This document contains confidential or proprietary information of GE-Lunar Corp. neither the
document nor the information is to be reproduced, distributed, used or disclosed, either in whole or
in part, except as specifically authorized by GE-Lunar Corp.
GE-LUNAR Corporation makes no warranty of any kind with regard to this material, and shall not be
held liable for errors contained herein or for incidental or consequential damages in connection
with the furnishings or use of this manual.
Read through this manual thoroughly before attempting to service any components. Unauthorized
service may void system warranties or service contracts. Consult the GE-LUNAR Customer
Support Department prior to attempting any servicing:
608-828--2663
608-826-7107 (Fax)
is a registered trademark of General Electric Company.
g
PRODIGY® is a trademark of GE-LUNAR Corporation.
Windows NT® is a trademark of Microsoft Corporation
A person who will be performing service work on the PRODIGY should use this manual in the
following manner:
Read the Safety and Overview Chapters to familiarize yourself with the scanner as a whole and
with the general function of the circuit boards.
Chapter 3 should be understood completely as it explains the Diagnostics Software (built in –
requires a password for access).
The Chapter 4 and Chapter 5 contain common procedures and troubleshooting information and
can be read as needed, but are good sources of information.
When a problem arises, Chapter 4 should be referenced. Check the table of contents for Chapter 4
to see if the problem being experienced is described. If so, refer to the appropriate page. If not, try
to generalize the problem (e.g. the Detector is repeatedly running into the front of the scanner and
reversing and then running back into the front of the scanner. This is a mechanical problem in
general, specifically with Transverse Mechanics, check that subsection of Chapter 4 for the
subsystem experiencing the fault.
This manual commonly references other Sections and pages of the manual as needed, so often
procedures in the Chapter 5 Appendix are referred to as ways to solve problems described in
Chapter 4.
Information Requested By GE-LUNAR:
When requesting assistance from GE-LUNAR, please provide the following information:
System Number (DF+xxxxx)
Institution or Doctor's name
Location
Complete list of symptoms
Names and part numbers of parts needed for service
Prodigy Service Manualv
In addition to the information above, an Error Log printout and QA’s and QA history or other failing
diagnostic test printout (Alignment Test etc.) will also help to improve the accuracy of our
diagnosis:
For problems with specific patient scans, it is recommended that you copy the scan files to
diskette, and send it, with a description of the problem, to the Applications Department at GE LUNAR.
viProdigy Service Manual
11112
Safety
Chapter 1:Safety
This chapter highlights safety devices and features a Service
Engineer should know before servicing a PRODIGY system.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
1-14 SafetyPRODIGY Service Manual (Rev C - 2000)
1.0General Safety
• DO NOT attempt to service the PRODIGY without first reading this
manual.
• DO NOT attempt any repairs without prior instructions from authorized
LUNAR personnel.
• In order to maintain electrical safety and electromagnetic compatibility,
the Lunar PRODIGY is only to be connected to a computer, printer, and
peripherals that are certified to be compliant with IEC 950/EN 60950
Safety of information technology equipment, including electrical business
equipment and IEC 601-1-2 Medical electrical equipment, Part 1:
General requirements for safety, 2. Collateral Standard: Electromagnetic
compatibility - Requirements and tests.Emergency Stop Button
1.1Symbols and labels found on the PRODIGY
1.1.1 External Symbols
• The following symbols are found on the PRODIGY, in the Operators
manual, and in the Service Manual.
Attention: contains important safety information such as the
location of a pinch point.
Emergency Stop Button: shows the location of the
emergency stop button.
Power On: shows the location of the Power On indicator and the
switch position for Power On.
Power Off: shows the switch position for Power Off.
Laser On: shows the location of the Laser On indicator. It is found
only on systems delivered internationally.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)Safety1-15
Shutter Open: shows the location of the Shutter Open indicator.
X-ray On: shows the location of the X-Ray On indicator.
Type B Equipment: shows that the scanner has Type B
protection against electrical shock.
1.1.2 Internal Symbols
• The following symbols are found inside the PRODIGY, and in the Service
Manual.
Protective Earth: shows the location of a Protective Earth
Terminal.
Functional Earth: shows the location of a Functional Earth
Terminal.
1.1.3 Labels
• The following labels are found on the PRODIGY Scanner.
Laser Caution Label: Shows that the scanner uses
a Class II laser. This label is not found on systems
shipped to Canada.
Laser Caution Label: Canada only
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
1-16 SafetyPRODIGY Service Manual (Rev C - 2000)
Tube Head Assembly Label (system number
DF+12000 and greater): This label gives tube head
assembly and x-ray source characteristics
information. It is located on the tube head assembly
and the foot panel of the scanner.
Tube Head Assembly Label (system number
DF+11999 and lower): This label gives tube head
assembly and x-ray source characteristics
information. It is located on the tube head assembly
and the foot panel of the scanner.
A definition of each symbol on this label follows:
Inherent Filtration
Tube Insert
X-ray Source
Focal Point
System Label (system number DF+12000 and
greater): This label gives system input power
requirements and compliance information. It is
located on the foot panel of scanners. The CE
mark shows compliance with UL/CSA and the
Medical Device Directive 93/42/EEC.
System Label (system number DF+11999 and
lower): This label gives system input power
requirements and compliance information. It is
located on the foot panel of scanners. The CE
mark shows compliance with UL/CSA and the
Medical Device Directive 93/42/EEC.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)Safety1-17
High Voltage Power Supply: This label gives
high voltage power supply (x-ray generator)
information. It is located on the positive and
negative power supplies, and foot panel of the
scanner.
X-ray Controller: This label shows x-ray
controller compliance. It is located on the foot
panel of the scanner.
Collimator Assembly: This label gives
collimator assembly information. It is located on
the collimator and foot panel of the scanner.
Warning Label and Radiation Symbol: The
Warning label shows that the system uses
ionizing radiation. It is found only on systems
delivered in the United States. Always obey
instructions for safe operation.
Radiation Label: This label shows that the system uses
ionizing radiation.
Grounding Reliability Label: This label states that
grounding reliability can only be maintained when using a
“Hospital Grade” or “Hospital Only” receptacle. It is only found
on systems delivered in the United States.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
1-18 SafetyPRODIGY Service Manual (Rev C - 2000)
1.2Emergency Stop Button
• The Emergency Stop Button is a round red button located on the scanner
display panel (see Figure 1-1). When pressed, power is removed from
the X-ray tube head, the laser, and the shutter is closed. Power is also
removed from the scan arm motors, allowing the operator/patient to push
the scan arm out of the way.
Figure 1-1. The PRODIGY Display Panel
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)Safety1-19
1.3Laser Exposure
• The PRODIGY is equipped with a Class II Laser device. This laser is
used for patient positioning. A Class II rating indicates a low power visible
laser that is not normally hazardous to eyesight but has the potential to
be hazardous if viewed directly for an extended period of time. Because
of this potential hazard, DO NOT stare directly into the beam while the
laser is in operation, and DO NOT allow the beam to shine directly into
the patients' eyes. No specific eye protection is required with a Class II
laser.
• A amber laser-on indicator, located on the front of the scan arm, is lit
when the laser is on. The program activates the laser during positioning
for an image acquisition. The program then turns off the laser when you
begin the scan. The emergency stop button will turn off the laser.
• There is a caution label (Figure 1.2) on the scan arm near the Display
Panel.
Figure 1-2. Laser Warning Label (U.S. systems only)
Figure 1-3. Laser Warning Label (Canadian Systems only)
Figure 1-4. Laser Warning Label (International systems only)
Note:DO NOT STARE INTO THE BEAM while the laser is
operating.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
1-20 SafetyPRODIGY Service Manual (Rev C - 2000)
1.4Shutter Indicator
• This symbol is used to indicate an open-shutter condition in accordance
with the safety standards established by the International
Electrotechnical Commission (IEC).
Figure 1-5. Source (x-rays) off - Shutter closed (green)
Figure 1-6. Source (x-rays) on - Shutter open (yellow)
• This symbol appears near the yellow X-ray shutter-open indicator light.
The X-ray shutter-open indicator light is located on the Display Panel on
the scan arm near the front.
Note:When the x-ray on / shutter open symbol appears in literature
associated with the PRODIGY scanner, it will be used to
indicate that the procedure being described results in an openshutter condition. During these times personnel should
exercise caution to avoid excessive exposure to the X-rays.
1.5Cautions, Warnings, and Notes
• This manual contains warning and caution statements wherever
appropriate for your safety. The warnings and cautions used throughout
the manual are based on the safety standards established by the
International Electrotechnical Commission (IEC). In addition, the manual
uses notes to attract the reader's attention to important information.
1.5.1 Caution Statements
A caution statement reflects a condition
that, if not avoided, could cause
equipment or property damage.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)Safety1-21
1.5.2 Warning Statements
1.5.3 Note Statements
Note:This symbol turns the reader's attention to important
information which may otherwise be overlooked.
1.6Safety Concerns
1.6.1 Pinch points
The Warning label below identifies the location of possible pinch points
A warning statement reflects a potentially
hazardous condition that, if not avoided, could
result in serious injury.
Because the PRODIGY Densitometer
contains moving parts, there are places
on the scanner where there is a danger
of pinching. Operators should be aware
of these pinch points to avoid injury to
the patient or themselves. When the scanner arm is in
motion, make sure possible pinch point locations are
clear at all times. Labels applied at the LUNAR factory
indicate the location of the pinch points. The pinch
points and their labels are shown in the figure 1-7.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
1-22 SafetyPRODIGY Service Manual (Rev C - 2000)
Figure 1-7. Potential Pinch Points on the PRODIGY
1.6.2 Laser safety
Do not touch the AC Surge Suppressor
(located on the AC terminal block) it
may be hot.
Do not electrically connect the isolation
transformer bolt head to ground, doing
this will short out the transformer
Scan table isolated outlet strip must be
appropriately connected.
DO NOT STARE INTO THE LASER BEAM
during patient positioning and Quality
Assurance procedures.
The label that follows is located on the scanner arm and shows the location of
the laser aperture.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)Safety1-23
1.6.3 Radiation safety
X-ray exposure: The system makes radiation when electric voltage is
supplied to, and current flows through, the x-ray tube. During a measurement,
the shutter opens to let a beam of radiation pass through the scanner table
and patient. The radiation field at the table top is 19.2 mm x 3.3 mm. Lead
oxide shielding surrounds the x-ray tube insert inside the tube housing
assembly and reduces radiation levels around the scanner table.
Skin entrance dose: Table 1 of the PRODIGY Operators Safety and
Specifications manual shows irradiation times and skin entrance doses.
1.6.4 Scatter Radiation
• There is some scatter radiation from the PRODIGY when it is running.
Figure 1-8 shows the radiation dosages while the scanner is running at
3.00 mA at certain distances.
• These dosages are relatively insignificant as the allowed yearly dosage
for a person working with radiation emitting equipment is 5000 mRem.
Radiation however should be avoided when possible.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
1-24 SafetyPRODIGY Service Manual (Rev C - 2000)
Figure 1-8. PRODIGY Iso-Dose Diagram
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)Safety1-25
1.7Controlling Computer and Accessories
1.7.1 Electrical Safety
IEC and UL/CSA certification
IEC: To maintain electrical safety, all computer equipment and accessories
connected to the scanner must meet all IEC requirements for safety, such as
IEC 950, "Safety of information technology equipment, including electrical
business equipment," and IEC 801-5, "EMC Surge Immunity Requirements."
The computer and all accessories must have the CE label.
UL/CSA: To maintain electrical safety, all computer equipment and
accessories connected to the scanner must have safety agency approvals for
UL/CSA.
