GE Medical Systems MAC 1200 Operator's Manual

MAC® 1200
Operator’s Manual
Software Version 6
2012250-022 Revision B
Listed below are GE Med ica l Sy stem s Information Technologies trademarks. All other trademarks contained herein are the property of their respective owners.
900 SC, ACCUSKETCH, AccuVision, APEX, AQUA-KNOT, ARCHIVIST, Autoseq, BABY MAC, C Qwik Connect, CardioServ, CardioSmart, CardioSys, CardioWindow, CASE, CD TELEMETRY, CENTRA, CHART GUARD, CINE 35, CORO, COROLAN, COROMETRICS, Corometrics Sensor Tip, CRG PLUS, DASH, Digistore, Digital DATAQ, E for M, EAGLE, Event-Link, FMS 101B, FMS 111, HELLIGE, IMAGE STORE, INTELLIMOTION, IQA, LASER SXP, MAC, MAC-LAB, MACTRODE, MANAGED USE, MARQUETTE, MARQUETTE MAC, MARQUETTE MEDICAL SYSTEMS, MARQUETTE UNITY NETWORK, MARS, MAX, MEDITEL, MEI, MEI in the circle logo, MEMOPORT, MEMOPORT C, MINISTORE, MINNOWS, Monarch 8000, MULTI-LINK, MULTISCRIPTOR, MUSE, MUSE CV, Neo-Trak, NEUROSCRIPT, OnlineABG, OXYMONITOR, Pres-R-Cuff, PRESSURE-SCRIBE, QMI, QS, Quantitative Medicine, Quantitative Sentinel, RAC RAMS, RSVP, SAM, SEER, SILVERTRACE, SOLAR, SOLARVIEW, Spectra 400, Spectra-Overview, Spectra-Tel, ST GUARD, TRAM, TRAM-NET, TRAM-RAC, TRAMSCOPE, TRIM KNOB, Trimline, UNION STATION, UNITY logo, UNITY NETWORK, Vari-X, Vari-X Cardiomatic, VariCath, VARIDEX, VAS, and Vision Care Filter are trademarks of GE Medical Systems Information Technologies registered in the United States Patent and Trademark Office.
12SL, 15SL, Access, AccuSpeak, ADVANTAGE, BAM, BODYTRODE, Cardiomatic, CardioSpeak, CD TELEMETRY Cumulus, Event-Link Nimbus, HI-RES, ICMMS, IMAGE VAULT, IMPACT.wf, INTER-LEAD, IQA, LIFEWATCH, Managed Use, MARQUETTE PRISM, MARQUETTE MicroSmart, MMS, MRT, MUSE CardioWindow, NST PRO, NAUTILUS, O
®
-LAN, CENTRALSCOPE, Corolation, EDIC, EK-Pro, Event-Link Cirrus, Event-Link
®
RESPONDER, MENTOR,
SENSOR, Octanet, OMRS, PHi-
2
Res, Premium, Prism, QUIK CONNECT V, QUICK CONNECT, QT Guard, SMART-PAC, SMARTLOOK, Spiral Lok, S wee the art, UNITY, Univer s al, Wate r fa ll, and Walk mom ar e tra dema rks of GE Med ica l Syst ems Information Technologies.
© GE Medical Systems Information Technologies, 2004. All rights reserved.
T-2 MAC 1200 Revision B
2012250-022 21 July 2004
CE Marking Information
CE Marking Information
Compliance
The MAC 1200 bears the CE mark “CE-0459”, notified body GMED, indicating its conformity with the provisions of the Council Directive 93/ 42/EEC concerning medical devices and fulfills the essential requirements of Annex I of this directive.
Any other directive(s) and all the standards the product complies to are listed in the general information of the operator manual for the product following this page.
The country of manufacture can be found on the equipment labeling. The product is in radio-interference protection class A in accordance with
EN 55011. The product complies with the requirements of standard EN 60601-1-2
“Electromagnetic Compatibility - Medical Electrical Equipment”.
Exceptions
The safety and effecti veness of this device has been ve rified against previously distributed devices. Although all standards applicable to presently marketed devices may not be appropriate for prior devices (i.e. electromagnetic compatibility standards), this device will not impair the safe and effective use of those previously distributed devices. See user’s information.
The MAC 1200 EMC: Immunity Performance
Users should be aware of known RF sources, such as radio or TV stations and hand-held or mobile two-way radios, and consider them when installing a medical device or system.
Be aware that adding accessories or components, or modifying the medical device or system may degrade the EMI performance. Consul t with qualified personnel regarding changes to the system configuration.
Revision B MAC 1200 CE-1
2012250-022
General Information
CE Marking Information
The device is designed to comply with IEC 60601 requirements. It is a protection class I device.
The CE mark covers only the accessories listed in t he chapter “Orde r Information”.
The information contained in th is ma nual d escr ibes sof tware ve rsion
6.
CE-2 MAC 1200 Revision B
2012250-022
Contents
1 The Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Manual Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
MAC 1200 Resting ECG Analysis System Option Codes . . . . . . . . . . . . . . . . . . . 1-4
Intended Use and Functional Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Intended Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Illustrations and Names . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
Underwriters Laboratories, Inc. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-12
Biocompatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Literature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Service Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-14
2 Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
MAC 1200 Control Panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
MAC 1200 Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
3 Operating and Performance Tests . . . . . . . . . . . . . . . . . 3-1
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Installation and Mains Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Performance Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Contrast Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
System Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Revision B MAC 1200 i
2012250-022
Connecting Peripheral Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
4 Preparing for ECG Recording . . . . . . . . . . . . . . . . . . . . . 4-1
Connecting the Patient Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Electrode Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Applying Plate (Limb) Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Applying Suction Electrodes (Chest) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Electrode Placement for Standard Leads (l, II, III, aVR, aVL, aVF, V1...V6) . . . . .4-6
Artifact Due to Poor Electrode Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Entering Patient Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
New Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-12
Last Name, First Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-12
Date of Birth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-12
Patient ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-12
Chest Pain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
Pacemaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
Gender/Race . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
Height/Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
Systolic BP/Diastolic BP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
Ordering Physician / Referring Physician / Technician . . . . . . . . . . . . . . . . . . . .4-14
Phone Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-14
Medication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-14
Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-14
ID Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-14
Secondary ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-14
Secondary ID Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
Last/First Name Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
Location Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
Room . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
Order Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
Prompts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
5 Recording in 12 Lead Mode . . . . . . . . . . . . . . . . . . . . . . . 5-1
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Recording . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
The Storage Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Report Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Detailed Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-11
ECG Transmission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
ii MAC 1200 Revision B
2012250-022
General Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Transmission via Modem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Batch Transmission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-15
Transmission Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-15
Transmitting Data to a MUSE CV System Via Modem . . . . . . . . . . . . . . . . . . . .5-16
Receiving Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-19
Cart to Cart Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-20
Modem Setup (for Modem Æ Other) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-20
Adjusting Measurement Points/QT Dispersion . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
Global Measurement Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-22
Local T Offset Measurement Point/QT Dispersion . . . . . . . . . . . . . . . . . . . . . . .5-24
Brief Operating Instructions – 12 Lead Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-26
6 Recording in 6 Lead Mode . . . . . . . . . . . . . . . . . . . . . . . . 6-1
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Recording . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
7 Arrhythmia M ode Recording . . . . . . . . . . . . . . . . . . . . . . 7-1
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Recording . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
During the Recording . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-6
Final Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-8
8 Pacemaker Patients / Recording During Defibrillation . 8-1
Recording ECGs of Pacemaker Patients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
ECG Recording During Defibrillation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
9 System Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
12 Lead Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Report Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
Rhythm Leads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
Gain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
Revision B MAC 1200 iii
2012250-022
Report Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
Detailed Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
Continuous Rhythm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-6
Muscle Filter/AC Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-6
Muscle Filter Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-6
Manual Copy To . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-7
Number of Copies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-7
Interpretation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-7
Print Interpretation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-7
Override Function [no] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-8
6 Lead Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9
Report Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-9
Gain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-10
Speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-10
Muscle Filter/AC Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-10
Muscle Filter Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-10
Anti-Drift System (ADS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-11
Auto Paper Feed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-11
Arrhythmia Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12
Report Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-12
Gain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-12
Muscle Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-12
AC Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-12
Muscle Filter Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-13
Trend Recording . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-13
Arrhythmia Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-13
Episodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-13
Pharma . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-14
Patient Data Customization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-14
Project Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-15
Trial ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-15
Investigator ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-15
Visit Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-16
Visit Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-16
Dose Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-16
Extra Questions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-16
High Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-17
Device Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-17
Don’t Allow Record Edits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-18
Delete Only After Transmission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-18
Force Technician Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-19
System Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-20
Ordering Physician/Referring Physician/Technician . . . . . . . . . . . . . . . . . . . . . .9-20
Institution Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-20
Cart Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-20
Site Number * . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-20
iv MAC 1200 Revision B
2012250-022
Location* . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-21
Date/Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-21
Lead Fail Beep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-21
High HR Beep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-21
Lead Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-21
Pace Enhancement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-21
Baseline Roll Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-21
Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-22
Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-22
Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-22
Mains . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-22
LCD Light Off After . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-22
Low Battery Beep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-22
Default mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-22
Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-22
Enable Password Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-23
Test Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-23
Restore Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-23
Print Setup Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-23
Transmission Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-24
Check Record Retxn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-24
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-25
Baud Rate (PC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-25
Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-25
Modem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-25
PIN Dialing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-26
Patient Data Menu Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-27
Required Data Fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-28
Extra Questions 1 to 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-28
Option Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-29
ECG Transmission via Modem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-30
Selecting the Communication Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-30
Direct ECG Transmission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-31
Selecting the Communication Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-31
10 Loading Chart Paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
End-of-Paper Indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-5
Aging Stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-6
Revision B MAC 1200 v
2012250-022
11 Cleaning, Disinfection and Maintenance . . . . . . . . . . . 11-1
Cleaning and Disinfecting the Recorder Housing . . . . . . . . . . . . . . . . . . . . . . . . 11-3
Cleaning and Disinfecting the Patient Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-3
Cleaning and Disinfecting the Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-3
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
Checks Before Each Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-4
Technical Inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-4
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-4
12 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-1
Troubleshooting Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-3
13 Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . 13-1
Recording . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-3
Printer Paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-3
Paper Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-3
Membrane Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-4
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-4
Indicators (LEDs) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-4
Lead Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-4
Automatic Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-4
Detection of Pacer Pulses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-5
Heart Rate Indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-5
Signal Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-5
Data Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-6
Transfer of ECGs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-6
Receiving Data with the CSI Communication Protocol from the Following Units 13-6
Sending ECGs to the Following Units with the A5 Protocol . . . . . . . . . . . . . . . . .13-6
Remote Start (Hardware) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-7
Signal Transmission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-8
14 Order Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-1
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-3
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-3
General Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-3
vi MAC 1200 Revision B
2012250-022
Entering Special Characters . . . . . . . . . . . . . . . . . . . . . . .A-1
Special Characters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Index-1
Revision B MAC 1200 vii
2012250-022
For your notes
viii MAC 1200 Revi sion B
2012250-022

1 The Basics

Revision B MAC 1200 1-1
2012250-022
For your notes
1-2 MAC 1200 Revision B
2012250-022

About This Manual

Revision History

This manual has a revision letter, locate d at the bottom of each page. This revision letter changes whenever t he manual is upd ated. Revisi on A is the initial release of the document.

