GE Medical Systems LUNAR User Manual

g
DOC. ID: DXSE0012
MANUFACTURING PROCEDURES
GE Medical Systems
TITLE: DEXA PM Work Instructions
LUNAR
REVISION: G DATE: 2/28/02 WRITTEN BY: Jeff Franz APPROVED ON CO: L10173

PURPOSE

This procedure describes the steps necessary to do a Planned Maintenance Check on a DEXA scanning system at a customer's site, and to assure that the system is operating within GE-LUNAR's specifications and in compliance with applicable Federal and State regulations.

APPLICATION

This procedure applies to all employees performing Planned Maintenance or Site Certification of DPX Series or PRODIGY, and DPX-NT scanning systems.

SAFETY PRECAUTIONS

When power is applied this equipment generates ionizing radiation. Take precautions that no part of your body passes through the x-ray beam when the equipment is energized.

ASSOCIATED MATERIALS

DXAP0008 DPX-Series Installation Test Procedures DXAP2007 Prodigy Hacksaw Test DXAP3002 DPX-NT Installation Test Procedures DXQC0022 Aluminum Spine Phantom Scan Procedure DXQC2022 enCORE software Aluminum Spine Phantom Scan Procedure DXSE0036 OINK Daughter board and Limit Switch Field Replacement Instructions DXSE0038 DEXA PM Quality record

DEFINITIONS

DPX Series scanners are all non- Prodigy, non- DPX-NT/MD+ scanners. enCORE based systems are all PRODIGY and DPX-NT/MD+ scanners. PRODIGY I systems are defined as PRODIGY systems with system numbers 11999 and lower. DPX-NT (A) systems are defined as DPX NT systems with system numbers 71999 and lower.
QUALITY SYSTEM DOCUMENT Page 1 of 10
It is the responsibility of the user to verify this document originated from the controlled master and is the active revision prior to use.
This document contains confidential or proprietary information of GE Lunar. Neither the document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE Lunar.
GE Medical Systems LUNAR
DOC ID: DXSE0012
REV: G DATE: 2/28/02 CO: L10173

PROCEDURE

A. QA HISTORY AND ERRORLOG.

1. Perform a Quality Assurance test to insure the system is operating. a. Verify that all tests on the QA are passing.
2. DPX Series scanners:
The QA values needed may be found by viewing the QA database (F4 Database options – F2 QA History) and / or in the Daily QA printouts in the Customer’s QA binder.
a. Verify that the Detector has not lost significant resolution by checking that
the high and low energy air counts have not decreased by more than 10% from the date of installation or the date of the last tube head or detector change).
b. Verify that the air ratio has not changed by more than .02. from the date of
installation or last tube head or detector change (calculate from earliest QA or lookup on QA printouts).
3. Systems with enCORE software a. Copy the QA.mdb file to a diskette (Tools / Send Configuration / Select
Error Log and QA Database, make sure compress Files is selected and select Span Multiple Disks if necessary).
b. Select the appropriate drive (3 ½ floppy A: ) insert disk into drive and click
OK.
c. Label with system number and PM, the file may be mailed to Service
Engineering or email file to lunar.support@med.ge.com
NOTE: This is only to gather statistical in formation on enCORE ba sed systems performance, there will be no follow up regarding these QA histories.
4. Analysis of the QA History file a. DPX Series
i. Verify % Spillover is between 6.3 and 9.5
ii. Check the Air Ratio – it cannot deviate more than .02 since the last
detector / tube change
iii. Check the low air counts – a change greater than 10% since the
date of last detector change indicates a failing detector
iv. Check the Large standard Mass – should not deviate more than
2%
QUALITY SYSTEM DOCUMENT Page 2 of 10
It is the responsibility of the user to verify this document originated from the controlled master and is the active revision prior to use.
This document contains confidential or proprietary information of GE Lunar. Neither the document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE Lunar.
GE Medical Systems LUNAR
DOC ID: DXSE0012
v. If any of the above parameters are failing, call GE-Lunar Customer
Support for help diagnosing the system.
1. To aid in diagnosis, email the QA.MDB file to lunar.sppport@med.ge.com
5. Review the error log(s). b. DPX Series systems print from DOS prompt
i. C:\> ERRDUMP –L –N >PRN
1. See DPX Series CSB Vol 9 for DPX Series Error Log
Interpretation c. enCORE based systems Tools\ Error Log – add to instructions d. Verify that there are no hardware subsystem failures (X-ray, Detector,
Mechanical)
e. Contact Lunar Service Department if any unexplained or unusual errors
are discovered in the error log.
i. Error Log may be faxed to 608-826-7107 to aid in interpretation.
REV: G DATE: 2/28/02 CO: L10173
f. Delete error log for DPX Series (C:\LUNAR> del err_320.bin or C:\LUNAR
del errlog.bin).
g. Start a new error log for enCORE-based systems by clicking on Initialize
Log.

