GE Druck PC6-IDOS User Manual

Page 1
Druck PC6-IDOS
Pressure calibrator (Bench/panel mount version)
User manual - K0354
GE
Sensing
A-8010 Graz, Riesstraße 146
Sensorik
Messtechnik
Handelsgesellschaft m.b.H.
Page 2
Publication
K0354, Issue 1
Copyright
© 2006 General Electric Company. All rights reserved.
Trademarks
All product names are trademarks of their respective companies.
Page 3
CONTENTS
Welcome ( Indicator Note ) Safety Warning Quick Reference
Specifications
1.0
1.1
1.2
1.3
1.4
1.4.1
1.4.2
1.4.3
1.4.4
1.4.5
PC6-IDOS Arrangement Functions (menu 1) Functions (menu 2)
1.5
1.6
1.6.1
1.6.3
1.6.5
1.6.6
1.6.7
1.7.1
1.7.2
1.7.3
1.7.4
1.7.5
1.7.6
1.7.7
1.7.8
1.7.9
1.7.10
1.7.11
1.7.12
1.7.13
1.7.14
1.7.15
1.7.16
1.7.17
1.7.18
1.7.19
1.8
1.9
1.10
1.11
1.12
1.13
2.0
2.1
2.2
2.3
2.4
2.5
2.6
2.10
2.12
3.0
3.1
3.2
3.3
Menu 1, Menu 2 & Menu 3 Menu 1
Menu 2
Menu 3
...................... ii
........................................... iii
iv
............................................. v
Introduction ...................................... 2
General ............................................. 2
Keypad ............................................. 2
LCD .................................................. 3
Connection Panels ( .......... 3
4 mm Sockets ............................ . 3
Quick connect coupling ........... .. 3
RS232 ..................................... .... 3
Charger ...................................... . 3
EPM ........................................... . 3
............................... 4
....................................
.................................... 6
Introduction ......................................... 10
Connections
Pressure Input .................................... 10
Voltage/Current Input ...................... 10
RS232 ................................................ 11
Battery charger ................................... 11
Voltage Output .................................. 11
Messages & Parameters
Retained parameters .... .. 12
Battery low state .............................. . 12
Tare Value .......................................... 12
Maximum/Minimum ......................... 12
Void leak test ..................................... 12
Switch Off .......................................... 12
Auto Switch Off .................................. 12
Pressure Overload ............................ 12
Low Battery Detect ............................ 12
Pressure Switch State Change .......... 12
Printer Busy ........................................ 12
Outside User Set Alarm Value ........... 12
Display Hold ...................................... 12
Connection Fault ............................... 13
Short Detected .................................. 13
Voltage / Current Overscale .............. 13
RS232 Output Format ....................... 13
Alarm Flag ......................................... 13
Symbols ............................................. 13
Units ................................................... 15
Max,Min, Switch, Temp.,
Tare .................................................... 15
V / mA ................................................. 15
Leak Test ............................................ 17
File (logged files) ................................ 18
Alarm Settings - High / Low ............... 21
V / mA (Measure mode)............................ 22
V-Out, Display light, Port ................... 26
Resolution ........................................... 26
Event ................................................... 26
Log ..................................................... 27
Zero .................................................... 30
Port
Power_Off ........................................... 31
Span ................................................... 33
History (overload, zero/span, ...)........... 35
Language .................................... 35
Units ................................................... 35
..........................
Unit, Leak ....................................... 38
V / mA ............................................ 39
File ................................................. 40
High & Low .................................... 41
V / mA (measure) ........................... 42
V / mA (source) .............................. 43
V-out, Display light ....... 48
Resolution, Event, RS232 ............. 48
Log ................................................ 49
Date&Time, Language, Zero, Port,
Power_Off ...................................... 50
Span .............................................. 51
History ............................................ 52
Language ........................................ 53
Units ................................................ 54
(Keys, Activating functions, Exit,
Menu selection, Data Entry, Tare) ............................
Maintenance
Functions (menu 3)
1.6.2
1.6.4
2.7
2.8
2.9
................................................ vi
Front/Rear)
........
..........
........
........
........
.................................... 8
......................................... 10
.
EPM ..................................................... 10
Switch Test........................................... 11
.
.
.................... 12
.................. ....
.
Date& Time,% 15
.
V / mA (Source mode)...............................23
..
.
RS232 ................................................. 29
Date & Time ....................................... 30
Language ............................................. 30
....................................................... 31
.
.........
..
.................
........
5
37
1.7
2.11
Description 1
Operating Instructions 9
How to Select Functions (Menu 1) 14
Set-menu (menu 2) 20
Cal-menu (menu 3) 32
Flowcharts 36
Appendices 55
Index 62
1 External Pressure Module (EPM) 55
23Calibration Procedures......................... 58
Panel Mount Dimensions...................... 60
Page i
K0354, Iss. 1
Page 4
WELCOME
This detailed operating manual will help you to become familiar
with the many features of the PC6-IDOS Calibrator. The simple step by step
instructions will quickly guide you through the procedures for using the
calibrator and other items needed for accurately testing and calibrating
numerous types of instruments. Please t
efore you begin to use the PC6-IDOS.
ake time to carefully read
the whole manual b
Two versions of the PC6-IDOS are available:
PC6-IDOS Calibrator:
PC6-IDOS Indicator:
All of the functions shown in this manual apply to this unit. The Calibrator can be supplied with a PV211 pneumatic test system for pressures up to 40 bar (600 psi) or PV212 hydraulic test system for pressures up to 700 bar (10 psi). Refer to the instructions supplied with the applicable pressure system.
This is a simplified version of the Calibrator and does not include the Voltage/Current input/output.
PC6-IDOS
PV210 Low pressure pneumatic test system for pressures up to 3 bar (45 psi),
,000
PC6-IDOS
Page ii
K0354, Iss. 1
Page 5
SAFETY WARNING!
HIGH PRESSURE
RECHARGEABLE Ni-MH BATTERIES
:
Uncontrolled release of high pressure is hazardous to personnel and may cause damage to equipment. Before connection of any pressure component to the PC6­IDOS ensure that the component(s) is/are isolated from the pressure supply and any internal pressure is released slowly.
Rechargeable Ni-MH batteries must be recycled or disposed of properly.
CAUTION: Use charger supplied by GE Sensing only.
May explode if damaged or disposed of in fire. DO NOT short-circuit.
Ni-MH
Page iii
K0354, Iss. 1
Page 6
QUICK REFERENCE
Keys
Press to turn PC6-IDOS on and off
Press to access the menus
Press to accept functions or settings
Press to move cursor up/left or increase values
Press to move cursor down/right or decrease values
Press to print/log/start event depending on previous setting
Select menu position cursor press the Enter key
under function
To exit out of any menu select 'Exit' option, or press key.
Activating Functions
Exit
Menu Selection
Pressure Display
1st Press for Menu 1
2nd Press for Menu 2
3rd Press back to Pressure Display /
Calibration Menu (see page 33)
Press or to position cursor and press
to select required alphanumeric. Press
key to release cursor.
For next data entry, repeat the above procedures then press once all data is entered.
Note: To quickly scroll through alphanumeric lists hold down the or key.
Data Entry
Tare
This is a temporary function and will be cancelled when the unit is switched off.
