Refer to the User's Reference manual for step-by-step instructions. Read
each component's User's Reference manual before using this system. All
Warnings and Cautions are in the User's Reference Manual.
WARNING
Before using this system:
•Complete all of the preoperative tests.
•Test all other system components.
Introduction
Introduction
Welcome
Thank you for choosing the GE Healthcare CARESCAPE® R860. Our
goal is to provide you with the highest quality product and services
available. This ventilator features a user interface specifically designed to
streamline workflow while providing exceptional insight into patient
needs.
Measured data definitions
Patient monitoring views show patient data measured by the ventilator
and accessories.
Note
Some measured data can be viewed with different units. Set unit
preferences on the Configuration > Units menu. See "Configuring units"
in the "System configuration (Super User) and service" section.
Note
Some data is only available when an airway module with the required
capabilities, such as spirometry and metabolics, is installed and warmed
up.
.
Gases DataDefinitionUnit
FiO2The percentage of oxygen that the
ventilator delivers to the patient.
EtO2The percentage of oxygen exhaled,
measured at the end of expiration.
EtCO2The percentage of carbon dioxide exhaled,
measured at the end of expiration.
FI-ET O2The difference between inspiratory and
expiratory concentrations of oxygen.
.
Pulmonary Data DefinitionUnit
CThe compliance of the patient’s respiratory
system measured during the breath cycle.
CstatThe static compliance of the patient’s
respiratory system measured during an
inspiratory hold.
2065492-0011
%
%
%, kPa, or mmHg
%
ml/cmH2O, ml/
kPa, or ml/mbar
ml/cmH2O, ml/
kPa, or ml/mbar
Pulmonary Data DefinitionUnit
RawThe average inspiratory and expiratory
airway resistance measured during the
breath cycle.
PEEPe+iThe sum of extrinsic and intrinsic positive
end expiratory pressures.
Time ConstantThe time needed for the lungs to deflate by
cmH2O/l/s,
kPa/l/s, or
mbar/l/s
cmH2O, kPa, or
mbar
ms
a certain amount or a percentage of
volume.
•One Time Constant allows 63% of
volume to be exhaled.
•Two Time Constants allow for 86% of
volume to be exhaled.
•Three Time Constants allow for 95%
of volume to be exhaled.
•Four Time Contants allow for 98% of
volume to be exhaled.
Static PEEPiThe pressure above PEEPe that remains
in the patient’s lungs, measured at the end
cmH2O, kPa, or
mbar
of the expiratory phase during an
expiratory hold.
.
Mechanical/
DefinitionUnit
Spontaneous
Data
MVexp spontThe volume of gas the patient exhales per
l/min
minute with spontaneous breaths.
RR spontThe number of spontaneous breath cycles
/min
the patient completes per minute.
VTexp spontThe volume of gas the patient exhales with
ml
a spontaneous breath.
MVexp mechThe volume of gas the patient exhales per
l/min
minute with mechanical breaths.
RR mechThe number of mechanical breath cycles
/min
the patient completes per minute.
VTexp mechThe volume of gas the patient exhales with
ml
a mechanical breath.
22065492-001
Mechanical/
DefinitionUnit
Spontaneous
Data
RSBIThe rapid shallow breathing index is
/min/l
calculated by dividing the spontaneous
breath rate by the tidal volume, averaged
over one minute. The RSBI reflects the
frequency and depth of the patient’s breath
cycles. A high RSBI value indicates that
patient’s breath cycles are more frequent
and shallow. RSBI is calculated in
spontaneous breathing modes (CPAP/PS,
VS, NIV, and SBT).
.
Per Weight Data DefinitionUnit
WeightThe calculated ideal body weight for adult
kg
patients. The entered weight of the patient
for pediatric patients.
MVexp/kgThe volume of gas the patient exhales per
l/min/kg
minute per the patient’s ideal body weight.
VTexp/kgThe volume of gas the patient exhales in a
ml/kg
breath per the patient’s ideal body weight.
MVexp spont/kgThe volume of gas the patient exhales per
l/min/kg
minute with spontaneous breaths per the
patient’s ideal body weight.
VTexp spont/kgThe volume of gas the patient exhales in a
ml/kg
spontaneous breath per the patient’s ideal
body weight.
C/kgThe dynamic compliance of the patient’s
lungs per the patient’s calculated ideal
body weight.
VO2/kgThe volume of oxygen a patient inhales
ml/kPa/kg, ml/
cmH2O/kg, or ml/
mbar/kg
ml/min/kg
(consumes) per minute per the patient’s
set weight.
