GE Remote Alarm Box User manual

Remote Alarm Box RAB &
Remote Alarm Box for Remote Light RAB RL
Service Manual
Version 1.0
2006406-010 SA (eng) Revision A
GE Medical Systems Information Technologies
NOTE: Due to continuing product innovation, specifications in this manual are subject to change without notice.
Trademarks Trademarked names appear throughout this document. Rather than list
the names and entities that own the trademarks or insert a trademark symbol with each mention of the trademarked name, the publisher states that it is using the names only for editorial purposes and to the benefit of the trademark owner with no intention of improperly using that trademark.
ACCUSKETCH, AccuVision, APEX, AQUA-KNOT, ARCHIVIST, Autoseq, BABY MAC, CardioServ, CardioSmart, CardioSys, CardioWindow, CASE, CD TELEMETRY, CENTRA, CHART GUARD, CINE 35, CORO, COROMETRICS, CRG PLUS, Digistore, Digital DATAQ, E for M, EAGLE, Event-Link, HELLIGE, IMAGE STORE, INTELLIMOTION, LASER SXP, MAC, MAC-LAB, MACTRODE, MARQUETTE, MARQUETTE MAC, MARQUETTE UNITY NETWORK, MARS, MAX, MEDITEL, MEI, MEI in the circle logo, MEMOPORT, MEMOPORT C, MINISTORE, MINNOWS, Monarch 8000, MULTI-LINK, MULTISCRIPTOR, MUSE, MUSE CV, Neo-Trak, OnlineABG, OXYMONITOR, Pres-R-Cuff, PRESSURE-SCRIBE, QMI, QS, Quantitative Medicine, Quantitative Sentinel, Qwik Connect Spiral, RAMS, RSVP, SAM, SEER, SOLAR, SOLARVIEW, Spectra 400, Spectra-Tel, ST GUARD, TRAM, TRAM-NET, TRAM-RAC, TRAMSCOPE, TRIM KNOB, UNITY logo, UNITY NETWORK, Vari-X, Vari-X Cardiomatic, VariCath, VAS, and Vision Care Filter are trademarks of GE Medical Systems Information Technologies registered in the United States Patent and Trademark Office.
12SL, 15SL, AccuSpeak, ADVANTAGE, BAM, BODYTRODE, Cardiomatic, CardioSpeak, CD TELEMETRY .-LAN, CENTRALSCOPE, Corolation, Corometrics Sensor Tip, Dash, EDIC, HI-RES, IMAGE VAULT, IMPACT.wf, INTER-LEAD, IQA, LIFEWATCH, MARQUETTE MEDICAL SYSTEMS, MARQUETTE . RESPONDER, MENTOR, MicroSmart, MMS, MRT, MUSE CardioWindow, O 2 SENSOR, OMRS, Premium, RAC, SILVERTRACE, SMART-PAC, SMARTLOOK, Spectra-Overview, Trimline, UNITY, and Universal are trademarks of GE Medical Systems Information Technologies
© GE Medical Systems Information Technologies, 2001. All rights reserved.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 2 / 45
GE Medical Systems Information Technologies
CONTENTS
1 INTRODUCTION.................................................................................................................. ................................4
1.1 M
1.2 S
1.3 S
1.4 T
1.5 H
ANUAL INFORMATION ....................................................................................................................................4
AFETY INFORMATION.......................................................................................................................................6
ERVICE INFORMATION.....................................................................................................................................8
ERMINOLOGY ..................................................................................................................................................9
OW TO REACH US .........................................................................................................................................10
2 EQUIPMENT OVERVIEW................................................................................................................................12
2.1 P
2.2 LED I
2.3 D
2.4 S
RODUCT DESCRIPTION...................................................................................................................................12
NDICATOR DESCRIPTION........................................................................................................................13
ESCRIPTION OF OPERATION ...........................................................................................................................14
PECIFICATION MECHANICA L, ELECTRICAL....................................................................................................15
3 MAINTENANCE .................................................................................................................................................16
3.1 M
3.2 V
3.3 G
3.4 C
3.5 E
AINTENANCE SCHEDULE ..............................................................................................................................16
ISUAL INSPECTION ........................................................................................................................................17
ENERAL CLEANING .......................................................................................................................................18
HECKOUT PROCEDURES ................................................................................................................................19
LECTRICAL SAFETY TEST ..............................................................................................................................21
3.5.1 Ground (Earth) Wire Leakage Current Tests.........................................................................................22
3.5.2 Enclosure Leakage Current Test............................................................................................................24
3.6 M
AINTENANCE BLINK MODE ..........................................................................................................................25
4 TROUBLESHOOTING.......................................................................................................................................26
4.1 E
4.2 RAB / RAB RL
4.3 S
4.4 T
LECTROSTATIC DISCHARGE (ESD)................................................................................................................26
FOR YOUR SAFETY.................................................................................................................27
ERVICE MODE RAB / RAB RL .....................................................................................................................27
ROUBLESHOOTING FAQ’S.............................................................................................................................28
5 CONFIGURATION.............................................................................................................................................30
5.1 CONFIGURATION J
5.2 C
5.3 C
5.4 C
5.5 C
ONNECTING THE CABLE TO THE HOSPITAL ALARM SYSTEM..........................................................................31
ONNECTING THE ALARM LIGHT (RAB RL)...................................................................................................32
ONNECTING THE POWER SUPPLY UNIT (RAB RL) ........................................................................................33
ONNECTING RAB / RAB RL TO THE PATIENT MONITOR ..............................................................................35
UMPER..........................................................................................................................30
6 CONNECTORS....................................................................................................................................................36
6.1 H
6.2 RAB R
6.3 RAB P
6.4 C
OSPITAL ALARM SYSTEM CONNECTOR .........................................................................................................37
EMOTE LIGHT (RAB RL ONLY) CONNECTOR ......................................................................................39
OWER SUPPLY (RAB RL ONLY) CONNECTOR......................................................................................40
ABLING RAB RL INTERNAL ..........................................................................................................................41
7 ASSEMBLY DRAWINGS...................................................................................................................................42
7.1 S
PAREPARTS....................................................................................................................................................43
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 3 / 45
GE Medical Systems Information Technologies
1 INTRODUCTION
1.1 Manual Information
Scope of the Manual
The content of this field service manual is aimed primarily at biomedical equipment technicians and field service personnel. The user of this field service manual is expected to have a solid background in electronics, including strong backgrounds in analog and digital electronics, as well as microcomputer technology familiarity.
