Gefran 40B96 User Guide

40B96
POSITION, FORCE and PRESSURE INDICATOR-INTERCEPTOR with INPUT
for STRAIN-GAUGE and POTENTIOMETER
Software Version 3.2x
code 81652D / Edit. 04 - 04/05
INSTRUCTIONS FOR USE AND WARNINGS
GENERAL INDEX
page
Graphic symbol used 31
1 Preliminary instructions 31
General description 31 Alarm unit in basic version 31 Options 32 Operator interface 32 Electrical interface 32 Preliminary warnings 32
2 Installation and connection 33
Electrical power supply 33 Notes concerning electrical safety and Electromagnetic compatibility 33 Advice for correct installation for EMC 34 Instrument power supply 34 Inputs and outputs connection 34 Dimensions and cut-out 35 Installation with panel mounting 35 Warnings and instructions for mounting to the panel 35 Nominal ambient conditions 35 Electrical connections 36 Example of connection with input from Melt Sensor 38
3 Functions 39
Operator interface 39 General operating notes 40 Navigating through the instrument menu 41
page
4 Configuration and programming 42
Configuration/Programming 42
Application notes 50 Eb function 50 HOLD function 50 FLASH function 50 Alarms 50 Filter 51 UCAL strain-gauge calibration 51 Virtual Instrument 52
5 Technical specifications 53
6 Maintenance 54
Cleaning the instrument 54 Repairs 54 Checking the jumpers 54 Troubleshooting Guide 55
7 Technical-Commercial information 55
Order code 55
Accessories
RS232/TTL interface for Gefran Instrument configuration 56
Appendix 56
The contents of each section are summarized immediately following the section heading
Graphic symbols used
To distinguish between the type and importance of the information provided in these instructions for use, graphic
symbols have been used as a reference to make interpreting the information clearer .
Indicates the contents of the various manual sections,
the general warnings, notes, and other points to
which the reader's attention should be drawn.
Indicates a particularly delicate situation that
could affect the safety and correct working
operation of the instrument, or a rule that must
be strictly observed to avoid dangerous situa-
tions
Indicates a condition of risk for the safety of the
user, due to the presence of dangerous voltages
at the points shown
Indicates a suggestion based on the expe-
rience of the GEFRAN Technical Staff,
which could prove especially useful under
given circumstances
In the programming and configuration flows
for the controller, indicates all the parame-
ters that can be set in the configuration.
AL.2
1 • PRELIMINARY INSTRUCTIONS
This section contains information and warnings of a general nature which should be read befo­re proceeding with instrument installation, configuration and use.
General Description
GEFRAN series 40B96 digital controllers have been designed for temperature control in any applications involving heating or cooling processes. They represent an exclusive combination of performance, reliability and applicational flexibility. In particular, this new line of Gefran temperature controllers is the ideal solution for application in sectors where performance and service continuity are important, including:
• extrusion lines
• presses for rubber
• test benches
• honing machines
• processing plant for the food industry
• weighing
• manostat
• positioner
• motopotentiometer
• etc.
The 40B96 series interceptors are made on an extremely versatile hardware and software platform, that allows the most suitable I/O composition for the plant to be chosen from a series of options, up to a maximum of:
- variable input
- digital input
- transmitter power supply
- analog retransmission output
- a maximum of 4 outputs
- RS485 interface
Alarm unit in basic version
- 1 input that allows connection of most popular sensor
types
• potentiometer with 100 ohm minimum resistance
• load cells with autoranging sensitivity from 1.5 to 3.3 mV/V
• strain-gauge pressure probes accuracy better than 0.2% f.s.
- output for sensor or transmitter power supply
• 1.2 Vdc for potentiometer
• 5Vdc, 10Vdc max 120 mA for strain-gauge
• 15Vdc, 24Vdc max 50 mA for transmitter
- 2 standard outputs: one relay type and the other relay/logic, or
1 triac output
- display functions with setting of engineering scale
and possible linearization, settable decimal point position, sampling time from 30 to 120 msec
- trip point values can be set in a range dif fering from scale limits, different modalities, hysteresis settable in scale points or time, with or without latch.
- service serial line for configuration via PC (Winstrum)
Options
- 3rd relay output
- 4th relay output (alternative to analog retransmission
output)
- 1 digital input
- 1 analog retransmission output in current
0/4…20mA or voltage 0…10V (alternative to 4th relay output)
- RS485 optically isolated serial interface
Operator Interface
All the operator interface devices are concentrated on
the controller faceplate, suitably protected by a
membrane in Lexan that guarantees IP65 level
protection.
• 3 buttons to be used for manual configuration/selection
• 1 red five-digit displays (Process Variable)
• 4 red LEDs for status indication of same number of
relay/logic outputs
Electrical Interface
All connection terminals (power supply, inputs, outputs,
options) are grouped together on the back of the instru-
ment.
For technical specifications and performance details
refer to Section 5 "Technical Specifications”.
Preliminary Warnings
The following preliminary warnings should be read before installing and using the series 40B 96 interceptor . This will allow the controller to be put into service more quickly and will avoid certain problems which may mistakenly be inter­preted as malfunctions or limitations of the inter­ceptor.
• Immediately after unpacking the controller, make a note of the order code and the other identification data given on the label affixed to the outside of the container and copy them to the table below. These details must always be kept close at hand and referred to the personnel involved in the event of help from Gefran Customer Service Assistance.
• Check also that the controller is complete and has not been damaged at all during transit, and that the package contains not only the instrument and these Instructions for Use, but also the two brackets for fixing to the panel and the dust protection seal - see: Installation with Panel Fixing in Section 2.
Any inconsistencies, omissions or evident signs of damage should be reported immediately to your Gefran sales agent.
• Check that the order code corresponds with the configuration requested for the application the controller is needed for, referring to Section 7: “Technical - Commercial Information”.
• No. and Type of Inputs/Outputs available
• Presence of the necessary options and
accessories
• Mains voltage supply
Ex: 40B – 24 – RR – 00 – 1 – 2 – 1
Model 40B96 alarm unit 24 V output for transmitter power supply Output 1 - Relay; Output 2 - Relay, Output 3 - none Digital Input - no analog retransmission RS485 Digital Communication Power supply 100...240Vac/dc
Before installing the series 40B96 instrument on the control panel of the machine or host system, refer to the paragraph "Dimensions and Cut-out" in Section 2 "Installation and Connection”.
• Where configuration by PC is provided for, make sure the interface RS232 cable is available and the CD- ROM containing the WINSTRUM software.
