General Eastern E4-1111H-GE Operating Manual

E4-1111H-GE
DEW POINT MONITOR
Operator’s Manual
The information in this manual is believed to be reliable. However, General Eastern makes no warranty, ex­pressed or implied, as to its accuracy and assumes no liability arising out of its use by others. We reserve the right to change this manual without prior notice.
Document No. A40232856/A00 ECO #5521 October 31, 2001 Printed in U.S.A. Copyright
2001 General Eastern Instruments
The Humidity Experts
General Eastern Instruments
20 Commerce Way, Woburn, MA 01801-1057
Tel: (781) 938-7070 Fax: (781) 938-1071
Safety:
This manual as well as safety labels posted on the instrument use the following safety alerts to draw your attention to special safety instructions that should be followed.
CAUTION
CAUTION: Hazards or unsafe practices could result in electrical shock, minor
injury, or product damage.
WARNING
WARNING: Refer to accompanying document for additional information.
ALERT:
ALERT
Earth ground connection; removal could result in electrical shock,
minor injury, or product malfunction.
CONTENTS
CHAPTER 1 — INTRODUCTION
About General Eastern - - - - - - - - - - - - - - - - - - - - - - 1 Overview of the E4-1111H-GE - - - - - - - - - - - - - - - - - - 1 The humidity sensor - - - - - - - - - - - - - - - - - - - - - - - 3 Theory of operation - - - - - - - - - - - - - - - - - - - - - - - - 3
CHAPTER 2 — INSTALLATION
Mounting - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7 Input power - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8 Wiring the outputs - - - - - - - - - - - - - - - - - - - - - - - - 9 Sensor information - - - - - - - - - - - - - - - - - - - - - - - 12 Sensor replacement - - - - - - - - - - - - - - - - - - - - - - - 13
CHAPTER 3 — OPERATION
General information - - - - - - - - - - - - - - - - - - - - - - - 15 Control switches - - - - - - - - - - - - - - - - - - - - - - - - 15 RS-232C data outputs - - - - - - - - - - - - - - - - - - - - - - 17 Helpful hints for operating the unit - - - - - - - - - - - - - - - 17 The PACER Cycle - - - - - - - - - - - - - - - - - - - - - - - - 21
CHAPTER 4 — PROGRAMMING
General information - - - - - - - - - - - - - - - - - - - - - - - 23 RS-232C programming - - - - - - - - - - - - - - - - - - - - - 23
CHAPTER 5 — SERVICE
Minor maintenance of sensor optics - - - - - - - - - - - - - - - 31 Field replacement of sensor mirror - - - - - - - - - - - - - - - 32 Test and calibration - - - - - - - - - - - - - - - - - - - - - - - 35 Troubleshooting - - - - - - - - - - - - - - - - - - - - - - - - - 35
- - - - - - - - - - - - - - - - - - - - - - - 31
- - - - - - - - - - - - - - - - - - - - 1
- - - - - - - - - - - - - - - - - - - - 7
- - - - - - - - - - - - - - - - - - - - - 15
- - - - - - - - - - - - - - - - - - - 23
APPENDIX A — SPECIFICATIONS APPENDIX B — WARRANTY AND RETURN PROCEDURE
- - - - - - - - - - - - - - - - - - 39
- - - - - - 41
APPENDIX C — HUMIDITY EQUATIONS AND CONVERSION CHART APPENDIX D — CONFIGURING THE RS-232 INTERFACE APPENDIX E — GLOSSARY
- - - - - - - - - - - - - - - - - - - - - 53
- - - - - - 49
45
FIGURES
Figure 1 Dimensions Figure 2 The chilled-mirror hygrometer Figure 3 Front view Figure 4 Side view Figure 5 Rear view Figure 6 PC Board Figure 7 Wiring and sensor replacement Figure 8 DIP Switches Figure 9 The PACER cycle Figure 10 Results of the PACER cycle Figure 11 Scaling parameters Figure 12 Balance adjustment screw locations Figure 13 Using a Resistance Decade Box
- - - - - - - - - - - - - - - - - - - - - - - 3
- - - - - - - - - - - - - - 4
- - - - - - - - - - - - - - - - - - - - - - - 7
- - - - - - - - - - - - - - - - - - - - - - - - 8
- - - - - - - - - - - - - - - - - - - - - - - - 9
- - - - - - - - - - - - - - - - - - - - - - - 10
- - - - - - - - - - - - - 11
- - - - - - - - - - - - - - - - - - - - - 16
- - - - - - - - - - - - - - - - - - - 21
- - - - - - - - - - - - - - - 22
- - - - - - - - - - - - - - - - - - - 25
- - - - - - - - - - - - - 37
- - - - - - - - - - - 32
Chapter 1 — Introduction
About General Eastern
General Eastern is devoted solely to the design and manufacturing of accurate, reliable and rugged humidity measuring equipment. We specialize in providing solutions for applications where humidity measurements are critical.
There are many ways to make humidity measurements, and no one humidity sensor meets all requirements for all applications. Our variety of sensor types — including chilled mirror, lithium chloride, resistance polymer, and wet bulb — can precisely determine dew point, parts per million by volume, percent relative humidity, and other parameters.
In keeping with General Eastern’s philosophy of providing the best solutions to humidity measurement problems, we offer the following products and services:
• high quality state-of-the-art instrumentation to assure excel­lent performance
• a broad range of humidity instruments capable of covering vir­tually any humidity measurement application
• full applications assistance to help you choose the sensor that is best for your needs
• full and superior service, should it ever be needed
If you have questions about a particular measurement problem, we invite you to call and discuss your application with one of our engi­neers. Call 800-225-3208 (if you’re calling from Massachusetts or outside the United States, call 781-938-7070). Our fax number is 781-938-1071.
Overview of the E4-1111H-GE
The General Eastern E4-1111H is a multi-purpose chilled-mirror hygrometer, suitable for use in a variety of applications. The unit measures dew/frost point from –25°C to +65°C (–13°F to +149°F).
Chapter 1 — Introduction
1
The E4-1111H is designed for wall or surface mounting. It is housed in a gasketed NEMA-4 aluminum enclosure, suitable for industrial environments.
A 4–20mA analog output is provided as well as an RS-232C communications port.
A red indicator light shows when indicator light shows that the unit’s tion (see “The PACER Cycle” beginning on page 21).
The monitor utilizes advanced microprocessor control and includes General Eastern’s patented Programmable Automatic Contaminant
Error Reduction (PACER) system to insure accuracy.
Detailed specifications are given in Appendix A, beginning on page 39.
POWER
BALANCE
is applied and a green
function is in opera-
System
Components
Physical
description
The complete system is packaged with the following items:
• The electronic monitor with integrated dew point sensor and filter
• Operator’s manual
• Certification that the unit is traceable to the National Institute of Standards and Technology
The E4-1111H is designed for surface mounting on a vertical wall or panel. The unit’s dimensions are shown in Figure 1 in inches (cm).