Revision History
Each page of this manual has a revision letter located at the bottom of the page. It identifies the revision level of the entire manual. This may be important if you have different manuals and you don’t know which is the most current. For the initial release, all pages have the revision letter A. For the second update, all pages receive the revision letter B. The latest letter of the alphabet added to the table below corresponds to the most current revision.
Revision History
Revision Date Comment
A 04.05.2001 Initial release of this document
Manual Purpose
This field service manual has been prepared by the technical publications staff at GE Medical System - IT, Inc. It is intended for use by biomedical electronic technicians or other qualified service personnel responsible for installation, maintenance or repair of the RAB or RAB RL.
Chapter Content
The field service manual is organized into sections, as follows:
Introduction
Chapter one, “Introduction”, describes the revision history, manufacturer responsibility, notes/cautions/warnings and abbreviations.
Equipment Overview
Chapter two, “Equipment Overview”, describes the product, technical specifications, preparation for use and theory of operation.
Maintenance
Chapter three, “Maintenance”, describes the maintenance schedule, visual inspection, cleaning the monitor, checkout procedures, leakage current tests and “Blink Mode speciality”.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 4 / 45
GE Medical Systems Information Technologies
Troubleshooting
Chapter four, “Troubleshooting”, describes electro-static discharge, service mode and FAQ’s
Configuration
Chapter five, “Configuration”, describes jumper configurations, installing the Remote Alarm Box
Connectors
Chapter six, “Connectors”, describes pin assignment of the connectors.
Assembly Drawings
Chapter seven, “Assembly Drawings”, provides mechanical diagrams, reference diagrams, schematic diagrams and parts lists.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 5 / 45
GE Medical Systems Information Technologies
1.2 Safety Information
Responsibility of the Manufacturer
GE Medical Systems is responsible for the effects of safety, reliability, and performance only if:
Assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized by GE Medical Systems.
The electrical installation of the relevant room complies with the requirements of the appropriate regulations.
The equipment is used in accordance with the instructions for use.
Intended Use
This device is intended for use under the direct supervision of a licensed health care practitioner. To ensure patient safety, use only parts and accessories manufactured or recommended by GE Medical Systems. Contact GE Medical Systems for information before connecting any devices to this equipment that are not recommended in this manual.
Equipment Symbols
The following symbols appear on the equipment.
NOTE
Some symbols may not appear on all equipment. ATTENTION: Consult accompanying documents before using the equipment.
Consult accompanying documents
AC voltage
DC voltage
LED indicator
Protection class II (Caution the power supply unit for the alarm light of RAB RL is
a protection class I device)
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 6 / 45
GE Medical Systems Information Technologies
Notes, Cautions, and Warnings
The safety statements presented in this section apply to the components of the RAB and RAB RL Remote Alarm Box. Look for additional safety information throughout the rest of this manual. The order in which safety statements are presented in no way implies order of importance.
The terms WARNING, CAUTION, and NOTE are used throughout this manual to point out hazards and to designate a degree or level or seriousness. Familiarize yourself with their definitions and significance. Hazard is defined as a source of potential injury to a person.
WARNING
indicates a potential hazard or unsafe practice which, if not avoided, could result in death or serious injury.
CAUTION
indicates a potential hazard or unsafe practice which, if not avoided, could result in minor personal injury or product/property damage.
NOTE
provides application tips or other useful information to assure that you get the most from your equipment.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 7 / 45
GE Medical Systems Information Technologies
1.3 Service Information
Service Requirements
Follow the service requirements listed below.
Refer equipment servicing to GE Medical Systems authorized service personnel only.
Any unauthorized attempt to repair equipment under warranty voids that warranty.
It is the user’s responsibility to report the need for service to GE Medical Systems or to one of their
authorized agents.
Failure on the part of the responsible individual, hospital, or institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards.
Regular maintenance, irrespective of usage, is essential to ensure that the equipment will always be functional when required.
Equipment Identification
Every GE Medical Systems device has a unique serial number for identification.
Warranty
1 year
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 8 / 45
GE Medical Systems Information Technologies
1.4 Terminology
RAB Remote Alarm Box RAB RL Remote Alarm Box with Remote Light UTP CABLE Unshielded Twisted Pair Cable PFD Power Failure Detection Mode of RAB / RAB RL PWA Printed Wire Assembly DMM Digital Multi Meter GEMS-IT GE Medical Systems Information Technologies
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 9 / 45
GE Medical Systems Information Technologies
1.5 How to Reach Us
Service Calls and Product Support
To open a service call or obtain product support call the numbers below:
or contact your representative or distributor For other product information please contact one of the offices listed on
the next page.
Ordering Supplies &Service Parts
Order supplies (leadwires, electrode paste, thermal paper, etc.)or service parts (manuals, circuit boards, cables, software, etc.)from:
800.558.7044 (US &Canada)
561.575.5000 (outside US)
Accessories
GE Medical Systems Accessories and Supplies 2607 North Grandview Blvd. Mail Code:SN-471 Waukesha,WI 53188
Telephone: 800.558.5102 (US only)
262.521.6856 (outside U.S.)
Fax: 800.232.2599 (US only)
262.521.6855 (outside US)
Service Parts
GE Clinical Services P.O.Box 9100,100 Marquette Drive Jupiter,FL 33468-9100
Telephone: 800.558.7044 (US only)
561.575.5000 (outside US)
Fax: 800.421.6841 (US only)
561.575.5050 (outside US)
Have the following information available before calling:
part number of the defective part, or model and serial number of the equipment, part number/name of the assembly where the item is used, item name, and where applicable, reference designation (eg,R13,S12).