2065492-0013
Per Weight Data DefinitionUnit
VCO2/kgThe volume of carbon dioxide a patient
ml/min/kg
exhales (produces) per minute per the
patient’s set weight.
.
Metabolics Data DefinitionUnit
EEThe amount of energy the patient expends
kcal/d or kJ/d
per day in calories.
RQThe ratio between the amount of carbon
N/A
dioxide the patient produces and oxygen
the patient consumes.
VO2The volume of oxygen a patient inhales
ml/min
(consumes) per minute.
VCO2The volume of carbon dioxide a patient
ml/min
exhales (produces) per minute.
VO2/m2The volume of oxygen a patient inhales
ml/min/m2
(consumes) per minute per square meter
of body surface area.
VCO2/m2The volume of carbon dioxide a patient
ml/min/m2
exhales per minute per square meter of
body surface area.
.
Spirometry Data DefinitionUnit
PpeakThe highest pressure level measured
during the inspiratory phase.
PplatThe pressure level measured after the
inspiratory phase and before the expiratory
cmH2O, kPa, or
mbar
cmH2O, kPa, or
mbar
phase (during an inspiratory pause).
PmeanThe average pressure level measured
during the breath cycle.
PEEPeThe pressure on the patient’s airway at the
end of the expiratory phase.
PEEPiThe pressure that remains on the patient’s
airway at the end of the expiratory phase
cmH2O, kPa, or
mbar
cmH2O, kPa, or
mbar
cmH2O, kPa, or
mbar
due to incomplete expiration. PEEPi is
measured above PEEPe.
42065492-001
Spirometry Data DefinitionUnit
VTinspThe volume of gas the patient inhales per
ml
breath.
MVinspThe volume of gas the patient inhales per
l/min
minute.
VTexpThe volume of gas the patient exhales per
ml
breath.
MVexpThe volume of gas the patient exhales per
l/min
minute.
LeakThe percentage of volume leaked from the
%
patient circuit.
.
Timing DataDefinitionUnit
I:EThe ratio of inspiratory time to expiratory
N/A
time.
TinspThe duration of the inspiratory phase of the
s
breath cycle.
TexpThe duration of the expiratory phase of the
s
breath cycle.
RRThe number of breath cycles a patient
/min
completes per minute.
Cycle TimeThe sum of the duration of inspiratory and
s
expiratory phases.
2065492-0015
Navigation
Navigation
Note
Shared information section for adult, pediatric, and neonatal patient
types.
Ventilator display
The 15-inch touchscreen display provides audible and visual alarms,
integrated key pad, and a Trim Knob control. The display unit uses the
Panasonic CR2477/BN battery (1000 mAh and 3V). To select menu
options or settings, touch only one touch point at a time to make sure the
correct selection is made. Touch the setting or press the Trim Knob to
confirm settings.
The touchscreen allows swipe gestures to move from one workspace to
another workspace.
Do not use pencils, pens, or other objects to activate the touchscreen.
The touchscreen will not function properly if tape or paper is stuck to the
display surface.
WARNING
Liquids on the display may degrade the performance of the touchscreen.
If liquids come in contact with the display, lock the touchscreen and clean
the display. Unlock the touchscreen once the display has been cleaned to
resume use of the touchscreen.
CAUTION
Do not apply excessive force to the touchscreen as damage may occur.
6
2065492-001
AB.100.016
713
2
4
Figure 1 • Display controls and indicators
.
1. Alarm lightThe integrated alarm light provides a visual alarm
when an alarm condition occurs. The alarm light also
provides a visual indicator when Audio Pause is
active and alarm audio is silenced.
2. Trim knob controlTurn the Trim Knob clockwise or counterclockwise to
change a setting. Press the Trim Knob to confirm a
setting.
3. Hard keys (key pad)Press the Audio Pause, Increase O2, Snapshot, Lock/
Unlock, or Home hard keys to access the associated
features.
•Audio Pause: Press to silence alarms for 120
seconds.
2065492-0017
4. LED indicatorThe green LED illuminates when the ventilator is
connected to the main power supply. The internal
battery is charging when the LED is lit.
Display user interface
The user interface incorporates the Menu, Current Patient menu, alarm
management, and Favorites procedures at the top of the display. The
patient status (airway pressure bar) and workspace/monitoring area are
located in the middle of the display. The navigation bar, message areas,
battery status, standby, and quick keys are located at the bottom of the
display.