2
E4-1111H-GE Operator’s Manual
)
n
Ø 0.50 (12.7
4 plcs
3
Removable filter
17.36 (44.1)
2.75 (70)
8 (20.3) 10 (254)
10.5 (267)
The humidity sensor
The monitor is configured with a General Eastern 1111H single-stage chilled-mirror dew point sensor. The sensor provides data to the electronics unit that calculates humidity.
18.4 (468)
dimensions in inches (mm)
4.25 (108)
Sensor Probe
1/2" Conduit fittings
Figure 1 — Dimensions
9.2 (237)
Figure 1 — Dimensions
1.5 (39)
dimensions i inches (mm)
Theory of operation
What is Optical
Condensation
Hygrometry?
Chapter 1 — Introduction
Optical condensation hygrometry is a precise technique for deter­mining the water vapor content in gases by directly measuring dew point or frost temperatures. Using this technique, a metallic mirror is cooled until it reaches a temperature at which a thin layer of condensation begins to form on it. The dew layer is detected opti­cally, and the mirror is held at that temperature. The mirror temperature, measured with a platinum resistance thermometer, is
r
an accurate indicator of the dew or frost point. Because these hygrometers are so accurate, they are widely used as a standard in many of the world’s metrology laboratories.
How do General
Eastern
hygrometers
function?
Figure 2 illustrates how General Eastern hygrometers detect and measure dew point. The condensate mirror is illuminated with a high-intensity, solid state, light emitting diode (LED). A photode­tector monitors the LED light reflected from the mirror. The photo­detector is fully illuminated when the mirror is clear of dew, and it receives less light as dew forms. A separate LED and photodetector pair are used as a known reference to compensate for any thermally induced changes in the optical components. The photodetectors are arranged in an electrical bridge circuit, the output current of which is proportional to the light reflected from the mirror. The bridge output controls the electrical current to the thermoelectric cooler.
Figure 2 — The chilled-mirror hygromete
A large bridge current develops when the mirror is dry, causing the mirror to cool toward the dew point. As dew begins to form on the mirror, less light is reflected, and the bridge output decreases. This, in turn, causes a decrease in cooling current. A rate feedback loop
4
E4-1111H-GE Operator’s Manual
5
within the amplifier ensures critical response, causing the mirror to stabilize quickly at a temperature that maintains a thin dew or frost layer on the mirror surface. A precision thermometer element embedded within the mirror directly monitors this dew point temperature.
Using General
Eastern
Hygrometers as a
standard for
calibration
An optical condensation hygrometer such as the E4-1111H can be sent to the National Institute of Standards and Technology (NIST) in Gaithersburg, Maryland for calibration against the NIST stan­dard. A calibrated instrument can then be used as a transfer stan­dard in local laboratories to calibrate lower echelon instruments. Hygrometers used as calibration standards must have the following characteristics:
• The mirror thermometer must have suitable long-term accu­racy (such as that obtained with a platinum resistance ther­mometer).
• A means must be provided for viewing the dew or frost forma­tion on the mirror.
Chapter 1 — Introduction
6
E4-1111H-GE Operator’s Manual
Mounting
)
Chapter 2 — Installation
Mount the E4-1111H on a vertical surface using the four mounting holes at the corners of the unit. Provide adequate space below the unit for cabling, and to the left of the unit to allow the door to open fully. The sensor module protrudes from the rear of the unit into the measurement space. Mounting dimensions are shown in Figures 3, 4 and 5.
Ø 0.50 (12.7
4 plcs
17.36 (44.1)
18.4 (468)
dimensions in inches (mm)
Figure 3 — Front view
8 (20.3)
10 (254)
10.5 (267)
2.75 (70)
Chapter 2 — Installation
7
n
Removable filter
Sensor Probe
1.5 (39)
Input power
1/2" Conduit fittings
dimensions i inches (mm)
4.25 (108)
9.2 (237)
Figure 4 — Side view
The E4-1111H is available in versions wired for 115, 230, and 100 VAC. The 230 VAC version is double-fused.
The unit’s voltage and frequency rating and tolerances, as well as fusing data, are listed inside the unit.
Input power wiring enters through the bottom edge of the unit and connects to the right-hand terminal block as shown in Figure 7.
8
E4-1111H-GE Operator’s Manual
n
Ø
0.50 (12.7)
4 plcs
5.63
(143)
I+
Wiring the outputs
The E4-1111H’s output terminals are located inside the unit’s front door. Cabling is brought in through the bottom edge of the unit and connected to terminal blocks as shown in Figure 7.
Analog humidity
output
The humidity output provides a 4–20mA analog signal (500 Ohms maximum load). Connect to the terminals labelled the left-hand terminal block.
4 (10.15)
8 (20.3)
4 (10.15)
11.73
(298)
17.36 (441)
dimensions i inches (mm)
Figure 5 — Rear view
and
RTN
on
Chapter 2 — Installation
9
d
The humidity units and scaling may be set using the RS-232C port. See “Chapter 4 — Programming” beginning on page 23 for details.
.
DIP switches
RS-232 Connector
Switch 1
Right Side (closed)Left Side (open)
Switch 8
PC Boar
Figure 6 — PC Board
Service output
10
The
SRVC
output provides a TTL-compatible signal that indicates
that the mirror needs cleaning (see page 31).
Connect to the terminals labelled
SRVC
and
Figure 7.
E4-1111H-GE Operator’s Manual
RTN
as shown in
RS-232C
connector
The RS-232C connector is located at the bottom edge of the PC board (see Figure 6). It provides RS-232C digital communications between the unit and an RS-232C terminal or a PC running in terminal emulation mode.
Sensor mounting plate
Sensor cable connection
Wing nuts (4 plcs)
SRVC
RTN
L (line)
I+
Ground
N (neutral)
AC power
Chapter 2 — Installation
Outputs
Figure 7 — Wiring and sensor replacement
The connector is a standard 25-pin D connector. For connection to another RS-232C device, the cable is wired as shown below. Pin connections are given for both 25-pin and 9-pin devices.
11
E4-1111H
25-pin device 9-pin device
232
323
775
The baud rate of the data is 1200. The format of the data is 8 data bits, 1 stop bit, and no parity.
The unit can communicate with a handheld RS-232C terminal or a personal computer running in terminal emulation mode. See Appendix D for a BASIC computer program that can be used for simple RS-232C communications between the unit and your PC.
Handshaking.
to
NO HANDSHAKE
In order to enable handshaking, the RS-232C menu must be accessed through the RS-232C port. See “Handshake control” on page 28 for details. To implement handshake control using Clear-To-Send and Ready-To-Send, use pins 4 and 5. Applying 8 volts ( transmit.
