GE Sievers 500 RL Operation And Maintenance Manual

Sievers 500 RL On-Line Total Organic Carbon Analyzer
Operation and Maintenance Manual
Firmware Version 2.10 or later
GE Analytical Instruments 6060 Spine Road Boulder, CO 80301 USA phone 800.255.6964 • 303.444.2009 fax 303.444.9543 DLM 74001-04 Rev. A
www.geinstruments.com
Printed in USA ©2010
Identification Records
Default Administrator User ID: ADMIN
Default Administrator Password: GEAI
Analyzer serial number: _______________________
(This is located on the side of the Analyzer.)
Date of receipt and installation of Analyzer: _______________________ (This is the warranty start date.)
GE Analytical Instruments ©2010 2 of 226 DLM 74001-04 Rev. A
Table of Contents
Identification Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
List of Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
List of Figures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Trademarks and Patents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Confidentiality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Standard Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Limitation of Remedies and Liability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Limitación de remedios y responsabilidad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Limites de correction et de fiabilité . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Beschränkte Ansprüche und Haftung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Limitazione di rimedi e responsabilità . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
赔偿与责任限制
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Chapter 2: System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
System Specifications* . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Sample Flow Path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Additional System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Microprocessor Controller and Electronics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Data Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Chapter 3: Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Step 1: Unpack and Inspect the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Additional Installation Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Step 2: Complete the Identification Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Step 3: Select a Location for the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Step 4: Install Power and Control Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
GE Analytical Instruments ©2010 3 of 226 DLM 74001-04 Rev. A
Connecting to a Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Installing the Analog Outputs and Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Wiring the 4-20 mA Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Wiring the Remote Start (Binary Input) Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Step 5: Installing the Printer, USB, and Serial Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Installing the Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Installing the USB Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Installing the Serial Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Wiring the Ethernet Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Step 6: Connect the Sample Inlet and Outlet Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Step 7: Fill the DI Water Cartridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Step 8: Configure Basic Analyzer Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Enabling DataGuard or Password Protection (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Enabling Password Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Enabling DataGuard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Establishing a New Administrator Account for DataGuard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Setting the Clock and Time Zone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Naming the Analyzer Location (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Setting the Analyzer Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Setting Up the Data History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Setting Up the Printer (Optional). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Exporting and Printing System Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Set Up Data I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Step 9: Rinsing the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Step 10: Install DataShare 500 Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Chapter 4: Basic Analyzer Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
The Main Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Taking TOC Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
The Menu Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Setting the Analyzer Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Grab Mode Measurement Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Managing Data History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Setting up Data History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Archiving Data History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Viewing Data History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Graphing Data History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Printing Data History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Exporting Data History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Setting Up Analyzer Input and Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
GE Analytical Instruments ©2010 4 of 226 DLM 74001-04 Rev. A
Setting Up Analog Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Adjusting Analog Output Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Setting Up Serial Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Interpreting Serial Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Issuing Serial Commands to the Analyzer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Configuring Printer Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Activating Binary Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Setting Alarm Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Using the Ethernet Connection and Modbus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Managing Maintenance Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Displaying Consumables Status. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Setting the Analyzer Clock and Time Zone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Reviewing Warnings and Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Displaying System Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Advanced Analyzer Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Exporting or Printing the Warning and Error History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Saving System Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Printing the Constants Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Adjusting Display Contrast Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Naming the Analyzer Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Changing the Display Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Specifying the Pharmacopeia. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Pharmacopeia Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Pass/Fail Reporting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Selecting the Program Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Configuring the Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Specifying Auto Restart Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Programming the TOC Autozero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Draining the iOS System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Activating Additional Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Basic Hardware Operation Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Opening the Analyzer Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
The Touch Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Recommended Operational State. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
The iOS System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
The Sample Inlet Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Vial Set Cartridges. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Steam Exposure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Chapter 5: Password Protection and DataGuard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .101
Enabling Password Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
GE Analytical Instruments ©2010 5 of 226 DLM 74001-04 Rev. A
