GE RT-FINO_RT3200 ADVANTAGE II_RT-X200 SERVICE MANUAL_SM_P9030GX_25 Clay Adams Compact II Centrifuge Model 420225 & 420227

Page 1
Technical Publications
P9030GX
Revision 25
RTfino/RT3200 Advantage–II/RT–x200 Service Manual
Copyright1991, 1992, 1993, 1994, 1995, 1996, 1998 by General Electric Company
Operating Documentation
Page 2
All packages should be closely examined at time of delivery. If damage is apparent, have notation of ”bad order” placed by the delivering driver on all copies of the freight or express bill. If damage is of a concealed nature, notify transportation agent as soon as possible to make an ”inspection report of damage” but in any event not later than 15 days after delivery. A trans­portation company usually will not pay a claim for concealed damage if an inspection is not requested within this 15 day period. Complete instructions regarding claim procedure are found in section ”S” of the Policy & Proce­dure Bulletins.
If shipment was handled by moving van service - uncrated - call Traffic ­Milwaukee immediately when any damage is found. Do not attempt to call any local agent. At this time be ready to describe type of damage, type of equipment, serial numbers and if possible the order number.
The above paragraph is in regard to equipment requiring installation only, and does not apply to supply items. The F.O.B. point for these items is as shown in the Price Book.
GE Medical Systems: Telex 3797371 P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A. (Asia, Pacific, Latin America, North America)
GE Medical Systems – Europe: Telex 698626 283, rue de la Miniere, B.P. 34, 78533 Buc Cedex France
Page 3
GE YOKOGAWA MEDICAL SYSTEMS, LTD. SERVICE ENGINEERING TECHINICAL SUPPORT US
7–127, Asahigaoka 4–Chome Hino–shi, Tokyo, 191 JAPAN
Page 4
Stick along this line
(DO NOT TEAR) FOLD HERE AND SEAL (DO NOT TEAR)
REPORT ON TECHNICAL PUBLICATION – RT
(OMISSIONS,ERRORS,SUGGESTIONS)
PRODUCT PUBLICATION # PAGE # PAGE REV DAT A REPORTED SERVICE RECORD SUBMITTED YES NO
PLEASE BE SPECIFIC IN YOUR CORRECTIONS AND SUGGESTIONS
FOLD FOLD
IF ADDITIONAL P AGES ARE INCLUDED, FOLD AND STAPLE T O LOWER PART OF THIS SHEET BEFORE MAILING.
SERVICE REP
PHONE
FAX
COUNTRY
CITY
P9030CB REV2
Page 5
US Advanced Document Group Of GEYMS
FAX Number: 81–42–585–5950
REV. 4
Page 6
GE MEDICAL SYSTEMS
WARNING
AVERTISSEMENT
THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE TRANSLATION SERVICES.
DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
F AILURE TO HEED THIS W ARNING MAY RESULT IN INJUR Y TO THE SER VICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL OR OTHER HAZARDS.
CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE TRADUIRE.
NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.
WARNUNG
AVISO
DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN
ENGLISCHER SPRACHE.
FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG ZU SORGEN.
VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST–HANDBUCH NICHT ZU RATE GEZOGEN UND VERST ANDEN WURDE.
WIRD DIESE W ARNUNG NICHT BEACHTET , SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICIT A UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE OTRA NATURALEZA.
P9030CD REV. 0
Page 7
GE MEDICAL SYSTEMS
ATENÇÃO
AVVERTENZA
ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA
DISPONÍVEL EM INGLÊS.
SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS, MECÂNICOS OU OUTROS.
IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE
SOLTANTO IN INGLESE.
SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE.
SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL CONTENUTO.
NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.
Page 8
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 25

LIST OF EFFECTIVE PAGES

REV DATE PRIMARY REASON FOR CHANGE
0 Feb. 14, 1991 Initial Release.. . . . . . . .
1 Feb. 20, 1991. . . . . . . .
2 Feb. 27, 1991. . . . . . . .
3 . Jun. 21, 1991. . . . . . .
4 Aug. 23, 1991. . . . . . . .
5 Dec. 12, 1991. . . . . . . .
6 Mar. 3, 1992. . . . . . . . .
7 Mar. 18, 1992. . . . . . . .
8 May 25, 1992. . . . . . . .
9 Aug. 26, 1992. . . . . . . .
10 Dec. 1, 1992. . . . . . . .
11 Dec. 29, 1992 CRT monitor is mounted onto the system as one unit shipping. . . . . . .
12 Feb. 18, 1993 Style C (version 3) Console for Asia. . . . . . .
13 Apr. 23, 1993 Style C (version 3) Console for America and Europe. . . . . . .
14 Aug. 20, 1993 Style D (version 4) Console. . . . . . .
15 Nov. 25, 1993 TRC board modification for Europe (console P9515CG) and VDO Board modification for America. . . . . . .
16 Feb. 10, 1994 Additional of the rating plate for the console made in India. . . . . . .
17 June. 15, 1994 Style E (version 5) Console. . . . . .
18 July. 26, 1994 Part number correction. . . . . . .
19 Aug. 11, 1994 Part number correction. . . . . . .
20 Oct. 1, 1994 Part number correction. . . . . . . .
21 Jan. 17, 1995 Addition of leakage current measurement procedure. . . . . . .
22 Jun. 2, 1995 Part number correction and Additional descriptions for Japanese console. . . . . . . .
23 Dec. 4, 1995 Additional descriptions for EMC, peripherals and others. . . . . . . .
24 July 15, 1996 Style F (Version 6) Console. . . . . . .
25 Aug. 19, 1998 Additional descriptions for new parts and error correction. . . . . . .
P9030GX
P AGE REV P AGE REV P AGE REV P AGE REV P AGE REV
Title page 25. . . . . . . . .
GE Logo page . . . . . .
P9030CB 2. . . . . . . . . . .
P9030CC 4. . . . . . . . . .
P9030CD 0. . . . . . . . . .
A to B 25. . . . . . . . . . . . .
i to vi 23. . . . . . . . . . . . .
vii 24. . . . . . . . . . . . . . . .
viii 23. . . . . . . . . . . . . . . .
INTRODUCTION
1–1 23. . . . . . . . . . . . . . .
1–2 0. . . . . . . . . . . . . . . .
1–3 to 1–12 23. . . . . . . .
PREINSTALLATION
2–1 23. . . . . . . . . . . . . . .
2–2 1. . . . . . . . . . . . . . . .
2–3 2. . . . . . . . . . . . . . . .
2–4 1. . . . . . . . . . . . . . . .
2–5 23. . . . . . . . . . . . . . .
2–6 2. . . . . . . . . . . . . . . .
2–7 1. . . . . . . . . . . . . . . .
2–8 to 2–12 23. . . . . . . .
2–13 to 2–14 24. . . . . .
2–15 to 2–16 23. . . . . .
INSTALLATION AND REPLACEMENT
3–1 21. . . . . . . . . . . . . . .
3–2 0. . . . . . . . . . . . . . . .
3–3 to 3–4 11. . . . . . . . .
3–5 to 3–52 21. . . . . . . .
FUNCTIONAL CHECKS
4–1 1. . . . . . . . . . . . . . . .
4–2 to 4–3 0. . . . . . . . . .
4–4 5. . . . . . . . . . . . . . . .
4–5 0. . . . . . . . . . . . . . . .
THEORY
5–1 to 5–3 0. . . . . . . . . .
5–4 2. . . . . . . . . . . . . . . .
5–5 to 5–8 0. . . . . . . . . .
5–9 1. . . . . . . . . . . . . . . .
5–10 0. . . . . . . . . . . . . . .
SERVICE ADJUSTMENT
6–1 2. . . . . . . . . . . . . . . .
6–2 0. . . . . . . . . . . . . . . .
6–3 4. . . . . . . . . . . . . . . .
6–4 to 6–5 2. . . . . . . . . .
SCHEMATICS
7–1 2. . . . . . . . . . . . . . . .
7–2 to 7–3 0. . . . . . . . . .
7–4 8. . . . . . . . . . . . . . . .
7–5 1. . . . . . . . . . . . . . . .
7–6 to 7–7 0. . . . . . . . . .
7–8 8. . . . . . . . . . . . . . . .
DIAGNOSTICS
8–1 2. . . . . . . . . . . . . . . .
8–2 1. . . . . . . . . . . . . . . .
8–3 2. . . . . . . . . . . . . . . .
8–4 5. . . . . . . . . . . . . . . .
RENEWAL PARTS
9–1 22. . . . . . . . . . . . . . .
9–2 to 9–13 24. . . . . . . .
9–14 to 9–16 22. . . . . .
9–17 24. . . . . . . . . . . . . .
9–18 to 9–25 14. . . . . .
9–26 to 9–27 24. . . . . .
9–28 to 9–29 25. . . . . .
9–30 to 9–32 24. . . . . .
9–33 to 9–35 25. . . . . .
9–36 to 9–37 17. . . . . .
9–38 to 9–39 24. . . . . .
9–40 to 9–41 25. . . . . .
9–42 to 9–44 24. . . . . .
9–45 to 9–46 25. . . . . .
9–47 24. . . . . . . . . . . . . .
9–48 to 9–49 17. . . . . .
9–50 to 9–52 24. . . . . .
9–53 to 9–54 25. . . . . .
9–55 24. . . . . . . . . . . . . .
9–56 to 9–57 25. . . . . .
9–58 to 9–59 24. . . . . .
9–60 to 9–63 17. . . . . .
9–64 to 9–65 25. . . . . .
9–66 to 9–68 17. . . . . .
9–69 24. . . . . . . . . . . . . .
9–70 to 9–75 17. . . . . .
9–76 to 9–77 25. . . . . .
9–78 to 9–80 17. . . . . .
9–81 to 9–82 25. . . . . .
9–83 to 9–85 17. . . . . .
9–86 to 9–87 22. . . . . .
9–88 25. . . . . . . . . . . . . .
9–89 to 9–92 22. . . . . .
9–93 25. . . . . . . . . . . . . .
9–94 to 9–99 22. . . . . .
A
Page 9
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 25
P9030GX
LIST OF EFFECTIVE PAGES (continued)
PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
9–100 to 9–101 25. . . .
9–102 to 9–103 22. . . .
9–104 to 9–105 25. . . .
9–106 to 9–1 11 22. . . . .
9–112 to 9–113 25. . . . .
9–114 to 9–115 22. . . . .
9–116 to 9–117 25. . . . .
9–118 to 9–136 22. . . .
9–137 25. . . . . . . . . . . . .
9–138 22. . . . . . . . . . . . .
9–139 25. . . . . . . . . . . . .
9–140 to 9–148 22. . . .
9–149 24. . . . . . . . . . . . .
9–150 to 9–186 22. . . .
PERIODIC MAINTENANCE
10–1 to 10–2 0. . . . . . .
10–3 to 10–6 24. . . . . .
OPTION MANUALS
B
Page 10
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23
P9030GX

TABLE OF CONTENTS

SECTION TITLE PAGE
T ABLE OF CONTENTS i. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 1 – INTRODUCTION
1–1 SERVICE MANUAL CONTENTS 1–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–2 SAFETY 1–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–2–1 Warnings 1–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–2–2 Specifications 1–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3 EMC (Electromagnetic Compatibility) 1–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–1 EMC Performance 1–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–2 Notice upon Installation of Product 1–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–3 General Notice 1–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–4 Countermeasures against EMC–related Issues 1–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–3–5 Notice on Service 1–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1–4 ADDRESS 1–12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 2 – PREINSTALLATION
2–1 FOREWORD 2–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2 GENERAL DESCRIPTION 2–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–1 Ultrasound Imaging 2–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2–2 Imaging Functions 2–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–3 ELECTRICAL SPECIFICATIONS 2–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–3–1 Power Supply 2–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–3–2 Leakage Current 2–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–3–3 Power Line Requirements 2–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–3–4 Transient Protection 2–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–3–5 Facility Power Receptacle 2–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–4 PHYSICAL SPECIFICATIONS 2–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–4–1 Operating Conditions 2–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–4–2 Shipping and Storage Requirements 2–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–5 RECOMMENDED ULTRASOUND ROOM LAYOUT 2–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–5–1 Floor Loading 2–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–6 PERIPHERALS/ACCESSORIES CONNECTOR PANEL 2–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
i
T ABLE OF CONTENTS
Page 11
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23
SECTION TITLE PAGE
P9030GX
CHAPTER 2 – PREINSTALLATION (Continued)
2–7 OPTIONAL PERIPHERALS 2–13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–7–1 Probes 2–13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–7–2 Recording Devices 2–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–7–3 Connecting Cables 2–16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 3 – INSTALLATION AND REPLACEMENT
3–1 INTRODUCTION 3–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–2 AVERAGE INSTALLATION TIME 3–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–3 W ARNINGS 3–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–4 UNPACKING RTfino/RT3200 Advantage–II/RT–x200 COMPONENTS 3–4. . . . . . . . . . . . . . . . . . . . . . .
3–4–1 Normal Unpacking Procedure 3–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–5 ASSEMBLING THE RTfino/RT3200 Advantage–II/RT–x200 SYSTEM 3–6. . . . . . . . . . . . . . . . . . . . . . .
3–6 MOVING THE RTfino/RT3200 Advantage–II/RT–x200 SYSTEM 3–7. . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–6–1 Moving the RTfino/RT3200 Advantage–II/RT–x200 To Another Room 3–7. . . . . . . . . . . . . . .
3–6–2 Transporting the RTfino/RT3200 Advantage–II/RT–x200 System 3–7. . . . . . . . . . . . . . . . . . .
3–7 ELECTRICAL SAFETY TESTS 3–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–7–1 Outlet Test Wiring Arrangement 3–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–7–2 Grounding Continuity 3–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–7–3 Chassis Leakage Current Test 3–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–7–4 Probe Leakage Current Test 3–13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–7–5 ECG Leakage Current Test 3–17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–7–6 When There’s Too Much Leakage Current 3–21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–8 REPLACEMENT 3–22. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–8–1 Gas Spring Replacement 3–22. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–8–2 Keyboard (Track Ball, TGC Assy) Replacement 3–28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–8–3 Monitor Replacement 3–30. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–9 FRENCH ROM 3–31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–10 GERMAN ROM 3–42. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ii
T ABLE OF CONTENTS
Page 12
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23
SECTION TITLE PAGE
P9030GX
CHAPTER 4 – FUNCTIONAL CHECKS
4–1 INTRODUCTION 4–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–1–1 Required Equipment 4–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–2 FUNCTIONAL CHECK PROCEDURES 4–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–2–1 Power Up 4–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4–2–2 REMAINING FUNCTIONAL CHECK 4–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 5 – THEORY
5–1 INTRODUCTION 5–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–2 PRINCIPLES OF ULTRASOUND IMAGING 5–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–2–1 Ultrasound Emission 5–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–2–2 Reflection 5–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–2–3 Ultrasound Displays 5–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–2–4 Beam Focusing 5–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–2–5 Echo Signal Processing 5–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–2–6 Time Gain Compensation (TGC)/Time Frequency Compensation (TFC) 5–6. . . . . . . . . . . . .
5–3 RTfino/R T3200 Advantage–II/RT–x200 SYSTEM 5–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–3–1 Echo Signal Flow 5–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–3–2 Circuit Board Nomenclature 5–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 6 – SERVICE ADJUSTMENTS
6–1 INTRODUCTION 6–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2 LOW VOLTAGE (+/– 5 VDC, +12 VDC) POWER SUPPLY ADJUSTMENT 6–3. . . . . . . . . . . . . . . . . . .
CHAPTER 7 – SCHEMATICS
7–1 OPERATOR CONSOLE ASSY (RTfino/RT3200 Advantage–II/RT–x200) 7–4. . . . . . . . . . . . . . . . . . . . .
iii
T ABLE OF CONTENTS
Page 13
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23
SECTION TITLE PAGE
P9030GX
CHAPTER 8 – DIAGNOSTICS
8–1 INTRODUCTION 8–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–2 SERVICE SOFTWARE 8–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3 THE SERVICE SOFTWARE (SS) PROGRAM 8–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–1 Entering Service Software Program 8–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–2 Leaving Service Software Program 8–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–3 Stopping Looping Service Software Routines 8–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8–3–4 Sample Command Summary 8–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 9 – RENEWAL PARTS
RTfino/RT3200 Advantage–II/RT–x200 MATERIAL LIST 9–2. . . . . . . . . . . . . . . . . . . . . . . . . .
OPERATOR CONSOLE ASSY 9–18. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
VIDEO DISPLAY MONITOR ASSY 9–122. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
FAN ASSY 9–126. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
FAN CABLE ASSY 9–126. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
PIPE ASSY 9–128. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
XFMR ASSY 9–130. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
KEYBOARD ASSY 9–132. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SHELF B SET FOR POLAROID CAMERA 9–134. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SHELF C SET FOR 380mm VTR 9–134. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SHELF D SET 9–136. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ADAPTOR SPRT SET 9–136. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
PRINTER TRAY SET 9–138. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
POWER SW ASSY 9–138. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
FRONT PANEL ASSY 9–140. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
UPPER PANEL ASSY 9–146. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CABLE ASSY 9–152. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
RUBBER CAP SET 9–186. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TGC ASSY 9–186. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 10 – PERIODIC MAINTENANCE
10–1 INTRODUCTION 10–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10–1–1 Periodic Maintenance 10–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10–1–2 Reference Chapters 10–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10–2 PERIODIC MAINTENANCE PROCEDURE 10–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 11 – OPTION MANUAL
iv
T ABLE OF CONTENTS
Page 14
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23
P9030GX