Ordering Manuals
When ordering additional operator manuals, be sure to include the software version of the product.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 10 / 45
GE Medical Systems Information Technologies
Other Questions or Problems
For additional information contact one of the offices listed below or see our website at www.gemedicalsystems.com
Headquarters
GE Medical Systems Information Technologies 8200 West Tower Avenue Milwaukee,WI 53223,USA Telephone: 414.355.5000 or 800.558.5120 (US only) Fax: 414.355.3790
Europe
GE Medical Systems Information Technologies Postfach 60 02 65 D-79032 Freiburg, Germany Telephone: +49.761.4543.0 Fax: +49.761.4543.233
Australia
GE Medical Systems (Australia)Pty Ltd. 13 South Street Rydalmere NSW 2116,Australia Telephone: +61.2.9975.5501 Fax: +61.2.9975.5503
Japan
GE Medical Systems,Japan 67-4 Takakura-cho Hachiojii-shi,Tokyo 192-003,Japan Telephone: +81.42.648.2944
+0120-055-919 (toll free inside Japan only)
Fax: +81.24.648.2902
Hong Kong
GE Medical Systems Hong Kong Limited 11th Floor,The Lee Gardens 33 Hysan Avenue Causeway Bay,Hong Kong Telephone: +852.2100.6300 Fax: +852.2100.6292
Southeast Asia
GE Pacific 298 Tion Bahru Road #15-01/06 Central Plaza,Singapore 168730 Telephone: +65.277.7620 Fax: +65.277.7600
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 11 / 45
GE Medical Systems Information Technologies
2 EQUIPMENT OVERVIEW
2.1 Product Description
Remote Alarm Box (RAB)
The Remote Alarm Box is a microprocessor controlled isolation relay (secondary alarm device). It is used to connect the device listed below to the hospital alarm system:
Dash patient monitors
Dash Port Docking Station
Figure 2.1 Remote Alarm Box
Solar 8000M patient monitor
Remote Alarm Box with Remote Light (RAB RL)
The RAB RL is used to connect an alarm light (Remote Light, P/N 8500 AAX) to the devices listed above (secondary alarm light). This RAB RL does not connect to the hospital alarm system. A special, hospital grade power supply unit is required to supply power to the alarm light; This unit is included in the shipment.
Figure 2.2 Remote Alarm Box RAB RL
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 12 / 45
GE Medical Systems Information Technologies
Both, RAB and RAB RL will function correctly only in conjunction with monitors with the software versions listed below:
Monitor Software Version
Dash 2000 V3A or later Dash 3000 Dash 4000 Solar8000M V2A or later
1)
not via Dash Port Docking Station
V2B V2B
1)
or later
1)
or later
Table 2.1 RAB Software compatibility
2.2 LED Indicator Description
Figure 2.3 RAB LED Interface
Explanation of the LED Indicator
Indicator Function
indicate the selected alarm
A, B, C
mode: A : continuous mode
B : intermittent mode C : blink mode
D
mode with power failure detection
E
lights up in alarm situation
Table 2.2 LED Indicator
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 13 / 45
GE Medical Systems Information Technologies
2.3 Description of Operation
Figure 2.4 RAB Block Diagram
Block diagram of RAB :
The interface to the connected Patient Monitoring System is realized by a standard serial RS232 communication interface. So, in different to the existing Isolation Boxes („Nurse Call Boxes“) a small microprocessor is needed to realize the communication part of the box. The RAB / RAB RL operation mode could be configured by three configuration jumpers. The µProcessor is requesting every 500ms the alarm status of the patient monitor via serial communication protocol. If the Monitor is in “ALARM”-state, the RAB / RAB RL toggles the alarm relay according to the configured alarm mode. If the Monitor is not in this state, the RAB / RAB RL switches the relay back (if it was in alarm state) or the relay remain the “non-alarm” position according to the configured operating mode. After power up of the RAB / RAB RL there is a onetime switch delay of approx. 20 seconds to let the patient monitor startup to normal operation. An independent hardware watchdog / power supervisor chip is provided to guarantee proper powerup/ powerdown and software behavior. The configured device status and also the detected alarm status are displayed by 5 LED’s. On Patient Monitor side the communication interface is realized as a M-Port interface with an 8-pin RJ-45 connector. To connect the box to a hopsital alarm call system a 3 terminal screwing connector is provided and to connect the RL version of the RAB to the EAGLE Remote Light and external IEC601-1 power supply, a 9­pin SUB-D and a 3-pin power connector are used .
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 14 / 45
GE Medical Systems Information Technologies
2.4 Specification Mechanical, Electrical
Mechanical Specification:
Size of the Housing : 114mm x 72mm x 45mm
Electrical Specification :
RAB / RAB RL
Operating Voltage +5V (+-10%), max. 0.1A
Relay Switching Capacity 24V DC / 1A
250V AC / 1A
(Blink Mode : 24V DC / 24V AC, 0.1 A !)
Relay Isolation Voltage 4000V - RAB / RAB RL to Relay Output
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 15 / 45
GE Medical Systems Information Technologies
3 MAINTENANCE
3.1 Maintenance Schedule
Manufacturer Recommendation
To make sure the RAB / RAB RL remains in proper operational and functional order, a good maintenance schedule must be adhered to. The manufacturer's recommendations in this regard are as follows:
Visual Inspection: These should be performed by qualified service personnel upon receipt of the equipment, every 12 months thereafter, and each time the monitor is serviced.
General Cleaning: Service personnel should perform this after servicing the monitor.
Checkout Procedures: These should be performed by qualified service personnel upon receipt of the
equipment, every 12 months thereafter, and each time the monitor is serviced.
Leakage Current Tests: These should be performed by qualified service personnel upon receipt of the equipment, every 12 months thereafter, and each time the monitor is serviced.
Maintenance “Blink Mode” : If Blink mode is used a special maintenance must be performed by qualified service personnel every 12 months.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 16 / 45
GE Medical Systems Information Technologies
3.2 Visual Inspection
Inspecting the Monitor
The RAB / RAB RL should be carefully inspected. Follow these guidelines when inspecting the equipment:
Carefully inspect the RAB / RAB RL for obvious physical damage to the outer case. Do not use the RAB / RAB RL if physical damage is determined. Refer damaged equipment to qualified service personnel for repair before using it again on a patient.
Inspect all external connectors, front and rear, for degraded pins, prongs and connector housings. Refer damaged equipment to qualified service personnel for repair before using it again on a patient.