To enable Password Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104
To change the password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104
To configure password settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104
Disabling Password Protection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Using DataGuard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .105
Enabling DataGuard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Adding User IDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Editing User Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Configuring Login Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Changing User Passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Dealing with Forgotten Passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Reactivating Inactivated User Accounts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Archiving User Accounts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Viewing, Exporting, and Printing Audit Trails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Logging Out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Chapter 6: Calibration and System Protocols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113
Required Calibration Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .114
Preparing for Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .115
Save System Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Perform Annual Maintenance Tasks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Handling Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Disable Daily Download and Binary Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Performing a Single-Point Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Performing a Multi-Point Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .120
Accuracy, Precision, and Calibration Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Performing Pharmacopeia Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .127
System Suitability Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
JP Protocol. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Reviewing Calibration and Verification History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
High Blank TOC Warnings/Errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Single-Point Calibration Worksheet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Multi-Point Calibration Worksheet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Accuracy, Precision, and Verification Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
System Suitability Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
JP Protocol Worksheet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Chapter 7: Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Replacing Consumables and Maintenance Items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Checking and Refilling the DI Water Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .140
Replacing the UV Lamp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Replacing the Sample Pump Tubing and Pump Heads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
GE Analytical Instruments ©2010 6 of 226 DLM 74001-04 Rev. A
Replacing the DI Water Cartridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Replacing the In-Line Filter Element . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
Setting the Installation or Date for New Consumables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Cleaning the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Chapter 8: Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .153
Step 1: Review Warnings and Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .153
Warnings and Error Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
Step 2: Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .159
External Inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
Internal Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
Step 3: Review Solutions for Basic Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .159
The Analyzer Will Not Power On. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
No Flow Through the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
Checking the Sample Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .160
Confirm Flow of Water to the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .160
Backflushing the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .160
Negative, High, or Erratic Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .163
Conductivity Autozero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .163
Sample Conductivity Cell Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .163
Sample Cell Zero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .164
TOC Autozero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .164
TOC Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .165
Problems Changing the Program Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .165
Problems with Vial Set Cartridges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .166
Problems with The DI Water Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166
Step 4: Contact Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .167
Step 5: Return the Analyzer to GE Analytical Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .167
Appendix A: Analyzer Dimension Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171
Appendix B: Connecting to a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179
Using HyperTerminal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .179
Determining Your Communications Port Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
GE Analytical Instruments ©2010 7 of 226 DLM 74001-04 Rev. A
GE Analytical Instruments ©2010 8 of 226 DLM 74001-04 Rev. A
List of Tables
Table 1 Alarm Outputs (TB2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Table 2 Alarm Outputs (TB1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Table 3 Serial and 4-20 mA Outputs (TB3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Table 4 Data Fields Output to the Serial (RS-232) Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Table 5 Modbus Register Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Table 6 Audit Trail Output Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .110
Table 7 Record of User IDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Table 8 Standards Required for Single-Point Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .114
Table 9 Standards Required for Multi-Point Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Table 10 Standards Required for Accuracy, Precision, and Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .115
Table 11 Standards Required for System Suitability Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .115
Table 12 Maximum Stored Results by Protocol Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .132
Table 13 Replacement Schedule for Consumables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .140
Table 14 Connections to the DI Water Cartridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .148
Table 15 Sievers 500 RL On-Line TOC Analyzer 3-Year Maintenance Worksheet . . . . . . . . . . . . . . . . . . . . . . . . .151
GE Analytical Instruments ©2010 9 of 226 DLM 74001-04 Rev. A
GE Analytical Instruments ©2010 10 of 226 DLM 74001-04 Rev. A
List of Figures
Figure 1 Analyzer Schematic (Shown with Standard iOS). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Figure 2 Wiring AC Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Figure 3 Wiring Diagram for the 4-20 mA Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Figure 4 Wiring Option for Binary Input Using Analyzer’s Internal Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Figure 5 Wiring Option for Binary Input Using External Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Figure 6 Input and Output Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Figure 7 Locating The DI Water Cartridge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Figure 8 Filling the DI Water Cartridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Figure 9 Removing the John Guest Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Figure 10 The Sievers 500 RL On-Line TOC Analyzer Main Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Figure 11 The Setup Tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Figure 12 The Data Tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Figure 13 Specifying a Start Date on the View Data Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Figure 14 Graphing the Data History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Figure 15 The I/O Tab (With Ethernet Port Installed) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Figure 16 The Modbus Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Figure 17 The Maintenance Tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Figure 18 The Advanced Menu with One Program Language Loaded . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Figure 19 The Standard iOS System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Figure 20 The Super iOS System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Figure 21 The Sample Inlet Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Figure 22 Menu Map — Menu Lockout Off (Requires Calibration Passwords) . . . . . . . . . . . . . . . . . . . . . . . . . . . .102
