GE Invenia Operating Manual

Technical Publicatio
n
Invenia
Invenia
TM
ABUS 2.0/InveniaTM ABUS 2.0 Original/
TM
ABUS 2.0 Clinical
4700-0043-00
Rev. 7
Operating Documentation
Copyright © 2018-2022 by General Electric Company Inc. All Rights Reserved.
Product Information
This manual is a reference for the Invenia ABUS 2.0/Invenia ABUS 2.0 Original/ Invenia ABUS 2.0 Clinical . It applies to version 2.0 software or later for the Invenia
ABUS 2.0 ultrasound system.
Change history
Revision History
Date
Revision
Rev. 1 2018-07-19 Initial release
Rev. 2 2019-01-04 Update package information and unpacking procedure
Rev. 3 2019-01-18 Update revision number
Rev. 4 2019-04-07 Update device label information
(YYYY-MM-DD) Reason for change
Rev. 5 2020-09-04 Update Common Documentation Library website
Rev. 6 2021-01-14 Add Content related to Sync Weight Tower
Rev. 7 2022-06-20 Add Invenia ABUS 2.0 Original/Invenia ABUS 2.0
Added the content related to SW R2.2.0
Clinical in the manual
List of Effected Pages (LOEP)
Pages Revision Pages Revision
Front Rev. 7 Chapter 6 Rev. 7
Front matter Rev. 7 Chapter 7 Rev. 7
TOC Rev. 7 Chapter 8 Rev. 7
Chapter 1 Rev. 7 Chapter 9 Rev. 7
Chapter 2 Rev. 7 Chapter 10 Rev. 7
Chapter 3 Rev. 7 Index Rev. 7
Chapter 4 Rev. 7 Rear Cover Rev. 7
Chapter 5 Rev. 7
Please verify that you are using the latest revision of this document. Information pertaining to this document is maintained in the Engineering Central portion in MyWorkshop. If you need to know the latest revision, contact your distributor, local GE Sales Representative or in the USA call the GE Ultrasound Clinical Answer Center at 1 800 682 5327 or 1 262 524 5698.
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
i-1
Important precautions
Translation policy
i-2 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
i-3
i-4 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
i-5
i-6 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
i-7
i-8 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
i-9
i-10 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
i-11
i-12 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
i-13
i-14 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7
Damage in transportation
All packages should be closely examined at time of delivery. If damage is apparent, write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested within this 14 day period.
Certified electrical contractor statement - For USA Only
All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be performed by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall be performed by qualified personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
i-15
Omission and errors
If there are any omissions, errors or suggestions for improving this documentation, please contact the GE Global Ultrasound Documentation Group with specific information listing the system type, manual title, part number, revision number, page number and suggestion details.
Mail the
information to:
GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC 9900 Innovation Drive
Wauwatosa, WI 53226
GE employees should use Post-market Quality Management (PQM) to report service documentation issues.
These issues will then be in the internal problem reporting tool and communicated to the writer.
i-16 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7
Service Safety Considerations
DANGER
WARNING
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety glasses, and kneeling pad, to reduce the risk of injury.
For a complete review of all safety requirements, refer to Chapter 1 in the Service Manual.
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
i-17
Legal notes
Trademarks
Copyrights
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission.
GE may revise this publication from time to time without written notice.
All products and their name brands are trademarks of their respective holders.
All Material Copyright © 2018-2022 by General Electric Company Inc. All Rights Reserved.
i-18 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7
Translation policy - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-2 Damage in transportation - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-15 Certified electrical contractor statement - For USA Only - - - - - - - - - - - - i-15 Omission and errors - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-16 Service Safety Considerations - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-17 Legal notes - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-18 Trademarks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-18 Copyrights - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-18