For the order code refer to Section 7 "Technical ­Commercial Information”..
SN: ......................... (Serial no.)
CODE: ......................... (Finished product code)
TYPE: ......................... (Order Code)
SUPPLY: ......................... (Type of electrical power supply)
VERS: ......................... (Software version)
2 • INSTALLATION AND CONNECTION
This section contains the instructions necessary for correct installation of the 40B96 instrument into the machine control panel or the host system and for correct connection of the inter­ceptor power supply, inputs, outputs and interfa­ces.
Before proceeding with installation read the following warnings carefully! Remember that lack of observation of these warnings could lead to problems of electrical safety and electromagnetic compatibility, as well as invalidating the warranty.
Electrical power supply
• the instrument is NOT equipped with an On/Off switch: the user must provide a two-phase disconnecting switch that conforms to the required safety standards (CE mar­king), to cut off the power supply upstream of the instru­ment. The switch must be located in the immediate vicinity of the instrument and must be within easy reach of the ope­rator. One switch may control more than one instrument.
• if the instrument is connected to NOT isolated electri­cal equipment (e.g. thermocouples), the earth connec­tion must be made with a specific conductor to prevent the connection itself from coming directly through the machine structure.
• if the instrument is used in applications with risk of damage to persons, machinery or materials, it is essen­tial to connect it up to auxiliary alarm equipment. It is advisable to make sure that alarm signals are also triggered during normal operation. The instrument must NOT be installed in flammable or explosive environments; it may be connected to equip­ment operating in such atmospheres only by means of appropriate and adequate types of interface, conforming to the applicable safety standards.
Notes Concerning Electrical Safety and Electromagnetic Compatibility:
CE MARKING: EMC Conformity (electromagnetic compatibility)
in accordance with EEC Directive 89/336/CEE modified by Directive 93/68. Series 40B96 interceptor are mainly designed to opera­te in industrial environments, installed on the switch­boards or control panels of productive process machi­nes or plants. As regards electromagnetic compatibility, the strictest generic standards have been adopted, as indicated in the table below. BT Conformity (low tension) in accordance with Directive 73/23/CEE modified by Directive 93/68.
EMC conformity has been tested with the following connections.
Function Cable type Length
Power supply cable 1mm
2
1m
Relay output cables 1mm
2
3.5m
Serial connection wire 0,35mm
2
3.5m
Input 1mm
2
3m
Generic standards, emission standard for residential commer­cial and light industrial environments Emission enclosure Emission AC mains Radiated emission
CEI EN 61000-6-3
CEI EN 61000-6-3 CEI EN 61000-6-3 CEI EN 61326 CISPR 16-2
Gruppo1 Classe B Gruppo1 Classe B Classe B
EMC Emission
EMC Immunity
Generic standards, immunity standard for industrial environ­ments Immunity ESD
Immunity RF interference
Immunity conducted disturbance
Immunity burst
Immunity pulse
Immunity Magnetic fields Voltage dips, short interruptions and voltage immunity tests
CEI EN 61000-6-2
CEI EN 61000-4-2
CEI EN 61000-4-3 /A1
CEI EN 61000-4-6
CEI EN 61000-4-4
CEI EN 61000-4-5
CEI EN 61000-4-8 CEI EN 61000-4-11
4 kV contact discharge level 2 8 kV air discharge level 3 10 V/m amplitude modulated 80 MHz-1 GHz 10 V/m amplitude modulated
1.4 GHz-2 GHz 10 V/m amplitude modulated
0.15 MHz­80 MHz (level 3) 2 kV power line (level 3) 2 kV I/O signal line (level 4) Power line-line 1 kV (level 2) Power line-earth 2 kV (level 3) Signal line-earth 1 kV (level 2) 100 A/m (level 5) 100%U, 70%U, 40%U,
LVD Safety
Safety requirements for electrical equipment for measurement, control and laboratory use
CEI EN 61010-1
Advice for Correct Installation for EMC
Instrument power supply
The power supply to the electronic equipment on the
switchboards must always come directly from an
isolation device with a fuse for the instrument part.
The electronic instruments and electromechanical
power devices such as relays, contactors, solenoid
valves, etc., must always be powered by separate
lines.
When the electronic instrument power supply is
strongly disturbed by the commutation of transistor or
power units or motors, an isolation transformer
should be used for the controllers only, earthing the
screen.
It is essential that the plant has a good earth
connection:
- the voltage between neutral and earth must not be <1V
- the Ohmic resistance must be <6Ω;
If the mains voltage fluctuates strongly, use a voltage
stabilizer.
In the proximity of high frequency generators or arc
welders, use adequate mains filters.
The power supply lines must be separate from the
instrument input and output ones.
Inputs and outputs connection
The externally connected circuits must be doubly
isolated.
To connect the analogue inputs the following is
necessary:
- physically separate the input cables from those of the
power supply, the outputs and the power connections.
- use woven and screened cables, with the screen
earthed in one point only
To connect the regulating and alarm outputs
(contactors, solenoid valves, motors, fans, etc.), fit
RC groups (resistance and condensers in series) in
parallel to the inductive loads that operate in
Alternating Current.
(Note: all the condensers must conform to VDE
(class X2) standards and withstand a voltage of at
least 220V AC. The resistances must be at least 2W).
Fit a 1N4007 diode in parallel with the coil of the
inductive loads that operate in Direct Current.
GEFRAN S.p.A. declines all responsibility for
any damage to persons or property caused
by tampering, neglect, improper use or any
use which does not conform to the characte-
ristics of the instrument and to the indica-
tions given in these Instructions for Use.
Dimensions and cut-out
Installation with panel mounting
As well as the actual instrument and these instructions
for use, the package also contains:
• 2 panel fixing brackets (A)
• 1 protective seal against dust and water spray ( B)
Fit the instrument to the panel as shown in the figure.
Warnings and instructions for mounting to the panel
Instructions for installation category II,
pollution level 2, double isolation.
• only for low power supply: supply from Class 2 or low voltage limited energy source.
• the power supply lines must be separate from the controller input and output ones
• group the instruments together keeping them separate from the powered part of the relay
• do not install high-power remote switches, contactors, relays, thyristor power units (especially the "phase angle" type), motors, etc. in the same switchboard
• avoid dust, humidity, corrosive gasses and heat sources
• do not block the ventilation holes: the working temperature must be between 0...50°C
• surrounding air: 50°C
• use 60/75°C copper (Cu) conductor only, wire size range 2x No 22 - 14AWG, Solid/Stranded
• use terminal tightening torque 0.5N m
Nominal ambient conditions
Before supplying the Controller with power, make sure that the mains voltage is the same as that shown in the last number of the order code.
