Ge Ups-ii User Manual

GE
OUT
INT
LOOP
PWR
READ
EXT
UPS - II
mA
%
*
VALVE
LIN
FLO
W
%
M
A
STEPS
FIXED
loop calibrator
RANGES: OUTPUT& REA
D0 - 22.2 mA
PRES
S* 2 SEC.
TODISPLAY MENU
SCROLLFUNCTIONS WITH KEY
PRESS* TO ENTER FUNCTION
PRESS* TO SELECT FIXED STEPS USE OR TOADJUST OUTPUT
a
Druck
a
ADruckCompany
Sensing & Inspection Technologies
Druck UPS-II
Loop Calibrator
User manual - K0434
Approved Service Agents
For the list of service centres visit our web site:
www.gesensinginspection.com
Symbols
This equipment meets the requirements of all relevant European safety directives. The equipment carries the CE mark.
Do not dispose of this product as household waste. Use an approved organisation that collects and/or recycles waste electrical and electronic equipment. For more information:
Contact us at www.gesensinginspection.com
This instrument is fitted with four size AA batteries either rechargeable (nickel cadmium) or non-rechargeable (alkaline).
Before storing this instrument remove the batteries.
When fitting batteries make sure the electrical contacts are clean and observe the correct polarity.
The battery compartment should be inspected for corrosion caused by leaking batteries. Corrosion must be removed using approved methods*.
When storing and transporting batteries make sure that they cannot be short circuited. A short-circuited battery can become very hot and can, in certain circumstances, explode. It is recommended that a suitable container is used for storing and transporting batteries.
Dispose of old batteries using a safe, approved method.*
*Refer to the manufacturer for this information:
© General Electric Company. All rights reserved.
Battery Safety
Introduction
The Druck UPS-II Loop Calibrator can supply power (source mode) and produce readings (measure mode) to perform field calibrations on 2-wire devices.
Operation
For different applications place switches in position as indicated below:
MODE OUT READ EXT. INT. Milliamp. source • Transmitter sim. • Milliamp measure • Transmitter cal.
(transmitters are 4-20mA 2 wire)
The UPS-II has special functions for fixed steps and readings in %. To open menu press contents with key and make your choice. Press confirm your choice.
MENU CHOICE SWITCH DISPLAY READING IN: 4-20mA lin # out 0 to 22mA or % span 0-20mA lin out 0 to 22mA or % span 4-20mA flow out 0 to 22mA or % span 0-20mA flow out 0 to 22mA or % span 4-20mA valve out 0 to 22mA 4-20mA lin # read 0 to 22mA or % span 0-20mA lin read 0 to 22mA or % span 4-20mA flow read 0 to 22mA or % span 0-20mA flow read 0 to 22mA or % span
Note: Functions marked with # are directly available after
selection. Flow = SQ.RT in %.
Press mA/% key to read in mA or %. Press * to select "continuous" or "fixed steps" in output mode.
Fixed Steps
To output fixed calibration currents in series as indicated below: choose range from the menu and select "fixed steps". Press or to advance one step
4-20mA lin# 4 - 8 - 12 - 16 - 20mA 0-20mA lin 0 - 5 - 10 - 15 - 20mA 4-20mA flow 4 - 5 - 8 - 13 - 20mA 0-20mA flow 0 - 1,25 - 5 - 11,25 - 20mA 4-20mA valve 3,8-4-4,2 - 12 - 19 - 20 - 21mA
* for 2 seconds. Scroll menu
* again to
out
or
read
1 K0434 Issue No. 2
To output these currents automatically select "fixed steps" and press keys simultaneously. Time between steps is 10 seconds. Stepping starts at the low end of span. To return to manual press or .
Valve mode
In the valve mode, stepping will be 3,8 - 4 - 4,2 or 19 - 20 - 21 mA. Select appropriate current to start auto-stepping.
Ramp mode
To output a continuous up/down ramp cycle press keys simultaneously. Ramp travel time between range limits is 60
seconds cycle starts at displayed value. Press or first to select start direction. To return to manual press or .
Batteries
Use alkaline or rechargeable NiCad batteries only. The display shows LO BAT when battery capacity goes low.
CAUTION DO NOT LEAVE DISCHARGED BATTERIES IN THE INSTRUMENT . O
LD BATTERIES CAN LEAK AND CAUSE CORROSION.
W
HEN FITTING THE BATTERIES MAKE SURE THE ELECTRICAL CONTACTS
ARE CLEAN AND OBSERVE THE CORRECT POLARITY
1Centre screw. 2Battery
compartment cover.
3Batteries size AA
x4.
Replacement
Unscrew the centre screw (1).
Remove the cover (2).
Make sure the polarity of the new batteries (3) is correct.
UPS-II Loop Calibrator Technical Notes
Current reading
Switch positions
Function
Loop power
Read
Ext.
Transmitter Calibration
Switch positions
Function
Loop power
Read
Int.
3 K0434 Issue No. 2
Transmitter Simulation
Switch positions
Function
Loop power
Out
Ext.