712713
11
10
9
8
77
6
1
2
3
4
5
AB.100.010
82065492-001
Figure 2 • Display user interface components
.
1. FavoritesProvides short-cuts for up to four procedures (as
selected by the user). Use to select specific
procedures such as Increase O2, Inspiratory Hold,
Expiratory Hold, and Manual Breath. See "SettingFavorites" in the "Operation" section.
2. Patient statusThe airway pressure bar shows a dynamic view of
the patient airway pressure, Pmax, Ppeak, PEEP,
FiO2, and VTexp. Use the tab on the pressure bar
to collapse (hide) from view when available.
3. NavigationSelect an icon to open the corresponding view. See
"Navigating the user interface" for detailed
information.
4. Additional InformationShows current time and additional setting
information.
5. Main powerIndicates whether the ventilator is connected to the
main power supply or is running on battery. Also
shows battery status when running on battery.
6. StandbySelect the Standby quick key to go into Standby
(pause/stop ventilation). See "Standby" in the
"Operation" section.
7. Quick KeysSelect to change the corresponding ventilator
setting. Turn the Trim Knob to make a change.
Select the quick key or press the Trim Knob to
activate the change. When a quick key setting is
selected, a Trim Knob visual cue indicates the
change may be made by turning the Trim Knob and
pressing to confirm the setting.
8. Current Mode and Mode
Settings
9. General MessagesShows notices, procedure status, and system
10.MonitoringThis area is used to view waveforms, measured
Shows the active ventilation mode. Select to
access ventilation modes, and change mode
settings.
status information to the user. See "Generalmessages" in the "Alarms and troubleshooting"
section.
data, and settings.
2065492-0019
11.MenuSelect to quickly access options such as: System
menu, Procedures, Lung Mechanics, Suction, and
Nebulizer. See "Main menu" in the "Navigation"
section.
12.Current Patient menuSelect to enter the Current Patient menu. This
menu allows entry of the patient ID using an alphanumeric keyboard. Entered values for patient
gender, height, and weight are used to calculate
BSA (body surface area), and IBW (ideal body
weight). This menu also allows the selection of tube
type and diameter. See"New Patient" and "CurrentPatient" in the "Operation" section.
13.Alarm managementSelect to view alarms, alarm history, alarm setup,
and alarm help. See "Alarms and troubleshooting"
section.
Navigating active alarms
When an alarm occurs for measured data, the number and alarm limits
are shown with a border around them. The color of the border and the
alarm limit shows the priority of the alarm. Select within the border of the
active alarm to open the Alarm Setup menu. Select the alarm limit that
needs adjustment, then use the Trim Knob to adjust the setting and
confirm changes. See "Alarm management" in the "Alarms andtroubleshooting" section for additional information.
AB.100.187
Figure 3 • Select inside of the border to open the Alarm Setup menu.
Standby
Standby is displayed upon system startup or when the Standby quick key
is selected. When the system is in Standby, the Standby quick key and
the patient status (airway pressure) bar are colored tan. A “Standby”
10
2065492-001
message is displayed in the navigation bar when in the Present/Patient
4
1
2
5
9
7
68
3
AB.100.186
Status workspace. Standby is used to stop ventilation to the patient,
select a New or Previous patient, perform a System Check, and Park/
Unpark the patient circuit. The Setup button accesses the password
protected Configuration (Super User) and Service menus.
Figure 4 • Standby menu
.
1. New PatientSelect New Patient to enter patient information.
2. Circuit SetupSelect Circuit Setup to select HME or Humidifier for
adult and pediatric patient types.
3. StandbyWhen Standby (hand icon) is selected the Standby
menu displays. If the Patient detected alarm
occurs, the Standby menu automatically displays.
4. Park/Unpark CircuitWhen the circuit is parked a message displays:
Patient circuit is occluded and ventilator is in
Standby.
5. Start VentilationSelect to start patient ventilation.
6. System CheckSelect System Check to perform a ventilator
system check.
7. InformationSelect to access information regarding the system
check status and troubleshooting.
2065492-00111
8. Previous Patient/Current
AB.100.136
1
2
3
4
5
Patient
Select Previous Patient to use the previous
patient’s ventilator settings and patient information.
Select Current Patient to open the Current Patient
menu and use the current patient’s ventilator
settings and patient information.
9. SetupSelect to access the Configuration (Super User)
and Service menus. A password is required to
enter these menus. Contact a training
representative to obtain the password.