Ex: 40B96-5 – xx – xx – xx – x – x – 1 = 100..240V ac/dc 40B96-5 – xx – xx – xx – x – x – 0 = 11..27Vac/dc
A
B
A
B
Altitude Up to 2000m Working/storage 0..50°C/-20...70°C temperature Non condensing 20...85% relative humidity
A
108
96
48
44,5
92
105
115
70
10
Electrical Connections
PWR
Outputs Out1 - Out2 - Out3
Inputs
Serial line
Digital inputs
Output Out4 /
Retransm. Output
Always make the connections using cable types suitable for the voltage and current limits given in
Section 5 - Technical Specifications.
If the instrument has faston terminals these must be protected and isolated.
If it has screw terminals, the wires must be attached, at least in pairs
Power Supply
Inputs
Standard: 100...240V ac/dc ±10% Optional: 1 1...27Vac7dc ±10% Power: max 8VA; 50/60 Hz
Potentiometer input
4-wires strain gauge input
PWR
strain-gauge
24V
- IN
+ IN
+
-
-
+
+S
-S
Input for 0...20/4...20mA transmitter
Terminal 2 (IN -) can be connected to terminal 4 in case of 3-wire transmitter
1,2V
+
- IN
+ IN
~ ~
Pot
- E
+ E
- IN
+ IN
+-
Inputs
Strain gauge input 6 wires to connect to Melt transducers
Outputs Out1, Out 2, Out3 User configurable generic outputs
- Relay 5A cosϕ = 1 250Vac/30Vdc
- Triac
20...240Vac, max. 1A ± 10%
Outputs Out 1
- Relay 5A cosϕ = 1 250Vac/30Vdc
- Logic 24V 10V a 20mA
Outputs Out 2
(absent on Out1 and of Triac type)
Outputs Out 3
Outputs Out 4 / Retransmission output User configurable generic outputs
- Relay 5A cosϕ = 1 250Vac/30Vdc
- Analogue 0...10V, 0/4...20mA
- 0/2...10V (S1-ON), 0/4...20mA (S1-OFF) S1 is a jumper on the board for continuous or analogic output
Digital inputs User configurable generic inputs
- Digital input 24V 5mA (Jumpers S1, S2 in position P) or from non-powered terminal
(Jumpers S1, S2 in position N) Hrd Menù configuration: Parameter diG or di2 = +16
Serial line
RS485 isolated serial line
Modbus 2 wires (Standard)
RS485 isolated serial line
For Modbus 4 wires/Cencal configuration see technical sheet of the serial board.
Modbus 4 wires / Cencal
strain-gauge
Brown or Orange
Green
White
Black or Yellow
Red
Blue
* For internal calibration
- Relay 5A cosϕ = 1 250Vac/30Vdc
S1
S1 S2
NP
12
*
13
-
E
+
E
- INS
+ IN
S
CAL
CAL
+-
NO
C
NO +
C -
NO
C
C
-
+
NO
-
+
GND
B (data -)
A (data +)
+
Rx
+
Tx
Example of Connection with Input from Melt Sensor
Analog retransmission of pressure value, logic and relay alarm outputs
Shows the process variable, the menu identification, the parameters identification and the error codes
Increases/Decreases the value of the parameter displayed until the max/min. value is reached.
Held down: progressively increases the speed of increasing/decreasing the value displayed.
Used to move between the various menus and parameters of the instrument. Confirms the value of the current parameter (or parameter edited using
) and selects the next parameter..
Output status indicators
Position where to apply the label with united ingegneristica
3 • FUNCTIONS
This section illustrates the functions and operating modes of the displays, the indicator lights and the
buttons that make up the operator interface of series 1200/1300 controllers. It is therefore an essential
requirement for programming and configuring the controllers correctly.
Operator Interface
ID Symbol Function
General Operating Notes
• Immediately after switching on the instrument carries out a self-diagnostic test.
During the test, all the display segments and the 4 indicator lights will flash,
when finished, enters normal work mode (Level 1) and displays variable value.
Displays the value of the Process Variable.
• By pressing briefly it is possible to see in sequence (and if necessary edit)
the significant values that condition the way the instrument works in Level 1
(Thresholds of interception)
• Keeping pressed down for 3 seconds we enter the
Programming/Configuration menu - see Navigation in the instrument Menus for
further details..
In the event of errors during normal working:
LO process variable < min. scale limit (param. LS in the IN)
KI process variable > max. scale limit (param. KS in the IN)
BR broken probe or input values higher than maximum limits
ER input values lower than minimum limits
EB Probe power supply failure (function enabled via parameters ( (.I. on IN
menu)
To solve the problem, refer to the paragraph: Troubleshooting Guide in Section 6 Maintenance
Switching on and using the intrument
Self-diagnostics
Normal Working - Level 1
Errors while working
399
Navigating through the instrument Menu
Keep pressed down to scroll through the menus in sequence and release it when the required menu appears.
Press to access the parameters of the selected menu.
Level 1 Display
Menu
Alarm point 1 (scale points)
Alarm point 2 (scale points)
Alarm point 3 (scale points)
Alarm point 4 (scale points)
3 sec.
Information display
OK
Jumper S4 on soldered side of CPU board See Section 6 - Maintenance
S4 ON ?
NO
OK
PAS=99 ?
NO
Configuration
Serial Communication
Inputs setting
Outputs setting
Password
Protection Code
Input Linearization
User Calibration
The parameters and menus not significant for a given configuration are NOT displayed
If the keys , are not pressed within approx. 15 seconds, the display returns to level 1
PV process variable
39.9
O.I
o
o
o
.
2
.
3
.
4
I
F
(
F
R
S
n
I
v
0
A
P
P
r
L
n
.
U
(
4 • CONFIGURATION / PROGRAMMING
This section contains the instructions necessary to configure the instrument according to the needs of the application..
Optimal working operation of the 40B96 in the field of applica­tion it is intended for depends largely on correct configuration and programming of the relevant control parameters. The flexibility and high performance level of these instruments is in fact based on the numerous parameters that can be programmed directly by the user with the buttons on the control panel, or transferred from a PC, in the form of configuration files, by means of the RS485 interface available as an option.
Configuration
Access to all the configuration / programming menus and to all the parameters available in configuration, allows every detail of the Controller to be configured, to satisfy all application require­ments.