3) to Pin 5 of the RS-232C connector allows the unit to
±
The default handshaking protocol for the unit is set
.
Sensor information
Filter
Requirements
Flow Rate
12
The E4-1111H is equipped with a submicron Teflon filter to avoid particulate contamination and to protect the sensor from the pres­ence of liquid water. In many applications, the filter requires peri­odic inspection and replacement. Typical applications might require replacement twice a year.
It is important to have adequate flow through the sensor. Too little flow can slow the response (particularly at very low frost points).
E4-1111H-GE Operator’s Manual
Too much flow can cause instability of the control system at high dew points and can reduce the depression capability of the cooling pump at very low dew points. Too much flow also accelerates the rate of system contamination. A flow rate of 2 to 2.5 ft over 1 liter/min) is ideal for most applications. In many cases, flow rates between 0.2 and 5 ft
Sensor replacement
If it necessary to remove the sensor module, follow the steps below:
1. Disconnect the line power from the unit.
2. Open the front door.
3. Unplug the sensor cable from the end of the sensor probe as shown in Figure 7.
4. Unscrew the four wing nuts shown in Figure 7.
5. Carefully remove the sensor module.
3
/h (a little
3
/h (0.1 and 2.5 liter/min) may be used.
13
Chapter 2 — Installation
14
E4-1111H-GE Operator’s Manual
Chapter 3 — Operation
General information
Operating instructions fall into three categories:
• normal operation
• setup and programming
• system maintenance
Normal operation of the unit’s controls is described in the next section.
Setup and programming involves customizing the unit for special­ized applications, and is not required for many conventional appli­cations. The unit is shipped properly programmed to meet the customer’s requirements (see Appendix A). Complete programming instructions are given in Chapter 4, beginning on page 23.
Maintenance involves manually testing the unit’s cooling capacity, clearing the mirror, and other operations that might be required on a regular basis, or when a problem is suspected, depending on the application.
An RS-232C-compatible device, such as a laptop computer, can be connected to the unit to perform programming and configuration functions. The RS-232C port can also be used to send data measured by the unit to an external computer or other device.
Control switches
Control switches are located at the top edge of the PC board, within the red switch module containing eight DIP switches. Switches are number from 1 to 8, with Switch 1 at the top and Switch 8 at the bottom as shown in Figure 8.
Chapter 3 — Operation 15
)
s
Switch 1
Switch 2
Switch 3
Left Side
(contacts open)
Switch 4
Switch 5
Switch 6
Switch 7
Switch 8
Right Side
(contacts closed
Figure 8 — DIP Switche
All switches are normally set to the open position, with the left-hand side depressed. Press the right hand side to activate a function as described below. (Switches 5–8 are not used).
Switch Function Left side pressed Right side pressed
1 Mirror cooling Normal Cool the mirror.
2 Mirror heating Normal Heat the mirror.
3 Initiate PACER
cycle
4 Automatic
Normal Initiate one PACER
cycle.
Normal Periodic PACER cycles.
PACER cycles
The PACER cycle for automatically cleaning the sensor mirror is described in “The PACER Cycle” beginning on page 21.
To initiate a single PACER cycle, press Switch 3 to the right and then return it to the normal position.
To cause automatic PACER cycles at a programmable periodic rate, set Switch 4 to the right. See “Setting the balance interval.” beginning on page 26 for programming the rate.
16 E4-1111H-GE Operator’s Manual
RS-232C data outputs
See “RS-232C connector” on page 11 for details on hardware, cabling and protocols for establishing RS-232 communications with the monitor. RS-232 communications can be divided into two sections: Operate Mode and Programming Mode.
Operate Mode
While the unit is in Operate Mode, the RS-232C port can provide any data that is available. The data available for this model is as follows:
DP C=xxxx DP F=xxxx
The data output by the unit can be modified in two ways:
• the user can customize the data output to a subset of the
above (See “RS-232C programming” on page 23.)
• the user can enter commands via the RS-232C port to obtain
immediate output of a single parameter (See “Outputting data ‘on command’” on page 28.)
Programming is described in “Chapter 4 — Programming” beginning on page 23.
Helpful hints for operating the unit
Time response
At dew points above 0°C, the system stabilizes within a few seconds at a consistent dew or frost layer. Once the system is stable valid readings may be taken.
When the system is operating at very low frost points (below –40°C), extra care may be required when interpreting readings because of the slower response of the system. Time response depends on a number of factors including dew/frost point, slew rate, upstream filtering, and flow rate.
• As the dew/frost point becomes lower, water molecules in the air sample become scarcer, and it takes longer to condense a frost layer on the mirror sufficiently thick to establish an equi­librium condition.
Chapter 3 — Operation 17
• Temperature slew rate is dependent on dew point and depres­sion (the temperature difference between the mirror and the sensor body); at higher dew points and moderate depressions, it is typically 1.5°C/second. At lower dew points and/or larger depressions, slew rate becomes progressively slower.
• Flow rate affects response by determining the rate at which water vapor is supplied or carried off.
There is, of course, a trade-off between response time, control system stability, and sensitivity to contamination.
Supercooled dew
points
Contamination
Slightly below the freezing point, water can exist in a supercooled liquid state for extended periods of time. Extra care may be needed when making measurements in the frost point region of 0 to –20°C, because the mirror temperature may temporarily stabilize at the supercooled dew point, 0.5 to 1°C below the actual frost point.
To assure that the unit is operating in the ice phase within this temperature range, allow the instrument to operate continuously. Before manually clearing a frost layer, take a reading, and after­wards allow sufficient time to reform a stable frost layer before taking further readings.
Mirror Cleanliness. Proper operation of a condensation hygrometer
depends on the condition of the mirror surface. In general, accuracy is reduced when contaminants accumulate on the mirror.
However, the mirror does not have to be microscopically clean. In fact, the mirror performs best a few hours after cleaning, when nucleation sites have formed. On an unscratched, freshly cleaned mirror, there are relatively few nucleation sites on which dew or frost deposits can form, and more time is required to collect a condensation layer at low frost points. Also, overshoot may occur, which can cause oscillations as the temperature stabilizes.
Particulate contaminants.
water may accumulate on the mirror surface, but does not affect the instrument accuracy until the mirror reflectance is reduced substantially. In many cases, particulates improve instrument response by providing condensation sites.
Particulate matter that is insoluble in
18 E4-1111H-GE Operator’s Manual
Water-soluble contaminants.
Contaminants which readily dissolve in water, such as naturally occurring salts, are detrimental to accurate vapor concentration measurement by any condensa­tion method. These materials readily go into solution with the water condensate on the mirror surface, and then reduce the vapor pressure in accordance with Raoult’s Law. As the concentration increases with time, the saturation vapor pressure of the liquid solution decreases.
