GE moisture.iq User Manual

Page 1
moisture.IQ
User’s Manual
Moisture and Oxygen
bhge.com
910-298 Rev. E
September 2017
Page 2
Page 3
moisture.IQ
Multi-Channel Moisture and Oxygen Analyzer
User’s Manual
910-298 Rev. E September 2017
bhge.com
© 2017 Baker Hughes, a GE company – All rights reserved.
Page 4
[no content intended for this page]
Page 5
Contents
Product Registration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Terms and Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Typographical Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Safety Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Auxiliary Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
Environmental Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ix
Chapter 1. Installation
1.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Unpacking the moisture.IQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.3 Installing the moisture.IQ. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.4 Choosing an Installation Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.4.1 General Site Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.4.2 Low Voltage Directive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.5 Moisture Probe Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.5.1 Temperature Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.5.2 Moisture Condensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.5.3 Static or Dynamic Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.5.4 Pressure Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.5.5 Long-Term Storage & Operational Stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.5.6 Interference Effects. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.5.7 Corrosive Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.6 Sample System Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
1.6.1 Moisture Sample Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.6.2 Oxygen Sample Systems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.7 Mounting the Hygrometer System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.7.1 Mounting the Electronics Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.7.2 Mounting a Sample System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.7.3 Mounting the Oxygen Cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.8 Installing the Probes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.8.1 Moisture Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.8.2 Adapting an Existing TF-Series Probe Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
1.8.3 Adapting an Existing M-Series Probe Cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
1.8.4 Pressure Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
1.8.5 Delta F Oxygen Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
1.9 Making Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
1.9.1 Using the Lever to Make Wiring Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
1.9.2 Connecting the Input Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
1.9.3 Connecting Moisture Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
1.9.4 Connecting the Delta F Oxygen Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
moisture.IQ User’s Manual iii
Page 6
Contents
1.10 Establishing a Gas Flow Through the Oxygen Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
1.11 Connecting an Auxiliary Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
1.12 Connecting Analog Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
1.13 Connecting Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Chapter 2. Operation
2.1 Powering Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
2.2 Measurement Display Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
2.3 The Help Button. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
2.4 The Data Element Editor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
2.5 Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
2.6 Restarting the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
2.7 Shutting Down the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Chapter 3. Using the Settings Menu
3.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
3.2 Display Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
3.3 System Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
3.4 File Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
3.5 Screen Alignment (Weatherproof and Explosion-Proof) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
3.6 Screen Configuration (Weatherproof and Explosion-Proof). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
3.7 Notifications Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
3.8 Fault Alarm Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
3.9 Module Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
3.10 Service Menu Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
3.10.1 Software Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
3.10.2 Restart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
3.10.3 Shut Down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Chapter 4. Using the Outputs, Alarms and Logger Menus
4.1 Setting Up the Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
4.1.1 Configuring an Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
4.1.2 Testing the Selected Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
4.1.3 Trimming the Selected Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
4.1.4 Setting the Output Range Error Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
4.2 Setting Up the Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
4.2.1 Configuring an Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
4.2.2 Testing the Selected Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
4.2.3 Setting the Alarm Range Error Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
4.3 Setting Up and Running Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
Chapter 5. Using the Configuration Menu
5.1 Configuring the Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
iv moisture.IQ User’s Manual
Page 7
Contents
5.1.1 The Probe Configuration Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
5.1.2 Configuring Delta F Oxygen Sensor Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
5.1.3 Calibrating Individual Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64
5.1.4 Setting the Probe Auto-Cal Schedule. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65
5.2 Calibrating the Probes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
5.2.1 Entering Calibration Data Automatically . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
5.2.2 Entering Calibration Data Manually . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
5.3 Tagging the Inputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
5.4 Entering User Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
5.4.1 Entering User Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
5.4.2 Entering User-Defined Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
5.4.3 Entering User Constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
5.4.4 Entering Saturation Constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
Chapter 6. Setting Up Communications
6.1 Setting Up moisture.IQ Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
6.2 Configuring the Serial Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
6.3 Configuring the Modbus Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
6.4 Connecting to an Ethernet LAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
6.4.1 Configuring the Ethernet TCP/IP Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
6.4.2 Capabilities of the Ethernet Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
6.5 Configuring a VNC Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .82
6.6 Configuring the Web Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
6.7 User Management Operations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
6.8 Setting Up a Remote Connection to a PC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
Chapter 7. Maintenance
7.1 Delta F Oxygen Cell Electrolyte . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87
7.1.1 Checking the Electrolyte Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .88
7.1.2 Replenishing the Electrolyte . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .88
7.2 Replacing and Recalibrating Moisture Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89
7.3 Calibrating the Delta F Oxygen Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89
7.3.1 Displaying Oxygen Content in PPMv and μA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89
7.3.2 Checking the Oxygen Cell Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90
7.4 Delta F Oxygen Cell Background Gas Correction Factors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
7.4.1 Correcting for Different Background Gases. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
7.4.2 Entering the Background Gas Correction Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92
Chapter 8. Troubleshooting
8.1 Screen Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .93
8.2 Common Problems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .96
moisture.IQ User’s Manual v
Page 8
Contents
Chapter 9. Specifications
9.1 Electronics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .99
9.2 Moisture Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
9.3 Temperature Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
9.4 Pressure Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
9.5 Oxygen Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
9.6 General Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
9.7 Output Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
9.8 Input Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
9.9 Probe Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Appendix A. Menu Maps
A.1 The Settings Menu Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
A.2 The Outputs, Alarms and Logger Menu Map. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
A.3 The Configuration > Probe & User Menu Maps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
A.4 The Configuration > Comms Menu Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Appendix B. Wiring Diagrams
B.1 Terminal Blocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
B.2 Rear View Channel Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
B.3 Power Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
B.4 RS-485 Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Appendix C. Updating the moisture.IQ Firmware
C.1 Updating the Chassis Firmware. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
C.2 Updating the moisture.IQ Module Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Appendix D. Modbus Register Map
vi moisture.IQ User’s Manual
Page 9
Preface
Product Registration
Thank you for purchasing your moisture.IQ from Baker Hughes, a GE Company. Please register your product at
www.gemeasurement.com/productregistration for product support such as the latest software/firmware
upgrades, product information and special promotions.
Services
BHGE provides customers with an experienced staff of customer support personnel ready to respond to technical inquiries, as well as other remote and on-site support needs. To complement our broad portfolio of industry-leading solutions, we offer several types of flexible and scalable support services including: Training, Product Repairs, Extended Warranties, Service Agreements and more. Please visit
www.gemeasurement.com/services for more details.
Terms and Conditions
Sales Terms and Conditions for your recent purchase of a BHGE product, including the applicable product Warranty, can be found on our website at the following link:
www.gemeasurement.com/sales-terms-and-conditions
Typographical Conventions
Note: These paragraphs provide information that provides a deeper understanding of the situation, but is not
essential to the proper completion of the instructions.
IMPORTANT: These paragraphs provide information that emphasizes instructions that are essential to proper
setup of the equipment. Failure to follow these instructions carefully may cause unreliable performance.
CAUTION! This symbol indicates a risk of potential minor personal injury and/or severe damage to the equipment, unless these instructions are followed carefully.
WARNING! This symbol indicates a risk of potential serious personal injury or death, unless these instructions are followed carefully.
Safety Issues
WARNING! It is the responsibility of the user to make sure all local, county, state and national codes,
regulations, rules and laws related to safety and safe operating conditions are met for each installation.
Attention European Customers! To meet CE Marking requirements for all units intended for use in the EU, all electrical cables must be installed as described in this manual.
moisture.IQ User’s Manual vii
Page 10
Preface
Auxiliary Equipment
Local Safety Standards
The user must make sure that he operates all auxiliary equipment in accordance with local codes, standards, regulations, or laws applicable to safety.
Working Area
WARNING! Auxiliary equipment may have both manual and automatic modes of operation. As
equipment can move suddenly and without warning, do not enter the work cell of this equipment during automatic operation, and do not enter the work envelope of this equipment during manual operation. If you do, serious injury can result.
WARNING! Make sure that power to the auxiliary equipment is turned OFF and locked out before you perform maintenance procedures on this equipment.
Qualification of Personnel
Make sure that all personnel have manufacturer-approved training applicable to the auxiliary equipment.
Personal Safety Equipment
Make sure that operators and maintenance personnel have all safety equipment applicable to the auxiliary equipment. Examples include safety glasses, protective headgear, safety shoes, etc.
Unauthorized Operation
Make sure that unauthorized personnel cannot gain access to the operation of the equipment.
viii moisture.IQ User’s Manual
Page 11
Preface
Environmental Compliance
RoHS
The moisture.IQ fully complies with RoHS regulations (Directive 2011/65/EU).
Waste Electrical and Electronic Equipment (WEEE) Directive
BHGE is an active participant in Europe’s Waste Electrical and Electronic Equipment (WEEE) take-back initiative (Directive 2012/19/EU).
The equipment that you bought has required the extraction and use of natural resources for its production. It may contain hazardous substances that could impact health and the environment.
In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural resources, we encourage you to use the appropriate take-back systems. Those systems will reuse or recycle most of the materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please contact your local or regional waste administration.
Please visit more information about this initiative.
http://www.gemeasurement.com/environmental-health-safety-ehs for take-back instructions and
moisture.IQ User’s Manual ix
Page 12
Preface
[no content intended for this page]
x moisture.IQ User’s Manual
Page 13

Chapter 1. Installation

Chapter 1. Installation

1.1 Introduction

Users typically install the moisture.IQ as part of a complex process system, which includes components such as filters, pumps, and pressure regulators. In such an environment, probes and other parts of the system may be subjected to environmental hazards, such as high temperature, pressure extremes, corrosive elements and mechanical vibrations.
This section contains information and instructions for installing the moisture.IQ into a process system, while taking into account all of the above factors. The following section describes how to set up and connect the moisture.IQ.
If you have questions about the installation procedures, contact our technical support department. See the rear cover of this manual for contact information.

1.2 Unpacking the moisture.IQ

Before beginning the installation, unpack the unit and make sure all the parts and documentation listed on the packing slip are included. Be sure to inspect each piece of equipment, including the sample system, for evidence of mishandling. If anything has been damaged or is missing, report this to the carrier and to BHGE immediately.

1.3 Installing the moisture.IQ

To install the moisture.IQ, refer to the drawing for your specific configuration (benchtop, rack mount, panel mount, weatherproof or explosion-proof) in the Certification and Safety Statements at the end of this manual.
WARNING! To ensure the safe operation of this unit, you must install and operate the moisture.IQ as described in this user’s manual. In addition, be sure to follow all applicable safety codes and regulations for installing electrical equipment in your area.
WARNING! For Zone 2 Stainless Steel: Do not rub front face with a dry cloth within the hazardous area as a spark may occur.
moisture.IQ User’s Manual 1
Page 14
Chapter 1. Installation

1.4 Choosing an Installation Site

You should have discussed environmental and installation factors with a BHGE applications engineer or field sales person by the time you receive the meter. The equipment provided should be suited to your application and installation site.
The moisture.IQ is available in rack-mount, bench top, or panel-mount versions that are suitable for most indoor installations. BHGE also provides weatherproof and explosion-proof housings for outdoor and hazardous area locations. See the drawings in the Certification and Safety Statements at the end of this manual for details of each enclosure.
Before installing the unit, read the guidelines below to verify that you have selected the best installation site.

1.4.1 General Site Guidelines

Choose an installation site for the probes and sample systems that is as close to the process line as
possible. Avoid long lengths of connecting tubing. If long distances are unavoidable, a fast sampling by-pass loop is recommended. Do not install any other components, such as filters, upstream from the probes or sample system unless instructed to do so by BHGE.
Observe all normal safety precautions. Use the probes within their maximum pressure and
temperature ratings.
Although the moisture.IQ may not need to be accessed during normal operation, install the electronics
unit at a convenient location for programming, testing and servicing. A control room or instrument shed are typical locations.
Locate the electronics unit away from high temperatures, strong electrical transients, mechanical
vibrations, corrosive atmospheres, and any other conditions that could damage or interfere with the meter operation. See Chapter 9. “Specifications” on page 99 for environmental limits.
Protect the probe cables from excessive physical strain (e.g., bending, pulling, twisting, etc.).
Observe the proper cable restrictions for the probes. You can locate Moisture Image Series probes up to
915 m (3000 ft) from the electronics unit with unshielded twisted-pair cable. M-Series probes can be located up to 600 m (2000 ft) from the unit with the use of specially shielded cable.

1.4.2 Low Voltage Directive

If the power mains switch on the unit remains accessible to the user after installation a disconnect device is not required. However, if installation blocks access to the power mains switch, compliance with the EU Low Voltage Directive (IEC 61010) requires an external power disconnect device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly visible, directly accessible, and located within
1.8 m (6 ft) of the unit.
2 moisture.IQ User’s Manual
Page 15
Chapter 1. Installation

1.5 Moisture Probe Considerations

The M-Series and Moisture Image Series probes consist of an aluminum oxide sensor mounted on a connector head. Standard probe mounts include a protective stainless steel shield.
The probe sensor materials and housing maximize durability and insure a minimum of water absorbing surfaces in the vicinity of the aluminum oxide surface. A sintered stainless steel shield is used to protect the sensor from high flow rates and particulate matter. The end cap should not be removed except upon advice from BHGE.
The sensor has been designed to withstand normal shock and vibration. You should make sure that the active sensor surface is never touched or allowed to come into direct contact with foreign objects, since this may adversely affect performance.
Observing these few simple precautions will result in a long and useful probe life. BHGE recommends that probe calibration be checked routinely, at 12-month intervals, or as recommended by our applications engineers for your particular application.
The probe will measure the water vapor pressure in its immediate vicinity. Therefore, readings will be influenced by its proximity to the system walls, materials of construction, and other environmental factors.The sensor can be operated under vacuum or pressure and under flowing or static conditions.
See the following sections for the recommended environmental precautions.

