Page 1

GEHealthcare
MUSE™v9
CardiologyInformationSystem
DevicesandInterfacesInstructionManual
2059568-013J
MUSEv9CardiologyInformationSystem
English
©2015-2019GeneralElectricCompany.
AllRightsReserved.
Page 2

PublicationInformation
TheinformationinthisdocumentappliesonlytoMUSECardiologyInformationSystemVersion9.0.Itdoesnotapplytoearlierproduct
versions.Duetocontinuingproductinnovation,specicationsinthisdocumentaresubjecttochangewithoutnotice.
MUSE,MARS,CASE,andMACCRAaretrademarksownedbyGEMedicalSystemsInformationTechnologies,Inc.,aGeneralElectricCompany
goingtomarketasGEHealthcare.Allothertrademarkscontainedhereinarethepropertyoftheirrespectiveowners.
Thedocumentpartnumberandrevisionareoneachpageofthedocument.Therevisionidentiesthedocument’supdatelevel.Therevision
historyofthisdocumentissummarizedinthefollowingtable.
Revision
A
B
C
Date
18March2015Initialrelease.
17July2015Customerrelease.
1September2015ChangemadetosupportHCSDM00336108.
Comments
D
E
F
G30June2017
H
J
ToaccessotherGEHealthcareDiagnosticCardiologydocuments,gototheCommonDocumentationLibrary(CDL),locatedat
www.gehealthcare.com/documents
ToaccessOriginalEquipmentManufacturer(OEM)documents,gotothedevicemanufacturer'swebsite.
ThisdocumentdescribestheMUSECardiologyInformationSystem,alsoreferredtoasthe”product”,“system”,or“device”.Thisdocumentis
intendedtobeusedbyqualiedGEHealthcareserviceengineersandthird-partyserviceengineersauthorizedbyGEHealthcare.
NOTE:
DonotattempttoinstalltheMUSEsystemdevicesandinterfacesifyouarenotpartoftheintendedaudienceorhavenotreadand
understoodtheseinstructionsintheirentirety.
16October2015UpdatedtoincludeinformationfortheMUSEv9DICOM.
23May2016
1December2016UpdatestosupportHCSDM00433985.
5October2018UpdatedtosupportHCSDM00528040.
13March2019UpdatedtosupportHCSDM00547818
,andclickCardiology.
UpdatestosupportHCSDM00401589.
UpdatedtheConguretheSystemDICOMModalityWorklistClientService
section.
UpdatedtosupportHCSDM00436636.
TheMUSECardiologyInformationSystemisintendedtobeusedunderthedirectsupervisionofalicensedhealthcarepractitioner,by
trainedoperatorsinahospitalorfacilityprovidingpatientcare.
Thisdocumentprovidesinformationrequiredfortheproperuseofthesystem.Familiarizeyourselfwiththisinformation,andreadand
understandallinstructionsbeforeattemptingtousethissystem.Keepthisdocumentwiththeequipmentatalltimesandperiodically
reviewit.
NOTE:
Allillustrationsinthisdocumentareprovidedasexamplesonly.Dependingonsystemconguration,screensinthedocumentmay
differfromthescreensonyoursystem.
Allpatientnamesanddataarectitious.Anysimilaritytoactualpersonsiscoincidental.
2MUSE™v92059568-013J
13March2019
Page 3

ServiceManualLanguageInformation
WARNING
(EN)
ПРЕДУПРЕЖДЕНИЕ
(BG)
警告
(ZH-CN)
警告
(ZH-TW)
UPOZORENJE
(HR)
VAROVÁNÍ
(CS)
ThisservicemanualisavailableinEnglishonly.
•Ifacustomer'sserviceproviderrequiresalanguageotherthanEnglish,itisthecustomer's
responsibilitytoprovidetranslationservices.
•Donotattempttoservicetheequipmentunlessthisservicemanualhasbeenconsulted
andisunderstood.
•Failuretoheedthiswarningmayresultininjurytotheserviceprovider,operator ,orpatient,
fromelectricshock,mechanicalorotherhazards.
Товаупътванезаработаеналичносамонаанглийскиезик.
•Акодоставчикътнауслугатанаклиентаизискадругезик,задължениенаклиентае
даосигурипревод.
•Неизползвайтеоборудването,предидастесеконсултиралииразбралиупътването
заработа.
•Неспазванетонатовапредупреждениеможедадоведедонараняваненадоставчика
науслугата,оператораилипациентврезултатнатоковударилимеханичнаили
другаопасност.
本维修手册仅提供英文版本。
•如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
•未详细阅读和完全理解本维修手册之前,不得进行维修。
•忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的伤害。
本維修手冊只提供英文版。
•如果客戶的維修人員有英語以外的其他語言版本需求,則由該客戶負責提供翻
譯服務。
•除非您已詳閱本維修手冊並了解其內容,否則切勿嘗試對本設備進行維修。
•不重視本警告可能導致維修人員、操作人員或病患因電擊、機械因素或其他因素
而受到傷害。
Oveuputezaservisiranjedostupnesusamonaengleskomjeziku.
•Ukolikokorisničkiserviszahtijevanekidrugijezik,korisnikovajeodgovornostosigurati
odgovarajućiprijevod.
•Nemojtepokušavatiservisiratiopremuukolikonistekonzultiraliirazumjelioveupute.
•Nepoštivanjeovogupozorenjamožerezultiratiozljedamaservisnogosoblja,korisnikaili
pacijentaprouzročenimelektričnimudaromtemehaničkimilinekimdrugimopasnostima.
Tentoprovoznínávodexistujepouzevanglickémjazyce.
•Vpřípadě,žeexterníslužbazákazníkůmpotřebujenávodvjinémjazyce,jezajištění
překladudoodpovídajícíhojazykaúkolemzákazníka.
•Nesnažteseoúdržbutohotozařízení,anižbystesipřečetlitentoprovoznínávoda
pochopilijehoobsah.
•Vpřípaděnedodržovánítétovarovánímůžedojítkporaněnípracovníkaprodejního
servisu,obslužnéhopersonálunebopacientůvlivemelektrickéhoproudu,respektive
vlivemmechanickýchčijinýchrizik.
2059568-013JMUSE™v9
3
Page 4

ServiceManualLanguageInformation(cont'd.)
ADVARSEL
(DA)
Denneservicemanualndeskunpåengelsk.
•Hvisenkundesteknikerharbrugforetandetsprogendengelsk,erdetkundensansvar
atsørgeforoversættelse.
•Forsøgikkeatservicereudstyretmedmindredenneservicemanualharværetkonsulteret
ogerforstået.
•Manglendeoverholdelseafdenneadvarselkanmedføreskadepågrundafelektrisk,
mekaniskellerandenfareforteknikeren,operatørenellerpatienten.
WAARSCHUWING
(NL)
DezeservicemanualisalleeninhetEngelsverkrijgbaar.
•Indienhetonderhoudspersoneeleenanderetaalnodigheeft,danisdeklant
verantwoordelijkvoordevertalingervan.
•Probeerdeapparatuurnietteonderhoudenvoordatdezeservicemanualgeraadpleegd
enbegrepenis.
•Indiendezewaarschuwingnietwordtopgevolgd,zouhetonderhoudspersoneel,de
gebruikerofeenpatiëntgewondkunnenrakenalsgevolgvaneenelektrischeschok,
mechanischeofanderegevaren.
HOIATUS
(ET)
Käesolevteenindusjuhendonsaadavalainultinglisekeeles.
•Kuiklienditeeninduseosutajanõuabjuhenditinglisekeelesterinevaskeeles,vastutab
klienttõlketeenuseosutamiseeest.
•Ärgeüritageseadmeidteenindadaenneeelnevaltkäesolevateenindusjuhendiga
tutvumistjasellestarusaamist.
•Käesolevahoiatuseeiraminevõibpõhjustadateenuseosutaja,operaatorivõipatsiendi
vigastamistelektrilöögi,mehaanilisevõimuuohutagajärjel.
VAROITUS
(FI)
Tämähuolto-ohjeonsaatavillavainenglanniksi.
•Josasiakkaanhuoltohenkilöstövaatiimuutakuinenglanninkielistämateriaalia,tarvittavan
käännöksenhankkiminenonasiakkaanvastuulla.
•Äläyritäkorjatalaitteistoaennenkuinoletvarmastilukenutjaymmärtänyttämän
huolto-ohjeen.
•Mikälitätävaroitustaeinoudateta,seurauksenavoiollahuoltohenkilöstön,laitteiston
käyttäjäntaipotilaanvahingoittuminensähköiskun,mekaanisenviantaimuun
vaaratilanteenvuoksi.
ATTENTION
(FR)
Cemanueltechniquen'estdisponiblequ'enanglais.
•Siunservicetechniqueclientsouhaiteobtenircemanueldansuneautrelangueque
l'anglais,ildevraprendreenchargelatraductionetlaresponsabilitéducontenu.
•Nepastenterd'intervenirsurleséquipementstantquelemanueltechniquen'apasété
consultéetcompris.
•Lenon-respectdecetavertissementpeutentraînerchezletechnicien,l'opérateuroule
patientdesblessuresduesàdesdangersélectriques,mécaniquesouautres.
4
MUSE™v92059568-013J
13March2019
Page 5

ServiceManualLanguageInformation(cont'd.)
WARNUNG
(DE)
ΠΡΟΕΙΔΟΠΟΙΗΣΗ
(EL)
FIGYELMEZTETÉS
(HU)
AÐVÖRUN
(IS)
PERINGATAN
(ID)
DieseServiceanleitungistnurinenglischerSpracheverfügbar.
•FallsderKundendiensteineandereSprachebenötigt,musserfüreineentsprechende
Übersetzungsorgen.
•KeineWartungdurchführen,ohnedieseServiceanleitunggelesenundverstandenzu
haben.
•BeiZuwiderhandlungkanneszuVerletzungendesKundendiensttechnikers,des
AnwendersoderdesPatientendurchStromschläge,mechanischeodersonstigeGefahren
kommen.
Τοπαρόνεγχειρίδιοσέρβιςδιατίθεταιστααγγλικάμόνο.
•Εάντοάτομοπαροχήςσέρβιςενόςπελάτηαπαιτείτοπαρόνεγχειρίδιοσεγλώσσαεκτός
τωναγγλικών,αποτελείευθύνητουπελάτηναπαρέχειυπηρεσίεςμετάφρασης.
•Μηνεπιχειρήσετετηνεκτέλεσηεργασιώνσέρβιςστονεξοπλισμόεκτόςεάνέχετε
συμβουλευτείκαιέχετεκατανοήσειτοπαρόνεγχειρίδιοσέρβις.
•Εάνδενλάβετευπόψητηνπροειδοποίησηαυτή,ενδέχεταιναπροκληθείτραυματισμός
στοάτομοπαροχήςσέρβις,στοχειριστήήστονασθενήαπόηλεκτροπληξία,μηχανικούς
ήάλλουςκινδύνους.
Ezaszervizkézikönyvkizárólagangolnyelvenérhetőel.
•Haavevőszervizellátójaangoltóleltérőnyelvretartigényt,akkoravevőfelelőssége
afordításelkészíttetése.
•Nepróbáljaelkezdenihasználniaberendezést,amígaszervizkézikönyvbenleírtakat
nemértelmeztékésértettékmeg.
•Ezengyelmeztetésgyelmenkívülhagyásaaszervizellátó,aműködtetővagyapáciens
áramütés,mechanikaivagyegyébveszélyhelyzetmiattisérüléséteredményezheti.
Þessiþjónustuhandbókereingöngufáanlegáensku.
•Efaðþjónustuveitandiviðskiptamannsþarfnastannarstungumálsenensku,erþaðskylda
viðskiptamannsaðskaffatungumálaþjónustu.
•Reyniðekkiaðafgreiðatækiðnemaþessiþjónustuhandbókhefurveriðskoðuðogskilin.
•Brotáaðsinnaþessariaðvörungeturleitttilmeiðslaáþjónustuveitanda,stjórnandaeða
sjúklingifráraosti,vélrænumeðaöðrumáhættum.
ManualservisinihanyatersediadalambahasaInggris.
•JikapenyediajasaservispelangganmemerlukanbahasalainselaindariBahasaInggris,
merupakantanggungjawabdaripenyediajasaservistersebutuntukmenyediakan
terjemahannya.
•Janganmencobamelakukanservisterhadapperlengkapankecualitelahmembacadan
memahamimanualservisini.
•Mengabaikanperingataninibisamengakibatkancederapadapenyediaservis,operator,
ataupasien,karenaterkenakejutlistrik,bahayamekanisataubahayalainnya.
2059568-013JMUSE™v9
5
Page 6

ServiceManualLanguageInformation(cont'd.)
AVVERTENZA
(IT)
警告
(JA)
CẢNHBÁO
(VI)
ЕСКЕРТУ
(KK)
BRĪDINĀJUMS
(LV)
ĮSPĖJIMAS
(LT)
IlpresentemanualedimanutenzioneèdisponibilesoltantoinInglese.
•Seunaddettoallamanutenzionerichiedeilmanualeinunalinguadiversa,ilclienteè
tenutoaprovvederedirettamenteallatraduzione.
•Siprocedaallamanutenzionedell'apparecchiaturasolodopoaverconsultatoilpresente
manualeedavernecompresoilcontenuto.
•Ilnonrispettodellapresenteavvertenzapotrebbefarcompiereoperazionidacuiderivino
lesioniall'addetto,allamanutenzione,all'utilizzatoreedalpazienteperfolgorazione
elettrica,perurtimeccaniciodaltririschi.
このサービスマニュアルは英語版しかありません。
•サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業
者の責任で行うものとさせていただきます。
•このサービスマニュアルを熟読し、十分に理解をした上で装置のサービスを
行ってください。
•この警告に従わない場合、サービスを担当される方、操作員あるいは患者が、感電
や機械的又はその他の危険により負傷する可能性があります。
TàiLiệuHướngDẫnSửaChữachỉcóbảntiếngAnh.
•Nếucácđơnvịcungcấpdịchvụchokháchhàngyêucầumộtngônngữnàokháctiếng
Anh,thìkháchhàngsẽcótráchnhiệmcungcấpcácdịchvụdịchthuật.
•KhôngđượcsửachữathiếtbịtrừkhiđãthamkhảovàhiểuTàiliệuHướngdẫnSửachữa.
•Khôngtuânthủnhữngcảnhbáonàycóthểdẫnđếncáctổnthươngchongườithực
hiệnsửachữa,ngườivậnhànhhaybệnhnhân,dosốcđiện,cácrủirovềcơkhíhay
cácrủirokhác.
Бұлқызметкөрсетубойыншанұсқаулығытекағылшынтіліндеқолжетімді.
•Тұтынушыныңқызметпровайдеріағылшынтіліненбасқатілдегінұсқаныталапетсе,
аударубойыншақызметтеріменқамтамасызетутұтынушыжауапкершілігіндеболуы
тиіс.
•Бұлқызметкөрсетубойыншанұсқаулығынназарғаалып,түсінбегенше,жабдыққа
қызметкөрсетуденбастартыңыз.
•Бұлескертудіелемеуқызметпровайдері,операторнемесеемделушініңэлектр
шогынан,механикалықнемесебасқақауіптернәтижесіндежарақаталуынаәкелуі
мүмкін.
Šīapkalpotājurokasgrāmatairpieejamatikaiangļuvalodā.
•Jaapkalpošanassniedzējamnepieciešamainformācijacitā,nevisangļu,valodā,klienta
pienākumsirnodrošināttāstulkošanu.
•Neveicietaprīkojumaapkopi,neizlasotunnesaprototapkalpotājurokasgrāmatu.
•Šībrīdinājumaneievērošanavarradītelektriskāsstrāvastrieciena,mehāniskuvaiciturisku
izraisītutraumuapkopessniedzējam,operatoramvaipacientam.
Šiseksploatavimovadovasyraprieinamastikanglųkalba.
•Jeiklientopaslaugųtiekėjasreikalaujavadovokitakalba-neanglų,numatytivertimo
paslaugasyraklientoatsakomybė.
•Nemėginkiteatliktiįrangostechninėspriežiūros,nebentatsižvelgėteįšįeksploatavimo
vadovąirjįsupratote.
•Jeineatkreipsitedėmesioįšįperspėjimą,galimisužalojimaidėlelektrosšoko,mechaninių
arkitųpaslaugųtiekėjui,operatoriuiarpacientui.
6MUSE™v92059568-013J
13March2019
Page 7

ServiceManualLanguageInformation(cont'd.)
ADVARSEL
(NO)
Denneservicehåndbokennnesbarepåengelsk.
•Hviskundensserviceleverandørtrengeretannetspråk,erdetkundensansvaråsørge
foroversettelse.
•Ikkeforsøkåreparereutstyretutenatdenneservicehåndbokenerlestogforstått.
•Manglendehensyntildenneadvarselenkanføretilatserviceleverandøren,operatøren
ellerpasientenskadespågrunnavelektriskstøt,mekaniskeellerandrefarer.
OSTRZEŻENIE
(PL)
Niniejszypodręcznikserwisowydostępnyjestjedyniewjęzykuangielskim.
•Jeślidostawcausługklientawymagajęzykainnegoniżangielski,zapewnienieusługi
tłumaczeniajestobowiązkiemklienta.
•Nienależyserwisowaćwyposażeniabezzapoznaniasięizrozumienianiniejszego
podręcznikaserwisowego.
•Niezastosowaniesiędotegoostrzeżeniamożespowodowaćurazydostawcyusług,
operatoralubpacjentawwynikuporażeniaelektrycznego,zagrożeniamechanicznego
bądźinnego.
AVISO
(PT-BR)
Estemanualdeassistênciatécnicasóseencontradisponíveleminglês.
•SeoserviçodeassistênciatécnicadoclientenãoforGE,eprecisardeoutroidioma,será
daresponsabilidadedoclientefornecerosserviçosdetradução.
•Nãotenterepararoequipamentosemterconsultadoecompreendidoestemanualde
assistênciatécnica.
•Onãocumprimentodesteavisopodeporemperigoasegurançadotécnico,operadorou
pacientedevidoachoqueselétricos,mecânicosououtros.
AVISO
(PT-PT)
Estemanualtécnicosóseencontradisponíveleminglês.
•Seaassistênciatécnicadoclientesolicitarestesmanuaisnoutroidioma,éda
responsabilidadedoclientefornecerosserviçosdetradução.
•Nãotenterepararoequipamentosemterconsultadoecompreendidoestemanual
técnico.
•Onãocumprimentodesteavisopodeprovocarlesõesaotécnico,aoutilizadorouao
pacientedevidoachoqueseléctricos,mecânicosououtros.
AVERTISMENT
(RO)
Acestmanualdeserviceestedisponibilnumaiînlimbaengleză.
•Dacăunfurnizordeserviciipentruclienţinecesităoaltălimbădecâtceaengleză,estede
datoriaclientuluisăfurnizezeotraducere.
•Nuîncercaţisăreparaţiechipamentuldecâtulteriorconsultăriişiînţelegeriiacestui
manualdeservice.
•Ignorareaacestuiavertismentarputeaducelarănireadepanatorului,operatoruluisau
pacientuluiînurmapericolelordeelectrocutare,mecanicesaudealtănatură.
ПРЕДУПРЕЖДЕНИЕ
(RU)
Настоящееруководствопообслуживаниюпредлагаетсятольконаанглийскомязыке.
•Еслисервисномуперсоналуклиентанеобходиморуководствоненаанглийском,ана
каком-тодругомязыке,клиентуследуетобеспечитьпереводсамостоятельно.
•Преждечемприступатькобслуживаниюоборудования,обязательнообратитеськ
настоящемуруководствуивнимательноизучитеизложенныевнемсведения.
•Несоблюдениетребованийданногопредупрежденияможетпривестиктому,что
специалистыпообслуживанию,операторыилипациентыполучатударэлектрическим
током,механическуютравмуилидругоеповреждение.
2059568-013JMUSE™v9
7
Page 8

ServiceManualLanguageInformation(cont'd.)
UPOZORENJE
(SR)
VAROVANIE
(SK)
OPOZORILO
(SL)
ADVERTENCIA
(ES)
VARNING
(SV)
UYARI
(TR)
Ovoservisnouputstvojedostupnosamonaengleskomjeziku.
•Akoklijentovserviserzahtevanekidrugijezik,klijentjedužandaobezbediprevodilačke
usluge.
•Nepokušavajtedaopraviteuređajakonistepročitaliirazumeliovoservisnouputstvo.
•Zanemarivanjeovogupozorenjamožedovestidopovređivanjaservisera,rukovaocaili
pacijentausledstrujnogudara,ilimehaničkihidrugihopasnosti.
Tentonávodnaobsluhujekdispozíciilenvangličtine.
•Akzákazníkovposkytovateľslužiebvyžadujeinýjazykakoangličtinu,poskytnutie
prekladateľskýchslužiebjezodpovednosťouzákazníka.
•Nepokúšajtesaoobsluhuzariadeniaskôr,akosineprečítatenávodnaobsluhua
neporozumietemu.
•Zanedbanietohtovarovaniamôževyústiťdozraneniaposkytovateľaslužieb,obsluhujúcej
osobyalebopacientaelektrickýmprúdom,mechanickýmaleboinýmnebezpečenstvom.
Taservisnipriročnikjenavoljosamovangleškemjeziku.
•Čeponudnikstoritvestrankepotrebujepriročnikvdrugemjeziku,morastrankazagotoviti
prevod.
•Neposkušajteservisiratiopreme,četegapriročnikanistevcelotiprebraliinrazumeli.
•Četegaopozorilaneupoštevate,selahkozaradielektričnegaudara,mehanskihalidrugih
nevarnostipoškodujeponudnikstoritev,operateralibolnik.
Estemanualdeserviciosóloexisteeninglés.
•Sielencargadodemantenimientodeunclientenecesitaunidiomaquenoseaelinglés,el
clientedeberáencargarsedelatraduccióndelmanual.
•Nosedeberádarserviciotécnicoalequipo,sinhaberconsultadoycomprendidoeste
manualdeservicio.
•Lanoobservanciadelpresenteavisopuededarlugaraqueelproveedordeservicios,el
operadoroelpacientesufranlesionesprovocadasporcausaseléctricas,mecánicaso
deotranaturaleza.
Denhärservicehandbokennnsbaratillgängligpåengelska.
•Omenkundsserviceteknikerharbehovavettannatspråkänengelskaansvararkunden
föratttillhandahållaöversättningstjänster.
•Försökinteutföraservicepåutrustningenomduinteharlästochförstårdenhär
servicehandboken.
•Omduintetarhänsyntilldenhärvarningenkandetresulteraiskadorpåserviceteknikern,
operatörenellerpatiententillföljdavelektriskastötar ,mekaniskafarorellerandrafaror.
Buserviskılavuzununsadeceİngilizcesimevcuttur.
•Eğermüşteriteknisyenibukılavuzuİngilizcedışındabirbaşkalisandantalepederse,bunu
tercümeettirmekmüşteriyedüşer.
•Serviskılavuzunuokuyupanlamadanekipmanlaramüdahaleetmeyiniz.
•Buuyarıyauyulmaması,elektrik,mekanikveyadiğertehlikelerdendolayıteknisyen,
operatörveyahastanınyaralanmasınayolaçabilir.
8MUSE™v92059568-013J
13March2019
Page 9

