GE Medical Systems Information Technologies SHU-WMTS Users Manual

ApexPro
Telemetry System
Operator’s Manual
Software Version 3
2001989-134 Revision D
The information in this manual only applies to ApexPro Telemetry System software version 3. It does not apply to earlier software versions. Due to continuing product innovation, specifications in this manual are subject to change without notice.
Listed below are GE Medical Systems Information Technologies trademarks. All other trademarks contained herein are the property of their respective owners.
APEX, ApexPro, DINAMAP Pro, and UNITY NETWORK are trademarks of GE Medical Systems Information Technologies registered in the United States Patent and Trademark Office.
CD Telemetry
®
-LAN, CIC Pro, IMPACT.wf, PRN 50 are a trademarks of GE Medical Systems Information
Technologies.
© GE Medical Systems Information Technologies, 2003, 2004. All rights reserved.
T-2 ApexPro Telemetry System Revision D
2001989-134 12 March 2004
CE Marking Information
CE Marking Information
Compliance
The ApexPro telemetry system bears CE mark CE-0459 indicating conformity with the provisions of the Council Directive 93/42/EEC concerning medical devices, and fulfills the essential requirements of Annex I of this directive. The product is radio-interference protection class A in accordance with EN 55011.
The country of manufacture can be found on the equipment labeling.
The product complies with the requirements of standard EN 60601-1-2 “Electromagnetic Compatibility-Medical Electrical Equipment.”
The safety and effectiveness of this device has been verified against previously distributed devices. Although all standards applicable to presently marketed devices may not be appropriate for prior devices (i.e. electromagnetic compatibility standards), this device will not impair the safe and effective use of those previously distributed devices. See user's information.
Revision D ApexPro Telemetry System CE-1
2001989-134
CE Marking Information
Exceptions
The CIC Pro and ApexPro server is suitable for use in the specified electromagnetic environment. The customer and/or the user of the CIC Pro and ApexPro server should assure that it is used in an electromagnetic environment as described below:
CE Exception Table
EN60601-1-2 Clause 36 Exception Electromagnetic Environment Guidance
36.202.1 Immunity: ESD Direct - Discharges of 6 KV or greater to the rear I/O connector area may cause the system to lock up, thus experiencing loss of data and loss of functionality. Operator intervention may be required.
Likelihood of occurrence: Remote During testing there were 2 occurrences out of 1,920 discharges.
The rear I/O connector area is not considered to be user accessible during normal operation.
36.202.3.1 Immunity: Fast Transient Transients on the AC power line of +/- 1 KV or higher may cause momentary network packet loss (i.e. waveform and/or numeric data), thus experiencing momentary loss of data at the time of the surge.
36.202.3.2 Immunity: Fast Surges Surges on the AC power line of +/- 1 KV or higher may cause momentary network packet loss (i.e. waveform and/or numeric data), thus experiencing momentary loss of data at the time of the surge.
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial and/or hospital environment.
Mains power quality should be that of a typical commercial and/or hospital environment.
CE-2 ApexPro Telemetry System Revision D
2001989-134
CE Marking Information
CE Exception Table
EN60601-1-2 Clause 36 Exception Electromagnetic Environment Guidance
ApexPro and ApexPro CH Transmitters
36.202.2 Immunity: Radiated Fields If operated in the midst of the conditions outlined in EMC standard EN60601-1-2 (Radiated Immunity 3 V/m), fields in excess of 1 V/m may cause waveform distortions and erroneous numeric data at various electromagnetic interference (EMI) frequencies.
ApexPro Antenna System
36.202.2 Immunity: Radiated Fields If operated in the midst of the conditions outlined in EMC standard EN60601-1-2 (Radiated Immunity 3 V/m), fields in excess of 0 V/m in the frequency ranges of 520-534 MHz and 645-660 MHz may cause loss of telemetry.
