• Optional analog input for pressure compensation
or analog transmitter (voltage or current)
• One, two or three analog outputs; 0/4 to 20 mA or
0/1 to 5 VDC
• Three congurable relay outputs plus system
alarm relay
• Two line LCD and ve push button user interface
• RS485 serial output
• Full complement of pre-programmed moisture
measurement units
The HygroGuard MMY2650 is a moisture analyzer
designed to operate with the GE DY5 or DY55
planar gold aluminum oxide dew point sensor. The
unit oers an optional input that can be used for
pressure compensation, an external temperature
probe, or other analog transmitter. The standard
product comes with one current output (0/4 to
20 mA), a system alarm relay, and three adjustable
relays, all with single-pole double-throw 2.5 A
contacts, selectable failsafe mode, and adjustable
hysteresis. One or two additional analog outputs are
available. A display and ve push buttons from the
user interface for data display, unit selection, output
ranges, alarm setpoints and special functions.
The HygroGuard MMY2650 is available in both wall
mount and panel mount congurations. When used
with the DY5 dew point probe, the system is capable
of providing dew point readings in the range of
-112°F to 68°F (-80°C to 20°C) dew point over a
temperature range of -4°F to 140°F (-20°C to 60°C).
When congured with approved Zener Barriers the
model DY5xx probe is approved as intrinsically safe
for use in hazardous (classied) locations.
g
MMY2650 Specications
Standard Inputs
Two (moisture and temperature) from DY5 probe,
connects up to two DY55 probes.
Optional Input
For pressure transducer providing live pressure
compensation or monitoring one other analog
transmitter. Signals: 0/1 to 5 V, 0/4 to 20 mA loop
powered, or 4 to 20 mA/source.
Moisture Probe
Interconnects with DY5 probe (US) or DY55 (European)
probe
User Interface
Five push-buttons, easy conguration using a matrix
Display
Alphanumeric LCD displays measured value with units
of measure, matrix location programming instructions,
error indication and error codes; user selectable
scanning feature alternating the display every ve
seconds through active channels (three maximum)
European Compliance
Complies with EMC directive 2004/108/EC and
LVD 2006/95/EC
Digital Outputs
Four relays (SPDT dry contacts rated at 250 VAC, 2.5 A,
PCAC = 300 VA, cos phi > 0.7, PDC 100 W, 100 VDC). One
relay is system alarm. Three relays are congurable to any
input; failsafe mode: energized/de-energized selectable.
programmable hysteresis; high/low alarm selectable
Serial Output
RS485, streaming data, update rate once per second
Program
Non-volatile memory
Units Of Measure (Moisture)
Dew point °C, °F, ppmv (requires pressure measurement
or programmed constant), lb/MMSCF, g/m3, g/kg,
vapor pressure in hPa, or mmHg, %RH, process
pressure calculated dew point °C, °F (needs pressure
measurement using the optional input for live calculation
or pressure constant)
Units of Measure (DewPoint)
°F, °C, instrument range -202°F to 86°F (-130°C to 30°C)
Units of Measure (Pressure)
Optional input used with a pressure transducer to
selected unit; bara, barg, psia, psig, hPaa, hPag
Analog Outputs
Three, assigned to input 1, 2 and 3, 0/4 to 20 mA, load
resistance <500 ohms, 0/1 to 5 V, source resistance
249 ohms, user selectable range, user selectable
condition in case of error to 110%, -10% or hold at last
measured value
Data
EEPROM
Operating/Storage Temperature
14°F to 122°F (-10°C to 50°C)
Supply Voltage
90 to 240 VAC, optional 18 to 36 VDC
Power Consumption
5.8 VA for line voltage units, 2.2 W for DC powered units
Enclosures
Wall Mount: Type 12, Type 4X, Type 7, and Panel Mount.
