Baker Hughes reserves the right to make changes in specifications and features shown herein, or discontinue the product
described at any time without notice or obligation. Contact your BHGE representative for the most current information. The
Baker Hughes logo is a trade mark of Baker Hughes, a GE company. The GE Monogram is a trademark of the General Electric
Company.
Thank you for purchasing your moisture.IQ from Baker Hughes, a GE Company. Please register your product at
www.gemeasurement.com/productregistration for product support such as the latest software/firmware
upgrades, product information and special promotions.
Services
BHGE provides customers with an experienced staff of customer support personnel ready to respond to
technical inquiries, as well as other remote and on-site support needs. To complement our broad portfolio of
industry-leading solutions, we offer several types of flexible and scalable support services including: Training,
Product Repairs, Extended Warranties, Service Agreements and more. Please visit
www.gemeasurement.com/services for more details.
Terms and Conditions
Sales Terms and Conditions for your recent purchase of a BHGE product, including the applicable product
Warranty, can be found on our website at the following link:
www.gemeasurement.com/sales-terms-and-conditions
Typographical Conventions
Note: These paragraphs provide information that provides a deeper understanding of the situation, but is not
essential to the proper completion of the instructions.
IMPORTANT: These paragraphs provide information that emphasizes instructions that are essential to proper
setup of the equipment. Failure to follow these instructions carefully may cause unreliable
performance.
CAUTION! This symbol indicates a risk of potential minor personal injury and/or severe damage to the
equipment, unless these instructions are followed carefully.
WARNING! This symbol indicates a risk of potential serious personal injury or death, unless these
instructions are followed carefully.
Safety Issues
WARNING! It is the responsibility of the user to make sure all local, county, state and national codes,
regulations, rules and laws related to safety and safe operating conditions are met for each installation.
Attention European Customers! To meet CE Marking requirements for all units intended for use in
the EU, all electrical cables must be installed as described in this manual.
moisture.IQ User’s Manualvii
Preface
Auxiliary Equipment
Local Safety Standards
The user must make sure that he operates all auxiliary equipment in accordance with local codes, standards,
regulations, or laws applicable to safety.
Working Area
WARNING! Auxiliary equipment may have both manual and automatic modes of operation. As
equipment can move suddenly and without warning, do not enter the work cell of this equipment during
automatic operation, and do not enter the work envelope of this equipment during manual operation. If
you do, serious injury can result.
WARNING! Make sure that power to the auxiliary equipment is turned OFF and locked out before you
perform maintenance procedures on this equipment.
Qualification of Personnel
Make sure that all personnel have manufacturer-approved training applicable to the auxiliary equipment.
Personal Safety Equipment
Make sure that operators and maintenance personnel have all safety equipment applicable to the auxiliary
equipment. Examples include safety glasses, protective headgear, safety shoes, etc.
Unauthorized Operation
Make sure that unauthorized personnel cannot gain access to the operation of the equipment.
viiimoisture.IQ User’s Manual
Preface
Environmental Compliance
RoHS
The moisture.IQ fully complies with RoHS regulations (Directive 2011/65/EU).
Waste Electrical and Electronic Equipment (WEEE) Directive
BHGE is an active participant in Europe’s Waste Electrical and Electronic Equipment (WEEE) take-back initiative
(Directive 2012/19/EU).
The equipment that you bought has required the extraction and use of natural resources for its production. It
may contain hazardous substances that could impact health and the environment.
In order to avoid the dissemination of those substances in our environment and to diminish the pressure on
the natural resources, we encourage you to use the appropriate take-back systems. Those systems will reuse or
recycle most of the materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please contact your local or
regional waste administration.
Please visit
more information about this initiative.
http://www.gemeasurement.com/environmental-health-safety-ehs for take-back instructions and
moisture.IQ User’s Manualix
Preface
[no content intended for this page]
xmoisture.IQ User’s Manual
Chapter 1. Installation
Chapter 1.Installation
1.1Introduction
Users typically install the moisture.IQ as part of a complex process system, which includes components such as
filters, pumps, and pressure regulators. In such an environment, probes and other parts of the system may be
subjected to environmental hazards, such as high temperature, pressure extremes, corrosive elements and
mechanical vibrations.
This section contains information and instructions for installing the moisture.IQ into a process system, while
taking into account all of the above factors. The following section describes how to set up and connect the
moisture.IQ.