Review the AAMI EMC Committee technical
information report (TIR-18) titled Guidance on electromagnetic compatibility of medical devices for clinical/biomedical engineers ­Part 1: Radiated radio-frequency electromagnetic energy. This TIR provides
a means to evaluate and manage the EMI environment in the hospital.
Manage (increase) distance between
sources of EMI and susceptible devices.
Manage (remove) devices that are highly
susceptible to EMI.
Lower power from internal EMI sources
under hospital control (i.e., paging systems).
Label devices susceptible to EMI.Educate staff (nurses and doctors) to be
aware of, and to recognize, potential EMI related problems.
Review the AAMI EMC Committee technical
information report (TIR-18) titled Guidance on electromagnetic compatibility of medical devices for clinical/biomedical engineers ­Part 1: Radiated radio-frequency electromagnetic energy. This TIR provides
a means to evaluate and manage the EMI environment in the hospital.
Manage (increase) distance between
sources of EMI and susceptible devices.
Manage (remove) devices that are highly
susceptible to EMI.
Lower power from internal EMI sources
under hospital control (i.e., paging systems).
Label devices susceptible to EMI.Educate staff (nurses and doctors) to be
aware of, and to recognize, potential EMI related problems.
Revision D ApexPro Telemetry System CE-3
2001989-134
CE Marking Information
CE Exception Table
EN60601-1-2 Clause 36 Exception Electromagnetic Environment Guidance
Apex Oximeter
36.202.2 Immunity: Radiated Fields If operated in the midst of the conditions outlined in EMC standard EN60601-1-2 (Radiated Immunity 3 V/m), fields in excess of
0.5 V/m may cause waveform distortions and erroneous numeric data at various electromagnetic interference (EMI) frequencies.
Xpod Oximeter
36.202.2 Immunity: Radiated Fields If operated in the midst of the conditions outlined in EMC standard EN60601-1-2 (Radiated Immunity 3 V/m), fields in excess of 1 V/m may cause waveform distortions and erroneous numeric data at various electromagnetic interference (EMI) frequencies.
Review the AAMI EMC Committee technical
information report (TIR-18) titled Guidance on electromagnetic compatibility of medical devices for clinical/biomedical engineers ­Part 1: Radiated radio-frequency electromagnetic energy. This TIR provides
a means to evaluate and manage the EMI environment in the hospital.
Manage (increase) distance between
sources of EMI and susceptible devices.
Manage (remove) devices that are highly
susceptible to EMI.
Lower power from internal EMI sources
under hospital control (i.e., paging systems).
Label devices susceptible to EMI.Educate staff (nurses and doctors) to be
aware of, and to recognize, potential EMI related problems.
Review the AAMI EMC Committee technical
information report (TIR-18) titled Guidance on electromagnetic compatibility of medical devices for clinical/biomedical engineers ­Part 1: Radiated radio-frequency electromagnetic energy. This TIR provides
a means to evaluate and manage the EMI environment in the hospital.
Manage (increase) distance between
sources of EMI and susceptible devices.
Manage (remove) devices that are highly
susceptible to EMI.
Lower power from internal EMI sources
under hospital control (i.e., paging systems).
Label devices susceptible to EMI.Educate staff (nurses and doctors) to be
aware of, and to recognize, potential EMI related problems.
CE-4 ApexPro Telemetry System Revision D
2001989-134
CE Marking Information
CE Exception Table
EN60601-1-2 Clause 36 Exception Electromagnetic Environment Guidance
Accutracker DX NBP Monitor
36.202.1 Immunity: ESD Air — Discharges in excess of ±6 Kv may cause the cuff to deflate and the unit to lock up. By turning the power switch off, then back on (manual reset), the unit will be restored to the user-defined settings and normal operation.
The Accutracker DX blood pressure monitor