1.5.1 Temperature Range

See “Probe Specifications” on page 105, for details.

1.5.2 Moisture Condensation

Be sure the temperature is at least 10°C higher than the dew/frost point temperature. If this condition is not maintained, moisture condensation could occur on the sensor or in the sample system, which will cause reading errors. If this happens, dry out the probe following the procedures outlined in BHGE document #916-064, Basic Hygrometry Principles.

1.5.3 Static or Dynamic Use

The sensor performs equally well in still air or where considerable flow occurs. Its small size makes it ideal for measuring moisture conditions within completely sealed containers or dry boxes. It will also perform well under gas flow conditions as high as 10,000 cm/sec and liquid flow conditions up to 10 cm/sec. Refer to BHGE document #916-064, Basic Hygrometry Principles, for maximum flow rates in gases and liquids.
moisture.IQ User’s Manual 3
Page 16
Chapter 1. Installation

1.5.4 Pressure Range

The moisture probe always senses the correct water vapor pressure regardless of the total ambient pressure. The moisture sensor measures water vapor under vacuum or high pressure conditions from as little as a few microns of Hg to as high as 5000 psi total pressure.

1.5.5 Long-Term Storage & Operational Stability

Sensors are not affected by continuous abrupt humidity changes or damaged by exposure to saturation conditions even when stored. However, you should store probes in their original shipping containers in a clean, dry location. If the probe is saturated during storage, refer to “Moisture Condensation” on page 3 before installing the probe. For best performance, do not store probes longer than one to two years from their calibration date.

1.5.6 Interference Effects

The sensor is completely unaffected by the presence of a wide variety of gases or organic liquids. Large concentrations of hydrocarbon gases, Freon effect on sensor water vapor indications. The sensor will operate properly in a multitude of gaseous or non-conductive liquid environments.
, ozone, carbon dioxide, carbon monoxide, and hydrogen have no

1.5.7 Corrosive Materials

Avoid all materials that are corrosive or otherwise damaging to aluminum or aluminum oxide. These include strongly acidic or basic materials and primary amines.
4 moisture.IQ User’s Manual
Page 17
Chapter 1. Installation

1.6 Sample System Guidelines

A sample system is required for oxygen measurement and, although not mandatory, is highly recommended for moisture measurement. The purpose of a sample system is to condition or control a sample stream to within the specifications of a probe. The application requirements determine the design of the sample system. BHGE applications engineers will make recommendations based on the following general guidelines.

1.6.1 Moisture Sample Systems

Typically, sample systems should be kept very simple. They should contain as few components as possible and all or most of those components should be located downstream of the measurement probe. Figure 1 below shows an example of a basic sample system consisting of an explosion-proof housing with a sample cell, a filter, a flow meter, a vent valve, an inlet shut-off valve and an outlet shut-off valve.
The sample system components should not be made of any material that will affect measurements. A sample system may include a filter to remove particulates from the sample stream or a pressure regulator to reduce or control the pressure of the stream. However, most common filters and pressure regulators are not suitable for sample systems because they have wetted parts that may absorb or release components (such as moisture) into the sample stream. They may also allow ambient contamination to enter the sample stream. In general, you should use stainless steel construction for all wetted parts. Contact BHGE for further instructions.
Sample Ce ll
Sample Inle t
Vent
Figure 1: A Typical Moisture Sample System
Sample Outlet
Note: The actual sample system design is dependent on the application requirements.
moisture.IQ User’s Manual 5
Page 18
Chapter 1. Installation

1.6.2 Oxygen Sample Systems

Oxygen sample systems are required and can be ordered from BHGE for bench or wall mounting. You can also build your own sample system by using the following guidelines.
IMPORTANT: The BHGE warranty will be voided if the sample system does not have a relief valve.
The basic sample system requirements are as follows (see Figure 2 below):
The oxygen cell requires a sample gas flow of 0.5 to 1 LPM (1 to 2 SCFH).
The gas pressure in the sample cell must be between 0.0 and 1.0 psig. The pressure must not exceed
1.0 psig.
A 10 psig pressure relief valve installed upstream of the oxygen cell is required to prevent
over-pressure.
A flow meter is required to measure the flow.
A pressure gauge is required to measure the pressure.
A flow regulating or needle valve is required and should be located upstream of the sample cell.
A pressure regulator is required for sample gas supplies of 50 psig or greater.
If a sample pump is required to draw a sample to the oxygen cell, the pump should be installed downstream from the oxygen cell. You must also install a vacuum relief valve set at 1.0 psig between the oxygen cell and the pump.
Sam ple
Sam ple Inlet
Outlet
Figure 2: A Typical Oxygen Cell Sample System
Note: The actual sample system design is dependent on the application requirements.
6 moisture.IQ User’s Manual
Page 19
Chapter 1. Installation

1.7 Mounting the Hygrometer System

Mounting the hygrometer system consists of mounting the electronics unit, the probes, and the sample systems.

1.7.1 Mounting the Electronics Unit

Use the outline and dimension drawings in the Certification and Safety Statements at the end of this manual to mount the moisture.IQ. These drawings provide clearance and other mounting dimensions needed to prepare the site for mounting.
WARNING! If the power mains switch on the unit remains accessible to the user after installation, a disconnect device is not required. However, if the installation blocks access to the power mains switch, compliance with the EU's Low Voltage Directive (IEC 61010) requires an external power disconnect device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly visible, directly accessible, and located within 1.8 m (6 ft) of the unit.
Be sure to follow the guidelines outlined in “Choosing an Installation Site” on page 2 before mounting the enclosure.
Note: If the installation location does not provide enough room for connections to be made easily after
installation, you may want to make connections to the MIS or M-Series probes, Delta F Oxygen Cell, Outputs and Alarms before mounting the instrument.

1.7.2 Mounting a Sample System

BHGE sample systems are normally fastened to a metal plate that has four mounting holes. BHGE can also provide the sample system in an enclosure, if requested. In either case, fasten the sample system plate or enclosure with bolts in the mounting holes at each of the four corners. If you ordered sample system outline and dimension drawings, they will be included in your shipment.
Connect the sample system inlet to the process and the outlet to the return, using the appropriate fittings or an appropriate NPT adapter.
CAUTION! Do not start a flow through the sample system until all probes and transmitters are properly installed.
moisture.IQ User’s Manual 7
Page 20
Chapter 1. Installation

1.7.3 Mounting the Oxygen Cell

If your oxygen cell is not mounted in a sample system, refer to the Figure 3 below for mounting details.
5.90
(149.9)
TEMPERATURE SENSOR
MAX
MIN
CONN
.40
5.00
(127.0 MAX)
CONN
0.25 (6.4)
MAX
0.55
(14.0)
0.62
(15.8)
1.06
(26.9)
1.50
(38.1)
0.37
(9.4)
1.23
(31.2)
1.05
(26.7)
2.05
(52.1)
Electrolyte
Electrolyte
2.10
(53.3)
reservoir
level
window
Sample
gas
inlet
Sample
gas
outlet
MAX
MIN
SERVOMEX
2.80
(71.1)
4.91
(124.7)
Figure 3: Oxygen Cell Dimensions
8 moisture.IQ User’s Manual
Page 21
Chapter 1. Installation

1.8 Installing the Probes

After you mount the sample system, you must insert the moisture probes into the sample cells. In addition, you must check, prepare, and connect any installed oxygen cells to the gas line.

1.8.1 Moisture Probes

BHGE Moisture Image Series (MIS) and M-Series moisture probes have 3/4 inch-16 straight threads, with an O-ring seal, to the secure probes into the sample system or directly into the process line. Other fittings are available for special applications.
CAUTION! If mounting the moisture probes directly into the process line, you must consult BHGE for proper installation instructions and precautions.
Moisture probes are usually installed into a sample system. The sample system protects the probes from coming into contact with damaging elements in the process. Moisture probes are installed into a cylindrical shaped container, called the sample cell, which is included as part of your sample system. The sample cell is identified by a label on the sample system plate.
To install a moisture probe that has a 3/4 inch-16 straight thread, insert it into the sample cell so it is perpendicular to the sample inlet. Screw the probe into cell, making sure not to cross thread it, and tighten it securely. Figure 4 below shows a typical moisture probe installation, with the probe mounted in a BHGE sample cell. Install any moisture probes with different fittings in the appropriate manner.
Probe
Inlet
Sample Cell
Outlet
Figure 4: Typical Moisture Probe in Sample Cell
Note: Standard moisture probes have a sintered stainless steel shield that protects the aluminum oxide sensor.
Be sure to leave the shield in place for maximum protection.
It is important to eliminate all gas or liquid leaks for safety reasons and to ensure that measurements are not affected by ambient contamination. For gas applications, you should check for leaks using a soap bubble solution.
IMPORTANT: Refer to the Calibration Data Sheets to connect the probes to their corresponding channels. If a
probe is connected to the wrong channel, the meter will display incorrect data.
moisture.IQ User’s Manual 9
Page 22
Chapter 1. Installation

1.8.2 Adapting an Existing TF-Series Probe Cable

If an existing moisture measurement system uses a TF-Series moisture probe, the cable for that probe may be adapted for use with an MISP2 probe. To retrofit the existing system, refer to Figure 5 below and complete the following steps:
MISP2 Connector
[Total Cable Length = 6.50”]
TF Connector
Figure 5: TF-MISP2 Converter Cable
1. Locate the existing TF-Series probe and disconnect the cable from the probe.
2. Replace the TF-Series probe with an MISP2 probe.
3. Using BHGE adapter cable 704-1362-00, shown in Figure 5 above, connect one end of the cable to the
MISP2 probe and connect the other end to the existing TF-Series cable connector.
4. At the moisture analyzer end of the TF-Series cable, locate the same TF-Series cable and disconnect all flying leads from the pluggable connector.
5. Locate the BLACK and WHITE wires of the TF-Series cable and connect them to the moisture.IQ at the
MISP2 probe gray pluggable connector. See “Moisture Image Series Probes (MISP and MISP2)” on
page 19 for help.
6. Trim all other flying leads from the TF-Series cable flush with the jacket.
7. Plug the newly-wired MISP2 probe pluggable connector into the moisture.IQ.
10 moisture.IQ User’s Manual
Page 23
Chapter 1. Installation

1.8.3 Adapting an Existing M-Series Probe Cable

If an existing moisture measurement system uses a M-Series moisture probe, the cable for that probe may be adapted for use with an MISP2 probe. To retrofit the existing system, refer to Figure 6 below and complete the following steps:
MISP2 Connector
[Total Cable Length = 6.50”]
M Series Connector
Figure 6: M-MISP2 Converter Cable
1. Locate the existing M-Series probe and disconnect the cable from the probe.
2. Replace the M-Series probe with an MISP2 probe.
3. Using BHGE adapter cable 704-1649-00, connect one end of the cable to the MISP2 probe and connect
the other end to the existing M-Series cable connector.
4. At the moisture analyzer end of the M-Series cable, locate the same M-Series cable and disconnect all
flying leads from the pluggable connector.
5. Locate the GREEN and RED wires of the M-Series cable and connect them to the BLACK and WHITE wires
at the MISP2 probe gray pluggable connector. (The GREEN wire is connected to the MISP connector
"BLK" label and RED wire is connected to the MISP connector "WHT" label.) See “Moisture Image Series
Probes (MISP and MISP2)” on page 19 for help.
6. Trim all other flying leads from the M-Series cable flush with the jacket.
7. Plug the newly-wired MISP2 probe pluggable connector into the moisture.IQ.
moisture.IQ User’s Manual 11
Page 24
Chapter 1. Installation

1.8.4 Pressure Sensors

If a pressure input is required and if an optional pressure signal is not available from a Moisture Image Series probe, you can connect a separate pressure transmitter to an auxiliary input.
The moisture.IQ accepts any 0 to 20-mA, 4 to 20-mA, or 0 to 2-V pressure transmitters.
Always mount the pressure transmitter directly downstream from the moisture probe to ensure that the pressure is measured at the same point as the moisture measurement.