1.7.2 Electromagnetic interference
Although the scanner meets safety standards regarding electromagnetic
interference (EN60601-1-2), you may still experience a loss of performance
under extreme electromagnetic conditions. Maximize the distance between
the scanner and other equipment, and use a dedicated power line, to avoid
interference to and from the scanner.
1.7.3 Peripheral Configurations
The correct connection of the computer and all
peripherals is necessary to maintain electrical
safety. The signal cable of the scanner is
intended only for connection to an approved
computer. Call LUNAR Support or your LUNAR
distributor before adding peripherals.
Note:The scanner's output power strip can be used to supply the
Host PC with isolated power. If it is to be used the following
conditions must be met. If the conditions cannot be met, the
scanner's output power strip cannot be used.
Note:See also Peripheral Block Diagrams - Section 2.2.3
1.7.4 Standard room configuration (system no. DF+12000 and
greater)
The computer, peripherals, and all other equipment must be located more
than 1.83 m from the scanner. Use an outlet strip to power the computer and
all peripherals. The outlet strip must be mounted off the floor so that it does
not touch other equipment. If your outlet strip was provided by LUNAR, it has
a maximum output of 15A, 120VAC. Only system-related equipment should
be powered by the outlet strip.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
1-26 SafetyPRODIGY Service Manual (Rev C - 2000)
A modem and/or network connection can be made at any time if you are using
the standard room configuration.
1.7.5 Small room configuration (system no. DF+12000 and higher)
You must power the computer, peripherals, and all other equipment with an
isolating transformer if the room is too small to maintain at least 1.83 m of
separation between the scanner and all other equipment.
The isolation transformer supplied by LUNAR has a maximum output of 400/
500VA. Only system-related equipment should be powered by the isolation
transformer. Failure to use an isolation transformer can cause leakage
currents in excess of 100 microamperes.
A modem and/or network connection can only be made in the small room
configuration if all exposed metal surfaces of the computer and peripherals
are out of the patient environment.
1.7.6 Scanner power output configuration (system no. DF+11999
and lower)
LUNAR recommends that you use scanner power output to provide isolated
power to the computer and all peripherals. The power strip must be mounted
off the floor such that it does not touch other equipment. The computer and
ALL peripherals must be powered by the scanner. All other equipment must
not be powered by the scanner and must be located more than 1.83 m from
the scanner. Failure to use scanner power output can cause leakage currents
in excess of 100 microamperes.
If a network and/or modem connection is needed, refer to the wall outlet
configuration.
1.7.7 Wall outlet configuration (system no. DF+11999 and lower)
As an option to scanner power output, a wall outlet can be used to power the
computer and peripherals. Isolated power from the scanner must not be used
to power any equipment if a wall outlet is used. All exposed metal surfaces of
the computer, peripherals, and other equipment must be located more than
1.83 m from the scanner.
A network and/or modem connection can be made to the computer if power is
supplied from a wall outlet as described above.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)Safety1-27
This page left blank intentionally.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
1-28 SafetyPRODIGY Service Manual (Rev C - 2000)
21112
System Overview
Chapter 1:System Overview
This chapter provides an overview of the PRODIGY system.
• In addition the chapter contains a brief discussion of major subsystems and illustrations of the PRODIGY power system.
• This Chapter contains the PRODIGY Block Diagrams
Table 2-10. FDA certified components (system no. DF+12000 and higher).189
Table 2-11. FDA certified components (system no. DF+11999 and lower).1-90
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-32 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.0PRODIGY System
The PRODIGY includes the patient table and frame, X-ray tube, X-ray
generator, detector, and arm. Its physical specifications are summarized in
section 2.15.
The PRODIGY has a mechanical design with two separate motion systems
that are capable of simultaneous operation. These are transverse, and
longitudinal. Both motion systems are driven by stepper motors.
2.0.1 PRODIGY Electronics
The internal components of the scanner are safely secured by a number of
panels, including the scanner's tabletop.
Figure 2-9. PRODIGY System Exploded View of External Covers and
associated hardware
• The front and side panels are secured by screws from the inside.
• The rear panel is secured by screws from the outside.
• The table top is screwed down from the top.
Note:Primary Service access to the electronics of the scanner is
through the table top.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-33
• The Detector electronics (in the scan arm) are secured by an upper and
lower shroud, held in place by screws.
• Each metal panel is grounded to the electronics pan.
It is not usually necessary to remove the front and back panels for most
service needs. However, if access is needed to the Front and Rear
Longitudinal Carriages, these can be removed.
The back panel is secured by hex socket head-head screws and must be slid
out of the way, for it is between the Arm Column and the frame.
If access is needed to the detector, Transverse Limit Switches or the other
components mounted above in the arm, the covers of the arm must be
removed.
• The upper scan arm shroud can be removed by loosening the two
screws holding it in place (on the back of the arm column) and tipping it
forward.
• The lower cover is held in place by four screws, two in the front and two
in the back, be sure to remove the ground wire for the metal portion of
the lower cover as well
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-34 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.1Electronics
2.1.1 Cautions
PRODIGY I electronics use Negative Temperature Coefficient (NTC)
thermistors to limit the in rush current of the AC isolation transformer and the
Detector Mother Board. These devices have a high resistance when cold and
decrease in resistance when warm.
PRODIGY electronics are static sensitive,
take static control precautions before
servicing scanner circuitry.
Figure 2-10. PRODIGY I NTC thermistor on the Primary Terminal Block
• A cool down period of 30 seconds is required before power is turned on
on the system.
PRODIGY I ONLY: Failure to allow the
system to “cool down” may cause the
circuit breaker for the AC line to trip and /
or the + 12VDC for the Detector Mother
Board (DMB) will not come up to +12 VDC.
Note:The error log entry "DMB Power Cycle observed" is an
indicator that the DC power to the Detector Mother Board has
been interrupted and the system should be powered down for
30 seconds to restore power to the Detector Mother Board.
PRODIGY I ONLY: The NTC’s can get hot - they
are located on the AC terminal block and on the
DC terminal block be careful when servicing these
areas of the scanner.
2.1.2 Electronics Pan
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-35
The electronic components of the PRODIGY are mounted on the grounded
Electronics Pan which is horizontally fastened inside the frame.
Note:Appendix 2 A contains a layout drawing of the Electronics Pan
for PRODIGY I and PRODIGY II systems.
PRODIGY I Systems (System numbers DF+11999 and lower):
• There are four low-voltage linear DC power supplies (under 30VDC), and
two high-voltage DC power supplies (to supply 76kV to the x-ray tube) on
the pan.
• In addition to the power supplies, the electronics mounting chassis holds
four printed circuit boards, a stepper motor controller, the audible alarm,
an AC entrance/line filter/fuse holder, an isolation transformer and a
terminal strip for AC power distribution to the Host PC and peripherals.
PRODIGY II Systems (System numbers DF+12000 and greater):
• There is one low-voltage DC switching power supply (under 30VDC), and
one high-voltage DC power supply (x-ray generator - supplies 76kV to
the x-ray tube) on the pan.
• In addition to the power supplies, the electronics mounting chassis holds
one printed circuit board, a stepper motor controller, and an AC entrance/
line filter/fuse holder.
2.1.3 Scan Arm
• The scan arm contains one high-voltage power supply (1000VDC / +12
VDC input) is located in the upper arm near the X-ray detector and
provides power to the Detector Array.
• The scan arm also houses the detector and 5 associated printed circuit
boards, and a stepper motor contoller.
2.1.4 Power specifications
Leakage current
• Total System with Isolation Transformer: <100 microamperes.
• Scanner Table alone: <100 microamperes.
Scanner input power
• PRODIGY I ONLY: The scanner has 12 different nominal inputs: 100,
110, 115, 120, 125, 127, 200, 220, 230, 240, 250, and 254 VAC. During
installation, the scanner is configured for the nominal input which best
matches the voltage on site.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-36 System OverviewPRODIGY Service Manual (Rev C - 2000)
• PRODIGY II ONLY: The scanner can accept any AC input between 100
and 254 VAC.
• Voltage may fluctuate ±10% from the nominal value without a loss of
scanner performance.
• The nominal input (range of inputs) can be found on the system label.
• The rated power input is 1500 VA.
• The input power must meet IEEE 519-1992 for power quality and total
harmonic distortion (THD <5%).
Scanner output power (PRODIGY I system numbers DF+11999 and lower
only)
• The scanner has 3 different nominal outputs: 100, 120, 240 VAC.
• The nominal voltage output of the scanner is shown on the system label.
• The computer and all peripherals which use the scanner output power
must be rated for this voltage.
• The maximum power output is 400 VA.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-37
2.2PRODIGY Block Diagrams
The block diagrams for the PRODIGY system follow:
2.2.1 PRODIGY I (System numbers DF+11999 and lower) Power
Distribution Block Diagram (AC entrance)
Figure 2-11. PRODIGY I Power Distribution Block Diagram
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-38 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.2.2 PRODIGY System Block Diagram PRODIGY I System Block
Diagram
Figure 2-12. PRODIGY I (systems DF+11999 and lower) Block Diagram
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-39
2.2.3 PRODIGY I Peripheral Configuration Block Diagrams
Figure 2-13. PRODIGY I (Systems DF+11999 and lower) Peripheral
Configuration Block Diagrams
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-40 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.2.4 PRODIGY II System / Power Block Diagram
Figure 2-14. PRODIGY II System / Power Block Diagram
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-41
2.3PRODIGY Fusing
PRODIGY I (system numbers DF+11999 and lower) Fuses
FuseRatingType*
F1 (Computer AC)T 2.0 AL5x20MM
F2 (Computer AC)T 2.0 AL5x20MM
F3 (+5, +/-12 VDC
PS (1))
F4 (+5, +/-12 VDC
PS (2))
F5 (+26VDC PS)T 2.5 AL5x20MM
F6 (+28VDC PS)T 0.63 AL5x20MM
F7 (+5, +/-12VDC
PS (1))
F8 (+5, +/-12VDC
PS (2))
F9 (+26 VDC PS)T 2.5 AL5x20MM
F10 (+28 VDC PS)T 0.63 AL5x20MM
MAX PCB F1F 0.5 AL1/4x1 1/4 in.
All fuses are 250V, low breaking capacity (25A minimum)
Table 2-1. PRODIGY I (Systems DF+11999 and lower) Fuses
T 1.25 AL5x20MM
T 1.25 AL5x20MM
T 1.25 AL5x20MM
T 1.25 AL5x20MM
PRODIGY II (systems DF+12000 and greater) Fuses
FuseRatingType
Condor PS
F1
Condor PS
F2
COndor PS
F3
All fuses are 250V, low breaking capacity (25A minimum)
Table 2-2. PRODIGY II (systems DF+12000 and greater) Fuses
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-42 System OverviewPRODIGY Service Manual (Rev C - 2000)
F3.15 AH5x20 mm
F3.15 AH5x20 mm
F3.15 AH5x20 mm
2.4PRODIGY I (system numbers DF+11999 and
lower) Single Board Controller
2.4.1 SBC Functions
The microprocessor-based Single Board Controller (SBC) provides overall
operation and control of the scan table.
Control Functions
• Control of x-ray source power supplies, and shutter / collimator in a fail
safe manner.
• Provides control signals for two external stepper motor drives (Centent)
to scan patient in a fail safe manner and senses limit switch actuation at
the limits of travel (via FOINK see 2.11).
• Controls the detector array (communicates with the Detector Mother
Board, section 2.10.2).
• Responds to external scanner failure signals (interrupts generated by
SBC, FOINK see 2.11, and Detector Mother Board see 2.10).
Communication
• Communicates with PC host via an optically isolated RS-422 interface.
• Collects data from the detector array.
