UPS III Loop Calibrator

Introduction
The Druck UPS III Series of loop calibrators can supply power (source mode) and produce readings (measure mode) to perform f ield calibrations on 2-wire devices. The set-up menu enables the user to “source” or “measure” in either voltage or current and to perform continuity tests. These user instructions include the operation, safety instructions and installation requirements for the loop calibrator.

Specifications

Accuracy
Values in table below includes temperature effects 17°C to 27 °C
Outside these limits .................. 0.003%/°C(0.0015%/°F)
Calibration Reference................ 22°C ±1°C/RH45%±15%
Mode Range Accuracy Remarks
Source mA 0 to 24 mA* 0.01% rdg + 2 lsd V-max. 75V Source mA +
24V Measure mA 0 to 24 mA* 0.01% rdg + 2 lsd V-max. 75V Measure mA
+ 24V Measure V 0 to 60V* 0.02% rdg + 4 lsd R-measure
Continuity <100Ω ** -1 mA
* Resolution 0.001 lsd least significant digits ** Audio + visual rdg reading
Hart® communications...............menu selectable 220Ωloop resistor
Operating Temperature...... -10°C to 50°C (-14°F to 122°F)
Storage Temperature......... -20°C to 70°C (-4°F to 158°F)
Conforms to ......... EN61010, EN 61326-1(1997)+ A1(1998)
Electrical Power Supply
.................................... 4 x 1.5 V alkaline size AA or
Universal power supply {see accessories}
0 to 24 mA* 0.01% rdg + 2 lsd R-max 1k
0 to 24 mA* 0.01% rdg + 2 lsd R-measure
This loop calibrator meets the essential protection requirements of the relevant EEC directives.
at 20 mA
15
Ω
1M
Ω
Ω
1 K0317 Issue No. 3
Page 4
Physical
Dimensions......................... 77 x 129 x 24 mm(3” x 5” x 1”)
Weight .................................................. 275 grams (9.7 oz.)
Terminals ...................................4 mm sockets {gold plated}
Case ....................................................... High impact ABS
Relative Humidity....................................................0 to 90%

Safety

This symbol, on the loop calibrator, indicates that the user should refer to the user guide or manual.
Batteries
Remove batteries from the loop calibrator immediately when discharged and before storage.
Dispose of batteries in accordance with local regulations and battery manufacturers' instructions.
When storing and transporting batteries make sure they cannot be short circuited.
Power Supply
The power supply for this loop calibrator can be the internal non-rechargeable batteries or the external Universal power supply unit (see accessories).
Battery life >
The display shows with low battery power.
Battery Replacement
Unscrew and remove the securing screw from the battery panel.
Replace the batteries, check the polarity of the batteries.
Refit and secure the battery panel.
Accessories
Assy 305 Test lead set 191-129 Power supply, Universal, 100-240 V a.c. 47-63Hz 38016 Carrying case 38023 Protective rubber boot
75 hours in measure mode
>
18 hours at 12 mA (source mode)
K0317 Issue No. 3 2
Page 5

OPERATION

Keys
The key switches the calibrator on and off. Press and hold for 2 seconds.
The key changes the measure or source operating mode. Pressing the
keys makes menu selections, sets numerical values and controls step and ramp functions (up/down).
The select advanced functions shown on the bottom of the display. When no key is pressed for 10 minutes, the calibrator times out and switches off. To disable this automatic time out, select autpower down in the set-up menu.
3 K0317 Issue No. 3
Page 6

Operating Modes

F1
F1
Serial number
VERSION
CAL. DATE
HART
BATTERY
Information
F2
F2
Resolution
adjust
F2
Contrast
adjust
F2
Range
4-20 mA/0-20mA
F2
Source unit
mA/%
F2
Hart® ON/OFF
F2
Aut power down
ON/OFF
F2
Decimal
.,/
F2
Access code
menu
F2
Calibration
menu
F2
F1
F2
F2
Pressing
the start-up sequence. Pressing , at this time, the display shows the information screen:
Pressing , at this time, the display shows the set-up screen:
switches the instrument on and the display shows
The calibrator can be used in two modes
measu re
or
source

Measure mode

The display shows the measured value; depending on the settings made in set-up and advanced settings:
When measuring current pressing enables linear or flow,
pressing enables mA or % (value of 4 to 20 mA or 0 to 20 mA).
When measuring voltage pressing changes the resolution between 0.00V and 0.000V.
To measure continuity the displays shows an open or closed switch symbol with an audible signal on switch closure.
K0317 Issue No. 3 4
.
Page 7
Connect the loop calibrator to the device to be tested:
and Measure mA Press the mode key and select [Measure mA]. External power supplies Vmax = 60 V for the loop. The calibrator measures the current flow of the loop.
Closed loop current measurement from transmitter test terminal.
* test terminal
Measure mA with 24 V Press mode key and select [Measure mA and 24V]. The calibrator supplies 24 V for the loop, maximum 24 mA
5 K0317 Issue No. 3
Page 8
Measure Volts Press mode key and select [Measure V], measure range 60V, maximum impedance 1 Mohm.
Continuity Test Press mode key and select [Continuity Test].
Pressing switches the audible signal on/off.
K0317 Issue No. 3 6
Page 9

