Gebruder Heyl Analysentechnik Testomat 2000 DUO Operating Instructions Manual

Online analysis instrument for water hardness, carbonate hard­ness, p-value or minus m-value
Operating Instructions
Contents
2
Contents
Contents ............................................................................................... 2
Important safety information ............................................................. 4
Intended use ......................................................................................... 4
Qualification of the staff ........................................................................ 4
Warning notices in these instructions ................................................... 5
Further documents ................................................................................ 5
Pay particular attention to .................................................................. 5
General instructions .............................................................................. 5
Installation ............................................................................................. 6
Operation .............................................................................................. 6
Cleaning ................................................................................................ 6
De-installation ....................................................................................... 6
Disposal ................................................................................................ 6
Scope of delivery ................................................................................ 7
Performance specifications ............................................................... 7
Indicators for Testomat 2000® DUO instruments .................................. 8
Application instructions ..................................................................... 9
Installation ......................................................................................... 10
Operating Testomat 2000® DUO in the pressure range 0.3 to 1 bar .. 10
Installing Testomat 2000® DUO .......................................................... 10
Connecting the water inlet and outlet ................................................. 11
Water inlet ........................................................................................... 11
Water outlet ......................................................................................... 11
Connecting the power supply and devices ......................................... 12
Block diagram Testomat 2000® DUO ................................................ 12
Internal design Testomat 2000® DUO................................................. 13
Connecting the mains voltage............................................................. 14
Connecting the plant components ...................................................... 15
Connecting the inputs and outputs ..................................................... 16
Commissioning ................................................................................. 17
Inserting the indicator bottle ................................................................ 17
Extracting the indicator ....................................................................... 17
Opening the water inlet ....................................................................... 17
Instrument settings and data input ...................................................... 18
Functions of the operating and display elements ......................... 18
Switching Testomat 2000® DUO on/off ............................................... 18
Display functions ................................................................................. 19
Operating elements and function keys ............................................... 20
Operating system ................................................................................ 21
Password protection and basic program ....................................... 22
Entering basic program data ............................................................... 22
Selecting the indicator and the bottle size .......................................... 22
Selecting the operating mode ............................................................. 23
Selecting the display unit(s) ................................................................ 24
Measuring point or parameter selection ............................................. 24
Entering further basic program data .............................................. 25
Internal flushing ................................................................................... 25
External flushing ................................................................................. 25
Interval pause ..................................................................................... 26
Contents
3
Limit value monitoring ......................................................................... 26
Hysteresis ........................................................................................... 27
Switch functions of the limit value outputs LV1 and LV2 .................... 28
Switch function 0, duration .................................................................. 28
Switch function 1, impulse .................................................................. 28
Switch function 2, interval ................................................................... 28
Function IN1 ........................................................................................ 28
Water meter ........................................................................................ 28
Alarm / Message ................................................................................. 29
Function AUX ...................................................................................... 29
Service II ............................................................................................. 30
Reset operating time ........................................................................... 30
Maintenance interval ........................................................................... 30
Description of the signal inputs/outputs ........................................ 31
Interfaces (optional) ............................................................................ 33
Monitoring a measuring point.............................................................. 33
Monitoring of two measuring points .................................................... 33
Calculating the output currents ........................................................... 34
Serial interface RS232 ........................................................................ 35
Description of the relay outputs ...................................................... 35
Flushing (external flush valve) ............................................................ 35
LV1 and LV2 limit value outputs ......................................................... 35
Measuring points 1 or 2 (measuring point switch-over) ...................... 36
Alarm (fault message output) .............................................................. 37
Maintenance (output for maintenance message) ............................... 38
Information menu "i" ........................................................................ 39
Program menu "M" ........................................................................... 40
Structure of the basic program............................................................ 42
Error messages / Troubleshooting ................................................. 43
Further information .............................................................................. 44
Maintenance ...................................................................................... 45
Description of maintenance work ........................................................ 45
Service instructions ............................................................................. 46
Testomat 2000® spare parts and accessories ................................ 47
Accessories ......................................................................................... 48
Technical data ................................................................................... 49
Important safety information
4
Important safety information
Please read these operating instructions carefully and completely
prior to working with the instrument.
Ensure that these operating instructions are always available for all
users.
These operating instructions must always be passed on to the new
owner should Testomat 2000® DUO change hands.
Always adhere to hazard warnings and safety tips when using
reagents, chemicals and cleaning agents. Please adhere to the respective safety data sheet! Download the safety data sheets for the supplied reagents at http://www.heyl.de .
Intended use
Testomat 2000® DUO is used for the automatic determination and monitoring of residual total hardness (water hardness), residual car­bonate hardness, as well as the minus m-value and the p-value in water. The measurement parameters and the respective measuring range are determined by the indicator selection and according to the user programming.
Always adhere to the performance limits stated in the section enti-
tled “Technical data”.
Always observe the application areas/application limits of the indi-
cators and the requirements of the medium being measured.
To ensure correct and intended usage, always read and understand
these instructions, especially the section entitled “Important safety information”, prior to use.
The instrument is not used as intended if
it is used in areas not specified in these instructions, it is used in areas which do not correspond to the ones described
in these instructions.
Qualification of the staff
Assembly and commissioning require fundamental electrical and process engineering knowledge as well as knowledge of the respec­tive technical terms. Assembly and commissioning should therefore only be carried out by a specialist or by an authorised individual su­pervised by a specialist.
A specialist is someone who due to his/her technical training, know­how and experience as well as knowledge of relevant regulations can assess assigned tasks, recognise potential hazards and ensure ap­propriate safety measures. A specialist should always adhere to the relevant technical regulations.
Pay particular attention to
5
Warning notices in these instructions
The warning notices in these instructions warn the user about poten­tial dangers to individuals and property resulting from incorrect han­dling of the instrument. The warning notices are structured as follows:
Description of the type or source of danger
Description of the consequences resulting from non-observance
Preventive measures. Always adhere to these preventive
measures.
DANGER” indicates an immediate hazardous situation which, if not
avoided, will result in death or serious injury.
WARNING” indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION” indicates a potentially hazardous situation which, if not
avoided, could result in minor or moderate injuries or property dam­age.
NOTE” indicates important information. If this information is not observed, it may result in an undesirable result or state.
Further documents
Testomat 2000® DUO is a plant component. Therefore, always ob­serve the maintenance manual of Testomat 2000® / Testomat 2000 ECO® and the documentation of the plant manufacturer.
Pay particular attention to
General instructions
Please adhere to health and safety regulations, electrical equip-
ment safety regulations, and environmental protection regulations valid in the country of use and at the installation site.
Adhere to national and local regulations during installation and
commissioning.
Always protect the instrument against moisture and humidity. It
should never come into contact with condensation or splash water.
Do not carry out any changes or actions at the instrument which are not described in these instructions;
failure to adhere to these in­structions will negatively affect any warranty claims that you make thereafter.
SIGNAL WORD!
DANGER
!
WARNING
!
CAUTION
!
NOTE
WARNING
!
Pay particular attention to
6
Installation
Always completely disconnect the relevant plant part before in-
stalling the instrument or connecting/disconnecting it to/from the power supply. Secure the plant against reconnection.
Only connect the instrument to the mains voltage specified on the
rating plate.
Always observe technical data and ambient parameters.
Testomat 2000® DUO requires an interference free and stable
power supply. If necessary, use a mains filter to protect Testomat 2000® DUO against interference voltages caused, e.g., by solenoid valves or large motors. Never lay connecting cables parallel to power cables.
Operation
Ensure that the maximum electrical load capacity of the relay out-
puts is never exceeded.
Immediately switch off Testomat 2000® DUO and contact service
staff if malfunctioning occurs. The warranty will be void if you tam­per with or attempt to repair Testomat 2000® DUO. Repairs must be carried out by authorised service staff.
Cleaning
Only use a dry, lint-free cloth for cleaning.
De-installation
Prior to de-installing a defective instrument, always write down a
description of the error (failure effect). It is only possible to repair a defective instrument (irrespective of the warranty period) if it has been de-installed and returned to us with a description of the error.
Disposal
Dispose of the instrument in accordance with national regulations.
WARNING
!
NOTE
Scope of delivery
7
Scope of delivery
1 Testomat 2000® DUO 1 Plastic bag with two screw caps with a hole and an insert for the
screw caps of the indicator bottles 1 operating instructions
Performance specifications
Testomat 2000® DUO is used for the automatic determination and monitoring of residual total hardness (water hardness), residual car­bonate hardness, as well as the minus m-value and the p-value in water. The instrument can be used to alternately measure two pa­rameters at one measuring point or two measuring points with differ­ent (or the same) parameters. The measurement parameters and the respective measuring ranges are determined by the indicator selec­tions and according to the user programming.
