This Service manual provides information needed to service the Responder™ 2000. This manual should only be used by
technical personnel trained to service the Responder 2000.
This chapter contains general information for servicing the Responder 2000.
This Service manual assumes familiarity with the controls and basic operation of the Responder 2000. Detailed
information regarding controls, operation, set-up, and regular maintenance procedures are found in the Operator’s
manual. If necessary, review the Operator’s Manual before servicing the Responder 2000.
TOPIC
Description 9
Controls and Indicators 10
General Service Information 11
PAGE
2025653-048 Revision B Responder™ 2000 Page 8
SECTION 1: Introduction
Description
The Responder 2000 is a defibrillator/monitor/pacemaker intended for use by personnel trained in its operation. The
device is lightweight, portable, easy to use and reliable. It incorporates a 320 x 240 transmissive color TFT color display
for wide viewing angles in all light conditions. The device operates using either an AC power supply or internal
rechargeable Li-Ion battery. The device provides continuous ECG monitoring and three types of therapies: defibrillation,
cardioversion and external pacing. Defibrillation can be applied manually or semi-automatically. Pacing therapy can be
either fixed or demand. The device employs patented RHYTHMx
Biphasic waveform delivers impedance-compensated energy ranging from 2-270 Joules. Features and options include
external paddles, spoons, disposable pads, 3- and 5-lead ECG, pulse oximetry (SpO2), built-in 60 mm thermal printer,
internal storage of event history and remote synchronization to bedside monitor.
®
software which provides ECG rhythm analysis. STAR®
2025653-048 Revision B Responder™ 2000 Page 9
Controls and Indicators
SECTION 1: Introduction
2025653-048 Revision B Responder™ 2000 Page 10
SECTION 1: Introduction
General Service Information
Refer to the Operator’s Manual for setup (and configuration options) required before placing the Responder 2000 into
service.
Navigation
Use the Rotary Selection knob to navigate the Responder 2000 user interface. The Rotary Selection knob is used for:
• Scrolling through menus and sub-menus
• Selecting soft keys
• Setting values
Passwords
The default service password is “HEART1”. See the note under Maintenance Mode Tests in Section 5 (on page 54) for
detailed password instructions.
Battery Maintenance
The Responder 2000 has a rechargeable battery requiring periodic calibration. Calibration consists of a full charge, full
discharge, and full charge of the battery. The cycle may take up to 20 hours to complete.
To calibrate the battery using the optional external charger:
1. Insert the battery into the external charger.
2. Press the calibration button on the battery. The Mode light turns red to indicate calibration in progress.
To stop calibration, press the calibration button again. The Mode light turns green and the battery begins to
charge.
To calibrate the battery without the external charger:
1. Insert the battery into the Responder 2000.
2. Plug the Responder 2000 into an AC outlet.
3. Allow the battery to fully charge.
4. Disconnect the AC power cord and leave on the Responder 2000 until the battery is fully discharged (full
discharge takes about four hours).
5. Reconnect the AC power cord and allow the battery to fully charge.
2025653-048 Revision B Responder™ 2000 Page 11
SECTION 2: Maintenance
Overview
Use the following procedures to upgrade the Responder 2000 software:
• Upgrade Software to update one or more software files
• Upgrade Printer Software to upgrade printer software and fonts
TOPIC PAGE
Equipment Setup 13
Upgrade Software 14
Upgrade Printer Software 16
Update Serial and Model Numbers 17
SECTION 2: Maintenance
2025653-048 Revision B Responder™ 2000 Page 12
SECTION 2: Maintenance
Equipment Setup
Required Tools
• Serial Cable
• PC (Windows 2000 or XP)
Procedure
1. Plug AC power cord.
Note: When performing a software upgrade, the Responder 2000 must be running on AC power.