LIST OF ILLUSTRATIONS

SECTION TITLE PAGE
CHAPTER 1 – INTRODUCTION
1–2 1–1 REAR SIDE MARKINGS OF RTfino/RT3200 Advantage–II/RT–x200 1–5. . . . . . . . . . . . . .
1–2 RIGHT SIDE MARKINGS OF RTfino/RT3200 Advantage–II/R T–x200 1–6. . . . . . . . . . . . .
1–3 LEFT SIDE MARKINGS OF RT–x200 1–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 2 – PREINSTALLATION
2–2 2–1 RTfino/RT3200 Advantage–II/RT–x200 PICT ORIAL VIEW 2–4. . . . . . . . . . . . . . . . . . . . . . .
2–4 2–2 RTfino/RT3200 Advantage–II/RT–x200 DIMENSIONS 2–7. . . . . . . . . . . . . . . . . . . . . . . . . . .
2–6 2–3 UPPER CONNECTOR PANEL 2–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–4 LOWER CONNECTOR PANEL 2–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 3 – INSTALLATION
3–4 3–1 UNPACKING OF RTfino/RT3200 ADVANTAGE–II/RT–x200 3–5. . . . . . . . . . . . . . . . . . . . .
3–7 3–2 OUTLET TEST 3–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–3 GROUND CONTINUITY TEST 3–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–4 SETUP FOR CHASSIS SOURCE LEAKAGE CURRENT 3–10. . . . . . . . . . . . . . . . . . . . . . .
3–5 GROUND AND CHASSIS LEAKAGE CURRENT TEST 3–1 1. . . . . . . . . . . . . . . . . . . . . . . .
3–6 SETUP FOR PROBE LEAKAGE CURRENT 3–13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–7 TRANSDUCER SOURCE LEAKAGE CURRENT TEST 3–14. . . . . . . . . . . . . . . . . . . . . . . .
3–8 TEST CIRCUIT FOR MEASURING NON–ISOLATED PATIENT LEADS 3–17. . . . . . . . . .
3–9 ECG LEAKAGE CURRENT TEST 3–18. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–8 3–10 ASSEMBLED MONITOR BOTTOM VIEW 3–22. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–11 UNPACKED MONITOR 3–23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–12 CONSOLE SIDE VIEW 3–23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–13 BOTTOM VIEW – MONITOR ARM 3–23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–14 BOTTOM VIEW – MONITOR ARM 3–24. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–15 RAISED MONITOR ARM 3–24. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–16 SYSTEM REAR VIEW (REAR PLATES REMOVED) 3–25. . . . . . . . . . . . . . . . . . . . . . . . . . .
3–17 REAR VIEW 3–26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–18 BOTTOM VIEW – GAS SPRING 3–26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–19 GAS SPRING ASSEMBLY 3–27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–20 KEYBOARD – BOTTOM COVER 3–29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
v
T ABLE OF CONTENTS
Page 15
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23
P9030GX
LIST OF ILLUSTRATIONS (Continued)
SECTION TITLE PAGE
CHAPTER 3 – INSTALLATION (Continued)
3–8 3–21 KEYBOARD – PC BOARD 3–29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–9 3–22 PROCEDURE FLOW CHART 3–32. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–23 OPERATION KEYS 3–32. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–24 MANUFACTURE’S PRESET PARAMETER 3–34. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–25 ANATOMICAL SURVEY TABLE 3–35. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–26 USER PROGRAMMABLE COMMENTS LIBRARY 3–36. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–27 BACK PANEL REMOVAL 3–37. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–28 A3 MST ASSY COMPONENT SIDE 3–38. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–29 FRENCH SOFTW ARE 3–39. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–10 3–30 PROCEDURE FLOW CHART 3–43. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–31 OPERATION KEYS 3–43. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–32 MANUFACTURE’S PRESET PARAMETER 3–45. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–33 ANATOMICAL SURVEY TABLE 3–46. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–34 USER PROGRAMMABLE COMMENTS LIBRARY 3–47. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–35 BACK PANEL REMOVAL 3–48. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–36 A3 MST ASSY COMPONENT SIDE 3–49. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–37 GERMAN SOFTW ARE 3–50. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 4 – FUNCTIONAL CHECKS
4–2 4–1 POWER UP CRT DISPLAY 4–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
vi
T ABLE OF CONTENTS
Page 16
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 24
P9030GX
LIST OF ILLUSTRATIONS (Continued)
SECTION TITLE PAGE
CHAPTER 5 – THEORY
5–2 5–1 UL TRASOUND DISPLAY 5–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–3 5–2 (a) RTfino/RT3200 Advantage–II/RT–x200 SYSTEM BLOCK DIAGRAM 5–8. . . . . . . . . . . . . .
5–2 (b) RTfino/RT3200 Advantage–II/R T–x200 SYSTEM BLOCK DIAGRAM 5–9. . . . . . . . . . . . . .
CHAPTER 6 – SERVICE ADJUSTMENTS
6–2 6–1 REAR CONSOLE COVER 6–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–2 POWER SUPPLY UNIT 6–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–4 6–3 KEYBOARD – BOTTOM COVER 6–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6–4 KEYBOARD – PC BOARD 6–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 7 – SCHEMATICS
7–1 7–1 (a) RTfino/RT3200 Advantage–II/RT–x200 WIRING SCHEMATIC (1) 7–4. . . . . . . . . . . . . . . .
7–1 (b) RTfino/RT3200 Advantage–II/R T–x200 WIRING SCHEMATIC (1) 7–5. . . . . . . . . . . . . . . .
7–2 (a) RTfino/RT3200 Advantage–II/R T–x200 WIRING SCHEMATIC (2) 7–6. . . . . . . . . . . . . . . .
7–2 (b) RTfino/RT3200 Advantage–II/R T–x200 WIRING SCHEMATIC (2) 7–7. . . . . . . . . . . . . . . .
7–3 RTfino/RT3200 Advantage–II/R T–x200 WIRING SCHEMATIC (3) 7–8. . . . . . . . . . . . . . . .
CHAPTER 10 – PERIODIC MAINTENANCE
10–2 10–1 INSTALLING CRT FIL TER AND CLAMPS 10–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
vii
T ABLE OF CONTENTS
Page 17
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23
P9030GX

LIST OF TABLES

SECTION TITLE PAGE
CHAPTER 2 – PREINSTALLATION
2–4 2–1 SHIPPING REQUIREMENTS 2–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–2 STORE AND OPERATION REQUIREMENTS 2–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–5 2–3 UL TRASOUND ROOM REQUIREMENTS 2–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–7 2–4 LIST OF TRANSDUCERS (a) 2–13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–5 LIST OF TRANSDUCERS (b) 2–14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–6 LIST OF PROBE ADAPTERS (a) 2–14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–7 LIST OF PROBE ADAPTERS (b) 2–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–8 LIST OF RECORDING DEVICES 2–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2–9 POWER CONSUMPTION OF OPTIONAL RECORDING DEVICES 2–16. . . . . . . . . . . . . .
2–10 LIST OF CONNECTING CABLES 2–16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 3 – INSTALLATION
3–3 3–1 RTfino/RT3200 Advantage–II/RT–x200 INSTALLATION CHECK LIST 3–3. . . . . . . . . . . . .
3–7 3–2 CHASSIS LEAKAGE CURRENT TEST CONDITIONS 3–1 1. . . . . . . . . . . . . . . . . . . . . . . . . .
3–3 MAXIMUM ALLOWANCE LIMIT FOR CHASSIS SOURCE LEAKAGE CURRENT 3–12. .
3–4 DATA SHEET FOR CHASSIS SOURCE LEAKAGE CURRENT 3–12. . . . . . . . . . . . . . . . . .
3–5 MAXIMUM ALLOWANCE LIMIT FOR PROBE LEAKAGE CURRENT 3–16. . . . . . . . . . . .
3–6 DATA SHEET FOR TRANSDUCER SOURCE LEAKAGE CURRENT 3–16. . . . . . . . . . . . .
3–7 TESTING POWER CONDITIONS 3–19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3–8 MAXIMUM ALLOWANCE LIMIT FOR ECG LEAKAGE CURRENT 3–20. . . . . . . . . . . . . . .
3–9 DATA SHEET FOR ECG LEAKAGE CURRENT 3–20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 5 – THEORY
5–2 5–1 REFLECTION PERCENTAGES 5–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5–3 5–2 CIRCUIT BOARD NOMENCLATURE 5–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHAPTER 8 – DIAGNOSTICS
8–3 8–1 SS COMMAND CODE SUMMARY 8–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
viii
T ABLE OF CONTENTS
Page 18
REV 0
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
This page is left blank intentionally.
1–2
INTRODUCTION
Page 19
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23

1–1 SERVICE MANUAL CONTENTS

This manual provides service information on this Ultrasound Scanning System. It contains the following chapters:
1. Chapter 2, Preinstallation: Contains physical and electrical requirements that should be considered prior to installation;
2. Chapter 3, Installation and Replacement: Contains a complete installation procedure with checklist;
3. Chapter 4, Functional Checks: Contains functional checks that should be performed as part of the installation, or as required during servicing and periodic maintenance;
4. Chapter 5, Theory of Operation: Contains functional explanations of the RTfino/RT3200 Advantage–II/RT– x200 electronics;
5. Chapter 6, Service Adjustments: Contains adjustment procedures for all field adjustable RTfino/RT3200 Ad- vantage–II/RT–x200 potentiometers and other adjustable elements;
6. Chapter 7, Schematics: Contains copies of the RTfino/RT3200 Advantage–II/RT–x200 schematics;
7. Chapter 8, Diagnostics: Contains troubleshooting information for the analog and digital sections of the RTfino/ RT3200 Advantage–II/RT–x200;
P9030GX
8. Chapter 9, Parts List: Contains a complete list of replacement parts for the RTfino/RT3200 Advantage–II/R T– x200;
9. Chapter 10, Periodic Maintenance: Provides periodic maintenance procedures for the RTfino/RT3200 Advan­tage–II/RT–x200.
10. Chapter 11, Option Manuals: Provides option manuals for the RTfino/RT3200 Advantage–II/RT–x200.
1–3
INTRODUCTION
Page 20
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23

1–2 SAFETY

1–2–1 Warnings

P9030GX
W ARNING!
CAREFULLY READ ALL THE WARNINGS LISTED BELOW!
1. The operator manual should be fully read and understood before operating the RTfino/RT3200 Advantage–II/ RT–x200 and kept nearby for quick reference.
2. Although the ultrasound energy transmitted from the RTfino/RT3200 Advantage–II/R T–x200 transducer is within AIUM/NEMA standards, unnecessary exposure should be avoided. Only trained personnel should operate the RTfino/RT3200 Advantage–II/RT–x200.
3. To prevent electrical shock, the RTfino/RT3200 Advantage–II/RT–x200 should be connected to a properly grounded power receptacle. Do not use a three prong to two prong adapter. This defeats safety grounding.
4. Do not use with Defibrillator when RTfino/RT3200 Advantage–II/RT–x200 is operated .
5. Probes are fragile, handle with care.
6. Concerning Outside Markings, refer to Illustrations 1–1 and 1–2.
7. For the cleaning, disinfection, and sterilization, refer to Probe section in RTfino/RT3200 Advantage–II/RT–x200 User Manual and Caution Sheet supplied with each probe.
NOTICE
This medical equipment is approved, in terms of the prevention of radio wave interference, to be used in hospitals, clinics and other institutions which are environmentally qualified. The use of this equip­ment in an inappropriate environment may cause some electronic interference to radios and televi­sions around the equipment. Proper handling of this equipment is required in order to avoid such trouble according to the operator and service manuals. This equipment can be used in residential areas only under the supervision of physicians or qualified technicians.
1–4
INTRODUCTION
Page 21
REV 23
1–2–1 Warnings (Continued)
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
FOR PAL CONSOLE
FOR NTSC CONSOLE
FOR Latin America CONSOLE
FOR Latin America CONSOLE
FOR PAL CONSOLE
FOR NTSC CONSOLE
REAR SIDE MARKINGS OF RTfino/RT3200Advantage–II/RT–x200
ILLUSTRATION 1–1
1–5
INTRODUCTION
Page 22
REV 23
1–2–1 Warnings (Continued)
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
RIGHT SIDE MARKINGS OF RTfino/RT3200Advantage–II/RT–x200
ILLUSTRATION 1–2
1–6
INTRODUCTION
Page 23
REV 23
1–2–1 Warnings (Continued)
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
LEFT SIDE MARKINGS OF RT–x200
ILLUSTRATION 1–3
Note
The labels indicated in ILLUSTRA TION 1–3 are attached onto the RT–x200 from 1996, January ship­ment.
1–7
INTRODUCTION
Page 24
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23
P9030GX
1–2–1 Warnings (Continued)
The following table describes the purpose and location of safety labels and other important information provided on the equipment.
Label/Symbol Purpose/Meaning Location
Identification and Rating Plate
Type/Class Label
Device Listing/ Certification Labels
”DANGER – Risk of explosion used in...”
Manufacturer’s name and address
Date of manufacture
Model and serial numbers
Electrical ratings
Used to indicate the degree of safety or protection.
Equipment Type BF (man in the box symbol) IEC 878–02–03 indicates B Type equipment having a floating ap­plied part.
Laboratory logo or labels denoting conformance with industry safety standards such as UL or IEC.
The system is not designed for use with flammable anesthetic gases.
”ATTENTION – Consult accompa­nying documents ” is intended to alert the user to refer to the operator manual or other instructions when complete information cannot be pro­vided on the label.
”Mains OFF” Indicates the power off position of the mains power switch.
Rear of console near power inlet
Probe connectors and PCG connector
Rear of console
Rear of console
Various
Adjacent to On–Off Switch
”Mains ON” Indicates the power on position of the mains power switch.
”Protective Earth” Indicates the protective earth (grounding) terminal.
1–8
Adjacent to On–Off Switch
Various
INTRODUCTION
Page 25
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23

