GE MUSE v8.0 Service Manual_SM_2034539-043_D Invenia ABUS System Setup and Basic Service Manual

Page 1
GEHealthcare
MUSE™CardiologyInformationSystem
ServiceManual
SoftwareVersion8.0 2034539-043D
MUSECardiologyInformationSystem English ©2011-2013GeneralElectricCompany. AllRightsReserved.
Page 2
PublicationInformation
TheinformationinthismanualonlyappliestoMUSE™CardiologyInformationSystemVersion8.0.Itdoesnotapplytoearlierproduct versions.Duetocontinuingproductinnovation,specicationsinthismanualaresubjecttochangewithoutnotice.
MUSEandMarquettearetrademarksownedbyGEMedicalSystemsInformationTechnologies,Inc.,aGeneralElectricCompanygoingto marketasGEHealthcare.Allothertrademarkscontainedhereinarethepropertyoftheirrespectiveowners.
Thedocumentpartnumberandrevisionareoneachpageofthedocument.Therevisionidentiesthedocument’supdatelevel.Therevision historyofthisdocumentissummarizedinthefollowingtable.
Revision
A
B
C
D
ToaccessotherGEHealthcareDiagnosticCardiologymanuals,gototheCommonDocumentationLibrary(CDL),locatedat
www.gehealthcare.com/documents,andclickCardiology.
ToaccessOriginalEquipmentManufacturer(OEM)manuals,gotothedevicemanufacturer'sWebsite.
Date
20March2011InitialReleaseofthismanual.
14April2011UpdatestoresolveSPR66469.
22April2012UpdatedtheHL7SystemStateBackupand
5March2013UpdatedfortheHPrp5800platform.
Comment
RecoverychapterwithinstructionsfortheG7 server.UpdatesperSPR133402totheHL7System StateBackupandRecoverychapterinsections: BeforeYouBegin(rstparagraph3rdsentence change),RAIDConrguration(changestosteps6 and14),andApplyingtheServerImage(changed step10).
UpdatedcontraindicationsperSFDArequest.
ServiceManualLanguageInformation
WARNING (EN)
ПРЕДУПРЕЖДЕНИЕ (BG)
警告
ZH-CN
ThisservicemanualisavailableinEnglishonly.
Ifacustomer'sserviceproviderrequiresalanguageotherthanEnglish,itisthecustomer's
responsibilitytoprovidetranslationservices.
Donotattempttoservicetheequipmentunlessthisservicemanualhasbeenconsulted
andisunderstood.
Failuretoheedthiswarningmayresultininjurytotheserviceprovider ,operator ,orpatient,
fromelectricshock,mechanicalorotherhazards.
Товаупътванезаработаеналичносамонаанглийскиезик.
Акодоставчикътнауслугатанаклиентаизискадругезик,задължениенаклиентае
даосигурипревод.
Неизползвайтеоборудването,предидастесеконсултиралииразбралиупътването
заработа.
Неспазванетонатовапредупреждениеможедадоведедонараняваненадоставчика
науслугата,оператораилипациентврезултатнатоковударилимеханичнаили другаопасност.
本维修手册仅提供英文版本。
如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
未详细阅读和完全理解本维修手册之前,不得进行维修。
忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的伤害。
2
MUSE™CardiologyInformationSystem
2034539-043D
5March2013
Page 3
ServiceManualLanguageInformation(cont'd.)
警告
(ZH-TW)
UPOZORENJE (HR)
VAROVÁNÍ (CS)
ADVARSEL (DA)
WAARSCHUWING (NL)
HOIATUS (ET)
本維修手冊只提供英文版。
如果客戶的維修人員有英語以外的其他語言版本需求,則由該客戶負責提供翻
譯服務。
除非您已詳閱本維修手冊並了解其內容,否則切勿嘗試對本設備進行維修。
不重視本警告可能導致維修人員、操作人員或病患因電擊、機械因素或其他因素
而受到傷害。
Oveuputezaservisiranjedostupnesusamonaengleskomjeziku.
Ukolikokorisničkiserviszahtijevanekidrugijezik,korisnikovajeodgovornostosigurati
odgovarajućiprijevod.
Nemojtepokušavatiservisiratiopremuukolikonistekonzultiraliirazumjelioveupute.
Nepoštivanjeovogupozorenjamožerezultiratiozljedamaservisnogosoblja,korisnikaili
pacijentaprouzročenimelektričnimudaromtemehaničkimilinekimdrugimopasnostima.
Tentoprovoznínávodexistujepouzevanglickémjazyce.
Vpřípadě,žeexterníslužbazákazníkůmpotřebujenávodvjinémjazyce,jezajištění
překladudoodpovídajícíhojazykaúkolemzákazníka.
Nesnažteseoúdržbutohotozařízení,anižbystesipřečetlitentoprovoznínávoda
pochopilijehoobsah.
Vpřípaděnedodržovánítétovarovánímůžedojítkporaněnípracovníkaprodejního
servisu,obslužnéhopersonálunebopacientůvlivemelektrickéhoproudu,respektive vlivemmechanickýchčijinýchrizik.
Denneservicemanualndeskunpåengelsk.
Hvisenkundesteknikerharbrugforetandetsprogendengelsk,erdetkundensansvar
atsørgeforoversættelse.
Forsøgikkeatservicereudstyretmedmindredenneservicemanualharværetkonsulteret
ogerforstået.
Manglendeoverholdelseafdenneadvarselkanmedføreskadepågrundafelektrisk,
mekaniskellerandenfareforteknikeren,operatørenellerpatienten.
DezeservicemanualisalleeninhetEngelsverkrijgbaar.
Indienhetonderhoudspersoneeleenanderetaalnodigheeft,danisdeklant
verantwoordelijkvoordevertalingervan.
Probeerdeapparatuurnietteonderhoudenvoordatdezeservicemanualgeraadpleegd
enbegrepenis.
Indiendezewaarschuwingnietwordtopgevolgd,zouhetonderhoudspersoneel,de
gebruikerofeenpatiëntgewondkunnenrakenalsgevolgvaneenelektrischeschok, mechanischeofanderegevaren.
Käesolevteenindusjuhendonsaadavalainultinglisekeeles.
Kuiklienditeeninduseosutajanõuabjuhenditinglisekeelesterinevaskeeles,vastutab
klienttõlketeenuseosutamiseeest.
Ärgeüritageseadmeidteenindadaenneeelnevaltkäesolevateenindusjuhendiga
tutvumistjasellestarusaamist.
Käesolevahoiatuseeiraminevõibpõhjustadateenuseosutaja,operaatorivõipatsiendi
vigastamistelektrilöögi,mehaanilisevõimuuohutagajärjel.
2034539-043D
MUSE™CardiologyInformationSystem
3
Page 4
ServiceManualLanguageInformation(cont'd.)
VAROITUS (FI)
ATTENTION (FR)
WARNUNG (DE)
ΠΡΟΕΙΔΟΠΟΙΗΣΗ (GR)
FIGYELMEZTETÉS (HU)
AÐVÖRUN (IS)
Tämähuolto-ohjeonsaatavillavainenglanniksi.
Josasiakkaanhuoltohenkilöstövaatiimuutakuinenglanninkielistämateriaalia,tarvittavan
käännöksenhankkiminenonasiakkaanvastuulla.
Äläyritäkorjatalaitteistoaennenkuinoletvarmastilukenutjaymmärtänyttämän
huolto-ohjeen.
Mikälitätävaroitustaeinoudateta,seurauksenavoiollahuoltohenkilöstön,laitteiston
käyttäjäntaipotilaanvahingoittuminensähköiskun,mekaanisenviantaimuun vaaratilanteenvuoksi.
Cemanueltechniquen'estdisponiblequ'enanglais.
Siunservicetechniqueclientsouhaiteobtenircemanueldansuneautrelangueque
l'anglais,ildevraprendreenchargelatraductionetlaresponsabilitéducontenu.
Nepastenterd'intervenirsurleséquipementstantquelemanueltechniquen'apasété
consultéetcompris.
Lenon-respectdecetavertissementpeutentraînerchezletechnicien,l'opérateuroule
patientdesblessuresduesàdesdangersélectriques,mécaniquesouautres.
DieseServiceanleitungistnurinenglischerSpracheverfügbar .
FallsderKundendiensteineandereSprachebenötigt,musserfüreineentsprechende
Übersetzungsorgen.
KeineWartungdurchführen,ohnedieseServiceanleitunggelesenundverstandenzu
haben.
BeiZuwiderhandlungkanneszuVerletzungendesKundendiensttechnikers,des
AnwendersoderdesPatientendurchStromschläge,mechanischeodersonstigeGefahren kommen.
Τοπαρόνεγχειρίδιοσέρβιςδιατίθεταιστααγγλικάμόνο.
Εάντοάτομοπαροχήςσέρβιςενόςπελάτηαπαιτείτοπαρόνεγχειρίδιοσεγλώσσαεκτός
τωναγγλικών,αποτελείευθύνητουπελάτηναπαρέχειυπηρεσίεςμετάφρασης.
Μηνεπιχειρήσετετηνεκτέλεσηεργασιώνσέρβιςστονεξοπλισμόεκτόςεάνέχετε
συμβουλευτείκαιέχετεκατανοήσειτοπαρόνεγχειρίδιοσέρβις.
Εάνδενλάβετευπόψητηνπροειδοποίησηαυτή,ενδέχεταιναπροκληθείτραυματισμός
στοάτομοπαροχήςσέρβις,στοχειριστήήστονασθενήαπόηλεκτροπληξία,μηχανικούς ήάλλουςκινδύνους.
Ezaszervizkézikönyvkizárólagangolnyelvenérhetőel.
Haavevőszervizellátójaangoltóleltérőnyelvretartigényt,akkoravevőfelelőssége
afordításelkészíttetése.
Nepróbáljaelkezdenihasználniaberendezést,amígaszervizkézikönyvbenleírtakat
nemértelmeztékésértettékmeg.
Ezengyelmeztetésgyelmenkívülhagyásaaszervizellátó,aműködtetővagyapáciens
áramütés,mechanikaivagyegyébveszélyhelyzetmiattisérüléséteredményezheti.
Þessiþjónustuhandbókereingöngufáanlegáensku.
Efaðþjónustuveitandiviðskiptamannsþarfnastannarstungumálsenensku,erþaðskylda
viðskiptamannsaðskaffatungumálaþjónustu.
Reyniðekkiaðafgreiðatækiðnemaþessiþjónustuhandbókhefurveriðskoðuðogskilin.
Brotáaðsinnaþessariaðvörungeturleitttilmeiðslaáþjónustuveitanda,stjórnandaeða
sjúklingifráraosti,vélrænumeðaöðrumáhættum.
4
MUSE™CardiologyInformationSystem
2034539-043D
5March2013
Page 5
ServiceManualLanguageInformation(cont'd.)
PERINGATAN (ID)
AVVERTENZA (IT)
警告
(JA)
경고
(KO)
BRĪDINĀJUMS (LV)
ĮSPĖJIMAS (LT)
ManualservisinihanyatersediadalambahasaInggris.
JikapenyediajasaservispelangganmemerlukanbahasalainselaindariBahasaInggris,
merupakantanggungjawabdaripenyediajasaservistersebutuntukmenyediakan terjemahannya.
Janganmencobamelakukanservisterhadapperlengkapankecualitelahmembacadan
memahamimanualservisini.
Mengabaikanperingataninibisamengakibatkancederapadapenyediaservis,operator ,
ataupasien,karenaterkenakejutlistrik,bahayamekanisataubahayalainnya.
IlpresentemanualedimanutenzioneèdisponibilesoltantoinInglese.
Seunaddettoallamanutenzionerichiedeilmanualeinunalinguadiversa,ilclienteè
tenutoaprovvederedirettamenteallatraduzione.
Siprocedaallamanutenzionedell'apparecchiaturasolodopoaverconsultatoilpresente
manualeedavernecompresoilcontenuto.
Ilnonrispettodellapresenteavvertenzapotrebbefarcompiereoperazionidacuiderivino
lesioniall'addetto,allamanutenzione,all'utilizzatoreedalpazienteperfolgorazione elettrica,perurtimeccaniciodaltririschi.
このサービスマニュアルは英語版しかありません。
サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業
者の責任で行うものとさせていただきます。
このサービスマニュアルを熟読し、十分に理解をした上で装置のサービスを
行ってください。
この警告に従わない場合、サービスを担当される方、操作員あるいは患者が、感電
や機械的又はその他の危険により負傷する可能性があります。
본서비스지침서는영어로만이용하실수있습니다.
고객의서비스제공자가영어이외의언어를요구할경우,번역서비스를제공하는것
은고객의책임입니다.
본서비스지침서를참고했고이해하지않는한은해당장비를수리하려고시도하
지마십시오.
이경고에유의하지않으면전기쇼크,기계상의혹은다른위험으로부터서비스제
공자,운영자혹은환자에게위해를가할수있습니다.
Šīapkalpotājurokasgrāmatairpieejamatikaiangļuvalodā.
Jaapkalpošanassniedzējamnepieciešamainformācijacitā,nevisangļu,valodā,klienta
pienākumsirnodrošināttāstulkošanu.
Neveicietaprīkojumaapkopi,neizlasotunnesaprototapkalpotājurokasgrāmatu.
Šībrīdinājumaneievērošanavarradītelektriskāsstrāvastrieciena,mehāniskuvaiciturisku
izraisītutraumuapkopessniedzējam,operatoramvaipacientam.
Šiseksploatavimovadovasyraprieinamastikanglųkalba.
Jeiklientopaslaugųtiekėjasreikalaujavadovokitakalba-neanglų,numatytivertimo
paslaugasyraklientoatsakomybė.
Nemėginkiteatliktiįrangostechninėspriežiūros,nebentatsižvelgėteįšįeksploatavimo
vadovąirjįsupratote.
Jeineatkreipsitedėmesioįšįperspėjimą,galimisužalojimaidėlelektrosšoko,mechaninių
arkitųpaslaugųtiekėjui,operatoriuiarpacientui.
2034539-043D
MUSE™CardiologyInformationSystem
5
Page 6
ServiceManualLanguageInformation(cont'd.)
ADVARSEL (NO)
Denneservicehåndbokennnesbarepåengelsk.
Hviskundensserviceleverandørtrengeretannetspråk,erdetkundensansvaråsørge
foroversettelse.
Ikkeforsøkåreparereutstyretutenatdenneservicehåndbokenerlestogforstått.
Manglendehensyntildenneadvarselenkanføretilatserviceleverandøren,operatøren
ellerpasientenskadespågrunnavelektriskstøt,mekaniskeellerandrefarer .
OSTRZEŻENIE (PL)
Niniejszypodręcznikserwisowydostępnyjestjedyniewjęzykuangielskim.
Jeślidostawcausługklientawymagajęzykainnegoniżangielski,zapewnienieusługi
tłumaczeniajestobowiązkiemklienta.
Nienależyserwisowaćwyposażeniabezzapoznaniasięizrozumienianiniejszego
podręcznikaserwisowego.
Niezastosowaniesiędotegoostrzeżeniamożespowodowaćurazydostawcyusług,
operatoralubpacjentawwynikuporażeniaelektrycznego,zagrożeniamechanicznego bądźinnego.
AVISO (PT-BR)
Estemanualdeassistênciatécnicasóseencontradisponíveleminglês.
SeoserviçodeassistênciatécnicadoclientenãoforGE,eprecisardeoutroidioma,será
daresponsabilidadedoclientefornecerosserviçosdetradução.
Nãotenterepararoequipamentosemterconsultadoecompreendidoestemanualde
assistênciatécnica.
Onãocumprimentodesteavisopodeporemperigoasegurançadotécnico,operadorou
pacientedevidoachoqueselétricos,mecânicosououtros.
AVISO (PT-PT)
Estemanualtécnicosóseencontradisponíveleminglês.
Seaassistênciatécnicadoclientesolicitarestesmanuaisnoutroidioma,éda
responsabilidadedoclientefornecerosserviçosdetradução.
Nãotenterepararoequipamentosemterconsultadoecompreendidoestemanual
técnico.
Onãocumprimentodesteavisopodeprovocarlesõesaotécnico,aoutilizadorouao
pacientedevidoachoqueseléctricos,mecânicosououtros.
AVERTISMENT (RO)
Acestmanualdeserviceestedisponibilnumaiînlimbaengleză.
Dacăunfurnizordeserviciipentruclienţinecesităoaltălimbădecâtceaengleză,estede
datoriaclientuluisăfurnizezeotraducere.
Nuîncercaţisăreparaţiechipamentuldecâtulteriorconsultăriişiînţelegeriiacestui
manualdeservice.
Ignorareaacestuiavertismentarputeaducelarănireadepanatorului,operatoruluisau
pacientuluiînurmapericolelordeelectrocutare,mecanicesaudealtănatură.
ПРЕДУПРЕЖДЕНИЕ (RU)
Настоящееруководствопообслуживаниюпредлагаетсятольконаанглийскомязыке.
Еслисервисномуперсоналуклиентанеобходиморуководствоненаанглийском,ана
каком-тодругомязыке,клиентуследуетобеспечитьпереводсамостоятельно.
Преждечемприступатькобслуживаниюоборудования,обязательнообратитеськ
настоящемуруководствуивнимательноизучитеизложенныевнемсведения.
Несоблюдениетребованийданногопредупрежденияможетпривестиктому,что
специалистыпообслуживанию,операторыилипациентыполучатударэлектрическим током,механическуютравмуилидругоеповреждение.
6
MUSE™CardiologyInformationSystem
2034539-043D
5March2013
Page 7
ServiceManualLanguageInformation(cont'd.)
UPOZORENJE (SR)
VAROVANIE (SK)
OPOZORILO (SL)
ADVERTENCIA (ES)
VARNING (SV)
UYARI (TR)
Ovoservisnouputstvojedostupnosamonaengleskomjeziku.
Akoklijentovserviserzahtevanekidrugijezik,klijentjedužandaobezbediprevodilačke
usluge.
Nepokušavajtedaopraviteuređajakonistepročitaliirazumeliovoservisnouputstvo.
Zanemarivanjeovogupozorenjamožedovestidopovređivanjaservisera,rukovaocaili
pacijentausledstrujnogudara,ilimehaničkihidrugihopasnosti.
Tentonávodnaobsluhujekdispozíciilenvangličtine.
Akzákazníkovposkytovateľslužiebvyžadujeinýjazykakoangličtinu,poskytnutie
prekladateľskýchslužiebjezodpovednosťouzákazníka.
Nepokúšajtesaoobsluhuzariadeniaskôr,akosineprečítatenávodnaobsluhua
neporozumietemu.
Zanedbanietohtovarovaniamôževyústiťdozraneniaposkytovateľaslužieb,obsluhujúcej
osobyalebopacientaelektrickýmprúdom,mechanickýmaleboinýmnebezpečenstvom.
Taservisnipriročnikjenavoljosamovangleškemjeziku.
Čeponudnikstoritvestrankepotrebujepriročnikvdrugemjeziku,morastrankazagotoviti
prevod.
Neposkušajteservisiratiopreme,četegapriročnikanistevcelotiprebraliinrazumeli.
Četegaopozorilaneupoštevate,selahkozaradielektričnegaudara,mehanskihalidrugih
nevarnostipoškodujeponudnikstoritev,operateralibolnik.
Estemanualdeserviciosóloexisteeninglés.
Sielencargadodemantenimientodeunclientenecesitaunidiomaquenoseaelinglés,el
clientedeberáencargarsedelatraduccióndelmanual.
Nosedeberádarserviciotécnicoalequipo,sinhaberconsultadoycomprendidoeste
manualdeservicio.
Lanoobservanciadelpresenteavisopuededarlugaraqueelproveedordeservicios,el
operadoroelpacientesufranlesionesprovocadasporcausaseléctricas,mecánicaso deotranaturaleza.
Denhärservicehandbokennnsbaratillgängligpåengelska.
Omenkundsserviceteknikerharbehovavettannatspråkänengelskaansvararkunden
föratttillhandahållaöversättningstjänster.
Försökinteutföraservicepåutrustningenomduinteharlästochförstårdenhär
servicehandboken.
Omduintetarhänsyntilldenhärvarningenkandetresulteraiskadorpåserviceteknikern,
operatörenellerpatiententillföljdavelektriskastötar,mekaniskafarorellerandrafaror.
Buserviskılavuzununsadeceİngilizcesimevcuttur.
Eğermüşteriteknisyenibukılavuzuİngilizcedışındabirbaşkalisandantalepederse,bunu
tercümeettirmekmüşteriyedüşer.
Serviskılavuzunuokuyupanlamadanekipmanlaramüdahaleetmeyiniz.
Buuyarıyauyulmaması,elektrik,mekanikveyadiğertehlikelerdendolayıteknisyen,
operatörveyahastanınyaralanmasınayolaçabilir .
2034539-043D
MUSE™CardiologyInformationSystem
7
Page 8
ServiceManualLanguageInformation(cont'd.)
ЗАСТЕРЕЖЕННЯ (UK)
CNHBÁO (VI)
Данекерівництвозсервісногообслуговуванняпостачаєтьсявиключноанглійськоюмовою.
Якщосервіснийінженерпотребуєкерівництвоіншоюмовою,користувачзобов'язаний
забезпечитипослугиперекладача.
Ненамагайтесяздійснюватитехнічнеобслуговуванняданогообладнання,якщо
винечитали,абонезрозумілиінформацію,наданувкерівництвізсервісного обслуговування.
Недотриманняцьогозастереженняможепризвестидотравмуваннясервісного
інженера,користувачаданогообладнанняабопацієнтавнаслідокелектричногошоку, механічногоушкодженняабозіншихпричинневірногообслуговуванняобладнання.
TàiLiuHướngDnSaChachỉcóbntiếngAnh.
Nếucácđơnvịcungcpdchvụchokháchhàngyêucumtngônngữnàokháctiếng
Anh,thìkháchhàngsẽcótráchnhimcungcpcácdchvụdchthut.
KhôngđượcsachathiếtbịtrừkhiđãthamkhovàhiuTàiliuHướngdnSacha.
Khôngtuânthủnhngcnhbáonàycóthểdếncáctnthươngchongườithc
hinsacha,ngườivnhànhhaybnhnhân,doscđin,cácrirovềcơkhíhay cácrirokhác.
8
MUSE™CardiologyInformationSystem
2034539-043D
5March2013
Page 9
Contents
1Introduction
IndicationsforUse....................................................................................15
Contraindications.....................................................................................15
SystemAccuracy.......................................................................................15
PrescriptionDeviceStatement.................................................................15
RegulatoryandSafetyInformation...........................................................16
SafetyConventions..................................................................................16
SafetyHazards........................................................................................16
RFCaution.............................................................................................19
EMI/EMCRequirements.............................................................................19
EquipmentCompliance.............................................................................20
PartsandAccessoriesInformation................................................................21
ResponsibilityoftheManufacturer................................................................22
ResponsibilityofthePurchaser/Customer.......................................................22
Symbols................................................................................................22
Training.....................................................................................................25
ServiceInformation...................................................................................25
ServiceRequirements...............................................................................25
Customer-suppliedHardware......................................................................26
HardwareSuppliedbyGEHealthcare............................................................26
SecurityUpdates.....................................................................................26
AdditionalAssistance................................................................................26
ManualInformation..................................................................................26
IntendedAudience...................................................................................27
ManualPurpose......................................................................................27
DocumentConventions.............................................................................27
RelatedDocuments...................................................................................28
2ProductOverview
GeneralOperation....................................................................................29
DataAcquisition........................................................................................30
DeviceInterfaces......................................................................................31
MACCarts..............................................................................................31
MARS...................................................................................................32
CASE/CardioSoftStressSystems..................................................................32
MonitoringSystems..................................................................................32
Modems................................................................................................32
2034539-043D
MUSE™CardiologyInformationSystem
9
Page 10
HL7Interface............................................................................................33
OptionalFeatures.....................................................................................34
HardwareSpecications...........................................................................37
SoftwareSpecications.............................................................................37
TheMUSESysteminVirtualEnvironments...............................................37
MUSEv8DriveContentsandSupportingFolders.....................................38
MUSEServices...........................................................................................40
RequiredNetworkPorts............................................................................41
3SystemSetup
SettingUpModems...................................................................................43
CustomerResponsibilities...........................................................................38
DataAcquisition......................................................................................38
HL7InterfaceVirtualMachine.....................................................................38
MuseApplicationandDatabaseonSameServer..............................................39
MUSEApplicationandDatabaseonSeparateServers........................................39
Verifying/InstallingtheMUSEModemServiceandWireless/LAN
Communication.......................................................................................43
SettingUpaModemDevice........................................................................44
RestartingModems..................................................................................45
TestingCSINetworkModemConnections.......................................................47
InstallingandConguringtheXMLImportOption...................................48
InstallingtheXMLImportOptionandMUSEXMLService.....................................48
Requirements.........................................................................................48
UsingXMLCONFIG.EXEtoConguretheOptiononMUSE.....................................48
SettingUptheMUSEFileServerforFileCopy...........................................50
SettingUpaFolder...................................................................................51
MUSEAuthentication—EnableorDisable.................................................51
ConguringWindowsHighContrastColorSchemeonMUSE
Client.........................................................................................................52
ChangingMUSEServiceAccounts.............................................................53
MUSEUserServiceAccounts.......................................................................53
MUSEWindowsUserServiceAccounts..........................................................53
ChangingtheMUSEUserAccountPasswords..................................................53
4SystemAdministration
RemoteDatabases....................................................................................55
SQLPorts..............................................................................................55
SQLRemoteConnections...........................................................................56
AutoShutdown..........................................................................................56
ActivatingSystemShutdown.......................................................................56
CancelingtheSystemShutdown..................................................................58
10
MUSE™CardiologyInformationSystem
2034539-043D
Page 11
5Maintenance
ModifyingtheMUSEInstalledConguration............................................58
MovingtheMUSEDatabases....................................................................59
PreparingtheNewDatabaseServer..............................................................60
DetachingandCopyingtheMSUEDatabasesfromtheCurrentServer....................60
AttachingtheMUSEDatabasestotheNewSQLServer.......................................60
Uninstalling/ReinstallingtheMUSEApplication.......................................61
UninstallingtheMUSEApplication................................................................61
ReinstallingtheMUSEApplication.................................................................61
RenamingtheMUSEServer.......................................................................62
ChangingthePortNumberoftheFileServer...........................................63
ChangingthePortNumberontheMUSEServer...............................................63
ChangingthePortNumberontheMUSEClients...............................................64
BIOSUpdates............................................................................................64
UpgradingtheG4ServerBIOSFirmware........................................................64
MaintenanceGuidelines............................................................................67
OEMMaintenance....................................................................................68
RequiredToolsandSupplies........................................................................68
InspectionandCleaning...........................................................................68
SafeShutdownProcedures.........................................................................69
Precautions............................................................................................69
VisualInspection......................................................................................70
CheckCoolingFansandVentilation..............................................................70
ExteriorCleaning.....................................................................................71
FunctionalCheckoutProcedures..............................................................71
HardwareFRURepairs..............................................................................71
Non-FRURepairs.....................................................................................76
6HL7SystemStateBackupandRecovery
CopyingtheSystemStateBackup............................................................83
DisasterRecovery.....................................................................................83
BeforeYouBegin.....................................................................................83
RebuildingtheHPDL360G5Server..........................................................84
ConguringtheServer...............................................................................84
ConguringtheRAID.................................................................................85
InstallingWindowsServer2003...................................................................86
ConguringDisplaySettings.......................................................................87
SettingUptheWindowsEnvironment............................................................87
InstallingWindows2003Servicepack2.........................................................87
RebuildingtheHPDL360G7Server..........................................................87
VerifytheBootSequence...........................................................................88
RAIDConguration...................................................................................88
ApplyingtheServerImage..........................................................................89
VerifyHelpandSupportService...............................................................90
2034539-043D
MUSE™CardiologyInformationSystem
11
Page 12
ConguringSNMP.....................................................................................90
InstallingCCG............................................................................................91
InstallingInSiteExC...................................................................................91
RestoringSystemState.............................................................................91
7MUSESystemBackupandRecovery
Introduction..............................................................................................93
BackupOptions.........................................................................................94
OverviewoftheMUSEDiskandDataStructures...............................................94
OverviewoftheNetworkandTapeBackupOptions...........................................95
SettingUpAutomaticBackups.................................................................97
CreatingBackupJobs...............................................................................98
ConguringtheBackupJobs.......................................................................99
ChangingtheBackupSchedule.................................................................102
CreatingNotications.............................................................................103
TestingtheBackupJobs..........................................................................107
BackingUptheDatabaseManually........................................................109
DatabaseRecovery.................................................................................109
CopyingtheCurrentDatabase...................................................................110
RecoveringtheMUSEDatabase.................................................................110
RevertingtotheCurrentDatabase..............................................................115
AdditionalInformation............................................................................115
SystemShutdownandRestartProcedure.....................................................115
InitializingaNewTape.............................................................................116
ASystemRecovery
UsingRecoveryConsole..........................................................................117
ServerRecoveryUsingaNetworkBackup..............................................118
ReimagingtheHPML370G5Server............................................................118
ReimagingtheHPDL370G6Server............................................................119
RecoveringtheSystemPartition(C:Drive).....................................................122
AddingtheServertotheDomain................................................................123
RecoveringtheDataPartition(D:Drive)........................................................124
RestoringtheMUSEDatabases..................................................................125
VerifyingtheMUSECommunications...........................................................125
RestoringOtherFunctionality....................................................................125
ServerRecoveryUsingaTapeBackup....................................................125
UsingAutomatedSystemRecovery.............................................................126
ConguringtheFileServerRAIDArray.........................................................126
RecoveringtheSystemPartition.................................................................128
RecoveringtheMUSEDatabases................................................................129
OtherTaskstoPerformAfterSystemRestore.................................................130
ClientRebuild.........................................................................................130
SettingSATAEmulationandBootOrder(AllCongurations)................................131
LoadingtheImage.................................................................................132
12
MUSE™CardiologyInformationSystem
2034539-043D
Page 13
ConguringtheOperatingSystem..............................................................133
ActivatingtheOperatingSystem................................................................136
ReinstallingtheMUSEApplication...............................................................137
BElectromagneticCompatibility
ElectromagneticCompatibilityforDL370G6Server...............................139
ElectromagneticCompatibility(EMC)Requirements.........................................139
GuidanceandManufacturer’sDeclaration–ElectromagneticEmissions.................141
GuidanceandManufacturer’sDeclaration–ElectromagneticImmunity.................141
GuidanceandManufacturer’sDeclaration–ElectromagneticImmunity.................143
RecommendedSeparationDistancesBetweenPortableandMobileRF
CommunicationsEquipmentandtheMUSEDL370G6Server.............................144
EMCException(s)Disclosure......................................................................145
ElectromagneticCompatibilityforML370G5Server..............................146
ElectromagneticCompatibilityRequirements(EMC).........................................146
GuidanceandManufacturer’sDeclaration—ElectromagneticEmissions................147
GuidanceandManufacturer’sDeclaration—ElectromagneticImmunity................148
GuidanceandManufacturer’sDeclaration—ElectromagneticImmunity................150
RecommendedSeparationDistancesBetweenPortableandMobileRF
CommunicationsEquipmentandtheMUSEDL360G5Server.............................152
EMCException(s)Disclosure......................................................................152
CNationalHealthServiceofGreatBritain(NHS)Patient
Identiers
Purpose...................................................................................................155
Overview.................................................................................................155
NumberValidation..................................................................................155
NumberVerication...............................................................................156
SearchingbyPatientID(PID)...................................................................157
DisplayingthePatientID........................................................................157
UpdatingLegacySystemData...............................................................158
InstallingtheNHSNumberFeature........................................................159
ModulesandFilesAffected.....................................................................159
DGlossary
................................................................................................................163
2034539-043D
MUSE™CardiologyInformationSystem
13
Page 14
14
MUSE™CardiologyInformationSystem
2034539-043D
Page 15

