GE K0470 User Manual

Page 1
GE
Digital Solutions
Pressure Automated Calibration Equipment
User manual - K0470 Revision A
PACE Indicators
© 2010 General Electric Company. All Rights Reserved. Specifications are subject to change without notice. GE is a registered trademark of General Electric Company. Other company or product names mentioned in this document may be trademarks or registered trademarks of their respective companies, which are not affiliated with GE.
Page 2
PACE Pressure Indicators User Manual
Introduction
This manual contains Installation and operating instructions for PACE Pressure Indicators.
Safety
The manufacturer has designed this equipment to be safe when operated using the procedures detailed in this manual. Do not use this equipment for any other purpose than that stated, the protection provided by the equipment may be impaired. This publication contains operating and safety instructions that must be followed to make sure of safe operation and to maintain the equipment in a safe condition. The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage. Use qualified * technicians and good engineering practice for all procedures in this publication.
Pressure
Do not apply pressures greater than the maximum working pressure to the equipment.
Toxic Materials
There are no known toxic materials used in construction of this equipment.
Maintenance
The equipment must be maintained using the procedures in this publication. Further manufacturer’s procedures should be done by an authorized service agents or the manufacturer’s service departments.
Technical Advice
For technical advice contact the manufacturer. * A qualified technician must have the necessary technical knowledge, documentation,
special test equipment and tools to carry out the required work on this equipment.
[EN] English i K0470 Revision A
Page 3
General Specification
Display LCD: Colour display with touch-screen Operating temperature 10°C to 50°C (50° to 122°F) Storage temperature -20°C to 70°C (-4° to 158°F) Ingress protection IP20 (EN60529) Operating humidity 5% to 95% RH (non-condensing) Vibration MIL-PRF-28800 Type 2 class 5 style E/F Operating altitude Maximum 2000 metres (6560ft) EMC EN 61326 Electrical safety EN 61010-1, UL61010-1, CSA 22.2, No. 61010-1 and IEC61010-1 Power adaptor Input range: 100 - 240VAC, 50 to 60Hz, 660mA. Installation category II
Pressure safety
Pressure Equipment Directive - class: sound engineering practice (SEP)
for group 2 fluids. Pollution degree 2 Operating Environment Indoor use only. Do NOT use in potentially explosive environments
Abbreviations
The following abbreviations are used in this manual; abbreviations are the same in the singular and plural.
a Absolute min Minute or minimum a.c Alternating current mm millimetre ALT Altitude mV millivolts ASCII American Standard Code for Information Interchange MWP Maximum working pressure BSP British pipe thread No Number CAS Calibrated airspeed NPT National Pipe Thread
CSK Countersunk PACE
d.c. Direct current Para. Paragraph DPI Digital Pressure Instrument PDCR Pressure transducer etc. And so on PED Pressure equipment directive e.g. For example psi Pounds per square inch Fig. Figure PTX Pressure transmitter ft Foot ROC Rate of climb (vertical speed) g Gauge RS232 Serial communications standard GPIB General purpose interface bus Rt CAS Rate of Calibrated airspeed Hg Mercury Rt MACH Rate of MACH Hz Hertz Rx Receive data
IAS Indicated airspeed SCPI
IDOS Intelligent digital output sensor (GE product) SDS Sales data sheet
Pressure automated calibration equipment
Standard commands for programmable instruments
K0470 Revision A ii [EN] English
Page 4
PACE Pressure Indicators User Manual
i.e. That is SELV
IEEE 488
in Inch UUT Unit under test kg kilogram V Volts
kts knots +ve Positive m Metre -ve Negative mA milliampere °C Degrees Celsius max Maximum °F Degrees Fahrenheit mbar Millibar
Institute of Electrical and Electronic Engineers standard 488 (for programmable devices with a digital interface)
Tx Transmit data
Separated (or Safety) extra low voltage
Related publications
K0467 User Guide and Safety Instructions
K0469 PACE Heritage Communications Manual
K0450 PACE Series Calibration Manual
K0472 PACE Series SCPI Manual
[EN] English iii K0470 Revision A
Page 5
Symbols
This equipment meets the requirements of all relevant European safety directives. The equipment carries the CE mark.
This symbol, on the equipment, indicates that the user should read the user manual.
This symbol, on the equipment, indicates a warning and that the user should refer to the user manual.
Ce symbole, sur l’instrument, indique que l’utilisateur doit consulter le manuel d’utili­sation. Ce symbole, dans le manuel, indique une situation dangereuse.
This symbol warns the user of the danger of electric shock. Ce symbole alerte l’utilisateur sur le danger de choc électrique.
Do not dispose of this product as household waste. Use an approved organisation that collects and/or recycles waste electrical and electronic equipment. For more information, contact one of these:
- Our customer service department: www.gemeasurement.com
- Your local government office.
WARNINGS TURN OFF THE SOURCE PRESSURE(S) AND CAREFULLY VENT THE PRESSURE LINES BEFORE
DISCONNECTING OR CONNECTING THE PRESSURE LINES. PROCEED WITH CARE. ONLY USE EQUIPMENT WITH THE CORRECT PRESSURE RATING. BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE.
REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT. DO NOT USE ANY DAMAGED FITTINGS AND EQUIPMENT.
DO NOT EXCEED THE MAXIMUM WORKING PRESSURE OF THE INSTRUMENT. THIS EQUIPMENT IS NOT RATED FOR OXYGEN USE.
K0470 Revision A iv [EN] English
Page 6
PACE Pressure Indicators User Manual
THE GROUND LEAD OF THE INSTRUMENT MUST BE CONNECTED TO THE AC SUPPLY PROTECTIVE SAFETY GROUND.
ISOLATE THE POWER SUPPLY BEFORE MAKING ANY ELECTRICAL CONNECTIONS TO THE REAR PANEL.
Pressure units and conversion factors
Pressure units Factor (hPa) Pressure units Factor (hPa)
mbar 1.0
bar 1000.0
cmH
mH
O @ 20°C
2
O @ 20°C
2
0.978903642
97.8903642
Pa (N/m2)
hPa 1.0
kPa 10.0 torr 1.333223684
MPa 10000.0 atm 1013.25
mmHg @ 0°C 1.333223874 psi 68.94757293
cmHg @ 0°C 13.33223874
mHg @ 0°C 1333.223874
inHg @ 0°C 33.86388640341
mmH
cmH2O @ 4°C
mH
O @ 4°C
2
O @ 4°C
2
0.01
0.0980665
0.980665
98.0665
2
kg/m
kg/cm
lb/ft
inH
inH
inH2O @ 60°F
ftH2O @ 4°C
ftH2O @ 20°C
2
2
O @ 4°C
2
O @ 20°C
2
0.0980665
980.665
0.4788025898
2.4908891
2.486413
2.487641558
29.8906692
29.836983
mmH
Unit Conversion
O @ 20°C
2
0.097890364
Convert FROM pressure VALUE 1 in pressure UNITS 1 TO pressure VALUE 2 in pressure UNITS 2, calculate as follows:
[EN] English v K0470 Revision A
ftH2O @ 60°F
29.8516987
VALUE 2 = VALUE 1 x FACTOR 1
FACTOR 2
Page 7
CONTENTS
Section Title Page
1 Description 1-1
1.1 Introduction 1-1
2 Installation 2-1
2.1 Packaging 2-1
2.2 Packaging for Storage and Transportation 2-1
2.3 Preparation for Use 2-1
2.4 Connecting the Instrument 2-2
2.5 Mounting kits 2-8
2.6 Electrical connections 2-10
3 OPERATION 3-1
3.1 Preparation 3-1
3.2 Power-up Sequence 3-1
3.3 Measure Mode 3-2
3.4 Operation and Example Procedures 3-8
3.5 Global Set-up Selections 3-11
3.6 Supervisor Set-up 3-12
3.7 Instrument Status 3-13
4 MAINTENANCE 4-1
4.1 Introduction 4-1
4.2 Visual inspection 4-1
4.3 Cleaning 4-1
K0470 Revision A vi [EN] English
Page 8
PACE Pressure Indicators User Manual
4.4 Test 4-1
4.5 Calibration 4-1
5 TESTING AND FAULT FINDING 5-1
5.1 Introduction 5-1
5.2 Standard Serviceability Test 5-1
5.3 Ethernet testing 5-1
5.4 Fault Finding 5-4
5.5 Approved Service Agents 5-4
6 REFERENCE AND SPECIFICATION 6-1
6.1 Installation notes 6-1
6.2 Reference port 6-2
6.3 Icons 6-3
6.4 Measure Set-up 6-6
6.5 Status 6-7
6.6 Global Set-up 6-8
6.7 Supervisor Set-up 6-9
6.8 Calibration 6-27
6.9 Specification 6-27
6.10 Options 6-28
6.11 Installation and Ancillary Equipment 6-31
6.12 Return Goods/Material Procedure 6-31
6-13 Packaging Procedure 6-31
[EN] English vii K0470 Revision A
Page 9
Intentionally blank
K0470 Revision A viii [EN] English
Page 10
PACE Pressure Indicators User Manual
1Description
1.1 Introduction
The PACE Pressure Indicator measures both pneumatic and hydraulic pressures and displays, on a colour touch-screen, the measured pressure and instrument status. The touch-screen enables selections and settings in measuring modes. The instrument can be operated remotely through communication interfaces.