If you have questions about the installation procedures, contact our technical support department. See the rear
cover of this manual for contact information.
1.2Unpacking the moisture.IQ
Before beginning the installation, unpack the unit and make sure all the parts and documentation listed on the
packing slip are included. Be sure to inspect each piece of equipment, including the sample system, for
evidence of mishandling. If anything has been damaged or is missing, report this to the carrier and to BHGE
immediately.
1.3Installing the moisture.IQ
To install the moisture.IQ, refer to the drawing for your specific configuration (benchtop, rack mount, panel
mount, weatherproof or explosion-proof) in the Certification and Safety Statements at the end of this manual.
WARNING! To ensure the safe operation of this unit, you must install and operate the moisture.IQ as
described in this user’s manual. In addition, be sure to follow all applicable safety codes and regulations
for installing electrical equipment in your area.
WARNING! For Zone 2 Stainless Steel: Do not rub front face with a dry cloth within the hazardous
area as a spark may occur.
moisture.IQ User’s Manual1
Chapter 1. Installation
1.4Choosing an Installation Site
You should have discussed environmental and installation factors with a BHGE applications engineer or field
sales person by the time you receive the meter. The equipment provided should be suited to your application
and installation site.
The moisture.IQ is available in rack-mount, bench top, or panel-mount versions that are suitable for most
indoor installations. BHGE also provides weatherproof and explosion-proof housings for outdoor and
hazardous area locations. See the drawings in the Certification and Safety Statements at the end of this manual
for details of each enclosure.
Before installing the unit, read the guidelines below to verify that you have selected the best installation site.
1.4.1General Site Guidelines
•Choose an installation site for the probes and sample systems that is as close to the process line as
possible. Avoid long lengths of connecting tubing. If long distances are unavoidable, a fast sampling
by-pass loop is recommended. Do not install any other components, such as filters, upstream from the
probes or sample system unless instructed to do so by BHGE.
•Observe all normal safety precautions. Use the probes within their maximum pressure and
temperature ratings.
•Although the moisture.IQ may not need to be accessed during normal operation, install the electronics
unit at a convenient location for programming, testing and servicing. A control room or instrument
shed are typical locations.
•Locate the electronics unit away from high temperatures, strong electrical transients, mechanical
vibrations, corrosive atmospheres, and any other conditions that could damage or interfere with the
meter operation. See Chapter 9. “Specifications” on page 99 for environmental limits.
•Protect the probe cables from excessive physical strain (e.g., bending, pulling, twisting, etc.).
•Observe the proper cable restrictions for the probes. You can locate Moisture Image Series probes up to
915 m (3000 ft) from the electronics unit with unshielded twisted-pair cable. M-Series probes can be
located up to 600 m (2000 ft) from the unit with the use of specially shielded cable.
1.4.2Low Voltage Directive
If the power mains switch on the unit remains accessible to the user after installation a disconnect device is
not required. However, if installation blocks access to the power mains switch, compliance with the EU Low
Voltage Directive (IEC 61010) requires an external power disconnect device such as a switch or circuit breaker.
The disconnect device must be marked as such, clearly visible, directly accessible, and located within
1.8 m (6 ft) of the unit.
2moisture.IQ User’s Manual
Chapter 1. Installation
1.5Moisture Probe Considerations
The M-Series and Moisture Image Series probes consist of an aluminum oxide sensor mounted on a connector
head. Standard probe mounts include a protective stainless steel shield.
The probe sensor materials and housing maximize durability and insure a minimum of water absorbing
surfaces in the vicinity of the aluminum oxide surface. A sintered stainless steel shield is used to protect the
sensor from high flow rates and particulate matter. The end cap should not be removed except upon advice
from BHGE.
The sensor has been designed to withstand normal shock and vibration. You should make sure that the active
sensor surface is never touched or allowed to come into direct contact with foreign objects, since this may
adversely affect performance.
Observing these few simple precautions will result in a long and useful probe life. BHGE recommends that
probe calibration be checked routinely, at 12-month intervals, or as recommended by our applications
engineers for your particular application.
The probe will measure the water vapor pressure in its immediate vicinity. Therefore, readings will be
influenced by its proximity to the system walls, materials of construction, and other environmental factors.The
sensor can be operated under vacuum or pressure and under flowing or static conditions.
See the following sections for the recommended environmental precautions.
1.5.1Temperature Range
See “Probe Specifications” on page 105, for details.
