PRESS* TO SELECT FIXED STEPS
USE OR TOADJUST OUTPUT
a
Druck
a
ADruckCompany
Sensing & Inspection Technologies
Druck UPS-II
Loop Calibrator
User manual - K0434
Approved Service Agents
For the list of service centres visit our web site:
www.gesensinginspection.com
Symbols
This equipment meets the requirements of all
relevant European safety directives. The
equipment carries the CE mark.
Do not dispose of this product as household
waste. Use an approved organisation that
collects and/or recycles waste electrical and
electronic equipment. For more information:
Contact us at www.gesensinginspection.com
This instrument is fitted with four size AA batteries either
rechargeable (nickel cadmium) or non-rechargeable
(alkaline).
Before storing this instrument remove the batteries.
When fitting batteries make sure the electrical contacts are
clean and observe the correct polarity.
The battery compartment should be inspected for corrosion
caused by leaking batteries. Corrosion must be removed
using approved methods*.
When storing and transporting batteries make sure that they
cannot be short circuited. A short-circuited battery can
become very hot and can, in certain circumstances, explode.
It is recommended that a suitable container is used for
storing and transporting batteries.
Dispose of old batteries using a safe, approved method.*
The UPS-II has special functions for fixed steps and readings
in %. To open menu press
contents with key and make your choice. Press
confirm your choice.
MENU CHOICESWITCHDISPLAY READING IN:
4-20mA lin #out0 to 22mA or % span
0-20mA linout0 to 22mA or % span
4-20mA flowout0 to 22mA or % span
0-20mA flowout0 to 22mA or % span
4-20mA valveout0 to 22mA
4-20mA lin #read0 to 22mA or % span
0-20mA linread0 to 22mA or % span
4-20mA flowread0 to 22mA or % span
0-20mA flowread0 to 22mA or % span
Note: Functions marked with # are directly available after
selection. Flow = SQ.RT in %.
Press mA/% key to read in mA or %. Press * to select
"continuous" or "fixed steps" in output mode.
Fixed Steps
To output fixed calibration currents in series as indicated
below: choose range from the menu and select "fixed steps".
Press or to advance one step
To output these currents automatically select "fixed steps"
and press keys simultaneously. Time between steps is
10 seconds. Stepping starts at the low end of span.
To return to manual press or .
Valve mode
In the valve mode, stepping will be 3,8 - 4 - 4,2 or 19 - 20 - 21
mA. Select appropriate current to start auto-stepping.
Ramp mode
To output a continuous up/down ramp cycle press keys
simultaneously. Ramp travel time between range limits is 60
seconds cycle starts at displayed value. Press or first
to select start direction. To return to manual press or .
Batteries
Use alkaline or rechargeable NiCad batteries only. The
display shows LO BAT when battery capacity goes low.
CAUTION
DONOTLEAVEDISCHARGEDBATTERIESINTHEINSTRUMENT .
O
Make sure the
polarity of the
new batteries (3)
is correct.
UPS-II Loop Calibrator Technical Notes
Current reading
Switch positions
Function
Loop power
Read
Ext.
Transmitter Calibration
Switch positions
Function
Loop power
Read
Int.
3 K0434 Issue No. 2
Transmitter Simulation
Switch positions
Function
Loop power
Out
Ext.
Valve Stroking
Switch positions
Function
Loop power
Out
Int.
Calibration Instructions
General
A calibration period of 12 months is recommended. The
actual calibration interval depends on instrument usage and
the total measurement uncertainty acceptable for the
specified application.
UPS-II
is a precise measuring instrument and the test
The
equipment and conditions of test must be suitable for the
type of work. The calibration check and calibration
adjustment should be carried out in a controlled
environment by a calibration technician*.
The manufacturer offers a comprehensive and, if required,
UKAS accredited calibration service.
*
A calibration technician must have the necessary technical knowledge,
documentation, special test equipment and tools to carry out the
calibration work on this equipment.
Calibration Equipment
The following tables give the accuracy requirements for the
calibration equipment and the UPS-II.
Calibration requires a stable temperature of 21° ±1°C (70° ±2°F).
UPS-II measure mode
Applied mAPermitted UPS-II
00.0070
40.0070.00014
120.0070.00030
200.0070.00046
UPS-II source mode
Applied mAPermitted UPS-II
00.010
40.010.00012
120.010.00011
200.010.00015
Table 1
error (mA)
Table 2
error (mA)
Calibrator error
(mA)
Calibrator error
(mA)
5 K0434 Issue No. 2
Calibration Check
Calibration Procedure mA Input
2 seconds
Apply zero - open circuit test leads
Connect test leads to mA source
apply full-scale
Adjust mA source to any
value within range - check reading
example
serial number
READ
and
+
Enter
*
Enter
*
Enter
*
VALVE
LIN
FLOW
MA
VALVE
LIN
FLOW
MA
and
+
-
*
adjust each
digit
adjust each digit
adjust each digit
Enter
*
Input calibration complete new offset values stored
continue with output calibration.
VALVE
LIN
FLOW
MA
1. Check the instrument for accuracy in measure and
source modes, if there is an error greater than the
permitted error, the instrument requires a calibration
adjustment.
Calibration Adjustment
Note:
For complete accuracy, the measure mode and source mode adjustment
must be carried together and in the following order.
Measure mode
1. Connect the UPS-II to the electrical calibrator. Switch on
the electrical calibrator and allow it to thermally
stabilise.
2. Switch on the UPS-II and allow the instrument to
thermally stabilise.
3. Select EXT on the LOOP PWR slide switch. Use the
display menu to select the calibration values:
Source mode
Calibration Procedure mA Output
2 seconds
Connect test leads to mA-meter
compare values.
serial number
OUT
and
+
Enter
*
Enter
*
Enter
*
Enter
*
Enter
*
and
+
-
*
adjust each
digit
adjust each digit
adjust each digit
Calibration complete new output values stored
Compare values.
Check reading.
Check reading.
VALVE
LIN
FLOW
MA
VALVE
LIN
FLOW
MA
VALVE
LIN
FLOW
MA
VALVE
LIN
FLOW
MA
1. Connect the UPS-II to the electrical calibrator. Switch on
the electrical calibrator and allow it to thermally
stabilise.
2. Switch on the UPS-II and allow the instrument to
thermally stabilise.
3. Select INT on the LOOP PWR slide switch. Use the
display menu to select the calibration values:
Completion
1. Switch off and disconnect the calibration equipment.
2. Update the calibration records for the instrument.
7 K0434 Issue No. 2
Range........................................................................................ 0 to 22.2 mA
Specifications
Accuracy........................................................................ 0,05% of range