2. Connect serial cable between the communications port of the PC and the serial port of the Responder 2000.
3. At the desktop on the PC, double-click the Responder 2000 Upgrade icon to open the CodeLink 2000 program.
4. Select the appropriate Com port and click OK.
5. Complete the following on-screen instructions and click OK.
2025653-048 Revision B Responder™ 2000 Page 13
SECTION 2: Maintenance
Upgrade Software
Caution: Procedure Failure
Do not use this procedure to update the printer files. Updating the printer files incorrectly may cause the printer to fail. Use
the Upgrade Printer Software procedure on page 16.
1. Perform the Equipment Setup procedure starting on page 13 until the CodeLink 2000 selection window is
displayed:
2. Click Choosefolder to select all the update files in a folder or Choose file to select a single update file.
3. Browse to the folder (or file) to be uploaded and click OK (or Open). While the files are updating, the status of
each file is displayed.
As the files are updated, the Shock button and then the Manual button flash on the Responder 2000. When all
files are updated, a confirmation screen is displayed
4. Verify Pass is displayed for each file and click OK.
Note: If a file fails to update, retry updating the file individually.
The Upgrade Complete Notification is displayed:
5. Click Yes to upgrade more files (and repeat this procedure from step 2) or No to exit.
2025653-048 Revision B Responder™ 2000 Page 14
SECTION 2: Maintenance
6. Click Exit.
7. Verify the file upgrade on the Responder 2000.
A text file with the software version numbers is included with the upgrade folder. Compare those software
version numbers to the versions listed on the Responder 2000.
a. Press and hold the Power button for 5 seconds to reboot the Responder 2000.
b. Highlight and click the System Menu.
c. Click About.
d. Click Next to scroll through the version screens. Verify all software versions are updated as listed in the
Version.txt file.
2025653-048 Revision B Responder™ 2000 Page 15
SECTION 2: Maintenance
Upgrade Printer Software
When installing the printer software, the files must be upgraded in the following order without reboot:
• printerflash.bin
• printerfont.bin
• printerflash.bin again
Caution: Procedure Failure.
The printer files must be updated without reboot. If the unit is rebooted during the installation, the printer may not work
(the Main Board firmware must then be reinstalled at a repair center).
1. Perform the Equipment Setup procedure starting on page 13 until the CodeLink 2000 selection window is
displayed:
2. Click Choose file to select a single update file.
3. Navigate to the Printer folder and select the printerflash.bin file.
4. Click Open. While the file is updating, the status of the file is displayed.
As the file is updated, the Shock button and then the Manual button flash on the Responder 2000.
5. When the file is updated, a confirmation screen is displayed:
6. Verify Pass is displayed for the file and click OK.
Note: If the file fails to update, try updating the file again.
The Upgrade Complete Notification is displayed:
7. Click Yes and repeat this procedure for the printerfont.bin file and then the printerflash.bin again.
8. Click Exit.
9. Verify the file upgrade on the Responder 2000.
a. Press and hold the Power button for 5 seconds to reboot the Responder 2000.
b. Highlight and click the System Menu.
c. Click About.
d. Click Next to scroll through to the printer version screen. Verify the Printer and Font software versions are
updated as listed in the Version.txt file.
2025653-048 Revision B Responder™ 2000 Page 16
SECTION 2: Maintenance
Update Serial and Model Numbers
When replacing the Main Board, the Serial and Model numbers must be updated.
1. Before replacing the Main Board, record the Serial and Model numbers from the About screen.
Note: If the Responder 2000 cannot be booted, record the model and serial number from the back panel label.
2. Replace the Main Board.
3. Perform the Equipment Setup procedure starting on page 13 until the CodeLink 2000 selection window is
displayed:
4. Click Update Factory Data and enter the previously model and serial numbers.
Note: If the Main Board is not replaced, the fields are not editable.
5. Click OK to accept.
2025653-048 Revision B Responder™ 2000 Page 17
SECTION 3: Troubleshooting
SECTION 3: Troubleshooting
Overview
This section describes how to troubleshoot the Responder 2000. These instructions are intended for use only by service
providers who are specifically trained to service the Responder 2000.
