Gefran HME SERIES Specification Sheet

SMART HART MERCURY FILLED MELT PRESSURE
TRANSMITTERS - HME SERIES
CURRENT OUTPUT AND PERFORMANCE LEVEL ‘c’
MAIN FEATURES
• Pressure ranges from: 0-17 to 0-2000 bar / 0-250 to 0-30000 psi
• Accuracy: < ±0.25% FSO (H); < ±0.5% FSO (M)
• Fluid-filled system for temperature stability
• Mercury filling volume:
HME0 (30mm³); HME1, HME2, HME3 (40mm³)
• 1/2-20UNF, M18x1.5 standard threads; other types avai­lable on request
• Autozero function on board / external option
• Standard diaphragm is 15-5 PH stainless steel with GTP+ coating
• 17-7 PH corrugated diaphragm with GTP+ coating for ranges below 100 bar-1500 psi
GTP+ (advanced protection) Coating with high resistance against corrosion, abrasion and high temperature
AUTOZERO FUNCTION
All signal variations in absence of pressure can be elimina­ted by using the Autozero function. This function is activated by closing a magnetic contact loca­ted on the transmitter housing. The procedure is permitted only with pressure at zero. This function can be activited via HART as well.
4...20mA Output
The HME series of Gefran are pressure transmitters with HART communication protocol for using in high tempera­ture environment. The main characteristic of this series is the capability to read temperature of the media up to 400°C. The constructive principle is based on the hydraulic tra­smission of the pressure. The fluid-filled system assures the temperature stability. The physical measure is transformed in a electrical mea­sure by means of strain-gauge technology.
TECHNICAL SPECIFICATIONS
Accuracy (1)
Resolution 16 bit
Measurement range
Rangeability 3:1
Maximum overpressure (without degrading performances)
Measurement principle Extensimetric
Power supply 13...30Vdc
Maximum current absorption
Output signal Full Scale (FSO) 20mA
Zero balance (tollerance ± 0.25% FSO)
Calibration signal
Power supply polarity reverse protection
Compensated temperature range housing
Operating temperature range housing
Storage temperature range housing -40...+125°C
Thermal drift in compensated range: Zero / Calibration / Sensibility
Diaphragm maximum temperature 400°C / 750°F
Zero drift due to change in process temperature (zero)
Standard material in contact with process medium
Thermocouple (model HME2) STD: type “J” (isolated junction)
Protection degree (with 6-pole female connector)
FSO = Full scale output (1) BFSL method (Best Fit Straight Line): includes combined effects of Non-Linearity, Hysteresis and Repeatability
H <±0.25%FSO (100...2000 bar) M <±0.5%FSO (17...2000 bar)
0..17 to 0..2000bar
0..250 to 0..30000psi
2 x FS
1.5 x FS above 1000bar/15000psi
23mA
(40mA with relay optional)
4mA
80% FSO
YES
0...+85°C
-30...+85°C
< 0.02% FSO/°C
< 0.02 bar/°C
Diaphragm:
• 15-5 PH with GTP+ coating
• 17-7 PH corrugated diaphragm with GTP+ coating for ranges <100bar (1500psi) Stem: • 17-4 PH
IP65