should be kept in the carrying pouch supplied with each unit.
Care should be taken to minimize the ESD
potential when the Accutracker DX blood pressure monitor is removed from the pouch. This includes:
Handling the unit in an ESD-protected
area.
Maintaining humidity levels of 50%
relative humidity or greater.
Discharging ESD potentials on human
hands prior to handling the unit out of the pouch.
Radio and Telecommunication Terminal Equipment Directive
The ApexPro telemetry system transmitters bear the CE mark CE 0123 indicating conformity with the provisions of the Council Directive 1999/5/ EC of 9 March 1999 concerning R&TTE as tested by MKES BABT Services GmbH Notified Body TUV (0123).
The product complies with the requirements of standard EN 300 220-1 [ETSI 300 220-1 v1.3.1]: “Electromagnetic Compatibility and Radio Spectrum Matters (ERM); Short Range Devices (SRD); Part 1: Technical Characteristics and Test Methods”.
Revision D ApexPro Telemetry System CE-5
2001989-134
Restrictions
CE Marking Information
The Radio and Telecommunication Terminal Equipment Directive (R&TTE) states that radio equipment operating in frequency bands for which the use has not been harmonized across the European Community have to be identified by an equipment symbol. ApexPro is classified as Class 2 Equipment and bears the following equipment symbol (also
called the Alert Mark): .
Not all European member states have approved the frequencies within
433. 25 MHz to 434.75 MHz transmitted by the ApexPro telemetry system for medical telemetry applications.
The following European member states have approved the frequencies within 433. 25 MHz to 434.75 MHz transmitted by the ApexPro telemetry system for medical telemetry application:
Austria: Approved for 433.25 MHz to 434.75 MHz; individual
approval from local authorities necessary.
Belgium
Denmark: Individual registration by user necessary in accordance
with “Danish Radio Interface Regulation for radio equipment for medical telemetry No. 00024.”
Finland
France
Germany
Ireland
Italy: Approved providing compliance with Art. 344 p. to 8 of the
Codice P.T.
Luxembourg
Netherlands
Norway: Approved for 441.750 MHz to 441.975 MHz.
Spain
Sweden
United Kingdom: Approved for 458.975 MHz to 459.100 MHz.
CE-6 ApexPro Telemetry System Revision D
2001989-134
General Information
CE Marking Information
This manual is an integral part of the product and describes its
intended use. It should always be kept close to the equipment. Observance of the manual is a prerequisite for proper product performance and correct operation and ensures patient and operator safety.
The symbol means ATTENTION: Consult accompanying
documents.
Information which refers only to certain versions of the product is
accompanied by the model number(s) of the product(s) concerned. The model number is given on the nameplate of the product.
The warranty does not cover damages resulting from the use of
accessories and consumables from other manufacturers.
GE Medical Systems Information Technologies is responsible for the
effects on safety, reliability, and performance of the product, only if:
assembly operations, extensions, readjustments, modifications,
or repairs are carried out by persons authorized by GE Medical Systems Information Technologies;
the electrical installation of the relevant room complies with the
requirements of the appropriate regulations; and,
the device is used in accordance with the instructions for use.
All publications conform with the product specifications and
applicable IEC publications on safety and essential performance of electromedical equipment as well as with applicable UL and CSA requirements and AHA recommendations valid at the time of printing.
The GE Medical Systems Information Technologies quality
management system complies with the international standards EN ISO 9001 and EN 46001, and the Council Directive on Medical Devices 93/42/EEC.
Revision D ApexPro Telemetry System CE-7
2001989-134
For your notes
CE Marking Information
CE-8 ApexPro Telemetry System Revision D
2001989-134