Figure 1-1 PACE1000 General view
The rear of the instrument houses all the electrical and pressure input connections. The electrical connections provide a power supply, serial and parallel communication interfaces, pressure ports and option ports.
The instrument can be used as follows:
Free-standing instrument positioned on a horizontal surface.
Rack-mounted in a standard 19 inch rack using the rack-mount option kit.
Panel mount using the panel-mount option kit.
[EN] English 1 - 1 K0470 Revision A
Page 11
1 Description
Options available for the PACE1000 refer to the data sheet.
Information and notes on applications (Ref: Reference and Specification, Section 6) or www.gemeasurement.com
K0470 Revision A 1 - 2 [EN] English
Page 12
PACE Pressure Indicators User Manual
2 Installation
2.1 Packaging
Check the contents of the PACE1000 packaging with the list that follows:
Packaging List - PACE1000
i) PACE1000 Pressure Indicator. ii) Adaptor, power supply (GE part number 191-370). iii) User guide and safety instructions, and CD containing the full documentation suite. iv) Calibration certificate.
2.2 Packaging for Storage or Transportation
To store or return the instrument for calibration/repair do the procedures that follow:
1. Pack the instrument (Ref: Reference and Specification, Section 6.13).
2. Return the instrument for calibration/repair complete the return goods procedure (Ref: Reference and Specification, Section 6.12).
2.3 Preparation for Use
The instrument can be used as a:
Free-standing instrument positioned on a horizontal surface.
Panel-mounted using the panel-mount option kit (Ref: Section 2.5).
Rack-mounted in a standard 19 inch rack using the rack-mount option kit (Ref: Section
2.5).
For free-standing instruments, the feet on the front of the base can be used elevate the instrument to a better viewing angle.
Note: Allow a free flow of air around the instrument, especially at high ambient temperatures.
[EN] English 2 - 1 K0470 Revision A
Page 13
2 Installation
ISO 228 G1/8 ISO 228 G1/8
bonded seal bonded seal
Recommended method Alternative method below 100 bar
2.4 Connecting the Instrument
WARNINGS
TURN OFF THE SOURCE PRESSURE(S) AND CAREFULLY VENT THE PRESSURE LINES BEFORE DISCONNECTING OR CONNECTING THE PRESSURE LINES. PROCEED WITH CARE.
ONLY USE EQUIPMENT WITH THE CORRECT PRESSURE RATING.
BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE. REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT. DO NOT USE ANY DAMAGED FITTINGS AND EQUIPMENT.
DO NOT EXCEED THE MAXIMUM WORKING PRESSURE OF THE INSTRUMENT.
THIS EQUIPMENT IS NOT RATED FOR OXYGEN USE.
Pneumatic Pressure (Figure 2-1)
1. Refer to the Data sheet for the correct pressure mediums to be used.
2. Connect the Unit Under Test (UUT) to the required connection port.
Note: For instruments with NPT connections, use applicable bonded sealing as shown in figure
below.
Figure 2-1, Sealing Pneumatic Connections
K0470 Revision A 2 - 2 [EN] English
Page 14
PACE Pressure Indicators User Manual
Parallel threads must be used. Female thread type is parallel thread to ISO228/1 (DIN ISO228/1, JIS B0202) G1/8
Tapered threads NOT allowed.
PACE pressure port
PACE rear panel
9
8
PACE rear panel
PACE pressure port
Pneumatic connections
WARNING
PARALLEL THREADS MUST BE USED. FEMALE THREAD TYPE IS PARALLEL THREAD TO ISO228/1 (DIN ISO228/1, JIS B0202) G1/8.
TAPERED THREADS NOT ALLOWED.
For examples of adaptors (Ref: page 2 - 4).
Input Pressure (Figure 2)
Note: For instruments with NPT connections, use applicable bonded sealing as shown in figure
[EN] English 2 - 3 K0470 Revision A
Connection
Input ISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202) Reference ISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)
1. Make sure the user systems can be isolated and vented.
2. Connect the Unit Under Test (UUT) to the output connection port.
below.
Page 15
2 Installation
2) Bonded seal
1) Connector
2
1
Figure 2-2, Pneumatic Connections
Installation
The instrument connects to the Unit Under Test.
Input Pressure and Equipment
The pressure should not exceed 1.25 x full-scale or MWP stated on the rear panel of the instrument.
To protect the instrument from over-pressure a suitable protection device (such as a relief valve or bursting disc) must be fitted to prevent over pressurization.
Pneumatic Connection
WARNING
PRESSURE RANGES > 210
BAR (3000 PSI) ARE ONLY RATED FOR HYDRAULIC USE.
CAUTIONS
Do not exceed the maximum pressures stated in the appropriate Component Manual for the unit under test.
Reduce pressure at a controlled rate when venting to atmosphere.
Carefully de-pressurize all pipes to atmospheric pressure before disconnecting and connecting to the unit under test.
Connections
1. Switch off the power supply before connecting or disconnecting the instrument.
2. Use the appropriate sealing method for all pressure connections.
Note: For instruments with NPT connections, use applicable bonded sealing as shown in figure
below.
K0470 Revision A 2 - 4 [EN] English
Page 16
PACE Pressure Indicators User Manual
ISO 228 G1/8 ISO 228 G1/8
bonded seal bonded seal
Recommended method Alternative method below 100 bar
IO-SNUBBER-1
IO-ADAPT-9/16AUTOC ISO 228 Male to 9/16 Autoclave
IO-DIFFUSER-1
IO-ADAPT-G1/4 ISO 228 Male to G1/4 Female
IO-ADAPT-1/8NPT ISO 228 Male to 1/8 NPT Female
IO-ADAPT-1/4NPT ISO 228 Male to 1/4 NPT Female
IO-ADAPT-7/16UNF ISO 228 Male to 7/16 UNF Female
IO-ADAPT-AN4 ISO 228 Male to AN4 37° Male
IO-ADAPT-AN6 ISO 228 Male to AN6 37° Male
IO-ADAPT-BARB ISO 228 Male to 1/4 Hose
Method of connection
Adaptors
Refer to the data sheet for the range of adaptors.
[EN] English 2 - 5 K0470 Revision A
Page 17
2 Installation
Hydraulic Pressure
WARNINGS
HYDRAULIC LIQUID IS DANGEROUS. OBSERVE RELEVANT HEALTH AND SAFETY PRECAUTIONS. USE APPROPRIATE PROTECTIVE BARRIERS AND EYE PROTECTION.
BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE AND ENSURE THAT ALL EQUIPMENT IS TO THE CORRECT PRESSURE RATING.
DO NOT EXCEED THE MAXIMUM WORKING PRESSURE OF THE INSTRUMENT.
PURGE ALL AIR FROM THE HYDRAULIC LIQUID.