Introduction

Thischapterprovidesgeneralinformationrequiredfortheproperuseofthesystem andthismanual.Familiarizeyourselfwiththisinformationbeforeusingthesystem.

IndicationsforUse

TheMUSECardiologyInformationSystemisintendedtostore,accessandmanage cardiovascularinformationonadultandpediatricpatients.Theinformationconsists ofmeasurements,text,anddigitizedwaveforms.TheMUSECardiologyInformation Systemprovidestheabilitytoreviewandeditelectrocardiographicprocedureson screen,throughtheuseofreviewing,measuring,andeditingtoolsincludingECGserial comparison.TheMUSECardiologyInformationSystemisintendedtobeusedunder thedirectsupervisionofalicensedhealthcarepractitioner,bytrainedoperatorsina hospitalorfacilityprovidingpatientcare.
1

Contraindications

WARNING:
CONTRAINDICATION:Thesystemisnotintendedforpediatricserialcomparison.
WARNING:
CONTRAINDICATION:Thesystemisnotintendedforreal-timepatientmonitoring.

SystemAccuracy

AccuracyoftheMUSEsystemis100%datareproduction,dependentonzoom settings,andresolutionofdisplayand/orprinterbeingused.

PrescriptionDeviceStatement

CAUTION:
UnitedStatesfederallawrestrictsthisdevicetosalebyorontheorderofa physician.
2034539-043D
MUSE™CardiologyInformationSystem
15
Page 16
Introduction

RegulatoryandSafetyInformation

Thissectionprovidesinformationaboutthesafeuseandregulatorycomplianceof thissystem.Familiarizeyourselfwiththisinformation,andreadandunderstandall instructionsbeforeattemptingtousethissystem.Thesystemsoftwareisconsidered medicalsoftware.Assuch,itwasdesignedandmanufacturedtotheappropriate medicalregulationsandcontrols.
NOTE:
Disregardingthesafetyinformationprovidedinthismanualisconsidered abnormaluseofthissystemandcouldresultininjury,dataloss,oravoided warranty.

SafetyConventions

AHazardisasourceofpotentialinjurytoaperson,property,orthesystem. ThismanualusesthetermsDANGER,WARNING,CAUTION,andNOTICEtopointout
hazardsandtodesignateadegreeorlevelofseriousness.Familiarizeyourselfwith thefollowingdenitionsandtheirsignicance.
DenitionsofSafetyConventions

SafetyHazards

Thefollowingmessagesapplytothesystemasawhole.Specicmessagesmayalso beprovidedelsewhereinthemanual.
DANGER:
Safety Convention
DANGER
WARNING
CAUTION
NOTICE
EXPLOSIONHAZARD:Flammableanestheticvaporsorliquidscancause explosions.
DoNOTuseinthepresenceofammableanestheticvaporsorliquids.
Denition
Indicatesanimminenthazard,which,ifnotavoided,willresultindeath orseriousinjury.
Indicatesapotentialhazardorunsafepractice,which,ifnotavoided, couldresultindeathorseriousinjury.
Indicatesapotentialhazardorunsafepractice,which,ifnotavoided, couldresultinmoderateorminorinjury.
Indicatesapotentialhazardorunsafepractice,which,ifnotavoided, couldresultinthelossordestructionofpropertyordata.
WARNING:
ACCIDENTALSPILLS:Ifliquidshaveenteredthesystem,takeitoutofserviceand haveitcheckedbyaservicetechnicianbeforeitisusedagain.
Toavoidelectricshockordevicemalfunction,liquidsmustnotbeallowedto enterthesystem.
16
MUSE™CardiologyInformationSystem
2034539-043D
Page 17
WARNING:
CONNECTIONTOMAINS:ThisisClassIequipment. Themainsplugmustbeconnectedtoanappropriatepowersupply. Theleservermaycontainaninternalbatterypack.Thereisriskofreandburns
ifthebatterypackisnothandledproperly.Toreducetheriskofpersonalinjury:
•Donotattempttorechargethebattery.
•Donotexposetotemperatureshigherthan60°C(140°F).
•Donotdisassemble,crush,puncture,shortexternalcontactsordisposeofin reorwater.
•Replaceonlywiththeappropriatesparepartsdesignatedforthisproduct.
WARNING:
DATALOSSORCORRUPTION:DONOTloadanysoftwareotherthanthatspecied byGEHealthcareontotheMUSEsystem.Installationofsoftwarenotspeciedby GEHealthcaremaycausedamagetotheequipment,orlossorcorruptionofdata.
WARNING:
DELAYINTREATMENT:Thisdeviceisnotintendedforreal-timemonitoring.
Introduction
WARNING:
ELECTRICSHOCK:Toreducetheriskofelectricshock,doNOTremovecover(or back).
Referservicingtoqualiedpersonnel.
WARNING:
ELECTRICSHOCK:Improperuseofthisdevicepresentsashockhazard.Strictly observethefollowingwarnings.Failuretodosomayendangerthelivesofthe user,andbystanders.
Whendisconnectingthedevicefromthepowerline,removetheplugfromthe walloutletrst,beforedisconnectingthecablefromthedevice;otherwise,thereis ariskofcomingincontactwithlinevoltagebyinadvertentlyintroducingmetal partsinthesocketsofthepowercord.
Devicesmaybeconnectedtootherdevicesortopartsofsystemsonlyafter makingcertainthatthereisnodangertothepatient,theoperators,orthe environmentasaresult.StandardsEN/IEC60601-1-1mustbecompliedwithin allcases.
WARNING:
EMIPERFORMANCEHAZARD:UsersshouldbeawareofknownRadioFrequency (RF)sources,suchas:
•radioandTVstations
•portableandmobileRFcommunicationdevices(cellphones,two-wayradios)
andconsiderthemwheninstallingthemedicaldeviceorsystem. Seethe"Electromagneticcompatibility"sectionfoundeitherinthismanualand/or
theservicemanualforrecommendedseparationdistances. Addingaccessoriesorcomponents,ormodifyingthemedicaldeviceorsystem
maydegradetheElectromagneticInterference(EMI)performance.Consultwith qualiedpersonnelregardingchangestothesystemconguration.
2034539-043D
MUSE™CardiologyInformationSystem
17
Page 18
Introduction
WARNING:
EXPLOSIONHAZARD:Flammableanestheticvaporsorliquidscancause explosions.
DoNOTuseinthepresenceofammableanestheticvaporsorliquids.
WARNING:
INCORRECTTREATMENT:FailuretohaveauniquePatientID(PID)inpatient demographicsmaycauseincorrectpatientdataassociatedwiththePID.
AlwaysassigntheproperPIDandnamebeforetransmissiontotheMUSEsystem. DonotconrmpatientrecordscontainingdefaultPIDs.Formoreinformation concerningPIDs,seetheMUSECardiologyInformationSystemOperator's Manual.
WARNING:
INCORRECTTREATMENT:Someofthecommunicationsprotocolsusedinthis product(CSIandDCP)donotprovideencryptionorauthenticationatthistime. TheseprotocolsareusedtosendclinicaldatatothesystemfromECGcartsand otherclinicaldevices.
Youshouldtakeappropriatestepstosecuretheprivacyofcommunicationson yournetworkwhenusingthisproduct.
WARNING:
LOSSOFDATA:Databasebackupistheresponsibilityofthecustomer. GEMedicalSystemsInformationTechnologies,Inc.isnotresponsiblefordataloss
ofanykindasaresultofcustomer'sfailuretobackupdata.
WARNING:
PATIENTSAFETY:Allcomputer-generatedtracingsshouldbeoverreadbya qualiedphysician.
WARNING:
DATALOSSORCORRUPTION:InstallationofsoftwarenotspeciedbyGE Healthcaremaycausedamagetotheequipment,lossorcorruptionofdata.
DONOTloadanysoftwareotherthanthatspeciedbyGEHealthcareontothe system.
WARNING:
LOSSOFDATA:Changingsettingswithoutknowinghowtheyaffectthesystem cancauselossofdata.
Donotchangeanycurrentsettingsunlessyouunderstandhowthechange affectsthesystem.
WARNING:
PERSONALINJURYFallingobjectsmaycausesevereinjury. Besurethattheserverisadequatelystabilized.
CAUTION:
DATALOSSORSYSTEMFAILURE:Datalossorsystemfailurecanresultdueto ingressofliquids.Thesystemdoesnotprovideprotectionagainstingressofliquids.
Ensureinstallationinacool,dryenvironment.
18
MUSE™CardiologyInformationSystem
2034539-043D
Page 19
Introduction
CAUTION:
EQUIPMENTINTERFERENCEEquipmentorsystemshouldnotbeusedadjacent to,orstackedwith,otherequipment.
Ifadjacentorstackeduseisnecessary,theequipmentorsystemshouldbetested toverifynormaloperationinthatconguration.
NOTE:
MUSEsystemoperationmaybeaffectediflargemachineswithhighcurrent drawsareconnectedtothesameelectricalcircuitastheMUSEsystem.Itis recommendedthattheMUSEsystembeconnectedtoapowersourceawayfrom thesemachines.
NOTE:
IfaUPSwasprovidedbyGEHealthcare,donotconnectyourlaserprintertothe UPS.Insomeenvironments,thecyclingofpowerinthelaserprintercantrigger thelowvoltagealarmontheUPS.

RFCaution

RadioFrequency(RF)devicesmayinterferewiththeuseoraccuracyofthedevice orsystem.Wheninstallingorusingthedeviceorsystem,youshouldconsiderthe proximityofknownRFsources,suchas:
•RadioandTVstations
•PortableandmobileRFcommunicationdevices(cellphones,two-wayradios)
•X-ray,CT,orMRIdevices Thesedevicesarealsoapossiblesourceofinterferenceastheymayemithigher levelsofelectromagneticradiation.
WARNING:
EQUIPMENT:Donotusethedeviceorsystemadjacenttoorstackedwithother equipment.Ifadjacentorstackeduseisnecessary,observethedeviceorsystem toverifynormaloperationinthecongurationinwhichitisbeingused.
SeetheElectromagneticCompatibilitysectionfoundintheservicemanualfor recommendedseparationdistances.
WARNING:
ACCESSORIES/COMPONENTS:Addingaccessoriesorcomponents,ormodifying themedicaldeviceorsystem,mayresultinincreasedEMISSIONSordecreased IMMUNITYofthedeviceorsystem.
•Seethesuppliesandaccessoriesmanualforyoursystemforqualiedaccessories and/orcomponents,ifapplicable.
•Consultwithqualiedpersonnelregardingchangestothedeviceorsystem conguration.

EMI/EMCRequirements

2034539-043D
Thisdeviceorsystemislabeledundertheoriginalequipmentmanufacturerslabel (forexample,USAFCC47CFR15,CEEUEMC2004/108/EC),anddeemedsufcientby GEHealthcaretobeincompliancewithEN/IEC60601-1-2whenusedaccordingto thedeviceorsystem'sintendeduse.HardwaresuppliedbyGEHealthcaremeetsthe applicablecountryrequirements.
MUSE™CardiologyInformationSystem
19
Page 20
Introduction
ChangesormodicationstothissystemnotexpresslyapprovedbyGEHealthcare couldcauseEMCissueswiththisorotherequipment.Thissystemisdesignedand testedtocomplywithapplicableregulationsregardingEMC,andmustbeinstalled andputintoserviceaccordingtotheEMCinformationstatedintheEMCchapter locatedintheservicemanualforthissystem.
WARNING:
EQUIPMENTMALFUNCTION–Theuseofportablephonesorotherradiofrequency (RF)emittingequipmentnearthesystemmaycauseunexpectedoradverse operation.
Donotuseportablephonesorotherelectronicequipmentthatmayemitradio frequency(RF)nearthissystem.
WARNING:
EQUIPMENTMALFUNCTION–Donotusetheequipmentorsystemadjacentot,or stackedwith,otherequipment.
Ifadjacentorstackeduseisnecessary,testtheequipmentorsystemtoverify normaloperationinthecongurationinwhichyourareusingit.
Classication
ClassAThedeviceorsystemissuitableforuseinallestablishments,otherthan
ClassBThedeviceorsystemissuitableforuseinallestablishments,including
Description
domesticandthosedirectlyconnectedtothepubliclow-voltagepower supplynetworkthatsuppliesbuildingsusedfordomesticpurposes. Mainspowershouldbeatypicalcommercialorhospitalenvironment.
domesticestablishmentsandthosedirectlyconnectedtothepublic low-voltagepowersupplynetworkthatsuppliesbuildingsusedfor domesticpurposes.
NOTE:
Complianceprovidesreasonableprotectionagainstradio-frequencyinterference. However,thereisnoguaranteethatinterferencewillnotoccurinaparticular installation.Youcantellwhetherthisdeviceorsystemiscausinginterferenceby turningitoff.Iftheinterferencestops,itwasprobablycausedbythedeviceor system.

EquipmentCompliance

Thedeviceorsystemislabeledundertheoriginalequipmentmanufacturer’slabel(for example,UL,CEEULVD2006/95/EC)anddeemedsufcientbyGEHealthcaretobe incompliancewithEN/IEC60601-1(clause3.201.2foruseofnon-medicaldevices inamedicalsystem),whenusedaccordingtothedeviceorsystem’sintendeduse. HardwaresuppliedbyGEHealthcaremeetstheapplicablecountryrequirementsfor InformationTechnologyEquipment(ITE).
InformationTechnologyEquipmentRequirements
Thehardwarecomponentsofthesystemarenotconsideredmedicalequipment. TheyareconsideredInformationTechnologyEquipment(ITE).Thesystem’sindividual componentscomplywiththestandardsforSafetyofInformationTechnology Equipment(forexample,UL60950-1,EN/IEC60950-1).
20
MUSE™CardiologyInformationSystem
2034539-043D
Page 21
Ifyouusethesysteminapatient’svicinity,itmustcomplywiththestandard requirementsformedicalsystems(thatis,EN/IEC60601-1).
•Tocomplywiththisstandard,youmustconnectthecomponentsandallattached accessoriestoamedicalgrade(EN/IEC60601-1)powersource(forexample, MedicalgradeUPSorIsolationTransformer).
•SystemInputs/Outputsconnectedtonon-medicalequipmentmustalsobeisolated foroverallsystemcompliance.
•Patientvicinityisdenedasaspace,withinalocationintendedfortheexamination andtreatmentofpatients,extending1.83m(6ft.)beyondthenormallocationof thebed,chair,table,treadmill,orotherdevice(s)supportingthepatientduring examinationandtreatment,andextendingverticallyto2.5m(8ft.2.4in.)above theoor.
Inaddition,anynon-medicalelectricalequipmentthatyouusewiththesystem (outsidethepatientvicinity)mustcomplywithapplicablesafetystandardsforthat equipment(thatis,EN/IEC60950-1).
NOTE:
IftheequipmentisinstalledintheU.S.A.using240Vratherthan120V,thesource mustbeacenter-tapped,240V,single-phasecircuit.
Introduction

PartsandAccessoriesInformation

WARNING:
PATIENTSAFETY—Toensurepatientsafety,useonlypartsandaccessories manufacturedorrecommendedbyGEHealthcare.
ContactGEHealthcareforinformationbeforeconnectinganydevicestothis equipmentthatarenotrecommendedinthismanual.
IftheinstallationofthisequipmentintheU.S.A.uses240Vratherthan120V,the sourcemustbeacenter-tapped,240V,single-phasecircuit.
Partsandaccessoriesmustmeettherequirementsoftheapplicable60601safety standards,and/orthesystemcongurationmustmeettherequirementsofthe 60601-1-1MedicalElectricalSystemsstandard.
Usingaccessoryequipmentthatdoesnotcomplywiththeequivalentsafety requirementsofthisequipmentmayleadtoareducedlevelofsafetyofthe resultingsystem.Considerationrelatingtothechoiceshallinclude:
•UseoftheaccessoryinthePatientVicinity. Patientvicinityisdenedasaspace,withinalocationintendedforthe examinationandtreatmentofpatients,extending1.83m(6ft.)beyondthe normallocationofthebed,chair,table,treadmill,orotherdevice(s)supporting thepatientduringexaminationandtreatment,andextendingverticallyto2.5m (8ft.2.4in.)abovetheoor.
•Evidencethatthesafetycerticationoftheaccessorywasperformedin accordancewiththeappropriate60601-1and/or60601-1-1standard(s).
2034539-043D
MUSE™CardiologyInformationSystem
21
Page 22
Introduction

ResponsibilityoftheManufacturer

GEHealthcareisresponsibleforthesafety,reliability,andperformanceofhardware suppliedbyGEHealthcareonlyifthefollowingconditionsaremet:
•Assemblyoperations,extensions,readjustments,modications,orrepairsare performedbypersonsauthorizedbyGEHealthcare.
•Theelectricalinstallationoftheroomwherethedeviceisusedcomplieswiththe requirementsoftheappropriatelocal,state,andothergovernmentregulations.
•Theequipmentisusedinaccordancewiththeinstructionsforuse.

