viHygroGuard MMY 2650 Installation and Operation Manual
Preface
Information Paragraphs
• Note paragraphs provide information that provides a deeper understanding of the
situation, but is not essential to the proper completion of the instructions.
• Important paragraphs provide information that emphasizes instructions that are
essential to proper setup of the equipment. Failur e to follow these instructions car efully
may cause unreliable performance.
• Caution! paragraphs provide information that alerts the operator to a hazardous
situation that can cause damage to property or equipment.
• Warning! paragraphs provide information that alerts the operator to a hazardous
situation that can cause injury to personnel. Cautionary information is also included,
when applicable.
Safety Issues
WARNING! It is the responsibility of the user to make sure all local, county, state and
national codes, regulations, rules and laws related to safety and safe
operating conditions are met for each installation.
Auxiliary Equipment
Local Safety Standards
The user must make sure that he operates all auxiliary equipment in accordance with local
codes, standards, regulations, or laws applicable to safety.
Working Area
WARNING! Auxiliary equipment may have both manual and automatic modes of
operation. As equipment can move suddenly and without warning, do
not enter the work cell of this equipment during automatic operation,
and do not enter the work envelope of this equipment during manual
operation. If you do, serious injury can result.
WARNING! Make sure that power to the auxiliary equipment is turned OFF and
locked out before you perform maintenance procedures on the
equipment.
HygroGuard MMY 2650 Installation and Operation Manualix
Preface
Qualification of Personnel
Make sure that all personnel have manufacturer-approved training applicable to the
auxiliary equipment.
Personal Safety Equipment
Make sure that operators and maintenance personnel have all safety equipment applicable
to the auxiliary equipment. Examples include safety glasses, protective headgear, safety
shoes, etc.
Unauthorized Operation
Make sure that unauthorized personnel cannot gain access to the operation of the
equipment.
Environmental Compliance
Waste Electrical and Electronic Equipment (WEEE) Directive
GE Measurement & Control Solutions is an active participant in Europe’ s Waste Electrical
and Electronic Equipment (WEEE) take-back initiative, directive
2002/96/EC.
The equipment that you bought has required the extraction and use of natural resources for
its production. It may contain hazardous substances that could impact health and the
environment.
In order to avoid the dissemination of those substances in our environment and to diminish
the pressure on the natural resources, we encourage you to use the appropriate take-back
systems. Those systems will reuse or recycle most of the materials of your end life
equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please
contact your local or regional waste administration.
Visit http://www.gesensing.com/environment/weee.htm
and more information about this initiative.
xHygroGuard MMY 2650 Installation and Operation Manual
for take-back instructions
Chapter 1. Features
Chapter 1.Features
1.1 Overview
The MMY 2650 is a moisture analyzer designed to operate with trace moisture probes
DY5x7 and DY5x2. The unit offers an optional input that can be used for pressure
compensation or any other sensor. The standard product comes with one current output
(0–20 mA or 4–20 mA), a system alarm relay, and three adjustable relays, all with
single-pole double-throw contacts, selectable fail-safe mode, and adjustable hysteresis.
One or two additional current outputs are available as options. A display and five push
buttons form the user interface for data display, and selection of units, ranges, setpoints,
and special functions.
The MMY 2650 is to be used only as specified by the factory. Use other than as specified
may impair the safety precautions of the system.
1.2 Available Versions
The MMY 2650 is available in two mounting versions: wall mount and panel mount.
This manual generally describes the installation and wiring of the wall-mount version.
Specific instructions for the panel-mount version are given where necessary.
The panel-mount version is available with or without a protective and lockable door
covering the unit’s controls.
HygroGuard MMY 2650 Installation and Operation Manual1
Chapter 1. Features
1.3 Unpacking and Inspection
Examine the shipping carton for broken or open packing, distortion, o r any other eviden ce
of mishandling. If inspection indicates damage to the unit or any of its components, notify
the carrier promptly and request an inspection.
Move the carton to a clean work area and unpack the unit. The carton should contain:
•MMY 2650 analyzer•CD with Manual
•DY5xx probe with its cable•Calibration certificate
Figure 1: Unpacking the Components
2HygroGuard MMY 2650 Installation and Operation Manual
Chapter 1. Features
1.3.1Connection Cable
Configuration information for the cable connecting the MMY 2650 to the DY5 probe is
shown below:
7C2550-X
2C2550-X
7C2550-0Model Number Example
Cable with multi-pin connector and tinned ends; for use
with probe DY5-X7 25XXX
Cable with spade lugs and ferrule ends; for use with
probe DY5-X2 25XXX
Cable Length
05 ft.
Other length; specify r equir ed le ngth (Be lden 8723 o r
L
equivalent)
For probes with a junction box (DY-502), the factory provides five feet of cable at no
charge.