PRESSURE RANGES > 210
BAR (3000 PSI) ARE ONLY RATED FOR HYDRAULIC USE.
DO NOT USE A SENSOR FOR GAS THAT HAS BEEN USED WITH HYDRAULIC LIQUID.
CAUTIONS
Do not exceed the maximum pressures stated in the appropriate component manual for the unit under test.
Reduce pressure at a controlled rate when venting to atmosphere.
Carefully de-pressurize all pipes to atmospheric pressure before disconnecting and connecting to the unit under test.
Observe absolute cleanliness when using the instrument.
Severe damage can be caused if equipment connected to this instrument is contaminated.
Connect only clean equipment to the instrument.
To avoid any contamination, an external filter is recommended.
K0470 Revision A 2 - 6 [EN] English
Page 18
PACE Pressure Indicators User Manual
ISO 228 G1/8 ISO 228 G1/8
bonded seal bonded seal
Recommended method Alternative method below 100 bar
Installation
The instrument connects to the Unit Under Test.
Input Pressure and Equipment
1. The pressure should not exceed 1.25 x full-scale or MWP stated on the rear panel of the instrument.
2. To protect the instrument from over-pressure a suitable protection device (such as a relief valve or bursting disc) must be fitted to limit the pressure to below the MWP.
Note: For instruments with NPT connections, use applicable bonded sealing as shown in figure
below.
Figure 2-3, Sealing Hydraulic Connections
Hydraulic connections
Connection
Input ISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202) Reference ISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)
Note: Pressure connections greater than 210 bar are 9/16” 18UNF Male Autoclave.
1. Switch off the power supply before connecting or disconnecting the instrument.
2. Use the applicable sealing method for all pressure connections.
3. Isolate the hydraulic pressures and de-pressurise the pipes before connecting or disconnecting the instrument.
Pressure input (Figure 2)
1. Make sure the user systems can be isolated and vented.
2. Use the applicable sealing method for all pressure connections.
3. The hydraulic liquid must be clean, refer to specification given in the Data sheet.
4. Connect the Unit Under Test (UUT) to the appropriate connection port.
[EN] English 2 - 7 K0470 Revision A
5. Fill and bleed the UUT and connecting pipes.
Page 19
2 Installation
2.5 Mounting kits
Rack-mount option (Figure 2-5)
There must be enough space at the rear of the instrument for all the cables and pipes. The length of the cables and pipes must allow for the removal and installation of the instrument. The cooling air of the instrument must not be obstructed. Allow a free flow of air through the equipment rack and around the instrument, especially at high ambient temperatures.
Procedure
1. Locate instrument in rack mount assembly
2. Secure with the four M3 x 6 screws , (maximum length M3 x 8).
3. Support the instrument and connect the cables and pipes.
4. Refer to the electrical connections below before fitting the instrument into the equipment rack.
5. Temporarily locate the two spigots* to each side of the equipment rack.
6. Locate and slide the instrument into the rack.
7. Locate the instrument on the spigots*.
K0470 Revision A 2 - 8 [EN] English
Figure 2-5 Rack-mounting
.
Page 20
PACE Pressure Indicators User Manual
8 Secure the instrument in the equipment rack with two of the screws and washers
(supplied).
9. Remove the two spigots* and replace with the remaining two screws and washers (supplied).
Panel-mount option (Figure 2-6)
There must be enough space at the rear of the instrument for all the cables and pipes. The length of the cables and pipes must allow for the removal and fitment of the instrument. The cooling air of the instrument must not be obstructed. Allow a free flow of air through the equipment rack and around the instrument, especially at high ambient temperatures.
Procedure
1. Remove the four screws
2. Locate the instrument in panel mount assembly.
3. Secure with the four screws
4. Support the instrument and connect the cables and pipes.
5. Refer to the electrical connections below before fitting the instrument into the panel.
6. Secure the instrument in the panel with four screws and washers
[EN] English 2 - 9 K0470 Revision A
Figure 2-6 Panel-mounting
from the instrument.
.
.
Page 21
2 Installation
2.6 Electrical connections
WARNINGS
THE GROUND LEAD OF THE INSTRUMENT MUST BE CONNECTED TO THE AC SUPPLY PROTECTIVE SAFETY GROUND.
ISOLATE THE POWER SUPPLY BEFORE MAKING ANY ELECTRICAL CONNECTIONS TO THE REAR PANEL.
CAUTIONS
Use the power adaptor supplied with the instrument (GE part no. 191-370). Using other power adaptors may cause over-heating, this can result in a fire.
Do not let the power adaptor come into contact with any moisture or liquids.
Connecting (Figure 2-7)
1. Before use, make sure the SELV power adaptor supplied with the instrument is used (GE part number 191-370).
2. Install an accessible power isolator to use as the disconnecting device in the power adaptor supply circuit.
3. The power adaptor input power supply range: 100 - 240VAC, 50 to 60Hz 660mA, Installation Category II.
Note: The power adaptor must be supplied by a fused or overload-protected power
supply.
4. Connect the power adaptor to the instrument.
5. Switch the power supply on.
6. Check that the front panel display shows the power-up sequence (Ref: section 3.2).
Note:. After the power-up sequence, the instrument shows the default display on the
touch screen. The touch screen divides into a number of mimic keys.
Requirements for rack-mounted and panel-mounted instruments
1. Install an accessible power isolator to use as the disconnecting device in the power adaptor supply circuit.
2. Set the power supply isolator to OFF.
3. Connect the power adaptor before sliding the instrument into the rack.
4. Set the power supply isolator to ON.
5. Check that the front panel display shows the power-up sequence (Ref: section 3.2).
K0470 Revision A 2 - 10 [EN] English
Page 22
PACE Pressure Indicators User Manual
2
4
5
6
7
22V to 26V
ANALOGUE 1
CONSULT HANDBOOK
1
Communication Connections
Connect the applicable connectors into the rear panel communications ports and, if appropriate, secure with the captive screws.
Note: The RS232 and IEEE 488 interfaces are both enabled at power-up. Set the required parameters
in Supervisor Setup/communications menu, see Section 3.6.
Figure 2-7, Communication Connectors
1
Power supply adaptor 2 RS232 4 IEEE488
5
USB B 6 USB A 7 Ethernet
RS232 Interface
When using the RS232 interface, a cable must be connected directly from the instrument to a suitable port on the computer in a ‘point to point’ link. The pin connections for the 9-pin D-type, RS232 connector and the relationship between the instrument and the RS232 control signals, together with device interconnection interface is shown in Table 2-1. The instrument is configured as Data Circuit Terminating Equipment (DCE).
[EN] English 2 - 11 K0470 Revision A
Page 23
2 Installation
Instrument Control Line Computer/Printer
Instrument
Function
RxD (I/P) 3 TxD 3 2
TxD (O/P) 2 RxD 2 3
GND 5  GND 5 7
CTS (I/P) 7 RTS 7 4
RTS (O/P) 8 CTS 8 5
Pulled
high
internally
Not
connected
Pulled
high
internally
Equipment
chassis
Connector
9-way D-type Pin No.
1
4 DTR 4 20
6 
Connector
shell
Signal
Direction
 Cable Screen - 1
RS232
Terminology
RLSD
(DCD)
DSR
DCE Ready
Connector Type
9-way D-type Pin No.
18
66
25-way
D-type Pin No.
Table 2-1, RS232 Connections
Handshaking connections
Software handshaking use: TXD, RXD and GND. Hardware handshaking use: TXD, RXD, GND, CTS, RTS and DTR.
K0470 Revision A 2 - 12 [EN] English
Page 24
PACE Pressure Indicators User Manual
2
3
IEEE 488 Interface
The interface complies with IEEE 488 standard. The IEEE 488 parallel interface connects a computer/controller to one or more PACE1000
instruments and other instruments. Up to 30 instruments can be connected through a high-speed data bus to the computer/controller.
Note: The length of each IEEE 488 cable must be less than 3 metres to comply with the EMC
requirements (Ref: Data sheet).