Hold down the key for 4 seconds while in the pressure display mode. The unit will bleep after two seconds when the Max. and Min. values reset. After another two seconds a second bleep will sound as the displayed pressure now resets to zero.
MENU
MENU
MENU
MENU
MENU
ON/OFF
MENU
MENU
Page iv
K0354, Iss. 1
Page 7
SPECIFICATIONS
* Temperature coefficient ± 0.005% of reading / °C w.r.t. 20°C (68°F)
15% FS 6 Digits External Pressure Modules. See Appendix 1 for EPM details. bar, mbar, MPa, kPa, psi, kg/cm², atm, inH 0, mH 0, mmH 0, inHg, mmHg, UNIT1, UNIT2 (user defined)
emperature effects are included in the accuracy.
1011 records into a maximum of 20 files, 64k memory
Backlit LCD, 16 character x 2 line alpha numeric
Overload Pressure Resolution EPM
Pressure Units
Data Capacity
Display
2
2 2
Overload Warning Temperature Display
Pressure Switch Test
Operating Temperature Calibration Temperature Calibrated Temp. range Storage Temperature Temperature Effects
RS232 Parameters
Zero Reset
Humidity Resolution Select Power Supply
Voltage Output
Flashing display and audible tone at 115% of FS Temperature of pressure sensor (Accuracy: ±1 °C (1.8 °F) ) Status OPEN/CLOSED
0 to 50 °C (32 to 122 °F) 20 ±2 °C (68 ±3.6 °F) 5 to 45 °C (41 to 113 °F)
-20 to 70°C (-4 to 158 °F) T
baud rate 1200, 2400, 4800, 9600, 19200 and
38400. Stop bits 1 or 2, status ON/OFF Manual keyboard operation
5 to 95% Relative Humidity non condensing Increase or decrease by a factor of 10 6V Ni-MH rechargeable battery pack (see warning on page iii)/ mains operation via charger (supplied).
Regulated DC supply 10V ± 0.1% max current 10mA Unregulated DC supply 24V ± 10% max current 50mA
Pressure Range
Bar psi
-0.07 to 0.07 -1.0 1.0
-1 to 1 15
-1 to 2 30
-1 to 3.5 50
-1 to 7 100
-1 to 10 150
-1 to 20 300
-1 to 35 500
0 to 70 1,
0 to 100 1,500
0 to 200 3,
0 to 350 5,
0 to 700
to
-0.2 to 0.2 -3.0 to 3.0
-0.35 to 0.35 -5.0 to 5.0
-0.7 to 0.7 -10 to 10
-15 to
-15 to
-15 to
-15 to
-15 to
-15 to
-15 to
0 to 000
0 to
0 to 000
0 to 000
0 to 10,000
Accuracy (FS)
up to
0.05%
0.025%
0.025%
0.025%
0.025%
0.025%
0.05%
0.05%
0.05%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
Model
-2-C
-10-C
-20-C
PC6-B-350-C SG
PC6-B-700-C SG
PC6-B-0.07-C G
PC6-B-0.2-C G
PC6-B-0.35-C G
PC6-B-0.7-C G
PC6-B-1-C G
PC6-B G or A
PC6-B-3.5-C G
PC6-B-7-C G or A
PC6-B G
PC6-B G or A
PC6-B-35-C G
PC6-B-70-C G
PC6-B-100-C G
PC6-B-200-C G
Resolution
1 V
0.1mV 1mV
1 Aµµ
Measurement
Voltage *
Current *
Auto Range
(Auto Range)
Accuracy
0.025% FS
0.025% FS
0.025% FS
0.025% FS
0.025% FS
Range
0 to 100mV
0 to 5V
0 to 50V 0 to 50mA 0 to 25mA
Input resistance
1 M
1 M
1 M
6
Resolution
0.2mV
1 Aµ
Sourcing Voltage * Current *
Accuracy
0.025% FS
0.025% FS
Range
0 to 10V
0 to 50mA
Max. Load
20mA
330 Ohm
Other Pressure ranges are available to special order. SG = Sealed Gauge,G = Vented Gauge, A = Absolute
° (32 to 122 °F)† 0.075% FS accuracy, 0 to 50 C
Page v
K0354, Iss. 1
Page 8
SPECIFICATIONS / MAINTENANCE
Battery Life Recharge Time Low Battery
Dimensions (l: w: h)
Weight
Microprocessor Recommended Recalibration Period
Electrical Connections
Pressure Media
Pressure Connection
Software (optional)
System requirements:Fully charged: up to 8 hours
1
14-16 hours for full charge Continuous check, audio and visual warning
x 180 x 85 mm (7.9 x 7.1 x 3.3 in)
Approximately kg (4 lb)
H8S/2138
1 Year (Max)
Ground Standard 4 mm socket Voltage/Current Input/Output Standard 4 mm sockets Pressure Switch Input Standard 4 mm sockets Battery 2 pole miniature round connector RS232 8 pole miniature round connector EPM (P2) 5 pole miniature round connector
Water, Mineral Oil, Non-corrosive gases (for other media, contact GE). ‘quick-fit’ connector / 1/4”BSP or 1/4”NPT female adaptors.
The ‘ ’ software will present downloaded data in various graphical forms such as gauges or a scrolling bar & remotely control the PC6-IDOS via the RS232 cable linked to a PC. The downloaded data can be stored in a variety of file types compatible with most analysis database and word processing programs. Calibration procedures can be uploaded to the PC6-IDOS.
200
1.8
Minimum Intel Pentium with Windows 95 or
higher.
1
The battery usage time specified here is heavily dependent upon the functions selected. For instance, using the back-light function will reduce the battery usage time, as will source mode functions.
Clean the case with a moist, lint-free cloth and a weak detergent. Do not use solvents or abrasive materials.
Do not do local repairs. Return the equipment to the manufacturer or an approved service agent.
Do not dispose of this product as household waste. Use an approved organisation that collects and/or recycles waste electrical and electronic equipment.
For more information, contact one of these:
• our customer service department
(Contact us at www.gesensing.com)
• your local government office
SiCal PRO
® ®
Ordering Codes:
Maintenance
Page vi
For Panel Mount version add a 'P' to the order code i.e. PC6-BP-XXXX-XXXX-C-1. See page 61 for panel dimensions.
PC6 - B - XXXX - XXXX - C - 1
Range 1 (R1)
(R1 > R2)
C Ior (Calibrator or Indicator)
1 2 3 4-UK / -Europe / -USA / -AUS
Range 2 (R2)
(optional)
K0354, Iss. 1
Page 9
DESCRIPTION
1Page
K0354, Iss. 1
Page 10
1.0 Introduction
1.1 General
1.2 Keypad
The Pressure Calibrator is a microprocessor controlled, precision instrument powered by an internal 6V rechargeable battery pack. It is portable and capable of accurate pressure measurement, voltage or current measurements and sourcing when calibrating transducers, pressure switches, etc. This section describes the equipment from an external viewpoint, allowing the user to become familiar with the various controls and connections provided.
The PC6-IDOS is housed in a painted metal
The Keypad is a non-tactile membrane keyboard with six buttons. The keyboard layout is shown below.
An audible tone accompanies
every key operation.
PC6-IDOS
case with a carry handle/stand, a liquid crystal display unit at the front and a
six button keyboard. The main body contains the circuitry, connections, internal transducer(s) and battery pack.