1.5.2Moisture Condensation
Be sure the temperature is at least 10°C higher than the dew/frost point temperature. If this condition is not
maintained, moisture condensation could occur on the sensor or in the sample system, which will cause
reading errors. If this happens, dry out the probe following the procedures outlined in BHGE document
#916-064, Basic Hygrometry Principles.
1.5.3Static or Dynamic Use
The sensor performs equally well in still air or where considerable flow occurs. Its small size makes it ideal for
measuring moisture conditions within completely sealed containers or dry boxes. It will also perform well
under gas flow conditions as high as 10,000 cm/sec and liquid flow conditions up to 10 cm/sec. Refer to BHGE
document #916-064, Basic Hygrometry Principles, for maximum flow rates in gases and liquids.
moisture.IQ User’s Manual3
Chapter 1. Installation
1.5.4Pressure Range
The moisture probe always senses the correct water vapor pressure regardless of the total ambient pressure.
The moisture sensor measures water vapor under vacuum or high pressure conditions from as little as a few
microns of Hg to as high as 5000 psi total pressure.
1.5.5Long-Term Storage & Operational Stability
Sensors are not affected by continuous abrupt humidity changes or damaged by exposure to saturation
conditions even when stored. However, you should store probes in their original shipping containers in a clean,
dry location. If the probe is saturated during storage, refer to “Moisture Condensation” on page 3 before
installing the probe. For best performance, do not store probes longer than one to two years from their
calibration date.
1.5.6Interference Effects
The sensor is completely unaffected by the presence of a wide variety of gases or organic liquids. Large
concentrations of hydrocarbon gases, Freon
effect on sensor water vapor indications. The sensor will operate properly in a multitude of gaseous or
non-conductive liquid environments.
, ozone, carbon dioxide, carbon monoxide, and hydrogen have no
1.5.7Corrosive Materials
Avoid all materials that are corrosive or otherwise damaging to aluminum or aluminum oxide. These include
strongly acidic or basic materials and primary amines.
4moisture.IQ User’s Manual
Chapter 1. Installation
1.6Sample System Guidelines
A sample system is required for oxygen measurement and, although not mandatory, is highly recommended
for moisture measurement. The purpose of a sample system is to condition or control a sample stream to
within the specifications of a probe. The application requirements determine the design of the sample system.
BHGE applications engineers will make recommendations based on the following general guidelines.
1.6.1Moisture Sample Systems
Typically, sample systems should be kept very simple. They should contain as few components as possible and
all or most of those components should be located downstream of the measurement probe. Figure 1 below
shows an example of a basic sample system consisting of an explosion-proof housing with a sample cell, a
filter, a flow meter, a vent valve, an inlet shut-off valve and an outlet shut-off valve.
The sample system components should not be made of any material that will affect measurements. A sample
system may include a filter to remove particulates from the sample stream or a pressure regulator to reduce or
control the pressure of the stream. However, most common filters and pressure regulators are not suitable for
sample systems because they have wetted parts that may absorb or release components (such as moisture) into
the sample stream. They may also allow ambient contamination to enter the sample stream. In general, you
should use stainless steel construction for all wetted parts. Contact BHGE for further instructions.
Sample
Ce ll
Sample
Inle t
Vent
Figure 1: A Typical Moisture Sample System
Sample
Outlet
Note: The actual sample system design is dependent on the application requirements.
moisture.IQ User’s Manual5
Chapter 1. Installation
1.6.2Oxygen Sample Systems
Oxygen sample systems are required and can be ordered from BHGE for bench or wall mounting. You can also
build your own sample system by using the following guidelines.
IMPORTANT: The BHGE warranty will be voided if the sample system does not have a relief valve.
The basic sample system requirements are as follows (see Figure 2 below):
•The oxygen cell requires a sample gas flow of 0.5 to 1 LPM (1 to 2 SCFH).
•The gas pressure in the sample cell must be between 0.0 and 1.0 psig. The pressure must not exceed
1.0 psig.
•A 10 psig pressure relief valve installed upstream of the oxygen cell is required to prevent
over-pressure.
•A flow meter is required to measure the flow.
• A pressure gauge is required to measure the pressure.
• A flow regulating or needle valve is required and should be located upstream of the sample cell.
• A pressure regulator is required for sample gas supplies of 50 psig or greater.