Inspect all cable insulation, cable strain-reliefs and cable connectors for damage, cracks or degradation. Refer damaged equipment to qualified service personnel for repair before using it again on a patient. For RAB RL check the fixing of the power cable at power supply and the RAB RL.
Safety labels and inscriptions on the device are clearly legible.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 17 / 45
GE Medical Systems Information Technologies
3.3 General Cleaning
Clean External Surfaces
The exterior surfaces may be cleaned with a lint-free cloth dampened with one of these approved solutions:
ammonia (diluted),
Cidex,
mild soap (dissolved), or
sodium hypochlorite bleach (diluted).
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 18 / 45
GE Medical Systems Information Technologies
3.4 Checkout Procedures
Checkout Procedures
The following pages contain the checkout procedures for the RAB / RAB RL. The purpose of the checkout procedures is to provide service personnel with a method which can be used to verify operational and functional performance of the RAB / RAB RL. Failure to attain any of the listed results indicates a potential malfunction of the RAB / RAB RL. Perform the checkout procedures upon receipt of the RAB / RAB RL, every twelve months thereafter, and each time the circuit board is removed or replaced. The checkout procedures are based on the assumption that the RAB / RAB RL being tested is used with known good cables and test equipment. It also requires that the user be somewhat familiar with the operation of all test equipment required for the checkout procedures. For more information concerning the operation of these components, refer to the respective operator manual.
Functional Test
It’s recommended that the RAB / RAB RL will be checked connected to a complete system with DASH2000, DASH 3000, DASH4000 or SOLAR8000M as a bedside.
For the functional test switch the monitor to a “NO Alarm” state.
RAB / RAB RL Mode Relay
Intermittent Mode with PFD closed Intermittent Mode open Continuous Mode with PFD closed Continuous Mode open Blink Mode with PFD closed Blink Mode open Test Mode for Service only Open for 0.5 seconds and
Closed for 0.5 seconds, periodical
Table 3.1 RAB Function Test
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 19 / 45
GE Medical Systems Information Technologies
After this change the monitor to a “Alarm “ state.
RAB / RAB RL Mode Relay
Intermittent Mode with PFD Open for 100 seconds and
Closed for 1 seconds, intermittent
Intermittent Mode Closed for 100 seconds and
Open for 1 seconds,
intermittent Continuous Mode with PFD open Continuous Mode closed Blink Mode with PFD Open for 2 seconds and
Closed for 2 seconds,
intermittent Blink Mode Closed for 2 seconds and
Open for 2 seconds,
intermittent Test Mode for Service only Open for 0.5 seconds and
Closed for 0.5 seconds,
periodical
Table 3.2 RAB RL Function Test
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 20 / 45
GE Medical Systems Information Technologies
3.5 Electrical Safety Test
General
Electrical safety test provide a method of determining potential electrical health hazards to the patient or operator of the device exist.
Recommendations
To help you establish a systematic maintenance routine, GEMS-IT recommends that you perform all safety tests presented in this chapter
upon receipt of the device,
every twelve months thereafter,
each time the main enclosure is disassembled or a circuit board is removed, tested, repaired, or
replaced, and
record the date and results on the “Maintenance/Repair Log” included at the end of this chapter.
CAUTION
Failure to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards. Unless you have an Equipment Maintenance Contract, GE Medical Systems does not in any manner assume the responsibility for performing the recommended maintenance procedures. The sole responsibility rests with the individual or institution using the equipment. GEMS-IT service personnel may, at their discretion, follow the procedures provided in this manual as a guide during visits to the equipment site.
Test Conditions
Electrical safety tests may be performed under normal ambient conditions of temperature, humidity, and pressure.
Test Equipment
The manufacturer recommended test equipment required to perform electrical safety tests is listed below (see table 3.3). Equivalent equipment may be substituted as necessary.
Required Tools/Special Equipment
Leakage Current Tester 120 V 240 V (equivalent to the figures 3.1 - 3.2) Multimeter 0 – 1000mV
Table 3.3 RAB Safety Test special Equipment
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 21 / 45
GE Medical Systems Information Technologies
Wall Receptacle Test
Before starting the tests, the wall receptacle from which the monitoring device will get electrical power must be checked. This test checks the condition of the wall receptacle to ensure correct results from leakage tests. For international wall receptacles, refer to the internal standards agencies of that particular country. Use a digital multimeter to ensure the wall receptacle is wired properly. If other than normal polarity and ground is indicated, corrective action must be taken before proceeding. The results of the following tests will be meaningless unless a properly wired wall receptacle is used.
3.5.1 Ground (Earth) Wire Leakage Current Tests
This test needs to be performed only with RAB RL. Perform this test to measure current leakage through the ground (earth) wire of the equipment during
normal operation.
1. Set the leakage tester switches as follows:
GND switch - OPEN,
Polarity switch – NORM and RVS
Power switch - OFF.
2. Connect the DMM to the METER jacks on the leakage tester. Set the DMM to measure AC millivolts.
3. Connect the power cord of the device under test to the power receptacle on the rear of the leakage tester.
NOTE
The device under test is to be tested at its normal operating voltage.
4. Set the leakage tester power switch to ON.
5. Read the current leakage indicated on DMM. If the reading is greater than the appropriate
specification below, the device under test fails and should be repaired and tested again.
300 microamperes (0.3 volts on the DMM), and the device under test is powered from 100 - 120 V / 50 - 60 Hz
300 µA (0.3 volts on the DMM), and the device under test is powered from a centered-tapped 200 - 240 V / 50 - 60 Hz, single- phase circuit
500 µA (0.5 volts on the DMM), and the device under test is powered from a non-center-tapped, 200 - 240 V / 50 - 60 Hz, single-phase circuit
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 22 / 45
GE Medical Systems Information Technologies
NOTE
Center-tapped and non-center-tapped circuits produce different leakage currents and the UL and IEC limits are different.
6. Set the leakage tester power switch to OFF.
Figure 3.1 RAB Ground (Earth) Wire Leakage Current Tests
NOTES
The MD (measuring device) is the circuitry defined by the appropriate standard for measuring leakage current. The measuring devices, defined by various standard organizations (IEC, UL, etc.), produce almost identical test measurement results.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 23 / 45
GE Medical Systems Information Technologies
3.5.2 Enclosure Leakage Current Test
This test needs to be performed only with RAB RL. Perform this test to measure current leakage through exposed
conductive surfaces on the device under test during normal operation.