B. HOST PC MAINTENANCE

1. Clean the Printer according to the manufacturer’s instructions a. For Hewlett Packard Printers go to www.hp.com and search the support
pages for the customer’s printer model.
2. DPX series systems: a. From a DOS prompt
i. Delete contents of C:\SCRATCH and C:\PRNTMNGR directories.
ii. Delete all *.CHK files in C:\ directory.
3. Software updates a. DPX Series systems: all DPX-IQ densitometers may be updated to
version 4.7e software free of charge.
b. All other (non-IQ) DPX Series Densitometers should be running version
3.65 or 4.7e (if upgraded to IQ software)
c. If the customer is running a version lower than 4.7e, or 3.65 ask them if
they are interested in this free update, they may get by contacting the GE­Lunar Apps line (USA) or you may order it from GE-Lunar at
lunar.support@med.ge.com
or at 866-225-4771.
QUALITY SYSTEM DOCUMENT Page 3 of 10
It is the responsibility of the user to verify this document originated from the controlled master and is the active revision prior to use.
This document contains confidential or proprietary information of GE Lunar. Neither the document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE Lunar.
GE Medical Systems LUNAR
DOC ID: DXSE0012
d. enCORE based systems:
i. Contract and W a rranty customers are entitled to free software
updates (all software versions <6.0 are updates, > 6.0 is considered an upgrade and all customers must purchase it)
ii. Billable customers may purchase software updates by contacting
GE-Lunar at 866-225-4771.

