GE PACE5000, PACE6000, PACE1000 Calibration Manual

Page 1
GE
Digital Solutions
Pressure Automated Calibration Equipment
Calibration manual - K0450 Revision B
PACE5000
PACE6000
PACE1000
Page 2
© 2010 General Electric Company. All Rights Reserved. Specifications are subject to change without notice. GE is a registered trademark of General Electric Company. Other company or product names mentioned in this document may be trademarks or registered trademarks of their respective companies, which are not affiliated with GE.
Page 3
PACE Series Calibration Manual
Introduction
This technical manual provides calibration instructions for the PACE Pressure Controllers and Indicators.
The features shown and described in this manual may not be available on some models.
Safety
The manufacturer has designed this equipment to be safe when operated using the procedures detailed in this manual. Do not use this equipment for any other purpose than that stated.
This publication contains operating and safety instructions that must be followed to ensure safe operation and to maintain the equipment in a safe condition. The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage.
Use suitably qualified* calibration technicians and good engineering practice for all procedures in this publication.
Pressure
Do not apply pressures greater than the maximum working pressure to the equipment. It is the responsibility of the calibration technician to apply pressures within the published pressure range and to only use external pressure equipment with correctly rated fittings and components.
Toxic Materials
There are no known toxic materials used in construction of this equipment.
Maintenance
The equipment must be correctly maintained, the manufacturer’s procedures should be carried out by authorized service agents or the manufacturer’s service departments.
Technical Advice
For technical advice contact the manufacturer.
* A qualified calibration technician must have the necessary technical knowledge, documentation, special calibration/test equipment and tools to carry out the required work on this equipment.
[EN] English i K0450 Revision B
Page 4
Abbreviations
The following abbreviations are used in this manual; the abbreviations are the same in the singular and plural.
abs Absolute a.c. Alternating current ALT Altitude BSP British pipe thread CAS Calculated airspeed CSK Countersunk d.c. Direct current DPI Digital Pressure Instrument e.g. For example etc. And so on Fig. Figure ft Foot g Gauge Hg Mercury HTS High tensile steel Hz Hertz IAS Indicated airspeed i.e. That is IEC International Electrotechnical Commission IEEE 488 Institute of Electrical and Electronic Engineers standard 488 data in Inch kg Kilogram kts/kn knot LCD Liquid crystal display m Metre mA Milliampere max Maximum mbar Millibar min Minute or minimum mm Millimetre mV Millivolts MWP Maximum working pressure No. Number NPT National Pipe Thread PACE Pressure Automated Calibration Equipment Para. Paragraph PDCR Pressure transducer PED Pressure Equipment Directive psi Pounds per square inch PTX Pressure transmitter ROC Rate of Climb RS232 Serial communications data standard RtCAS Rate of calculated airspeed
K0450 Revision B ii [EN] English
Page 5
RtMach Rate of Mach SCPI Standard Commands for Programmable Instruments UUT Unit under test V Volts VFC Volts-free contact +ve Positive
-ve Negative °C Degrees Celsius
Associated Publications
K0447 PACE5000/6000 User Guide and Safety Instructions K0467 PACE1000 Indicator User Guide and Safety Instructions K0443 PACE5000/6000 Controller User Manual K0470 PACE1000 Indicator User Manual K0476 Pressure Control Module User Guide and Safety Instructions K0472 SCPI Communications Manual K0469 Heritage Communications Manual - Instrument Emulation
Symbols
PACE Series Calibration Manual
This equipment meets the requirements of all relevant European safety directives. The equipment carries the CE mark.
This symbol, on the equipment, indicates that the user should read the user manual.
This symbol, on the equipment, indicates a warning and that the user should refer to the user manual. Ce symbole, sur l’instrument, indique que l’utilisateur doit consulter le manuel d’utilisation. Ce symbole, dans le manuel, indique une situation dangereuse.
This symbol warns the user of the danger of electric shock. Ce symbole alerte l’utilisateur sur le danger de choc électrique.
Do not dispose of this product as household waste. Use an approved organisation that collects and/or recycles waste electrical and electronic equipment. For more information, contact one of these:
- Our customer service department: www.gemeasurement.com
- Your local government office.
[EN] English iii K0450 Revision B
Page 6
WARNINGS TURN OFF THE SOURCE PRESSURE(S) AND CAREFULLY VENT THE PRESSURE LINES BEFORE
DISCONNECTING OR CONNECTING THE PRESSURE LINES. PROCEED WITH CARE. ONLY USE EQUIPMENT WITH THE CORRECT PRESSURE RATING. BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE.
REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT. DO NOT USE ANY DAMAGED FITTINGS AND EQUIPMENT.
DO NOT EXCEED THE MAXIMUM WORKING PRESSURE OF THE INSTRUMENT. THIS EQUIPMENT IS NOT RATED FOR OXYGEN USE.
THE GROUND LEAD OF THE INSTRUMENT MUST BE CONNECTED TO THE AC SUPPLY PROTECTIVE SAFETY GROUND.
ISOLATE THE POWER SUPPLY BEFORE MAKING ANY ELECTRICAL CONNECTIONS TO THE REAR PANEL.
K0450 Revision B iv [EN] English
Page 7
Pressure Units and Conversion Factors
Pressure units Factor (hPa) Pressure units Factor (hPa)
mbar 1.0
bar 1000.0
cmH
mH
O @ 20°C
2
O @ 20°C
2
PACE Series Calibration Manual
0.978903642
97.8903642
Pa (N/m
2
)
0.01
hPa 1.0
kg/m
kg/cm
2
2
0.0980665
980.665
kPa 10.0 torr 1.333223684
MPa 10000.0 atm 1013.25
mmHg @ 0°C 1.333223874 psi 68.94757293
cmHg @ 0°C 13.33223874
mHg @ 0°C 1333.223874
inHg @ 0°C 33.86388640341
mmH
cmH
mH
mmH
O @ 4°C
2
O @ 4°C
2
O @ 4°C
2
O @ 20°C
2
0.0980665
0.980665
98.0665
0.097890364
2
lb/ft
O @ 4°C
inH
2
inH
O @ 20°C
2
inH20 @ 60°F
ftH2O @ 4°C
ftH2O @ 20°C
ftH20 @ 60°F
0.4788025898
2.4908891
2.486413
2.487641558
29.8906692
29.836983
29.8516987
Unit Conversion
To convert FROM pressure VALUE 1 in pressure UNITS 1 TO pressure VALUE 2 in pressure UNITS 2, calculate as follows:
VALUE 2 VALUE 1
FACTOR 1
=
---------------------­FACTOR 2
Note: The PACE instrument contains selectable pressure units and user defined units. Use the
conversion factors to calculate a user defined unit from the table above. Refer to the data sheets for the list of selectable pressure units.
[EN] English v K0450 Revision B
Page 8
Table of Contents
1 Introduction......................................................................................................................................................1
2 Calibration Status..........................................................................................................................................1
3 Calibration Equipment ................................................................................................................................1
4 Preliminary Operations...............................................................................................................................1
5 Notes on Calibration ....................................................................................................................................1
5.1 PACE Controller Connection .........................................................................................................2
Gauge Reference...............................................................................................................................2
5.2 PACE Indicator Connection...........................................................................................................3
Gauge Reference...............................................................................................................................3
6 Calibration Check (All Ranges).................................................................................................................4
6.1 Procedure..............................................................................................................................................4
7 Calibration Adjustment...............................................................................................................................5
Appendix A PACE Menus and Screens..................................................................................................6
A.1 PACE Controller Menus ...................................................................................................................6
A.2 PACE Controller Screens.................................................................................................................7
A.3 PACE Indicator Menus.....................................................................................................................8
A.4 PACE Indicator Screens ..................................................................................................................9
K0450 Revision B vi [EN] English
Page 9
PACE Series Calibration Manual