Manual Purpose

This manual describes the safe and effective operation of the MAC 1200 unit.
The Basics: About This Manual
Revision Date Comments
A 31 March 2003 Initial release of manual, describes version 6.0. B 21 July 2004 Revised CE compliance information, High Security
“memory full” information added.
127(
This document describes the functionality of the U.S. interface for the MAC 1200 unit.
Information in this manual differs from operating information for MAC 1200 units developed for use internationally. Please refer to PN 2012250-021 for information on using the international interface.
Revision B MAC 1200 1-3
2012250-022
The Basics: MAC 1200 Resting ECG Analysis System Option Codes

MAC 1200 Resting ECG Analysis System Option Codes

In addition to the software supplied with the unit, optional programs may be purchased to upgrade the MAC 1200 performance features. In order to use a new option, you need to activate it by entering the option code number (refer to “Option Co de” on pag e 9-29 for details). The option codes are entered into the MAC 1200 prior to shipping.
Software Package Functionality Option Code
MEAS Measurement (measurement of the 10-second resting ECG) _ _ _ _ _ _ _ _ _ _ _ _ DIAG Interpretation (interpretation of the 10-second resting ECG) _ _ _ _ _ _ _ _ _ _ _ _ MEMO Memory (storage of a maximum of 40 10-second resting ECGs) _ _ _ _ _ _ _ _ _ _ _ _ C100 Activates the three options MEAS, DIAG, MEMO for a maximum of 100 ECGs _ _ _ _ _ _ _ _ _ _ _ _ C500 Activates the three options MEAS, DIAG, MEMO for a maximum of 500 ECGs _ _ _ _ _ _ _ _ _ _ _ _ EVAL Activates the three options MEAS, DIAG, MEMO for a maximum of 4 weeks _ _ _ _ _ _ _ _ _ _ _ _
Serial No: _ _ _ _ _ _ _ _ _
1-4 MAC 1200 Revision B
2012250-022
The Basics: MAC 1200 Resting ECG Analysis System Option Codes

Intended Use and Functional Description

The MAC 1200 is an ECG acquisition and recording system designed and manufactured by GE Medical Systems Information Technologies.
It is intended to be used for resting ECG recording and realtime ECG recording with or without arrhythmia detection.
It is not intended for use as a vital signs physiological monitor.
The arrhythmia detection porti on of the M AC 1200 is pr ovided to the customer for the convenience of automatic documentation. It is not designed to provide alarms for arrhythmia detecti on .
The MAC 1200 offers no diagnostic opinion to the user. Instead it provides ana lytical statement s when configured wi th the appropriate options.
It is intended to be used by trained operators under direct physician supervision when ECG records are required.
It is not suitable for intracardiac application.
It is designed for continuous operati on.
It is not intended for home use.
The MAC 1200 is designed as a portable device and can easily be moved from one patient to another or to different locations. It is not intended to be used during patient transport.
Equipped with the standard software, the MAC 1200 supports the following operating modes.
12 Lead Mode (acquisition of 12 leads of ECG for a period of 10 seconds),
6 Lead Mode ( realtime recording of 6 ECG leads), and
Arrhythmia Mode (continuous ECG analysis for arrhythmias).
The graphics display shows 3 leads at a time. Resting ECGs can be transferred to the CardioSys/CardioSoft or MUSE
CV Information System via the RS232 interface. The device operates from b oth AC and DC (recharg eable batt eries) power
sources.
Revision B MAC 1200 1-5
2012250-022
The Basics: MAC 1200 Resting ECG Analysis System Option Codes
The unit’s performance features can be upgraded with the followi ng optional programs.
MEAS — measurement (measurement of t he 10-s eco nd rest ing E CG)
DIAG — interpretation (int erpre tat io n of the 1 0-seco nd re st ing ECG)
MEMO — memory (storage of a maximum of 40 10-second resting ECGs)
C100 — activates the three options MEAS, DIAG, MEMO for a maximum of 100 ECGs
C500 — activates the three options MEAS, DIAG, MEMO for a maximum of 500 ECGs
EVAL — activates the three options MEAS, DIAG, MEMO for a period of 4 weeks
The MAC 1200 resting ECG analysis system has a setup menu to customize th e sy st em pa r am e te r s.
Patient and user data can be entered for reliable and safe archiving of patient records. The patient name is annotated on each printed report page. All other data is printed on request.
1
2
1
y
6
p o c
/ t a
y
m
d
d
r
h
a
r
e
o
r
e
f
e
l
a p s
/
e
t
l
e
r
r
c
p
o
a
s
u
t
t e s
0
P
9
X
O
)
8
I
L
(
7
K
U
6
%
Y
J
5
>
*
H
T
M
4
+
<
R
G
N
3
-
.
E
F
B
2
=
,
D
W
V
1
?
/
Q
S
C
!
:
A
X ; Z
on
off
alt
1200
C
A
M
The MAC 1200 units are designed to comply with IEC 60601 / EN 60601 requirements. They are protection class I devices/devices with an internal power source. They are classified as MDD class IIa devices. They are designed for continuous ope ration. The units are not suitable for intracardiac application. The units are not intended for use as vital signs physiological monitors.
t
u
r
s
s
e
m
t l i
f
retrieve
n
i a g
p
o
t
s
t a p
o
f
n
i
w
lo y r
e
t t a
b
y
b d n a
t s
102A
1-6 MAC 1200 Revision B
2012250-022
The Basics: MAC 1200 Resting ECG Analysis System Option Codes

Intended Audience

Definitions

Black text Indicates keys on the keyboard, text to be entered, or hardware items such as buttons or
This manual is geared for clinical professionals. Clinical professionals are expected to have working knowledge of medical procedures, practices, and terminology as required for monitoring of critically ill patients.
&$87,21
PATIENT HAZARD — Medical technical equipment such as the MAC 1200 must only be used by qualified and trained personnel.
The following formats are used in this manual to highlight various web viewer features and functions.
switches on the equipment.
Italicized text Indicates software terms that identify menu items, buttons, or options in various windows.
Ctrl+Esc Indicates a keyboard operation. A (+) sign between the names of two keys indicates that
you must press and hold the first key while pressing the second key once. For example, “Press Ctrl+Esc” means to press and hold down the Ctrl key while pressing the Esc key.
<Space> Indicates you must press the space bar. When instructions are given for typing a precise
text string with one or more spaces, the point where the space bar must be pressed is indicated as: <Space>. The purpose of the < > brackets is to ensure you press the space bar when required.
Enter Indicates you must press the “Enter” or “Return” key on the keyboard. Do not type “enter”.

Illustrations and Names

All illustrations in this manual are provided as examples only. They may not necessarily reflect your monitoring setup or data displayed on your monitor.
In this manual, all names appearing in examples and illustrations are fictitious. The use of any real person’s name is purely coincidental.
Revision B MAC 1200 1-7
2012250-022

Safety Information

This manual is an integral part of the device. It should always be kept near the device. Close observance of the informa tion given in t he manual is a prerequisite for proper device performance and correct operation and ensures patient and operator safety. Please n ote that information pertinent to several chapters is given only once. Therefore, carefully read the manual o nce in its entirety.
The symbol means: Consult accompanying documents. It indicates points whi ch are of particul ar importance in the operation of the device.
This manual is in conformity with the devi ce specifications and standards on safety of electromedical equipment valid at th e time of printing. All rights are reserved for devices, circuits, techniques, software programs, and names appearing in this manual.
On request GE will provide a service manual. The GE quality management system complies with the standards DIN
EN ISO 9001 and EN 46001.
The Basics: Safety Information
To ensure patient safety, the specified measuring accuracy, and interference-free operation, we recommend to use only original GE components. The user is responsible for application of accessories from other manufacturers.
The warranty does not cover damage resulting f rom the use of unsuitab le accessories and consumables from other manufacturers.
GE is responsible for the effects on safety, reliability, and performance of the device, only if
assembly operations, extensions, readjustments, modifications, or repairs are carried out by GE or by persons authorized by GE, and
the device is used in accordance with the instructions given in this operator’s manual.
1-8 MAC 1200 Revision B
2012250-022