Figure 23 Menu Map — DataGuard (Optional Upgrade) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .103
Figure 24 Reviewing TC and IC Values from a Single-Point Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .119
Figure 25 Reviewing Conductivity Values from a Single-Point Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .119
Figure 26 The First Summary Screen in a Multi-Point Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .122
Figure 27 The Second Summary Screen in a Multi-Point Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Figure 28 The Third Summary Screen in a Multi-Point Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .123
Figure 29 The Accuracy, Precision, and Verification Summary Screen, Part 1 . . . . . . . . . . . . . . . . . . . . . . . . . . .126
Figure 30 The Accuracy, Precision, and Verification Summary Screen, Part 2 . . . . . . . . . . . . . . . . . . . . . . . . . . .126
Figure 31 System Suitability Summary Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GE Analytical Instruments ©2010 11 of 226 DLM 74001-04 Rev. A
. . . . . . . . . .129
Figure 32 Interior Overview of the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .140
Figure 33 Filling the DI Water Cartridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .141
Figure 34 Relative Positioning of Components in the UV Lamp Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .143
Figure 35 UV Lamp Ferrule Orientation Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .143
Figure 36 Replacing the Sample Pump Heads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .145
Figure 37 John Guest fittings on the DI Water Cartridge (Detail) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .147
Figure 38 Replacing the In-Line Filter Element . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .149
Figure 39 The Consumables Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .150
Figure 40 Disconnecting the Tubing from the Front Solenoid Valve. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .161
Figure 41 The Inlet and Outlet Connections Inside the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .162
Figure 42 The Sample Pump Tubing Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .162
Figure 43 The Sample Cell Calibration Results Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .164
Figure 44 Removing the John Guest Fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169
Figure 45 Analyzer Dimensions, with Standard iOS System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .171
Figure 46 Right-Side Analyzer Dimensions, with Standard iOS System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .172
Figure 47 Analyzer Dimensions, with Super iOS System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .173
Figure 48 Right-Side Analyzer Dimensions, with Super iOS System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .174
Figure 49 Analyzer Dimensions, with Sample Inlet Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .175
Figure 50 Right-Side Analyzer Dimensions, with Sample Inlet Block. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .176
Figure 51 Left-Side Analyzer Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .177
GE Analytical Instruments ©2010 12 of 226 DLM 74001-04 Rev. A
Revision History
Document Version Software Version/Enhancement Date
DLM 74000 Rev. A 1.0 September 2005 DLM 74000 Rev. B Editorial Corrections September 2005 DLM 74000 Rev. C 1.01 October 2005 DLM 74000 Rev. D 1.03 January 2006 DLM 74000 Rev. E 1.10 (Product Release) February 2006 DLM 74001-01 Rev. A 1.30 December 2007 DLM 74001-02 Rev. A Editorial Corrections May 2008 DLM 74001-02 Rev. B Editorial Corrections June 2008 DLM 74001-03 Rev. A 2.0 April 2009 DLM 74001-04 Rev. A 2.10 May 2010
GE Analytical Instruments ©2010 13 of 186 DLM 74001-04 Rev. A
Trademarks and Patents
Sievers®, DataShare™, DataGuard™, iOS System™, and Super iOS System™ are trademarks of General Electric Company and may be registered in one or more countries.
Norprene® is a registered trademark of Norton Performance Plastics Corporation; Swagelok® is a registered trademark of the Swagelok Company. Windows® and Excel® are registered trademarks of Microsoft Corporation. John Guest® is a registered trademark of John Guest International Ltd.
The Analyzer described in this manual is covered by one or more patents issued to and owned or pending by General Electric Company, including the following:
US 6,271,043
US 6,228,325
US 5,976,468
US 5,902,751
US 5,837,203
US 5,820,823
US 5,798,271
US 5,750,073
US 5,443,991
US 5,132,094
EP 0 897 530
FR 0 897 530
GB 0 897 530
DE 697 02 516 0-08
and other patents pending
GE Analytical Instruments ©2010 14 of 186 DLM 74001-04 Rev. A
Confidentiality
The information contained in this manual may be confidential and proprietary and is the property of GE Analytical Instruments. Information disclosed herein shall not be used to manufacture, construct, or otherwise reproduce the goods disclosed herein. The information disclosed herein shall not be disclosed to others or made public in any manner without the express written consent of GE Analytical Instruments.
Declaration of Conformity
A copy of the Declaration of Conformity for this product is available on our Web site at http://www.geinstruments.com (Products >500 RL > Specifications > Safety).
GE Analytical Instruments ©2010 15 of 186 DLM 74001-04 Rev. A
Standard Limited Warranty
GE Analytical Instruments warrants its products (Sievers®, GE Analytical™, and Leakwise™) for defects in materials and work­manship. GE Analytical Instruments will, at its option, repair or replace instrument components that prove to be defective with new or remanufactured components (i.e., equivalent to new). The warranty set forth is exclusive and no other warranty, whether written or oral, is expressed or implied.
Warranty Term
The GE Analytical Instruments warranty term is thirteen (13) months ex-works, or twelve (12) months from installation or start up by GE Analytical Instruments certified service personnel. In no event shall the standard limited warranty coverage extend beyond thirteen (13) months from original shipment date.
Warranty Service
Warranty Service is provided to customers through telephone support (1-800-255-6964), Monday - Friday, from 8:00 a.m. to 5:00 p.m. (Mountain Time), excluding all company and legal holidays. Telephone support is provided for troubleshooting and determination of parts to be shipped from GE Analytical Instruments to the customer in order to return the product to oper­ation. If telephone support is not effective, the product may be returned to GE Analytical Instruments for repair or replace­ment. In some instances, suitable instruments may be available for short duration loan or lease.
GE Analytical Instruments warrants that any labor services provided shall conform to the reasonable standards of technical competency and performance effective at the time of delivery. All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative, or designate. GE Analytical Instruments warrants these services for 30 days after the authorization and will correct any qualifying deficiency in labor provided that the labor service deficiency is exactly related to the originating event. No other remedy, other than the provision of labor ser­vices, may be applicable.
Repair components (parts and materials), but not consumables, provided in the course of a repair, or purchased individually, are warranted for 90 days ex-works for materials and workmanship. In no event will the incorporation of a warranted repair component into an instrument extend the whole instrument’s warranty beyond its original term.
Consumables (e.g., dilution standards, verification solutions, and UV lamps, etc.) are warranted to the extent of their stated shelf life, provided these items are maintained within the stated environmental limitations. Warranty claims for consumables and verification standards are limited to the replacement of the defective items, prorated from the time of claim to the expi­ration of shelf life.