1–2–2 Specifications

Type of protection against electric shock : Class I EQUIPMENT (System) *1 Degree of protection against electric shock : Type BF EQUIPMENT *2 Power source voltage
:US : 115V ~ ±10% :Europe : 220–240V ~ ±10% :Far East : 100V ~ ±10%
Line Frequency : 50/60Hz (±2Hz)
*1. Class 1 EQUIPMENT
EQUIPMENT in which protection against electric shock does not rely on BASIC INSULATION only, but which includes an additional safety precaution in such a way that means are provided for the connection of ACCESSI­BLE CONDUCTIVE P ARTS to the PROTECTIVE(EARTH) CONDUCTOR in the fixed wiring of the installation in such a way that ACCESSIBLE CONDUCTIVE P ARTS cannot become LIVE in the event of a failure of the BASIC INSULA TION.
*2. Type BF EQUIPMENT
Type B EQUIPMENT with a F-TYPE isolated applied part providing a degree of protection against electric shock to such a degree that the allowable P A TIENT LEAKAGE CURRENT under SINGLE F AUL T CONDITIONS is not exceeded when 1.1 times the highest rated MAINS VOL TAGE is applied between the APPLIED PAR T and earth.
P9030GX
– allowable LEAKAGE CURRENT;
Patient leakage current Less than 100µA Less than 500µA
Normal mode Single failure mode
1–9
INTRODUCTION
Page 26
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23

1–3 EMC (Electromagnetic Compatibility)

1–3–1 EMC Performance

All types of electronic equipment may characteristically cause electromagnetic interference with other equipment, ei­ther through air or connecting cables. The term EMC (Electromagnetic Compatibility) indicates capability of the equip­ment, which curbs electromagnetic influence from other equipment and at the same time does not affect other equip­ment with similar electromagnetic radiation from itself.
This product is designed to fully comply with the EN60601–1–2 (IEC601–1–2), in Medical electrical equipment EMC regulations.
Proper installation following this service manual is required in order to achieve the full EMC performance of the prod­uct.
The product must be installed as stipulated in 1–3–2, Notice upon Installation of Product. In case of issues related to EMC, please follow procedures stated in 1–3–4, Countermeasures against EMC-related
Issues.
P9030GX

1–3–2 Notice upon Installation of Product

1) Use either power supply cords provided by GEMS or ones designated by GEMS. Products equipped with
power source plug should be plugged into the fixed power socket which has the protective grounding conduc­tor.
Connect a three-pole plug to a three-pole socket without using a three-pole-to-two-pole converter.
2) Locate the equipment as far as possible from other electronic equipment.
3) Be sure to use either any cables provided by GEMS or ones designated by GEYMS. Wire these cables follow-
ing these installation procedures. (Example) Wire power cables separately from signal cables.
4) Lay out the main equipment and other peripherals following the installation procedures described in Chapter3,
INST ALLATION.
1–10
INTRODUCTION
Page 27
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23

1–3–3 General Notice

1) Designation of Peripheral Equipment Connectable to This Product
The equipment which conforms to EN60601–1–2 (IEC601–1–2), can be hooked up to the product without compromising its EMC performance.
Avoid using non-standardized equipment. Failure to comply with this instruction may result in poor EMC per­formance of the product.
2) Notice against User Modification
Never modify this product. Unilateral user modification may cause degradation in EMC performance. Modification of the product includes:
a) Changes in cables (length, material, wiring etc.) b) Changes in system installation/layout c) Changes in system configuration/components
P9030GX
d) Changes in means of fixing system/parts (cover open/close, cover screwing)
3) Operate the system with all covers closed. If you open any cover for some reason, be sure to shut it before
starting/resuming operation. Operating the system with any cover open may affect EMC performance.

1–3–4 Countermeasures against EMC-related Issues

Generally it is very difficult to grapple with issues related to EMC. It may take much time and cost. General countermeasures
Electromagnetic interference with other equipment
1) Electromagnetic interference may be alleviated by positioning other equipment far from the system.
2) Electromagnetic interference may be mitigated by changing the relative location (installation angle) between
the system and other equipment.
3) Electromagnetic interference may be eased by changing wiring locations of power/signal cables of other
equipment.
4) Electromagnetic influence may be reduced by altering the path of power supply for other equipment.

1–3–5 Notice on Service

1) Ensure all screws are tight after servicing. Loose screws may cause degradation in EMC performance.
2) In case the high frequency gasket of this system is broken, replace it with a new one immediately.
1–1 1
INTRODUCTION
Page 28
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23

1–4 ADDRESS

This system is not repairable by the customer. If this equipment does not work as indicated in the Operator Manual, please contact your service support center. If the service engineer needs additional information to repair this equip­ment, please contact the following address (The necessary information will be provide to the Service Engineer as needed):
GE Medical Systems
Ultrasound Business Group 4855 W. Electric Ave., Milwaukee, WI 53219 USA TEL: (1) 800–437–1171 FAX: (1) 414–647–4090
CANADA TEL: (1) 800–668–0732
LATIN & SOUTH AMERICA TEL: (1) 305–735–2304
P9030GX
GE Medical Systems Europe
Ultrasound Business Division 283 rue de la miniere BP 34 78533 BUC CEDEX FRANCE TEL: (33) (1) 30.70.40.40 FAX: (33) (1) 30.70.42.50
GE YOKOGAWA MEDICAL SYSTEMS
Ultrasound Group 7 –127 Asahigaoka 4–Chome, Hino–Shi, Tokyo, 191 JAPAN TEL: (81) 426–56–0014 FAX: (81) 426–56–0002
1–12
INTRODUCTION
Page 29
REV 1
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
This page is left blank intentionally.
2–2
PREINSTALLATION
Page 30
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 2

2–1 FOREWORD

This chapter describes various general electrical, operational, and environmental considerations that should be con­sidered before installing the RTfino/RT3200 Advantage–II/RT–x200 Ultrasound Unit.

2–2 GENERAL DESCRIPTION

This unit is an ultrasound scanner designed to aid the physician in diagnosing diseases. See Illustration 2–1 for a pictorial view of the machine.

2–2–1 Ultrasound Imaging

This unit allows the physician to “see into” the patient’s body, using high frequency ultrasound imaging. When the physician places the transducer on a patient’s skin, acoustic pulses are transmitted into the body, reflected, and re­turned to the transducer. These returns, called echoes, are processed, amplified, and projected on a video monitor to form an image of the area scanned in the patient’s body.

2–2–2 Imaging Functions

P9030GX
All necessary imaging functions are performed by this console and its transducers. The console front panel contains operator controls for optimizing the image and a keyboard for entering data onto the image. Selected key LEDs illumi­nate when their respective key is pressed. Unacceptable keyboard entries trigger a beeper alarm. In addition, a graphics controller central processing unit automatically superimposes certain machine parameters over the edges of the image.
2–3
PREINSTALLATION
Page 31
REV 1
2–2–2 Imaging Functions (continued)
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
RTfino/RT3200 Advantage–II/RT–x200 PICTORIAL VIEW
ILLUSTRATION 2–1
2–4
PREINSTALLATION
Page 32
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23

2–3 ELECTRICAL SPECIFICATIONS

Electrical conduit, junction boxes, outlets, circuit breakers, and switches should be in place before installing this Console. It is customer’s responsibility to ensure proper environmental and electrical Power Requirements are satisfied prior to installation.

2–3–1 Power Supply

This unit operates on 1 15 or 220 VAC, single phase, 50/60 Hz power . Maximum current drawn is 5.0 amps at 120 VAC, and 2.0 amps at 220/240 V AC. The 115 VAC main circuit breaker is rated at 15A; the 220/240 VAC version at 10A. Voltage setup is performed in the factory, because different rear panels containing different power cables and circuit breakers are used for the115 V AC and 220 VAC versions.

2–3–2 Leakage Current

T otal ground leakage from any exposed conductive surface to ground is less than 100 microamps (For UL 544). T otal ground leakage is less than 500 microamps (For 220V). Patient leakage current is less than 10 microamps. (These parameters are checked during installation). This unit is classified as Class I, type BF equipment in accordance with IEC–601–1.
P9030GX

2–3–3 Power Line Requirements

The following power line parameters should be monitored for one week before installation. We recommend that a Dranetz Model 605–3 Power Line Monitor with option 101 be used:
PARAMETER : LIMITS
Voltage Range : U.S. : 103–127 VAC
: Europe : 206–254 VAC : Far East : 90–110 VAC
Power : U.S. : MAX. 500 VA
: Europe : MAX. 500 VA
: Far East : MAX. 500 VA Line Frequency : All applications : 50/60Hz (+/– 2Hz) Power Transients : Less than 25 % of nominal peak voltage for less than 1 millisecond for any type of
transient, including line frequency, synchronous, asynchronous, or aperiodic transients.
Decaying Oscillation : Less than 15 % of peak voltage for less than 1 millisecond.

2–3–4 Transient Protection

If testing indicates excessive transients occur, we recommend that a transient suppressor be installed, such as the Isolatrol Model LAI–115 made by Control Concept Corporation.
2–5
PREINSTALLATION
Page 33
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 2

2–3–5 Facility Power Receptacle

A separate power outlet with a 15 amp circuit breaker for 1 15 VAC units, or a 10 amp circuit breaker for 220 V AC units, is recommended. The specific power receptacle used depends on your country’s power line standards.
For U.S. applications, use an Underwriters Laboratory (UL) approved Hospital Grade receptacle. For Canadian ap­plications, the receptacle should be approved by the Canadian Standards Association (CSA). For other countries, the receptacle should have International Electrotechnical Commission (IEC) approval.

2–4 PHYSICAL SPECIFICA TIONS

The RTfino/RT3200 Advantage–II/RT–x200 (excluding accessories) weighs 86 kg (190 lbs). See Illustration 2–2 for dimensions.

2–4–1 Operating Conditions

This unit is designed to operate within a temperature range of 50 degrees F to 104 degrees F (15.6 degrees C to 40 degrees C), and a relative humidity range of 30 % to 75 %. (Non–condensing).
P9030GX
Patient Comfort
Despite permissible operating temperature and humidity tolerances, we recommend that ambient room temperature be maintained between 68 and 79 degrees F (20 to 26 degrees C), Humidity should be maintained between 50 % and 70 % for patient comfort during ultrasound scanning.
Electromagnetic Interference (EMI)
In general, the maximum EMI noise Field Strength Intensity should be no more than 100 dB uV/m within the 10 kHz to 100 MHz frequency range, as measured three meters from the this unit. If there is reason for concern, we recommend that you measure EMI using a spectrum analyzer.
1. EMI sources should be located and their effects minimized. These noise sources include microwave ovens, cer­tain types of electronic surgical equipment, and digital equipment such as desk top computers and terminals, gel warmers, and AM radio broadcast sources.
2. Keep the unit at least 15 feet (approximately five meters) away from the above noise sources; preferably in anoth­er room since wood or plaster walls, doors, floors, and ceilings provide some EMI shielding. Special EMI shield­ing may be required to eliminate noise problems caused by high frequency , high powered radio or video broad­cast signals.
2–6
PREINSTALLATION
Page 34
REV 1
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
WEIGHT 86 kg (89 kg)
190 lbs (196 lbs)
320mm
12.6in
240mm
9.44in 405mm
15.94in
1200mm
(1300mm)
400mm
15.74in
31.81in 808mm
720mm
28.34in
47.24in
(51.18in)
795mm
(895mm)
31.30in
(35.24in)
162mm
6.38in
NOTE:
MEASUREMENTS IS PARENTHESIS REFERS TO TALL MODEL.
RTfino/RT3200 Advantage–II/RT–x200 DIMENSIONS
ILLUSTRATION 2–2
2–7
PREINSTALLATION
Page 35
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23
P9030GX

2–4–2 Shipping and Storage Requirements

This unit excluding PROBES is shipped in a single container. Dimension of two types of models, namely Short and Tall,,are given below.
TABLE 2–1
SHIPPING REQUIREMENTS
MODEL TYPE WEIGHT kg (lbs)
Short Monitor Container = 12.5 kg (27.5)
Console Container = 98 kg (215.6)
Tall
Monitor Container = 12.5 kg (27.5)
Console Container = 100 kg (220)
DIMENSION (cm)
49 x 45 x 43
53 x 94 x 120
49 x 45 x 43
53 x 94 x 130
T able 2–2 provides a summary of temperature, atmospheric pressure, and humidity tolerances for shipping and opera­tion:
TABLE 2–2
STORAGE AND OPERATION REQUIREMENTS
PARAMETER STORAGE OPERATION
TEMPERATURE (DEG.C)
(DEG.F)
0to50
32 to 122
10 to 40 50 to 104
ATOMOSPHERIC PRESSURE
(hPa)
HUMIDITY (%) (Non–condensing)
700 to 1060 700 to 1060
10 to 85 30 to 75
2–8
PREINSTALLATION
Page 36
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23

2–5 RECOMMENDED ULTRASOUND ROOM LAYOUT

Table 2–3 provides the requirements for an ultrasound room:
TABLE 2–3
ULTRASOUND ROOM REQUIREMENTS
MIN. FLOOR AREA 130 SQ FT (12 SQ M)
P9030GX
POWER SOURCE
CURRENT RATING
RADIATION SHIELDING
TEMPERATURE
HUMIDITY
HEAT DISSIPATION
FLOOR LOADING
115VAC, 60HZ or 220VAC, 50HZ, SINGLE PHASE
15A (115V) ; 10A (220V) CIRCUIT BREAKER
NONE REQUIRED for ULTRASOUND ENERGY
68–79 DEG F (20–26 DEG. C) for PATIENT COMFORT
50% to 70% for PATIENT COMFORT
650 BTU/Hr for RTfino/RT3200 Advantage–II/RT–x200 ; 300 BTU/Hr per PERSON
about 2.4 kg/cm
2
/wheel

2–5–1 Floor Loading

The floor must be capable of supporting the RTfino/RT3200 Advantage–II/RT–x200 Console, which weighs more than 86kg(190lbs), depending on options added. The floor load is approximately 2.4kg/cm2/wheel.
2–9
PREINSTALLATION
Page 37
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23
P9030GX