2.4.2 SBC Reset
All SBC circuitry resets when the microprocessor resets. This is done during
power up, and can also be done over the communication port (via the host),
through connections to other circuit boards (especially the Detector Mother
Board, as a fail-safe shutdown), the by pressing the reset button on the SBC,
by pressing the Emergency Stop Button on the arm, or when a fault is
detected by the SBC or the FOINK (section 2.11.1).
Error Conditions Sensed by the SBC:
• Loss of Communications with the Detector Mother Board (DMB).
• Loss of Communications with the Host PC.
• Limit Switch tripped when shutter is open.
• Limit Switch depressed when commands are being sent.
• Failure in x-ray source kV programming
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-43
• Interrupt sent by FOINK
The DMB will reset the SBC if it detects an error.
Error Conditions of the DMB sensed by the SBC:
• PLD programming failure
• Loss of +5 or +12 VDC
• Loss of Communications with SBC
• Corrupt FIRMWARE
•FLASH RAM Failure
• DTR Reset button on the DMB is pressed
2.4.3 SBC / Host PC Interface
The SBC communicates with the Host PC via an RS-422 interface. This is a
serial connection capable of transmitting more data than a standard RS-232
port. As it is not a standard serial port a RS-485 card must be installed into
one of the Host PCs expansion slots and the port must be configured correctly
for the PRODIGY to operate correctly (see DXPC2000 chapter 5 appendices).
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-44 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.5PRODIGY II Combined Single Board Controller
(cSBC) (systems DF+12000 and greater)
The cSBC printed wiring board (PWB) is an eight layer rectangular board
measuring 7.400" x 8.100". The PWB is mounted in the via four mounting
holes located 1/4" from each corner and 2 additional interior mounting holes.
The components are primarily surface mount, with board connectors, headers
and a few single-style ICs being the only exceptions.
Four layers were dedicated for routing layers. Sensitive signals were noted
and routed manually and isolated from more powerful signals to reduce signal
interference and crosstalk on the same routing layer. The main power and
ground planes were stacked adjacently on the central inner layers to increase
inter-planar capacitance thus reducing ground bounce and power supply
noise. Traces on the top and bottom layers were kept as short as reasonably
possible and tapped down to an internal trace layer through vias.
Component placement is arranged to separate analog from digital circuitry.
Further isolation was achieved by segmenting the power and ground planes
into analog and digital sections and denying analog/digital plane overlap, thus
preventing digital noise from coupling into the analog section. All scanner
control I/O is run via connectors located on the +24V plane section. The +24V
plane is fully optically isolated from both the analog and digital plane areas to
prevent motor noise from coupling into the analog section, to prevent DC
switching noise from radiating on scanner cables, and to prevent ESD
presented at cable inputs from reaching the digital IC's.
2.5.1 cSBC System Architecture
The cSBC employs an Intel 80C251 micro-controller as its processor. This
processor provides 1K of on-board RAM and no on-board ROM. The
controller is clocked at 16 MHz using a crystal.
cSBC Memory Space
The cSBC is designed to support a JEDEC-standard, non-volatile FLASH
memory device up to 512K x 8 bits in size for code and fixed data. The board
supports either 128K or 512K SRAM memory device as needed for program
volatile memory. Complete address decoding is provided via the MAX PLD,
the CBSC bus master, allowing the address space to be arbitrary and
changed via the PLD code. The 80251 can address four 64K segments,
referred to as 0x00, 0x01, 0xFE and 0xFF as per Intel literature. The firmware
has the ability to map any FLASH or SRAM segment to any CPU segment via
SFR’s in the MAX PLD.
FLASH RAM
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-45
At startup the CPU executes the boots code which programs the FLEX PLD
and then maps in either NT or Prodigy runtime firmware as appropriate based
on the most significant bit of the CCA REV register. To switch from boot code
to run code the firmware jumps to SRAM and executes a code snippet which
pages the boot code out of 0xFF and the desired firmware into 0xFF. The
snippet then jumps from SRAM back to 0xFF to execute the firmware. A boot
jumper, JP4 is provided to optionally force the CPU to remain in the boot
code. When the boot jumper is installed the boot code runs the host port at
115.2KB.
SRAM
The cSBC contains a single 128K x 8 bit SRAM which provides read/write
memory. The SRAM's segments are arbitrarily mapped to any CPU segment
by the CPU mapping registers.
2.5.2 cSBC Functions
The microprocessor-based Single Board Controller (cSBC) provides overall
operation and control of the scan table.
FLEX PLD Peripherals
The majority of scanner related programmable logic functionality is contained
in the FLEX PLD, an Altera EPF6024AQC208-3 device. Device programming
is handled exclusively by the CPU. On each cold boot the CPU reprograms
the FLEX devices from an image stored in it's FLASH. As such a firmware
download of a new FLEX image is required to permanently upgrade the PLD
code.
The functional components of the programmable logic are discussed in the
following subsections. Polarity of operation can be inferred from bit names
and use of preceding slash for inverted logic bits.
Note that ports A-F are reloaded with default values at time of CPU reset and
remain in the default state until SCANNER_RESET has been cleared and
new values are written by the firmware. Defaults for port F and all other
registers are invoked at power up only.
PORT A
BitNameR/
W
0trans_enable R/W0Transverse motor enable – low blocks trans motor pulses and
1/trans_fwd R/W0Transverse motor direction control.
R/W0Switches mA DAC from 2.048V to 0.500V reference.
Def.Description
Def.Description
trans_motor_accel
1
low_range_adc
2
long_motor_accel
3
hvps_vendor_id
4
iq_hvps
5
/hvps_enable_status
6
/power_up
7
R/W0Enables motor interrupt on every micro step.
R/W0Switches ADC from 5.000V to 0.500V reference.
R/W0Enables motor interrupt on every micro step.
RN/AFor 7681 supply, 0 = Spellman, 1 = Bertan.
RN/ASet by resistor placement to indicate 0311/0312 supplies.
RN/AEnable status monitor from 7681 supply.
RN/ASet to indicate cold boot.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-49
PORT F
BitNameR/
W
/motion_fail_enable
0
long_motor_fail_axis
1
/hvps_enable
2
flex_diag_2
3
/arm_estop_sense
4
spare_jmp_[1]
5
spare_jmp_[0]
6
cpu_p1_2
7
R/W1Arm scanner shutdown if OMI pulses w/o step pulses.
R/W0Motor fail circuitry axis control, clear for transverse.
R/W1Enable output to 7681 supply.
R/W0Firmware controlled diagnostic LED.
RN/AEmergency stop sense bit.
RN/AUnused input, resistor or jumper selectable on CCA.
RN/AUnused input, resistor or jumper selectable on CCA.
RN/AFirmware controlled diagnostic LED.
PORT G
BitNameR/
W
adc_mux_[0]
0
R/W0ADC analog MUX input selection control bit.
Def.Description
Def.Description
adc_mux_[1]
1
adc_mux_[2]
2
adc_mux_[3]
3
adc_mux_enable
4
8ms_clock
5
unused
6
unused
7
R/W0ADC analog MUX input selection control bit.
R/W0ADC analog MUX input selection control bit.
R/W0ADC analog MUX input selection control bit.
R/W0ADC MUX output enable control.
R/W0Clock output provided to MAX PLD.
N/AN/AFor expansion.
N/AN/AFor expansion.
2.5.3 TRANS / LONG MOTOR Control and Status
Dual axis stepper motor control is provided entirely by the FLEX PLD. To
make a typical move the firmware loads a starting velocity into the 16 bit
VELOCITY register, the total number of steps for the move into the 16 bit
TARGET register, and step at which to next interrupt the CPU into the 16 bit
STEP register. Velocity is in terms of periods of the 2.0MHz fundamental clock
per micro step pulse to the stepper drive. The drives provide 10 micro steps
per full step. The firmware can track move status by reading the 16 bit READ
register.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-50 System OverviewPRODIGY Service Manual (Rev C - 2000)
As part of the setup for a move the host and/or firmware must enable the
motors via the /motor_power, trans_enable, and long_enable outputs and
setup the trans_lsw_override, long_lsw_override, /motion_fail_enable, /
motor_fail_enable, and long_motor_fail_axis outputs as desired. If the system
is in scanner reset for any reason the FLEX PLD will over-ride the /
motor_power output and prevent 24V power from reaching the motor drives.
Addressing for the motor control interface is provided below.
2.5.4 AGS ROLL
This is a read only 8 bit register which returns the count of AGS roll-over
events since the previous read of the register. The AGS roll counter is reset
on read only - it is not tied to the PIT's sample clock.
2.5.5 AGS DAC
This port provides R/W access to the AGS circuit's 8 bit U/D counter. The
counter is tied via a dedicated 8 bit bus to the AGS DAC. The DAC's analog
voltage is tied to the gain control input of the variable gain amplifier (VGA)
used to control gain of the detector input signal. As such the firmware can
read this counter to determine the current DAC voltage level and hence gain
level. If ags_enable is low this port gives the firmware direct control of the
AGS DAC as a parallel R/W device. If ags_enable is high, the firmware can
write to the port but the DAC will continue to respond to UP/DOWN requests
from the AGS DCA circuitry and hence quickly return to the AGS current
operating voltage.
HE/LE COUNTERS
These read only ports provide access to the 16 bit event counters which are
incremented each time the DCA circuitry detects an input pulse within the HE
or LE windows (as defined by the LEL, LEH, HEL, and HEH DAC settings).
These counters are read in two 8 bit bus cycles, MSB then LSB. The event
counters themselves consist of a counting element and a bus element. On the
rising edge on the PIT output pulse the counting elements are latched to the
bus element. The PIT output is also tied to CPU external INT 1 and as such
the firmware interrupt handler then has until the next rising PIT edge to read
the counters before the bus elements are latched over with the next sample
count and data is lost.
PIT MSB/LSB
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-51
The PLD provides a programmable interval timer (PIT) to the CPU. The PLD
prescales it's input clock to generate a PIT base clock of 100KHz. The CPU
writes a 2 byte word to the PIT reload register MSB/LSB. The CPU then raises
the pit_enable bit to start the timer. In response to the rising edge on the
pit_enable, the PIT loads the reload word into it's counting element and
begins counting down. When the count rolls under the PIT asserts /
SAMPLE_INT, reloads the counting element, and begins another count down
sequence. The /SAMPLE_INT line is tied internally to the HE and LE event
counters and causes a synchronous latch of both counting elements. The /
SAMPLE_INT line is also tied to the processor's 2nd external interrupt line, /
INT1. The CPU interrupt handler reads the latched event counters and ships
the data to the host.
2.5.6 SCANNER RESET
The scanner reset register is used to provide failsafe shutdown operation of
the scanner. A falling edge on any of the inputs to this register will latch the
current value of the register and drop the /SCAN_FAIL_ANY output. The MAX
PLD latches the master reset register and raises CPU_RESET in response to
the falling edge on /SCAN_FAIL_ANY. The MAX PLD also provides
SCANNER_RESET as the logical OR of CPU_RESET and!/
SCAN_FAIL_ANY. The FLEX PLD uses it's SCANNER_RESET input as the
enable bit to the tri-state buffers used to drive all safety critical output lines
including shutter control, HVPS relay control, motor relay control, etc. As such
the scanner is locked into a fail-safe mode whenever SCANNER_RESET is
asserted.
The MAX's master reset register will remain latched until the next rising edge
on the HOST_RTS input. When the cSBC is latched into reset by a scanner
error it will remain in CPU reset until the host drops the RTS line and reasserts it. It will remain in scanner reset until the CPU reads the scanner reset
register following the next raising edge of the RTS line at which the condition
causing the /SCAN_FAIL_ANY has been cleared. The firmware passes the
value of the reset registers to the host to allowing the host to display
appropriate error messages to the operator. The host will be unable to
perform any scanner related operations until the SCANNER_RESET has
been cleared. Red diagnostic LED's (see 2.5.33) are provided for both
scanner and CPU reset lines. The CPU reset line is tied to the host CTS
output such that the host sees a CTS event when the cSBC enters CPU reset.