Note:
PRINT
LOG
ON/OFF
MENU
ENTER
ZERO
2Page
K0354, Iss. 1
Page 11
1.3 Liquid Crystal Display (LCD)
1.4 Connection Panels (Front/Rear)
The LCD is a 16 character x 2 line alphanumeric LED backlit display which is capable of displaying special characters.
Pressure reading Unit of measurement
Function Function Value
Front connection panel: (See Page 4, Fig 1)
This connection panel provides input / output connections as detailed in Fig.1
on the front panel provide
Pressure input
connections are made to
1.4.1 4 mm sockets.
1.4.3 RS232.
SiCal PRO
1.4.4 Charger.
1.4.5 EPM.
Six standard 4 mm electrical sockets mounted input or output current and voltage and switch status.
Voltage outputs of 10 and 24V are available for use with sensors, transducers, transmitters, etc.
An 8 way circular miniature connector is used for communication with RS232 standard communication protocols/equipment. The user software ‘ ’ (optional) can be used to remotely monitor or download/upload data from the PC6-IDOS via this connector.
(Optional RS232 cable - Part No. 01-0625)
The 5 way circular miniature connector is used to connect to the EPM (External Pressure Module). These optional modules cover pressure ranges from 0.07 to 700 bar (1.0 to 10,000 psi).
Rear connection panel:
(See Page 4, Fig 2)
1.4.2 ‘Quick Fit’ coupling.
'Quick-fit' connectors / 1/4BSP or 1/4NPT female adaptors.
The 2 way circular miniature connector is used for the 9V charger/power supply (provided).
This connector will not be available
on a dual (internal) sensor unit.
Note:
Special Symbols
3Page
COUPLING
P2
9V DC
RS232
K0354, Iss. 1
Page 12
+10V
V
+24V
mA
GND
SWITCH
Front View:
LCD, Keypad & Connections
Fig 1
Front Connection Panel
Display.
Keypad
See Page 2
ARRANGEMENT OF PC6-IDOS
4Page
+24 V Volts output. +10 V Volts output. GND Common Ground Socket
mA Milliamp Input (controls input current
in source mode). V Volt Input (output in source mode). SWITCH Pressure Switch Input.
(See Page 10
for V/mA
connections)
EPM
Rear View:
Pressure / RS232 / Power connections
Fig 2
9V DC / Battery Charger
Input Socket
See Page 11
9V DC
RS232
P1
PRESSURE INP UT
External Pressure Module Socket
See Note on page 3
RS232 Communication Socket See Page 11
Pressure Input connections
See Page 3
P2
Druck PC6-IDOS
K0354, Iss. 1
Page 13
PRESSURE DISPLAY -
Selection of pressure unit.
inches of mercury (at 0°C) of mercury user defined unit second user defined unit Returns to menu 1
Displays the maximum pressure measured since the last reset (60 samples per second).
Displays the pressure as a percentage between user pre-set high and low alarm values. Displays the pressure at the time of tare. Volts / milliamps functions
Displays the following input measurements.
Displays actual input measurements. Displays converted input measurements. Returns to Pressure Display.
Indicates the status of the pressure switch under test (OPEN/CLOSED). Allows the PC6-IDOS to measure pressure changes over time.
Starts the leak test. Sets the leak test time.
Returns to Pressure Display. Allows the user to monitor the temperature around the selected pressure transducer in pressure display mode. Allows the user to monitor the date & time of the on board real time clock in pressure display mode. Allows the user to view or manipulate a file previously stored in the filing system.
Allows the user to see logged file.
Outputs a logged file
Imports a file (refer to ‘ instructions for exporting files to PC6-IDOS)
Deletes a single file or all logged files.
Returns to Pressure Display. Returns to Pressure Display mode.
Allows the user to operate the PC6-IDOS for pressure measurements. (Display Functions)
SiCal PRO’
(SHORTCUT: To abort any menu level, simply press the MENU key).
Menu 1
Unit
inHg mmHg UNIT1 UNIT2 Exit
Max
% Tare V / mA
Measure
Actual Converted Exit
Switch Leak
Start Set-Time
Exit Temp C Date & Time File
View
Export
Import
Delete
Exit Exit
bar millibar megapascals kilopascals pounds force per square inch kilograms force per square centimetre atmospheres i inches of water (at 4°C) metres of water (at 4°C) millimetres of water (at 4°C)
millimetres (at 0°C)
Displays the minimum pressure measured since the last reset (60 samples per second).
Displays the following outputs (status must be ON - see ‘Setup’ in Menu 2)
Displays actual output measurements. Displays converted output measurements. Displays Pressure to Volts / milliamps conversion. Returns to Pressure Display.
Returns to pressure display
bar mbar
MPa kPa
psi kg/cm²
atm nH O
mH O mmH O
Min
Source
Actual Converted P to V / P to mA Exit
Exit
2
2 2
FUNCTIONS (MENU 1)
5Page
K0354, Iss. 1
Page 14
6Page
Continued on
Page 7
}
FUNCTIONS (MENU 2)
SETTING MENU Menu 2
- To set user specific functions. (Set Functions)
High
Low
Measure
Zero FS Text Root
Source
Setup
Status Root Step Ramp
Status Rate Dwell_time Exit
Exit
Exit
Allow the user to set a High or Low Pressure Alarm Limit. (by using the increase/decrease facility above which an alarm sounds.)
Voltage / Current input settings for Converted Measurement.
Set Zero. Set Full scale value. User defined measurement text. Switch measurement root function ‘on’ or ‘off’.
Voltage / Current output settings.
Sets the parameters for voltage/current sourcing.
Switches output on or off. Switch sourcing root function ‘on’ or ‘off’. Set step value (%FS) Settings for Ramp mode
Switches ramp mode on or off. Sets the ramp rate (V or mA/Sec). Sets the time for which the output remains at the same level. Returns to Source sub-menu.
Returns to Source sub-menu.
Returns to Pressure Display.
V / mA
Exit
Range
Zero FS Exit
Convert
Zero FS Text Exit
P to V / mA
Zero FS Exit
Exit
Reverts back to Pressure Display.
Sets the following values:-
Sets Zero value Sets Full Scale value Returns to Source sub-menu.
Sets the following converted values:-
Sets converted Zero values. Sets converted Full Scale values. User defined converted text. Returns to Source sub-menu.
Sets Pressure to V / mA conversion factors.
Set zero values. Set Full Scale values. Returns to Source sub-menu.
K0354, Iss. 1
Page 15
FUNCTIONS (MENU 2) Continued.
Page 7
Switches the Power Supply outputs +10V & +24V ON or OFF
(separate pc software utility needed -contact supplier).
Vout
Switches ON or OFF the LED back light. Selects low, normal or high resolution. Sets up the event status for RS232 output and for logging functions.
Sets key driven output or logging. Sets timed output defined by the user. Reverts back to Pressure Display.
Allows the user to setup, amend and view files for logging.
To select a log file and start logging. To look at a file. Deletes a file. To enable or disable logging. To export logged files. To import files from a PC. (see ‘ ’ software instructions for exporting files to the PC6-IDOS) Reverts back to Pressure Display.
Settings for RS232 communications.
