GE ImageQuant LAS 500 Site Preparation Manual

Page 1
ImageQuant
LAS 500
Site Preparation Guide
Page 2
Table of Contents
31 Introduction ..........................................................................................................
103 Site preparation checklist ..................................................................................
2 ImageQuant LAS 500 Site Preparation Guide 29-1358-76 AA
Table of Contents
Page 3
1 Introduction
Purpose of this document
The purpose of this document is to give the information you need to fully prepare your site for the installation and testing of the system.
A Site preparation checklist is included in this document.
Scope of this document
This Site Preparation Guide covers the ImageQuant LAS 500 imager.
Preparation
Make sure to read the entire Site Preparation Guide and complete the Site preparation checklist in Chapter 3.
If GE support is needed, returnthecompletedSite preparation checklist to the GE service office in your region. You will find E-mail address and fax number to the GE service office in your region under www.gelifesciences.com/contact. If your region is not listed, select the nearest region.
ImageQuant LAS 500 Site Preparation Guide 29-1358-76 AA 3
1 Introduction
Page 4
Assign personnel for site preparation tasks
The site preparation tasks listed in the following table can be divided between two or three people from the customer.
Site preparation tasksPersonnel
Site preparation/ installation coordinator
Review the site preparation guide for safety information and system re­quirements.
Choose the site.
Coordinate personnel and tasks.
Coordinate the installation timing of the system with any domestic safety procedures and testings.
Order required materials.
Schedule the installation and inform personnel of the installation date.
Laboratory personnel/ Primary users
Review the safety information.
Make sure that all customer-provided materials for the installation are present at the site.
Facilities personnel
Make sure that the installation requirements are met for:
-
Space at the installation site
-
Environmental conditions
-
Biosafety
-
Ventilation and waste collection
-
Electrical supply
-
Safety and installation materials
4 ImageQuant LAS 500 Site Preparation Guide 29-1358-76 AA
1 Introduction
Page 5
2 Site preparation overview
Introduction
Before the product is delivered, you must prepare the site according to the instructions in this chapter.
Select the site
Select a site that complies with national and international requirements for biosafety.
WARNING
Spread of biological agents. The operator has to take all necessary
actions to avoid spreading hazardous biological agents. The facility must comply with the national code of practice for biosafety.
General site requirements
The following requirements must be fulfilled at the location where the ImageQuant LAS 500 imager will be installed:
There must not be a heat source such as a radiator near the imager
There must be no risk that the imager can get wet or flooded
The imager must not be exposed to corrosive gas
The imager must not be installed in a dusty environment
The imager must not be installed in a place which is constantly or exessively exposed to vibration or impacts.
The imager must not be exposed to direct sunlight
ImageQuant LAS 500 Site Preparation Guide 29-1358-76 AA 5
2 Site preparation overview
Page 6
Space requirements
Prepare a clean working area on a stable laboratory bench. The bench must comply with the specifications in the following table.
SpecificationParameter
60 x 60 cmMinimum bench area for operating
ImageQuant LAS 500
At least 20 cm free space in front of the imager.
Free space required around the imager
10 cm free space on all other sides.
490 N (50 kg) or higherLoad capacity
Horizontal ±2°Inclination of bench surface
WARNING
Access to power cord. Do not block access to the power cord. The
power cord must always be easy to disconnect.
There must be additional space on the bench to place samples on the tray.
Note:
6 ImageQuant LAS 500 Site Preparation Guide 29-1358-76 AA
2 Site preparation overview
Page 7
Dimensions and weight
The following illustration shows the dimensions of the ImageQuant LAS 500 imager.
570 mm
280 mm
300 mm
The ImageQuant LAS 500 imager weighs 15 kg.
Environmental requirements
ImageQuant LAS 500 is intended for indoor use only.