The host code provides a CTS event handler which reads the reset registers
and prompts the user accordingly.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-52 System OverviewPRODIGY Service Manual (Rev C - 2000)
The firmware can also initiate a reset sequence in response to fatal error
conditions by writing a 'death code' to the suicide reset register. Resets can
also be initiated by the manual push button on the cSBC and by a low 5VDC
power condition as sensed by the MAX705 supervisor. The scanner register is
also latched at the end of read cycles such that current status can be
ascertained by a double read. A bit map of the scanner reset register is
provided below
BitNameR/
W
0/
thermostat_open_sense
1/external_estop_senseRN/AEmergency stop input from external options block.
2/dc_power_failRN/ALoss of one or more of +5VDC,+12VDC,-12VDC, or +24VDC.
3/long_motor_failRN/AMotor failure detected on longitudinal axis.
4/trans_motor_failRN/AMotor failure detected on transverse axis.
5/dmb_errorRN/ADMB dropped it’s CTS indicating a DMB reset event.
6/motion_failRN/AOMI pulses detected without step pulse (manual arm motion).
The HVPS error register is used to monitor the status of the 7681 X-ray
source HVPS. If the register value is not equal to 0xF when /hvps_enable is
low, the FLEX will raise the HVPS_ERROR_INT output to the MAX PLD. The
MAX PLD latches this into the IIR register and issues an interrupt to the CPU.
As such status of the HVPS is monitored when the unit is enabled. The
handler for HVPS interrupt reads this register to determine the cause of the
interrupt. The HVPS register is also latched at the end of read cycles such
that current status can be ascertained by a double read. A bit map of the
register is provided below.
BitNameR/
W
0/hvps_error_0RN/AError code bit from 7681 supply.
1/hvps_error_1RN/AError code bit from 7681 supply.
2/hvps_error_2RN/AError code bit from 7681 supply.
3hvps_enable_statusRN/ASet when /hvps_enable == /hvps_eanble_status
4UnusedN/AN/AExpansion room.
5UnusedN/AN/AExpansion room.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
Def.Description
PRODIGY Service Manual (Rev C - 2000)System Overview 2-53
6UnusedN/AN/AExpansion room.
7UnusedN/AN/AExpansion room.
2.5.8 DC FAIL
The DC fail error register latches the status of the DC power monitors at the
time of reset. If scanner reset code indicates /dc_power_fail the firmware can
read this register to identify the specific DC source failure. The register is also
latched at the end of read cycles such that current status can be ascertained
by a double read.
BitNameR/
W
0/plus_scanner_failRN/ALoss of +24V power input.
1/plus_analog_failRN/ALoss of +12V power input.
2/minus_analog_failRN/ALoss of -12V power input.
3UnusedN/AN/AExpansion room.
4UnusedN/AN/AExpansion room.
5UnusedN/AN/AExpansion room.
6UnusedN/AN/AExpansion room.
7UnusedN/AN/AExpansion room.
Def.Description
2.5.9 DCA / AGS / BIAS DAC's
The cSBC uses a single 10 bit octal DAC, the Linear Technology LTC1660, to
generate the AGS and DAC window reference voltages and the bias program
voltage. The DAC utilizes a serial interface and as such is not accessed with a
traditional CPU write cycle. To load the device the firmware writes to the DAC
address listed above, in response to which the MAX drops the /CS line to the
device. The firmware then manipulates the local serial bus clock and data
lines to load the DAC setting. The desired DAC channel address in encoded
into the first 4 bits of the output data word. A read is then made to the DAC
address, in response to which the MAX raises the /CS line. See device data
sheet and analog section below for further DAC details.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-54 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.5.10 KV/mA DAC
The cSBC uses a single 12 bit dual DAC, the Linear Technology LTC1454, to
generate the HVPS kV and mA program voltages. The DAC utilizes a serial
interface and as such is not accessed with a traditional CPU write cycle. To
load the device the firmware writes to the DAC address listed above, in
response to which the MAX drops the /CS line to the device. The firmware
then manipulates the local serial bus clock and data lines to load the DAC
setting. Both DAC channels must be written together, CHA (kV) first followed
by CHB (mA) in a 24 bit stream packet. A read is then made to the DAC
address, in response to which the MAX raises the /CS line. See device data
sheet and analog section below for further DAC details.
2.5.11 ARC/FIL DAC
The cSBC uses a single 10 bit dual DAC, the Linear Technology LTC1661, to
generate the HVPS filament limit and arc detect threshold voltages. The DAC
utilizes a serial interface and as such is not accessed with a traditional CPU
write cycle. To load the device the firmware writes to the DAC address listed
above, in response to which the MAX drops the /CS line to the device. The
firmware then manipulates the local serial bus clock and data lines to load the
DAC setting. The desired DAC channel address in encoded into the first 4 bits
of the output data word. A read is then made to the DAC address, in response
to which the MAX raises the /CS line. See device data sheet and analog
section below for further DAC details.
2.5.12 PEAK DAC
The cSBC uses a 12 bit DAC, the Linear Technology LTC8043, to generate
the detector peak gain voltage. The DAC utilizes a serial interface and as
such is not accessed with a traditional CPU write cycle. To load the device the
firmware manipulates the local serial bus clock and data lines to output load
the DAC setting and then performs a write/read cycle to the DAC address
listed above to pulse the DAC's load line low. See device data sheet and
analog section below for further DAC details.
2.5.13 MAX PLD Peripherals
The programmable logic section is based on an Altera MAX
EPM7128STC100-15 device. The configuration pins for this device are taken
to a JTAG style 10 pin header to allow for in-circuit programming of the device
from the Altera Byte-Blaster. The MAX device is FLASH based (non-volatile)
and is programmed at the time of CCA assembly. The functional components
of the programmable logic are discussed in the following subsections.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-55
2.5.14 Interrupts
The CPU's interrupt capacity is effectively increased by running several
interrupt signals to a register in the MAX PLD and tying the register output to
the CPU external INT 0 input. The firmware interrupt handler for INT 0 then
reads this register to identify the source of the interrupt and handles it
accordingly. The firmware then writes a bit masked '1' back to the IIR to clear
the bit of the interrupt it has serviced (the R/C in the table stands for READ/
CLEAR).
BitNameR/
W
0HOST_UART_INTR/CN/AHost UART interrupt.
1DEBUG_UART_INTR/CN/ADebug UART interrupt.
2DMB_UART_INTR/CN/ADMB UART interrupt.
3HVPS_ERROR_INTR/CN/AHVPS error interrupt.
48mS_CLOCKR/CN/A8ms clock tick interrupt from FLEX PLD.
5POWER_FAIL_INTR/CN/APower down pending in 5ms interrupt from DC supply.
6UnusedN/AN/AExpansion room.
7UnusedN/AN/AExpansion room.
Def.Description
2.5.15 MASTER RESET
The master reset register will force a CPU and scanner reset condition on the
falling edge of any of its listed inputs. The contents of the register will be
latched at the time of reset such that when the CPU next comes out of reset
the firmware can read the register to determine what caused the preceding
reset and report the appropriate code to the host. If the reset was cause by
the CPU_RST_WR input, the suicide reset register contains the specific error
code. If the reset was cause by the /SCAN_FAIL_ANY input, the scanner
reset register contains the specific error code.
The CPU and scanner resets will remain latched until the next rising edge of
the RTS input. At this time the CPU reset will be cleared if /POWER_RESET
bit is not asserted and the scanner reset will be cleared if /SCAN_FAIL_ANY
is high.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-56 System OverviewPRODIGY Service Manual (Rev C - 2000)
1/HOST_RTSRN/ARTS reset request from host via comm line.
2/MANUAL_RESETRN/APush button pressed.
3UnusedRN/AExpansion room, reads as ‘1’.
4UnusedRN/AExpansion room, reads as ‘1’.
5UnusedRN/AExpansion room, reads as ‘1’.
6CPU_RST_WRRN/AWrite to the suicide register, read suicide reg for error code.
7/SCAN_FAIL_ANYRN/AScanner reset register latched, read scanner reg for error code.
Def.Description
2.5.16 SUICIDE RESET
The CPU Reset SFR is a byte register into which the CPU can write a failure
code. In response to the write the MAX PLD will store the failure code and
assert the CPU_RESET line. The CPU_RESET line will be released on the
next rising edge of the host RTS, at which time the CPU will be able to read
the bit code from this SFR to determine the cause of the previous reset.
2.5.17 MISC OUT
The misc. output register is used to control the misc. output functions listed in
the following table.
2DMB HWPTR/W0Enable direct connect of host and DMB XCVR’s, bypassing
3RESET OVERRIDER/W1Enable override of CPU_RESET signal. Set to 1 on power-up such
4UnusedN/AN/AExpansion room.
5UnusedN/AN/AExpansion room.
6UnusedN/AN/AExpansion room.
7UnusedN/AN/AExpansion room.
Def.Description
UART’s, for maximized scan data bandwidth DMB to host.
that firmware can load the FLEX PLD at power-up regardless of
the host RTS state.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-57
2.5.18 MISC IN
The misc. output register is used to control the misc input functions listed in
the following table
BitNameR/
W
0BOOT JUMPERRN/AJP4, placed to force firmware to remain in boot code.
1CPU_P1_2RN/AInput from CPU port 1, pin 2 (diagnostic use only).
2UnusedN/AN/AExpansion room.
3UnusedN/AN/AExpansion room.
4UnusedN/AN/AExpansion room.
5UnusedN/AN/AExpansion room.
6UnusedN/AN/AExpansion room.
7UnusedN/AN/AExpansion room.
Def.Description
2.5.19 Stepper Motor Control
The stepper motors use the same interface design as used on IQ and Prodigy.
The FET exhibits a lower voltage drop and hence provides more available
power at the drives for a given source voltage - essentially recovering the
voltage lost in dropping the DC supply from the 26V used on IQ/Prodigy to the
24V which is readily available in commercial switchers. Diodes are placed in
series on the Centent power lines to prevent back EMF generated when the
arm is moved manually from reaching the 24V planes and damaging the
cSBC. A dual FET circuit is used to minimize the voltage drop on the 24V
plane due to in-rush current when the motors are enabled. A circuit in the
FLEX implements a 2 stage turn on when the firmware lowers /motor_enable.
Q15 is enabled first to provide 24V to the Centents through R381 and R384,
thus limiting the in-rush current. After the current pulse has stabilized Q14 is
enabled to bypass the inrush limiting resistors. Q15 is then disabled and the
motors are at full power. An PTC is provided on the source of Q15 such that
the current limiting resistors will not be smoked if Q15 is inadvertently left
enabled.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-58 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.5.20 OMI Input
The optical motion interrupt (OMI) sub-system connects to the transverse and
longitudinal OMI CCA's. The OMI CCA's are located on the far end (the gear
end farthest from the motor) of each drive axis. When the axis is in motion a
small toothed wheel spins through the OMI opto's beam and pulses are
sensed back on the cSBC. In this manner the system can sense a drive
circuit, motor, or belt failure which might otherwise result is a concentrated
exposure point during a patient scan. The OMI inputs are similar to those
used on the DPX-IQ. 74LS14's are added for hysteresis which provides a
clean direct interface to the PLD. As on Prodigy, logic in the FLEX PLD is
used to qualify the CH A and CH B inputs into a single 'valid motion' output.
Based on the phases of the square wave inputs on CH's A&B, the FLEX is
able to sense a change in direction. The circuit provides hysteresis to reject
false motion inputs resulting from scanner vibration when a wheel edge stops
in the center of the opto beam at the end of a move.
