GE ISOVOLT 3003 Operating Manual

GE Inspection Technologies
Technical Data (2)
Exposure Time
The equipment has one programmable timer with a non-volatile memory.
Preselection and setting digital setting from 0.1 to 99.9 minutes
Setpoint value display digital, 3 digits
i.e. after a mains failure exposure can be continued without any time error.
Prewarning Time
Preselection and setting digital setting from 2 to 250 seconds or de-activated
Programmed Mode
Number of storable programs 250
Tube Parameters (Limit Data)
Number of memory locations
for tube parameters 8
with digitally assigned tube limit data, e.g. nominal voltage,
limit filament current, anode power dissipation
Xray
Technology
20 kHz IGBT technology for high voltage circuit and filament current circuit
Emitted interference as per EN 55011A, Class A
Immunity to interference as per IEC 801-2/1991, 801-3/1984, 801-4/1988
Connected Loads and Mechanical Dimensions
Power connection 1/N~ 230 V ±10%, 50/60 Hz
Power consumption 4150 W at 60 kV, 58 mA
Max. mains current 24 A
Mains fuse 32 A (as per VDE)
Ground terminal 6 mm² Cu
Type of protection as per EN 60529, IP00 (rack mount), IP32 (desk-top)
Cooling water connection 3/4 inch
Cooling water consumption minimum 3.5 l/min, minimum 4.5 bar, maximum 7 bar,
at outlet point pressure-free
Cooling water temperature > dew point < 25° C
Dimensions overall 483 mm x 266 mm x 680 mm (WxHxD), 19“ (48 cm) rack mount
Weight approximately 57 kg
Specifications are subject to change without notice.
GEInspectionTechnologies.com
©2004 General Electric Company. All rights reserved.
GEIT-30052AE (7/04)
GE Inspection Technologies
SEIFERT ISOVOLT 3003
Xray
Highlystabilized X-Ray equipment ISOVOLT 3003
for Radiographic and Radioscopic Materials Inspection
Characteristics of the ISOVOLT equip-
ment are a medium-frequency convert-
er, IGBT technology in the power mod-
ule, SMD technique, LCD display, soft-
keys and the possibilities of integration
into computer-controlled X-ray systems;
a memory for operating data is a further
feature.
Features
• High stability of voltage and current through modern power electronics
• Voltage divider for exact high voltage measurements
• Proven medium-frequency converter technology
• High user comfort through microprocessor technology
• SMD technology
• Automatic warm-up program (real-time clock)
• Possibility of integration into existent systems (via RS232C-interface)
• Fail-safe circuit for “X-RAY ON” lamp
• Produced under ISO 9001 certified quality management system
• CE conformity
The use of the medium-frequency con-
verter technique entails high stability,
low weight, small dimensions and low
power consumption.
The employment of a microprocessor
offers increased operation safety, dia-
logue capability, user-comfort and easy
servicing (history memory and modem
support).
Loading...