1. Set the leakage tester switches as follows:
GND switch - OPEN,
Polarity switch – NORM
2. Connect a meter lead between the GND of the tester accordingly to the circuit and the metal housing of the RL connector and one of the metal screws of the remote light..
3. Set the leakage tester power switch to ON.
4. Read the current leakage indicated on DMM. If the reading is greater than the
appropriate specification below, the device under test fails and should be repaired and tested again.
300 microamperes (0.3 volts on the DMM), and the device under test is powered from 100 - 120 V / 50 - 60 Hz
300 µA (0.3 volts on the DMM), and the device under test is powered from a centered-tapped 200 - 240 V / 50 - 60 Hz, single phase circuit
500 µA (0.5 volts on the DMM), and the device under test is powered from a non-center-tapped, 200 - 240 V / 50 - 60 Hz, single-phase circuit
NOTE
Center-tapped and non-center-tapped circuits produce different leakage currents and the UL and IEC limits are different.
5. Set the polarity switch to RVS and observe the same meter readings as in the previous step.
6. Set the GND switch on the leakage tester to CLOSED.
7. Read the current leakage indicated on DMM. If the reading is greater than the appropriate specification below, and the device under test is powered from 100-240 V/50-60 Hz,
the device under test fails and should be repaired and tested again.
100 microamperes (0.1 volts on the DMM), and the device under test is powered from 100-240 V/50-60 Hz
8. Set the polarity switch to RVS and observe the same meter readings as in the previous step.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 24 / 45
GE Medical Systems Information Technologies
9. Set the leakage tester power switch to OFF and remove the meter lead connected in step 2.
Figure 3.2 Enclosure Leakage Current Test
Test Completion
Disconnect all test equipment from the device. Disconnect the device power cord plug from the leakage tester power receptacle. Disconnect the leakage tester from the wall receptacle.
3.6 Maintenance Blink Mode
Due of the maximum switching cycle of the relay contacts there is a reduced relais life time in Blink Mode. If the Blink mode is used, the PWA must be replaced every 12 months.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 25 / 45
GE Medical Systems Information Technologies
4 TROUBLESHOOTING
4.1 Electrostatic Discharge (ESD)
CMOS Components
The monitor makes extensive use of CMOS components because they are more immune to noise and consume less power than standard TTL or NMOS components. However, CMOS components are inherently more susceptible to electrostatic discharge (ESD) damage than other types of semiconductor materials. ESD damage, causing a weakening or complete breakdown of p-n junctions within multilayer semiconductor substrates, can range from slight degradation to catastrophic failure. Slight degradation usually results in intermittent failure of the affected component catastrophic failure results in rendering the affected component permanently unusable. Although CMOS components may be more sensitive to ESD, all semiconductor devices are susceptible to ESD damage. All external connector inputs and outputs of the monitor are designed with protection from ESD damage. However, if the monitor requires service, exposed components and assemblies contained within are susceptible to ESD damage. This includes human hands, non-ESD protected work stations and/or improperly grounded test equipment. The following guidelines help make a service workstation more resistant to the ESD damage:
Discharge any static charge you may have built up before handling semiconductors or assemblies containing semiconductors.
A grounded, antistatic wristband (3M part number 2046 or equivalent) or heel strap should be worn at all times while handling or repairing assemblies containing semiconductors.
Use properly grounded soldering and test equipment.
Use a static-free work surface (3M part number 8210 or equivalent) while handling or working on
assemblies containing semiconductors.
Do not remove semiconductors or assemblies containing semiconductors from antistatic containers (Velo-stat bags) until absolutely necessary.
Make sure power to an assembly is turned off before removing or inserting a semiconductor.
Do not slide semiconductors or electrical/electronic assemblies across any surface.
Do not touch semiconductor leads unless absolutely necessary.
Semiconductors and electrical/electronic assemblies should be stored only in antistatic bags or boxes.
These guidelines may not guaranty a 100% static-free workstation, but can greatly reduce the potential for failure of any electrical/electronic assemblies being served.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 26 / 45
GE Medical Systems Information Technologies
4.2 RAB / RAB RL for your Safety
Before opening the RAB, please disconnect the cable to the hospital alarm system from the wall inlet to prevent the risk of higher voltage inside the RAB.
For RAB RL also disconnect the power supply cable before opening the RAB RL. After configuration of the device to the alarm mode used in the specific hospital and closing / mounting of
the device, always a functional test of the complete alarm path (Patient Monitor -> RAB / RAB RL -> Hospital Alarm System / Remote Alarm Light with external power supply) must be performed!
4.3 Service Mode RAB / RAB RL
For a functional control of the RAB / RAB RL installation without using a Patient Monitor, a special service mode is implemented. After open the RAB / RAB RL box please configure the configuration jumpers to SERVICE MODE (J1:2-3, J2:2-3, J3:2-3). Please remember the original settings of the jumpers to reconfigure the device to the original settings.
This mode is only for service purposes. It will toggle the relay output every half second. This mode toggles also LED E every half second from ON to OFF and vice versa . It helps to ensure the functionality of the RAB / RAB RL microprocessor system , the relay output and the connected hospital alarm system / remote alarm light.
NOTE
When in working in this mode there will be no failure detection or alarm processing! Using this mode there will be no test of the alarm communication path from the patient monitor to the RAB / RAB RL! (Patient Monitor, cable, RAB / RAB RL communication interface are not tested).
NOTE
After reconfiguration of the device to the alarm mode used in the specific hospital and closing / mounting of the device, always a functional test of the complete alarm path must be performed!
