Gefran AM Data Sheet

AM
MINIATURE FORCE TRANSDUCER FOR COMPRESSION APPLICATIONS
MECHANICAL DIMENSIONS
Main features
• Range of measurement: from 5 to 20 kN
• Accuracy class: 1%
• All stainless steel construction
• Grade of protection: IP65 (DIN 40050)
• Small size
The AM force transducers series have been designed to mea­sure static and dynamic compression forces. They are particularly suitable for monitoring pounding opera­tions in compression which require a rugged transducer, insensitive to high resonance frequencies caused by non­homogeneous leads in dynamic sequences. The accuracy and the stability are not affected by continuous cycling under harsh conditions even with dynamic loads. The small size of the AM force transducers makes them ideal for retrofitting in existing equipment.
Dimensions mm. (± 0,1)
TECHNICAL DATA
1%
5...20 kN 2mV/V
<± 5% FSO
< ± 1% FSO
< ± 0,2% FSO
< ± 1% FSO
< ± 0,02% FSO°C < ± 0,04% FSO°C
-
350 Ohm
> 10 GOhm
10 V 15 V
-20...+50°C
-20...+60°C
-30...+80°C 130% Ln 100% Ln 150% Ln
> 300% Ln
40% Ln
< 0,2 mm
IP65
4x0,15 / 2 m.
Stainless steel
Accuracy Nominal full scale load (Ln)
Nominal output at FSO Output tolerance at Ln
Combined errors: Non linearity Histeresis, Repeatibility
Creep (after 30 min. at Ln) Zero load out of balance signal Thermal drift in Sensitivity
compensated Zero range Calibration
Nominal bridge resistance Isolation resistance Nominal supply voltage Maximum supply voltage Compensated temperature range Maximum temperature range Storage temperature range
Permitted static load Permitted dynamic load Maximum applicable load Rupture load
Maximum static lateral load Maximum elastic deformation at Ln Grade of protection (DIN40050) Electr. connections screened cable Elastic element material
LOAD INTRODUCTION DIRECTION
BALL R 40
SCREENED CABLE
DTS_AM_0709_ENG
GEFRAN spa reserves the right to make any kind of design or functional modification at any moment without prior notice.
ELECTRICAL CONNECTIONS
CONVERSION TABLE
ORDER CODE
Ex.: AM - KN5U
AM force transducer with range of measurement 0 - 5 kN.
If request, it is possible to sup­ply models with non-standard mechanical and/or electrical features.
Force transducer AM
0 - 20 KN2D
MEASUREMENT RANGE (kN)
0 - 5 0 - 10
KN5U KN1D
Cells connected in parallel
In systems that use several cells, the parallel connection automatically sums the loads on each individual cell. Using this method of measurement, the maximum load will be the sum of the loads on the individual cells and the sensitivity will be the average value of these cells. It is important that the user ensures that no cell is stessed beyond its maximum rating under any load condition.
Kg
1
0.102
0.454
N
9.807 1
4.448
Lb
2.205
0.225 1
OPTIONAL ACCESSORIES
* The screen is isolated from the transducer body. It is recommended that the ground is connected at the instrument end.
4x0.25
Screened cable
SIGNAL
SIGNAL EXCITATION EXCITATION SHIELD
CABLE EXIT
Red Yellow or Black
White Green
Yellow-Green
Shield
White
Green
Red
Yellow
Red Yellow or Black
White Green
Yellow-Green
Red Yellow or Black
White Green
Yellow-Green
Yellow-Green
Red Yellow or Black
White Green
SHUNT BOX