ServiceManualLanguageInformation(cont'd.)
ЗАСТЕРЕЖЕННЯ
(UK)
CẢNHBÁO
(VI)
Данекерівництвозсервісногообслуговуванняпостачаєтьсявиключноанглійськоюмовою.
•Якщосервіснийінженерпотребуєкерівництвоіншоюмовою,користувачзобов'язаний
забезпечитипослугиперекладача.
•Ненамагайтесяздійснюватитехнічнеобслуговуванняданогообладнання,якщо
винечитали,абонезрозумілиінформацію,наданувкерівництвізсервісного
обслуговування.
•Недотриманняцьогозастереженняможепризвестидотравмуваннясервісного
інженера,користувачаданогообладнанняабопацієнтавнаслідокелектричногошоку,
механічногоушкодженняабозіншихпричинневірногообслуговуванняобладнання.
TàiLiệuHướngDẫnSửaChữachỉcóbảntiếngAnh.
•Nếucácđơnvịcungcấpdịchvụchokháchhàngyêucầumộtngônngữnàokháctiếng
Anh,thìkháchhàngsẽcótráchnhiệmcungcấpcácdịchvụdịchthuật.
•KhôngđượcsửachữathiếtbịtrừkhiđãthamkhảovàhiểuTàiliệuHướngdẫnSửachữa.
•Khôngtuânthủnhữngcảnhbáonàycóthểdẫnđếncáctổnthươngchongườithực
hiệnsửachữa,ngườivậnhànhhaybệnhnhân,dosốcđiện,cácrủirovềcơkhíhay
cácrủirokhác.
2059568-013JMUSE™v99
Page 10

10MUSE™v92059568-013J
13March2019
Page 11

Contents
1Introduction
2CASEtoMUSECommunication
ManualPurpose........................................................................................17
RelatedDocuments...................................................................................17
TheoryofOperation..................................................................................19
InformationTransmission...........................................................................19
CommunicationLevels..............................................................................20
CongurationRequirements.....................................................................20
CustomerRequirements............................................................................20
CommunicationLevelRequirements.............................................................20
ConsiderationsWhenUsingMultipleCASESystems..........................................21
ProcessOverview......................................................................................21
3MARStoMUSECommunication
AddingtheStressExerciseOptiontotheMUSESystem...........................22
MUSEUserandSharedAcquisitionFolder................................................23
CreatingtheMUSEAcqUsersLocalWindowsGroup..........................................23
CreatingtheCASE8000WindowsUser...........................................................24
CreatingtheCASE8000WindowsShare.........................................................24
AddingtheUsertotheMUSEWebUsersGroup.......................................25
CreatingaMUSEUser...............................................................................25
ConguringtheCASESystemNetworkSettings.......................................26
ConguringMUSESystemSettingsontheCASESystem..........................27
ConguringCASEReportsontheMUSEFileServer..................................31
SystemCheckout......................................................................................33
VerifyingRecordTransfer...........................................................................33
VerifyingRequestsforData.........................................................................34
VerifyingtheRetrievalofPatientDemographics...............................................34
VerifyingtheRetrievalofOrders...................................................................34
Troubleshooting........................................................................................35
TheoryofOperation..................................................................................37
MUSEServices........................................................................................37
2059568-013JMUSE™v911
Page 12

InformationTransmission...........................................................................38
CustomerRequirements...........................................................................38
ConguringMARStoMUSECommunication.............................................38
ConguringtheMARSSystemNetworkSettings...............................................39
VerifyingtheMARSSoftwareVersion.............................................................39
RecordingtheIPAddressorHostNameoftheMARSSystem...............................39
VerifyingtheMARSReportsShare.................................................................39
ActivatingtheMARStoMUSEOptionontheMARSSystem(s)................................40
SettingUptheSiteInformationontheMARSSystem(s).......................................40
AddingtheHolterDataStorageOptiontotheMUSESystem................................41
AddingtheMARSSystemtotheMUSEGeneracqConguration............................42
InstallingtheMARSPrintFormatter...............................................................43
Creatingsite.inifortheMARSPrintFormatter...................................................45
Copyingtherusty.iniCongurationFilefromtheMARSSystemtotheMUSE
System..................................................................................................46
SystemCheckout......................................................................................46
SavingaReportontheMARSSystem............................................................46
SendingaMARSHolterReporttotheMUSESystem...........................................47
ViewingaMARSHolterReportintheMUSEEditor.............................................47
PrintingMARSHolterReportsfromtheMUSESystem.........................................48
Troubleshooting........................................................................................48
4MUSEMonitoringGateway
TheoryofOperation..................................................................................51
InformationTransmission...........................................................................52
InstallingtheMUSEMonitoringGateway.................................................52
PreparingfortheInstallationoftheMUSEMonitoringGateway.............................53
VerifyingandConguringNetworkConnections...............................................53
FirewallConsiderations..............................................................................54
CreatingaShareontheMUSEMonitoringGateway...........................................54
InstallingtheMUSEMonitoringGatewaySoftware............................................55
ConguringtheMUSEMonitoringGatewaySoftware.........................................55
ConguringtheMUSEApplicationServer........................................................56
ConguringBedsideMonitors......................................................................57
SystemCheckout......................................................................................57
Troubleshooting........................................................................................58
UninstallingMUSEMonitoringGatewayv1.1............................................59
5MACECGSystemstoMUSECommunication
TheoryofOperation..................................................................................61
CustomerRequirements...........................................................................62
ConguringMACECGtoMUSECommunication.......................................62
SettingUpModems...................................................................................63
AddingtheWireless/LANCommunicationOptiontotheMUSESystem....................63
Verifying/InstallingtheMUSEModemService..................................................64
12MUSE™v92059568-013J
Page 13

SettingUpDCPInboundCommunication.................................................69
SystemCheckout......................................................................................71
6MUSEeDocConnect
TheoryofOperation..................................................................................73
PreparingtoCongureeDocConnect......................................................74
CustomerRequirements...........................................................................74
InstallingandConguringMUSEeDocConnect.......................................74
SettingUpaModemDevice........................................................................67
RestartingModems..................................................................................68
Verifying/InstallingtheDCPInboundServiceandDCPCommunicationOption...........69
SettingUptheDCPServerCongurationintheMUSESystem...............................70
CSITransmissiontotheMUSESystem...........................................................71
MUSEOrderDownloadviaCSI.....................................................................72
DCPTransmissiontotheMUSESystem..........................................................72
MUSEOrderDownloadviaDCP....................................................................72
InstallingtheeDocConnectOptionintheMUSESystem.....................................75
AddingorEnablingaTestTypeintheMUSESystem..........................................75
SettinguptheAcquisitionProleintheMUSESystem........................................76
SettinguptheFileShare............................................................................77
SystemCheckout......................................................................................78
Troubleshooting........................................................................................79
7MUSEXMLImportOption
TheoryofOperation..................................................................................81
CustomerRequirements...........................................................................81
InstallingandConguringtheMUSEXMLImportOption.........................81
InstallingtheXMLImportOptionandtheMUSEXMLParserService.......................82
SettinguptheXMLSharedFolder.................................................................83
UsingXMLCONFIG.EXEtoAdd,Update,orDeleteXMLDevices..............................83
SystemCheckout......................................................................................87
Troubleshooting........................................................................................87
8DICOMCommunication
TheoryofOperation..................................................................................89
TransmissionFlowCharts...........................................................................90
MUSEServices........................................................................................91
CustomerRequirements............................................................................91
ConguringDICOMCommunication.........................................................91
AddingtheDICOMService(s)andOptiontotheMUSESystem...............................92
ConguringtheMUSESystemtoReceiveDICOMTests.......................................93
ConguringtheSystemtoSendDICOMTests...................................................95
2059568-013JMUSE™v913
Page 14

9MUSEAPI3Installation
ConguringtheSystemtoQueryforDICOMOrders...........................................99
SystemCheckout....................................................................................102
ReceivingDICOMTestsintotheMUSESystem................................................102
SendingDICOMTestsfromtheMUSESystem.................................................102
QueryingforDICOMOrders.......................................................................102
Troubleshooting......................................................................................102
TheoryofOperation................................................................................106
Pre-InstallationInstructions...................................................................106
DeterminingWhetherMUSEAPI3isAlreadyInstalled........................................106
DeterminingtheCommunicationProtocol(s)thatMUSEAPI3Uses.......................107
DeterminingthePortAssignmentsforMUSEAPI3............................................107
LocatingtheMUSEApplicationFolderontheMUSEServer................................107
InstallingMUSEAPI3................................................................................107
ChangingtheMUSEAPI3ServiceProtocolConguration........................112
RemovingMUSEAPI3...............................................................................113
RestoringtheMUSEAPI3Conguration..................................................114
MUSEAPITestClient...............................................................................114
RunningtheMUSEAPITestClient...............................................................114
UsingtheMUSEAPITestClient...................................................................114
ConguringSSLCerticatefortheMUSEAPI3Port.................................115
10MUSEWebCompatibilityLayer
TheoryofOperation................................................................................117
SystemRequirements.............................................................................117
RequiredInternetInformationServices(IIS)fortheMUSEWebCompatibility
Layer..................................................................................................118
InstallingtheMUSEWebCompatibilityLayer.........................................119
CreatingaMUSEUserfortheMUSEWebCompatibilityLayerWebsite..................120
InstallingandConguringtheMUSEWebCompatibilityLayer............................121
AddingUserstotheMUSEWebCompatibilityLayerWebsite................................124
SystemCheckout....................................................................................124
ChangingtheMUSEWebCompatibilityLayerConguration
Settings...................................................................................................127
ManualCorrectionsforMUSEWebCompatibilityLayerInstaller...........128
NologinpromptorHTTPError401.2whenaccessingtheMUSEWebCompatibility
Layerwebsite.......................................................................................128
HTTPError404.0WhenAccessingtheMUSEWebCompatibilityLayer
Website...............................................................................................128
14MUSE™v92059568-013J
Page 15

11MACCRACompatibility
Troubleshooting......................................................................................129
UninstallingtheMUSEWebCompatibilityLayer....................................130
CreatinganAdditionalMUSEWebCompatibilityLayerWebsite............131
DisablingDirectoryBrowsingforMUSEWebCompatibilityLayer...........133
IISUnlistedFileNameExtensionCongurationforMUSEWeb
CompatibilityLayer.................................................................................133
ManuallyInstallingandVerifyingIISRelatedRoles,RoleServices,and
Features..................................................................................................134
InstallingIIS7RoleServicesonWindowsServer2008or2008R2........................134
InstallingIIS8RoleServicesonWindowsServer2012or2012R2........................136
TheoryofOperation................................................................................139
InstallationRequirements.......................................................................140
BeforeYouBegin...................................................................................140
CustomerRequirements..........................................................................140
MACCRACompatibilitywithCVWebv1.xandv2.x.................................140
CreatingaMUSEUserforMACCRACompatibility...................................141
InstallingMACCRACompatibility............................................................142
VerifyingtheMACCRACompatibilityInstallation....................................146
SystemCheckout....................................................................................146
ChangingMACCRACompatibilityConguration.....................................148
MOCKRACongWriter.exe.........................................................................148
MOCKRAServerProvisioner.exe...................................................................149
UninstallingMACCRACompatibility........................................................150
12CardioDaytoMUSESystemInterface
TheoryofOperation................................................................................151
CardioDaytoMUSEHolterPatientADT/OrderInterface....................................151
CardioDayHolterReportExporttotheMUSESystem.......................................152
DeterminingMUSEeDocConnectParameters...............................................153
CardioDaytoMUSEInterfacePrerequisites...................................................153
VerifyingtheCardioDaySoftwareVersion.....................................................154
VerifyingtheMUSESystemSoftwareVersion.................................................154
CustomerRequirements..........................................................................155
AdditionalResources..............................................................................155
ConguringCardioDayandMUSEInterfaceSettings.............................155
EnablingtheHolterDataStorage,eDocConnect,andHISInterfaceOptiononthe
MUSESystem........................................................................................155
EnablingMUSEHolterTestTypeforEachSite.................................................156
2059568-013JMUSE™v915
Page 16

EnablingMUSEADTOrdersandHolterOrdersforEachSite...............................156
CreatingtheCardioDayAcquisitionProleontheMUSESystem..........................157
SettingUptheCardioDayShareFolderintheMUSESystem...............................157
CreatingaDedicatedMUSEUserAccountforCardioDayHolterOrders
Query.................................................................................................158
ConguringtheCardioDaySystemforMUSEHolterOrders................................159
ConguringtheCardioDaySystemtoExportHolterReportstotheMUSE
System................................................................................................159
DisablingtheeDocConnectOption.............................................................159
CardioDay/MUSESystemCheckout......................................................160
CardioDayHolterReportExporttoMUSECheckout.........................................160
MUSEOrderOptionCheckout.................................................................160
Troubleshooting......................................................................................160
TroubleshootingtheCardioDayHolterReportExporttotheMUSESystem.............160
TroubleshootingtheCardioDayMUSEOrderOptionandeDocConnect.................161
13MUSECongurationforVAVistAImaging
TheoryofOperation................................................................................163
InformationTransmission.........................................................................164
CustomerRequirements.........................................................................164
ProcessOverview....................................................................................164
ConguringtheVOL000Share................................................................164
ConguringVAVistAImagingFormats...................................................165
InstallingMUSEAPI3................................................................................167
InstallingMACCRACompatibility............................................................167
SystemCheckout....................................................................................168
CreatingEnhancedMetaleUsingMUSEAPIDCOMTester................................168
ViewingEnhancedMetaleUsingMUSEImageViewer.....................................169
16MUSE™v92059568-013J
Page 17

1
Introduction
ManualPurpose
Thisdocumentprovidesinformationoninstallingandconguring:
•CASEtoMUSEcommunication
•MARStoMUSEcommunication
•MUSEMonitoringGateway
•MACECGsystemstoMUSEcommunication
•MUSEeDocConnect
•MUSEXMLImportoption
•MUSEDICOMCommunication
•MUSEAPI3Installation
•MUSEWebCompatibilityLayer
•MACCRACompatibility
•CardioDaytoMUSESystemInterface
•MUSECongurationforVAVistAImaging
RelatedDocuments
Thefollowingdocumentsprovideadditionalinformationthatmaybehelpfulinthe
installation,conguration,maintenance,anduseofthissystem.
PartNumberDocumentTitle
2020299–021
2020299–025
2059568–009
2059568–011
2059568–016
2059568–017
2059568-013JMUSE™v917
MobileLinkInstallationManual
LANOptionforMACInstallationandTroubleshooting
MUSEv9CardiologyInformationSystemOperator’sManual
MUSEv9CardiologyInformationSystemInstallationandUpgrade
Manual
MUSEv9CardiologyInformationSystemRegulatoryandSafety
Guide
MUSEv9CardiologyInformationSystemServiceManual
Page 18

Introduction
PartNumberDocumentTitle
2059568–018
2059568–023
MUSEv9CardiologyInformationSystemPre-InstallationGuide
MUSEv9CardiologyInformationSystemDICOMConformance
Statement
2059568–025
MUSEv9CardiologyInformationSystemTransactionalXML
Developer’sManual
2092513-001
2092513-002
2107900–003
2107900–004
CardioDayV2.5Pre-InstallationManual
CardioDayV2.5InstallationandFieldServiceManual
CVWebv3.0InstallationManual
MUSEEnterpriseIntegrationReferenceManual
18MUSE™v92059568-013J
Page 19

2
CASEtoMUSECommunication
ThischapterdescribeshowtocongureCASEsystemstocommunicatewithMUSEv9
systems.Intheseinstructions,thetermCASEdenoteseitherCASE,CardioSoft,orCS.
NOTE:
MUSEv9systemssupportonlyWindowsServer2008andneweroperating
systems.WindowsServer2008andneweroperatingsystemsarenotcompatible
withWindowsNTbasedCASEsystems.
TheoryofOperation
CASEsystemtoMUSEsystemcommunicationallowsyoutotransferECGsandstress
exercisetestsfromtheCASEsystemtotheMUSEsystemforviewing,editing,printing,
andstorage.CASEsystemtoMUSEsystemcommunicationalsoallowsfortheCASE
systemtoretrievepatientinformationandordersviatheMUSEWebCompatibility
Layer.
ExampleNetworkDiagram
InformationTransmission
InCASEsystemtoMUSEsystemcommunication,theCASEsystemtransfersteststoa
sharedfolderontheMUSEapplicationserver.TheMUSEGeneracqservicesearches
thissharedfolderforlesandpullsthemtotheMUSEsystemforprocessing.Testsare
thennormalizedontheMUSEsystemandstoredinthedatabase.
TransmissionFlowChart
2059568-013JMUSE™v919
Page 20

CASEtoMUSECommunication
TheCASEsystemalsomakesrequestsforpatientororderinformation.Inthese
cases,theMUSEWebCompatibilityLayerreturnsthepatientororderinformationto
theCASEsystem.
MUSEWebRequestFlowChart
CommunicationLevels
TheCASEsystemsandMUSEsystemssupportthreecommunicationlevels.Each
levelbuildsonthefeaturesandrequirementsofthepreviouslevel,asshowninthe
followingtable:
CommunicationLevels
CommunicationLevel
Record
Transfer
MUSEWeb
CompatibilityLayer
HISOrdersOrderData
TheHISOrderslevelincludesthefeaturesandrequirementsoftheMUSEWeb
CompatibilityLayerlevel,whichinturnincludesthefeaturesandrequirementsof
theRecordTransferlevel.
CongurationRequirements
Therequirementsspeciedinthissectionmustbesatisedbeforeyoubegin
conguringCASEsystemtoMUSEsystemcommunication.
CustomerRequirements
Data
Exchanged
CASEReportsCASEMUSE
MUSE
Reports
PatientData
Directions
FromTo
MUSECASE
MUSECASE
Transfer
Method
Sharedfolder
NetworkMUSEAPI3and
NetworkHISOrders
Required
Components
MUSEExercise
TestingData
Storageoption
MUSEWeb
Compatibility
Layer
Interface
option
Thecustomerisresponsibleforprovidingtheappropriatenetworkconnectivity,
includingnameresolution,betweentheCASEsystemsandtheMUSEapplication
server.
CommunicationLevelRequirements
IfyouareusingtheMUSEWebcommunicationlevel,youmustinstalltheMUSEWeb
CompatibilityLayerbeforeproceeding.ForinformationoninstallingtheMUSEWeb
CompatibilityLayer,seeChapter10“MUSEWebCompatibilityLayer”onpage117.
20MUSE™v92059568-013J
Page 21

IfyouareusingtheHISOrderscommunicationlevel,youmustinstallandfully
conguretheHISOrdersInterfaceoptionbeforeproceeding.TheHISOrders
InterfaceisconguredbyatrainedHL7implementationperson.ContacttheHL7
implementationpersontoconrmwhethertheHISOrdersInterfaceiscongured.
ConsiderationsWhenUsingMultipleCASESystems
IfyouareusingmultipleCASEsystems,reviewtheitemsbelowforadditional
information.
•WhenusingmultipleCASEsystemswithasingleMUSEapplicationserver,the
sameWindowsCASE8000usernameandpasswordmaybeusedwitheveryCASE
systemifdesired.
•IfyouareaddinganewCASEsystemtoanexistingMUSEsystemwithCASE
systemsalreadyconguredtocommunicatewithit,someofthestepsinthis
chaptermayhavealreadybeencompleted.
ProcessOverview
CASEtoMUSECommunication
Whilethesetupprocessdiffersdependingonthecommunicationlevelyouare
conguring,sometasksareperformedformorethanonelevel.Thefollowing
tableidentiesallsetuptasksintheorderinwhichtheyareperformedandwhich
communicationlevelsrequirethem.Usethetabletodeterminewhichtasksyouneed
toperformtosetuptherequiredcommunicationlevels.
SetupTasks
Record
Transfer
√
√
√√√
√√√
√√√
√√√
MUSE
Web
√√
√√
HIS
Orders
Tasks
“AddingtheStressExerciseOptiontotheMUSESystem”on
page22
“CreatingtheMUSEAcqUsersLocalWindowsGroup”onpage
23
“CreatingtheCASE8000WindowsUser”onpage24
“CreatingtheCASE8000WindowsShare”onpage24
“AddingtheUsertotheMUSEWebUsersGroup”onpage25
“CreatingaMUSEUser”onpage25
“ConguringtheCASESystemNetworkSettings”onpage26
“ConguringMUSESystemSettingsontheCASESystem”on
page27
√√
“ConguringCASEReportsontheMUSEFileServer”onpage31
Inaddition,atthebeginningofeachtask,atable(similartothefollowing)indicates
whichcommunicationlevelsrequirethecompletionofthattask.Ifthecommunication
2059568-013JMUSE™v921
Page 22