Shipping
A Repair Authorization Number (RA) must be obtained from the Technical Support Group before any product can be returned to the factory. GE Analytical Instruments will pay freight charges, exclusive of any taxes and duties, for replacement or re­paired products shipped to the customer site. Customers shall pay freight charges, including all taxes and duties, for all prod­ucts returning to GE Analytical Instruments. Any product returned to the factory without an RA number will be returned to the customer.
GE Analytical Instruments ©2010 16 of 186 DLM 74001-04 Rev. A
Limitation of Remedies and Liability
The foregoing warranty shall not apply to defects resulting from impr oper or inadequate installation, maintenance, adjust­ment, calibration, or operation by customer. Installation, maintenance, adjustment, calibration, or operation must be per­formed in accordance with instructions stated in the Operation and Maintenance Manual. Usage of non-recommended maintenance materials may void a warranty claim.
The remedies provided herein are the customer's sole and exclusive remedies. In no event shall GE Analytical Instruments be liable for direct, indirect, special, incidental or consequential damages (including loss of profits) whether based on contract, tort, or any other legal theory. The Operation and Maintenance Manual is believed to be accurate at the time of publication and no responsibility is taken for any errors that may be present. In no event shall GE Analytical Instruments be liable for in­cidental or consequential damages in connection with or arising from the use of the manual and its accompanying related materials. Warranty is valid only for the original purchaser. This Limited Warranty is not transferable from the original pur­chaser to any other party without the express written consent from GE Analytical Instruments. GE Analytical Instruments
specifically disclaims the implied warranties of merchantability and fitness for a particular purpose.
Limitación de remedios y responsabilidad
La garantía anterior no se aplicará a los defectos que resulten de la realización incorrecta o inadecuada de la instalación, el mantenimiento, el ajuste, la calibración o el manejo por parte del cliente. La instalación, el mantenimiento, el ajuste, la cal­ibración o el manejo deberán llevarse a cabo de acuerdo con las instrucciones indicadas en el manual de funcionamiento y mantenimiento. El uso de materiales de mantenimiento que no sean los recomendados puede anular una reclamación de garantía.
Los remedios que aquí se indican serán los únicos los remedios para el cliente. En ningún caso GE Analytical Instruments será responsable de daños directos, indirectos, especiales, incidentales o consecuentes (incluida la pérdida de beneficios) ya sean contractuales, extracontractuales o basado en cualquier otra teoría legal. Se considera que el manual de funcionamiento y mantenimiento es exacto en el momento de su publicación y no se acepta ninguna responsabilidad por los errores que pueda contener. En ningún caso será GE Analytical Instruments responsable de los daños incidentales o consecuentes que resulten o estén relacionados con el uso del manual y los materiales que lo acompañan. La garantía es únicamente válida para el comprador original. El comprador original no puede transferir esta garantía limitada a ninguna otra parte sin el consenti­miento expreso por escrito de GE Analytical Instruments. GE Analytical Instruments renuncia específicamente a las garantías
implícitas de comercialización e idoneidad para un determinado propósito.
GE Analytical Instruments ©2010 17 of 186 DLM 74001-04 Rev. A
Limites de correction et de fiabilité
La garantie susdite ne s’applique pas aux défauts résultants d’une installation, d’une maintenance, d’un réglage, d’un cal­ibrage ou d’un fonctionnement inapproprié, opéré par l’utilisateur. L’installation, la maintenance, le réglage, le calibrage ou le fonctionnement doit être réalisé conformément aux instructions du manuel de l’utilisateur et de maintenance. La mise en œuvre de procédures de maintenance non recommandées peut annuler toute disposition de garantie.
Les procédures de correction indiquées dans le présent document sont les seuls remèdes du client. Le groupe GE Analytical Instruments ne saurait en aucun cas être tenu pour responsable de tout préjudice direct, indirect ou spécial de quelque na­ture que ce soit (y compris, les pertes de bénéfices), qu’il soit fondé sur un contrat, sur un acte dommageable ou sur une autre théorie légale. Le manuel de l’opérateur et de maintenance est aussi précis que possible au moment de la publication et la responsabilité du groupe ne saurait être engagée pour les éventuelles erreurs qu’il pourrait contenir. Le groupe GE Analytical Instruments ne saurait en aucun cas être tenu pour responsable des préjudices accidentels ou de quelque nature que ce soit, dus à l’utilisation du manuel ou de la documentation connexe. La garantie ne s’applique qu’à l’acquéreur d’origine. La garan­tie limitée ne peut être transférée par l’acquéreur d’origine à une autre partie sans l’autorisation expresse écrite du groupe GE Analytical Instruments. GE Analytical Instruments exclut tout particulièrement les garanties implicites de commerciali-
sation et d’adaptabilité dans un but spécifique.
Beschränkte Ansprüche und Haftung
Die vorangehende Garantie gilt nicht für Schäden, die aus unsachgemäßer oder unzureichender Installation, Wartung, An­passung, Kalibrierung oder Betrieb durch den Kunden resultieren. Installation, Wartung, Anpassung, Kalibrierung oder Betrieb müssen gemäß den Anweisungen in der Bedienungsanleitung durchgeführt werden. Durch die Verwendung von nicht empfohlenen Wartungsmaterialien kann der Garantieanspruch erlöschen.