ResponsibilityofthePurchaser/Customer

Thecustomerisresponsibleforprovidingappropriatedesks,chairs,electricalwall outlets,networkconnections,andanalogphonelines,andforlocatinganyofthe systemcomponentsdescribedinthismanualincompliancewithalllocal,state,and nationalcodes.

Symbols

Thefollowingsymbolsmayappearonthedeviceoritspackaging.Familiaritywith thesesymbolsassistsinthesafeuseanddisposaloftheequipment.Forequipment symbolsnotshown,refertotheoriginalequipmentmanufacturers(OEM)manuals.
TheOEMdevicesincludedwiththesystemareasfollows(seetheMUSECardiology InformationSystemServiceManualforacompletelistandnamesofOEMdevices):
•Monitor(s)
•FileServer
•Client
•Modem(s)
NOTE:
Allsymbolsindicated/denedinthismanualmayormaynotappearonthe product.
EquipmentSymbols
Symbol
CAUTION:
Consultaccompanyingdocuments.
Indicatesthepresenceofhazardousenergycircuitsorelectricshockhazards.
WARNING:
ELECTRICSHOCKToreducetheriskofelectricshock,doNOTremove frontorbackcover.Referservicingtoqualiedpersonnel.
Description
CAUTION:
SAFETYGROUNDPRECAUTIONRemovepowercordfromthemains sourcebygraspingtheplug.DONOTpullonthecable.
22
MUSE™CardiologyInformationSystem
2034539-043D
Page 23
EquipmentSymbols(cont'd.)
Introduction
Indicatesthatthewasteofelectricalandelectronicequipmentmustnotbe disposedasunsortedmunicipalwasteandmustbecollectedseparately. Pleasecontactanauthorizedrepresentativeofthemanufacturerfor informationconcerningthedecommissioningofyourequipment.
DateofManufacture(Year-Month).
Indicatesthepresenceofahotsurfaceorhotcomponent.Ifthissurfaceis contacted,thepotentialforinjuryexists.
WARNING:
HOTSURFACE/COMPONENTToreducetheriskofinjuryfromahot component,allowthesurfacetocoolbeforetouching.
ThissymbolonanRJ-45receptacleindicatesaNetworkInterface Connection.
WARNING:
ELECTRICSHOCKToreducetheriskofelectricshock,reordamage totheequipment,donotplugtelephoneortelecommunications connectorsintothisreceptacle.
Indicatesthepresenceofelectricshockhazards.Theareacontainsnouser oreldserviceableparts.Donotopenforanyreason.
WARNING:
ELECTRICSHOCKToreducetheriskofelectricshockhazards,donot openthisenclosure.
Thesesymbolsonpowersuppliesorsystemsindicatetheequipmentis suppliedbymultiplesourcesofpower.
WARNING:
ELECTRICSHOCKToreducetheriskofinjuryfromelectricshock,remove allpowercordstocompletelydisconnectpowerfromthesystem.
Indicatesthecomponentexceedstherecommendedweightforone individualtohandlesafely.
WARNING:
INJURY/DAMAGEToreducetheriskofpersonalinjuryordamagetothe equipment,observelocaloccupationalhealthandsafetyrequirements andguidelinesformanualmaterialhandling.
Humiditylimitations.
Atmosphericpressurelimitation.
2034539-043D
MUSE™CardiologyInformationSystem
23
Page 24
Introduction
EquipmentSymbols(cont'd.)
Batchcodeofpaperorbattery.
Manufacturernameandaddress.
CEMark-SymbolizesconformitywithapplicableEU(EuropeanUnion) directives
PCT(GOST-R)Mark-Symbolizescompliancewithapplicablerequirements forRussia
Serialnumber.
Catalognumber.
Europeanauthorizedrepresentative.
Thepackagingofthisproductcanberecycled.
IndicatestheproductisclassiedasCLASS1LASERPRODUCT.Locatedon thesurfaceofyourlaserdevice.
CDheadphonejack.
Adjustmentcontroldial.
CDejectbutton.
Scanningthebarcodewiththebarcodereaderallowsyoutoacquire recordsfromadiskette.
Indicatesthepresenceofasharpedgeorobjectthatcancausecutsor otherbodilyinjury.
WARNING:
BODILYINJURYTopreventcutsorotherbodilyinjury,donotcontact sharpedgeorobject.
24
MUSE™CardiologyInformationSystem
2034539-043D
Page 25
EquipmentSymbols(cont'd.)
Introduction
Indicatesthepresenceofmechanicalpartsthatcanresultinpinching, crushingorotherbodilyinjury.
WARNING:
BODILYINJURYToavoidriskofbodilyinjury,keepawayfrommoving parts.
Indicatesthepresenceofapotentialtip-overhazardthatcanresultinbodily injury.
WARNING:
BODILYINJURYToavoidriskofbodilyinjury,followallinstructionsfor maintainingstabilityoftheequipmentduringtransport,installationand maintenance.

Training

ServiceInformation

ServiceRequirements

Thismanualisintendedasasupplementto,notasubstitutefor,thoroughproduct training.Ifyouhavenotreceivedtrainingontheuseofthesystem,youshouldrequest trainingassistancefromGEHealthcare.
Toseeavailabletraining,gototheGEHealthcaretrainingWeb site(www.gehealthcare.com/training).SelectEducation>Product Education-Technical>DiagnosticCardiology.
Formoreself-pacedcourseofferings,tools,andreferenceguidesyou maynduseful,pleasevisittheGEHealthcareEducationStoreat
www.gehealthcare.com/educationstore
Thissectionprovidesinformationpertainingtothemaintenanceandservicingofthe system.FamiliarizeyourselfwiththisinformationbeforerequestingservicefromGE Healthcareoritsauthorizedrepresentatives.
Failureonthepartoftheresponsibleindividual,hospital,orinstitutionusingthis equipmenttoimplementasatisfactorymaintenanceschedulemaycauseundue equipmentfailureandpossiblesafetyhazards.
.
Regularmaintenance,irrespectiveofusage,isessentialtoensurethatthecomponents ofthissystemarealwaysfunctionalwhenrequired.
2034539-043D
MUSE™CardiologyInformationSystem
25
Page 26
Introduction

Customer-suppliedHardware

Customerswhopurchasetheirownequipmentareresponsibleforanyrepair ormaintenance,andforcompletingequipmentcheckoutafteranyrepairor maintenance.

HardwareSuppliedbyGEHealthcare

ForhardwaresuppliedbyGEHealthcare,onlyauthorizedGEHealthcareservice personnelshouldservicetheequipment.Anyunauthorizedattempttorepair equipmentunderwarrantyvoidsthatwarranty.Itistheuser'sresponsibilitytoreport theneedforservicetoGEHealthcareortooneoftheirauthorizedagents.

SecurityUpdates

AlistofvirusesthatposeasignicantthreattoGEHealthcarecustomers'system securityispostedontheGEHealthcareProductSecuritywebsite.
Asnewvulnerabilitiesandpotentialsecurityissuesarise,GEHealthcaremakesevery efforttoquicklyidentifyandnotifycustomersofapprovedxes.Timeisrequired forGEHealthcaretoidentifythevulnerability,testthex,andrunavalidationtest onthesystemforsafetyandfunctionality.OnlyafterthisrigorousprocessdoesGE Healthcarereleasetheofcialpatch.Whilewerecognizetheurgencytocorrectthese problems,wemustensurethattheintegrityofthesystemisnotcompromised.
AftersecuritypatchesarevalidatedforspecicGEHealthcaresystems,the informationisaddedtotheProductSecuritywebsite.Youcandownloadthepatch directlyfromthewebsiteofthesoftwaremanufacturer(Microsoft,andsoforth)and applyittoyourGEHealthcaresystem.Tocheckonthelatestinformationregarding validatedsecuritypatches:
1.BrowsetotheGEHealthcareProductSecuritywebsite:http://
prodsecdb.gehealthcare.com
TheSingleSignOn(SSO)windowopens.
2.EnteryourSSOnumberandpasswordandclickLogIn.
IfyoudonothaveanSSOnumber ,clicktheSignUplinktoobtainone.
3.UsethefeaturesontheGEHealthcareProductSecurityDatabaseWebsiteto
identifysecuritypatchesthatyoucanapplytoyoursystem.

AdditionalAssistance

GEHealthcaremaintainsatrainedstaffofapplicationandtechnicalexpertstoanswer questionsandrespondtoissuesandproblemsthatmayariseduringtheinstallation, maintenance,anduseofthissystem.
ContactyourlocalGEHealthcarerepresentativetorequestadditionalassistance.

ManualInformation

Thissectionprovidesinformationforthecorrectuseofthismanual.
26
MUSE™CardiologyInformationSystem
2034539-043D
Page 27
Introduction
Keepthismanualwiththeequipmentatalltimesandperiodicallyreviewit.You shouldrequesttrainingassistancefromGEHealthcare,ifneeded.

IntendedAudience

ThismanualisintendedtobeusedbytrainedGEHealthcareservicepersonnel, GEHealthcare-approvedthirdpartyservicepersonnel,orlocalbiomedicalorIT personnel,responsibleforadministrationandmaintenanceoftheMUSEsystemand thehardwareandnetworkenvironmentonwhichitisrunning.
Thismanualdoesnotprovideinstructionforclinicaluseofthesystem.Forclinicaluse oftheMUSEsystem,refertotheMUSECardiologyInformationSystemOperator's Manual.

ManualPurpose

Thismanualprovidestechnicalinformationtoserviceandtechnicalpersonnelso theycanserviceandmaintaintheMUSEsystem.TheMUSEsystem,asdenedin thismanual,includestheMUSEApplicationserver,MUSEdatabase,MUSEclients, andHL7Interfaceserver.

DocumentConventions

TypographicalConventions
Wherenecessary,themanualidentiesadditionalsourcesofrelevantinformation and/ortechnicalassistance.
ThismanualisintendedonlyforusewithMUSEv8.xForearlierversionsofMUSE,use theMUSECardiologyInformationSystemServiceManualthatoriginallyshipped withyourproduct.SeetheMUSECardiologyInformationSystemOperator'sManual forinstructionnecessarytooperatetheequipmentsafelyinaccordancewithits functionandintendeduse.
Thismanualusesthefollowingconventions.
ConventionDescription BoldTextIndicateskeysonthekeyboard,texttoenter,orhardwareitemssuchas
buttonsorswitchesontheequipment.
Italicized-Bold
Text
CTRL+ESC
<space>
Indicatessoftwaretermsthatidentifymenuitems,buttonsoroptionsin variouswindows.
Indicatesakeyboardoperation.Aplus(+)signbetweenthenamesoftwo keysindicatesthatwhileholdingtherstkey,youshouldpressandrelease thesecondkey.Forexample,PressCTRL+ESCmeanstopressandholdthe CTRLkeyandthenpressandreleasetheESCkey.
Indicatesthatyoumustpressthespacebar .Wheninstructionsaregiven fortypingaprecisetextstringwithoneormorespaces,thepointwhere youmustpressthespacebarisindicatedas<space>.Thisensuresthat thecorrectnumberofspacesisinsertedinthecorrectpositionswithin theliteraltextstring.Thepurposeofthe<>bracketsistodistinguishthe commandfromtheliteraltextwithinthestring.
2034539-043D
MUSE™CardiologyInformationSystem
27
Page 28
Introduction
ConventionDescription Enter
>
IndicatesthatyoumustpresstheEnterorReturnkeyonthekeyboard.Do nottypeEnter.
Thegreaterthansymbol,orrightanglebracket,isaconcisemethodto indicateasequenceofmenuselections.
Forexample,thestatement“Fromthemainmenu,selectSystem>Setup> OptionstoopentheOptionActivationwindow”replacesthefollowing:
1.Fromthemainmenu,selectSystemtoopentheSystemmenu.
2.FromtheSystemmenu,selectSetuptoopentheSetupmenu.
3.FromtheSetupmenu,selectOptionstoopentheOptionActivation window.
Illustrations
Allillustrationsinthemanualareprovidedasexamplesonly.Dependingonsystem conguration,screensinthemanualmaydifferfromthescreensonyoursystem.
Allpatientnamesanddataarectitious.Anysimilaritytoactualpersonsis coincidental.
Notes

RelatedDocuments

Notesprovideapplicationtipsoradditionalinformationthat,whileuseful,arenot essentialtothecorrectoperationofthesystem.Theyarecalledoutfromthebody textthroughaagwordandindentation,asfollows:
NOTE:
ThetiporadditionalinformationisindentedbelowtheNOTEagword.
Thefollowingdocumentsprovideadditionalinformationthatmaybehelpfulinthe installation,conguration,maintenance,anduseofthissystem.
DocumentsRelatedtotheMUSECardiologyInformationSystemServiceManual
PartNumberDocumentTitle
2034539–042 2034539–172 2034539–044 2034539–180 2034539–050 2020299–021 2020299–025
2034539–048
MUSECardiologyInformationSystemOperator'sManual MUSECardiologyInformationSystemHardwareManual MUSECardiologyInformationSystemPreInstallationManual MUSECardiologyInformationSystemDevicesandInterfacesManual MUSECardiologyInformationSystemClientInstallationManual MobileLinkWirelessCommunicationInstallationManual LANOptionforMACRestingECGSystemsInstallationand
TroubleshootingGuide MUSECardiologyInformationSystemAdvancedSecurityGuide
28
MUSE™CardiologyInformationSystem
2034539-043D
Page 29

ProductOverview

Thischapterprovidesageneraldescriptionoftheproduct,itsconnectivitytoother devicesandinterfaces,andadescriptionofavailableoptions.

GeneralOperation

TheMUSEsystemconsistsofanapplication/databaseserverthatstoresthesystem congurationandpatienttests(ECG,Stress,Holter,andHiRes)inarelationaldatabase.
ClientcomputersruntheMUSEapplicationandcommunicatewiththeMUSEdatabase viaamiddle-tierservicethatrunsontheMUSEserver.
Dependingonthedeviceandthesystemconguration,youcanacquiretestsovera:
•wirednetwork
•wirelessnetwork
•modem
•oppydiskette
•securedigital(SD)card
•serialdownloadcable
•anycombinationofthese
2
Youacquirethetestsfromassociateddevicesforaparticulartypeofstudy.These datatypescomefromdevicesorsystemssuchas:
•ECGElectrocardiographCarts
•CASEStresssystems
•MARSHoltersystems
•Patientmonitors
TheMUSEapplicationconnectsclientworkstationstotheserveroverthenetwork. Theseworkstationsaccesstheservertoperformsystemfunctionssuchasediting, testretrieval,systemsetup,runningdatabasesearches,andcheckingsystemstatus.
Theeditorapplicationallowsyoutocompleteoverreading,retrieving,andconrming tests.Youcanmanuallyorautomaticallyroutetheteststooutputdevicessuchas laserprinters,networklaserprinters,faxmachines,anHL7(HealthLevel7)interface, ortodestinationssuchasemailaddressesandsharedfolders.Youcanmanuallyor automaticallyroutetheteststooutputdevicessuchassharedprinters,faxmachines, anHL7Resultsdevice,ortodestinationssuchase-mailaddressesandsharedfolders.
2034539-043D
MUSE™CardiologyInformationSystem
29
Page 30
ProductOverview

DataAcquisition

YoucanacquiredataintotheMUSEsystemdatabasefromseveraldifferentinterfaces anddatasources.Followingisasummaryofthedifferentinterfaces:
CSI ECGcartssenddataeitherthroughaserialcableconnectedtoaworkstation,aCSI modem,a802.11b/gwirelessconnection,orthroughaLANconnection.
DirectMediaAcquisition TheMUSEsystemcanacquiredatathatwasstoredtolocalmediaonanECGcart, suchasaoppydisk(onoldercarts),oraSecureDigital(SD)card.
GeneralAcquisition TheMUSEsystemcanacquiredatafromeitheralocalfolderorfromanetwork share.Usethismethodtoacquiredatafrombedsidemonitors,MARSHolter systems,andCASEstresssystems.TheMUSEBkgndaccountbeingusedtorunthe MUSEservicesneedsaccesstothefolderinordertoacquirethedata.
NOTE:
MUSEv8doesnotsupportacquiringnewpatientdatafromCASEsystems runningWindowsNTorolderoperatingsystems.
XMLImport YoucansendECGdatafromothervendordevicestotheMUSEsystemthrough theXMLImportfeature,whichisapurchasableoption.MUSEsupportstheXML formatsentfromtheLifePak®12/15debrillators,DatamedFT®formattranslator software,andGetemedCardioday®Holter\ECGsystem.
DataacquiredataworkstationiscopiedthroughthenetworktotheMUSEserverfor processingandstorage.Dependingonthetypeoftest,thesizeoftheacquireddata varies.AcquireddataisplacedintheNewlyAcquiredQueueinthedatabase.From therethesystemprocessesthedataintothesystemandmakesitavailabletousers forviewing,editing,orprinting.YoucanautomaticallyrouteteststoMUSE-dened devicesuponacquisitionintothesystem,whensavedasDemographicsComplete,or whenconrmedbyanoverreader.ThefollowingDataTypestableliststhedifferent datatypesyoucanstoreontheMUSEsystem,alongwiththeacquisitiondevice, andtypicalrecordsize.
DataTypes
DataType
RestingECG—12Lead(250Hz)Electrocardiograph RestingECG—12Lead(500Hz)Electrocardiograph ECG(240Hz) HiResolutionECG—15LeadElectrocardiograph StressCASE750KB—3MB Holter
AcquisitionDevice
Clinical(monitoring)device
MARS1MB—2.5MB
TypicalRecordSize
7KB 10KB—20KB 7KB 10KB—20KB
30
MUSE™CardiologyInformationSystem
2034539-043D
Page 31

DeviceInterfaces

MACCarts

ProductOverview
TheMUSEsystemcaninterfacewiththefollowingsystemsanddevices:
•MACCarts
•MARS
•CASE/CardioSoftStressSystems
•MonitoringSystems
YoucansendECGdataacquiredattheMACcartsintotheMUSEsystemforlong-term storage.Dependingonthetypeofcartandtheoptionsinstalled,youcansenddata oneofseveralways:
•Removethedisketteorsecuredigital(SD)cardfromthecartanddownloaditonto aMUSEclient
•CSImodem
•CSIdirect(directserialconnection)
•CSInetwork(wireless/LAN)
CartssupportingRemoteQueryallowthecarttoretrievearecordfromtheMUSE system.YoumustpurchaseRemoteQueryasacartoption.TheAdvancedSecurity featuresthatarepartoftheMUSEsoftwareallowyoutoturnoffRemoteQueryona per-sitebasiswithintheMUSEsystem.
ThefollowingtableprovidesthemethodsavailablefortransferringdatafromtheCart totheMUSEsystem.Manyofthemethodslistedrequirethepurchaseandactivation ofanoptiononthecart.RefertoyourGEHealthcaresalesrepresentativeformore information.
ECGCartInterfacestoMUSE
ECGCartCSI
Modem
MAC600 MAC800 MAC
1200 MAC
1600 MAC
3500 MAC
5000 MAC
5500
1
ThisoptionisavailableontheMAC3500onlyoutsideofNorthAmerica.
XXXX XX
XXXX
XX
XXXXX
XXXXXX
CSINetwork Wireless
1
LAN
XXX
CSI Direct
XX
DisketteSDCard
Remote Query
2034539-043D
MUSE™CardiologyInformationSystem
31
Page 32
ProductOverview

MARS

YoucancongureMARSHoltersystemstostoresavedreportstotheMUSEsystem forlong-termstorage.FullDisclosureisnotsenttotheMUSEsystem.Toaccomplish thedatatransfer ,theMUSEsystemconnectstoanetworkshareontheMARSsystem overthenetworkandcopiesthelestotheMUSEapplication/databaseserverfor processing.TheMUSEsystemrequirestheHolterDataTypeoption.UNIX-based MARSsystemsrequiretheenterprisenetworkcardoption.

CASE/CardioSoftStressSystems

YoucanconguretheCASE/CardioSoftStresssystemstostoresavedreportstothe MUSEsystemforlong-termstorage.CASE/CardioSoftsystemsrequireanetwork optiontosenddatatotheMUSE.TheCASE/CardioSoftsystemcopiesthelestoa sharedfolderontheMUSEsystem,andthentheMUSEsystempicksuptheleslocally forprocessing.YoumustenabletheExerciseTestingoptionfortheMUSEsystem.

MonitoringSystems

Youcancongurebedsidemonitorstosend12–leadECGdatatotheMUSEsystem forlong-termstorage.Youaccomplishthisthroughtheuseofamonitoringgateway, whichisacomputerthathastwonetworkinterfaces:oneonthemonitoringnetwork, andoneonthehospitalnetworkonwhichtheMUSEapplication/databaseserver islocated.WhenyousendatesttoMUSEfromabedsidemonitor,themonitoring gatewayreceivesitandstoresitlocallyinasharedfolder.TheMUSEserveris conguredtolookfordataonthatnetworkshare,andwhenitndsnewdata,moves ittotheMUSEserverandprocessesintothesystem.