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 27 / 45
GE Medical Systems Information Technologies
4.4 Troubleshooting FAQ’s
This section covers (when not specifically mentioned) both devices, RAB and RAB RL
No reaction of the RAB / RAB RL after power up
If the RAB / RAB RL will not react on connecting it to a patient monitor / docking station there could be the following listed reasons for this behavior: When used with DASH / DASH PORT docking station a AUX-Port to M-Port Adapter must be used at patient monitor / DASH PORT docking station side. When used with SOLAR8000M, no adapter must be used. Cable to patient monitor could be defective or has no contact to the patient monitor / docking station connector or the RAB / RAB RL connector AUX-Port to M-Port Adapter could be defective or has no contact At patient monitor side, cable is plugged into wrong connector (DASH ­> Ethernet) or the RAB could be defective. Please replace the RAB / RAB. Patient Monitor’s or Docking Station’s M-port / AUX-Port has a defect.
After power up, not all LED’s are flashing ON for a short time
After powering the device, all five LED’s must flash ON for nearly a second. This is a visual control of the LED’s circuits functionality. If this error occurs, LED, LED transistor or specific microcontroller output could have a malfunction. Please replace the RAB / RAB.
Error detected, all five LED’s are blinking
There are several tests done by the internal microprocessor during startup and normal operation of the device. Because all tests check functional critical parts of the device, a detected error means that the correct functionality of the device can not be guaranteed. Please replace the RAB / RAB.
RAB / RAB RL works normal, but no relay toggle / LED “E” ON when alarm
There can be the following listed reasons for this behavior: Cable to patient monitor / docking station or AUX-Port to M-Port adapter is defective or RAB is defective Please replace the RAB / RAB. Patient Monitor’s or Docking Station’s M-port / AUX-Port has a defect.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 28 / 45
GE Medical Systems Information Technologies
RAB / RAB RL works normal, also in alarm state, but no reaction of Alarm System:
Check complete cabling to the hospital Alarm System Check hospital Alarm System functionality Check the fuse inside the RAB. If the fuse was burned, please replace the RAB. Check the functionality of the RAB relay. To do this, disconnect the RAB from the Alarm System and use an Ohm Meter to check the correct functionality of the relay. If the relay / LED E will not work properly, please replace the RAB.
RAB RL:
Check the 12 VDC of the medical power supply. If the power supply is defective, please replace it. Check the wiring inside the RAB RL for broken connections. When the power supply is connected to the RAB RL, verify the voltages at the RAB RL remote light connector. (See 4.3.3 pinning). If the voltages are OK (also when in alarm state -> +12VDC on pin ALARM) and the remote light will not work, please replace the alarm light. Check the fuse inside the RAB. If the fuse was burned, please replace the RAB RL. This should never happen, if it will, also the power supply and the remote light will surely be defective. Check the functionality of the RAB relay. To do this, disconnect the RAB RL from the Alarm System and use an Ohm Meter to check the correct functionality of the relay. If the relay / LED E will not work properly, please replace the
RAB RL.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 29 / 45
GE Medical Systems Information Technologies
5 CONFIGURATION
5.1 CONFIGURATION Jumper
Alarm Mode Configuration (RAB / RAB RL)
Figure 5.1 Jumper Description (Example)
Table 5.1 Alarm Mode Jumper
The factory default setting for the alarm mode is the “Intermittent Mode (100 seconds) with Power Failure Detection”. If your application requires a different alarm mode, configure this mode as follows before connecting the RAB / RAB RL.
Remove the 4 screws on the underside of the housing.
Take off the upper part of the housing.
You will see the three jumpers J 1, J 2 and J 3 on the PWA. The jumper configuration determines the alarm mode (Figure 5.1). Upon delivery, all three jumpers are in position 1-2 (intermittent mode – 100 seconds, with power failure detection). Refer to Table 5.1 for all possible jumper configurations.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 30 / 45
GE Medical Systems Information Technologies
5.2 Connecting the Cable to the Hospital Alarm System
Warning
Patient Safety — Choose cables and connectors with the appropriate voltage and current carrying capacity.
Note
When operating the hospital alarm system in the current monitoring mode, please connect the cables appropriately.
Loosen the screw of the cable collar on the PWA.
Guide the cable through the sleeve.
Connect the cable to the relay contacts as required for the configured alarm mode:
- terminals REL 2 and REL 3: break contact
(open when relay attracts)
- terminals REL 1 and REL 2: make contact
(closed when relay attracts)
Secure the cable with the collar. To ensure that thin cables are securely held in place by the cable collar, wrap them in insulating tape or reinforce them with shrink-down tubing.
Place the upper part of the housing on the lower part, checking the correct position of the optical light guides.
Screw tight the four screws on the underside of the housing.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 31 / 45
GE Medical Systems Information Technologies
5.3 Connecting the Alarm Light (RAB RL)
The factory default wiring of the alarm light connector (1, Figure 5.2) is for an alarm mode with power failure detection, i.e., when the relay is released, the light is on. To select a mode without power failure detection, change the wiring internally as follows:
Figure 5.2 Connecting RAB RL
Open the housing and disconnect the cable to terminal REL 3 and connect it to terminal REL 1.
Close the housing.
Connect the alarm light to connection (1, Figure 5.2).
Secure the connection with the screws provided on the plug.
4
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 32 / 45
GE Medical Systems Information Technologies
5.4 Connecting the Power Supply Unit (RAB RL)
Note
You can install the power supply unit either by means of a Velcro strip attached to the back side of the support, or mount it to a GCX system. When using a GCX system, the appropriate fixation elements must be mounted on the support before the power supply unit can be inserted in the support.
Remove the upper part of the cable collar (6, Figure 5.3).
Plug the power supply cable into the socket (5, Figure 5.3).
Place the cable in the lower part of the cable collar
(6, Figure 5.3) and screw tight the upper part of the collar to secure the able.
Insert the power supply unit (the legs pointing down) in
Screw the bracket with the cable collar on to the support
Remove the upper part of the cable collar (9 Figure 5.5).
Place the power cord in the lower part of the cable collar and
Connect the power supply unit to the power line.
Figure 5.3 RAB RL connecting power supply
the support (7 Figure 5.4).
(8 Figure 5.5).
Connect the lower cord to the power supply unit.
screw tight the upper part of the collar to secure the cable (10, Figure 5.6).
