Series USB-Port (Bluetooth)
Olivetti Olivetti Job Jet 210 USB-Port (Bluetooth)
Hewlett Packard HP 5940 USB-Port (Bluetooth)
Canon Pixma MP600 / MP610 USB-Port (Bluetooth)
SVO-9500MD
SVO-9500-MDP
HS-MD3000U HS-MD3000E
NTSC
PAL
NTSC
PAL
Chapter 3 - Setup Instructions 3-43
Page 94
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-8-3 External I/O Connection Panel (GES)
1 2 3 4 5 6
Figure 3-49 External I/O Panel Connectors
Table 3-8 External I/O Connector Descriptions
Item Connector Name Table Number Description
1
VGA OUT
2
MODEM
3
NETWORK
4
USB-1
5
USB-2
6
COMP VIDEO OUT
Table 3-9 print out VGA signal with monitor/printer
Table 3-10 RJ-11 with global adapter kit for modem connection
Table 3-11
Table 3-12 USB-2.0 port
Table 3-12 USB-2.0 port
Table 3-13 BNC Connector, Color Video Output
DICOM input/output
twisted pair RJ-45 10/100 megabit/s
3-44 Section 3-8 - System Configuration
Page 95
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-8-3-1 External I/O Pin Outs
Table 3-9 VGA OUT Connector, Sub-D 15 Pin
Pin No Output Signal Description
1 VGA OUT1 R Red
2 VGA OUT1 G Green
3 VGA OUT1 B Blue
4, 9,11,12,15 N/C N/C
5, 6, 7, 8, 10 GND GND
13 VGA OUT1 HS H Sync
14 VGA OUT1 VS V Sync
Table 3-10 Modem, RJ-11 Modular 6 Pin
Pin No Output Signal Description
2 TEL L4 Telephone L4
3 TEL L2 Telephone L2
4 TEL L1 Telephone L1
5 TEL L3 Telephone L3
Others NC Non-connection
Table 3-11 Network Connector, RJ45 Modular 8 Pin
Pin No Output Signal Description
1 ETHER TD Ethernet RD+
2 ETHER TD Ethernet RD-
3 ETHER RD Ethernet TD+
6 ETHER RD Ethernet TD-
Others NC Non-connection
Table 3-12 USB 1, 2 Connectors
Pin No Output Signal Description
1 VCC USB Power Supply
2 - Data USB Data (-)
3 + Data USB Data (+)
4 GND USB Power Ground
Chapter 3 - Setup Instructions 3-45
Page 96
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Table 3-13 COMP VIDEO OUT Connector
Pin No Signal Description
1 (Center Pin) Signal
2 (Coax Pin) Signal GND
Table 3-14 Footswitch Connector (located at Power Supply Module CPN - rear side)
Pin No Input Signal Description
1 Signal GND
2 right switch
3 left switch
4 not connected
3-8-4 Video Specification
Video specifications may be needed to be able to connect laser cameras or other devices to the Voluson® 730Expert.
Table 3-15 Video Specifications VGA Connector
Timing Parameter Value
Visible Resolution
Horizontal Rate [kHz]
H Total cycle time [µs]
H Sync width [µs]
H Back Porch [µs]
normally open
normally open
not connected
800 x 600
47.20
21.20
1.50
2.94
H Active Video Time [µs]
H Front Porch [µs]
Horizontal +/-
Vertical Rate [Hz]
Vertical Total cycle time [ms]
V Sync Width [ms]
V Back Porch [ms]
V Active Video Time [ms]
V Front Porch [ms]
Dot Clock [MHz]
16.15
0.616
pos
75.00
13.30
0.170
0.276
12.68
0.174
49.54
3-46 Section 3-8 - System Configuration
Page 97
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-8-4-1 Electrical Specifications on VGA Connector
Signal Level: 700 mV at 75 Ohm
H/V Sync: TTL Level
Table 3-16 Video Specifications for Composite, Video Connector
S-Video Output set to:
Timing Parameter
Visible Resolution
Pixel Clock
Horizontal Total Line
Horizontal Active Display
Horizontal Front Porch
Horizontal Sync Width
Horizontal Back Porch
Vertical Total Lines
Vertical Active Lines
Vertical Front Porch
Vertical Sync
Vertical Back Porch
Serration Pulses
Interlaced
Aspect Ratio pixel size
PAL
50Hz
800 x 600 800 x 600
17.734475 MHz = 4* ft 14.318 MHz = 4* ft
64µs / 1135 Pixel 63.56µs / 910 Pixel
50.50µs / 902 Pixel 52.50µs / 752 Pixel
1.96µs / 35 Pixel 1.62µs / 23 Pixel
4.62µs / 82 Pixel 4.68µs / 67 Pixel
6.52µs / 116 Pixel 4.76µs / 68 Pixel
20ms / 312.50 Lines 16.68ms / 262.50 Lines
18.18ms / 284 Lines 15.22ms / 239.50 Lines
256us 4 Lines 381us / 6 Lines
160us / 2.50 Lines 190us 7 3 Lines
1408us / 22 Lines 890us / 14 Lines
56
yes yes
14.75 (H) : 17.734475 (V) 14.75 (H) : 14.318 (V)
NTSC
60Hz
Video levels on 75 Ohm:
white level
black level
blanking level
sync level*
Section 3-9 Available Probes
See Chapter 9 - Probes, for part numbers to be used when ordering new or replacement service probes.