TOPIC PAGE
Safety Precautions 19
Who Should Perform Repairs 19
Service/Replacement Parts 19
Disposal 19
Repair Tools and Equipment 20
Troubleshooting Guide 20
2025653-048 Revision B Responder™ 2000 Page 18
SECTION 3: Troubleshooting
Safety Precautions
WARNING: Shock Hazard
The Responder 2000 is designed to deliver high-voltage therapeutic shock. Before performing any service on the
equipment, read and follow all safety precautions and instructions in the Operator’s Manual.
WARNING: Shock Hazard or Equipment Damage
Before servicing the Responder 2000, disconnect the AC power cord and remove the battery.
WARNING: Shock Hazard or Equipment Damage
Internal components of the Responder 2000 may still contain high voltages even after the AC power cord and battery are
removed. Before working on any internal component, verify high voltages are not present.
WARNING: Shock Hazard or Equipment Damage
Some service activities require the Responder 2000 to be energized with covers removed. Ensure all personnel and
equipment is clear while the equipment is energized.
WARNING: Biological Contamination
During normal operation, the Responder 2000 may be contaminated by blood, body fluids, or other biological agents.
Always assume the Responder 2000 is contaminated and use appropriate safety procedures until decontamination is
performed. Always decontaminate the Responder 2000 in accordance with hospital or facility procedures before servicing
or returning to service. Refer to the Responder 2000 Operator's Manual for recommended cleaning agents and
instructions
Caution: Electrostatic Damage
Always use a wrist grounding strap and anti-static mat while performing service on any internal components.
Caution: Voiding Product Warranty
Any service performed on the Responder 2000 must be provided by authorized service representatives only.
Unauthorized repair voids the product warranty.
Who Should Perform Repairs
Repair and service of the Responder 2000 must be performed by qualified service technicians trained in safe and proper
servicing of the Responder 2000.
Service/Replacement Parts
For service, please contact your local GE agency. For additional information, please visit our Web site at:
http//www.gehealthcare.com.
Please have the serial and model numbers available when contacting Customer Service. (The serial and model numbers
are located on the back (and on the inside) of the Responder 2000.)
Disposal
Always dispose of the Responder 2000, any unserviceable parts or accessories, or batteries in accordance with any local
disposal regulations for equipment containing electronic parts. Note the following precautions:
WARNING: Shock Hazard
Disposal of the Responder 2000 with the battery inserted presents a potential shock hazard.
WARNING: Environmental Contamination
Disinfect the Responder 2000 appropriately prior to disposal. Also recycle or dispose of the lithium-ion battery in
accordance with applicable local regulations.
WARNING: Fire or explosion hazard
Do not burn or incinerate the battery. Recycle or dispose of the lithium-ion battery in accordance with applicable local
regulations.
Caution: Environmental Contamination
Dispose of the pads or electrodes in accordance with any local disposal regulations.
2025653-048 Revision B Responder™ 2000 Page 19
SECTION 3: Troubleshooting
Repair Tools and Equipment
The following tools are needed to perform the procedures in this section.
• TR15 Safety Torx
• TR25 Safety Torx
• Serial cable for software upgrades (with Windows 2000 or XP PC)
• Software upgrade kit
Troubleshooting Guide
General Troubleshooting
Before performing any troubleshooting, check the following:
Ensure all external cables and connections are tight and undamaged.
•If possible, ensure the unit is plugged in to a receptacle with appropriate voltage available (see SECTION 8:
Specifications and Safety for voltage requirements). If no receptacle is available, use a known good, charged
battery to perform any tests.
• Visually inspect the unit for any obvious external damage, including cracks in the display, case, or connectors.
• Ensure all connector pins and sockets are clean, free of debris, and intact.
• Discuss the issues with the operator:
o Have the operator explain and demonstrate the problem.
o Ask about any previous repairs or problems.
o Has the unit been stressed (including extreme heat/cold, submersion, falls, etc.)?
2025653-048 Revision B Responder™ 2000 Page 20
SECTION 3: Troubleshooting
No Boot
Use this procedure to troubleshoot the Responder 2000 when plugged into an AC outlet. For battery troubleshooting,
including an operation failure while using the battery, see “Battery Troubleshooting”.