1.8.5 Delta F Oxygen Cell

Although the moisture.IQ accepts other oxygen devices as auxiliary inputs, it is designed to accept oxygen inputs directly from the Delta F Oxygen Cell. There are three steps for installing the Delta F Oxygen Cell:
1. Prepare the oxygen cell for operation
2. Calibrate the oxygen cell.
3. Connect the oxygen cell to the gas line.
The Delta F Oxygen Cell is available with various types of mounting enclosures. However, the cell itself will always look like the one shown in Figure 7 below.
Electrolyte Level
x
x
a
a
M
M
n
i
M
Figure 7: Delta F Oxygen Cell Electrolyte Level
12 moisture.IQ User’s Manual
Page 25
Chapter 1. Installation
1.8.5.1 Preparing the Oxygen Cell
To prepare the oxygen cell for operation, you must fill it with the electrolyte that was supplied in a plastic bottle.
WARNING! The electrolyte contains potassium hydroxide that is harmful if it comes in contact with eyes or skin. Consult your company safety personnel for proper procedures for handling the electrolyte.
1. Unscrew the top cover on the oxygen cell reservoir.
2. Slowly add the entire contents of the electrolyte bottle, approximately three ounces (90 ml), making
sure not to spill any on the outside of the cell. Be especially careful that the electrolyte does not come in contact with any of the cell electrical connections.
3. Using the min/max window on the oxygen cell, check the electrolyte level. The electrolyte should cover
about 60% of the window (see Figure 7 on page 12). The cell is now ready to be connected to the gas line.
4. Replace the top cover on the oxygen cell.
Note: After you add the electrolyte, DO NOT add additional electrolyte to the reservoir. If the level falls below the
minimum level, refer to “Delta F Oxygen Cell Electrolyte” on page 87 to replenish the cell.
5. Calibrate the oxygen cell as described in “Calibrating the Delta F Oxygen Cell” on page 89. After you
calibrate the oxygen cell, connect it to the gas line as described in the following section.
1.8.5.2 Connecting the Oxygen Sample System to the Gas Line
To connect the oxygen sample system to the process line, attach a 1/8 inch O. D. (outside diameter) tube to the 1/8 inch sample gas inlet fitting, using a Swagelokor equivalent mating connector. Avoid using plastic or rubber for any tubings or fittings which are included in the inlet gas lines.
WARNING! Do not connect the oxygen cell outlet to flow restricting devices, pressure lines, or vacuum lines. Pressure differentials across the oxygen cell sensor in excess of 1 psig could be harmful or damage the cell.
If the gas being monitored does not create a safety hazard, vent it to atmosphere at the sample system outlet. If venting the gas to atmosphere causes a safety hazard, vent the gas to a safe location. Make sure the venting system does not create a back pressure on the oxygen cell.
Note: The vented sample will not be corrosive if you install and operate the cell properly.
moisture.IQ User’s Manual 13
Page 26
Chapter 1. Installation

1.9 Making Electrical Connections

WARNING! To ensure the safe operation of this unit, you must install and operate the moisture.IQ as
described in this user’s manual. In addition, be sure to follow all applicable safety codes and regulations for installing electrical equipment in your area.
WARNING! Turn off the moisture.IQ before making any electrical connections.
WARNING! The stainless steel Weatherproof enclosure must be properly grounded.
To open a Weatherproof moisture.IQ unit, use a large, flat-bladed screwdriver to turn the latch on the enclosure cover.
To open an Explosion-Proof moisture.IQ unit, remove all 24 bolts from around the periphery of the enclosure cover. Do not open an Explosion-Proof moisture.IQ unit that is energized.
To facilitate wiring, Weatherproof and Explosion-Proof moisture.IQ units are equipped with a hinging Chassis that can be swung out and rested on a rod support. During normal use, the Chassis is fixed in place by a latch mechanism located at the center of the mounting panel.
To release the Chassis from its locked position, the latch pin at the center of the mounting panel is pushed upwards (in the direction indicated by the surrounding label). To lock the Chassis in place, the Chassis is pushed down towards the mounting panel until the latch engages. In order to safely move the Chassis after cables are installed, service loops for the Weatherproof units shall be at least 20 inches. Service loops for the Explosion-Proof units shall be at least 18 inches
Make all connections to the back of the meter (see Figure 8 on the next page):
The probe connections in the left two sections are for Module A - Channels 1, 2, 3.
The probe connections in the right two sections are for Module B -Channels 4, 5, 6.
For each Module, the left group of probe connections (MIS, M-SERIES, O2T and OXYGEN) are rated for
hazardous area use, while the right group of probe connections ( non-hazardous area use.
AUX IN/OUT and ALARMS) are rated for
The connections along the bottom of the panel are for power input and remote devices.
14 moisture.IQ User’s Manual
Page 27
1.9 Making Electrical Connections (cont.)
Chapter 1. Installation
Module A Hazardous Area Connections
ON/OFF Switch
Power Cord Receptacle
Module A Non-Hazardous Area Connections
Module B Hazardous Area Connections
Module B Non-Hazardous Area Connections
Figure 8: Rear Panel Connections

1.9.1 Using the Lever to Make Wiring Connections

Use the included levers to assist with wiring the cables to the terminal blocks, as shown in Figure 9 below:
1. Press and hold the lever against the terminal block.
2. Insert the wire into the terminal block.
3. Release the lever to secure the connection.
Figure 9: Using the Press Lock Lever
IMPORTANT: To maintain good contact at each terminal block and to avoid damaging the pins on the connector,
pull the connector straight off (not at an angle). Make the cable connections while the connector is away from the unit, and then push the connector straight on (not at an angle) when the wiring is complete.
Proper connections and cabling are extremely important for accurate measurement. Use the correct cable type for each probe and make sure the cables are not damaged during installation. See the sections that follow for specific connection instructions.
moisture.IQ User’s Manual 15
Page 28
Chapter 1. Installation

1.9.2 Connecting the Input Power

To connect the input power, simply plug the power cable into the receptacle next to the ON/OFF switch at the lower left corner of the back panel (see Figure 8 on page 15). DC wiring is shown in Figure 9 below.
WARNING! If the power mains switch on the unit remains accessible to the user after installation, a disconnect device is not required. However, if the installation blocks access to the power mains switch, compliance with the EU's Low Voltage Directive (IEC 61010) requires an external power disconnect device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly visible, directly accessible, and located within 1.8 m (6 ft) of the unit.
DC Power Supply Connector
21
3
+
-
Figure 10: DC Power Wiring

1.9.3 Connecting Moisture Probes

The moisture.IQ uses M-Series and Moisture Image Series probes to measure moisture. If you need to connect a different probe type, please contact our technical support department. If you ordered one or more M-Series probes with the moisture.IQ, BHGE has entered the necessary probe setup data on a pre-assigned channel.
IMPORTANT: See the Calibration Data Sheets, shipped with the probes, to connect the probes to their
corresponding channels. If a probe is connected to the wrong channel, the meter will display incorrect data.
Probes are identified on the Calibration Data Sheet by a serial number. The serial number is also engraved on the hex nut of the probe, as shown in Figure 11 below.
Probe Serial Number
Figure 11: Moisture Probe Serial Number
The Moisture Image Series probe does not require any pre-programming, because it stores all the necessary setup data in its electronics module. Therefore, you can install a Moisture Image Series probe on any available channel. After you install the probe, you must activate the probe on the installed channel, as described in “The Probe Configuration Screen” on page 60. Proceed to the following sections to properly connect your moisture probes.
16 moisture.IQ User’s Manual
Page 29
Chapter 1. Installation
1.9.3.1 M-Series Probes
M-Series probes are primarily used for moisture measurement, but they can be used to measure temperature as well. If ordered, an optional temperature thermistor is included in the moisture probe and requires an additional connection.
If the M-Series probe does not have a thermistor, then you may use a two-wire shielded cable with a bayonet-type connector to connect the probe to the electronics unit. If an optional thermistor is installed, then use a four-wire shielded cable with a bayonet-type connector to connect the M-Series Probe to the electronics unit. The M-Series Probe may be located up to 600 m (2000 ft) from the moisture.IQ.
Before making the electrical connections, connect the cable to the probe by inserting the bayonet-type connector onto the probe and twisting the shell clockwise until it snaps into a locked position (approximately 1/8 of a turn). Wire the probe cable per the wiring diagram in Figure 12 below. The M-Series terminal block on the back panel of the moisture.IQ (see Figure 13 on page 18) also notes the corresponding color from the probe cable.
M-SERIES
54321
Note: The label
used to terminate the cable shield.
SHD
BLACK
WHITE
GREEN
RED
.
Temp
M-SERIES
Figure 12: M-Series Probe Cable Wiring Diagram
SHD in Figure 12 above corresponds to the SH label on the connector, and these terminals are
Moisture
moisture.IQ User’s Manual 17
Page 30
Chapter 1. Installation
1.9.3.1 M-Series Probes (cont.)
M-Series Probe Connections
Module A - CH1
Module A - CH2
Module A - CH3
Figure 13: M-Series Probe Connections
Note: Additional M-Series probes may be connected to CH4, CH5 and CH6 in the Module B connector group on
the right side of the back panel.
If you connect a probe to the wrong channel, you can either reconnect the probe to the assigned channel or reconfigure the current channel as described in “The Probe Configuration Screen” on page 60.
WARNING! The M-Series probe may not be capable of withstanding the 500V insulation rating. This must be taken into account in any installation in which it is used.
18 moisture.IQ User’s Manual
Page 31
Chapter 1. Installation
1.9.3.2 Moisture Image Series Probes (MISP and MISP2)
Connect the Moisture Image Series probes (MISP) to the moisture.IQ using the cable provided. You can locate a Moisture Image Series probe up to 915 m (3000 ft) from the electronics unit.
Note: Sale of the MISP probe was discontinued by BHGE in 2010 and it was replaced by the MISP2 probe.
Although the MISP probe can be used successfully with the moisture.IQ analyzer, there may be communication issues when using long cables.
Before making any electrical connections, you must assemble the probe. A Moisture Image Series probe may be shipped in two parts: a probe and an electronics module, each of which has its own serial number. If you ordered more than one probe, be sure to match each probe to its matching electronics module using the serial numbers listed on the probe Calibration Data Sheet. To assemble a Moisture Image Series probe, insert the probe into the probe connector on the matching electronics module and turn the probe counterclockwise until it locks in place (see Figure 14 below).
Connector
Figure 14: Moisture Image Series Probe (MISP) Assembly
CAUTION! The MISP2 probe has an integral set of electronics permanently attached (see Figure 15
below). Do not attempt to separate an MISP2 probe from its electronics.
Figure 15: MISP2 Probe
moisture.IQ User’s Manual 19
Page 32
Chapter 1. Installation
1.9.3.2 Moisture Image Series Probes (cont.)
If the Moisture Image Series probe requires assembly, complete that task before proceeding. Then, connect the probe cable to the terminal block labeled MIS on the back panel of the electronics unit (see Figure 16 below).
You can connect the Moisture Image Series Probe to any channel. However, if you are also using other sensors such as M-Series probes, be sure to connect the Moisture Image Series probe to an unassigned channel.
IMPORTANT: Check the Calibration Data Sheets of all installed sensors to determine which channels already
have probes assigned to them.
MIS Probe Connections
Module A - CH1
Module A - CH2
Module A - CH3
Figure 16: MIS Probe Connections
Note: Additional MIS probes may be connected to CH4, CH5 and CH6 in the Module B connector group on the
right side of the back panel.
After you make the back panel connections, connect the other end of the probe cable to the Moisture Image Series probe per the wiring diagram in Figure 17. Trim all other flying leads from the cable flush with the
jacket.
After you complete the Moisture Image Series probe connections, you must activate the probe on the installed channel as described in “The Probe Configuration Screen” on page 60.
MIS
21
WHITE
BLACK
MISP 1
or MISP 2
Figure 17: MIS Probe Cable Wiring Diagram
20 moisture.IQ User’s Manual
Page 33
Chapter 1. Installation

