This manual contains Installation and operating instructions for PACE Pressure Indicators.
Safety
The manufacturer has designed this equipment to be safe when operated using the
procedures detailed in this manual. Do not use this equipment for any other purpose than
that stated, the protection provided by the equipment may be impaired.
This publication contains operating and safety instructions that must be followed to make
sure of safe operation and to maintain the equipment in a safe condition. The safety
instructions are either warnings or cautions issued to protect the user and the equipment
from injury or damage.
Use qualified * technicians and good engineering practice for all procedures in this
publication.
Pressure
Do not apply pressures greater than the maximum working pressure to the equipment.
Toxic Materials
There are no known toxic materials used in construction of this equipment.
Maintenance
The equipment must be maintained using the procedures in this publication. Further
manufacturer’s procedures should be done by an authorized service agents or the
manufacturer’s service departments.
Technical Advice
For technical advice contact the manufacturer.
* A qualified technician must have the necessary technical knowledge, documentation,
special test equipment and tools to carry out the required work on this equipment.
[EN] EnglishiK0470 Revision A
Page 3
General Specification
DisplayLCD: Colour display with touch-screen
Operating temperature10°C to 50°C (50° to 122°F)
Storage temperature-20°C to 70°C (-4° to 158°F)
Ingress protectionIP20 (EN60529)
Operating humidity5% to 95% RH (non-condensing)
VibrationMIL-PRF-28800 Type 2 class 5 style E/F
Operating altitudeMaximum 2000 metres (6560ft)
EMCEN 61326
Electrical safetyEN 61010-1, UL61010-1, CSA 22.2, No. 61010-1 and IEC61010-1
Power adaptorInput range: 100 - 240VAC, 50 to 60Hz, 660mA. Installation category II
Pressure safety
Pressure Equipment Directive - class: sound engineering practice (SEP)
for group 2 fluids.
Pollution degree2
Operating EnvironmentIndoor use only. Do NOT use in potentially explosive environments
Abbreviations
The following abbreviations are used in this manual; abbreviations are the same in the singular and plural.
aAbsoluteminMinute or minimum
a.cAlternating currentmmmillimetre
ALTAltitudemVmillivolts
ASCIIAmerican Standard Code for Information InterchangeMWPMaximum working pressure
BSPBritish pipe threadNoNumber
CASCalibrated airspeedNPTNational Pipe Thread
CSK CountersunkPACE
d.c.Direct currentPara.Paragraph
DPIDigital Pressure InstrumentPDCRPressure transducer
etc.And so onPEDPressure equipment directive
e.g.For examplepsiPounds per square inch
Fig.FigurePTXPressure transmitter
ftFootROCRate of climb (vertical speed)
gGaugeRS232Serial communications standard
GPIBGeneral purpose interface busRt CASRate of Calibrated airspeed
HgMercuryRt MACH Rate of MACH
HzHertzRxReceive data
IASIndicated airspeedSCPI
IDOSIntelligent digital output sensor (GE product)SDSSales data sheet
Institute of Electrical and Electronic Engineers standard
488 (for programmable devices with a digital interface)
TxTransmit data
Separated (or Safety) extra low
voltage
Related publications
K0467 User Guide and Safety Instructions
K0469 PACE Heritage Communications Manual
K0450 PACE Series Calibration Manual
K0472 PACE Series SCPI Manual
[EN] EnglishiiiK0470 Revision A
Page 5
Symbols
This equipment meets the requirements of all relevant European safety directives.
The equipment carries the CE mark.
This symbol, on the equipment, indicates that the user should read the user manual.
This symbol, on the equipment, indicates a warning and that the user should refer to
the user manual.
Ce symbole, sur l’instrument, indique que l’utilisateur doit consulter le manuel d’utilisation. Ce symbole, dans le manuel, indique une situation dangereuse.
This symbol warns the user of the danger of electric shock.
Ce symbole alerte l’utilisateur sur le danger de choc électrique.
Do not dispose of this product as household waste. Use an approved organisation
that collects and/or recycles waste electrical and electronic equipment. For more
information, contact one of these:
- Our customer service department: www.gemeasurement.com
- Your local government office.
WARNINGS
TURN OFF THE SOURCE PRESSURE(S) AND CAREFULLY VENT THE PRESSURE LINES BEFORE
DISCONNECTING OR CONNECTING THE PRESSURE LINES. PROCEED WITH CARE.
ONLY USE EQUIPMENT WITH THE CORRECT PRESSURE RATING.
BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE.
REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT. DO NOT USE ANY DAMAGED
FITTINGS AND EQUIPMENT.
DO NOT EXCEED THE MAXIMUM WORKING PRESSURE OF THE INSTRUMENT.
THIS EQUIPMENT IS NOT RATED FOR OXYGEN USE.
K0470 Revision Aiv[EN] English
Page 6
PACE Pressure Indicators User Manual
THE GROUND LEAD OF THE INSTRUMENT MUST BE CONNECTED TO THE AC SUPPLY
PROTECTIVE SAFETY GROUND.
ISOLATE THE POWER SUPPLY BEFORE MAKING ANY ELECTRICAL CONNECTIONS TO THE
REAR PANEL.
Convert FROM pressure VALUE 1 in pressure UNITS 1 TO pressure VALUE 2 in pressure UNITS
2, calculate as follows:
[EN] EnglishvK0470 Revision A
ftH2O @ 60°F
29.8516987
VALUE 2=VALUE 1 x FACTOR 1
FACTOR 2
Page 7
CONTENTS
SectionTitlePage
1Description1-1
1.1Introduction1-1
2Installation2-1
2.1Packaging2-1
2.2Packaging for Storage and Transportation2-1
2.3Preparation for Use2-1
2.4Connecting the Instrument2-2
2.5Mounting kits2-8
2.6Electrical connections2-10
3OPERATION3-1
3.1Preparation3-1
3.2Power-up Sequence3-1
3.3Measure Mode3-2
3.4Operation and Example Procedures3-8
3.5Global Set-up Selections3-11
3.6Supervisor Set-up3-12
3.7Instrument Status3-13
4MAINTENANCE4-1
4.1Introduction4-1
4.2Visual inspection4-1
4.3Cleaning4-1
K0470 Revision Avi[EN] English
Page 8
PACE Pressure Indicators User Manual
4.4Test4-1
4.5Calibration4-1
5TESTING AND FAULT FINDING5-1
5.1Introduction5-1
5.2Standard Serviceability Test5-1
5.3Ethernet testing5-1
5.4Fault Finding5-4
5.5Approved Service Agents5-4
6REFERENCE AND SPECIFICATION6-1
6.1Installation notes6-1
6.2Reference port6-2
6.3Icons6-3
6.4Measure Set-up6-6
6.5Status6-7
6.6Global Set-up6-8
6.7Supervisor Set-up6-9
6.8Calibration6-27
6.9Specification6-27
6.10Options6-28
6.11Installation and Ancillary Equipment6-31
6.12Return Goods/Material Procedure6-31
6-13Packaging Procedure6-31
[EN] EnglishviiK0470 Revision A
Page 9
Intentionally blank
K0470 Revision Aviii[EN] English
Page 10
PACE Pressure Indicators User Manual
1Description
1.1Introduction
The PACE Pressure Indicator measures both pneumatic and hydraulic pressures and
displays, on a colour touch-screen, the measured pressure and instrument status. The
touch-screen enables selections and settings in measuring modes. The instrument can be
operated remotely through communication interfaces.
Figure 1-1 PACE1000 General view
The rear of the instrument houses all the electrical and pressure input connections. The
electrical connections provide a power supply, serial and parallel communication interfaces,
pressure ports and option ports.
The instrument can be used as follows:
•Free-standing instrument positioned on a horizontal surface.
•Rack-mounted in a standard 19 inch rack using the rack-mount option kit.