To enable RS232 communication. To select the baud rate. To select the number of stop bits. Reverts back to Pressure Display.
Sets up Date & Time parameters
Sets the date & time. To look at date & time.
Reverts back to Pressure Display. Allows user to change language Will reset pressure reading to Zero. ( This will remove Zero offset permanently / NOT available for absolute units ). Selects current pressure module (P1, P2) Allows the user to set Power_Off time or to switch the function on or off.
Switches the auto-power function off.
Allows the user to set the auto-power off time in minutes.
Reverts back to Pressure Display. Returns to Pressure Display
Display Light Resolution Event
Key Timed Exit
Log
Log View Delete Status Export Import Exit
RS232
Status Baud-rate Stop-bits Exit
Date/Time
Set View
Exit Language Zero
Port Power_Off
Key
Timed
Exit Exit
SiCal PRO
K0354, Iss. 1
Page 16
FUNCTIONS (MENU 3)
8Page
CAL MENU
Menu 3
Span
Continue Exit
History
Overload Zero-Span
Exit Language Units
View
Alter
Exit Exit
Allows user to set span.
Starts span calibration procedure. Returns to Cal Menu.
Allows user to see overload and zero-span history.
Displays period of time over which overload occurred. Displays date last calibrated, last span set, range and overload value.
Returns to Cal Menu.
Allows user to download different languages.
Displays the two current user-defined units. Allows user to modify the conversion factor and/or text for the two user-defined units. Returns to Cal Menu
Returns to Pressure Display.
PC6S/No Displays the unit’s serial number.
Displays the serial number of the currently selected pressure module.PM S/No
K0354, Iss. 1
Page 17
OPERATING
INSTRUCTIONS
9Page
K0354, Iss. 1
Page 18
10Page
Fig.3
Input
0V
Output
Three Wire Transmitter
0 to 5V
+10V
V
V
+24V
+24V
mA
mA
GND
GND
SWITCH
SWITCH
+10V
EPM
+V
-V
Two Wi re
Transmitter
4 to 20 mA
1.5 Introduction
The PC6-IDOS enables the user to carry out checks on pressure transducers, transmitters, pressure switches, pressure gauges and process indicators quickly and accurately. It is intended for use as a portable, on-site calibrator, used in close proximity with the equipment under test and without the need to remove the unit from its parent system. This section describes the connections to be made when checking switches, transmitters, etc. It also lists menu options, selection procedures and purpose of options available.
1.6 Connections
1.6.1 Pressure Input The PC6-IDOS measures pressure
directly
Alternatively, connect the item under test to the PC6-
IDOS with the adaptors and seals provided.
via the 'Quick-Fit' connectors found at the rear of the
case.
1.6.2 EPM (External Pressure Module)
Different pressure ranges can be read by attaching an External Pressure Module to the 5-way socket (only available on single range instruments - ).See Fig.2
Warning: The PC6-IDOS must be switched off prior to
connecting or disconnecting the EPM .
1.6.3 Voltage / Current Input / Output Voltage and Current
readings can be taken by connecting the outputs from the item under test to the 'V' or 'mA' and 'GND' input sockets on the connection panel using the connection leads provided.
Warning: When measuring voltage or current, make sure 'GND' is connected (where applicable, see fig. 3)
P2
K0354, Iss. 1
Page 19
1.6.4 Switch Test Activate ‘Switch’ function in menu 1. The switching pressure of a pressure switch can now be monitored by connecting the contacts of the pressure switch between the INPUT 'GND' and 'SWITCH' sockets on the connection panel and pressurizing the pressure switch.
1.6.5 RS232
If the RS232 facility is to be used then connections to the 8­way socket must be made to the specifications given in the table below. Alternatively an RS232 communication cable can be purchased from GE suitable for most RS232 applications as shown below.
To RS232 Standard 9-way socket on PC6- 'D'type female IDOS connector
Pin 6 and pin 8 should not be used.
RS232 Connections Table
Note:
1.6.5.1 RS232 Output
Warning:
1.6.6 Battery Charger
CAUTION:
1.6.7 Voltage Output
Connect the PC6-IDOS to a printer or computer using a suitable cable (not supplied) as detailed above and set the parameters as detailed in Para. 2.5 & 2.7. The output can now be obtained as ‘single shot' or ‘timed' as follows:
- This output is available in Display mode when RS232 status is ON & Log status is OFF. The output cycle is initiated by pressing the key. Pressing the key cancels the output. Once initiated, the output is repeated at the selected time interval (see ‘Event' function, page 26 ) until the key is pressed.
The internal battery pack may be recharged (when indicated by the battery low message) by plugging the charger (supplied) into the '9V' socket.
When recharging internal batteries, only the charging
unit supplied with the PC6-IDOS should be used.
A voltage of +10V or +24V may be sourced from the PC6-IDOS by connecting to the 4 mm sockets. A typical application for this voltage output would be to power a pressure transmitter. ( see Fig. 3. page 10 ) To switch on voltage output function:­a) Press the key twice to activate 'Set' menu. b) Select 'V-out' and press the key. c) Select 'ON' and press the key.
1.6.5.2 Single Shot - This output is always available in Display mode when RS232 status & Log status are OFF. The output is obtained each time the key is pressed (provided any previous output is complete).
The PC6-IDOS must be switched off prior to
connecting or disconnecting the RS232 cable.
1.6.5.3 Timed
SIGNAL
Data Terminal Ready (DTR)
Transmit Data (TxD)
Receive Data (RxD)
Request To Send (RTS)
Clear To Send (CTS)
Do not use
Ground
Do not use
STATE
Held High
Active High
INPUT/OUTPUT
Output Output
Input
Output
Input
Output
Common
Input
PIN No.
1 2 3 4 5 6 7 8
PRINT
LOG
PRINT
LOG
MENU
MENU
MENU
Cable Part No: 01-0625
11Page
ENTER
ZERO
ENTER
ZERO
K0354, Iss. 1
Page 20
1.7 Messages & Parameters
The PC6-IDOS registers pressure input at switch on after the initial start­up message. The pressure reading is displayed on the top line of the LCD (see Page 3). Pressure values are monitored, alarm values are compared, alarm warnings and other information messages displayed until the unit is switched off.
Any read / write errors to memory detected at switch on will be reported as: Memory Error ‘Error No.’
When switched on, the unit registers the same state as when last switched off. The display mode (Operating/Set), units of measurement, function and parameter settings, alarm limits and RS232 output control, Port (P1 or P2) selection, logged parameters are all retained at switch off, with the exception of the following conditions:
A battery low state occurring prior to the switch off. When switched on, the unit registers the latest retained set of parameters.
Any TARE value set during use is not retained and
therefore should be set, if required, each time the unit is switched on.
Maximum and Minimum values are not retained. Current values reflect pressure monitored from the start of the pressure input.
If the unit is switched off during a LEAK test then the test will be void. When switched on again the unit will start up in LEAK test mode.
When the key is pressed the 'switch off' message is displayed for two seconds accompanied by an audible bleep. Normal operation may be resumed by pressing any key during switch off.
The PC6-IDOS will automatically turn itself of
after a set period (See Page 31 for instructions)
(Bleep rate 4 Hz) ‘OVERLOAD' occurs when
input pressure exceeds overload pressure level.