CASEtoMUSECommunication
levelyouaresettingupischecked,youneedtoperformthetask.Ifthecommunication
levelyouaresettingupisnotchecked,skipthetask.
ExampleofTaskstobeCompleted
RecordTransfer
MUSEWeb
HISOrders
√
√
√
AddingtheStressExerciseOptiontotheMUSE
System
Completethistaskonlyifthefollowingtablehasacheckmarknexttothe
communicationlevelyouaresettingup.Ifthecommunicationlevelyouaresetting
upisnotchecked,skipthistask.
AddStressExerciseOptiontotheMUSESystem
RecordTransfer
MUSEWeb
HISOrders
UsethefollowingproceduretoaddthestressexerciseoptiontotheMUSEsystem.This
proceduremayonlybeperformedbyaqualiedGEHealthcareservicerepresentative.
1.LogontotheMUSEapplicationserverastheMUSEAdministratoruser.
√
2.PerformafullorpartialshutdownoftheMUSEsystemfollowingtheauto
shutdownproceduresdescribedinthe“SystemAdministration”chapterofthe
MUSEv9CardiologyInformationSystemServiceManual.
NOTE:
Ifyouarejustcheckingtoseeiftheoptionisalreadyenabled,ashutdownis
notrequired.If,however,youhavetoenableordisabletheoption,theMUSE
serviceswillberestartedandashutdownisrequired.
3.GotoControlPanel>ProgramsandFeatures.
4.SelectMUSE9andclickChange.
TheWelcomewindowopens.
5.ChooseModifyandclickNext.
TheSelectFeaturewindowopens.
6.ContinuetoclickNextoneachwindowuntilyoureachtheChooseMUSE
Optionswindow.
7.VerifythatExerciseTestingDataStorageischecked.
Ifitisnot,checkitnow.
8.ClickNext.
TheMUSESerialNumberwindowopens.
22MUSE™v92059568-013J
Page 23

9.IfyouaddedtheExerciseTestingDataStorageoptioninstep7,enterthe
OptionsCongurationPassword.
NOTE:
OnlyqualiedGEHealthcareservicerepresentativeshaveaccesstothis
password.Thispasswordcannotbeprovidedtocustomers.
10.ClickNextuntilyourchangesareappliedandtheMaintenanceComplete
windowopens.
11.ClickFinish.
12.IfyouperformedafullorpartialshutdownoftheMUSEsysteminstep2,cancel
theshutdownfollowingtheproceduresdescribedinthe“SystemAdministration”
chapteroftheMUSEv9CardiologyInformationSystemServiceManual.
MUSEUserandSharedAcquisitionFolder
TheMUSEv9InstallShieldmayhavebeenconguredtocompletethefollowingactions
whentheMUSEsystemwasrstinstalled:
CASEtoMUSECommunication
•CreatingtheMUSEAcqUserslocalWindowsgroup
•CreatingtheCASE8000Windowsuser
•CreatingtheCASE8000Windowsshare
RefertotheMUSEv9CardiologyInformationSystemInstallationandUpgradeManual
fordetailsontheMUSEv9InstallShieldconguration.
IftheMUSEv9InstallerwasnotconguredtocreatetheCASEcongurationitems,
theymustnowbecompletedmanually.
NOTE:
IftheCASEcongurationitemswerecreatedwhentheMUSEsystemwasinitially
installed,proceedtothenexttaskforyourselectedcommunicationlevel.The
CASEcongurationitemsdonotneedtoberecreated.
CreatingtheMUSEAcqUsersLocalWindowsGroup
Completethistaskonlyifthefollowingtablehasacheckmarknexttothe
communicationlevelyouaresettingup.Ifthecommunicationlevelyouaresetting
upisnotchecked,skipthistask.
CreateMUSEAcqUsersLocalWindowsGroup
RecordTransfer
√
MUSEWeb
HISOrders
TotransferMUSEreportsfromaCASEsystemtotheMUSEsystem,youneedtogive
CASEusersaccesstotheMUSEsystem.Therstpartofthisaccessisthecreationof
theMUSEAcqUserslocalWindowsgroup.
Ifonedoesnotalreadyexist,createalocalgroupontheMUSEApplicationserver
namedMUSEAcqUsers.
2059568-013JMUSE™v923
Page 24

CASEtoMUSECommunication
CreatingtheCASE8000WindowsUser
Completethistaskonlyifthefollowingtablehasacheckmarknexttothe
communicationlevelyouaresettingup.Ifthecommunicationlevelyouaresetting
upisnotchecked,skipthistask.
CreateCASE8000WindowsUser
RecordTransfer
MUSEWeb
HISOrders
√
√
√
TotransferMUSEreportsfromaCASEsystemtotheMUSEsystem,inadditionto
givingCASEusersaccesstotheMUSEsystem,youalsoneedtocreateaCASE8000
Windowsuser.
1.IfaCASE8000userdoesnotalreadyexistontheMUSEsystem,createalocal
usernamedCASE8000ontheMUSEApplicationserverandassignthisusera
passwordofcase!8000.
ItisrecommendedthattheWindowsuserbeconguredwiththefollowing
optionsorequivalent:
•Usercannotchangepassword
•Passwordneverexpires
NOTE:
Youmayuseadifferentpasswordifdesired,however,thepassword
speciedontheCASEsystemmustmatchthepasswordusedhere.
2.Ifnotalreadydone,addthenewlycreatedCASE8000usertothelocalMUSE
AcqUsersgroupcreatedontheMUSEApplicationserverinthe“Creatingthe
MUSEAcqUsersLocalWindowsGroup”onpage23section.
CreatingtheCASE8000WindowsShare
Completethistaskonlyifthefollowingtablehasacheckmarknexttothe
communicationlevelyouaresettingup.Ifthecommunicationlevelyouaresetting
upisnotchecked,skipthistask.
CreateCASE8000WindowsShare
RecordTransfer
MUSEWeb
HISOrders
Lastly,totransferMUSEreportsfromaCASEsystemtotheMUSEsystem,youneedto
createaCASE8000share.
24MUSE™v92059568-013J
√
√
√
Page 25

CASEtoMUSECommunication
Bydefault,theMUSEGeneracqserviceontheMUSEApplicationserverisconguredto
checktheacqfolder(defaultisd:\muse\acq)forincominglestoprocess.Inorderto
transferrecordsfromaCASEsystemtotheMUSEsystem,thislocationmustbeshared.
1.Ifitisnotalreadyshared,sharetheacqfolder(defaultisd:\muse\acq)with
asharenameofCASE8000.
2.ProvideChangeandReadSharePermissionstotheMUSEAcqUsersgroup.
3.IftheEveryonegroupislistedwithsharepermissions,removeit.
AddingtheUsertotheMUSEWebUsersGroup
Completethistaskonlyifthefollowingtablehasacheckmarknexttothe
communicationlevelyouaresettingup.Ifthecommunicationlevelyouaresetting
upisnotchecked,skipthistask.
AddUsertoMUSEWebUsersGroup
RecordTransfer
MUSEWeb
HISOrders
ToaccessMUSEreportsthroughMUSEWeb,youmustaddtheCASE8000usertothe
MUSEWebUsersgroupontheMUSEapplicationserver.
NOTE:
YoumusthavetheMUSEWebCompatibilityLayerinstalledontheMUSE
Applicationserverbeforeproceedingwiththisstep.SeeChapter10“MUSEWeb
CompatibilityLayer”onpage117forinstructions.
Ifnotalreadydone,addthepreviouslycreatedlocalCASE8000usertothelocal
MUSEWebUsersgroupontheMUSEApplicationserver.
CreatingaMUSEUser
Completethistaskonlyifthefollowingtablehasacheckmarknexttothe
communicationlevelyouaresettingup.Ifthecommunicationlevelyouaresetting
upisnotchecked,skipthistask.
CreateMUSEUser
RecordTransfer
√
√
MUSEWeb
HISOrders
TorequestdatafromtheMUSEsystemviaMUSEWeb,youmustsetupaCASEuser
accountintheMUSEapplication.
2059568-013JMUSE™v925
√
√
Page 26

CASEtoMUSECommunication
Ifonehasnotalreadybeencreated,createanewuseraccountinMUSEUserSetup
withthefollowingvalues.
MUSEUser
Properties
General
General
GeneralWindowsUser
General
General
FieldValue
LastName8000
FirstName
Name
Accountis
Enabled
ActiveSites
CASE
<servername_or_domain>\CASE8000
NOTE:
Replace<servername_or_domain>withtheactual
nameoftheMUSEleserverforlocalaccountsor
thedomainfordomainaccounts.
Checkthebox.
Selectthesitestowhichtheuserneedsaccess.
NOTE:
TheCASEusercanonlyrequestdataforthesitesto
whichtheuseraccountisgrantedaccess.Therefore
itisimportantthatyouselectallthesitesforwhich
theCASEsystemmayrequestdata.
Advanced
AdvancedRoleViewOnly
AdvancedJobTitlesUncheckanyboxes.
UserID
EnteranavailableUserIDfortheMUSEsystemonwhich
youareworking.
AdvancedDisplayUserin
PersonnelLists
Uncheckthebox.
Ifnecessary,refertotheMUSEv9CardiologyInformationSystemOperatorManual
forinformationoncreatingusersintheMUSEsystem.
ConguringtheCASESystemNetworkSettings
Completethistaskonlyifthefollowingtablehasacheckmarknexttothe
communicationlevelyouaresettingup.Ifthecommunicationlevelyouaresetting
upisnotchecked,skipthistask.
CongureCASESystemNetworkSettings
RecordTransfer
MUSEWeb
HISOrders
FortheCASEsystemsandMUSEsystemtocommunicate,theCASEsystemsmust
conformtothenetworkingsettingsineffectattheinstallationsite.Contactthe
site’ssystemadministratortoobtainrequiredinformationortoassistinconguring
thesystem.
√
√
√
26MUSE™v92059568-013J
Page 27

CASEtoMUSECommunication
EnsurethatnetworkingisenabledonallCASEsystemsthatneedtocommunicatewith
theMUSEsystem.Additionally,conrmthatWindowsnetworksettingsontheseCASE
systemsareappropriatelyconguredtocommunicatewiththeMUSEapplication
andtheMUSEwebsiteviaTCP/IP.
ConguringMUSESystemSettingsontheCASE
System
Completethistaskonlyifthefollowingtablehasacheckmarknexttothe
communicationlevelyouaresettingup.Ifthecommunicationlevelyouaresetting
upisnotchecked,skipthistask.
CongureMUSESettingsontheCASESystem
RecordTransfer
MUSEWeb
HISOrders
BeforeyoucanfullyintegrateCASEsystemswiththeMUSEsystem,youmust
completetheMUSEcongurationontheCASEsystems.
NOTE:
NotalloftheMUSEcongurationmustbecompletedforallinstallations.Referto
thefollowingtabletodeterminewhichsectionstocomplete.
CASESystemCongurationWindow—
MUSEtab
SetupforMUSEsectionRequiredforMUSEWebandHISOrders
StoreprocedureforMUSEsectionRequiredforRecordTransfer
EverythingelseRequiredforRecordTransfer
1.StarttheCASEapplication.
2.SelectSystemConguration.OncetheSystemCongurationwindowopens,
selecttheMUSEtab.
√
√
√
SectionCompletionRequirements
2059568-013JMUSE™v927
Page 28

CASEtoMUSECommunication
3.CompletetheSetupforMUSEsectionbyperformingthetasksinthetable
followingthescreencapture.
FieldTask
RequestMUSEData
MUSEWebServerEnterthenameorIPaddressoftheMUSEapplication
Checkthisbox.
server.
28MUSE™v92059568-013J
Page 29

CASEtoMUSECommunication
FieldTask
MUSEUserName
EntertheusernameoftheuserwithaccesstoMUSE
Web.Thisistypicallycase8000.
NOTE:
IftheMUSEwebsiteisconguredwithadefault
domain,andalocalcase8000userisbeing
used,youmayneedtospecifytheMUSEUser
Nameas<muse_server_name>\case8000where
<muse_server_name>isthecomputernameof
theMUSEapplicationserver.
MUSEPasswordEnterthepasswordfortheuserdenedpreviously.This
istypicallycase!8000.
PortnumberEntertheportforthewebsiteontheMUSEsystem.This
istypically80.
PortnumberandSSL
Connection
InternetBrowser
IfanSSLconnectiontotheMUSEWebServeris
used,entertheSSLPortnumberandchecktheSSL
Connectionbox.TheSSLportistypically443.
Usethebrowsebuttontondtheexecutablefor
InternetExplorer.InternetExploreristypicallylocated
inthefollowingpath:C:\ProgramFiles\Internet
Explorer\Iexplore.exe.Adjustthepathaccordinglyfor
non-Englishoperatingsystems.
4.CompletetheStoreprocedureforMUSEsectionbyperformingthetasksin
thetablefollowingthescreencapture.
FieldTask
DatatransfertoMUSEviaShared
Directory
SharedDirectoryEntertheUNCpathofthe
2059568-013JMUSE™v929
Selectthisradiobutton.
sharedfolderontheMUSE
applicationserver.Thisistypically
\\<muse_server_name>\case8000
where<muse_server_name>isthe
computernameoftheMUSEapplication
server.
Page 30

CASEtoMUSECommunication
FieldTask
DirectoryUserName
MUSEPasswordEnterthepasswordfortheuserdened
Entertheusernameoftheuserwith
accesstotheshareddirectory.Thisis
typicallycase8000.
above.Thisistypicallycase!8000.
5.PopulatetheLocationNumber,Cart/DeviceNumber,andMUSESoftware
Versioneldsasspeciedinthetablefollowingthescreencapture.
FieldTask
LocationNumberEnterorselectthedesiredMUSElocation
Cart/DeviceNumberChooseauniquedevicenumberfor
MUSESoftwareVersionSelect7.2andabovefromthedrop-down
6.ClickOK.
NOTE:
ForinformationregardingallotheroptionsontheMUSEtaboftheSystem
Congurationwindow,pleaserefertotheappropriateCASEsystem
documentation.
fortheCASEsystem.Thislocationshould
matchtheappropriatelocationdened
intheMUSEsystem.
thisCASEsystem(andallfutureCASE
systems),toensurethatallrecordssent
totheMUSEsystemhaveuniquele
names.
list.
30MUSE™v92059568-013J
Page 31

ConguringCASEReportsontheMUSEFile
Server
Completethistaskonlyifthefollowingtablehasacheckmarknexttothe
communicationlevelyouaresettingup.Ifthecommunicationlevelyouaresetting
upisnotchecked,skipthistask.
CongureCaseReports
CASEtoMUSECommunication
RecordTransfer
MUSEWeb
HISOrders
InorderfortheMUSEsystemtocorrectlyformatCASEreportssenttotheMUSEshare,
theMUSEsystemmustusethesamereporttemplates.CASEreporttemplatesare
createdontheCASEsystemsandthencopiedtotheMUSEsystem.
RefertoyourCASEsystemdocumentationforinstructionsoncreatingthereport
templates.
1.AftercreatingthetemplatesontheCASEsystem,copytheNARRATIVfolder
fromtheCASEsystemtoremovablemediaoranetworkshare.
ThedefaultlocationisC:\CASE\NARRATIVorC:\CARDIO\NARRATIV.
2.LogontotheMUSEApplicationserverastheMUSEAdministratoruser.
3.OntheMUSEsystem,copytheNARRATIVfolderfromtheremovablemediaor
networksharetoatemporarylocationontheMUSEapplicationserversuchas
C:\CASE1\NARRATIV.
NOTE:
IfyouarecopyingreporttemplatesfrommultipleCASEsystems,createa
separatetemporarylocationforeachsystem,suchasC:\CASE1\NARRATIV,
C:\CASE2\NARRATIV,andsoon.
√
√
4.OntheMUSEsystem,openaCommandPrompt.
5.ChangetothelocationoftheMUSEapplicationles(defaultisC:\Program
Files(x86)\MUSE).
6.Typethefollowingcommand:
loadtemplate-path:"<pathtonarrativfolder>"-test:4-lang:<language>
-db:<dbname>
Where<pathtonarrativfolder>,<language>,and<dbname>areappropriately
populated.Usethefollowingtablestodeterminethevaluesoftheseitems.
ExamplecommandlineforanEnglishlanguageCASEsystemwithalocalMUSE
database:
2059568-013JMUSE™v931
Page 32

CASEtoMUSECommunication
loadtemplate-path:"C:\CASE1\NARRATIV"-test:4
NOTE:
Noticethatinthisexampleneitherthelanguagenorthedatabasename
wasspecied.BecausethedefaultlanguageisEnglishandthedatabaseis
localwithadefaultprexofMUSE,itwasnotnecessarytospecifyeitherof
theseinthecommand.
ExamplecommandlineforaSpanishlanguageCASEsystem,aremoteMUSE
databasenamedSQLSERVER1,andadefaultprexofMUSE:
loadtemplate-path:"C:\CASE1\NARRATIV"-test:4-lang:es
-db:SQLSERVER1\.MUSE
ExamplecommandlineforaSwedishlanguageCASEsystem,aremoteMUSE
database,aninstancenameofSQLSERVER1\MUSE,andadefaultprexof
MUSE:
loadtemplate-path:"C:\CASE1\NARRATIV"-test:4-lang:sv-db:SQLSERVER1\
MUSE.MUSE
CommandOptions
CommandOption
-pathRequired.
-langOptional.
-dbOptional.
LanguageCodes
Code
daDanish
de
en
es
Description
ThepathtotheCASEreporttemplates.
Forexample:C:\CASE1\NARRATIV.
Thereportlanguage.
Ifyoudonotincludethisswitch,thelanguagedefaultstoen
(English).Thelanguageyouentershouldmatchthelanguage
oftheCASEsystem.
Seethefollowingtableforotherlanguagecodes.
DatabaseNameandprex(forexampleServer\
Instance.Prex).
Thedefaultis\.MUSE.
Language
German
English
Spanish
Finnish
frFrench
it
32MUSE™v92059568-013J
Italian
Page 33

LanguageCodes(cont'd.)
CASEtoMUSECommunication
Code
ja
nlDutch
no
ru
sv
zh-chsSimpliedChinese
zh-chtTraditionalChinese
Language
Japanese
Norwegian
Russian
Swedish
7.PressEnter.
ThetemplatesareloadedintotheMUSEdatabase.
8.RepeatthisproceduretoloadtemplatesfromeachCASEsystemasdesired.
NOTE:
IfmorethanoneCASEsystemhasthesameReportTemplatename,the
lastReportTemplatewiththatnameloadedintotheMUSEdatabasewill
betheReportTemplateusedintheMUSEdatabase.UseuniqueReport
TemplatenamesattheCASEsystemstoavoidoverwritingreporttemplates
alreadyloadedintotheMUSEdatabase.
9.VerifythereporttemplatesareloadedintotheMUSEdatabase.
SystemCheckout
AftersettinguptheMUSEsystemandallconnectedCASEsystems,completethe
followingvericationprocedurestoensuretheCASEsystemtoMUSEsystem
communicationissetupproperly.
VerifyingRecordTransfer
UsethefollowingproceduretoverifythattheCASEsystemstransferreportstothe
MUSEshareandthattheMUSEsystemimportstheCASEreports.
1.FromtheinitialwindowoftheCASEapplication,clickLocalDatabase.
2.HighlightapatientnameandclickSelect.
3.SelectatestrecordandclickTransfertoMUSE.
a.OpenastressexercisetestintheMUSEEditor.
b.GototheClericaltabandclickthedownarrowintheReportTemplate
drop-downlist.
c.VerifythereporttemplatesloadedfromtheCASEsystemsarelistedthere.
TheSelectTestwindowopens.
4.ClickSave.
2059568-013JMUSE™v9
33
Page 34

CASEtoMUSECommunication
5.LogontotheMUSEsystem.
6.VerifythetestrecordisdisplayedintheMUSEsystemEditList.
VerifyingRequestsforData
UsethefollowingproceduretoverifythattheWebconnectionbetweentheCASE
systemandtheMUSEsystemfunctionsproperlyandsuccessfullysupportsdata
requests.
1.FromtheinitialwindowoftheCASEapplication,clickMUSEBrowser.
TheCASEsystemlaunchestheMicrosoftInternetExplorerandconnectstothe
MUSEhomepage.Ifyouareusingthedefaultconguration,youareprompted
toenterausernameandpassword.
2.Entertheusernameandpassword.
NOTE:
Ifyouchangethedefaultpasswordforthecase8000accountontheCASE
system,youmustalsochangeitontheMUSEsystem.
3.SubmitarequestqueryandviewtheresultstoverifythattheWebconnection
functionsproperly.
VerifyingtheRetrievalofPatientDemographics
UsethefollowingproceduretoverifythattheCASEsystemisabletoviewthepatient
listfromtheMUSEsystemdatabase.
1.FromtheinitialwindowoftheCASEapplication,clickNewTest.
Alistofavailablepatientsisdisplayed.
2.VerifythatthepatientlistisfromtheMUSEsystemdatabaseandnotthelocal
CASEsystemdatabase.
VerifyingtheRetrievalofOrders
UsethefollowingproceduretoverifythattheCASEsystemisabletoviewtheopen
ordersthatexistontheMUSEsystem.
1.FromtheinitialwindowoftheCASEapplication,clickNewTest.
2.FromtheNewTestscreen,clickOrderList.
3.VerifythatopenordersontheMUSEsystemaredisplayed.
34
MUSE™v92059568-013J
Page 35