Definitions

The Basics: Safety Information
The terms danger, warning, and caution are used throughout this manual to point out hazards and to designate a degree or level of seriousness. Familiarize yourself with their definitions and significance.
Hazard is defined as a source of potential injury to a person. '$1*(5 indicates an imminent hazard which, if not avoided, will
result in death or serious injury. :$51,1* indicates a potential hazard or unsafe practice which, if not
avoided, could result in death or serious injury. &$87,21 indicates a potential hazard or unsafe practice which, if not
avoided, could result in minor personal injury or product/property damage.
127( provides application tips or other useful information to assure that you get the most from your equipment. The safety information given in this manual is classified as follows.
'$1*(5
EXPLOSION HAZARD — The device is not designed for use in areas of medically used rooms where an explosion hazard may occur. An explosion hazard may result from the use of flammable anesthetics, skin cleansing agents and disinfectants.
Revision B MAC 1200 1-9
2012250-022
The Basics: Safety Information
:$51,1*6
SHOCK HAZARD — Strictly observe the following warnings. Failure to do so may endanger the lives of the patient, the user and bystanders.
Before using the device, the operator must ascertain that it is in correct working order and operating condition. In particular, all connectors, electrodes as well as sensors and probes must be checked for signs of damage. Damaged parts must be replaced immediately, before use.
When disconnecting the device from the power line, remove the plug from the wall outlet first, before disconnecting the cable from the device. Otherwise there is a risk of coming in contact with line voltage by inadvertently introducing metal parts in the sockets of the power cord.
The mains plug must be connected to an appropriate power supply with a non-fused grounded-to-earth wire. If these requiremen ts cannot be met, operate the device on battery power.
Do not use multiple portable socket outlets (MPS O) to connect the device to the power line.
Devices may be connected t o other devices or to parts of systems only when it has been made certain that there is no danger to the pat ient, the opera tors, or the environment as a result. In those instances where there is any element of doubt concerning the sa fety of connected devices, the user must contact the manufacturers concerned or other informed experts as to whether there is any possible danger to the patient, the operator, or the environment as a result of the proposed combination of devices. Standards IEC 60601-1-1/EN60601-1-1 must be complied with in all cases.
All devices of a system must be connected to the same electric circuit. Devices which are not connected to the same circuit must be electrically isolated (isolated RS232 interface).
1-10 MAC 1200 Revision B
2012250-022
The Basics: Safety Information
:$51,1*6
EQUIPMENT FAILURE — Magnetic and electrical fields are capable of interfering with the proper performance of the device. For this reason make sure that all peripheral devices operated in the vicinity of the recorder comply with the relevant EMC requirements. X­ray equipment, MRI devices, radio systems (cellular telephones) etc. are possible sources of interference as they may emit higher leve ls of electr omagneti c radiati on. Keep the recorder away from t hese devices and verify the recorder performance before use.
SUFFOCATION HAZARD — Dispose of the packaging material, observing the applicable waste-control regulations. Keep the packaging material out of children's reach.
&$87,216
EQUIPMENT DAMAGE — Devices intended for emergency application must not be exposed to low temperatures during storage and transport to avoid moisture condensation at the application site. Wait until all moisture has vaporized before using the device.
EQUIPMENT DAMAGE — Before connecting the device to the power line, verify that the ratings of your local power line are those indicated on the device nameplate.
RESTRICTED SALE — U.S. federal law restricts this device to sale by or on the order of a physician.
Revision B MAC 1200 1-11
2012250-022
The Basics: Safety Information
4P41
Classification
The unit is classified, according to IEC 60601-1, as:
Type of protection against electrical shock Class I internally powered equipment Degree of protection against electrical
shock Degree of protection against harmful
ingress of water Degree of safety of application in the
presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide
Method(s) of sterilization or disinfection recommended by the manufacturer
Mode of operation Continuous operation
Type CF defibrillation-proof applied part
Ordinary Equipment (enclosed equipment without protection against ingress of
water) Equipment not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide
Not applicable

Underwriters Laboratories, Inc.

Medical Equipment With respect to electric shock, fire and mechanical hazards only in accordance with UL 2601-1, and CAN/CSA C22.2 NO. 601.1.
1-12 MAC 1200 Revision B
2012250-022

Biocompatibility

Literature

The Basics: Biocompatibility
The parts of the product described in this operator manual, including all accessories, that come in contact with the patient during the intended use, fulfill the biocompatibility requirements of the applicable standards. If you have questions in this matter, please contact GE Medical Systems Information Technologies or its representatives.
Medical Device Directive of August 2, 1994 EN 60601-1: 1990 + A 1: 1993 + A 2: 1995
Medical electrical equipment. General requirements for safety. EN 60601-1-1: 9/1994 + A1: 12/1995
General requirements for safety. Requirements for the safety of medical electrical syst e ms.
IEC-Publication 513/1994: Fundamental aspects of safety standards for medical equipment.
Revision B MAC 1200 1-13
2012250-022

Service Information

Requirements

Refer equipment servicing to GE Medical Systems Information Technologies’ authorized service personnel only. Any unauthorized
attempt to repair equipment under warranty voi ds that warranty. It is the user’s responsibility to report the need for service to GE Medical
Systems Information Technologies or to one of their authorized agents. Every GE Medical Systems Information Technologies device has a
unique serial number for identification. The serial number appears on the device label.
The Basics: Service Information
A
B
F E D CG
Table 1. Equipment Identification
Item Description
A name of device B manufacturer C location code D serial number E unique product code F last digit of year manufactured G month manufactured
105A
1-14 MAC 1200 Revision B
2012250-022

2 Controls and Indicators

Revision B MAC 1200 2-1
2012250-022
For your notes
2-2 MAC 1200 Revision B
2012250-022
Controls and Indicators: General Information

General Information

Controls and indicators of the MAC 1200 electrocardiograph are shown in this chapter.

MAC 1200 Control Panels

1
12 345
2 1
y
6
p o
c
/ t a
y
m
d
d
r
h
a
r
e
o
r
e
f
e
l
a p s
/
e
t
l
e
r
r
c
p
o
s
u
t
t e s
0
P
9
X
O
)
8
I
L
(
7
K
U
6
%
Y
J
5
>
*
H
T
M
4
+
<
R
G
N
3
-
.
E
F
B
2
=
,
D
W
V
1
?
/
Q
S
C
!
:
A
X ; Z
on
off
alt
1200
C
A
M
ta
u
r
s
s
e
m
t l i
f
retrieve
n i a
g
p
to
s
t a
p
o f
in
w
o l y r
e t
t a b
y
b d n a
t s
Description
1 Power input 2 Paper door, windows allows you to check the paper supply 3 Patient cable connector 4 Connection for electrode application system KISS (option) 5 Serial interface (See Chapter 13, “Technical Specifications” for details.)
001A
Revision B MAC 1200 2-3
2012250-022
Controls and Indicators: General Information

MAC 1200 Keyboard

1234567891011
12
on
stdby
Q
1
!
A
2
W
S
Z;X
alt
4
R
T
-
F
/
V,B.N
5
3
E
=
?
D
:
C
6
Y
+
*
G
H
8
7
I
U
(
%
J
K
<
>
M
O9P
)
L
standby battery low
0
X
format/
speed
muscle
filter
gain
pat
info
store/
retrieve
copysetup
lead
12
6
arrhy
start
stop
13 14 15
23 22 21 20 19 18 17 16
Description
1 Power switch (ON/OFF) 2 Keys to select a higher or lower HR alarm limit 3 Confirms entered data 4 Correction key (entry of data) 5 Displays the configuration menu 6 Enables/disables the muscle filter (elimination of muscle artifact) 7 Selects the writer speed 25, 50 or 5 mm/s in 6-Lead Mode and the report formats in 12-Lead Mode 8 Selects the gain (5, 10, 20, 40 mm/mV)
9 Press to print the report or additional copies of the ECG, or to send/receive ECGs 10 Selects the ECG leads displayed and recorded in 6-Lead Mode and displayed 12-Lead Mode 11 Sends ECG to memory/retrieves ECG from memory 12 Selects the 12-Lead Mode 13 Selects the 6-Lead Mode 14 Selects the Arrhythmia Mode 15 Starts and stops the recorder, clears the setup menu and terminates patient data entry
098A
16 Indicators:
Green: recording in selected mode started; Yellow: recording in selected mode stopped
2-4 MAC 1200 Revision B
2012250-022
Controls and Indicators: General Information
Description
17 Enables entry of patient data 18 Indicator lights up when battery needs to be recharged 19 Indicator is illuminated when unit is connected to the power line 20 Cursor control keys 21 Space bar 22 Shift key 23 Press to access special characters
Revision B MAC 1200 2-5
2012250-022

Symbols

Controls and Indicators: Symbols
Consult accompanying documents
Signal input
Type CF signal input, highly insulated, defibrillation-proof
Start
Stop
2-6 MAC 1200 Revision B
2012250-022
3 Operating and
Performance Tests
Revision B MAC 1200 3-1
2012250-022
For your notes
3-2 MAC 1200 Revision B
2012250-022

Power Supply

Operating and Performance Tests: Power Supply
8
7
I
U
(
%
J
K
<
>
M
0
O9P
)
L
standby battery low
X
format/
speed
muscle
filter
gain
pat
info
copysetup
lead
store/
retrieve
12
6
arrhy
start
stop
Indicator 2 Indicator 1
106A
Indicators
The units are powered from the power line or from the rechargeable battery.
The battery charges automatically when the unit is connected to the power line and the green indicator 2 is illuminated as shown above. It is not necessary to switch on the device for charging. To ensure that the battery is alway s fully charged, leave the e lectrocardiogra ph connected to the power line whenever possible. The battery regains its full capacity after being connected to the power line for four hours.
Indicator 1 is illuminated when the battery needs to be charged. The unit can also be set up to emit an additional audio signal when the battery requires charging.
With a full battery, about 50 ECGs (1 page each) can be recorded in the 12 Lead mode. When its capacity drops to about 25 recordings, the battery is used up and must be replaced by a service specialist.
127(
To prolong the battery life, fully discharge the battery at least once per month (by operating the electrocardiograph on battery power).
127(
In standby mode, a fully charged battery is drained within approximately 4 hours. Therefore, when operating the device on battery power, be sure to turn it off when it is not in use.
Revision B MAC 1200 3-3
2012250-022
Operating and Performance Tests: Installation and Mains Connection

Installation and Mains Connection

The figure below shows a practical arrangement of patient and recorder. For interfer ence-f ree oper atio n, it is import ant th at the patient cable a nd the power cord do not run parallel.
Arranging the Electrocardiograph and the Examination Couch
Use the power cord to connect the device to the power line (see figure below). Use only the original power cord or an equivalent cable.
003A
004A
AC Power Input
Indicator 2 will illuminate.
Check the paper supply (the window in the paper door allows you to look inside the paper compartment). Refer to Chapter 10, “Loading
Chart Paper” for instructions on inserting a new paper pad.
3-4 MAC 1200 Revision B
2012250-022
Operating and Performance Tests: Performance Check

Performance Check

Press the power bu tton to switch the device on.
The amber indicator will illuminate. After power-up, the electrocardiograph runs an automatic self-test. The
display indicates the memories currently being tested. The self-test takes about 15 to 20 seconds. When no problem is detected, the device defaults to the 12 Lead mode. If a malfunction is identified, the dis p lay will show an error message Error... In this situation, notif y ser vi ce to che ck an d repair the device.
on
stdby
Q
1
!
A
Z;X
2
3
W
E
?
S
D
:
alt
Power Button
4
R
=
/
C
5
T
+
-
G
F
V,B.N
6
Y
*
H
005A
The self-test can be skipp ed with the button. In this case, the device immediately activates the default mode confugured in the Setup menu.
127(
The backlighting of the display switches off automatically when no key is activated for 20 minutes (adjustable). The illumination is turned on again by activation of any key.
Run the full self-test at least once a day to ensure that the device is functioning properly.
Press and hold the power button for several seconds to turn the device off.
Revision B MAC 1200 3-5
2012250-022
Operating and Performance Tests: Performance Check