Main menu
Select Menu to quickly access ventilator features and options.
Figure 5 • The main Menu accesses the System menu, Procedures, Lung
Mechanics, Nebulizer, and Suction options.
.
1. SystemUse the System menu to access data source,
module type and version, calibrations (Paux Zero
and Purge Flow), and display brightness. The
System menu shows the software version, running
hours, altitude, O2 supply pressure, air supply
pressure, and battery status. See "System menu"
located in this section.
122065492-001
2. ProceduresUse the Procedures menu to access the Assign
Favorites menu and the following procedures:
Manual Breath, Increase O2, Inspiratory Hold,
Expiratory Hold, and Auto PEEP. See "SettingFavorites" in the "Operation" section.
3. Lung MechanicsUse the Lung Mechanics menu to access the
Assign Favorites menu and the following
procedures: P0.1, Negative Inspiratory Force (NIF),
and Vital Capacity. See "Setting Favorites" in the
"Operation" section.
4. NebulizerUse the Nebulizer menu to access the Aerogen
and Pneumatic Nebulizer procedures. See "SettingFavorites" section in the "Operation" section.
5. SuctionUse the Suction menu to access the Assign
Favorites menu and the Suction procedure. See
"Setting Favorites" in the "Operation" section.
System menu
The System menu contains settings for data source selection, calibration
options, display brightness, and system information.
1.Select Menu > System.
The Airway Module type and software version number are shown
under data source.
2.Select Data Source (Ventilator or Airway Module).
•For Neonatal; select Ventilator or NFS. See "System menu" in
the "Neonatal Operation" section.
3.Select Calibrations (Airway Module, Paux Zero, or Purge Flow).
•Select Airway Module to calibrate the airway module.
•Select Paux Zero. A green check mark indicates Paux Zeroing
calibration was successful.
•Select Purge Flow. The Purge Flow check box may be checked
or unchecked when performing a Paux Zero. Continuous purge
flow will come from the Paux outlet when the Purge Flow check
box is selected. A white check mark indicates Purge Flow is
active.
Note
See "Purging the auxiliary pressure tubing" and "Zeroing
auxiliary pressure" in the "Setup and connections" section.
--
4.Select Display Brightness to adjust the brightness level of the user
interface.
2065492-001
13
Select brightness level of 1 (low) to 5 (high).
5.View system information: software version, service packet version,
running hours, altitude, O2 supply pressure, air supply pressure, and
battery status.
Changing a setting
1.Touch the setting.
2.Change the value by turning the Trim Knob or selecting a menu item.
3.Touch the setting or push the Trim Knob to confirm the setting.
Figure 6 • The Trim Knob graphic is used to indicate that the use of the Trim Knob
is necessary to change or confirm a setting
Note
To cancel or back out of a setting change, select X in the lower right
corner of the menu, touch outside of the setting twice, select the
Home hard key, or wait for the selection to time out. For example,
ventilation and alarm setting changes can be cleared by selecting the
Home hard key prior to confirming a setting.
--
Navigating the user interface
The ventilator user interface uses three different workspaces: Past/
Historical trends, Present/Patient status, and Future/Clinical decision
support. Each workspace (rectangle icon) contains views (circle icons)
that contain different configurations of data and functions.
When a workspace is selected, the correlating view icons are displayed.
•Use a swipe gesture or touch a workspace icon to go to a new
workspace (swipe gesture: touch display and move finger tip left or
right).
•When you navigate away from a workspace and then navigate back,
the display will show the last view that was displayed from the
workspace.
•If a view is not supported by the current patient type or software is
not installed, it will not display.
14
2065492-001
AB.100.185
Figure 7 • Navigation example; select a workspace (rectangle) to see correlating
views (circle).
Present/Patient Status workspace and views
The Present/Patient Status workspace shows the following views: Basic,
Basic Waveform, Advanced Waveform, Splitscreen, and Charting. This
workspace allows the user to choose the view in which they would like to
see patient data displayed. See "Measured data definitions" in the
"Patient monitoring" section for information on the numerics displayed in
the Present views. See "Neonatal measured data definitions" in the
"Neonatal patient monitoring" section for information on the numerics
displayed in the Present views for a neonatal patient type.
•Touch the icon to display the corresponding view.
•Use a swipe gesture to view and move to Past (Historical trends) and
Future (Clinical decision support) workspaces.