Simple, menu-driven operating and programming via a plain
text display
Determinable measuring of residual hardness, total hard-
ness, carbonate hardness, minus m-value, p-value via indica­tor selection
Freely selectable hardness unit in °dH, °f, ppm CaCO3 or
mmol/l
High measuring accuracy provided by a precise piston-
dosing pump
Analysis initiation:
- Automatic interval operation (interval pause can be set from 0-99 minutes)
- External control
- Quantity dependent via impulse water meter
Two independent limit values with hysteresis (1, 2 or 3 bad
analyses) and settable switch functions
Monitoring of two measuring points (change-over via external
solenoid valves)
Internal error documentation  Programmable service address  Programmable maintenance interval for a maintenance re-
quest
Extended operating periods due to 500 ml indicator storage
bottle
optional:
Interface card (0/4-20 mA or 0/2-10 V) Interface RS 232 (for a protocol printer)
Performance specifications
8
Indicators for Testomat 2000® DUO instru­ments
Parameter/Indicator type
Water hardness
TH 2005
TH 2025
TH 2100
TH 2250
Unit
°dH
(resolution)
0.05 – 0.50
(0.01)
0.25 – 2.50
(0.05)
1.0 – 10.0
(0.2)
2.5 – 25.0
(0.5)
°f
(resolution)
0.09 – 0.89
(0.02)
0.45 – 4.48
(0.1)
1.8 – 17.9
(0.4)
4.5 – 44.8
(1.0)
ppm
CaCO
3
(resolution)
0.89 – 8.93
(0.2)
4.5 – 44.8
(0.9)
18 - 179
(3.8)
45 - 448
(10)
mmol/l
(resolution)
0.01 – 0.09
(0.002)
0.04 – 0.45
(0.01)
0.18 – 1.79
(0.04)
0.45 – 4.48
(0.1)
Parameter
/Indicator
type
Carbonate
hardness
minus m-
value
p-value
p-value
TC 2050
TC 2100
TM 2005
TP 2010
TP 2100
Unit
°dH
(resolution)
0.5 – 5.0
(0.5)
1.0 – 20.0
(1.0)
- - -
°f
(resolution)
0.90 – 8.96
(0.9)
1.8 – 35.8
(1.79)
- - -
ppm
CaCO
3
(resolution)
8.9 – 89.5
(8.9)
18 - 358
(18)
- - -
mmol/l
(resolution)
0.18 – 1.79
(0.18)
0.36 – 7.16
(0.36)
0.05 – 0.50
(0.01)
0.1 – 1.5
(0.1)
1 - 15
(1)
Display, if measuring value is outside of range (for example TH2005): At falling below: < 0,05°dH At exceeding: > 0,50°dH
Application instructions
9
Application instructions
Wait at least 5 seconds before switching the instrument on and
then off again at the main switch.
In order for Testomat 2000® DUO to operate reliably, use Heyl
Testomat 2000® indicators in the pH-range 4 – 10.5!
With Testomat® instruments for water hardness monitoring, larger
quantities of heavy metal ions in the softened water might influ­ence the colour reaction, especially iron above 0.5 mg/I, copper above 0.1 mg/I and aluminium above 0.1 mg/l (brownish-red colour display).
If the measuring water contains more than 20 mg/I CO2 (carbonic
acid), incorrect evaluations cannot be excluded.
The concentration of influencing contents can be determined by
using our colourimetric TESTOVAL® test kit.
Careful handling of the instrument increases both its operational
reliability and service life! Therefore, carry out a visual inspection at regular intervals as described below:
- Has the use-by date of the indicator expired?
- Are the hose connections of the dosing pump free of leaks?
- Is there air inside the dosing hoses?
- Are all the water connections free of leaks?
- Are the doors of the instrument closed properly?
- Is the instrument heavily soiled?
- Are the measuring chamber and the drain duct/drain hose clean?
Trouble-free operation is only possible when maintenance is car-
ried out on a regular basis! For more information, please refer to
the section entitled “Maintenance” and the "Maintenance manual
of Testomat 2000® / Testomat 2000 ECO®"
If problems occur, please refer to the section entitled "Error mes-
sages/Troubleshooting”.
Installation
10
Installation
Risks resulting from incorrect installation!
Install Testomat 2000® DUO at a location where it is protected
against dripping or splash water, dust and aggressive substances – e. g. in a switch cabinet or on a suitable wall.
Information for trouble-free operation
Install Testomat 2000® DUO vertically and without mechanical
stress.
Install Testomat 2000® DUO at a vibration-free site.
Operating Testomat 2000® DUO in the pres­sure range 0.3 to 1 bar
Prior to installation, please check whether lower operating pressure is required. The instrument is factory set for the operating range 1 to 8 bar. Remove the flow controller valve body to operate the instru­ment in the operating range 0.3 to 1 bar (e.g. when using an aerator type R). This involves removing the retaining pin from the control­ler / filter receiver . Subsequently use the metal bracket to remove the controller plug from the borehole. Subsequently remove the flow controller valve body and reinsert the controller plug and the retaining pin.
Installing Testomat 2000® DUO
Select an installation site where the water inlet hose can be kept as short as possible (max. 5 m).
Please leave sufficient space on the left-hand side of the instru-
ment to open the door.
Drill the mounting holes as shown in the drawing on the left. Use three screws to attach the instrument at a suitable position in
the switch cabinet or on a wall.
NOTE
Remove for pressure range 0.3 to 1 bar
WARNING
!
Installation
11
Connecting the water inlet and outlet
Information for trouble-free operation
The water pressure must be between 0.3 bar and 8 bar Avoid strong pressure fluctuations The measuring water temperature must be between 10 °C and
40 °C
For temperatures above 40 °C, the KCN type cooler should be
installed in the branch line of Testomat 2000® DUO.
Water inlet
The measuring water is taken from the main water line of the water treatment plant and fed to the inlet connection of Testomat 2000
®
DUO. The instrument is equipped with a plug connector for plastic hoses 6/4 x 1 (external diameter 6 mm/ internal diameter 4 mm, wall thickness 1 mm) as standard.
Install the connection for the branch line of Testomat 2000® DUO
directly at the main water line directly after the water treatment
plant
It is important that the branch line connection is laid vertically up-
wards in order to prevent dirt particles from entering the instru­ment from the main water line
Install a manually operated shut-off valve Testomat 2000® DUO  Use an opaque plastic hose 6/4 x 1 (max. length 5 m) for the water
inlet
Flush the inlet to remove any dirt particles
When operating within a pressure range of 0.3 to 1 bar or with a sup­ply via a booster pump, please remove the valve body from the con­troller and the filter housing. The pump should have a feeding capaci­ty of between 25 and 35 litres/hour and be resistant to the medium being measured.
When using a cooler
The hot water can cause burns and damage wetted parts of
Testomat 2000® DUO.
Water outlet
The feed water flows through the measuring chamber to the drain via the outlet hose.
Connect the outlet connection of Testomat 2000® DUO to an
opaque outlet hose (internal diameter 14 mm).
Lay this hose without backwater development and any syphoning
effect, e.g. via an open funnel, to the drain
NOTE
    
NO !!
"Sagging"
causes back-
water!
CAUTION
!
Installation
12
Connecting the power supply and devices
Risk of electric shocks during installation!
If the power supply is not disconnected prior to installation, it may result in personal injuries, destruction of the product or damage to plant parts.
Always disconnect the relevant plant parts before installing Tes-
tomat 2000® DUO.
Only use tested cables with sufficient cross-sections for the con-
nections.
Risk of damages caused by electromagnetic fields!
If Testomat 2000® DUO or the connecting cables are installed
parallel to power cables or in close proximity to electromagnetic fields, the instrument may be damaged or measurements incor­rect.
Ensure that connecting cables are as short as possible. Always install connecting cables and power cables separately. Connect the instrument to the protective earth conductor (for
230/115 VAC).
Protect Testomat 2000® DUO against interference voltages e.g.
via a mains filter.
Shield the instrument against strong electromagnetic fields.
Block diagram Testomat 2000® DUO
Drawn relay positions: Instrument de-energised
WARNING
!
NOTE
Installation
13
Internal design Testomat 2000® DUO
Terminal strip for inputs Start, Stop, IN1, IN2, and output OUT
Mains switch
Terminal strip for power input and power output
Terminal strip for relay outputs
Dosing pumps
Water connections, inlet and outlet
Flow controller
Measuring chamber
Installation
14
Connecting the mains voltage
Only connect the instrument to the specified mains voltage. Refer to the rating plate for the appropriate mains voltage. Connect the cables as follows:
Loosen both fastening screws and open the upper door. The
terminal box is now accessible.
Pierce the required rubber cable glandswith a screwdriver and
insert the cable through the bush into the terminal box (1)
Subsequently pull back the cable until the bush has been turned
over (2)
Connect the power supply to terminals PE,N,L or for 24 V instru-
ments to terminals U,V
Connect the conductor to the terminal block as shown on the left
.
Ensure that the leads are held securely in the terminals Proceed as shown in figure to loosen the connection.
Terminal description
Type
Function
Comment
PE
IN
Mains – protective earth (5x)
Only with mains 115/230 V !
N (U)
L (V)
IN
Mains, N=neutral (U=24V) Mains, L=live (V=24V)
Mains input 24 V / 115 V / 230 V
n
l
OUT
Neutral, switched (8x) Live, switched (8x)
Mains for consumers, max. 4 A
Insert the conductor with ferrule or the solid conductor into the round input.
1. Insert a screwdriver into the square opening without force in order to open the terminal.
2. Once the terminal has been opened, remove the conductor.
 
Installation
15
Connecting the plant components
Connect the plant components to the output terminals of relays 1
to 19 (e.g. valves)
If the plant components require mains voltage, connect the
switched mains voltage (l) to the common contact of the re­spective relay (see the connection example for 230 VAC on the
left)
Connect the neutral conductor of the plant component to one of the
terminals (n)
For components with a protective earth conductor connection,
connect it to the PE connection
Ensure that the leads are held securely in the terminals
No.