2–6 PERIPHERALS/ACCESSORIES CONNECTOR PANEL

RTfino/RT3200 Advantage–II/RT–x200 peripherals and accessories can be properly connected using the upper and lower connector panels.
The video input/output connectors, audio input/output connectors, camera expose connectors, foot switch connector, and power outlet connectors for peripherals are located on the upper and lower panels.
This section indicates the pin assignment for each connector ( 2–4) on pages 2–11 through 2–12.
Note
The following illustrations show the example of RT–x200 upper and lower connector panels. There are some differences depending on the model. For the connector specifications of other models, con­tact your service support center.
1–4
in ILLUSTRA TION 2–3 and ILLUSTRATION
UPPER CONNECTOR PANEL
ILLUSTRATION 2–3
LOWER CONNECTOR PANEL
ILLUSTRATION 2–4
2–10
PREINSTALLATION
Page 38
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23
2–6 PERIPHERALS/ACCESSORIES CONNECTOR PANEL (Continued)
Note
Each outer (case) ground line of peripheral/accessory connectors are protectively grounded. Signal ground lines are Not Isolated.
P9030GX
Note
Output level of control signals indicated in the above tables are TTL level.
2–1 1
PREINSTALLATION
Page 39
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 23
2–6 PERIPHERALS/ACCESSORIES CONNECTOR PANEL (Continued)
P9030GX
Note
Output level of RS232C signals:
High +3V to +15V Low –15V to 0V
Note
Output level of control signals indicated in the above tables are TTL level.
2–12
PREINSTALLATION
Page 40
REV 24

2–7 OPTIONAL PERIPHERALS

2–7–1 Probes

1. Probes applied to CE Mark:
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
TABLE 2–4
LIST OF TRANSDUCERS (a)
PROBE MATERIAL OF AREA OF CATALOG
NAME HEADSHELL USING NO.
A PBT ABDOMINAL LINEAR H4061A B PBT ABDOMINAL LINEAR H4061B D PES ABDOMINAL LINEAR H4061D With Probe Adapter E PBT ABDOMINAL LINEAR H4061E H PBT SUPERFICIAL LINEAR H4061H LP PES SUPERFICIAL LINEAR H45202LP CA PES ABDOMINAL CONVEX H4061CA CB PES ABDOMINAL CONVEX H4061CB CTV ABS INTRACABITY CONVEX H4061TV S PES ABDOMINAL CONVEX H40612S TRR7 ABS INTRACABITY CONVEX H4061TR ATF (SH) ABS INTRACABITY CONVEX H44201TT ATF (BH) ABS INTRACABITY CONVEX H44201BH Wide MZ PES INTRACABITY CONVEX H45202FM Wide CK PES ABDOMINAL CONVEX H45202FK
TYPE PART No. NOTE
B9714AK B9714AA B9714AL B9714AJ B9714AN B9714AD B9714AP B9714AE B9714AS B9714AH P9601AP P9601AN B9719AN B9719AE B9719AL B9719AC 46–285221G1 B9719UM B9719AR B9719AQ 46–285515G1 2118380 2114530 2118378 2101 125 2118505 2120311 2120310 2120305 2120304
For V5 and V6 Console
Note
Probes whose part number are indicated by “
”, are applied to CE Mark.
Note
Some probes on the above list have two different part numbers. The upper raw shows the part num­bers of probes for regions other than Japan. The lower row shows the part numbers of probes for Japan.
2–13
PREINSTALLATION
Page 41
REV 24
2–7–1 Probes (Continued)
2. Other Probes:
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
TABLE 2–5
LIST OF TRANSDUCERS (b)
PROBE MATERIAL OF AREA OF CATALOG
NAME HEADSHELL USING NO.
C PBT ABDOMINAL LINEAR H4061C F PBT INTRAOPERATION LINEAR H4061F With Probe Adapter G PBT INTRACABITY LINEAR H4061G With Probe Adapter I PBT INTRAOPERATION LINEAR H4061I With Probe Adapter J PBT INTRAOPERATION LINEAR H4061J With Probe Adapter SB(L) PVDF INTRAOPERATION LINEAR H44201SB CC PES ABDOMINAL CONVEX H4061CC ATV ABS INTRACABITY CONVEX H44201AT SV ABS INTRACABITY CONVEX H4061SV ATRR7 NORYL INTRACABITY CONVEX H4061RA MZ PES INTRACABITY CONVEX H45022MZ CK PES ABDOMINAL CONVEX H45202CK CD PVDF NEONETAL CONVEX H44201CD
TYPE PART No. NOTE
B9714AM B9714AC B9714AQ B9714AF B9714AR B9714AG B9719AJ B9719AA B9719AK B9719AB 45310649
B9719AM B9719AD 46–285644G1
46–285499G1 46–285613G1 B9719A V
B9719AU P9603A V For V1, V2, V3, and P9603AC V4 console 45310651
Note
Some probes on the above list have two different part numbers. The upper raw shows the part num­bers of probes for regions other than Japan. The lower row shows the part numbers of probes for Japan.
3. Probe Adapters applied to CE Mark:
LIST OF PROBE ADAPTERS (a)
PROBE CATALOG
NAME NO.
D Probe only B9714GS B9714GT
DESCRIPTION NO. FOR NO. FOR
Probe + Adapter B9714AD B9714AN H4061D Adapter only B9714GC B9714GD Probe + Adapter + Attachment P9515PA P9515QA H45052E Adapter + Attachment P9515PJ P9515QJ H45062F
Probes whose part number are indicated by “
TABLE 2–6
PART PART
JAPAN OTHERS
Note
”, are applied to CE Mark.
2–14
PREINSTALLATION
Page 42
REV 23
2–7–1 Probes (Continued)
4. Other Probe Adapters:
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
TABLE 2–7
LIST OF PROBE ADAPTERS (b)
PROBE CATALOG
NAME NO.
F Probe only B9714HS B9714HT
G Probe only B9714JS B9714JT
I Probe only B9719AA B9719AJ
J Probe only B9719AB B9719AK
I, J Adapter only B9714LC B9714LD

2–7–2 Recording Devices

DESCRIPTION NO. FOR NO. FOR
PART PART
JAPAN OTHERS
Probe + Adapter B9714AF B9714AQ H4061F Adapter only B9714HC B9714HD Probe + Adapter + Attachment P9515PB P9515QB H45072G Adapter + Attachment P9515PK P9515QK H45082H
Probe + Adapter B9714AG B9714AR H4061G Adapter only B9714JC B9714JD Probe + Adapter + Attachment P9515PC P9515QC H45042D Adapter + Attachment P9515PL P9515QL H45032C
Probe + Adapter B9719JV Probe + Adapter + Attachment P9515PD P9515QD H45092J
Probe + Adapter B9719JW Probe + Adapter + Attachment P9515PE P9515QE H45022B
Adapter + Attachment P9515PM P9515QM H45012A
1. Devices: The following recording devices are suggested to connect with RT–x200.
TABLE 2–8
LIST OF RECORDING DEVICES
DEVICE MANUFACTURER MODEL VIDEO SIGNAL
Video Cassette Recorder SONY SVO–9500MDP PAL Video Graphic Printer SONY UP890CE PAL
2–15
PREINSTALLATION
Page 43
REV 23
2–7–2 Recording Devices (Continued)
2. Power Consumption: The table 2–9 shows the power consumption of each optional peripheral for RTfino/RT3200 Advantage–II/
RT–x200.
POWER CONSUMPTION OF OPTIONAL RECORDING DEVICES
DEVICE MODEL POWER CONSUMPTION
Video Cassette Recorder SVO–9500MDP 64W B/W Video Printer UP890CE 110W

2–7–3 Connecting Cables

RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
TABLE 2–9
CAUTION
Equipment damage possibility . Be sure to use the following recommended connecting cables to connect recording devices with RTfino/RT3200 Advantage–II/RT–x200 console.
TABLE 2–10
LIST OF CONNECTING CABLES
NAME PART NO. FIGURE NOTE
POWER CABLE P9515UD For B/W Video Printer
BNC CABLE P9509BD Video signal cable for B/W Video
Printer
MINI PLUG CABLE P9509BC Control signal cable for B/W
Video Printer
2–16
PREINSTALLATION
Page 44
REV 0
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
This page is left blank intentionally.
3–2
INST ALLA TION AND REPLACEMENT
Page 45
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 11
P9030GX

3–1 INTRODUCTION

This chapter contains many of the procedures required to install this Console. The Functional Checks chapter con­tains the remaining installation procedures. All these procedures are included in T able 3–1, the Installation Checklist. Use this checklist to check off each installation step. This ensures a thorough installation and functional checkout of the unit.

3–2 AVERAGE INSTALLA TION TIME

The RTfino/RT3200 Advantage–II/RT–x200 has been designed to be installed and checked out by an Ultrasound ser­vice technician in approximately four hours. RTfino/RT3200 Advantage–II/RT–x200 Consoles with optional equip­ment may take slightly longer.

3–3 WARNINGS

1. Since the RTfino/RT3200 Advantage–II/R T–x200 weighs approximately 190 lbs (86 kg) without options, prefer­ably two people should unpack it.
2. The operator manual should be fully read and understood before operating the RTfino/RT3200 Advantage–II/ RT–x200 and kept nearby for quick reference.
3. Although the ultrasound energy transmitted from the RTfino/RT3200 Advantage–II/RT–x200 transducer is within AIUM/NEMA standards, unnecessary exposure should be avoided.
4. To prevent electrical shock, the RTfino/RT3200 Advantage–II/RT–x200 should be connected to a properly grounded power receptacle. Do not use a three prong to two prong adapter. This defeats safety grounding.
5. This System has no operator serviceable components. To prevent shock, do not remove any covers or panels. Should problems or malfunctions occur, unplug the power cord. Only service personnel should carry out servic­ing and troubleshooting.
TABLE 3–1
RTfino/RT3200 Advantage–II/RT–x200 INSTALLATION CHECKLIST
PROCEDURES PARAGRAPH INITIALS
1. OBSERVE WARNINGS 3–3 ______
2. UNPACK RTfino/RT3200 Advantage–II/RT–x200 COMPONENTS 3–5 ______
3. ASSEMBLE THE RTfino/RT3200 Advantage–II/RT–x200 3–6 ______
4. PERFORM ELECTRICAL SAFETY TESTS 3–8 ______
5. PERFORM FUNCTIONAL CHECKS
(SEE CHAPTER 4 FOR PROCEDURES)
3–3
INST ALLA TION AND REPLACEMENT
Page 46
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 11

3–4 UNPACKING RTfino/RT3200 Advantage–II/RT–x200 COMPONENTS

Note
Check the shipping container for special instructions. Verify that the container is intact. In some cases a secondary container may be used. If so, ask the carrier for unpacking instructions.

3–4–1 Normal Unpacking Procedure

1. Leaving the bottom of the container intact, cut the plastic bands. (See Illustration 3–0)
2. Pull the cap up and off of the container.
3. Pull out the three plastic joints by hand or with pliers; turn the plastic joints counterclockwise and pull. 4 Open the outer sleeve and cut the corners of the bottom tray.
5. Remove protective packing material; save it for future shipments, if required.
6. Preferably with a helper, carefully roll the RTfino/RT3200 Advantage–II/RT–x200 off its shipping platform.
P9030GX
CAUTION
The RTfino/RT3200 Advantage–II/RT–x200 System weighs approximately 86kg (190lbs). Be prepared for a surge as the RTfino/RT3200 Advantage–II/RT–x200 is rolled off of its shipping platform.
3–4
INST ALLA TION AND REPLACEMENT
Page 47
REV 21
3–4–1 Normal Unpacking Procedure (Continued)
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
PLASTIC BANDS
CAP
INNER SLEEVE
REAR
BOTTOM TRAY
FRONT
OUTER SLEEVE
OPEN
PLASTIC JOINT
PALLET
UPPER CUSHION
LOWER CUSHION
UNPACKING OF RTfino/RT3200 ADVANTAGE-II/RT-x200
ILLUSTRATION 3–1
3–5
INST ALLA TION AND REPLACEMENT
Page 48
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21

3–5 ASSEMBLING THE RTfino/RT3200 Advantage–II/RT–x200 SYSTEM

3–5–1 Installation

1. Locate remove, and unpack the transducers located in the separate container.
2. Secure any of transducers to either of the two receptacle as follows. a. Ensure transducer twist lock lever points towards the 2 o’clock position. b. Install transducer connector on the receptacle guide pin until it touches the receptacle mating surface. c. Twist the transducer connector twist lock lever to the 6 o’clock position to lock it in place. T wist the lever to the 2
o’clock position to disconnect the transducer.
3. It is not necessary to turn off to RTfino/RT3200 Advantage–II/R T–x200 system power to connect or disconnect a transducer.
4. Unpack the foot switch. Connect the foot switch to the Connector on the lower RTfino/RT3200 Advantage–II/RT– x200 front panel.
P9030GX
5. Connect the RTfino/RT3200 Advantage–II/RT–x200 power cable to a hospital grade power receptacle of the proper voltage. Never use a three–to–two prong adapter; this defeats the safety ground.
3–6
INST ALLA TION AND REPLACEMENT
Page 49
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21

3–6 MOVING THE RTfino/RT3200 Advantage–II/RT–x200 SYSTEM

3–6–1 Moving the RTfino/RT3200 Advantage–II/RT–x200 to Another Room

In general, a single adult can move the system along an even surface with no steep grades. At least two people should move the machine when large humps, grooves, or grades will be encountered. (It is better to pull the console from the rear rather than to push the console). Before moving, store all loose parts in the unit. Wrap transducers in soft cloth or foam to prevent damage.