Single Unit Installation (Figure 2-8)
1. Connect an IEEE 488 connector/cable assembly to the rear panel of the instrument.
2. Connect the other end of the connector/cable assembly to the IEEE 488 connector on the controller/computer.
3. Change the IEEE 488 communication parameters (Ref: Supervisor set-up, Section 6.7).
Multiple Unit Installation (Figure 2-8)
To install multiple units use stacking plugs to link the first instrument and second instrument as follows:
1 Connector to rear panel of first
instrument (Ref Illustration).
2 Connector from controller/computer
(Ref Illustration).
3 Connector to rear panel of second
instrument (Ref Illustration).
4. Connect the IEEE 488 connector on the controller/computer and the other connector into the next instrument.
5. Repeat this procedure for all the instruments in the system.
6. Use the Supervisor set-up (communications) menu on each instrument to set-up the required communication parameters (Ref: Section 3.8).
1
[EN] English 2 - 13 K0470 Revision A
Page 25
2 Installation
IEEE 488
ADDRESS 1
ADDRESS 2
ADDRESS N
(30 maximum)
6
7
8
DAV (DATA VALID)
NRFD (NOT READY FOR DATA)
NDAC (NO DATA ACCEPTED)
5
9
101117
EOI (END OF IDENTIFY)
IFC (INTERFACE CLEAR)
SRQ (SERVICE REQUEST)
ATN (ATTENTION)
REN (REMOTE ENABLE)
1
2
3
4
13
DIO1
DIO2
DIO3
DIO4
DIO5
14
15
16
DIO6
DIO7
DIO8
1819202122
GND (6)
GND (7)
GND (8)
GND (9)
GND (10)
23
24
GND (11)
GND
DATA /
STATUS
BAR
0V (GND)
12 CHASSIS/FRAMECHASSIS/FRAME
K0470 Revision A 2 - 14 [EN] English
Figure 2-8 - IEEE 488 Connection
Page 26
PACE Pressure Indicators User Manual
Do not touch the display screen during power-up
3Operation
This section contains quick reference charts detailing all the available functions and the set-up menu.
3.1 Preparation
Make sure the electrical cables and pneumatic pipes comply with the installation requirements (Ref: Section 2).
Before use do the following:
1. If necessary, do the maintenance task (Ref: Section 4).
2. For bench-top, single instrument operation do the following:
a. Connect the instrument to the electrical supply.
b. Inspect the pneumatic hoses for damage, ingress of dirt and moisture.
3. Before use, the instrument should be tested.
4. Review and become familiar with the procedure before starting a process on a component or system.
3.2 Power-up sequence
The following sequences of operation shows the instrument display.
Note: The following sequence is an example, the values and selections displayed depend on the
range(s) and options enabled in the instrument.
1. Set the power supply to ON.
2. The display shows the power-up sequence.
3. The instrument carries out a self-test. a. If the test finds a fault, the display
shows an error (Ref: Fault Finding and Testing, Section 5).
4. If the self-test is successful the system enables the touch screen and changes to measure mode.
5. The touch screen shows the measured pressure in the parameters selected in set-up.
6. The instrument is now ready for use.
[EN] English 3 - 1 K0470 Revision A
Page 27
3 Operation
1
2
3
4
5
6
3.3 Measure mode
Touch screen areas
1 Pressure reading 2 Functions enabled 3 Zero key (vent system before starting zero sequence) 4 Function area 5 Status area 6 Current pressure range
Display Icons
Tare enabled Filter pressure reading
Percentage Ethernet not connected
Reference level difference (gas head correction)
Ethernet connected
K0470 Revision A 3 - 2 [EN] English
Page 28
PACE Pressure Indicators User Manual
1
Function area Min/Avg/Max is performed on the reading selected for display in the top screen.
The Ethernet LAN Status indication (1) shows the following:
• Colour red - not connected
• Colour green - connected
[EN] English 3 - 3 K0470 Revision A
Page 29
3 Operation
1
Data Logging
The “Datalog” icon (1) is present when a memory card is fitted:
K0470 Revision A 3 - 4 [EN] English
Page 30
PACE Pressure Indicators User Manual
[EN] English 3 - 5 K0470 Revision A
Page 31
3 Operation
K0470 Revision A 3 - 6 [EN] English
Page 32
PACE Pressure Indicators User Manual
current pressure reading to zero, offset stored for current range.
allows selection of available pressure ranges. Ê
-
-
select from list of available pressure measurement units.
-
Absolute range selection available when barometric option installed.
Ê
% Filter pressure reading Tare
Pressure zero
Range
Process
Task
Units
Global set-up
Status
Measure set-up
More
Next page of menu options. Loops from last page to first page.
Stores settings and exits set-up.
see 3.6:-
next page:-
Exit set-up.
Peak
Measure Menu Set-up
[EN] English 3 - 7 K0470 Revision A
Page 33
3 Operation
Selecting a task exits the menu and changes the display to the task selected.
Ê
Exit set-up.
Option refer to data sheet
Ê
Basic
Airfield
Task
Return to home.
3.4 Operation and Example Procedures
Introduction
Before operation, the instrument must be connected to the correct electrical and pneumatic/ hydraulic) supplies, (Ref: Installation,Section 2).
When the instrument is switched ON the display shows measured pressure mode and the task set before the power-off.
Measure Mode
The instrument works as a precision pressure indicator and shows the pressure measured at the output port.
Pressing Task enables pre-determined functions:
Task
The display shows the task screen (Ref: Illustration above). When selected, e.g. Basic, the screen changes to show the selected task. To measure pressure in the task proceed as follows:
1. Select the required units of pressure measurement from the measure set-up menu.
K0470 Revision A 3 - 8 [EN] English
Page 34
PACE Pressure Indicators User Manual
QNE QNH QFE
29.9212 inHg (1013.25 kPa) Datum
Sea Level Reference Datum
(adjusted)
Runway
(on Earth’s surface)
Note: In the Airfield Task Q codes can be used. These units are standardised three letter codes and
are available in aeronautical units (feet and metres). The codes used are:
QFE - Atmospheric pressure at sea level, corrected for temperature and adjusted to airfield elevation. When set on the altimeter it reads height.
QNE - Atmospheric pressure at Sea level in International Standard Atmosphere (ISA)
1013.25 mbar.
QFF - Barometric pressure at a place, reduced to Mean Sea Level (MSL) using the actual temperature at time of observation as mean temperature.
QNH - Atmospheric pressure at Mean Sea Level (MSL) (may be local, measured pressure or a Regional Forecast Pressure (RFP). When set on altimeter it reads altitude.
[EN] English 3 - 9 K0470 Revision A
Page 35
3 Operation
Start
Test time
Leak test
Settings
Leak rate
/second
/minute
Leak testing option
This task, measures the leak rate over the measure dwell time. At the start of the test, the instrument measures the test pressure of the user system. The
instrument then records the pressure change during measure dwell time. On completion, the display shows the leak rate results with leak rate per second or per minute in
the current pressure units selected in measure set-up.
K0470 Revision A 3 - 10 [EN] English
Page 36
PACE Pressure Indicators User Manual
1
2
Supervisor set-up
Calibration
Global set-up
Save/recall user set-up
Save current user set-up
Recall current user set-up
Display
PIN - enter four digit code (4321)
- % brightness
- Timer (time-out) setting
Resolution - Measure display resolution
Back light
Audio volume - % of full volume Satus area
Display Mode
Function area
Reading
Graph
3.5 Global Set-up Selections
Global set-up selections provide access to the instrument‘s settings for both measure and control modes.
This set-up menu provides PIN-protected access to the supervisor set-up and calibration. Pressing Global Set-up changes the touch-screen display to show available selections.
Supervisor Set-up, Calibration, Save/Recall User Set-up and Display.
1 Selections 2 Escape Key
[EN] English 3 - 11 K0470 Revision A
Page 37
3 Operation
enter four digit code (0268)
-
-
Alarms
Communications
Restore last settings
Supervisor
More
next page:-
PIN
Pressing the escape key stores settings and returns to global set-up.