Correct setting of the parameters involved in the
configuration presumes a high level of knowledge regarding of use problems and techniques, and so it is recommended that these parameters are not chan­ged unless the user is fully aware of the consequen­ces, that could arise form incorrect settings.
It is the user's responsibility to check that the parameters are set correctly before putting the instrument into service, in order to avoid damage to persons or property.
The following pages describe one by one the various menus and show for each parameter the concise description of the function performed, any default values and the range of values that can be set.
Supplementary notes on consulting the Configuration/ Programming pages
To set some particularly complex parameters certain tables or detailed explanatory notes must be referred to. These tables or explanatory notes are shown directly on the right of the page next to the parameter in question.
Application notes
The detailed explanations of certain functioning modes or special techniques the result Gefran's lengthy experience in the field of temperature control are instead contained at the end of the Configuration/Programming Section and can represent a valuable reference tool for the user. Where necessary, in the configuration / programming flows the appropriate referrals to these Application Notes are provided.
Password:
PPAA
When scrolling through the menu (keeping pressed down), following the 0V menu, the word PA appears. Access to the next menus is only possible if the PA = 99, para­meter is set to 99 by pressing .
Having set the value to 99, press and keep it pressed
down to access the next menus.
Protection code:
PPRR
The PR parameter is used to enable or disable the display and/or the possibility of editing certain parameters. For further details refer to the description of the PR parameter in the confi­guration flows.
Jumper S4 on CPU Board
The absence of jumper S4 on the CPU board of the instrument prevents access to all the menus when the hardware configuration of the instrument does not require the pre-set parameters to be changed. This jumper is engaged or disengaged during production and normally does not need to be modified by the final user. For further information, refer to Section 6 - Maintenance.
IIFF
Information display Programming configuration
Software version
This menu provides information on the status and hardware configuration of the instrument (number and type of inputs/out-
puts, software version, etc.).
Instrument code
hrd 1 configuration
hrd 2 configuration
00 11 00 11 11
OUTPUT 1
0 = None 1 = Present
OUTPUT 2
0 = None 1 = Present
OUTPUT 3
0 = None 1 = Present
OUTPUT 4
0 = None 1 = Present
00 22 55 00 11
ANALOG OUTPUT
0 = None 5 = Present
DIGITAL 1 INPUT
0 = None 7 = Present
SERIAL
0 = None 1 = RS 485
((FF
Configuration
This menu makes it possible to configure the interception parameters.
Hysteresis for trip point 1
-999 ... +999 scale points
Hysteresis for trip point 2
-999 ... +999 scale points
Hysteresis for trip point 3
-999 ... +999 scale points
IHysteresis for trip point 4
-999 ... +999 scale points
I
F
U
d
(
o
K
d
K
2
(
F
K.
1
K.
2
K.
3
.
4
K
SSrr
Serial communication
This menu makes it possible to configure the various parameters that control serial communication between the instrument and the supervisor.
Instrument identification code
[0 ... 247]
Serial interface protocol
[0 ... 1]
Parity selection
[0 ... 2]
Select Baudrate
[0 ... 4]
Virtual instrument inputs
[0 ... 63]
Virtual instrument outputs
[0 ... 31]
Virtual instrument user interface
[0 ... 127]
SS..PP
Serial protocol
0 CENCAL Gefran 1 MODBUS RTU
BBAA
Baudrate
0 1200 1 2400 2 4800 3 9600 4 19200
PPtt
Parity
0 No parity 1 Odd 2 Even
Inputs IN PV
oo
.4
oo
.3
oo
.2
oo
.1
Bit 543210
Ex. 0 1 1 0 0 0
Set code 24 on S.I in to manage PV and oo.4 alarm, via serial line
Outputs OUTW OUT4 OUT3 OUT2 OUT1
Bit 4 3 2 1 0
Ex. 1 0 0 1 1
Set code 19 in S.0 to manage OUT1, 2 and W out­puts via serial line
Interf KEYB DIS LED LED LED LED
OUT4 OUT3 OUT2 OUT1
Bit 7654 3 210
Ex 0100 0 001 If you want to control KEYB interface elements and OUT1 LED via serial line, set S.U to 65
R
S
(
o
P
S
b
A
P
t
S.
I
24
S.
0
19
S.
U.
65
IInn
Input settings
This menu makes it possible to configure the parameters for the instrument input signals.
Probe type, signal and main input scale
TTPP
Probe type Signal polarization Main inputs scale limits
0 Potentiometer Positive (ex. 0/1V) -1999/9999 -19990/99990 -1999/28000 1 Potentiometer Positive (ex. 0/1V) Linear custom Linear custom Linear custom 2 Strain gauge Positive (ex. 0/10mV) -1999/9999 -19990/99990 -1999/28000 3 Strain gauge Symmetrical (ex.-10/+10mV) -1999/9999 -19990/99990 -1999/28000
Digital filter on input
[0.0 ... 20.0] sec
If set to "0", the medium filter on the sample value is excluded
Digital filter on input display
[0 ... 9.9] punti scala
Select sampling time (resolution)
0 120ms 120ms > 14bit (campionamento Eb 240ms) 1 120ms 60ms > 14bit; 16000 points (sampling Eb 120msec) 2 60ms 30ms > 13bit; 8000 points (samplingEb 60msec) 3 30ms 15ms > 12bit; 4000 points (sampling Eb 30msec)
+4 to disable filter (average of the last eight values sampled)
Strain gauge only +8 disables Eb (sampling time is halved) N.B.: maximum sampling frequency and minimum intercept time is obtained with code 15 (15 msec,
11bit resolution, filter off)
Maximum limit of main input scale and retransmission analog output
Minimum limit of main input scale and retransmission analog output
Decimal point position for input scale
DP Format
0 xxxxx 1 xxxx.x 2 xxx.xx 3 xx.xxx
+4 to display Lo below value L.S -(k.S - L.S) /256
(only for TP = 0 or 2)
+8 select input scale -19990/99990
(value of least significant digit set to 0)
+16 select input scale -1999/28000
Min... Max value associated with the input selected with the TP parameter
Min... Max value associated with the input selected with the TP parameter
For Strain gauge with Sensor power supply control (Eb) For potentiometer Resolution
n
I
t.
P
1.
G.