Table of content

Chapter 1 — Introduction
Manual Overview
Contents in this manual - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-2 Typical users of the System Setup and Basic Service Manual- - - - - - - - 1-4 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
models covered by this manual - - - - - - - - - - - - - - - - - - - - - - - - - - 1-4
Instrument Overview
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical 1-6 View-on PACs - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-9
Important conventions
Conventions used in book - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-10 Standard hazard icons - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-12 Standard Icons that indicate that a special procedure is to be used - - - 1-13 Product icons - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-13 Label locations - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-14
Safety considerations
Introduction - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-15 Dangerous procedure warnings - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-19 Electrical safety - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-19 Mechanical safety - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-21
Electromagnetic compatibility (EMC)
What is EMC? - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-26 Compliance - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-26 Electrostatic discharge (ESD) prevention - - - - - - - - - - - - - - - - - - - - - 1-27
Customer assistance
Contact information - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-28 Phone numbers for Customer Assistance - - - - - - - - - - - - - - - - - - - - - 1-29 System manufacturer - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-30
Returning parts
Returning the transducer and repair parts - - - - - - - - - - - - - - - - - - - - - 1-31
Table of content
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
i-19
Chapter 2 — Site Preparations
General requirements
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
environmental requirements- - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-2
Electrical requirements - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-4
Facility needs
Purchaser responsibilities - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-8 Time and manpower requirements - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-9 Facility needs - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-9
Networking setup requirements
Invenia ABUS 2.0 connected to hospital’s network- - - - - - - - - - - - - - - 2-17 Purpose of the DICOM network function - - - - - - - - - - - - - - - - - - - - - - 2-17 DICOM setup requirements - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-18
Environmental Dangers
Patient Vicinity UL60601-1 (USA) - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-20
Chapter 3 — System Setup
Setup Time and Warnings
Average setup time - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2 Setup warnings- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2
Receiving and Unpacking the Equipment
Warnings for receiving and unpacking - - - - - - - - - - - - - - - - - - - - - - - - 3-4 Receiving the Invenia ABUS 2.0/Invenia ABUS 2.0 Original/
Invenia ABUS 2.0 Clinical - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-5
Unpacking the Invenia ABUS 2.0/Invenia ABUS 2.0 Original/
Invenia ABUS 2.0 Clinical - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-9
Preparing for Setup
Verify customer order - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-34 Physical inspection - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-34 EMI protection - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-34 Warning Label Locations - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-35 Setup Warnings and Cautions - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-40 Electrical specifications - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-41 Connections on the Rear Panel - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-43 Switch ON the AC Power to Invenia ABUS 2.0/Invenia ABUS 2.0 Original/
Invenia ABUS 2.0 Clinical - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-46
Power shut down - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-46
Setting up the Invenia ABUS 2.0/Invenia ABUS 2.0 Original/
Invenia ABUS 2.0 Clinical
Powering on the device - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-47 Logging in as Admin - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-48 Configuration Wizard (for R2.2.0 and above) - - - - - - - - - - - - - - - - - - - 3-49
Customizing the Invenia ABUS 2.0
General Configuration - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-57 Scanning Protocols - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-67 Worklist Configuration - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-68 Worklist Configuration (continued) - - - - - - - - - - - - - - - - - - - - - - - - - - 3-69
i-20 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7
System Configuration (Administrator) - - - - - - - - - - - - - - - - - - - - - - - - 3-70 Setting up the Windows network (for R2.1.1 and below) - - - - - - - - - - - 3-82 Storage Configuration - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-102 Service Configuration Menu - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-103 About Page - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-105
Paperwork after setup
User’s Manual(s) - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-107 Product Locator Installation Card - - - - - - - - - - - - - - - - - - - - - - - - - - 3-108
Chapter 4 — Functional Checks
General procedures
General Precautions - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-2 Overview - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-3 Power ON/Boot Up - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-4 Moving and Transporting the Invenia ABUS 2.0/Invenia ABUS 2.0 Original/
Invenia ABUS 2.0 Clinical - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-7
Where are the User Manuals and the Service Manual? - - - - - - - - - - - 4-12
Functional Checks
Overview - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-13 Performance Functional Checks- - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-13 Mechanical Functions Checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-16 Connectivity Functional Checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-17
Chapter 5 — Components and Functions (Theory)
Introduction
Overview - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-2 General physical design of the ABUS 2.0 system- - - - - - - - - - - - - - - - - 5-5 Physical / mechanical main elements - - - - - - - - - - - - - - - - - - - - - - - - - 5-5
General theory of operation
Introduction - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-7 Image handling- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-8 Image file management - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-8 System elements and their operation - - - - - - - - - - - - - - - - - - - - - - - - - 5-9 PC box - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-11 Weight Tower - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-14 . - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-16 Scan Arm Assembly - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-24 Scan Head - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-26 Display arm assembly with Touchscreen Monitor- - - - - - - - - - - - - - - - 5-30 Controls, displays, and ports - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-31 Electrical / Electronic - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-34 Service-Related - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-34
Chapter 6 — Service Adjustments and Lubrication
Service Adjustments
Scan Head Drive Screw Damper Adjustment- - - - - - - - - - - - - - - - - - - - 6-3 Ball Joint Lubrication - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 6-4 Calibrating the Touch Screen Display - - - - - - - - - - - - - - - - - - - - - - - - - 6-5
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
i-21
Chapter 7 — Diagnostics/Troubleshooting
Service safety considerations Gathering troubleshooting data
Collecting a screen capture with logs - - - - - - - - - - - - - - - - - - - - - - - - - 7-3
Service Desktop
Introduction - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-4 Connecting to Service Desktop - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-5 Color status - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-6 Licenses - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-7 Basic Class Access- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-8 Utilities- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-27
Troubleshooting and System Messages
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
Troubleshooting and System Messages - - - - - - - - - - - - - - - - - - - 7-29
Invenia ABUS 2.0 Messages - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-38
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
Error Log
Important daily error logs - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-41 Common Error Codes for the Invenia ABUS 2.0/Invenia ABUS 2.0 Original/
Invenia ABUS 2.0 Clinical - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-44 Error Code Identifications - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-45 Error Code List - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-47
Chapter 8 — Replacement Procedures
Warnings and important information
Warnings - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-2 Manpower - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-4 Returning/shipping transducers and repair parts - - - - - - - - - - - - - - - - - 8-5
Parts Disposal
Used Media and Used Parts Disposal- - - - - - - - - - - - - - - - - - - - - - - - - 8-6
Cleaning the Filter / Replacement
Filter Replacement Functional Check - - - - - - - - - - - - - - - - - - - - - - - - - 8-8
Replacing the Transducer
Replacing the Transducer (continued) - - - - - - - - - - - - - - - - - - - - - - - 8-11 Transducer Functional Check - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-12 B-Mode Functional Checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-12 3D Volume Functional Checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-12 Performing the Exam - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-13
Replacing the Power Cord System Backup and Software Load
Notice against user modification- - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-15
i-22 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7
Internal Component Replacement
Chapter 9 — Renewal Parts
List of Abbreviations Reordering Supplies
Chapter 10 — Care & Maintenance
Why Do Maintenance?
Keeping Records - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-2 Customer Quality Assurance Program - - - - - - - - - - - - - - - - - - - - - - - 10-2
Care and Maintenance Schedule Cleaning
Cleaning Scan Head Assembly - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-6 Cleaning Air Filter - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-9
Index
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
i-23
i-24 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7
Chapter 1