3–6–2 Transporting the RTfino/RT3200 Advantage–II/RT–x200 System

All shipping should be done by General Electric representatives, or their assignees. Follow the instructions in para­graph 3–6–1, with the following additional precautions:
1. The original packaging should be used when reshipping the unit. General Electric representatives or assignees shall repackage the console before shipment.
2. Carefully load the RTfino/RT3200 Advantage–II/RT–x200 system aboard the van. Use four straps to secure the unit and prevent lateral movement. Set the unit over the van’s center of gravity . Do not lay the unit down for
shipment!
P9030GX
3. Prevent shock by driving at 45–55 mph on paved roads. Avoid unpaved roads and drive carefully at road and railroad crossings.
4. Repeat the installation procedure, completing the installation checklist, at the new site.
3–7
INST ALLA TION AND REPLACEMENT
Page 50
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
P9030GX

3–7 ELECTRICAL SAFETY TESTS

The following tests are performed at the factory and should be performed at the site. These tests are: grounding conti­nuity , chassis leakage current, probe leakage current, and ECG leakage current. All measurements are made with an Electrical Safety Analyzer Model Dale 600/600E built by Dale Technology Corporation or equivalent device.
Recommended Tool
P ART NUMBER NAME DESCRIPTION
46–285652G1 Electric Safety Analyzer DALE 600 For 120V Units 46–328406G2 Electric Safety Analyzer DALE 600E For 220V Units 2113015 Leakage Current Ultrasound Kit For 120V and 220V Units

3–7–1 Outlet Test Wiring Arrangement

T est all outlets in the area for proper wiring arrangement by plugging in the Dale 600/600E and noting the combination of lights that are illuminated (Refer to ILLUSTRATION 3–2).
OUTLET TEST
ILLUSTRATION 3–2
Note
No outlet tester can detect the condition where the Neutral (grounded supply) conductor and the Grounding (protective earth) conductor are reversed. If later tests indicate high leakage currents, this should be suspected as a possible cause and the outlet wiring should be visually inspected.
3–8
INST ALLA TION AND REPLACEMENT
Page 51
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
P9030GX

3–7–2 Grounding Continuity

CAUTION
Electric Shock Hazard. The patient must not be contacted to the equipment during this test.
Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case (ILLUSTRA­TION 3–3). The ground wire resistance should be less than 0.2 ohms.
CONSOLE
ACCESSIBLE MET AL PARTS
1. Monitor Housing
GROUND PIN
2. Rear Panel Connector
3. Any Caster/Wheel Support
OHMMETER
GROUND CONTINUITY TEST
ILLUSTRATION 3–3
Meter Procedure
Follow these steps to test the ground wire resistance.
1. Turn the RTfino/RT3200 Advantege-II/RT-x200 unit OFF.
2. Plug the unit into the meter, and the meter into the tested AC wall outlet (Refer to ILLUSTRATION 3–5).
3. Plug the black chassis cable into the meter’s ”CHASSIS” connector and attach the black chassis cable clamp to an exposed metal part of the RTfino/RT3200 Advantege-II/RT-x200 unit (Refer to ILLUSTRATION 3–5).
4. Set the meter’s ”FUNCTION” switch to the RESISTANCE position.
5. Set the meter’s ”POLARITY” switch to the OFF (center) position.
6. Measure the ground wire resistance and keep a record of the results with other hard copies of PM data kept on site.
3–9
INST ALLA TION AND REPLACEMENT
Page 52
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
P9030GX

3–7–3 Chassis Leakage Current Test

Definition
Leakage current is the electrical current that could flow through the patient or sonographer should a ground wire break. The unit, the probes, the ECG module, and all external peripherals must be tested.
Generic Procedure
The test verifies the isolation of the power line from the chassis. The testing meter is connected from accessible metal parts of the case to ground. Measurements should be made with the unit ON and OFF, with the power line polarity Normal and Reversed. Record the highest reading of current.
POWER OUTLET
H (BLACK)
N (WHITE)
G (GREEN)
POLARITY REVERSING SWITCH
MOMENT ARY SWITCH
SETUP FOR CHASSIS SOURCE LEAKAGE CURRENT
ILLUSTRATION 3–4
CONSOLE
ACCESSIBLE METAL PARTS
LEAKAGE
TEST
METER
3–10
INST ALLA TION AND REPLACEMENT
Page 53
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
P9030GX
3–7–3 Chassis Leakage Current Test (Continued)
Meter Procedure
Follow these steps to test the unit for leakage current.
1. Turn the RTfino/RT3200 Advantege-II/RT-x200 unit OFF.
2. Plug the unit into the meter, and the meter into the tested AC wall outlet (Refer to ILLUSTRATION 3–5).
GROUND AND CHASSIS LEAKAGE CURRENT TEST
ILLUSTRATION 3–5
3. Plug the black chassis cable into the meter’s ”CHASSIS” connector and attach the black chassis cable clamp to an exposed metal part of the RTfino/RT3200 Advantege-II/RT-x200 unit (Refer to ILLUSTRATION 3–5).
4. Set the tester’s ”FUNCTION” switch to CHASSIS position.
5. Follow the test conditions described in the following table for every test point on the unit.
TABLE 3–2
CHASSIS LEAKAGE CURRENT TEST CONDITIONS
TEST CONDITION
1 Mounting screw for transducer receptacle 2 Wheel support 3 Mounting screw for CRT housing 4 Mounting screw for peripheral plugged into unit 5 Mounting screw for other peripheral powered by unit
6. Keep a record of the results with other hard copies of PM data kept on site.
3–1 1
INST ALLA TION AND REPLACEMENT
Page 54
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
P9030GX
3–7–3 Chassis Leakage Current Test (Continued)
Data Sheet for Chassis Source Leakage Current
Follow the foregoing test procedure. The test passes when all readings measure less than the value shown in the table below (TABLE 3–3).
TABLE 3–3
MAXIMUM ALLOWANCE LIMIT FOR CHASSIS SOURCE LEAKAGE CURRENT
COUNTRY NORMAL OPEN REVERSE OPEN
CONDITION GROUND POLARITY NEUTRAL
USA N/A 0.1mA 0.1mA N/A
OTHERS 0.1mA 0.5mA 0.5mA 0.5mA
CAUTION
Electric Shock Hazard. When the meter’s ground switch is OPEN, don’t touch the unit!
TABLE 3–4
DATA SHEET FOR CHASSIS SOURCE LEAKAGE CURRENT
Unit Tester Tester Test1 Test2 Test3 Test4 Test5
Power POLARITY NEUTRAL Probe wheel CRT
Switch Switch connector
Enter name of tested peripheral here:
ON NORM OPEN ON NORM CLOSED ON REV OPEN
ON REV CLOSED OFF NORM OPEN OFF NORM CLOSED OFF REV OPEN OFF REV CLOSED
3–12
INST ALLA TION AND REPLACEMENT
Page 55
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
P9030GX

3–7–4 Probe Leakage Current Test

Definition
This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
Generic Procedure
Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the probe must be active to find the worst case condition.
POWER OUTLET
H (BLACK)
POLARITY REVERSING SWITCH
PROBE
N (WHITE)
CONSOLE
MOMENT AR Y
G (GREEN)
SWITCH
LEAKAGE TEST METER
SETUP FOR PROBE LEAKAGE CURRENT
ILLUSTRATION 3–6
Note
Each probe will have some amount of leakage current, dependent on its design. Small variations in probe leakage currents are normal from probe to probe. Other variations will result from differences in line voltage and test lead placement.
3–13
INST ALLA TION AND REPLACEMENT
Page 56
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
P9030GX
3–7–4 Probe Leakage Current Test (Continued)
Meter Procedure Using Probe Adapter
The Dale 600/600E provides a method for testing probes independently from the system. The meter utilizes a probe adapter to apply a test potential commonly to all connector pins.
The probe’s imaging area is immersed in a saline solution along with a grounding probe from the meter to complete the current path. Saline solution is a mixture of water and salt. The salt adds a free ion to the water, making it conductive. Normal saline solution is 0.9% salt or 1/2 gram salt per 1 liter of water. If saline is not available, a mixture of 1 quart wat er with one or more grams of table salt, mixed thoroughly, will substitute.
CAUTION
T o avoid probe damage and possible electric shock, do not immerse probes into any liquid beyond the level indicated in the probe users manual. Do not touch the probe, conductive liq­uid or any part of the unit under test while the ISO TEST switch is depressed.
Follow these steps to test each transducer for leakage current.
1. Turn the RTfino/RT3200 Advantege-II/RT-x200 unit OFF.
2. Plug the unit into the test meter, and the meter into the tested AC wall outlet.
3. Connect the probe for test with the meter’s appropriate adapter (Refer to ILLUSTRATION 3–7).
4. Plug the saline probe into the meter’s ”CHASSIS” connector (Refer to ILLUSTRATION 3–7).
TRANSDUCER SOURCE LEAKAGE CURRENT TEST
ILLUSTRATION 3–7
3–14
INST ALLA TION AND REPLACEMENT
Page 57
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
3–7–4 Probe Leakage Current Test (Continued)
5. Plug the probe adapter into the meter’s connector marked ”EXTERNAL” (Refer to ILLUSTRATION 3–7).
6. Set the meter’s ”FUNCTION” switch to EXTERNAL position.
7. Add the saline probe and the imaging area of the probe into the saline bath.
8. Have unit power ON for the first part; turn it OFF for the second half.
9. Depress the ISO TEST rocker switch and record the highest current reading.
10. Follow the test conditions described in the following table (TABLE 3–6) for every transducer.
11. Keep a record of the results with other hand copies of PM data.
Meter Procedure
Follow these steps to test each transducer for leakage current.
P9030GX
1. Turn the RTfino/RT3200 Advantege-II/RT-x200 unit OFF.
2. Plug the unit into the test meter, and the meter into the tested AC wall outlet.
3. Plug the external probe into the meter’s ”EXTERNAL” connector.
4. Set the meter’s ”FUNCTION” switch to EXTERNAL position.
5. Connect the probe for test with the connector of the console.
Note
Do not test all the probes on the same XDIF connector. Use all three slots eventually.
6. Apply a liberal amount of gel to the transducer’s lens.
7. Wrap the transducer in aluminum foil, making sure there is good contact with its acoustic aperture.
8. Either suspend the transducer by its cable or place it on an insulated surface.
Note
Lift only one probe at a time from its resting hook.
9. Have unit power ON for the first part; turn it OFF for the second half.
10. Hold the external probe against the aluminum foil on the transducer when the unit is ready to scan.
11. Depress the ISO TEST rocker switch and record the highest current reading.
12. Follow the test conditions described in the following table (TABLE 3–6) for every transducer.
13. Keep a record of the results with other hand copies of PM data.
3–15
INST ALLA TION AND REPLACEMENT
Page 58
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
P9030GX
3–7–4 Probe Leakage Current Test (Continued)
Data Sheet for Transducer Source Leakage Current
Follow the foregoing test procedure for every transducer. The test passes when all readings measure less than the value shown in the table below (TABLE 3–5).
TABLE 3–5
MAXIMUM ALLOWANCE LIMIT FOR PROBE LEAKAGE CURRENT
1. Type BF Applied Part Leakage Current Limits [Non–conductive (floating) Surface and Cavity Transducers]
COUNTRY NORMAL OPEN REVERSE OPEN
CONDITION GROUND POLARITY NEUTRAL
USA 0.05mA 0.05mA 0.05mA 0.05mA
OTHERS 0.1mA 0.5mA 0.5mA 0.5mA
2. Type CF Applied Part Leakage Current Limits [Surgical Transducer and ECG Connections]
COUNTRY NORMAL OPEN REVERSE OPEN
CONDITION GROUND POLARITY NEUTRAL
USA 0.01mA 0.01mA 0.01mA 0.01mA
OTHERS 0.01mA 0.05mA 0.05mA 0.05mA
TABLE 3–6
DATA SHEET FOR TRANSDUCER SOURCE LEAKAGE CURRENT
Transducer tested:
RTfino/RT3200 Advan- Tester Power Tester Ground Measurement
tage-II/RT-x200 Unit Power Polarity Switch
ON NORM OPEN ON NORM CLOSED ON REV OPEN ON REV CLOSED OFF NORM OPEN OFF NORM CLOSED OFF REV OPEN OFF REV CLOSED
3–16
INST ALLA TION AND REPLACEMENT
Page 59
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
P9030GX

3–7–5 ECG Leakage Current Test

Definition
This test measures the current which would flow to ground from any of the isolated ECG leads. The meter simulates a patient who is contacted to the monitoring equipment and is grounded by touching some other grounded surface.
Generic Procedure
Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the operating controls such as the lead switch should be operated to find the worst case condition.
The connection is at service entrance or on the supply side of a separately derived system
POWER OUTLET
Building ground
Grounding contact switch (use in both ”open” and ”closed” positions)
Appliance power switch (use both ”off” and ”on” positions)
Patient lead selector switch (if any) (activated as required)
Polarity reversing switch
H (Black)
N (White)
(use both positions)
H N
Appliance
Internal
circuitry
G
G (Green)
Insulating surface
H = Hot
Current meter
TEST CIRCUIT FOR MEASURING NON–ISOLATED PATIENT LEADS (PROBES & ECG)
ILLUSTRATION 3–8
N = Neutral (grounded) G = Grounding conductor
Patient connected leads
Between each patient lead and ground
Between combined patient leads and ground
3–17
INST ALLA TION AND REPLACEMENT
Page 60
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
P9030GX
3–7–5 ECG Leakage Current Test (Continued)
Meter Procedure
The Dale 600/600E provides five snap type ECG buttons for testing patient leads. Snap on all patient leads to the meter and assure that the ground clip is connected to the system’s ground terminal.
CAUTION
Line voltage is applied to the ECG leads during this test. T o avoid possible electric shock haz­ard, the system being tested must not be touched by patients, users or anyone while the ISO TEST switch is depressed.
When the meter’s ground switch is OPEN, don’t touch the unit!
Follow these steps to test the ECG module for leakage current.
1. Turn the RTfino/RT3200 Advantege-II/RT-x200 unit OFF.
2. Plug the unit into the meter, and the meter into the tested AC wall outlet (Refer to ILLUSTRATION 3–9).
3. Plug the black chassis cable into the meter’s ”CHASSIS” connector and attach the black chassis cable clamp to an exposed metal part of the RTfino/RT3200 Advantege-II/RT-x200 unit (Refer to ILLUSTRATION 3–9).
4. Connect the patient leads to the corresponding snaps located at the upper front of the Dale 600/600E. Lead no­menclature for this test is not important (Refer to ILLUSTRATION 3–9).
ECG LEAKAGE CURRENT TEST
ILLUSTRATION 3–9
3–18
INST ALLA TION AND REPLACEMENT
Page 61
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
P9030GX
3–7–5 ECG Leakage Current Test (Continued)
5. Set the meter’s ”FUNCTION” switch to LEAD TO GROUND position to measure the patient lead to ground leak­age current.
6. Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the power condition combinations shown in TABLE 3–7.
TABLE 3–7
TESTING POWER CONDITIONS
ECG Meter’s Meter’s
Power Polarity Neutral
Switch Switch
ON NORM CLOSED ON NORM OPEN ON REVERSE CLOSED
ON REVERSE OPEN OFF NORM CLOSED OFF NORM OPEN OFF REVERSE CLOSED OFF REVERSE OPEN
7. Record the patient lead to ground leakage current measured on the data sheet (Refer to TABLE 3–9).
8. Set the meter’s ”FUNCTION” switch to LEAD TO LEAD position to measure the lead to lead leakage current.
9. Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the power condition combinations shown in TABLE 3–7.
10. Record the lead to lead leakage current measured on the data sheet (Refer to TABLE 3–9).
11. Set the meter’s ”FUNCTION” switch to LEAD ISO position to measure the patient lead isolation current.
12. Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the power condition combinations shown in TABLE 3–7.
13. Depress the rocker switch to ISO TEST and read the isolation current. T o apply the voltage to the lead safely , the voltage is only applied when the rocker switch is depressed to ISO TEST.
14. Record the patient lead isolation current measured on the data sheet (Refer to TABLE 3–9).
3–19
INST ALLA TION AND REPLACEMENT
Page 62
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
P9030GX
3–7–5 ECG Leakage Current Test (Continued)
Data Sheet for ECG Leakage Current
Follow the foregoing test procedure. The test passes when all readings measure less than the value shown in the table below (TABLE 3–8).
TABLE 3–8
MAXIMUM ALLOWANCE LIMIT FOR ECG LEAKAGE CURRENT
MAXIMUM
AC POWER
SOURCE
ALLOWANCE LIMIT
GROUND GROUND
OPEN CLOSED
Patient Lead to Ground Leakage Current Test
and
Patient Lead to Lead Leakage Current Test
115V 10 µA 10 µA
220/240V 500 µA 10 µA
AC POWER
SOURCE
1 15V 20 µA
Patient Lead Isolation Current Test
TABLE 3–9
DATA SHEET FOR ECG LEAKAGE CURRENT
220/240V 5mA
Patient Lead Isolation Current Test
Patient Lead to Lead Leakage Current Test
ECG Tester Tester Tester Lead Selector
Power Polarity Neutral
Patient Lead to Ground Leakage Current Test
Switch Switch RL RA LA LL C
ECG Tester Tester Tester Lead Selector
Power Polarity Neutral
ON NORM CLOSED ON REVERSE CLOSED ON NORM OPEN
ON REVERSE OPEN OFF NORM CLOSED OFF REVERSE CLOSED OFF NORM OPEN OFF REVERSE OPEN
ON NORM CLOSED ON REVERSE CLOSED ON NORM OPEN
ON REVERSE OPEN OFF NORM CLOSED OFF REVERSE CLOSED OFF NORM OPEN OFF REVERSE OPEN
Switch Switch RL RA LA LL C
ECG Tester Tester Tester Lead Selector
Power Polarity Neutral
Switch Switch RL RA LA LL C
ON NORM CLOSED ON REVERSE CLOSED ON NORM OPEN
ON REVERSE OPEN OFF NORM CLOSED OFF REVERSE CLOSED OFF NORM OPEN
MAXIMUM
ALLOWANCE LIMIT
OFF REVERSE OPEN
3–20
INST ALLA TION AND REPLACEMENT
Page 63
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21

3–7–6 When There’s Too Much Leakage Current...