1.9.4 Connecting the Delta F Oxygen Cell

The Delta F Oxygen Cell is available in a general-purpose model with standard or VCR® fittings. The oxygen cell can also be mounted in a weatherproof enclosure (R4) for outdoor applications or an explosion-proof enclosure (R7) for use in hazardous areas.
CAUTION! Do not power up the moisture.IQ without first establishing a gas flow through the Delta F Oxygen Cell (see “Establishing a Gas Flow Through the Oxygen Cell” on page 26).
Each Delta F Oxygen Cell has a set of sensor electrodes and a set of secondary electrodes. For proper operation, make connections to each set of electrodes using a six-wire shielded cable. BHGE provides the 704-1357-B-Z cable with 22 AWG conductors for use with Delta F Oxygen cells.
Note: Cable with 16 AWG conductors is the largest cable size that can be easily installed into the moisture.IQ
terminal blocks and the Delta F sensor terminal posts.
Cable error is a function of cable resistance/foot, length of cable, and maximum sensor current output. Since higher range sensors have a greater current output, they have shorter acceptable cable lengths. Larger gauge cable results in longer acceptable cable lengths. Use Table 1 below to determine acceptable installation lengths for your application.
Table 1: Acceptable Cable Lengths for Delta F Ranges
Delta F Sensor Range Cable Gauge Max Length
0-50 ppm and 0-100 ppm 22 AWG 1300 ft
0-1000 ppm 22 AWG 400 ft
0-10,000 ppm and greater 22 AWG 100 ft
0-50 ppm and 0-100 ppm 20 AWG 2100 ft
0-1000 ppm 20 AWG 630 ft
0-10,000 ppm and greater 20 AWG 160 ft
0-50 ppm and 0-100 ppm 18 AWG 3300 ft
0-1000 ppm 18 AWG 1000 ft
0-10,000 ppm and greater 18 AWG 250 ft
0-50 ppm and 0-100 ppm 16 AWG 6600 ft
0-1000 ppm 16 AWG 2000 ft
0-10,000 ppm and greater 16 AWG 500 ft
moisture.IQ User’s Manual 21
Page 34
Chapter 1. Installation
1.9.4 Connecting the Delta F Oxygen Cell (cont.)
Instructions for connecting each type of oxygen cell are described in the following sections. If you are installing the oxygen cell in an intrinsically safe area, refer to the following sections for special installation requirements.
WARNING! The Delta F Oxygen Cell is BASEEFA approved for use in intrinsically safe areas when connected to a BASEEFA approved moisture.IQ hygrometer. Install the apparatus so that the terminals are protected to at least the requirements of IP20. Copies of official BASEEFA documentation (certificates of compliance, licenses, etc.) are to be made in their entirety.
O2T Connections
Oxygen Probe Connections
Module A - CH1
Module A - CH2
Module A - CH3
Figure 18: Delta F Oxygen Cell Connections
Note: Additional Delta F Oxygen cells may be connected to
CH4, CH5 and CH6 in the Module B connector group
on the right side of the back panel.
Note: For moisture.IQ Fiberglass Weatherproof installation with optional Delta F sensors, clamp a ferrite bead
GE P/N 222-031 to each Delta F cable on the inside of the enclosure.
O
T
2
12
OXYGEN
1234
BLUE
YELLOW
BLACK
WHITE
RED
GREEN
.
-
-
+
*Terminated at sensor side
Temp (Optional)
Stab-el
DELTA F
Figure 19: Delta F Oxygen Cell Wiring
22 moisture.IQ User’s Manual
+
Oxygen
Page 35
Chapter 1. Installation
1.9.4.1 Standard Delta F Oxygen Cells
Figure 20 below shows a standard Delta F Oxygen Cell and identifies the sensing and secondary electrodes. Make oxygen cell connections from the electrodes on the oxygen cell to the OXYGEN terminal block on the back of the moisture.IQ. Refer to Figure 18 on page 22 and Table 2 below for the required connections.
IMPORTANT: To maintain good contact at each terminal block and to avoid damaging the pins on the connector,
pull the connector straight off (not at an angle), make cable connections while the connector is away from the unit, and push the connector straight on (not at an angle) when the wiring is complete.
Table 2: Standard Delta F Oxygen Cell Connections
To Delta F
Connect:
Oxygen Cell:
red wire + sensing electrode pin 4 RD
green wire – sensing electrode pin 3 GR
white wire + secondary electrode pin 2 WT
black wire – secondary electrode pin 1 BK
To moisture.IQ OXYGEN
Terminal Block
shield Earth Ground Connection ---
Secondary Electrodes
Sensing
Electrodes
Figure 20: Standard Delta F Oxygen Cell
Note: The temperature reading of the sensor is used for internal temperature compensation and does not
require any user setup.
moisture.IQ User’s Manual 23
Page 36
Chapter 1. Installation
1.9.4.2 Weatherproof Delta F Oxygen Cells
The weatherproof oxygen cell has a set of sensing and secondary electrodes that are wired to a terminal strip in the weatherproof enclosure. Connect the weatherproof oxygen cell using a four-wire shielded cable with a mating bayonet-type connector. Fasten the bayonet-type connector to the mating connector on the bottom of the weatherproof enclosure. Connect the other end of the cable to the OXYGEN terminal block on the back of the moisture.IQ. Refer to Figure 18 on page 22 and Table 3 below for the required connections.
Table 3: Weatherproof Delta F Oxygen Cell Connections
To Delta F Enclosure
Connect:
Terminal Block
red wire (+) pin 1 pin 4
green wire (–) pin 2 pin 3
white wire (+) pin 3 pin 2
black wire (–) pin 4 pin 1
To moisture.IQ OXYGEN
Terminal Block
24 moisture.IQ User’s Manual
Page 37
Chapter 1. Installation
1.9.4.3 Explosion-proof Delta F Oxygen Cells
Figure 21 below shows the explosion-proof oxygen cell. The oxygen cell has a set of sensing and a set of secondary electrodes that are wired to a terminal strip in the explosion-proof enclosure. Connect the explosion-proof oxygen cell using a four-wire shielded cable. Connect one end of the cable to the OXYGEN terminal block on the back of the moisture.IQ and the other end to the terminal strip in the oxygen cell enclosure. Feed the cable wires through one of the cable ports in the side of the explosion-proof enclosure. Refer to Figure 18 on page 22 and Table 4 below for the required connections.
Table 4: Explosion-proof Oxygen Cell Connections
To Oxygen Cell
Connect:
Terminal Block
red wire (+) pin 1 pin 4
green wire (–) pin 2 pin 3
white wire (+) pin 3 pin 2
black wire (–) pin 4 pin 1
To moisture.IQ OXYGEN
Terminal Block:
MAX
MIN
Figure 21: Explosion-Proof Delta F Oxygen Cell
moisture.IQ User’s Manual 25
Page 38
Chapter 1. Installation

1.10 Establishing a Gas Flow Through the Oxygen Cell

CAUTION! Establish a gas sample flow through the Delta F Oxygen Cell before you power up the
system or damage may occur to the oxygen cell.
Note: If you are not using a Delta F Oxygen Cell in your system, skip this section and proceed to the following
chapters to configure your system.
The Delta F Oxygen Cell requires a gas flow rate of 2 to 2.5 SCFH through the cell. The oxygen cell inlet pressure should be between 0.2 and 1.0 psig. Refer to Figure 22 on page 27 when establishing a gas sample flow.
CAUTION! Do not operate the Delta F oxygen cell for extended periods of time at oxygen concentrations that are over range. Trace and low percent range sensors may be damaged if exposed to high levels of oxygen, such as air, for long periods (>1 hour) while the moisture.IQ is in operation. If exposure is unavoidable, either disconnect the oxygen cell from the moisture.IQ or equip the sample system with a valve that allows the cell to be switched to a purge gas.
1. Close the flow control valve and adjust the upstream pressure as required. BHGE recommends a pressure of about 2 to 10 psig pressure upstream from the flow control valve, depending on what type of valve is installed in the sample system.
2. To safeguard against over pressurizing the oxygen cell, install a relief valve rated at 10 psig in the gas flow system. There should be no restrictions downstream from the oxygen cell. Use 1/4 inch tubing or larger on the oxygen cell outlet and the relief valve outlet. Both outlets should vent to atmosphere if possible.
CAUTION! Do not connect the relief valve and oxygen cell outlets to a common outlet line smaller than 1/4 inch. This pressure restriction will damage the oxygen cell. In addition, a relief valve should be installed in the oxygen sample system. If either of these conditions are not met, the Delta F Oxygen Cell warranty will be voided.
26 moisture.IQ User’s Manual
Page 39
Chapter 1. Installation
1.10Establishing a Gas Flow Through the Oxygen Cell (cont.)
3. Slowly open the flow control valve until the recommended flow of 2 to 2.5 SCFH is established on the
flow meter.
4. When the proper flow is achieved, make sure the relief valve is closed by placing an object (e.g., your
finger, if the gas is non-corrosive) over the relief valve vent. Cover and uncover the relief valve vent and verify that the flow meter shows no change in the flow rate. Keep the relief valve closed during operation to minimize leakage in the sample system.
Sample
Gas In
Optimum Inlet
Pressure Limits
0.2 – 1.0 psig
Flow Control
Tee
Relief
Valve
Assembly
Valve
Optimum Sample
2.0 – 2.5 SCFH
Electrolyte
Reservoir
In
Out
Oxygen Cell
Vent
Sample Gas Outlet and Relief Valve
should vent to atmosphere.
Flowmeter
Flow Range
Sample
Gas Out
Figure 22: Gas Flow Schematic
moisture.IQ User’s Manual 27
Page 40
Chapter 1. Installation

1.11 Connecting an Auxiliary Device

moisture.IQ can power an auxiliary device (such as a loop power transmitter) in which the connection is +24V to Aux 1 or 2. For external power devices with output directly connected to the auxiliary device, you can use aux 1 or 2 with respect to RTN (pin 5). Make the connections to the right-side group of probe connections (AUX
IN/OUT
input and the electronics will automatically switch the circuits, so the moisture.IQ has no physical switch to toggle. Figure 23 and Figure 24 below show the connections and wiring for auxiliary devices.
and ALARMS) that are rated for non-hazardous area use. Users set the voltage/current for the auxiliary
Auxiliary Connections
Figure 23: Auxiliary Connections
ANALOG OUT
AUX IN
External Power Mode
-
++
.
++++
----
Analog Out Volt/Current A
Analog Out Volt/Current B
Aux In 1 Aux In 2
OR
Loop Powered Mode
56 78
-+-
87654321
.
+
--
+
Aux In 2Aux In 1
Figure 24: Auxiliary Device Wiring Diagram
28 moisture.IQ User’s Manual
Page 41
Chapter 1. Installation
Note: If you are connecting the oxy.IQ to AUX-IN, the oxy.IQ's shield needs to be terminated to the moisture.IQ
via RTN (pin 5). In order to expose the shield wire, strip back the cable jacket.

1.12 Connecting Analog Outputs

The moisture.IQ has two Analog Outputs per channel. Connect the outputs to the terminal blocks, labeled REC A and REC B. See Figure 23 for the terminal block location and Figure 25 for the wiring diagram. Make connections for recorders using Table 5.
Table 5: Output Connections
To REC Terminal
Connect Output A
return (-) pin A-
out (+) out (+)
Connect Output B
return (-) pin B-
out (+) pin B+
Block
To REC Terminal
Block
AUX OUT
123
+
Analog Out Volt/Current 1
-
Figure 25: Output Wiring
+
Analog Out Volt/Current 2
4
-
moisture.IQ User’s Manual 29
Page 42
Chapter 1. Installation

1.13 Connecting Alarms

Note: Customers must provide their own cables for connecting alarm relays. Acceptable cables range from 12 to
24 AWG.
The moisture.IQ has two high/low alarm relays per channel. There is also one system fault alarm relay, which is not channel specific. Each alarm relay has single-pole, double-throw contacts that consist of the following:
Normally Open (NO)
Common (C)
Normally Closed (NC)
Table 6 and Figure 26 below list alarm wiring connections.
Table 6: Pin Designations for Relay Contacts
Description Fault Alarm A Alarm B
Normally Open 1 1 6
Common 2 2 7
Normally Closed 3 3 8
123
CC
NO
Alarm A
NC
Alarms
5
4
RTN RTN
67 8
NO
Alarm B
NC
Fault Alarm Connector
12
C
NO
Figure 26: Alarm Wiring
3
NC
30 moisture.IQ User’s Manual
Page 43

Chapter 2. Operation

Chapter 2. Operation

2.1 Powering Up

After completing the installation of your system, as described in Chapter 1. “Installation” on page 1, power up the moisture.IQ with the ON/OFF switch located at the lower left corner of the back panel. You will observe the following display screens in sequence:
1. A black screen with a large GE monogram in the center.
2. A black screen with a GE monogram and the “moisture.IQ” name, along with a blue progress bar along
the right edge of the screen.
3. A measurement display screen, similar to that shown in Figure 27 below.
Figure 27: Measurement Display Screen
Note: The moisture.IQ is equipped with an automatic cooling fan. This fan will activate as needed to maintain
the proper temperature inside the enclosure. The periodic sound of this fan running is normal.

2.2 Measurement Display Components

The measurement display screen shown in Figure 27 above is a touch screen which includes the following items:
Either 6 or 12 measurement displays per screen (depending on your display configuration settings)
The main menu for programming the moisture.IQ, consisting of the 6 options shown
Left and Right arrows, with a page number between them, which are used to scroll through the pages of
measurement displays (6 pages of 6 measurements/page or 3 pages of 12 measurements/page)
To access any of the screen components, simply tap on the desired item with your finger tip.
moisture.IQ User’s Manual 31
Page 44
Chapter 2. Operation

2.3 The Help Button

In the main menu on the right side of the measurement display screen and on every other moisture.IQ programming screen, a Help button is available to provide the user with detailed information on that specific screen or menu option. A sample Help screen from the Probe Configuration menu is shown in Figure 28 below.
Figure 28: Data Display Help Screen
To access the context-sensitive help information for any screen, just tap the
Help button on that screen.

2.4 The Data Element Editor

The display of any of the 36 data measurements available from the main screen may be individually configured by tapping on the desired measurement display. This will open a Data Element Editor screen similar to that shown in Figure 29 below.
Figure 29: The Data Element Editor Screen
32 moisture.IQ User’s Manual
Page 45
Chapter 2. Operation
moisture.IQ User’s Manual 33
Page 46
Chapter 2. Operation
2.4 The Data Element Editor (cont.)
As shown in Figure 29 on page 32, the following settings are available for the display:
Channel: Select the channel (1-6) to be displayed.
Mode: Select the measurement type (Hygro, Pressure, Temperature, Oxygen, Aux 1, Aux 2 or Function)
to be displayed.
Unit: Select the units of measure to be displayed.
123/Graph: Toggle between a Numeric and Graphical display.
Note: The error message is only available in numeric mode — not in graphical mode.
12.23/1.2E5: Toggle between Floating Point Decimal and Scientific numeric displays.
Slider Bar: Use the Left and Right arrows to select the number of decimal places for the numeric display.
Copy and Paste: Use these buttons to duplicate display settings from one measurement display to
another.
Clear All: Use this button reset the display to the default state.
Note: The current selection for the toggled items is highlighted in blue.
As with all of the moisture.IQ programming screens, tap the
tap the
Cancel button to keep the previous settings.
Check button to save your new settings, or

2.5 Initial Setup

Before you can obtain accurate measurement data, you must properly configure the moisture.IQ and all attached probes. This is accomplished by using the Main Menu options along the right side of the measurement display screen. Simply tap the desired menu item on the touch screen to access that particular menu, and proceed to the appropriate chapter in this manual for detailed instructions.
After your initial setup is complete, these same menus can be used to program the moisture.IQ with your personal preferences and to reconfigure the unit whenever there are any changes in your system.
Note: Initially, all menus are locked at startup. After you enter a password, these menus will be unlocked, but
will timeout (and relock) after 15 minutes with no activity.
34 moisture.IQ User’s Manual
Page 47
Chapter 2. Operation

2.6 Restarting the System

To restart the moisture.IQ, complete the following steps:
1. Tap the Settings Menu button on the right side of the touch screen.
2. Tap the Restart button in the Service section of the touch screen.
3. At the confirmation screen shown in Figure 30 below, tap the Yes button to restart the unit or tap the
No button to cancel the request.
Figure 30: Restart Confirmation Screen
moisture.IQ User’s Manual 35
Page 48
Chapter 2. Operation

2.7 Shutting Down the System

To shut down the moisture.IQ, complete the following steps:
1. Tap the Settings Menu button on the right side of the touch screen.
2. Tap the Shut Down button in the Service section of the touch screen.
3. At the confirmation screen shown in Figure 31 below, tap the Yes button to shut down the unit or tap
the No button to cancel the request.
4. After the message “It is now safe to shut down the unit” is displayed on the touch screen, use the
On/Off power switch at the lower left corner of the back panel to shut down the moisture.IQ.
[no content intended for this page]
Figure 31: Shut Down Confirmation Screen
36 moisture.IQ User’s Manual
Page 49
Chapter 2. Operation
moisture.IQ User’s Manual 37
Page 50
Chapter 2. Operation
38 moisture.IQ User’s Manual
Page 51