Enable/disable high pressure
Enable/disable low pressure
IEEE488 RS232 USB Ethernet
Gas head correction
Enable/
disable
Head - in or m Local gravity - m/s
2
Barometric - mbar Reference port medium* Test port medium* Temperature (gas) Humidity (gas)
*air or nitrogen
Lock
Lock/unlock individual tasks
Lock/unlock all tasks
Lock tasks
Change PIN
Enter new PIN
Re-enter new PIN
Enter new supervisor PIN
User defined units
Enter the name of the defined units
Enter the equivalent value in Pascals
User defined units ­user unit 1
Instrument alias name
Language
Restore previous defaults Y/N?
Rate set-up
/second
/minute
Area of Use
Europe North America Japan Asia Rest of World
Additional selections for options enabled: Aeronautical etc.
3.6 Supervisor Set-up
K0470 Revision A 3 - 12 [EN] English
Page 38
PACE Pressure Indicators User Manual
Instrument
Software installed
Hardware build
Instrument Status
IEEE488*
Instrument Main Code Instrument OS Build Instrument Boot ROM Analogue O/P Main Code Analogue O/P Boot ROM VFC Main Code VFC Boot ROM
Additions to the standard instrument
Pressing the escape key stores settings and returns to global set-up.
History
Calibration Zero Software Hardware Message
Communications
RS232*
* current settings, see 6.7
Current set-up
Support
Contact details
Ethernet (optional)
USB Ethernet
More
next page:-
3.7 Instrument Status
The control set-up menu provides access to the status of the instrument:
[EN] English 3 - 13 K0470 Revision A
Page 39
3 Operation
Exit set-up.
Instrument Main Code
Software history
Analogue O/P Main Code
Instrument OS Build
Instrument Boot ROM
Analogue O/P Boot ROM
VFC Main Code
VFC Boot ROM
Example
Additional selections for options
enabled: Analgue etc.
Software
Software history, in the status menu, provides read only information on the current software in the instrument.
K0470 Revision A 3 - 14 [EN] English
Page 40
PACE Pressure Indicators User Manual
4 Maintenance
4.1 Introduction
This section contains procedures for routine maintenance and the replacement of components (Ref: Testing and Fault Finding, Section 5).
Table 4.1 - Maintenance Tasks
Task Period
Visual Inspection Before use Test Before use Cleaning Weekly* Calibration 12 months †
* may change depends on usage (e.g., rack mounted, bench top) and environment (e.g., humidity, dust). † may change depends on the required accuracy.
4.2 Visual Inspection
Inspect for obvious signs of damage and dirt on the following:
a. External of the instrument. b. Power supply adaptor c. Associated equipment.
Damaged parts must be replaced contact GE Service.
For cleaning (Ref: Cleaning Section 4.3).
4.3 Cleaning
CAUTION Do not use solvents for cleaning.
Clean the front panel with a damp lint-free cloth and mild detergent.
4.4 Test
Do a standard serviceability test (Ref: Standard Serviceability Test, Section 5.2).
4.5 Calibration
The instrument should be returned to the manufacturer or calibration facility, (Ref: Section 6.14).
To find the date of the last calibration, press Measure set-up/Status/Calibration history.
[EN] English 4 - 1 K0470 Revision A
Page 41
4 Maintenance
Intentionally blank
K0470 Revision A 4 - 2 [EN] English
Page 42
PACE Pressure Indicators User Manual
5 Testing and Fault Finding
5.1 Introduction
This section details the standard serviceability test. Table 5.1 lists possible faults, and the response.
The PACE1000 contains a self-test and diagnosis system that continuously monitors the performance of the unit. At power-up, the system performs a self-test.
5.2 Standard Serviceability Test
The following procedure shows if the unit is serviceable and checks functions and facilities of the PACE1000.
Procedure
CAUTION Always release pressure before disconnecting pressure equipment from the outlet port.
1. Connect the instrument (Ref: Installation, Section 2) Connect a UUT.
2. After power-up, select measure set-up. a. Select the required units of pressure measurement from the measure set-up
menu.
b. Apply a known pressure to one of the sensors. Make sure the instrument
pressure reading is within tolerance, stated in the specification (Ref: SDS). c. Carefully release the applied pressure to atmospheric pressure. d. Make sure the instrument pressure reading shows atmospheric or ambient
pressure. e. Test complete.
After a successful serviceability test the instrument is ready for use.
5.3 Ethernet Testing
1. Connect PACE Ethernet port to a PC.
2. Check the PACE Ethernet LAN indication turns green after a few seconds.
3. Set the PACE Ethernet address to Auto IP.
4. Record the PACE auto IP address (The test screen below has auto IP address 3.115.21.237)
5. Open the PC command prompt screen.
6. Using the “ping” command, ping the PACE IP address (see screen capture below). The PACE will reply if operating correctly.
[EN] English 5 - 1 K0470 Revision A
Page 43
5 Testing and Fault Finding
Ethernet open ports and their usage: 80/tcp PACE Web server (http) 111/tcp ONC RPC port mapper 372/tcp VXI-11 communication (Dynamically allocated) 443/tcp Web server (https) 5025/tcp PACE SCPI communication socket 111/udp ONC RPC port mapper
K0470 Revision A 5 - 2 [EN] English
Page 44
PACE Pressure Indicators User Manual
Ethernet testing using a web browser
[EN] English 5 - 3 K0470 Revision A
Page 45
5 Testing and Fault Finding
Ethernet testing using NI
http://www.ni.com/download/ni-daqmx-14.1/4953/en/
5.4 Fault Finding
Check the faults and responses (Ref: Table 5.1 Fault Diagnosis) before contacting www.gemeasurement.com or a recommended Service Agent.
Fault Response
Power supply connected, display not lit. Check electrical power supply fuse or circuit breaker. Display pressure reading in red Over-range, carefully de-pressurize.
Vent system pressure.
Instrument will not zero.
Check for blockage. Contact approved service agent for repair.
Table 5.1 - Fault Diagnosis
5.5 Approved Service Agents
For the list of service centres logon to www.gemeasurement.com
K0470 Revision A 5 - 4 [EN] English
Page 46
PACE Pressure Indicators User Manual
6 Reference and Specification
6.1 Installation notes
The PACE1000 pressure indicator requires a set of connections with the exception of the
reference connection, this provides a reference to atmosphere for gauge sensors and barometric sensors.
The gas density and type does not affect the accuracy of pressure measurement, assuming
that the UUT is at the same level (height) as the indicator or gas head correction is accurately set.
-3
Values of air density (kg m
dioxide by volume.
) for air of relative humidity 50% and containing 0.04% carbon
Table 6-1 Air Density Values
Air
pressure
(kPa)
87 88 89 90 91 92 93 94 95 96 97 98 99
100
Air temperature (°C)
14 16 18 20 22 24 26
1.052 1.045 1.037 1.029 1.021 1.014 1,006
1.064 1.057 1.049 1.041 1.033 1.025 1.018
1.077 1.069 1.061 1.053 1.045 1.037 1.029
1.089 1.081 1.073 1.065 1.057 1.049 1.041
1.101 1.093 1.085 1.077 1.069 1.061 1.053
1.113 1.105 1.097 1.089 1.080 1.072 1.064
1.125 1.117 1.109 1.100 1.092 1.084 1.076
1.137 1.129 1.121 1.112 1.104 1.096 1.088
1.149 1.141 1.133 1.124 1.116 1.108 1.099
1.162 1.153 1.145 1.136 1.128 1.119 1.111
1.174 1.165 1.156 1.148 1.139 1.131 1.123
1.186 1.177 1.168 1.160 1.151 1.143 1.134
1.198 1.189 1.180 1.172 1.163 1.154 1.146
1.210 1.201 1.192 1.184 1.175 1.166 1.158
101 102 103 104 105 106
1.222 1.213 1.204 1.196 1.187 1.178 1.169
1.234 1.225 1.216 1.207 1.199 1.190 1.181
1.247 1.237 1.228 1.219 1.210 1.201 1.193
1.259 1.249 1.240 1.231 1.222 1.213 1.204
1.271 1.261 1.252 1.243 1.234 1.225 1.216
1.283 1.274 1.264 1.255 1.246 1.237 1.228
Note: 100 kPa = 1 bar
[EN] English 6 - 1 K0470 Revision A
Page 47
6 Reference and Specification
6.2 Reference Port
The reference port provides the negative pressure to the gauge sensor and to the barometric reference (option). Gauge sensors use this port identified as “REF”. For gauge sensors (without a barometric reference) small pressures can be applied (Ref: Specification, Section 6-7). All other pressure measurement requires the port to be opened to atmosphere. When in gauge mode, the instrument shows and controls the pressure difference between the reference port and the output port.