Contrast Adjustment

Simultaneously press and the appropriate cursor key:
Alt
more contrast ,
less contrast .
3-6 MAC 1200 Revision B
2012250-022

System Setup

Operating and Performance Tests: System Setup
The table below shows the system setup parameters that can be modified, as well as the the fact ory defa ults . Refer to Chapter 9, “System
Setup” section for details.
Table 1. System Setup Menu
Parameter Factory Default Options
Ordering Physician none selection from a list of 10 names Referring Physician none selection from a list of 10 names Technician none selection from a list of 10 names Institution Name empty text box text box for 40 characters Cart # 1 1 ... 9999
1
Site # Location # 1 ... 600 Ease of Clock Setting No Yes Date (dd.mm.yyyy) current date Time (hh:mm) current time Lead fail beep No Yes High HR beep No Yes Lead Labels IEC AAMI Pace Enhancement Yes No Baseline roll filter 0.08 Hz 0.04 Hz, 0.16 Hz Date dd.mm.yyyy mm/dd/yyyy Time 24 12 Units cm, kg in, lb Mains 50 Hz 60 Hz LCD light off after 20 min 1 ... 99 min Low battery beep 0 (Off) 5 s to 60 s (5 s = beep at 5 second
Default Mode 12-Lead 6-Lead, Arrhythmia (MAC 1200 only) Language German all available languages Enable password No Yes Test DATA No Yes Restore defaults No Yes Print Configuration Lists No Yes Print ECG Tx Log No Yes Check Record Retxn No Yes
1 1 ... 255
interval)
Revision B MAC 1200 3-7
2012250-022
Operating and Performance Tests: System Setup
1. only for CSI communication protocol
3-8 MAC 1200 Revision B
2012250-022
Operating and Performance Tests: Connecting Peripheral Equipment

Connecting Peripheral Equipment

:$51,1*
SHOCK HAZARD — Strictly observe the following warnings. Failure to do so may endanger the lives of the patient, the user and bystanders.
Connecting peripheral devices to the RS2 32 interfa ce of the electrocardiograph creates a medical system. This system must meet the requirements of IEC 60601-1-1.
Use only the original GEMS IT connection cables.
All non-medical devices of a system must be connected to the same electric circuit. Devices which are not connected to the same circuit must be electrically isolated (use isolated RS232 interface as per IEC 60601-1).
A PC connected to the electrocardiograph should meet the requirements of EN 60601. If it doesn't, it must be set up outside the patient environment. If the PC fulfills the requirements of EN 60950, it must be set up within the medically used area, but outside the patient environment.
Do NOT connect PCs to the electrocardiograph that fulfill neither EN 60601 nor EN 60950.
Modems connected to the electrocardiograph must meet the requirements of EN 60950 or UL1950 (all modems recommended by GEMS IT meet these requirements). The specific regulations valid in your country must also be observed.
The modem must be set up within the medically used area, but outside the patient environment.
Revision B MAC 1200 3-9
2012250-022
Operating and Performance Tests: Connecting Peripheral Equipment
The electrocardiograph can be directly connected via the serial interface to a PC (CardioSoft), to the CardioSys system, or to a MUSE CV system.
Resting ECGs acquired in the 12 Lead mode, and the corresponding data, can then be transmitted to these peripheral devices. See Chapter 5,
“ECG Transmission” section for details.
The table below shows the factory defaults and all possible adjustments. For instructions on changing the default setup, see “Communication” on
page 9-25 for details.
Table 2. Modem Configuration Menu
Parameter Factory Default Options
Choices for Modem —> Other
no other
MultiTech 19.2 MultiTech 56k Elsa 14.4 Elsa 28.8 Elsa 33.6
Elsa 56k Choices for Modem —> other Phone Number
Initial. modem
AT&FM1X3S
0=1V0 dial string hang­up
ATDT
+++ATH Choices for Modem —> MultiTech, Elsa Dial mode
Tone Pulse Phone Number Outside line
0 to 9 (20 digits) 0 to 9 (20 digits)
3-10 MAC 1200 Revision B
2012250-022
For your notes
Operating and Performance Tests: Connecting Peripheral Equipment
Revision B MAC 1200 3-11
2012250-022
Operating and Performance Tests: Connecting Peripheral Equipment
3-12 MAC 1200 Revision B
2012250-022
4 Preparing for ECG
Recording
Revision B MAC 1200 4-1
2012250-022
For your notes
4-2 MAC 1200 Revision B
2012250-022
Preparing for ECG Recording: Connecting the Patient Cable

Connecting the Patient Cable

:$51,1*
SHOCK HAZARD — Strictly observe the following warnings. Failure to do so may endanger the lives of the patient, the user and bystanders.
For reasons of patient safety, use only the original GE Medical Systems Information Technologies patient cables. Before connecting the cable to the device, check it for signs of mechanical damage. Do not use a damaged cable.
Ensure that conductive parts (such as the patient, connectors, electrodes, transducers) that are connected to the isolated patient signal input do not come into contact with other grounded, conductive parts. This would bridge the patient's isolation and cancel the protection provided by the isolated input. The neutral electrode, in particular, must not come into contact with ground.
If your electrocardiograph is equipped with an integrat ed suction pump connector, you can connect the electrode application system KISS inst ead of the standard patient cable.
Use the 10-lead patient cable for acquisition of the standard ECG leads (Einthoven, Goldberger, Wilson).
Connect the patient cable to ECG signal input.
ECG Signal Input
!
006A
Revision B MAC 1200 4-3
2012250-022
Preparing for ECG Recording: Electrode Application

Electrode Application

Careful application of electrodes and skin prepara tion are the key to an interference-free ECG.
&$87,21
PATIENT HAZARD, Delayed ECG Display — Use only silver-silver chloride electrodes when recording the ECG of a patient who may have to be defibrillated. (Refer to Chapter 8, “Pacemaker Patients / Recording During
Defibrillation” for details.)
4-4 MAC 1200 Revision B
2012250-022
Preparing for ECG Recording: Electrode Application
R
A

Applying Plate (Limb) Electrodes

Plate electrodes are applied by means of a rubber strap, and electrode paper is the recommended contact medium.
Moisten the electrode pap er with tap water a nd place it bet ween skin and electrode.
Secure the electrode with the rubber strap, but do not hinder blood circulation.
A
L
RL LL
Applying Limb-Lead Electrodes
Lead Description
RA (white) electrode on right arm LA (black) electrode on left arm LL (red) electrode on left leg RL (green) electrode on right leg
&$87,21
Use only silver-silver chloride electrodes, if the patient may have to be defibrillated. (See “ECG Recording
During Defibrillation” on page 8-4 for details.)
V
103A

Applying Suction Electrodes (Chest)

Shave application points, if necessary.
Moisten the electrode paper with tap water and place it between skin and electrode. Electrode cream or gel can be used instead of paper. On hairy chests, the cream or gel improves adhesion of the electrodes.
Revision B MAC 1200 4-5
2012250-022
Preparing for ECG Recording: Electrode Application

Electrode Placement for Standard Leads (l, II, III, aVR, aVL, aVF, V1...V6)

For acquisition of the standard ECG leads, four electrodes must be applied on the limbs and six on the chest. The limb electrodes should be placed above the wrists and ankles. The figure below shows the chest electrode application points.
V1
V2
V2V3
V3
V4
V4
V5
V6
V6
V5
V1
Chest Electrode Placement
Description
V1 4th intercostal space at the right border of the sternum V2 4th intercostal space at the left border of the sternum V3 Midway between locations V2 and V4 V4 At the mid-clavicular line in the 5th intercostal space V5 At the anterior auxiliary line on the same horizontal level as V4 and V6 V6 At the mid-auxiliary line on the same horizontal level as V4
7A
V7* At the left posterior auxiliary line in the 5th intercostal space V8* At the left scapulary line in the 5th intercostal space V3R* Opposite V3, on the right side of the chest V4R* Opposite V4, on the right side of the chest
* additional standard leads
4-6 MAC 1200 Revision B
2012250-022
Preparing for ECG Recording: Electrode Application
Connect the 10-lead patient cable as shown below.
RL green RA white
right leg right arm
V1 red V2 yellow V3 green
V1
V2
V3
V4
V5
V6
V4 blue V5 orange V6 purple LA black LL red
left arm left leg
V1
V2V3
V4
V6
V5
Connecting the Patient Cable (10-Lead Cable, Standard ECG Leads)
Arrange the leadwires and patient cable as shown below.
009A
Correct
Incorrect
010A
Arranging the Patient Cable
Revision B MAC 1200 4-7
2012250-022
Preparing for ECG Recording: Electrode Application
A

Artifact Due to Poor Electrode Application

The electrocardiograph is equipped with state-of- the-art electronic utilities that ensure artifact-free recordings. Among these are the automatic baseline adjustment and the anti-drift system (cubic spline) (ADS).
At the beginning of the recording the automatic baseline adjustment algorithm verifies the incoming signal and adjusts the baseline position accordingly.
During the recording, the anti-drift system (cubic spline) continuously checks the baseline position and returns it to the normal level, if required (see Sample Recording figure).
For the Manual Mode, the anti-drift system (cubic spline) can be enabled and disabled from the setup menu, in the 12 Lead and Arrhythmia Modes, it is always enabled.
When electrodes are not properly applied, these measures may not fully compensate for artifact. High polarization voltages induced by electrodes applied without conductive gel may cause the amplifier to overrange, so that a straight line will be recorded instead of the ECG (see figure). In this situation the device will automatically block and return the baseline to its normal position. A baseline is then recorded for approximately 1 second. It is possible to block the amplifiers manually by disconnecting the R electrode.
approximately 1 second
013
Sample Recording
On the display th is condit io n i s i ndi cated b y **** i ns te ad of t he el ectrode label).
4-8 MAC 1200 Revision B
2012250-022
Remedy
Preparing for ECG Recording: Electrode Application
Apply the electrodes according to instructions.
Do not apply the electrodes on top of clothing.
Use a contact agent (e.g. mois tened el ectrod e paper, ele ctrode cream , spray, etc.).
Wait approximately 10 seconds before initiating a recording. After the 10-second period, the automatic functions are enabled and the polarization volta ges have stabilized, provided the electrodes are properly applied. In case of improper electrode application, an error message will appear on the display (RL, LL, LA, LL, V1 to V6).
If required, the ADS (cubic spline) and the filters (20/40 Hz, 50 Hz) can be disabled to verify the “raw” ECG signal.
Revision B MAC 1200 4-9
2012250-022
Preparing for ECG Recording: Entering Patient Data
pat
info