The unit responds to this lower vapor pressure by elevating the mirror temperature in order to maintain a vapor pressure that is in equilibrium with the partial pressure of atmospheric water vapor. The measured dew point, therefore, drifts upward above the true dew point. Because the measurement error increases gradually, it often goes undetected.
To determine whether dissolved contaminants are affecting dew point measurement, perform the following steps:
1. Note the indicated dew point
2. Clean the mirror
3. Balance the detector by initiating a PACER cycle
4. Measure the dew point again
If the new reading is lower than the first reading, it is likely that soluble material was present in sufficient quantity to cause a measurement error.
Gaseous contaminants.
When a gaseous material that has a higher condensation temperature than that of water is present (even in very low concentrations), the unit will eventually control on that material, rather than on water. The system then displays the condensation temperature of the contaminant, not of water. Such material accumulates on the mirror only when chilled. In the normal atmosphere, gaseous contaminants do not have a detect­able effect.
Minimizing the effects of contaminants.
1. Use the PACER feature to reduce the effect of contaminants on the unit’s performance.
2. Reduce the gas flow rate to reduce the rate of accumulation of contaminants on the mirror.
Chapter 3 — Operation 19
3. Clean the mirror according to the recommended optics clean-
ing procedure. See “Cleaning the sensor mirror” on page 31. To determine the proper cleaning interval for a given set of condi­tions, take a dew point reading before and after the cleaning. Any appreciable shift indicates that under these conditions, the mirror should be cleaned more often.
Mirror Flooding
Sample line
maintenance
Pressure effects
If there is an abrupt transition from dry to moist conditions (partic­ularly when accompanied by a transition from cold to warm temperatures), the mirror may accumulate an overload of moisture. It then may take several minutes before the sensor dries out and valid readings can be obtained. The drying process can be acceler­ated by setting DIP Switch 2 to sensor.
Contaminated sample lines slow the unit’s response time and can cause erroneous readings, usually on the high side. Clean the sample lines as often as necessary. To determine the required cleaning frequency, take dew point readings before and after cleaning the lines, sensor cavity, and mirror. If the two readings differ appreciably, the sampling lines should be cleaned more often. To reduce the rate of contamination, reduce flow and/or install a filter upstream.
If the pressure of the gas is increased or reduced from atmospheric pressure, but the mixing ratio (moisture content) stays constant, the dew point is correspondingly increased or decreased. The monitor displays the dew/frost point at the pressure to which the sensor chamber is exposed. The sensor location and hookup arrangement can influence the pressure. The dew point change due to pressure change can be calculated by using Dalton’s Law and the Smithsonian Tables or a proper nomograph. Appendix C contains basic data for these calculations. Request a copy of General Eastern’s Humidity Handbook for additional detailed information.
HEAT, to temporarily heat the
20 E4-1111H-GE Operator’s Manual
The PACER Cycle
General Eastern has developed and patented a compensation tech­nique called PACER (Programmable Automatic Contaminant Error Reduction) that is very effective in reducing the Raoult Effect error associated with soluble contaminants, particularly for near-ambient dew points. The PACER cycle replaces the AUTO balance cycle available on earlier General Eastern products.
Figure 9 — The PACER cycle
The PACER cycle, diagrammed in Figure 9, begins with a coales­cence period, during which the mirror is cooled well below the dew point of the sample gas, condensing out a large amount of water. This excess water easily dissolves any water-soluble contaminants. The mirror is then heated. During the heating phase, the large puddles of water gradually evaporate, carrying increasingly heavy concentrations of salt as the puddles become smaller. Finally, when all the puddles have evaporated, dry “islands” of crystallized salt are left on the mirror. The area between the islands (80-85% of the mirror surface) is now clean and shiny, whereas before the PACER cycle it may have been completely covered. The total amount of contamination has not been reduced, but instead redistributed as
Chapter 3 — Operation 21
shown in Figure 10, with more clean mirror surface available for dew formation.
Before
Pacer Cycle
After
Pacer Cycle
Figure 10 — Results of the PACER cycle
22 E4-1111H-GE Operator’s Manual
Chapter 4 — Programming
General information
The E4-1111H-GE is programmed at the factory to display and output the data required for most applications. In many cases, no further programming is required.
Programming allows control of two major items:
• the analog outputs
• the digital (RS-232C) outputs
Programming is most easily accomplished using a computer or data terminal connected to the RS-232C data port.
Note that commands issued to the unit through the RS-232C port would usually come from a manually operated computer or terminal as described below. For specialized automated operation, commands could also be issued autonomously by a properly programmed com­puter.
RS-232C programming
Programming of the unit using the RS-232C channel is accom­plished by a series of menus described in this section. The menus can be accessed by a computer or data terminal connected to the unit’s RS-232C data port. The Data Terminal Emulator included in Windows software provides a very convenient means of controlling the unit.
The programming menus are not available while BALANCE mode is active. (Switch 3 in closed position)
When programming menus are active, the sensor cooling is deacti­vated and the sensor mirror is allowed to rise to ambient tempera­ture.
Chapter 4 — Programming 23
Main menu
(software
version 2.0)
The main menu is the starting point for all programming functions. To access the main menu from normal operating mode, press ESC (hex 1B) twice on your terminal. To return to the main menu from any sub-menu, press E on your terminal one or more times until the main menu is displayed. The main menu will contain some or all of the following choices, depending on the model:
1) Scale Outputs 2) Set Balance Type 3) Set RS-232 Units
4) Display Option 5) Set Averaging 6) Calibrate Output1 E) Exit
Press the digit or letter corresponding to the item you wish to program.
When finished, press E from the main menu to return to OPERATE mode. Note that several seconds may be required for the system to respond and provide data on the RS-232 terminal or the main display.
Each menu item is described in detail in the sections following.
Function 1:
Scale outputs
From the main menu, press 1 to assign and scale the analog outputs. The current output selection(s) and scaling are displayed, followed by the Scale Outputs menu:
OUTPUT1: DP C: HIGH = 75 LOW = -45
1) OUTPUT1 E) EXIT
Scaling Output1. Press 1 to access the Output1 menu, which
controls the Humidity output. The text OUTPUT1: is displayed, followed by the menu listed below:
OUTPUT1:
0) DPC 1) DPF E) EXIT
24 E4-1111H-GE Operator’s Manual
Select a humidity parameter to be output, or press E to exit and return to the previous menu. If you select a parameter, the unit will also ask for the analog output scaling values (full-scale high and low values for the analog outputs).
HIGH =
Enter the value that should cause full-scale output (20mA), and press RETURN.
LOW =
Enter the value that should cause 4mA output, and press RETURN. The unit will again display the newly entered values and the Scale
Outputs
menu. Check the values, and press E to exit back to the main menu. When the unit returns to operating mode, the analog output voltage will vary with the measured humidity as shown in Figure 11.