1.3.2Theory of Operation
Moisture and temperature sensor signals are converted within the probe to frequencies,
allowing the use of shielded four-wire cable up to 1000 feet in length without noise
interference. The raw frequencies can be displayed for troubleshooting, calibration and
function control.
The DY5xx probe is calibrated against known dew point temperatures referenced by a
NIST-certified chilled mirror instrument. The sensor senses the water vapor pressure
independent of the carrier gas. Other units of measure are derived using the temperature
measurement, where necessary.
HygroGuard MMY 2650 Installation and Operation Manual3
Chapter 1. Features
(Frequency)
Moisture
(Frequency)
Temperature
Pt1000
f
V
–
f
V
+
Dew Point–80°C
1kHz
Frequency Hz
5kHz
2kHz
3kHz
4kHz
20°C
6kHz
1.3.2 Theory of Operation (cont.)
Figure 2: Theory of Operation
4HygroGuard MMY 2650 Installation and Operation Manual
HygroGuard MMY 2650 Installation and Operation Manual5
Chapter 1. Features
Ø1.0
(25)
3.8 (97)
Ø0.5
(13)
7.8 (198)
0.5 (13)
Dimensions are in inches (millimeters).
4.5 (114)
Dimensions are in inches (millimeters).
10.0 (254)
1/2" MNPT
1.0 (25)
Probe Mounting Adapter
1/2" FNPT
Conduit Connection
1.4 Equipment Drawings (cont.)
•The DY 507 probe uses a removable cable and connector.
•The DY 502 probe cable is wired to an attached junction box for use in hazardous
areas.
•The standard mounting adapter for the DY-series probe is a 1/2” tube compression
fitting by 1/2” MNPT.
Figure 5: DY 507 Probe
6HygroGuard MMY 2650 Installation and Operation Manual
Figure 6: DY 502 Probe
1.4 Equipment Drawings (cont.)
Connectors
Wiring
Figure 7: MMY 2650 Panel Mount Version
Chapter 1. Features
HygroGuard MMY 2650 Installation and Operation Manual7
Figure 8: MMY 2650 Panel Mount Version (rear view)
Chapter 1. Features
8HygroGuard MMY 2650 Installation and Operation Manual
Chapter 2. Installation
Pipe
Probe
1"
Probe
Pipe
1”
Chapter 2.Installation
2.1 Mounting the Probe
2.1.1In-Line Mounting
Mount the probe vertically so that the sensor tip points down. The probe is typically
mounted in a H” pipe thread connection. Insure that the tip of the probe does not touch the
inside wall of the pipe.
Adjust the stainless steel ferrule for a probe insertion length of at least 1” and tighten the
compression fitting as follows:
1. Hand-tighten the nut.
2. Using a wrench, tighten the nut one and one-half additional turns.
With this procedure, the fitting is sealed and can withstand pressure to 3000 psig.
Figure 9: Probe Mounting
HygroGuard MMY 2650 Installation and Operation Manual9
Chapter 2. Installation
Probe
Main
process
stream
Shutoff valves
the upstream valve and
Probe
Shutoff Valves
Main
Process
Stream
2.1.2Bypass Mounting
Using a bypass line assures no interruption of the process stream while the probe is being
serviced. T o mount the probe in a bypass, the above procedure for in-line mounting should
be followed. Exercise caution when reopening the bypass after a probe has been serviced
as this may contaminate the main stream.
CAUTION!
Before servicing the probe, close the upstream valve and the proces s
stream valve to depressurize the bypass; then close the system’s
downstream valve.
Figure 10: Bypass Mounting
10HygroGuard MMY 2650 Installation and Operation Manual
Chapter 2. Installation
2.1.3Mounting in a Sample System
If the gas velocity or temperature in your application exceeds the probe specifications in
Appendix A or if particulates are entrained in the stream, a sample system should be used.
In a sample system with low dew points, all wetted parts should be of stainless steel. (For
further details, refer to the General Hints for Trace Moisture Measurement manual.)
While the measured value of the probe is not flow-sensitive, the flow rate in a sample cell
influences the response time of the system. A flow rate of 1 to 5 SCFH through the flow
cell is ideal. In addition, the sample gas temperature in the flow cell must remain above the
dew point temperature to prevent condensation. (If the dew point is greater than 0°C,
maintaining a +10°C (18°F) difference is recommended.)
WARNING!
Before removing the probe, insure that the flow cell is not under
pressure!
Removing the probe when pressure exists in the flow cell could result
in an extremely dangerous situation, threatening to both life and
limb!
Figure 11 on page 12 illustrates a basic sample system.
HygroGuard MMY 2650 Installation and Operation Manual11
Chapter 2. Installation
Probe
Valve 1
Valve 2
Flow Cell
Flow
Intake towards the middle of the
p
ipe diameter is recommended.