Modems

ModemTypes
TheMUSEsystemusesmodemstoreceiveandsenddatatocartsandtofaxreports fromtheMUSEsystem.
ThefollowingtabledescribesthetypesofmodemstheMUSEsystemuses:
ModemTypes
Type
CSI
FAX
IfyoupurchaseaGEHealthcare-suppliedmodem,youreceivetheMultiTech MT9234ZBAmodem.FordetailedinformationabouttheMultiTechmodem,referto theMUSECardiologyInformationSystemHardwareManual.
Usage
Receive/SendwithECG cart
SendtofaxmachineNormallyinstalledontheserver,butaworkstationcan
WhereInstalled
Normallyinstalledontheserver,butaworkstationcan supportitbasedonthelocationofphonelinesandthe needtoavoidlongdistancecalls.
supportit,basedonthelocationofphonelinesandthe needtoavoidlongdistancecalls.OlderFaxmodemsare installedontheSMMclientonly.
32
MUSE™CardiologyInformationSystem
2034539-043D
Page 33
ProductOverview
CSIModem
TheCSImodemisanexternalmodemconnectedtotheMUSEserveroraMUSEclient. ItcancommunicatewiththefollowingECGcarts:
•MAC800
•MAC1200
•MAC1600
•MAC3500
•MAC5000
•MAC5500
CSIisaproprietaryprotocolrunningontopofSerialLineInternetProtocol(SLIP)and doesnotincludetheabilitytorespondtostandardIPnetworkconnections,such asPING.
CSImodemsaretypicallyusedtosendECGstotheMUSEsystemfromacart,butyou analsousethemtosendtestsorordersbacktothecart.
Therearenosecurityrisksassociatedwiththismodem.TheCSIprotocoldoesnot respondtoIP-basedprobing,anddoesnotallowanyaccesstosystemprocesses orthelestructure.SomeonewithacarthavingtheRemoteQueryoptioncould retrieveanECGiftheyknewthephonenumberofthesystem.Youcandisablethis functionalityonthesystemifyouusetheMUSEAdvancedSecurityOptions.
FaxModem
ThefaxmodemusesClassIIGroup3faxing,at9600Baud. ThefaxmodemsonMUSEsystemsareoutputonlyanddonotacceptincomingcalls.
ItisimpossibleforanyonetoaccesstheMUSEsystembycallingintoafaxmodem. Whensettingupafaxmodem,determiningtheappropriatelocationisimportant
regardingfaxmodemlocationrelativetoareacodes.Forinstance,aremotehospital sharingthesameMUSEserver,faxingrecordslocally,wouldbebetterservedby addingafaxmodemonalocalcomputertoreducephonecosts.
Therearenoassociatedsecurityriskswiththismodemsinceitdoesnotaccept incomingcallsandcannotaccesstheMUSEsystem.
Multi-portModemCard
YoucanequiptheMUSEserverwithamulti-portserialadapter.GEHealthcare providestheDigiAccelePort®8–portPCIadapter(Digipartnumber77000889),which canconnecteight,DB9connectionstosupportuptoeightmodems.Serialportsare conguredforCOM3—COM10.

HL7Interface

TheMUSEsystemsupportsHL7interfacestoexchangeADT,Order,Result,andBilling messagesbetweentheMUSEserverandtheHospitalInformationSystem(HIS).You mustpurchaseandenabletheappropriateHISoption(ADT,Orders,Results,Billing, ADTQuery,BatchProcessing)fortheMUSEsystem.
2034539-043D
MUSE™CardiologyInformationSystem
33
Page 34
ProductOverview
ThefollowingsoftwareisinstalledbyGEHealthcareserviceduringinstallation:
•HL7Interfacesoftware
•AdobeAcrobatReader
BoththeMUSEapplicationandSQL2005(singleuserlicense)areinstalledontheHL7 serverforconguringandtestingoftheHL7interfacewiththeHospitalInformation Systemduringinitialsysteminstallation.Theseapplicationsarenotneededfor operationoftheHL7interfaceandyoumayremovethemfromtheproductionHL7 serverafterinstallationandtestingarecompleted.
NOTE:
GEHealthcareconguresaweeklybackupoftheHL7congurationasaWindows scheduledtaskcalledSystemState_Recurring_Backup.Thisscheduledtasksaves thecongurationtothec:\SystemStatefolder.Youcanusethisbackuptorestore theHL7conguration.Thecustomerisresponsibleformakingacopyofthisfolder andsavingittoanetworkdriveorexternalmedia.

OptionalFeatures

Toviewalistofinstalledoptionsonthesystem,runtheMUSEv8applicationinmodify modeandviewtheMUSEOptionsscreen.Referto“ModifyingtheMUSEInstalled
Conguration”onpage58forinformationregardingrunninginModifymode.
CAUTION:
SYSTEMFAILURE:RunninginmodifymodeexposescriticalMUSEsystem congurationsettings.UnintendedorpoorlyimplementedchangestotheMUSE congurationcanresultintheMUSEsystemfailing.
DoNotlaunchtheMUSEsysteminmodifymodeunlessyouaretrainedand understandhowyourchangesaffecttheMUSEsystem.
ThefollowingfeaturesarestandardonallMUSEv8systems:
•DatabaseSearch
•ACITIPI
34
MUSE™CardiologyInformationSystem
2034539-043D
Page 35
SystemFeaturesOptions
ProductOverview
•Wireless/LANcommunication
•HiRESDataStorage
NOTE:
TheMUSEoptionslistedinthefollowingtablecanonlybeinstalledbyan authorizedGEHealthcareservicerepresentative.
Thefollowingtablelistsavailableoptions.
PurchasedOptionMUSEOption(s)Installed
StressModule
HolterModuleHolterDataStorageIncludes:
ECGSerialComparisonSerialComparison
ExerciseTestingData Storage
Description
Includes:
•LANinterfacetoCASEExerciseSystems
•Abilitytoeditdiscreetdataandreportelements
•AbilitytoviewsavedstripsfromCASE
•Abilitytoattachordersifyouareusingthe MUSEHL7HISOrdersoption
•LANinterfacetoMARSHolterSystems
•Abilitytoeditdiscreetdataandreportelements
•AbilitytoviewsavedstripsfromMARS
•Abilitytoattachordersifyouareusingthe MUSEHL7HISOrdersoption
•Providesautomaticcomparisontoidentify clinicallysignicantchangesinthecurrentECG totherstpreviousECG
•UsingGEMarquette12SLECGanalysis statements
•measurementsforQRS,ST
•morphologychanges
EnhancedECGEditorEnhancedEditorIncludes:
•fullscreenviewsfor12/15–leadadult/pediatric ECGandHiResECG
•SerialPresentationforcurrent,rstprevious, andoldestECGs
•Magnication,March-outCalipers,andSerial ComparisonOn—Demand
ECGReanalysisRe-analysisReanalysesincomingECGtestswiththelatest
versionof12SLalgorithmsoftware.
2034539-043D
MUSE™CardiologyInformationSystem
35
Page 36
ProductOverview
SystemFeaturesOptions(cont'd.)
PurchasedOptionMUSEOption(s)Installed
ThirdpartyXMLImport
XMLImport
Interface
Third-partyXMLExport
XMLExport
Interface
21CFRPart11Module(for
21CFRPart11
FDAGuidelines)
Description
•Providescapabilityforautomatedacquisition andreportroutingofECGsformmulti-vendor devicesorsystemutilizingXMLformatteddata.
•SupportsPhysio-ControlLIFEPAKandGetemed CardioDayECGs,aswellas12–leadECG datasenttotheMUSEsystemthroughthe DatamedFTsoftware.
Provides
•10–second12/15–leadECGdata
•patientdemographics
•testinformation
•globalmeasurementsQRStimesandtypes
•Automatedsinglerecordorbulktransferin XMLformatforexternalsystems
Provides:
•audittrail
•electronicsignature
•passwordpromptonconrmation
•reasonforchange
•enumeratedlistorfreetext
•loggingchangestoECGandsubjectdata
•user-denabletextforsignoff
ECGIntervalEditorModuleIntervalEditorProvides:
•beat-by-beatrawormedianbeatediting
•Automatedrecalculationincludingglobaland lead-by-leadQTintervalcalculations
•Semi-automatedandmanualmodes
•Trialsettinglock-down
•E14tools
FDAFormatXMLExport
XMLExport
ProvidesECGdataexportinFDAformat
Interface CVWeb2.0MUSEAPI(MACCRA)and
MUSEWeb
ProvidesonlineaccesstotheMUSEv8systemvia asecureonlineURLlinktoretrieveandviewECG, stress,andHolterreports.
VAVistAImageInterfaceMUSEAPI(MACCRA)and
MUSEWeb
MUSEWebConnectivityto GEHealthcareDevices
MUSEAPI(MACCRA)and MUSEWeb
ProvidesresultlinktoVAVistAImagedisplay system
ProvideswebinterfacetoGEHealthcaredevices, suchasCASEandPanel.
EmailModuleEmailProvidesdistributionofMUSEreportstocustomer
emailsystems
36
MUSE™CardiologyInformationSystem
2034539-043D
Page 37
SystemFeaturesOptions(cont'd.)
ProductOverview
PurchasedOptionMUSEOption(s)Installed
ADTModuleADTInterfaceReceivesadmission,discharge,andtransfer
OrdersModuleHL7OrdersDownload
Interface HISOrdersInterface
ECG/Stress/HolterText ResultsModule
WaveformImageResults Module
ADTQueryModuleHL7ADTQueryInterface
BillingModuleHL7BillingInterfaceProvidesreal-timetransferofnancialtransaction
HL7ResultsInterfaceSendsECG,stress,orHoltertextualresultsto
HL7WaveformInterface
Description
informationfromfacilityADTprovider(forexample HIS).
•Receivesordersfromfacilityorderentrysystem
•Providesabilitytodownloadorderinformation toselectedMACcartsandCASEstresssystems conguredwithMUSEconnectivity
•RequiresADT
facilityElectronicMedicalRecordsystem.
•SendswaveformresultstofacilityElectronic MedicalRecordsystem.
•RequiresECG/Stress/HolterTextResults module.
•Queriesadmission,dischargeandtransfer informationfromfacilityADTprovided(for exampleHIS).
•RequiresADT.
informationtofacilityHIS.
BatchDataModuleHL7BatchInterface
HardwareSpecications
Forhardwarerequirements,refertotheMUSEv8CardiologyInformationSystem PreInstallationManual.
ForinformationregardingreplacementpartsforGEHealthcare-suppliedhardware, refertotheMUSECardiologyInformationSystemHardwareManual.
SoftwareSpecications
Forinformationregardingsoftwarerequirementsandspecications,refertotheMUSE CardiologyInformationSystemPreInstallationManual.

TheMUSESysteminVirtualEnvironments

Customer-suppliedvirtualmachinesmustmeetorexceedthevirtualhardware requirementsdescribedintheMUSEv8CardiologyInformationSystemPre-Installation Manual.Softwarerequirements(operatingsystemandservicepacks)remainthe
•Transferstextualresultsornancialtransaction informationinbatchmodetofacilityHIS.
•RequiresBilling.
2034539-043D
MUSE™CardiologyInformationSystem
37
Page 38
ProductOverview

CustomerResponsibilities

sameasthoseforaphysicalserver.GEHealthcarereservestherighttorequest thevirtualmachine'svirtualhardwarecongurationsandtorequestallocationof additionalCPUormemorytoresolveperformance-relatedissues.
ThefollowingarecustomerresponsibilitieswhenoperatingtheMUSEsystemina virtualenvironment:
•Completionofthesetupandcongurationofthevirtualmachine(s)onwhichthe MUSEsystemisbeinginstalledpriortothearrivaloftheGEHealthcareinstallation engineer.
•CompletionofaMUSEdatabasebackupandrecoveryplan,includingits implementationandtesting.
•ContactwiththeVMsoftwarevendortoresolveissuesspecictothevirtual environment.
GEHealthcareserviceengineerssupporttheMUSEsystemrunningontheWindows operatingsystemsasstatedintheMUSEv8CardiologyInformationSystem Pre-Installationmanual,regardlessofwhetherthatoperatingsystemisinstalledon aphysicalorvirtualmachine.However,GEHealthcaredoesnotprovidesupportfor installationorcongurationofthevirtualenvironment,oranyfeaturesetprovidedas partofthatenvironment,suchashighavailability,fail-overclusters,orSAN.

DataAcquisition

CSImodems,usedtoacquiredatafromsomeMACcarts,cannotconnectdirectlytoa virtualMUSEserver,butcanconnecttoaMUSEclient.

HL7InterfaceVirtualMachine

ThesoftwareusedfortheHISinterfaceisamulti-threadedapplicationandis optimizedforusewithtwoCPUs.TheuseofonlyoneCPUaffectsperformanceand slowsdowntransactionprocessing.
Licensing
ThesoftwareusedtoruntheHL7interfaceislicensedusingtheMACaddressof theserver.AnychangetotheMACaddresswithinthevirtualmachinecausesthe interfacestofail.

MUSEv8DriveContentsandSupportingFolders

Thefollowingdrivelettersaretherecommendedinstallationcongurations.MUSE supportsinstallationonotherdrivepartitions.
38
MUSE™CardiologyInformationSystem
2034539-043D
Page 39

MuseApplicationandDatabaseonSameServer

Thisisthesimplestcongurationandavoidstheneedtoadjustrewallsettingsand setpermissionsbetweentheMUSEserverandSQLdatabases.Itisthestandard congurationwhenMUSEisinstalledonGEHealthcare-suppliedservers.
C:Drive
ProductOverview
Application
WindowsOS AdobeReader9 SQLServer2005 SQLManagementStudio MUSE InSiteExC
Description
MUSEapplicationfolderlocatedinProgramFiles EnablesGEHealthcareremotesupport
D:Drive(MUSE)
Folders
dbMUSEdatabases acq
backupSupportinglesforGEHealthcaretapeornetworkbackup mars
xmlTemporarilystoresinboundxmldata(requiresXMLimport
Description
NetworksharetotemporarilystoreECGsbeingacquired formMARSandMonitoringGateway
TemporarilystoresformattedHolterreportsreadyfor printing
option)

MUSEApplicationandDatabaseonSeparateServers

MUSEApplicationServer
Application/Folders
WindowsOS AdobeReader9 SQLServer2005 SQLManagementStudioRequiredforaccesstoMUSEdatabase MUSE InSiteExC MUSEdatafolderThenameandlocationoftheMUSEdatafolderisspecied
\acq
2034539-043D
MUSE™CardiologyInformationSystem
Description
MUSEApplicationfolderlocatedinProgramFiles EnablesGEremotesupport
duringMUSEsysteminstallation.Inatypicalinstallation,the foldernameisMUSEandincludesthefollowingfolders:
NetworksharetotemporarilystoreECGsbeingacquired fromMARSandMonitoringGateway
39
Page 40
ProductOverview
MUSEApplicationServer(cont'd.)

MUSEServices

Application/Folders
\backup \mars
\xml
Description
SupportinglesfroGEtapeornetworkbackup TemporarilystoresformattedHolterreportsreadyfor
printing Temporarilystoresinboundxmldata(requiresXMLimport
option)
MUSEDatabaseServer
Application/Folders
SQLServer2005 muse\db
Description
MUSEDatabases
MUSEusesagroupofservicestoperformcertainfunctionswithintheMUSE application.ServicesareinstalledtousethedefaultMuseBkgndasthelogonaccount. ThisaccountrequiressysadminprivilegestotheMUSEdatabases.
ThefollowingtableprovidesalistofMUSEserviceswithabriefdescriptionofeach. Whentroubleshootingspecicproblems,itissometimesusefultoverifythatthe correspondingserviceisrunning.
CAUTION:
STOPPINGASERVICEDISABLESTHATFUNCTION:Donotstopservicesunlessyou understandhowitaffectsthesystem,orunlessallusersareloggedoffthesystem.
MUSEv8includesanautoshutdownfeaturetonotifyusersinadvanceofa shutdown.Referto“AutoShutdown”onpage56formoreinformationonusing autoshutdown.
MUSEServices
ServiceDescription MUSESCM
MUSEEmailHandlesemailtransmission MUSEFileCopyCopiesoutputtofolderdevices MUSEformatformatsoutputtoMUSEdevicesotherthan
MUSEFTPCopy MUSEGeneric
MUSEHL7Outbound MUSEHL7Parser1–4
ServiceControlManager:turnsallMUSE servicesoffandon
HL7 copiesoutputtoFTPfolderdevices AcquiresHilltopdatafromshares(ECG,
Stress,andHolter) supportsOutboundResultsand/orBilling SupportsindividualinboundfeedsFROM
HospitalInformationSystem(HIS)
40
MUSE™CardiologyInformationSystem
2034539-043D
Page 41
MUSEServices(cont'd.)
ProductOverview
MUSEPrint
MUSEMACCRA MUSEmodemSupportsCSI,Fax,LAN,andwireless
MUSEMTHost
MUSENormalNormalizesacquiredECGs MUSESchedulerRunsMUSEscheduledtasks:database
MUSEXMLParser

RequiredNetworkPorts

ThefollowingportinformationfortheMUSEsystemisprovidedasaguidelinetohelp youunderstandthesystem’snetworkingrequirementsandtoassistinsituations whereyoumayneedtoconsidereithersoftwareorhardwarerewallcongurations. Notallsystemsuseeachconnection.Theportslistedaredefaultvaluesandinsome casescanbechanged.
NetworkingPorts
HandlesprintingtoPSandPCLprinter devices
SupportsMUSEWebandMUSEAPI
communication HandlesMiddleTiercommunications
betweentheclientandserver
search,SQLjobs,logandqueuemaintenance SupportsXMLacquisition
PurposePortType
GeneralAcquisition
SQLServer
MUSEApplication
MUSEWeb
CSINetwork
RemoteSupport
137 138 139TCP 445 1433TCP 1434
8001TCP
80TCP
3001TCP
443
Notes
UDP UDP
TCP
UDP
TCP
Requiredtoexchangedataviaanetworkshareor withexternalsources,suchasaMARSsystemor MonitoringGateway.
IfNetBIOSoverTCP/IPisdisabled,onlyport445is required.
Listensforincomingconnections.Canbechanged. AllowsadministratorstocheckthestatusofSQL
databases. TheseportsmustbeopentotheMUSEserver.
DefaultportusedbyMUSEuserinterface applications(Editor,Setup,Status,DatabaseSearch).
Youcanchangethisport,butitmustbethesame onboththeMUSEserversandclients.
Port80isthedefaultportforHTTPtrafc.Youmay changethisintheWebSiteproperties.
TheMACcartsusewhentransmittingdatatothe MUSEsystemoverLANorwireless.Thisportcanbe changed.
InSiteExCusesthistocommunicatewithGE Healthcaresupport.
2034539-043D
MUSE™CardiologyInformationSystem
41
Page 42
ProductOverview
42
MUSE™CardiologyInformationSystem
2034539-043D
Page 43

SystemSetup

SettingUpModems

3
ThischapterprovidesadditionalinstructionsforcongurationoftheMUSEsystem afterinstallationiscomplete.
TheMUSEsystemusesthefollowingmodemstosendandreceivedata:
MUSEModemTypes
FAXModemSupportsoutgoingfaxtransmissions CSIModemSupportscartmodemconnectionsfordataupload,order
download,andreversetransmissiontoacart CSIDirect CSINetwork
TosupportanyoftheavailablemodemsonMUSEv8.0,theMUSEModemfeature mustbeinstalledandtheMUSEModemservicemustberunning.
NOTE:
DuringtheupgradefromanearlierversionofMUSEtoMUSEv8.0,allCSIWireless modemswereconvertedtoCSINetworkmodems.IfyouseeaCSIWirelessentry intheModemSetuplist,themodemserverforthatmodemwasnotupgraded oritismissing.Inanormalsituation,youshouldneverseeaCSIWirelessdevice afterupgradingtoMUSEv8.0.
Supportsdirectserialcablecartconnections
Supportswirelessand/orLANcartconnections
Verifying/InstallingtheMUSEModemServiceand Wireless/LANCommunication
NOTE:
MuseoptionscanonlybeinstalledbyanauthorizedGEHealthcareservice representative.
2034539-043D
MUSE™CardiologyInformationSystem
43
Page 44
SystemSetup
UsethefollowinginstructionstoverifyandinstalltheMUSEModemserviceand wireless/LANcommunication.
1.RuntheMUSEInstallerinModifymode.See“ModifyingtheMUSEInstalled
Conguration”onpage58.
2.OntheSelectFeatureswindow,selectModem. ThisinstallstheMUSEModemservice.Thisserviceisrequiredformodem,as
wellas,WirelessandLANtransmission.
3.ClickNextthroughthefollowingwindowsuntiltheChooseMUSEOptions windowopens.
4.IfyouaresettingupwirelessorLANcommunication,verifythattheoptionis selected.
Ifyouneededtoinstalltheoption,youareaskedonthenextscreenforthe OptionActivationPassword.
5.EnterthepasswordandclickNextthroughtheremainingwindows. TheMUSEModemserviceandselectedmodemoptionareinstalled.

SettingUpaModemDevice

Usethefollowinginstructionstosetupamodemdevice.
1.LogonasauserwithprivilegestochangeMUSESetups.
2.OpentheMUSESetupapplication.
3.IntheNavigationpane,selectModems.
4.Right-clickintheModempane,andselectNew.
5.Selectthemodemtypeyouareadding. IfyouselectedFaxModem,CSIModem,orCSIDirect,theappropriateModem
Propertieswindowopens.
IfyouselectedCSINetwork,theModemProperties—CSINetworkwindow opens.
44
MUSE™CardiologyInformationSystem
2034539-043D
Page 45
6.Entertheappropriatevaluesdescribedinthefollowingtables.
Fax,CSI,orCSIDirectModemProperties
FieldDescription
ComputerName
Port BaudRate115.2KBaud(CSIDirect)9600Baud(CSI,Fax)
1
IftheFaxmodemencountersproblemsat9600baud,use4800baud.
Nameofthecomputerwherethemodemisphysically installed.Typically,thisistheMUSEleserver.
PortNumber
CSINetworkModemProperties
SystemSetup
1

RestartingModems

Field
ComputerName IPAddressorHostnameTheIPAddressorHostnameassignedtothecart.
Port RetryIntervalinSecondsDenestheupperlimitofthetimedelaybetween
Description
Namewheretheconnectionissupported.
Portthisconnectionisusing.
attemptsforthecarttocommunicatewiththeMUSE system. Thedefaultis30seconds.
7.SelectOKtosaveyourchanges. TheModemPropertieswindowcloses.
NOTE:
Whenanewmodemissetup,theMUSEModemserviceisnotiedand automaticallystartsanewthreadtosupporttheconnection.Youdonot needtorestarttheMUSEModemserviceafterdeninganewmodem.
Theindividualthreadsthatarerunningtosupporteachconnectionaredesignedto automaticallyrestartiftheystopforsomereason.Usethefollowingprocedureifyou wanttorestartthemmanually:
1.SelectSetupModems.
2.Selectthemodem(s)thatyouwanttorestart,right-clickonit,andselectRestart Modem.
2034539-043D
MUSE™CardiologyInformationSystem
45
Page 46
SystemSetup
NOTE:
Thereisalsoarestartmodemicononthetoolbar.
Amessagedisplaysthatthemodemsweresuccessfullyrestarted.
NOTE:
Ifyoudonotreceivethismessage,theMUSEMTHostservicewasnotable tocommunicatetotheMUSEModemservices.Therearetwocausesforthis:
•TheMUSEModemserviceisnotrunning.
•TherewallsettingsontheMUSEModemservice'shostsystemarenot conguredcorrectly.
3.Formoreinformationaboutthemodems,whentheystartorrestart,select Status>ApplicationLog.
TheApplicationLogPropertieswindowopens
Iftheconnectiontoacartfails,theMUSEModemserviceimmediatelyattemptsto restartit.ThefailureandrestartareloggedintheApplicationLog.
Iftherestartfailswithinoneminute,theservicedoesnotwaituntiltheoneminute intervalisupbeforetryingagain.
46
MUSE™CardiologyInformationSystem
2034539-043D
Page 47
Iftherearethreeconsecutivefailureswithinoneminute,amessageislogged indicatingthaterrormessageloggingforthismodemisstoppeduntilthemodemis workingagain.ThispreventstheApplicationLogfromllingupwithrepetitiveerror messages.Whilenomessagesarebeinglogged,theservicecontinuestorestartthe modeminthebackground.Oncethemodemisrestartedandcontinuesrunningforat leastoneminute,loggingresumesforthismodem.Manuallyrestartingthemodem fromtheuserinterfacealsoresumeslogging.