Analog
output
20mA
4mA
Humidity
Low High
Figure 11 — Scaling parameters
Function 2:
Set balance type
Chapter 4 — Programming 25
From the main menu, press 2 to change the type of balance used, to adjust the balance interval, or to remotely initiate a balance
cycle. The current balance state is displayed, followed by the
Balance Type menu.
BALANCE TYPE = Pacer DAY = 0 HOUR = 12
1) Pacer Bal 2) Auto Bal 3) Set Interval
4) Initiate Balance E) Exit
Selecting the balance type. Press 1 to set the type of balance to
PACER (the default) or 2 to set the type of balance to AUTO.
Note: In the E4-1111H, only the PACER cycle type is available, even though the older AUTO cycle can be selected. Do not use the AUTO setting. See “The PACER Cycle” on page 21.
Setting the balance interval. To set the balance interval, press 3.
The balance interval is specified in days and hours, and the interval will be the sum of the days and the hours entered. The following will appear:
DAY =
Type a value for the number of complete days between balance cycles and press RETURN. The following will appear:
HOUR =
Type a value from 1 to 24 for the number of hours between balance cycles, and press RETURN.
The new interval will be displayed, and the unit will return to the
Balance Type menu.
26 E4-1111H-GE Operator’s Manual
Examples:
Day = 0 Balance every 12 hours Hour = 12
Function 3:
Set RS-232 Units
Day = 1 Hour = 0
Day = 5 Hour = 12
Balance every 24 hours
Balance every 5½ days
Initiating a single balance cycle. Press 4 to manually initiate a
balance cycle. The unit will return to normal operating mode at the completion of the cycle.
Press E to return to the main menu.
From the main menu, press 3 to customize the RS-232C data output. This menu selects the units to be output and the hand­shaking status. The currently selected units and handshaking status are displayed, followed by the Set RS-232 Units menu.
Selected Units = 1,2 Handshake = Disabled
1) Data 2) Handshake (Toggle) 3) Exit
Selecting output data. Press 1 to configure the RS-232C data
output units. The following menu is displayed:
1) Select Units 2) User Prompt E) Exit
Press 1 to select the parameters to be displayed. The following menu is displayed:
1) DPC 2) DPF 3) None E) Exit
Press numbers individually to add the designated parameter to the RS-232C data output. Press the digit for “None” to turn off all data outputs, or to clear the list before adding parameters.
Chapter 4 — Programming 27
Outputting data ‘on command’. A feature of the E4-1111H allows
the unit to output data only when commanded or “prompted” by the user or by an external device. When a command code is received by the unit while operating in this mode, the data associ­ated with that code will be output.
To configure the unit to output data on command, press 2 from the above menu to select User Prompt. The current user-prompt status is displayed, and a sub-menu allows the status to be changed.
Status = Disabled ALLOW USER TO PROMPT FOR DATA
1) Enable 2) Disable E) Exit
Press 1 to enable the user prompt for data. Press E twice to return to OPERATE Mode. To return to normal operation, access the above menu and press 2 to disable this function.
While the unit is set to output data ‘on command,’ data will be output only when requested by the user prompt.
To output dew point in °C, enter the prompt “D1”.
Handshake control. From the Set RS-232C units menu, press 2
to change the handshake protocol. The default is “no handshake.” The unit will ask for a password before allowing the handshaking protocol to be changed.
Enter Access Code:
Type the code word “ACCESS”. The status will be changed and the new status will be displayed. Press E to return to the main menu.
Function 5:
Set Averaging
28 E4-1111H-GE Operator’s Manual
From the main menu, press 5 to set the data averaging parameter. This number determines how many dew point readings are aver­aged to determine the reading that is output. The number may
range between 1 and 20. The factory default is 20. The current value is displayed, followed by a menu:
Number of Points = 20
1) Set Average E) Exit
Press 1 to change the number of points to be averaged. The current averaging value is displayed:
VALUE =
Type a value between 1 and 20 and press RETURN. The new value is displayed. Press E to Exit.
Function 6:
Calibrate
Output1
To calibrate the unit’s Output1 (humidity), the following equipment is required:
•4½-digit voltmeter
• 0.01%, 100-Ohm precision resistor
CAUTION
Please consult the factory before attempting to calibrate the outputs of the unit. All analog outputs are factory calibrated and normally do not require calibration or field adjustment.
From the main menu, press 6 to calibrate the HUMIDITY output. The analog outputs are calibrated at the factory and normally never require recalibration. To protect against inadvertent calibra­tion, this function is protected by a password. The following is displayed, allowing the password to be entered:
Enter Access Code:
Chapter 4 — Programming 29
Type the code “ACCESS”. The Calibrate Output1 menu is displayed:
1) Calib Output 2) Check Output E) Exit
Calibrating the current output. Disconnect all devices from the
unit’s humidity output. Connect the 100-Ohm precision resistor across I+ and RTN terminals. Connect the voltmeter across the resistor leads. Set the voltmeter to 4 volts full scale.
From the above menu, press 1 to calibrate the humidity current output. The unit passes a current through the resistor, generating a voltage measured by the voltmeter. Multiply this voltage by 10 to convert it to milliamps of current in the resistor, and enter the result:
VALUE =
For example, if the voltmeter reads 1.6421 volts, type 16.421 and press RETURN.
The unit passes a new current through the resistor, and asks for a new reading to be entered. Type the new voltage (times 10) and press RETURN. The unit recalibrates its HUMIDITY output and returns to the Calibrate Output1 menu.
30 E4-1111H-GE Operator’s Manual
Chapter 5 — Service
Minor maintenance of sensor optics
Periodically inspect and maintain the sensor optics as described in the following paragraphs.
Cleaning the
sensor mirror
Balancing the
sensor optics
Under normal conditions, the system is self-checking and self-balancing. However, there are occasions when particulate matter and water-soluble contaminants reduce sensor mirror reflectance and system accuracy. See “Contamination” on page 18. When necessary, clean the sensor mirror, following the procedure below.
1. Deactivate the sensor cooler, by placing DIP Switch 2 to HEAT (press the right-hand side).
2. Open the sensor by removing the sensor filter.
3. Moisten a cotton swab with a cleaning solution suitable for mirrors, such as the blue cleaning solution in the General East­ern maintenance kit, or dilute methanol or alcohol. Clean the mirror with a few light wipes. If the sensor has been exposed to significant contamination, clean the other optical surfaces in the sensor and the sensor cavity itself.
4. Replace the sensor filter.
5. Return switches to normal.
6. Initiate a PACER cycle by closing Switch 3 and then returning it to the open position
If the service “SRVC” TTL output reappears after a PACER cycle (even after performing the mirror cleaning procedure above) check the sensor balance adjustment. Improper adjustment of the optical balance is the most common cause of instrument malfunction. In addition, new systems may require an optical balance adjustment after one or two months of operation. Always clean the mirror as described above before performing an optical balance.