Section 3-10 Software/Option Configuration
For description refer to:
Section 3-8-1 "System Setup" on page 3-41
NOTICE
!! NOTICE:
More detailed information pertaining System Setup and Measure Setup adjustments is found in the Voluson® 730Expert Basic User Manual, which is available in different languages.
Chapter 3 - Setup Instructions 3-47
1020mV 1020mV
370mV 370mV
320mV 320mV
20mV 20mV
Page 98
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Section 3-11 Network IP Address Configuration
NOTE: Following Information must be provided by customer or hospital engineer before you can start:
A Station name, AE Title, IP address and Port Number for the Voluson® 730Expert. The IP addresses for the default gateway and other routers at the site for ROUTING INFORMATION. Only if necessary (e.g. for Internet access).
1.) Touch the UTILITIES
2.) Select SYSTEM SETUP
3.) Click the NETWORK CONFIGURATION
4.) The “Internet Protocol (TCP/IP) Properties” dialog page appears.
key on the Touch Panel once to display the Utilities menu.
and open the NETWORK page on the System Setup desktop screen.
button, read the message and confirm with YES.
click Network Configuration button, read message and afterwards confirm with Yes
Figure 3-50 Network Configuration
You can get IP settings assigned automatically if your network supports this capability.
You need to ask your network administrator for the appropriate settings.
Type in:
• IP address
• Subnet mask
• Default gateway
NOTE:This example shows fictional numbers!
Figure 3-51 Internet Protocol (TCP/IP)
To specify a DICOM Address, follow the instructions of described in the Basic User Manual, Chapter 17 of the Voluson® 730Expert.
3-48 Section 3-11 - Network IP Address Configuration
Page 99
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-11-1 Map Network Drive
1.) Select the MAP NETWORK DRIVE button (in the System Setup - Network page, see: Figure 3-50)
to open a dialog where the system can be connected to a shared network drive of another server.
Figure 3-52 Map Network Drive window
2.) Enter the name of the shared network folder in the „Network Folder Name“ field.
3.) Supply a valid user name and a password for this folder.
NOTE: If you check the „Automatic Reconnect“ box, the system tries to establish the connection
again when starting up. Otherwise, the connection must be re-established manually after a shutdown or reboot.
!! NOTICE:
!! NOTICE:
!! WARNING:
NOTICE
NOTICE
WARNINGWARNING
4.) Select the CONNECT
If successful, the DISCONNECT
If there is an error during the connection, a warning message appears inside the dialog. In this case, please verify the data in the dialog.
If there already is a connection to the remote server, the CONNECT button is grayed. To change the existing connection, first click on DISCONNECT
Please make sure that the server you are connecting to is trustworthy and reliable. For details, contact your local system administrator.
button to establish the connection to the remote machine.
button becomes active.
and then enter the new settings.
If you backup Sonoview data to this server, all the patients’ demographic data will be copied to this server!
Chapter 3 - Setup Instructions 3-49
Page 100
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03) D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Section 3-12 Connectivity Setup Worksheet
Site System Information
Site:
Dept:
Serial #:
CONTACT INFORMATION
Name Title
Type:
TCP/IP Settings
Name - AE Title:
Floor:
Room:
REV:
Phone E-Mail Address
Comments:
IP Settings
IP Address:
Subnet Mask:
Default Gateway:
Services (Destination Devices)
Device Type
1 2 3 4 5 6 7 8 9 10 11 12
Manufacturer
Name
Remote Archive Setup
Remote Archive IP:
Remote Archive Name:
IP Address
Port
AE Title
3-50 Section 3-12 - Connectivity Setup Worksheet
Loading...