Chapter 3. Using the Settings Menu

Chapter 3. Using the Settings Menu

3.1 Introduction

After completing the installation of your moisture.IQ system (see Chapter 1) and powering up the system (see Chapter 2), you must configure the moisture.IQ before reliable data can be collected.
The Settings menu is used to configure the display and to program the desired system settings. Use the menu map in Figure 76 on page 109 and the display touch screen shown in Figure 32 below as guides while programming the Settings menu. To begin, tap the Settings menu button on the main measurement display screen, and proceed to the following sections for instructions.
Figure 32: Settings Menu Screen
Figure 33: Settings Menu Screen for Explosion-Proof and Weatherproof
moisture.IQ User’s Manual 39
Page 52
Chapter 3. Using the Settings Menu

3.2 Display Settings

The Display Settings menu (see Figure 34 below) allows you to configure the measurement screen.
Figure 34: The Display Settings Screen
Tap the 6 Items or 12 Items button to toggle between a main screen with 6 measurement displays or
one with 12 measurement displays. The current setting is highlighted in blue.
To adjust the screen brightness, tap the button at the right end of the Brightness slider to increase the
screen brightness or the button at the left end of the slider to decrease the screen brightness, until the desired setting is achieved.
The blue indicator bar on the slider indicates the current setting location.
After completing your settings, either tap the Check button to save your changes, or tap the Cancel
button to discard the changes and keep the original settings.
40 moisture.IQ User’s Manual
Page 53
Chapter 3. Using the Settings Menu

3.3 System Settings

The System Settings menu (see Figure 35 below) allows you to configure the global moisture.IQ settings.
Figure 35: The System Settings Screen
The system Serial Number is displayed as a read-only value at the top of the screen.
Tap the Date button and select the current date from the pop-up calendar. Tap the Date Format button
and select the desired format (
MM/DD/YYYY, DD/MM/YYYY or YYYY-MM-DD) from the drop-down list.
Tap the Time button and set the current time with the pop-up Hours and Minutes counters. Tap the
Time Format button to toggle between the available formats (24 Hr or AM/PM).
Tap the Number Format button and select either 1,234.567 or 1.234,567 from the drop-down list to
specify the format for displaying numbers.
After completing your settings, either tap the Check button to save your changes, or tap the Cancel
button to discard the changes and keep the original settings. Then, tap the the Settings Menu.
Note: If the Number Format is set to 1.234,567, it is best to avoid using a Comma for the field separator in data
logs.
Return button to return to
moisture.IQ User’s Manual 41
Page 54
Chapter 3. Using the Settings Menu

3.4 File Manager

The File Manager menu (see Figure 36 below) enables you to view and manage any files stored in the moisture.IQ memory or in a connected USB drive. The most current site file is always named moistureIQ.xml, while the previous file is renamed to moistureIQ.xml.backup.
Figure 36: The File Manager Screen
Tap the File Type button to open a drop-down list of options for selecting the type of files (Log Files, Site
Files, Cal Files, User Manual
or Audit Log) to be displayed.
Tap the Sort By button to open a drop-down list of options for sorting the list of files (Name, Date or
).
Size
Tap the name of any listed file to display detailed information about the file in the File Information box.
With a file name highlighted, tap either the Int or Ext button under Transfer Options to specify the
location for storing the selected file. (available only if an external USB drive is connected) stores the file to a USB drive connected to the moisture.IQ. The current location is highlighted in blue.
Int stores the file to the internal moisture.IQ memory, while Ext
As needed, tap the buttons in the upper right corner to Delete a file altogether, Copy it to another
location or
Transfer it to another location.
The User Manual and Audit Log are read-only files that can be transferred to an external device, but not
deleted.
After completing your file management activities, tap the Return button to return to the Settings menu.
42 moisture.IQ User’s Manual
Page 55
Chapter 3. Using the Settings Menu

3.5 Screen Alignment (Weatherproof and Explosion-Proof)

The Screen Alignment feature is enabled for Weatherproof and Explosion-Proof versions of the moisture.IQ. This feature allows the user to calibrate the LCD screen to their touch and angles of usage to optimize touch screen performance.
Figure 37: Screen Alignment
Touch the “+” symbol in its five locations: the middle and all four corners of the screen.
After you have completed touching the locations, simply tap anywhere in the screen to complete the
calibration.
Note: Ignore the Enter/ESC instructions.
moisture.IQ User’s Manual 43
Page 56
Chapter 3. Using the Settings Menu

3.6 Screen Configuration (Weatherproof and Explosion-Proof)

The Screen Configuration feature is enabled for Explosion-Proof and Weatherproof versions of the moisture.IQ. This feature allows the user to increase or decrease the touch screen's sensitivity based on the thickness of the protective window. Two other features allow for the touch screen's configuration in terms of X-axis and Y-axis to be flipped, but are only used in extremely rare cases. To adjust the screen's sensitivity based on thickness:
Figure 38: Screen Configuration
Tap Thick.
The existing setting will be highlighted in blue. Tap a setting above/below the current setting to
decrease/increase the sensitivity of the screen.
44 moisture.IQ User’s Manual
Page 57
Chapter 3. Using the Settings Menu

3.7 Notifications Setup

The Notifications Setup menu (see Figure 39 below) allows you to specify how often the moisture.IQ will display calibration reminders.
Figure 39: The Notifications Setup Screen
Tap the Calibration Reminder button to open a drop-down list of options for selecting the interval
(
None,
6 Months, 12 Months, 18 Months or 24 Months
) between automated reminder notifications.
After completing your selection, either tap the Check button to save your changes, or tap the Cancel
button to discard the changes and keep the original settings. In either case, you will be returned to the Settings Menu.
moisture.IQ User’s Manual 45
Page 58
Chapter 3. Using the Settings Menu

3.8 Fault Alarm Setup

The Fault Alarm Setup menu (see Figure 40 below) allows you to specify how the moisture.IQ dedicated fault alarm relay responds to a fault condition and it lists the source of any current fault conditions.
Figure 40: The Fault Alarm Setup Screen
Tap the In Case of Fault button to toggle between the Energized and DeEnergized options. This selection
determines whether the relay is detected. The source of a fault is displayed in the fault list in Figure 40 above.
Note: The fault alarm relay is tripped whenever any sensor is open, shorted or disconnected. It will also trip
when an overheating condition is detected.
Energized or DeEnergized (fail-safe operation) when a fault condition is
After completing your selection, either tap the Enter button to save your changes, or tap the Cancel
button to discard the changes and keep the original settings. In either case, you will be returned to the Settings Menu.
46 moisture.IQ User’s Manual
Page 59
Chapter 3. Using the Settings Menu

3.9 Module Setup

The Module Setup menu (see Figure 41 below) is primarily a read-only screen. It provides detailed information on the currently installed sensor modules.
Figure 41: The Module Setup Screen
The Serial Number and the current Firmware Version for the installed modules are listed at the top of
the screen.
Below the data boxes, the Module Letter and the Available Channels for each sensor module are listed.
When an external USB drive containing a sensor module firmware update file is connected to the
moisture.IQ, the
Module to be updated and then tap the Update Firmware button. The module firmware will be updated
and the system will reboot automatically.
Update Firmware button is enabled. To update the sensor module firmware, tap the
After you have finished reading the information on this screen, tap the Return button to return to the
Settings Menu.
moisture.IQ User’s Manual 47
Page 60
Chapter 3. Using the Settings Menu

3.10 Service Menu Options

This menu includes the three options described below.

3.10.1 Software Update

When an external USB drive containing a moisture.IQ software update file is connected, the Software Update button will be enabled. Proceed as follows:
1. Tap the Software Update button.
2. Tap Ye s at a screen similar to Figure 42 below to confirm your choice, or tap Cancel to abort the
update.
3. The new software will be installed automatically and you will be prompted to tap a Restart button to complete the process.
Note: Your system configuration settings will not be affected by the software update.
Figure 42: Software Update Confirmation Screen

3.10.2 Restart

To restart (i.e., reboot) the moisture.IQ system, tap the Restart button. Then, tap either Yes or No at a confirmation screen similar to the one shown in Figure 30 on page 35.

3.10.3 Shut Down

To shut down the moisture.IQ system, tap the Shut Down button. Then, tap either Yes or No at a confirmation screen similar to the one shown in Figure 31 on page 36.
48 moisture.IQ User’s Manual
Page 61

Chapter 4. Using the Outputs, Alarms and Logger Menus

Chapter 4. Using the Outputs, Alarms and Logger Menus

4.1 Setting Up the Outputs

4.1.1 Configuring an Output

Note: An active output is indicated by a “Play” symbol, and the alarm selected for editing is highlighted in
yellow. A dashed border around an output indicates that the channel is not installed.
The moisture.IQ features two isolated analog outputs (A and B) for each of its six channels. To configure these outputs, refer to the menu map in Figure 77 on page 110 and tap the Outputs button on the right side of the main touch screen to open a screen similar to that shown in Figure 43 below.
Figure 43: Outputs Selection Screen
1. Tap the button for the desired
2. In the Measurement Selection section, tap the Mode, Unit and Type (Current or Voltage) buttons to enter
the desired settings for the output. (For available modes and units, refer to Table 8 on page 61.)
3. Tap the Zero box and enter the zero value via the keypad.
4. Tap the Span box and enter the span value via the keypad.
5. Tap the Check button to save your settings.
Channel (1-6) and Output (A or B).
moisture.IQ User’s Manual 49
Page 62
Chapter 4. Using the Outputs, Alarms and Logger Menus

4.1.2 Testing the Selected Output

Make sure the output device is connected to the moisture.IQ as described in Chapter 1, “Installation” and complete the following steps to test the output:
1. Tap the Channel (1-6) and Output (A or B) to be tested, and then tap the Test button to open a screen similar to Figure 44 below.
Figure 44: Output Test Screen
2. Tap the text box and enter the desired Percent of Scale test value (0-120). Then, tap the
Check button to
send the test value to the output.
3. After about 5 seconds, this reading should be as shown in Table 7 below.
Table 7: Expected Multimeter Test Readings
Output Range Multimeter Reading
0 to 20 mA 20 x Test %/100 mA
4 to 20 mA 4 + 16 x Test %/100 mA
0 to 2V 2 x Test %/100 V
4. Repeat steps 1-3 for as many different test values as desired. When you have finished testing the output, tap the
Exit button to return to the Output Selection screen.
50 moisture.IQ User’s Manual
Page 63
Chapter 4. Using the Outputs, Alarms and Logger Menus

4.1.3 Trimming the Selected Output

The measured value of the outputs can vary from the programmed value due to the effects of load resistance. The Output Selection option provides a trimming feature you can use to compensate for any such variations. To accurately trim the outputs, you will need a digital multimeter capable of measuring 0 to 2 V with a resolution of ±0.0001 VDC (0.1 mV) or 0 to 20 mA with a resolution of ±0.01 mA, depending on your output settings. Most good quality 3 1/2-digit multi-meters are adequate for output trimming. Complete the following steps to trim output:
1. Make sure the Output Mode has been set for the desired output (Current or Voltage).
2. Temporarily disconnect the load from the output signal wires. The, attach the digital multimeter to the
signal wires either in series (for Current mode) or in parallel (for Voltage mode) with the load.
3. Tap the Channel (1-6) and Output (A or B) to be trimmed, and then tap the Trim button to open a screen
similar to Figure 45 below.
Figure 45: Output Trim Screen
4. Tap the
Reset Trim button to clear the current trim values.
5. Tap the Trim Zero button to output the zero value, and enter the value shown on your multimeter in the
Trim Zero text box.
6. Tap the Trim Span button to output the span value, and enter the value shown on your multimeter in
the Trim Span text box.
7. Tap the Exit button and test the output as described in “Testing the Selected Output” on page 50.
Note: The 0% of full scale value is: 1 mA for a 0-20 mA scale, 4 mA for a 4-20 mA scale or 0.1 V for a 0-2 V scale.
moisture.IQ User’s Manual 51
Page 64
Chapter 4. Using the Outputs, Alarms and Logger Menus

4.1.4 Setting the Output Range Error Response

A Range Error can occur when a measured value is within the capacity of the analyzer, but it exceeds the calibration range of the probe. They can be either Under Range or Over Range errors.
The moisture.IQ indicates Range Errors with an Over Range or Under Range message.The error condition applies to all measurements of the same mode. For example, if a dew point measurement is Over Range, then moisture in ppMv will also be Over Range. If multiple Range Errors occur simultaneously, the moisture.IQ responds to them in the following order: Oxygen Errors, Moisture Errors, Temperature Errors and then Pressure
Errors.
Complete the following steps to configure the range error response of the selected output:
1. Tap the desired Channel (1-6) and Output (A or B), and then tap the Range Error button to open a screen similar to Figure 46 below.
Figure 46: Range Error Setup Screen - Output Response
2. Set the output to be driven either
High on Under Range errors, Low on Under Range errors or to Ignore
Under Range errors (Ignore is the default setting).
3. Set the output to be driven either High on Over Range errors, Low on Over Range errors or to Ignore Over Range errors (Ignore is the default setting).
4. Tap the Save button to keep the new settings or tap the Cancel button to keep the previous settings.
52 moisture.IQ User’s Manual
Page 65
Chapter 4. Using the Outputs, Alarms and Logger Menus