A. If the AC Power LED is not lit
1. Verify power at the AC outlet.
2. Verify AC power available from the AC power cord.
3. Replace power supply.
4. Replace the Power ControlBoard.
B. If the AC Power LED is lit
1. Check Power button for mechanical operation.
2. Check cables from Power ControlBoard to MainBoard.
3. Ensure cables are properly seated, tight, and undamaged.
4. Replace Power ControlBoard.
5. Replace MainBoard.
Boot, but no Response
Use this procedure to troubleshoot the Responder 2000 when the unit seems to boot up normally (i.e., the Front Panel
LEDs flash and a speaker tone is heard when the Power button is pressed), but does not respond to any subsequent user
input.
A. If the ECG trace is not moving
1. Replace the Main Board.
B. The Rotary Selection knob works properly
1. Check buttons for mechanical operation.
2. Check cables from Power ControlBoard to MainBoard.
Ensure cables are properly seated, tight, and undamaged.
3. Replace Power ControlBoard.
4. Replace MainBoard.
C. The Rotary Selection knob does not work properly
1. Replace Rotary Selection knob.
Note: The figure below shows the area where the Rotary Selection knob cable can be easily damaged if flexed
several times or otherwise stressed.
2. Check cables from Power ControlBoard to MainBoard.
Ensure cables are properly seated, tight, and undamaged.
3. Replace Power Control Board.
4. Replace MainBoard.
2025653-048 Revision B Responder™ 2000 Page 21
SECTION 3: Troubleshooting
Bad Front Panel LEDs
Use this procedure to troubleshoot the Responder 2000 when the Front Panel LEDs are not displaying correctly. Normal
operation is indicated by:
At system boot, all LEDs will flash.
Green AC Power LED is lit when the unit is plugged in.
Yellow Battery Charging LED is lit when the battery is lit when the unit is plugged in and a discharged battery is inserted.
Red Service Required LED should be off.
Blue Manual button LED is lit.
1. Ensure the unit is plugged in and receiving power from the wall outlet.
2. If the battery charge LED is not lit when a battery is inserted, replace the battery with a known good battery.
3. Check cables from Power ControlBoard to MainBoard.
4. Ensure cables are properly seated, tight, and undamaged.
5. Replace Power ControlBoard.
6. Replace MainBoard.
Buttons do not Work
A. The Rotary Selection knob does not work properly
1. Replace Rotary Selection knob.
Note: The figure below shows the area where the Rotary Selection knob cable can be easily damaged if flexed
several times or otherwise stressed.
2. Check cables from Power ControlBoard to Main Board.
3. Ensure cables are properly seated, tight, and undamaged.
4. Replace Power Control Board.
5. Replace Main Board.
B. The Rotary Selection knob works properly
Perform Button Tests (refer to the Button Test procedure on page 54).
If Manual, Charge, Shock fails:
1. Check cables from Power Control Board to Main Board.
2. Ensure cables are properly seated, tight, and undamaged.
3. Replace Power Control Board.
4. Replace Main Board.
If Paddles, Charge, Shock fails:
1. Replace paddles.
2. Check cable from rear connector to Therapy Board.
3. Check cable between Therapy Board and Main Board.
4. Replace Therapy Board.
5. Replace Main Board.
2025653-048 Revision B Responder™ 2000 Page 22
SECTION 3: Troubleshooting
Bad Speaker
Use this procedure to troubleshoot the Responder 2000 when the Speaker is not working correctly.
Note: The Speaker cannot be removed from the Front Panel. If the speaker must be replaced, the entire Front Panel must
be replaced.
Normal operation is indicated by:
At system boot, the speaker sounds a brief tone (listen carefully because the tone is easy to miss in a noisy environment).
When audio indication is enabled in the Settings menu, the speaker should emit loud, non-distorted tones.