Terminal
description
Type
Function
Comment
1 2
Flush
OUT
External flush valve
Isolated relay output, max. 240 VAC, 4 A
3 4 5 LV1
OUT
Limit value output 1 – Normally closed Limit value output 1 – Normally open Limit value output 1 – Common
Isolated relay output, max. 240 VAC, 4 A
6 7 8 GW2
OUT
Limit value output 1 – Normally closed Limit value output 1 – Normally open Limit value output 1 – Common
Isolated relay output, max. 240 VAC, 4 A
9 10 11
Measur-
ing points
1/2
OUT
Measuring point 1 – Normally closed Measuring point 2 – Normally open Measuring point switch-over - Common
Isolated relay output, max. 240 VAC, 4 A
12 13
AUX
OUT
Universal output
Isolated relay output, max. 240 VAC, 4 A
14 15 16
Alarm
OUT
Fault message output – Normally closed Fault message output – Normally open Fault message output – Common
Isolated relay output, max. 240 VAC, 4 A
17 18 19
Mainte-
nance
OUT
Maintenance message - Normally closed Maintenance message - Normally open Maintenance message - Common
Isolated relay output, max, max. 240 VAC, 4 A
Connection example:
Limit value contact LV 1 switches mains voltage
n l
3 4 5
e.g.
solenoid
valve
PE
Installation
16
Connecting the inputs and outputs
Testomat 2000® DUO has the following connections for control and monitoring functions.
Do not connect external voltage to these connections! Ensure that the leads are held securely in the terminals. Use the two fastening screws to close the upper door once installa-
tion has been completed.
For more information, please refer to the section entitled "Description of the signal inputs/outputs".
No.
Terminal
description
Typ e
Function
Comment
20 21
Start
IN
External analysis start Common earth for inputs
Only for isolated normally open!
22 23
Stop
IN
External analysis stop Common earth for inputs
Only for isolated normally closed/normally open!
24 25
IN1
IN
Universal input 1 Common earth for inputs
Only for isolated normally closed/normally open!
26 27
IN2
IN
Universal input 2 (water meter) Common earth for inputs
Only for isolated normally open!
28
29
OUT
OUT
OUT
OUT
OUT
OUT
IN
RS 910: Earth
SK910: (+) UK910: (+) RS910: (TxD) SK910: (-) UK910: (-) RS232: (RxD)
RS232 interface for printer
Galvanically separated current interface 0/4 - 20mA
Galvanically separated voltage interface 0/2 – 10V
RS232 interface for printer
Galvanically separated current interface 0/4 - 20mA
Galvanically separated voltage interface 0/2 – 10V
RS232 interface for printer
Commissioning
17
Commissioning
In order for Testomat 2000® DUO to operate reliably, only use Heyl Testomat
®
indicators
!
Inserting the indicator bottle
Open the lower housing door by pulling on the right-hand side. Remove the cap from the indicator bottle Remove the plastic bag from inside the lower housing door. The
plastic bag contains the screw cap with hole and the insert
for the screw cap.
Connect the parts as shown on the left. Screw the hose connector of the intake hose hand-tight into
the insert .
Place the insert with the screwed-in intake hose into the indicator
bottle.
Now screw the screw cap with hole hand-tight onto the
indicator bottle.
Extracting the indicator
During operation, the pump (DOSIClip) automatically extracts
the indicator.
To ensure that indicator is available for the initial analyses, the
intake hose and the transport hose must be filled with indica­tor from the pump up to the measuring chamber.
Press the "manual" key several times until the intake hose
and the transport hose are filled with indicator up to the meas­uring chamber (always switch on the instrument at the mains
switch first!)
Opening the water inlet
Open the lower housing cover. Slowly open the manually operated shut-off valve to prevent the
measuring chamber overflowing. The flow regulator requires a few seconds to function correctly.
Make sure that the water conducting parts are not leaky.
     
CAUTION
!
Functions of the operating and display elements
18
If water sprays from the vent hole of the measuring chamber
, reduce the amount of inlet water via the manually operated
shut-off valve. It should take 2 to 6 seconds to fill the measuring chamber!
Instrument settings and data input
Please read the following information before carrying out settings
and entering data for operating the instrument.
Functions of the operating and dis­play elements
The Testomat 2000® DUO display shows operating statuses and measured values. The input keys for programming (cursor block) and the function keys are located underneath the display.
Switching Testomat 2000® DUO on/off
(1) Mains switch
Use this switch to switch the instrument on or off.
(2) Instrument fuse (inside the instrument)
This fuse protects Testomat 2000® DUO and the outputs against overloads and short circuits.
(1)
(2)
Functions of the operating and display elements
19
Display functions
1 Status of limit value displays (red/green)
The display 1 illuminates red if limit value 1 has been reached or exceeded. The display 1 illuminates green if the value falls below the limit value. The same principle applies to limit value 2 and dis­play 2.
2 Text display (4 lines)
Displays the current analysis result as well as all important status­es and programming data:
Line 1 = the set indicator types for measuring point 1 (M1:) and 2 (M2:) are displayed. (P1:/P2:) is displayed for for a measuring point. Line 2 = measured value display measuring point (M1) or (P1) Line 3 = measured value display measuring point (M2) or (P2)
Value falls below the measuring range = "<" e.g. M1: < 0.05 °dH Value exceeds the measuring range = ">" e.g. M1: > 5.0 °dH
Line 4 = The set limit values LV1 and LV2 are
displayed 3 Alarm (red) Indicates malfunctioning/error message or warning message.
4 Maintenance message (yellow)
Indicates current maintenance requests
5 Status display of the active instrument components (line)
Six displays indicate the current instrument and analysis status
Cancelling error messages/warning messages
Press to acknowledge the message and, if necessary,
eliminate the cause of the fault.
All error and warning messages are alternately shown in line 1 of the standard display!
NOTE
Inlet
valve
Analysis
running
Outlet valve
Dosing
message
Testomat
2000
®
only
Analysis stop
Functions of the operating and display elements
20
  
Programming keys (cursor block)
Operating elements and function keys
If you wish to carry out settings or enter data, or if alterations are necessary, press the "M" key to open the programming mode. Use this key when in the menu to go to the higher order menu items or to exit the programming mode. Please note: The basic program is password protected!
Programming keys (cursor block)
Use the programming keys (cursor block) to navigate in the menu, to select the desired functions and to enter necessary instrument and plant specific data. Press the "ENTER" key to select the submenu item and to confirm and accept the selection or data input.
Function keys
Manually start an analysis via the "Manual" key
Set the instrument to standby mode via the "STANDBY"
key (automatic analyses are not carried out: analysis
stop)
Acknowledge error and warning messages via the "Horn" key
Open the program menu for user-specific and instrument­specific settings via the "M" key
Retrieve all instrument information and settings via the
"i" key
Cursor block
(M)enu key
Functions of the operating and display elements
21
Operating system
Instrument settings and data input Date, time and weekday input
Press the "M" key
The basic menu " " appears
Use the cursor block to select the desired menu item
" "
The selection appears in CAPITAL LETTERS.
Press "ENTER" to confirm your selection
The selected submenu " " appears The menu item " has already been selected (capital letters)
Press "ENTER" to confirm the menu item " "
The cursor flashes in the date field:
Use the cursor keys to select the
desired number
Use the keys to move the cursor to the next input field Repeat this input process until the year has been entered
Press "ENTER" to confirm the entry
The date has now been entered.
Exit the menu item " " in order to set the time.
Use the cursor block to select the desired menu item
" "
Press "ENTER" to confirm your selection
The cursor flashes at the first position of the time:
Use the cursor keys to select the desired number Use the keys to move the cursor to the next input field
Repeat this input process until the seconds have been entered Press "ENTER" to confirm the entry
The time has now been entered.
Exit the menu item " " to set the weekday
Use the cursor block to select the desired menu item
"
Press "ENTER" to confirm your selection Use the cursor block to select the selected weekday Press "ENTER" to confirm the entry Press the "M" key twice to end programming
The standard measured value display appears on the display
Testomat instrument
(display) in display mode
Password protection and basic program
22
Password protection and basic program
A four-digit password is required to enter data and to carry out set­tings in the basic program. The password is the current time of Tes­tomat 2000® DUO in reverse order.
Password entry
Press the "M" key
The basic menu " " appears
Use the cursor block to select the desired menu item
" "
The selection appears in CAPITAL LETTERS.
Press "ENTER" to confirm your selection
The selected submenu " " appears
Press "ENTER" to confirm the menu item " "
The cursor flashes in the
Use the cursor keys to enter the
time in reverse order:
Press "ENTER" to confirm the entry
The selection menu for the basic program appears. You can now enter the plant specific data.
Entering basic program data
Selecting the indicator and the bottle size
In , select
=> =>
Press "ENTER" to confirm the selection
The menu shown on the left appears
Select the size of the indicator bottle
(A 500 ml bottle " * " is factory set)
Press "ENTER" to confirm the selection
(An asterisk " * " appears at the end of the line)
Select the indicator type
(The indicator type TH2005 " * " is factory set)
Press "ENTER" to confirm the selection
(An asterisk " * " appears at the end of the line) The asterisk " * " displays the active menu item.
The indicator has now been selected.
▼▲
▼▲
▼▲
Geben Sie hier die Uhrzeit rück­wärts ein:
=> 5201
Testomat instrument
(display) in display mode
▼▲
Password protection and basic program
23
Selecting the operating mode
Under the menu item it is possible to select the type of analysis controller. Testomat 2000® DUO provides nu­merous selection options: Time control, quantity control via water meter, dynamic analysis triggering and external analysis triggering.
Shortest interval = 0 minutes between analyses. Largest interval = 99 minutes.