If a sample pump is required to draw a sample to the oxygen cell, the pump should be installed downstream
from the oxygen cell. You must also install a vacuum relief valve set at 1.0 psig between the oxygen cell and the
pump.
Sam ple
Sam ple
Inlet
Outlet
Figure 2: A Typical Oxygen Cell Sample System
Note: The actual sample system design is dependent on the application requirements.
6moisture.IQ User’s Manual
Chapter 1. Installation
1.7Mounting the Hygrometer System
Mounting the hygrometer system consists of mounting the electronics unit, the probes, and the sample
systems.
1.7.1Mounting the Electronics Unit
Use the outline and dimension drawings in the Certification and Safety Statements at the end of this manual to
mount the moisture.IQ. These drawings provide clearance and other mounting dimensions needed to prepare
the site for mounting.
WARNING! If the power mains switch on the unit remains accessible to the user after installation, a
disconnect device is not required. However, if the installation blocks access to the power mains switch,
compliance with the EU's Low Voltage Directive (IEC 61010) requires an external power disconnect
device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly visible,
directly accessible, and located within 1.8 m (6 ft) of the unit.
Be sure to follow the guidelines outlined in “Choosing an Installation Site” on page 2 before mounting the
enclosure.
Note: If the installation location does not provide enough room for connections to be made easily after
installation, you may want to make connections to the MIS or M-Series probes, Delta F Oxygen Cell,
Outputs and Alarms before mounting the instrument.
1.7.2Mounting a Sample System
BHGE sample systems are normally fastened to a metal plate that has four mounting holes. BHGE can also
provide the sample system in an enclosure, if requested. In either case, fasten the sample system plate or
enclosure with bolts in the mounting holes at each of the four corners. If you ordered sample system outline
and dimension drawings, they will be included in your shipment.
Connect the sample system inlet to the process and the outlet to the return, using the appropriate fittings or an
appropriate NPT adapter.
CAUTION! Do not start a flow through the sample system until all probes and transmitters are
properly installed.
moisture.IQ User’s Manual7
Chapter 1. Installation
1.7.3Mounting the Oxygen Cell
If your oxygen cell is not mounted in a sample system, refer to the Figure 3 below for mounting details.
5.90
(149.9)
TEMPERATURE SENSOR
MAX
MIN
CONN
.40
5.00
(127.0 MAX)
CONN
0.25
(6.4)
MAX
0.55
(14.0)
0.62
(15.8)
1.06
(26.9)
1.50
(38.1)
0.37
(9.4)
1.23
(31.2)
1.05
(26.7)
2.05
(52.1)
Electrolyte
Electrolyte
2.10
(53.3)
reservoir
level
window
Sample
gas
inlet
Sample
gas
outlet
MAX
MIN
SERVOMEX
2.80
(71.1)
4.91
(124.7)
Figure 3: Oxygen Cell Dimensions
8moisture.IQ User’s Manual
Chapter 1. Installation
1.8Installing the Probes
After you mount the sample system, you must insert the moisture probes into the sample cells. In addition, you
must check, prepare, and connect any installed oxygen cells to the gas line.
1.8.1Moisture Probes
BHGE Moisture Image Series (MIS) and M-Series moisture probes have 3/4 inch-16 straight threads, with an
O-ring seal, to the secure probes into the sample system or directly into the process line. Other fittings are
available for special applications.
CAUTION! If mounting the moisture probes directly into the process line, you must consult BHGE for
proper installation instructions and precautions.
Moisture probes are usually installed into a sample system. The sample system protects the probes from
coming into contact with damaging elements in the process. Moisture probes are installed into a cylindrical
shaped container, called the sample cell, which is included as part of your sample system. The sample cell is
identified by a label on the sample system plate.
To install a moisture probe that has a 3/4 inch-16 straight thread, insert it into the sample cell so it is
perpendicular to the sample inlet. Screw the probe into cell, making sure not to cross thread it, and tighten it
securely. Figure 4 below shows a typical moisture probe installation, with the probe mounted in a BHGE
sample cell. Install any moisture probes with different fittings in the appropriate manner.
Probe
Inlet
Sample Cell
Outlet
Figure 4: Typical Moisture Probe in Sample Cell
Note: Standard moisture probes have a sintered stainless steel shield that protects the aluminum oxide sensor.
Be sure to leave the shield in place for maximum protection.
It is important to eliminate all gas or liquid leaks for safety reasons and to ensure that measurements are not
affected by ambient contamination. For gas applications, you should check for leaks using a soap bubble
solution.