1. Check speaker plug and cable.
2. Check speaker resistance. It should be between 4 and 10 ohms.
3. Replace FrontPanel.
4. Replace MainBoard.
Display is Dim, Dark, Fuzzy, or Unreadable
Normal operation is indicated by:
At system boot, all LEDs will flash.
The speaker sounds a brief tone (listen carefully because the tone is easy to miss in a noisy environment).
Green AC Power LED is lit when the unit is plugged in.
Yellow Battery Charging LED is lit when the unit is plugged in and a partially or fully discharged battery is inserted.
Red Service Required LED should be off.
Blue Manual button LED is lit.
A. If the Display is Dim
1. Check the software setting for display brightness.
2. Check the backlight cable
3. Run display test from the Maintenance menu to verify the display is working and there are no bad pixels.
If these checks are ok, continue with the following steps.
B. If the Display is Dark, Fuzzy, or Unreadable
1. Check the display cable.
2. Replace the Display.
3. Replace the MainBoard.
Printing Problems
Note: Check the paper roll to ensure the customer is using approved paper listed in the Accessories list. Non-approved or
generic paper may cause printer damage or failure.
1. Ensure paper roll is in good condition (dry and undamaged).
2. Ensure paper is loaded correctly
3. If the Print icon is not selectable, ensure the printer door is fully closed and the paper is properly inserted.
4. Replace the printer door.
5. Check printer cables.
6. Replace printer.
7. Replace MainBoard.
After troubleshooting, always print a test strip (page 55) to ensure proper operation.
2025653-048 Revision B Responder™ 2000 Page 23
SECTION 3: Troubleshooting
Battery Troubleshooting
Check battery statistics. Good batteries have a capacity of at least 4500 mAH and will be charging when the Responder
2000 is plugged into an AC outlet.
1. Ensure battery is fully charged.
2. Perform Battery Calibration on page11.
3. Replace the Battery.
4. Replace Power ControlBoard.
5. Replace MainBoard.
SpO2 Not Working
Verify the Responder 2000 has a SpO2 Main Board installed.
Check the About screen. “SpO2 - - - -” indicates a non-SpO
1. Connect the Responder 2000 to a SpO
If the Responder 2000 SpO
display does not agree with the simulator setting, substitute a known good SpO2
2
simulator with the appropriate sensor.
2
MainBoard is installed in the unit.
2
sensor.
2. Check SPO
cable.
2
3. Check cable from Front Panel to SPO2 Board.
4. Replace SPO
Board.
2
5. Replace Main Board.
Shock into Paddle Dock Fails
If shocking into the paddle dock does not work or shows high impedance.
Perform a button test for each Paddle.
• If the test fails, replace the Paddle.
• If the test does not fail, connect the Paddles to a simulator and attempt to shock.
If the shock is successful:
1. Check the cables from the TherapyBoard to the paddle dock (one on each side).
2. Replace the TherapyBoard.
If the shock is unsuccessful:
1. Ensure the paddles are securely paced in the paddle dock.
2. Replace the Paddles.
3. Check the cables between the Rear Panel and the Therapy Board.
4. Check the cable between the Therapy Board and the Main Board.
5. Replace the Therapy Board.
6. Replace the Main Board.
2025653-048 Revision B Responder™ 2000 Page 24
SECTION 3: Troubleshooting
Shock into Patient Fails
1. Perform a button test for each Paddle.
If the test fails, replace the Paddle.
If the test does not fail, connect the Paddles/Pads/Spoons to a simulator and attempt to shock.
2. If the shock is successful:
The unit is working correctly. Check patient preparation.
3. If the shock is unsuccessful:
1. Replace the Paddles/Pads/Spoons.
2. Check the cables between the RearPanel and the Therapy Board.
3. Check the cable between the TherapyBoard and the Main Board.
4. Replace the TherapyBoard.
5. Replace the MainBoard.
No ECG from Paddles/Pads/Spoons
1. Replace the paddles/pads/spoons.
2. Check the cables between the RearPanel and the Therapy Board.
3. Check the cable between the TherapyBoard and the Main Board.
4. Replace the TherapyBoard.
5. Replace the MainBoard.
Service Required LED is ON
If the Service Required LED is on and the fault has been corrected, perform this procedure to clear the fault.