C. SCAN TABLE MAINTENANCE

1. Remove the scanner tabletop. a. Power down the scan table. b. DPX Series tables – unlock the end panel doors and remove the tabletop
screws.
i. DPX-IQ
ii. Non DPX-IQ
c. DPX-NT – unlock the end panels and remove the tabletop screws.
REV: G DATE: 2/28/02 CO: L10173
i.
d. PRODIGY – lift up the table pad and remove the 4 counter sunk 1/8” hex
head screws in the 4 corners of the tabletop.
2. Clean the scanner interior a. Clean the electronics pan of dust and dirt. b. Wipe the dust and dirt off of the rear longitudinal rail and the transverse
rail.
3. Inspect mechanical parts and cables for wear and breakage. a. Check all cables on the cable trough for wear b. Check the v-wheels on the rear longitudinal carriage for appropriate
tension Insert V-wheel instructions.
4. X-Ray Generator a. Power up the scan table. b. Ramp the scanner to the appropriate kV / mA and read the test points on
the appropriate board as indicated in the X-Ray Settings table on DXSE0038.
i. DPX Series
1. Load the version specific diagnostic software from the Tools disk – from a DOS prompt type: A:/diags/diagnos load c:/lunar
2. Enter the diagnostic software – from a C:\LUNAR prompt type: diags
3. Choose option 5 Static Counter
QUALITY SYSTEM DOCUMENT Page 4 of 10
It is the responsibility of the user to verify this document originated from the controlled master and is the active revision prior to use.
This document contains confidential or proprietary information of GE Lunar. Neither the document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE Lunar.
GE Medical Systems LUNAR
DOC ID: DXSE0012
ii. EnCORE based systems
1. Place the scanner in Service Mode
a. Options / User Options / check the check box for
Service – Enter Password (Smile sw ver 1.0 – 2.05 or Award ver 2.10 +)
2. Go to Tools / Diagnostics / Scanner X-ray
3. Enter desired kV and ma – click on ramp
NOTE: Be sure to use your anti-static wrist strap when working on or near circuit board assemblies. .
5. Wedge Test (Rollover Test – DPX Series Only) a. Select the Static Counter from the initial menu. When the static counter
screen appears, use the arrow key to move the cursor to Operating Voltage, and enter 76.0.
b. Move the cursor to Tube Current and enter the number 150.
REV: G DATE: 2/28/02 CO: L10173
c. Set the "Repeat" number to 20 and the "Printer" to ON.
i. If the AGS is operating properly, the low channel counts will be
approximately 600,000 to 800,000 counts, and there will be fewer high channel counts in the range of approximately 400,000 to 700,000. Both counts will be stable to approximately +/- 3000 counts from one sample to the next.
d. Place the aluminum wedge test tool on the top of the collimator, with the
thin edge toward the collimator’s larger aperture, and the thick edge away from the aperture. Insure that the aluminum wedge is NOT in the beam.
e. Press <Esc> and the shutter will open. Allow a 7 samples to appear on
the monitor, and then slowly move the Aluminum Wedge through the radiation beam, starting with the thin edge of the wedge.
f. Move the wedge very slowly until the beam reaches an aluminum
thickness of approximately 2 mm.
g. In the results displayed on the monitor, note the sample number where
the high channel air counts begin to be reduced. At this point the counts in both channels should be reduced for the following samples. In other words: both the high and low channel counts must decrease consistently as the wedge enters the beam.
h. If the low channel counts increase at any point instead of decreasing, the
AGS is not properly adjusted or the detector is failing.
i. This may cause "bone points" or white pixels in the image around
the outside of a patient's Total Body scan and or white streaks in femur scans.
QUALITY SYSTEM DOCUMENT Page 5 of 10
It is the responsibility of the user to verify this document originated from the controlled master and is the active revision prior to use.
This document contains confidential or proprietary information of GE Lunar. Neither the document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE Lunar.
GE Medical Systems LUNAR
DOC ID: DXSE0012
REV: G DATE: 2/28/02 CO: L10173
i. Run the AGS wedge test and verify that it passes attach results to PM
Check sheet.
6. Alignment verification / Mechanical / Cabling check a. Run Air Matrix Scan
i. Remove the tabletop if it is on.
ii. In DIAGS select F4 QUA LITY ASSURANCE TESTS, select option
3 – PERFORM AIR MATRIX SCAN. Choose 1 - Perform A Scan. Change the number of lines to the maximum value, also change the number of steps to the maximum value.
iii. Note the filename here_____________________________. This