Cable Entry
Metric cable gland M12, 2x M16, M20
Weight
2.2 lb (1 kg)
1/2 FNPT
conduit
connection
4.5 in (114 mm)1 in (25 mm)4.5 in (114 mm)
DY502 probe dimensions
10 in (254 mm)
0.5 in (13 mm)
Operating Pressure
0 to 1750 psig (0 to 120 bar, 12 MPa)
Recommended Flow Rate
1 to 5 SCFH (in a bypass mode)
Gas Flow Velocity
• 165 ft/s (50.0 m/s) at 14.7 psig (1 bar)
• 16.5 ft/s (5.0 m/s) at 145 psig (10 bar)
• 1.65 ft/s (0.5 m/s) at 1450 psig (100 bar)
0.5 in (13 mm)
1 in
(25 mm)
0.5 in
(13 mm)
DY507 probe dimensions
3.5 in (89 mm)3.8 in (97 mm)
7.8 in (198 mm)
1 in (25 mm)
DY5xx (US) Moisture Probe
Sensor
Planar gold/aluminum oxide capacitance
Calibration Range
-112°F to 68°F (-80°C to 20°C)
Recommended Recalibration Cycle
Six to 12 months depending on the application
Accuracy
±3.6°F (±2°C) in the standard calibration range at 77°F
(25°C)
Calibration Data
Stored in EEPROM of analyzer DewPro MMY2650
Probe Tube
1/2 in (12.7 mm) diameter, 316 stainless steel
Mounting Adapter
1/2 in tube x 1/2 in NPTM, 316 stainless steel compression
tting standard, others available
Cable Connection
Rugged multi-pin connector; screw terminal in explosionproof junction box with 1/2 in NPTF conduit connection
Probe Cable
Four-conductor, AWG 22, stranded, shielded to maintain
EMI/RFI/ESD resistance, up to 1000 ft
Weight
1 lb (.45 kg)
1.67 in
(144 mm)
.71 in
(18 mm)
1.18 in
(30 mm)
G 1/2 DIN
ISO 228
DY55 probe dimensions
2.48 in (63 mm)
1.14 in
(29 mm)
.72 in
(18.2 mm)
Standard Operating Temperature
-4°F to 140°F (-20°C to 60°C)
Maximum Relative Humidity
50% at dew point > 32°F (0°C )
Temperature Sensor
Zener device, range -94°F to 158°F (-70°C to 70°C)
Signal Transmission
Frequency, generated by probe electronics
DY55 (European) Moisture Probe
Sensor
Planar gold/aluminum oxide, capacitance principle
Calibration Range
-112°F to 68°F (-80°C to 20°C) dew point temperature
Recommended Recalibration Cycle
Six to 12 months, depending on the application and
required accuracy
Accuracy
±3.6°F (±2°C) at 77°F (25°C)
Maximum Relative Humidity
50% at dew point > 32°F (0°C )
Ambient and Storage Temperature
-76°F to 122°F (-60°C to 50°C)
Standard Operating Temperature
-4°F to 140°F (-20°C to 60°C)
Maximum Operating Pressure
5075 psig (350 bar)
8.23 in
(209 mm)
5.67 in
(144 mm)
Helium Leak Rate
> 10-7 mbar 1/s
Gas Flow Velocity
• 165 ft/s (50.0 m/s) at 14.7 psig (1 bar)
• 16.5 ft/s (5.0 m/s) at 145 psig (10 bar)
• 1.65 ft/s (0.5 m/s) at 1450 psig (100 bar)
Signal Transmission
Standardized pulse frequency modulation (PFM) via
two-wire cable, up to 3280 ft (1000 m) length.
Probe Mounting Adapter
Stainless steel 1.4571 with G 1/2 thread (DIN/ISO 228) and
O-ring seal out of uor caoutchouc (SMK)
Rating
IP 40
Torque When Tightening
50 Nm
Wrench Width
1.2 in (30 mm)
HygroGuard MMY2650 panel mount enclosure dimensions
MMY2650 Panel Mount Enclosure
Material
Black anodized aluminum
Dimensions
5.67 in x 5.67 in (144 mm x 144 mm)
Depth
• Maximum protrusion at the rear of the panel: 8.23 in
(209 mm)
• Maximum protrusion at the front of the panel: 0.32 in
(8.25 mm) with bezel
• Maximum protrusion at the front of the panel with door:
1.26 in (32 mm)
Panel Cutout
5.43 in x 5.43 in (138 mm x 138mm)
Wiring
Same conguration as the wall mount unit, wired in the rear
Protection
Protective cap with 10 mm lter
Weight
0.5 lb (250 g)
Mounting
Insert from front into the panel, install the clamps, tighten
the clamp screws from the rear against the panel using a
long screwdriver
Front Panel
Overlay with membrane buttons integrated LEDs and
clear window for display