Contents

1 The Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Manual Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Intended Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Manual Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Product References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Illustrations and Names . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Common Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
“Clicking” the Mouse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Radio Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Check Boxes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Scroll Bars . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Popup Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Unit Defaults Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Telemetry Alarm Control Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Telemetry Unit Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-11
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
For Your Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
System Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Reference Literature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Underwriters Laboratories, Inc. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Equipment Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Revision D ApexPro Telemetry System i
2001989-134
3 Equipment Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
ApexPro Telemetry System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Compatibility with Bedside Monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
ApexPro Transmitters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
ApexPro CH Transmitter (not for sale outside of the U.S. and Canada) . . . . . . . 3-10
ApexPro FH Transceiver (not for sale outside of the U.S.) . . . . . . . . . . . . . . . . . 3-11
Apex Oximeter SpO2 Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Xpod™ Oximeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
Accutracker DX Noninvasive Blood Pressure (NBP) Monitor . . . . . . . . . . . . . . . 3-14
DINAMAP® PRO Series Monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Antenna System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Receiver System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Unity Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
CIC Pro Clinical Information Center . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
4 Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
ApexPro Transmitter Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
ApexPro Transmitter Battery Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Battery Functional Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
ApexPro Transmitter Leadwires . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Apex Oximeter Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Apex Oximeter Battery Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Battery Functional Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Apex Oximeter Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Xpod Oximeter Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Accutracker DX Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Tips for Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Battery Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-10
Accutracker DX Functional Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Accutracker DX Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
DINALink™ Serial Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
Interconnection Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
ii ApexPro Telemetry System Revision D
2001989-134
5 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Biocompatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
General Cleaning/Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
Cleaning the Transmitters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
More Intensive Disinfecting or Sterilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Cleaning the Oximeter and Accutracker DX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Cleaning the Power Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Transmitter and Leadwire Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Storage Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-10
Technical Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-11
6 Telemetry Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Telemetry Factory Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Service Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Telemetry Unit Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Viewing Telemetry Unit Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
CIC Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
Adjusting Alarm Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Current Telemetry Listings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Full Disclosure Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
Revision D ApexPro Telemetry System iii
2001989-134
7 Admit/View a Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
About Admitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Switching Transmitters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Terminology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Admit Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Admit Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Discharge Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
New Patient Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
Clearing a Patient Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Move Telemetry Patients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11
Moving Locked/Unlocked Beds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11
Viewing a Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
Relearn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-13
Viewing Another Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
Viewing in the Single Patient Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-14
Viewing in the Multiple Patient Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-15
Viewing Patients Through Alarm Condition Indicators . . . . . . . . . . . . . . . . . . . . 7-17
8 Alarm Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Alarm Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Patient Status Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
System Status Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
Alarm Control Tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
Accessing the Alarm Control Tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-5
Parameter Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
Parameter Alarm Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-7
Arrhythmia Alarm Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8
Alarms On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Recalling Unit Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12
Alarm Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12
Printing Alarm Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-13
Silencing Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13
Alarm Pause Breakthrough . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14
Unit Default Settings for Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-16
Telemetry Alarm Control Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-16
iv ApexPro Telemetry System Revision D
2001989-134
9 Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Initiating a Manual Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Automatic Alarm Graphs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Transmitter Initiated Graphs (Manual Graphs) . . . . . . . . . . . . . . . . . . . . . . . . . . .9-4
Graph Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Graph All Patients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
Initiating a Graph All Patients Request . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
Graph Location Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
Stopping a Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9
Graph Paper Out Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-10
Graph Setup Tab Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11
Graph Waveforms Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-12
Graph Location Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13
Graph Speed Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-15
Laser Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-16
10 Patient Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
Alarm Histories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Event Directory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-5
Event Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
Alarm Histories Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-6
Graphic Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8
Trend Directory Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-8
Cursor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-9
Time Resolution Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-10
Graphic Trends Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-10
Printing Graphic Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-12
Vital Signs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-13
Data Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-13
Sort Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-14
Increment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-14
Scroll Bars . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-15
Printing Vital Signs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-15
Data Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-16
Revision D ApexPro Telemetry System v
2001989-134
Time Focus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-17
Full Disclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-18
Full Disclosure Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-19
11 ECG Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3
Data Synchronization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3
Patient Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
Skin Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-4
Electrode Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
Special Considerations for 6-Lead Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . 11-10
V FAIL Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-10
Relearn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-10
ECG Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-11
ECG in the Multiple Patient Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-11
ECG in the Single Patient Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-12
ECG Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-13
ECG Artifact . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-14
ECG Tab Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-15
Accessing the ECG Tab Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-15
Display Lead . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-16
Relearn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-16
Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-16
Detect Pace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-17
Lead Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-22
Arrhythmia . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-24
Full Arrhythmia Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-26
Lethal Arrhythmia Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-29
PVC Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-29
ST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-29
Va Lead and Vb Lead . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-29
AFIB Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-30
ST Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-31
ST Deviation Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-32
Arrhythmia Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-33
Pacemaker Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-35
Interface Connector Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-35
ST Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-37
vi ApexPro Telemetry System Revision D
2001989-134
12 SpO2 Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-3
SpO2 Probe Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-4
Infants and Pulse Oximetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-5
Signal and Data Validity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-6
Signal Strength Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-6
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-7
SPO2 Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-8
SPO2 in the Multiple Patient Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-8
SPO2 in the Single Patient Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-9
SPO2 Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-9
SPO2 Tab Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-10
Accessing the SPO2 Tab Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-10
Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-11
Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-11
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-12
SpO2 Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-12
13 NBP Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-3
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-4
Programming the Blood Pressure Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-5
Setting the Measurement Interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-5
Setting Test Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-6
Setting Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-8
Software and Hardware Versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-8
Patient Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-9
Microphone Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-11
NBP Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-13
NBP in the Multiple Patient Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-13
NBP in the Single Patient Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-13
NBP Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-14
Revision D ApexPro Telemetry System vii
2001989-134
Pressures Tab Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-15
Accessing the Pressures Tab Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-15
Auto . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-16
Cuff Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-16
Clear Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-16
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-17
NBP Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-17
Message Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Message Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Alarm Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Graph Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-5
Transmitter-Related Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-6
System Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-6
Patient Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-7
Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
Contact Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
Abbreviations and Symbols . . . . . . . . . . . . . . . . . . . . . . . .C-1
Abbreviations and Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-3
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-7
viii ApexPro Telemetry System Revision D
2001989-134
ApexPro FH Transceiver . . . . . . . . . . . . . . . . . . . . . . . . . .D-1
The Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-3
About the ApexPro FH Transceiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-3
Programming Transceivers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-4
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-5
Warnings, Cautions, and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-5
Equipment Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-8
ApexPro FH Transceiver Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . .D-8
Push Button Function and Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-8
Attendant Present/Procedure Alarm Silence (PAS) Unlock Button . . . . . . . . . . . .D-9
Procedure Alarm Silence (PAS) Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-10
LED Indicators Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-12
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-14
Connecting the ApexPro FH Transceiver to Accessory Devices . . . . . . . . . . . . .D-14
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-18
General Cleaning/Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-18
Storage and Usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-23
Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-26
Compliance Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-26
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Index-1
Revision D ApexPro Telemetry System ix
2001989-134
For your notes
x ApexPro Telemetry System Revision D
2001989-134