F
F
dP
L .S
t
d
0.1
0.5
0
0
K .S
1000
o
Offset correction of main input
[-999 ... +999] scale points
Lower Limit for Alarm trip point setting
L.S ... K.S
Upper Limit for Alarm trip point setting
L.S ... K.S
Select function of digital input
Raise key function (active only in P.V.selection)
Lower key function (active only in P.V.selection)
0 None 1 Zero 2 Hold 3 Flash 4 Max. peak display 5 Min. peak display 6 Delta peak display 7 Peak memory reset 8 Zero + peak memory reset 9 Alarms reset 10 Peak + alarms reset 11 Zero + alarms reset 12 Peak + zero + alarms reset 13 Output status OUT1 /check CAL 14 Output status OUT2 /check CAL 15 Output status OUT3 /check CAL 16 Output statusOUT4 / check CAL 17 KEYLOCK
(d.I. only)
18 access to U.CAL for the selected
input probe type (t.U. and t.d. only)
19 OFFSET ( t.U. and t.d. only)
d. i. - t. u. - t. d.
Note 1 If you configure the "direct access to potentio­meter calibration menu (tP=0 or 1)" function with parameters t.U. = 18 (t.d. = 18), the calibration procedure is as follows:
- keep the raise (lower) key pressed until you see C.L.
- release the key. Mechanically bring the poten­tiometer to minimum stroke position.
- confirm with the F key. C.H. will appear on the display.
- mechanically bring the potentiometer to maxi­mum stroke position.
- keep the F key pressed to exit the procedure and return to normal display
Note 2 If you configure the "direct access to parameter o.F" function (offset correction of display) with parameters t.U. = 19 (t.d.=19), the calibration procedure is as follows:
- keep the raise (lower) key pressed until you see o.F.
- when the F key is released, the display will alternate between o.F and the previously set o.F value.
- set the offset value you want by means of the raise/lower keys.
- keep the F key pressed to exit the procedure
and return to normal display.
d
t
t
L L
F
0
.i
0
.u.
0
.d.
0
0
K .L
1000
00VV
Output settings
This menu makes it possible to configure the parameters of the instrument outputs.
Number of outputs
Alarm type 1 (absolute only)
Alarm type 3
Alarm type 4
Alarm type 2
11..TT
Direct
22..TT
(high limit) Absolute or Relative Normal or
33..TT
Inverse (low limit) to previous Symmetric
44..TT
absolute (window)
0 Direct Absolute Normal 1 Inverse Absolute Normal 2 Direct Relative Normal 3 Inverse Relative Normal 4 Direct Absolute Symmetrical 5 Inverse Absolute Symmetrical 6 Direct Relative Symmetrical 7 Inverse Relative Symmetrical
+8 to disable on power-up until first alarm +16 to memorize +32 to filter with F.O. mode (output filter mode) +64 output for calibration of 6 wire probe (Phase 2)
..FF..00
Functions
0 inactive, calculated state is shown directly on relays 1 delayed activation (DON) 2 delayed activation of output from time output is
deactivated (DBI) 3 delayed deactivation (DOF) 4 delayed activation only when instrument is switched on
(DPO)
0 ... 4
Output filter modalities
+ 8 time basis 99 min max. (default = 99 sec)
Delay relative to F.O
0 ... 99 min
o sec
Minimum output trip time
0 ... 99 sec
Fault action (definition of
output state in case of
broken probe) Er, br
Disabled by setting 0. Displayed if assigned to at least one output
Value Out 1 Out 2 Out 3 Out 4
0 OFF OFF OFF OFF 1 ON OFF OFF OFF 2 OFF ON OFF OFF 3 ON ON OFF OFF
4 OFF OFF ON OFF
5 ON OFF ON OFF 6 OFF ON ON OFF 7 ON ON ON OFF 8 OFF OFF OFF ON
9 ON OFF OFF ON 10 OFF ON OFF ON 11 ON ON OFF ON 12 OFF OFF ON ON 13 ON OFF ON ON 14 OFF ON ON ON 15 ON ON ON ON
Value Direct Absolute or Normal or
(high limit) Relative Symmetric
Inverse to previous (window)
(low limit) absolute 0 Direct Absolute Normal 1 Inverse Absolute Normal 2 Direct Relative Normal 3 Inverse Relative Normal 4 Direct Absolute Symmetrical 5 Inverse Absolute Symmetrical 6 Direct Relative Symmetrical 7 Inverse Relative Symmetrical
v
0
v
0.
.
1
T
0
.
2
T
0
.
3
T
0
.
4
T
0
0
F.
0
r
.
A
0
t.
A
0
r
E
0
PP RR
Protection code
This menu makes it possible to enable/disenable the display and/or modification of specific parameters
PR Displayed Modifiable
0 o1 o2 o3 o4 o1 o2 o3 o4 1 o1 o2 o1 o2 2 o1 o1 3 o1 none
By adding the following figures to the value in the table it is possible to enable a series of supplementary functions: +4: to disable IN, 0V +8: to disable [F, SR +16: to enable maintenance of reset latch at power-off +32: base configuration (the following parameters will not
be displayed: IN: Ft, Fd, oF, L.L,
KK
.l
0V: 0n [forced to no. outputs present], rE Ft, Fd, oF remain at set value L.L,
KK
.l are forced to L.s, KK.s
+64: Virtual instrument. +128: All pages access disabled, except PA (Password)
Protection code
LLNN
Input linearization
This menu makes it possible to carry out custom linearization for the main input. Only if t.P = Linear Custom
Step 0 (beginning of scale value) Display limits: (range within LS...HS)
[-1999 ... 9999 x 4 digit -19999...28000 x 5 digit]
Step 32 (full scale value) Display limits:
[-1999 ... 9999 x 4 digit -19999...28000 x 5 digit]
..........
The “n” step value corresponds to input: mV beginning scale + n* mV mV = (mV full scale - mV beginning scale) / 32
P
r
0
L
n
00
---
32
---
UU..[[
User calibration
This menu makes it possible to carry out user calibration.
UU..[[
Calibration function
1 analog retransmission output 2 potentiometer 3 strain-gauge positive polarization 4 strain-gauge symmetrical polarization
Minimum calibration (*)
Maximum calibration (*)
(*) Press keys ∆∇ to cali- brate the analog output
if
UU..[[
= 1
Zero acquisition phase, potentiometer with cursor in minimum voltage position
Maximum acquisition phase, potentiometer with cursor in maximum voltage position
if
UU..[[
= 2 if
UU..[[
= 3, 4
Tare zero acquisition phase with strain-gauge discharged (no weight or pressure)
Acquisition phase for automatic definition of sensitivity and full scale, load strain-gauge with sample reference or automatic
activation of configured output to check 6-lead sensor (only for version SW 2.0x). 80% f.s. is
set as default. The value can be changed to set the value in engineering units corresponding to the sam­ple used.