Note: This is not a true differential operation as there is no true differential calibration of the
sensor.
The transducer of the barometric reference option senses atmospheric pressure via the reference port, when enabled the port MUST be open to atmosphere. The reference connection should be actively used (differential connection option) for precision low pressure measurement . The instrument measures pressure relative to the pressure at the reference port.
An atmospheric pressure change causes the indicator to adjust the pressure and appears at the pressure output as apparent instability. To keep a stable controlled pressure, the reference port should be restricted. Using a reference port restrictor (snubber), short term ambient pressure variations can be prevented from affecting indicator performance. The indicator and UUT references should be connected together (using the optional differential connection kit) to provide a common reference to atmosphere.
K0470 Revision A 6 - 2 [EN] English
Page 48
PACE Pressure Indicators User Manual
6.3 Icons
The following icons are used in the PACE series of instruments, not all icons are used in every PACE instrument.
Display Icons in Set-up Menus
Icon Function Icon Function Icon Function
Active Aero set-up Aeronautical
Airspeed range Alarm Altitude range
Area of use Asterisk Auto range
Audio volume Auto zero Backlight
Barometer Basic
Calibration Calibration history
Change supervisor PIN
Control mode Copy
Correction SCM Correction sensor
Correction valve Current set-up Date & time
Delete
Communications Contrast
Diagnostic analogue output
Burst pressure control mode
Correction analogue output
Correction source sensor
Diagnostic barometric option
Diagnostic general Diagnostic RS232
[EN] English 6 - 3 K0470 Revision A
Diagnostic control sensor
Diagnostic controller
Diagnostic source sensor
Page 49
6 Reference and Specification
Icon Function Icon Function Icon Function
Diagnostic vacuum sensor
Display Divider Error
Escape Ethernet
Ethernet connected
Gas head pressure Gauge mode Global set-up
Go-to-ground Hardware build Home
Idle time-out IEEE488 Information
Diagnostic volt­free
Exclamation Fault history
Diagnostics
Ethernet not connected
In limits Instrument
Instrument alias name
Lock Lock tasks Logic output
Max-min Max peak Min peak
Nudge Passive mode Percentage
PIN Power-up Preset
Pressure Pressure filter Process
Language Leak test
Instrument accuracy
K0470 Revision A 6 - 4 [EN] English
Page 50
PACE Pressure Indicators User Manual
Icon Function Icon Function Icon Function
Protective vent Question Range
Recall user set-up Reset use log Resolution
Re-try Roughing RS232
Restore to as shipped settings
Save as shipped
Screen mode Screen saver SCM filter
SCM zero Select range
Set-point limits
Set date Set serial number Set time
Set-up zero Slew rate linear Slew rate max rate
Restore settings 2 Run
Save recall user set-up settings
Set-point higher limit
Save user set-up
Set-point disable/ enable
Set-point lower limit
Software build
Status Status area Step (single)
Stop Supervisor set-up Switch test
Tare Support Task
[EN] English 6 - 5 K0470 Revision A
Software upgrade history
Software upgrade
Page 51
6 Reference and Specification
Icon Function Icon Function Icon Function
Test program Test program copy
Timing Time out Timed zero
Units User defined units Use log
Use log history Vent Vent time out
Vent Yes/No Vent set-up Warning
Zero analogue output
6.4 Measure Set-up
Pressure zero
Test program delete
Zero history Zero
During use, the instrument pressure sensor can show small zero shifts caused by time and temperature changes. Regular “zeroing” increases measuring precision.
Process
Selects display processing features that change the reading, as follows: %: Pressure can be displayed reading as a percentage of full-scale or as a percentage
of a specified span.
Filter: The displayed reading can be filtered by a custom low pass filter or the filter can
be disabled (default disabled). The indicator works at a speed independent of the filter time constant.
Tare: A specific tare value can be selected or the current displayed pressure reading can
be “captured” as the tare value. The display shows the selected tare value in the pressure window.
Peak: Maximum, minimum and average display of pressure readings.
Task
Selecting Task enables a set of pre-determined functions and software enabled optional functions.
K0470 Revision A 6 - 6 [EN] English
Page 52
PACE Pressure Indicators User Manual
Units
Select the new units from the list of pressure measurement units. Special units can also be defined (Ref: Global set-up,
Global set-up
Section 6.6, supervisor set-up).
Ref: Global set-up,
Set-up zero
Zero from top level screen (main range only). If other ranges are fitted these can be zeroed by selecting the displayed reading.
Barometric Reference Option
The barometric reference option measures the barometric pressure at the reference port. Depending on the sensors fitted, it also permits the indicator to operate in either pseudo-
gauge or pseudo-absolute mode by the addition of barometric pressure.
6.5 Status
The display shows the following: a. Instrument status
Section 6.6.
•Model
Serial number
MAC address sensor(s)
Range
Last calibration date* b. Software build - read only data. c. Hardware build - read only data. d. History - read only data
•Calibration
•Zero
Software
Hardware
Message
Ethernet connection e. Communications, IEEE 488 and RS232 are fitted as standard. Additional communication
types are options - USB and Ethernet. f. Current set-up - read only data. g. Support
List contact information for support and advice.
[EN] English 6 - 7 K0470 Revision A
Page 53
6 Reference and Specification
6.6 Global set-up
Supervisor set-up
PIN protected menu (Ref: Supervisor set-up, Section 6.7).
Calibration
PIN protected menu (Ref: Calibration set-up, Section 6.8).
Save/recall user set-up
Save user set-up. Recall user set-up.
Display
a. Resolution b. Backlight c. Audio volume d. Status area e. Display Mode
•Reading (default)
Graph
* The instrument date and time must have been set correctly.
K0470 Revision A 6 - 8 [EN] English
Page 54
PACE Pressure Indicators User Manual
6.7 Supervisor Set-up
The Supervisor menu provides facilities for programming settings. These are usually made
during installation as follows:
Important Note: A PIN protects the Supervisor menu against unauthorised use. Each
instrument on delivery contains the factory set PIN (0268). To continue protecting the supervisor set-up menu the PIN should be changed as soon as possible.
Alarms
An alarm can be set to trigger when the pressure exceeds the high alarm or falls below the low alarm. A buzzer sounds when the alarm triggers and the alarm symbol (bell) appears on the display.
Comms
Selects the communication port parameters and simultaneous operation of the RS232, IEEE 488 interface and Ethernet.
Note: PACE1000 Lab view drivers are available for download from:
http://sine.ni.com/apps/utf8/niid_web_display.download_page?p_id_guid=B6F9A6B06AEA01F1E0440021287E65E6
The user can select appropriate settings for communicating with the control computer (PC) and the required command protocol.
Referring to the PACE SCPI communications user manual there are three commands that can be used to retrieve pressure readings from the PACE100X:
:INST:SENS[x]:READ? Where x = 1 to 8 set in the comms range menu
:SENS:PRES?
:DISP[x]:WIND?
Returns the reading top main display window with display filtering (2Hz update rate) applied
Where [x] = display (1=top main, 2=middle status area & 3= bottom function area) window
To obtain fastest readings you must use the following SCPI command :INST:SENS[x]:Read? This command returns two readings, a filtered reading and a raw reading (unfiltered
reading) from internal sensors, thus giving the user a choice of which reading to use which is most suitable for their application.
e.g.
TX> :Inst:Sens2:Read? RX> :INST:SENS2:READ "993.539148733033 993.543837356372"
Note: Raw reading returned from IDOS sensors is “0” as a raw readings cannot be obtained
from IDOS sensors.