•Panel mount using the panel-mount option kit.
[EN] English1 - 1K0470 Revision A
Page 11
1 Description
Options available for the PACE1000 refer to the data sheet.
Information and notes on applications (Ref: Reference and Specification, Section 6) or
www.gemeasurement.com
K0470 Revision A1 - 2[EN] English
Page 12
PACE Pressure Indicators User Manual
2Installation
2.1Packaging
Check the contents of the PACE1000 packaging with the list that follows:
Packaging List - PACE1000
i)PACE1000 Pressure Indicator.
ii)Adaptor, power supply (GE part number 191-370).
iii)User guide and safety instructions, and CD containing the full documentation suite.
iv)Calibration certificate.
2.2Packaging for Storage or Transportation
To store or return the instrument for calibration/repair do the procedures that follow:
1.Pack the instrument (Ref: Reference and Specification, Section 6.13).
2.Return the instrument for calibration/repair complete the return goods procedure (Ref:
Reference and Specification, Section 6.12).
2.3Preparation for Use
The instrument can be used as a:
•Free-standing instrument positioned on a horizontal surface.
•Panel-mounted using the panel-mount option kit (Ref: Section 2.5).
•Rack-mounted in a standard 19 inch rack using the rack-mount option kit (Ref: Section
2.5).
For free-standing instruments, the feet on the front of the base can be used elevate the instrument
to a better viewing angle.
Note: Allow a free flow of air around the instrument, especially at high ambient temperatures.
[EN] English2 - 1K0470 Revision A
Page 13
2 Installation
ISO 228 G1/8ISO 228 G1/8
bonded sealbonded seal
Recommended methodAlternative method below 100 bar
2.4Connecting the Instrument
WARNINGS
TURN OFF THE SOURCE PRESSURE(S) AND CAREFULLY VENT THE PRESSURE LINES BEFORE
DISCONNECTING OR CONNECTING THE PRESSURE LINES. PROCEED WITH CARE.
ONLY USE EQUIPMENT WITH THE CORRECT PRESSURE RATING.
BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE.
REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT. DO NOT USE ANY DAMAGED
FITTINGS AND EQUIPMENT.
DO NOT EXCEED THE MAXIMUM WORKING PRESSURE OF THE INSTRUMENT.
THIS EQUIPMENT IS NOT RATED FOR OXYGEN USE.
Pneumatic Pressure (Figure 2-1)
1.Refer to the Data sheet for the correct pressure mediums to be used.
2.Connect the Unit Under Test (UUT) to the required connection port.
Note: For instruments with NPT connections, use applicable bonded sealing as shown in figure
below.
Figure 2-1, Sealing Pneumatic Connections
K0470 Revision A2 - 2[EN] English
Page 14
PACE Pressure Indicators User Manual
Parallel threads must be used.
Female thread type is parallel thread to
ISO228/1 (DIN ISO228/1, JIS B0202)
G1/8
Tapered threads NOT allowed.
PACE pressure port
PACE rear
panel
9
8
PACE rear
panel
PACE pressure port
Pneumatic connections
WARNING
PARALLEL THREADS MUST BE USED. FEMALE THREAD TYPE IS PARALLEL THREAD TO
ISO228/1 (DIN ISO228/1, JIS B0202) G1/8.
TAPERED THREADS NOT ALLOWED.
For examples of adaptors (Ref: page 2 - 4).
Input Pressure (Figure 2)
Note: For instruments with NPT connections, use applicable bonded sealing as shown in figure
[EN] English2 - 3K0470 Revision A
Connection
InputISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)
ReferenceISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)
1.Make sure the user systems can be isolated and vented.
2.Connect the Unit Under Test (UUT) to the output connection port.
below.
Page 15
2 Installation
2) Bonded seal
1) Connector
2
1
Figure 2-2, Pneumatic Connections
Installation
The instrument connects to the Unit Under Test.
Input Pressure and Equipment
The pressure should not exceed 1.25 x full-scale or MWP stated on the rear panel of the
instrument.
To protect the instrument from over-pressure a suitable protection device (such as a relief valve or
bursting disc) must be fitted to prevent over pressurization.
Pneumatic Connection
WARNING
PRESSURE RANGES > 210
BAR (3000 PSI) ARE ONLY RATED FOR HYDRAULIC USE.
CAUTIONS
Do not exceed the maximum pressures stated in the appropriate Component Manual for
the unit under test.
Reduce pressure at a controlled rate when venting to atmosphere.
Carefully de-pressurize all pipes to atmospheric pressure before disconnecting and
connecting to the unit under test.
Connections
1. Switch off the power supply before connecting or disconnecting the instrument.
2.Use the appropriate sealing method for all pressure connections.
Note: For instruments with NPT connections, use applicable bonded sealing as shown in figure
below.
K0470 Revision A2 - 4[EN] English
Page 16
PACE Pressure Indicators User Manual
ISO 228 G1/8ISO 228 G1/8
bonded sealbonded seal
Recommended methodAlternative method below 100 bar
IO-SNUBBER-1
IO-ADAPT-9/16AUTOC
ISO 228 Male to 9/16 Autoclave
IO-DIFFUSER-1
IO-ADAPT-G1/4
ISO 228 Male to G1/4 Female
IO-ADAPT-1/8NPT
ISO 228 Male to 1/8 NPT Female
IO-ADAPT-1/4NPT
ISO 228 Male to 1/4 NPT Female
IO-ADAPT-7/16UNF
ISO 228 Male to 7/16 UNF Female
IO-ADAPT-AN4
ISO 228 Male to AN4 37° Male
IO-ADAPT-AN6
ISO 228 Male to AN6 37° Male
IO-ADAPT-BARB
ISO 228 Male to 1/4 Hose
Method of connection
Adaptors
Refer to the data sheet for the range of adaptors.
[EN] English2 - 5K0470 Revision A
Page 17
2 Installation
Hydraulic Pressure
WARNINGS
HYDRAULIC LIQUID IS DANGEROUS. OBSERVE RELEVANT HEALTH AND SAFETY
PRECAUTIONS. USE APPROPRIATE PROTECTIVE BARRIERS AND EYE PROTECTION.
BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE AND
ENSURE THAT ALL EQUIPMENT IS TO THE CORRECT PRESSURE RATING.
DO NOT EXCEED THE MAXIMUM WORKING PRESSURE OF THE INSTRUMENT.
PURGE ALL AIR FROM THE HYDRAULIC LIQUID.
PRESSURE RANGES > 210
BAR (3000 PSI) ARE ONLY RATED FOR HYDRAULIC USE.
DO NOT USE A SENSOR FOR GAS THAT HAS BEEN USED WITH HYDRAULIC LIQUID.
CAUTIONS
Do not exceed the maximum pressures stated in the appropriate component manual for
the unit under test.
Reduce pressure at a controlled rate when venting to atmosphere.
Carefully de-pressurize all pipes to atmospheric pressure before disconnecting and
connecting to the unit under test.
Observe absolute cleanliness when using the instrument.
Severe damage can be caused if equipment connected to this instrument is
contaminated.
Connect only clean equipment to the instrument.
To avoid any contamination, an external filter is recommended.
K0470 Revision A2 - 6[EN] English
Page 18
PACE Pressure Indicators User Manual
ISO 228 G1/8ISO 228 G1/8
bonded sealbonded seal
Recommended methodAlternative method below 100 bar
Installation
The instrument connects to the Unit Under Test.
Input Pressure and Equipment
1.The pressure should not exceed 1.25 x full-scale or MWP stated on the rear panel of
the instrument.
2.To protect the instrument from over-pressure a suitable protection device (such as a
relief valve or bursting disc) must be fitted to limit the pressure to below the MWP.
Note: For instruments with NPT connections, use applicable bonded sealing as shown in figure
below.