Zero recalculation phase, unload strain-gauge by removing sample reference.
P. V.
Process variable
(LEVEL 1)
.......
[[.. LL..
[[.. LL..
[[.. AA
[[.. AA
FFII
FF22
FF33
.
U
(
Application Notes
Eb Function
The standard device is produced with parameter C.I.= 8, with sampling time of 120 msec and Eb function disabled. The Eb function lets you detect if the probe power supply is interrupted. This function is valid for probe currents > 20mA (8mA in version 2.0x)
Example:
- probe voltage 10V
- strain-gauge resistance 350
V 10
- current = = 28mA
R 350
or for three probes connected in parallel
- probe voltage 10V
- strain-gauge resistance 350
V 10 30
- corrente = = = ≅ 85mA
R 350 350
3
HOLD function
The input value and alarms are frozen while the logic input is active. With the logic input active, a reset turns OFF both the relay outputs and the alarms latch.
FLASH function
Input value is sampled; state of alarms is not transferred to outputs; outputs are “frozen”. When the logic input is active the input value is “frozen” and the outputs are updated according to the calculated alarms state, including the ones latched.
Alarms (Interceptions)
time
AL1 + H1
AL2 + H2
AL2
AL1
alarm 1
alarm 2
(*)
For AL1 inverse absolute alarm (min.) with positive H1, 1 t = 1 (*) = OFF if disabling on power-on exists For AL2 direct absolute alarm (max) with negative H2, 2 t = 0
For AL1 inverse absolute, symmetrical alarm with hysteresis H1, 1 t = 5 For AL1 direct absolute, symmetrical alarm with hysteresis H1, 1 t = 4
Normal absolute alarm Symmetrical absolute alarm
inverse
direct
AL1
AL1 + [ H1 ]
AL1 - [ H1 ]
time
For AL1 direct absolute alarm (max) with negative H 1, 1 t = 0 For AL2 direct relative alarm (max) with negative H2, 2 t = 2
For AL1 direct absolute alarm (max) with negative H1, 1 t = 0 For AL2 symmetrical deviation alarm H2, 2 t = 6
time
AL1+AL2
AL1
alarm 1
alarm 2
AL1+AL2
AL1
alarm 1
alarm 2
time
AL1 + AL2 + H2
Normal deviation alarm
(AL1 absolute, AL2 relative)
Symmetrical deviation alarm
(AL1 absolute, AL2 relative)
AL1-AL2
AL1 + H1
AL1+AL2+H2
AL1+H1
Filter - outputs with reference to parameters F.0 and r.A
The diagrams refer to a normal absolute alarm with hysteresis H = 0
UCAL: strain-gauge calibration
a) Positive signal polarization
Suppose we test a probe (load cell) with sensitivity of 2mV/V , powered at 10V. The input signal range is 0 to 20mV. From unloaded to loaded cell you should see a value from 0 to 1000. Set scale limits: L.S. = 0; H.S. = 1000. Calibration procedure: U.C. = 3. Phase F1: unload the cell (equal to applying input voltage of 0mV). W ait 2 seconds for the signal to stabilize. Press key F. Phase F2: load the cell with a sample weight: for example, 80% of total weight (equal to applying input voltage of 80% of 20mV (2mV*10V) = 16mV). On display, set a value equal to 80% of H.S. = 800. Press key F. Phase F3: unload the cell (equal to applying input voltage of 0mV). W ait 2 seconds for the signal to stabilize. Press key F (calibration finished).
b) Suppose we test a probe (force sensor) with sensitivity of 2mV/V , powered at 10V. The input signal range is -20mV to 20mV. From maximum compression to maximum traction you should see a value from -1000 to 1000. Set scale limits: L.S. = -1000; H.S. = 1000. Calibration procedure: U.C. = 4. Phase F1: do not apply any traction or compression to the sensor (equal to applying input voltage of 0mV). Wait 2 seconds for the signal to stabilize. Press key F . Phase F2: apply compression equal to 80% of total compression (equal to applying input voltage of 80% of 20mV (2mV*10V) = 16mV). On display , set a value equal to 80% of H.S. = 800. Press key F. Phase F3: do not apply any traction or compression to the sensor (equal to applying input voltage of 0mV). Wait 2 seconds for the signal to stabilize. Press key F (calibration finished).
time
Alarm
setpoint
Output
Alarm
F.0 = 1
DON = Delayed activation
F.0 = 2
DBI = Delay in turning on output after output is
turned off
time
time
time
F.0 = 3
DOF = Delayed deactivation
F.0 = 4
DP0 = Delayed activation only at power-on
Variable
rA rA
t < rA
Alarm
setpoint
Output
Alarm
Variable
rA
t < rA
t > rA
Alarm
setpoint
Output
Alarm
Variable
Alarm
setpoint
Output
Alarm
Variable
rA
rA rA
rA
Virtual Instrument
Management of the virtual instrument via serial is activated with the Pr parameter . By setting the S.I, S.o and S.U parameters, you can enable writing of a few variables from the serial line to set input values, output state, LED state, and the value shown on displays. You have to enable alarm limits when the write operations are continuous and you don’t have to keep the last value in EPROM (applications requiring continuous variations of interceptions). Enabling the PV input lets you exclude local acquisition of the input, replacing it with the value written in the VALUE_F register. Enabling digital input lets you set the state of this input, for example, to perform the zero setting with writing of the appropriate bit in the V_IN_OUT register. In analog mode, you can set the ON/OFF state of outputs OUT1, ..., OUT4 by writing the bit in the V_IN_OUT register; the value of analog output OUTW in the V ALUE_W register; the state of LEDs OUT1,..., OUT4 by means of the bits in the V_X_LEDS register. * To use the Display resources for setting a value on the display and Keyboard to simulate key pressure, you also have to set bit 7 of the STATUS6_W register (Modbus address 345 / Cencal 690). This bit is reset at every power-up. The following table shows the register addresses:
Parameter bit Resource enabled Image register Format Register
address name
Modbus Cencal
S.I 0 Alarm limit o.1 341 682 word Interception 1
1 Alarm limit o.2 342 684 word Interception 2 2 Alarm limit o.3 343 686 word Interception 3 3 Alarm limit o.4 302 604 word Interception 4 4 PV input 347 694 word VALUE_F 5 IN digital input 344 688 word, bit 4 V_IN_OUT
S.o 0 Output OUT1 344 688 word, bit 0 V_IN_OUT
1 Output OUT2 344 688 word, bit 1 V_IN_OUT 2 Output OUT3 344 688 word, bit 2 V_IN_OUT 3 Output OUT4 344 688 word, bit 3 V_IN_OUT 4 Output OUTW 307 614 word VALUE_W
S.U 0 Led OUT1 351 702 word, bit 0 V_X_LEDS
1 Led OUT2 351 702 word, bit 1 V_X_LEDS 2 Led OUT3 351 702 word, bit 2 V_X_LEDS 3 Led OUT4 351 702 word, bit 3 V_X_LEDS 4 * Display high thousands 321 642 word ADD1 4 * Display high hundreds 322 644 word ADD2 4 * Display high tens 323 646 word ADD3 4 * Display high units 324 648 word ADD4 4 * Display high tens of
thousands 325 650 word ADD5 6 * Keyboard F key 320 640 word, bit 0 NEW_TAST 6 * Keyboard DOWN key 320 640 word, bit 1 NEW_TAST 6 * Keyboard UP key 320 640 word, bit 3 NEW_TAST
5 • TECHNICAL SPECIFICATIONS
This section contains a list of the Technical Specifications for the 40B96 instrument.