Die hier erwähnten Ansprüche beziehen sich auf die einzigen und ausschließlichen Ansprüche des Kunden. GE Analytical Ins­truments ist unter keinen Umständen verantwortlich für direkte, indirekte, besondere, zufällig entstandene oder Folge­schäden (einschließlich Verlust von Einkünften), die auf Vertrag, unerlaubten Handlungen oder andere Rechtstheorien basieren. Die Bedienungsanleitung ist zur Zeit der Veröffentlichung nach bestem Wissen korrekt, und es wird keine Verant­wortung für mögliche vorhandene Fehler übernommen. GE Analytical Instruments ist unter keinen Umständen haftbar für zu­fällige oder Folgeschäden, die in Verbindung mit oder durch die Verwendung der Bedienungsanleitung und begleitender Materialien entstehen. Die Garantie gilt nur für den ursprünglichen Käufer. Die beschränkte Garantie lässt sich nicht ohne aus­drückliche schriftliche Genehmigung von GE Analytical Instruments vom ursprünglichen Käufer auf eine andere Person über­tragen. GE Analytical Instruments schließt besonders die konkludente Garantie der Handelsüblichkeit und Eignung für
einen bestimmten Zweck aus.
GE Analytical Instruments ©2010 18 of 186 DLM 74001-04 Rev. A
Limitazione di rimedi e responsabilità
La precedente garanzia non è valida per difetti risultanti da installazione, manutenzione, regolazione, taratura o utilizzo improprio o inadeguato da parte dell’utente. L’installazione, la manutenzione, la regolazione, la taratura o l’utilizzo deve essere conforme alle istruzioni indicate nel manuale d’uso e manutenzione. L’utilizzo di materiali di manutenzione diversi da quelli consigliati rende nullo un reclamo in garanzia.
Gli unici rimedi spettanti all’utente sono quelli qui inclusi. In nessun caso GE Analytical Instruments sarà responsabile per dan­ni diretti, indiretti, speciali, accidentali o consequenziali (inclusa la perdita di profitti) risultanti dall’applicazione del contratto, atto illecito o altra teoria legale. Il manuale d’uso e manutenzione è accurato al momento della pubblicazione e l’azienda non si assume alcuna responsabilità per la presenza di eventuali errori. In nessun caso GE Analytical Instruments sarà responsa­bile per danni accidentali o consequenziali correlati o derivanti dall’utilizzo del manuale e di altro materiale di supporto cor­relato. La garanzia è valida solo per l’acquirente originale. La presente garanzia limitata non è trasferibile dall’acquirente originale a terzi senza l’espresso consenso scritto da parte di GE Analytical Instruments. GE Analytical Instruments declina
espressamente le garanzie implicite di commerciabilità e idoneità a un particolare scopo.
限定責任および救済
前述の保証は、お客様による不適切または不十分な取り付け、保守、調整、校正、あるいは操 作によって生じた不具合には適用されないものとします。 取り付け、保守、調整、校正、また は操作は、『Operation and Maintenance Manual ( 操作取扱説明書 )』に記載されている指示に 従って行う必要があります。 推奨外の保守資材を使用すると、保証請求が無効になる場合があ ります。
ここで提供される救済は、お客様の唯一の排他的救済となります。 GE Analytical Instruments は、いかなる場合におい ても、直接的、間接的、特別的、付随的、または派生的損害(利益の逸失を含む)に対し、それが契約、不法行為、 またはその他の法的理論に基づくものであるかどうかにかかわらず、一切責任を負いません。 『Operation and Maintenance Manual ( 操作取扱説明書 )』は、出版された時点で正確であるものと考えられており、万が一発生した誤 りに対する責任は一切負いません。 GE Analytical Instruments は、いかなる場合においても、マニュアルまたはそれに 付属の関連資料の使用に関連して、またはその使用が原因で発生した付随的または派生的損害にも一切責任を負いま せん。 保証は最初の購入者に対してのみ有効です。 本限定保証を、GE Analytical Instruments の書面による同意なしに、 最初の購入者から第三者に譲渡することはできません。 GE Analytical Instruments
的に対する適合性の黙示の保証を一切拒否します。
は、商品性および特定の目
GE Analytical Instruments ©2010 19 of 186 DLM 74001-04 Rev. A
赔偿与责任限制
上述保证不适用于因客户不正确或不恰当的安装、维护、调整、校准或操作导致的故障。安装、 维护、调整、校准或操作必须遵循操作与维护手册中的说明进行。使用非推荐的维护材料可能 会导致保证失效。
这里提供的赔偿为客户的唯一和独占赔偿。在任何情况下,
GE Analytical Instruments
不对任何 直接的、间接的、特殊的、偶发的或连带发生的损失(包括利润损失)负责,无论这些损害是依 据何种合同责任理论、侵权行为责任理论或其它法律理论进行推断的。操作与维护手册在出版时 被认为是准确的, 下,
GE Analytical Instruments
GE Analytical Instruments
均不对因使用该手册(或与其使用有关)或相关材料导致的偶发或 连带发生的损失负责。保证仅对原购买者有效。未经 限保证不可由原购买者转让给任何其他方。
不对其中可能存在的任何错误负责。在任何情况
GE Analytical Instruments
GE Analytical Instruments
特此声明不提供任何关于特殊
明确书面同意,此有
用途的适销性和适用性的暗示担保。
GE Analytical Instruments ©2010 20 of 186 DLM 74001-04 Rev. A
Warnings
English
This symbol on the instrument indicates that the user should refer to the manual for operating instructions
The vial port contains sharp needles designed to pierce the septa of sample vials. Do not put fingers or inappropriate materials into the vial port.