Entering Patient Data

It is possible to enter patient data and have them annotat ed on the recording for easy archiving of patient records.
Press to enter the patient data mode.
The recorder displays the menu items in a defined order. In the configuration menu (see “Pat ient Data Menu Setup” on page 9-27 for details) you determine the items to be included in the menu (In the table on the next page, the items that appear in the patient data menu in the default configurat ion are marked as Yes in the Menu item displayed column, the other menu items are marked as No.
To skip a menu item, press or the cursor key or . It is not possible to write capital and small letters (do not press
the Shift key). For entry of numbers (e.g. date of birth), it is not necessary to press the Shift key.
For fields that allow alphanumeric e ntries the NumLock function can be enabled with (the symbol appears in the upper
format/
speed
right corner of the display).
All entries must be confirmed with .
Press or to exit the patient data mode.
info
pat
127(
Please refer to the Appendix for instructions on entering special characters.
4-10 MAC 1200 Revision B
2012250-022
Preparing for ECG Recording: Entering Patient Data
The table below shows the menu items in the correct order. On the display, selected options are shown in brackets. After the table each Parameter is explained in detail.
Table 1. Patient Data Entry Menu
Parameter
Adjusted Menu Item Displayed
Factory Default
Options
New Patient No Yes Yes Last name Yes First name Yes Date of Birth 00.00.0000(dd.mm.yyyy) Yes Patient ID Yes
Chest pain
1
Unknown Yes Secondary Complaint, Chief Complaint
Pacemaker No Yes Yes Gender - Yes female, male Height Yes Weight Yes Race unknown Yes other Systolic BP 0 mmHg Yes Diastolic BP 0 mmHg Yes Ordering Physician Yes selection from a list of 10 names Referring Physician Yes selection from a list of 10 names Technician Yes selection from a list of 10 names Phone No. -- Yes Medication Yes
1. unknown Yes other
2. unknown Yes other Comments Yes ID Required Yes Patient ID Length Yes 3 to 16 characters; default is 9 characters Secondary ID No Secondary ID Required No Last Name Required No First Name Required No Location # No 1 ... 600 Room No Order number No Prompt 1 No
Revision B MAC 1200 4-11
2012250-022
Preparing for ECG Recording: Entering Patient Data
Parameter
Prompt 2 No Prompt 3 No Prompt 4 No
1. with 12SL program only. This will appear only if the ACI-TIPI option is enabled in the HARDWARETESTS menu. Refer to the MAC 1200/1200 ST Service Manual for more information on enabling this option.

New Patient

Last Name, First Name

Table 1. Patient Data Entry Menu
Factory Default
Adjusted Menu Item Displayed
Yes: existing patient data are deleted. No: entered data can be edited.
Last name: 16 characters. First name: 10 characters.
Options

Date of Birth

Patient ID

You do not need to enter separators between date, month, and year fields.
format/
The button has the function of the Shift-Lock key when entering
speed
data in alphanumeric text boxes. This function allows you to enter the characters shown in the upper part of the keys, without pressing the Shift key. The symbol appears in the display to indicate the Shift­Lock status.
This field accepts 3 to 16 characters. The exact length is determined in the patient data setup menu.
127(
When entering a patient ID whi ch consists of numerals only, the blanks preceding the number are replaced with 0. Example: If a 6­digit text box is configured and you enter the patient ID 123, the final ID number will read “000123”.
4-12 MAC 1200 Revision B
2012250-022

Chest Pain

Pacemaker

Preparing for ECG Recording: Entering Patient Data
127(
The chest pain option is for use with 12SL program only. This will appear only if the ACI-TIPI option is enabled in the HARDWARETESTS menu.
The entry for this menu item is pass ed on to th e 12SL AC I-TIPI program which considers it in the test interpretation. If you choose one of the options and, additionally, enter the patient’s gender and date of birth, and whether the chief complaint is of chest pain, the program will determine a percentage value indicating the probability of acute ischemia. This value alon g with a reason for this conclusion will appear in the interpretation.
Influences the identification of pacer pulses in Arrhythmia Mode. Enable the function Yes when recording the ECG of a pace maker patient. The recording will then be annotated with the messa g e Pacemaker Patient.

Gender/Race

These parameters influence the ECG. If you do not intend to enter all patient data, select the neutral entries “-” and unknown.

Height/Weight

Enter the patient's height (cm) and weight (k g). The weight can be entered with one decimal place.

Systolic BP/Diastolic BP

Enter the blood pressure readings in mmHg.
Revision B MAC 1200 4-13
2012250-022
Preparing for ECG Recording: Entering Patient Data

Ordering Physician / Referring Physician / Technician

When you choose Yes for New patient, the de fault names entered in the System Setup will appear here. When you choose other, you can pick a name from the list. It is also possible to choose none. You can press
to quit the list. The Referring Physician is only relevant if you send ECGs to the MUSE CV system. This name will not be annotated on the ECG recording.

Phone Number

Enter the patient’s telephone number.

Medication

Enter the patient’s medications and confirm entries with .

Comments

ID Required

Secondary ID

4 lines of 30 characters each
Default value is Yes.
This field accepts 3 to 16 characters. The exact length is determined in the patient data setup menu.
If Secondary ID Required is enabled, then Secondary ID must also be enabled.
4-14 MAC 1200 Revision B
2012250-022
Preparing for ECG Recording: Entering Patient Data

Secondary ID Required

Yes: patient secondary ID required. No: patient secondary ID not required.
If Secondary ID is enabled, then Secondary ID Required must also be enabled.

Last/First Name Required

Yes: patient last or first name required. No: patient last or first name not required.

Location Number

ID number for the location (3-place). When you select Yes for New patient, the default value from the System Setup will auto matically be
adopted, but you can overwri te this number.

Room

Order Number

Prompts

5 digits
9 digits for entry of the ECG order number.
Answer the prompts entered in the patient data setup menu. (See
“Patient Data Menu Setup” on page 9-27 for details).
Revision B MAC 1200 4-15
2012250-022
For your notes
Preparing for ECG Recording: Entering Patient Data
4-16 MAC 1200 Revision B
2012250-022
5 Recording in 12 Lead
Mode
Revision B MAC 1200 5-1
2012250-022
For your notes
5-2 MAC 1200 Revision B
2012250-022
Recording in 12 Lead Mode: General Information

General Information

In 12 Lead mode, 12 leads of ECG are acquired simultaneously for a period of 10 seconds. When initiated with , the recording proceeds automatically.
Depending on the implemented software options, the ECG is:
Units equipped with the optional memory function MEMO (MAC 1200 only) can save up to 40 resting ECG. These ECGs can be printed or sent to CardioSys/CardioSoft (A5 protocol) or to the MUSE CV system (CSI protocol) (see “The Storage Program” on page 5-8).
The unit offers different report formats for printout of t he ECG. With the system defaults, all 12 leads including the measurement and analysis results will be documented on a single sheet. (See “Report Formats” on page 5-10.)
only printed out (MAC 1200 without options MEAS, DIAG), measured and printed out with the measurement results (MAC 1200
with option MEAS), and measured, interpreted (analyzed) and printed out with the
interpretative statements (MAC 1200/1200 ST with option DIAG).
Some of the system settings can be customized. They are identified with “configurable.”
Revision B MAC 1200 5-3
2012250-022
Recording in 12 Lead Mode: General Information
The following information refers to a unit with the system defaults. For instructions on changing the default setup, see “12 Lead Mode” on page 9-5.
Table 1. Setup Menu for the 12 Lead Mode
Parameter Factory Default Options
Report Sequence STANDARD CABRERA Rhythm Leads II, V1, V5 I, III, aVR, aVL, aVF, V2, V3, V4,
V6 Gain 10 mm/mV “*auto”, 5, 20, 40 mm/mV Report Format 4x2.5R1 1x10R12, 2x5R1, 2x5_50,
4x2.5R1, 1x10R3, 4x2.5R3 Detailed Results No Yes Muscle Filter No Yes Muscle Filter Frequency 40 Hz 20 Hz AC Filter Yes No Manual copy to EKG HOST Number of copies 1 0 to 9 Delete ECG after
No Yes
Transmission Autosave ECG No Yes Use Screening criteria No Yes Suppress NORMAL
No Yes
statement Suppress ABNORMAL
No Yes
statement Interpretation Yes No Print Interpretation No Yes Override function No Yes
5-4 MAC 1200 Revision B
2012250-022

Recording

Recording in 12 Lead Mode: Recording
127(
When an electrode is off the patient or it has failed in some other manner, the unit will record an ECG unless the override function has been enabled in the 12 Lead Setup menu.For more information, see Chapter 9, “12 Lead Mode” .
When the unit is powered up, it defaults to the 12 Lead mode (this default setting can be modified).
1. Before recording the ECG, patient data can be entered ( ). It is
info
pat
recommended that you enter the patient's name so it appears on every report.
2. After applying the electrodes, wait about 10 seconds for the signal to stabilize (stabilization of polarization voltages, see Chapter 4,
“Artifact Due to Poor Electrode Application” ). If you initiate a
recording with immediately after selecting the 12 Lead mode, the message Collecting data displays, and you will need to
wait for 10 to 12 seconds.
Before initiating a recording, check the display for error messages. Check all electrodes; if the message persists, there must be a break in the patient cable. Replace the cable with a new one.
Table 1. Messages Indicating Disconnected Electrodes
Symbol Description
*RL*: Right leg electrode disconnected *RA*: Right arm electrode disconnected *LA*: Left arm electrode disconnected *LL*: Left leg electrode disconnected *V1*: Chest electrode V1 disconnected *V2*: Chest electrode V2 disconnected *V3*: Chest electrode V3 disconnected *V4*: Chest electrode V4 disconnected *V5*: Chest electrode V5 disconnected *V6*: Chest electrode V6 disconnected
Revision B MAC 1200 5-5
2012250-022
Recording in 12 Lead Mode: Recording
3. Initiate a recording with . The MAC 1200 acquires all 12 leads simultaneously for 10 seconds, and then generates a printout.
127(
The device can be set up to allow a recording only when specific patient information has been entered (last name, first name, ID, Secondary ID, for more information, see “Patient Data Menu
Setup” on page 9-27).
When you initiate a recording with , the unit prints the most recent 10 seconds of ECG data and analyzes it. Therefore, wait un til
the patient has been lying relaxed and motionless for about 10 seconds before starting the recording.
127(
Filters may suppress diagnostically relevant portions of the signal, because th ey limit the transmission range. Therefore, only enable filters if necessary.
With the system defaults unchanged, the unit will activate the following functions and settings after power-up (system defaults which can be changed are indicated with “configurable”).
12 Lead mode (configurable)
Standard report sequence (lead to channel correspondence): I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Rhythm leads II, V1 and V5 (configurable)
Gain of 10 mm/mV (configurable) (calibration pulse at the beginning of the recording)
AC filter is on (configurable)
Muscle filter is disabled ( ) (configurable)
AC line filter is enabled (configurable)
ADS (anti-drift system) is always enabled
Report format is 4x2.5R1, i.e. 12 leads and al l data are printed o n one
muscle
filter
page (configurable)
Interpretation is printed (configurable)
Detailed results page (including the median complexes and the measurement results) is not printed (configurable) (MAC 1200 only)
Pressing will produce one copy of the printed ECG
copy
(configurable)
Override function is disabled (configurable)
5-6 MAC 1200 Revision B
2012250-022
Recording in 12 Lead Mode: Recording
MAC 1200 with the MEMO (memo ry) option
ECGs are not automatically stored (configurable)
ECGs that were successfully transmitted to a host system are not automatically removed from memory (configurable)
QTC is calculated with the Bazett formula (only MAC 1200 with option MEAS (measureme nt) or DIA G ( mea sur em ent + i n terp re tati on) ).
The locations of all relevant device settings are shown below:
Operating mode Muscle filter
Report function
2
Failure message
1
AC line filter1Anti-drift system1“Patient Name”
Gain Report sequence
Three leads
display here
3
1. if enabled
2. or REC OFF when the recording function is disabled
3. if any
Heart rate
The display shows three leads at a time. With you can display all
lead
leads of the report sequence in groups of three.
The recording can be stopped with .
For a description of the different reports, see “Report Formats” on page 5-10.
Revision B MAC 1200 5-7
2012250-022
Recording in 12 Lead Mode: The Storage Program