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 33 / 45
GE Medical Systems Information Technologies
Figure 5.4 a) RAB RL mounting the power supply
Figure 5.5 b) RAB RL mounting the power supply
Figure 5.6 c) RAB RL mounting the power supply
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 34 / 45
GE Medical Systems Information Technologies
5.5 Connecting RAB / RAB RL to the Patient Monitor
Caution
Patient Safety — Only the GEMS IT supplied UTP cable or an equivalent cable may be used to connect the device to the patient monitor. Observe the following points to ensure correct functioning of the RAB / RAB RL: unshielded, twisted-pair cable, 8-pole, max. length 5 m, min. AWG 26, gold-plated contacts; do not use X-cables.
Plug the supplied UTP cable into socket.To do so, push the plug into the socket until you hear it click into place. Briefly pull on the cable to check that the connection is secure. When connecting the device to a Dash patient monitor or to a Dash Port Docking Station, the adapter ADPTR AUTOPORT M-PORT (2001973-001) must be connected to the AUX port of the monitor / Dash Port Docking Station first (Figures 5.7).
When connecting the device to a SOLAR 8000M patient monitor, the cable is directly plugged into one of the M-PORT sockets (Figures 5.8).
Connect the cable as shown in Figures 5.7 and 5.8. Ensure that the connector(s) engage properly.
Figure 5.7 RAB M-Port to Autoport Adapter
Figure 5.8 RAB M-Port UPT cable
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 35 / 45
GE Medical Systems Information Technologies
6 Connectors
General
The connector to the Patient Monitor is directly M-Port compatible, so the RAB could be connected directly to a SOLAR8000M patient monitor.
To connect the RAB to a DASH patient monitor, the Autoport adapter PN 2001973-001, Rev.B or newer must be used.
Monitor Connector RJ-45
Figure 6.1 RAB / RAB RL RJ-45 Connector
Pinning:
PIN No. Label Direction/ Power Description
1 ETH1 N.a. M-Port Signal, not used 2 ETH2 N.a. M-Port Signal, not used 3 SID I/O Serial Identification Pin,
1-Wire Protocol 4 M_PORT_GND Power Ground return line 5 TXD Output RS 232 Transmit line 6 SID_RETURN Power Serial Identification, Ground return line 7 RXD Input RS232 Receive line 8 M_PORT_VCC Power +5V Power Input
Table 6.1 RAB / RAB RL RJ-45 Connector
Input Data
SID Input signal to communicate from Patient Monitor (e.g. with M-Port,
like SOLAR8000M) with the DS2430A E²PROM to determine Serial ID number.
RXD Input signal to get patient monitor alarm status via RS232 communication
Output Data
TXD Output signal to request patient monitor alarm status via RS232 communication.
ETH1,ETH2 M-Port Ethernet differential lines, not connected, not used
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 36 / 45
GE Medical Systems Information Technologies
Power Signal M_PORT_VCC Power input signal (5V) M_PORT_GND Power Ground SID_RETURN Power Ground signal from Patient Monitor (e.g. with M-Port, like
SOLAR8000M) to determine Serial ID number
6.1 Hospital Alarm System Connector
Figure 6.2 RAB Hospital Alarm System Connector
REL1
REL2
REL3
Figure 6.3 RAB Hospital Alarm System Connector - Schematic
REL1 Contact to be connected with REL2 if relay current is switched on REL2 Middle contact of the relay REL3 Contact to be connected with REL2 if relay is not powered
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 37 / 45
GE Medical Systems Information Technologies
How to connect the cable to Hospital Alarm System
Loosen the screw of the cable collar on the PWA.
Guide the cable through the sleeve.
Connect the cable to the relay contacts as required for the configured alarm mode:
- terminals REL 2 and REL 3: break contact (open when relay activates)
- terminals REL 1 and REL 2: make contact (closed when relay activates)
Secure the cable with the collar. To ensure that thin cables are securely held in place by the cable collar, wrap them in insulating tape or reinforce them with shrink-down tubing.
Place the upper part of the housing on the lower part, checking the correct position of the optical light guides.
Screw tight the four screws on the underside of the housing.
Warning
Patient Safety — Choose cables and connectors with the appropriate voltage and current carrying capacity.
Note
The RAB can be connected to the hospital alarm system in different ways, depending on whether or not the power failure detection mode is configured and whether or not the hospital alarm system is using the current monitoring mode.
Power failure detection mode ENABLED
hospital alarm system using current monitoring mode: connect terminals REL 1 and REL 2 (break contact when power off/alarm)
hospital alarm system not using current monitoring mode: connect terminals REL 2 and REL 3 (make contact when power off/alarm)
Power failure detection mode DISABLED
hospital alarm system using current monitoring mode: connect terminals REL 2 and REL 3 (break contact when alarm )
hospital alarm system not using current monitoring mode: connect terminals REL 1 and REL 2 (make contact when alarm)
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 38 / 45
GE Medical Systems Information Technologies
6.2 RAB Remote Light (RAB RL only) Connector
Figure 6.4 RAB RL Remote Light Connector
A 9 pin Sub-D connector to connect the EAGLE Remote Light 8500AAX to the RAB RL Pinning of the 9-pin Sub-D connector:
PIN number
1 VS Power Output +12 VDC 2 - - not connected 3 - - not connected 4 - - not connected 5 - - not connected 6 - - not connected 7 COM Power Output Ground 8 ALARM Signal Output +12V: Alarm
9 - - not connected Shield Shield - connected to Alarm Light shield not connected to RAB
Table 6.2 RAB RL Remote Light Connector
Label Direction/ Power Description
NC/Ground: No Alarm
RL, no other electrical connection
Output Data
VS +12VDC to EAGLE Remote Light 8500AAX COM Ground signal to EAGLE Remote Light 8500AAX Alarm Alarm signal to EAGLE Remote Light 8500AAX
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 39 / 45
GE Medical Systems Information Technologies
6.3 RAB Power Supply (RAB RL only) Connector
Figure 6.5 RAB RL Power Supply Connector
3-pin male connector for the 12VDC Medical Power supply 2000300-001 Pinning of the 3-pin Power connector:
PIN number
1 +12VDC Power Input 2 GND Power Input 3 - - not connected
Table 6.3 RAB RL Power Supply Connector
Label Direction/ Power Description
Input & Power Data
VCC +12VDC from external medical power supply GND GND from external medical power supply
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 40 / 45
GE Medical Systems Information Technologies
6.4 Cabling RAB RL internal
Figure 6.6 RAB RL cabling
Wire 1 (red) is to be connected to REL2 and
Wire 2 (red/black) is to be connected according to the selected alarm mode:
REL3 when operating with power failure detection (default mode) or
REL1 when operating without power failure detection
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 41 / 45
GE Medical Systems Information Technologies
7 ASSEMBLY DRAWINGS
Introduction
Included in this section is a complete set of mechanical diagrams, Reference diagram and schematic diagrams.