4.2 Setting Up the Alarms

4.2.1 Configuring an Alarm

Note: An active alarm is indicated by a “Play” symbol, and the alarm selected for editing is highlighted in yellow.
The moisture.IQ features two optional alarms (A and B) for each of its six channels. To configure these alarms, refer to the menu map in Figure 77 on page 110 and tap the Alarms button on the right side of the main touch screen to open a screen similar to that shown in Figure 47 below.
Figure 47: Alarm Selection Screen
1. Tab the button for the desired
Channel (1-6) and Alarm (A or B).
2. In the Measurement Selection section, tap the Mode, Unit and Type (In Band, Out Band or Setpoint)
buttons to enter the desired settings for the alarm. (For available modes and units, refer to Table 8 on page 61.)
3. Tap the Upper text box and enter the upper value via the keypad. Repeat this procedure for the Lower
value.
moisture.IQ User’s Manual 53
Page 66
Chapter 4. Using the Outputs, Alarms and Logger Menus

4.2.2 Testing the Selected Alarm

Complete the following steps to test the alarm:
1. Tap the Channel (1-6) and Alarm (A or B) to be tested, and then tap the Test button to open a screen similar to Figure 48 below.
Figure 48: Alarm Test Screen
2. Tap the
Trip button to trip the alarm and verify that the alarm has been tripped.
3. Tap the Reset button to reset the alarm and verify that the alarm has been reset.
4. After you have finished testing the alarm, tap the Exit button to return to the Alarm Selection screen.
54 moisture.IQ User’s Manual
Page 67
Chapter 4. Using the Outputs, Alarms and Logger Menus

4.2.3 Setting the Alarm Range Error Response

A Range Error can occur when a measured value is within the capacity of the analyzer, but it exceeds the calibration range of the probe. They can be either Under Range or Over Range errors.
The moisture.IQ indicates Range Errors with an Over Range or Under Range message.The error condition applies to all measurements of the same mode. For example, if a dew point measurement is Over Range, then moisture in ppMv will also be Over Range. If multiple Range Errors occur simultaneously, the moisture.IQ responds to them in the following order: Oxygen Errors, Moisture Errors, Temperature Errors and then Pressure
Errors.
Complete the following steps to configure the range error response of the selected alarm:
1. Tap the Channel (1-6) and Alarm (A or B), and then tap the Range Error button to open a screen similar
to Figure 49 below.
Figure 49: Range Error Setup Screen - Alarm Response
2. Set the alarm to either
Trip on Under Range errors or to Ignore Under Range errors (Ignore is the default
setting).
3. Set the alarm to either Trip on Over Range errors or to Ignore Over Range errors (Ignore is the default
setting).
4. Tap the Save button to keep the new settings or tap the Cancel button to keep the previous settings.
moisture.IQ User’s Manual 55
Page 68
Chapter 4. Using the Outputs, Alarms and Logger Menus

4.3 Setting Up and Running Logs

To set up and run logs, refer to the menu map in Figure 77 on page 110 and tap the Logger button on the main touch screen to open a screen similar to Figure 50 below.
Figure 50: Logger Setup Screen
The following logging functions are available on the Logger Setup screen:
The pane on the left lists all current log files in memory. Tap the Sort By button and then select Name,
Date, Size
sorted.
or Running from the drop-down list to specify the manner in which the list of log files is
To see more details about any of the listed log files, highlight the desired files and the available
information for that log will be displayed in the
Log Information pane.
The two buttons above the Log Information pane are used to control a Running log. Tap the End button
to permanently end the log or tap the paused, that button changes to a Start button. Tap this button to resume a paused log or to start a new log.
Pause button to temporarily suspend the log. After a log is
The Transfer Log button is used to move a highlighted log file from the internal moisture.IQ memory to
a connected USB drive. Tap the
Transfer Log button and follow the instructions.
The Delete Log button is used to delete a highlighted log file. Tap the Delete Log button and then
confirm your choice.
The View Log button is used to view the measurements selected for a highlighted log file. Tap the View
button, and the measurements are shown in the Log Information pane.
Log
The Clone Log button is used to create a new log based on the settings of a highlighted existing log. If a
log has ended, you can create a new log with identical measurements and options. Tap the button, edit the filename for the cloned log, and modify the Start Time and End Time. If desired, you may also modify the Separator, Log Type or Measurements settings. When the setup is complete, tap the Return button and start the new log.
56 moisture.IQ User’s Manual
Clone Log
Page 69
4.3 Setting Up and Running Logs (cont.)
Create Log button:
Chapter 4. Using the Outputs, Alarms and Logger Menus
a. Tap the
b. Enter the
(Normal, Cyclic or Error) and the Logging Interval (in minutes:seconds).
c. Tap the Log Data button at the upper right corner of the Log Creator screen to open the Log Data
screen.
Create Log button to open the Log Creator screen, as shown in Figure 51 below.
Figure 51: Log Creator Screen
Log File Name, the Start Time, the End Time, the Separator (Comma or Ta b), the Log Type
d. Highlight one of the 16 log measurement boxes and use the buttons at the top of the screen to
specify the Channel, Mode and Unit for that measurement.
e. After you have finished configuring your log data, tap the Return button to return to the Log Creator
screen above. Then, tap the Return button again to return to the Logger Setup screen.
f. When you are ready, tap the Start button to begin running the new log.
moisture.IQ User’s Manual 57
Page 70
Chapter 4. Using the Outputs, Alarms and Logger Menus
[no content intended for this page]
58 moisture.IQ User’s Manual
Page 71

Chapter 5. Using the Configuration Menu

Chapter 5. Using the Configuration Menu

5.1 Configuring the Probes

After establishing the meter settings, you must configure and calibrate the attached probes. Although the probes are physically connected to the back of the electronics unit, you must program the moisture.IQ with the type of measurements you need the probe to make. In addition, if you want to use a constant value rather than a live input, or if you want to apply a user function, you must program the moisture.IQ accordingly. If you do not activate the probes, or if you activate them incorrectly, the meter will display No Probe or other error messages.
Refer to the menu map in Figure 78 on page 111 and tap the Configuration button on the main screen to open the Configuration Menu (see Figure 52 below).
Figure 52: Configuration Menu Screen
moisture.IQ User’s Manual 59
Page 72
Chapter 5. Using the Configuration Menu

5.1.1 The Probe Configuration Screen

Tap the Probe Config button to open the Probe Configuration screen (see Figure 53 below).
Figure 53: Probe Configuration Screen
From the Channel vs. Mode grid, you can activate the installed probes on each channel. See Table 8 on page 61 for the available measurement modes and units. The probe options vary with the Mode selected, as
follows:
Hygro - M-Series, MISP/MISP2 (MIS), Constant Value (kH) or M-Series with Computer-Enhanced
Response* (M-CER)
Note: Only M-Series PR probes will work with Computer-Enhanced Response. On the main display, the
measurement label will appear in blue.
Temp - M-Series, MISP/MISP2 (MIS), Constant Value (kT)
Pressure - MISP/MISP2 (MIS), Auxiliary Input 1 (Aux 1), Auxiliary Input 2 (Aux 2) or Constant Value
(
kP)
Oxygen - GE01-GE09, % O2 (Percent), PPM O2 (Parts per Million), PPB O2 (Parts per Billion)
[see “Configuring Delta F Oxygen Sensor Probes” on page 63 for details.]
Aux 1 - mA (current), Volts (voltage)
Aux 2 - mA (current), Volts (voltage)
IMPORTANT: Activating probes that are not physically connected to the meter will result in error messages and
slow performance.
60 moisture.IQ User’s Manual
Page 73
Chapter 5. Using the Configuration Menu
5.1.1 The Probe Configuration Screen (cont.)
Table 8: Available Measurement Modes and Units
Selected
Measurement
Mode
Unit Description
Oxygen % = Percent Oxygen (default) Oxygen %
PPM = Parts Per Million Oxygen PPM
PPB = Parts Per Billion Oxygen PPB
μ = Microamps (Diagnostic Mode) Oxygen μ
Hygrometry DP °C = Dew/Frost Point Degrees Celsius (default) Dew Point °C
DP °F = Dew/Frost Point Temperature Fahrenheit Dew Point °F
PPMv = Parts per Million of Water by Volume H
PPBv = Parts per Billion of Water by Volume H
PPMw = Parts per Million of Water by Weight H
RH % = Relative Humidity Rel. Humidity %
MMSCFig = Pounds of Water per Million Std. Cubic Feet in Ideal Gas
MMSCFng = Pounds of Water per Million Std. Cubic Feet in Natural Gas
Equiv DP°C NG = Equivalent Dew/Frost Point Degrees Celsius in Natural Gas
Displayed
Measurement Mode
O PPMv
2
O PPBv
2
O PPMw
2
O/MMSCFig Lbs
H
2
O/MMSCFng Lbs
H
2
Equiv DP °C
Displayed
Units
Equiv DP°F NG = Equivalent Dew/Frost Point Degrees
Equiv DP °F
Fahrenheit in Natural Gas
PPMv/ng = Parts Per Million by Volume in Natural Gas H
3
= grams per meter cubed
g/m
mg/m3 = milligrams per meter cubed
O/Natural Gas PPMv
2
Hygro
Hygro
g/m
mg/m
Pw/kPa = Vapor Pressure in KiloPascals Vapor Pressure kPa
Pw/mmHg = Vapor Pressure in Mercury Vapor Pressure mmHg
MH = MH* (Diagnostic Mode) H
FH = FH* (Diagnostic Mode) H
OMH
2
OFH
2
Temperature °C = Degrees Celsius (default) Temperature °C
°F = Degrees Fahrenheit Temperature °F
Kelvin = Kelvin Temperature K
°R = Degrees Rankine Temperature °R
3
3
moisture.IQ User’s Manual 61
Page 74
Chapter 5. Using the Configuration Menu
Table 8: Available Measurement Modes and Units (cont.)
Selected
Measurement
Mode Unit Description
Displayed
Measurement Mode
Displayed
Units
Pressure kPa(a) = KiloPascal Absolute (default) Pressure kPa(a)
mPa(a)= MegaPascal Absolute Pressure MPa(a)
Pa(a) = Pascal Absolute Pressure Pa(a)
kPa(g) = KiloPascal Gauge Pressure kPa(g)
mPa(g)= MegaPascal Gauge Pressure MPa(g)
Pa(g) = Pascal Gauge Pressure Pa(g)
PSI(a) = Pounds per Square Inch Absolute Pressure PSI(a)
PSI(g) = Pounds per Square Inch Gauge Pressure PSI(g)
ATM = atmospheres Pressure ATM
Bar(a) = Bars Absolute Pressure Bar(a)
Bar(g) = Bars Gauge Pressure Bar(g)
mmHg = Millimeters of Mercury Pressure mmHg
FP = FP* (Diagnostic Mode) Pressure FP
Auxiliary 1 mA = Milliamps (default) Aux1 mA
V = Volts Aux1 V
Scaled = scale defined by the user in the calibration menu Aux1 Scaled
Auxiliary 2 mA = Milliamps (default) Aux1 mA
V = Volts Aux1 V
Scaled = scale defined by the user in the calibration menu Aux1 Scaled
User Func1-6 = User defined functions for each channel TBD TBD
*The MH, FH and FP values are the moisture sensors’ response values and
are the values that are recorded during calibration.
62 moisture.IQ User’s Manual
Page 75
Chapter 5. Using the Configuration Menu

5.1.2 Configuring Delta F Oxygen Sensor Probes

There are 13 selections available for the Delta F Oxygen Sensor probes. These are highlighted by the box in the Probe Configuration window shown in Figure 54 below.
Figure 54: Delta F Probe Configuration Screen
The Delta F Oxygen Sensor probes are available in the two following types:
GE 1 through GE 9, which include temperature compensation
% O2, PPM O2 and PPB O2, which do not include temperature compensation
To properly configure your Delta F Sensor probe, use the information found on the label affixed to the sensor body. Figure 55 below shows a sample label for a
GE3 sensor type.
Figure 55: Sample Delta F Probe Label
moisture.IQ User’s Manual 63
Page 76
Chapter 5. Using the Configuration Menu

5.1.3 Calibrating Individual Channels

1. From the main screen, tap Configuration, then Probe Config, then Reference Values to access the Channel Calibration screen.
Figure 56: Calibrating Channels
2. In the Channel Calibration screen, enter the high and low values supplied with your dummy probes in their respective boxes under External References. Tap the
Check button on the Channel Calibration
screen after both values have been entered. Once this is completed, these values will be populated for every channel and do not need to be re-entered for subsequent calibrations.
3. Select the Channel Number for the desired channel to configure. Make sure this is the channel to which your calibration probes will be attached.
4. Tap the Lab Cal button to begin the low value calibration. The system will check for an active AutoCal before proceeding. The status should read “Connect probe to channel X, then click Read”. Connect your low value dummy probe to the M-Series input on the corresponding channel of the moisture.IQ module.
5. Tap the Read button. Wait up to a minute. When complete, the status will show “Reading complete on channel X*.”
6. Disconnect the low value dummy probe and connect the high value dummy probe. Do not change the external reference values or tap the Check button.
7. Tap Lab Cal again, and wait for the message, “Connect the probe to channel X, then click Read.” Tap
Read and wait 2-3 minutes.
8. When complete, the new High and Low values will appear in their respective boxes under Channel Calibration. You can now exit the Channel Calibration screen with the Return button, or repeat steps 3-8 for any other channel requiring calibration.
64 moisture.IQ User’s Manual
Page 77
Chapter 5. Using the Configuration Menu

5.1.4 Setting the Probe Auto-Cal Schedule

To set the frequency with which the probe Auto-Cal function is run, proceed as follows:
1. From the Probe Configuration screen, tap Auto-Cal Setup button to access the Set Auto-Cal Interval
screen for the desired probe (see Figure 57 below).
Figure 57: Set Auto-Cal Interval Screen
2. Enter the desired number of hours (1 - 730) between Auto-Cal runs.
3. Tap the
Save button to enter the new value into the moisture.IQ, or tap the Cancel button to retain the
previous value.
moisture.IQ User’s Manual 65
Page 78
Chapter 5. Using the Configuration Menu

5.2 Calibrating the Probes

For newly-purchased moisture.IQ systems, GE has already entered the required calibration data for all purchased moisture and oxygen sensors. However, you should verify this data before startup. You will also need to enter new calibration data if you install a probe from your stock or attach a transmitter to the auxiliary inputs.
Note: For hygrometry, you need to enter calibration data only for M Series probes. It is not necessary to enter
calibration data for the Moisture Image Series Probe unless you send the probe back to GE for calibration without its electronics module. If this is the case, enter the calibration data for the re-calibrated probe as described in this section. The moisture.IQ will automatically download the new calibration data into the Moisture Image Series Probe electronics module.