Display 5 red digits, digit height 14mm (5 digits) Keys 3 mechanical type (NC, DEC, F) Accuracy 0.2% f.s. ±1 digit a temperatura ambiente di 25°C t.s. 120msec Thermal drift 0,005% f.s. / °C Resolution function of settable sampling time:
>14bit, t.s. 120msec with sensor power control if strain gauge >13bit, t.s. 30msec (60msec with strain gauge power control) >12bit, t.s. 15msec (30msec with strain gauge power control)
Main input differential input for
- from strain-gauge 350(for pressure, force, etc.) sensitivity 5mV/V with strain-gauge power max 15V, (7.5mV/V con power supply max. 10V-15mV/V with power supply max 5V), positive or symmetrical polarization, calibration with automatic calibration of sensitivity, possible signaling of interrupted sensor power supply
- from potentiometer with power supply 1.2V, 100
- 0...20/4...20mA transmitter with 24Vdc supply (Ri = 50Ω)
Linear scale range -1999...9999 (with 4 digits),-1999...28000 (with 5 digits) settable
decimal point; a 32 section linearization can be inserted
Alarms (interception) max 4 configurable alarm types: absolute, deviation, symmetrical deviation.
Hysteresis setting
Alarm masking options:
- exclusion at switch-on
- latch, reset from key and/or contact
- insertion of delay filter (DON,DBI,DOF,DPO)
- activation of minimum time on trip Relay contact NO (NC), 5A, 250V/30Vdc cosϕ=1 Logic output 24V ±10% (10V min / 20mA) Triac output 20...240Vac ±10%, 1A max, snubberless, inductive and resistive load I2t = 128A Fault settings Alarm states can be configured in probe fault condition Sensor power 1,2Vdc for potentiometer > 100
5Vdc, 10Vdc, max 120mA (for strain-gauge) 15Vdc, 50mA max. 24Vdc ± 10% non stabilized 50mA, (100mA max for the
0...20/4...20mA transmitter input model only) Analogue retransmission 10V/20mA Rload max 500resolution 12 bit Digital inputs Ri = 4,7K(24V, 5mA) or from terminal not supplied. Serial interface (option) RS485, isolated Baudrate 1200, 2400, 4800, 9600, 19200 Protocol Gefran CENCAL / MODBUS Power supply (switching type) (standard) 100...240Vac/dc ±10%
(optional) 11...27Vac/dc ±10%
50/60Hz, 8VA max. Faceplate protection IP65 Working / Storage temperature range 0...50°C / -20...70°C Relative humidity 20...85% Ur non-condensing Environmental working conditions for indoor use, altitudes up to 2000m Installation panel, removable faceplate Installation specifications installation category II, pollution level 2, double isolation Weight 160 g complete version
6 • MAINTENANCE
This section gives the information and the necessary warnings for routine maintenance of the instrument and contains a Troubleshooting Guide which should be read before seeking help from the Gefran Customer Service Assistance, in the event of instru­ment malfunction.
If installed and configured correctly according to the instructions and the recommendations provided in Sections 2 and 4 of these Instructions for use, the instrument will work normally without any need for maintenance, apart from the usual operations of cleaning the faceplate, and if necessary the internal parts of the instrument.
To gain access to the inside of the instrument (for example for cleaning or to check the jumpers) just undo the screw at the bottom of the faceplate and take out the instrument without having to
disconnect the cables. Make sure that the power is turned off upstream of the instrument however. Remember that the 40B96 interceptor is not equipped with an ON/OFF switch.
Cleaning the Instrument
To clean the faceplate and the case use only a cloth dampened in water or ethyl alcohol. Do not use hydrocarbon-based solvents
(trichiorethylene, petrol, etc.). Do not use compressed air to remove dust from the elec­tronic circuit boards, if necessary use a clean brush with soft bristles.
Repairs
Repairs to the 40B96 instrument must only be carried out by qualified technicians, properly trai­ned and authorized by Gefran. Any attempts at repair or modification of the instrument hardware characteristics by unauthorized personnel will invalidate the warrantya.
Checking the jumpers
The solders side of the CPU board contains the jumper S4 which enables (if on) access to the controller menus.
The instrument contains components which are sensitive to electrostatic discharge, so the relevant precautions must be taken when handling the electronic circuit boards contained in it, in order to avoid permanent damage to components themselves.
Device structure: identification of boards
CPU BOARD +
POWER SUPPLY
BOARD
CPU BOARD + POWER SUPPLY BOARD
OUTPUT BOARD
S3
S2
S1
1V
5V 10V 15V 24V
You can select output state to have direct or reverse mode. Jumper S1, S2 and S3 are normally closed in position A. To change their state, the connection has to be removed. This operation must be done by trained technical personnel.
Example in case of relay output:
- Direct mode: energized relay and related closed contact correspond to active trip point.
- Reverse mode: de-energized relay and related open contact correspond to active trip point.
Power supply selection
Sensor or Transmitter
LS
S9 S1
S10
S6 S8
S4
S7 S5
ACCESS TO THE MENU (S4)
Jumper S4 is normally closed. To change its state, the connec­tion has to be removed. This operation must be done by trained technical personnel.
DISPLAY BOARD
OUTPUT
BOARD
Jumper Direct Inverse
AB AB
OUT1 S1 ON OFF OFF ON OUT2 S2 ON OFF OFF ON OUT3 S3 ON OFF OFF ON
7 • TECHNICAL/COMMERCIAL INFORMATION
This section contains information regarding the instrument order codes and the main accessories available.
