General Eastern MMY2650 Data sheet

GE
Measurement & Control Solutions
HygroGuard MMY2650
General Eastern Trace Moisture Analyzer
Features
• Optional analog input for pressure compensation or analog transmitter (voltage or current)
• One, two or three analog outputs; 0/4 to 20 mA or 0/1 to 5 VDC
• Three congurable relay outputs plus system alarm relay
• Two line LCD and ve push button user interface
• RS485 serial output
• Full complement of pre-programmed moisture measurement units
The HygroGuard MMY2650 is a moisture analyzer designed to operate with the GE DY5 or DY55 planar gold aluminum oxide dew point sensor. The unit oers an optional input that can be used for pressure compensation, an external temperature probe, or other analog transmitter. The standard product comes with one current output (0/4 to 20 mA), a system alarm relay, and three adjustable relays, all with single-pole double-throw 2.5 A contacts, selectable failsafe mode, and adjustable hysteresis. One or two additional analog outputs are
available. A display and ve push buttons from the user interface for data display, unit selection, output ranges, alarm setpoints and special functions.
The HygroGuard MMY2650 is available in both wall mount and panel mount congurations. When used with the DY5 dew point probe, the system is capable of providing dew point readings in the range of
-112°F to 68°F (-80°C to 20°C) dew point over a temperature range of -4°F to 140°F (-20°C to 60°C). When congured with approved Zener Barriers the model DY5xx probe is approved as intrinsically safe for use in hazardous (classied) locations.
g
MMY2650 Specications
Standard Inputs
Two (moisture and temperature) from DY5 probe, connects up to two DY55 probes.
Optional Input
For pressure transducer providing live pressure compensation or monitoring one other analog transmitter. Signals: 0/1 to 5 V, 0/4 to 20 mA loop powered, or 4 to 20 mA/source.
Moisture Probe
Interconnects with DY5 probe (US) or DY55 (European) probe
User Interface
Five push-buttons, easy conguration using a matrix
Display
Alphanumeric LCD displays measured value with units of measure, matrix location programming instructions, error indication and error codes; user selectable scanning feature alternating the display every ve seconds through active channels (three maximum)
European Compliance
Complies with EMC directive 2004/108/EC and LVD 2006/95/EC
Digital Outputs
Four relays (SPDT dry contacts rated at 250 VAC, 2.5 A, PCAC = 300 VA, cos phi > 0.7, PDC 100 W, 100 VDC). One relay is system alarm. Three relays are congurable to any input; failsafe mode: energized/de-energized selectable. programmable hysteresis; high/low alarm selectable
Serial Output
RS485, streaming data, update rate once per second
Program
Non-volatile memory
Units Of Measure (Moisture)
Dew point °C, °F, ppmv (requires pressure measurement or programmed constant), lb/MMSCF, g/m3, g/kg, vapor pressure in hPa, or mmHg, %RH, process pressure calculated dew point °C, °F (needs pressure measurement using the optional input for live calculation or pressure constant)
Units of Measure (DewPoint)
°F, °C, instrument range -202°F to 86°F (-130°C to 30°C)
Units of Measure (Pressure)
Optional input used with a pressure transducer to selected unit; bara, barg, psia, psig, hPaa, hPag
Analog Outputs
Three, assigned to input 1, 2 and 3, 0/4 to 20 mA, load resistance <500 ohms, 0/1 to 5 V, source resistance 249 ohms, user selectable range, user selectable condition in case of error to 110%, -10% or hold at last measured value
Data
EEPROM
Operating/Storage Temperature
14°F to 122°F (-10°C to 50°C)
Supply Voltage
90 to 240 VAC, optional 18 to 36 VDC
Power Consumption
5.8 VA for line voltage units, 2.2 W for DC powered units
Enclosures
Wall Mount: Type 12, Type 4X, Type 7, and Panel Mount.
Cable Entry
Metric cable gland M12, 2x M16, M20
Weight
2.2 lb (1 kg)