TestingCSINetworkModemConnections

Usethefollowingproceduretoviewtheconnectionstatusforeachnetwork connection.
1.OpenaCommandPromptwindow.
2.Typenetstat<space>–n. Thisdisplaysalistofallopensocketconnectionsonthesystem. IfaconnectionisestablishedbetweentheMUSEModemserviceandthecart,
thereisanentrycontainingthelocaladdresswiththeIPaddressoftheserver, theforeignaddresswiththeIPaddressofthecart,andtheportnumberafter thecart'sIPaddress.
SystemSetup
Item
1 2 3 4 *Inthisexamplethecartisasimulatorrunningonthesamesystem.Normallythe
serverandcartarenotthesameIPaddress.
Description
ServerIPaddress CartIPaddress* PortNumber Connectionstate
Ifthecartisnotconnected,youdonotseeanentryinthislist.TheMUSE Modemserviceopensaconnectiononcethenumberofsecondsequaltothe retryintervalisreached.Ifitfailstoconnect,itclosesthisconnectionuntilthe nextintervaltime.
2034539-043D
MUSE™CardiologyInformationSystem
47
Page 48
SystemSetup
InstallingandConguringtheXMLImportOption
TheXMLImportoptionisrequiredforsystemsthatareacquiringdatainXMLformat. TheMUSEsystemcanacquiredatainXMLformatfromthefollowingdevices:
•LifePak12throughtheLIFENETReceivingStation(referencedocument
2002783-040).
•FromaDatamedFT(FormatTranslator)providedbyEngineeringSolutions,Inc.
•CardioDayHolterECGSystem.
XMLImportrequiresthattheXMLImportoptionandtheMUSEXMLserviceare installedontheMUSEapplicationserveasfollows.
NOTE:
MUSEoptionscanonlybeinstalledbyanauthorizedGEHealthcareservice representative.

InstallingtheXMLImportOptionandMUSEXMLService

UsethefollowingproceduretoinstalltheXMLImportoptionandMUSEXMLservice
1.RuntheMUSEInstallerinModifymode.See“ModifyingtheMUSEInstalled
Conguration”onpage58.
2.OntheSelectFeatureswindow,selectXML. ThisinstallstheMUSEXMLParserservice.
3.ClickNextoneachofthenextwindowsuntiltheChooseMUSEOptionswindow opens.
4.VerifythattheXMLImportoptionisselected. Ifyouneededtoinstalltheoption,youareaskedonthenextscreenforthe
OptionActivationpassword.
5.EntertheOptionActivationpassword.
6.ClickNextoneachoftheremainingwindows. TheXMLImportoptionandMUSEXMLParserserviceareinstalled.

Requirements

XMLImportrequiresMSXML4.0.Thisprogramshouldhavebeeninstalledduringthe initialMUSEinstallation.
UsingXMLCONFIG.EXEtoConguretheOptiononMUSE
ThisutilityinsertsentriesintothecfgXmlInputtableintheMUSE_Systemdatabase.
1.RunthexmlcongutilitylocatedinthefolderwheretheMUSEapplicationis installed,typicallyC:\ProgramFiles\MUSE.
2.ClickNewDevice.
48
MUSE™CardiologyInformationSystem
2034539-043D
Page 49
3.Setthecorrectoptionsfortheappropriatedevice. ForLifePak12enter/selectthefollowing:
SystemSetup
ForaDatamedFT,enter/selectthefollowing:
2034539-043D
MUSE™CardiologyInformationSystem
49
Page 50
SystemSetup
ForGetemedC3012,enterorselectthefollowing:
4.ClickOk.
5.Createotherdevices,updateanexistingdevice,ordeleteadeviceasrequired.
6.Whenyouarenished.clicktheclosebox(X)toexittheutility.

SettingUptheMUSEFileServerforFileCopy

ThissectiondescribeshowtosetupaFileCopyDevice.Usethisifyouwanttocopy anECGtoafolderontheMUSEleserverortoanetworkshare.
50
MUSE™CardiologyInformationSystem
2034539-043D
Page 51

SettingUpaFolder

SystemSetup
YoumustinstalltheMUSEFileCopyservicetousethisfeature.Thisserviceisnormally installedwithallMUSEv8systemsandisinitiallysetuptousethesameservicelogon account(MUSEBkgnd)thatallMUSEservicesuse.
Ifyouaresendinglestoanothercomputeronadomain,youneedtouse domain\MUSEBkgndasthelogonfortheMUSEFileCopyservice.IftheMUSE serviceswereoriginallyinstalledtousealocalMUSEBkgndaccount,youneedto changethelogonservicefortheMUSEFileCopyservice.Workwiththehospital networkadministratortocreateadomain\MUSEBkgnduseraccountthatyoucan usefortheservice.
IftheMUSEsystemisinstalledonWindows2008,workwiththedatabase administrator(DBA)toprovidethedomain\MUSEBkgndaccountwithsysadmin privilegestotheMUSEdatabases.
UsethefollowingproceduretosetupafolderforFileCopy.
1.LogonasauserwithprivilegestochangeMUSESetups.
2.OpentheMUSESetupapplication.
3.Inthenavigationpane,clickDevices.
4.Right-clickintheDevicespane,andselectNew>Folder.
5.TypetheDeviceNameandFolderDestinationintheappropriateelds.
6.IntheFileExtensioneld,ensurethattheappropriateleextensionisdisplayed.
7.FromtheOutputTypedrop-downlist,selecttheappropriateletype.
8.SelecttheappropriateOutputOptions.
9.OntheHoursofOperationtab,setupthehoursofoperationforthiscustomer andclickOK.
10.OntheAdvancedtab,setuptheformatsettingsforthiscustomerandclickOK.

MUSEAuthentication—EnableorDisable

MUSEAuthenticationisenabledbydefaultduringtheinitialMUSEinstallation.
2034539-043D
MUSE™CardiologyInformationSystem
51
Page 52
SystemSetup
Usethefollowingproceduretoturnoff(disable)MUSEAuthentication.
1.LogontotheMUSEapplicationusinganaccountwithsystemsetupprivileges.
2.GotoSystem>Setup.
3.Fromthenavigationpanel,selectSystem.
4.Right-clickonthenameoftheMUSEsystemandselectProperties. TheSystemPropertieswindowopens.
5.DeselectAllowuserstologonusingMUSEauthentication.
6.ClickOKtosaveyourchanges.
ConguringWindowsHighContrastColor SchemeonMUSEClient
Foreachworkstationthatmaybeusedbyoneormoreindividualswithcolorvision deciency(colorblindness)youmayenabletheWindowsHighContrastColor SchemeforMUSEasfollows:
1.FromtheWindowsdesktopoftheclientthatisbeingusedbyacolorimpaired individual,selectStart>ControlPanel.
2.Double-clickDisplaytoopentheDisplayPropertieswindow.
3.SelecttheAppearancetab.
4.IntheWindowsandbuttonslist,selectWindowsClassicstyle.
5.IntheColorschemelist,selectoneoftheHighContrastoptions:
•HighContrast#1
•HighContrast#2
•HighContrastBlack
•HighContrastWhite
6.ClickApply.
7.FromtheWindowsdesktop,copyaMUSEshortcuticon,whichwascreated duringinstallation,andpasteitontheWindowsdesktop.
52
MUSE™CardiologyInformationSystem
2034539-043D
Page 53
8.RenametheshortcutbyaddingWindowsDefaultColorstotheendofthename.
9.Right-clickontheshortcutandselectProperties.
10.IntheTargeteld,addthefollowingtotheendofthestring:<space>—nocui.
11.ClickOKtosaveyourchanges.
12.Repeatstep7throughstep11foreachoftheMUSEshortcuticons.
NOTE:
InformcolorimpairedsystemuserstousetheshortcutswithWindows DefaultColorsappendedtotheendofthename.
Informsystemuserswhoarenotcolorimpairedtousetheoriginalshortcuts.

ChangingMUSEServiceAccounts

TheMUSEserviceaccountsareintegraltothecorrectoperationoftheMUSEsystem. Thefollowingtableidentiesthedefaultaccounts.
TobeabletologontoboththeMUSEsystemandWindows,theaccountsmustbe setupwiththefollowinguseraccounts.
SystemSetup

MUSEUserServiceAccounts

ThedefaultMUSEAdminandMUSEBkgndaccountsareautomaticallysetupasUsers withintheMUSEapplication.TheyarecriticaltotheinternalworkingoftheMUSE systemandcannotbechanged.However,toprovideaddedsecurity,youmaymodify theirpasswords.See“ChangingtheMUSEUserAccountPasswords”onpage53.

MUSEWindowsUserServiceAccounts

TheWindowsMUSEAdminandMUSEBkgndaccountsprovideaccesstotheMUSE application,MUSEservices,andSQLdatabase.TheyarelinkedtotheMUSEuser accounts.WhileyoucanchangethepasswordsfortheWindowsaccounts,changing thenameoftheaccountsrequiresthatyouchangetheservicesandoperations dependentonthemaswell.

ChangingtheMUSEUserAccountPasswords

TochangethepasswordoftheMUSEAdminorMUSEBkgndaccount,modifythe user'spasswordintheMUSEapplication.
1.LogontotheMUSEapplication.
2.GotoSetup>Users.
3.Right-clickontheaccountandselectProperties.
4.FromtheUserPropertieswindow,enteranewMUSEpassword.
5.Saveyourchange.
2034539-043D
MUSE™CardiologyInformationSystem
53
Page 54
SystemSetup
54
MUSE™CardiologyInformationSystem
2034539-043D
Page 55

SystemAdministration

RemoteDatabases

SQLPorts

4
Thischapterprovidesproceduresforadministrativefunctionsyoumayneedtodoon theMUSEsystem.
ThissectionprovidesadditionalconsiderationstoensureMUSEsysteminstallation andoperationonaremotedatabase.RemotedatabasemeansthattheSQLinstance runningtheMUSEdatabasesislocatedonaseparateserver.InasimpleMUSE systemconguration,theMUSEapplication,andtheSQLandMUSEdatabasesare allinstalledonthesameserver.
IftheMUSEsystemisusingaremoteMUSEdatabase,theMUSEapplicationserver requiresaccesstotheMUSEdatabases.Thisrequiresthefollowingportsbeopen totheSQLserver.
SQLListeningPort
Listensforincomingconnections.ForMUSE,theseincomingconnectionsoriginate attheMUSEapplicationserver.SQLassignsTCPport1433asthelisteningport whenthedefaultinstanceofSQLisused.Ifthecustomerwantsaddedsecurity,they canchangethisport.WhenusinganamedinstanceofSQL,SQLassignsadynamic port,whichthedatabaseadministrator(DBA)canchangetoaknownxedport.
SQLMonitorServer
ThemonitorserverserviceusesUDPport1434,whichallowsadministratorsand userstocheckthestatusoftheSQLdatabases.Likethelisteningportdescribed previously,thecustomercanassignthisportadifferentportnumberforincreased security.
IftheMUSEapplicationserverishavingproblemscommunicatingwiththedatabase, andtheMUSEserviceaccountshavesysadminprivileges,contactthelocalDBA andverifythatthecorrectportsareopen.
2034539-043D
MUSE™CardiologyInformationSystem
55
Page 56
SystemAdministration

SQLRemoteConnections

Wheninstalledonaseparatedatabase,theMUSEapplicationrequiresthattheSQL remoteconnectionsareenabledasfollows.ContactthelocalITdepartmentorDBA.
1.OntheSQLserver,gotoStart>MicrosoftSQLServer2005>Conguration Tools>SQLServerSurfaceAreaConguration.
2.ClickSurfaceAreaCongurationforServicesandConnections.
3.UndertheSQLinstancewheretheMUSEdatabasesarebeinginstalled,select RemoteConnections.
4.VerifythattheLocalandremoteconnections>UsingbothTCP/IPandName Pipescheckboxisselected.
Ifitisnotselected,selecthecheckboxandapplythechange.
5.ExittheSQLServer2005SurfaceAreaCongurator.
NOTE:
InadditiontotheRemoteConnectionsconguration,youmuststarttheSQL ServerBrowserservice.

AutoShutdown

TheMUSEAutoshutdownfeaturecannotifyuserscurrentlyloggedontotheMUSE systemuptoveminutesinadvancethatitisshuttingdown.
ThisfeatureisonlyavailabletoMUSEsystemuserswhohavetheShutdownprivilege enabledfortheirrole.ThisprivilegeissetunderRole>Properties>SetupPrivileges. Rolesthathavethisprivilegebydefaultinclude:
•MUSEService
•SystemOwner
•SiteManager
Therearetwotypesofshutdown:
Full
ShutsdownALLMUSEservers,includingallHL7servicesontheMUSEserver.MUSE cannotbelaunchedfromaclientorfromtheserver.
Partial
ShutsdownMUSEclientapplicationsrunningremotely.TheMUSEservicescontinue torunandtheMUSEapplicationcanbelaunchedattheserver.

ActivatingSystemShutdown

Usethefollowingproceduretoactivatesystemshutdown.
1.IntheMUSEapplication,gotoSystem>Setup.
2.Fromthenavigationpane,selectSystem.
56
MUSE™CardiologyInformationSystem
2034539-043D
Page 57
3.Right-clickontheProductnameandclickShutdownSystem. TheShutdownSystemwindowopens.
SystemAdministration
4.SelecttheShutdownTypeandShutdownDelay.
5.ClickActivateShutdown. Apop-upmessageopensindicatingthatthesystemwillbeshuttingdown.
6.Closethemessagewindow. ThetopoftheMUSEclientapplicationdisplaysacountdown.
Ifyouselectedafullshutdown,theMUSEclientapplicationclosesandallMUSE servicesareshutdown.
Ifyouselectedapartialshutdown,allremoteconnectionstotheMUSEclient applicationclose.TheMUSEclientapplicationontheserverandallMUSE servicesontheservercontinuetorun.
NOTE:
IftheMUSEapplicationisstillopenonremoteclient,itstopsresponding andthewindowautomaticallycloses.Autoshutdowndoesnotclosethe application,onlyitsconnectiontotheMUSEserver.
2034539-043D
MUSE™CardiologyInformationSystem
57
Page 58
SystemAdministration

CancelingtheSystemShutdown

Usethefollowingproceduretocancelsystemshutdown.
1.LogontotheMUSEapplicationontheserver.
NOTE:
Ifthesystemisinfullshutdownmode,manuallystarttheMUSEservice.This startstheMUSEMThostserviceandallowsyoutologon.
2.Theapplicationdisplaysthecurrentshutdownstatusatthetopofthewindow.
3.IntheMUSEapplication,gotoSystem>Setup.
4.Onthenavigationpane,selectSystem.
5.Right-clickontheProductnameandclickShutdownSystem. TheShutdownSystemwindowopens.
6.ClickCancelShutdown. IftheMUSEserviceswerestopped,theyarenowrestartedandremote
connectivityisrestored.
NOTE:
CancelingtheshutdowndoesnotautomaticallynotifyusersthattheMUSE systemisavailable.
ModifyingtheMUSEInstalledConguration
TovieworchangethecurrentMUSEinstalledconguration,runtheMUSEv8Installer inModifymode.ThereasonsforrunninginModifymodeinclude:
•Viewinginstalledoptions.
•Addinganewlypurchasedoption.
•Addingaservicerequiredforanewoption.
•AddinganewMARSorMonitoringGatewaysystemforgeneralacquisition.
•ChangingtheMUSEPortnumber.
Beforemakinganycongurationchanges,readandobservethefollowingcautions.
CAUTION:
LOSSOFDATA:Changingsettingswithoutknowinghowtheyaffectthesystem canresultindataloss.
Donotchangeanycurrentsettingsunlessyouunderstandhowthechange affectsthesystem.
CAUTION:
LOSSOFDATA:MUSEServicesaredesignedtorestartaftermakingchangesto theinstallationcongurations.
Toavoidhavinguserslosechangestoopenrecords,usetheMUSEautoshutdown featuretonotifyusersoftheshutdown.
58
MUSE™CardiologyInformationSystem
2034539-043D
Page 59
SystemAdministration
Usethefollowingproceduretochangeanexistingconguration.
1.PerformafullorpartialshutdownoftheMUSEsystemfollowingtheShutdown proceduresdescribedinthismanual,ornotifyusersthatthesystemisbeing shutdownformaintenance.
2.ClosetheMUSEclientapplicationontheserver.
3.GototheControlPanel
•OnWindows2003,openAdd/RemovePrograms.
•OnWindows2005,selectClassicviewandclickProgramsandFeatures.
4.SelecttheMUSEv8entryandclickChange. TheMUSEInstallShieldlaunches.
5.VerifythatModifyisselected,andclickNext.
6.Gothroughthesetupstepsandmakeyourcongurationchanges.
NOTE:
YoucanclickCancelatanytimetoclosetheprogramandavoidsaving anychanges.

MovingtheMUSEDatabases

7.Ifnochangesarerequiredforasetupwindow,clickNexttoadvancetothe nextwindow.
8.ThelastwindowtoopenisMARSandMonitoringGatewayConnectivity.
NOTE:
ClickingNextatthiswindowIMMEDIATELYappliesyourchanges.
9.ClickFinish. Ifyouareaddingaserviceoroption,adialogboxdisplaysamessageindicating
thattheservicesarerestarting.
10.Afterthemodicationsarecomplete,canceltheSystemShutdown.
MovingthedatabasesrequirestheassistanceandcooperationofthelocalIT departmentanddatabaseadministrator.ContactMUSETechnicalSupportbefore performinganydatabasemoves.Movingthedatabaseconsistsofthefollowingsteps:
1.Preparingthenewdatabaseserver.
2.DetachingandCopyingtheMUSEdatabasesformthecurrentserver.
3.AttachingthedatabasestothenewSQLserver.
2034539-043D
MUSE™CardiologyInformationSystem
59
Page 60
SystemAdministration
4.UninstallingtheMUSEapplication.
5.ReinstallingtheMUSEapplication.
NOTE:
YoumustuninstallandreinstalltheMUSEapplicationtoestablishthenewlocation ofthedatabase.UninstallingtheMUSEapplicationalsouninstallsMUSEWeb.Any customizationsthatweremadetotheMUSEWebsiteneedtoberecongured.

PreparingtheNewDatabaseServer

1.CreatetheSQLinstanceonwhichtheMUSEapplicationwillrun.
2.AssignsysadminrightstotheMUSEAdminandMUSEbkgndaccounts. Thesemustbethesameaccountsthatareintheadministratorsgrouponthe
MUSEApplicationserver.TheMUSEBkgndaccountmustbethesameaccount thatisrunningMUSEservices.RefertotheMUSEv8CardiolotyInformation SystemInstallationManualformoreinformation.
3.CreateafolderontheSQLserverwherethedatabaseswillbecopied. Whileyoucanuseanyfoldername,itisrecommendedthatyoucreatea
MUSE\dbfoldertoremainconsistentwithstandardMUSEcongurations.
NOTE:
Foradditionalinformation,see“RemoteDatabases”onpage55.

DetachingandCopyingtheMSUEDatabasesfromthe CurrentServer

1.StopallMUSEservices.
2.OpenSQLServerManagementStudioandconnecttothecurrentMUSE databaseserver.
3.ExpandtheDatabaseobject.
4.4.Right-clickontheMUSE_Systemdatabase,selectTasksand,clickDetach.
5.OntheDetachDatabasewindow,conrmthenameofthedatabaseyouare detachingandclickOK.
6.Repeatstep4andstep5foreachMUSEdatabase.(MUSE_SiteXXX,MUSE_Site Template)
7.CopytheMUSEdatabasestothenewserver.

AttachingtheMUSEDatabasestotheNewSQLServer

1.OpenSQLServerManagementStudioandconnecttotheSQLserverwherethe MUSEdatabaseswerecopied.
2.Right-clickontheDatabaseobjectandselectAttach.
3.OntheAttachDatabasewindow,clickAddandbrowsetothelocationofthe MUSEdatabases.
4.SelecttheMUSE_System.mdfle
5.ClickOKtoattachthele
60
MUSE™CardiologyInformationSystem
2034539-043D
Page 61
6.Repeatstep2throughstep5foreachMUSEdatabase.(MUSE_SiteXXX, MUSE_SiteTemplate).
7.AfterallMUSEdatabasesareattachedtothenewSQLinstance,followthe proceduresfor“Uninstalling/ReinstallingtheMUSEApplication”.

Uninstalling/ReinstallingtheMUSEApplication

UninstallingtheMUSEapplicationremovesthefollowingcomponentsandentries fromtheserver:
•MUSEdirectoryinProgramFiles
•MUSEservices
•MUSEdesktopshortcuts
•MUSEentryintheStartMenu
•MUSEWebfolders
ThefollowingMUSEcomponentsremainonthesystem:
SystemAdministration
•MUSEdatabases(dbfolder)
•MUSEdatabasesubfolders(acq,backup,xml,logs,andsoforth)
•.NetFramework
•AdobeReader
•MUSEWebsite(YouneedtomanuallyremovebeforereinstallingtheMUSE
NOTE:

UninstallingtheMUSEApplication

TouninstallMUSE:
1.LogontotheMUSEapplicationserverasadministrator.
2.IntheInstalledProgramslist,selecttheMUSEv8.xentry.
3.SelectMUSE8.xandclickUninstall.
application.
IfyouplanonreinstallingtheMUSEapplicationlater,itisagoodideatorunMUSE v8.0inModifymodeandmakenoteoftheexistingservicesthatareinstalled.
•InWindows2003,clickStart>ControlPanelandselectAddorRemove
Programs.
•InWindows2008,clickStart>ControlPanel>Classicview>Programs
andFeatures.

ReinstallingtheMUSEApplication

ProvidingthatallrequirementsandpermissionsforMUSEandMUSEdatabaseare met,youcanreinstalltheMUSEapplicationfollowingtheinstructionsintheMUSE
v8CardiologyInformationSystemInstallationGuideunderthesectionInstalling theMUSEApplicationandDatabase.
2034539-043D
MUSEisuninstalledfromthesystem.
MUSE™CardiologyInformationSystem
61
Page 62
SystemAdministration

RenamingtheMUSEServer

IftheMUSEapplicationisalreadyinstalledonthesever,usethefollowingprocedureto renametheserver.
1.HavethelocalITornetworkadministratorcompletethefollowingsteps: a.Renameandrestarttheserver. b.Verifythenamechangeandthatthecomputerhasjoinedthedomain.
2.LogontotheMUSEapplicationserverasadministrator.
3.IfSQLisinstalledonaseparateservergotostep4. IfSQLisinstalledonthesameserver,renamethelocalinstanceofSQLas
describedinthefollowingsteps. a.OpenSQLServerManagementStudio.
b.Connecttotheserverusingthenewname. c.ClickNewQuery. d.OntheQuerytabtypethefollowingcommands
execsp_dropserver‘oldname’
execsp_addserver‘newname'
execsp_helpserver
where'oldname'istheoldcomputernameand‘newname’isthenew computername.
e.Executethecommandandverifythatthenewnameisdisplayedonthe
Resultstab.
4.IntheInstalledProgramslistselecttheMUSEv8.xentry.
•InWindowsXP,clickStart>ControlPanelandselectAddorRemove
Programs.
•InWindows7,clickStart>ControlPanel>ProgramsandFeatures.
5.SelectModifyandclickNext.
6.ClickNextoneachwindowuntiltheMUSEClientCongurationwindowopens.
62
MUSE™CardiologyInformationSystem
2034539-043D
Page 63
Thenewnameoftheservershouldbedisplayed.
7.ClickNext. TheMUSEServerValidationwindowopens.
8.ClickYestoacceptthechange.
9.ClickNextontheremainingwindows.
10.OntheMaintenanceCompletewindow,clickFinish. TheMUSEservicesrestart.
11.Repeatstep4thoughstep10foreachMUSEclient.
12.VerifythatyoucanopenMUSEontheserverandonaclient.
SystemAdministration
13.IfMUSEWebisinstalledontheserver,notifyusersofthenewURL.