Figure 2-3, Sealing Hydraulic Connections
Hydraulic connections
Connection
InputISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)
ReferenceISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)
Note: Pressure connections greater than 210 bar are 9/16” 18UNF Male Autoclave.
1. Switch off the power supply before connecting or disconnecting the instrument.
2.Use the applicable sealing method for all pressure connections.
3.Isolate the hydraulic pressures and de-pressurise the pipes before connecting or
disconnecting the instrument.
Pressure input (Figure 2)
1.Make sure the user systems can be isolated and vented.
2.Use the applicable sealing method for all pressure connections.
3.The hydraulic liquid must be clean, refer to specification given in the Data sheet.
4.Connect the Unit Under Test (UUT) to the appropriate connection port.
[EN] English2 - 7K0470 Revision A
5.Fill and bleed the UUT and connecting pipes.
Page 19
2 Installation
2.5Mounting kits
Rack-mount option (Figure 2-5)
There must be enough space at the rear of the instrument for all the cables and pipes. The length
of the cables and pipes must allow for the removal and installation of the instrument. The cooling
air of the instrument must not be obstructed. Allow a free flow of air through the equipment rack
and around the instrument, especially at high ambient temperatures.
Procedure
1.Locate instrument in rack mount assembly
2.Secure with the four M3 x 6 screws , (maximum length M3 x 8).
3.Support the instrument and connect the cables and pipes.
4.Refer to the electrical connections below before fitting the instrument into the
equipment rack.
5.Temporarily locate the two spigots* to each side of the equipment rack.
6.Locate and slide the instrument into the rack.
7.Locate the instrument on the spigots*.
K0470 Revision A2 - 8[EN] English
Figure 2-5 Rack-mounting
.
Page 20
PACE Pressure Indicators User Manual
8Secure the instrument in the equipment rack with two of the screws and washers
(supplied).
9.Remove the two spigots* and replace with the remaining two screws and washers
(supplied).
Panel-mount option (Figure 2-6)
There must be enough space at the rear of the instrument for all the cables and pipes.
The length of the cables and pipes must allow for the removal and fitment of the instrument.
The cooling air of the instrument must not be obstructed.
Allow a free flow of air through the equipment rack and around the instrument, especially at high
ambient temperatures.
Procedure
1.Remove the four screws
2.Locate the instrument in panel mount assembly.
3.Secure with the four screws
4.Support the instrument and connect the cables and pipes.
5.Refer to the electrical connections below before fitting the instrument into the panel.
6.Secure the instrument in the panel with four screws and washers
[EN] English2 - 9K0470 Revision A
Figure 2-6 Panel-mounting
from the instrument.
.
.
Page 21
2 Installation
2.6Electrical connections
WARNINGS
THE GROUND LEAD OF THE INSTRUMENT MUST BE CONNECTED TO THE AC SUPPLY
PROTECTIVE SAFETY GROUND.
ISOLATE THE POWER SUPPLY BEFORE MAKING ANY ELECTRICAL CONNECTIONS TO THE
REAR PANEL.
CAUTIONS
Use the power adaptor supplied with the instrument (GE part no. 191-370). Using other
power adaptors may cause over-heating, this can result in a fire.
Do not let the power adaptor come into contact with any moisture or liquids.
Connecting (Figure 2-7)
1.Before use, make sure the SELV power adaptor supplied with the
instrument is used (GE part number 191-370).
2.Install an accessible power isolator to use as the disconnecting device
in the power adaptor supply circuit.
3.The power adaptor input power supply range: 100 - 240VAC, 50 to 60Hz
660mA, Installation Category II.
Note: The power adaptor must be supplied by a fused or overload-protected power
supply.
4.Connect the power adaptor to the instrument.
5.Switch the power supply on.
6.Check that the front panel display shows the power-up sequence (Ref:
section 3.2).
Note:. After the power-up sequence, the instrument shows the default display on the
touch screen. The touch screen divides into a number of mimic keys.
Requirements for rack-mounted and panel-mounted instruments
1.Install an accessible power isolator to use as the disconnecting device in the power
adaptor supply circuit.
2.Set the power supply isolator to OFF.
3.Connect the power adaptor before sliding the instrument into the rack.
4.Set the power supply isolator to ON.
5.Check that the front panel display shows the power-up sequence (Ref: section 3.2).
K0470 Revision A2 - 10[EN] English
Page 22
PACE Pressure Indicators User Manual
2
4
5
6
7
22V to 26V
ANALOGUE 1
CONSULT HANDBOOK
1
Communication Connections
Connect the applicable connectors into the rear panel communications ports and, if appropriate,
secure with the captive screws.
Note: The RS232 and IEEE 488 interfaces are both enabled at power-up. Set the required parameters
in Supervisor Setup/communications menu, see Section 3.6.
Figure 2-7, Communication Connectors
1
Power supply adaptor2RS2324IEEE488
5
USB B6USB A7Ethernet
RS232 Interface
When using the RS232 interface, a cable must be connected directly from the instrument to a
suitable port on the computer in a ‘point to point’ link.
The pin connections for the 9-pin D-type, RS232 connector and the relationship between the
instrument and the RS232 control signals, together with device interconnection interface is shown
in Table 2-1. The instrument is configured as Data Circuit Terminating Equipment (DCE).
[EN] English2 - 11K0470 Revision A
Page 23
2 Installation
InstrumentControl LineComputer/Printer
Instrument
Function
RxD (I/P)3TxD32
TxD (O/P)2RxD23
GND5GND57
CTS (I/P)7RTS74
RTS (O/P)8CTS85
Pulled
high
internally
Not
connected
Pulled
high
internally
Equipment
chassis
Connector
9-way
D-type
Pin No.
1
4DTR420
6
Connector
shell
Signal
Direction
Cable Screen-1
RS232
Terminology
RLSD
(DCD)
DSR
DCE Ready
Connector Type
9-way
D-type
Pin No.
18
66
25-way
D-type
Pin No.
Table 2-1, RS232 Connections
Handshaking connections
Software handshaking use: TXD, RXD and GND.
Hardware handshaking use: TXD, RXD, GND, CTS, RTS and DTR.
K0470 Revision A2 - 12[EN] English
Page 24
PACE Pressure Indicators User Manual
2
3
IEEE 488 Interface
The interface complies with IEEE 488 standard.
The IEEE 488 parallel interface connects a computer/controller to one or more PACE1000
instruments and other instruments.
Up to 30 instruments can be connected through a high-speed data bus to the computer/controller.
Note: The length of each IEEE 488 cable must be less than 3 metres to comply with the EMC
requirements (Ref: Data sheet).
Single Unit Installation (Figure 2-8)
1.Connect an IEEE 488 connector/cable assembly to the rear panel of the instrument.
2.Connect the other end of the connector/cable assembly to the IEEE 488 connector on
the controller/computer.
3.Change the IEEE 488 communication parameters (Ref: Supervisor set-up, Section 6.7).
Multiple Unit Installation (Figure 2-8)
To install multiple units use stacking plugs to link
the first instrument and second instrument as
follows:
1Connector to rear panel of first
instrument (Ref Illustration).
2Connector from controller/computer
(Ref Illustration).
3Connector to rear panel of second
instrument (Ref Illustration).
4.Connect the IEEE 488 connector on
the controller/computer and the other
connector into the next instrument.
5.Repeat this procedure for all the instruments in the system.
6.Use the Supervisor set-up (communications) menu on each instrument to set-up the
required communication parameters (Ref: Section 3.8).