The analysis interval (time between two analyses) is determined by the duration of the supplementary program AUX, the set flush times (internal and external), the programmed interval and the duration of the analysis. The analysis duration depends directly on the meas­ured value.
Selecting the time control
In , select
=> => =>
Press "ENTER" to confirm the selection
(An asterisk " * " appears at the end of the line) ("TIME CONTROLLED" " * " is preset)
Enter the interval pause and the flush times
In , select
=> => =>
Enter the " in minutes (m)
(1 minute is preset)
Enter the " in seconds (s)
(00 seconds (s) is preset)
Enter the " in seconds (s)
(00 seconds (s) is preset)
Press "ENTER" to complete all the entries
Duration of the analysis interval
The analysis interval is the addition of the "AUX before/after analy-
sis", "Internal flush" and "External flush" times and the measuring
value dependent analysis duration (see diagram on the left)
Time control
Internal start via timer.
▼▲ ▼▲
NOTE
Analysis interval
Sequence of times
Password protection and basic program
24
Minimum interval = 1 litre, maximum interval = 9999 litres. The analy­sis is carried out once the programmed water quantity has been measured. The line and the measuring chamber are flushed prior to the analysis (observe the programmed flush times).
Selecting the quantity control
In , select
=> => =>
Press "ENTER" to confirm the selection
The menu on the left appears.
Enter the respective flow rate in litres Press "ENTER" to confirm the entry
Selecting the type of water meter
Select the menu
=> => =>
Select the water meter constant (litre/impulse)
(100 litres/impulse " * " is factory set)
Press "ENTER" to confirm the selection
External analysis start
See section “Description of the signal inputs/outputs".
Selecting the display unit(s)
It is possible to program the units of the displayed values. The units dH, °f, ppm CaCO3 and mmol/l can be selected. All the following in­puts and displays will then be displayed in the programmed unit.
In , select
=> =>
Select the desired unit Press "ENTER" to confirm the selection
Measuring point or parameter selection
By selecting "1 measuring point" in the menu, parameter 1 and pa­rameter 2 are measured alternately at a measuring point (e.g. water hardness and carbonate hardness). "P" appears on the display. By selecting "2 measuring points", parameter 1 and parameter 2 are measured alternately at measuring point 1 and measuring point 2 respectively. “ ” appears on the display.
The measurement can be limited to one measuring point or parame-
ter via the inpiut IN1. See section “Description of the signal in- puts/outputs”. A change-over of the parameters or measuring points
also only occurs after completion of the "Hysteresis" function.
Quantity control
Start via water meter
▼▲
External
Analysis start
Start via input
"Start"
▼▲
▼▲
Entering further basic program data
25
Entering further basic program data
For selecting and entering data for these functions, please proceed as described under "Entering basic program data".
Internal flushing
To ensure that the analysed sample represents the current value, the sampling line must be sufficiently flushed. If the plant has been out of operation for a longer period or in case of long analysis intervals, we recommend you to select a flushing time greater than 60 seconds. Flushing starts by simultaneously opening the inlet and the outlet valve of Testomat 2000® DUO.
Duration of the analysis interval
The analysis interval depends directly on the programmed flushing
time. If, e.g., a flushing time of 90 seconds has been set, the actu-
al analysis interval cannot be less than 90 seconds.
In , select
=> =>
Enter the " in seconds (s) Press "ENTER" to confirm the entry
External flushing
If very short analysis intervals are required, or if a very long (several metres) sampling line or a line with a large diameter is used, an ex­ternal flush valve should be installed upstream of Testomat 2000® DUO. The external flush valve has to be connected to the "Flush” outlet. If the unit is used for monitoring two measuring points, external flushing prevents incorrect measurements caused by sample mixing. The external flushing time for the valve depends, just as the flushing time for unit flushing does, on the length and diameter of the supply line to Testomat 2000® DUO.
Setting the external flushing time
For connections longer than 3 m and with an internal hose diame-
ter of 6 mm a minimum internal flushing time of 10 seconds is re-
quired to ensure that a valid sample is taken from the sampling
line. The required quantity of flush water for one-minute internal
flushing is 0.5 litres.
In , select
=> =>
Enter the " in seconds (s) Press "ENTER" to confirm the entry
NOTE
NOTE
▼▲
▼▲
Entering further basic program data
26
Interval pause
If the analysis is triggered via a timer, the interval between two anal­yses (plus flushing time) is determined by the interval pause. The shortest interval can be 0 minutes. In this case, analyses are carried out continuously. The longest interval is 99 minutes.
In , select
=> =>
Enter the " in minutes (m) Press "ENTER" to confirm the entry
Limit value monitoring
It is possible to program the limit values on a continuous scale. The limit value range depends on the used indicator type and the pro­grammed unit. Two limit value outputs are available for monitoring. This ensures that two limit values or two measuring points can be monitored. The functions of the allocated relay outputs can be pro­grammed independently of each other.
If the unit is used for monitoring two limit values, the limit value out­puts are permanently allocated to these limit values!
1 LV1 = Limit value 1 2 LV2 = Limit value 2
If the unit is used for monitoring two measuring points, the limit value outputs are permanently allocated to these measuring points!
1 LV1 = Measuring point 1 2 LV2 = Measuring point 2
If the limit value LV1 has been exceeded, the limit value control dis­play 1 lights up RED and the relay output LV1 reacts as pro­grammed in the switch function. If the limit value has not been ex­ceeded, the display lights up GREEN. The same applies for the limit value LV2.
In , select
=> =>
Enter the values for " or " Press "ENTER" to confirm the entry
Monitoring of two limit values
Monitoring of two measuring points
▼▲
Entering further basic program data
27
Hysteresis
The respective limit value output only switches after the first, second or third bad analysis (suppression of the first or the second measured value). This increases the reliability of the analysis evaluation, e.g. after the measuring point has been switched over or if the sampling line has not been flushed sufficiently. The hystereses of the two out­puts LV1 and LV2 can be set independently of each other.
Function: A further analysis is carried out immediately after the limit value has been exceeded for a hysteresis of 2. The respective output only reacts if the limit value of this analysis is exceeded again. If a hysteresis of 3 has been set, the respective output only reacts after the limit value has been exceeded for the third time in succession. This setting is only reactivated once the value has fallen below the limit value!
(The basic setting for LV1 and LV2 is 1)
In , select
=> =>
Enter the number of analyses Press "ENTER" to confirm the entry
Hysteresis and measuring point or parameter change-over
The number of analyses entered in the menu item "Hysteresis" is
always executed with the same parameter or always at the same
measuring point. The parameter or measuring point change-over
only occurs after completion of the Hysteresis function.
Suppression of
bad analyses
NOTE
Entering further basic program data
28
Switch functions of the limit value outputs LV1 and LV2
Switch function 0, duration
If the limit value LV1 or LV2 has been exceeded, the output relay LV1 or LV2 switches. If the measured value falls below the limit value LV1 or LV2 without locking, the relevant relay drops out again.
Switch function 1, impulse
If the measured value exceeds the limit value LV1 or LV2, the rele­vant output switches for a settable time (t).
The respective output always remains switched for the set time, irre­spective of how long the limit value has been exceeded. A new im­pulse is only possible once the value has fallen below the limit value!
Switch function 2, interval
If the limit value has been exceeded, the respective output switches at intervals with the settable time (t) = impulse or interval as long as the limit value is exceeded.
Function IN1
This input is used to suppress change-over of the parameters or the measuring points. The measurement only occurs at parameter 1 or at measuring point 1 with parameter 1.
See section “Description of the signal inputs/outputs".
In , select
=> =>
Select a normally closed or normally open contact Press "ENTER" to confirm the entry
Water meter
It is necessary to connect a water meter to input IN2 for quantity­dependent analysis triggering. Program the corresponding water meter rating.
In , select
=> =>
Select the water meter rating Press "ENTER" to confirm the entry
▼▲
▼▲ ▼▲
Switch functions dia-
gram
Entering further basic program data
29
Alarm / Message
The instrument is equipped with an alarm relay output for signalling faults. The events which mean a fault at the instrument or are intend­ed to trigger a message, can either trigger an alarm "A" (continuous contact) or a message "M" (2-second impulse).
The faults are recorded and stored in the error history if the event has been programmed as an alarm or message. For example, if a low indicator level has not been programmed as an ALARM/MESSAGE, it is not registered in the error history. Up to 20 error messages can be stored. A list of these errors can be opened in the information menu. The information stored per event is the time (day, month, year and hour) and the type of the error.
In , select
=> =>
Select the type of monitoring A=alarm, M=message or - = no action
for the individual menu items
Press "ENTER" to confirm the respective entry
Error messages
All error messages are lost after a power failure. Certain instrument faults always trigger an alarm or a message!
Function AUX
The AUX relay output can be programmed for the following control functions:
- As a function output for the contact with programmable duration prior to and/or during the analysis, or after an analysis.
For example, it is possible to control the cooling water inlet of an up­stream cooler via a solenoid valve. This ensures that the cooling wa­ter only flows when required, i.e. when an analysis is being carried out.
In , select
=> =>
Select the program step at which the AUX contact is to be activat-
ed
Under "Time" enter the contact duration in minutes (m) and sec-
onds (s)
Press "ENTER" to confirm the entry
NOTE
▼▲
▼▲
A=Alarm, M=Message
- = no action
Ff.=Function fault
Mf.=Measuring fault
Entering further basic program data
30
Service II
The service II menu contains various functions for monitoring the operation of the instrument:
Programming of the maintenance interval, operation (reset) internal data/setting, e.g. operating time.