ChangingthePortNumberoftheFileServer

TheportnumbertheMUSEsystemusesissetduringinitialinstallationandyoushould notneedtochangeit.TheportnumberonallMUSEclientsmustmatchtheport numbersetontheserver.Ifaportnumberchangeisrequired,usethefollowing procedure
NOTE:
ChangingtheportnumbercausesallMUSEservicestorestart.Beforechanging theportnumber,usetheMUSEautoshutdownfeaturetonotifyusersofthe systemshutdown.

ChangingthePortNumberontheMUSEServer

UsethefollowingproceduretochangetheportnumberontheMUSEserver.
1.LogontotheMUSEapplicationserverasadministrator.
2.PerformaSystemShutdown.(See“AutoShutdown”onpage56.)
3.IntheInstalledProgramslist,selecttheMUSEv8.xentry.
•InWindows2003,clickStart>ControlPanel,andselectAddorRemove
Programs.
•InWindows2008,clickStart>ControlPanel>Classicview>Programs
andFeatures.
4.SelectMUSE8.xandclickChange. TheWelcomewindowopens.
5.SelectModifyandclickNext.
6.ClickNextuntiltheMUSEClientCongurationwindowopens.
2034539-043D
MUSE™CardiologyInformationSystem
63
Page 64
SystemAdministration
7.Entertheportnumberyouwanttouse. Theprogramcheckstoseeiftheportisfree.Iftheportisinuseyoureceive
amessage.
8.ClickYestocontinuewiththeinstallationusingtheportentered,orclickNoto gobacktotheMUSEClientCongurationwindowtochangetheportnumber orCanceltheinstallation.
9.ClickNextontheremainingwindowstosaveyourchange.
10.OntheMaintenanceCompletewindow,clickFinish. TheMUSEservicesrestart.
11.Afterthechangeiscomplete,canceltheSystemShutdown.

ChangingthePortNumberontheMUSEClients

UsethefollowingproceduretochangetheportnumberontheMUSEClient.
1.LogontotheMUSEclientasadministrator.
2.IntheInstalledProgramslist,selecttheMUSEv8.xentry.
•InWindowsXP,clickStart>ControlPanelandselectAddorRemove
Programs.
•InWindows7,clickStart>ControlPanel>ProgramsandFeatures.
3.SelectModifyandclickNext.
4.ClickNextuntiltheMUSEClientCongurationwindowopens.
5.Entertheportnumberyouwanttouse. ThismustbethesameportyouconguredfortheMUSEserver.Theprogram
checkstoseeiftheMuseserverisusingtheportyouentered.IftheMUSEserver isnotusingtheportyoureceiveamessage.
6.ClickYestocontinuewiththeinstallationusingtheportentered,orclickNoto gobacktotheMUSEClientCongurationwindowtochangetheportnumber orcanceltheinstallation.
7.ClickNextontheremainingwindowstosaveyourchange.
8.OntheMaintenanceCompletewindowclickFinish.

BIOSUpdates

UpgradingtheG4ServerBIOSFirmware

Ifyouneedtoreplacethe1MBprocessorthatoriginallyshippedwiththeG4 applicationserver,youmayneedtoreplaceitwitha2MBprocessor.
NOTE:
Afterreplacingtheprocessor,verifyifyouneedtoupgradetheBIOSrmware. YouneedtoupgradetheBIOSrmwareifthedatedisplayedwhentheapplication
serverbootsisearlierthan11/09/2005(November2005).IftheBIOSrmwaredate islater,youdonotneedtoupgradetheBIOSrmware.
64
MUSE™CardiologyInformationSystem
2034539-043D
Page 65
CreatingaBIOSUpgradeFloppyDisk
Youneedtocreateaoppydiskyoucanusetobootthesystem.
1.InsertablankdisketteintotheoppydriveandtheMUSESupportCDintothe CD-ROMdrive.
2.SelectStart>Run. TheRunwindowopens.
3.ClickBrowseandnavigateto:[CDdriveletter]:\BIOS\ML370G4\SP31871.exe.
4.SelecttheexecutableandpressOpen. TheRunwindowispopulatedwiththeexecutable.
5.ClickOK.
6.Followtheon-screenpromptstocreatetheBIOSupgradeoppydisk.
7.Whentheprocessisdone,removetheCDfromtheCD-ROMdrive.
UpgradingtheBIOSFirmware
SystemAdministration
1.Rebootthesystemwiththeoppydiskinthedrive. ThesystemshouldautomaticallybootfromtheoppydiskanddisplayROMPAQ
FirmwareUpgradeUtilityV4.20M.
NOTE:
Ifthesystemdoesnotbootfromtheoppy
1.RebootthesystemandpressF9duringthebootprocess.
2.Selecttheoppyandthebootdrive.
2.AttheSelectaDevicewindow,selectCompaqML370G4andpressEnter.
3.AttheSelectanImagewindow,select11/09/2005andpressEnter.
4.AttheCautionwindow,readandfollowthewarning:Donotrebootorturnoff yourmachinewhileprogrammingisinprocess;pressEnter.
5.Waitfortheupgradeprocesstocomplete(2–3minutes)anddisplay:Flash ProgrammingCompletedSuccessfully;pressEnter.
6.Atthemessage:DoyouwanttoreprogramanotherROM?pressESC.
7.Removetheoppydiskandpowercyclethesystem.
2034539-043D
MUSE™CardiologyInformationSystem
65
Page 66
SystemAdministration
66
MUSE™CardiologyInformationSystem
2034539-043D
Page 67

Maintenance

MaintenanceGuidelines

5
Aregularequipmentmaintenanceprogramhelpspreventunnecessaryequipment andpowerfailuresandalsoreducespossiblehealthhazards.Thischaptercontains instructionsforthefollowingrecommendedmaintenance:
•Cleaningtheequipment'saccessories
•Visualinspectionofequipment
•Exteriorcleaningofequipment
•Conductingrepairandreplacementcheckoutprocedures.
GEHealthcarerecommendsthatyouperformthetestsdescribedinthischapter:
•Everyyearaspartofroutinemaintenance.
•Wheneverinternalassembliesareserviced.
•Wheneveryourepairorreplaceequipment.
•AsprescribedintheOEMdocumentationthatcomeswiththeequipment.
NOTE:
UnlessyouhaveanEquipmentMaintenanceContract,GEHealthcaredoesnot assumeresponsibilityforperformingtherecommendedmaintenanceprocedures. Thesoleresponsibilityrestswiththeindividualorinstitutionusingtheequipment. GEHealthcareservicepersonnelmay,attheirdiscretion,followanyorallofthe proceduresasaguideduringvisitstotheequipmentsite.
Itemsrequiringmaintenanceinclude:
•MUSEserver(s)
•HL7InterfaceServer(CCG)
•MUSEClient
•Monitors
•DVD-RWorCD-RWDrive
•Printer
2034539-043D
MUSE™CardiologyInformationSystem
67
Page 68
Maintenance

OEMMaintenance

ForinformationonmaintenanceofOriginalEquipmentManufacturer(OEM) components,pleasereferencetheappropriateOEMmanualsfortherecommended maintenanceoftheirproduct.ForlinkstotheOEMmanuals,refertotheMUSE CardiologyInformationSystemHardwareManual.
Fortechnicalsupport,contactyourlocaltechnicalsupportteam. Fortechnicalsupport,contactyourlocalGEHealthcaretechnicalsupportteam.

RequiredToolsandSupplies

Inadditiontoastandardsetofhandtools,youneedthefollowingspecialtoolsand itemstomaintainorcheckoutthesystem:
RequiredToolsandSupplies

InspectionandCleaning

Item
DVOM
LeakageCurrentTesterusedfortheElectricalSafetyCheck CD-ROMCleaningKitanylocallyavailablekt
Theequipmentissealedbeforeitleavesthefactory.Thereshouldbenodustbuildup onthesurfacesoftheinteriorPCBassembliesandcomponentswhenyoureceiveit.
Ifdustisanenvironmentalproblem,useacommerciallyavailabledustremover (compressedair)tocleanthefollowinginternalcomponents.
•PCBassemblies
•Coolingfans
•Expansioncards
•Vents
Beforebeginninganyofthefollowinginspectionorcleaningtasks,shutdownthe systemanddisconnectitfromitspowersourceusingtheappropriateinstructions:
•ForinstructionsonshuttingdowntheMUSEsystem,referto“SafeShutdown
Procedures”.
Description
DigitalVolt/Ohmmeter
•Forinstructionsonshuttingdownindividualcomponentsanddevices,referto
theappropriateOEMmanuals.
68
MUSE™CardiologyInformationSystem
2034539-043D
Page 69

SafeShutdownProcedures

ShuttingDowntheServer
1.Beforeshuttingdownthesever,notifyallusersofthescheduledshutdown. Forhardwaremaintenance,doaFullSystemshutdown.(See“AutoShutdown”
onpage56.)
2.AftertheMUSEapplicationisshutdown,thenshutdowntheserverfollowing normalWindowsprocedures.
ShuttingDownMUSEClient
1.BeforeexitingtheMUSEapplication,closeallpatienttests.
2.Closeanybrowserwindowsorotherapplicationsthatareopen.
3.ShutdownthesystemfollowingnormalWindowsshutdownprocedures.

Precautions

Maintenance
•Turnofftheunitandremoveallpowerbeforeinspectingorcleaning.
WARNING:
ELECTRICSHOCK:Improperuseofthisdevicepresentsashockhazard.Strictly observethefollowingwarnings.Failuretodosomayendangerthelivesofthe user,andbystanders.
Whendisconnectingthedevicefromthepowerline,removetheplugfromthe walloutletrst,beforedisconnectingthecablefromthedevice;otherwise, thereisariskofcomingincontactwithlinevoltagebyinadvertentlyintroducing metalpartsinthesocketsofthepowercord.
Devicesmaybeconnectedtootherdevicesortopartsofsystemsonlyafter makingcertainthatthereisnodangertothepatient,theoperators,orthe environmentasaresult.StandardsEN/IEC60601-1-1mustbecompliedwith inallcases.
•Donotimmerseanypartoftheequipmentinwater.
•Donotdripwateroranyliquidonthewriterassembly,andavoidcontactwith
openvents,plugs,orconnectors.
WARNING:
ACCIDENTALSPILLS:Ifliquidshaveenteredthesystem,takeitoutofserviceand haveitcheckedbyaservicetechnicianbeforeusingitagain.
Toavoidelectricshockordevicemalfunction,donotallowliquidstoenter thesystem.
CAUTION:
DATALOSSORSYSTEMFAILURE:Datalossorsystemfailurecanresultdueto ingressofliquids.Thesystemdoesnotprovideprotectionagainstingressof liquids.
Ensureinstallationinacool,dryenvironment.
2034539-043D
MUSE™CardiologyInformationSystem
69
Page 70
Maintenance
•Donotuseorganicsolvents,ammonia-basedsolutions,orabrasivecleaningagents
thatmaydamageequipmentsurfaces.
WARNING:
EXPLOSIONHAZARD:Flammableanestheticvaporsorliquidscancause explosions.
DoNOTuseinthepresenceofammableanestheticvaporsorliquids.
•Donotusemetalarticles,suchasascrewdriver,tocleanthetapeheads.
•Donotbringanymagneticmaterialneartheheadassembly.
CAUTION:
EQUIPMENTDAMAGE/DATALOSSORCORRUPTION/PERSONALINJURY:This equipmentcontainsnouser-serviceableparts.Usingnon-approvedservice equipmentorprocedurescancausedamagetotheequipment,lossorcorruption ofdata,orpersonalinjury.
Onlyqualiedpersonnelshouldperformthemaintenanceprocedures.

VisualInspection

Performavisualinspectionregularly.Turnofftheunitandremovethecoverbefore inspectingtheunit.
•Checkthecaseanddisplayscreenforcracksorotherdamage.
•Inspectallcordsandcablesforfrayingorotherdamage.
Performsafetytestsonanyrepairedlinecords.
•Inspectallplugs,cables,andconnectorsforbentprongsorpins.
•Verifythatallcords,socketedcomponents,andconnectorsaresecurelyseated.
•Verifythatallkeys,knobs,andpowerswitchesfunctionproperlyanddonotstickin
oneposition.
Iftheinspectionfails,donotusetheunituntilthefailedconditioniscorrected.

CheckCoolingFansandVentilation

Usethefollowingproceduretocheckcoolingfansandventilationoftheunits.
1.Withtheunitoperating,verifythatallfanscontainedinthefollowingunitsare operatingproperly:
•Client
•Server
2.Checkthepositionoftheunittoensureadequateventilation:
•Verifytheserverisinapositionwheretheventilationholesinthefrontand
reararenotblockedorrestricted.
•Verifythattheserverisnotplacednearducts,pipes,orequipmentthat
generateheat.
CAUTION:
DAMAGETOSYSTEMCOMPONENTS:Poorventilationcancauseoverheating. Positiontheunittoensureadequateventilation.
70
MUSE™CardiologyInformationSystem
2034539-043D
Page 71

ExteriorCleaning

Cleantheexteriorsurfacesoncepermonthandmorefrequentlyifneeded.
1.Turnofftheunitandremovepowerbeforecleaning.
2.Dipaclean,softclothinamilddishwashingdetergentdilutedinwater .
3.Wringtheexcesswaterfromthecloth. Donotdripwateroranyliquidonopenvents,plugs,orconnectors.
4.Wipethesystemsurfaceswiththedampcloth.
5.Drythesurfaceswithacleanclothorpapertowel.

FunctionalCheckoutProcedures

Wheneverasystemisserviced,youmustperformcheckoutprocedurestocomply withFDAguidelinesandtoensurethatthesystemissafeandfunctioningproperly. Thespecicproceduresdependontheserviceperformed.Checkoutproceduresare separatedintotwocategories:
Maintenance
“HardwareFRURepairs”
“Non-FRURepairs”
Followthecheckoutprocedureappropriatefortherepairyouperformed.
NOTE:
Forcustomer-suppliedhardware,thecustomerisresponsiblefortroubleshooting, FRUreplacement,andcheckouts,astheyrelatetohardwarerepairsonthe systemservers.

HardwareFRURepairs

MakeFRUreplacementsfollowingtheappropriaterepairproceduresoutlinedin themanufacturer'smanualforthesystem.Forsystemsunderthemanufacturer's warranty,theOEMservicerepresentativemayperformtherepair.
NOTE:
FinalfunctionalcheckoutofthesystemistheresponsibilityoftheGEHealthcare servicerepresentativeonsiteatthetimeoftherepair.
TheFRUcheckoutprocedureforanylistedFRUalsoappliestoitsinternalPCBs andcomponents.
RequiredTools
Thefollowinglistidentiesallstandardtoolsthatyoumayneedtoperformcheckout procedures.
•Standardhandtools,includinga#10TorxDriver
•Antistaticwriststrap
•Canofcompressedair
•ApplicableServiceand/orOperatormanual,asneeded.(UseOEMmanualsas
reference.)
2034539-043D
MUSE™CardiologyInformationSystem
71
Page 72
Maintenance
Aftercompletinganyrepair,youneedtocompletethefollowingcheckoutprocedures:
•VisualInspection:See“VisualInspection”onpage72.
•Server:See“Server”onpage73.
•Client:See“Client”onpage73.
LocatetherepairedFRUinthefollowingtablestoidentifyadditionaltoolsand checkoutprocedurestouse:
FRUCheckoutProcedures
FRUDescription
HarddriveSee“HardDrive”onpage73.
PowersupplySee“ElectricalSafetyChecks”
FloppydriveSee“FloppyDrive”onpage
OpticaldriveSee“OpticalDrive”onpage
DisplaySee“Display”onpage73. MultimediaCardReader
(external) ModemSee“Modem”onpage74. DiskArrayControllerSee“DiskArrayController(GE
TapeDrive/TapeDrive Controller
AdditionalCheckout Procedures
onpage74.
73.
73.
See“MultimediaCard Reader”onpage74.
Healthcare-suppliedserver only)”onpage74.
See“TapeDrive/Tape Controller”onpage74.
AdditionalTools
•Leakagetester
•Multimeter Floppydiskette
CDorDVD
SDCard
Tape
VisualInspection
Toconductavisualinspection,dothefollowing:
•Checkthecaseanddisplayscreenforcracksorotherdamage.
•Inspectallcordsandcablesforfrayingorotherdamage.
Performsafetytestsonanyrepairedlinecords.
•Inspectallplugs,cables,andconnectorsforbentprongsorpins.
•Verifythatallcords,socketedcomponents,andconnectorsaresecurelyseated.
•Ifthesystemboxwasopened,checkthattheinteriorisfreeofexcessivedust
buildup. Ifnecessary,usecommerciallyavailablecompressedairtoclean,followingthe manufacturer'sinstructions.
ComponentCheckoutProcedures
Usetheappropriatecheckoutprocedures(identiedintheprevioustables)toverify thatthecomponentsareworkingproperlyafteraHardwareFRUreplacementor repair.Fordetailedinstructionsforperforminganyofthefunctionaltasks(suchas
72
MUSE™CardiologyInformationSystem
2034539-043D
Page 73
Server
Client
Maintenance
acquiringdata,selectingpatients,andsoforth),refertotheOperator'smanualfor yoursystem.
NOTE:
Verifythatrack-mountedsystemsarefullyseatedandfastenedtotherackbefore proceeding.
1.Restarttheserver.
2.Conrmthatthesystemstartsnormally(thatis,nounknownerrorsoccur).
3.ConrmyoucansuccessfullylogontotheMUSEapplicationattheserver.
4.ConrmthattheusercansuccessfullylogontotheMUSEapplicationfroma MUSEclient.
5.ConrmthatausercanopentheEditListandaccessapatientrecordata clientworkstation.
HardDrive
Display
FloppyDrive
1.Restartthesystem.
2.Conrmthatthesystemstartsnormally(thatis,nounknownerrorsoccur).
3.ConrmthattheusercansuccessfullylogontotheMUSEapplicationfroma MUSEclient.
4.ConrmthatausercanopentheEditListandaccessapatientrecordata clientworkstation.
OnasystemequippedwithRAID,conrmthatthereplacementdriveisoperational,as describedintheOEMservicemanual.Onasingledrivesystem,verifythatthedisaster recoveryprocedurerestoredthesystemandanyavailabledata.
1.Conrmthatthedisplayresolutionissettotherecommendedresolution.
2.StarttheMUSEapplicationandconrmitdisplayscorrectly.
1.Inserttheoppydiskintothediskdrive.
2.AccesstheoppydiskusingMyComputerorWindowsExplorer.
3.Conrmthatthesystemrecognizestheoppydiskandyoucanviewitscontents.
OpticalDrive
2034539-043D
4.Conrmthatyoucanopenlesfromandsavelestotheoppydisk.
5.ConrmyoucanacquiredataintotheMUSEapplicationfromtheoppydisk. Thisisrequiredonlyifthecustomerusestheoppydrivetoacquiredata.
1.InsertaCDorDVDintotheopticaldrive.
2.AccessthemediausingMyComputerorWindowsExplorer.
MUSE™CardiologyInformationSystem
73
Page 74
Maintenance
MultimediaCardReader
Modem
3.Conrmthatthesystemrecognizesthemediaandyoucanviewitscontents.
4.Conrmthatyoucanopenlesfromthemedia.
1.InsertanSDcardintothereaderandaccessitusingMYComputerorWindows Explorer.
2.Conrmthatthesystemrecognizesthecardandyoucanviewitscontents.
3.Conrmthatyoucanopenlesfromandsavelestothecard.
4.Conrmthatyoucanacquiredataintothesystemfromthecard.
1.TransmitanECGfromacarttothemodem.
2.IntheMUSEapplication,selectSystemStatus.
3.SelectNewlyAcquiredorAcquisitionLog.
4.VerifythePID/NameoftheECGtransmittedfromthecartwasacquired.
DiskArrayController(GEHealthcare-suppliedserveronly)
1.ConrmthattheC:andD:drivepartitionswerecreatedcorrectly.
2.Verifythatthedisasterrecoveryprocedurerestoredthesystemdatabase.
TapeDrive/TapeController
1.UseSQLServerManagementStudiotorunamanualsystembackup.
2.Conrmthebackupwassuccessful.
3.Verifythatalltapesweresuccessfullyusedaccordingtothebackupschedule.
4.Replacetapesthatarepasttheirintendedlifespan.
ElectricalSafetyChecks
CAUTION:
EQUIPMENTFAILURE/HEALTHHAZARD:Failuretoimplementasatisfactory maintenanceschedule.
YouneedtoperformthesetestsanytimetheACgroundisdisturbed,forexample whenyouserviceorreplacethepowersupplyormainsystemboard.
Theseproceduresverifythataproperlywiredoutletisfurnishingpowertothesystem.
1.Connecttheappropriateleakagetestertothepoweroutlet.
2.EnsurethattheGNDswitchontheleakagetesterisinthedown(closed)position.
74
MUSE™CardiologyInformationSystem
2034539-043D
Page 75
3.RefertothefollowingElectricalSafetyCheckstableforadescriptionofthe
teststoperform.
4.Recordthemeasurementvaluesinyourdebrief.
NOTE:
TheLeakageCurrentLimitsgiveninthefollowingtableassumethatyour workstationisoutsideofthepatientvicinity.Ifitislocatedwithinthepatient vicinity,contactGEHealthcareformoreinformation.
ElectricalSafetyChecks
Maintenance
Step
SafetyCheckConditionResult
LeakageCurrentLimitsin MicroAmps(µA)
4
UL60950–1
A
EarthLeakageCurrent(UUT-ONinµA)
1
ForwardPolarity
2
NeutralOpen,Forward
IEC60950–1 TypeITETypeITE
1
NC
2
SFC
Pass/Fail Pass/Fail
35003500 35003500
Polarity
3
NeutralOpen,Reverse
2
SFC
Pass/Fail
35003500
Polarity
4
ReversePolarity
B
EnclosureLeakageCurrent(UUT-ONinµA)
1
ForwardPolarity
2
NeutralOpen,Forward
1
NC
Pass/Fail
35003500
TypeITETypeITE
1
NC
2
SFC
Pass/Fail Pass/Fail
35003500 35003500
Polarity
3
GroundOpen,Forward
2
SFC
Pass/Fail
35003500
Polarity
4
GroundOpen,Reverse
2
SFC
Pass/Fail
35003500
Polarity
5
NeutralOpen,Reverse
2
SFC
Pass/Fail
35003500
Polarity
6
ReversePolarity
E
GroundContinuity(UUT-ONinΩ)
1
ACmainspowercord groundprongtoexposed
1
NC
3
N/A
Pass/Fail
Pass/Fail
35003500
Resistancein Ω
Lessthan200 mΩ
Resistance inΩ
Lessthan
200mΩ metalsurface(thatis, groundlug)
1
NC=Normalcondition
2
SFC=SingleFaultCondition
3
N/A=NotApplicable
4
CountriesrecognizingIECstandardsuseIECvalues.CountriesrecognizingULstandards
useULvalues.
4
2034539-043D
MUSE™CardiologyInformationSystem
75
Page 76
Maintenance