Chassis Fails
Check the ground on the power cord and plug for continuity . Ensure the ground is not broken, frayed, or intermittent. Replace any defective part.
Tighten all grounds. Ensure star washers are under all ground studs.
Inspect wiring for bad crimps, poor connections, or damage.
T est the wall outlet; verify it is grounded. Notify the user or owner to correct any deviations. As a work around, check the other outlets to see if they could be used instead.
Note
No outlet tester can detect the condition where the white neutral wire and the green grounding wire are reversed. If later tests indicate high leakage currents, this should be suspected as a possible cause and the outlet wiring should be visually inspected.
P9030GX
Probe Fails
Test the probe in another transducer interface connector.
If excessive leakage current is slot dependent, inspect the Transducer Interface board for bent pins, poor connections, and ground continuity.
If the problem still has stayed with the probe, replace the probe.
Peripheral Fails
Tighten all grounds. Ensure star washers are under all ground studs.
Inspect wiring for bad crimps, poor connections, or damage.
Still Fails
If all else fails, begin isolation by removing the probes, external peripherals, then the on board ones, one at a time while monitoring the leakage current measurement.
New Unit
If the leakage current measurement tests fail on a new unit and if situation can not be corrected, submit a Safety Failure Report to document the system problem. Remove unit from operation.
3–21
INST ALLA TION AND REPLACEMENT
Page 64
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21

3–8 REPLACEMENT

3–8–1 GAS SPRING REPLACEMENT

1. Raise the monitor to the highest position.
2. Unscrew and then remove the bottom (sub) cover. (See Illustration 3–10)
P9030GX
SCREW
BOTTOM (SUB) COVER
ASSEMBLED MONITOR BOTTOM VIEW
ILLUSTRATION 3–10
3. Pull out the monitor cable and disconnect the connector. (See Illustration 3–11)
4. Remove two rubber plugs and unscrew the two neck pins and then lower the monitor to the lowest position. (See Illustration 3–12)
5. Turn the monitor until the groove in the rotation slider (See Illustration 3–1 1) comes to the position of the pin hole. (See Illustration 3–12)
6. Pull the monitor out of the monitor arm. Ensure that the nylon washer remains in place in the monitor arm. (See Illustration 3–12)
7. Unscrew the three screws fixing the monitor arm cover under the monitor arm and remove monitor arm cover. (See Illustration 3–13)
3–22
INST ALLA TION AND REPLACEMENT
Page 65
REV 21
3–8–1 GAS SPRING REPLACEMENT (continued)
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
PIN HOLE
NYLON WASHER
MONITOR ARM
WIRE HARNESS
CONNECTOR
ROTATION SLIDER
GROOVE
UNPACKED MONITOR CONSOLE SIDE VIEW
ILLUSTRATION 3–11 ILLUSTRATION 3–12
MONITOR CABLE
CONNECTOR
RUBBER PLUG
NECK PIN
RUBBER PLUG
BELLOWS
MONITOR ARM
MONITOR ARM COVER
SCREWS
BOTTOM VIEW – MONITOR ARM
ILLUSTRATION 3–13
8. Unscrew the two screws fixing the push bracket and remove the push bracket from the monitor arm. (See Illustration 3–14)
9. Unscrew the one screw and stud fixing the clamp plate and remove the clamp plate. (See Illustration 3–14)
3–23
INST ALLA TION AND REPLACEMENT
Page 66
REV 21
3–8–1 GAS SPRING REPLACEMENT (continued)
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
CLAMP PLATE
STUD
SCREW
FIXING SCREWS
BOTTOM VIEW – MONITOR ARM
ILLUSTRATION 3–14
MONITOR ARM
PUSH BRACKET
10. Remove the bellows both from the monitor arm and the main unit. Some force may have to be applied to loosen the adhesive holding the bellows.
11. Unscrew the four Allen screws to remove the monitor arm from the pipe.
MONITOR ARM
ALLEN SCREWS
PIPE
BELLOWS
RAISED MONITOR ARM
ILLUSTRATION 3–15
12. Unscrew the screws from the front cover, front plate, front panel, rear cover , and remove them so that the inside of the main body will be exposed.
3–24
INST ALLA TION AND REPLACEMENT
Page 67
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
P9030GX
3–8–1 GAS SPRING REPLACEMENT (continued)
13. Remove the connectors of signal cable and power supply (used for monitor) cable from the mother board. At the same time, the tiewrap that bundles them.
RIGHT SIDE
SIGNAL CABLE
MOTHER BOARD
FIXING BOLT
PANEL
STOPPER BRACKET
POWER SUPPLY CABLE
SYSTEM REAR VIEW (REAR PLATES REMOVED)
ILLUSTRATION 3–16
14. Unscrew two screws fixing stopper bracket located on the left side to rear, and remove the stopper bracket.
15. Unscrew two screws fixing ground wire from rear to neck support and remove the earth wire.
16. Cut the tiewrap fixing cable from monitor to main board.
17. Unscrew the screw fixing gas spring and remove the bolt.
3–25
INST ALLA TION AND REPLACEMENT
Page 68
REV 21
3–8–1 GAS SPRING REPLACEMENT (continued)
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
FAN
NECK SUPPORT
BOLT
TIEWRAP
SCREWS
REAR VIEW
ILLUSTRATION 3–17
BOARDS
SIDE BRACKET
18. Hold the pipe and raise it until the pin welded pipe reaches plastic of main board. Pull the pipe with a little twist.
19. Unscrew the nut fixing gas spring to pipe and remove the gas spring from the pipe. Hollow pipe without gas spring looks as shown in Illustration 3–19.
HEXAGONAL NUT FIXING GAS SPRING WITH PIPE.
BOTTOM VIEW – GAS SPRING
ILLUSTRATION 3–18
20. Put a new gas spring through the pipe and fix it with the nut.
21. Wind the cable four times around the gas spring and fix the cable to the gas spring with tape.
3–26
INST ALLA TION AND REPLACEMENT
Page 69
REV 21
3–8–1 GAS SPRING REPLACEMENT (continued)
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
PIPE
BELLOW
GAS SPRING ASSEMBLY
ILLUSTRATION 3–19
TAPE
GAS SPRING
HOLLOW PIPE
22. Put the assembly from step 21 back to the same position pulled from the main position. Be careful not to get the cable in between bearing and pipe.
23. Remove the tape fixing cable in step 21 from the low part of the main body and fix the part of cable covered the heat compression tube on neck support with tiewrap.
24. Insert the screw removed in step 17 to original position and fix the gas spring.
25. Fix two earth cables removed in step 15 with two screws.
26. Install the side bracket removed in step 14.
27. Install the signal cable and power supply cable removed in step 13 to the mother board.
28. Install the front cover, front plate, front panel and rear cover removed in step 12.
29. Install the monitor arm removed in step 11 to pipe.
30. Install the push bracket removed in step 8 with insert from the upper part to the monitor arm.
31. Install the clamp plate removed in step 9. Be careful as stud is easy to break.
32. Install the monitor arm cover removed in step 7 to the monitor arm.
3–27
INST ALLA TION AND REPLACEMENT
Page 70
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
3–8–1 GAS SPRING REPLACEMENT (continued)
33. Ensure that the nylon washer is in place (see Illustration 3–12) on the monitor arm.
34. Position the arm wire harness over the pin hole (see Illustration 3–12).
35. Gently place the monitor on the monitor arm, aligning the groove of the rotation slider with the pin–hole of the monitor arm. Take care to route the wire harness through the groove. (see Illustration 3–11 and 3–12)
36. Connect the connector. (see Illustration 3–11 and 3–12)
37. Fix the neck pins and insert the rubber plugs. (see Illustration 3–12)
38. Ensure that the monitor arm does not pinch the wire harness. (see Illustration 3–12)
39. Fix the bottom (sub) cover. (see Illustration 3–10)
40. Install the bellows removed in step 10 to both the monitor arm and the main unit. Additional Adhesive may be used to secure Bellows. (see Illustration 3–15)

3–8–2 Keyboard (Track Ball, TGC Assy) Replacement

P9030GX
1. Remove the two rubber plugs and screws from the front cover. (see Illustration 3–20)
2. Remove the front cover by unscrewing its three screws and the bottom cover of keyboard by unscrewing its seven screws. (see Illustration 3–20)
3. Disconnect the three cables from keyboard unit to the main unit. (see Illustration 3–21)
4. Unscrew the eight keyboard screws. (see Illustration 3–21)
5. Remove the keyboard (pc boards and touch panel as one assembly).
6. If required, remove the track ball assembly by disconnecting the connector and then unscrewing the four nuts.
7. If required, remove the TGC assembly by disconnecting the connector and then unscrewing the four screws.
8. Install the new keyboard and fasten its eight screws.
9. If removed, install the new track ball assembly, connect its connector, and screw it with its four nuts.
10. If removed, install the new TGC assembly, connect its connector, and screw it with its four screws.
11. Install the bottom cover with its seven screws.
12. Install the front cover with its five screws and two rubber plugs.
3–28
INST ALLA TION AND REPLACEMENT
Page 71
REV 21
3–8–2 Keyboard Replacement (continued)
3 SCREWS FOR THE FRONT COVER
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
2 RUBBER PLUGS AND SCREWS
8 SCREWS FOR THE KEYBOARD
FRONT COVER
KEYBOARD – BOTTOM COVER
ILLUSTRATION 3–20
CONNECTOR (CABLE TO THE MAIN UNIT)
CONNECTOR (CABLE TO THE MAIN UNIT)
7 SCREWS FOR THE BOTTOM COVER
BOTTOM COVER
CONNECTOR FOR THE TRACK BALL
CONNECTOR (FLAT CABLE TO THE MAIN UNIT)
TRACK BALL
KEYBOARD
4 NUTS FOR THE TRACK BALL
KEYBOARD – PC BOARD
ILLUSTRATION 3–21
3–29
TGC ASSY
4 SCREWS FOR THE TGC ASSY
INST ALLA TION AND REPLACEMENT
Page 72
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21

3–8–3 MONITOR REPLACEMENT

1. Refer to step 2 to 6 of 3–8–1 GAS SPRING REPLACEMENT for disassembly of the monitor.
2. Refer to step 33 to 39 of 3–8–1 GAS SPRING REPLACEMENT for assembly of the monitor.
P9030GX
3–30
INST ALLA TION AND REPLACEMENT
Page 73
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21