1
[EN] English2 - 13K0470 Revision A
Page 25
2 Installation
IEEE 488
ADDRESS 1
ADDRESS 2
ADDRESS N
(30 maximum)
6
7
8
DAV (DATA VALID)
NRFD (NOT READY FOR DATA)
NDAC (NO DATA ACCEPTED)
5
9
101117
EOI (END OF IDENTIFY)
IFC (INTERFACE CLEAR)
SRQ (SERVICE REQUEST)
ATN (ATTENTION)
REN (REMOTE ENABLE)
1
2
3
4
13
DIO1
DIO2
DIO3
DIO4
DIO5
14
15
16
DIO6
DIO7
DIO8
1819202122
GND (6)
GND (7)
GND (8)
GND (9)
GND (10)
23
24
GND (11)
GND
DATA /
STATUS
BAR
0V (GND)
12CHASSIS/FRAMECHASSIS/FRAME
K0470 Revision A2 - 14[EN] English
Figure 2-8 - IEEE 488 Connection
Page 26
PACE Pressure Indicators User Manual
Do not touch the display screen during power-up
3Operation
This section contains quick reference charts detailing all the available functions and the set-up
menu.
3.1Preparation
Make sure the electrical cables and pneumatic pipes comply with the installation requirements
(Ref: Section 2).
Before use do the following:
1.If necessary, do the maintenance task (Ref: Section 4).
2.For bench-top, single instrument operation do the following:
a.Connect the instrument to the electrical supply.
b.Inspect the pneumatic hoses for damage, ingress of dirt and moisture.
3.Before use, the instrument should be tested.
4.Review and become familiar with the procedure before starting a process on a
component or system.
3.2Power-up sequence
The following sequences of operation shows the instrument display.
Note: The following sequence is an example, the values and selections displayed depend on the
range(s) and options enabled in the instrument.
1.Set the power supply to ON.
2.The display shows the power-up sequence.
3. The instrument carries out a self-test.
a.If the test finds a fault, the display
shows an error (Ref: Fault Finding and
Testing, Section 5).
4.If the self-test is successful the system
enables the touch screen and changes to
measure mode.
5.The touch screen shows the measured
pressure in the parameters selected in set-up.
6.The instrument is now ready for use.
[EN] English3 - 1K0470 Revision A
Page 27
3 Operation
1
2
3
4
5
6
3.3Measure mode
Touch screen areas
1 Pressure reading 2 Functions enabled3 Zero key (vent system before starting zero sequence)
4 Function area5 Status area6 Current pressure range
Display Icons
Tare enabledFilter pressure reading
PercentageEthernet not connected
Reference level difference
(gas head correction)
Ethernet connected
K0470 Revision A3 - 2[EN] English
Page 28
PACE Pressure Indicators User Manual
1
Function area Min/Avg/Max is performed on the reading selected for display in the top
screen.
The Ethernet LAN Status indication (1) shows the following:
• Colour red - not connected
• Colour green - connected
[EN] English3 - 3K0470 Revision A
Page 29
3 Operation
1
Data Logging
The “Datalog” icon (1) is present when a memory card is fitted:
K0470 Revision A3 - 4[EN] English
Page 30
PACE Pressure Indicators User Manual
[EN] English3 - 5K0470 Revision A
Page 31
3 Operation
K0470 Revision A3 - 6[EN] English
Page 32
PACE Pressure Indicators User Manual
current pressure reading to zero,
offset stored for current range.
allows selection of available pressure
ranges. Ê
-
-
select from list of available pressure
measurement units.
-
Absolute range selection available when
barometric option installed.
Ê
%
Filter pressure reading
Tare
Pressure zero
Range
Process
Task
Units
Global set-up
Status
Measure set-up
More
Next page of menu options. Loops
from last page to first page.
Stores settings and exits
set-up.
see 3.6:-
next page:-
Exit set-up.
Peak
Measure Menu Set-up
[EN] English3 - 7K0470 Revision A
Page 33
3 Operation
Selecting a task exits the menu and
changes the display to the task
selected.
Ê
Exit set-up.
Option refer to data sheet
Ê
Basic
Airfield
Task
Return to home.
3.4Operation and Example Procedures
Introduction
Before operation, the instrument must be connected to the correct electrical and pneumatic/
hydraulic) supplies, (Ref: Installation,Section 2).
When the instrument is switched ON the display shows measured pressure mode and the task set
before the power-off.
Measure Mode
The instrument works as a precision pressure indicator and shows the pressure measured at the
output port.
Pressing Task enables pre-determined functions:
Task
The display shows the task screen (Ref: Illustration above).
When selected, e.g. Basic, the screen changes to show the selected task.
To measure pressure in the task proceed as follows:
1.Select the required units of pressure measurement from the measure set-up menu.
K0470 Revision A3 - 8[EN] English
Page 34
PACE Pressure Indicators User Manual
QNEQNHQFE
29.9212 inHg (1013.25 kPa)
Datum
Sea Level Reference Datum
(adjusted)
Runway
(on Earth’s surface)
Note: In the Airfield Task Q codes can be used. These units are standardised three letter codes and
are available in aeronautical units (feet and metres). The codes used are:
•QFE - Atmospheric pressure at sea level, corrected for temperature and adjusted to
airfield elevation. When set on the altimeter it reads height.
•QNE - Atmospheric pressure at Sea level in International Standard Atmosphere (ISA)
1013.25 mbar.
•QFF - Barometric pressure at a place, reduced to Mean Sea Level (MSL) using the
actual temperature at time of observation as mean temperature.
•QNH - Atmospheric pressure at Mean Sea Level (MSL) (may be local, measured
pressure or a Regional Forecast Pressure (RFP). When set on altimeter it reads
altitude.
[EN] English3 - 9K0470 Revision A
Page 35
3 Operation
Start
Test time
Leak test
Settings
Leak rate
/second
/minute
Leak testing option
This task, measures the leak rate over the measure dwell time.
At the start of the test, the instrument measures the test pressure of the user system. The
instrument then records the pressure change during measure dwell time.
On completion, the display shows the leak rate results with leak rate per second or per minute in
the current pressure units selected in measure set-up.
K0470 Revision A3 - 10[EN] English
Page 36
PACE Pressure Indicators User Manual
1
2
Supervisor set-up
Calibration
Global set-up
Save/recall user set-up
Save current user set-up
Recall current user set-up
Display
PIN - enter four digit code (4321)
- % brightness
- Timer (time-out) setting
Resolution - Measure display resolution
Back light
Audio volume - % of full volume
Satus area
Display Mode
Function area
Reading
Graph
3.5Global Set-up Selections
Global set-up selections provide access to the instrument‘s settings for both measure and control
modes.
This set-up menu provides PIN-protected access to the supervisor set-up and calibration.
Pressing Global Set-up changes the touch-screen display to show available selections.
Supervisor Set-up, Calibration, Save/Recall User Set-up and Display.
1Selections2Escape Key
[EN] English3 - 11K0470 Revision A
Page 37
3 Operation
enter four digit code
(0268)
-
-
Alarms
Communications
Restore last settings
Supervisor
More
next page:-
PIN
Pressing the escape key stores
settings and returns to global
set-up.
Enable/disable high pressure
Enable/disable low pressure
IEEE488
RS232
USB
Ethernet
Gas head correction
Enable/
disable
Head - in or m
Local gravity - m/s
2
Barometric - mbar
Reference port medium*
Test port medium*
Temperature (gas)
Humidity (gas)
*air or nitrogen
Lock
Lock/unlock individual tasks
Lock/unlock all tasks
Lock tasks
Change PIN
Enter new PIN
Re-enter new PIN
Enter new
supervisor PIN
User defined units
Enter the name of the defined units
Enter the equivalent value in Pascals
User defined units user unit 1
Instrument alias name
Language
Restore previous defaults Y/N?
Rate set-up
/second
/minute
Area of Use
Europe
North America
Japan
Asia
Rest of World
Additional selections for options enabled: Aeronautical etc.