2.5.21 Patient Positioners
Four optically isolated inputs are provided for patient positioning. These are
used by the firmware to implement a joystick mode which is used in
conjunction with the laser to position the X-ray beam as desired over the
patient immediately prior to a scan. 74LS14 inverters are used to provide
hysteresis and isolate the FLEX PLD inputs from the slow rise times of the
opto output signals.
2.5.22 Limit Switches
Four optically isolated inputs are provided for limit switches. These are used
by the firmware to define the transverse and longitudinal table limits. 74LS14
inverters are used to provide hysteresis and isolate the FLEX PLD inputs from
the slow rise times of the opto output signals.
2.5.23 X-ray Source Control / Mechanical Interlocks
The cSBC is designed such that a high on the scanner reset net disables all
scanner functionality and assures a know, fail safe, state. All scanner control
outputs are driven by the FLEX PLD. The FLEX outputs drive the low side of a
PS2501 opto coupler emitter. The far side of all these opto circuits are
configured such that the opto has to be energized for the scanner function to
be active. As such the FLEX has to sink current to energize the opto and
activate the desired scanner function. Internally the FLEX code defines these
outputs as tri-state buffers, each of which has their enable line tied to the
scanner reset net. As such a scanner reset will force all these FLEX scanner
outputs to a high impedance state, de-energizing the opto's and disabling the
scanner.
The FLEX device is SRAM based and hence must be reprogrammed by the
CPU at power up. When the device is not programmed all I/O pins default to
the high impedance state. As such the scanner will also be in a fail safe state
when the FLEX is not programmed. The CPU port 1, pin 2 also runs directly to
the FLEX's ENABLE pin.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-59
The cSBC provide a failsafe mechanism independent of the programmable
logic via the +5V_IO circuit. If one or more of either the E-stop, external Estop, or tube thermostat is open, the FET driving the +5V_IO net from the
+5VDC plane will be disabled. The +5V_IO net provides power to the emitter
anode of all opto's which drive critical scanner functions. As such the scanner
will enter a failsafe state in response to these mechanical interlocks, even in
the event of a PLD device failure. The +5V_IO FET will also be disabled by
either a HOST_RTS or CPU_RESET.
2.5.24 Shutter / Collimator Drive
The shutter and collimator solenoid drive circuits are the same as that used on
Prodigy I. The FLEX drives an isolation opto which in turn switches a pair of
FET's to control solenoid current. The first FET is used for an initial 'hard hit'
on open commands. It presents 24V directly to the solenoid for several
hundred msec's, resulting in a large initial current pulse to the solenoid. The
second FET provides the 'hold' current through a pair of current limiting power
resistors. The hold FET is tied directly to the /shutter_open_ctrl bit. Note that
the collimator drive is populated only for NT-A and IQ upgrades which use the
old style IQ collimator assembly. The nominal values of the power resistors is
changed between the NT and Prodigy II BOM's to support both the traditional
linear solenoid of IQ, and the rotary solenoid of Prodigy. A jumper or DNP'd
resistor is used to drive the IQ_SHUTTER line to the FLEX such that both
mechanical and optical shutter limit switches are supported.
2.5.25 End of Exposure Alarm
An on-board end of exposure alarm is provided. The alarm chosen is the
board mount equivalent of that used on Prodigy I.
2.5.26 Panel LED's
The 4 panel LED's, power on, X-ray on, source exposed, and laser on, are all
driven PS2501-2 opto's through 750R0 / 1W current limiting resistors.
2.5.27 HVPS Control
The HVPS analog control interface is designed to provide maximum
performance at minimum cost. A single, 16 channel multiplexed, 16 bit, high
accuracy, ADC is used in conjunction with several lower cost, lower bit
resolution DAC's. Absolute accuracy of DAC's is poor but errors are calibrated
out by the firmware which monitors the actual DAC output via the ADC. In this
scenario DAC integral non-linearity (INL) specs are not important, sufficient
differential non-linearity (DNL) is all that is required. Serial DAC's and ADC's
are chosen to conserve board space and simplify routing. Serial parts are
typically also lower cost as their maximum bandwidth is limited by the serial
baud rate.
The HVPS AC is enabled via a relay controlled by the cSBC. The line is
primarially used to disable the HVPS by removing the AC to the HVPS via the
relay. AC Power to the HVPS can be left on for up to one hour after
generation of x-rays to prevent AC cycling between concurrent scans.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-60 System OverviewPRODIGY Service Manual (Rev C - 2000)
A jumper and/or DNP'd resistor pad is provide to drive the IQ_HVPS line,
which the firmware reads to determine which HVPS it is intended to operate.
The PWB provides lemo style connector pads for use with the traditional 0311/
0312 HVPS and MAX CCA. For the 7681 HVPS the lemos are DNP'd and a
single DB-25 connector is used to control the HVPS. Tranzorb pads are
provide on the PWB and are expected to be populated only if deemed
necessary by EMC testing. An opto bank is also provided to support the digital
interface to the 7681. See Lunar dwg 7681-SPC for details.
2.5.28 ADC
The Burr Brown ADS8320 16 bit, serial, single channel ADC was chosen for
the cSBC for its high resolution, and excellent accuracy.
An Analog devices AD586 +5V voltage reference is used. The part was
selected for it low noise and high accuracy
Overall the ADC and reference give the cSBC +/-5mV analog accuracy 11.2
DAC's
A single LTC1454 12 bit, serial, dual channel, DAC is used to provide the kV
and mA program voltages. The part was chosen for it low cost, ready
availability at national distributors, and excellent DNL specifications. The DAC
is used in the x2 configuration such that the full scale output is twice the
reference voltage. Voltage outputs feed back to the ADC MUX such that
firmware can calibrate out DAC INL errors.
A single LTC1661 10 bit, serial, dual channel, DAC is used to provide the arc
threshold and filament current limit input voltages to the new 7681 HVPS. The
part was chosen for it low cost, ready availability at national distributors, and
reasonable DNL specifications. Voltage outputs are not fed back to the ADC
MUX as high accuracy is not required on these threshold inputs.
2.5.29 mA Low Range
To support future scan modes, an mA low range circuit is provided. The FLEX
PLD provides a control bit by which the firmware can switch the mA DAC
reference voltage from 2.048V to 0.5V, hence decreasing the LSB size, hence
allowing the firmware to take smaller voltage steps when ramping to low uA
settings. A second control bit is provided to switch the ADC from 5.0 to 0.5V
reference.
2.5.30 Detector Interface
The cSBC provides a RS-422 communications post to the Detector Mother
Board. The CSBC does not participate in any Detector data manipulation, all
analysis and Detector control is performed at the DMB.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-61
2.5.31 Communications Ports
Host RS-422 Port
The host I/O port provides optical isolation per medical leakage requirements
of EN 60601-1-1, Annex BBB, section 7. QT's 6N136 opto's are used to
provide the required 115.2KB operational bandwidth and the required 2500
Vrms standoff at a reasonable price. Opto inputs and outputs are routed
through the MAX PLD to support the hardware pass though mode to the DMB.
A DB-9 female connector is provided on the host side of the isolation barrier.
RS-232 is supported by populating the XCVR U26. RS-422 is supported by
U21 and U37, the RS-422 XCVR's. A simple charge pump circuit is used to
generate +5V_ISO on the host side of the barrier. Prodigy II will populate the
RS-422 section and run at 115.2KB.
2.5.32 Debug RS-232 Port
The debug port is provided such that the firmware can echo status messages
and other information to a dumb terminal to support debugging, development,
testing, and servicing. The port is fully driven by a 16550 compatible UART is
fully connected to a DB-9 header using the standard PC comm port pinout.
The firmware can be configured to accept input from the debug port if
required. The port is expected to be depopulated in the final production BOM
to minimize cost.
2.5.33 Diagnostic LED's
16 diagnostic LED's are provided by the cSBC. Functionality is listed below
LEDColorFunctionDescription (status indicated when lit)
D13Green+24V+24V DC input from supply present.
D14Green+12V+12V DC input from supply present.
D15Green-12V-12V DC input from supply present.
D7Green+5V+5V DC input from supply present.
D16RedFLEX CONFIGFLEX PLD not programmed.
D20RedCPU RESETCPU in reset mode.
D19RedSCANNER RESETScanner in reset (failsafe) mode.
D27RedAGS ROLLAGS roll over or under detected.
D29AmberFLEX DIAG 1Diagnostic LED for misc use by firmware, control reg in FLEX.
D31AmberFLEX DIAG 2Diagnostic LED for misc use by firmware, control reg in FLEX.
D22AmberMAX DIAG 2Diagnostic LED for misc use by firmware, control reg in MAX.
D28AmberTRANS OMIValid transverse motion sensed by OMI circuitry.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-62 System OverviewPRODIGY Service Manual (Rev C - 2000)
D22AmberLONG OMIValid longitudinal motion sensed by OMI circuitry.
D25AmberHE COUNTValid high energy photon event sensed by DCA circuitry.
D23AmberLE COUNTValid low energy photon event sensed by DCA circuitry.
D21AmberHVPS ENABLE7681 HVPS enabled.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-63
2.6Power Distribution (PRODIGY II Systems
DF+1200 and greater)
The PRODIGY II system reduces scanner cost by using a single switching DC
power supply which provides +5,+/-12, and +24VDC. The supply is connected
directly to the cSBC via J14. J13 is provided as a power out connector to run
power to the DMB for Prodigy II. 0R0 bypass resistor are provided to allow
flexibility in grounding and filtering schemes. The PWB is divided into 4
ground plane regions - DGND for +5V digital returns, AGND for +/-12V analog
returns, SCNGND for +24V returns, and ISO_GND for the host I/O island.
Large zeners and rectifiers are provided for over-voltage and reverse-voltage
protection. Standard capacitor filter banks are provided. For EMC all cable
exits other than host and DMB I/O are located over the 24V planes. The 24V
plane is separated from the 5V plane via optical isolation to prevent 5V noise
from reaching cables (radiated EMC) and to prevent motor noise and ESD
pulses from reaching the digital and analog IC's. The 24V plane contains
primarily power FET switches and power resistors used to control scanner
motors, X-ray source, LED, etc. The +/-12V planes contain analog electronics
used for detector and HVPS control. A +12V regulator is used to produce
12VDC for the Centent drives of off +24VDC, maintaining isolation from the
+12V plane. An additional +5V regulator is used to make 5V for the OMI's of
off +12V motors. 4 opto channels are used to sense the presence of the 4 DC
input voltages. The opto outputs are used to drive 4 green power on indicator
LED's and to allow the CPU to detect power outs.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-64 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.7Tube Head and X-ray Insert
The Tube Head consists of an oil-filled metal housing which contains a fixedanode X-ray Insert (essentially a vacuum tube diode), lead-type shielding,
collimating devices, a filament transformer, electrical connectors, and a
Cerium (K-edge) filter.
THe X-ray generator provides a constant 76kV potential. mA control circuit
regulates the fialment current via filament temperature to amintain the
programmed current emission. The X-ray generator only provides mA
monitor and mA limiting. PRODIGY I systems (DF+11999 and lower) have
MAX CCA, PRODIGY II systems (DF+12000 and greater have mA control
internal to the x-ray generator.
2.7.1 X-ray generation and Spectrum
• The X-ray Insert converts current into x-rays by accelerating the
electrons across a large potential from the cathode into a tungsten target
on the anode.
• The x-rays produced by the x-ray insert are filtered by the cerium filter to
produce a dual energy spectrum. The x-ray spectrum has peaks at 62
and 38 kV.
• Insert potential is provided by a/the High Voltage Power Supply(ies).
• The tube current is set by the (c)SBC at 0.15 mA or 3.00 mA depending
on the acquisition type.
• Tube Current is set by the (c)SBC and regulated by the MAX Board
(PRODIGY I) or the Power Supply (PRODIGY II systems).