The Storage Program

With MAC 1200 units equipped with the optional storage program MEMO the ECG including patient, measurement and analysis data can
be saved with the button after ECG acquisition. A message informs the user that ECGs are being saved and indicates the number of stored
ECGs.
store/
retrieve
To retrieve an ECG from memory, simultaneously press and .
store/
retrieve
127(
With a fully charged battery and the unit turned off, ECGs will remain stored for approximately 4 weeks.
The function keys appear at the top of the menu:
Print (prints the selected ECG)
Send (see “ECG Transmission” on page 5-12)
Delete (deletes the selected ECG)
Change (enables modification of the patient data)
When you view a saved ECG, the Print function is selected. The list on the next page shows all patients for whom ECG are stored.
To perform an action with one or more ECGs, you must first select the ECG(s). Follow these steps:
1. Move the cursor down with . When the cursor reaches the Print directory, it moves on to the patient list. In the list, the bar cursor
changes to a line cursor.
2. To select multiple ECGs, select them with , while holding
down. Selected ECGs are marked with a black bar in the margin
of the display. Deselect the ECGs with the same key combination.
5-8 MAC 1200 Revision B
2012250-022
Recording in 12 Lead Mode: The Storage Program
127(
Due to the limited size of the display, you cannot view the entire list at a time. These are the columns that follow to the right of the patient name:
Date and time of the ECG recording
S (indicates that the ECG has been sent to another system)
Patient ID
Comments
To display the remaining columns, you can continuously scroll the display with or you can shift the list by the display’s
width with and .
Alt
3. When you have selected all ECGs to view, press the key to return to the function keys.
4. Use the cursor to select the function you need. Initiate it with .
Other Considerations
With the Print directory command, you can print a list of all stored ECGs. The printout includes all columns, except the Comments column.
If you try to save an ECG when the memory is full, a message informs you of the memory status. When you delete an ECG from memory, the new ECG will automatically be saved.
The unit may be set up to automatically save ECGs (without pressing
/
) and automatically remove ECGs from memory that were
C
successfully transmitted to the CardioSys, CardioSoft or MUSE systems.
The storage program can be terminated at any time with .
127(
If you plan to print a large number of stored ECGs, connect the unit to the power line or to check that the battery is fully charged.
When you terminate the storage program with , it is not possible to save the current ECG again.
Revision B MAC 1200 5-9
2012250-022

Report Formats

Recording in 12 Lead Mode: Report Formats
The length and scope of the reports depends on the implemented software (standard, MEAS ( measurement), DIAG (measurement + interpretation)).
The table below shows all of the 14 different report formats available with MAC 1200 units.
Format
4x2.5R1
(default format)
4x2.5R3 4x2.5 s/4x3 10 s/3 25 mm/s Yes Yes 1 4x2.5R3
2x5R1 2x5 s/2x6 10 s/1 25 mm/s Yes Yes 1 2x5R1
2x5_50 2x5 s/2x6 No 50 mm/s Yes Yes 2 2x5_50
1x10R12 10 s/1x12 No 25 mm/s No No 1 1x10R12
1x10R3 10 s/1x3 10 s/3 25 mm/s Yes Yes 1 1x10R3
4x2.5R1
(default format)
4x2.5R3 4x2.5 s/4x3 10 s/3 25 mm/s Yes Yes 1 4x2.5R3
2x5R1 2x5 s/2x6 10 s/1 25 mm/s Yes Yes 1 2x5R1
2x5_50 2x5 s/2x6 No 50 mm/s Yes Yes 2 2x5_50
1x10R12 10 s/1x12 No 25 mm/s No No 1 1x10R12
1x10R3 10 s/1x3 10 s/3 25 mm/s Yes Yes 1 1x10R3
4x2.5R1
(default format)
MUSE1 4x2.5 s/ 4x3 3x10 s / 2 25 mm/s yes yes 1 no
MUSE2
ECG Traces
Length/Leads
4x2.5s/4x3 10 s/1 25 mm/s Yes Yes 1 4x2.5R1
4x2.5s/4x3 10 s/1 25 mm/s Yes Yes 1 4x2.5R1
4x2.5s/4x3 10 s/1 25 mm/s Yes Yes 1 4x2.5R1
2
4x2.5 s/ 4x3 3x10 s / 2 25 mm/s yes yes 1 no
Rhythm Lead Length/Leads
Speed
Measurement
1
Interpretation* Pages Medians
(default format)
(default format)
(default format)
1. Measurement results and interpretative statements are only available from MAC 1200 with the appropriate software options
2. MUSE2 report format is only available on Pharma devices.
127(
Heart rate is calculated from all beats of the 10-second ECG. Printed reports are unconfirmed documents. They must be overread,
verified, and signed by a physic ian for confirmation.
5-10 MAC 1200 Revision B
2012250-022

Detailed Results

Recording in 12 Lead Mode: Report Formats
The MAC 1200 (with option MEAS or DIAG) setup menu allows you to choose the Detailed Results page. When selected, this page will be appended to the report. It contains patient data, measurement results, interpretative statements (rea sons are only annotated if the unit is equipped with the HEART interpretation program), medians and the tabular measurement values.
127(
To obtain a printout of the full patient data, select the 6 Lead mode and press .
copy
Revision B MAC 1200 5-11
2012250-022
Recording in 12 Lead Mode: ECG Transmission

ECG Transmission

127(
Resting ECGs acquired in 12 Lead mode can be transmitted to CardioSys/CardioSoft or to a MUSE CV system. The units can either communicate via modem or directly via a connection cable (see “Direct
Transmission” on page 5-16).

General Considerations

Observe the safety information given in Chapter 3, “Connecting
Peripheral Equipment” .
The status of transmission progress is indicated on the display with message text, e.g., “Transmitting 1 of 1,” or “Transmitting 2 of 6.”
With the transmission can be stopped. ECGs that were successfully transmitted are identified with the
letter S (for “Sent”). After the selected ECG records have been successf ully transmitted, a
confirmation message displays, e.g. “6 of 6 ECG records successfully
transmitted.”
All ECG records are time stamped (hours, minutes, and seconds). To

Transmission via Modem

Depending on the modem mo del u sed, th e m od em MUST be co nnect ed either with the 9-pole cable 223 378 01 or with the 25-pole cable 223 378
02. For transmission of the ECG, the unit must be set up as described in
“ECG Transmission via Modem” on page 9-3 0.
view a record’s time stamp, highlight it. Then press + . Use
to scroll right to see the time stamp.
5-12 MAC 1200 Revision B
2012250-022
Recording in 12 Lead Mode: ECG Transmission
1. After acquiring the ECG, start the transmission with .
copy
127(
Manual copy to: HOST must be enabled in the 12 Lead setup menu.
The MAC 1200 is also capable of transmitting stored ECGs (if memory option MEMO is installed).
Activate the storage program by simultaneously pressing and
store/
. (Press the button first and hold it depressed.)
retrieve
2. To select one or more ECGs for transfer, move the cursor down with .
When the cursor reaches Print directory, it moves on to the patient list. In the list, the bar cursor changes to a line cursor.
To select multiple ECGs, select them with , while holding
down. Selected ECGs are marked with a black bar in the margin
of the display. Deselect ECGs with the same key combination.
3. When you have selected the ECG(s) to transfer, press . You will
be returned to the function key menu.
4. Using the cursor, select the Send command. Confirm your choice
with . The transmission menu will be displayed, showing the phone number
and the options Start transmission, Modify settings, and Cancel.
5. Check the displayed telephone number and press to initiate the
transfer. If it is necessary to change the number, press to display the
setup
configuration menu.
6. As soon as you initiate the transmission with , the unit
automatically dials the number of the modem at the receiving end and establishes a connection.
Revision B MAC 1200 5-13
2012250-022
Recording in 12 Lead Mode: ECG Transmission
The message Initializing transmission is displayed, and the option to cancel the transmission is provided. Then it will send the ECG.
7. After the transmission, a message on the display indicates the
number of successfully transmitted ECGs. As soon as you acknowledge the message with , the 12 Lead mode acquisition screen appears.
8. If the ECG could not be transmitted (wrong modem setup, modem
off, etc.), the unit will display an error message, such as
Transmission Error! (A5).
In this situation you have the following choices:
Repeat the transmission with .
Error Messages
Change the settings with .
Stop the transmission with .
setup
You must select one of these choices to clear the transmission error message from the display.
127(
Unsuccessful transmission does not remove/delete the ECG record from the local database.
Transmission error messages are listed below.
ECG transmission error! (A5)/(CSI) (depending on selected protocol)
Check interface!
Dial locked! (temporarily)
No dial tone!
Busy!
No answer!
No carrier!
Check modem configuration!
5-14 MAC 1200 Revision B
2012250-022

Batch Transmission

Recording in 12 Lead Mode: ECG Transmission
The MAC 1200 unit can send all untransmitted records in a batch transmission.
1. Activate the storage program by simultaneously pressing and
store/
. (Press the button first and hold it depressed.)
retrieve
2. Select Send All Unsent Records. Confirm your choice with .
The message Initializing transmission is displayed, and the option to cancel the transmissi on is provided.
Then it sends batch of ECG records.
3. After the transmission, a message on the display indicates the
number of successfully transmitted ECGs. As soon as you acknowledge the message with , the 12 Lead mode acquisition screen appears.