will have to be entered to analyze the data (filename format: XM followed by the 4-digit system number followed by MO.01A. Subsequent scans have extensions 01B, 01C, etc.) Then Press Esc to continue.
iv. Remove the detec to r collimator (non-IQ / MD systems only) and
install the Air Matrix Piece (attenuation insert) or tape a 0.25 inch piece of aluminum over the detector (over on top of the collimator) to reduce the intensity of the beam.
v. Press Esc to begin the scan. The scan will take approx. 30
minutes to complete.
NOTE: If anything passes through the beam during this test, IT WILL FAIL.
vi. Analyze the Air Matrix scan in DIAGS by selecting option F4
QUALITY ASSURANCE TESTS then select option 4 ANALIZE AIR MATRIX SCAN.
vii. Enter the Air Matrix file name when prompted and press the enter
key.
viii. All categories mu st pass, see table 1 for Air Matrix values.
Category Limit
Chi Squared 45 Skew +/- 0.07 Kurtosis +/- 0.07
Table 1: Failure Limits for the Air Matrix Test
ix. Print the overall values by answering Y (yes) when the software
prompts
x. Return the printout of the passing Air Matrix with the PM
Checksheet report (DXSE0038).
QUALITY SYSTEM DOCUMENT Page 6 of 10
It is the responsibility of the user to verify this document originated from the controlled master and is the active revision prior to use.
This document contains confidential or proprietary information of GE Lunar. Neither the document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE Lunar.
GE Medical Systems LUNAR
xi. If any categories fail, check the deviation image for possible
xii. Run a Daily QA from the system software, or a Peak from the
xiii. Once an accepta ble Air Matrix scan is printed, delete the Air Matrix
b. Beam Wobble scan (Prodigy systems with software version 2.20 and
greater DXAP2011) or Alignment scan (Prodigy systems with software versions lower than 2.20 and DPX-NT/MD+ systems- DXAP3001) as appropriate and verify that it passes (attach the results).
7. Limit Switches a. Verify Limit Switch distances (DPX series Diags – F1 Limit Switch
Diagnostics / 1 Limit to Limit Distance; enCORE systems - tools / Diagnostics / scanner motion
DOC ID: DXSE0012
REV: G DATE: 2/28/02 CO: L10173
electronic or alignment problems, consult GE-Lunar OLC for assistance.
diagnostic software, before trying the Air Matrix. files (these files begin with "XM" or "XP" plus the System Number,
and they are generally the only files beginning with "X" in the C:\LUNAR subdirectory).
b. DPX Series Only - Home the scanner c. Mark the home position on the Home Jig
QUALITY SYSTEM DOCUMENT Page 7 of 10
It is the responsibility of the user to verify this document originated from the controlled master and is the active revision prior to use.
This document contains confidential or proprietary information of GE Lunar. Neither the document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE Lunar.
GE Medical Systems LUNAR
(4-3) Transverse Steps
DPXα
10920 – 10960
DOC ID: DXSE0012
DPX­IQ / MD­180 or DPXα w/
4.1d or greate r host
10780­10820
DPXL
1244 0 ­1248 0
DPX­IQ / MD­240 or DPXL w/
4.1d or greate r host
12430­12470
DPX­SF
1092 0 – 1096 0
REV: G DATE: 2/28/02 CO: L10173
DPX­SF w/
4.1d or greate r host
10780­10820
PRODIGY 240cm
Table
12475 +/- 25
PRODIGY 180cm
Table
12475 +/- 25
DPX­NT / MD+
240c m Table
12450 +/- 20
DPX­NT/MD+
180cm Table
12450 +/- 20
(1-2) Longitudina l Steps
(3-4) Transverse Steps
(2-1) Longitudina l Steps
13360 – 13400
10920 – 10960
13360 – 13400 w/
3.63 host
13280­13320
10780­10820
13280­13320
1988 0 ­1992 0
1244 0 – 1248 0
1988 0 – 1992 0 w/
3.63 host
19730­19770
12430­12470
19730­19770
8. Replace the table top and screw down. a. DPX Series Only – Insure Home position is correct by using the Home Jig b. If necessary adjust limit switches
9. Daily QA Tests
4225 – 4265
1092 0 – 1096 0
4225
­4265 w/
3.63 host
4195­4235
10780­10820
4195­4235
19890 +/- 35
12475 +/- 25
19890 +/- 35
13790 +/- 35
12475 +/- 20
13790 +/- 35
19605 +/- 20
12450 +/-20
19605 +/- 20
13530 +/-20
12450 +/- 20
13530 +/- 20
a. DPX Series only – check the peak setting on the QA last run
i. If peak is not 500 +/- 25 units adjust the peak
1. The voltage to the detector is controlled by the adjustment of the potentiometer on the SBC board labeled HV ADJ - R43.
2. The voltage at TP18 of the SBC is equal to the detector voltage divided by 1000. If R43 is adjusted to decrease the voltage at TP18 by 0.001, the peak will increase by 5 units.
3. The last peak value obtained during a Quality Assurance scan is stored in the Quality Assurance History database file.
QUALITY SYSTEM DOCUMENT Page 8 of 10
It is the responsibility of the user to verify this document originated from the controlled master and is the active revision prior to use.