IMPORTANT: Refer to the Calibration Data Sheets to connect the probes to their corresponding channels. If a
probe is connected to the wrong channel, the meter will display incorrect data.
moisture.IQ User’s Manual9
Chapter 1. Installation
1.8.2Adapting an Existing TF-Series Probe Cable
If an existing moisture measurement system uses a TF-Series moisture probe, the cable for that probe may be
adapted for use with an MISP2 probe. To retrofit the existing system, refer to Figure 5 below and complete
the following steps:
MISP2 Connector
[Total Cable Length = 6.50”]
TF Connector
Figure 5: TF-MISP2 Converter Cable
1. Locate the existing TF-Series probe and disconnect the cable from the probe.
2. Replace the TF-Series probe with an MISP2 probe.
3. Using BHGE adapter cable 704-1362-00, shown in Figure 5 above, connect one end of the cable to the
MISP2 probe and connect the other end to the existing TF-Series cable connector.
4. At the moisture analyzer end of the TF-Series cable, locate the same TF-Series cable and disconnect all
flying leads from the pluggable connector.
5. Locate the BLACK and WHITE wires of the TF-Series cable and connect them to the moisture.IQ at the
MISP2 probe gray pluggable connector. See “Moisture Image Series Probes (MISP and MISP2)” on
page 19 for help.
6. Trim all other flying leads from the TF-Series cable flush with the jacket.
7. Plug the newly-wired MISP2 probe pluggable connector into the moisture.IQ.
10moisture.IQ User’s Manual
Chapter 1. Installation
1.8.3Adapting an Existing M-Series Probe Cable
If an existing moisture measurement system uses a M-Series moisture probe, the cable for that probe may be
adapted for use with an MISP2 probe. To retrofit the existing system, refer to Figure 6 below and complete
the following steps:
MISP2 Connector
[Total Cable Length = 6.50”]
M Series Connector
Figure 6: M-MISP2 Converter Cable
1. Locate the existing M-Series probe and disconnect the cable from the probe.
2. Replace the M-Series probe with an MISP2 probe.
3. Using BHGE adapter cable 704-1649-00, connect one end of the cable to the MISP2 probe and connect
the other end to the existing M-Series cable connector.
4. At the moisture analyzer end of the M-Series cable, locate the same M-Series cable and disconnect all
flying leads from the pluggable connector.
5. Locate the GREEN and RED wires of the M-Series cable and connect them to the BLACK and WHITE wires
at the MISP2 probe gray pluggable connector. (The GREEN wire is connected to the MISP connector
"BLK" label and RED wire is connected to the MISP connector "WHT" label.) See “Moisture Image Series
Probes (MISP and MISP2)” on page 19 for help.
6. Trim all other flying leads from the M-Series cable flush with the jacket.
7. Plug the newly-wired MISP2 probe pluggable connector into the moisture.IQ.
moisture.IQ User’s Manual11
Chapter 1. Installation
1.8.4Pressure Sensors
If a pressure input is required and if an optional pressure signal is not available from a Moisture Image Series
probe, you can connect a separate pressure transmitter to an auxiliary input.
The moisture.IQ accepts any 0 to 20-mA, 4 to 20-mA, or 0 to 2-V pressure transmitters.
Always mount the pressure transmitter directly downstream from the moisture probe to ensure that the
pressure is measured at the same point as the moisture measurement.
1.8.5Delta F Oxygen Cell
Although the moisture.IQ accepts other oxygen devices as auxiliary inputs, it is designed to accept oxygen
inputs directly from the Delta F Oxygen Cell. There are three steps for installing the Delta F Oxygen Cell:
1. Prepare the oxygen cell for operation
2. Calibrate the oxygen cell.
3. Connect the oxygen cell to the gas line.
The Delta F Oxygen Cell is available with various types of mounting enclosures. However, the cell itself will
always look like the one shown in Figure 7 below.
Electrolyte Level
x
x
a
a
M
M
n
i
M
Figure 7: Delta F Oxygen Cell Electrolyte Level
12moisture.IQ User’s Manual
Chapter 1. Installation
1.8.5.1 Preparing the Oxygen Cell
To prepare the oxygen cell for operation, you must fill it with the electrolyte that was supplied in a plastic
bottle.