3.6Supervisor Set-up
K0470 Revision A3 - 12[EN] English
Page 38
PACE Pressure Indicators User Manual
Instrument
Software installed
Hardware build
Instrument Status
IEEE488*
Instrument Main Code
Instrument OS Build
Instrument Boot ROM
Analogue O/P Main Code
Analogue O/P Boot ROM
VFC Main Code
VFC Boot ROM
Additions to the
standard instrument
Pressing the escape key
stores settings and returns
to global set-up.
History
Calibration
Zero
Software
Hardware
Message
Communications
RS232*
* current settings, see 6.7
Current set-up
Support
Contact details
Ethernet (optional)
USB
Ethernet
More
next page:-
3.7Instrument Status
The control set-up menu provides access to the status of the instrument:
[EN] English3 - 13K0470 Revision A
Page 39
3 Operation
Exit set-up.
Instrument Main Code
Software history
Analogue O/P Main Code
Instrument OS Build
Instrument Boot ROM
Analogue O/P Boot ROM
VFC Main Code
VFC Boot ROM
Example
Additional selections for options
enabled: Analgue etc.
Software
Software history, in the status menu, provides read only information on the current software in the
instrument.
K0470 Revision A3 - 14[EN] English
Page 40
PACE Pressure Indicators User Manual
4Maintenance
4.1Introduction
This section contains procedures for routine maintenance and the replacement of components
(Ref: Testing and Fault Finding, Section 5).
Table 4.1 - Maintenance Tasks
TaskPeriod
Visual InspectionBefore use
TestBefore use
CleaningWeekly*
Calibration12 months †
* may change depends on usage (e.g., rack mounted, bench top) and environment (e.g., humidity, dust).
† may change depends on the required accuracy.
4.2Visual Inspection
Inspect for obvious signs of damage and dirt on the following:
a.External of the instrument.
b.Power supply adaptor
c.Associated equipment.
Damaged parts must be replaced contact GE Service.
For cleaning (Ref: Cleaning Section 4.3).
4.3Cleaning
CAUTION
Do not use solvents for cleaning.
Clean the front panel with a damp lint-free cloth and mild detergent.
4.4Test
Do a standard serviceability test (Ref: Standard Serviceability Test, Section 5.2).
4.5Calibration
The instrument should be returned to the manufacturer or calibration facility, (Ref: Section 6.14).
To find the date of the last calibration, press Measure set-up/Status/Calibration history.
[EN] English4 - 1K0470 Revision A
Page 41
4 Maintenance
Intentionally blank
K0470 Revision A4 - 2[EN] English
Page 42
PACE Pressure Indicators User Manual
5Testing and Fault Finding
5.1Introduction
This section details the standard serviceability test. Table 5.1 lists possible faults, and the
response.
The PACE1000 contains a self-test and diagnosis system that continuously monitors the
performance of the unit. At power-up, the system performs a self-test.
5.2Standard Serviceability Test
The following procedure shows if the unit is serviceable and checks functions and facilities of
the PACE1000.
Procedure
CAUTION
Always release pressure before disconnecting pressure equipment from the outlet port.
1.Connect the instrument (Ref: Installation, Section 2) Connect a UUT.
2.After power-up, select measure set-up.
a.Select the required units of pressure measurement from the measure set-up
menu.
b.Apply a known pressure to one of the sensors. Make sure the instrument
pressure reading is within tolerance, stated in the specification (Ref: SDS).
c.Carefully release the applied pressure to atmospheric pressure.
d.Make sure the instrument pressure reading shows atmospheric or ambient
pressure.
e.Test complete.
After a successful serviceability test the instrument is ready for use.
5.3Ethernet Testing
1. Connect PACE Ethernet port to a PC.
2. Check the PACE Ethernet LAN indication turns green after a few seconds.
3. Set the PACE Ethernet address to Auto IP.
4. Record the PACE auto IP address (The test screen below has auto IP address 3.115.21.237)
5. Open the PC command prompt screen.
6. Using the “ping” command, ping the PACE IP address (see screen capture below). The PACE
will reply if operating correctly.
[EN] English5 - 1K0470 Revision A
Page 43
5 Testing and Fault Finding
Ethernet open ports and their usage:
80/tcpPACE Web server (http)
111/tcpONC RPC port mapper
372/tcpVXI-11 communication (Dynamically allocated)
443/tcpWeb server (https)
5025/tcpPACE SCPI communication socket
111/udpONC RPC port mapper
K0470 Revision A5 - 2[EN] English
Page 44
PACE Pressure Indicators User Manual
Ethernet testing using a web browser
[EN] English5 - 3K0470 Revision A
Page 45
5 Testing and Fault Finding
Ethernet testing using NI
http://www.ni.com/download/ni-daqmx-14.1/4953/en/
5.4Fault Finding
Check the faults and responses (Ref: Table 5.1 Fault Diagnosis) before contacting
www.gemeasurement.com or a recommended Service Agent.
FaultResponse
Power supply connected, display not lit.Check electrical power supply fuse or circuit breaker.
Display pressure reading in redOver-range, carefully de-pressurize.
Vent system pressure.
Instrument will not zero.
Check for blockage.
Contact approved service agent for repair.
Table 5.1 - Fault Diagnosis
5.5Approved Service Agents
For the list of service centres logon to www.gemeasurement.com
K0470 Revision A5 - 4[EN] English
Page 46
PACE Pressure Indicators User Manual
6Reference and Specification
6.1Installation notes
The PACE1000 pressure indicator requires a set of connections with the exception of the
reference connection, this provides a reference to atmosphere for gauge sensors and
barometric sensors.
The gas density and type does not affect the accuracy of pressure measurement, assuming
that the UUT is at the same level (height) as the indicator or gas head correction is
accurately set.
-3
Values of air density (kg m
dioxide by volume.
) for air of relative humidity 50% and containing 0.04% carbon
Table 6-1 Air Density Values
Air
pressure
(kPa)
87
88
89
90
91
92
93
94
95
96
97
98
99
100
Air temperature (°C)
14161820222426
1.052 1.045 1.037 1.029 1.021 1.014 1,006
1.064 1.057 1.049 1.041 1.033 1.025 1.018
1.077 1.069 1.061 1.053 1.045 1.037 1.029
1.089 1.081 1.073 1.065 1.057 1.049 1.041
1.101 1.093 1.085 1.077 1.069 1.061 1.053
1.113 1.105 1.097 1.089 1.080 1.072 1.064
1.125 1.117 1.109 1.100 1.0921.084 1.076
1.137 1.129 1.121 1.112 1.104 1.096 1.088
1.149 1.141 1.133 1.124 1.116 1.108 1.099
1.162 1.153 1.145 1.136 1.128 1.119 1.111
1.174 1.165 1.156 1.148 1.139 1.131 1.123
1.186 1.177 1.168 1.160 1.151 1.143 1.134
1.198 1.189 1.180 1.172 1.163 1.154 1.146
1.210 1.201 1.192 1.184 1.175 1.166 1.158
101
102
103
104
105
106
1.222 1.213 1.204 1.196 1.187 1.178 1.169
1.234 1.225 1.216 1.207 1.199 1.190 1.181
1.247 1.237 1.228 1.219 1.210 1.201 1.193
1.259 1.249 1.240 1.231 1.222 1.213 1.204
1.271 1.261 1.252 1.243 1.234 1.225 1.216
1.283 1.274 1.264 1.255 1.246 1.237 1.228
Note: 100 kPa = 1 bar
[EN] English6 - 1K0470 Revision A
Page 47
6 Reference and Specification
6.2Reference Port
The reference port provides the negative pressure to the gauge sensor and to the barometric
reference (option). Gauge sensors use this port identified as “REF”. For gauge sensors
(without a barometric reference) small pressures can be applied (Ref: Specification, Section
6-7). All other pressure measurement requires the port to be opened to atmosphere. When in
gauge mode, the instrument shows and controls the pressure difference between the
reference port and the output port.
Note: This is not a true differential operation as there is no true differential calibration of the
sensor.