(Bleep rate 1 Hz). ‘Low Battery' occurs when internal battery pack drops below 5.5V. Message flashes for 4 seconds, after a further 30 seconds, low battery detection is re-enabled.
(Bleep rate 1 Hz)
If the receiving device is not ready for input (e.g. device not connected or handshake LOW), 10 seconds after starting an output cycle (single shot or timed output) the message 'Printer Busy' is flashed on the display, together with an audible warning. If the device does not accept input within the next 5 seconds then the output cycle is cancelled and output is stopped until the key is pressed again.
(Bleep rate 2 Hz)
Occurs when measured pressure is lower than the user set Low alarm value or greater than the user set High alarm value and the
message is flashed. Indication ceases when pressure is equal to
either alarm value or between the values.
(Bleep rate 2 Hz)
Occurs in pressure mode. If the key is pressed once (less than 1 second) the pressure value (upper line) is frozen and the lower line is flashed with the ‘Display Hold’ message at 1 Hz. Normal display update resumes when any top row key is pressed.
Memory Errors:
Retained parameters
Battery low state
Tare value
Maximum/Minimum
Void leak test
Error Numbers:-
0 8 9- Onboard eeprom; - RTC; - 2nd Onboard eeprom. Errors with Pressure Modules will be indicated by the message: "No P1 (or P2) Module" preceeded by "PM EEPROM ERROR" at power on.
1.7.1
1.7.2
1.7.3
1.7.4
1.7.5
1.7.6 Switch off
1.7.7 Auto Switch off
1.7.8 Pressure Overload
1.7.9 Low Battery Detect
1.7.11 Printer Busy
1.7.12 Outside User Set Alarm Values
1.7.13 Display Hold
1.7.10 Pressure Switch State Change (Bleep rate 2 Hz)
Occurs in pressure mode with the pressure switch state display being selected. When the connected pressure switch changes state, the pressure value (upper line) is frozen and the pressure switch state (lower line) display is flashed at 1 Hz unless a log file is open. Normal display update resumes when the key is pressed.
’outside
limits’
ON/OFF
12Page
ENTER
ZERO
ENTER
ZERO
PRINT
LOG
K0354, Iss. 1
Page 21
Note: n pressing the key.
The display contents can be printed out duri g display hold by
1.7.14 Connection Fault:
1.7.15 Short Detected:
1.7.16 Voltage / Current Overscale:
1.7.17 RS232 Output Format
1.7.18 Alarm Flags
In voltage / current source mode, if the PC6-IDOS is unable to control the voltage/current output (e.g. because there is no loop back in the current mode), the message “Connection Fault” will flash on the lower line accompanied by a bleep at 1 Hz. The message will continue until the user presses the or key ( after fault is rectified ) or the user exits from source mode.
If the 24V output is ON and it is shorted to 0V, the message “*Short detected*” will be displayed on the lower line accompanied by a bleep at 1 Hz. This message will be cleared after approximately 5 seconds. In this condition the 24V supply will be switched off for safety.
In Voltage measure mode, if a voltage in excess of 55V is connected to the PC6-IDOS the message “*OVERSCALE*” will be displayed on the lower line, until the voltage falls within the input limits. In Current measure mode, if a current in excess of 55mA enters the mA terminal, the PC6-IDOS will display “*OVERSCALE*” on the upper line and the lower line will scroll the message “Check connections: Press MENU to continue”. Pressing the key will remove the message provided the current measured is within range. This message may also appear in the source mode under connection fault conditions. Source mode will change to measure mode for safety. If the current overscale message occurs when the PC6-IDOS is in timed RS232 output mode, the unit will exit from the timed output mode. If there is an excess voltage at the mA input when the PC6-IDOS is placed into current measurement mode, the message “Connection Fault” will be scrolled on the lower line. Once the fault has been rectified, press the key to return to measuring mode.
The output format is as follows: LF Time & date ( 1100) +LF display top line (16 characters ) + LF display lower line (16 characters ) + LF alarm flag (1 character ) + LF
A single character is output to indicate the various
pressure alarm conditions as given below:
* overload pressure level exceeded
(pressure warnings, 1.7.8 & 1.7.12) > pressure exceeds user set high limit < pressure below user set low limit “ ” a space if no alarms apply. c connection fault s short detected
23:13:30 08
Note: If two (or more) conditions occur which require a message display, the highest priority condition in the above list will be displayed. The sequence of messages given above set the priority except 'Switch Off' which comes last.
If more than one symbol is applicable, they are displayed one after the other.
1.7.19 Display Symbols
Top line symbols
Bottom line symbols
13Page
or
or
Log Status ON
Pressure Port 1 selected
Pressure Port 2 (EPM) selected
Rs232 Output Status ON
Timed Log / output
Keyed Log / output
Source Step Mode
Measure Mode & P-V / P-mA
Root Function
Source Mode Ramp
Nudge Mode (UP)
Nudge Mode (DOWN)
Log Status ON File Open
MENU
MENU
PRINT
LOG
K0354, Iss. 1
Page 22
HOW TO SELECT
FUNCTIONS
(MENU 1)
14Page
K0354, Iss. 1
Page 23
MENU
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
15Page
1.8 Units
1.9 Max., Min., Switch, Temp C, Date & Time, %
1.10 Tare
To change the units of pressure, proceed as follows:
• Press the key once, then press key to choose the units sub-menu
• Select required unit by pressing the key and then press key.
The display will indicate the pressure in the selected unit and the pre-selected 2nd line function.
To display any one of these functions, proceed as follows:
• Press the key once and then scroll to the required
function using the or key and then press key to select it.
The display will show the pressure on the top line and the 2nd line will show the function that has now been selected.
°
For Max. & Min., the displayed values are based on a 60Hz update rate. However, since the upper line pressure value is updated every half a second, the Max./Min. values may not be seen here. Conversely, for switch, each time the switch changes state, the upper line pressure value is refreshed with the latest 60 Hz pressure value.
:
The Tare value can be displayed on the bottom line by selecting ‘Tare’ from the menu using the key and pressing the key.
To activate function: hold down the key for 4 seconds while in the pressure display mode. The unit will bleep after two seconds when the Max. and Min. values reset. At four seconds another bleep will sound as the displayed pressure now resets to zero. This is a temporary function and will be cancelled when the unit is switched off.
Tare
1.11 V/mA (procedures are the same unless otherwise
stated)
Two main options are available for ‘V’ and ‘mA’:-
• Select ‘Measure’ for V and mA input.
• Select ‘Source’ for V and mA output.
To measure input voltage or current select ‘Measure’ and press the key.
Two options are now available; ‘Actual’ and ‘Convert’.
• ‘Actual’ is selected by pressing the key. An example display below shows the Measure Mode symbol
on the second line.
• Select ‘Convert’ using the key and press the key. To setup the conversion parameters, i.e. zero, full scale or text; see 2.1 on page 22.
K0354, Iss. 1
Page 24
To output voltage or current:
• Select ‘Source’ using the key and press the key. Three options become available; ‘Actual’, ‘Convert’ and ‘P
V’ / mA.
• ‘Actual’ is selected by pressing the key. An example display below shows the Source Mode symbol on
the second line.
• Select ‘Converted’ using the key and then pressing the key. To set the converted parameters refer to section 2.2.1 on page 23.
