GE
Sensing & Inspection Technologies
Rightrax HT Specications
Online Corrosion Monitoring System
HT350x Sensor
Nominal frequency
Nominal bandwidth
Operating mode
Typical measuring range 3- 16 mm (0.120 - 0.63”) in mild steel
Contact diameter
Couplant
Coupling pressure
Mounting
Cabling CMX conguration – 5 m (16.4 ft) length cable to open leads (no connector)
Delay length
Delay material
5.0 Mhz
100%
Pulse echo
Measuring range is affected by material properties, coupling and other factors
7 - 21 mm (0.820”) depending on contact surface radius
Metal foil
Varies depending on test surface dimensions and nish
Adaptable clamping system
Stand alone conguration - 5 m (16.4 ft) length coax cable
17.7 mm (0.7”)
304 Stainless steel
CMX-HT Sensor Node
Probe system Single element 5Mhz probe with stainless steel delay line. The probe is clamped onto the pipe and coupled to the
System thickness
resolution
Temperature Range
Thickness resolution
single reading
Thickness resolution
with regression
± 0.0025 mm
(0.0001”)
-20 to +350°C
(-4° up to
662°F)
± 0.01 mm
(0.0004”)
± 0.0025 mm
(0.0001”)
surface with a gold foil.
The resolution of the thickness reading through the complete system is the smallest change that can be reliably
detected. This is inuenced by short-term temperature uctuations. These variations will result in changes in the
temperature prole across the pipe wall.
Skin surface temperature of test surface.
With temperature variations at rates less then 1°C (2°F) per minute over a 5°C (10°F) range, the resolution is
improved to the value obtained using regression analysis.
Reading errors due to short-term uctuations are removed after taking a series of readings over time and
performing a regression analysis.
www.gesensinginspection.com
© 2009 General Electric Company. All Rights Reserved. Specications subject to change without notice.
GEIT-20212EN (12/09)