Use of the Service II menu
The functions in the Service II menu directly influence the opera-
tion and monitoring functions of the instrument!
Reset operating time
After replacing a dosing pump or the measuring chamber holder, it is possible to reset the current operating time to 0 hours:
In , select
=> =>
Select "Reset" to reset the operating time Press "ENTER" to confirm the selection
The operating time "000000h" appears on the display
Maintenance interval
Observance of the maintenance intervals is monitored and displayed by Testomat 2000®. Program the desired maintenance interval in days here. (0 days equals no maintenance interval)
In , select
=> =>
Enter the maintenance interval in days (D) Press "ENTER" to confirm the entry
CAUTION
▼▲
▼▲
▼▲
▼▲
Description of the signal inputs/outputs
31
Description of the signal in­puts/outputs
Connecting the signal inputs
Only connect the signal inputs "Start", "Stop", "IN1" and "IN2" with
volt-free contacts! The connection of external voltages would damage the instrument!
Function
Test time
Action
Start
External analysis triggering (only normally
open)
None
In EXTERNAL operating mode, an analysis is started by triggering a contact at the input.
Permanent contact results in regular analyses.
External analysis triggering occurs via a contact at the start input. Note: The current analysis interval can be interrupted by triggering a
contact at the stop input.
Function
Test time
Action
Stop
External analysis stop (e.g. via flow controller or pro­cess controller)
None
As long as the contact at the input is 'open' or 'closed', no analyses are carried out
An active Stop input prevents an analysis start, e.g. via a current interval. This can be necessary if the plant does not supply water. A current analysis is stopped when the input valve is opened (while the measuring chamber is being flushed or filled). The water which has already entered the measuring chamber remains there. If the measuring chamber is already full, the analysis is executed. Manual start has priority over the Stop input, i.e. if the Stop input is active, an analysis can be started manually or a manually started analysis cannot be stopped by the stop signal. In the operating mode “time-controlled”, the interval time continues when the Stop input has been activated.
In , select
=> =>
Select the type of contact Press "ENTER" to confirm the selection
Start
Terminals 20,21
External
Analysis start
Stop
Terminals 22,23
▼▲
CAUTION
!
Description of the signal inputs/outputs
32
Function
Test time
Action
IN1
- Measurement limited to parameter 1 or measuring point 1
Fixed, 10 sec­onds
Selection "1 measuring point":
Only parameter 1 is measured as long as the contact is open or closed.
Selection "2 measuring point":
Only measuring point 1 with pa­rameter 1 is measured as long as the contact is open or closed.
In , select
=> =>
Select the type of contact Press "ENTER" to confirm the selection
Function
Test time
Action
IN2
Water meter input
None
Quantity recording for start­ing an analysis
Function
Connection
Action
OUT
Programmable interface 0-20 mA or 4-20 mA
max. load 500 Ohms
OR Programmable voltage
interface 0 - 10 V or 2 - 10 V
OR Serial interface RS 232
Serial bus
(2-wire cable)
See description of inter­face card RS 910
(See Technical info "T2000­ RS232")
The section "Interfaces" contains a detailed description
IN1
Terminals 24,25
▼▲
IN2
Terminals 26,27
OUT
Terminals ┴, 28,29
Description of the signal inputs/outputs
33
Interfaces (optional)
Current interface 0/4-20 mA
Current interface load
The maximum load of 500 Ohms should not be exceeded!
In case of possible faults and when using very long cables (approx. 20 m), a screened cable should be used, if possible.
Installing the interface cards SK910/RS910
Avoid static charges during installation! Personnel responsible for installation must be appropriately trained
and qualified!
Switch off the instrument. Insert the plug-in circuit board into the left-hand slot with the compo-
nent side on the left. (Contact no.1 is at the top) Switch on Testomat 2000®. Program the desired type of current (0/4-20mA) (for SK910).
Monitoring a measuring point
A printer can be connected to record the analysis results. The instru­ment is equipped with a programmable current output for this pur­pose (optional 0-20 mA or 4-20 mA).
The example on the left displays the current profile in the 0-20 mA range for one measuring point.
In , select
=> =>
Select the desired current range Press "ENTER" to confirm the selection
In , select
=> =>
Select the desired configuration Press "ENTER" to confirm the selection
Monitoring of two measuring points
The example on the left displays the current profile for 4-20mA and the use of two measuring points.
Measuring point 1 and measuring point 2 are measured alternatively. The measured value 1 is shown in display line 2 (M1:) and the meas­ured value 2 in line 3 (M2:). . The currently analysed measuring point is indicated by an asterisk on the right.
NOTE
▼▲
▼▲
Base circuit board Testomat
WARNING
!
Description of the signal inputs/outputs
34
The range of the current interface is divided. The range 4 – 12 mA is available for the measured value from measuring point 1 and range 12 – 20 mA for the measured value from measuring point 2.
In , select
=> =>
Select the desired current range Press "ENTER" to confirm the selection
In , select
=> =>
Select the desired configuration Press "ENTER" to confirm the selection
Calculating the output currents
The entire current range (0 – 20 mA or 4 – 20 mA) is available for a measuring point. The current range is divided for two measuring points. The value of measuring point 1 is displayed in the lower half ( 0 – 10 mA or 4 – 12 mA), and the value of measuring point 2 is dis­played in the upper half (10 – 20 mA or 12 – 20 mA).
Measured value Current = x 20 mA Maximum value
Measured value Current = x 16 mA + 4 mA Maximum value
Measured value 1 Current 1 = x 10 mA Maximum value
Measured value 2 Current 2 = x 10 mA + 10 mA Maximum value
Measured value 1 Current 1 = x 8 mA + 4 mA Maximum value
Measured value 2 Current 2 = x 8 mA + 12 mA Maximum value
The current is set to 0 or 4 mA. (for one measuring point)
The current is set to 20 mA.
Measured value = Value displayed in the selected hardness unit Maximum value = Final value of the used indicator
(e.g. indicator type 2005 = 0.5 °dH)
1 measuring point
0 - 20 mA
1 measuring point
4 - 20 mA
2 measuring points
0 - 20 mA
2 measuring points
4 - 20 mA
Measuring range not reached (e.g. <0.05 °dH)
Measuring range ex­ceeded (e.g. >0.5 °dH)
▼▲ ▼▲
How is the current
calculated for a cer-
tain measured value?
Description of the relay outputs
35
Serial interface RS232
Testomat 2000® DUO can also be connected to a log printer via the serial interface RS232 to enable the printout of measuring results and error messages. Analyses can then be continuously logged. This option is only possible in connection with the interface card RS232 (Art. no. 270310).
In , select
=> =>
Select the desired interface Press "ENTER" to confirm the selection
Description of the relay outputs
All relay outputs are neutral contacts. This ensures that all connection options are available. The switching of mains voltage and external voltage, and the direct switching of inputs, e.g. a process controller, can be realised.
Flushing (external flush valve)
Immediately before each analysis the external flush valve is opened for the programmed period allowing the line up to Testomat 2000® DUO to fill with measuring water. Please ensure that the programmed flush time is sufficient.
Please refer to "Entering further basic program data" "External flush" for programming details
LV1 and LV2 limit value outputs
Two volt-free relay contacts are available to signal that a limit value has been exceeded. The limit values, the hysteresis and the function can be freely programmed for both contacts:
Function
Contact
Action
LV1
Relay switches if limit value 1
or measuring point 1 have been exceeded
Volt-free change­over contact
programmable:
- Continuous contact
- Impulse (1 - 99 seconds/minutes)
- Interval (1 - 99 seconds/minutes)
- Hysteresis (1, 2 or 3 limit values exceeded)
▼▲
Flush valve
Terminals 1,2
Limit value 1
Terminals 3,4,5
Description of the relay outputs
36
Function
Contact
Action
LV2
Relay switches if limit value 1
or measuring point 2 have been exceeded
Volt-free change­over contact
programmable:
- Continuous contact
- Impulse (1 - 99 seconds/minutes)
- Interval (1 - 99 seconds/minutes)
- Hysteresis (1, 2 or 3 limit values exceeded)
For a detailed description and programming refer to the section enti­tled "Switch functions of the limit value outputs LV1 and LV2"!
Measuring points 1 or 2 (measuring point switch­over)
The Testomat 2000® DUO can be used to monitor two measuring points. Two external valves that can be controlled via the terminals 9, 10, 11 are required for this task. The menu item "2 measuring points" must be selected for this option.
Measuring point switch-over occurs automatically alternating with the analyses. The parameters are strictly allocated to the measuring points:
Measuring point 1 = Parameter 1 (indicator type 1) Measuring point 2 = Parameter 2 (indicator type 2)
In , select
=> =>
Select "2 measuring points" Press "ENTER" to confirm the selection
Measuring point switch-over can be surpressed by a contact (or nor­mally closed contact) at input IN1. The active status of IN1 has to be programmed accordingly.
IN1 active = Measurements only from measuring point 1 A switch-over of the measuring points also only occurs after comple-
tion of the "Hysteresis" function.
Limit value 2
Terminals 6,7,8
Measuring point switch-over
Terminals 9,10,11
▼▲
Description of the relay outputs
37
AUX (Programmable function output)
The functioning of this volt-free relay output is programmable:
1. For reporting a current analysis and/or
2. For contact prior to an analysis, e.g. to operate a cooler or
3. Contact after an analysis
In , select
=> =>
Select the program step at which the AUX contact is to be activat-
ed
Under "Time" enter the contact duration in minutes (m) and sec-
onds (s)
Press "ENTER" to confirm the entry
Alarm (fault message output)
The "Alarm" output is a volt-free change-over relay contact. During trouble-free operation, the contact between the terminals 15 – 16 is closed and the one between terminals 14 – 16 is open. In case of a voltage breakdown, the contact between the terminals 14 – 16 is closed and the one between terminals 15 – 16 is open.