• Select ‘P V’ /mA using the key and then press the
key. To set the parameters for this Pressure to Volts /
milliamps conversion refer to section 2.2.3 on page 23.
An external pressure source is required.
• By pressing and holding the key for more than two seconds whilst in ‘Source Mode’ you will activate a ‘ ’ function allowing the source value to be increased or decreased, the longer you hold down the or key, the larger the nudge increments will be (see Page 25 for more information).
By pressing and holding the key again for more than two seconds, the nudge function will be cancelled.
Note:
Nudge
MENU
MENU
ENTER ZERO
Page 16
Note: NOT Available in Ramp or P-V / P-mA modes
ENTER
ZERO
ENTER ZERO
ENTER
ZERO
K0354, Iss. 1
Page 25
1.12 Leak Test
To carry out leak tests, proceed as follows:
• Press the key once and scroll to the 'Leak' function using the key. Press key to select.
Presented with the leak test sub menu, the options 'Start' and 'Set Time' are given as shown below.
To set the duration of the leak test.
Scroll using the key to select 'Set Time' and press the
key.
With the first digit now flashing, press and then select the number required using the keys and press key. ( e.g. time set to 999 min.)
With the second digit flashing, press and again select the number required and press key.
Repeat procedure for third number and press key.
Now select the units of time measurement ( e.g. seconds, minutes, hours, days, etc. ) Scroll through the units with the
keys and press
Once the leak test time has been set press the key to return to the leak test sub-menu.
Select 'Start' using the keys and press to start the leak test.
The top line shows 0.0000 pressure and the bottom line
shows the current pressure.
Start the leak test by pressing the key.
The top line displays the difference in pressure while the bottom line shows the start pressure and the time counting down.
When the leak test has finished, to start the leak test again
press the key once to reset the time and again to start the test.
If no other leak test is to be carried out, press the key
to return to the leak test sub-menu.
Leak Sub-menu
1.12.1 Set Time
MENU
MENU
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER ZERO
ENTER
ZERO
ENTER ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
Page 17
K0354, Iss. 1
Page 26
1.13 File
This function allows the user to view, export, import and delete previously stored files.
To use the filing system proceed as follows:
Switch the unit on.
Press the key once and scroll to 'File' using the key and press .
To view a file, proceed as follows:
Select ‘View' using key and press .
The latest filename is shown in the bottom left hand side
of the display. The record count is shown in the bottom right hand side. To select this file press . If another file is required then scroll through the filenames using the keys and then press key.
To move through the record now selected, press the
key or the key. This will first show the ‘IPM / EPM' serial number from which data was read.
To view the following records in the selected file use the
or key to advance to the next screen.
Date & Time
Pressure Data
2nd line function
Next reading
Pressing the or key will enable the user to go through the records. Once the last record has been viewed an **END** message is displayed.
The records then 'wrap around' so the user can move to the first or last file record using the keys.
At the time of logging, if the pressure is above or below the set limits then a 'High' or 'Low' message will appear after the Pressure Data screen.
1.13.1
Time & Date
Note:
18Page
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
One Complete Reading
‘N’ = Normal Log ‘C’ = Calibration procedure
Number of records
Transducer accuracy
(e.g. 253 = 0.025%)
K0354, Iss. 1
Page 27
19Page
MENU
MENU
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER ZERO
Note: When exporting to ‘SiCal PRO’ do not press enter after selecting the file.
1.13 File (continued)
1.13.2
1.13.3
1.13.4
To a file, proceed as follows:
Scroll the cursor using the key and select ‘Export' and press the key.
Select a file to export using the keys and press the key
To files, proceed as follows:
Scroll to ‘Import’ using the key. Once you have
finished a calibration procedure in ‘SiCal PRO’, click on ‘Export’, press the key on the PC6-IDOS to download. The message “downloading Cal” will be displayed. Once SiCal PRO gives “All done” message, press the key
Choose ‘Yes’ when prompted to ‘Stop?’
To files, proceed as follows:
Scroll the cursor using the key and select ‘Delete' and press the key.
The following options are available:
If you want to delete then select ‘File’ using the key and then press the key.
Select a file to delete using the keys and press the key.
The unit will then display:
If you are sure that the file is to be deleted then the ‘Yes' option should be selected using the key and then entered using the key.
Once the operation is completed the display will return to the 'File' sub-menu.
If another file is to be deleted then repeat the procedure. If no more files are to be deleted then press the key to
return to the pressure display.
Pressing the key once at any time during the
procedure will take you back to the 'File' sub-menu.
If all the files in the PC6-IDOS are to be deleted then select ‘ALL’ in the Delete sub-menu and press the key.
The following message is displayed:
Select ‘Yes’ or 'No' and press the key.
Export
Import
Delete
a particular file
Note:
K0354, Iss. 1
Page 28
SET- MENU
(MENU 2)
20Page
K0354, Iss. 1
Page 29
2.0 Alarm settings ‘High’/‘Low’
To set High or Low pressure warning limits proceed as follows:
Press the key to switch on the PC6-IDOS.
Select the 'Set' menu by pressing the key twice.
Select 'High' or Low' by pressing the key and then
press the key.
Select the digits to be set by moving the cursor with the
keys and pressing the key.
Set the number to the required value using the
keys and then press the key.
Repeat the previous two steps to change other digits for
the alarm value.
Once finished setting the alarm limit value press the
key to return to the pressure display.
A bleeping warning ( 2 Hz ) and the ‘Outside Limits’ message will signal that the measured pressure is lower than the user set low alarm value or greater than the user set high alarm value. The bleeping ceases when the pressure is read between the two values.
The PC6-IDOS will not allow pressure values to be
entered greater than 5% above the maximum working range.
Note:
21Page
MENU
MENU
ON/OFF
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
K0354, Iss. 1
Page 30
2.1 V/mA (Measure mode)
The PC6-IDOS can measure voltage or current and simply display it or convert it using the ‘Converted, Setup’ options and then display the measured value via a conversion factor.
• Select menu 2 by pressing the key twice.
• Select V or mA by pressing the key and then press
(e.g. mA is selected).
• The following options will appear, as shown below:
• Select ‘Measure’ using the key and press the key
2.1.1 Zero:
2.1.2 FS (Full Scale):
2.1.3 Text:
• Select 'Zero' using the key and then press the key.
• Using the keys and the key, change the
value of the converted zero. Press the key to finish editing.
• From the setup menu select 'FS' using the key and then press the key.
• As with the 'Zero' above, set the conversion factor for a full scale input using the keys and the key. Press the key to finish editing.
The ‘Text’ function is used for changing the displayed units when the ‘converted’ (see 1.11 on page 15) option is selected. If 20 mA is being measured from a device then this could be converted and displayed as pressure in bar for example.
e.g. 4.00 mA will be converted to 0.00 bar
e.g. Full Scale 20mA will be converted to 10 bar
(6 characters max.)
2.1.4 Root:
• Select ‘Root’ using the key and press the key.
• Select ‘On’ or ‘Off’ using the key and press the key.
With ‘On’ selected, any measurements will be displayed having the root function applied until ‘Off’ is next selected.
Due to display constraints, the six character ‘Text’ readout will be shortened to display the first three characters when the ‘Root’ function is switched on. ‘Root’ function is available in both Actual and Converted modes.