1 The Basics

Revision D ApexPro Telemetry System 1-1
2001989-134
For your notes
1-2 ApexPro Telemetry System Revision D
2001989-134

About This Manual

Manual Purpose

This manual contains the instructions necessary to operate the ApexPro telemetry system safely and in accordance with its function and intended use.
This manual addresses the operation of the ApexPro telemetry system at the CIC Pro. Be sure to read the entire manual before using the equipment.

Intended Audience

This manual is geared for clinical professionals. Clinical professionals are expected to have a working knowledge of medical procedures, practices, and terminology, as required for monitoring of critically ill patients.
The Basics: About This Manual
This manual assumes that you are familiar with the operating procedures of the CIC Pro. If you would like more information about using the CIC Pro, refer to the CIC Pro Clinical Information Center Operator’s Manual.
This manual also assumes that you are familiar with the operation of a two-button computer mouse. If you would like more information about operating the mouse, refer to the CIC Pro Clinical Information Center Operator’s Manual, or to the documentation supplied with the mouse.
Revision D ApexPro Telemetry System 1-3
2001989-134

Revision History

The Basics: About This Manual
Each page of the document has the document part number and revision letter at the bottom of the page. The revision letter changes whenever the document is updated.
Revision Date Comments
A 25 June 2003 Initial release of this document, corresponding to
ApexPro telemetry system software version 3.
B 23 October 2003
C 21 January 2004
D 12 March 2004
Updated to support the ApexPro FH Transceiver Added additional CE Exceptions for the ApexPro
system.
Added FCC ComplianceAdded Industry Canada ComplianceChanged Nurse Graph to Event Marker Graph and
Transmitter Nurse to Event Marker
Added Data Source informationChanged recommended cleaning solutionsAdded information regarding time discrepancy
between waveforms
Added information on stopping manual graphsMade correction to INTFC connector ports 1 and 2Updated Equipment Symbols
Warning added regarding adjusting System Status
Alarms levels
Added verify lead information to the admit procedure
and electrode placement
Message alarm level added to System Status AlarmGraph Location Settings information revisedRemoved reference to alarm level settings for ARR
SUSPEND, LEADS FAIL, and OFF NETWORK
Changes made to Appendix D for the ApexPro FH
Transceiver
1-4 ApexPro Telemetry System Revision D
2001989-134