3–9 FRENCH ROM

P9030GX
Purpose
Installation of the French ROM on the MST Board.
Supply Parts List :
There are six kinds of French ROM. The one used depends on the MST Board’s Part number and Revision level. Install French EPROM(s) up carefully. See the Table below.
NO Version (Style) MST-E Assy French ROM LOCATION
1 Version 1 (style A) P9514XJ-02 P9514PF-01 U9A
Version 1 (style A) P9514XJ-03 P9514PF-01 U9A
2 Version 1 (style A) P9514XJ-04 P9514PF-02 U9A
P9514PL-02 U11A
3 Version 1 (style A) P9514XJ-05 P9514PF-03 U9A
P9514PL-03 U11A
4 Version 2 (style B) P9514HX-02 P9514PV-02 U9A
P9514PW-02 U11A
5 Version 4 (style D) P9514ZZ–00 P9514QT–00 U9A
P9514QU–00 U11A
6 Version 5 (style E) 2113953–3 2113969–3 U9A
2113970–3 U11A
Man Power
1.5~4 h 1 SVE Engineer (Depending on whether the system has already been used by customer or not; refer to Il-
lustration 3–22).
Required Tool
Standard Tolls
Procedure
This French ROM Installation Instruction is described for Version 2 (Style B) and Version 4 (Style D) mainly. If this procedure is used for version 1 (Style A) systems, be careful since there are some differences (skipped operations or different Illustrations) between this and Version 1 (Style A). This French ROM Installation Instruction consists of the following sections.
SECTION (A) : Data Recording SECTION (B) : Installation of the French ROM SECTION (C) : System Initialization SECTION (D) : Confirm the System is French Software.
SECTION (E) : Preset All Parameters reset by Initialization
If the system has already been used by the customer, it is necessary to record the user’s defined preset parameters prior to changing the EPROM(s). In this case, refer to the Operator Manual to input the system data.
3–31
INST ALLA TION AND REPLACEMENT
Page 74
REV 21
Procedure (continued)
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
System already
used by customer
Yes
SECTION (A)
SECTION (B)
SECTION (C)
SECTION (D)
SECTION (E)
No
SECTION (B)
SECTION (C)
SECTION (D)
Complete
PROCEDURE FLOW CHART
ILLUSTRATION 3–22
SECTION (A) Data Recording
The <Control> key , <Enter> key and Function key group are represented below:
<Control>
<Enter>
Function key group
OB
Tbl 1
Calc1
Calc3
OB
Tbl 2
Calc2
Right arrow
OPERATION KEYS
ILLUSTRATION 3–23
3–32
INST ALLA TION AND REPLACEMENT
Page 75
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (A) Data Recording (continued)
1. Preparation for Installation Perform functional checks to verify that the System is functioning properly . Refer to Functional Check Chapter 4
in this Service Manual.
2. Record or identify all parameters set by the user. Refer to Operator Manual to record or identify. 2–1. Record the Hospital Name.
i Record the Hospital Name, which is displayed in the upper–left corner of the monitor.
2–2. Record Output (Gamma Curve Selection).
i Identify whether the user was using video or polaroid. If the user was using polaroid , measure the Film
Development Time.
Measure the Film Development Time. i Press the <Control> ,R,P and then the <Enter> key. ii Press the <Record> key and then measure the Film Development Time and record it. iii Press the <Control> key.
2–3. Identify the Inverse Display Setting.
i Record the Monitor Display inverse setting (negative or positive). ii Press the <Record> key and then note the Inverse setting of the printout recorder. If the Monitor
Display changes polarity ( negative or positive) then the printout recorder is set opposite to the monitor.
P9030GX
2–4. Identify the Map Package.
i Read the Operator Manual carefully for this function to avoid confusion. ii Press the <Control> and N keys and then record whether the Map is A or B. iii Press the <Control> key.
2–5. Identify the Dotted line.
i Measure the length of something and then record whether the system Dotted Line is ON or OFF.
2–6. Identify the Frame Averaging.
i Watch the image carefully to compare the frame averaging condition before and after step ii is
performed.
ii Press the <Control>, W and <Enter> keys and then record whether the Frame Averaging is ON or
OFF.
iii Press the <Control> key.
2–7. Record the OB Table Data set by the user.
i. Press the <Control> key. ii Input ”G1” followed by the <Enter> key. iii Record all the OB Table Data set by the user and press the <Enter> key. iv Repeat steps ii and iii for each of G2,G3,G4 and G5. v Press the <Control> key.
2–8. Identify the Body Pattern Package in use.
i Press the < ii Record the Body Pattern which is displayed in the lower–left corner of the monitor.
iii Press the <
>key.
>key.
3–33
INST ALLA TION AND REPLACEMENT
Page 76
REV 21
SECTION (A) Data Recording (continued)
2–9. Record Probe Data.
The number of connectable probes differs depending on the Version (Style). The parameter setting image in Illustration 3–24 may not be the same in every case. Illustration 3–24 is shown as Version 2(Style B) . However the procedure is the same as the following:
i. Press the <Control>, S , ! and then the <Enter> key and verify that the CRT displays “SERVICE !” in
upper–left corner of the screen. ii Press the S , E and <Enter> keys. iii All the probe parameter settings appear on the screen. If they are different from those in the attached
list, record the screen (or write down any different items directly onto illustration 3–24). iv Press the <Control> key.
SERVICE! 13/12/91 13:55:26 CA 5MHz *1.0
PRB G DR FOCUS SCL DIR INV MAP HFR MZ 50 48 3 1.5 RVS NOR NOR OFF RA 50 54 3 1.0 NOR NOR NOR OFF L 50 54 3 1.0 NOR NOR NOR OFF S 50 54 3 1.0 NOR NOR NOR OFF K 50 54 3 1.0 NOR NOR NOR OFF A 50 54 3 1.0 NOR NOR NOR OFF B 50 54 3 1.0 NOR NOR NOR OFF C 50 54 3 1.0 NOR NOR NOR OFF D 50 54 3 1.0 NOR NOR NOR OFF E 50 54 3 1.0 NOR NOR NOR OFF F 50 54 3 1.0 NOR NOR NOR OFF G 50 54 3 1.0 NOR NOR NOR OFF H 50 54 3 1.0 NOR NOR NOR OFF I 50 54 3 1.0 NOR NOR NOR OFF J 50 54 3 1.0 NOR NOR NOR OFF R 50 54 3 1.0 NOR NOR NOR OFF CV 50 54 3 1.0 NOR NOR NOR OFF CA 50 54 3 1.0 NOR NOR NOR OFF CB 50 54 3 1.0 NOR NOR NOR OFF CC 50 54 3 1.0 NOR NOR NOR OFF
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
G50D54T60 CONTROL : SE
MANUFACTURE’S PRESET PARAMETER
ILLUSTRATION 3–24
3–34
GE
INST ALLA TION AND REPLACEMENT
Page 77
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (A) Data Recording (continued)
2–10.Record the additional anatomical features added by the user in the Anatomical Survey Summary (Version 2
Style B, Version 4 Style D or V ersion 5 Style E). Skip this step so that this image is not displayed for V ersion 1 (Style A). i. Press the <Control>, O and then the <Enter> key. ii Press the right arrow key at the bottom of the Function key group twice. iii The anatomical Survey Summary appears on the screen. If the operator added anatomical features to
the list, record the screen ( or write down any different items directly onto illustration 3–25). iv Press the <Control> key.
P9030GX
NAME : ID : REFERRAL : REFERRED FOR:
HEAD YES NO FOUR CHAMBER VIEW YES NO SPIN YES NO STOMACH YES NO KIDNEYS YES NO CORD INSERTION YES NO BLADDER YES NO UPPER EXTREMITIES YES NO LOWER EXTREMITIES YES NO
COMMENTS :
REPORTED BY :
EXAM DATE: 12/08/92
USER SETTING AREA
ANATOMICAL SURVEY TABLE
LMP :
ANATOMICAL SURVEY
IMAGED? APPEARANCE
YES NO YES NO YES NO YES NO
ILLUSTRATION 3–25
GRAVIDA: PARA :
REPORT DATE :
3–35
INST ALLA TION AND REPLACEMENT
Page 78
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (A) Data Recording (continued)
2–1 1.Record the User Programmable Comments Library (Version 2 Style B, V ersion 4 Style D or V ersion 5 Style
E). Skip this step so that this image is not displayed for Version 1 (Style A). i. Press the <Control>, O and then the <Enter> key. ii Press the right arrow key at the bottom of the Function key group three times. iii The User Programmable Comments Library appears on the screen. If the operator has entered a code
and text , record the screen ( or write down any different items directly onto illustration 3–26). iv Press the <Control> key.
USER PROGRAMMABLE COMMENTS LIBRARY
CODE TEXT
P9030GX
USER PROGRAMMABLE COMMENTS LIBRARY
ILLUSTRATION 3–26
3. Turn the power switch off and unplug the unit.
3–36
INST ALLA TION AND REPLACEMENT
Page 79
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (B) Installation of the French ROM
4. Remove the BACK PANEL to access the MST board. Refer to illustration 3–27. 4–1. Remove the six rubber caps. 4–2. Unscrew the six screws from the back panel of the RT–x200. 4–3. Remove the back panel from the console.
BACK PANEL
SCREWS
BACK PANEL REMOVAL
ILLUSTRATION 3–27
SCREWS
P9030GX
5. MST Board 5–1. Remove the MST Board (A3) from the card cage and confirm that the MST Board # is in the list below: 5–2. Remove the English ROM(s) at the location(s) listed in step 5–3.
CAUTION
Be sure the EPROM pins are straight before installation!
Be very careful to straighten pins if necessary.
5–3. Install the French ROM(s) in the following location(s) identified in the Table. Refer to illustration 3–28.
MST Board # Location English ROM Part # French ROM Part # P9514XJ-02
P9514XJ-03 P9514XJ-04
P9514XJ-05
P9514XH-02
U9A P9514PD-01 P9514PF-01 U9A P9514PD-02 P9514PF-02
U11A P9514PE-02 P9514PL-02 U9A P9514PD-03 P9514PF-03
U11A P9514PE-03 P9514PL-03 U9A P9514PT-02 P9514PV-02
U11A P9514PU-02 P9514PW-02
P9514ZZ–00
2113953
U9A P9514NW–00 P9514QT-00
U11A P9514NX–00 P9514QU–00 U9A 2113967–3 2113969–3 U11A 2113968–3 2113970–3
3–37
INST ALLA TION AND REPLACEMENT
Page 80
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (B) Installation of the French ROM (continued)
P9030GX
A3 MST ASSY COMPONENT SIDE
ILLUSTRATION 3–28
5–4. Reinstall the MST board. 5–5. Reinstall the back panel of the RT–x200 unit.
6. Put the English EPROM in the ROM case and then ask the customer to keep it.
7. Plug In the RT–x200 unit.
8. Turn the power on.
3–38
INST ALLA TION AND REPLACEMENT
Page 81
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (C) System Initialization
9. System Initialize 9–1. Press the <Control>, S, ! and then the <Enter> key. 9–2. Press the S, V and <Enter> keys. 9–3. Press the S, W and <Enter> keys. 9–4. Press the S, Z and <Enter> keys.
SECTION (D) Confirm the system is French ROM software.
10 Confirm the system is French ROM software.
Gestational Report Page of Version 1 (style A) has some dif ferent points from Illustration 3–29. Confirm image is displayed in French.
10–1.Press the <Control>, G, R and <Enter> keys. 10–2.Confirm that the Gestational Report Page is displayed in French. Refer to Illustration 3–29. 10–3.Press the <Control> key.
P9030GX
NOM :
NOM : ID :
ID : REFERENT :
MOTIF VISITE : AGE GESTATION
AG DDR :
BIP (HADLOCK)
DOF
LCC (ROBINSON)
LF (O BRIEN) PC (HADLOCK)
DAT DAP PA (HADLOCK)
(TBL2) (TBL2) (TBL2)
(TBL2) (TBL2)
AGC : COMMENT :
EXCRIT PAR : DATE :
DATE EXAM: 12/08/92
AGE : DDR :
AGC? N AGC? N
AGC? N AGC? N
AGC? N AGC? N AGC? N
AGC? N AGC? N
AGC? N
FRENCH SOFTWARE
ILLUSTRATION 3–29
CALCULS INDEX CEPHAL : %
LF/BIP: %
LF/PA : % PC/PA : %
POIDS:
#1 (BIP, PA) : GR #2 (LF, PA) : GR
#3 (LF, PA, PC) : GR BPM : BPM IFA : cm
TERME/DDR : TERME/AGC :
/
GRAVI DE :
PARA :
%
3–39
INST ALLA TION AND REPLACEMENT
Page 82
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (E) Preset all the parameters reset by the initialization.
11. Preset all the parameters reset by the initialization. 11–1.Hospital Name setting.
i. Press the <Control>, H and then the <Enter> key. ii Type “Hospital Name” and press the <Enter> key. Confirm that the Hospital Name is displayed in
upper–left corner on the monitor.
iii Press the <Control> key.
11–2.Record Output (Gamma Curve Selection) setting.
i. Preset the Record Output (Gamma Curve Selection) recorded in step 2–2.
Refer to the Operator Manual to preset the Record Output (Gamma Curve Selection). If user was using polaroid , preset the Film Development Time.
Film Development Time setting. i. Preset the Film Development Time recorded in step 2–2.
Refer to the Operator Manual to preset the Film Development Time.
11–3.Inverse setting.
i. Preset the Inverse setting recorded in step 2–3
Refer to the Operator Manual to preset the Inverse.
P9030GX
11–4.Map Package setting.
Press the <Control> key
Yes No
Press the <Control> key Press the <Enter> key
. 11–5.Dotted Line setting.
i. Preset the Dotted Line setting recorded in step 2–5.
Refer to the Operator Manual to preset the Dotted Line.
11–6.Frame Averaging setting.
i. Preset the Frame Averaging setting recorded in step 2–6.
Refer to the Operator Manual to preset the Frame Averaging.
11–7.OB Table Data settings.
i. Preset the OB Table Data recorded in step 2–7.
Refer to the Operator Manual to preset OB Table Data.
Press N key
Is the display the same as step 2–4?
1 1–8.Body Pattern Package settings.
i. Preset the Body Pattern Package recorded in step 2–8.
Refer to the Operator Manual to preset the Body Pattern Package.
3–40
INST ALLA TION AND REPLACEMENT
Page 83
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (E) (continued)
11–9.Probes parameter settings.
i. Connect one of the probes used by the doctor and modify the value of the probe parameter settings in
order to obtain the correct values (Gain, Dynamic, Focus, Scale, Direction, Inverse, Map and HFR).
Refer to the Operator Manual to preset the Probes parameter. ii Press the <Control>, B and then the <Enter> key. Now, all the settings for one probe are completed. iii Repeat the same setting procedure for the other probes. iv Press the <Control> key.
11–10.Additional anatomical features settings (Version 2 style B, Version 4 Style D or Version 5 Style E).
Skip this step for Version 1 (Style A). i. Enter the Additional anatomical features recorded in step 2–10.
Refer to the Operator Manual to enter Additional anatomical features.
11–1 1.User Programmable Comments Library (Version 2 style B, Version 4 Style D or Version 5 Style E)..
Skip this step for Version 1 (Style A). i Enter code and text recorded in step 2–11.
Refer to the Operator Manual to Enter code and text.
12. Dispose of excess material.
P9030GX
13. This completes the procedures.
3–41
INST ALLA TION AND REPLACEMENT
Page 84
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21