Data throughput. Baud rates slower than 115k2 baud will slow the new data update rate pro-rata. Even at a 115k2 baud rate the readings are slower than using IEEE. IDOS is a slower sensor than internal sensors.
[EN] English 6 - 9 K0470 Revision A
Page 55
6 Reference and Specification
RS232 Located on the rear panel an external RS232 connection requires the following:
Connector 9-way ‘D’ female wired as per Table 2-1 Communications RS232 point-to-point only (daisy chain is not supported) Baud Rate power-up default 9600, no parity & handshake = xon/xoff Baud rates selectable ** 2400, 4800, 9600, 19k2, 38k4, 57k6 & 115k2 Parity None, Odd & Even Flow control None, Hardware & xon/xoff Protocols PACE SCPI Heritage emulation DPI 142/150, DPI 141 Terminator CR or LF or CR/LF
** Selectable through the user interface.
RS232 Comms Range Setup Note: The RS232 Comms Range values are set up by the user because of the hardware the
user desires to operate with the PACE 1000 equipment. This procedure only allows
supervisor to select these user pre-installed values. To initially define / change or
delete the RS232 Comms Range values, refer to (Ref: K0472 PACE SCPI Remote
Communications Manual).
K0470 Revision A 6 - 10 [EN] English
Page 56
PACE Pressure Indicators User Manual
1. On the main screen, touch any of the three horizontal touch areas on the screen.
2. On the MEASURE SETUP screen, select GLOBAL SETUP.
[EN] English 6 - 11 K0470 Revision A
Page 57
6 Reference and Specification
3. Select SUPERVISOR SETUP.
4. Enter the Supervisor PIN and press the top touch area. Use the back arrow in the top
right corner of the screen to delete any incorrect data entries.
K0470 Revision A 6 - 12 [EN] English
Page 58
PACE Pressure Indicators User Manual
5. On the SUPERVISOR screen, select COMMS.
6. On the COMMUNICATIONS screen, select COMMS RANGE SETUP.
7. On the COMMS RANGE SETUP, Use the Up and Down arrows to highlight the desired
Range (Ranges 1 to 10).
[EN] English 6 - 13 K0470 Revision A
Page 59
6 Reference and Specification
Comms range setup is used to re-align the SCPI index (x) to the correct pressure range.
Where Range is index (1)......etc.
8. Press the top touch area on the screen to change the range value.
9. Use the Up and down arrow to highlight the new Range value.
10. Press the top Measure on the screen to change the range value. The new range value
is set and the screen returns to the COMMS RANGE SETUP screen.
11. If necessary, repeat Steps 7 to 10 to set up other range values.
12. Press the Escape icon to go back to the COMMS RANGE SET UP screen.
13. When complete, press the Escape icon as necessary to return to the Measure screen.
IEEE Located on the rear panel an external IEEE 488 connection requires:
Connector 24-way ‘D’ female wired as IEEE 488 standard Communications IEEE488 GPIB Default Address 16 Protocols PACE SCPI Heritage emulation DPI 142/150, DPI 141
Ethernet Located on the rear panel an external Ethernet connection requires the following:
Connector Ethernet RJ45 Protocol SCPI Terminator CR/LF Default Address Auto IP (0.0.0.0) Host name PACExxxxxx (where xxxxxx = serial number) Web Password 0268 Access control Open Reset LAN Settings Selected in Supervisor set-up menu
Ethernet Firewall A firewall protects the Ethernet connection. The firewall is always turned on. The following ports are unfiltered to allow remote communication and control.
Port Description
80/tcp http (Web server) 111/tcp rpcbind (RPC for VXI) 111/udp rpcbind (RPC for VXI) 443/tcp https (Web server)
K0470 Revision A 6 - 14 [EN] English
Page 60
PACE Pressure Indicators User Manual
5025/tcp SCPI socket *****/tcp VXI-11 communication (dynamic port allocated)
Ethernet Functionality
• PACE Ethernet supports both VXI-II and Sockets automatically.
• Socket port address 5025.
• Internet protocol IP4.
Ethernet Parameters Range Setup
1. Touch any of the three horizontal Measure touch pads on the home screen to open the CONTROL SET UP screen.
2. On the MEASURE SETUP screen, select GLOBAL SETUP.
3. Select SUPERVISOR SET UP.
[EN] English 6 - 15 K0470 Revision A
Page 61
6 Reference and Specification
4. Enter the Supervisor PIN and press ENTER SUPERVISOR PIN. Use the back arrow in the top right corner of the screen to delete any incorrect data entries.
Note: The factory set Supervisor PIN is 0268. If the Supervisor PIN has been locally changed,
make sure that the new PIN is kept in a safe place. If the new PIN is lost, it can only be reset at a GE Service Centre.
5. Press COMMUNICATIONS STATUS to open the COMMUNICATIONS STATUS screen.
6. Select ETHERNET to open the ETHERNET PARAMETER screen.
K0470 Revision A 6 - 16 [EN] English
Page 62
PACE Pressure Indicators User Manual
7. To change the ADDRESS parameter, complete the following:
a. On the ETHERNET PARAMETER screen, use the UP and DOWN arrows to
highlight the ADDRESS field.
b. Press the top touch pad on the screen to enter the ADDRESS TYPE screen.
c. Use the UP and Down arrows to highlight the desired address type (either
AUTO IP or STATIC).
[EN] English 6 - 17 K0470 Revision A
Page 63
6 Reference and Specification
d. Press the top touch pad on the screen to set the new address type. The
screen automatically returns to the ETHERNET PARAMETERS screen.
8. To change the host name, complete the following:
a. On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the HOST NAME field.
b. Press the top touch pad on the screen to enter the HOST NAME screen. c. Use the keyboard to input the new host name and then press the top
button on the screen to set the host name. The screen automatically returns to the ETHERNET PARAMETERS screen.
K0470 Revision A 6 - 18 [EN] English
Page 64
PACE Pressure Indicators User Manual
9. To change the web password, complete the following:
a. On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the WEB PASWORD field.
b. Press the top touch pad on the screen to enter the WEB PASSWORD
screen. The keyboard screen opens.
[EN] English 6 - 19 K0470 Revision A
Page 65
6 Reference and Specification
.
c. Use the keyboard to input the new web password and then press the top
touch area on the screen to set the new password. The screen automatically returns to the ETHERNET PARAMETERS screen.
10. To reset the LAN settings, complete the following:
a. On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the RESET LAN SETTINGS field.
b. Press the RESET THE LAN SETTINGS touch pad on the top touch pad of the
screen.
c. The RESET LAN SETTINGS sub-screen asking for confirmation of the reset.
Press YES to confirm reset of the LAN settings.
11. To turn the LAN status indicator on or off, complete the following:
a. On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the SHOW LAN STATUS field.
K0470 Revision A 6 - 20 [EN] English
Page 66
PACE Pressure Indicators User Manual
b. Use the UP and DOWN arrows to highlight the desired setting. The setting
is either ON or OFF.
c. Press the top touch pad on the screen to set the new setting.
12. To change ACCESS MODE, complete the following:
a. On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the ACCESS CONTROL field.
[EN] English 6 - 21 K0470 Revision A
Page 67
6 Reference and Specification
b. Press the ACCESS CONTROL OPEN touch pad on the top of the screen to
open the ACCESS MODE screen.
c. Use the UP and DOWN arrows to highlight the required parameter. The
choices are OPEN or RESTRICTED.
K0470 Revision A 6 - 22 [EN] English
Page 68
PACE Pressure Indicators User Manual
d. Press the ACESS MODE OPEN or ACCESS MODE RESTRICTED touch pad at
the top of the screen to set the required access mode.
13. To change CONTROLLER IP ADDRESS complete the following:
a. On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the ACCESS CONTROL field.
b. Press the ACCESS CONTROL touch pad on the top of the screen. c. Use the UP and DOWN arrows to highlight the CONTROLLER IP ADDRESS
field.
[EN] English 6 - 23 K0470 Revision A
Page 69
6 Reference and Specification
d. Use the number touch pad at the bottom of the screen to input the new IP
address and press the CONTROLLER IP ADDRESS touch pad at the top of the screen to set the new IP address.