Manual Conventions

This section describes terminology, standards, and other conventions that are used throughout this manual.

Product References

In this manual:
The ApexPro telemetry system is referred to as the ApexPro system,
or simply the system.
The CIC Pro Clinical Information Center is referred to as the clinical
information center, the CIC Pro or the central station.
The SunTech Medical Systems Accutracker DX noninvasive blood
pressure monitor is referred to as the blood pressure monitor or the Accutracker.
The PRN 50 Digital Writer is referred to as the digital writer or the
writer.
The Direct Digital Writers are referred to as DDWs or writers.
The laser printer is referred to as the printer.
The Basics: Manual Conventions

Definitions

The following terms are used in this manual:
Buttons — The word “button” is defined in two ways:
1. A button is a labeled gray or red rectangle on the CIC Pro. Clicking on a button with the mouse pointer opens a tab sheet or performs the specified action (such as Print). Red buttons are used to view beds in alarm.
2. A button is a labeled circle, square, or rectangle located on the ApexPro transmitter, the Apex Oximeter, or the Accutracker blood pressure monitor. Pressing the button with your finger activates it.
NOTE
The computer mouse used with the CIC Pro also has two buttons. Refer to the CIC Pro Clinical Information Center Operator’s Manual for information about using those buttons.
Revision D ApexPro Telemetry System 1-5
2001989-134
The Basics: Manual Conventions
Messages/Prompts — A message is text that appears on the CIC Pro. It informs you of conditions occurring that are not necessarily part of normal operating conditions. Prompts are text messages that appear, instructing you to perform a specific action.
Multiple patient viewer — The multiple patient viewer is the CIC Pro display in its normal state. Bed windows for admitted patients are shown, as well as a menu bar at the bottom of the display.
Screen text — Any text that appears on the CIC Pro display. In this manual, screen text is shown in italics (for example, ECG, SAVING, etc.).
Single patient viewer — When you click on a patient’s bed window at the CIC Pro, the display rearranges to accommodate a set of tab sheets (see definition below) in the bottom portion of the display. This set of tab sheets is referred to as the single patient viewer because it contains information specific to one patient.
Tab options — Tab options are the choices and text entry fields available on a tab sheet. The information presented as tab options may pertain to a patient’s data, or may be control information (such as alarm settings) that can be modified to meet the user’s specific needs.
Tab sheets — Tab sheets look like labeled index cards. The menu tab labels indicate the type of information to be viewed and/or changed on its tab sheet. Clicking on a tab brings its tab sheet to the front of the “index card” stack, or to the front of the currently viewed window.
Tabs — Tabs are the labeled section of the tab sheets. Clicking on a tab brings its tab sheet to the front. Tab and tab sheet are sometimes used interchangeably.

Illustrations and Names

All illustrations in this manual are provided as examples only. They may not necessarily reflect your telemetry monitoring setup or data displayed on your equipment.
In this manual, all names appearing in examples and illustrations are fictitious. The use of any real person’s name is purely coincidental.
1-6 ApexPro Telemetry System Revision D
2001989-134

Common Operations

Some operations are used repetitively at the CIC Pro or the ApexPro telemetry system. Rather than explaining how to perform each operation every time it appears in this manual, these operations are presented below. Please familiarize yourself with the proper procedure for each.