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-65
2.8X-ray Generator (High Voltage Power Supply(ies))
The PRODIGY X-ray production system combines a constant potential x-ray
generator (consisting of either one or two high voltage power supplies) and an
X-ray tube. The x-rya generator provides a continuous output of 38 kV each
and up to 5 mA (PRODIGY I) or 3.5 mA (PRODIGY II) to the X-ray tube. The
x-ray insert anode is composed of a tungsten alloy. The specifications of the
X-ray generation subsystem is outlined in section 2.16.7.
• The High Voltage Power Supply(ies) provide the anode/cathode potential
to the X-ray insert.
• PRODIGY I scanners use two high-voltage power supplies (±40kVDC).
During normal operation, 76kVp is applied (+38kV at the anode and 38kV at the cathode).
• PRODIGY II scanners use a single high-voltage power supply
(±40kVDC). During normal operation, 76kVp is applied (+38kV at the
anode and -38kV at the cathode).
• The High Voltage Power Supplies are controlled by the (c)SBC.
• These power supplies are powered by the AC line voltage and have their
own built in fusing.
• The AC power is routed thru the isolation transformer (PRODIGY I Only)
and the supply(ies) are enabled via a relay (see section 2.1.1).
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-66 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.9MAX Board (PRODIGY I systems DF+11999 and
lower only)
2.9.1 MAX Board Function
• The Tube Head and MAX Board work with the dedicated +28VDC power
supply to supply tube head current.
• The current is set by the SBC (see section 2.4).
• MAX board actually regulates current to the filament transformer in the
Tube Head.
2.9.2 Dedicated +28VDC power supply
• The +28 VDC dedicated power supply is switched on (for fail-safe
purposes) by a solid-state relay.
• When errors are detected by the SBC, (for example, loss of arm motion)
the relay will switch off and prevent production of x-rays.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-67
2.10 XORB Board (PRODIGY I Systems DF+11999 and
lower only)
The XORB printed circuit board provides protection to various circuits (low
voltage) from transients within the HVPS and tube head (see 2.1.1 for
location).
• There may be occasional static discharges within the Tube Head. The
transients caused by these static discharges (arcs) are shunted to
ground through the array of transorbs present on the XORB board.
• There is no safety hazard to personnel, however, the electronics could be
damaged were no protection provided.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-68 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.11Detector Sub System
2.11.1 Detector Overview
• 16 individual CZT (cadmium zinc telluride) crystals which convert X-ray
energy directly to electrical charge
• The charge pulse generated by the individual detectors is directly
proportional to the photon energy that generated the charge pulse (see
2.16.9 for spectrum information).
• The array of 16 CZT crystals each act as pixel detectors, from which the
data is combined to form the densitometry data set.
• The Detector electronics amplify, discriminate and counts x-rays for all 16
detector elements.
• The detector communicates with the SBC.
• The Detector has 5 PCB associated with it, the Detector Mother Board
(DMB) and 4 identical Detector Daughter Boards (DDB’s).
• The detector has its own dedicated High Voltage Power Supply, used for
supplying detector bias.
2.11.2 Detector Operation
Figure 2-15. Detector Block Diagram
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-69
The Detector is held at 450 VDC bias. When a X-ray strikes the detector an
electron pair is created (small black arrows in the block diagram). This charge
pair induces a small amount of current in the CZT crystal.
A charge sensitive pre-amp then converts the current to voltage and passes
the signal along to the Detector Daughter Board for processing and analysis.
The signal out of the pre-amp is a +5 to +10 millivolt DC signal, each event on
each channel creates a small “blip”.
2.11.3 Detector Daughter Board Overview
• Each DDB processes the data from 4 of the 16 detector elements
simultaneously.
• Each detector element is handled as a separate channel.
• Each DDB is responsible for:
• Pulse Shaping
• Gain Adjust
• Energy discrimination (high or low energy)
• Event counting
• Each DDB is electronically identical
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-70 System OverviewPRODIGY Service Manual (Rev C - 2000)
Figure 2-16. Detector Daughter Board Operation
2.11.4 Detector Daughter Board Operation
Gain Stage / Pulse Shaping
The Detector Daughter Board takes the +5 to 10 millivolt signal, strips off the
DC offset and shapes the pulse into a bi-polar pulse. This signal is further
amplified to 2.4 VDC and is then fed to a bank of 4 pairs of comparators.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-71
The four fine gain control comparators are set up in two windows around the
High Energy signal. If the signal falls into the lower window a signal is sent to
the programmable logic (PLD). The PLD then sends an amplify signal to a
DAC which in turn increases gain slightly at the pulse shaping amplifier, so
that the High Energy Signal is +2.4 VDC. If the High Energy signal falls into
the upper window, the fine gain is adjusted down slightly by the same
mechanism at the pulse shaping amplifier, again so that the High Energy
Signal is + 2.4 VDC.
The fine gain adjust is applied to the entire signal on a particular channel at
the pulse shaping amplifier, therefore the Low Energy Signal’s gain is
adjusted along with the High Energy Signal.
The 4 comparators of the Gain Stage are set by digital to analog converters
(DAC’s) which use reference values stored in the Flash RAM of the Detector
Mother Board.
Energy Discrimination and Pulse Counting
The amplified and gain adjusted pulse is simultaneously fed to a second bank
of 4 comparators which are divided again into two sets of windows. The
windows of the Discrimination stage are centered around the High and Low
Energy pulses. These comparators are also set by DAC’s which are set by the
DMB Flash RAM.
Each time a pulse falls into the High Energy window, a counter in the
programmable logic (PLD) is ticked, the same happens for the Low Energy
window. These two data signals are then passed to the DMB, which transmits
the data to the SBC.
2.11.5 Detector Mother Board Overview
• Contains 4 slots for the DDB’s (see figure 2-9)
• Contains a slot for the Test Point Board (part of the PRODIGY tool kit see
figure 2-9)
• The Detector Mother Board provides +5 V, ±12 V plus digital and analog
grounds to the Detector Daughter Boards.
• The Detector Mother Board communicates with the SBC via RS-422 link.
• The patient positioning laser is controlled by the SBC via the DMB (See
figure 2-9).
2.11.6 Detector Mother Board Operation
The Detector Mother Board (DMB) contains a processor and Flash RAM,
which supplies programming data to the detector subsystem. The data for all
comparator settings is stored here (set values), detector Bias programming
data, firmware to program the PLD’s of the 4 DDB’s, the Peak value
calculation operation for the detector (see section 2.17 for a discussion of the
Secondary Calibration (QA)).
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-72 System OverviewPRODIGY Service Manual (Rev C - 2000)
If the Flash RAM is corrupt, the DMB will read from the Firmware on the SBC.
The Flash RAM can be reprogrammed from the Host / SBC.
The DMB has a processor, which is responsible for performing the Peak
calculation during a Secondary Calibration (QA), programming the 4 DDB
PLD’s and handles all communications with the SBC. The DMB processor
sends the SBC status reports and echoes back all commands that are sent to
it.
The DMB will reset the SBC if it detects an error.
Error Conditions of the DMB sensed by the SBC:
• PLD programming failure
• Loss of +5 or +12 VDC
• Loss of Communications with SBC
• Corrupt FIRMWARE
•FLASH RAM Failure
• DTR Reset button on the DMB is pressed
Figure 2-17. PRODIGY Detector Module
1 = Test Point Board, 2 = Detector Daughter Boards (4), 3 = Patient Positioning laser and
Detector Mother Board Interface, The Detector Mother Board is the horizontal circuit board
that the DMB’s and Test Point Board are plugged into.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-73
2.12 FOINK (PRODIGY I Systems DF+11999 and lower
only)
2.12.1 FOINK Functions
• The FOINK Board optically isolates control circuitry (SBC) from the rest
of the scanners electronics (see pan layout Appendix 2B).
• Shutter solenoid
• Scanner Motor Controllers
• Display Panel indicators and controls
• Monitors 5 different interrupts, all of which are capable of resetting the
SBC (shutting down x-ray production)
• Emergency Stop Switch - open circuit - (loss of signal - failsafe)
• Tube Head Thermocouple - circuit open (loss of signal - failsafe)
• Longitudinal Motion Interrupt - source open no movement (OMI)
• Transverse Motion Interrupt - source open no movement (OMI)
• Failure of the +28 VDC Power Supply
2.12.2 Motion Control and Detection
• Motor pulses, RUN/HOLD signals and direction are generated by the
SBC.
• The RUN/HOLD signals are optically isolated on the FOINK board before
being sent to the Centent motor drivers. The Centents supply current to
the stepper motor windings.
• Both the transverse and longitudinal drive systems have slotted disks
attached which pass through optical interrupters (OMI) for motion
detection. Should motion stop during patient scanning, the FOINK sends
an interrupt to the SBC, which is reset and x-ray production is halted.
• Limit switches at each end of the scanner and each end of the arm signal
the SBC whenever motion reaches either end or side of the scanner. If a
limit switch is tripped during a patient scan, the motor associated with the
tripped limit switch is halted, the SBC is reset and the shutter is closed.
• There is a dedicated DC power supply for the motion control circuitry,
supplying +26VDC. The +26VDC ground is isolated from the logic
ground to reduce noise (see section 2.1.1 for location)
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-74 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.13 Display Panel
• The Display Panel of the PRODIGY, located on the front face of the
upper arm, is the main operator interface outside the host computer.
• Four indicators display system status.
• A green Light Emitting Diode (LED) indicates a power-on condition; it
lights when the +26VDC power supply comes on.
Figure 2-18. PRODIGY display panel
Power on LED is lit whenever scan table power is on.
• A yellow LED indicates if the laser is on
The laser on LED is lit when the SBC signals the DMB to turn
the laser on, the LED is switched on by the DMB.
• An amber X-ray On LED indicates that x-rays are being produced
(current is flowing through the X-ray Insert), though exposure is not
necessarily taking place.
The X-ray on LED is lit when more than X current is flowing
through the X-ray insert.
• The yellow Shutter Open LED comes on is the Shutter open and
exposure possible.
The Shutter open LED is controlled by the SBC, the SBC will
allow the shutter to open even if the lamp is not lit (shutter
open indicator is also on the PC controller screen).
Note:It is not recommended that scanner operation be continued if
These LED's signal the operator of exposure to x-rays and
must be replaced as soon as possible.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-75
• Rocker switches on the panel allow operator control of the beam
position.
The Rocker switches are enabled by the SBC via the FOINK.
Note:The rocker switches are also referred to as the joystick in the
PRODIGY Service Manual and the PRODIGY Service
Software.
• The Emergency stop button (red), is to be used only in emergency
situations; it resets the (c)SBC and halts x-ray production.
The FOINK or cSBC detects if this normally closed circuit is
opened (fail-safe operation).
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-76 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.14 Audible X-RAY OFF Signal
• An audible signal notifies the operator that there is a “safe” condition
when the Shutter is closed.
• The FOINK or cSBC board senses the closure of the Shutter or the
cessation of current through the X-ray Insert (+28VDC Power Supply or
Model 7861 HVPS) and switches on the audible signal if either of these
occur during an exposure.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-77
2.15 X-Ray Collimator Subsystem
• The Shutter is actuated by a rotary solenoid using +26 VDC, the voltage
is reduced to a holding voltage of 12VDC after the shutter is opened.
• This solenoid is controlled by signals generated on the SBC, and
optically isolated through the FOINK board on PRODIGY I systems.
• An optical interrupter (OMI) detects the shutter open condition and
reports shutter position information back to the (c)SBC.
• This assembly is located on top of the X-ray Tube Head.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-78 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.16 PRODIGY Specifications
2.16.1 Component specifications
Table 2-3 gives specifications for standard components shipped with the
PRODIGY system.