Spare Part List - These list shows the complete parts for service use. Mechanical diagrams - These diagrams shows the mechanical assembly of RAB
and RAB RL Reference diagrams - These diagrams shows provide a reference and view of the
used components in the schematic. Schematic diagrams - These diagrams shows the electrical wired connections
between the used analog and digital electronics components.
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 42 / 45
GE Medical Systems Information Technologies
7.1 Spareparts
Electronic Parts
Part Description Detailed Part Description Part No.
Remote Alarm Box Remote Alarm Box RAB (RAB Box
including Printed Circuit Board) Remote Ala rm Box Remote Light Spare PCB. Remote Alarm Box
Power Supply Power Supply SW172 (12V) 2000300-001 Assembly Remote Light Assembly Remote Light for RAB RL Box 8500AAX Fuse Fuse for Relay circuit 3.15A Slow Blow 91208440
Remote Alarm Box Remote Light (RAB
RL Box including Printed Circuit Board)
Spare PCB. Remote Alarm Box for RAB
and RAB RL Alarm Boxes (High Pot
tested)
2005693-001 2005693-006 2007067-001
Mechanical Parts
Part Description Detailed Part Description Part No.
Optical Light Guide Optical Light Pipe for RAB, RAB RL 2005478-001 Cable clamp Cable clamp for RAB 91614100 Plastic nozzle Plastic nozzle for cable of RAB 42103273 Cable clamp Cable clamp for RAB RL 91618976 Bottom plate Bottom plate metal for RAB RL 2006575-001 Adapter plate Adapter plate metal for GCX mounting. 2006576-001 Holder for Power Supply Metal Holder for Power Supply SW172 2006574-001 Bracket Metal Bracket for Power Supply Holder 2006573-001
Cables and Connectors
Part Description Detailed Part Description Part No.
Adapter Autoport to M-Port Adapter Autoport to M-Port 2001973-001 Cable Assembly RJ45 5ft Cable Assembly RJ45 to bedsides 5ft 418335-001 Cable cord 2 wire Cable cord 2 wire (has to be ordered in
meter)
91906200
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 43 / 45
GE Medical Systems Information Technologies
The following Power Cords are only used for RAB/RL units
Detailed Part Description Part No.
PWR CORD 10A 8FT CONT EURO STR 401855-101 PWR CORD 10A 8FT BRITISH STR 401855-102 PWR CORD ITALIAN 10A 8FT STR 401855-103 PWR CORD ISRAELI 10A 8FT STR 401855-104 PWR CORD SWISS 10A 8FT STR 401855-107 PWR CORD INDIAN 10A 8FT STR 401855-108 PWR CORD DANISH 10A 8FT STR 401855-109 PWR CORD AUSTALIAN 10A 8FT STR 401855-110 PWR CORD 125V 6FT STR (United States) 80274-006
Operation and Service Manuals
Part Description Detailed Part Description Part No.
Operation Manual GER Operation Manual German language 2006406-001 Operation Manual ENG Operation Manual English language 2006406-002 Operation Manual FRE Operation Manual French language 2006406-003 Operation Manual ITA Operation Manual Italian language 2006406-004 Operation Manual SPA Operation Manual Spanish language 2006406-005 Operation Manual DUT Operation Manual Dutch language 2006406-006 Operation Manual SWE Operation Manual Swedish language 2006406-007
Operation Manual RUS Operation Manual Russian language 2006406-009 Operation Manual POL Operation Manual Polish language 2006406-011 Operation Manual JAP Operation Manual Japanese language 2006406-012 Operation Manual POR Operation Manual Portuguese language 2006406-013 Operation Manual NOR Operation Manual Norwegian language 2006406-014 Operation Manual DAN Operation Manual Danish language 2006406-015 Operation Manual CHI Operation Manual Chinese language 2006406-016 Operation Manual HUN Operation Manual Hungarian language 2006406-017
Service Manual Service Manual 2006406-010
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 44 / 45
GE Medical Systems Information Technologies
Following sheets are att ached:
2004577-001 SCHEM INTFC NURSE CALL 2004576-002 DOC REF DW G INTFC NURSE CALL 1 2004576-003 DOC REF DW G INTFC NURSE CALL 2
RAB :
2005693-002 DOC MECH DWG ZM01
RAB RL:
2005693-007 DOC MECH DWG ZM01 2005693-008 DOC SCHEM SZ01
Service Manual Remote Alarm Box V1.0 Revision A 2006406-010 page 45 / 45
Marquette Hellige GmbH
Not equipped in Version :
x1
2004576-001 x2 x3 x4 x5
Revision-No
ER24685 MR27364 ECO-066622
REVISIONS
Index
1 A B
Date / Name
03.11.00/WMW
20.12.00/TYR
16.03.01/TYR
A03
DRAWN APPROVED
ISSUED
D-79007 Freiburg
2004575-001
Date / Name
03.11.00/WMW
03.11.00/SNA
W.Waltersberger
2004577-001 SCHEMATIC
INTFC NURSE CALL
Sheet:
11
of
$USER_CAD_B/2004575/design
[2000, 7, 19, 4]
gemedicalsystems.com
European Headquarters GE Medical Systems Information Technologies GmbH Postfach 60 02 65 D-79032 Freiburg • Germany T el. +49 761 45 43 - 0 Fax +49 761 45 43 - 233
World Headquarters GE Medical Systems Information Technologies, Inc. 8200 West T ower A venue Milwaukee, WI 53223 • USA T el . +1 414 355 5000 Fax +1 414 355 3790
Asia Pacific GE Medical Systems Hong Kong Ltd. 11th Floor , The Lee Gardens 33 Hysan Avenue Causeway Bay Hong Kong Te l: +852.2100.6300 Fax: +852.2100.6292
Loading...