3–10 GERMAN ROM

P9030GX
Purpose
Installation of the German ROM on the MST Board.
Supply Parts List :
There are two kinds of German ROM. The one used depends on the MST Boards Parts number. Install German EPROMs up carefully. See the Table below.
NO Version (Style) MST-E Assy German ROM LOCATION
1 Version 4 (style D) P9514ZZ–00 P9514QV U9A
P9514QW U11A
2 Version 5 (style E) 2113953–3 2113971–3 U9A
2113972–3 U11A
Man Power
1.5~4 h 1 SVE Engineer (Depending on whether the system has already been used by customer or not; refer to Il­lustration 3–30).
Required Tool
Standard Tolls
Procedure
This German ROM Installation Instruction consists of the following sections.
SECTION (A) : Data Recording SECTION (B) : Installation of the German ROM SECTION (C) : System Initialization SECTION (D) : Confirm the System is German Software.
SECTION (E) : Preset All Parameters reset by Initialization
If the system has already been used by the customer, it is necessary to record the user’s defined preset parameters prior to changing the EPROM(s). In this case, refer to the Operator Manual to input the system data.
3–42
INST ALLA TION AND REPLACEMENT
Page 85
REV 21
Procedure (continued)
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
System already
used by customer
Yes
SECTION (A)
SECTION (B)
SECTION (C)
SECTION (D)
SECTION (E)
No
SECTION (B)
SECTION (C)
SECTION (D)
Complete
PROCEDURE FLOW CHART
ILLUSTRATION 3–30
SECTION (A) Data Recording
The <Control> key , <Enter> key and Function key group are represented below:
<Control>
<Enter>
Function key group
OB
Tbl 1
Calc1
Calc3
OB
Tbl 2
Calc2
Right arrow
OPERATION KEYS
ILLUSTRATION 3–31
3–43
INST ALLA TION AND REPLACEMENT
Page 86
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (A) Data Recording (continued)
1. Preparation for Installation Perform functional checks to verify that the System is functioning properly . Refer to Functional Check Chapter 4
in this Service Manual.
2. Record or identify all parameters set by the user. Refer to Operator Manual to record or identify. 2–1. Record the Hospital Name.
i Record the Hospital Name, which is displayed in the upper–left corner of the monitor.
2–2. Record Output (Gamma Curve Selection).
i Identify whether the user was using video or polaroid. If the user was using polaroid , measure the Film
Development Time.
Measure the Film Development Time. i Press the <Control> ,R,P and then the <Enter> key. ii Press the <Record> key and then measure the Film Development Time and record it. iii Press the <Control> key.
2–3. Identify the Inverse Display Setting.
i Record the Monitor Display inverse setting (negative or positive). ii Press the <Record> key and then note the Inverse setting of the printout recorder. If the Monitor
Display changes polarity ( negative or positive) then the printout recorder is set opposite to the monitor.
P9030GX
2–4. Identify the Map Package.
i Read the Operator Manual carefully for this function to avoid confusion. ii Press the <Control> and N keys and then record whether the Map is A or B. iii Press the <Control> key.
2–5. Identify the Dotted line.
i Measure the length of something and then record whether the system Dotted Line is ON or OFF.
2–6. Identify the Frame Averaging.
i Watch the image carefully to compare the frame averaging condition before and after step ii is
performed.
ii Press the <Control>, W and <Enter> keys and then record whether the Frame Averaging is ON or
OFF.
iii Press the <Control> key.
2–7. Record the OB Table Data set by the user.
i. Press the <Control> key. ii Input ”G1” followed by the <Enter> key. iii Record all the OB Table Data set by the user and press the <Enter> key. iv Repeat steps ii and iii for each of G2,G3,G4 and G5. v Press the <Control> key.
2–8. Identify the Body Pattern Package in use.
i Press the < ii Record the Body Pattern which is displayed in the lower–left corner of the monitor.
iii Press the <
>key.
>key.
3–44
INST ALLA TION AND REPLACEMENT
Page 87
REV 21
SECTION (A) Data Recording (continued)
2–9. Record Probe Data.
The number of connectable probes differs depending on the Version (Style). The parameter setting image in Illustration 3–32 may not be the same in every case. Illustration 3–32 is shown as Version 4(Style D) . However the procedure is the same as the following:
i. Press the <Control>, S , ! and then the <Enter> key and verify that the CRT displays “SERVICE !” in
upper–left corner of the screen. ii Press the S , E and <Enter> keys. iii All the probe parameter settings appear on the screen. If they are different from those in the attached
list, record the screen (or write down any different items directly onto illustration 3–32). iv Press the <Control> key.
SERVICE! 13/12/91 13:55:26 CA 5MHz *1.0
PRB G DR FOCUS SCL DIR INV MAP HFR CK 50 54 3 1.0 NOR NOR NOR OFF ATV 50 54 3 1.0 NOR NOR NOR OFF MZ 50 48 3 1.5 RVS NOR NOR OFF RA 50 54 3 2.0 NOR NOR NOR OFF L 50 54 3 1.0 NOR NOR NOR OFF S 50 54 3 1.0 NOR NOR NOR OFF K 50 54 3 1.0 NOR NOR NOR OFF A 50 54 3 1.0 NOR NOR NOR OFF B 50 54 3 1.0 NOR NOR NOR OFF C 50 54 3 1.0 NOR NOR NOR OFF D 50 54 3 1.0 NOR NOR NOR OFF E 50 54 3 1.0 NOR NOR NOR OFF F 50 54 3 1.0 NOR NOR NOR OFF G 50 54 3 1.0 NOR NOR NOR OFF H 50 54 3 1.0 NOR NOR NOR OFF I 50 54 3 1.0 NOR NOR NOR OFF J 50 54 3 1.0 NOR NOR NOR OFF R 50 54 3 1.0 NOR NOR NOR OFF CV 50 54 3 1.0 NOR NOR NOR OFF CA 50 54 3 1.0 NOR NOR NOR OFF CB 50 54 3 1.0 NOR NOR NOR OFF CC 50 54 3 1.0 NOR NOR NOR OFF
G50D54T60 CONTROL : SE
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
GE
MANUFACTURE’S PRESET PARAMETER
ILLUSTRATION 3–32
3–45
INST ALLA TION AND REPLACEMENT
Page 88
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (A) Data Recording (continued)
2–10.Record the additional anatomical features added by the user in the Anatomical Survey Summary.
i. Press the <Control>, O and then the <Enter> key. ii Press the right arrow key at the bottom of the Function key group twice. iii The anatomical Survey Summary appears on the screen. If the operator added anatomical features to
the list, record the screen ( or write down any different items directly onto illustration 3–33). iv Press the <Control> key.
P9030GX
NAME : ID : REFERRAL : REFERRED FOR:
HEAD YES NO FOUR CHAMBER VIEW YES NO SPIN YES NO STOMACH YES NO KIDNEYS YES NO CORD INSERTION YES NO BLADDER YES NO UPPER EXTREMITIES YES NO LOWER EXTREMITIES YES NO
COMMENTS :
REPORTED BY :
EXAM DATE: 12/08/92
USER SETTING AREA
ANATOMICAL SURVEY TABLE
LMP :
ANATOMICAL SURVEY
IMAGED? APPEARANCE
YES NO YES NO YES NO YES NO
ILLUSTRATION 3–33
GRAVIDA: PARA :
REPORT DATE :
3–46
INST ALLA TION AND REPLACEMENT
Page 89
REV 21
SECTION (A) Data Recording (continued)
2–11.Record the User Programmable Comments Library .
i. Press the <Control>, O and then the <Enter> key. ii Press the right arrow key at the bottom of the Function key group three times. iii The User Programmable Comments Library appears on the screen. If the operator has entered a code
and text , record the screen ( or write down any different items directly onto illustration 3–34). iv Press the <Control> key.
USER PROGRAMMABLE COMMENTS LIBRARY
CODE TEXT
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
USER PROGRAMMABLE COMMENTS LIBRARY
ILLUSTRATION 3–34
3. Turn the power switch off and unplug the unit.
3–47
INST ALLA TION AND REPLACEMENT
Page 90
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (B) Installation of the French ROM
4. Remove the BACK PANEL to access the MST board. Refer to illustration 3–35. 4–1. Remove the six rubber caps. 4–2. Unscrew the six screws from the back panel of the RT–x200. 4–3. Remove the back panel from the console.
BACK PANEL
SCREWS
BACK PANEL REMOVAL
ILLUSTRATION 3–35
SCREWS
P9030GX
5. MST Board 5–1. Remove the MST Board (A3) from the card cage and confirm that the MST Board # is in the list below: 5–2. Remove the English ROM(s) at the location(s) listed in step 5–3.
CAUTION
Be sure the EPROM pins are straight before installation!
Be very careful to straighten pins if necessary.
5–3. Install the German ROM(s) in the following location(s) identified in the Table. Refer to illustration 3–36.
MST Board # Location English ROM Part # German ROM Part # P9514ZZ–00
2113953
U9A P9514NW–00 P9514QV
U11A P9514NX–00 P9514QW U9A 2113967–3 2113971–3 U11A 2113968–3 2113972–3
3–48
INST ALLA TION AND REPLACEMENT
Page 91
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (B) Installation of the French ROM (continued)
P9030GX
A3 MST ASSY COMPONENT SIDE
ILLUSTRATION 3–36
5–4. Reinstall the MST board. 5–5. Reinstall the back panel of the RT–x200 unit.
6. Put the English EPROM in the ROM case and then ask the customer to keep it.
7. Plug In the RT–x200 unit.
8. Turn the power on.
3–49
INST ALLA TION AND REPLACEMENT
Page 92
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (C) System Initialization
9. System Initialize 9–1. Press the <Control>, S, ! and then the <Enter> key. 9–2. Press the S, V and <Enter> keys. 9–3. Press the S, W and <Enter> keys. 9–4. Press the S, Z and <Enter> keys.
SECTION (D) Confirm the system is German ROM software.
10 Confirm the system is German ROM software.
10–1.Press the <Control>, G, R and <Enter> keys. 10–2.Confirm that the Gestational Report Page is displayed as shown in Illustration 3–37. 10–3.Press the <Control> key.
P9030GX
NAME : ID :
REFERRAL:
REFERRED FOR: GESTATIONAL AGE
GA BY LMP :
BPD(HANSMANN) CRL(HANSMANN) FL(HANSMANN) GSD(HANSMANN) ATD(HANSMANN)
(TBL2) (TBL2) (TBL2)
(TBL2) (TBL2)
CGA : COMMENT :
REPORTED BY :
EXAM DATE: 04 / 04 / 93
AGE : LMP :
CGA? N CGA? N CGA? N CGA? N CGA? N
CGA? N CGA? N CGA? N CGA? N CGA? N
GERMAN OB SOFTWARE
ILLUSTRATION 3–37
EGA:
CALCULATIONS FL/BPD : %
EFBW:
(BIP, ATD): 0,000 KG
HEART RATE: BPM AFI: cm
/
EDD BY LMP : EDD BY EGA : EDD BY CGA :
REPORT DATE :
GRAVI DA :
PARA :
%
3–50
INST ALLA TION AND REPLACEMENT
Page 93
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 21
SECTION (E) Preset all the parameters reset by the initialization.
11. Preset all the parameters reset by the initialization. 11–1.Hospital Name setting.
i. Press the <Control>, H and then the <Enter> key. ii Type “Hospital Name” and press the <Enter> key. Confirm that the Hospital Name is displayed in
upper–left corner on the monitor.
iii Press the <Control> key.
11–2.Record Output (Gamma Curve Selection) setting.
i. Preset the Record Output (Gamma Curve Selection) recorded in step 2–2.
Refer to the Operator Manual to preset the Record Output (Gamma Curve Selection). If user was using polaroid , preset the Film Development Time.
Film Development Time setting. i. Preset the Film Development Time recorded in step 2–2.
Refer to the Operator Manual to preset the Film Development Time.
11–3.Inverse setting.
i. Preset the Inverse setting recorded in step 2–3
Refer to the Operator Manual to preset the Inverse.
P9030GX
11–4.Map Package setting.
Press the <Control> key
Yes No
Press the <Control> key Press the <Enter> key
. 11–5.Dotted Line setting.
i. Preset the Dotted Line setting recorded in step 2–5.
Refer to the Operator Manual to preset the Dotted Line.
11–6.Frame Averaging setting.
i. Preset the Frame Averaging setting recorded in step 2–6.
Refer to the Operator Manual to preset the Frame Averaging.
11–7.OB Table Data settings.
i. Preset the OB Table Data recorded in step 2–7.
Refer to the Operator Manual to preset OB Table Data.
Press N key
Is the display the same as step 2–4?
1 1–8.Body Pattern Package settings.
i. Preset the Body Pattern Package recorded in step 2–8.
Refer to the Operator Manual to preset the Body Pattern Package.
3–51
INST ALLA TION AND REPLACEMENT
Page 94
REV 21
SECTION (E) (continued)
11–9.Probes parameter settings.
i. Connect one of the probes used by the doctor and modify the value of the probe parameter settings in
order to obtain the correct values (Gain, Dynamic, Focus, Scale, Direction, Inverse, Map and HFR).
Refer to the Operator Manual to preset the Probes parameter. ii Press the <Control>, B and then the <Enter> key. Now, all the settings for one probe are completed. iii Repeat the same setting procedure for the other probes. iv Press the <Control> key.
11–10.Additional anatomical features settings.
i. Enter the Additional anatomical features recorded in step 2–10.
Refer to the Operator Manual to enter Additional anatomical features.
11–1 1.User Programmable Comments Library.
i Enter code and text recorded in step 2–11.
Refer to the Operator Manual to Enter code and text.
12. Dispose of excess material.
13. This completes the procedures.
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
3–52
INST ALLA TION AND REPLACEMENT
Page 95
REV 0
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
This page is left blank intentionally.
4–2
FUNCTIONAL CHECKS
Page 96
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 0

4–1 INTRODUCTION

This chapter provides procedures for quickly checking of the RTfino/RT3200 Advantage–II/RT–x200. The other service chapters are Chapter 6, Service Adjustments.
Note
P9030GX
These procedures form part of the Installation Checklist found in Chapter 3, Installation, and
should be performed during installation.
However, these procedures are also the basic checks to use when service is required. In this
case, use this section’s table of contents to locate and exercise the specific function related to the service complaint.

4–1–1 Required Equipment

To perform these tests, you’ll need the following items:
3.5 MHz B Type Linear Transducer or a 3.5 MHz CB type convex transducer.
4–3
FUNCTIONAL CHECKS
Page 97
REV 5

4–2 FUNCTIONAL CHECK PROCEDURES

4–2–1 Power Up

RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
Test Setup/Operation
1. Connect the B type transducer or the CB type transducer to Probe Receptacle No. 1.
2. Turn on POWER switch.
Functional Check/Evaluation
a. Two beeps sound.
b. The following LED is ON continuously:
Preset
c. After about two seconds, the following LEDs light:
1. Focus – 3
2. Scale – X1
d. After about 3 seconds, these LEDs lightup :
Mode B / Probe / Sweep speed
e. After about four seconds, these displays appear on
CRT (See Illustration 4–1):
1. B–Mode Image
2. Gray Scale Display
3. Horizontal Scale Marker
4. Vertical Range Marker
5. Hospital Name (Optional)
6. Date (Day, Month, Year)
7. Time (Hour, Min., Second)
8. Probe Type/Frequency
9. Scale Factor
10. Gain (G50)
11. Dynamic Range (D54)
12. Transmit Focus (T60) (Items 10 through 12 form a continuous code in
lower left hand corner of CRT.)
4–4
FUNCTIONAL CHECKS
Page 98
REV 0
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
5
3
1
6
7
GE MED SYSTEMS 90 / 10 / 09 09 : 41 : 38 B 3.5MHz
x1.0
8
9
2
G50D54T60
10
11
12
4

4–2–2 REMAINING FUNCTIONAL CHECK

Please refer to OPERATOR MANUAL for the remaining functional checks.
POWER UP CRT DISPLAY
ILLUSTRATION 4–1
4–5
FUNCTIONAL CHECKS
Page 99
REV 0
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
P9030GX
This page is left blank intentionally.
5–2
THEORY
Page 100
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUALGE MEDICAL SYSTEMS
REV 0
P9030GX

5–1 INTRODUCTION

RTfino/RT3200 Advantage–II/RT–x200 is a compact ultrasound scanner. It meets the routine needs of both Obstet­rics and Gynecology investigations. Support of wide range of probes, gives the system added benefits to meet the diverse applications.

5–2 PRINCIPLES OF ULTRASOUND IMAGING

Ultrasound is produced by applying a short duration, high voltage pulse to the piezoelectric elements in a transducer. This causes these elements to vibrate at their resonant frequency . (This is determined by the element’s thickness and tuning.) If the element is making good mechanical contact with the skin, ultrasound waves will be transmitted into the tissue. (Ultrasound gel applied to the skin prior to ultrasound scanning maximizes transmission).

5–2–1Ultrasound Emission

Ultrasound imaging is based on the sonar principle. A short pulse of sound is transmitted from the transducer’s piezoelectric crystal element into a medium of known density and propagation velocity, in this case soft tissue.

5–2–2 Reflection

1. The product of density and velocity is called acoustic impedance. If any change in acoustic impedance is encoun­tered by the sound pulse, some of the sound will be reflected back to the piezoelectric element in the transducer. The remaining pulse will continue to greater body depths. The greater the change in acoustic impedance, the more sound will be reflected.
2. As the sound travels out and back, an electronic device is used to measure the travel time. Knowing the travel time (go–return time) and the velocity of the ultrasound wave, the distance traveled can be calculated. Thus, diagnostic ultrasound images consist of plots of distances between changes in acoustic impedances.
3. The propagation velocity of an ultrasound wave traveling through different types of body tissues is relatively con­sistent. An average of 1540 meters per second is used as the standard for normal diagnostic ultrasound applica­tions. Actually, the percentage of ultrasound wave reflected by soft tissue averages less than 1%. However, signal reflection increases to 41 to 48 percent when soft tissue–bone interfaces are encountered. T able 5–1 pro­vides reflection percentages for various types of materials.
TABLE 5–1
REFLECTION PERCENTAGES
MATERIAL TYPE
Fat–Muscle Interface (I/F) 1.08 Fat–Kidney I/F 0.64 Muscle–Blood I/F 0.07 Bone–Fat I/F 48.91 Bone–Muscle I/F 41.23 Soft Tissue–Water I/F 0.23 Soft Tissue–Air I/F 99.90 Reference: McDicken (1976)
PERCENT REFLECTION
5–3
THEORY
Loading...