CASEtoMUSECommunication
Troubleshooting
Usethefollowingtroubleshootingtipstohelpresolveissuesyoumayencounterwith
CASEsystemtoMUSEsystemcommunication.
TroubleshootingTips
Symptom
Unabletoviewpatientsor
ordersontheMUSEsystem
fromtheCASEsystem.
Unabletotransfertestsfrom
aCASEsystemtotheMUSE
system.
PossibleCauseRecommendations
TheMUSEUserNameand/or
MUSEPasswordspeciedin
theSetupforMUSEonthe
CASEsystemisincorrect.
TheMUSEAPI3service
isstoppedontheMUSE
applicationserver.
TheMUSEsystemisinan
AutoShutdownstate.
TheExerciseTestingData
Storageoptionisnotenabled
ontheMUSEsystem.
TheDirectoryUserName
and/orDirectoryPassword
denedintheStore
procedureforMUSEon
theCASEsystemisincorrect.
TheSharedDirectory
denedintheStore
procedureforMUSEon
theCASEsystemisincorrect.
TheDirectoryUserName
denedintheStore
procedureforMUSEon
theCASEsystemdoesnot
haveaccesstotheMUSE
share.
EnsuretheMUSEUserName
and/orMUSEPasswordare
enteredcorrectlyintheCASE
SystemConguration.
EnsuretheMUSEAPI3service
isstartedontheMUSE
applicationserver.
CanceltheAutoShutdown.
EnabletheExerciseTesting
DataStorageoptiononthe
MUSEsystem.
EnsuretheDirectoryUser
Nameand/orDirectory
Passwordareentered
correctlyintheCASESystem
Conguration.
EnsuretheSharedDirectory
isenteredcorrectlyinthe
CASESystemConguration.
EnsuretheDirectoryUser
NamedenedontheCASE
systemisamemberofthe
MUSEAcqUsersgroupon
theMUSEapplicationserver.
TheshareontheMUSE
applicationserverisnot
denedcorrectly.
2059568-013JMUSE™v9
Ensuretheacqfolder
(defaultisd:\muse\acq)is
correctlysharedontheMUSE
applicationserver.
35
Page 36

CASEtoMUSECommunication
36MUSE™v92059568-013J
Page 37

3
MARStoMUSECommunication
ThischapterdescribeshowtocongureMARSsystemsandMUSEsystemsinorderto
sendHolterreportsfromtheMARSsystemtotheMUSEv9system.
TheoryofOperation
MARSsystemtoMUSEsystemcommunicationallowsyoutotransferstoredMARS
reportsfromtheMARSsystemtotheMUSEsystemforviewing,editing,printing,and
storage.TheMARSsystemtransfersthecompleteHolterreport.AttheMUSEsystem,
youcanviewthestrippagesandeditpatientdemographics,diagnosisstatements,
andndings.
NOTE:
MARSstoredreportsdonotcontainfulldisclosureinformation.
ExampleNetworkDiagram
MUSEServices
MARSsystemtoMUSEsystemcommunicationusesthefollowingtwoMUSEservices:
MUSEGeneracqandMUSEFormatServices
ServiceDescription
MUSEGeneracqService
MUSEFormatService(s)
2059568-013JMUSE™v9
TheMUSEGeneracqServicehandlesacquisitionsfrom
othersystemsordevices.InMARSsystemtoMUSEsystem
communication,itsearchesthereportsshareontheMARS
systemforstoredreports(leswiththe*.mrsleextension)
andpullsthereportstotheMUSEsystemforprocessing.
TheMUSEFormatservice(s)launchtheMARSFormatter
program.TheMARSFormatterprogramformatstheoutput
tomatchtheformatfromtheMARSsystem.
37
Page 38

MARStoMUSECommunication
InformationTransmission
InMARSsystemtoMUSEsystemcommunication,aHoltertestisstoredinthereports
folderontheMARSsystemwitha.mrsextension.TheMUSEGeneracqservice
searchesthatfolderfor*.mrsles.
TestsarethennormalizedontheMUSEsystemandstoredinthedatabase.
TestsarealsoprocessedthroughtheMARSFormatterduringacquisitiontogeneratea
PDFthatcanbeviewedintheMUSEEditor.
NOTE:
WhiletestsarebeingprocessedbytheMARSFormatterduringacquisition,you
cannotvieworedittheminthetesteditor.Thefollowingmessageisdisplayed
whiletheHolterreportisbeingformattedafternormalization:
TheHolterrecordacquiredforpatientiscurrentlycheckedoutbyAdmin.
AcquisitionFlowChart
CustomerRequirements
Thecustomerisresponsibleforsupplyingappropriatenetworkconnectivity,including
nameresolution,betweentheMARSsystemsandtheMUSEapplicationserver.
ConguringMARStoMUSECommunication
AddingMARSsystemtoMUSEsystemcommunicationrequireschangestoboththe
MARSworkstationandtheMUSEapplicationserverandincludes:
•“ConguringtheMARSSystemNetworkSettings”onpage39
•“VerifyingtheMARSSoftwareVersion”onpage39
•“RecordingtheIPAddressorHostNameoftheMARSSystem”onpage39
•“VerifyingtheMARSReportsShare”onpage39
•“ActivatingtheMARStoMUSEOptionontheMARSSystem(s)”onpage40
•“SettingUptheSiteInformationontheMARSSystem(s)”onpage40
•“AddingtheHolterDataStorageOptiontotheMUSESystem”onpage41
•“AddingtheMARSSystemtotheMUSEGeneracqConguration”onpage42
•“InstallingtheMARSPrintFormatter”onpage43
•“Creatingsite.inifortheMARSPrintFormatter”onpage45
38MUSE™v92059568-013J
Page 39

•“Copyingtherusty.iniCongurationFilefromtheMARSSystemtotheMUSE
System”onpage46
•“SystemCheckout”onpage46
ConguringtheMARSSystemNetworkSettings
FortheMARSandMUSEsystemstocommunicate,theMARSsystemmustconform
tothenetworksettingsineffectattheinstallationsite.Contactthesite’ssystem
administratortoobtainrequiredinformationortoassistinconguringthesystem.
EnsurethatnetworkingisenabledonallMARSsystemsthatneedtocommunicate
withtheMUSEsystem.Additionally,conrmthatWindowsnetworksettingson
theMARSsystemsareappropriatelyconguredtocommunicatewiththeMUSE
applicationserver.
VerifyingtheMARSSoftwareVersion
TodetermineyourMARSsoftwareversion:
1.FromtheMARSsystemmenubar,selectHelp>About.
MARStoMUSECommunication
TheAboutwindowopens.
2.RecordthesoftwareversionlistedontheAboutwindow.
3.Findyourversioninthefollowingtabletodetermineyournextstep.
IfthecurrentMARS
softwareversionis...
5.xor6.x
7.0orlaterContinuetheMARStoMUSEcommunication
Then...
Upgradetoversion7.0orlater.
conguration.Goto“RecordingtheIPAddressorHost
NameoftheMARSSystem”onpage39.
RecordingtheIPAddressorHostNameoftheMARSSystem
RecordthecomputernameorIPaddressofeachMARSsystemyouwishtocongure
forcommunicationwiththeMUSEsystem.
IfyouwanttousetheIPaddress,itmustbestatic.
VerifyingtheMARSReportsShare
EnsurethattheMARSreportsfolder(defaultisc:\gemsit\reports)issharedandhasa
sharenameofReports.YoumustestablishandgiveleandsharepermissionsofFull
ControltotheuseraccountthatisconguredtostarttheMUSEGeneracqservice
ontheMUSEapplicationserver.
2059568-013JMUSE™v939
Page 40

MARStoMUSECommunication
ActivatingtheMARStoMUSEOptionontheMARSSystem(s)
UsethefollowingproceduretoactivatetheMARStoMUSEoptionontheMARSsystem.
1.LocatetheMARSsoftwareactivatorsheet.
2.FromtheMARSapplicationmenubar,selectSystem>SystemSetup>Software
Activators.
Alistoftasknamesopensalongwiththeircorrespondingmodesandstatuses.
3.IftheMARStoMUSEtasknameisdisabled,clickMARStoMUSE.
AlistofavailablemodesisdisplayedintheChangeModeTolistbox.
4.ClickActivate.
5.TypetheaccesscodefromtheactivatorsheetintotheEnterActivatorCode
Heretextbox.
6.ClickSaveChanges.
Ifthecodeisincorrectorincomplete,anerrormessageisdisplayed.
7.SelectOK.
8.ClickQuittoclosethewindow.
9.Repeatsteps2through8onallsystemsrequiringMARSsystemtoMUSEsystem
communication.
SettingUptheSiteInformationontheMARSSystem(s)
SitesetupisnecessarytotransferHolterdatatoaMUSEsystem.Thesitesand
locationsenteredineachMARSsystemmustmatchthesitesandlocationsusedon
theMUSEsystem.
NOTE:
ContacttheMUSEsystemownerforthesiteandlocationinformationyouneed
tousefortheMARSsystem.
RepeatthefollowingstepsforeachMARSsystemthatiscommunicatingwiththe
MUSEsystem.
1.FromtheMARSmainwindow,selectSystem>SystemSetup>Site.
TheSystem:SiteandLocationsSetupwindowopens.
2.SelecttheappropriateSite#.
3.EnterthecorrespondingSiteName.
4.SelecttheappropriateLocation#.
5.EnterthecorrespondingLocationName.
6.ClickAdd.
7.Repeatsteps2through6asnecessary.
8.ClickOK.
9.AttheChangesmade.Savethem?prompt,clickYestosaveyourchanges.
40MUSE™v92059568-013J
Page 41

MARStoMUSECommunication
AddingtheHolterDataStorageOptiontotheMUSESystem
UsethefollowingproceduretoaddtheHolterDataStorageoptiontotheMUSE
system.
NOTE:
ThisproceduremayonlybeperformedbyaqualiedGEHealthcareservice
representative.
1.LogontotheMUSEapplicationserverastheMUSEAdministratoruser.
2.PerformafullorpartialshutdownoftheMUSEsystemfollowingtheauto
shutdownproceduresdescribedinthe“SystemAdministration”chapterofthe
MUSEv9CardiologyInformationSystemServiceManual.
NOTE:
Ifyouarejustcheckingtoseeiftheoptionisalreadyenabled,ashutdownis
notrequired.If,however,youhavetoenableordisabletheoption,theMUSE
serviceswillberestartedandashutdownisrequired.
3.IntheWindowsControlPanel,selectProgramsandFeatures.
4.SelectMUSE9andclickChange.
TheWelcomewindowopens.
5.ChooseModifyandclickNext.
TheSelectFeaturewindowopens.
6.ContinuetoclickNextoneachwindowuntilyoureachtheChooseMUSE
Optionswindow.
7.VerifythatHolterDataStorageischecked.
Ifitisnot,checkitnow.
8.ClickNext.
TheMUSESerialNumberwindowopens.
9.IfyouaddedtheHolterDataStorageoptioninstep7,youneedtoenterthe
OptionsCongurationPassword.
NOTE:
OnlyqualiedGEHealthcareservicerepresentativeshaveaccesstothis
password.Thispasswordcannotbeprovidedtocustomers.
10.ClickNextuntilyourchangesareappliedandtheMaintenanceComplete
windowopens.
11.ClickFinish.
12.IfyouperformedafullorpartialshutdownoftheMUSEsysteminstep2,cancel
theshutdownfollowingtheproceduresdescribedinthe“SystemAdministration”
chapteroftheMUSEv9CardiologyInformationSystemServiceManual.
2059568-013JMUSE™v941
Page 42

MARStoMUSECommunication
AddingtheMARSSystemtotheMUSEGeneracq
Conguration
ToensuretheMUSEsystemcanlocateandcommunicatewiththeMARSsystem(s),
usethefollowingprocedurestoadd,modify,orremovepathstotheMARSsystem(s)
asneeded.
AddingthePathofEachMARSSystemtotheMUSEDatabase
CompletethefollowingproceduretoaddthepathofeachMARSsystemtotheMUSE
database.
1.FromwithintheMUSEapplication,gotoSetup.
2.SelectShareFolder.
ThelistofexistingShareFoldersisdisplayed.
TheShareFolderoptionwithinMUSEiswhereMUSEGeneracqfoldersandle
nameltersarecongured.
3.SelectAction>New.
TheShareFolderPropertiesdialogopens.
4.Completetheeldsasdescribedinthefollowingtable:
FieldTask
Entry
FileNameFilter
ProleNameSelectNone.
5.ClickOK.
EntertheUNCpathoftheMARSreportsshareontheMARS
system,forexample\\MARS001\reports.
Enter*.MRS.
ModifyinganExistingShareFolderEntry
ThefollowingprocedurecanbeusedtomodifyanexistingShareFolderentry.
1.FromwithintheMUSEapplication,gotoSetup.
2.SelectShareFolder.
Thelistofexistingsharefoldersisdisplayed.
3.Right-clickontheShareFolderyouwanttomodifyandchooseProperties.
TheShareFolderPropertieswindowopens.
4.Completetheeldsasdescribedinthefollowingtable:
FieldValue
Entry
FileNameFilter
ProleNameSelectNone.
5.ClickOK.
42MUSE™v92059568-013J
EntertheUNCpathoftheMARSreportsshareontheMARS
system,forexample\\MARS001\reports.
Enter*.MRS.
Page 43

RemovinganExistingShareFolderEntry
ThefollowingprocedurecanbeusedtodeleteanexistingShareFolderentryif
necessary.
1.FromwithintheMUSEapplication,gotoSetup.
2.SelectShareFolder.
ThelistofexistingShareFoldersisdisplayed.
3.Right-clickontheShareFolderyouwanttoremoveandchooseDelete.
InstallingtheMARSPrintFormatter
TheMARSPrintFormatterallowsHolterreportsthatareviewedandprintedfromthe
MUSEsystemtolookthesameaswhentheyareprintedfromtheMARSsystem.
StartingwiththeMUSEv9system,theMARSPrintFormatterisautomaticallyinstalled
whentheMUSEapplicationisinstalledontheMUSEapplicationserver.
TheversionoftheMARSPrintFormatterinstalledwiththeMUSEv9systemisMARS
v8.0SP4.TheprintformatterisbackwardcompatiblewithpreviousversionsofMARS,
howeveritmaynotbeforwardcompatible.IftheMARSPrintFormatterversionislater
thanv8.0SP4,youmayneedtoinstallittoensurecompatibilitywithMARSHolter
reportsfromMARSversionsnewerthanMARSv8.0SP4.
MARStoMUSECommunication
Thefollowingstepsdescribehowtoinstallanduninstalltheprintformattersoftware.
NOTE:
DonotperformthesestepsunlessyouareinstallingaversionoftheMARSPrint
FormatterthatislaterthanMARSv8.0SP4.
UninstallinganOlderMARSPrintFormatter
1.LogontotheMUSEapplicationserverasanadministrator.
2.GotoWindowsControlPanel>ProgramsandFeatures.
3.SelectMUSE-MARSandchooseUninstall.
TheMARSInstallShieldWizardisdisplayedandpromptsyoutoconrmthe
uninstall.
4.ClickOK.
5.AttheInstallShieldWizardCompletescreen,clickFinish.
6.Conrmthatthe<drive>:\gemsitfoldernolongerexists,where<drive>is
theistheletterofthedriveonwhichtheMARSPrintFormattersoftwarewas
previouslyinstalled.
Ifthe<drive>:\gemsitfolderstillexists,renameordeleteit.
InstallingaNewerMARSPrintFormatter
1.LogontotheMUSEapplicationserverasanadministrator.
2.Ensureallantivirussoftwareisturnedoffduringtheinstallation;itcanbeturned
onagainaftertheinstallationiscompleted.
2059568-013JMUSE™v9
43
Page 44

MARStoMUSECommunication
3.InsertormounttheMARSPrintFormatterinstallationmediaintotheoptical
drive.
TheInstallShieldWizardwindowmayopen.Ifnecessary,browsetotheoptical
driveandrunsetup.exe.
IfaUserAccountControlpromptappears,chooseYesorAllow.
4.ChooseyourdesiredinstallerlanguageandclickNext.
NOTE:
TheinstallerlanguageselectiononlysetsthelanguageoftheInstallShield
Wizard.
5.AttheWelcometotheInstallShieldWizardforMUSE-MARSscreen,clickNext.
6.AttheLicenseAgreementscreen,clickYes.
7.AttheDestinationDiskscreen,ensuretheC:driveisselectedandclickNext.
8.AttheReadytoInstalltheProgramscreen,clickInstall.
9.AttheInstallShieldWizardCompletescreen,clickFinish.
ReinstallingtheMARSPrintFormatterSoftwarethatComeswiththeMUSE
v9System
IfforsomereasonyouneedtoreinstalltheMARSFormattersoftwarethatcomeswith
theMUSEv9system,completethefollowingprocedure.
1.Performthestepsin“UninstallinganOlderMARSPrintFormatter”onpage43to
removetheexistingMARSPrintFormattersoftware.
2.Ensureallantivirussoftwareisturnedoffduringtheinstallation;itcanbeturned
onagainaftertheinstallationiscompleted.
3.InsertormounttheMUSEv9installationmediaintotheopticaldrive.
IftheMUSEv9installationoptionswindowopens,closeit.
4.Browsetothe\MUSESetup\MARS-MUSE_FormatterfolderontheMUSEv9
installationmedia.
5.Runsetup.exe.
IfaUserAccountControlpromptappears,chooseYesorAllow.
TheInstallShieldWizardopens.
6.ChooseyourdesiredinstallerlanguageandclickNext.
NOTE:
TheinstallerlanguageselectiononlysetsthelanguageoftheInstallShield
Wizard.
7.AttheWelcometotheInstallShieldWizardforMUSE-MARSscreen,clickNext.
8.AttheLicenseAgreementscreen,clickYes.
9.AttheDestinationDiskscreen,ensuretheCdriveisselectedandclickNext.
10.AttheReadytoInstalltheProgramscreen,clickInstall.
11.AttheInstallShieldWizardCompletescreen,clickFinish.
44
MUSE™v92059568-013J
Page 45

Creatingsite.inifortheMARSPrintFormatter
Usethefollowinginstructionstocreatethesite.inileontheMUSEApplicationserver.
Thesite.inileisusedbytheMARSPrintFormatterforoutputtingtheMUSEsiteand
locationnameonMARSformattedreports.Ifthesestepsarenotperformed,theMARS
formattedreportsviewedandprintedfromtheMUSEsystemwillshowUnknownfor
sitenamesandlocations.
NOTE:
ThisprocesswillneedtoberepeatedeachtimeaMUSEsiteorlocationthatis
usedbyaMARSformattedHolterreportisaddedtoMUSEsystem.
1.LogontotheMUSEapplicationserverastheMUSEAdministratoruser.
2.InserttheMUSEv9installationmediaintotheopticaldriveofthesystem.
IfanyAutorunorAutoPlayscreensappear,closeorcancelthem.
3.BrowsetheopticaldriveinWindowsExplorerandperformoneofthefollowing:
•IftheMUSEv9ApplicationandSupportDVDisinserted,navigatetothe
\MUSESupport\MARSSiteINIUpdatefolder.
MARStoMUSECommunication
•IftheMUSEv9SupportISOisbeingused,navigatetothe\MARSSiteINI
Updatefolder.
4.CopySiteIniUpdate.exefromtheMUSEv9supportmediatothelocationwhere
theMUSEapplicationisinstalled.ThedefaultlocationoftheMUSEapplicationis:
C:\ProgramFiles(x86)\MUSE.
5.RunSiteiniupdate.exefromtheMUSEinstallationfolderusingRunas
Administrator.
WhenSiteiniupdate.exeruns,acommandpromptwindowwillopenandclose.
Therewillbenoon-screenmessagesifitcompletessuccessfully.
6.VerifytheC:\gemsit\var\MarsNT\system\site.iniiscreatedorupdated.
ThecontentsofthelewillreecttheMUSEsiteandlocationcongurationof
theMUSEsystem.Seethefollowingsite.inilebasics.
site.iniFileBasics
Thesite.inileismadeupofatleastthreesections.
•The[SiteList]sectionlistsallofthesitenumbers.
•Eachsitewillhavea[Site]sectionthatliststhelocationnumbersandthesite
name.Iftherearemultiplesites,therewillbemultiple[Site]sections.
•Eachsiteandlocationcombinationwillhaveitsown[SiteLocation]section
containingthelocationname.Iftherearemultiplelocationsforeachsite,there
willbemultiple[SiteLocation]sections.
Followingisanexampleofasite.inile:
[SiteList]
SiteNumbers=12
[Site1]
LocationNumbers=12
SiteName="MemorialHospital"
[Site1Location1]
2059568-013JMUSE™v9
45
Page 46

MARStoMUSECommunication
LocationName="HolterScanning"
[Site1Location2]
LocationName="ECGDepartment"
[Site2]
LocationNumbers=12
SiteName="GeneralHospital"
[Site2Location1]
LocationName="Mary’sOfce"
[Site2Location2]
LocationName="John’sOfce"
Copyingtherusty.iniCongurationFilefromtheMARS
SystemtotheMUSESystem
TomaintainproperMARSHolterreportformattingontheMUSEv9system,youmust
copytheRusty.inicongurationlefromtheprimaryMARSServer(orstandalone
workstation)totheMUSEsystem.
ThisleislocatedinthefollowingdefaultinstallationfolderontheMARSsystem:
C:\gemsit\var\MarsNT\system
NOTE:
RepeatthisstepifchangesaremadetotheMARSsystemconguration.
SystemCheckout
ToensurethattheMARSsystemtoMUSEsystemcommunicationisfunctioning
properly,youneedtosaveareportoneachMARSworkstation,sendthosereportsto
theMUSEapplicationserver,andthenretrieveandprintthosereportsontheMUSE
applicationserver.
SavingaReportontheMARSSystem
CompletethefollowingstepstosaveareportontheMARSsystem.Thesestepsare
performedontheMARSsystem.
1.ClickthePatientSelecticon.
2.InthePatientSelectwindow,selectHolterfromtheDataTypelist.
3.Selectapatientinthelist.
4.ClickthePatientInformationicon.
5.VerifythattheSiteandLocationinformationislledinforthispatient.
6.ClicktheReportReviewicon.
7.ClickSaveReport.
Thefollowingmessageisdisplayed:Reportsuccessfullystored.
8.ClickOK.
9.ClickClose.
TheReportReviewtoolcloses.
46MUSE™v92059568-013J
Page 47