The transducer of the barometric reference option senses atmospheric pressure via the
reference port, when enabled the port MUST be open to atmosphere.
The reference connection should be actively used (differential connection option) for
precision low pressure measurement . The instrument measures pressure relative to the
pressure at the reference port.
An atmospheric pressure change causes the indicator to adjust the pressure and appears at
the pressure output as apparent instability. To keep a stable controlled pressure, the
reference port should be restricted. Using a reference port restrictor (snubber), short term
ambient pressure variations can be prevented from affecting indicator performance.
The indicator and UUT references should be connected together (using the optional
differential connection kit) to provide a common reference to atmosphere.
K0470 Revision A6 - 2[EN] English
Page 48
PACE Pressure Indicators User Manual
6.3Icons
The following icons are used in the PACE series of instruments, not all icons are used in every PACE instrument.
Display Icons in Set-up Menus
IconFunctionIconFunctionIconFunction
ActiveAero set-upAeronautical
Airspeed rangeAlarmAltitude range
Area of useAsteriskAuto range
Audio volumeAuto zeroBacklight
BarometerBasic
CalibrationCalibration history
Change supervisor
PIN
Control modeCopy
Correction SCMCorrection sensor
Correction valveCurrent set-upDate & time
Delete
CommunicationsContrast
Diagnostic
analogue output
Burst pressure
control mode
Correction
analogue output
Correction source
sensor
Diagnostic
barometric option
Diagnostic generalDiagnostic RS232
[EN] English6 - 3K0470 Revision A
Diagnostic control
sensor
Diagnostic
controller
Diagnostic source
sensor
Page 49
6 Reference and Specification
IconFunctionIconFunctionIconFunction
Diagnostic
vacuum sensor
DisplayDividerError
EscapeEthernet
Ethernet
connected
Gas head pressureGauge modeGlobal set-up
Go-to-groundHardware buildHome
Idle time-outIEEE488Information
Diagnostic voltfree
ExclamationFault history
Diagnostics
Ethernet not
connected
In limitsInstrument
Instrument alias
name
LockLock tasksLogic output
Max-minMax peakMin peak
NudgePassive modePercentage
PINPower-upPreset
PressurePressure filterProcess
LanguageLeak test
Instrument
accuracy
K0470 Revision A6 - 4[EN] English
Page 50
PACE Pressure Indicators User Manual
IconFunctionIconFunctionIconFunction
Protective ventQuestionRange
Recall user set-upReset use logResolution
Re-tryRoughingRS232
Restore to as
shipped settings
Save as shipped
Screen modeScreen saverSCM filter
SCM zeroSelect range
Set-point limits
Set dateSet serial numberSet time
Set-up zeroSlew rate linearSlew rate max rate
Restore settings 2Run
Save recall user
set-up settings
Set-point higher
limit
Save user set-up
Set-point disable/
enable
Set-point lower
limit
Software build
StatusStatus areaStep (single)
StopSupervisor set-upSwitch test
TareSupportTask
[EN] English6 - 5K0470 Revision A
Software upgrade
history
Software upgrade
Page 51
6 Reference and Specification
IconFunctionIconFunctionIconFunction
Test programTest program copy
TimingTime outTimed zero
UnitsUser defined unitsUse log
Use log historyVentVent time out
Vent Yes/NoVent set-upWarning
Zero analogue
output
6.4Measure Set-up
Pressure zero
Test program
delete
Zero historyZero
During use, the instrument pressure sensor can show small zero shifts caused by time and
temperature changes. Regular “zeroing” increases measuring precision.
Process
Selects display processing features that change the reading, as follows:
%:Pressure can be displayed reading as a percentage of full-scale or as a percentage
of a specified span.
Filter:The displayed reading can be filtered by a custom low pass filter or the filter can
be disabled (default disabled). The indicator works at a speed independent of the
filter time constant.
Tare:A specific tare value can be selected or the current displayed pressure reading can
be “captured” as the tare value. The display shows the selected tare value in the
pressure window.
Peak: Maximum, minimum and average display of pressure readings.
Task
Selecting Task enables a set of pre-determined functions and software enabled optional
functions.
K0470 Revision A6 - 6[EN] English
Page 52
PACE Pressure Indicators User Manual
Units
Select the new units from the list of pressure measurement units. Special units can also be
defined (Ref: Global set-up,
Global set-up
Section 6.6, supervisor set-up).
Ref: Global set-up,
Set-up zero
Zero from top level screen (main range only).
If other ranges are fitted these can be zeroed by selecting the displayed reading.
Barometric Reference Option
The barometric reference option measures the barometric pressure at the reference port.
Depending on the sensors fitted, it also permits the indicator to operate in either pseudo-
gauge or pseudo-absolute mode by the addition of barometric pressure.
6.5Status
The display shows the following:
a. Instrument status
Section 6.6.
•Model
•Serial number
•MAC address sensor(s)
•Range
•Last calibration date*
b. Software build - read only data.
c. Hardware build - read only data.
d. History - read only data
•Calibration
•Zero
•Software
•Hardware
•Message
•Ethernet connection
e. Communications, IEEE 488 and RS232 are fitted as standard. Additional communication
types are options - USB and Ethernet.
f. Current set-up - read only data.
g. Support
•List contact information for support and advice.
[EN] English6 - 7K0470 Revision A
Page 53
6 Reference and Specification
6.6Global set-up
Supervisor set-up
PIN protected menu (Ref: Supervisor set-up, Section 6.7).
Calibration
PIN protected menu (Ref: Calibration set-up, Section 6.8).
Save/recall user set-up
Save user set-up.
Recall user set-up.
Display
a.Resolution
b.Backlight
c.Audio volume
d.Status area
e.Display Mode
•Reading (default)
•Graph
* The instrument date and time must have been set correctly.
K0470 Revision A6 - 8[EN] English
Page 54
PACE Pressure Indicators User Manual
6.7Supervisor Set-up
The Supervisor menu provides facilities for programming settings. These are usually made
during installation as follows:
Important Note: A PIN protects the Supervisor menu against unauthorised use. Each
instrument on delivery contains the factory set PIN (0268). To continue
protecting the supervisor set-up menu the PIN should be changed as soon
as possible.
Alarms
An alarm can be set to trigger when the pressure exceeds the high alarm or falls below the
low alarm. A buzzer sounds when the alarm triggers and the alarm symbol (bell) appears on
the display.
Comms
Selects the communication port parameters and simultaneous operation of the RS232, IEEE
488 interface and Ethernet.
Note: PACE1000 Lab view drivers are available for download from:
The user can select appropriate settings for communicating with the control computer (PC)
and the required command protocol.
Referring to the PACE SCPI communications user manual there are three commands that
can be used to retrieve pressure readings from the PACE100X:
:INST:SENS[x]:READ?Where x = 1 to 8 set in the comms range menu
:SENS:PRES?
:DISP[x]:WIND?
Returns the reading top main display window with display filtering
(2Hz update rate) applied
Where [x] = display (1=top main, 2=middle status area & 3= bottom
function area) window
To obtain fastest readings you must use the following SCPI command :INST:SENS[x]:Read?
This command returns two readings, a filtered reading and a raw reading (unfiltered
reading) from internal sensors, thus giving the user a choice of which reading to use which is
most suitable for their application.
Note: Raw reading returned from IDOS sensors is “0” as a raw readings cannot be obtained
from IDOS sensors.
Data throughput. Baud rates slower than 115k2 baud will slow the new data update rate
pro-rata. Even at a 115k2 baud rate the readings are slower than using IEEE. IDOS is a slower
sensor than internal sensors.