The instrument is equipped with a range of monitoring functions. You can define the individual statuses as a fault and program the corre­sponding message either as a continuous contact (A) or as a mes­sage impulse (M).
Functions/Behaviour of the "Alarm" output:
- With a continuous contact, the "Alarm" output remains activated
(terminals 15 – 16 closed) as long as the fault persists.
- With a message impulse, the output is switched 'on' for 2 sec-
onds and then switched 'off' for 5 seconds.
- If several faults with differently programmed messages are sig-
nalled simultaneously, the output is switched to continuous con­tact.
- The red LED "Alarm" and the text on the display indicate a fault.
- The fault message signal at the "Alarm" output is deleted by con-
firming the fault via the "Horn" key.
- The error message can only be deleted if the fault has been elim-
inated.
- Exception: The maintenance date has been exceeded. This mes-
sage is confirmed in the M menu, see below (Maintenance).
- Each new fault is entered into the error history (also see "i
menu").
- There is no additional alarm via the fault message output when
the limit value is exceeded!
The error messages are described under "Error messages / Trouble­shooting"
Alarm
Terminals 14,15,16
Always fault message for:
Power failure Low water level Function fault optics Measuring fault analysis Function fault dosing pump Function fault drain out­let Function fault failure 24V
Programmable fault messages for:
Low indicator level Function fault dosing error Function fault soiling Measuring fault turbid Transmission error Measuring range ex­ceeded Maintenance exceeded
The following faults activate the "Alarm" output and are displayed:
AUX
Terminals 12,13
▼▲
Description of the relay outputs
38
Maintenance (output for maintenance message)
The "Maintenance" output is a volt-free change-over contact. During trouble-free operation without a programmed maintenance interval, the contact between the terminals 17 – 19 is closed and the one be­tween terminals 18 – 19 is open.
The instrument is equipped with a range of monitoring functions and a programmable maintenance interval. The respective maintenance message is always a continuous contact.
A maintenance request is displayed via the yellow "Maintenance" LED. The maintenance display can only be deleted once the status has been corrected or after the maintenance request has been con­firmed.
Please refer to the section "Password protection and basic program­ming" for further programming details.
Maintenance
Terminals 17,18,19
Activation of the maintenance output for:
Low indicator level Function fault dosing error Function fault soiling Maintenance date reached
Information menu "i"
39
Information menu "i"
In the information menu, it is possible to request active settings and statuses of the instrument, the error history, the date for the next maintenance and the customer service address.
Call (1)
Use the key to open the information menu "i".
Request options: Customer service, operating values, program val­ues, error history, maintenance
Please refer to the section "Password protection and basic program­ming" for further programming and setting details for the individual menu items.
Operating values (3)
Display of the current values.
Program values (4)
Use the arrow keys to open the menu item "Program values". Press the "ENTER" key to open the list of the set values. Press the "ENTER" key to request the current setting of a parameter
An asterisk indicates the se­lected functions. (There are no active lines.)
Error history (5)
Use the "i" and "ENTER" keys to open the error history. The error history is a list of errors or statuses which have occurred during operation. The list is deleted after a power failure and a new list is started.
If no errors have occurred since start-up, the last switch­on time of the unit is displayed, e.g.:
POWER FAILURE from 16.06.09 06:56 to 16.06.09 07:09
Maintenance (6)
Display of the next mainte­nance date and the pro­grammed maintenance interval. It is possible to set the mainte­nance interval in the basic program (password protected). Refer to the section entitled "Maintenance" for further de­tails on maintenance intervals.
Customer service (2)
Display of the customer service address or e.g. a service hot­line number. You can freely program these three lines in the basic progam (password protected).
Program menu "M"
40
Program menu "M"
Call: (1)
Use the key to open the program menu "M". It is possible to call up all the functions without password protection
except for the basic program.
Programming of: Indicator, manual mode, flushing, flush chamber, drain chamber, fill chamber, self-test, confirm maintenance, diagno­sis, date, time, basic program with password
Availability of functions
All manual functions can only be selected during an analysis
pause. Analyses are not carried out during manual operation. All signal inputs and outputs are locked.
Service I (2)
Input indicator (3)
Enter the new filling levels after each refill or indicator bottle change. Once you have select­ed the menu item for entering the filling level "Indicator filling (0 - 100%)" via the "ENTER" key, the value is preset to 100%. If you have connected a full bottle, press "ENTER" to confirm the value. If the filling level of the bottle differs, enter the corresponding value.
Manual operation (4)
After confirming the information message (4) via the "ENTER" key, it is possible to select and activate the desired function by using the arrow keys and pressing the "ENTER" key. These functions are used for checking the functions and for commissioning.
Fill chamber (8)
Press the "ENTER" key to fill the measuring chamber.
Flush (5)
Press the "ENTER" key to start the flushing of the sampling line via the internal valves. Press the "ENTER" key again to cancel this function.
Flush chamber (6)
Press the "ENTER" key to flush the measuring chamber once.
Drain chamber (7)
Press the "ENTER" key to open the outlet valve in order to drain the water from the measuring chamber. Press the "ENTER" key again to cancel this func­tion.
NOTE
Program menu "M"
41
Self-test (9)
Press "ENTER" to start the functional test of Testomat 2000® DUO. The program checks all relevant instrument functions and carries out an analysis. A respective message appears after an error-free test.
Press the "ENTER" key again to cancel this function and to return to the "MANUAL MODE" menu.
Confirm maintenance (10)
After maintenance has been carried out, confirm it by pressing the "ENTER" key and exit this item via the "M" key. The maintenance interval is restarted.
Confirm a maintenance request once the maintenance interval has expired in the M menu. The displayed message is deleted and the “maintenance” output reset.
Refer to the section entitled "Maintenance" for further details on maintenance intervals.
Diagnosis (11)
It is possible to request a list of current statuses of the signal inputs and outputs. Active statuses are marked with an *. (see "Structure of the basic program").
The current interface can be checked under "OUT output". Press the "Enter" key to toggle between minimum and maximum current. Change-over occurs between 000 and 200 at 0-20 mA!
Time/Date (12)
Set the time and date by selecting and activating the desired function
via the arrow keys and the "ENTER” key. Subsequently press the "M"
key again to save the setting and to return to the display function. Refer to "Functions of the operating and display elements" "Oper-
ating system" for more details
Basic program
This menu item can only be accessed after entering the password! After entering the password and confirming it via the "ENTER" key, it
is possible to carry out basic programming of the instrument and to select various service functions (e.g. calibration).
In the basic program, the following abbreviations are used in the re­spective menu items:
s = seconds; m = minutes; h = hours; T = days; l = litres
▼▲ ▼▲
Call the factory de­fault setting:
Press the "M" and "i" key and switch on Testomat 2000® DUO. Caution: All previously entered data is overwrit­ten!
The values and settings of the basic default set­ting are described in the "structure of the basic program".
Program menu "M"
42
Structure of the basic program
Call the basic factory program by simultaneously pressing and holding down the "M" and "i" keys while switching on the instrument. CAUTION, the most recent programming will be lost!
Error messages / Troubleshooting
43
Error messages / Troubleshooting
Displayed message (flashes at selected display)
Instrument result functions
Possible causes
Remedies
Ff. POWER FAILURE 24 V
CANCEL WITH HORN KEY
- After programming: Continuous alarm or message impulses
- Standby
- Internal power failure of the
24 V supply
Replace fuse F4 or F8
(The control lamp "Power"
of the dosing pump should illuminate)
Ff. DOSING PUMP
CANCEL WITH HORN KEY
- After programming: Continu­ous alarm or message im­pulses
- Standby
- Dosing pump is defective
- No dosing message from
dosing pump
Replace dosing pump Check cable to the dosing
pump for correct connec tion .