14. To view the recent IP addresses, complete the following:
a. On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on
the right of the screen to highlight the ACCESS CONTROL field. b. Press the ACCESS CONTROL button on the top of the screen. c. Use the UP and DOWN arrows to highlight the SELECT FROM RECENT field.
K0470 Revision A 6 - 24 [EN] English
Page 70
PACE Pressure Indicators User Manual
Reference
Level
d. Press the SELECT FROM RECENT button at the top of the screen to view
the IP address history.
Head Correction
Corrects pressure reading for the height difference between instrument reference level and UUT. For accuracy, head correction must be enabled and the parameters set for each sensor:
for UUT positioned higher than the reference level of the PACE1000 enter a positive height correction.
for UUT positioned lower than the reference level of the PACE1000 enter a negative height correction.
[EN] English 6 - 25 K0470 Revision A
Page 71
6 Reference and Specification
Lock Tasks
Individual tasks
:
Allows any combination of individual tasks to be disabled.
Note: Restricts operation of the instrument to specific tasks or functions, recommended for production procedures.
All
:
Disables all tasks.
Change PIN
Changes the Supervisor PIN: enter the existing PIN, then the new PIN and confirmation of the new PIN.
Note: Confirmation of the new PIN permanently
replaces the old PIN. Record this new PIN and keep in a safe place. If new PIN is lost it can only be reset by returning the instrument to a GE service centre.
User defined units
Permits the user to define a set of units. Following the on-screen prompts special units may be set by selecting a Pascal multiplier and assigning a five character name.
Instrument alias name
Permits the user to define a 20 character alias name for the instrument. The instrument returns this name through the communications interfaces.
Language
Operation in any of the languages refer to the SDS. Further languages can be up-loaded.
Restore as shipped settings
Restores instrument settings to factory default.
Note: Does not affect PIN settings.
K0470 Revision A 6 - 26 [EN] English
Page 72
PACE Pressure Indicators User Manual
6.8 Calibration
The calibration menu provides facilities for programming settings for maintenance as follows:
Note: A PIN protects the Calibration menu against unauthorised use. Each instrument, on
delivery, contains the factory set PIN (4321). To continue protecting the supervisor set­up menu, the PIN should be changed as soon as possible.
sensor correction
Selects the range for a three-point calibration routine.
screen calibration
Selects touch screen calibration routine.
Time & Date
Sets instrument clock and date.
Change PIN
Changes the Calibration PIN.
a. Enter the existing PIN b. Enter the new PIN c. Confirm the new PIN
Note: Confirmation of the new PIN permanently
keep in a safe place. If new PIN is lost it can only be reset by returning the instrument to a GE service centre. For more information regarding calibration, refer to PACE Calibration Manual K0450.
6.9 Specification
Refer to the PACE1000 data sheet for details.
Note: The data sheet is contained in the CD shipped with the product.
replaces the old PIN. Record this new PIN and
[EN] English 6 - 27 K0470 Revision A
Page 73
6 Reference and Specification
Selects Analogue Output Range
On/Off
Update rate of Analogue Output option from the control module.
Analogue Output Bandwidth = 0.5 x Update Rate (Hz)
30V maximum with respect to chassis. Rated output = 24V
30Vmax
To maintain PACE product safety, external circuits connected to the instrument must meet Safety Extra-Low Voltage (SELV) requirements.
15-way female D connector
Front view of Analogue Option PCB connector
123
4
5
6
7
8
15 14 13 12 11 10 9
30Vmax
22Vto 26V
ANALOGUE1
ANALOGUE2
15
CONSULTHANDBOOK
analogue connections
6.10 Options
Option enable process
To enable soft options on a PACE instrument, use the following:
1. Touch the top Measure area of the screen.
2. Select Global Setup.
3. Select Calibration.
4. Enter a Calibration PIN 1234.
5. Enter new option key xxxxxxxxxx (10 digits).
6. After entry of this key PACE confirms the options have been enabled. Note: Hardware options automatically enable post installation
K0470 Revision A 6 - 28 [EN] English
Page 74
PACE Pressure Indicators User Manual
Pin number Function Pin number Function
1 not used 9 not used
2 not used 10 0V return
3 not used 11 +24V DC OUT @ 100mA
4 not used 12 SW IN 1
5 not used 13 SW IN 2
6 not used 14 analogue +
7 not used 15 analogue -
8not used
[EN] English 6 - 29 K0470 Revision A
Page 75
6 Reference and Specification
Each selection has three Volts-free Contacts.
Selection of trigger conditions.
Selection of trigger conditions.
30V maximum with respect to chassis. Rated output = 24V
30Vmax
To maintain PACE product safety, external circuits connected to the instrument must meet Safety Extra-Low Voltage (SELV) requirements.
0V
+V
event
trigger
24 Vdc maximum 30 Vdc wrt chassis 1 A resistive maximum
Typical Volts-free schematic
Relay contacts rated at 30 Vdc, 1 A resistive, 200 mA inductive.
22Vto 26V
ANALOGUE1
ANALOGUE2
15
CONSULTHANDBOOK
volts-free connections
Volts-free Contact Option
The Volts-free Contact option provides a selectable relay contact toggle depending on conditions set in the PACE instrument.
Pin number Function Pin number Function
1 Relay 1 normally CLOSED 9 Relay 3 common
2 Relay 1 normally OPEN 10 0V return
3 Relay 1 common 11 +24V DC OUT @ 100mA
4 Relay 2 normally CLOSED 12 SW IN 1
5 Relay 2 normally OPEN 13 SW IN 2
6 Relay 2 common 14 not used
7 Relay 3 normally CLOSED 15 not used
8 Relay 3 normally OPEN
K0470 Revision A 6 - 30 [EN] English
Page 76
PACE Pressure Indicators User Manual
6.11 Installation and Ancillary Equipment Kit
Refer to the PACE1000 data sheet for details.
6.12 Return Goods/Material Procedure
If the unit requires calibration or is unserviceable return it to the nearest GE Service Centre listed at www.gemeasurement.com
Contact the Service Department to obtain a Return Authorisation (Worldwide excluding USA). In the USA obtain a Return Material Authorization [RMA]. Providing the following information on either a RGA or RMA:
Product (i.e. PACE1000)
Serial number
Details of defect/work to be undertaken
Calibration traceability requirements
Operating conditions
Safety Precautions
You must inform GE if the product has been in contact with any hazardous or toxic substance.
The relevant COSHH or in the USA, MSDS, references and precautions to be taken when handling.
Important notice
Service or calibration by unauthorized sources will affect the warranty and may not guarantee further performance.
6.13 Packaging Procedure
1 The instrument should be at zero/ambient pressure.
2. Switch off and isolate the electrical power supply to the instrument.
3. Shut off the pneumatic pressure and vacuum supplies to the instrument.
4. Remove the instrument from the equipment rack to access the rear panel.
5. Disconnect the power supply cable and the pneumatic supply hose assemblies.
6. Stow the power supply cable in the packaging below.
7. Remove any pressure adaptors, diffusers and restrictors.
If available, use the original packing material. When using packing materials other than the original, do the following:
8. Fit protection to all the ports to prevent ingress of moisture and dirt.
Note: Use the original red plastic plugs or low tack masking tape.
9. Wrap unit in polyethylene sheeting.
[EN] English 6 - 31 K0470 Revision A
Page 77
6 Reference and Specification
10. Select a double-wall cardboard container.
Inside dimensions must be at least 15 cm (6”) greater than the equipment
The carton must meet test strength requirements of >
11. Protect all sides with shock-absorbing material to prevent equipment movement within the container.
12. Seal carton with approved sealing tape.
13. Mark carton “FRAGILE” on all sides, top, and bottom of shipping container.
Environment
The following conditions apply for both shipping and storage:
Temperature range-20° to +70°C (-4° to +158°F)
125 kg (275 lbs).
K0470 Revision A 6 - 32 [EN] English
Page 78
www.gemeasurement.com
Loading...