Introduction

This chapter describes important issues related to safely servicing the Invenia ABUS 2.0/ Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical. The service provider must read and understand all the information presented here before installing or servicing the Invenia ABUS 2.0/Invenia ABUS 2.0 Original/ Invenia ABUS 2.0 Clinical.
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
1-1
Introduction
Contents in this manual
This manual provides installation and service information for the Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical.
The manual is divided into ten chapters.
In the beginning of the manual, before Chapter 1, you will find the Revision overview, the Important precautions including
Translation policy, Damage in transportation, Certified electrical contractor statement, Omission & errors, Service safety considerations and Legal notes, and the Table of Contents.
An Index has been included after Chapter 10.

Manual Overview

Table 1-1: Contents in this manual
Chapter Number Chapter Title Description
1. Introduction Contains a content summary and warnings.
2. Site preparations Contains pre-setup requirements for the Invenia ABUS 2.0/ Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical.
3. System Setup Contains setup procedure with procedure checklist.
4. General Procedures and
Functional Checks
5. Components and
Functions (Theory)
6. Service Adjustments Contains instructions on how to make any available adjustments
7. Diagnostics/
Troubleshooting
8. Replacement procedures Provides disassembly procedures and reassembly procedures
9. Renewal Parts Contains a list of replacement parts for Invenia ABUS 2.0/
Contains functional checks that must be performed as part of the installation, or as required during servicing and periodic maintenance.
Contains functional explanations of the electronics.
to the Invenia ABUS 2.0/Invenia ABUS 2.0 Original/ Invenia ABUS 2.0 Clinical.
Provides procedures for running diagnostic or related routines for the Invenia ABUS 2.0/Invenia ABUS 2.0 Original/ Invenia ABUS 2.0 Clinical.
for all changeable FRU.
Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical.
1-2 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7
Manual Overview
Table 1-1: Contents in this manual (Continued)
Chapter Number
10. Care & Maintenance Provides maintenance procedures for Invenia ABUS 2.0/
Index Index A quick way to the topic you’re looking for.
Chapter Title Description
Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical.
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical – System Setup and Basic Service Manual 4700-0043-00 Rev. 7
1-3
Introduction
Typical users of the System Setup and Basic Service Manual
GE Service Personnel (setup, maintenance, etc.)
• Invenia ABUS 2.0/Invenia ABUS 2.0 Original/ Invenia ABUS 2.0 Clinical User
Licensed Hospitals’ Service Providers
• Technical Trainers
Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0
Clinical models covered by this manual
Table 1-2: Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
Models and Hardware/Software Compatibility
Model
Number Description System SW Application SW
H5018SS Invenia ABUS 2.0, Global Windows Embedded Standard 10
64-bit
H5018SC Invenia ABUS 2.0, China Windows Embedded Standard 10
64-bit
NOTE: When not otherwise specified, the contents of this manual
applies to all Invenia ABUS 2.0/Invenia ABUS 2.0 Original/ Invenia ABUS 2.0 Clinical models.
Field Replaceable Units (FRUs)
Please refer to Chapter 8 in this manual.
How to turn the Invenia ABUS 2.0/Invenia ABUS 2.0 Original/
Invenia ABUS 2.0 Clinical ON and OFF
Please refer to Chapter 4 in this manual.
How to check for software version
To verify the software version on the instrument, please refer to Chapter 4 in this manual.
Version 2.0.3 or later
Version 2.0.3 or later
1-4 Invenia ABUS 2.0/Invenia ABUS 2.0 Original/Invenia ABUS 2.0 Clinical
System Setup and Basic Service Manual 4700-0043-00 Rev. 7