Present/Patient Status Workspace
2065492-00115
Basic View
AB.100.129
Use the Basic view to see measured data in a large format that
can be easily viewed from a distance. Note: The patient status
(airway pressure) bar is permanently displayed to easily view
patient airway pressure, tidal volume, and FiO2.
Basic Waveform View
AB.100.119
Use the Basic Waveform view to see patient waveforms and
measured data. Note: the airway pressure bar may be collapsed
to expand the monitoring area when the Paw and Flow
waveforms are displayed.
162065492-001
Advanced Waveform View
AB.100.120
AB.100.121
Use the Advanced Waveform view to see additional measured
data associated with the patient waveforms. Note: The airway
pressure bar may be collapsed to expand the monitoring area
when the Paw and Flow waveforms are displayed.
Splitscreen View
Use the Splitscreen view to see spirometry, measured data, and
waveforms. Select the upper right corner of the spirometry
waveform to change settings. Note: The airway pressure bar may
be collapsed to expand the monitoring area when the Paw and
Flow waveforms are displayed.
2065492-00117
Charting View
AB.100.122
AB.100.123
Use the Charting view to see a complete list of patient data.The
airway pressure bar is permanently displayed to easily view
patient airway and pressure settings, tidal volume and FiO2.
Past/Historical Trends workspace and views
The Past/Historical trends workspace shows information for the following
views: Graphical trends, Numerical trends, Trends log, and Snapshot
trends.
•Touch the icon to display the corresponding view.
•Use a swipe gesture to view and move to Present/Patient status or
Future/Clinical Decision Support workspaces.
Past/Historical Trends Workspace
Graphical Trends View
182065492-001
Graphical Trends View
AB.100.124
AB.100.125
Use the Graphical trends view to review historical waveforms and
patient trends. See "Graphical trends view" and "Graphical trendsview - Neonatal" in the Patient Monitoring section.
Numerical Trends View
Use the Numerical trends view to review patient ventilation
modes and settings, measured data, and alarm settings. See
"Numerical trends view" and "Numerical trends - Neonatal" in the
Patient Monitoring section.
Trends Log View
Use the Trends Log to review patient alarms and settings, and
events that have occurred during ventilation. See "Trends logview" and "Trends log view - Neonatal" in the Patient Monitoring
section.
2065492-00119
Snapshot Trends View
AB.100.126
AB.100.127
Use Snapshot trends to view saved patient data. See "Snapshottrends view" and "Snapshot trends view - Neonatal" in the Patient
Monitoring section for more information.
Future/Clinical Decision Support workspace and views
The Future/Clinical Decision Support workspace shows the following
views (if software is installed): SBT, FRC, Spirometry, Metabolics, and
Calculations.
•Touch the view icon to display the corresponding view.
•Use a swipe gesture to move to Present/Patient Status or Past/
HistoricalTrends workspaces and associated views.
Future/Clinical Decision Support
SBT View
202065492-001
SBT View
AB.100.128
AB.100.130
Use the SBT view to evaluate spontaneous breathing trial data.
See "SBT view" and "SBT view - Neonatal" in the Clinical
Decision Support section.
FRC View
Use the FRC view to evaluate and review patient respiratory
data. The FRC view includes three tabs: Evaluate, FRC INview
(FRC procedure), and PEEP INview (PEEP INview procedure).
See "FRC INview procedures" in the Clinical Decision Support
section.
Spirometry View
Use the Spirometry view to evaluate and review graphs and data
from spirometry and spirodynamic measurements. The
Spirometry view includes the Spirometry tab and SpiroDynamics
tab. See "Spirometry view" and "Spirometry view - Neonatal" in
the Clinical Decision Support section.
2065492-00121
Metabolics View
AB.100.131
AB.100.132
Use the Metabolics view to evaluate and review Metabolics
measurements. See "Metabolics view" in the Clinical Decision
Support section.
Calculations View
Use the Calculations view to calculate and review data based on
the ventilator, measured data, and laboratory blood gas analysis
data. See "Calculations view" in the Clinical decision support
section.
222065492-001
Setup and connections
EXP
AB.100.088
Menu
Adult
Current Mode
FiO2
30
%
Pinsp
33
cmH2O
Rate
10
/min
Tinsp
1.7
s
PEEP
5
cmH2O
PS
5
cmH2O
Standby
Insp
Hold
Exp
Hold
Manual
Breath
14:38
FiO2
No Alarms
Airway
Pressure
cmH2O
60
40
20
0
-5
Pmax
Ppeak
PEEP
VTexp
508
30
%
40
22
PEEPe
5
cmH2O
10
3
Peak Pressure
40
cmH2O
50
10
Minute Volume
6.2
l/min
12
4
Tidal Volume
508
ml
600
400
Respiratory Rat e
14
/min
20
5
1
2
4
5
6
7
8
3
9
11
12
17
10
13
1415
16
18
1
Setup and connections
Note
Shared information section for adult, pediatric, and neonatal patient
types.