SendingaMARSHolterReporttotheMUSESystem
CompletethefollowingstepstosendaMARSreporttotheMUSEsystem.Thesesteps
areperformedontheMARSsystem.
1.ClickthePatientSelecticon.
2.InthePatientSelectwindow,selectStoredReportsfromtheDataTypelist.
3.Selectthestoredreportyousavedusingthestepsin“SavingaReportonthe
MARSSystem”onpage46.
4.ClickTools.
5.ClickStoretoMUSE.
Thefollowingmessageisdisplayed:Youhaveselected1le(s)forMUSE
storage.Areyousureyouwanttostoretheselectedle(s)toMUSE?
6.ClickYes.
Thefollowingmessageisdisplayed:1report(s)queuedforstoragetoMUSE.
7.ClickOK.
MARStoMUSECommunication
8.Afterabriefdelay,verifythatthereportislistedasStoredtoMUSE.
NOTE:
IftheDeleteReportsAfterTransfertoMUSEoptionisenabledontheMARS
system,thepatientreportwillautomaticallyberemovedfromtheStored
Reportslist.
9.ClickClose.
ViewingaMARSHolterReportintheMUSEEditor
WiththeMUSEv9systemthereisaReportPreviewtabavailablewithintheMUSE
Editor.ThisReportPreviewdisplaystheHolterReportasitwouldlookifprintedfrom
theMARSsystem.
1.OpentheHolterreportstoredtotheMUSEsystemusingthestepsin“Sendinga
MARSHolterReporttotheMUSESystem”onpage47.
2.VerifytheReportPreviewtabdisplaystheMARSHolterreport.
NOTE:
BeawareofthefollowingwhenviewingHolterreportsontheMUSEsystem:
•CreatedreportsarenotdisplayedintheReportPreviewtabuntilthe
associatedelectronicdocumentdataismanuallyimported.
•HolterreportsacquiredbyearlierversionsoftheMUSEsystemcannotbe
previeweduntiltheyhavebeenopenedintheMUSEEditoratleastonce.
ThefollowingmessagedisplaysthersttimeyouopenaHolterreportin
theMUSEsystemthatdoesnothaveaReportPreview:TheFullReportis
currentlynotpresentforthistest,butwillbeavailablethenexttime
thestudyisopenedintheeditor.
2059568-013JMUSE™v9
47
Page 48

MARStoMUSECommunication
PrintingMARSHolterReportsfromtheMUSESystem
1.SelecttheHolterreportfromtheMUSEEditList.
2.Onthetoolbar,clickPrintTest.
3.FromtheAvailablePrinterslist,selectaLaserprinterorPDFFolderdevice.
4.ClickOK.
5.Verifythatthereportwasgenerated,thenconrmthatitisformattedproperly
basedontheconguration:
•IftheMARSFormatHolterspecicformatsettingisenabled,thereportshould
looklikeaMARSreport.
•IftheMARSFormatHolterspecicformatsettingisnotenabled,thereport
shouldlooklikeaMUSEreport.
Troubleshooting
Usethefollowingtroubleshootingtipstohelpresolveissuesyoumayencounterwith
MARSsystemtoMUSEsystemcommunication.
TroubleshootingTips
Symptom
Unabletostoretestsfrom
theMARSsystemtothe
MUSEsystem.
PossibleCauseRecommendations
TheHolterDataStorage
optionisnotenabledonthe
MUSEsystem.
TheMARStoMUSEoptionis
notactivatedontheMARS
system.
Theuseraccountcongured
tostarttheMUSEGeneracq
serviceontheMUSE
applicationserverdoes
nothaveaccesstothe
reportsshareontheMARS
system.
Thereportsshareonthe
MARSsystemisnotdened
correctly.
TheShareFolderisnotset
upcorrectlyfortheMARS
systemintheMUSESystem
Setup.
EnabletheHolterData
StorageoptionontheMUSE
system.
ActivatetheMARStoMUSE
optionontheMARSsystem.
TheMARStoMUSEoption
mustbeactivatedoneach
MARSsystemthatneeds
tostoredatatotheMUSE
system.
Ensuretheuseraccount
conguredtostarttheMUSE
Generacqserviceonthe
MUSEapplicationserverhas
accesstothereportsshare
ontheMARSsystem.
Ensurethereportsshareon
theMARSsystemissetup
correctly.
EnsuretheShareFolder
setupinMUSESystemSetup
iscorrectlydened.
TheSiteand/orLocation
onMARSreportsdisplayas
Unknownwhenviewedon
theMUSEsystem.
48MUSE™v92059568-013J
Thesite.inihasnotbeen
correctlyconguredonthe
MUSEsystemfortheMARS
PrintFormatter.
Ensurethesite.iniexistsand
iscorrectlydenedonthe
MUSEsystem.
Page 49

TroubleshootingTips(cont'd.)
MARStoMUSECommunication
Symptom
Thefollowingmessage
appearswhenattempting
toviewarecentlyacquired
HolterreportintheMUSE
application:TheHolter
recordacquiredforpatient
iscurrentlycheckedoutby
Admin.
Thefollowingmessage
appearswhenattemptingto
viewaHolterreportthatwas
acquiredbeforetheMUSE
systemwasupgradedtov9:
TheFullReportiscurrently
notpresentforthistest,but
willbeavailablethenext
timethestudyisopenedin
theeditor.
PossibleCauseRecommendations
TheHolterreportisbeing
processedbytheMARSPrint
Formatter.
TheHolterreporthasnot
beenprocessedbytheMARS
PrintFormatteryet.
Waitforthetesttobe
processedbytheMUSE
systemandMARSPrint
Formatterandthenopenthe
testintheMUSEapplication.
Waitforthetesttobe
processedbytheMARS
PrintFormatterandthen
openitagainintheMUSE
application.
2059568-013JMUSE™v949
Page 50

MARStoMUSECommunication
50MUSE™v92059568-013J
Page 51

MUSEMonitoringGateway
ThischapterdescribeshowtoInstalltheMUSEMonitoringGatewaytoallowbedside
monitorstotransmitECGteststotheMUSEv9system.
TheoryofOperation
TheMUSEMonitoringGatewaycomputerallowsbedsidemonitorsonareal-time
monitoringnetworktotransmitECGstotheMUSEsystem.Thisisfacilitatedbythe
MUSEMonitoringGatewayhavingtwoNetworkInterfaceCards(NICs),eachona
differentnetwork.InmanycasesthereisalsoaGEHealthcareconguredrouter
whichwillallowsomebedsidemonitorstoretrieveECGsbackfromMUSEsystem.
WhenanECGistransmittedfromthebedsidemonitor,thetestisreceivedbytheMUSE
MonitoringGateway.TheMUSEapplicationthenretrievesthetestfromtheMUSE
MonitoringGatewayandacquiresitintotheMUSEdatabase.
4
ExampleNetworkDiagram—MUSEMonitoringGatewayontheMissionControland
InformationExchangeNetworks
2059568-013JMUSE™v951
Page 52

MUSEMonitoringGateway
ExampleNetworkDiagram—MUSEMonitoringGatewayontheUnityandHospital
Networks
InformationTransmission
InMUSEMonitoringGatewaytoMUSEsystemcommunication,anECGistransmitted
fromthebedsidemonitortotheMUSEMonitoringGateway.TheMUSEGeneracq
servicesearchestheACQMONshareontheMUSEMonitoringGatewayforlesand
pullsthemtotheMUSEsystemforprocessing.Testsarethennormalizedonthe
MUSEsystemandstoredinthedatabase.
AcquisitionFlowChart
InstallingtheMUSEMonitoringGateway
FollowtheinstructionsinthischaptertocompletetheinstallationoftheMUSE
MonitoringGateway.Thischaptercovers:
•“PreparingfortheInstallationoftheMUSEMonitoringGateway”onpage53
•“VerifyingandConguringNetworkConnections”onpage53
•“FirewallConsiderations”onpage54
•“CreatingaShareontheMUSEMonitoringGateway”onpage54
•“InstallingtheMUSEMonitoringGatewaySoftware”onpage55
•“ConguringtheMUSEMonitoringGatewaySoftware”onpage55
•“ConguringtheMUSEApplicationServer”onpage56
•“ConguringBedsideMonitors”onpage57
•“SystemCheckout”onpage57
•“Troubleshooting”onpage58
•“UninstallingMUSEMonitoringGatewayv1.1”onpage59
NOTE:
TheMUSEMonitoringGatewaysoftwareversion(v1.1)hasnotchangedbetween
theMUSEv8andMUSEv9releases.Therefore,thisprocedureshouldnotneed
tobeperformedforMUSEv8toMUSEv9systemupgrades,iftheMonitoring
Gatewayexistedpriortotheupgrade.
52MUSE™v92059568-013J
Page 53

PreparingfortheInstallationoftheMUSEMonitoring
Gateway
Inordertoensureasuccessfulinstallation,thepersoninstallinganewMUSE
MonitoringGatewaysystemmustbeawareofthefollowingitemspriortobeginning
theinstallation.
MUSEMonitoringGateway
Item
IsthemonitoringnetworkconguredbyGE
Healthcareorthecustomer?
Routerconguration,IPAddressesofthe
NICs,NATAddressoftheNIC,andphysical
networkconnections
Action
WorkwiththeProjectManagerand/orthe
customertodeterminethis.
IfitisaGEHealthcarecongurednetwork,
theGEHealthcareND&I(NetworkDesignand
Implementation)teamcongureseverything.
TheGEHealthcareND&Iengineerperforming
thecongurationprovidesinformationtothe
FE(FieldEngineer)installerandtotheProject
Manager(ifapplicable).
Ifitisacustomercongurednetwork,the
FEinstallermustworkwiththecustomer
todetermineappropriateconguration
information.
Ensureyoumeetthefollowingrequirementsbeforecontinuingwiththeinstallation:
System
MonitoringGateway
Requirements
•Windows7Professional,Enterprise,orUltimate
NOTE:
Only32-bitEnglish-languageversionsofWindows7are
supported.
•C:drivepartition
•Twonetworkinterfacecards(NICs)
•NopreviousMUSEclientinstallations(thatis,thesystemshould
havenoleftovertracesofapreviousMUSEinstallation,suchas
MUSEspecicentriesinthewin.inile)
MUSEApplication
Server
AuserstartingtheMUSEGeneracqserviceontheMUSEApplication
isabletoconnecttosharesontheMonitoringGateway.
VerifyingandConguringNetworkConnections
CongureandverifythattheMonitoringGatewayhastwoNetworkInterfaceCards
(NICs)withthefollowingnetworkconnections:
TypicalNewInstallationsonGEHealthcare-suppliedMonitoringNetworks
NetworkIPAddressSubnetMask
GEHealthcareMonitoring
MC(MissionCritical)Network
GEHealthcareIX(Information
Exchange)Network
2059568-013JMUSE™v9
172.16.0.1255.255.0.0
AppropriateIPaddressandappropriatesubnetmask
53
Page 54

MUSEMonitoringGateway
LegacyorCustomer-installedMonitoringNetworks
NetworkIPAddressSubnetMask
UnityNetwork
126.8.8.1255.0.0.0
LANonwhichtheMUSE
applicationserverresides
NOTE:
Thenetworksettingslistedintheprevioustablesaredefaults.TheIPaddressand
subnetmaskmaybedifferentforcustomercongurednetworks.
AfterconguringtheNICsitisagoodpracticetorenametheLocalNetwork
ConnectionsinWindowstomatchthenetworkstowhichtheyareconnected.
AppropriateIPaddressandappropriatesubnetmask
FirewallConsiderations
IftheMUSEMonitoringGatewayhasarewallinplace,youmustmakethefollowing
exceptions:
•TFTP(UDPport69):allowscommunicationfromtheBedsideMonitorstotheMUSE
MonitoringGateway.
•WindowsFileSharing:allowstheMUSEsystemtoaccessthefoldershareonthe
MUSEMonitoringGateway.
CreatingaShareontheMUSEMonitoringGateway
UsethefollowingproceduretosetupafolderontheMonitoringGatewaythatthe
MUSEapplicationservercanaccess.
1.OntheMUSEapplicationserver,determinewhatuseraccountisrunningthe
MUSEGeneracqservice.
Typically,thisistheMUSEBackgroundaccount.However,forsecurityreasons
thecustomermaybeusingadifferentaccount.
2.CreateaC:\ACQMONfolderontheMonitoringGatewaysystem.
3.Sharethefoldercreatedinstep2,givingtheaccountdeterminedinstep1,
fullpermissionstotheshare.
IfalocalaccountisbeingusedtostarttheMUSEGeneracqservice,create
anidenticaluseraccountlocallyontheMonitoringGateway,usingthesame
password.
ThekeyrequirementhereisthattheaccountconguredtostarttheMUSEGeneracq
serviceontheMUSEapplicationservermusthavefullaccesstotheACQMONshare
ontheMUSEMonitoringGateway.
NOTE:
DonotchangetheaccountusedtostarttheMUSEGeneracqserviceontheMUSE
applicationserver.ChangingthisaccountcouldimpactotherMUSEGeneracq
ShareFoldercongurations.
54
MUSE™v92059568-013J
Page 55

MUSEMonitoringGateway
InstallingtheMUSEMonitoringGatewaySoftware
1.InserttheMUSEMonitoringGatewayCDintotheopticaldrive.
2.Browsetotheopticaldriveandrunsetup.exe.
3.Followthepromptsuntiltheinstallercompletes.
IfaUserAccountControlprompttorunMuseGatewaySetup.msiisdisplayed,
chooseYesorAllow.
4.WhentheInstallationCompletescreendisplays,clickClose.
5.VerifythatthefollowingWindowsservicesareinstalledandstarted:
•MUSEGatewayRWHAT
•MUSEGatewayTFTP
6.Right-clickontheMUSEGatewayRWHATserviceandchooseProperties.The
MUSEGatewayRWHATPropertieswindowopens.
7.SelecttheLogOntab.
8.VerifytheAllowservicetointeractwithdesktopboxischecked.Ifitisnot
checked,checkit.
9.ClickOK.
10.IfyouenabledtheAllowservicetointeractwithdesktopoptioninstep8,
restarttheMUSEGatewayRWHATservice.
ConguringtheMUSEMonitoringGatewaySoftware
IftheIPaddressusedbytheNetworkInterfaceCard(NIC)connectedtothe
MC/Unitynetworkisnotthedefaultof126.8.8.1,theMUSEGatewayRWHATservice
congurationneedstobemodied.PerformthefollowingstepstomodifytheMUSE
GatewayRWHATserviceconguration.
1.UsingNotepad,openc:\programles\monitorgateway\monitorgateway.ini.
NOTE:
YoumayneedtouseRunasAdministratorwhenopeningNotepadto
ensureyoucansavetheleaftermakingchanges.
2.DeletethesemicolonintheRWhat=line.
3.ChangetheIPaddressintheRWhat=linetotheIPaddressoftheMonitoring
GatewayNICconnectedtotheMC/Unitynetwork.
Forexample,iftheIPaddressoftheMonitoringGatewayNICconnectedtothe
MC/Unitynetworkis172.16.0.1,thelineshouldbeRWhat=172.16.0.1.
4.DeletethesemicolonintheRWhat_Subnet=line.
5.ChangethesubnetaddressintheRWhat_Subnet=linetothesubnetaddressof
theMonitoringGatewayNICconnectedtotheMC/Unitynetwork.
Forexample,ifthesubnetaddressoftheMonitoringGatewayNIC
connectedtotheMC/Unitynetworkis255.255.0.0,thelineshouldbe
RWhat_Subnet=255.255.0.0.
6.SaveyourchangesandexitNotepad.
2059568-013JMUSE™v9
55
Page 56

MUSEMonitoringGateway
7.RestarttheMonitoringGatewaycomputer.
8.VerifythatbothMUSEGatewayRWHATandMUSEGatewayTFTPservices
arestarted.
ConguringtheMUSEApplicationServer
ToensuretheMUSEsystemcanlocateandcommunicatewiththeMUSEMonitoring
Gatewaysystems,usethefollowingprocedurestoadd,modify,orremovethepaths
totheMUSEMonitoringGatewaysystem(s)asneeded.
AddingthePathofEachMUSEMonitoringGatewaySystemtotheMUSE
Database
CompletethefollowingproceduretoaddthepathofeachMUSEMonitoringGateway
systemtotheMUSEdatabase.
1.FromwithintheMUSEapplication,gotoSetup.
2.SelectShareFolder.
ThelistofexistingShareFoldersisdisplayed.
TheShareFolderoptionwithinMUSEiswhereMUSEGeneracqfoldersandle
nameltersarecongured.
3.SelectAction>New.
TheShareFolderPropertieswindowopens.
4.Completetheeldsasdescribedinthefollowingtable:
FieldTask
Entry
FileNameFilter
ProleNameSelectNone.
5.ClickOK.
EntertheUNCpathoftheACQMONshareontheMUSE
MonitoringGatewaysystem,forexample\\MMG001\ACQMON.
Enter*.*.
ModifyinganExistingShareFolderEntry
ThefollowingprocedurecanbeusedtomodifyanexistingShareFolderentryif
necessary.
1.FromwithintheMUSEapplication,gotoSetup.
2.SelectShareFolder.
Thelistofexistingsharefoldersisdisplayed.
3.Right-clickonthesharefolderyouwanttomodifyandchooseProperties.
TheShareFolderPropertieswindowopens.
56MUSE™v92059568-013J
Page 57

4.Completetheeldsasdescribedinthefollowingtable:
FieldTask
Entry
FileNameFilter
ProleNameSelectNone.
5.ClickOK.
EntertheUNCpathoftheACQMONshareontheMUSE
MonitoringGatewaysystem,forexample\\MMG001\ACQMON.
Enter*.*.
RemovinganExistingShareFolderEntry
ThefollowingprocedurecanbeusedtodeleteanexistingShareFolderentryif
necessary.
1.FromwithintheMUSEapplication,gotoSetup.
2.SelectShareFolder.
Thelistofexistingsharefoldersisdisplayed.
MUSEMonitoringGateway
3.Right-clickonthesharefolderyouwanttoremoveandchooseDelete.
ConguringBedsideMonitors
Thebedsidemonitorsmayrequireadditionalcongurationoroptionstosuccessfully
transmitECGstotheMUSEsystemthroughtheMUSEMonitoringGateway.Usethe
followinginformationasahigh-levelreference,andrefertotheappropriatebedside
monitorservicedocumentationforcompletedetails.
•Allbedsidemonitors:SiteandLocationneedtobedenedappropriatelytoensure
thetestsareassociatedwiththecorrectMUSEsiteandlocation.
•Bx50Monitors:InadditiontoSiteandLocationconguration,Bx50monitors
needtohaveaMUSEWebURLconguredinWebmin.AvalidURLwouldbe
inthefollowingformat:http://<ip_or_name_of_muse_app_server>:<port>/
musescripts/museweb.dll
SystemCheckout
ThisvericationisonlyfortransmittingECGsfrombedsidemonitorstotheMUSE
systemviatheMUSEMonitoringGateway.
NOTE:
ThisproceduredoesnottesttheretrievalofECGsfromtheMUSEsystemviaMUSE
Web.ThecongurationandvericationofMUSEWebareoutsidethescopeof
thisdocument.
1.Fromabedsidemonitor,transmita12SLECGtotheMUSEsystem.
2.IntheMUSEsystem,selectSystemStatus.
3.SelecttheAcquisitionLog.
4.ConrmthattheECGwasacquiredsuccessfullyintotheMUSEsystemby
locatingthePID/NameofthetransmittedECGintheAcquisitionLog.
2059568-013JMUSE™v9
57
Page 58

MUSEMonitoringGateway
Troubleshooting
UsethefollowingtroubleshootingtipsiftheMonitoringGatewaywasinstalledand
conguredcorrectly,butthebedsidemonitorisunabletosenddatatoit.
TroubleshootingTips
Symptom
MonitoringGateway
isnotavailableonthe
bedsidemonitor.
MonitoringGateway
isavailablefromthe
bedsidemonitorbut
cannotreceivedata.
ECGtestssentfrom
bedsidemonitorsare
notshowingupinthe
MUSEsystem.
Condition
Somebedsidemonitors
maynotseeanew
MonitoringGateway
onthenetwork.This
isespeciallycommon
whentheMonitoring
Gatewayhasrecently
changedIPaddresses.
TheMonitoring
Gatewaysystemhasa
rewallinplace.
Theuseraccount
conguredtostart
theMUSEGeneracq
serviceontheMUSE
applicationserverdoes
nothaveaccesstothe
ACQMONshareonthe
MonitoringGateway
system.
Action
Restartingthebedsidemonitorshould
resolvethisissue.
Youneedtoaddthefollowing
exceptionstotherewall:
•TFTP(UDPport69):
Allowscommunicationfromthe
monitorstotheMonitoringGateway.
•WindowsFileSharing:
Allowscommunicationfromthe
MUSEsystemtotheMonitoring
Gateway.Dependingonhowthe
routerandnetworksarecongured,
youmayneedtospecifytheMUSEIP
addressintheexceptions.
Ensuretheuseraccountconguredto
starttheMUSEGeneracqserviceonthe
MUSEapplicationserverhasaccessto
theACQMONshareontheMonitoring
Gatewaysystem.
TheACQMONshare
ontheMonitoring
Gatewaysystemisnot
denedcorrectly.
TheShareFolderisnot
setupcorrectlyforthe
MonitoringGateway
systemintheMUSE
SystemSetup.
TheECGtesthasan
invalidsite.
58MUSE™v92059568-013J
EnsuretheACQMONshareonthe
MonitoringGatewaysystemissetup
correctly.
EnsuretheShareFoldersetupinMUSE
SystemSetupiscorrectlydened.
Ensurethebedsidemonitoriscongured
withavalidMUSEsitenumber.Check
theMUSEDiscardedDataListforthe
missingtest(s).
Page 59