WARNING! The electrolyte contains potassium hydroxide that is harmful if it comes in contact with
eyes or skin. Consult your company safety personnel for proper procedures for handling the electrolyte.
1. Unscrew the top cover on the oxygen cell reservoir.
2. Slowly add the entire contents of the electrolyte bottle, approximately three ounces (90 ml), making
sure not to spill any on the outside of the cell. Be especially careful that the electrolyte does not come in
contact with any of the cell electrical connections.
3. Using the min/max window on the oxygen cell, check the electrolyte level. The electrolyte should cover
about 60% of the window (see Figure 7 on page 12). The cell is now ready to be connected to the gas
line.
4. Replace the top cover on the oxygen cell.
Note: After you add the electrolyte, DO NOT add additional electrolyte to the reservoir. If the level falls below the
minimum level, refer to “Delta F Oxygen Cell Electrolyte” on page 87 to replenish the cell.
5. Calibrate the oxygen cell as described in “Calibrating the Delta F Oxygen Cell” on page 89. After you
calibrate the oxygen cell, connect it to the gas line as described in the following section.
1.8.5.2 Connecting the Oxygen Sample System to the Gas Line
To connect the oxygen sample system to the process line, attach a 1/8 inch O. D. (outside diameter) tube to the
1/8 inch sample gas inlet fitting, using a Swagelokor equivalent mating connector. Avoid using plastic or
rubber for any tubings or fittings which are included in the inlet gas lines.
WARNING! Do not connect the oxygen cell outlet to flow restricting devices, pressure lines, or
vacuum lines. Pressure differentials across the oxygen cell sensor in excess of 1 psig could be harmful or
damage the cell.
If the gas being monitored does not create a safety hazard, vent it to atmosphere at the sample system outlet. If
venting the gas to atmosphere causes a safety hazard, vent the gas to a safe location. Make sure the venting
system does not create a back pressure on the oxygen cell.
Note: The vented sample will not be corrosive if you install and operate the cell properly.
moisture.IQ User’s Manual13
Chapter 1. Installation
1.9Making Electrical Connections
WARNING! To ensure the safe operation of this unit, you must install and operate the moisture.IQ as
described in this user’s manual. In addition, be sure to follow all applicable safety codes and regulations
for installing electrical equipment in your area.
WARNING! Turn off the moisture.IQ before making any electrical connections.
WARNING! The stainless steel Weatherproof enclosure must be properly grounded.
To open a Weatherproof moisture.IQ unit, use a large, flat-bladed screwdriver to turn the latch on the enclosure
cover.
To open an Explosion-Proof moisture.IQ unit, remove all 24 bolts from around the periphery of the enclosure
cover. Do not open an Explosion-Proof moisture.IQ unit that is energized.
To facilitate wiring, Weatherproof and Explosion-Proof moisture.IQ units are equipped with a hinging Chassis
that can be swung out and rested on a rod support. During normal use, the Chassis is fixed in place by a latch
mechanism located at the center of the mounting panel.
To release the Chassis from its locked position, the latch pin at the center of the mounting panel is pushed
upwards (in the direction indicated by the surrounding label). To lock the Chassis in place, the Chassis is
pushed down towards the mounting panel until the latch engages. In order to safely move the Chassis after
cables are installed, service loops for the Weatherproof units shall be at least 20 inches. Service loops for the
Explosion-Proof units shall be at least 18 inches
Make all connections to the back of the meter (see Figure 8 on the next page):
•The probe connections in the left two sections are for Module A - Channels 1, 2, 3.
•The probe connections in the right two sections are for Module B -Channels 4, 5, 6.
•For each Module, the left group of probe connections (MIS, M-SERIES, O2T and OXYGEN) are rated for
hazardous area use, while the right group of probe connections (
non-hazardous area use.
AUX IN/OUT and ALARMS) are rated for
•The connections along the bottom of the panel are for power input and remote devices.
14moisture.IQ User’s Manual
1.9 Making Electrical Connections (cont.)
Chapter 1. Installation
Module A Hazardous Area Connections
ON/OFF Switch
Power Cord Receptacle
Module A Non-Hazardous Area Connections
Module B Hazardous Area Connections
Module B Non-Hazardous Area Connections
Figure 8: Rear Panel Connections
1.9.1Using the Lever to Make Wiring Connections
Use the included levers to assist with wiring the cables to the terminal blocks, as shown in Figure 9 below:
1. Press and hold the lever against the terminal block.
2. Insert the wire into the terminal block.
3. Release the lever to secure the connection.
Figure 9: Using the Press Lock Lever
IMPORTANT: To maintain good contact at each terminal block and to avoid damaging the pins on the connector,
pull the connector straight off (not at an angle). Make the cable connections while the connector is
away from the unit, and then push the connector straight on (not at an angle) when the wiring is
complete.