Note:
22Page
MENU
MENU
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
K0354, Iss. 1
Page 31
2.2 V/mA (Source mode)
The second option for V and mA is ‘Source’ (output) mode.
Select ‘Source’ using the key and press the key. The following options are available:-
(sets the upper and lower limits in Source
mode)
2.2.1 Range:
Select ‘Range’ using the key and press the key.
The following options are available:- ‘Zero’ or ‘FS’.
‘Zero’ is the first option, press the key to select it.
Change the value as required using the keys and press the key to enter.
To alter the FS value, select ‘FS’ using the key and
press the key.
Change the value as required using the keys and
press the to enter.
Note: The ‘Zero’ and ‘FS’ values have been limited with respect
to each other as follows:
The lower range is limited to between 0 and the user’s upper
range minus 1% of the maximum source range available (e.g. 10 V or 50mA).
The upper range is limited to between the lower range plus
1% of the maximum source range and the maximum source range available.
To set up conversion factors for Zero, Full Scale or a user defined (Text) value.
Select ‘Zero’ by pressing the key.
Change the zero conversion values as required using the keys and the key.
Select ‘FS’ (full scale) using the key and press the
key.
Change the ‘FS’ conversion values as required using the keys and the key.
Select ‘Text’ using the key and press the key.
Select a three letter or number text value using the
keys and the key.
• Once you have completed altering any or all of the convert
functions select ‘Exit’ or press the key to return to the ‘Source’ menu.
To setup a pressure to voltage / current
conversion
• Select ‘P->V/mA’ using the key and press the key.
• For ‘Zero’, press the key to enter.
• Change the V / mA value as required using the
keys and the key.
• Change the pressure value as required using the
keys and the key.
• Select ‘FS’ using the key and press the key.
• Change the V / mA value using the keys and the
key.
• As with the ‘Zero’, change the pressure value as required
using the keys and the key.
2.2.2 Convert:
2.2.3 P->V / mA:
23Page
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
MENU
K0354, Iss. 1
Page 32
24Page
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
2.2.4 Setup: Sets the parameters for V / mA sourcing
• Select ‘Status’ using the key and press the key.
• Select ‘Off’ or ‘On’ and press the key.
When switched on, this provides a voltage or current output with the application of square root function.
• Select ‘Root’ using the key and press the key.
• Switch the function ‘Off’ or ‘On’ using the and key.
To set the voltage or current output increments (%FS):
Select ‘Step’ with the key and press the key.
Change the value as required using the and key.
To set the ‘Status’, Rate’ and ‘Dwell_time’ proceed as follows:-
Select ‘Status’ by pressing the key.
Select ‘Off’ or ‘On’ using the key and press the key
Select ‘Rate’ with the key and press the key.
Change the V/sec or mA/sec value using the keys and press the key.
Select ‘Dwell_time with the key and press the key.
Change the time if necessary using the and keys..
2.2.4.1 Status:
2.2.4.2 Root:
2.2.4.3 Step: (Not available in P to V / mA mode)
2.2.4.4 Ramp:
2.2.4.4.1 Status
2.2.4.4.2 Rate
2.2.4.4.3 Dwell_Time
(Explained on Page 25.)
K0354, Iss. 1
Page 33
25Page
Ramp Mode: (Not available in P to V / mA mode)
Ramp mode consists of two parts, a ramp and level. The ‘Step’ and ‘Range’ (zero and full scale) determine the values for voltage / current at each of the levels (see Fig.4).
Rate is the Ramp rate or increments in V/sec or mA/sec.
Dwell_time is the time for which the output (source) stays on each level, when the Dwell_time has expired the output will ramp to the next level.
With the Ramp status on, the unit will operate in Ramp Mode, with the Ramp status off, the unit will operate in Step Mode ( using the same steps as the Ramp Mode but with no ramping between levels). In Step Mode, pressing the or key will cause the output to increase or decrease to the next level. In Ramp Mode, the key will start the ramping, the key will return the unit to Measure Mode.
Pressing the together for one second returns the Ramp Mode to its starting point. In Step Mode, pressing the
Keys together returns to the first Step level.
To enter ‘Nudge Mode’ (where the output increments in smaller steps) press and hold down the key for two seconds (operates in source ‘Step Mode’ only). The unit will bleep to indicate that the Nudge Mode has been activated. In Nudge Mode the source symbol changes to either or
depending upon the nudge direction. To exit from
Nudge Mode press and hold the key for 2 seconds
Nudge increments are as follows:
mA range = 1uA Lower V range = 0.2mV (up to 5.5V) Upper V range = 2mV (5.5 to 10V)
If the or key is held down for more than 4 seconds the nudge increment is increased by a factor of 5 accompanied by an audible bleep every 2 seconds whilst either or key is pressed until the user’s end limits are reached.
MENU
MENU
Ramp Rate
Dwell_Time
V / mA
Seconds
Fig. 4 Ramp mode.
K0354, Iss. 1
Page 34
26Page
2.3 V-out, Display Light
2.4 Resolution
These functions may be selected as follows:
Switch on PC6-IDOS.
Press the key twice to select the Set menu.
Select the required function using the keys and then press the key. Each of these functions has an On or Off setting.
Select the setting using the key and then press the
key.
To change the resolution proceed as follows:
Press the key twice to select the Set menu.
Select 'Resolution' using the keys and press the key.
Select the required resolution ( Low, Normal or High ) using the keys and then press the key.
Resolution will change by a factor of 10.
Note: The display light can also be set On or Off by holding down the key for 2 seconds. The display light status is retained at power-off.
more than
Note:
2.5 Event
Press the key twice to select the Set menu.
Select 'Event' using the key and then press the
key. The Event menu is shown below:
This results in RS232 (status ON) output / data storage triggered by key press.
Select 'Key' using the key and then press the key.
This results in RS232 (status ON) output / data storage at fixed time intervals set by the user.
Select 'Timed' using the key and then press the key.
Select the digit to be modified using the key and press the key. Use the keys to change the value and press the key.
Repeat for next digit and press the key.
Change the time measurement units in the same way and press the key.
Press the key to return to the pressure display.
2.5.1 Key:
2.5.2 Timed:
MENU
ENTER ZERO
ENTER
ZERO
MENU
MENU
ON/OFF
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
K0354, Iss. 1
Page 35
27Page
2.6 Log
Switch on PC6-IDOS.
elect 'Log' and press .
The display gives the option to log a file, view a file or change the log status as shown below:
Select 'Log' using the key and then press the key.
Presented with the sub-menu, enter a filename and record count if no files are present, or select a file to log into.
(max. 8 characters)
Using the keys, move to the filename position and press the key.
Select the characters/numbers to be used to name the file by scrolling up or down at each position and press the key when completed.
Press the key to move on to the record count.
As with the filename, define the record count and then press the key when completed.
If a file has already been set up with the same name then the ‘Overwrite' option screen will be displayed.
Select ‘View' from the log menu using the key and then press the key.
Select the file to view using the keys and then press the key.
* Please see Appendix 2, page 58 for details of Calibration Procedures.
Press the key twice, scroll to and s
Start logging by pressing the key. (see 2.5 ‘Event’ on page 26).
The symbol will appear to show file is open.