UninstallingMUSEMonitoringGatewayv1.1
UsethefollowingproceduretouninstallMonitoringGatewayv1.1shoulditneedto
beuninstalledforanyreason.
1.Saveacopyofthec:\programles\monitorgateway\monitorgateway.inile
toadifferentlocationforfuturereference.
2.GotoControlPanel>ProgramsandFeatures.
3.SelectMonitoringGateway1.1andchooseUninstall.
4.ChooseYeswhenyoureceivethefollowingprompt:“Areyousureyouwantto
uninstallMUSEMonitoringGateway1.1?”
IfyoureceiveapromptfortheUserAccountControltoconrmtheremovalof
theMonitoringGateway1.1,chooseYesorAllow.
ThisremovestheMUSEGatewayRWHATandMUSEGatewayTFTPservices
anddeletestheC:\ProgramFiles\monitorgatewayfolder.
MUSEMonitoringGateway
2059568-013JMUSE™v959
Page 60

MUSEMonitoringGateway
60MUSE™v92059568-013J
Page 61

MACECGSystemstoMUSE
Communication
ThischapterdescribeshowtocongureMACECGsystemstosendteststoMUSEv9
systems.
TheoryofOperation
MACECGtoMUSEcommunicationallowsyoutotransfertestsfromMACECGsystems
totheMUSEsystemforviewing,editing,printing,andstorage.ItalsoallowsMACECG
systemstoreceiveordersand/orpatientdemographicsinformationfromtheMUSE
system.Additionally,remotequeryallowsaMACECGsystemtoqueryforandreceive
testsfromtheMUSEsystem,whilereversetransmissionallowsMUSEtosenddata
toaMACECGsystem.
5
ExampleNetworkDiagram
2059568-013JMUSE™v961
Page 62

MACECGSystemstoMUSECommunication
MACECGdevicescancommunicatewiththeMUSEsystemviadirectserialcable,
modem,wirelessnetwork,orlocalareanetwork.
DataFlowBetweenMACECGSystemsandtheMUSESystem
CustomerRequirements
TransmissionFlowChartOrderDownloadFlowChart
RemoteQueryFlowChartReverseTransmissionFlowChart
Thecustomerisresponsibleforappropriateserial,telephony,ornetworkconnectivity
betweentheMACECGsystemsandtheMUSEsystemsthathavebeenconguredfor
MACsystemtoMUSEsystemcommunication.
ConguringMACECGtoMUSECommunication
UsethefollowingsectionsforconguringMACECGtoMUSEsystemcommunication.
•DirectSerialCableandModem
See“SettingUpModems”onpage63fordetailsonsettingupmodemsintheMUSE
system.
•WirelessNetworkviaCSI
See“SettingUpModems”onpage63forinformationonsettinguptheCSI
Networktypeofmodemrequiredforwirelessnetworkcommunication.Formore
detailedinformationonconguringMACECGsystemstocommunicatewithMUSE
wirelessly,refertotheMobileLinkInstallationManual.
•LocalAreaNetworkviaCSI
See“SettingUpModems”onpage63forinformationonsettinguptheCSINetwork
typeofmodemrequiredforlocalareanetworkcommunication.Fordetailed
informationonconguringMACECGsystemstocommunicatewithMUSEvialocal
areanetwork,refertotheLANOptionforMACInstallationandTroubleshooting
Guide.
•WirelessorLocalAreaNetworkviaDCP(DCARCommunicationProtocol)
SomeGEHealthcareMACECGsystemscanuseDCPtocommunicatewiththeMUSE
system.See“SettingUpDCPInboundCommunication”onpage69forinformation
onsettinguptheMUSEDCPServer.FordetailedinformationonconguringMAC
ECGsystemstocommunicatewiththeMUSEsystemvialocalareanetwork,referto
theLANOptionforMACInstallationandTroubleshootingGuide.
62MUSE™v92059568-013J
Page 63

SettingUpModems
MACECGSystemstoMUSECommunication
TheMUSEsystemusesthefollowingmodemtypestosendandreceivedata:
MUSEModemTypes
ModemType
FAXModemSupportsoutgoingfaxtransmission.Requiresphysical
CSIModemSupportsPlainOldTelephoneService(POTS)modem
CSIDirect
CSINetwork
TosupportanyoftheavailablemodemtypesonMUSEv9,theMUSEModemfeature
mustbeinstalledandtheMUSEModemservicemustberunning.
NOTE:
DuringanupgradefromMUSEv7.x,allCSIWirelessmodemsareconvertedto
CSINetworkmodems.IfyouseeaCSIWirelessentryintheModemSetuplist,
themodemserverforthatmodemwasnotupgradedorismissing.Refertothe
MUSEv9CardiologySystemInstallationandUpgradeManualforinformationon
performingawirelessmodemmigrationduringaMUSEv7upgrade.
Description
modem.
communicationtocompatibleMACsystems.Requires
physicalmodem.
Supportsdirectserialcablecommunicationwith
compatibleMACsystems.Requiresphysicalserialcable.
Supportswirelessand/orLANcartconnectionswith
compatibleMACsystems.
AddingtheWireless/LANCommunicationOptiontothe
MUSESystem
TheWireless/LANCommunicationoptionisrequiredforWireless/LANCSI
Communication.UsethefollowingproceduretoaddtheWireless/LANCommunication
optiontotheMUSEsystem.ThisproceduremayonlybeperformedbyaqualiedGE
Healthcareservicerepresentative.
1.LogontotheMUSEapplicationserverastheMUSEAdministratoruser.
2.PerformafullorpartialshutdownoftheMUSEsystemfollowingtheauto
shutdownproceduresdescribedinthe“SystemAdministration”chapterofthe
MUSEv9CardiologyInformationSystemServiceManual.
NOTE:
Ifyouarejustcheckingtoseeiftheoptionisalreadyenabled,ashutdownis
notrequired.If,however,youhavetoenableordisabletheoption,theMUSE
serviceswillberestartedandashutdownisrequired.
3.GotoControlPanel>ProgramsandFeatures.
4.SelectMUSE9andclickChange.
TheWelcomewindowopens.
5.ChooseModifyandclickNext.
TheSelectFeatureswindowopens.
2059568-013JMUSE™v963
Page 64

MACECGSystemstoMUSECommunication
6.ContinuetoclickNextoneachwindowuntilyoureachtheChooseMUSE
Optionswindow.
7.VerifythatWireless/LANCommunicationischecked.
Ifitisnot,checkitnow.
8.ClickNext.
TheMUSESerialNumberwindowopens.
9.IfyouaddedtheWireless/LANCommunicationoptioninstep7,enterthe
OptionsCongurationPassword.
NOTE:
OnlyqualiedGEHealthcareservicerepresentativeshaveaccesstothis
password.Thispasswordcannotbeprovidedtocustomers.
10.ClickNextuntilyourchangesareappliedandtheMaintenanceComplete
windowopens.
11.ClickFinish.
12.IfyouperformedafullorpartialshutdownoftheMUSEsysteminstep2,cancel
theshutdownfollowingtheproceduresdescribedinthe“SystemAdministration”
chapteroftheMUSEv9CardiologyInformationSystemServiceManual.
Verifying/InstallingtheMUSEModemService
TheMUSEModemfeatureandservicearetypicallyinstalledduringtheinitial
installationoftheMUSEclient.Usethefollowinginstructionstoverifywhetherthe
MUSEModemfeatureandserviceareinstalledontheMUSEsystemandtoinstall
themiftheyarenot.ThesestepscanbeperformedontheMUSEapplicationserveror
MUSEworkstation.
1.LogontotheMUSEsystemyouwanttocongureastheMUSEAdministrator
user.
2.GotoControlPanel>ProgramsandFeatures.
3.SelectMUSE9andclickChange.
TheWelcomewindowopens.
4.ChooseModifyandclickNext.
TheSelectFeatureswindowopens.
5.IntheSelectFeatureswindow,performoneofthefollowing:
•IfthecheckboxnexttoModemisalreadychecked,clickCancel.TheModem
featureisalreadyinstalled.
Proceedto“SettingUpaModemDevice”onpage67".
•IfthecheckboxnexttoModemisnotchecked,checktheboxandclickNext.
NOTE:
OntheMUSEapplicationservertheModemfeaturewillbeasub-itemof
Server>Services.OntheMUSEclientworkstation,theModemfeature
willbeasub-itemofMUSEClient.
64MUSE™v92059568-013J
Page 65

MACECGSystemstoMUSECommunication
6.ClickNextuntiltheSelectDatabaseServerwindowopens.
7.IntheDatabaseServereld,clickBrowseortypethenameoftheSQLServer
wheretheMUSEdatabasesareinstalled.Besuretoincludetheinstancename
ifanon-defaultSQLServerinstanceisbeingused.
Thefollowingisadefaultinstanceexample:
SQLSERVER
Namedinstanceexample:
SQLSERVER\INSTANCE
NOTE:
YoumustspecifytheSQLServerbyname.DonotusetheIPAddressof
theSQLServer.
8.ClickNext.
Theinstallervalidatesthatthedatabaseisavailable.Ifthedatabasecannotbe
found,thefollowingSQLServerValidationwarningmessagewillbedisplayed,
Databasenotfound,pleasemanuallyverifythatthedatabaseserveris
correct.OncethedatabaseserverisavailabletheMUSEServicesinstalledon
thisboxmustberestarted.Doyoustillwanttocontinue?ClickNoandverify
thenameandinstanceoftheSQLServerbeforeproceeding.
2059568-013JMUSE™v965
Page 66

MACECGSystemstoMUSECommunication
Ifthedatabasevalidationissuccessful,theMUSEServicesConguration
windowopens.
9.EntertheMUSEBackgroundUserNameandPasswordfortheMUSE
Backgroundaccount.
•ThedefaultusernamefortheMUSEBackgroundaccountisMUSEBkgnd.
•Ifyouareusingadomainaccount,enterthenamein<domainname>\<user
name>format.
•Ifyouareusingalocalaccount,entertheusernamein.\<username>format.
TheMUSEserviceswillbeinstalledusingthespeciedMUSEServiceManager
accountandpassword.
NOTE:
TheShowAdvancedOptionsbuttoncanbeusedtospecifyService
CommandLineArguments.Donotusethisbuttonunlessspecically
directedtodosobyGEHealthcare’sMUSEEngineeringorMUSETechnical
Support.
10.ClickNext.
Theinstallervalidatesthattheuseraccountyouchoseexistsonthesystem.
Ifyoureceiveawarningmessagethattheaccountwasnotfoundoruser
validationfailed,chooseNototheprompttoreturntotheMUSEServices
Congurationandcheckthefollowing:
•Theuseraccountandpasswordarecorrect
•Theaccountexists
Iftheuseraccountisvalidated,theMaintenanceCompletewindowopens.
11.ClickFinish.
12.VerifythattheMUSEandMUSEModemservicesarestarted.
66MUSE™v92059568-013J
Page 67

SettingUpaModemDevice
Usethefollowinginstructionstosetupamodemdevice.
1.LogontotheMUSEsystemasauserwithprivilegestomodifysettingsinMUSE
Setup.
2.GotoSystem>Setup.
3.IntheNavigationpane,selectModems.
4.Performoneofthefollowing:
a.Tocreateanewmodem,gotoAction>Newandselectoneofthefollowing:
MACECGSystemstoMUSECommunication
Modem
FaxModemUsedforphysicalFAXmodems
CSIModemUsedforphysicalCSImodems
CSIDirect
CSINetworkUsedforCSILANorWirelessCarts
Description
UsedfordirectserialcableataMUSE
client
TheappropriateModemPropertieswindowopens.
b.Tomodifyanexistingmodem,right-clickonanexistingentryandchoose
Properties.
TheappropriateModemPropertieswindowopens.
5.Entertheappropriatevaluesdescribedinthefollowingtables.
FAX,CSI,orCSIDirectModemProperties
Field
ComputerName
Port
Baud
1
IftheFAXmodemencountersproblemsat9600baud,use4800baud.
Description
Nameofthecomputerwherethemodem
isphysicallyinstalled.
ForFAXmodemsandCSImodems,thisis
theportwherethemodemisphysically
connected.ForCSIDirect,thisisthe
portwheretheserialcableisphysically
connected.
115.2K(CSIDirect),9600(CSI,FAX)
1
CSINetworkModemProperties
Field
ComputerName
IPAddressorHostnameTheIPaddressorhostnameassignedto
2059568-013JMUSE™v967
Description
Nameofthecomputerwherethe
connectionissupported.
thedevice.
Page 68

MACECGSystemstoMUSECommunication
CSINetworkModemProperties(cont'd.)
RestartingModems
Field
Port
RetryIntervalinSecondsDenestheupperlimitofthetime
Description
Portthisconnectionisusing.
delaybetweenthecart’sattemptsto
communicatewiththeMUSEsystem.The
defaultis30seconds.
NOTE:
Refertotheappropriatedocumentreferencedbelowfordetailedinstallation
andcongurationinformationfortheCSINetworkmodemtypeswhenused
withcompatibleMACECGsystems:
•MobileLinkWirelessCommunicationInstallationManual
•LANOptionforMACRestingECGSystemsInstallationandTroubleshooting
Guide
6.ClickOKtosaveyourchangesorClose/Canceltoignoreyourchanges.
NOTE:
Whenanewmodemissetup,theMUSEModemserviceisnotiedand
automaticallystartsanewthreadtosupporttheconnection.Youdonot
needtorestarttheMUSEModemserviceafterdeninganewmodem.
Theindividualthreadsrunningtosupporteachconnectionaredesignedto
automaticallyrestartiftheystopforanyreason.Usethefollowingprocedureifyou
needtorestartthemmanually.
1.LogontotheMUSEapplication.
2.SelectSetup>Modems.
3.Selectthemodem(s)thatyouwanttorestart,right-clickonit/them,andselect
RestartModem.
NOTE:
ThereisalsoaRestartModemicononthetoolbarthatcanbeusedto
restartamodem.
68MUSE™v92059568-013J
Page 69

MACECGSystemstoMUSECommunication
Amessagedisplaysthatthemodemsweresuccessfullyrestarted.
NOTE:
Ifyoudonotreceivethismessage,theMUSEMTHostservicewasnotable
tocommunicatetotheMUSEModemservices.Therearetwopossible
causesforthis:
•TheMUSEModemserviceisnotrunning.
•TherewallsettingsontheMUSEModemservice'shostsystemarenot
conguredcorrectly.
Formoreinformationaboutthemodems,includingwhentheystartorrestart,referto
theMUSEApplicationlog.
Iftheconnectiontoacartfails,theMUSEModemserviceimmediatelyattemptsto
restartit.ThefailureandrestartareloggedintheApplicationLog.
Iftherestartfailswithinoneminute,theservicedoesnotwaituntiltheoneminute
intervalisupbeforetryingagain.
Iftherearethreeconsecutivefailureswithinoneminute,amessageislogged
indicatingthaterrormessageloggingforthismodemhasbeenstoppeduntilthe
modemisworkingagain.HaltingerrormessageloggingpreventstheApplicationLog
fromllingupwithrepetitiveerrormessages.Whilenomessagesarebeinglogged,
theservicecontinuestorestartthemodeminthebackground.Oncethemodemis
restartedandcontinuesrunningforatleastoneminute,loggingresumesforthis
modem.Manuallyrestartingthemodemfromtheuserinterfacealsoresumeslogging.
SettingUpDCPInboundCommunication
TheMUSEsystemcanreceiveinboundtestsandrequestsforordersviaDCPInbound
communication.CompatibleGEHealthcareMACECGsystemscanusetheprotocolto
communicatedirectlywiththeMUSEapplicationserverwirelesslyorviaLAN.
Verifying/InstallingtheDCPInboundServiceandDCP
CommunicationOption
TheMUSEDCPInboundserviceandDCPCommunicationoptionaretypicallyinstalled
duringtheinitialinstallationoftheMUSEsystem.Usethefollowinginstructionsto
verifywhethertheMUSEDCPInboundserviceandDCPCommunicationoptionare
installedontheMUSEsystemandtoinstallthemiftheyarenot.Thisprocedureis
performedontheMUSEapplicationserver.ItmayonlybeperformedbyaqualiedGE
Healthcareservicerepresentative.
1.LogontotheMUSEapplicationserverastheMUSEAdministratoruser.
2.PerformafullorpartialshutdownoftheMUSEsystemfollowingtheauto
shutdownproceduresdescribedinthe“SystemAdministration”chapterofthe
MUSEv9CardiologyInformationSystemServiceManual.
NOTE:
Ifyouarejustcheckingtoseeiftheoptionisalreadyenabled,ashutdownis
notrequired.If,however,youhavetoenableordisabletheoption,theMUSE
serviceswillberestartedandashutdownisrequired.
3.GotoControlPanel>ProgramsandFeatures.
2059568-013JMUSE™v969
Page 70

MACECGSystemstoMUSECommunication
4.SelectMUSE9andclickChange.
TheWelcomewindowopens.
5.ChooseModifyandclickNext.
TheSelectFeatureswindowopens.
6.IntheSelectFeatureswindow,gotoServer>Services>DCPandverifythatDCP
Inboundischecked.
Ifitisnot,checkitnowandclickNext.
7.ContinuetoclickNextoneachwindowuntilyoureachtheChooseMUSE
Optionswindow.
8.VerifythatDCPCommunicationischecked.
Ifitisnot,checkitnow.
9.ClickNext.
TheMUSESerialNumberwindowopens.
10.IfyouaddedtheDCPcommunicationoptioninstep8,entertheOptions
CongurationPassword.
NOTE:
OnlyqualiedGEHealthcareservicerepresentativeshaveaccesstothis
password.Thispasswordcannotbeprovidedtocustomers.
11.ClickNextuntilyourchangesareappliedandtheMaintenanceComplete
windowopens.
12.ClickFinish.
13.IfyouperformedafullorpartialshutdownoftheMUSEsysteminstep2,cancel
theshutdownfollowingtheproceduresdescribedinthe“SystemAdministration”
chapteroftheMUSEv9CardiologyInformationSystemServiceManual.
14.VerifythattheMUSEDCPInboundserviceisstarted.
SettingUptheDCPServerCongurationintheMUSE
System
Bydefault,theDCPInboundservicehasaDeviceFriendlyNameofMUSEandlistens
onport9240ofallnetworkinterfacesontheMUSEapplicationserver.
Performthefollowingstepstomodifythesedefaults:
1.LogontotheMUSEsystemasauserwithprivilegestomodifysettingsinMUSE
Setup.
2.GotoSystem>Setup.
3.IntheNavigationpane,selectSystem.
4.Right-clickontheMUSEentryandchooseProperties.
TheSystemPropertieswindowopens.
5.SelectDCPConguration.
70MUSE™v92059568-013J
Page 71

MACECGSystemstoMUSECommunication
6.Modifytheeldsusingtheinformationinthefollowingtable.
Field
DeviceFriendlyNameThisisthenamethecompatibledevicewillsee
ServerPort
NetworkInterfacesThisiswhereyoucanspecifywhichnetwork
ServerAddressesThisisaread-onlyoutputindicatingtheServer
Description/Action
whenndingDCPservers.ThedefaultisMUSE.
Changethisifdesired.
ThisistheportonwhichtheDCPInbound
serviceislisteningforinboundconnections.
Thedefaultis9240.Changethisifnecessary.
interfacetheDCPServershouldlistenon.This
eldisblankbydefaultsoitwilllistenonall
networkinterfacesontheMUSEapplication
server.ToconguretheDCPServertolisten
onlyonasinglenetworkinterface,forexample
IPv4,youcantypetheIPv4IPaddressintothis
eld.
Address(es)thattheDCPInboundserviceis
currentlylisteningon.ThisisthefullDCPURL
thatcanbeusedtodenethisMUSEsystemon
acompatibleDCPclientdevicesuchasaMAC
2000.Multipleserveraddressesmaybelistedif
theNetworkInterfaces’eldisblank.
7.ClickOKtosaveyourchangesorClose/Canceltoignoreyourchanges.
NOTE:
RefertotheLANOptionforMACInstallationandTroubleshootingGuidefor
detailedinstallationandcongurationinformationregardingtheuseofthe
DCPProtocoloncompatibleMACECGsystems.
8.Ifanycongurationchangesweremade,restarttheMUSEDCPInboundservice
ontheMUSEApplicationserver.
SystemCheckout
CompletethefollowingvericationprocedurestoensuretheMACECGdevicescan
successfullytransmitteststotheMUSEsystemanddownloadordersfromtheMUSE
system.
CSITransmissiontotheMUSESystem
1.TransmitanECGtestfromtheMACECGdevicetotheMUSEsystemusingthe
CSIprotocol.
2.VerifythetestissuccessfullyacquiredintotheMUSEsystem.
2059568-013JMUSE™v971
Page 72

MACECGSystemstoMUSECommunication
MUSEOrderDownloadviaCSI
1.FromtheMACECGdevice,downloadanorderfromtheMUSEsystemusing
theCSIprotocol.
2.VerifytheorderissuccessfullydownloadedtotheMACECGdevice.
DCPTransmissiontotheMUSESystem
1.TransmitanECGtestfromtheMACECGdevicetotheMUSEsystemusingthe
DCPprotocol.
2.VerifythetestissuccessfullyacquiredintotheMUSEsystem.
MUSEOrderDownloadviaDCP
1.FromtheMACECGdevice,downloadanorderfromtheMUSEsystemusing
theDCPprotocol.
2.VerifytheorderissuccessfullydownloadedtotheMACECGdevice.
72MUSE™v92059568-013J
Page 73