Warning
Warning
Warning
Water in the iOS System may be hot. Before inserting a vial into the iOS System, slide the door open and wait 30 seconds to allow sample to completely drain. Inserting a vial before draining can result in potentially hot water spray projecting upward out of the iOS System.
Warning
This symbol indicates the protective earth terminal (ground) for the Analyzer.
Warning
This symbol indicates that to comply with European Union Directive 2002/96/EC for waste electrical and electronic equipment (WEEE), the Analyzer should be disposed of separately from standard waste.
GE Analytical Instruments ©2010 21 of 220 DLM 74001-04 Rev. A
Warning
This symbol on the instrument indicates that the user should refer to the manual for instructions on filling the DI water cartridge, as found in the Installation and Maintenance chapters.
Warning
This is a Safety Class I product. It must be must be attached to a grounded power source. To avoid damage to the Analyzer and the chance of injury, make sure the Analyzer is connected to a power line of the proper voltage.
Warning
Any operation requiring access to the inside of the Analyzer, including installation of maintenance items, could result in injury. To avoid potentially dangerous shock, turn off power and disconnect from the power supply before opening the Analyzer.
Warning
The UV lamp and the display screen contain mercury and may be considered hazardous material in your local area. Dispose of these items in accordance with federal, state, or local government regulations.
Warning
Should the UV lamp become broken or damaged it should be handled in accordance with your organization’s toxic waste handling procedure and disposed of in accordance with federal, state, or local government regulations.
Warning
To protect against accidental exposure to ultra-violet radiation, do not operate the UV lamp outside of its protective housing. The UV lamp housing may be hot; use extreme caution whenever touching the UV lamp housing.
Warning
Always stop analysis before turning off the Analyzer.
GE Analytical Instruments ©2010 22 of 220 DLM 74001-04 Rev. A
Warning
This is a Class A product. In a domestic environment, this product may cause electromagnetic interference in which case the user may be required to take adequate measures to correct the interference.
Warning
To avoid false TOC readings and possible damage to the Analyzer, always make sure the sample inlet is open and the DI water cartridge is filled before starting analysis. Always close the door of the Standard iOS System when not sampling from a vial.
Warning
Make sure the DI water cartridge is full, particularly when running samples with high TOC or high salt concentrations. Always “clean-up” the Analyzer by running low-TOC DI water after running high TOC or salt samples.
Warning
If this instrument is used in a manner not specified by GE Analytical Instruments, the protection provided by the instrument may be impaired.
Warning
The glass window in the UV reactor indicates proper lamp operation and should not be viewed too closely (closer than 6 inches) or for extended periods (greater than 1 minute). Protective eyewear is recommended.
Warning
Consult the MSDS that accompanies each standards set to identify proper disposal techniques.
Warning
To protect the Analyzer from damage, always drain the water from the Analyzer prior to shipping.
GE Analytical Instruments ©2010 23 of 220 DLM 74001-04 Rev. A
Español
Advertencia
Este símbolo del instrumento indica que el usuario debe consultar el manual para ver las instrucciones de manejo
Advertencia
El sistema iOS y el puerto del vial contienen dos agujas afiladas diseñadas para perforar los tabiques de los viales que contienen las muestras. No coloque los dedos ni ningún material que no sea adecuado en el sistema iOS ni en el puerto del vial.
Advertencia
El agua del sistema iOS puede estar caliente. Antes de insertar un vial en el sistema iOS, abra la puerta deslizándola y espere durante 30 segundos a que la muestra se vacíe por completo. Si inserta un vial antes de que se vacíe es posible que agua potencialmente caliente se proyecte fuera del sistema iOS.
Advertencia
Este símbolo indica el terminal de protección a tierra (masa) del analizador.
Advertencia
Toda operación que requiera el acceso al interior del analizador, incluida la instalación de los elementos de mantenimiento, puede causar daños personales. Para evitar descargas potencialmente peligrosas, apague el analizador y, si es posible, desconéctelo de la fuente de alimentación antes de abrirlo.
Advertencia
La lámpara UV y la pantalla de visualización contienen mercurio, por lo que es posible que se consideren materiales peligrosos en su zona local. Deseche estos elementos de acuerdo con la normativa federal, del estado o del gobierno local.