The installation site must compliy with the following specifications:
SpecificationParameter
18°C to 28°CAmbient temperature, operation
Temperature fluctuation < 10°C per hour
15% to 70% RH, non-condensingRelative humidity, operation
-25°C to 70°CAmbienttemperature, transportation/stor-
age
5% to 95% RH, non-condensingAmbient humidity, transportation/storage
ImageQuant LAS 500 Site Preparation Guide 29-1358-76 AA 7
2 Site preparation overview
Page 8
SpecificationParameter
(1) Operating time Oscillation: 0.03G (5 to 60 Hz)
Floor vibration conditions
Impact :1G
(2) Non-operating time Oscillation: 0.4G (5 to 60 Hz)
Impact :2G
Up to 2000 mAltitude, operation
Pollution degree 2Pollution degree of the environment
CAUTION
Do not use the instrument in a room with a temperature above +28°C. Higher temperature does not allow the CCD to cool down properly (to -25°C).
Electrical requirements
SpecificationParameter
100 to 240 V AC ±10%Supply voltage
50/60 HzFrequency
200 WMax power
Overvoltage category IITransient overvoltages
WARNING
Protective ground. The product must always be connected to a
grounded power outlet.
8 ImageQuant LAS 500 Site Preparation Guide 29-1358-76 AA
2 Site preparation overview
Page 9
User-supplied accessories and consumables
The following accessories must be supplied by the user:
Personal Protective Equipment (PPE) as required (gloves, lab coats, etc.).
Other product specifications
SpecificationParameter
70 dB(A) or lowerNoise
Fast Lmax ≤60dB(A), and average Leq ≤ 54dB(A) (1 m from the instrument)
IP21Protection provided by the enclo-
sure
ImageQuant LAS 500 Site Preparation Guide 29-1358-76 AA 9
2 Site preparation overview
Page 10
3 Site preparation checklist
Introduction
Read the previous chapters in this Site Preparation Guide before completing the checklist.
Complete the checklist in this chapter to make sure that all necessary preparations have been made for the installation of ImageQuant LAS 500.
Checklist
Comment/specifyN/ANOYESRequirement
General
If GE support is needed for installation, will the GE service personnel be required to take site EHS (Environment, Health and Safety) or security training before doing the installation?
Assigned personnel
Personnel are assigned as specified.
Site selection
Site complies with requirements for biosafety
General site requirements are fulfilled
Space requirements
Space is available as specified.
Imager
Accessories/additional components
Environmental requirements
The proper environmental conditions are fulfilled as specified.
Ambient temperature
Relative humidity
10 ImageQuant LAS 500 Site Preparation Guide 29-1358-76 AA
3 Site preparation checklist
Page 11
Comment/specifyN/ANOYESRequirement
Atmospheric pressure (altitude)
Pollution degree of the environment
Electrical requirements
Electrical requirements are fulfilled as specified.
Computer requirements
Will the ImageQuant LAS 500 imager be connected to a network?
Is there on-site IT support available if needed?
User supplied equipment
Personal Protective Equipment (PPE) as required, (e.g., gloves, lab coats, etc.)
ImageQuant LAS 500 Site Preparation Guide 29-1358-76 AA 11
3 Site preparation checklist
Page 12
GE and GE monogram are trademarks of General Electric Company.
ImageQuant is a trademark of General Electric Company or one of its sub­sidiaries.
All other third party trademarks are the property of their respective owner.
© 2014 General Electric Company – All rights reserved. First published Dec. 2014
All goods and services are sold subject to the terms and conditions of sale of the company within GE Healthcare which supplies them. A copy of these terms and conditions is available on request. Contact your local GE Healthcare repre­sentative for the most current information.
GE Healthcare Europe GmbH Munzinger Strasse 5, D-79111 Freiburg, Germany
GE Healthcare UK Limited Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA, UK
GE Healthcare Bio-Sciences Corp. 800 Centennial Avenue, P.O. Box 1327, Piscataway, NJ 08855-1327, USA
GE Healthcare Japan Corporation Sanken Bldg. 3-25-1, Hyakunincho Shinjuku-ku, Tokyo 169-0073, Japan
For local office contact information, visit www.gelifesciences.com/contact
GE Healthcare Bio-Sciences AB Björkgatan 30 751 84 Uppsala Sweden
www.gelifesciences.com/quantitative_imaging
29-1358-76 AA 12/2014 a2094
Loading...