Valve Stroking
Switch positions
Function
Loop power
Out
Int.
Calibration Instructions
General
A calibration period of 12 months is recommended. The actual calibration interval depends on instrument usage and the total measurement uncertainty acceptable for the specified application.
UPS-II
is a precise measuring instrument and the test
The equipment and conditions of test must be suitable for the type of work. The calibration check and calibration adjustment should be carried out in a controlled environment by a calibration technician*.
The manufacturer offers a comprehensive and, if required, UKAS accredited calibration service.
*
A calibration technician must have the necessary technical knowledge, documentation, special test equipment and tools to carry out the calibration work on this equipment.
Calibration Equipment
The following tables give the accuracy requirements for the calibration equipment and the UPS-II. Calibration requires a stable temperature of 21° ±1°C (70° ±2°F).
UPS-II measure mode
Applied mAPermitted UPS-II
0 0.007 0
4 0.007 0.00014 12 0.007 0.00030 20 0.007 0.00046
UPS-II source mode
Applied mAPermitted UPS-II
00.01 0
4 0.01 0.00012 12 0.01 0.00011 20 0.01 0.00015
Table 1
error (mA)
Table 2
error (mA)
Calibrator error
(mA)
Calibrator error
(mA)
5 K0434 Issue No. 2
Calibration Check
Calibration Procedure mA Input
2 seconds
Apply zero - open circuit test leads
Connect test leads to mA source apply full-scale
Adjust mA source to any value within range - check reading
example
serial number
READ
and
+
Enter
*
Enter
*
Enter
*
VALVE
LIN
FLOW
MA
VALVE
LIN
FLOW
MA
and
+
-
*
adjust each digit
adjust each digit
adjust each digit
Enter
*
Input calibration complete new offset values stored continue with output calibration.
VALVE
LIN
FLOW
MA
1. Check the instrument for accuracy in measure and source modes, if there is an error greater than the permitted error, the instrument requires a calibration adjustment.
Calibration Adjustment
Note:
For complete accuracy, the measure mode and source mode adjustment must be carried together and in the following order.
Measure mode
1. Connect the UPS-II to the electrical calibrator. Switch on the electrical calibrator and allow it to thermally stabilise.
2. Switch on the UPS-II and allow the instrument to thermally stabilise.
3. Select EXT on the LOOP PWR slide switch. Use the display menu to select the calibration values:
Source mode
Calibration Procedure mA Output
2 seconds
Connect test leads to mA-meter compare values.
serial number
OUT
and
+
Enter
*
Enter
*
Enter
*
Enter
*
Enter
*
and
+
-
*
adjust each digit
adjust each digit
adjust each digit
Calibration complete new output values stored
Compare values.
Check reading.
Check reading.
VALVE
LIN
FLOW
MA
VALVE
LIN
FLOW
MA
VALVE
LIN
FLOW
MA
VALVE
LIN
FLOW
MA
1. Connect the UPS-II to the electrical calibrator. Switch on the electrical calibrator and allow it to thermally stabilise.
2. Switch on the UPS-II and allow the instrument to thermally stabilise.
3. Select INT on the LOOP PWR slide switch. Use the display menu to select the calibration values:
Completion
1. Switch off and disconnect the calibration equipment.
2. Update the calibration records for the instrument.
7 K0434 Issue No. 2
Range........................................................................................ 0 to 22.2 mA
Specifications
Accuracy........................................................................ 0,05% of range
Resolution...................................................................... 10 microamps
Temperature effect.............................. 0,003%/°C - 0,0015%/°F
Internal loop power...................................................24 VDC stabilized
External loop power............................................................48 VDC max.
Measurement input................................................... 12 ohms fused
Drive capability..................................................................... 900 ohms
Operating temperature............ -10°C to 50°C - 14°F to 122°F
Storage temperature .............. -20°C to 60°C - -4°F to 140°F
Battery......................... 4 x 1.5V alkaline or rechargeable size AA
Battery life (alkaline)............................. source 12mA; 18 hours
................................................................................. measure 900 hours
Loop mismatch........................................................... flashes "LOOP"
Low battery................................................................... flashes "LOBAT"
Over-range.................................................................... flashes "EEEE"
Reversed polarity....................................................... flashes "POL"
Adaptor line/6VDC
230V/50Hz............................ part numbers - UNO-13600-1 (UK)
...................................................................... - UNO-13600-2 (Europe)
110V/60Hz............................ part number - UNO-13600-3 (USA)
Housing.......................................................................... high impact ABS
Size................................................................................... 3" x 5" x 0,83"
................................................................................. 77 x 124 x 21 mm
Weight ............. ..................... ..................... ........... 212 grams - 7.5 oz
................................................................................. (including batteries)
Traceability and quality
UPS-II
Serial No: ...........................................................................................
Month: ...........................................................................................
Year: ...........................................................................................
This notification serves to certify that the unit described above has been inspected and tested in accordance with the specifications.
The unit calibration is traceable to international standards.
Loading...