Proper connections and cabling are extremely important for accurate measurement. Use the correct cable type
for each probe and make sure the cables are not damaged during installation. See the sections that follow for
specific connection instructions.
moisture.IQ User’s Manual15
Chapter 1. Installation
1.9.2Connecting the Input Power
To connect the input power, simply plug the power cable into the receptacle next to the ON/OFF switch at the
lower left corner of the back panel (see Figure 8 on page 15). DC wiring is shown in Figure 9 below.
WARNING! If the power mains switch on the unit remains accessible to the user after installation, a
disconnect device is not required. However, if the installation blocks access to the power mains switch,
compliance with the EU's Low Voltage Directive (IEC 61010) requires an external power disconnect
device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly visible,
directly accessible, and located within 1.8 m (6 ft) of the unit.
DC Power Supply Connector
21
3
+
-
Figure 10: DC Power Wiring
1.9.3Connecting Moisture Probes
The moisture.IQ uses M-Series and Moisture Image Series probes to measure moisture. If you need to connect a
different probe type, please contact our technical support department. If you ordered one or more M-Series
probes with the moisture.IQ, BHGE has entered the necessary probe setup data on a pre-assigned channel.
IMPORTANT: See the Calibration Data Sheets, shipped with the probes, to connect the probes to their
corresponding channels. If a probe is connected to the wrong channel, the meter will display
incorrect data.
Probes are identified on the Calibration Data Sheet by a serial number. The serial number is also engraved on
the hex nut of the probe, as shown in Figure 11 below.
Probe Serial Number
Figure 11: Moisture Probe Serial Number
The Moisture Image Series probe does not require any pre-programming, because it stores all the necessary
setup data in its electronics module. Therefore, you can install a Moisture Image Series probe on any
available channel. After you install the probe, you must activate the probe on the installed channel, as
described in “The Probe Configuration Screen” on page 60. Proceed to the following sections to properly
connect your moisture probes.
16moisture.IQ User’s Manual
Chapter 1. Installation
1.9.3.1 M-Series Probes
M-Series probes are primarily used for moisture measurement, but they can be used to measure temperature
as well. If ordered, an optional temperature thermistor is included in the moisture probe and requires an
additional connection.
If the M-Series probe does not have a thermistor, then you may use a two-wire shielded cable with a
bayonet-type connector to connect the probe to the electronics unit. If an optional thermistor is installed, then
use a four-wire shielded cable with a bayonet-type connector to connect the M-Series Probe to the electronics
unit. The M-Series Probe may be located up to 600 m (2000 ft) from the moisture.IQ.
Before making the electrical connections, connect the cable to the probe by inserting the bayonet-type
connector onto the probe and twisting the shell clockwise until it snaps into a locked position (approximately
1/8 of a turn). Wire the probe cable per the wiring diagram in Figure 12 below. The M-Series terminal block
on the back panel of the moisture.IQ (see Figure 13 on page 18) also notes the corresponding color from the
probe cable.
M-SERIES
54321
Note: The label
used to terminate the cable shield.
SHD
BLACK
WHITE
GREEN
RED
.
Temp
M-SERIES
Figure 12: M-Series Probe Cable Wiring Diagram
SHD in Figure 12 above corresponds to the SH label on the connector, and these terminals are
Moisture
moisture.IQ User’s Manual17
Chapter 1. Installation
1.9.3.1 M-Series Probes (cont.)
M-Series Probe Connections
Module A - CH1
Module A - CH2
Module A - CH3
Figure 13: M-Series Probe Connections
Note: Additional M-Series probes may be connected to CH4, CH5 and CH6 in the Module B connector group on
the right side of the back panel.
If you connect a probe to the wrong channel, you can either reconnect the probe to the assigned channel or
reconfigure the current channel as described in “The Probe Configuration Screen” on page 60.
WARNING! The M-Series probe may not be capable of withstanding the 500V insulation rating. This
must be taken into account in any installation in which it is used.
18moisture.IQ User’s Manual
Loading...
+ 152 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.