Non-FRURepairs

Systemfunctionalchecksarerequiredfornon-FRUrepairsandtypicallyinvolve systemsetupandcongurationsthatyoucanperformremotelyoronsite.Ifyou performthemremotely,theremotesupportengineercanconrmthemthrough remoteaccess,orverifythemwiththecustomercontact.Foradditionalinstructions, refertotheOperator'smanualforyoursystem.
Printing
Usethefollowingprocedurestoverifythattheprintingfunctionsareworkingcorrectly.
ManualPrintCheckout
1.HighlightarecordontheEditlistandclickPrint.
2.Fromthedrop-downmenu,selecttheprinteryouwanttotestandclickOK.
3.Verifythattherecordprintedoutcorrectlyattheselectedprinter.
AutomaticPrintCheckout
1.Acquiredatafromacart/stressdevicewiththelocationnumberyouaretesting.
2.Verifythattherecordprintsoutcorrectlyattheprintersetupforautomatic distribution.
FormatSettingChangeCheckout
1.HighlightarecordontheEditlistandclickPrint.
2.Fromthedrop-downmenu,selecttheprinteryouwanttotestandclickOK.
3.Comparethenewprintouttoapreviousprintouttoverifyformatchangestook effect.
Faxing
Usethefollowingprocedurestoverifythatthefaxingfunctionsareworkingcorrectly.
ManualFaxtoTemporDenedDeviceCheckout
1.HighlightarecordontheEditlistandclickPrint.
2.Sendtherecordtoatemporaryfaxmachine.
3.Checkthatthefaxprintedcorrectlyatthereceivingfaxmachine.
AutomaticFaxtoExistingFaxDeviceCheckout
1.AcquiredatafromanECGcartwiththelocationsetupforthecorrectreport distribution,orconrmdatawithcorrectphysicianmappingforauto-fax.
2.Checkthatthefaxprintedcorrectlyatthereceivingfaxmachine.
Editing/Conrming
Usethefollowingprocedurestoverifythattheeditingandconrmingfunctionsare workingcorrectly.
EditRecordCheckout
1.HavethecustomeropenarecordontheEditlist.
2.UpdatetherecordintheEditlist.
3.Re-opentherecordtoverifythechangesweresaved.
76
MUSE™CardiologyInformationSystem
2034539-043D
Page 77
ConrmRecordCheckout
1.HavethecustomeropenarecordontheEditlist.
2.Havethecustomerconrmtherecord.
3.Doaretrievaloftherecordandverifythatitisnowinthedatabase.
ReportDistributionCheckout
Usethefollowingprocedurestoverifythatthereportdistributionfunctionisworking correctly.
1.Acquireorconrmdataforthelocationofanewreportdistributionsetup.
2.Checktheoutputdevicesetupinthereportdistributionforthereport.
DeviceSetup
Usethefollowingprocedurestoverifythatthedevicesaresetupcorrectly.
OutputDevice(Printer,Fax,HL7Results)
Maintenance
UserSetup
CartsTransmission/Acquisition
1.HighlightarecordontheEditlistandclickPrint.
2.Fromthedrop-downmenu,selectthenewlycreateddeviceandclickOK.
3.Checkthedeviceforthereport.
InboundDevice(Modem,Share)
1.Transmitarecordfromasourcedevice.
2.Verifythattherecordisdisplayedintheapplication.
Usethefollowingprocedurestoverifythattheuserissetupcorrectly.
1.Haveanewlycreateduserstarttheapplication.
2.Havetheuserlogonandconrmapplicationaccessbasedonroles/privileges givenduringsetup.
UsethefollowingprocedurestoverifythattheCartstransmissionsandacquisitions functionsareworkingcorrectly.
ModemTransmissionCheckout
1.TransmitanECGfromthecarttoamodem.
2.Intheapplication,selectSystemStatus.
3.SelectNewlyAcquiredorAcquisitionLog.
4.CheckthatthePID/NameoftheECGtransmittedfromthecartwasacquired intheapplication.
WirelessTransmissionCheckout
1.TransmitanECGfromacartwhileinrangeofawirelessantennae.
2.Intheapplication,selectSystemStatus.
2034539-043D
MUSE™CardiologyInformationSystem
77
Page 78
Maintenance
3.SelectNewlyAcquiredorAcquisitionLog.
4.CheckthatthePID/NameoftheECGtransmittedfromthecartwasacquired intheapplication.
LANTransmissionCheckout
1.TransmitanECGfromacartthroughtheLAN.
2.Intheapplication,selectSystemStatus.
3.SelectNewlyAcquiredorAcquisitionLog.
4.CheckthatthePID/NameoftheECGtransmittedfromthecartwasacquired intheapplication.
RemovableMediaAcquisition
1.Insertthemediaintotheclient.
2.AcquirethemediathroughMUSEAcquisition.
3.Intheapplication,selectSystemStatus.
4.SelectNewlyAcquiredorAcquisitionLog.
MARSTransmission
CASE/CardiosoftTransmission
5.CheckthatthePID/NameoftheECGtransmittedfromthecartwasacquired intheapplication.
MARStoMUSESetupCheckout
1.FromaMARSworkstation,sendaHolterreporttotheMUSEsystem.
2.Intheapplication,selectSystemStatus.
3.SelectNewlyAcquiredorAcquisitionLog.
4.CheckthatthePID/NameoftheECGtransmittedfromtheMARSsystemwas acquiredintheMUSEsystem.
MARSFormattedReportPrintCheckout
1.IntheMUSEsystem,highlightaHolterreportontheEditlistorretrievearecord fromthedatabase.
2.ClickPrint.
3.ComparethenewlyprintedHolterreporttotheprintoutpriortotheMARS formatterinstallationandcheckthattheHolterformatchanged.
1.SendaStressreporttotheMUSEsystem.
2.Intheapplication,selectSystemStatus.
3.SelectNewlyAcquiredorAcquisitionLog.
4.CheckthatthePID/NameoftheECGtransmittedfromthecartwasacquired intheapplication.
78
MUSE™CardiologyInformationSystem
2034539-043D
Page 79
MonitoringTransmission
DatabaseSearch
Maintenance
Manualtransmissionfromabesidemonitor
1.FromtheBedside/Dashmonitor,transmit12SLtotheMUSEsystem.
2.IntheMUSEsystem,selectSystemStatus.
3.SelectNewlyAcquiredorAcquisitionLog.
4.CheckthatthePID/NameoftheECGtransmittedfromthecartwasacquired intheapplication.
Usethefollowingprocedurestoverifythatthedatabasesearchfunctionisworking correctly.
ManualSearchCheckout
1.GototheDatabaseSearchfunctionoftheapplication.
2.Createasearch.
3.Choosecriteriaandrunthesearch.
4.Verifythatthesearchresultsmatchtheselectedcriteria.
AutomaticSearchCheckout
1.GototheDatabaseSearchfunctionoftheapplication.
2.Createasearch.
3.Savethecriteria.
4.Schedulethesearch.
5.Verifythatthesearchrunswhenscheduledandexpectedresultsaregenerated.
WebRetrievalCheckout
Usethefollowingprocedurestoverifythattheretrievalcheckoutfunctionsare workingcorrectly.
MUSEWebDataRetrievalCheckout
1.StartInternetExplorer.
2.IntheURLeld,typetheMUSEservername(forexample,http://museserver).
3.ChooseFramesorNoFrames.
TheMUSEv8systemrunsallscheduledsearchesonlyonceperday.
Ifprompted,loginasauserwhohasaccesstoMUSEWeb.
4.InthePIDeld,typethepatientID,orintheLastNameeld,typethepatient name.
5.FindtherecordtoverifythatMUSEWebisworkingproperly.
CVWebDataRetrievalCheckout
1.StartInternetExplorer.
2.IntheURLeld,typetheCVWebservername(forexample,http://cvweb). Ifprompted,loginasauserwhohasaccesstoMUSEWeb.
2034539-043D
MUSE™CardiologyInformationSystem
79
Page 80
Maintenance
3.InthePIDeld,typethepatientID,orintheLastNameeld,typethepatient
4.VerifythattheappropriateMUSEserver/siteisselected.
5.FindtherecordtoverifythatCVWebisworkingproperly.
RemoteSupport
Usethefollowingprocedurestoverifythattheyoucanaccessthesystemforremote support.
1.Logontothecustomer'ssystemusingtheremoteconnectionconguredfor
2.Conrmthatyoucanaccessthecustomer'sdesktopviatheremoteconnection.
3.Conrmthatyoucanuploadanddownloadlestothecustomer'ssystem.
HIS/CCG
UsethefollowingprocedurestoverifythattheHISandCCGfunctionsareworking correctly.
name.
thatsystem(pcAnywhereorInSiteExC).
XMLImport
HISADT/OrdersInboundCheckout
1.FromtheMUSEsystem,gototheHISEventViewer.
2.VerifythatInboundADTandOrdereventsexist.
3.HavethecustomerverifythattheMUSEsystemisreceivingthedatatheyare sendingfromtheirinterface.
HISRealtimeResults/BillingOutboundCheckout
1.FromtheMUSEsystem,gototheHISEventViewer.
2.VerifythatOutboundResults/Billingeventsexist.
3.VerifythatoutbounddataisleavingtheMUSEresults/billingqueue.
HISBatchBilling
1.FromtheMUSEsystem,gotoSystem>Status.
2.ViewtheHL7BatchLogandverifythattheBatchwassentanditsucceeded.
OrderDownloadfromMACCartCheckout
1.FromaMACcart,usetheOrderDownloadfunction.
2.VerifythattheMACcartisabletosuccessfullydownloadtheorderfromthe MUSEsystem.
UsethefollowingprocedurestoverifythatimportXMLdatacorrectly.
1.CopyXMLdatalesfromthemanufacturer'sdeviceintotheXMLfolderset upontheMUSEserver.
2.Intheapplication,selectSystemStatus.
3.SelectNewlyAcquiredorAcquisitionLog.
4.VerifythattheXMLdatawasacquiredintotheMUSEsystem.
80
MUSE™CardiologyInformationSystem
2034539-043D
Page 81
Backup
Login
Maintenance
Usethefollowingprocedurestoverifythatthebackupproceduresworkcorrectly.
1.UseSQLServerManagementStudiotorunamanualMUSEbackup.
2.Conrmthatthebackupwassuccessful.
UsethefollowingprocedurestoverifythattheMUSEandWindowsAuthentication proceduresworkcorrectly.
MUSEAuthentication
1.HavetheuserlogontotheMUSEsystemusingMUSEAuthentication.
2.VerifytheuserisabletosuccessfullylogontotheMUSEsystem.
WindowsAuthentication
1.HavetheuserlogontotheMUSEsystemusingWindowsAuthentication.
2.VerifytheuserisabletosuccessfullylogontotheMUSEsystem.
ViaCitrixUsingMUSEAuthentication
1.HavetheuserlogontotheMUSEsystemusingMUSEAuthentication.
2.VerifytheuserisabletosuccessfullylogontotheMUSEsystem.
ViaCitrixUsingWindowsAuthentication
1.HavetheuserlogontotheMUSEsystemusingWindowsAuthentication.
2.VerifytheuserisabletosuccessfullylogontotheMUSEsystem.
Discarding/Recovering/DeletingData
Usethefollowingprocedurestoverifythattheusercandiscard,recoveranddeletea test.
DiscardingData
1.HavetheuserdiscardatestfromaMUSEsitedatabase.
2.VerifythetestwassuccessfullydiscardedandisdisplayedontheDiscarded DataListandasalogentryontheDiscardLog.
RecoveringData
1.HavetheuserrecoveratestfromtheDiscardedDataList.
2.Verifythatthetestwassuccessfullyrecoveredandisdisplayedonthe appropriateMUSEsitedatabaseandasalogentryontheDiscardLog.
DeletingData
1.HavetheuserdeleteatestfromtheDiscardedDataList.
2.Verifythetestwassuccessfullydeletedformthesystemandisdisplayedasa logentryontheDiscardLog.
2034539-043D
MUSE™CardiologyInformationSystem
81
Page 82
Maintenance
82
MUSE™CardiologyInformationSystem
2034539-043D
Page 83

HL7SystemStateBackupand Recovery

AbackupandrecoveryplanfortheHL7serveriscrucialtoensurethatyoucan recovertheHL7congurationshouldtheserverfailorlesbecomecorrupted.
CoordinaterecoveryoftheHL7SystemStatewiththeGEHealthcareHL7engineer.

CopyingtheSystemStateBackup

AfterinstallationandtestingoftheHL7interfaceiscomplete,theGEHealthcare HL7engineercreatesaweeklybackupjobthatsavestheHL7congurationinthe C:\SystemstatefolderontheHL7server.Itisthecustomer'sresponsibilitytocopy thisfoldertoanetworkdriveorexternalmediatoensurethatyoucanrestorethe HL7congurationsifyouneedtofullyrebuildtheserver.Alocalcopyofthefolder shouldremainontheserver.
6

DisasterRecovery

Thereareseveralreasonswhyyoumightneedtoperformadisasterrecovery.This proceduredescribesasituationwhereyouneedtorebuildtheserverandreturnitto itsoriginalmanufacturedstate.Usingthisrecoverymethoddeletesallinformationon thesystem'sharddrive.
Beforeperformingafulldisasterrecovery,considerothermethodsavailabletorecover thesystem,suchasstartinginSafeMode,rollingbacknewlyinstalleddrivers,using theLastKnownGoodCongurationoption,andotherconventionalWindowsrestore options.Formoreinformation,refertotheMicrosoftWebsite.

BeforeYouBegin

CheckifacopyoftheSystemStatefolderexistsonthenetworkorexternalmedia. ThisfolderiscreatedonceaweekandcontainsabackupoftheHL7conguration. Ifanexternalcopydoesnotexistandyouarestillabletologontotheserver,make acopyoftheSystemStatelelocatedintheC:\gehc-it\backup\datafolderbefore startingthedisasterrecoveryprocess.
Ifyouareabletologontotheserver,savetheInSiteExCcongurationasdescribedin theInSiteExCInstallationandCongurationGuide.
2034539-043D
MUSE™CardiologyInformationSystem
83
Page 84
HL7SystemStateBackupandRecovery

RebuildingtheHPDL360G5Server

ConguringtheServer
ThisprocedureisonlyfortheGEHealthcare-suppliedHPDL360G5server.Itrestores thesystemtoitsoriginalmanufacturedstate.
VerifythatyouhavethefollowingsoftwareCDs:
•2003801–012HPSmartStartCDv7.60
•2030821–001MUSEHISHP#325367–004
•2038348–001MUSEWin2003ServerSP2
1.StartthesystemusingtheSmartStartCDv7.60. Thistakesapproximatelythreeminutes.
2.PressF1tocontinue,ifapplicable.
3.AttheLanguageselectionwindow,selecttheEnglishoption.
4.OntheSelectthekeyboardtobeusedwiththesystemdrop-downlist,select EnglishandclickContinue.
5.ClickAgreetoaccepttheLicenseAgreement.
6.SelectMaintainServer.
7.SelectEraseSystem.
8.ConrmallboxesareselectedandclickContinue.
9.WhentheAreyousure…?windowopens,clickOK. Thesystempowersdownandthenrestarts.(Thistakesapproximatelythree
minutes.)
10.Duringstartup,whenprompted,pressF9fortheROM-basedSetupUtility.
11.SelectBootControllerOrderandmovePCIslot1HPSmartArrayControllerto thetopofthelist.
ThecontrollerisidentiedasE200i.
12.PressEnter.
NOTE:
Ifnecessary,pressEsctoreturntotheSelectionmenu.
13.SelectDateandTimeandpressEnter.
14.EnterthecurrentdateandtimeandpressEnter.
15.SelectServerAvailabilityandpressEnter.
16.SelectPOSTF1PromptandpressEnter.
17.SelectDisabledandpressEnter.
18.PressEsctoexitthewindow.
19.PressEsctoexittheutility.
20.PressF10toConrmExitUtility.
84
MUSE™CardiologyInformationSystem
2034539-043D
Page 85
ConguringtheRAID
1.RestartthesystemwithHPSmartStartCDv7.60intheopticaldrive.
2.AttheLanguageselectionwindow,selecttheEnglishoptionandclickContinue.
3.IntheSelectthekeyboardtobeusedwithsystemdrop-downlist,select
4.SelecttheAgreeoptiontoaccepttheLicenseAgreementandclickOK.
5.ClickMaintainServertobegincongurationofthearraycontrollers.
6.ClickCongureArray.
7.UnderCommonTasks,selectClearCongurations.
8.SelecttheSelectAllcheckboxandclickOK.
9.SelectCreateLogicalDrive.
10.ClickOKtousethedefaultsettings.
11.IntheCongurationViewpanel,conrmtheSCSIcontrollerhasonelogical
HL7SystemStateBackupandRecovery
EnglishandclickContinue.
TheCreateLogicalDrivewindowopens.
driveconguredtoRAID1+0.
12.ClickExitACU.
13.ClickSmartstartHome….
14.ClickDeployServer.
15.ClickContinue.
16.ClicktheMicrosoftWindowsfolder.
17.SelectMicrosoftWindowsServer2003,StandardEdition(HP-branded),and clickContinue.
18.IntheOperatingsystemmediasource,selecttheCD-ROMandFlatFiles options,andclickContinue.
NOTE:
IttakesseveralminuteswhileSmartStartpreparesthesystemforinstallation oftheoperatingsystem.
19.IntheFileSystemdrop-downbox,selectNTFS.
20.UnderthesectionSelectBootPartitionSize,verifythatMaximumisselected, andclickContinue.
21.IntheUserNameeld,type<HospitalName>.
22.IntheOrganizationeld,type<HospitalName>.
23.IntheLicenseTypeeld,selectPerSeatandclickContinue.
24.OntheSNMPCongurationwindow,acceptthedefaultentriesandclick Continue.
NOTE:
IttakesseveralminuteswhileSmartStartpreparesthesystemfor installationoftheoperatingsystem.
WhentheSmartStartCDejectsfromtheopticaldriveandyouarepromptedtoinsert theoperatingsystemmedia,proceedwith“InstallingWindowsServer2003”.
2034539-043D
MUSE™CardiologyInformationSystem
85
Page 86
HL7SystemStateBackupandRecovery

InstallingWindowsServer2003

1.OnthePleaseinserttheOperatingSystemmediaandclickContinuewindow, removetheSmartStartCDandinserttheWindows2003ServerCD(GEPN 2030821–001;HPPN325367–004)andclickContinue.
Thesystemrestartsafterthreetoveminutes.
2.WhentheCDdraweropens,removetheWindows2003ServerCD. Thesystemrestartsseveraltimesoverthreetoveminutes.
3.WhentheWelcometoWindowsSetupWizardwindowopens,clickNext. TheInstallingDeviceswindowopensforseveralminutes.
4.OntheRegionalandLanguageOptionswindow,clickNext.
5.OntheComputerNameandAdministratorPasswordwindow,followthe screeninstructionsandusethefollowinginformationwhereapplicable.Case sensitiveinputsarerequired.
FieldValues
ComputerName
Revisepercustomeroracceptthedefault
AdministratorAccount Password
ConrmPasswordEntersamepasswordenteredinthepreviouseld.
EntertheGEHealthcare-recommendedAdministrator password,orenterthecustomer-providedAdministrator password.
6.ClickNext.
7.SettheDate&TimeandTimeZonetoappropriatevaluesandclickNext. ThesystemdisplaysInstallingNetworkfor30seconds.
8.SelectNo,thiscomputer …orYes,makethiscomputer…perthecustomers' networkenvironment.
9.Inthetextbox,typeHIS_WG,orenterthedomainorworkgroupnameperthe customerorderandclickNext.
Thesysteminstallsdriversandprogramsforapproximately20minutesand thenrestarts.
10.OntheWelcometoWindowswindow,pressCTRL+ALT+DELandenterthe password.
11.OntheWindowsServerPost-SetupSecurityUpdateswindow,clickFinish.
12.OntheWhenyouclosethispage…window,selectYes.
13.OntheManageYourServerwindow,selecttheDon'tdisplaythispageat logoncheckbox.
14.ExittheManageyourserverwindow.
86
MUSE™CardiologyInformationSystem
2034539-043D
Page 87
ConguringDisplaySettings
1.ClickStart>ControlPanel>Display.
2.ClickontheSettingstabandconguretheSettingsinDisplayPropertiesas follows:
a.SettheScreenResolutionto1024x768pixels. b.SettheColorPalettetoHighest32bit. c.ClickAdvancedandselecttheMonitortab. d.SettheScreenrefreshrateto75HertzandclickOK. e.ClickApply. f.OntheTestingModedialogbox,clickYes. g.ClickOKtoexittheDisplayPropertieswindow.

SettingUptheWindowsEnvironment

HL7SystemStateBackupandRecovery
1.FromtheControlPanel,clicktheSystemicon. TheSystemPropertieswindowopens.
2.ClicktheAdvancedtab.
3.ClickStartupandRecoverySettings.
4.IntheTimetodisplaylistofoperatingsystemstextbox,enter5seconds.
5.intheWritedebugginginformationdrop-downbox,select(none).
6.ClickOKtoexittheStartupandRecoverywindow.
7.ClickOKtoexittheSystemPropertieswindow.

InstallingWindows2003Servicepack2

1.InserttheWin2003ServerSP2CDintotheserveropticaldrive. ThisCDshipswiththeMUSEv8CoreSoftwareKit.
2.ClicktheWindowsserver2003–KB914961–SP2–x86–enu.exele. TheServicePack2lesareextractedandtheWindows2003ServicePack
2installeropens.
3.FollowthepromptstoinstalltheServicePack.
4.Restarttheserverafterinstallationiscomplete.
Aftercompletingtherebuildprocessgoto“VerifyHelpandSupportService”on
page90.

RebuildingtheHPDL360G7Server

ThisprocedureisonlyfortheGEHealthcare-suppliedHPDL360G7server.Itrestores thesystemtoitsoriginalmanufacturedstate.
Verifythatyouhavethefollowingsoftware:
2034539-043D
MUSE™CardiologyInformationSystem
87
Page 88
HL7SystemStateBackupandRecovery
2062765-019DVDIMAGEDL360G7CCGSERVER2003
NOTE:
Beforestarting,disconnectthemousefromtheUSBport.Thisconnectioncan causeproblemswhenselectingthefunctionkeysusingthekeyboardduringthe rebuildprocess.

VerifytheBootSequence

1.Restarttheserver
2.Duringtherestart,whenpromptedtoPressanykeytoviewOptionROM messagespressF9Setup.
TheROM-basedSetupUtilityopens.
3.SelectStandardBootOrder(IPL)andpressEnter. TheStandardBootOrder(IPL)windowopens.
4.VerifythatCD-ROMislistedastherstbootsequence.Thisisnecessaryto ensurethatthesystemcanstartfromtheMUSEImageCD.Ifthebootsequence iscorrectskipthissectionandgoto“RAIDConguration”onpage88.IfCD-ROM isnotlistedastherstbootsequence,dothefollowing:
a.UsethearrowkeystoselectCD-ROMandpressEnter. b.SelectSettheIPlDeviceBootOrderto1andpressEnter.
TheCD-ROMmovestotherstpositioninthebootsequence.
c.PressEsctwicetoclosethemenuandtheROM-basedSetupUtility. d.PressF10toconrm.
Thesystemwillreboot.
RAIDConguration
1.Powerontheserver.
2.AllowtheservertoproceedthroughthePowerandThermalCalibrationin Progressscreen.
3.Whenprompted,pressanykeytoviewOptionROMmessages.
4.Whenprompted,pressF8toruntheOptionROMCongurationForArrays Utility.
5.SelectDeleteLogicalDriveandpressEntertoproceed. a.Iftherearenologicaldrivesavailable,proceedtostep6. b.PressF8todeletethelogicaldrive. c.PressF3toconrmthedeletionofthelogicaldrive. d.PressEntertocontinue.
6.SelectCreateLogicalDriveandpressEntertoproceed. TheArrayCongurationscreenopens,selectthefollowing: a.AvailablePhysicalDrives:Bothdrivesselected
b.RAIDCongurations:RAID1+0 c.ParityGroupCount:Noneselected
88
MUSE™CardiologyInformationSystem
2034539-043D
Page 89
d.Spare:Noneselected e.MaximumBootpartition:Disable(4GBmaximum)
7.PressEntertocreatealogicaldrive.
8.ReviewthecongurationandpressF8tosavetheconguration.
9.PressEntertocontinue.
10.SelectViewLogicalDriveandpressEnter.
11.Verifythatthenewlogicaldrivehasbeencreatedwiththefollowingsettings:
LogicalDrive#1,RAID1+0,600.09GB
12.PressESCtoreturntotheMainMenu.
13.PressEsctoexittheutility.
14.Ifthemousewasdisconnectedearlierintheprocess,reconnectitnow.