Transmission Log

The MAC 1200 unit can be set up to automatically print a log of the recently completed transmission. See “Transmission Log” on page 9-24 for more information.
If the Transmission log is enabled, after tran smission is complete, the message Print transmission log? will display.
Press to print the log.
The message Printing Transmission Log displays as the log is printing.
Revision B MAC 1200 5-15
2012250-022
Recording in 12 Lead Mode: ECG Transmission

Transmitting Data to a MUSE CV System Via Modem

127(
Pacemaker information, telephone number and comments entered in the patient data are not transmitted to the MUSE CV system.
Before sending data to the MUSE CV system, the MAC 1200 automaticall y logs on to th e datab ase. Th en the d ata tra nsfer is ini tiated. If the transfer is stopped, the MAC 1200 takes a few seconds before cancelling the connection because it has to log off the database first. Then the communication link with the receiving modem is interrupted and the standard display reappears.
ECG Record Serial Number
Every ECG record sent to the MUSE CV system contains a unique serial number for tracking purposes. It is made up of the MAC 1200 serial number + date + time.
127(
The ECG record serial number is printed on the transmission log only (if enabled). It d oes no t a pp ear on th e ECG report p rint ed a t the MAC 1200 unit. See “Transmission Log” on page 9-24 for more information.
Direct Transmission
The unit must be connected to the PC or to the MUSE CV system by means of the connection cable (pn 223 362 03).
For transmission of the ECG, the unit must be set up as described in
“Direct ECG Transmission” on page 9-31.
1. After acquisition of the ECG, the transmission is started with .
copy
127(
Manual copy to: HOST must be enabled in the 12 Lead setup menu.
The MAC 1200/1200 ST is also capable of transmitting stored ECGs (if the memory option MEMO is installed).
Activate the storage program by simultaneously pressing and
store/
. (Press the button first and hold it depressed.)
retrieve
5-16 MAC 1200 Revision B
2012250-022
Recording in 12 Lead Mode: ECG Transmission
2. To select one or more ECGs for transfer, move the cursor down with . When the cursor reaches Print directory, it moves on to the
patient list. In the list, the bar cursor changes to a line cursor. To select multiple ECGs, select them with , while holding down . Selected ECGs are marked with a black bar in the margin
of the display. Deselect the ECGs with the same key combination.
3. When you have selected the ECG(s) to transfer, press and you
are back in the function key menu.
4. Using the cursor, select the Send command. Confirm your choice
with . The message Initializing transmission is displayed, and the option to
cancel the transmissi on is provided. Then it starts. The message ECG Transmission (A5) is displayed.
5. After the transmission, a message on the display indicates the
number of successfully transmitted ECGs. As soon as you acknowledge the message with , the 12 Lead mode acquisition screen appears.
Revision B MAC 1200 5-17
2012250-022
Recording in 12 Lead Mode: ECG Transmission
6. If the ECG could not be transmitted (e.g. wrong baud rate,
connection error), the unit will display the error message
Transmission Error! (A5)/(CSI).
The message depends on the s elected protocol.
127(
Pacemaker information, telephone number and comments entered in the patient data are not transmitted to the MUSE CV system.
In this situation you have the following choices:
Repeat the transmission with .
Change the settings with .
Stop the transmission with .
You must select one of these choices to clear the transmission error message from the display.
127(
Unsuccessful transmission does not remove/delete the ECG record from the local database.
5-18 MAC 1200 Revision B
2012250-022

Receiving Data

Recording in 12 Lead Mode: ECG Transmission
See Chapter 13, “Technical Specifications” for additio nal informat io n. Receiving ECGs is only possible with MAC 1200 units. The units must be
in the 12 Lead mode.
1. Use the key combination and to display the screen for
copy
receiving ECGs. The connected modem is automatically initialized. The procedure can be aborted wi th .
2. Press to enable the receive data mode.
3. When you have enabled the receive data mode, the standard screen
display of the 12 Lead mode displays. The message 12 Lead (REC) indicates that the unit is ready to receive data.
4. A message displays on th e screen when the unit is receiving data.
The reception of data can be aborted with .
5. The ECG which has just been received is processed for the printout.
The report is printed in the selected format. Multiple ECGs are received and printed one after the other.
After printout of the last ECG, the receive data mode is automatically disabled. The mode is also disabled when you select another operating mode .
The following information is annotated in the bottom line of each report.
Sender
Software version and analysis program version used at the sending unit (e.g. ACQ-DEV: V5.1M12i HEART V5.1)
127(
Received records are printed by the MAC 1200, but are not stored by the MAC 1200.
Revision B MAC 1200 5-19
2012250-022
Recording in 12 Lead Mode: ECG Transmission

Cart to Cart Communication

Via modem, ECG data can be transmitted between two MAC 1200 units or between a MAC 1200 and any ECG recorder using the CSI protocol (see “Transmission via Modem” on page 5-12 and “Transmitting Data to
a MUSE CV System Via Modem” on page 5-16).
Modem Setup (for Modem Other)
If you prefer to use another modem than the standard models listed in the Setup menu (MultiTech, Elsa), you will have to enter a few parameters required for communication between the MAC 1200 and the modem.
For the AT commands which your modem understands, please refer to the modem user instructions. Three command sequences have to be entered, each of which d efines a specific modem operating state.
1. The modem is initialized (init string).
2. A communication link is established (dial string).
3. The communication is terminated (hangup string).
These three strin gs are entered in the modem setup menu. (See Chapter 3, “Connecting Peripheral Equipment” )
The example below shows the command strings for the MultiTech ZDX modem.
AT Command for Modem Initialization
Symbol Description
AT Prefix that precedes every command line &F Fetch factory configuration (loads the factory configuration from ROM into the
active configuration memory (RAM))
M1 Speaker is always on
X3 Call progress signal monitoring enabled
S0=1 Auto answer after one ring
V0 Digit result codes selected (0 to 999)
init string: AT&FM1X3S0=1V0
5-20 MAC 1200 Revision B
2012250-022
Recording in 12 Lead Mode: ECG Transmission
AT Command for Establishing a Communication Link
The following is an example of a dial string for a modem connected to a branch (PBX system) and dialing a modem via the public telephone network, using the touch tone mode.
Symbol Description
AT Pr efix that prec edes every command line DT Touch tone dial mode xxx After DT, enter the characters for access to the public telephone network
(e.g. 0)
W W, placed after a number, tells the modem in a PBX system to wait for the
dial tone of an outside telephone line
dial string: ATDTOW
AT Command for Termination of the Communication
The communication is terminated in two steps. First, the MAC 1200 sends an escape command to return from the on-
line state to the command state. Then the hangup command follows.
Symbol Description
+++ Escape command AT Prefix that precedes every command line H Hangup command hangup string: +++ATH
Revision B MAC 1200 5-21
2012250-022
Recording in 12 Lead Mode: Adjusting Measurement Points/QT Dispersion
A

Adjusting Measurement Points/QT Dispersion

This feature is only available with MAC 1200 units equipped with the MEAS or DIAG option. The HEART interpretation program is also required.

Global Measurement Points

After acquisition of an ECG in the 12 Lead mode, the global measurement points for P onset, P offset, QRS onset, QRS offset, and T offset can be adjusted manually.
4
After acquisition of the ECG, press to display the screen for verification of the global measurement point markers.
123 4 5
R
$
1 mV
Verification of the Global Measurement Points
1 Amplitude in [mV] 2 Selected lead 3 Active marker (large) 4 Inactive marker (small) 5 Time in [s]
1 s
028
Description
5-22 MAC 1200 Revision B
2012250-022
Recording in 12 Lead Mode: Adjusting Measurement Points/QT Dispersion
On this displa y you wi ll see all 12 ECG lead s; the ac tive le ad is bl ack an d displayed in the foregr ound, the inactive leads appear dimmed in the background.
The active measurement point marker is large, the four inactive markers are small. Use these keys to adjust the markers.
moves the active marker right or left
selects the next or previous marker
lead
activates the next lead
changes the gai n
terminates the adjustment, saving the changes
terminates the adjustment without saving the changes
space bar for the P onset and P offset values, th e space bar toggles
between definite value and approximate value.
Revision B MAC 1200 5-23
2012250-022
Recording in 12 Lead Mode: Adjusting Measurement Points/QT Dispersion

Local T Offset Measurement Point/QT Dispersion

When you exit the screen for verification of the global measurement points, the screen for verification of the T offset measurement point appears automatically.
12 3 4
1 mV
1 s
Verification of the T Offset Measurement Point
Description
1 Amplitude in [mV] 2 Selected lead 3 Time in [s] 4 Active marker
Therefore, this screen always shows only one lead at a time and the T offset point. Changing the local T offset point also affects the QT dispersion value.
029A
5-24 MAC 1200 Revision B
2012250-022
Recording in 12 Lead Mode: Adjusting Measurement Points/QT Dispersion
Use these keys to adjust the marker.
moves the marker right or left
lead
displays the next lead
changes the gai n
mm/mV
terminates the adjustment, saving the changes
terminates the adjustment without saving the changes
When you exit the screen, the acquisition screen for the 12 Lead mode reappears.
The corrected ECG can be printed with the button.
lead
If the unit is equipped with the MEMO memory option, the co rrected data can be saved (or it will be saved automatically when the corresponding function i s enab led). If the ori gin al ECG had alrea dy be en saved, the corrected data will overwrite this ECG.
127(
Changing the local T offset measurement point does not affect the global T offset point.
Revision B MAC 1200 5-25
2012250-022
Recording in 12 Lead Mode: Brief Operating Instructions – 12 Lead Mode