1 Introduction

The PACE controller incorporates a calibration facility; for the PACE to stay accurate, a calibration check should be carried out at chosen intervals. If the accuracy of the PACE is not within the permissible deviation, carry out a calibration adjustment.

2 Calibration Status

Using the Measured Pressure/Instrument Status menu, the calibration status of the calibrator can be displayed on the front panel screen. The status menu includes Calibration History which gives a list of dates of the stored calibration corrections.
Note: The Date and Time must be set correctly using the Measured Pressure/Global Set-up/
Calibration menu.

3 Calibration Equipment

The original GE Calibration Certificate shows the measurement uncertainty of the original calibration standard. To preserve uncertainty of the PACE calibration, checks and adjustments must be performed using a calibrator uncertainty of less than or equal to the original calibration standard.

4 Preliminary Operations

Review and become familiar with the whole procedure before beginning a calibration process. Allow at least one hour for the PACE to thermally stabilize in a thermally stable environment after switching on and before calibration.
Before starting a calibration procedure:
Carry out a leak test as detailed in PACE user manual K0443.

5 Notes on Calibration

The pressure standard output port and the reference level must be at the correct level or use height­corrected applied pressure.
To prevent applied calibration pressure “back feed” in PACE controllers, fit blanking plugs to both positive and negative supply ports on the manifold.
Set the PACE units of pressure to one of the required units for calibration.
Reference
Level
[EN] English 1 K0450 Revision B
Page 10

5.1 PACE Controller Connection

WARNING ON COMPLETION OF CALIBRATION, OPEN THE ON/OFF VALVE TO VENT TRAPPED
PRESSURE FROM THE SUPPLY + PORT TO ATMOSPHERE.
Connect the calibrator for each pressure range as follows:
Connect the output of the pressure standard to the PACE module output port.
Fit an on/off valve to the supply + input of the PACE controller. Leave the other side of the valve open to atmosphere. Turn the valve to the closed position.