ApplyingtheServerImage

HL7SystemStateBackupandRecovery
Theimagewill:
•InstallWindows2003ServerR2Standardedition
•InstallWindows2003ServicePack2
•Setscreenresolutionto1280x1024pixels
•Assignapasswordtothebuilt-inadministratoraccount
Toimagetheserverusethefollowingsteps:
1.Powerontheserver.
2.Immediatelyplace2062765-019DVDIMAGEDL360G7CCGSERVER2003into theDVD-ROMdriveoftheserver.
3.AllowtheservertoboottheimageDVD.
4.Allowtheimagingutilitytostart.
5.Whenpromptedtocontinueimagethedrive,pressY.
6.TheGhostutilitywillapplytheimagetotheserverharddrives.
7.Allowtheimagingprocesstonishandverifythattheimagehasbeenapplied successfully.
8.Oncetheimagehasbeenappliedandveried,aSuccessmessagewillbe displayed.
9.RemovetheimageDVDandrestarttheserver.
10.AllowtheservertobootuntiltheWindowsSetupWizardopens.
11.OntheWelcometotheWindowsSetupWizardscreendothefollowing: a.ClickNext.
TheLicenseAgreementscreenopens. b.SelectIacceptthisagreement. c.ClickNext.
2034539-043D
MUSE™CardiologyInformationSystem
89
Page 90
HL7SystemStateBackupandRecovery
12.OnDateandTimeSettingsscreensetthefollowing: a.Date b.Time c.TimeZone d.Automaticallyadjustclocktodaylightsavingschanges
13.ClickNext.
14.Windowswillnishthecongurationandreboottheserver.
15.Verifyyoucanlogontheoperatingsystemusingthebuilt-inadministrators account,IfyoudonotremembertheadministratorpasswordcontactGE HealthcareTechnicalSupport.
Proceedto“VerifyHelpandSupportService”onpage90.

VerifyHelpandSupportService

MicrosoftreportedcasesofHelpandSupportnotrunningafterWindows2003SP2 wasinstalled.Verify,andifnecessary,installHelpandSupportasfollows.
ConguringSNMP
1.SelectStart>Helpandsupport.
2.Ifthefollowingmessageisdisplayedgotostep3.IfHelpandSupportopens ignorethissectionandproceedto“ConguringSNMP”onpage90.
WindowscannotopenHelpandSupportbecauseasystemserviceisnot running.
3.Toxthisproblem,starttheservicenamedHelpandSupportbyopeninga commandprompt
a.Typecd\Windows\PCHealth\HelpCtr\Binaries b.Typecommand:start/whelpsvc/svchostnetsvcs
/regserver/install
c.Whenthecommandcompletes,theServicesshoulddisplaytheHelp
andSupportservice.
4.StarttheHelpandSupportservice.
Proceedto“ConguringSNMP”onpage90.
SNMPisenabledbydefault.Somecustomersmaywantitdisabledinordertoreduce securityrisksinthesystem.TodisableSNMPusethefollowingsteps.
1.SelectStart>AllPrograms>AdministrativeTools>Services.
2.Intherightpane,openSNMPService.
3.SelecttheGeneraltab.
4.AtStartuptype,clickDisabled.
5.AtServicestatus,clickStop,ifitisnotgrayedout.
6.OntheStopOtherServiceswindow,clickYes.
90
MUSE™CardiologyInformationSystem
2034539-043D
Page 91
7.AftertheStoppingServiceswindowcloses,clickOKtoclosetheSNMPService Propertieswindow.
8.ClosetheServiceswindow.
Aftertheserverhasbeenrestored,completetherecoveryprocessbyInstallingCCG, InstallingInSiteExC,andRestoringtheSystemStateasdescribedinthenextsections.

InstallingCCG

FollowtheinstructionsintheMUSECardiologyInformationSystemInstallation ManualtoinstallandverifyCCGontheserver.

InstallingInSiteExC

InstallInSiteExContheserverusingtheInSiteExCInstallationandConguration Guide.Congureitusingthesameserialnumberthatwasoriginallyusedtoregister
theserverwiththebackofce.IfyouwereabletosavetheInSitecongurationbefore rebuildingtheserver,youcanusetheInSiteRestoreCongurationfunctiontorestore theoriginalcongurationvalues.
HL7SystemStateBackupandRecovery

RestoringSystemState

Aftertheserverisrestarted,contacttheGEHealthcareHL7engineertorestorethe HL7congurationfromtheSystemStatebackup.
2034539-043D
MUSE™CardiologyInformationSystem
91
Page 92
HL7SystemStateBackupandRecovery
92
MUSE™CardiologyInformationSystem
2034539-043D
Page 93

MUSESystemBackupandRecovery

Introduction

Abackupandrecoveryplaniscrucialtopreventdatalossandtominimizeservice interruptionintheeventofsystemfailureordisaster.Thischapterdescribesthe backupandrecoveryoptionsprovidedbyGEHealthcare.Theseoptionsareavailable onlytousersrunningtheMUSEsystemonGE-suppliedhardware.Usersrunningthe software-onlyversionoftheMUSEsystem,orwhochoosenottousetheGE-supplied options,areresponsibleforimplementingandmaintainingtheirownbackupand recoveryplans.
GEHealthcareprovidestwobackupoptions.
•AnetworkbackupoptionisavailableforusersrunningMUSEsystemonGE-supplied
G5orG6platforms.
7
•AtapebackupoptionisavailableforusersrunningtheMUSEsystemonGE-supplied
G4orG5platformsequippedwiththeAITTapeDrive. Referto“BackupOptions”onpage94formoreinformation.
Afterdeterminingwhichbackupoptionyouwilluse,referto“SettingUpAutomatic
Backups”onpage97forinstructionsonhowtocreatethebackupjobs,howto
congurethebackupjobs,howtochangethebackupschedules,howtocreate automaticnotications,andhowtotestthebackupjobs.
Inadditiontorunningautomaticbackupjobsatregularlyscheduledtimes,youcan manuallyinitiatethebackupjobstorunoutsideoftheirnormalschedules.Usethis featureifyouneedtogenerateaspecialbackup,forexample,beforeupgradingor performingserviceonthesystem.Referto“BackingUptheDatabaseManually”on
page109formoreinformation.
Intheeventthatthedatabasebecomescorrupted,orifyouwanttoreturnthe databasetoapreviousstateforanyreason,youcanrestorejusttheMUSEdatabase. Fordetails,referto“DatabaseRecovery”onpage109.
NOTE:
Intheeventofasystem-widefailureornaturaldisaster,orifyouwanttoreturn thesystemtoapreviousstateforanyreason,youcanrestoretheentireMUSE system.Fordetails,refertoAppendixA“SystemRecovery”onpage117.
Duringthebackupandrecoveryprocess,itmaybenecessarytoshutdownorrestart theMUSEsystemortoinitializeanewbackuptape.Fordetails,referto“Additional
Information”onpage115.
2034539-043D
MUSE™CardiologyInformationSystem
93
Page 94
MUSESystemBackupandRecovery

BackupOptions

ThissectionprovidesthefollowinginformationabouttheMUSEbackupoptions:
•OverviewoftheMUSEdiskanddatastructures.
•Overviewofthenetworkandtapebackupoptions.

OverviewoftheMUSEDiskandDataStructures

KnowledgeoftheMUSEdiskanddatastructureshelpsyouunderstandthebackup optionsprovidedbyGEHealthcare.Ifyoupurchasedasoftware-onlysystem,orif youchoosenottousetheGE-suppliedbackupoptions,thisinformationisusefulin planningandimplementingyourownbackup/recoveryplan.
MUSEserversrunningonGE-suppliedhardwareareconguredwithtwopartitions,as showninthefollowingdiagram:
TheC:driveisfairlystaticandtypicallyneedstobebackeduponlyforoperating systemupdates,softwareupgrades,orcongurationchanges.Itcontainsthe followingcomponents:
•TheWindowsoperatingsystem
•SQLServersoftware-includingthemasterandmsdbsystemdatabases
•TheMUSEclientapplication
•GERemoteAccessFiles(InSiteExC)
TheD:driveisdynamicandtypicallyneedstobebackeduponaregularbasis.It containsthefollowingcomponents:
•TheMUSEuserdatabases(d:\muse\db)
•Thebackupreportandsupportingles(d:\muse\backup\)
•ThetemporarylesbeingsenttotheMARSformatterwhenHolterreportsare
printed(d:\muse\mars).
•TheXMLinputles(d:\muse\xml).
•Theacquisitiondatasentthroughanetworkshare(d:\muse\acq).
TheMUSEsystem,version7.0.2andabove,usestheSQLServer2005Database ManagementSystem(DBMS)andincludesthefollowingdefaultuserdatabases:
MUSE_Site0001
MUSE_SiteTemplate
MUSE_System
Ifadditionalsiteshavebeensetuponthesystem,additionalsitedatabaseswillexist; forexample,MUSE_Site0002,MUSE_Site0003,andsoon.
94
MUSE™CardiologyInformationSystem
2034539-043D
Page 95

OverviewoftheNetworkandTapeBackupOptions

ForcustomersrunningonGEHealthcare-suppliedhardware,GEHealthcareprovides anetworkoptionandatapeoptionforbackinguptheMUSEsystemanddatabases. Yourhardwareplatformdetermineswhichoptionsareavailable,asshowninthe followingtable.
PlatformTapeBackupNetworkBackup
HPML370G4 HPML370G5 HPDL370G6
CAUTION:
LOSSOFDATA/DATACORRUPTION:Thebackupproceduresinthischapterwere testedonlyontheGE-suppliedhardwareoutlinedintheprevioustable.GE Healthcarecannotguaranteedatabaseorsystemrecoveryfollowingthese proceduresoncustomer-suppliedhardware.
Customersimplementingthenetworkbackupmodeldescribedinthisdocument ontheirownhardwareareresponsiblefortesting,verifying,andmaintaining theprocedure.
x xx
MUSESystemBackupandRecovery
x
GEServicesupportsboththetapeandnetworkbackupoptions,ascoveredbyany applicablewarrantiesorserviceagreements.GEHealthcareServicealsoassistswith theinitialsetupandcongurationoftheselectedoption.
TheMUSEbackupoptionsaresetfortheSimpleRecoveryModel.Customerswho requireaFullRecoveryModel(sothattheyhavetransactionlogsthatprovidethe abilitytorecovertoaspecicpointintime)needtoimplementtheirownbackupand recoveryplans.Customer-implementedbackupplansarenotcoveredbyapplicable warrantiesorserviceagreements;however,GEHealthcareServicecanstillassistin thesecasesonatime-and-materialbasis.
NOTE:
EventhoughGEHealthcareServicemayinstallandcongurethebackupoptions, thecustomerisresponsibleforverifyingthateachbackupcompletedsuccessfully. YoucannotusethesuccessorfailurestatusoftheSQLbackupjobitselfasproofof thejob’ssuccess;insomecircumstances,theSQLjobcanindicateafalsesuccess.
Youhavetwomethodsforverifyingthesuccessofabackupjob:
•Reviewthebackuploglegeneratedatthecompletionofeachbackup. Theloglecanbeemailedtoaselecteduserforreview.Referto“Creating
Notications”onpage103fordetails.
•Verifythedatabasebackuplesandtheirdates. Referto“NetworkBackupOption”onpage95foralistofthebackuples createdonthenetworkdrive.
NetworkBackupOption
2034539-043D
ThenetworkbackupoptionisavailabletousersrunningMUSEv7.1.1orhigheronGE Healthcare-suppliedG5orG6platforms.Itprovidesbothnightlyandweeklybackup jobs.
MUSE™CardiologyInformationSystem
95
Page 96
MUSESystemBackupandRecovery
NightlyBackup
Thenightlybackupjob,7Nightly_Backups,doesthefollowing:
Createsadailybackupleforeachdatabase. Thelesarenamedbycombiningthedayoftheweekwiththenameofthe database,asseeninthefollowingexamples:
•MondayMaster.bak
•Mondaymsdb.bak
•MondayMUSE_Site0001.bak
•MondaySiteTemplate.bak
•MondaySystem.bak
Thisresultsinsevensetsofbackuples.Eachleoverwritesthecorrespondingle fromthepreviousweek.
PerformsafullbackupofalltheMUSEdatabases. Theadvantageofafullbackupisthatyoucanrestorethedatabasesusingasingle backupinsteadofaseriesofdifferentialorincrementalbackups.Thedisadvantage ofafullbackupisthatitrequiresmorestoragespacethanadifferentialor incrementalbackup.Becausethebackupjobcreatessevensetsofbackups(one foreachdayoftheweek),thedestinationdriveforthebackupsmustbeatleast seventimesgreaterthanthecombinedsizeofthedatabases.Forexample,ifthe MUSEdatabasestotalled20GB,thedestinationdrivewouldneedtobeatleast140 GB(7x20GB)toholdafullseven-daybackupcycle.
Savesthebackuptoalocalfolderandthencopiesittoanetworkshare. Youcaneditthescriptle,7Nightly_Backups_NetCopyScript.bat,tocustomize thisbehaviorintwoways:
•First,youcanmodifythewayinwhichthebackuplesarecopiedtothenetwork
drivebysettingtheCOPYOPTIONparametertooneofthefollowingvalues:
MIRROR-Copiesthenightlydatabasebackupsfromthelocaldrivetothe networkshare,leavingtheoriginalsonthelocaldrive.Thisprovidesafullset ofbackupsonboththelocaldriveandthenetworkshare.
MIRRORSOME-Copiesadenednumberofbackupsetsfromthelocaldrive tothenetworkshare,leavingthedenednumberoforiginalsonthelocal driveandremovingtheremainder.Thisprovidesapartialsetofbackupson thelocaldriveandafullsetofbackupsonthenetworkshare. Todenethenumberofrecordstoretainonthelocaldrive,setthe LOCALDAYSTOMAINTAINparameterwithanumberfrom1to7.Forexample, settingLOCALDAYSTOMAINTAIN=3retainsthethreemostrecentnightly backupsonthelocaldrive.
MOVE-Copiesthenightlydatabasebackupsfromthelocaldrivetothe networkshareandremovestheoriginalsfromthelocaldrive.Thisprovides onlyonesetofbackupsonthenetworkshare.
•Second,youcanchangethelocationofthenetworksharewherethebackups willbecopied.
Attemptstoreconnectifthenetworkconnectionisinterrupted. Thebackupjobusestherobocopycommandtocopyormovethebackuplesfrom thelocaldirectorytothenetworkshare.Ifthenetworkconnectionisinterrupted duringthetransfer,robocopynoteswherethecopystopped,attemptstoreconnect tothenetwork,andresumesthetransfer.
96
MUSE™CardiologyInformationSystem
2034539-043D
Page 97
LocalTapeBackupOption
MUSESystemBackupandRecovery
WeeklyBackup
Theweeklybackupjob,WeeklyCDriveNetworkBackup,backsupthecompleteC: drive,includingthesystemstate,toasharednetworkdrive.Theweeklybackupjob savesthelesdirectlytothenetworkdrive;unlikethenightlybackupjob,itdoesnot rstsavetheleslocallyandthencopythemtothenetworkdrive.
Bydefault,theweeklybackupjobsavesthebackuple,CDrivebackup.bkf,tothe locald:\muse\backupfolder.Tochangethedestinationtoanetworkshare,editthe locationintheweeklybackupbatchle,Cbackup_NO_TAPE.bat.Forinstructions, referto“ConguringtheBackupJobs”onpage99.Thebackupleisoverwritteneach timetheweeklybackupruns.
Unliketheweeklytapebackup(see“LocalTapeBackupOption”onpage97),the weeklynetworkbackupisnotintendedforuseaspartofanAutomatedSystem Recovery(ASR).Instead,whenusingtheweeklynetworkbackup,recoveryinvolves theusingtheserverimageCDandweeklybackuple.Fordetails,refertoAppendixA
“SystemRecovery”onpage117.
ThelocaltapebackupoptionisavailableonlyonGEHealthcare-suppliedG4servers, forwhichatapedrivewassuppliedasstandardequipment,andG5servers,forwhich atapedrivewasavailableasoptionalequipment.Thetapedriveisnolongeravailable forpurchase.AnyG5serverthatdoesnotalreadyhaveatapedrivemustusethe networkbackupoption“NetworkBackupOption”onpage95).
ThelocaltapebackupoptionincludesnineAITtapesandprovidesanightlybackup,a weeklybackup,andamonthlybackup.
NightlyBackup
ThenightlybackupcopieseachoftheMUSEuserdatabasestoanAITtape.This backupjobrequiressevenofthenineAITtapes,eachlabeledwithadayoftheweek fromSundayBackupthroughSaturdayBackup.
WeeklyBackup
TheweeklybackupcopiestheentireC:drivetoanAITtape.Thisbackupjobrequires oneofthenineAITtapes,labeledWeeklyC:Backup.
MonthlyBackup
ThemonthlybackupcopiestheentireMUSEsystemtoanAITtape.Thisbackupjob requiresoneofthenineAITtapes,labeledMonthlySystemBackup.

SettingUpAutomaticBackups

Thissectionprovidesthefollowinginformationforcreatingandconguringautomatic backups:
•Creatingthebackupjobs
•Conguringthebackupjobs
•Changingthebackupschedules
•Creatingnotications
•Testingthebackupjobs
2034539-043D
MUSE™CardiologyInformationSystem
97
Page 98
MUSESystemBackupandRecovery

CreatingBackupJobs

Theprocedureforinstallingthenetworkbackupjobsandtheprocedureforinstalling thetapebackupjobsarenearlyidentical.Theonlydifferencebetweenthemisthe nameandlocationofthelestoinstall.Thatdifferenceisnotedwhereapplicable.
NOTE:
Ifyouarecreatingbackupjobsonanexistingsystem,dothefollowingbefore proceeding:
•Deleteanybackupleslocatedinthed:\muse\backupfolder.
•DeleteanySQLbackupjobsinSQLServerManagementStudio.
Tocreatethebackupjobs,dothefollowing:
1.Createthed:\muse\backupfolderifitdoesnotalreadyexist.
2.InserttheMUSEsupportCDintotheopticaldrive.
3.Copyallthelesfrom\LocalTapeBackuponthesupportCDtod:\muse\backup
ontheMUSEserver. TheSQLbackupjobsutilizebatchlesstoredinthed:\muse\backupfolder.If
thebackupfolderiscreatedonadifferentdrivepartition,orusesadifferent name,thebackupjobsandemailnoticationswillfail.
4.VerifythattheSQLServerAgentserviceisrunningusingtheMUSEBkgnd
account. TheSQLServerAgentservicerunsSQLjobs,suchasthebackupjobs.Itisalso
requiredfordatabasesearchesandthelogandqueuemaintenancejobs.Use thefollowingproceduretoverifytheserviceisrunning:
a.SelectStart>AdministrativeTools>ServicestodisplaytheServiceslist. b.Right-clickSQLServerAgentandselectProperties. c.VerifythatthestartuptypeissettoAutomaticandchangeitifnecessary. d.StarttheSQLServerAgentserviceifitisnotalreadyrunning. e.ClosetheServiceswindow.
5.Dooneofthefollowingtobuildtheappropriatebackupjobs:
Tobuildthenetworkbackupjobs: a.OpenaCommandPromptwindow. b.Rund:\muse\backup\build_7Day_job.bat.
Thiscreatesthefollowingbackupjobsandschedules
98
MUSE™CardiologyInformationSystem
2034539-043D
Page 99
MUSESystemBackupandRecovery
YoureceivethemessageWarning:Non-existentstepreferencedby @on_fail_step_id.Thiswarningisnormalandyoucanignoreit.
c.IntheSQLServerManagementStudio,expandSQLServerAgent>Jobs
andverifythetwobackupjobswerecreated.
Tobuildthelocaltapebackupjobs: a.OpenaCommandPromptwindow. b.Rund:\muse\backup\buildjobs.bat.
Thiscreatesthefollowingbackupjobsandschedules:
ItalsoaddsabackupdeviceunderServerObjects>BackupDevices.If thedevicealreadyexists,youreceiveanerrormessagethatthedevice alreadyexists.Youcanignorethiserror.
c.IntheSQLServerManagementStudio,expandSQLServerAgent>Jobs
andverifythebackupjobswerecreated.
NOTE:
Ifthebackupjobsalreadyexist,thebatchlesdonotoverwritethem.This preventsyoufromaccidentallyremovinganycustomcongurations.To replacetheexistingjobswiththedefaults,runSQLServerManagement Studio,openSQLServerAgent>Jobs,anddeletetheexistingjobsbefore yourunthebatchles.
ConguringtheBackupJobs
Afterthebackupjobsareinstalled(“SettingUpAutomaticBackups”onpage97), congurethemtorunatatimeandinamannerthatmeetsyouroperationalneeds.
ConguringtheNetworkBackupJobs
Beforeconguringthenetworkbackupjobs,gatherthefollowinginformation:
•TheCOPYOPTIONparametertouse. Referto“NightlyBackup”onpage96fordetails.
•ThesizesoftheMUSEdatabaseandC:drive.
2034539-043D
MUSE™CardiologyInformationSystem
99
Page 100
MUSESystemBackupandRecovery
ThesizeoftheMUSEdatabaseandC:drivedeterminetherequiredamountoffree spaceonthedestinationdrive.Forexample:
Allowadditionalspacetoaccommodateanticipatedgrowth.
NOTE:
Becausethebackupprocedurerstbacksuptothelocaldriveandthenmoves thebackeduplestothenetworkshare,thelocaldriveneedsenoughfreespace toaccommodateatleastonebackup.Forexample,iftheMUSEdatabasestotal 20GB,thelocaldrivemusthaveatleast20GBoffreespaceorthebackupfails.
•Thelocationofthedestinationnetworkshare. Youneedthisinformationtoeditthe7Nightly_Backups_NetCopyScript.batand Cbackup_NO_TAPE.batles.Thepathtothenetworksharecannotcontainspaces; ifitcontainsspaces,thebackupfails.
NOTE:
Afteryouhaveidentiedthelocationofthedestinationnetworkshare,dothe following:
a.CreateafoldercalledMUSEbackupsonthesharedrive. b.Shareitonthenetwork. c.GrantFullControltotheMUSEBkgndaccount.
ThisshouldbethesameaccountastheSQLServerAgentserviceaccount logonontheMUSEserver.
ConguringtheNightlyNetworkBackup
1.Createthefolderd:\muse\musedbbackups.
2.UsinganASCIIeditorsuchasNotepad,opend:\MUSE\Backup\
7Nightly_Backups_NetCopyScript.bat.
3.SetCOPYOPTIONtoMIRROR,MIRRORSOME,orMOVE.
Forexample:SETCOPYOPTION=MOVEReferto“NightlyBackup”onpage96 fordetails.
4.IfyousettheCOPYOPTIONtoMIRRORSOME,settheLOCALDAYSTOMAINTAIN
parametertothenumberofbackupstokeeponthelocaldrive. Forexample:SETLOCALDAYSTOMAINTAIN=3retainsthethreemostrecent
backupsonthelocaldrive.
5.SettheNETWORKBACKUPDIRtotheUNCpathofthenetworksharewherethe
backupswillbecopied. Forexample:SETNETWORKBACKUPDIR=\\metropolis\musebackups.
6.SettheMUSEBACKUPDIRtotheUNCpathofthelocaldrivewherethebackups
willbesaved. Forexample:SETMUSEBACKUPDIR=d:\muse\musedbbackups.
7.Savethele.
100
MUSE™CardiologyInformationSystem
2034539-043D