Brief Operating Instructions – 12 Lead Mode

Switch on the unit and wait for self-test to end.
Apply electrodes to patient.
Enter the following patient data .
Check device settings.
Report sequence
Report for m at
AC filter
Override function
12SL interpretation configuration
Modify device settings, if required .
Wait for patient to lie motionless and for the unit to collect 10
info
pat
setup
seconds of ECG data.
Check that no lead failure message is displayed.
Start recording with .
5-26 MAC 1200 Revision B
2012250-022
6 Recording in 6 Lead
Mode
Revision B MAC 1200 6-1
2012250-022
For your notes
6-2 MAC 1200 Revision B
2012250-022
Recording in 6 Lead Mode: General Information

General Information

In 6 Lead mode, the system acquires six leads of ECG in real time. Recordings are started and stopped with . Some of the system settings can be customized. They are identified with “configurable”.
The following information refers to a unit with the system defaults. For instructions on changing the default setup, refer to “6 Lead Mode” on page 9-9 for details.
Report Sequence STANDARD CABRERA, SEQ.NO.4 Gain 10 mm/mV “*auto”, 5, 20, 40 mm/mV Speed 25 mm/s 5, 50 mm/s Muscle Filter No Yes
Table 1. Setup Menu for the 6 Lead Mode
Parameter Factory Default Options
Muscle Filter Frequency
AC Filter Yes No Anti-Drift System No Yes Automatic Paper Feed Yes No
40 Hz 20 Hz
Revision B MAC 1200 6-3
2012250-022

Recording

Recording in 6 Lead Mode: Recording
127(
In 6 Lead mode, messages indicating disconnected electrodes are also annotated on the recording, e.g. Lead fail C1.
After switching on the unit, press to select 6 Lead mode.
6
Before recording the ECG, patient data can be entered ( ). It is
info
pat
recommended that you enter the patient’s name to annotate it on every report.
Before initiating a recording, check the display for error messages (see table below). Check all electrodes; if the mess age persi sts, there must be a break in the patient cable. Replace the cable with a new one.
The recording is started and stopped with .
Table 2. Messages Indicating Disconnected Electrodes
*RL*: Right leg electrode disconnected *RA*: Right arm electrode disconnected *LA*: Left arm electrode disconnected *LL*: Left leg electrode disconnected *V1*: Chest electrode V1 disconnected *V2*: Chest electrode V2 disconnected *V3*: Chest electrode V3 disconnected *V4*: Chest electrode V4 disconnected *V5*: Chest electrode V5 disconnected *V6*: Chest electrode V6 disconnected *RL*: Right leg electrode disconnected *RA*: Right arm electrode disconnected
6-4 MAC 1200 Revision B
2012250-022
Recording in 6 Lead Mode: Recording
format/
speed
127(
In 6 Lead mode, report sequences can also be selected with these shortcuts.
= Standard
= CABRERA
= Seq. 4
With the system defaults, the MAC 1200 will activate the following functions and settings.
The Standard report sequence (lead to channel correspondence) : I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 (configurable); also available: CABRERA, SEQ. NR. 4 (custom report sequence).
A gain of 10 mm/mV (configurable) (calibration pulse at the beginning of the recording The unit can be set up to automatically adapt the gain to the ECG signal (see “6 Lead Mode” on page9-9).
Also, the ga in setting can be changed with (5, 10, 20 and 40
gain
mm/mV).
127(
Please note that filters may suppress diagnostically relevant portions of the signal, because they limit the transmission range. Filters should therefore only be enabled if necessary.
lead
selects the other 3 leads of the selected report sequence.
display the other 3 leads of the selected group of 6.
The writer prints at a speed of 25 mm/s, the speed can be changed with .
The muscle filter is disabled (configurable).
The AC line filter is enabled (configurable).
The anti-drift system (cubic spline) is disabled (configurable).
Pressing will not advance the paper to the next fold (Auto Paper Feed) (configurable).
Pressing after the ECG recording will print the patient data.
The unit advances the paper to the beginning of a new page each
copy
time a recording is initiated (configurable).
Revision B MAC 1200 6-5
2012250-022
Recording in 6 Lead Mode: Recording
lead
The locations of all relevant device settings are shown below:
Operating mode Muscle filter
Report function
2
Failure message
1
AC line filter1Anti-drift system1“Patient Name”
Gain Report sequence
Three leads
display here
3
1. if enabled
2. or REC OFF when the recording function is disabled
3. if any
Heart rate
If you change the writer speed, lea d group or any filter settings during a recording, the unit will briefly stop.
With you advance to the next group of 6 leads of the selected report sequence.
When the anti-drift system is enabled, there will be a short delay of
2.2 seconds before the recording starts.
The heart rate limit is automatically calculated from the date of birth (WHO 100% = 220 - age). When the date of birth is not entered, the unit
will set the limit at 180 bpm. This value can be changed with and
(in steps of 5 bpm).
6-6 MAC 1200 Revision B
2012250-022
7 Arrhythmia Mode
Recording
Revision B MAC 1200 7-1
2012250-022
For your notes
7-2 MAC 1200 Revision B
2012250-022
Arrhythmia Mode Recording: General Information
format/
speed

General Information

In Arrhythmia mode, the MAC 1200 continuously scans the ECG for arrhythmias. From six simultaneously acquired leads, the MAC 1200 automatically selects the two that provide the best signal for analysis.
When the analysis algorithm detects an arrhythmia, the event is recorded with “context” (see figure below). The length of the recording varies with the duration of the event episode. In the setup menu (Chapter 9, “System Setup” ) you determine the conditions for a recording.
The recorder starts each time it detects a single-beat event . The recorder starts each time i t detects an event different from the
previous event. The recorder does not start at all.
GE marquette MAC 1200 John Doe
Event Recording
127(
After starting the program, press to select a continuous
format/
speed
recording at 5 mm/s (c). If the unit identifies an arrhythmia event, it will automatically switch to the faster paper speed. With the same
key , the trend recording can be stopped. The unit can be set up to automatically start a trend recording when the Arrhythmia mode
is initiated.
031A
Revision B MAC 1200 7-3
2012250-022
Arrhythmia Mode Recording: General Information
Some of the system settings can be customize d. They a re labe led wi th “configurable”. The following informatio n refe rs to a unit wi th th e system defaults. For instructions on changing the default setup, refer to “Arrhythmia Mode” on page 9-12.
Table 1. Setup Menu for the Arrhythmia Mode
Parameter Factory Default Options
Report sequence STD_C (chest
leads V1 through V6)
STD_RED (I, II, III, V2, V4, V6) STD_LI, (I, II, III, aVR, aVL, aVF) CABR_LI (aVL, I, -aVR, II, aVF, III)
HIGH_C (V1’ through V6’) Gain 10 mm/mV “*auto”, 5, 20, 40 mm/mV Muscle Filter No Yes Muscle Filter Frequency 40 Hz 20 Hz AC Filter Yes No Trend Rec. No Yes Arrhythmia Data unequal all, no Episodes chron. prio, ventr., no
7-4 MAC 1200 Revision B
2012250-022

Recording

Arrhythmia Mode Recording: Recording
1. After switching on the unit, press to select the Arrhythmia
arrhy
mode. When working with the MAC 1200 ST equipped with the Stress Test
option ERGO, you must press the button twice.
2. Before recording the ECG, patient data can be entered ( ). It is
pat info
recommended that you enter the patient’s name to annotate it on every report.
Before initiating a recording, check the display for error messages. Check all electrodes; if the message persists, there must be a break in the patient cable. Replace the cable with a new one.
3. Start and stop the recording with the key.
Table 2. Messages Indicating Disconnected Electrodes
Symbol Definition
*RL*: right leg electrode disconnected *RA*: right arm electrode disconnected
*LA*: left arm electrode disconnected *LL*: left leg electrode disconnected *V1*: chest electrode V1 disconnected *V2*: chest electrode V2 disconnected *V3*: chest electrode V3 disconnected *V4*: chest electrode V4 disconnected *V5*: chest electrode V5 disconnected
*V6*: chest electrode V6 disconnected *RL*: right leg electrode disconnected *RA*: right arm electrode disconnected
Revision B MAC 1200 7-5
2012250-022

During the Recording

Arrhythmia Mode Recording: Recording
When the program starts, the unit records 6 leads of ECG (1 page ). During the following learn phase, the analysis algorithm learns the patient’s typical QRS complex.
After the learn phase, the recorder prints a report where the QRS complexes acquired in the learn phase are labeled L and the complex found to be the patient’s typical complex is labeled QRSL.
When the learn phase is complete, the MAC 1200 is ready to identify arrhythmias.
127(
With a single-page recording can be initiated after program
copy
start.
127(
Filters may suppress diagnostically relevant portions of the signal, because they limit the transmission range. Only enable filters if necessary.
With the system defaults, the MAC 1200 will activate the following functions and settings.
STD_C report sequence (lead to channel correspondence) (V1 to V6) (configurable).
Gain of 10 mm/mV (configurable) (calibration pulse at the beginning of the recording). The unit can be set up to automatically adapt the gain to the ECG signal (*auto).
Muscle filter is disabled (configurable).
AC line filter is enabled (configurable).
Slow trend recording is disabled (configurable).
Event episodes are recorded at a speed of 25 mm/s.
Unit documents all events that are different from the previous event (configurable). The unit can be configured to document all events or no events at all.
In the final report, the event episodes are printed in chronological order.
7-6 MAC 1200 Revision B
2012250-022
Arrhythmia Mode Recording: Recording
The locations of all relevant device settings are shown below:
Operating mode Muscle filter
Report function
2
Failure message
1
AC line filter1Anti-drift system1“Patient Name”
Gain Report sequence
Three leads
display here
3
1. if enabled
2. or REC OFF when the recording function is disabled
3. if any
Heart rate
Refer to Table 3, “Arrhythmia Codes,” on page 7-9 for an explanation of the arrhythmia codes annotated on the recording.
The heart rate limit is automatically calculated from the date of birth (WHO 100% = 220 - age). When the date of birth is not entered, the unit
will set the limit at 180 bpm. This value can be changed with and
(in steps of 5 bpm).
127(
The letter A on a recording indicates the presence of artifact which does not allow the algorit hm to identi fy arrhyt hmias. Caus es include wandering baselines. The anti-drift system largely prevents these disturbances. Check the electrodes and leadwires.
Revision B MAC 1200 7-7
2012250-022

Final Report

Arrhythmia Mode Recording: Recording
The arrhythmia recording can be stopped with .
Then the final report can be printed with . The final report consists
copy
of the following:
Patient ID sheet (with all patient data as well as with all analyzed QRS complexes, type and number of det ected events and the analys is duration in tabular form)
Episode report (3 sheets maximum with 2 episodes each Arrhythmia codes)
7-8 MAC 1200 Revision B
2012250-022
Loading...