1. Go to Maintenance > Stored Data Management > Clear Service Required Indicator.
2. Press Accept.
3. Cycle power.
If the fault does not clear, go to History > Event Log to find the fault log entry.
Note: All service required Error codes are displayed in the History with a red exclamation mark.
2025653-048 Revision B Responder™ 2000 Page 25
SECTION 4: Repair
SECTION 4: Repair
Overview
This section describes how to assemble and disassemble the Responder 2000. These instructions are intended for use
only by service providers who are specifically trained to service the Responder 2000.
This section is divided into two parts:
How to Replace Specific Components: Provides high-level disassembly steps with references to the assembly steps.
Assembly: Lists production assembly steps applicable to field repair.
TOPIC PAGE
Required Tools 27
Disassembly Overview 27
How to Replace Specific Components 28
Assembly 31
2025653-048 Revision B Responder™ 2000 Page 26
SECTION 4: Repair
Required Tools
The following tools are necessary to Assembly/ Disassembly the Responder 2000.
Note: The Security Torx is also known as Tamper Resistant Torx.
In the event of equipment failure, the two main capacitors may retain dangerous voltages even if the Responder
2000 is disconnected from AC power and the battery is removed. Normally, the capacitors are discharged when
power is shut off; however, it is possible for equipment damage to prevent the capacitors from discharging
properly.
Always assume the capacitors are fully charged.
Caution: Shock Hazard or Equipment Damage.
Before opening the case, ensure the AC power cord is disconnected and the battery is removed.
Caution: Equipment Damage.
Always wear a grounding wrist strap and use an anti-static mat when handling parts.
Caution: Procedure Failure.
Even though some assembly steps may not be applicable to a specific replacement procedure, all other steps
must be performed in the order listed.
The Responder 2000 has two access points:
Rear Cover: Provides access to Capacitors, Therapy Board, optional SPO2 Board, and ECG Board, and Front Bezel.
Removing the Bezel allows access to the Main Board, Printer, and Display.
Base Cover: Provides access to the Power Supply, Power Control Board, Fan, and Switches.
Note: The Speaker is part of the Front Body Assembly.
2025653-048 Revision B Responder™ 2000 Page 27
SECTION 4: Repair
How to Replace Specific Components
Use the following procedures as a guide to replace specific internal components in the Responder 2000. Not all
components are listed. (For example, if the Front Body is damaged and must be replaced, the entire unit must be
disassembled and then reassembled using the more detailed Assembly instructions).
Capacitors (see page 46)
WARNING: Lethal Shock Hazard.
In the event of equipment failure, the two main capacitors may retain dangerous voltages even if the Responder
2000 is disconnected from AC power and the battery is removed. Normally, the capacitors are discharged when
power is shut off; however, it is possible for equipment damage to prevent the capacitors from discharging
properly.
Always assume the capacitors are fully charged.
1. Remove Rear Cover (see Figure 67: Rear Body Installed).
Note: Two cover screws are located under the Bed Hook Covers.
2. Cut tie wraps to free capacitor leads.
3. Disconnect Capacitor leads (see Figure 58: Capacitors Installed) from Therapy Board.
Caution: Equipment Damage.
Carefully note the capacitor lead connection points and double check the lead and jack labels.
Connecting the Capacitors to the wrong jack could cause equipment damage.
4. Reconnect Capacitors. Ensure lead connections are tight.
5. Tie wrap leads and cables. Refer to the assembly instructions for proper cable routing.
6. If necessary replace or reseat rubber tubing along the edge of the Front Body.
7. Replace the Rear Cover. Ensure the cover seats properly—check for gaps or bulges around the entire edge.
Note: One screw is located under the label of the Rear Cover. (See Figure 67: Rear Body Installed)
8. Replace Bed Hook Covers. Check for proper orientation (see Figure 68: Bed Hook Cover Improperly Installed).
Therapy Board (see page 45)
1. Follow the Capacitor instructions (above) to remove the Capacitors.
2. Disconnect all other leads from the Therapy Board (see Figure 56: Therapy Board Cables Installed).
3. Remove the screws holding the Therapy Board (see Figure 54: Therapy Board Installed).
4. Exchange the board and replace screws.
5. Reconnect all cables and leads.
Caution: Equipment Damage.