As stated in the Preliminary Warnings of these Instructions for Use, correct interpretation of the instrument order code allows the hardware configuration to be identified immediately and so it is essential to quote the order code each time the Gefran Customer Care Service is contacted for assistance with any problems.
Order code – 40B96 alarm unit
For information on the availability of codes please contact your Gefran dealer .
The instrument display and Led do not come on
The characters shown on the display are incomplete or illegible
When pressing down none of the configuration menus can be accessed
When pressing down not all of the parameters and/or configuration menus can be accessed Instead of the process variable the PV display shows one of the following:
LO - KI - BR - ER - Eb
Instrument power supply problem. Check that power is being supplied to terminals 10-1 1. make sure the power supply corresponds with the one stated in the order code:
40B96 5 xx – xx – xx – x.x – 1 = 100..240V ac/dc
40B96 5 xx – xx – xx – x.x – 0 = 1 1..27Vac/dc Possible fault with one of the display segments. Check that all the segments are working pro­perly by switching the instrument off and then on again. When it is switched on again a self­diagnostic test is performed that checks intermittent start up of all the segments (displays the value 8888). If one or more segments do not light up contact your Gefran dealer . If the problem appears during the first installation, it probably means that the hardware configuration doesn't allow changing of the default parameters beyond the alarm setpoint value. (Parameter change is enabled by jumper S4 on the CPU board). If on the other hand the problem occurs on a instrument that previously gave access to the configuration parameters, this probably means that there is a false contact on the jumper S4. In this case check the continuity of the jumper referring to the previous paragraph.
Access to some menus and/or parameters is protected by a password (P A) and a code (PR) that limits configuration mode. To set the password and the protection code correctly refer to Section 4 "Configuration/Programming”. In the first four cases it means that an input error has been found (for details refer to Section 3
- Functions). In the last case, it means probe power supply failure. This function has to be enabled via parameter c.I. on the In menu.
Symptom Cause and Recommended remedy
Troubleshooting Guide
OUT 1, OUT 2
Relay, Relay R R
Relay, Logic R D
Triac, None T O
5Vdc 0 5
1,2Vdc potentiometer 0 1
24Vdc * 2 4
10Vdc
15Vdc
1 0
1 5
None0
Both **3
DIGITAL INPUT / RETRANSMISSION OUTPUT
100...240Vac/dc1
POWER SUPPLY
11...27Vac/dc0
40B
None0
RS 4852
TRANSMITTER
POWER SUPPLY
SERIAL LINE
Digital input1
96
NR. DIGITS
55
5
OUT 3, OUT 4
None 0 0
Relay, None R 0
Relay, Relay ** R R
0/4...20mA (0...10V) retansmision output **2
* Specific model for use with transmitter 20mA (es for position
magnetostrittive sensor) excludes Output 3, Output 4 and
retransmission
** Out 4 alternative to retransmission output
*** Selectable (standard 24Vdc)
ALL *** 99
ACCESSORIES
APPENDIX
The appendix contains the list of all the abbreviations of parameters which appear in the various configu­ration/programming menus with the respective default values and meanings. The CONF column can be used to indicate the user's modified values with respect to the default configu­ration, on the basis of application requirements.
Display Default CONF Acronym Description
Level 1
o.1 40 Output 1 Setting of alarm setpoint (Scale points) o.2 50 Output 2 Setting of alarm setpoint (Scale points o.3 60 Output 3 Setting of alarm setpoint (Scale points) o.4 70 Output 4 Setting of alarm setpoint (Scale points)
Menu
IIFF UD 3.20 UPdate Software version identification (O 1 Code Instrument code identification .KD Conf Hardware 1 Hardware outputs configuration .K2 Conf Hardware 2 Hardware inputs configuration
Menu
((FF K.I -1 HYsteresis 1 Hysteresis for setpoint 1 K.2 -1 HYsteresis 2 Hysteresis for setpoint 2 K.3 -1 HYsteresis 3 Hysteresis for setpoint 3 K.4 -1 HYsteresis 4 Hysteresis for setpoint 4
Menu
SSRR (O 1 Instrument Code Instrument identification code S.P 1 Serial Protocol Serial interface protocol BA 4 bAudrate Baudrate selection Pt 0 PArity Parity selection S.I 0 S. Input Virtual instrument inputs S.o 0 S. Output Virtual instrument outputs S.U 0 S. User Interface Virtual instrument user interface
Menu
IINN TP 0 type of Probe Probe type, signal, enable linearization, etc. (i 0 Sample time Select sampling time FT 0.1 FiLter time Digital filter on input FD 0.5 FiLter display Digital filter on display DP. 0 Decimal point Decimal point position for input scale L.S 0 Low Scale Minimum limit input scale K.S 1000 High Scale Maximum limit input scale OF 0 oFfSet Offset correction of main input d.I 0 Digital input Digital input function tU 0 UP key Raise key function t.d 0 DOWN key Lower key function L.L 0 Low Limit Lower limit for setting SP and absolute alarms K.L 1000 High Limit Upper limit for setting SP and absolute alarms
• RS232 / TTL interface for GEFRAN instrument configuration
WSK-0-0-0
Cable interface + CD Winstrum
• ORDER CODE
N.B. RS232 interface for PC configuration is supplied with the WINSTRUM program-
ming software. Make connection with instrument powered but with inputs and outputs disconnected.
Side insertion cable
Display Default CONF Acronym Description
Menu
00VV
00NN
0 Output number Number of trip point outputs
IITT
0 Output type 1 Trip point type for Out 1
22TT
0 Output type 2 Trip point type for Out 2
33TT
0 Output type 3 Trip point type for Out 3
44TT
0 Output type 4 Trip point type for Out 4
FF00
0 Output filter Filter mode on trip points
rraa
0 Output delay Trip point delay
TTMM
0 Minimum type Minimum output trip point time
rrEE
0 Fault action Definition of output states with broken sensor
Menu
LLNN
- Input linearization 00 – 32
Default CONF
.00 0
.01 31
.02 62
.03 94
.04 125
.05 156
.06 187
Default CONF
.07 219
.08 250
.09 281
.10 312
.11 344
.12 375
.13 406
Default CONF
.14 437
.15 469
.16 500
.17 531
.18 562
.19 594
.20 625
Default CONF
.21 656
.22 687
.23 719
.24 750
.25 781
.26 812
.27 844
Default CONF
.28 875
.29 906
.30 937
.31 969
.32 1000
Loading...