Gauge Reference

If the pressure standard has a reference connection, then connect this to the PACE reference port on the module manifold. Otherwise the calibrator reference port should be open to atmosphere.
PACE Controller
Module 2
SUPPLY
+-
OUTLET VENT REF
On/off valve fitted to supply +
SUPPLY
+-
Rear view of instrument
Module 1
OUTLET VENT REF
Pressure Standard
Note: Module 2 reference port open to atmosphere or to reference port of the pressure
standard.
Note: Optimum performance is achieved by connecting the PACE reference port to the
pressure standard.
K0450 Revision B 2 [EN] English
Page 11
PACE Series Calibration Manual

5.2 PACE Indicator Connection

Connect the indicator for each pressure range as follows:
Connect the output of the pressure standard to the indicator port.

Gauge Reference

If the pressure standard has a reference connection, then connect this to the indicator reference port. Otherwise the indicator reference port should be open to atmosphere.
PACE Indicator
SENSOR
PORT
REFERENCE
PORT
Pressure Standard
Rear view of instrument
Note: Indicator reference port open to atmosphere or to reference port of the pressure
standard.
Note: Optimum performance is achieved by connecting the PACE reference port to the
pressure standard.
[EN] English 3 K0450 Revision B
Page 12

6 Calibration Check (All Ranges)

Note: The PACE adds the barometric reading to a gauge range to produce an absolute range.

6.1 Procedure

Set the calibrator to measure mode:
1 Connect calibration standard for the pressure range to be checked.
2 Press Task and select Basic.
3 With the pressure standard connected to the correct pressure port, select Measured
Pressure and press Range to select the gauge pressure range to be checked.
4 Barometric pressure can be displayed in the status area.
5 Set maximum displayed resolution.
Gauge ranges should be zeroed immediately prior to a calibration check, as follows:
1 Press Measured Pressure/Zero to zero the selected range.
2 On completion of the zero operation, the display shows Zero completed successfully.
3 Adjust calibration pressure to the first pressure value and wait until this pressure, as
displayed on PACE, is stable to less than 10ppm (0.001%). Display filtering may be required.
4 Compare the pressure value on the calibration standard to the value displayed and
record the difference.
5 Repeat (3) and (5) for each pressure, as prompted by the PACE.
6 If the recorded difference exceeds the permissible deviation for the selected range, the
calibrator requires a calibration adjustment for that range. Refer to PACE sales data sheets for permissible precision deviation. Precision specification is used if less than 24 hours since adjustment using the same calibration equipment. If greater than 24 hours, then the PACE specification is the sum of precision and long term stability.
7 Select the next pressure range for a calibration check.
8 After completing all calibration checks, adjust calibration standard to atmospheric
pressure.
9 Disconnect calibration standard from the output.
If no further calibration is required, switch off the PACE.
K0450 Revision B 4 [EN] English
Page 13
PACE Series Calibration Manual

7 Calibration Adjustment

To adjust a calibration range of the calibrator, proceed as follows:
1 Connect the calibrator for the range to be adjusted, as detailed in Calibration Check.
Note: The calibration adjustments may be carried out in any order.
2 Select Measured Pressure/Global Set-up/Calibration, enter the Calibration PIN
(4321).
3 Select Sensor Correction.
4 Select the pressure range to be corrected.
5 Select New Calibration.
6 The display shows the first value to be set on the pressure standard and to press OK
when the applied pressure is stable. Use the numeric keys to enter the precise applied pressure.
Note: The display also shows throughout this procedure the message Calibrating and the
selected pressure range.
7 Select Accept to store the first value and the display goes to the next pressure value to
be set.
8 Select Repeat to re-apply the same pressure and Quit Calibration to exit the
calibration of this pressure range.
9 Repeat steps (5) to (7) for the next value.
10 Carry out a calibration check to verify this procedure.
11 After completing the calibration procedures, adjust the calibration standard to
atmospheric pressure. Disconnect calibration standard from the PACE.
[EN] English 5 K0450 Revision B
Page 14

Appendix A PACE Menus and Screens

A.1 PACE Controller Menus

Calibration procedure next page
K0450 Revision B 6 [EN] English
Page 15

A.2 PACE Controller Screens

PACE Series Calibration Manual
[EN] English 7 K0450 Revision B
Page 16

A.3 PACE Indicator Menus

Calibration procedure next page
K0450 Revision B 8 [EN] English
Page 17

A.4 PACE Indicator Screens

PACE Series Calibration Manual
[EN] English 9 K0450 Revision B
Page 18
Page 19
Page 20
www.gemeasurement.com
Loading...