Chapter 5 — Service 31
1. Clean the mirror as described in “Cleaning the sensor mirror”
on page 31, if you have not already done so.
2. Set DIP Switch 2 to HEAT (right-hand position) and wait one minute for any condensation to evaporate.
Be sure to cover the optical cavity with your hand to prevent ambient light from affecting the sensor balance.
3. Observe the green BALANCE LED at the bottom of the unit. If the LED is illuminated, the sensor is properly balanced. If not, balance the sensor using the following steps.
4. Turn the balance screw on the sensor COUNTERCLOCKWISE until the BALANCE light goes out. The location of the balance adjustment screw is shown in Figure 12.
5. Slowly turn the balance screw CLOCKWISE until the BALANCE light just comes on.
6. Replace the sensor filter and verify that the BALANCE LED remains illuminated.
7. Return DIP Switch 2 to the normal (left) position to stop heat­ing the mirror.
8. Initiate a PACER cycle by switching DIP Switch 3 to the right, and then returning it to the left.
Sensor balance adjustment screw
(shown with filter removed)
Figure 12 — Balance adjustment screw locations
At the completion of the PACER cycle, the system is properly balanced. For more information on the PACER cycle, see “The PACER Cycle” on page 21.
Field replacement of sensor mirror
One advantage of the General Eastern chilled mirror dew point sensor is that the mirror is user-replaceable. The sensor does not have to be returned to the factory for replacement of the reflective
32 E4-1111H-GE Operator’s Manual
surface, but of course you can return it for factory service, if desired.
A mirror may require replacement for any of the following reasons:
• The mirror is constructed of silver/rhodium plated copper. Copper provides excellent thermal conductivity to the plati­num thermometer. However, some gas constituents, such as sulfur dioxide (SO2) may react with the copper and eventually pit the surface or form a copper sulfate coating.
• The reflective surface may be gradually abraded by sharp dirt particles in the gas being measured.
• The mirror surface may be accidentally scratched or gouged during use or cleaning.
If the sensor mirror has reacted with a corrosive material in the gas sample, such as an acid or sulfur compound, it should be replaced with a solid platinum mirror to remove any possibility of copper corrosion.
In extreme cases, a solid platinum mirror can make possible a successful application of chilled mirror technology. For example, measurements in tobacco factories and malting houses have drasti­cally improved after this change, since both locations have sample gas constituents that attack copper.
In addition, fine scratches can be buffed out of a solid platinum surface, restoring the mirror to new condition, since it is of solid construction.
Replacing the sensor mirror
Chapter 5 — Service 33
Required equipment: torque driver, set to 20-30 inch-ounces of torque. General Eastern’s type TW-1 is recommended.
The kit supplied by the factory contains the replacement mirror, a container of white thermal compound for proper heat transfer, and (in some models) a mylar washer that is to be placed under the mirror.
1. Open the sensor by removing the sensor filter.
2. Unscrew and discard the old mirror, using a 3/16-inch (0.187) hex socket.
3. Use a toothpick or similar tool to place a small amount of ther­mal compound in the hole supporting the mirror.
CAUTIONS:
Do not apply thermal compound to the mirror stem. Do not use an amount large enough to leak out when
the mirror is tightened. Do not allow any compound to get on the mirror sur-
face, as it is very difficult to remove completely.
4. Carefully screw in the new mirror and tighten to the proper torque as specified for the particular sensor.
5. Carefully clean the mirror surface, using a cotton swab and the General Eastern cleaning solution supplied with the mainte­nance kit. Distilled alcohol or diluted alcohol is also accept­able.
6. Replace the filter and return the sensor to normal operation. Under some circumstances, a new mirror may operate in a somewhat unstable manner for the first hour or two.
Modification for platinum mirror. If you choose to upgrade from
the standard plated copper mirror to the solid platinum mirror in the field, a circuit board change must also be made within the E4-1111H. Capacitor C4 must be increased from 33 microfarads to approximately 68 microfarads to avoid instability in the control loop. This change can be accomplished either by replacing C4 with a larger capacitor, or by adding a 33 microfarad capacitor in parallel with C4 using the designated pads on the circuit board.
If you received a system that had a platinum mirror installed at the factory, an additional 33 microfarad capacitor has already been added in parallel with C4.
If you purchase a platinum mirror to replace a standard mirror, you can return the unit to the factory for a no-charge circuit board modification. If you prefer to install the modification, General Eastern will supply a field modification kit at no charge. A trained electronic technician with soldering skills is required for proper circuit board modification.
34 E4-1111H-GE Operator’s Manual
Test and calibration
The procedures in this section effectively test and/or calibrate the following aspects of the monitor:
• Startup and power supply voltage
• Normal sensor operation
• Digital and analog outputs
• No heating or cooling
The unit has been completely tested and calibrated at the factory, and is ready to plug in and operate. As shipped, it meets all of our published specifications and has been checked out at a number of points against a dew point system that has been certified by the U.S. National Institute of Standards and Technology (NIST). A certif­icate of Compliance is supplied with the unit to indicate trace­ability.
There are no calibration adjustments in the instrument that affect the measured readings. The reading is determined directly from the mirror temperature measured by the platinum Resistance Tempera­ture Detector (RTD), which is controlled automatically at the dew point.
The analog outputs provided for recording or controlling are produced by a digital-to-analog converter. This converter can be adjusted as specified in the analog output calibration procedures. See “Function 6: Calibrate Output1” on page 29.
Troubleshooting
Red POWER
indicator doesn’t
light when
powered up
Incorrect dew
point measured
Chapter 5 — Service 35
1. Check the AC line supply. Make sure both ends are connected
and that the unit is connected to a proper source of AC volt­age.
2. Check the fuse. Make sure the proper fuse size is installed.
Make sure the fuse is not open.
If the dew/frost point reads incorrectly, first check the standard preventive maintenance items:
1. Clean the mirror. See “Cleaning the sensor mirror” on page 31.
2. Balance the sensor optics. See “Balancing the sensor optics” on page 31.
If the above procedures do not correct the problem, verify that the platinum resistance thermometer in the sensor is reading correctly with the following steps:
1. Disconnect the 6-pin J9 connector from the main circuit board in the unit. After J9 has been disconnected, and after the sen­sor mirror has stabilized at ambient temperature, the mea­sured dew point should equal the ambient temperature.
2. Check this temperature reading against another reliable ther­mometer placed near the dew point sensor.
3. If the platinum resistance thermometer is faulty, return the sensor to General Eastern for factory service.
An alternative method for checking the accuracy of the unit’s elec­tronics is to use a precision resistance decade box in place of the platinum thermometer. Wire the decade box to the unit’s sensor connector as shown in Figure 13, and verify that the resistance settings shown in the table produce the temperatures shown.