5.2.1 Entering Calibration Data Automatically

For moisture and oxygen probes included with the analyzer, GE pre-installs the corresponding calibration data files. To enter this calibration data, proceed as follows:
1. On the Probe Calibration screen, select the Channel Number where a probe requiring calibration is attached. Then, tap the FIND button to display a list of all included probes by serial number.
2. Scroll the list to find and select the data file matching the probe to be calibrated, and tap the Save button. The calibration data for that probe will be automatically entered into the calibration table.
3. Tap the Check button to use the new data.

5.2.2 Entering Calibration Data Manually

Be sure you have the Calibration Data Sheets that are supplied with each GE probe. Each Calibration Data Sheet consists of a list of data points that must be entered or verified. Each Calibration Data Sheet lists the corresponding probe serial number, as well as the preassigned channel number. Calibration Data Sheets are usually packed inside the case of the corresponding probe.
To enter your calibration data, refer to the menu map in Figure 78 on page 111 and complete the following steps:
1. From the Configuration Menu, tap the Probe Calibration button to open a Probe Calibration screen similar to Figure 58 on page 67. Note that the serial number of the corresponding probe is listed at the
top of the screen for reference.
66 moisture.IQ User’s Manual
Page 79
5.2.2 Entering Calibration Data Manually (cont.)
Figure 58: Probe Calibration Screen
Chapter 5. Using the Configuration Menu
2. Just below the probe serial number, tap the Channel button to select the channel to which the probe is
connected.
3. To the right of the Channel button, tap the Type button to select the type of input connected to that
channel (Hygro, Pressure, Oxygen, Aux 1 or Aux 2).
4. For each input selected, the table below the two buttons includes data entry boxes for 2 to 16 data
points. The four buttons to the right of the table are used to edit and arrange the calibration data as required:
Sort - Tap this button to sort the data rows in ascending order by the left column.
Edit Value - Tap this button and then tap a text box and enter the values from the relevant Calibration Data
Sheet for each point specified, using the calculator keys on the pop-up keypad. Continue entering or checking values for each input until you have completed each channel.
Insert Row - Use this button along with the Remove Row button to organize the rows in the table as desired.
Remove Row - Use this button along with the Insert Row button to organize the rows in the table as desired.
5. After you have finished entering your probe calibration data, tap the
Configuration Menu.
Return button to return to the
moisture.IQ User’s Manual 67
Page 80
Chapter 5. Using the Configuration Menu

5.3 Tagging the Inputs

The moisture.IQ enables users to assign tags to customize the display of the input parameters. The instrument accepts a separate 9-character label for each input. To tag inputs, complete the following steps:
Figure 59: Tag Setup
1. From the Configuration Menu screen, tap the
Tagging button to open the Tag Setup screen.
2. Tap the Channel button and specify the channel for which the Tag should be applied.
3. Tap the Mode button and select either Hygro, Pressure, Temperature, Oxygen, Aux 1 or Aux 2 from the
drop-down list.
4. Tap the text bar to enter the tag for the selected Channel and Mode. Tap the Save button after entering the text, then tap Save to complete the process.
Note: To remove a tag, tap Clear All, then tap Save.
68 moisture.IQ User’s Manual
Page 81
Chapter 5. Using the Configuration Menu

5.4 Entering User Information

Refer to the menu map in Figure 78 on page 111 and to Figure 52 on page 59 (the Configuration Menu screen) to enter the following types of user information:
User Functions (see “Entering User Functions” on page 69)
User Tables (see “Entering User-Defined Tables” on page 71)
User Constants (see “Entering User Constants” on page 72)
Saturation Constants (see “Entering Saturation Constants” on page 73)

5.4.1 Entering User Functions

User Functions permit the programming up to four mathematical equations on each channel. Any parameter on any channel can be used to calculate a different parameter. To enter a new user function or to edit an existing function, complete the following steps:
1. Tap the User Function button on the Configuration Menu screen (see Figure 52 on page 59) to open the
User Function Setup screen (see Figure 60 below).
Figure 60: User Function Setup Screen
2. Tap the
moisture.IQ User’s Manual 69
Channel button to select the channel where the function will be applied.
Page 82
Chapter 5. Using the Configuration Menu
5.4.1 Entering User Functions (cont.)
3. Tap the User Function button to select the function name (Func 1, Func 2, Func 3 or Func 4). Then, tap the large text box just below the Channel button to open the Channel x:User Function y screen, as shown in Figure 61 below, to enter the function equation. When the equation is complete, tap the Check button to return to the User Function Setup screen.
Figure 61: User Function Equation Screen
The buttons in Figure 61 above permit the inclusion of standard mathematical functions, lookup data from predefined user tables (see “Entering User-Defined Tables” on page 71). The
Check Math button is used to find
any errors in the equation definition. Also, you may tap the Clear All button at any time to start over. The
Copy and Paste buttons at the top right corner of the screen may be used to duplicate information from one
user function to another to avoid entering the same information multiple times.
Note: To assist in entering the user function, the current definition is shown at the bottom center of the User
Function Setup screen as the following selections are made.
4. Tap the Function Label text box to open the Edit Function Label screen and use the keyboard to enter the
desired label.
5. Tap the Unit text box to open the Edit Unit Label screen and use the keyboard to enter the desired units.
6. Tap the Decimals text box and enter the desired number of decimal places for the function (1 - 6).
7. Tap the Max and Min text boxes to enter the Valid Range for the function in digits from the keypad.
8. After the functions has been defined, tap the Return button to return to the Configuration Menu.
70 moisture.IQ User’s Manual
Page 83
Chapter 5. Using the Configuration Menu

5.4.2 Entering User-Defined Tables

To support user-defined functions, the moisture.IQ allows up to six user-defined tables (designated as Table A through Table F) of non-linear or empirical data. Up to 10 X-Y pairs can be entered in each table. If a user function inputs an X value, the meter interpolates the corresponding Y value and outputs that value to the function. (The results are extrapolated if the X value exceeds the range of the table.)
From the Configuration Menu screen (see Figure 52 on page 59), tap the Tables button to open a User Table Setup screen similar to Figure 62 below.
Figure 62: User Table Setup Screen
To set up a user table, complete the following steps:
1. Tap the button at the upper left of the screen to select the table name (
Table A through Table F) from the
drop-down list.
2. The table may contain from 2-10 rows. Use the Insert Row and Delete Row buttons to configure the
table with the desired number of rows.
3. To enter or edit the data in the table, tap the desired table cell to highlight it in blue. Then, tap the Edit
Value
button to enter the data for that cell. Repeat this process until all of the data has been entered.
Note: The Copy and Paste buttons can be used to copy data from another table into this new table.
4. After all of the data has been entered, tap the Sort button to be sure the data points are arranged in
ascending order of X values. Then, tap the Check button to save the table and tap the Return button to return to the Configuration Menu.
moisture.IQ User’s Manual 71
Page 84
Chapter 5. Using the Configuration Menu

5.4.3 Entering User Constants

User-defined constants can be substituted for Dewpoint (°C), Temperature (°C) and Pressure (Pa) measurements in all meter calculations. In addition, you can multiply all measured moisture PPMv values by a specified constant (the default multiplier is 1.000).
From the Configuration Menu screen (see Figure 52 on page 59), tap the Constants button to open a User Constant Setup screen similar to Figure 63 below.
Figure 63: User Constant Setup Screen
To enter user constants, complete the following steps:
1. Tap the
Channel button to select the channel for which the constant should be applied.
2. Tap the text box for the Hygro, Temperature and/or Pressure values and enter the desired constant for each measurement type. The specified values will then be used as a constant multiplier for that measurement type in all future meter calculations.
IMPORTANT: Before the moisture.IQ can use the constants entered above, the probe type on the specified
channel must be properly configured for Constant Value (kH, kT or kP), as described in “The Probe Configuration Screen” on page 60.
3. If desired, tap the k x ppmv text box to enter a constant multiplier for all PPMv moisture measurements. Note that no special probe configuration setting is required for the constant multiplier to be used.
4. If you are using a Delta F Oxygen Cell probe and your background gas is not Nitrogen, you must apply a correction factor to all measurements (see “Delta F Oxygen Cell Background Gas Correction Factors” on page 91). To do so, tap the O2 Background Correction text box and change the multiplier from the default value of 1.00 to the required value.
5. After all constants have been entered, tap the Return button to return to the Configuration Menu screen.
72 moisture.IQ User’s Manual
Page 85
Chapter 5. Using the Configuration Menu

5.4.4 Entering Saturation Constants

Henry's Law is applicable to ppmw moisture measurements in all organic liquids. It states that, “At a constant temperature, the amount of a given gas that dissolves in a given type and volume of liquid is directly proportional to the partial pressure of that gas in equilibrium with that liquid.” In other words, PPMw = (Pw/Ps) x Cs. To
calculate ppmw moisture values for an organic liquid with the moisture.IQ, saturation values (Cs), as a function of temperature, must be entered in the Cs Table shown in Figure 64 below.
Figure 64: Cs Table (Henry’s Law) Screen
To enter the Cs values for your application, complete the following steps:
1. Tap the
Channel button and specify the channel for which the Cs curve should be applied.
2. Use the Insert Row and Remove Row buttons to configure the table with up to 10 rows.
3. To enter or edit the data in the table, tap the desired table cell to highlight it in blue. Then, tap the Edit
Value
button to enter the data for that cell.
Figure 65: Selecting the Chemical
moisture.IQ User’s Manual 73
Page 86
Chapter 5. Using the Configuration Menu
5.4.4 Entering Saturation Constants (cont.)
4. Select the Chemical and enter its percentage of the overall composition.
5. Repeat steps 2-4 until all of the data has been entered. After you select Save, the Cs Table will appear.
Note: At least 90% of the total composition must be entered. For any compositions that total less than 100%,
each chemical will be extrapolated to bring the total to 100%.
5.4.4.1 Entering Saturation Constants Manually
Figure 66: Entering Constants Manually
To manually enter the Cs values for your application, complete the following steps:
1. Select Manual Entry.
2. Tap the Channel button and specify the channel for which the Cs curve should be applied.
3. Use the Insert Row and Remove Row buttons to configure the table with up to 10 rows.
4. To enter or edit the data in the table, tap the desired table cell to highlight it in blue. Then, tap the Edit
Value
button to enter the data for that cell. Repeat this process until all of the data has been entered.
Note: The Copy and Paste buttons can be used to copy data from another table into this new table.
5. After all of the data has been entered, tap the Sort button to be sure the data points are arranged in ascending order of Temp C values. Then, tap the Check button to save the table and tap the Return button to return to the Configuration Menu.
74 moisture.IQ User’s Manual
Page 87

Chapter 6. Setting Up Communications

Chapter 6. Setting Up Communications

6.1 Setting Up moisture.IQ Communications

The Comms section of the Configuration Menu is used to set up all communications for the moisture.IQ (see Figure 79 on page 112). It includes the following options:
Serial Port (see “Configuring the Serial Port” below)
Modbus (see “Configuring the Modbus Connection” on page 76)
TCP/IP (see “Connecting to an Ethernet LAN” on page 77)
VNC (see “Configuring a VNC Connection” on page 82)
Web Server (see “Configuring the Web Server” on page 83)
User Management (see “User Management Operations” on page 84)

6.2 Configuring the Serial Port

IMPORTANT: For proper communications, the serial port settings for the moisture.IQ must be the same as those
for the device connected to the moisture.IQ serial port.
To configure the moisture.IQ serial port, complete the following steps:
1. From the Configuration Menu screen (see Figure 52 on page 59), tap the Serial Port button to open a
Serial Port Setup screen similar to Figure 67 below.
Figure 67: Serial Port Setup Screen
2. Tap the
3. Tap the Baud Rate button and select either 9600, 19200, 38400, 57600 or 115200 from the drop-down
list.
moisture.IQ User’s Manual 75
Mode button and select either RS-232 or RS-485 from the drop-down list.
Page 88
Chapter 6. Setting Up Communications
6.2 Configuring the Serial Port (cont.)
4. Tap the Word Size button and select either 7 bits or 8 bits from the drop-down list.
5. Tap the Parity button and select either Even, Odd or None from the drop-down list.
6. Tap the Stop Bits button and select either 1 or 2 from the drop-down list.
7. After all of the settings have been configured, tap the Check button to save your settings and then tap
the Return button to return to the Configuration Menu.
Note: For RS-485 wiring, refer to Figure 83 on page 116.