Carefully note the capacitor lead connection points and double check the lead and jack labels.
Connecting the Capacitors to the wrong jack could cause equipment damage.
6. Follow the Capacitor instructions (above) to complete the reassembly.
2025653-048 Revision B Responder™ 2000 Page 28
SECTION 4: Repair
SPO2 Board (see page 44)
Note: Most SP02 failures are the result of bad sensors. Before replacing the SPO2 Board, be sure to complete the
troubleshooting procedure on page 24 to verify the board is the cause of the fault.
1. Follow the Capacitor and Therapy Board instructions (above) to remove the Capacitors and Therapy Board.
2. Remove the screws holding the SPO2 Board (see Figure 51: SPO2 Board Installed).
3. Lift the SPO2 Board out to disconnect it from the Main Board.
4. Disconnect the SPO2 cable.
5. Exchange the board and reconnect the SPO2 cable.
6. Reconnect the SPO2 Board to the Main Board.
7. Replace the screws.
8. Follow the Therapy Board and Capacitor instructions (above) to replace the Therapy Board and Capacitors.
ECG Board (see page 40)
Note: Most ECG failures are the result of bad cables or sensors. Before replacing the ECG Board, be sure to complete
the troubleshooting procedure on page 25 to verify the board is the cause of the fault.
1. Follow the Capacitor and Therapy Board instructions (above) to remove the Capacitors and Therapy Board.
2. Remove the ECG Board EMI shield.
3. Disconnect the ECG cable (see Figure 41: ECG Cable Installation Detail).
4. Remove the screws holding the ECG Board.
5. Exchange the board and replace the screws.
6. Reconnect the ECG cable.
7. Replace the ECG Board EMI shield.
8. Follow the Therapy Board and Capacitor instructions (above) to replace the Therapy Board and Capacitors.
2025653-048 Revision B Responder™ 2000 Page 29
SECTION 4: Repair
Display/Main Board/Printer (see page 42)
1. Follow the Capacitor, Therapy Board, SPO2 Board, and ECG Board instructions (above) to remove the
Capacitors Therapy Board, SPO2 Board, and ECG Board.
2. Remove the screws holding the Front Bezel (see Figure 45: Front Bezel Installed).
3. Remove the screws holding the Display (see Figure 43: LCD Screw Locations).
At this point the Display can be replaced (disconnect Display cables).
If the Main Board or Printer must be replaced, perform the following steps:
Main Board
a. Remove the screws (see Figure 11: Main Board Installed).
b. Disconnect all cables
c. Exchange the Main Board.
d. Reconnect all cables.
e. Replace the screws.
Printer
a. Remove the screws (see Figure 14: Printer Screws).
b. Disconnect the ribbon and Power cables.
c. Exchange the Printer.
d. Reconnect ribbon and Power cables.
Note: Removing the printer bracket may help when connecting the Printer ribbon cable to the Main Board
(see Figure 10: Printer Bracket Removal).
e. Replace the screws.
4. Connect the Display cables.
5. Replace the Display screws.
6. Follow the ECG Board, SPO2 Board, Therapy Board, and Capacitor instructions (above) to replace the ECG
Board, SPO2 Board, Therapy Board, and Capacitor.
Power Supply and Power Control Board (see page 36)
Remove the Base Cover, disconnect all cables, and remove screws.
Note: The Power Supply is held in place only by the two larger Base Cover screws.
Switches and Fan (see page 36)
Remove Base Cover and Power Control Board.
2025653-048 Revision B Responder™ 2000 Page 30
Loading...
+ 108 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.