36 E4-1111H-GE Operator’s Manual
Figure 13 — Using a Resistance Decade Box
No analog output
No digital
RS-232C output
Chapter 5 — Service 37
If there is no analog output, check the analog output scaling. See “Function 1: Scale outputs” on page 24.
Check the output scaling for the RS-232C port. See “Function 1: Scale outputs” on page 24.
No cooling and/
or heating
The RS-232 or analog output can be used to indicate whether or not cooling and heating are operating correctly. If DIP Switch 2 is set to HEAT, the displayed temperature should increase. If Switch 1 is set to COOL, the temperature should decrease.
The following procedure can be used to determine whether or not the sensor thermoelectric cooler has failed.
1. Disconnect the sensor from the sensor cable.
2. Using a jumper wire, connect pins A and B of the 1123HK cable or 1 and 9 of the D-2K cable at the sensor end of the cable connector.
3. Connect a digital voltmeter across resistor R28 on the main circuit board.
4. Set DIP Switch 1 to COOL (press the right-hand side). The volt­age should read approximately 0.22 (±0.02) VDC. Return the switch to its normal position.
5. Set DIP Switch 2 to HEAT (press the right-hand side). The volt­age should read 0.06 (±0.01) VDC with the opposite sign from the previous reading. Return the switch to its normal position.
If the above voltages read correctly, the instrument is operating correctly and the sensor thermoelectric cooler is likely to have failed. Return the system to General Eastern for factory service.
38 E4-1111H-GE Operator’s Manual
Appendix A — Specifications
PERFORMANCE
Accuracy: Dew/Frost point:
[complete system at 25°C (77°F)]
Sensitivity:
Repeatability:
Hysteresis:
Measurement range:
Measurement capability:
(at 25°C ambient air temp.)
Full-Scale analog output range: –45°C to +75°C (–49°F to +167°F) dew/frost point
Response time:
Dew/frost point cooling rate: 1.5°C (2.7°F)/sec [typical, above 0°C (32°F)]
Update time:
± 0.2°C (± 0.36°F)
> 0.05°C (0.09°F)
± 0.05°C (±0.09°F)
None
–25°C to +65°C (–13°F to +149°F) dew/frost point
2 sec
Appendix A — Specifications 39
FUNCTIONAL
Analog output:
4–20mA DC, 500 Ohm maximum load
Standard Analog Scaling (field or factory programmable):
–45 to +75°C (–49 to +167°F)
Digital output:
RS-232C
Alarms: Service alarm: TTL compatible. Service flag also available on
RS-232C output.
Balance status:
Power:
Green indicator light. Balance (PACER) flag also available on RS-232C output.
110, 115 or 230 VAC (+/– 10%), 50-60 Hz, 35 Watts maximum, 250V 3AG 2A SB double fuses
Operating ranges:
Dew point sensor
Ambient temperature: –20°C to +80°C (–4°F to + 176°F)
pressure: –14.7 to 300 psig (0 to 22 bar)
Electronics:
Ambient temperature: –20°C to +75°C (–4°F to +167°F)
relative humidity 85% maximum
PHYSICAL
Dimensions:
Weight:
Shipping Weight:
Environmental:
10.5"W x 18.4"H x 4.25"D (267 x 468 x 108 mm) not including sensor
12 lbs (5 kg)
14 lbs (5.5 kg)
Surface-mount, industrial environment (NEMA-4)
40 E4-1111H-GE Operator’s Manual
Appendix B — Warranty and return
procedure
Warranty
General Eastern (the Seller) warrants equipment of its manufacture against defective materials or workmanship for a period of one year from date of shipment. Liability of the Seller under this warranty is limited, at Seller’s option, to:
• Repair or replacement of defective parts at no charge
• Credit adjustment, not to exceed original sales price
This warranty is subject to the following conditions:
• Prompt notification to Seller upon discovery of defects or miss­ing items
• Obtaining a Return Authorization Number from Seller to return defective items to plant as directed
• Return of equipment with freight charges prepaid, or as other­wise agreed
Defects caused by negligence, misuse, improper installation, acci­dent or unauthorized repair or alteration by buyer or user, or any modification, such as changing range resistors, may void this warranty.
This warranty does not include mechanical parts failing from normal usage, nor does it cover limited-life electrical components which deteriorate with age.
This warranty is in lieu of all other warranties, expressed or implied, including the implied warranty of fitness for a particular purpose to the original Purchaser or to any other person. Seller shall not be liable for consequential damages of any kind.
Damaged
Shipments
Appendix B — Warranty and return procedure 41
In case of shipping damage, it is the Buyer’s responsibility to file a claim. The Buyer should inspect the shipping container upon receipt and note any evidence of damage on the freight waybill. If concealed damage is found after opening the container, the
customer should file a claim with the carrier at once. The customer must retain the shipping container and all materials during the life­time of the warranty.
Repaired
Equipment
Instrument
Return Procedure
All repairs are warranted for 90 days. Only the repairs and compo­nents replaced as part of these repairs are covered by this warranty. Other repairs or defective parts are covered by the original warranty, if applicable.
The aforementioned provisions do not extend the original warranty of any article which has been either repaired or replaced by the Seller.
All General Eastern instruments are fully tested and calibrated prior to shipment. Should a problem with the operation of the equip­ment arise, follow the procedure below:
1. Contact the factory to discuss the problem. In countries other than the U.S., the local agent can also be contacted. Sometimes a problem can be resolved by a change in operating procedure or an adjustment to the equipment.
2. If the equipment must be returned to the factory, obtain a return authorization number from General Eastern, and refer­ence the number on the return shipping papers. A written description of the problem should also be included with the instrument.
3. If equipment is not covered by General Eastern’s Warranty Pol­icy, a purchase order should be submitted with the equipment returned. The order should cover one of the following:
a. Open order, authorizing repair of equipment to meet pub-
lished specs. Repair costs will be billed on an actual basis, but will not exceed 50% of the replacement cost without prior customer approval.
b. Order that is not to exceed $500.00 or 30% of the
replacement cost, whichever is higher. If repair costs exceed this amount, the customer will be quoted costs before the work is done.
c. Order to cover cost of test and evaluation only. Amount
based on type of equipment returned. General Eastern will
42 E4-1111H-GE Operator’s Manual
evaluate but not repair the unit. General Eastern will call the customer to discuss the evaluation and quote the cost of repair or replacement.
To expedite repairs and reduce costs, General Eastern recom­mends options (a) or (b).
4. After receiving a Return Authorization Number, the equip-
ment must be returned freight prepaid.
5. General Eastern reserves the right to apply a minimum service
charge in cases where an instrument is returned for repairs or recalibration, but does not require service.
Returning equipment without a Return Authorization number and Purchase Order significantly delays turnaround time and incurs additional costs. To expedite repairs and reduce costs, please follow the above instructions.