GE Analytical Instruments ©2010 24 of 220 DLM 74001-04 Rev. A
Advertencia
Si se rompiese la lámpara UV o resultase dañada, deberá tratarse de acuerdo con el procedimiento de tratamiento de residuos tóxicos de la organización y desecharse de acuerdo con la normativa federal, del estado o del gobierno local.
Advertencia
Como protección frente a la exposición accidental a la radiación ultravioleta, no maneje la lámpara UV fuera de su alojamiento protector.
Advertencia
Este producto es de clase de seguridad I. Debe conectarse a una fuente de alimentación con toma de masa.
Advertencia
Si este instrumento se utiliza de una manera no especificada por GE Analytical Instruments USA, la protección ofrecida por el instrumento puede verse reducida.
Advertencia
Detenga siempre el análisis antes de apagar o desconectar el analizador.
Advertencia
Asegúrese de que el depósito de agua DI está lleno, en especial al procesar muestras con altas concentraciones de sal o carbono orgánico total (TOC). “Limpie” siempre el analizador procesando agua DI con bajos niveles de TOC después de procesar muestras con niveles altos de sal o de TOC.
Advertencia
Este producto es de clase A. En entornos domésticos, puede producir interferencias electromagnéticas en cuyo caso puede que se le requiera al usuario que tome las medidas oportunas para corregir la interferencia.
Advertencia
Para evitar falsas lecturas de TOC y posibles daños al analizador, asegúrese siempre de que la entrada de muestras está abierta y de que el depósito de agua DI está lleno antes de iniciar el análisis.
GE Analytical Instruments ©2010 25 of 220 DLM 74001-04 Rev. A
Français
Avertissement
Ce symbole placé sur l’instrument indique que l’utilisateur doit se rapporter au manuel pour les instructions de fonctionnement.
Avertissement
Le système iOS et le porte-fiole contiennent deux aiguilles acérées conçues pour percer les septa des fioles. N’introduisez ni vos doigts, ni aucun objet dans le système iOS System ou le porte-fiole.
Avertissement
L’eau contenue dans le système iOS peut être brûlante. Avant d’insérer une fiole dans le système iOS, ouvrez la porte en la faisant coulisser et attendez 30 secondes que l’eau s’écoule
complètement. L’insertion d’une fiole avant le séchage complet peut entraîner une projection d’eau chaude en dehors du système iOS.
Avertissement
Ce symbole spécifie la borne de protection (mise à la terre) de l’analyseur.
Avertissement
Toute opération nécessitant d’accéder à l’intérieur de l’analyseur, y compris l’installation d’éléments de maintenance, peut entraîner des blessures. Afin d’éviter tout choc électrique potentiellement dangereux, mettez l’analyseur hors tension et, si possible, débranchez-le de la prise d’alimentation avant de l’ouvrir.
Avertissement
La lampe UV et l’écran contiennent du mercure et peuvent ainsi être considérés comme des éléments dangereux dans votre secteur. Jetez ces éléments conformément aux réglementations locales en vigueur.
GE Analytical Instruments ©2010 26 of 220 DLM 74001-04 Rev. A
Avertissement
Si la lampe UV venait à être cassée ou endommagée, elle devrait être remplacée conformément à la procédure en vigueur dans votre entreprise pour le remplacement de produits toxiques.
Avertissement
Afin d’éviter toute exposition accidentelle aux rayons ultra-violets, ne sortez pas la lampe UV de sa coque de protection.
Avertissement
Ce produit est de sécurité – classe I. Il doit être relié à une source d’alimentation mise à la terre.
Avertissement
Si cet instrument est utilisé de manière non conforme à ce qui est spécifié par le groupe GE Analytical Instruments aux USA, la protection fournie par l’instrument risque d’être réduite.
Avertissement
Arrêtez toujours l’analyse avant de mettre l’analyseur hors tension ou de le débrancher.
Avertissement
Assurez-vous que le réservoir d’eau DI est plein, tout particulièrement lors de l’utilisation d’échantillons comportant des concentrés fortement salés ou à TOC élevé. Nettoyez toujours l’analyseur à l’aide d’eau DI à faible TOC après avoir utilisé des concentrés salés ou à TOC élevé.
Avertissement
Ce produit fait partie de la classe A. Dans un environnement domestique, il peut entraîner des interférences électromagnétiques. Dans ce cas, l’utilisateur doit prendre les mesures appropriées pour corriger ces interférences.
Avertissement
Afin d’éviter les mesures TOC erronées et d’éventuels dommages de l’analyseur, assurez-vous toujours que l’entrée de l’échantillon est ouverte et que le réservoir d’eau DI est plein avant de commencer l’analyse.
GE Analytical Instruments ©2010 27 of 220 DLM 74001-04 Rev. A
Deutsch
Warnung
Dieses Symbol auf dem Instrument zeigt an, dass der Benutzer die Bedienungsanleitung beachten sollte.
Warnung
Das iOS-System und der Vial-Port enthalten zwei scharfe Nadeln, mit denen die Septa der Proben-Vials durchstochen werden. Bringen Sie Ihre Finger oder ungeeignete Materialien nicht mit dem iOS-System oder dem Vial-Port in Berührung.