Table 2-3. PRODIGY Component Specifications
ComponentSpecifications
Dimensions: 262.3 cm x 109.3 cm x 128.3 cm
Scanner table*
Console table78.5 cm x 63.3 cm x 48.1 cm
Computer
Windows
Internet Explorer version 4.01 with service pack 2
Fast Serial I/O board (LUNAR part number 7151)
Weight: approximately 272.16 kg
Maximum patient weight supported: 136 kg
Greater than 266Mhz Pentium
128 MB RAM
Greater than 1GB hard disk
17” SVGA monitor (800x600x16-bit color)
LS-120 Super drive
CD ROM
Modem
®
NT version 4.0 operating system with service
pack 6
Printer
W x D x H—44.0 x 40.0 x 19.6 cm
*Depth is measured from the front edge of the scanner table to the back edge of the
scanner arm. Height is measured from the top of the scanner arm to the bottom of the
scanner arm.
*Width is measured from the front edge of the scanner table to the back edge of the
scanner arm. Height is measured from the top of the scanner arm to the bottom of the
scanner arm.
#Some languages may require 1024x768x16-bit color to fit the translated software text
on the screen
@GE-LUNAR Recommends Diskeeper Lite (Executive Software - http://
www.execsoft.com - Glendale, California, USA)
HP DeskJet 930C
2.16.2 Functional specifications
General specifications
Focal spot to image receptor distance is 67 cm. Attenuation equivalence of
patient support table is 0.7 mm Al.
2.16.3 Maximum scan area (long x transverse)
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-79
• AP Spine Measurements
40.3 cm x 18 cm
• Femur Measurements
20.2 cm x 18 cm
• Dual Femur Measurements
20.2 cm x 18 cm (each femur - two scan areas)
• Total Body Measurements
197.5 cm x 60 cm measurement field
2.16.4 Programs
Note:Depending on the number of options purchased, not all of the
options listed below may be included with the PRODIGY
system software:
• AP spine measurement and analysis
• Femur measurement and analysis
• Total Body measurement and analysis.
• Forearm measurement and analysis
• Lateral spine measurement and analysis
• Quality Assurance
2.16.5 Environmental specifications
Operational environment
Adhere to the specifications that follow during scanner operation:
••Ambient Space (Interior Subcomponents)–For scanner operation
and servicing, do not block the area around the scanner table. Make sure
there is a minimum clearance of 30.5 cm at the head and foot ends of the
scanner table, at least 15.2 cm for the arm side, and 45.7 cm for the
operator side.
• Ambient Space (Ventilation)–Do not block the cooling vents on the
computer and scanner table. Make sure there is 15.2 cm from the
console table to the wall for cable clearance and computer plugs.
• Dust, Fumes and Debris–Install the system in a clean, ventilated area.
Dust and other airborne debris can cause the diskette drive heads and
other sensitive mechanical components to malfunction. GE-LUNAR
recommends that smoking is not permitted in the scanner room.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-80 System OverviewPRODIGY Service Manual (Rev C - 2000)
• Humidity–Make sure the humidity for the scanner area is 20%–80%,
non-condensing.
• Static Electricity–Install and operate the system in a static-free area.
Adhere to minimum humidity requirements to prevent malfunctions
caused by static electricity.
• Shock and Vibration–Make sure the scanner table does not receive
shock greater than 1 G for more than 1 millisecond. Make sure the
scanner table does not receive vibrations greater than 0.25 G at 5 Hz.
• Temperature–Make sure the temperature during system operation is
65°F–81°F (18°C–27°C).
Note:When the system is turned off, or there is a power failure,
allow the system to be on and let it warm for one hour. After
one hour, complete a Quality Assurance procedure.
2.16.6 Storage and transport environment
Adhere to the specifications that follow for scanner storage and
transportation:
• Humidity, 0% to 95% non-condensing.
• Atmospheric pressure, 500 to 1060 hPa.
• Temperature, -40° to 70° C.
2.16.7 X-ray generator (system no. DF+12000 and higher)
Table 2-4. X-ray generator technical information.
ClassificationClass I EquipmentIEC 601-2-7 5.1
Degree of protection against electrical
shock
Protection against ingress of liquids
Connection to supply mainsPower supply cordIEC 601-2-7 6.1g)
Mode of operationContinuousIEC 601-2-7 6.1m)
Maximum X-ray tube voltage76 kVIEC 601-2-7 6.1m)
Maximum X-ray tube current3 mAIEC 601-2-7 6.1m)
Rated mains voltage100-240 VACIEC 601-2-7 6.1j)1
Type B equipmentIEC 601-2-7 5.2
Ordinary medical electrical equip-
ment
IEC 601-2-7 5.3
Number of phases in mains1IEC 601-2-7 6.1j)2
Mains frequency50/60 HertzIEC 601-2-7 6.1j)3
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-81
Table 3-9 gives the specifications for the GE-LUNAR laser.
Table 2-9. Laser specifications.
Output Power<1mW
Wavelength635nm
Beam Diameter at
aperture
4x1mm
Aspect Ratio4.0 to 1
Divergence24 degrees
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-88 System OverviewPRODIGY Service Manual (Rev C - 2000)
Table 2-9. Laser specifications. (continued)
Radiant Exposure0.0001 W
Integrated Radiance46 W
Current Draw105 mA
Voltage Input4-6 VDC
Safety RatingClass II
2.16.13 Compatible components
For customers located internationally, make sure the computer is certified to
local requirements. The computer must meet the minimum requirements that
follow:
• Greater than 266MHz Pentium
• 128 MB RAM
• Greater than 1GB Hard Disk
• 8X CD ROM
• 14” SVGA monitor with at least 800x600x16-bit color
• Windows NT version 4.0 operating system with service pack
• Internet Explorer version 4.01 with service pack 2
2.16.14 FDA Certified Components (USA Only)
Tables 2-9 and 2-10 list components certified to the FDA for use with
PRODIGY scanners and is updated periodically. Contact GE-LUNAR for a
current listing of compatible components.
Table 2-10. FDA certified components (system no. DF+12000 and higher).
ComponentDescription
X-ray ControllerLUNAR single board controller7635
High Voltage Power
Supplies
Bertan
SBD40PN280X2890
Tube Head AssemblyLUNAR X-Ray Tube Head Assembly8743
Collimator
LUNAR PRODIGY Collimator Assem-
1
Model: 2907
Spellman
bly
2
Model:
LUNAR Model
#
7681
7681
8915
1
Bertan High Voltage Corp., 121 New South Road, Hicksville, NY
2
Spellman High Voltage Electronics Corp., 475 Wireless Boulevard, Hauppauge, NY
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-89
Table 2-11. FDA certified components (system no. DF+11999 and lower).
ComponentDescriptionLUNAR Model #
X-ray Controller
High Voltage Power
Supplies
LUNAR PRODIGY single board con-
troller
Spellman
1
Models:
PTV40N200X2113
PTV40P200X2112
2
Bertan
Models:
2411 N
2411 P
5447
0311
0312
0311
0312
Tube Head AssemblyLUNAR X-Ray Tube Head Assembly6838
Collimator
1
Spellman High Voltage Electronics Corporation Hauppauge, NY
2
Bertan Associates, 121 New South Road, Hicksville, NY
LUNAR PRODIGY Collimator Assem-
bly
6893
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-90 System OverviewPRODIGY Service Manual (Rev C - 2000)
2.17 Secondary Calibration / Daily QA
2.17.1 Secondary Calibration overview
• Daily QA (Quality Assurance) tests the integrity of the scanner so
operator knows it is operating within specifications.
• Daily QA (Secondary Calibration) adjusts the calibration of the scanner to
offset any changes due to the tube aging, temperature varying, or when
certain components are changed.
• A third type of function, QC Phantom, is another method for testing
system performance. A QC phantom is scanned using standard software
and results are stored (aluminum spine in lucite).
• Daily QA ensures the system is operating properly by running 3 types of
tests:
• A peak test to adjust photon counting electronics.
• A functional test that checks the indicator lights, motion system,
shutter, tube output, and detector performance.
• A test that measures a secondary calibration object with different
bone chambers and tissue plugs which correct for system aging and
prevents long-term drifts.
• On many tests, the results must be calculated for all 16 detectors
separately. The display for the test will consist of the mean value.
• All tests are run all of the time. The tests are executed and test results
reported as each test completes. Test results include pass/fail and any
quantitative information when appropriate.
2.17.2 Starting the Daily QA (secondary calibration)
Operator presses [F5] to start [Daily QA] which presents a plot of the last used
parameter from the last Daily QA run. (Default display is the medium bone
chamber.) At this point operator can interactively manipulate the data in QA
database to create tables and graphs that can be printed.
Tech presses [Start] to start running Daily QA tests.
Software displays instructions to position calibration block on the table. Tech
places block on tabletop silk screen and click [OK] to start.
The UI locks out all other operations except Stop button.
Scan arm moves to the "Home" position. In this case, Home is always at head
of scanner for QA independent of user setting for Home.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-91
2.17.3 Tests Performed in the Secondary Calibration
Detector Mother Board Self Test
The DACs and Flash RAM of the Detector Mother Board are tested.
Find Block
The scanner performs a scout scan to find the block. If block is not properly
positioned, the scanner will prompt the user to Reposition standard more
accurately.
The software automatically finds landmarks within the secondary phantom.
This reduces the need for exact positioning on the table top.
Peak Test
During this test, the shutter opens with the x-rays on.
This test adjusts the sensitivity of the photon counting electronics. The test
determines the optimal voltage setting for the detector amplifier so the
maximum number of photons are detected. With the shutter open the software
adjusts the voltage setting for the detector amplifier, acquiring counts at the
different amplifier settings.
The detector peak test then determines the optimal voltage setting for the
detector amplifiers based on the results. The actual peak setting is taken from
the high energy count rates, the low energy channel is essentially blocked by
the brass spillover piece in the QA standard.
Beam Stop test
Measures the movement of the source shutter. A functional test checks the
shutter LSW and ensures the shutter stops all photons.
Motor motion and limit switches
OMI Interrupt function - the OMI signal is tested by verifying the OMI Interpol
signal is present when the scan arm is stationary.
Each motor (transverse, longitudinal) will be run between hard limits and back
to home position. If the hard limits do not engage, this is a failure. Also
compare steps against predetermined scan window size.
If a failure occurs the number of steps taken, the expected value and the
direction the arm was traveling is noted in the error log.
Spillover test
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-92 System OverviewPRODIGY Service Manual (Rev C - 2000)
Measures the effect of high energy photons registering as low energy
photons in the detector system. The beam passes through a brass piece that
stops all low energy photons. The detector then measures the number of low
energy photons detected as a measure of spillover. Spillover should be less
than 13%. Spillover with the CZT detector will degrade about 0.2% per year.
Spillover Stability test
As part of the secondary calibration, the change in spillover must be
calculated on a per element basis and smoothed over the last 3 passing QA's.
Daily QA measures actual spillover and stores it in QA database. As part of
primary calibration, the spillover value used during that calibration is stored.
The delta spillover calculation using the spillover from primary calibration and
the running average of the last five spillover measurements. This information
is in the scan file.
Reference Value Test
The output of the x-ray tube is measured as part of a Reference Value (old
term was air) measurement. This is measured at various currents between
3mA and 0.150mA.
Note:Not all currents that are ramped to / tested are used to perform
patient scans.
Reference Value measurement
The HE/LE counts are measured as the number of photons counted by
detector. This value is used in all bone density calculations. It must be stored
in the systems data file. Limits for low/high counts -
• Reference value measurements will be performed at various currents.
These reference values are stored separately for all detector elements.
• Reference Value measurements for all current settings are actually made
through lucite sections of the secondary calibration block.