6.3 Configuring the Modbus Connection

To configure the moisture.IQ Modbus connection, complete the following steps:
1. From the Configuration Menu screen (see Figure 52 on page 59), tap the Modbus button to open a Modbus Setup screen similar to Figure 68 below.
Figure 68: Modbus Setup Screen
2. If the
3. Tap the second button to choose Modbus/RTU or Modbus/TCP.
4. For RTU, tap the Slave Address button and use the keypad to enter an address from 1 to 247 (the
5. Tap the Disabled button to Enable the Modbus connection.
6. After the Modbus connection has been configured, tap the Check button to save your settings. Then, tap
Note: For a complete Modbus register map, see Table 13 on page 127.
76 moisture.IQ User’s Manual
Enabled/Disabled button indicates that the Modbus connection is currently Enabled, tap this
button to Disable the Modbus connection (changes to the settings can only be made when the connection is Disabled).
default setting is 1). For TCP, tap the Port button and use the keypad to enter a port from 81 to 65535 (the default setting is 502).
the Return button to return to the Configuration Menu.
Page 89
Chapter 6. Setting Up Communications

6.4 Connecting to an Ethernet LAN

The moisture.IQ Ethernet port supports the 10Base-T and 100Base-TX twisted-pair Ethernet standards. An
RJ-45 female modular connector is provided on the moisture.IQ rear panel for connection to the LAN.
WARNING! he Ethernet capability provided in the moisture.IQ is designed for use on a limited-access
Local Area Network (LAN) protected by a suitable firewall. It should not be operated if exposed to the Internet or other unmanaged Wide Area Network (WAN). Check with your network administrator to determine whether it is safe to connect the moisture.IQ to your network infrastructure.
WARNING! As shipped from the factory, there are two accounts enabled, with default passwords assigned. It is highly recommended that the default passwords be changed before placing the moisture.IQ into service.
Note: This chapter assumes that you have read Chapters 1 and 2, and are familiar with the operation and
installation of the moisture.IQ electronics and user interface.
Note: Contact your network administrator to obtain the proper information to use in the following section.

6.4.1 Configuring the Ethernet TCP/IP Connection

The moisture.IQ can either be configured with a static Internet Protocol (IP) address or it can use the Dynamic Host Configuration Protocol (DHCP) to obtain an address from a DHCP server or router. No further address
configuration is required. To configure the moisture.IQ TCP/IP connection, complete the following steps:
1. From the Configuration Menu screen (see Figure 52 on page 59), tap the TCP/IP button to open a TCP/IP
Setup screen similar to Figure 69 below.
Figure 69: TCP/IP Setup Screen
moisture.IQ User’s Manual 77
Page 90
Chapter 6. Setting Up Communications
6.4.1 Configuring the Ethernet TCP/IP Connection (cont.)
2. If the LAN Access button indicates that the TCP/IP connection is currently Enabled, tap this button to
Disable the TCP/IP connection (changes to the settings can only be made when the connection is Disabled). Note that the moisture.IQ MAC Address is shown as a read-only value.
3. Tap the IP Address button and select either Automatic or Static from the drop-down list. If Static is selected, enter values for the IP Address in dotted decimal notation (e.g., 192.168.1.123), the Subnet mask and the Gateway address in the text boxes provided.
4. Tap the DNS button and select either Automatic or Static from the drop-down list. If Static is selected, enter values for the DNS and Alternate DNS in the text boxes provided.
Note: For troubleshooting purposes, automatically assigned addresses are visible on this screen.
5. After the TCP/IP connection has been configured, tap the Check button to save your settings. Then, tap the Return button to return to the Configuration Menu.

6.4.2 Capabilities of the Ethernet Connection

IMPORTANT: The Web Server opens Port 80 when it is Enabled.
1
VNC
IMPORTANT: The
The moisture.IQ allows two methods of remote access and control:
connection opens port 5900 (user-configurable) when Enabled.
An embedded Web Server allows access to the moisture.IQ status and files using a standard web
browser.
A Virtual Network Computing (VNC) server allows full remote control of the moisture.IQ using any of
several VNC clients.
Note: Both the Web Server and the VNC Server are disabled by default.
Access to the web server requires a user name and password. The moisture.IQ provides two accounts for configuration and general administration. Up to three additional accounts can be created for use as needed. Each account has a set of permissions that can be configured to limit network access to moisture.IQ features. The two pre-defined accounts are:
Admin
Operator
1 VNC® is a registered trademark of RealVNC Ltd
78 moisture.IQ User’s Manual
Page 91
Chapter 6. Setting Up Communications
6.4.2.1 The Admin Account
The Admin account allows complete configuration of the Ethernet option. This account should be used only by experienced network administrators. Misuse of the Admin account could prevent connection to the moisture.IQ, enable unauthorized access to the moisture.IQ or possibly expose the network to unauthorized users.
The default login credentials for the Admin account are:
User name: Admin
Password: password
IMPORTANT: It is
Only the Admin can add, modify, or remove the other user accounts. To log in as the Admin, select
Configuration > User Management. to open a Log In screen similar to Figure 70 below.
IMPERATIVE that the Admin password be changed as soon as possible!
Figure 70: Log In Screen
moisture.IQ User’s Manual 79
Page 92
Chapter 6. Setting Up Communications
6.4.2.1 The Admin Account (cont.)
Enter your password and tap the Log In button. On a successful login, the Web User Management screen shown in Figure 71 below is opened.
Tap the
Figure 71: Web User Management Screen
Change Password button to enter and confirm a new password for the Admin account:
CAUTION! Record the new Admin password immediately and keep it in a safe and secure location. It
is NOT POSSIBLE to obtain a lost or forgotten Admin password!
80 moisture.IQ User’s Manual
Page 93
Chapter 6. Setting Up Communications
6.4.2.2 The Operator Account
The Operator account is provided to for day-to-day management of the moisture.IQ. By default, the Operator has all the same privileges as the Admin, but cannot create or modify other accounts. The Operator account allows the user to:
Enable/Disable DHCP
Specify a static IP, subnet, and gateway IP address
Modify some of the serial protocol settings
View the serial port connection status
View the TCP/IP connection status
View the Ethernet option firmware version, serial port and network diagnostics.
The default login credentials for the
Operator account are:
User name: Operator
Password: password
The
Operator password should be changed at the earliest opportunity. The steps for changing the Operator
password are the same as for changing the Admin password (see previous section). Select the Operator account from the Web User Management screen (see Figure 71 on page 80), then enter and confirm the new password.
Record the new operator password immediately, and keep it in a safe and secure location.
Note: If the operator password is lost, it can be reset using the Admin account.
moisture.IQ User’s Manual 81
Page 94
Chapter 6. Setting Up Communications

6.5 Configuring a VNC Connection

IMPORTANT: The VNC1 connection opens port 5900 (user-configurable) when Enabled.
Note: The moisture.IQ system cycle time may be increased by VNC usage.
A Virtual Network Console (VNC) connection allows remote access to the moiture.IQ by replicating the touchscreen on a PC monitor, tablet or smart phone. To configure a VNC connection, complete the following steps:
1. From the Configuration Menu screen (see Figure 52 on page 59), tap the VNC button to open a VNC Setup screen similar to Figure 72 below.
Figure 72: VNC Setup Screen
2. If the
VNC Server button indicates that the VNC Server is currently Enabled, tap this button to Disable
the VNC Server (changes to the settings can only be made when the VNC Server is Disabled).
3. Tap the Listen on Port button and enter the appropriate Port Number. Then, tap the Check button to save the value and tap the Return button to return to the VNC Setup screen. In most cases, the default
Port 5900 should be used.
Note: The Password and Accept Connections From buttons are not functional in this software version.
When a VNC client connection is established, the analyzer will display a Log In prompt. Log in using the credentials for the Admin, Operator or additional account that is configured to permit VNC connections. After a
VNC connection has been established, the analyzer will display a “flashing eye” cursor that tracks the
remote user’s actions. This indicates to a local user that a remote user is connected. When the remote user disconnects, the “flashing eye” cursor is removed.
1 VNC® is a registered trademark of RealVNC Ltd
82 moisture.IQ User’s Manual
Page 95
Chapter 6. Setting Up Communications

6.6 Configuring the Web Server

IMPORTANT: The Web Server opens Port 80 when it is Enabled.
The moisture.IQ can be remotely accessed using a web browser by enabling the Web Server. To configure the
Web Server, complete the following steps:
1. From the Configuration Menu screen (see Figure 52 on page 59), tap the Web Server button to open a
Web Server Setup screen similar to Figure 73 below.
Figure 73: Web Server Setup Screen
2. If the
Web Server button indicates that the Web Server is currently Enabled, tap this button to Disable
the Web Server (changes to the settings can only be made when the Web Server is Disabled).
3. Tap the Port button and enter the appropriate Port Number. Then, tap the Check button to save the
value and tap the Return button to return to the Web Server Setup screen. In most cases, the default
Port 80 should be used.
4. Tap the Web Server button so that it indicates that the server is either Enabled or Disabled, as desired.
5. After the Web Server status has been configured as desired, tap the Return button to return to the
Configuration Menu.
When the Web Server is Enabled, it can be used to monitor the instrument status and to perform file downloads.
moisture.IQ User’s Manual 83
Page 96
Chapter 6. Setting Up Communications

6.7 User Management Operations

The User Management option allows you to set up and control access to the moisture.IQ for Web Server and VNC users (see previous sections). Tap the button for the designated communication option to enter
parameters. to use this option, proceed as follows:
1. From the Configuration Menu screen (see Figure 52 on page 59), tap the User Management button to open the Log In screen (see Figure 70 on page 79).
2. Tap the User button and select either Admin or Operator from the drop-down list.
3. Tap the Password text box and use the keypad to enter your assigned password. Then, tap the Check
button to enter your password.
4. Tap the Log In button to open the Remote User Management screen (see Figure 71 on page 80).
5. To add a new user (up to three total) tap the Add User button in the top right corner of the screen. Then,
complete the User, Password and Confirm Password text boxes. When finished, tap the Check button and then the Return button.
6. To Set User Access, tap the User button and select the desired user from the drop-down list. Then, use the Remove and Change Password buttons to configure this user account as desired.
7. Use the two buttons at the bottom of the screen (Change Settings and Remote Access) to set the permissions for the selected user. Buttons highlighted in blue indicate that the user has access to those operations.
8. When you are finished, tap the Return button twice to return to the Configuration Menu.
84 moisture.IQ User’s Manual
Page 97
Chapter 6. Setting Up Communications

6.8 Setting Up a Remote Connection to a PC

As a practical example of using the communications features described in the previous sections, the procedure for connecting the moisture.IQ to a remote PC via a Local Area Network (LAN) is described below.
1. Ensure that the PC is properly connected to the LAN and that the user has logged in.
2. Refer to “Configuring the Ethernet TCP/IP Connection” on page 77, and ensure that the TCP/IP
connection is Enabled and that the IP Address is set to Automatic.
3. Refer to “Configuring a VNC Connection” on page 82, and ensure that the VNC Server is Enabled.
4. Use a standard ethernet cable to connect the ethernet port on the back of the moisture.IQ to the LAN.
Then, open the TCP/IP Setup screen and record the IP Address shown in the text box.
5. From the PC, go to the RealVNC web site and download the VNC Viewer. Then, launch the VNC Viewer
and enter the recorded moisture.IQ IP Address when prompted (see Figure 74 below). Click the
Connect button.
Figure 74: VNC Viewer Screen
6. A Log In screen should now be open on the moisture.IQ. Enter your
Password and Log In. The
moisture.IQ display should now be replicated on your PC monitor.
7. You can now control the moisture.IQ remotely by using your mouse to click on the display buttons. Note
that keyboard input from the PC is not accepted by the moisture.IQ.
moisture.IQ User’s Manual 85
Page 98
Chapter 6. Setting Up Communications
[no content intended for this page]
86 moisture.IQ User’s Manual
Page 99

Chapter 7. Maintenance

Chapter 7. Maintenance
WARNING! Do not attempt maintenance of the moisture.IQ beyond the instructions in this section. If
you do, you may damage the unit and void the warranty.
This chapter covers the following topics:
“Delta F Oxygen Cell Electrolyte” on page 87
“Replacing and Recalibrating Moisture Probes” on page 89
“Calibrating the Delta F Oxygen Cell” on page 89
“Delta F Oxygen Cell Electrolyte” on page 87

7.1 Delta F Oxygen Cell Electrolyte

As a result of operating the moisture.IQ, particularly when monitoring dry gases, there may be a gradual loss of water from the electrolyte. The electrolyte level should be checked at regular intervals to ensure that the cell is always operating properly. This section describes how to check and replenish the electrolyte in the oxygen cell.
Note: Some applications require that the electrolyte be changed periodically. Consult BHGE.
moisture.IQ User’s Manual 87
Page 100
Chapter 7. Maintenance

7.1.1 Checking the Electrolyte Level

Using the level indicator window on the oxygen cell, check the electrolyte level. The electrolyte should cover about 60% of the window. The electrolyte level should be between the Min and Max lines, as shown in Figure 75 below.
When the Delta F Oxygen Cell connections are made to the moisture.IQ, the analyzer can detect the electrolyte level and indicate when it is low. When the electrolyte level is low, all oxygen readings are highlighted in amber and an “Oxygen Fluid Low” error message is displayed beneath the measurements.
Level
Indicator
x
x
a
a
M
M
n
i
M
Figure 75: Electrolyte Level for the Delta F Oxygen Cell

7.1.2 Replenishing the Electrolyte

Once the oxygen cell receives the initial charge of electrolyte, you should monitor the level regularly. DO NOT let the fluid level drop below the MIN level mark on the window.
WARNING! The electrolyte contains a strong caustic component and can be harmful if it comes in contact with the skin or the eyes. Follow proper procedures for handling the caustic (potassium hydroxide) solution. Consult your company safety personnel.
To raise the fluid level in the reservoir, add add the distilled water, making sure you do not overfill the reservoir. The electrolyte mixture should cover approximately 60% of the level indicator window.
88 moisture.IQ User’s Manual
DISTILLED WATER slowly in small amounts. Check the level as you
Loading...