6.VGA (not for clinical use)19.Patient circuit support arm
7.Module bay connection20.Port 1 (neonatal flow sensor
8.Main power inlet and fuse holder21.Port 2 (not supported)
9.Power switch22.Port 3 (exhalation valve heater
10.Equipotential stud23.Port 6 (RS232 Serial
11.Module bay mounting
thumbscrews
12.Oxygen supply connection
(pipeline)
13.O2 high-pressure inlet filter
(optional)
15.Air supply connection (pipeline)
17.Ventilator unit fan filter
connection)
connection)
communication port)
24.Port 5 (RS232 Serial
communication port)
25.Display unit fan filter
Connecting the breathing circuit
WARNING
Do not use antistatic or electrically conductive breathing tubes or masks.
Check all connections to the breathing circuit to make sure that there are
no unintended connections made to other equipment, especially
equipment that delivers fluids, as the patient could be harmed.
The inspiratory safety guard is required to connect the breathing circuits
to the ventilator. The inspiratory safety guard must be used at all times
during ventilation.
Note
See "Cleaning and maintenance" for information on the replacement of
the inspiratory safety guard. See "Parts and accessories" for ordering
information.
The exhalation valve heater should be used when an active humidifier
with a heated expiratory limb is used.
Important
Consult your hospital guidelines for proper use of expiratory filters in
conjunction with heated humidifiers.
2065492-001
25
Connecting a HME (heat and moisture exchanger)
Note
To prevent excessive resistance in the breathing circuit, the HMEF500
should not be used for Adult patients.
Note
If using optional accessories see Figure in "Connecting the Pedi-lite(+)
and D-lite(+) sensors".
1.Connect the inspiratory safety guard to the inspiratory port.
2.Attach the inspiratory limb of the patient circuit to the inspiratory
safety guard.
3.Attach the expiratory limb of the patient circuit to the expiratory port
or expiratory filter (if used).
4.Connect the Pedi-lite(+) or D-lite(+) sensor to the patient wye (if
used). Use a 5 ml (minimum) spacer and elbow when using the Pedilite(+) or D-lite(+) sensor.
5.Connect the HME.
•Place the HME between the SpiroDynamics catheter (if used),
but after the Pedi-lite(+) and D-lite(+) sensor (if used).
•The HME should be removed when a nebulizer is active.
Replace the HME when the nebulizer is not in use.
6.Connect the circuit elbow to the HME (if used).
Note
To disconnect, follow instructions in reverse order.
--
26
2065492-001
Adult
Menu
FiO2
30
Minute Volume
6.2
T
o
t
a
S
p
o
n
t
Current Mode
INSP
1
6
2
5
Figure 10 • Overview of setup and connections with a HME
1.Inspiratory safety guard
2.Inspiratory limb
3.Expiratory port/expiratory filter if used
4.Expiratory limb
5.Patient wye
6.HME
No Alarms
Insp
Exp
Manual
Hold
Hold
Breath
14:38
Airway
Pressure
PEEPe
Peak Pressure
cmH2O
cmH2O
cmH2O
%
10
50
40
22
l/min
12
4
l
FiO2
30
%
508
60
3
10
40
5
Tidal Volume
Pinsp
33
cmH2O
Pmax
Ppeak
40
Respiratory Rate
20
ml
/min
PEEP
600
20
0
400
5
14
-5
VTexp
T
o
t
a
l
S
p
o
508
n
t
Rate
PEEP
PS
Tinsp
Standby
10
5
5
1.7
/min
cmH2O
cmH2O
s
EXP
3
4
AB.100.188
Connecting the humidifier
The ventilator is designed to work with active humidification. GE
Healthcare recommends the use of the Fisher & Paykel MR850
humidifier (refer to humidifier instructions for detailed information on
humidifier connections and use).
WARNING
Never position any filter in the inspiratory limb downstream of a
humidifier.
When adding attachments or other components to the ventilator, the
pressure gradient across the breathing circuit may change.
1.Slide the humidifier heater onto the accessory rail (do not plug in).
2065492-001
27
Loading...
+ 66 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.