[EN] English6 - 9K0470 Revision A
Page 55
6 Reference and Specification
RS232
Located on the rear panel an external RS232 connection requires the following:
Connector9-way ‘D’ female wired as per Table 2-1
CommunicationsRS232 point-to-point only (daisy chain is not supported)
Baud Rate power-up default 9600, no parity & handshake = xon/xoff
Baud rates selectable **2400, 4800, 9600, 19k2, 38k4, 57k6 & 115k2
ParityNone, Odd & Even
Flow controlNone, Hardware & xon/xoff
ProtocolsPACE SCPI
Heritage emulationDPI 142/150, DPI 141
TerminatorCR or LF or CR/LF
** Selectable through the user interface.
RS232 Comms Range Setup
Note: The RS232 Comms Range values are set up by the user because of the hardware the
user desires to operate with the PACE 1000 equipment. This procedure only allows
supervisor to select these user pre-installed values. To initially define / change or
delete the RS232 Comms Range values, refer to (Ref: K0472 PACE SCPI Remote
Communications Manual).
K0470 Revision A6 - 10[EN] English
Page 56
PACE Pressure Indicators User Manual
1.On the main screen, touch any of the three horizontal touch areas on the screen.
2.On the MEASURE SETUP screen, select GLOBAL SETUP.
[EN] English6 - 11K0470 Revision A
Page 57
6 Reference and Specification
3.Select SUPERVISOR SETUP.
4.Enter the Supervisor PIN and press the top touch area. Use the back arrow in the top
right corner of the screen to delete any incorrect data entries.
K0470 Revision A6 - 12[EN] English
Page 58
PACE Pressure Indicators User Manual
5.On the SUPERVISOR screen, select COMMS.
6.On the COMMUNICATIONS screen, select COMMS RANGE SETUP.
7.On the COMMS RANGE SETUP, Use the Up and Down arrows to highlight the desired
Range (Ranges 1 to 10).
[EN] English6 - 13K0470 Revision A
Page 59
6 Reference and Specification
Comms range setup is used to re-align the SCPI index (x) to the correct pressure range.
Where Range is index (1)......etc.
8.Press the top touch area on the screen to change the range value.
9.Use the Up and down arrow to highlight the new Range value.
10.Press the top Measure on the screen to change the range value. The new range value
is set and the screen returns to the COMMS RANGE SETUP screen.
11.If necessary, repeat Steps 7 to 10 to set up other range values.
12.Press the Escape icon to go back to the COMMS RANGE SET UP screen.
13.When complete, press the Escape icon as necessary to return to the Measure screen.
IEEE
Located on the rear panel an external IEEE 488 connection requires:
Connector24-way ‘D’ female wired as IEEE 488 standard
CommunicationsIEEE488 GPIB
Default Address16
ProtocolsPACE SCPI
Heritage emulationDPI 142/150, DPI 141
Ethernet
Located on the rear panel an external Ethernet connection requires the following:
ConnectorEthernet RJ45
ProtocolSCPI
TerminatorCR/LF
Default AddressAuto IP (0.0.0.0)
Host namePACExxxxxx (where xxxxxx = serial number)
Web Password0268
Access controlOpen
Reset LAN Settings Selected in Supervisor set-up menu
Ethernet Firewall
A firewall protects the Ethernet connection. The firewall is always turned on. The following
ports are unfiltered to allow remote communication and control.
PortDescription
80/tcphttp (Web server)
111/tcprpcbind (RPC for VXI)
111/udprpcbind (RPC for VXI)
443/tcphttps (Web server)
K0470 Revision A6 - 14[EN] English
Page 60
PACE Pressure Indicators User Manual
5025/tcpSCPI socket
*****/tcpVXI-11 communication (dynamic port allocated)
Ethernet Functionality
• PACE Ethernet supports both VXI-II and Sockets automatically.
• Socket port address 5025.
• Internet protocol IP4.
Ethernet Parameters Range Setup
1.Touch any of the three horizontal Measure touch pads on the home screen to open the
CONTROL SET UP screen.
2.On the MEASURE SETUP screen, select GLOBAL SETUP.
3.Select SUPERVISOR SET UP.
[EN] English6 - 15K0470 Revision A
Page 61
6 Reference and Specification
4.Enter the Supervisor PIN and press ENTER SUPERVISOR PIN. Use the back arrow in the
top right corner of the screen to delete any incorrect data entries.
Note: The factory set Supervisor PIN is 0268. If the Supervisor PIN has been locally changed,
make sure that the new PIN is kept in a safe place. If the new PIN is lost, it can only be
reset at a GE Service Centre.
5.Press COMMUNICATIONS STATUS to open the COMMUNICATIONS STATUS screen.
6.Select ETHERNET to open the ETHERNET PARAMETER screen.
K0470 Revision A6 - 16[EN] English
Page 62
PACE Pressure Indicators User Manual
7.To change the ADDRESS parameter, complete the following:
a.On the ETHERNET PARAMETER screen, use the UP and DOWN arrows to
highlight the ADDRESS field.
b.Press the top touch pad on the screen to enter the ADDRESS TYPE screen.
c.Use the UP and Down arrows to highlight the desired address type (either
AUTO IP or STATIC).
[EN] English6 - 17K0470 Revision A
Page 63
6 Reference and Specification
d.Press the top touch pad on the screen to set the new address type. The
screen automatically returns to the ETHERNET PARAMETERS screen.
8.To change the host name, complete the following:
a.On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the HOST NAME field.
b.Press the top touch pad on the screen to enter the HOST NAME screen.
c.Use the keyboard to input the new host name and then press the top
button on the screen to set the host name. The screen automatically
returns to the ETHERNET PARAMETERS screen.
K0470 Revision A6 - 18[EN] English
Page 64
PACE Pressure Indicators User Manual
9.To change the web password, complete the following:
a.On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the WEB PASWORD field.
b.Press the top touch pad on the screen to enter the WEB PASSWORD
screen. The keyboard screen opens.
[EN] English6 - 19K0470 Revision A
Page 65
6 Reference and Specification
.
c.Use the keyboard to input the new web password and then press the top
touch area on the screen to set the new password. The screen
automatically returns to the ETHERNET PARAMETERS screen.
10.To reset the LAN settings, complete the following:
a.On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the RESET LAN SETTINGS field.
b.Press the RESET THE LAN SETTINGS touch pad on the top touch pad of the
screen.
c.The RESET LAN SETTINGS sub-screen asking for confirmation of the reset.
Press YES to confirm reset of the LAN settings.
11.To turn the LAN status indicator on or off, complete the following:
a.On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the SHOW LAN STATUS field.
K0470 Revision A6 - 20[EN] English
Page 66
PACE Pressure Indicators User Manual
b.Use the UP and DOWN arrows to highlight the desired setting. The setting
is either ON or OFF.
c.Press the top touch pad on the screen to set the new setting.
12.To change ACCESS MODE, complete the following:
a.On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the ACCESS CONTROL field.
[EN] English6 - 21K0470 Revision A
Page 67
6 Reference and Specification
b.Press the ACCESS CONTROL OPEN touch pad on the top of the screen to
open the ACCESS MODE screen.
c.Use the UP and DOWN arrows to highlight the required parameter. The
choices are OPEN or RESTRICTED.
K0470 Revision A6 - 22[EN] English
Page 68
PACE Pressure Indicators User Manual
d.Press the ACESS MODE OPEN or ACCESS MODE RESTRICTED touch pad at
the top of the screen to set the required access mode.
13.To change CONTROLLER IP ADDRESS complete the following:
a.On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the ACCESS CONTROL field.
b.Press the ACCESS CONTROL touch pad on the top of the screen.
c.Use the UP and DOWN arrows to highlight the CONTROLLER IP ADDRESS
field.
[EN] English6 - 23K0470 Revision A
Page 69
6 Reference and Specification
d.Use the number touch pad at the bottom of the screen to input the new IP
address and press the CONTROLLER IP ADDRESS touch pad at the top of
the screen to set the new IP address.