Mf. TURBID
CANCEL WITH HORN KEY
- After programming: Con­tinuous alarm or message impulses or no message
- Continue measurements
- The water is turbid / soiled
MEASURING RANGE EXCEEDED
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses or
no message
- Continue measurements
- The measuring range has been exceeded
Select another type of
indicator (basic program)
LOW WATER LEVEL
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses
- Standby
- no water input although
LED "IN" lights up
- Inlet pressure too low
- Overflow detection does not react
Check water inlet Connector at the inlet
valve oxidised
Clean filter strainer Replace valve block
Extract pressure control ler valve body
Replace fuse F6
Ff. OUTLET TO DRAIN
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses
- Standby
- Water remains in the meas­uring chamber although LED "OUT" illuminates
Check water outlet Connector at the outlet
valve oxidised
Replace valve block
LOW INDICATOR LEVEL
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or mes
sage impulses or no message
- LED and output "mainte
nance" on
- Continue measurements
- Minimum indicator quantity
not reached Without BOB: 50 ml (10%), With BOB: According to
calculation
Check indicator level and,
if necessary, refill (enter the filling quantity!):
Mf. SOILING
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses or no message
- LED and output "mainte-
nance" on
- Continue measurements
- Sight-glass windows are soiled
Clean sight-glass windows
Ff. OPTICS
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses
- Standby
- Plug-in circuit board defec-
tive
- Error at the optical compo­nent (transmitter or receiver defective)
Replace plug-in circuit
board
Measuring chamber
holder Abbreviations: Ff.: = function fault, Mf. = Measuring fault
Error messages / Troubleshooting
44
Displayed message (flashes at selected display)
Instrument result functions
Possible causes
Remedies
Mf. ANALYSIS
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses
- Standby
- Air inside the dosing hoses
- Incomplete mixing
- Indicator out of date or use of a third-party indicator
Retighten connections of
the dosing pump(s)
Replace suction insert in
the bottle(s)
Replace stirring bar Replace indicator, only
use HEYL Testomat 2000® indicator
Ff. DOSING ERROR
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses or no message
- LED and output "maintenance" on
- Continue measurements
- Dosing accuracy of the dosing pump
Replacing dosing pump or
return for calibrating
MAINTENANCE INTERVAL EXCEEDED BY
XXX DAYS
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses or no message
- LED and output "mainte nance" on
- Continue measurements
- Programmed maintenance date reached or exceeded
Carry out maintenance
and subsequently cancel or confirm
Abbreviations: Ff.: = function fault, Mf. = Measuring fault
Further information
Error
Possible causes
Remedies
Current interface functions incorrectly
- Incorrect measured value at the output or
no power supplied
Replace fuse F7 Replace the interface circuit board
Unit is not functioning, even though it is switched on No display
- Fuse F9, F5 or F2 (240 V: F1) defective
- Power switch defective
- Ribbon cable at display circuit board or
base circuit board is loose
- Error at display circuit board or base
circuit board
Replace fuses Replace power switch Reconnect ribbon cable Replace display or base circuit board
Response of a protective circuit
After a protective circuit (fuse) has been tripped, attempt to eliminate the cause of malfunctioning (e.g. replace a defective valve) before reactivating the protective circuit. Frequent tripping is usually due to power overload which, in certain circumstances, may also damage the instrument.
Malfunctioning/Repairing a defective instrument: The repair of a defective instrument – irrespective of the warranty period - is only possible after the instrument has been dismantled and returned to us with a description of the error. Furthermore, please inform us of the indicator type being used and the measured medium. Before you return the instrument for repair work, remove the bottle and ensure that the measuring chamber has been flushed out and is empty.
Maintenance
45
Maintenance
Required maintenance measures
Regular maintenance is necessary to ensure trouble-free opera-
tion of the instrument!
Please regularly carry out the maintenance work described in the following section when
the programmed maintenance date has been reached (display
"maintenance date exceeded")
the instrument displays the following error messages:
"Mf. soiling" or "Low indicator level"
the last maintenance was carried out max. 6 months ago
Cleaning measures
Never use organic solvents to clean the measuring chamber or
other plastic parts!
Please observe the safety regulations when handling cleaning
agents!
If the measuring range of the instrument is exceeded over a longer
period of time, a coloured film may form on the sight-glass win-
dows. Use alcohol to remove this sticky film.
Description of maintenance work
The maintenance manual Teastomat 2000®/Testomat ECO® contains a detailed description of maintenance work. The measures described here provide a brief overview.
Cleaning the measuring chamber and sight-glass windows
Switch off the instrument or press the "STANDBY" key . If re-
quired, remove any water from the measuring chamber:
Close the manually-operated valve of the branch line to Testomat
2000® DUO.
Unhook the toggle type fastener , tip the measuring chamber
upwards and remove it.
Release both sight-glass window holders , remove and clean the
sight-glass windows.
Use alcohol to clean off the film on the sight-glass windows. If the
instrument has been used with hard water for a longer period of
time (measuring range exceeded!), a hard-to-remove film may
have formed on the sight-glass windows. In this case, clean the
sight-glass windows as described below for cleaning the measur-
ing chamber.
NOTE
Scan the code to download the latest service manuals!
CAUTION
!
Maintenance
46
The measuring chamber can be cleaned with a cleaning agent
suitable for decalcification and rust removal. Flush the measuring
chamber thoroughly after cleaning.
After cleaning, re-insert the sight-glass windows and secure them
with the sight-glass window holders (do not forget the flat seals
and ensure correct fitting in the groove).
Re-insert the measuring chamber and use the toggle type fastener
to secure it.
Cleaning the filter housing
Close the manually-operated valve of the branch line to Testomat
2000® DUO.
Depressurise the lines of Testomat 2000® via the “Flush” function:
Switch off the instrument and loosen the hose connections at the
filter housing.
Unscrew the inlet connection using an open ended spanner (size
22) and remove the seal, spring and filter.
Remove the retaining pin and extract the flow controller, subse-
quently remove the flow controller valve body.
Clean the filter housing with water or alcohol; then reassemble the
unit.
If required, replace the seals. Insert the filter strainer with the cone facing downwards! Re-attach the hose connections at the filter housing.
Important maintenance information
Water leakage at sealed points can damage parts of the instrument! Please check the instrument for leaks prior to the first analysis.
Switch the instrument to STANDBY Manually fill the measuring chamber Manually dose the indicator ("Manual" key)
Check the connections and seals for leaks
Service instructions
The surface of the instrument has not been treated. Therefore, avoid any soiling caused by indicators, oil or grease. However, if the hous­ing becomes soiled, please clean the surface with isopropanol (never use other solvents).
Zulaufanschluß Inlet connector
Regler-/Filteraufnahme Regulator/filter
CAUTION
!
Testomat 2000® spare parts and accessories
47
Testomat 2000® spare parts and ac­cessories
Art. No.
Pressure controller
40125
Controller / Filter receiver, complete
40120
Controller / Filter receiver
40129
Controller plug T2000, complete
11225
Flow controller valve body, complete
11230
Retaining pin 3x38 90°
11217
Inlet filter 19.5dx25
11218
Spring for inlet filter
40121
Inlet connector
40153
Screw-in connector G 1/4” -6
40157
Angled plug-in connector G 1/8”
Measuring chamber
40173
Sight-glass window with seal, T2000
40170
Sight-glass window 30x3
40176
Sight-glass holder, countersink and thread.
33253
Bolt M3x40, A2, DIN 965
40032
Latch fastener TL-17-201-52
11203
Plastic plug 5.3dx5 PE natural
40022
Measuring chamber T2000, complete
Measuring chamber holder
40370
Measuring chamber holder, complete (without valves)
40050
Magnetic stirrer
40186
Screw-in connector 3/8" -10
40018
Solenoid valve, 2/2-way
40181
Rear guide bar for measuring chamber 5 x 60
Dosing pump DOSIClip®
40001
Jet pump SP
40011
Suction hose, complete
40016
Pressure hose, complete
37232
Base circuit board TI (3)
40040
Valve set
32046
Plastic cover CNH 45 N
Bottle connection / Suction device
40131
Screw cap with bottle insert T2000
40130
Screw cap GL32 - hole
40135
Bottle insert for screw cap with push-fit suction tube
Art. No.
Unit spare parts list
31582
Fuse GS-M 5x20E 4A
40294
Base circuit board T2000, complete 230V
40092
Control circuit board T2000, complete
40091
Plug in circuit board driver/receiver SE­T2000 (6)
40190
Cable sleeve 5-7, grey
40191
Cable sleeve 7-10, grey
31713
Ribbon cable 10 pole with EMI filter clamp
40096
Ribbon cable 26 pole with EMI filter clamp
40060
Cable loom 2V for T2000
40062
Cable loom 2P for T2000
40200
Cable loom complete with power switch and cover
31596
Fuse, soldered T0.16A
31585
Fuse, soldered T0.315A
31595
Fuse, soldered T0.1A
31622
Fuse, soldered T0.16A
31592
Fuse, soldered T1.016A
Spare parts requirement for 2 - 3 years of
operation
40173
Sight-glass window with seal, T2000
11217
Inlet filter 19.5dx25
40124
Gasket set T2000
31585
Fuse, soldered T0,315A
31592
Fuse, soldered T1.016A
Testomat 2000® spare parts and accessories
48
Accessories
Indicator type
Range
Art. no.:
TH2005
Water hardness 0.05 – 0.5 °dH
152005
TH2025
Water hardness 0.25 – 2.5 °dH
152025
TH2100
Water hardness 1.0 – 10.0 °dH
152100
TH2250
Water hardness 2.5 - 25 °dH
152250
TC2050
Carbonate hardness
153050
TC2100
Carbonate hardness
153100
TM2005
minus m-value
154005
TP2010
p-value
155010
TP2100
p-value
155100
Please refer to our delivery programme for an up-to-date overview of available accessories.
*) retrofit kit for water inlet (Art. no. 040123)
If fabric-reinforced pressure hoses (e.g. for existing installations) are used, please replace the plug connector at the controller and filter housing with a plug for the quick-release coupling (not included
).
Art. no.
Description
040123
Retrofit kit for water inlet T2000 *)
040315
Discharge hopper for Testomat 2000/ECO
270305
Current interface card 0/4-20 mA, SK 910
270310
RS232 interface card for log printer, RS 910
270315
Voltage interface card 0/2 - 10 V, UK 910
270500
Log printer
270501
Power supply unit for log printer
270410
Booster pump
270335
Maintenance case T2000 Heyl
Technical data
49
Technical data
Power supply:
230 VAC, 115 VAC or 24 VAC ± 10%, 50 ­60 Hz
Instrument fuse 230 V: T0.1A Instrument fuse 115 V: T0.2A Instrument fuse 24 V: T1.0A
Power consumption:
max. 30 VA, without external load
Protection class:
I
Degree of protection:
IP 65
Conformity:
EN 50081-1, EN 50082-2, EN 61010-1
Ambient temperature:
10 - 45°C
Measuring range:
See section "Performance specifications"
Current interface:
0/4 - 20 mA, max. load 500 Ohm
Protocol printer:
See section "Accessories"
Dimensions:
W x H x D = 380 x 480 x 280 mm
Weight:
Approx. 9.5 kg
Other:
The instrument is non-volatile
Water connection
Operating pressure:
1 to 8 bar / 1x105 to 8x105 Pa or
0.3 to 1 bar / 0.3x105 to 1x105 Pa
(Remove flow controller valve body 11225!)