“Clicking” the Mouse

The term “click” refers to positioning the mouse pointer on a selection and pressing the left mouse button one time.
In situations where the right mouse button should be pressed, this is specifically called out. In all other cases, assume that you should press the LEFT mouse button.
Pressing the mouse button two times in a row is called double clicking. In situations where the mouse needs to be double clicked to perform a function, this is specifically called out. In all other cases, assume that you only need to click the mouse button ONE time.
The Basics: Common Operations

Radio Buttons

To use a radio button control, click on the white circle (radio button) or the text next to it. When selected, a black dot is shown in the white circle. To deselect a radio button control, click again on the label text or in the white circle. When it is not selected, no black dot is shown.
Radio Buttons — “Dot” in center indicates active selection.
007A
Revision D ApexPro Telemetry System 1-7
2001989-134

Check Boxes

The Basics: Common Operations
To use check box controls, click on the square or the text next to it. When selected, a check mark is shown in this square. To deselect a check box control, click again on the text or in the square. When deselected, no check mark is shown.
Check Boxes —
indicates active
selection.
078B

Scroll Bars

Use horizontal and vertical scroll bars to move a window’s contents left/ right and up/down. Place the mouse pointer on the appropriate arrow to move the scroll bar, or click and hold the mouse button down while dragging the scroll bar until the desired information is displayed.
Vertical Scroll Bar
Horizontal Scroll Bar
359A
1-8 ApexPro Telemetry System Revision D
2001989-134

Popup Lists

The Basics: Common Operations
Clicking in a text field may produce a down arrow button on the right side of the field. This arrow button is used to open a popup list. A popup list is a list of options available for that particular field. Use the mouse to click on the arrow button, which opens the popup list.
Arrow Button — indicates there is a popup list.
Popup List
332A
Once the popup list is open, use the mouse to click on an option. This selects the option and closes the popup list.
NOTE
If you click on the right side of a field, the down arrow button and the popup list of selections may appear simultaneously.
Revision D ApexPro Telemetry System 1-9
2001989-134
The Basics: Unit Defaults Worksheet

Unit Defaults Worksheet

This worksheet has been provided as an optional reference tool to record your care unit’s default settings. Fill out the information and keep it in a prominent place to refer to your setup. For your convenience, the factory default settings appears dimmed. Fill in only those settings that differ from the factory default ones.
NOTE
Before you fill it out, you may want to make additional copies of the worksheet for future use.
Date: Unit:

Telemetry Alarm Control Defaults

Parameter Limits and Alarm Levels Low High Level Arrhythmia Alarm Levels Levels
HR bpm 50 150 Warning Asystole Crisis
ST-I mm -2.0 2.0 Warning VFIB/VTAC Crisis
ST-II mm -2.0 2.0 Warning V Tach Crisis
ST-III mm -2.0 2.0 Warning VT > 2 Crisis
ST-V mm -2.0 2.0 Warning V Brady Crisis
ST-V2 mm -2.0 2.0 Warning Acc Vent Advisory
ST-V3 mm -2.0 2.0 Warning Pause Advisory
ST-V4 mm -2.0 2.0 Warning Tachy Advisory
ST-V5 mm -2.0 2.0 Warning Brady Advisory
ST-V6 mm -2.0 2.0 Warning R on T Message
ST-AVR mm -2.0 2.0 Warning Couplet Message
ST-AVL mm -2.0 2.0 Warning Bigeminy Message
ST-AVF mm -2.0 2.0 Warning Trigeminy Message
NBP-S mmHg 80 200 Warning PVC Message
NBP-D mmHg 20 120 Warning Irregular Message
NBP-M mmHg 40 140 Warning Atrial Fib Message
SPO2 % 90 105 Warning
SPO2-R bpm 50 150 Warning System Alarm Levels Levels
RR breaths/min 5 30 Warning Change Battery Sys Warning
RR-APNEA seconds 30 Warning Off Network Sys Warning
PVC #/min 6 Advisory Arr Suspend Sys Warning
Leads Fail Sys Warning
Probe Off Sys Warning
1-10 ApexPro Telemetry System Revision D
2001989-134
Loading...
+ 95 hidden pages