Note: General Eastern guarantees NIST traceability and operation within stated specifications. However, claims regarding accuracy or traceability will be covered under warranty only when verified at General Eastern, or by a fully independent testing laboratory. Exam­ples of independent labs are: National Institute of Standards and Technology in the U.S., and the National Physical Laboratory (NPL) in the U.K.
Appendix B — Warranty and return procedure 43
44 E4-1111H-GE Operator’s Manual
Appendix C — Humidity equations and
conversion chart
The following symbols appear in the equations below:
e
= Vapor Pressure, millibars
e
= Vapor Pressure with respect to ice, millibars
i
e
= Vapor Pressure with respect to water, millibars
w
e
= Saturation vapor pressure, ice, millibars
is
e
= Saturation vapor pressure, water, millibars
ws
P
= Total Pressure, millibars
T
= Temperature, °C
T
= Ambient temperature, °C
a
T
= Dew point temperature, °C
d
T
= Frost point temperature, °C
f
Vapor Pressure
Appendix C — Humidity equations and conversion chart 45
Saturation vapor pressure with respect to water is a function of temperature only and is given by the following:
17.502T
e
Saturation vapor pressure with respect to ice requires a minor adjustment of the constants as given by the following:
e
In addition to yielding saturation vapor pressure as a function of ambient temperature, the above equations also yield ambient vapor pressure as a function of dew/frost point.
The total pressure of a gas mixture is equal to the sum of the partial pressure each gas would exert, were it to occupy the same total volume, according to Dalton’s law.
ws
is
6.1121
6.1115
exp=
exp=
--------------------------
240.97 T+
22.452T
--------------------------
272.55 T+
(2)
(1)
Humidity
Relative Humidity is defined as the ratio of the water vapor pres­sure (e) to the saturation vapor pressure (es) at the prevailing ambient or dry bulb temperature (Ta):
()
e

%RH 100
Absolute humidity is expressed as water vapor density: water vapor mass per unit volume of dry air, according to the following:
------
m
Water vapor content expressed as parts per million by volume is given by the following:
PPM
Expressing water vapor content as parts per million by weight (or mixing ratio) requires multiplication of the above by the ratio of the molecular weight of water to that of air as given by the following:
PPM
==
216.7eT
g
----------------------------=
3
T 273.16+
=
V
W
----

e
()
6
-----------------------
10
PeT
0.622 10
=
s
d
eTd()
()
×
100
d
6
------------
Pe
e
wTd
-------------------
e
()
wsTa
e
(3)
(4)
(5)
(6)
A graphical humidity conversion chart is given below.
46 E4-1111H-GE Operator’s Manual
Appendix C — Humidity equations and conversion chart 47
48 E4-1111H-GE Operator’s Manual
Appendix D — Configuring the RS-232
interface
The E4-1111H is configured as Data Terminal Equipment (DTE). The following pins are used on the RS-232 interface:
2 - Transmitted data (TXD)
3 - Received data (RXD)
7 - Signal ground (GND)
To send the output of the monitor to a terminal or a terminal emulator, use the cable arrangement shown below.
Some communications protocols require that the following pins also be used:
4 - Request to send (RTS)
5 - Clear to send (CTS)
6 - Data set ready (DSR)
20 - Data terminal ready (DTR)
The unit, as shipped from the factory, has pins 4 and 5 connected and pins 6 and 20 connected via jumpers JPR4 and JPR5, gener­ating the appropriate control signals for devices requiring these
Appendix D — Configuring the RS-232 interface 49
lines. If you want independent control over these signals, JPR4 and JPR5 can be removed. These jumpers are located on the printed circuit board and are clearly marked.
If you wish to interface the unit to Data Communications Equip­ment (DCE), use the cable configuration shown below.
In order to communicate with the unit, you need a standard serial interface card installed in your computer. If your computer has one serial card, it is addressed as COM1. If there is a second serial card, it is addressed as COM2. You must also have a cable with the following connectors:
• A 25-pin male D connector connecting to the monitor
• A 25-pin female D connector connecting to the computer
Only pins 2, 3, and 7 are used by the monitor. They are connected straight through: 2 to 2, 3 to 3, and 7 to 7.
After you have installed the serial card and connected the cable, you can use the Terminal program in Windows 3.1 to easily communicate with the unit. If Windows is not available, you can use the following BASIC program to interface to the instrument:
10 ON ERROR GOTO 90 20 OPEN COM1: 1200,N,8,1,CS,DS,CD AS #1 30 OPEN SCRN: FOR OUTPUT AS #2 40 B$ = INKEY$: IF B$ <> THEN GOTO 100 CHECK KEYBOARD 50 IF EOF(1) THEN 40 CHECK END OF TRANSMISSION 60 A$ = INPUT$(LOC(1),#1) GET DATA FROM PORT 70 PRINT #2,A$; DISPLAY DATA FROM SERIAL PORT
50 E4-1111H-GE Operator’s Manual
80 GOTO 40 REPEAT LOOP 90 CLOSE: GOTO 20 ERROR, CLOSE AND REPEAT 100 IF B$<>CHR$(27) THEN PRINT $1,B$;: GOTO 50 CHECK FOR ESCAPE 110 FOR I=1 TO 5 SEND BURST OF ESCAPE CHAR UNTIL RECOGNIZED 130 PRINT #1, CHR$(27) 150 LINE INPUT #1, A$ GET STRING 160 IF INSTR(A$, VERSION ) THEN 210 170 IF INSTR(A$, SCALE ) THEN 210 180 IF INSTR(A$, CALIBRATE ) THE 210 190 NEXT I 210 GOTO 70
Notes:
Line 20: Serial input = COM1, baud rate = 1200, parity = NONE, no.
of data bits = 8, no. of stop bits = 1, timeout = 10
SECONDS
, set to file #1.
Line 30: File #2 set as the video display terminal.
Line 40: Input from keyboard. If there is no input then continue. If
there is input, send it to E4-1111H and to video display terminal.
To send data to your printer, insert the following line in the program:
75 LPRINT A$
For further information on interfacing, refer to the factory or to the Electronic Industries Association (EIA) standard for interfacing.
Appendix D — Configuring the RS-232 interface 51
52 E4-1111H-GE Operator’s Manual
Appendix E — Glossary
depression
capability
PACE R
parameter
scaling
The temperature difference by which the chilled mirror can be lowered from the ambient temperature.
General Eastern’s patented Programmable Automatic Contaminant Error Reduction system, which consolidates soluble contaminants to reduce their effect on system accuracy. See “The PACER Cycle” on page 21.
A measured quantity available for display by the unit, such as Dew Point in °C, Humidity in Grams/Kilogram, or Pressure in Bar.
The process of selecting the maximum and minimum output values of a chosen parameter.
Appendix E — Glossary 53
54 E4-1111H-GE Operator’s Manual
Loading...