Water inlet:
Opaque pressure hose with an external diameter of 6/4x1 mm
Water outlet:
Opaque hose with an internal diameter of 14 mm
Water temperature:
10 to 40 °C
* When using Testomat 2000® at a pre-pressure of 0.3 bar, it must
be ensured that the flow rate through the measuring chamber is at least 400 ml/min.
We reserve the right to make technical changes without notice in the interest of constantly improving our products! Our manuals are updated regularly. If you have an older version (see version at the back of the manual) you will find the current manual on our website www.heyl.de on the download page.
Technical data
50
Conformity Declaration
Check List Testomat 2000®
Dear customers and service technicians, This check list cannot replace your expertise or extensive experience in fault resolution. It is intended to support fast and systematic error diagnosis and error documentation. This list does not claim to be complete. We are therefore always grateful for any advice and information you may be able to provide. General user instructions can be found on the rear of this check list.
The Instrument Manufacturer Block 1 / Plant and instrument data
Testomat 2000®
Testomat® ECO
Plant type
Instrument type
Instrument no.
Indicator type
Software status
Pump no.
Block 2 / Error message and error history Please mark appropriately (X)
What does your instrument’s error history display? (“i” and “Enter” key => operating instructions)
( Error history text )
Does an error message appear on the display?
For example, “Mf. analysis”, “Low water level”, etc. (See operating instructions: “Error messages / Troubleshoot- ing”)
Yes No
( Error message text )
Block 3 / Visual inspection and functional test Please mark appropriately (X) If applicable, values / comments
Is the instrument connected to the mains voltage specified on the rating plate?
Yes No
Does a message appear on the display?
Yes No
Does the instrument display a plausible measured value? (possible manual measurement ___________ value)
Yes No
Measured value: Are the measuring chamber and sight-glass windows clean?
Yes No
Are the measuring chamber and the water-carrying hoses free of leaks?
Yes No
Is the indicator’s expiry date still valid? (See expiry date printed on the indicator bottle)
Yes No
Expiry date:
Has the correct indicator type been programmed? (TH 2025 => 0.25 to 2.5 °dH = factory setting)
Yes No
Type:
Is the water pressure within the specified range (400 ml/min)? (See the type plate on the instrument)
Yes No
Plant pressure:
Does the drain hose prevent the risk of backwater?
(No “syphoning effect”!!)
Yes No
Is the drain hose free of blockages? (Microorganisms caused by contamination, etc.)
Yes No
Does the set flushing time/quantity of flush water ensure that only freshwater is measured?
Yes No
Flushing time:
Are the hoses at the dosing pump free from air bubbles? (Operate the pump manually / Carry out a manual analysis)
Yes No
CARRYING OUT A (MANUAL) ANALYSIS
Does the water column rise evenly up to the overflow borehole when filling the measuring chamber (5 mm below the upper edge of the measuring chamber)? (If not: check the water pressure, water throughput/flow controller)
Yes No Does the indicator pump dose correctly when starting an analysis? (LED at the pump illuminates!)
Yes No
No. of dosing strokes:
Have the indicator and water been mixed properly in the measuring cham­ber after the dosing process?
Check the magnetic stirring bar! =>see maintenance manual “Adjust mode”
Yes No
PROGRAMMING DATA / OPERATING CONDITIONS
Have the limit values been set correctly? (Within the measuring range/according to the performance limit of the plant?)
Yes No
Limit values:
Is the Testomat instrument always supplied with mains voltage – except during maintenance work/emergency situations?
(Temporary shutdown only via the “Standby” key or the “Stop” input!)
Yes No
See the “General instructions for operat­ing Testomat 2000® and Testomat®
ECO”
Please refer to “Error messages / Troubleshooting” in the operating instructions for further information on error messages and possible causes
of faults.
Further functional tests (e.g. overflow detection and amplification setting => “Special function Adjust mode”) and service instructions can be found in the maintenance manual.
After completing these checks, experience shows that it can be assumed that the checked functions (Block 3) are in effective working order if you have answered all the questions with “Yes”. We recommend you to carry out these checks during each inspection or if faults occur.
Instrument settings
Caution!
Your settings may be deleted if repairs are carried out. Therefore, note down your instrument settings in the table below before sending the instrument to our service team for repairs. Please enclose a copy of the table with the instrument. If you have noted down the settings, they can be easily re-entered by your service staff once any repairs have been completed.
Menu
Setting
MODE OF OPERATION
Time-controlled
Volume interval
External (Start)
DISPLAY UNIT 1
Display in °dH
Display in °f
Display in ppm CaCO3
Display in mmol/l
DISPLAY UNIT 2
Display in °dH
Display in °f
Display in ppm CaCO3
Display in mmol/l
TYPE OF REAGENT 1
500ml-bottle
100ml-bottle
TH2005 Water hard.
TH2025 Water hard.
TH2100 Water hard.
TH2250 Water hard.
TC2050 Carbon. hardn.
TC2100 Carbon. hardn.
TM2005 minus-m-value
TP2010 p-value
TP2100 p-value
TYPE OF REAGENT 2
500ml-bottle
100ml-bottle
TH2005 Water hard.
TH2025 Water hard.
TH2100 Water hard.
TH2250 Water hard.
TC2050 Carbon. hardn.
TC2100 Carbon. hardn.
TM2005 minus-m-value
TP2010 p-value
TP2100 p-value
LIMIT VALUES
Limit val. 1:
Limit val. 2:
FLUSH TIMES/INTERVAL
Internal flush time
External flush time
Interval pause
MEASURING POINTS
1 Measuring point
2 Measuring points
TYPE OF WATER METER
1 Litre/Impulse
2,5 Litres/Impulse
5 Litres/Impulse
10 Litres/Impulse
100 Litres/Impulse
500 Litres/Impulse
1000 Litres/Impulse
PLANT CONTROL
Min. Res. Quant.
Limit val. 1:
Limit val. 2:
FUNCTION LV1
Duration
Impulse
Interval
Time:
FUNCTION LV2
Duration
Impulse
Interval
Time:
HYSTERESIS LV1
Analysis (1,2,3)
HYSTERESIS LV2
Analysis (1,2,3)
ALARM/MESSAGE
Reagent low level
Low water pressure
Mf. analysis
Ff. optics
Ff. dosing fault
Ff. dosing pump
Ff. outlet to drain
Mf. dirtiness
power failure 24 V
Mf. turbid
Transfer error
Meas. range exceeded
Maint. int. exceeded
FUNCTION IN1
Normally open contact
Normally closed contact
FUNCTION STOP
Normally open contact
Normally closed contact
INTERFACES
Type 0-20 mA
Type 4-20 mA
Type RS232
FUNCTION AUX
Contact before analysis
Contact during analysis
Contact after analysis
Time:
OPERATING TIME
MAINTENANCE INTERV.
CUSTOMER SERVICE
Technical data
Product overview Testomat 2000®-Instruments
Model/Type
Measuring Parame­ter
Measuring Range
Applications/Functions
Testomat 2000®
• Water hardness
• Carbonate hard-
ness
• p-value
• minus-m-value
0.05-25 °dH 0,5-20 °dH 1-15 mmol/l
0.05-0.5 mmol/l
• Universal for water treatment
plants
• allowed for boiler houses
Testomat 2000® Antox
as Testomat 2000®
as Testomat 2000®
• dosing a reducing agent
Testomat 2000® CAL
as Testomat 2000®
as Testomat 2000®
• Automatic calibration function
Testomat 2000® CLF
• Free Chlorine
0-2.5 mg/l
• DPD-method for swimming pool and drinking water control
Testomat 2000® CLT
• Total Chlorine
0-2.5 mg/l
• DPD-method for swimming pool and drinking water control
Testomat 2000® CrVI
• Chromate
• Chrome-VI
0-2.0 mg/l 0-1.0 mg/l
• process control of waste water
in galvanic industry
Testomat 2000® Duo
as Testomat 2000®
as Testomat 2000®
• Controlling of two measuring
points
Testomat 2000
®
Fe
• Iron-II and Iron-III
0-1.0 mg/l
• De-Ironing plants
Testomat 2000
®
SO3
• Sulphite
0-20 mg/l
• Control of the Oxygen-binding by Sulphite in boiler feed wa­ter
Testomat 2000
®
self clean
as Testomat 2000®
as Testomat 2000®
• Automatic cleaning of the
measuring chamber
Testomat 2000 THCL®
• Total Chlorine
• Water hardness
0-2.5 mg/l
0.25-2.5 °dH
• DPD-method for swimming pool and drinking water control
• combination system for hard­ness and chlorine
Testomat 2000
®
V
• Water hardness
• Carbonate
hardness
1.0-25.0 °dH
1.0-20.0 °dH
blending water
54
Gebrüder Heyl Analysentechnik GmbH & Co. KG Orleansstraße 75b D 31135 Hildesheim www.heyl.de
Testomat_2000_DUO_GB_130705.doc
Scan the code and visit us on our website!
Loading...