Is
Ia
Ntao
Probe
Flow
Valve 1
Valve 2
Flow Cell
Intake toward the middle of the
pipe diameter is recommended.
See note below.
2.1.3 Mounting in a Sample System (cont.)
Measurement Options
•If measurement is desired under system pressure:
Valve 1 is fully open and the sample flow is regulated by Valve 2.
•If measurement is desired under atmospheric pressure:
Valve 2 is fully open and the sample flow is regulated by Valve 1.
Note:At least five feet of 1/4” tubing, if vented to the atmosphere, will ensure accuracy
of sample in the flow cell.
12HygroGuard MMY 2650 Installation and Operation Manual
Figure 11: Basic Sample System
Chapter 2. Installation
2.1.4Probe Mounting Hardware
The following fittings are available for the DY 5xx probe. A 316 stainless steel ferrule is
provided for use in pressurized systems. For non-pressurized systems, a nylon ferrule is
available.
•½” tube by ½” or ¾” MNPT compression fitting; ferrule
•¾” x 16 “O” ring fitting x ½” tube 316 DD compression fitting
•For metric systems, a compression fitting with gasket and
G ½ thread, DIN-ISO 228, is available.
Figure 12: Compre ssion Fitting
HygroGuard MMY 2650 Installation and Operation Manual13
Chapter 2. Installation
2.2 Opening the Wall-Mount Enclosure
To access the mounting points and electrical connections, open the case as follows (refer
to Figure 13 below):
1. Ensure that no electrical power is present at the analyzer.
2. Loosen the four plastic screws securing the cover to the enclosure and remove the
cover.
Note:The cover contains electronic components and is connected to the body of the
analyzer.
Figure 13: Opening the Wall-Mount Enclosure
14HygroGuard MMY 2650 Installation and Operation Manual
Chapter 2. Installation
2.3 Mounting the Wall-Mount Enclosure
The MMY 2650 analyzer is for indoor use only. It should be mounted with appropriate
hardware, and all wiring should conform to local electrical codes and standards.
A drilling template for locating the 2650’s mounting holes is supplied with the unit.
Choose mounting hardware appropriate for your application.
Locate the analyzer module so that the probe cable length will not exceed 1000 feet.
1. Remove the cover as described in the previous section.
2. Mount the analyzer as shown in Figure 14 below.
Note:Insert the upper screw into the mounting hole and tighten it. Finally, insert the two
lower screws.
3. Replace the cover (Maximum torque = 1.2 N-m (12.2 Kg-cm)).
T o dismount the instrument, remov e the cover. Remove the lower mounting screws. Then,
remove the top mounting screws.
Figure 14: Mounting the Wall-Mount Enclosure
HygroGuard MMY 2650 Installation and Operation Manual15
Chapter 2. Installation
2.4 Mounting the Panel-Mount Enclosure
The panel-mount version of the MMY 2650 is for indoor use only and is designed to be
mounted in a square cutout in an equipment panel. The unit accommodates a wide range
of panel thicknesses. Mount the unit as described below:
1. Make a cutout in the panel measuring 5.4 3” by 5.43”
(138 by 138 mm).
2. Slide the unit into the cutout from the front of the panel.
3. From the rear side of the panel, slide the two mounting clamps into place on each side
of the unit (see Figure 7 on page 7).
4. Tighten each clamp screw from the rear using a long screwdriver.
2.5 Wiring
Figure 15 below shows the MMY 2650 wiring compartment, showing typical connections
for power and the DY 5xx probe.
Figure 15: Wiring Compartment
Figure 16 on pag e 17 shows the complete wiring diagram for the 2650. Wiring for the
probe, power and outputs are shown in the following sections.
Wiring for the Panel Mount version is the same, except that the wiring connections are
accessed from the rear of the unit.
16HygroGuard MMY 2650 Installation and Operation Manual
2.5 Wiring (cont.)
249 Ohms
–
Current Out 2
32
16
CRelay 2
Relay 1
NC
NO
NC
C
Current Out 1
Relay 4
Relay 3
Outputs
NC
NC
NO
NO
C
+
+
NO
C
–
249 Ohms
(–) (Probe Black)
FRQ2 (Probe White)
FRQ1 (Probe Green)
+8.2V (Probe Red)
51 2 34
19212027
6 79810
24
22 232625
13121114 15
292830 31
Inputs
Ground Wire
Earth Ground
Current Out 3
Power
L1 or (+)
+
–
Cable Shield
L2/Neutral or( –)
249 Ohms
+Ch. 3 Input 0–5V
1817
3533 3436
213
+15 VDC Out
(–)
Chapter 2. Installation
Figure 16: Wiring Connections
HygroGuard MMY 2650 Installation and Operation Manual17
Loading...
+ 57 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.