MUSEeDocConnect
TheoryofOperation
6
ThischapterdescribeshowtoconguretheMUSEeDocConnectoptionusedto
processincomingelectronicdocumentlesontheMUSEv9system.
TheMUSEeDocConnectoptionallowstheMUSEsystemtoacquireelectronic
documents.Oncetheelectronicdevicetypesandsharedfoldersareappropriately
denedintheMUSEsystem,devicestransferelectronicdocumentstoashared
folder.TheMUSEGeneracqservicesearchesthissharedfolderforlesandstores
themintheMUSEdatabase.Electronicdocumenttestsmayalsobeacquiredvia
theMUSEAcquisitionmodule.RefertotheMUSEv9CardiologyInformationSystem
OperatorManualforinstructionsonhowtoacquireelectronicdocumentsviathe
MUSEAcquisitionmodule.
eDocConnectAcquisitionFlowChart
TheMUSEsystemcanacquirethefollowingtypesofelectronicdocuments:
•PDF–PortableDocumentFormatFile
•PDF/A–PortableDocumentFormatFile/Archive
•JPG–JPEGImage
•DOC–MicrosoftWordDocument
•DOCX–MicrosoftWordOpenXMLDocument
•TIFF–TaggedImageFileFormat
•PNG–PortableNetworkGraphic
•TXT–PlainTextFile
2059568-013JMUSE™v9
73
Page 74

MUSEeDocConnect
PreparingtoCongureeDocConnect
ThefollowingstepsareperformedbyGEHealthcarepersonneltoprepareforthe
installationandcongurationoftheeDocConnectoption:
1.AMUSEsystemcustomercontactsGEHealthcaretoconguretheirMUSE
systemtoreceiveandstoreelectronicdocumenttestles.
2.AGEHealthcarerepresentativeasksforseveralexamplesofthedocumentsthe
customerwantstoimportandinquiresabouttheabilityofthedevicestoexport
thesedocuments.TheGEHealthcarerepresentativealsogathersinformation
regardingthetesttypeandtargetMUSEsiteandlocationforthesetestresults.
3.GEHealthcarecreatesandtestsanAcquisitionProlethatallowstheelectronic
documentstobeautomaticallyimportedintothetargetMUSEsiteandlocation,
withthecorrectTestTypeandPatient/Testdemographics.
NOTE:
TheAcquisitionProleiscreatedusingtheAcquisitionProleTool.Useof
theAcquisitionProleTooltocreatetheAcquisitionProleisoutsidethe
scopeofthisdocument.
AfterGEHealthcarecollectsthenecessaryinformationandcreatestheAcquisition
Proles,theeDocConnectoptioncanbeinstalledandcongured,asdescribedin
theremainderofthischapter.
CustomerRequirements
Thecustomerisresponsibleforthefollowing:
•SupplyingelectronicdocumentsthatmeettheconguredeDocConnectprole(s).
•Establishingnetworkconnectivitybetweenthedevicethatoutputstheelectronic
documentsandtheMUSEsystem.
•GrantingtheuseraccountconguredtostarttheMUSEGeneracqservicewith
ReadandDeletepermissionstothelesandsharelocationoftheelectronic
documents.
InstallingandConguringMUSEeDocConnect
InstallingandconguringeDocConnectconsistsofthefollowingtasks:
•“InstallingtheeDocConnectOptionintheMUSESystem”onpage75
•“AddingorEnablingaTestTypeintheMUSESystem”onpage75
•“SettinguptheAcquisitionProleintheMUSESystem”onpage76
•“SettinguptheFileShare”onpage77
74
MUSE™v92059568-013J
Page 75

InstallingtheeDocConnectOptionintheMUSESystem
UsethefollowingproceduretoaddtheeDocConnectoptiontotheMUSEsystem.
NOTE:
ThisproceduremayonlybeperformedbyaqualiedGEHealthcareservice
representative.
1.LogontotheMUSEapplicationserverastheMUSEAdministratoruser.
2.PerformafullorpartialshutdownoftheMUSEsystemfollowingtheauto
shutdownproceduresdescribedinthe“SystemAdministration”chapterofthe
MUSEv9CardiologyInformationSystemServiceManual.
NOTE:
Ifyouarejustcheckingtoseeiftheoptionisalreadyenabled,ashutdownis
notrequired.If,however,youhavetoenableordisabletheoption,theMUSE
serviceswillberestartedandashutdownisrequired.
3.FromtheWindowsControlPanel,gotoProgramsandFeatures.
4.SelectMUSE9andclickChange.
MUSEeDocConnect
TheWelcomewindowopens.
5.ChooseModifyandclickNext.
TheSelectFeaturewindowopens.
6.ContinuetoclickNextoneachwindowuntilyoureachtheChooseMUSE
Optionswindow.
7.VerifythateDocConnectischecked.
Ifitisnot,checkitnow.
8.ClickNext.
TheMUSESerialNumberwindowopens.
9.IfyouaddedtheeDocConnectoptioninstep7,youneedtoentertheOptions
CongurationPassword.
NOTE:
OnlyqualiedGEHealthcareservicerepresentativeshaveaccesstothis
password.Thispasswordcannotbeprovidedtocustomers.
10.ClickNextuntilyourchangesareappliedandtheMaintenanceComplete
windowopens.
11.ClickFinish.
12.IfyouperformedafullorpartialshutdownoftheMUSEsysteminstep2,cancel
theshutdownfollowingtheproceduresdescribedinthe“SystemAdministration”
chapteroftheMUSEv9CardiologyInformationSystemServiceManual.
AddingorEnablingaTestTypeintheMUSESystem
Usethefollowingtwoprocedurestocreateanewtesttypeortoenable,disable,or
modifyanexistingtesttype,asneeded.
2059568-013JMUSE™v9
75
Page 76

MUSEeDocConnect
CreatingaNewTestType
1.FromwithintheMUSEapplication,gotoSetup.
2.SelectTestTypes.
Thelistofexistingtesttypesisdisplayed.
3.SelectActionandclickNew.
4.Completetheeldsasindicatedinthefollowingtable:
FieldTask
TestTypeName
TestTypeAbbreviationEnterthethreecharacterabbreviationforthetesttype.
EnabledforSitesPutacheckintheboxnexttositesforwhichthistesttype
5.ClickOKtosaveyourchanges.
Enterthenameofthetesttype.
shouldbeenabled.
Enabling,Disabling,orModifyinganExistingTestType
1.FromwithintheMUSEapplication,gotoSetup.
2.SelectTestTypes.
Alistofexistingtesttypesisdisplayed.
3.Right-clickonthetesttypeyouwanttomodifyandchooseProperties.
TheTestTypePropertieswindowopens.
4.Checkthecheckboxnexttoeachsiteyouwanttoenableforthistesttype.
Iftheboxisuncheckedforasite,itmeansthetesttypeisdisabledforthatsite.
Ifthetesttypeisnotasystem-denedtesttype,thentheTestTypeNameand
TestTypeAbbreviationmayalsobechangedatthistime.
5.ClickOKtosaveyourchanges.
Ifnochangesweremade,clickClosetoclosetheTestTypeProperties.
SettinguptheAcquisitionProleintheMUSESystem
Usethefollowingtwoprocedurestocreateanewacquisitionproleortomodifyan
existingone,asneeded.
CreatingaNewAcquisitionProle
Tocreateanewacquisitionprole:
1.FromwithintheMUSEapplication,gotoSetup.
2.SelectAcquisitionProle.
Thelistofexistingacquisitionprolesisdisplayed.
Ifthisistherstacquisitionprole,thelistisempty.
3.SelectActionandclickNew.
76MUSE™v92059568-013J
Page 77

4.Completetheeldsasindicatedinthefollowingtable.
FieldValue
Name
TestType
Enterthenameoftheprole.
Itisrecommendedthatthenamereectthetesttype,site
andlocation.Forexample:CATH-SITE1-LOC50
Choosethetesttypefromthedrop-downlist.
MUSEeDocConnect
Site
Location
ProleClickImport,browsetothelecreatedintheAcquisition
5.ClickOK.
ModifyinganExistingAcquisitionProle
1.FromwithintheMUSEapplication,gotoSetup.
2.SelectAcquisitionProle.
Thelistofexistingacquisitionprolesisdisplayed.
3.Right-clickontheacquisitionproleyouwanttomodifyandchooseProperties.
4.Modifytheeldsasneeded.
5.ClickOK.
SettinguptheFileShare
TotransferelectronicdocumentsfromadevicetotheMUSEsystem,youneedto
createashare.Usethefollowinginstructionstocreateashare.
ChoosetheMUSEsitefromthedrop-downlist.
ChoosetheMUSEsitelocationfromthedrop-downlist.
ProleTool,andclickOpen.
CreatingtheWindowsFileShare
ThesharecanbealocalfolderontheMUSEleserverthatissharedtothedevice,
orashareonadevicethattheMUSEapplicationservercanaccess.Thesharecan
alsobeagateway-typecomputertowhichboththeMUSEapplicationserverand
thedevicehaveaccess.Inanycase,theuseraccountconguredtostarttheMUSE
GeneracqservicemusthaveaccesstothelocationthroughFullControlleand
sharepermissions.
SettinguptheShareFolderinMUSE
Usethefollowingtwoprocedurestocreateanewsharefolderortomodifyanexisting
one,asneeded.
CreatingaNewShareFolder
1.FromwithintheMUSEapplication,gotoSetup.
2.SelectShareFolder.
Thelistofexistingsharefoldersisdisplayed.
3.SelectActionandclickNew.
2059568-013JMUSE™v9
77
Page 78

MUSEeDocConnect
4.Completetheeldsasindicatedinthefollowingtable:
FieldValue
Entry
FileNameFilterEnterthelemasktolteron.Forexample*.PDFensures
ProleNameChoosetheappropriateMUSEprolenamefromthe
EntertheUNCorLocalPathfromwhichtheelectronic
documentsareparsed.
thatonlyleswithanextensionof.PDFareprocessed.
drop-downlist.AlleDocConnectShareFoldersmusthave
aprolenameselected.
5.ClickOK.
ModifyinganExistingShareFolder
1.FromwithintheMUSEapplication,gotoSetup.
2.SelectShareFolder.
Thelistofexistingsharefoldersisdisplayed.
3.Right-clickonthesharefolderyouwanttomodifyandchooseProperties.
4.Modifytheeldsasneeded.
5.ClickOK.
SystemCheckout
CompletethefollowingproceduretoensurethattheMUSEsystemcansuccessfully
acquireelectronicdocuments.
1.Fromadevicecapableoftransferringelectronicdocuments,transferan
2.LogontotheMUSEsystem.
3.VerifythetestisdisplayedintheMUSEsystemEditList.
4.VerifythetestcanbeopenedintheMUSEEditor.
electronicdocumenttotheappropriatesharedfolderontheMUSEsystem.
78MUSE™v92059568-013J
Page 79

MUSEeDocConnect
Troubleshooting
Symptom
Documentimportedwith
incorrectacquisitionprole.
Cannotacquireelectronic
documents.
Condition
Anelectronicdocumentwas
copiedtoanincorrectfolder
orbulkacquiredusingthe
wrongacquisitionprole.If
thishappensthenitismost
likelyacquiredwithincorrect
attributes(testtype,site,
location,etc.)
Theuseraccountthatis
conguredtostarttheMUSE
Generacqserviceonthe
MUSEapplicationserver
doesnothaveaccesstothe
folderorsharedenedfor
theShareFolderintheMUSE
system.
TheShareFolderisnotset
upcorrectlyforthelocation
oftheelectronicdocument
lesinMUSESystemSetup.
Action
Tocorrectthis,discardthe
test.OnceontheDiscarded
DataList,youcanreacquire
theelectronicdocument
usingthecorrectacquisition
prole.Right-clickonthetest
ontheDiscardedDataList,
chooseCorrectAcquisition
Prole,andchoosethe
correctacquisitionprole
withwhichtoreimportthe
test.
Ensuretheuseraccount
conguredtostarttheMUSE
Generacqserviceonthe
MUSEapplicationserverhas
accesstothefolderorshare
denedfortheShareFolder
intheMUSEsystem.
EnsuretheShareFolderset
upinMUSESystemSetupis
correctlydened.
2059568-013JMUSE™v979
Page 80

MUSEeDocConnect
80MUSE™v92059568-013J
Page 81

MUSEXMLImportOption
ThischapterdescribeshowtoconguretheMUSEXMLImportoptiontoprocess
incomingXMLlesontheMUSEv9system.
TheoryofOperation
TheXMLImportoptionallowstheMUSEsystemtoacquireXMLlesthatmeet
theMUSETransactionalXMLspecication.Devicesthatareabletooutputthe
appropriatelystructuredXMLlessendtheirXMLlesintoasharedfolderontheMUSE
system.TheMUSEXMLParserservicesearchesthissharedfolder,acquirestheXML
lesfromthesharedfolder,andmovesthemintotheMUSEsystemforprocessing.
ThetestisnormalizedandstoredintheMUSEdatabase.
7
XMLAcquisitionFlowChart
CustomerRequirements
Thecustomerisresponsibleforsupplyingthefollowing:
•XMLlesthatmeettheMUSETransactionalXMLspecication.RefertotheMUSE
CardiologyInformationSystemTransactionalXMLDeveloper’sGuidefordetails.
•NetworkconnectivitybetweenthedevicethatcanoutputXMLlesandtheMUSE
applicationserver.
InstallingandConguringtheMUSEXMLImport
Option
AddingtheXMLImportoptionconsistsofthefollowingtasks:
•“InstallingtheXMLImportOptionandtheMUSEXMLParserService”onpage82
•“SettinguptheXMLSharedFolder”onpage83
•“UsingXMLCONFIG.EXEtoAdd,Update,orDeleteXMLDevices”onpage83
2059568-013JMUSE™v981
Page 82

MUSEXMLImportOption
InstallingtheXMLImportOptionandtheMUSEXMLParser
Service
UsethefollowingproceduretoinstalltheXMLImportoptionandMUSEXMLParser
service.
NOTE:
ThisproceduremayonlybeperformedbyaqualiedGEHealthcareservice
representative.
1.LogontotheMUSEapplicationserverastheMUSEAdministratoruser.
2.PerformafullorpartialshutdownoftheMUSEsystemfollowingtheauto
shutdownproceduresdescribedinthe“SystemAdministration”chapterofthe
MUSEv9CardiologyInformationSystemServiceManual.
NOTE:
Ifyouarejustcheckingtoseeiftheoptionisalreadyenabled,ashutdownis
notrequired.If,however,youhavetoenableordisabletheoption,theMUSE
serviceswillberestartedandashutdownisrequired.
3.FromtheWindowsControlPanel,gotoProgramsandFeatures.
4.SelectMUSE9andclickChange.
TheWelcomewindowopens.
5.ChooseModifyandclickNext.
TheSelectFeaturewindowopens.
6.IntheSelectFeatureswindow,gotoServer>ServicesandverifythatXMLis
checked.
ThisinstallstheMUSEXMLParserservice.
7.ContinuetoclickNextoneachwindowuntilyoureachtheChooseMUSE
Optionswindow.
8.VerifythatXMLImportischecked.
Ifitisnot,checkitnow.
9.ClickNext.
TheMUSESerialNumberwindowopens.
10.IfyouaddedtheXMLImportoptioninstep8,youneedtoentertheOptions
CongurationPassword.
NOTE:
OnlyqualiedGEHealthcareservicerepresentativeshaveaccesstothis
password.Thispasswordcannotbeprovidedtocustomers.
11.ClickNextuntilyourchangesareappliedandtheMaintenanceComplete
windowopens.
12.ClickFinish.
13.IfyouperformedafullorpartialshutdownoftheMUSEsysteminstep2,cancel
theshutdownfollowingtheproceduresdescribedinthe“SystemAdministration”
chapteroftheMUSEv9CardiologyInformationSystemServiceManual.
82MUSE™v92059568-013J
Page 83

SettinguptheXMLSharedFolder
TotransferXMLtestsfromanXML-capabledevicetotheMUSEsystem,youneedto
createasharedfolderontheMUSEsystem.
TheMUSEXMLParserserviceontheMUSEapplicationserverisonlyconguredto
checktheMUSEXMLfolder(defaultisd:\muse\xml)forincomingXMLlestoprocess.
ThisfoldermayneedtobesharedtoallowXML-capabledevicestotransferrecords
intoit.Ifthisfolderisshared,theXML-capabledevicethatwillbewritingdatatothe
sharemusthaveaccesstoit.Customersareresponsibleforensuringthisconnectivity
betweentheXMLdeviceandtheMUSEsystem.
NOTE:
TherearenineDTDlesintheMUSEXMLfolder.These.DTDlesarerequired
fortheMUSEsystemtoacquireXMLrecords.Theselesmustnotbealteredor
deleted.
MUSEXMLImportOption
UsingXMLCONFIG.EXEtoAdd,Update,orDeleteXML
Devices
UsethefollowingtwoprocedurestoaddXMLdevicesortoupdateordeleteXML
devicesasneeded.SettingsforknownXMLdevicesareprovidedafterthese
procedures.
UsingXMLCONFIG.EXEtoAddaNewXMLDevice
TheXMLCONFIGutilityinsertsentriesintothecfgXmlInputtableintheMUSE_System
database.
1.LogontotheMUSEapplicationserverastheMUSEAdministratoruser.
2.Runthexmlcong.exeutilitylocatedinthefolderwheretheMUSEapplication
isinstalled(defaultisc:\ProgramFiles(x86)\MUSE).
TheXMLInputDeviceswindowopens.
2059568-013JMUSE™v983
Page 84

MUSEXMLImportOption
3.ClickNewDevice.
TheAddaNewXMLDevicewindowopens.
4.Enterthedevicecongurationusingthefollowingtable.
FieldValue
DeviceName
DeviceType
ManufacturerSelecttheappropriatemanufacturer.Anyoneotherthan
ImageType
AnalysisFunctionAlwayssetthistoXML_AnalyzeWaveform.Thisiscase
Enterauniquenameforeachdevice.Thevalueentered
heremustmatchthe<AcquisitionDevice>valueinthe
incomingXMLle.
SelectECG.Nootherdatatypesaresupportedatthistime.
PhysioControlsorGetemedisacceptableunlessthe
deviceisactuallyaPhysioControls(exceptforLP15-GL)or
Getemeddevice.ThespeciccongurationsforthePhysio
ControlLP12andGetemeddevicesaredocumentedin
“KnownXMLDeviceCongurations”onpage85.Whenin
doubt,chooseOther.
AlwayssetthistoImageasWaveformDataPoints.
sensitiveandisauto-populatedintheXMLCONFIGutility;
donotchangethisvalue.
5.ClickOK.
6.VerifytheDeviceNamehasbeenaddedtothelistofXMLDevicesinthe
XMLCONFIGutility.
84MUSE™v92059568-013J
Page 85

7.ClosetheXMLCONFIGutility.
8.RestarttheMUSEXMLParserservice.
UsingXMLCONFIG.EXEtoUpdateorDeleteaDevice
1.LogintotheMUSEapplicationserverastheMUSEAdministratoruser.
2.Runthexmlcong.exeutilitylocatedinthefolderwheretheMUSEapplicationis
installed(defaultisc:\ProgramFiles(x86)\MUSE).
MUSEXMLImportOption
KnownXMLDeviceCongurations
TheXMLInputDeviceswindowopens.
3.Toupdateadevice,selectthedeviceandclickUpdateDeviceandusethe
informationinstep4ofthe“UsingXMLCONFIG.EXEtoAddaNewXMLDevice”
onpage83.
NOTE:
Whenupdatingadevice,thedevicenamecannotbechanged.Tochange
thenameofadevice,thedevicemustrstbedeletedandthenrecreated
withanewname.
Todeleteadevice,clickDeleteDevice.
4.ClickOK.
5.ClosetheXMLCONFIGutility.
6.RestarttheMUSEXMLParserservice.
ThefollowingwindowsdisplaytheXMLdevicecongurationsyoumustuseto
congurethefollowingspecicdevices:
•PhysioControlsLifePak12
•PhysioControlsLifePak15
•GetemedCM3012
2059568-013JMUSE™v985
Page 86

MUSEXMLImportOption
•DataMedFormatTranslator
•ZollRescueNet
PhysioControlsLifePak12 GetemedCM3012
PhysioControlsLifePak15 DataMedFormatTranslator
86MUSE™v92059568-013J
Page 87

MUSEXMLImportOption
ZollRescueNet
SystemCheckout
CompletethefollowingproceduretoensurethattheMUSEXMLImportoptionis
setupproperly.
1.TransferanXMLletotheXMLfolderontheMUSEsystem.
2.LogontotheMUSEsystem.
3.VerifythetestisdisplayedintheMUSEsystemEditList.
4.VerifythetestcanbeopenedintheMUSEEditor.
Troubleshooting
Usethefollowingtroubleshootingtipsifsomethingisnotworkingcorrectly.
Symptom
TheXMLleisnotprocessed
fromtheXMLfolder.
XMLisrenamedtoBADin
theMUSEXMLfolder .
Condition
TheMUSEXMLParser
serviceisnotstarted.
TheXMLCONFIG.EXE
congurationisincorrect.
MUSEXMLParserisunable
toprocesstheXMLle.
Action
StarttheMUSEXMLParser
service.
EnsuretheXMLCONFIG.EXE
congurationiscorrect.
ChecktheMUSEapplication
logfordetails.
2059568-013JMUSE™v987
Page 88

