GE Rightrax HT Operating Manual

GE
Sensing & Inspection Technologies
Rightrax HT Specications
Online Corrosion Monitoring System
HT350x Sensor
Nominal frequency Nominal bandwidth Operating mode Typical measuring range 3- 16 mm (0.120 - 0.63”) in mild steel
Contact diameter Couplant Coupling pressure Mounting Cabling CMX conguration – 5 m (16.4 ft) length cable to open leads (no connector)
Delay length Delay material
5.0 Mhz 100% Pulse echo
Measuring range is affected by material properties, coupling and other factors 7 - 21 mm (0.820”) depending on contact surface radius Metal foil
Varies depending on test surface dimensions and nish
Adaptable clamping system
Stand alone conguration - 5 m (16.4 ft) length coax cable
17.7 mm (0.7”)
304 Stainless steel
CMX-HT Sensor Node
Probe system Single element 5Mhz probe with stainless steel delay line. The probe is clamped onto the pipe and coupled to the
System thickness resolution
Temperature Range
Thickness resolution single reading Thickness resolution with regression
± 0.0025 mm (0.0001”)
-20 to +350°C
(-4° up to
662°F) ± 0.01 mm
(0.0004”)
± 0.0025 mm (0.0001”)
surface with a gold foil. The resolution of the thickness reading through the complete system is the smallest change that can be reliably
detected. This is inuenced by short-term temperature uctuations. These variations will result in changes in the temperature prole across the pipe wall.
Skin surface temperature of test surface.
With temperature variations at rates less then 1°C (2°F) per minute over a 5°C (10°F) range, the resolution is improved to the value obtained using regression analysis. Reading errors due to short-term uctuations are removed after taking a series of readings over time and
performing a regression analysis.
www.gesensinginspection.com
© 2009 General Electric Company. All Rights Reserved. Specications subject to change without notice.
GEIT-20212EN (12/09)