Source Mode

The display shows the source value in mA or % value of 4 to 20 mA or 0 to 20 mA, linear or flow depending on the settings made in set-up and advanced settings.
Source mA
Press mode key and select [Source mA]. The calibrator supplies maximum output of: 24 mA; Vmax = 60; receiver input Rmax = 1kOhm.
Source mA with 24V
Press mode key and select [Source mA and 24V]. The calibrator supplies maximum loop power of: 24 V and 24 mA.
7 K0317 Issue No. 3
Page 10
Advanced Options in a Source mode
MODE
F1
Lin
Flow
Valve
Step
Auto-step
Span Check
Ramp
Time (M:SS)
Time (M:SS)
Adjust +1/ -1 second
Adjust +1/ -1 second
F2
Mode
mA source
mA source & 24V
select
Enter
Advanced
select
F2
select
F2
F2
F2
Press the key and select mA Source or mA Source & 24V.
Use and
(Enter) to select the function.
Press the key (Advanced) and the display shows: Linear simulates linear transmitters. Flow simulates flow transmitters. Valve simulates valve control signals.
Use and
(Enter) to select the Advanced option:
Advanced
Step 25% steps for linear and flow - fixed values for
valve. Auto-step The same as step with a timed step interval. Span Check Step between 4 (or 0) mA and 20 mA. Ramp Automatic ramp between 4 (or 0) mA and 20 mA.
Note: Ramp function not available for valve selection.
Use
to quit.
The display returns to the selected source
mode with the advanced setting available.
K0317 Issue No. 3 8
Page 11
Operation of Advanced Options
Advanced
%
+100.0%
mA
mA(24V )
COM
V
+24.000
mA
Hart® communicator
W
220W
menu selectable loop resistor
Press the key to switch the advanced setting on and off:
on or off
e.g. Press or to:
step the output up or down. step the span check maximum or minimum start the “ramp”.
Press then to start:
continuous auto-st ep. or continuous ramp cycle.
Hart® Application
This application allows mA measure and source modes to be used through the Hart® communicator.
9 K0317 Issue No. 3
Page 12
blank page
K0317 Issue No. 3 10
Page 13

Maintenance

Return the loop calibrator to an authorised repair centre for any repairs, it cannot be repaired on-site.
To keep the loop calibrator accurate a calibration check should be carried out once per year.
Cleaning
Clean the loop calibrator case with a moist, lint-free cloth and weak detergent.
11 K0317 Issue No. 3
Page 14

Battery Replacement

Only use the battery type listed on page one. Unscrew and remove the securing screw from the battery panel. Replace the batteries, check the polarity of the batteries. Refit and secure the battery panel.
K0317 Issue No. 3 12
Page 15

Calibration Instructions

General

The instrument is supplied by the manufacturer, complete with calibration certificate(s). A calibration period of 12 months is recommended. The actual calibration interval depends on instrument usage and the total measurement uncertainty acceptable for the specified application.
The
UPS-III
is a very precise measuring instrument and the test equipment and conditions of test must be suitable for the type of work. The calibration check and calibration adjustment should be carried out in a controlled environment by a calibration technician*.
The manufacturer offers a comprehensive and, if required, UKAS accredited calibration service.
*
A calibrati on technician must have the necessary technical k nowledge, documentation, special test equipment and tools to carry out the calibration work on this equipment.
Calibration Equipment
The following tables give the accuracy requirements for the calibration equipment and the UPS-III. Calibration requires a stable temperature of 21° ±1°C (70° ±2°F).
UPS-III measure mode
Applied mAPermitted UPS-III error
0 0.002 0
4 0.002 0.00014 12 0.002 0.00030 20 0.002 0.00046
Table 1
mA measure
(mA)
Calibrator error
(mA)
13 K0317 Issue No. 3
Page 16
Table 2
Applied V Permitted UPS-III error
0 0.004 0.00040 20 0.004 0.00014 40 0.005 0.00064 50 0.005 0.00070
V measure
Calibrator error
(mV)
(mV)
UPS-III source mode
Applied mAPermitted UPS-III error
0 0.002 0
4 0.002 0.00012 12 0.002 0.00011 20 0.002 0.00015
Table 3
mA source
(mA)
Calibrator error
(mA)

Calibration Check

1.Connect the UPS-III to the electrical calibrator. Switch on the electrical calibrator and allow it to thermally stabilise.
2.Switch on the UPS-III and allow the instrument to thermally stabilise.
3.Set the UPS-III to mA measure, adjust the electrical calibrator to apply the first value in the table 1. Record the reading of the UPS-III.
4.Repeat step 3 for all the values in the table 1.
5.Compare the recorded values and the applied values. If the difference is greater than the permitted error, the instrument requires a calibration adjustment.
6.Repeat this procedure for V measure (table 2) and mA source (table 3).
K0317 Issue No. 3 14
Page 17

Calibration Adjustment

1.Connect the UPS-III to the electrical calibrator. Switch on the electrical calibrator and allow it to thermally stabilise.
2.Switch on the UPS-III and press , within two seconds to select Calibration. Enter the access code [9410 factory setting] and allow the instrument to thermally stabilise.
3.Select the parameter required for calibration. Use the display menu to select the calibration values. After a successful calibration enter the new calibration date.
15 K0317 Issue No. 3