MUSEXMLImportOption
88MUSE™v92059568-013J
Page 89

8
DICOMCommunication
ThischapterdescribeshowtocongurethesystemforDICOMcommunication.
FordetailedinformationonDICOMconformance,refertotheMUSEv9Cardiology
InformationSystemDICOMConformanceStatement.
TheoryofOperation
TheMUSEsystemsupportsDICOMfunctionalityinthreedifferentways.Thefollowing
tableprovidesadescriptionofeach.
Feature
DICOMStorageServiceClass
Provider(SCP)
DICOMStorageServiceClass
User(SCU)
DICOMModalityWorklist(MWL)
ServiceClassUser(SCU)
Description
AllowsDICOMECGteststobereceivedbytheMUSE
system.StorageCommitmentissupported.TheMUSE
systemactsasaDICOMStorageServiceClassProvider.
NOTE:
InboundDICOMtestscanonlybeacquiredinto
MUSESite1andwillbeacquiredwithoutanyMUSE
locationinformation.Thismeansthattestsacquired
viaDICOMandintendedforaMUSESiteotherthan
1mustbediscardedandrecoveredtothecorrect
site.Furthermore,non-defaultreportdistribution
locationscannotbeusedforroutingthetestwhenit
isacquired.Thelocationofthetest,however ,canbe
changedintheMUSEEditor.
AllowsteststobesentfromtheMUSEsystemtoDICOM
devices.StorageCommitmentissupported.Thesystem
actsasaDICOMStorageServiceClassUser.
AllowsthesystemtoreceiveordersfromDICOMModality
WorklistServiceClassProviders.Thesystemqueries
ModalityWorklistServiceClassProvider(SCP)forModality
Worklistordersandupdatesthesystemordersdatabase.
ThesystemactsasaDICOMModalityWorklistService
ClassUser .
2059568-013JMUSE™v989
Page 90

DICOMCommunication
ExampleNetworkDiagram
TransmissionFlowCharts
Thefollowingdescriptionsandowchartsexplaintheowofdatabetween:
•SendingDICOMSCUandthereceivingMUSEDICOMSCP
•SendingMUSEDICOMSCUandthereceivingDICOMSCP
•MUSEMWLSCUandDICOMMWLSCP
DICOMSCUtoMUSEDICOMSCP
TheDICOMdeviceStorageServiceClassUsertransmitsaDICOMtestacrossthe
networktotheMUSEDICOMStorageServiceClassProviderservice.Thetestisthen
normalizedontheMUSEsystemandstoredinthedatabase.Ifstoragecommitmentis
enabled,aDICOMSCUsendsastoragecommitmentrequesttotheMUSEsystemand
theMUSEsystemreturnsastoragecommitmentresponsetotheDICOMSCU.
MUSEDICOMSCUtoDICOMSCP
TheMUSEDICOMdeviceStorageServiceClassUsertransmitsaDICOMtestacrossthe
networktoareceivingDICOMStorageServiceClassProvider.Ifstoragecommitment
isenabled,theMUSEsystemsendsastoragecommitmentrequesttotheDICOMSCP,
andtheDICOMSCPreturnsastoragecommitmentresponsetothesystem.
MUSEModalityWorklistSCU
TheDICOMModalityWorklistServiceClassUserservicequeriestheDICOM
ModalityWorklistServiceClassProviderfororders.Ordersthatarereturnedare
90MUSE™v92059568-013J
Page 91

MUSEServices
DICOMCommunication
created/updatedinthesystemordersdatabase.Theseordersbehavethesameas
ordersreceivedviatheMUSEHL7Parser.
EachsystemDICOMfunctionhasitsownMUSEserviceonthesystemapplication
server,asdescribedinthefollowingtable.
MUSEServiceNameDICOMFunction
MUSEDICOMModalityWorklistClientDICOMModalityWorklist(MWL)ServiceClass
User(SCU)
MUSEDICOMStorageProviderDICOMStorageServiceClassProvider(SCP)
MUSEDICOMStorageUser
DICOMStorageServiceClassUser(SCU)
CustomerRequirements
ConguringDICOMCommunication
Inadditiontotheseservices,theMUSENormalserviceisusedtonormalizeall
inboundtestdata,includingDICOM,andtheMUSEFormatserviceisusedtoformat
alloutbounddata,includingDICOMIODandDICOMEncapsulatedPDF.
Thecustomerisresponsibleforsupplyingthefollowing:
•Networkconnectivitybetweenthesystemandthenon-systemDICOMservices
anddevices.
•AEtitles,IPaddresses,andportsforallnon-systemDICOMservicesanddevices.
FollowtheinstructionsinthissectiontocompletethecongurationofMUSEDICOM
Communication.
•“AddingtheDICOMService(s)andOptiontotheMUSESystem”onpage92
•“ConguringtheMUSESystemtoReceiveDICOMTests”onpage93
•“ConguringtheSystemtoSendDICOMTests”onpage95
•“ConguringtheSystemtoQueryforDICOMOrders”onpage99
2059568-013JMUSE™v991
Page 92

DICOMCommunication
AddingtheDICOMService(s)andOptiontotheMUSESystem
TherearethreeDICOMservicesandoneDICOMoption.Usethefollowingprocedure
toaddtheDICOMservice(s)andoptiontotheMUSEsystem.
NOTE:
ThisprocedurecanonlybeperformedbyaqualiedGEHealthcareservice
representative.
1.LogontotheMUSEapplicationserverastheMUSEAdministratoruser.
2.PerformafullorpartialshutdownoftheMUSEsystemfollowingtheauto
shutdownproceduresdescribedinthe“SystemAdministration”chapterofthe
MUSEv9CardiologyInformationSystemServiceManual.
NOTE:
Ifyouarejustcheckingtoseeiftheoptionisalreadyenabled,ashutdownis
notrequired.If,however,youhavetoenableordisabletheoption,theMUSE
serviceswillberestartedandashutdownisrequired.
3.GotoControlPanel>ProgramsandFeatures.
4.SelectMUSE9andclickChange.
TheWelcomewindowopens.
5.SelectModifyandclickNext.
TheSelectFeatureswindowopens.
6.GotoServer>Services.
7.ScrolldowntothebottomofthelistandselectthecheckboxnexttoDICOM.
SelectingDICOMautomaticallyselectsallDICOMservices.
8.ClickNext.
TheMUSEClientCongurationwindowopens.
9.ContinueandclickNextoneachwindowuntilyoucometotheChooseMUSE
Optionswindow.
10.ScrolltothebottomofthelistandverifythatDICOMisselected.
IfDICOMisnotselected,selectitnow.
11.ClickNext.
TheMUSESerialNumberwindowopens.
12.IfyouaddedtheDICOMoptioninstep10,youneedtoentertheOptions
ConrmationPassword.
NOTE:
OnlyqualiedGEHealthcareservicerepresentativeshaveaccesstothis
password.Thispasswordcannotbeprovidedtocustomers.
13.ClickNext.
TheMaintenanceCompletewindowopens.
14.ClickFinish.
92MUSE™v92059568-013J
Page 93

IfyouperformedafullorpartialshutdownoftheMUSEsysteminstep2,cancelthe
shutdownfollowingtheproceduresdescribedinthe“SystemAdministration”chapter
oftheMUSEv9CardiologyInformationSystemServiceManual.
ConguringtheMUSESystemtoReceiveDICOMTests
ConguringtheMUSEsystemtoreceiveDICOMtestsconsistsofthefollowing
procedures:
•“ConguringtheMUSEDICOMStorageProviderService”onpage93
•“ConguringtheInboundDICOMDevice(s)intheMUSESystem”onpage94
•“ConguringtheTransmittingDICOMDevice(s)toSendDICOMTeststotheMUSE
System”onpage95
ConguringtheMUSEDICOMStorageProviderService
TheMUSEDICOMStorageProviderserviceisaDICOMStorageServiceClassProvider
andreceivesinboundDICOMtestsfromDICOMStorageServiceClassUsers.The
serviceisconguredwithadefaultAETitleandport.
DICOMCommunication
UsethefollowingstepstoconguretheMUSEDICOMStorageProviderservice.
1.LogontotheMUSEapplication.
2.GotoSystem>Setup.
3.SelectDICOMServices.
4.Right-clickonDICOMSTORESCPandselectProperties.
5.SelectGeneralandcompletethefollowingeldsasappropriate:
Field
IPAddressThenameorIPaddress
AETitleTheDICOMApplication
Port
DescriptionAction
oftheMUSEapplication
server.
Thedefaultislocalhost.
EntityTitleoftheMUSE
StorageServiceClass
Provider.
Thedefaultis
MuseStoreSCP.
TheportthattheMUSE
StorageProviderservice
willlistenon.
Thedefaultportis104.
Youshouldnotneedto
changethis.
Typeanewtitleifneeded.
Typeanewportvalueif
needed.
EnableStorage
Commitment
2059568-013JMUSE™v993
Ifthisboxischecked,the
MUSEsystemwillprovide
storagecommitment
responses.
Checkthisboxasneeded.
TheMUSEStorage
Providerservicelistens
forstoragecommitment
messagesonthesame
IPaddressandportas
DICOMtests.
Page 94

DICOMCommunication
Field
RetryIntervalforsending
N-Report(inseconds)
Maximumwaittime
beforefailure
DescriptionAction
Thetimebetween
attemptstosendaStorage
Commitmentstatustothe
DICOMStorageService
ClassUser.
Thedefaultis60.
hemaximumtimethe
MUSEStorageProvider
serviceshouldwaitfor
aStorageCommitment
response.
Thedefaultis180.
6.SelectDICOMAssociationSettingstocongurethesesettings.
Defaultvaluesexistforthesesettingsandtypicallydonotneedtobechanged.
7.ClickOKtosaveyourchanges.
Toignoreyourchanges,clickClose.
Ifanychangesweremade,theMUSEDICOMStorageProviderservicewillneedtobe
restartedonthesystemapplicationserver.
ConguringtheInboundDICOMDevice(s)intheMUSESystem
Changethisvalueif
needed.
Changethisvalueif
needed.
EachinboundDICOMStorageServiceClassUserDeviceneedstobeconguredin
theMUSEsystempriortothedevicebeingabletoassociatewithandsendDICOM
teststoit.
UsethefollowingstepstocongureaninboundDICOMDeviceintheMUSEsystem.
RepeatthesestepsforeachinboundDICOMDevicethatneedstosendteststothe
MUSEsystem.
1.LogontotheMUSEapplication.
2.GotoSystem>Setup.
3.SelectDICOMDevices.
4.Performoneofthefollowingsteps:
a.Tocreateanewdevice,gotoAction>New.
b.Tomodifyanexistingdevice,right-clickontheentryforthedeviceand
selectProperties.
5.SelectGeneralandcompletethefollowingeldsasappropriate.
Field
AETitleTheDICOMApplication
IPAddressThenameorIPaddressof
DescriptionActions
EntityTitleoftheMUSE
StorageServiceClassUser
device.
theDICOMdevice.
TypetheAETitleofthe
DICOMDevicethatsends
teststotheMUSEsystem.
TypetheIPaddressofthe
DICOMDevicethatsends
teststotheMUSEsystem.
Description
94MUSE™v92059568-013J
Thiseldcanbeusedto
describethedevice.
Typeadescriptionofthe
deviceintothiseld.
Page 95

DICOMCommunication
Field
StorageCommitment
AETitleTheDICOMApplication
IPAddressTheIPAddressofthe
Port
6.ClickOKtosaveyourchange.
Toignoreyourchanges,clickClose.
DescriptionActions
Ifstoragecommitment
isenabled,theMUSE
systemsendsstorage
commitmentresponsesto
theDICOMdevicestorage
commitmentserviceusing
informationcongured
here.
EntityTitleofthestorage
commitmentserviceon
theDICOMdevice.
storagecommitment
serviceontheDICOM
device.
Theportnumberofthe
storagecommitment
serviceontheDICOM
device.
IftheDICOMdevice
supportsstorage
commitment,selectthis
boxtoenableit.
TypetheAETitleofthe
storagecommitment
serviceontheDICOM
Device.
TypetheIPAddressof
thestoragecommitment
serviceontheDICOM
Device.
Typetheportnumberof
thestoragecommitment
serviceontheDICOM
Device.
ConguringtheTransmittingDICOMDevice(s)toSendDICOMTeststothe
MUSESystem
TosendDICOMteststotheMUSEsystem,conguretheDICOMStorageServiceClass
UserDevicetousetheMUSEDICOMStorageServiceClassProviderAETitle,IPAddress,
andPortasdenedintheDICOMSTORESCPentryinDICOMServicesinMUSESetup.
Usethissameinformationforstoragecommitment,ifthatoptionisenabledforthe
MUSEStorageProviderservice.
ConguringtheSystemtoSendDICOMTests
ConguringthesystemtosendDICOMtestsconsistsofthefollowingprocedures:
•“ConguringtheDICOMStorageUserService”onpage95
•“ConguringtheOutboundDICOMDevice(s)intheSystem”onpage96
•“ConguringtheReceivingDICOMDevice(s)toReceiveDICOMTestsfromthe
System”onpage98
ConguringtheDICOMStorageUserService
TheDICOMStorageServiceClassUserservicesendsoutboundDICOMteststoDICOM
StorageServiceClassProviders.TheserviceisconguredwithadefaultAEtitleand
storagecommitmentport.
UsethefollowingstepstoconguretheDICOMStorageUserservice.
1.Logintothesystemapplication.
2.GotoSystem>Setup.
2059568-013JMUSE™v995
Page 96

DICOMCommunication
3.SelectDICOMServices.
4.Right-clickonDICOMSTORESCUandselectProperties.
5.SelectGeneral.
Completethefollowingeldsasappropriate:
Field
AETitleTheDICOMApplication
Storagecommitmentport
(forreceivingresponses)
MaximumWaitTime
Maximumretries
DescriptionAction
EntityTitleoftheSystem
StorageServiceClassUser.
Thedefaultis
MuseStoreSCU.
TheportthattheSystem
StorageUserservice
listensonforstorage
commitmentresponses.
Thedefaultportis105.
Themaximumtime
insecondsthatthe
systemwaitsforstorage
commitmentresponses.
Defaultis60.
Themaximumnumber
ofstoragecommitment
requeststhatthesystem
makesifpreviousrequests
fail.
Defaultis3.
Ifrequired,changethe
defaultbytypinganew
title.
Ifrequired,changethe
defaultbytypinganew
port.
Ifrequired,changethe
defaultbytypinganew
waittime.
Ifrequired,changethe
defaultbytypinganew
maximumnumberof
retries.
6.SelectDICOMAssociationSettingstocongurethesesettings.
Defaultvaluesexistforthesesettingsandtypicallydonotneedtobechanged.
7.ClickOKtosaveyourchanges.
ClickCloseorCanceltoignoreyourchanges.
Ifanychangesweremade,theMUSEDICOMStorageUserservicewillneedto
berestartedonthesystemapplicationserver.
ConguringtheOutboundDICOMDevice(s)intheSystem
EachoutboundDICOMdeviceneedstobeconguredinthesystempriortosending
atestfromthesystemtotheoutboundDICOMdevice.
UsethefollowingstepstocongureanoutboundDICOMdeviceinthesystem.Repeat
thesestepsforeachoutboundDICOMdevicethatneedstobecongured.
1.Logintothesystemapplication.
2.GotoSystem>Setup.
3.SelectDevices.
96MUSE™v92059568-013J
Page 97

4.Performoneofthefollowing:
•Tocreateanewdevice,gotoAction>Newandselectoneofthefollowing
devicetypes.
•DICOMIOD
DICOMInformationObjectDenition.OnlyECGtesttypesmaybesentto
DICOMIODdevices.
•DICOMPDF
DICOMEncapsulatedPDF
Thedevicetypesarebasedonthecompatibilityorpreferenceofthereceiving
DICOMstorageserviceclassprovider.
•TomodifyanexistingDICOMdevice,right-clicktheentryandselect
Properties.
5.SelectGeneralandsetupthefollowingeldsasappropriate:
DICOMCommunication
Field
DeviceName
DeviceAETitleTheDICOMApplicationEntityTitle
IPAddressTheIPaddressofthereceiving
Port
SendOriginalIODif
available
DescriptionAction
Thenameofthesystemdevice.Typethenameofthe
ofthereceivingDICOMstorage
serviceclassprovider.Thisis
theDICOMstorageserviceclass
providerthatreceivestestsfrom
thesystem.
DICOMstorageserviceclass
provider.ThisistheDICOMstorage
serviceclassproviderthatreceives
testsfromthesystem.
Theportnumberofthereceiving
DICOMstorageserviceclass
provider.ThisistheDICOMstorage
serviceclassproviderthatreceives
testsfromthesystem.
Flagthatdetermineswhetherthe
originalunediteddataortheedited
testisexported.
Ifthisoptionisenabled,theoriginal,
initiallyacquireddataisexported.
NOTE:
Thisoptionisonlyavailablefor
DICOMIODdevicetypes.
device.
TypetheAETitleofthe
DICOMstorageservice
classprovider.
TypetheIPaddress.
Typetheportnumber.
Checktheboxto
enablethisoption.
Unchecktheboxto
disabletheoption.
2059568-013JMUSE™v997
Page 98

DICOMCommunication
Field
SupportsStorage
Commitment
StorageCommitment
AETitle
StorageCommitment
IPAddress
StorageCommitment
Port
DescriptionAction
Aagthatenables/disables
storagecommitmentontheMUSE
system.IftheDICOMstorage
serviceclassprovidersupports
storagecommitment,thisboxcan
becheckedtoenableit.
Whenstoragecommitment
isenabled,thesystemsends
storagecommitmentrequests
totheDICOMdevicestorage
commitmentserviceusing
informationconguredhere.
TheDICOMApplicationEntityTitle
ofthestoragecommitmentservice
ontheDICOMdevice.
TheIPaddressofthestorage
commitmentserviceontheDICOM
device.
Theportnumberofthestorage
commitmentserviceontheDICOM
device.
Checktheboxto
enablethisoption.
Unchecktheboxto
disabletheoption.
TypetheAETitleofthe
storagecommitment
serviceontheDICOM
device.
TypetheIPaddress
ofthestorage
commitmentservice
ontheDICOMdevice.
Typetheportnumber
ofthestorage
commitmentservice
ontheDICOMdevice.
6.AfterenteringthedeviceAEtitle,IPaddress,andport,clickDICOMEchotoverify
anassociationcanbemadewiththeDICOMstorageserviceclassprovider.
Ifanassociationcanbeestablished,thefollowingmessageisdisplayed:DICOM
ECHOSuccessful
Ifanassociationcannotbeestablished,thefollowingmessageisdisplayed:
DICOMECHOFailed.Ifafailureoccurs,checkthesettingsandtryagain.
7.SelectHoursofOperationandcongurethemasdesired.
8.SelectAdvancedandcongureValidSitesasdesired.
9.ClickOKtosaveyourchanges.
ClickCloseorCanceltoignoreyourchanges.
ConguringtheReceivingDICOMDevice(s)toReceiveDICOMTestsfrom
theSystem
ToreceiveDICOMtestsfromthesystem,thereceivingDICOMstorageserviceclass
providermightneedtobeconguredwiththeDICOMstorageserviceclassuserAE
titleandIPaddressdenedintheDICOMSTORESCUentryinDICOMServicesinthe
systemsetup.Ifstoragecommitmentisused,specifytheMUSESTORESCUstorage
commitmentportonthereceivingDICOMstorageserviceclassprovider.
98MUSE™v92059568-013J
Page 99

DICOMCommunication
ConguringtheSystemtoQueryforDICOMOrders
UsethefollowingprocedurestocongurethesystemtoqueryforDICOMorders.
1.CongurethesystemDICOMmodalityworklistclientservice.
2.ConguretheDICOMmodalityworklistserviceclassprovidertoreceiveDICOM
orderqueriesfromthesystem.
ConguretheSystemDICOMModalityWorklistClientService
ThesystemDICOMmodalityworklistclientservicequeriesaDICOMmodalityworklist
serviceclassproviderforDICOMorders.
PerformthefollowingtocongurethesystemDICOMmodalityworklistclientservice
toqueryforDICOMorders.Repeatthesestepsforeachsystemsitethatneedsto
queryforDICOMorders.
1.Logontothesystemapplication.
2.GotoSystem>Setup.
3.SelectDICOMServices.
4.Performoneofthefollowingsteps:
a.TocreateanewDICOMMWLSCUservice,gotoAction>New>MWL
Server.
b.TomodifyanexistingDICOMMWLSCUservice,right-clickontheentry
andselectProperties.
5.SelectMWLCongandsetupthefollowingeldsasappropriate:
2059568-013JMUSE™v999
Page 100

DICOMCommunication
ModalityWorklistUserService
Field
AETitleTheDICOMApplicationEntity
DefaultSiteThesystemsitenumber .OrdersreceivedviaDMWLwithavalueinthe
QueryDefaultSite
Only
DescriptionAction
TypetheAETitle.
Titleofthesystemmodality
worklistserviceclassuser .
DICOMInstitutionName(0008,0080)thatmatches
aconguredMUSESitenamearestoredunder
thatsite.
IftheDICOMInstitutionNameiseitherblank,or
doesnotmatchaconguredMUSESitename,the
OrderisstoredinthisDefaultSite.
CheckboxtoenableIfenabled,querieswillbemadeforonlyordersthat
haveaDICOMInstitutionNamethatmatchedthe
nameoftheDefaultMUSESite.
ModalityWorklistProviderConguration
Field
AETitleTheDICOMApplicationEntity
IPAddressTheIPAddressoftheDICOM
Port
DescriptionAction
TitleoftheDICOMmodality
worklistserviceclassprovider
thatistoqueryfororders.
ModalityWorklistServiceClass
Providerthatthesystemisto
queryfororders.
ThePortoftheDICOMModality
WorklistServiceClassProvider
thatthesystemistoqueryfor
orders.
TypetheAETitle.
TypetheIPaddressoftheDICOMModalityWorklist
ServiceClassProvider.
TypethePortoftheDICOMModalityWorklist
ServiceClassProvider.
DefaultQuery
Field
QueryInterval(In
Minutes)
DaysintheFutureThenumberofdaysinthefuture
DaysinthePastThenumberofdaysinthepast
100MUSE™v92059568-013J
DescriptionAction
Thefrequencythesystem
queriesfororders.
(fromcurrenttime)toquery
forScheduledProcedureStart
Date/Time.
Thedefaultisve.
(fromcurrenttime)toquery
forScheduledProcedureStart
Date/Time.
Thedefaultisve.
Typeanumberinminutestospecifythefrequency.
Enterthenumberofdaysinthefuture(from
currenttime)toqueryforScheduledProcedure
StartDate/Time.
Enterthenumberofdaysinthepast(fromcurrent
time)toqueryforScheduledProcedureStart
Date/Time