This document contains confidential or proprietary information of GE Lunar. Neither the document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE Lunar.
GE Medical Systems LUNAR
DOC ID: DXSE0012
REV: G DATE: 2/28/02 CO: L10173
If the next peak value obtained differs from the last value stored in the database by more than 50 units, the Quality Assurance scan will fail. If the peak must be shifted by more than 50 units to return it to 500 units +25, the peak must be adjusted in steps of less than 50 units (40 unit steps are recommended).
4. A complete QA must be run
5. between each step to update the Quality Assurance History database file.
6. If the software peak setting corresponds to where both the high and low channel counts are at a maximum, (see fig.
6.3) follow the steps outlined below. The graph on the QA will look similar to the one shown in fig 6.3 with the dashed vertical line (where the software thinks the peak is) on the graph at a point where both the high and low energy plots are maximized (the dashed curved lines). If the graph on the QA is unlike fig 6.3, go to part B of this procedure.
7. select "Peak / Set AMP Gain", then select "Perform Peak" and [Esc] to run the peak. Observe the results generated by the QA to determine if the software has selected the Amplifier Setting where both high and low channel counts are at a maximum. If not go to part B.
8. 3. If the peak requires adjustment, (is not within +/- 25 units of 500) connect a voltage meter to the SBC at TP18 (see fig
6.2). Put the meter to the VDC scale and record the voltage reading (e.g. 0.703VDC
9. Using the voltage recorded in step 3 and the peak value recorded in step 2 determine the appropriate voltage adjustment for R43.
10. To determine voltage adjustment:
a. If Peak >500: New
/5 * + Old
b. Peak <500 New
VTP18
Old
VTP18
VTP18 = (500 – Current Peak) / 5 * –
VTP18 = (Current Peak – 500)
c. Do not attempt to adjust the peak by more than 40
units as an adjustment without a complete passing QA in the LUNAR software will cause a peak failure on the QA. For Peak settings that are more than 50 units away from 500 multiple adjustments are required.
QUALITY SYSTEM DOCUMENT Page 9 of 10
It is the responsibility of the user to verify this document originated from the controlled master and is the active revision prior to use.
This document contains confidential or proprietary information of GE Lunar. Neither the document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE Lunar.
GE Medical Systems LUNAR
b. Run two Daily QA’s and verify that they all pass (attach passing QA’s).
c. Start an additional Quality Assurance test and verify the function of the
Emergency Stop button by using it to stop the system
DOC ID: DXSE0012
REV: G DATE: 2/28/02 CO: L10173
11. With your voltmeter attached to TP 18, adjust R43 (see fig
6.2) until your voltmeter reads what you determined in step
6.
12. From the main menu press F3 System Quality Assurance and place the System Standard on the table silk-screen with brass insert up and at head end of table. Press[Esc] and allow the QA to run in its entirety and to print out. Verify that QA passes.
13. If peak value is not within 25 units of 500, record peak value from the QA that was just run and the current voltage at TP
18. Using these values repeat steps A. 3-8 until Peak is within 25 units of 500 and a complete passing QA has been run.
i. Verify that all four-indicator lamps are functioning.
10. Perform an AP Spine scan of the aluminum phantom (attach results) see DXQC0022 (DPX Series) or DXQC2022 (enCORE based systems).
a. Verify that the scanner is not “Hacksawing” – the edges on the spine
phantom image should be straight – if they are not
i. Verify that the front and rear transverse pulleys are tight
ii. Check for transverse binding
iii. If the transverse belt is loose tighten it.
iv. Prodigy Scanners – Run the Hacksaw test (DXAP2007).
b. The L2-L4 result should be within the expected range for the individual
phantom. c. % Fat result in (water based phantoms) should be within 4 to 5%. d. Verify that the Patient Positioning Switches and the Patient Locator Laser
are functioning.
11. Run a final Daily QA
12. Re check the last session in the error log and make sure it is free of hardware errors.
13. Make a copy of the DEXA PM Check sheet for the customer, discuss any service concerns with the system and explain work performed.

GENERATED RECORDS

DXSE0038 DEXA PM Quality record
QUALITY SYSTEM DOCUMENT Page 10 of 10
It is the responsibility of the user to verify this document originated from the controlled master and is the active revision prior to use.
This document contains confidential or proprietary information of GE Lunar. Neither the document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by GE Lunar.
Loading...