14.To view the recent IP addresses, complete the following:
a.On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the ACCESS CONTROL field.
b.Press the ACCESS CONTROL button on the top of the screen.
c.Use the UP and DOWN arrows to highlight the SELECT FROM RECENT field.
K0470 Revision A6 - 24[EN] English
Page 70
PACE Pressure Indicators User Manual
Reference
Level
d.Press the SELECT FROM RECENT button at the top of the screen to view
the IP address history.
Head Correction
Corrects pressure reading for the height difference
between instrument reference level and UUT. For
accuracy, head correction must be enabled and the
parameters set for each sensor:
•for UUT positioned higher than the reference level
of the PACE1000 enter a positive height correction.
•for UUT positioned lower than the reference level of the PACE1000 enter a negative
height correction.
[EN] English6 - 25K0470 Revision A
Page 71
6 Reference and Specification
Lock Tasks
Individual tasks
:
Allows any combination of individual tasks to be disabled.
Note: Restricts operation of the instrument to specific tasks or functions, recommended for
production procedures.
All
:
Disables all tasks.
Change PIN
Changes the Supervisor PIN: enter the existing PIN, then the new PIN and confirmation of the
new PIN.
Note: Confirmation of the new PIN permanently
replaces the old PIN. Record this new PIN and
keep in a safe place. If new PIN is lost it can only be reset by returning the instrument to
a GE service centre.
User defined units
Permits the user to define a set of units. Following the on-screen prompts special units may
be set by selecting a Pascal multiplier and assigning a five character name.
Instrument alias name
Permits the user to define a 20 character alias name for the instrument. The instrument
returns this name through the communications interfaces.
Language
Operation in any of the languages refer to the SDS. Further languages can be up-loaded.
Restore as shipped settings
Restores instrument settings to factory default.
Note: Does not affect PIN settings.
K0470 Revision A6 - 26[EN] English
Page 72
PACE Pressure Indicators User Manual
6.8Calibration
The calibration menu provides facilities for programming settings for maintenance as
follows:
Note: A PIN protects the Calibration menu against unauthorised use. Each instrument, on
delivery, contains the factory set PIN (4321). To continue protecting the supervisor setup menu, the PIN should be changed as soon as possible.
sensor correction
•Selects the range for a three-point calibration routine.
screen calibration
•Selects touch screen calibration routine.
Time & Date
•Sets instrument clock and date.
Change PIN
•Changes the Calibration PIN.
a.Enter the existing PIN
b.Enter the new PIN
c.Confirm the new PIN
Note: Confirmation of the new PIN permanently
keep in a safe place. If new PIN is lost it can only be reset by returning the instrument to
a GE service centre. For more information regarding calibration, refer to PACE
Calibration Manual K0450.
6.9Specification
Refer to the PACE1000 data sheet for details.
Note: The data sheet is contained in the CD shipped with the product.
replaces the old PIN. Record this new PIN and
[EN] English6 - 27K0470 Revision A
Page 73
6 Reference and Specification
Selects Analogue
Output Range
On/Off
Update rate of Analogue Output
option from the control module.
Analogue Output Bandwidth = 0.5 x Update Rate (Hz)
30V maximum with respect to
chassis.
Rated output = 24V
30Vmax
To maintain PACE product safety, external
circuits connected to the instrument must meet
Safety Extra-Low Voltage (SELV) requirements.
15-way female D connector
Front view of Analogue
Option PCB connector
123
4
5
6
7
8
1514131211109
30Vmax
22Vto26V
ANALOGUE1
ANALOGUE2
15
CONSULTHANDBOOK
analogue connections
6.10 Options
Option enable process
To enable soft options on a PACE instrument, use the following:
1.Touch the top Measure area of the screen.
2.Select Global Setup.
3.Select Calibration.
4.Enter a Calibration PIN 1234.
5.Enter new option key xxxxxxxxxx (10 digits).
6.After entry of this key PACE confirms the options have been enabled.
Note: Hardware options automatically enable post installation
K0470 Revision A6 - 28[EN] English
Page 74
PACE Pressure Indicators User Manual
Pin numberFunctionPin numberFunction
1not used9not used
2not used100V return
3not used11+24V DC OUT @ 100mA
4not used12SW IN 1
5not used13SW IN 2
6not used14analogue +
7not used15analogue -
8not used
[EN] English6 - 29K0470 Revision A
Page 75
6 Reference and Specification
Each selection has
three Volts-free
Contacts.
Selection of trigger
conditions.
Selection of trigger
conditions.
30V maximum with respect to
chassis.
Rated output = 24V
30Vmax
To maintain PACE product safety, external
circuits connected to the instrument must meet
Safety Extra-Low Voltage (SELV) requirements.
0V
+V
event
trigger
24 Vdc maximum
30 Vdc wrt chassis
1 A resistive maximum
Typical Volts-free
schematic
Relay contacts rated at 30 Vdc, 1 A resistive,
200 mA inductive.
22Vto26V
ANALOGUE1
ANALOGUE2
15
CONSULTHANDBOOK
volts-free connections
Volts-free Contact Option
The Volts-free Contact option provides a selectable relay contact toggle depending on conditions
set in the PACE instrument.
Pin numberFunctionPin numberFunction
1Relay 1 normally CLOSED9Relay 3 common
2Relay 1 normally OPEN100V return
3Relay 1 common11+24V DC OUT @ 100mA
4Relay 2 normally CLOSED12SW IN 1
5Relay 2 normally OPEN13SW IN 2
6Relay 2 common14not used
7Relay 3 normally CLOSED15not used
8Relay 3 normally OPEN
K0470 Revision A6 - 30[EN] English
Page 76
PACE Pressure Indicators User Manual
6.11 Installation and Ancillary Equipment Kit
Refer to the PACE1000 data sheet for details.
6.12 Return Goods/Material Procedure
If the unit requires calibration or is unserviceable return it to the nearest GE Service Centre
listed at www.gemeasurement.com
Contact the Service Department to obtain a Return Authorisation (Worldwide excluding USA).
In the USA obtain a Return Material Authorization [RMA].
Providing the following information on either a RGA or RMA:
•Product (i.e. PACE1000)
•Serial number
•Details of defect/work to be undertaken
•Calibration traceability requirements
•Operating conditions
Safety Precautions
You must inform GE if the product has been in contact with any hazardous or toxic
substance.
The relevant COSHH or in the USA, MSDS, references and precautions to be taken when
handling.
Important notice
Service or calibration by unauthorized sources will affect the warranty and may not
guarantee further performance.
6.13 Packaging Procedure
1The instrument should be at zero/ambient pressure.
2.Switch off and isolate the electrical power supply to the instrument.
3.Shut off the pneumatic pressure and vacuum supplies to the instrument.
4.Remove the instrument from the equipment rack to access the rear panel.
5.Disconnect the power supply cable and the pneumatic supply hose assemblies.
6.Stow the power supply cable in the packaging below.
7.Remove any pressure adaptors, diffusers and restrictors.
If available, use the original packing material. When using packing materials other than the
original, do the following:
8.Fit protection to all the ports to prevent ingress of moisture and dirt.
Note: Use the original red plastic plugs or low tack masking tape.
9.Wrap unit in polyethylene sheeting.
[EN] English6 - 31K0470 Revision A
Page 77
6 Reference and Specification
10.Select a double-wall cardboard container.
•Inside dimensions must be at least 15 cm (6”) greater than the equipment
•The carton must meet test strength requirements of >
11.Protect all sides with shock-absorbing material to prevent equipment movement
within the container.
12.Seal carton with approved sealing tape.
13.Mark carton “FRAGILE” on all sides, top, and bottom of shipping container.
Environment
The following conditions apply for both shipping and storage:
•Temperature range-20° to +70°C (-4° to +158°F)
125 kg (275 lbs).
K0470 Revision A6 - 32[EN] English
Page 78
www.gemeasurement.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.