Warnung
In den Analyzer kann das Wasser im iOS-System heiß sein. Bevor Sie das Vial in das iOS-System einsetzen, öffnen Sie die Tür und warten Sie 30 Sekunden, damit die Probe vollständig ablaufen kann. Das Einsetzen eines Vial, bevor das Wasser abgelaufen ist, könnte dazu führen, dass heißes Sprühwasser aus dem iOS-System herausspritzt.
Warnung
In Analyzers gibt dieses Symbol schützenden Erdungskontakt (Boden) für das Analysegerät an.
Warnung
Jeder Betrieb, der Zugriff auf das Innere des Analysegeräts erfordert, einschließlich Installation von Wartungsteilen, kann zu Verletzungen führen. Um gefährliche Verletzungen durch einen Stromschlag zu vermeiden, schalten Sie den Strom ab und trennen Sie, wenn möglich, das Analysegerät vor dem Öffnen vom Stromnetz.
Warnung
Die UV-Lampe und der Display enthalten Quecksilber und können regional als Gefahrgut angesehen werden. Entsorgen Sie diese Objekte gemäß den staatlichen oder regionalen Vorschriften.
GE Analytical Instruments ©2010 28 of 220 DLM 74001-04 Rev. A
Warnung
Sollte die UV-Lampe zerbrochen oder beschädigt sein, sollte sie gemäß den Vorschriften Ihres Unternehmens im Umgang mit Giftmüll gehandhabt und gemäß den staatlichen oder regionalen Vorschriften entsorgt werden.
Warnung
Betreiben Sie die UV-Lampe zum Schutz vor unbeabsichtigter ultravioletter Strahlung nicht außerhalb des schützenden Gehäuses.
Warnung
Dies ist ein Produkt der Sicherheitsstufe I. Es muss an eine geerdete Stromquelle angeschlossen werden.
Warnung
Wenn dieses Instrument in einer Art und Weise verwendet wird, die nicht von GE Analytical Instruments USA festgelegt ist, kann der durch dieses Instrument gebotene Schutz beeinträchtigt werden.
Warnung
Beenden Sie stets die Analyse, bevor Sie das Analysegerät ausschalten oder den Stecker herausziehen.
Warnung
Stellen Sie sicher, dass der DI-Wasserbehälter voll ist, besonders wenn Proben mit hohem Gesamtkohlenstoffgehalt (TOC) oder hohen Salzkonzentrationen analysiert werden. „Reinigen” Sie stets das Analysegerät, indem Sie DI-Wasser mit geringem Gesamtkohlenstoffgehalt durchlaufen lassen, nachdem Sie Proben mit hohem Gesamtkohlenstoffgehalt oder hoher Salzkonzentration verwendet haben.
Warnung
Dies ist ein Produkt der Klasse A. In einer häuslichen Umgebung kann das Produkt elektromagnetische Störungen verursachen. In diesem Fall muss der Benutzer möglicherweise entsprechende Maßnahmen ergreifen.
Warnung
Stellen Sie vor dem Beginn der Analyse stets sicher, dass der Probeneinlauf offen und der DI-Wasserbehälter gefüllt ist, um falsche Gesamtkohlenstoffmesswerte und mögliche Schäden am Analysegerät zu verhindern.
GE Analytical Instruments ©2010 29 of 220 DLM 74001-04 Rev. A
Italiano
Avvertenza
Questo simbolo posto sullo strumento indica che l’utente deve consultare il manuale per istruzioni sul funzionamento.
Avvertenza
Il sistema iOS e il sito per vial presentano due aghi creati per bucare le vial con campioni. Non mettere le dita o materiali non adatti nel sistema iOS o nel sito per vial.
Avvertenza
L’acqua nel sistema iOS può essere calda. Prima di inserire una vial nel sistema iOS, aprire lo sportello e attendere secondi, in modo da consentire al campione di scolare completamente. Inserire una vial prima che sia trascorso questo tempo può portare alla fuoriuscita di uno spruzzo di acqua calda diretto verso l’alto dal sistema.
Avvertenza
Qualsiasi funzionamento che richieda accesso all’interno dell’analizzatore, inclusa l’installazione di componenti per la manutenzione, può portare a lesioni. Per evitare uno shock potenzialmente pericoloso, spegnere l’analizzatore e, se possibile, scollegarlo dall’alimentazione prima di aprirlo.
Avvertenza
La lampada UV e il display contengono mercurio. Questo elemento può essere considerato materiale pericoloso nell’area di utilizzo dell’apparecchio. Smaltire questi componenti nel rispetto delle normative di governo locali o nazionali in vigore.
Avvertenza
Se la lampada UV dovesse danneggiarsi o rompersi, maneggiarla come indicato dalla procedura di gestione delle sostanze tossiche e smaltirla nel rispetto delle normative di governo locali o nazionali in vigore.
GE Analytical Instruments ©2010 30 of 220 DLM 74001-04 Rev. A
Loading...
+ 156 hidden pages