Reference value ratio test
The HE/LE ratio for each reference value is calculated for each tube current
tested in reference value measurement.
Detector Pre-AMP test
The detector Pre-amplifiers are tested by removing detector bias and verifying
that the signal out lines of the detector are not noisy.
Tissue Value Test
Tissue secondary calibration
• The system measures two tissue eqivilent materials (lucite and Acetron).
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)System Overview 2-93
• Store measured values for each chamber per element in QA database.
• The system determines the%fat for two different materials with different
compositions.
• Three%fat values are displayed - lean, mid and fat
• These calculated%fat values are compared against expected values and
checked against limits for the mean, standard deviation.
BM Chamber Measurements
The system acquires point measurements on the 3 bone chambers (approx
12 seconds / measurement) using the 3ma medium scan mode. The system
determines the bone mineral content for the three chambers of different sizes
and densities.
• Measured values for each chamber are stored on a per element basis in
the QA database.
• These calculated BM values are compared against expected values and
checked against limits for the mean, standard deviation.
• A slope is calculated and this adjustment is used as a secondary BMD
calibration to offset long term drifts in system performance.
Daily QA Results
At the end of Daily QA, a report is automatically printed. If printer is not on line
or fails, post an error to screen.
If Daily QA fails, post an error message that says "Unsuccessful QA Recommend repeating procedure."
• If a QA fails the error log will contain detailed information on the
failure.
Test results are saved automatically to a database.
Appropriate secondary calibration and Reference Value information is
recalculated from QA database using an average of the 3 previous passing
test results. This information is stored in each scan file.
A view with plot of medium bone chamber is generated, and user the user is
allowed to manipulate data interactively.
2.17.4 QA Database
A historical record of system performance is maintained by storing key QA
parameters in a database.
• 3mA Results - fat is hidden - only bone results are viewable
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
2-94 System OverviewPRODIGY Service Manual (Rev C - 2000)
Lunar Corporation
_
726 Heartland Trail
Madison, WI 53717
1.051 Date 3/20/00
Time 9:07:32 AM
1.007 BMD Medium (High mA) 0.971g/cm
0.971g/cm Deviation -0.82%
0.957 Mean BMD Medium (High 0.971g/cm
CV
Total Sessions 3
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
3-102Service SoftwarePRODIGY Service Manual (Rev C - 2000)
3.0Diagnostic Software
Note:GE-LUNAR Diagnostic software options are not available to
the customer. This software is available to GE-LUNAR
authorized representatives only.
the customer. The software will return to operators
mode as soon as the software is exited and restarted.
The operator mode may also be changed under the
Tools/User Options menu.
• The Diagnostic Software is available through the User Options Tab
When the software is used in the
diagnostic / service mode, the Engineer
performing service must verify that the
scanner is taken out of service mode
before returning control of the scanner to
The Service software gives the Technician
access to all areas of the
PRODIGYSoftware, DO NOT adjust AGS
subsystem settings.
3.0.1 To Access the Service Software:
• Start the PRODIGYSoftware by double clicking the PRODIGY Icon on
the desk top
• Double click the Too l s option on the menu bar
• When the drop down menu appears double click on User Options
• Double click on the Systems Tab on the dialogue box
• Click on Service under Display Mode
• When prompted enter the password Award (versions 2.05 and greater)
Smile versions lower than 2.05)
Note:the password is case sensitive
• The Tools option on the menu bar will now contain Service Options, and
Service scanning options will also be available under the F2 Measure
Option
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)Service Software3-103
3.1The Tools Menu
3.1.1 Spectrum
Allows the service engineer to view and print the raw energy spectrum for the
detector. The detector can also be repeaked from Spectrum.
3.1.2 Stability Run
Used for the detection of arcing and testing the detectors performance
3.1.3 Signal Monitor
View count rates for low and high-energy channels for the detector.
3.1.4 Primary Calibration
Used to calibrate the PRODIGY scanner See DXAP@002 PRODIGY Primary
Calibration Procedure Chapter 5 Appendix C.
3.1.5 Secondary Verification
Used in the manufacturing process
Figure 3-1. The Tools/Diagnostics Menu
3.1.6 Pileup
Used in the manufacturing process
3.1.7 Set Download Parameters
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
3-104Service SoftwarePRODIGY Service Manual (Rev C - 2000)
Download and verify the Firmware versions for the (c)SBC.
3.1.8 enCORE Composer
This is not a service tool - (Physical and Business Report Generator - see the
enCORE Operator’s Manual for information).
3.1.9 Error Log
Ability to view and print the contents of the diagnostic error log (see section
3.4)
3.1.10 Copy Configuration
Ability to copy the scanners Error log, Configuration Files and / or QA
Database to a drive on the host PC - useful when a host PC is replaced or
when diagnostic information needs to be sent to GE-LUNAR.
3.1.11 User Options
Set up user configured options for the scanner.
3.1.12 Service Options
Service Mode scanner configurations (see section 3.5).
3.1.13 Outbox
Configure Printing, E-mail and Faxing options.
3.1.14 System File Editor
Edit system settings for the enCORE software - do not alter these settings
without GE-LUNAR Customer Support authorization.
3.2Tools / Diagnostics Menu
3.2.1 Pileup
Determines Detector parameters for multiple events occurring simultaneously
- not used for service.
3.2.2 Calibration Pileup
Not used for service.
3.2.3 Spillover
Perform and Acquire a Spillover Measurement (test of source spectrum and/
or detector resolution).
3.2.4 Count Rate
Acquire and print Reference Counts (count rate of detector must be
performed through lucite).
3.2.5 Scanner Motion
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)Service Software3-105
Control manually and perform scanner motion diagnostics and motion
configuration.
•Motion Tests Tab
•Limit Switch Monitor - used to verify limit switch status
•Limit to Limit Test - used to verify limit switch positions
• Configure check box- this box must be checked when limit to limit is
run to establish the scanners range of motion (scan window).
•Cable Stress Test - used to cycle Transverse and Longitudinal
Mechanics - can be used to check new cables / check for binding
•Home scanner - moves mechanics to Home position
•Motion Commands Tab
•Home Position - change Arm Parking position from head to foot does not effect where the daily QA (Secondary calibration) is
performed - Move to Home - sends scan arm to home position.
Note:When the Home position is changed from the head end of the
table to the foot or vice versa, the Limit to Limit test (see
Motion Tests tab above) must be run with the Configure box
selected.
•Move Steps - move the scan arm to a specific location on the table move Absolute moves an absolute distance from home, Move
Relative - move relative to last position
•Joystick and Laser - Enable and Control Scanner Motion manually,
change motor speeds, toggle laser on and off
•Lower portion of the screen displays the status of limit switches, scan
arm position, and positioning switch (joystick) status.
•Find Block Tab
•Secondary Calibration - can be used to find / verify that the scanner is
able to find the different chambers on the secondary calibration (daily
QA) block. THis tool can be useful for troubleshooting “cannot find
Block errors” and for Transverse binding issues.
3.2.6 Scanner Detector
•Detector Tests Tab
•Gain Control - adjusts the gain on the AGS amplifier used in the
detector peak setting - peak adjusts only the AGS amplifier.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
3-106Service SoftwarePRODIGY Service Manual (Rev C - 2000)
•Bias Control - adjusts the bias of the detector. PRODIGY Systems
run with detector bias on demand (when x-rays are present) of
450VDC.
•DCA window settings - view DCA window settings - these settings
should not be edited.
•Detector status - allows the service engineer to view the detector
status real time - information displayed includes Peak setting Bias
setting, and DCA window settings.
•Detector AGS Tab
•AGS DAC control (verify AGS DAC settings), check on detector
rollover, and check on channel count rates.
DO NOT adjust the detector Bias, DCA
windows or AGS DAC settings changes in the Detector bias, DCA
windows or AGS DAC’s can change the
scanner’s results.
•AGS Mode - the operate / calibrate signal (disables the AGS systems
so that the detector can be peaked correctly) can be toggled.
3.2.7 Scanner X-ray
•X-ray Tests Tab-
•Manual Test - Control Manually the Shutter, X-ray on Indicator and
patient Positioning laser
•Automatic test - automatically toggle the Shutter, Laser and X-Ray
Relay
•X-Ray Source - Turn on source (ramp and make x-rays), warm up a
tube head at install or replacement.
•X-ray cycle test - cyclically ramp the source - (can be used to test
ramping failures or for arcing)
•X-rays Tab
•Ramp the kV and / or the mA of the x-ray generation system and view
feedback real-time. Faults are displayed and polled real-time in the
window on the right.
•Ports Tab - view status of and send commands to (c)SBC ports (see
section 2.4 for (c)SBC port definitions).
•Download Tab - identical to Set Download Parameters (section
3.1.7)
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)Service Software3-107
•Misc. Tab- not used for service.
3.2.8 Lin/Rep
Not Used for Service
3.2.9 Limit Switch Adjustment
Verify and Adjust Limit switch positions.
Note:When using Limit Switch Adjustment, DO NOT alter the
default settings.
3.2.10 Scanner Disconnect
Disconnect the scanner from the PRODIGY software (reset the (c)SBC).
3.3Diagnostic Scan Modes -
Performed under F2 Measure - only visible when the scanner is in service
mode.
3.3.1 Table Top Scan -
Not used in Service
3.3.2 Alignment Scan
Verify that the Source and Detector maintain a constant count rate in all areas
of the scan table. Also useful for testing cables.
3.3.3 Beam Wobble scan
Quantifies beam movement relative to the detector when the scanners
mechanics are moving (see DXAP2010 Chapter 5 Appendices).
3.3.4 Hacksaw
Calculates and sets correcton factors to align bone edges in scans (see
DXAP 2007 Chapter 5 Appendices).
3.4Error Log
• The PRODIGYwill maintain an error log file whenever the scan table is in
• The error log is one of the best troubleshooting tools for scanner
operation. All errors, fatal or nonfatal will be logged here. The Error log
gives a brief description of the error and if possible the origin of the fault.
malfunction.
Note:To aid in a quick and accurate response when requesting
assistance from GE-LUNAR, please have an error log
available for review by GE-LUNAR Customer Support.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
3-108Service SoftwarePRODIGY Service Manual (Rev C - 2000)
The error log file is named errorlog.bin, this file is the current error log session.
Older sessions are named errorlog.1, errorlog.2 and so on up to errorlog.4.
When requesting help from GE-LUNAR, E-mailing the current error log file
and a description of the symptoms will aid in a rapid diagnosis (see Copy
Configuration in section 3.1.10 for information on copying the error log to an
E-mailable file).
3.4.1 Printing the Error Log.
Note:The scanner does not have to be in Service mode to view or
print the Error Log.
• The Error Log is located under the Tools dropdown menu. Left click on
Tools / and then on Error Log.
Figure 3-2. The Error Log Dialogue Box
• Session - The PRODIGYsoftware will maintain up to 10 “sessions” each session is up to 100 errors.
•A new session is created by shutting down and restarting the
enCORE software. A new session may also be started by clicking on
the button
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
PRODIGY Service Manual (Rev C - 2000)Service Software3-109
•To view a session, left click on it, all errors (if present) will be shown in
the box below.
•The Errors list may be printed (click on print errors) or it may be
searched with the Find Error button.
•The details of the failure are shown when the error is highlighted and
then right clicked on.
Figure 3-3. Error log entry details
3.4.2 Troubleshooting Help software
This feature is found in software versions 2.26 and greater. When an error is
highlighted and the Troubleshoot button is clicked on, it opens an online
enCORE troubleshooting guide to assist the service engineer in
troubleshooting the system by error code.
This document contains confidential or proprietary information of GE-Lunar Corp. Neither the document nor the information therein is
to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE-Lunar Corp.
3-110Service SoftwarePRODIGY Service Manual (Rev C - 2000)
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.