2.6.1 Log:
To enter a filename:
2.6.2 View:
filename record count
MENU
MENU
MENU
ENTER ZERO
ENTER
ZERO
PRINT
LOG
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER ZERO
‘N’ = Normal Log ‘C’ = Calibration procedure log (requires ‘SiCal PRO’ software) *
K0354, Iss. 1
Page 36
28Page
2.6 Log continued
Note:
2.6.3 Delete:
2.6.3.1 File:
2.6.3.2 All:
2.6.3.3 Exit:
With the file now selected, use the or key to show the file contents in the sequence shown below.
Use the or key to scroll the other way. The
direction in which the scrolling moves when using the
key depends on the last key pressed.
Press twice to return to the 'Log' menu.
To delete a logged file, proceed as follows:
Select ‘Delete' from the log menu using the key and then press the key.
Select ‘File', ‘All' or ‘Exit' using the key and then press the key.
Using the keys, select a file and then press the key.
A confirmation to delete file prompt is given. Select ‘Yes' or ‘No' and then press the key.
Select the ‘All' option using the key and then press the key.
A confirmation to delete all prompt is given. Select ‘Yes' or 'No' and then press the key.
All Files stored on the PC6-IDOS will be deleted.
Select the ‘Exit' option using the key and then press the key.
ENTER
ZERO
ENTER
ZERO
ENTER ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
One Record
Last Record
File Name Number of Records
MENU
K0354, Iss. 1
Page 37
29Page
2.6 Log (continued)
2.6.4 Status:
2.6.5 Export and Import:
Use this option to enable or disable logging function.
Select 'Status' from the log menu using the key and then press the key.
Select 'On' or 'Off' and then press the key.
The display returns to the log menu.
Once a file has been set using the instructions given, the logging process is shown as active with a disc shaped icon in the top left hand corner of the display.
If you should press the key whilst logging is in progress then the following screen appears:
Select 'Yes' or 'No' and then press the key.
Pressing the key allows direct access to the log menu. This is only possible if the log status icon is flashing in the top left hand corner of the display.
See 1 13.2 and 1.13.3 on page 19 for and procedures.
Note:
. Export Import
2.7 RS232
RS232 standard communication protocol is available on the PC6-IDOS. Select the RS232 functions as follows:
Connect the PC6-IDOS to a printer or computer as described in paragraph 1.6.5 on page 11.
Switch on PC6-IDOS.
Select the Set menu by pressing the key twice.
Select 'RS232' option using the key, then press the
key. The RS232 menu is displayed as below:
Select 'Status' using the key. Choose 'On' or 'Off' option with the key and then press the key. The display returns to the RS232 menu.
Select 'Baud-rate' using the key and then press the key. Select the required baud rate (1200, 2400, 4800, 9600, 19200 or 38400) using the keys and then press the key.
Select the 'Stop bits' using the key and then press the key. Select either 1 or 2 bits with the keys and then press the key.
2.7.1 Status:
2.7.2 Baud Rate:
2.7.3 Stop Bits:
ENTER ZERO
ENTER
ZERO
ENTER
ZERO
PRINT
LOG
MENU
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
K0354, Iss. 1
Page 38
30Page
2.8 Date & Time 2.9 Language
To select Date & Time proceed as follows:
Switch on PC6-IDOS.
Press the key twice to select Set menu.
Select 'Date & Time' option using the key and then press the key.
Select 'Set' using the key, then press the key.
Select the digit you want to change using the key and
press key. Use the keys to change the value and then press key.
Repeat the above step to change the other digits to set
time and proceed to the next set of digits to change the date.
Press the key to return to the 'Date & Time' menu.
2.8.1 To set the time:
2.8.2 To view date & time:
Select 'View' using the key and then press the
key.
Press the key to return to the Date & Time menu.
Press the key twice.
Scroll to ‘Language’ using either arrow key and press key to select.
Note
2.10 Zero (for absolute transducers, please see page 31)
Select the desired language from the options provided and press the key.
: To download different languages, see 3.2 on page 35.
Note: Unlike the 'Tare' function (page 15), 'Zero' reset is a permanent setting. To reset pressure reading to zero:
Press the key twice to select the Set menu.
Select ‘Zero’ option using the key and then press the key.
Press the key as instructed.
The Pressure Display will now show the pressure value as zero. If however the pressure reading is not within 5% Full Scale the following message will be displayed and the zero reset will not be possible.
MENUMENU
MENU
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER ZERO
ENTER ZERO
ENTER
ZERO
ENTER
ZERO
MENU
MENU
K0354, Iss. 1
Page 39
31Page
MENU
MENU
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
For absolute transducers it is possible to adjust the pressure reading at barometric pressure. This function should only be performed when there is access to a high accuracy barometer, and both the barometer’s and ’s pressure sensing
To adjust the barometric reading of an absolute transducer, proceed as follows:
elect the zero function as described on the previous page. You will now be presented with the current barometric
pressure reading. When the pressure has stabilised, press the key.
Press the key twice.
Scroll to 'Power_Off' function using the key and press the key.
Select 'Key' to turn the auto-power function off.
Select 'Timed' to set the auto-power off time (in minutes) and press the key. This automatically turns the auto­power function on.
Set to desired time using the keys and press the
key.
PC6-IDOS
ports must be at the same height. The pressure reading can
only be altered by up to 5% of Full Scale pressure.
S
The upper line display is the current pressure reading; the lower line display is the desired pressure reading.
Adjust the lower line pressure value using the keys and the key. Once you are satisfied with the pressure value, press the key.
Provided that the desired change is acceptable, the adjustment will be performed, and you will see the message:
This function allows the user to select the current (displayed) pressure module.
Press the key twice to select the Set Menu.
Scroll to 'Port' using the key and press the key
Select a module using the keys and press the key to activate. When selecting a different
pressure module, the serial number and range of the module will be displayed briefly before the unit returns to pressure display.
2.11 Port
2.12 Power_Off
The PC6-IDOS has an auto-power off function. The time after which the unit automatically switches off is adjustable from 1 to 999 minutes. However, the auto-power off will not operate under the following conditions:
If a key has been pressed during the auto-power off time period. If a leak test is running. If the RS232 Status is On. If the Log Status is On. If the unit is online with SiCal PRO.
±
K0354, Iss. 1
Page 40
CAL-MENU
(MENU 3)
Important note:
Calibration Procedure:
.
Should you wish to change zero or span settings at any time, the original calibration made at time of manufacture by GE will no longer apply. We therefore take no responsibility for any false or inaccurate readings that occur after any changes to span have been made. If the temperature of the pressure sensor is outside the range of 10 to 30 °C (50 to 86 °F) it will not be possible to perform the span operation.
If necessary then the span calibration should be made using a deadweight tester with an 0.01% of reading or better, and carried out in a temperature controlled environment at 20 ±2 ° (68 ±3.6 °F).
To maintain the accuracy of the calibrator, do not increase the value of ‘tol-ppm’ above 50 ( see flow charts ).
For full calibration which includes Temperature Compensation between 0-50 °C (32-122 °F) the PC6-IDOS should be sent back to GE
accuracy of
C
Page 32
K0354, Iss. 1
Page 41
To access the Cal Menu, press all four top row buttons at the same time. A beep will accompany this action. Press the key three times. The functions available are as follows:
To gain access to the ‘Span’ and ‘Units’ functions, you will be asked to enter a password. Enter the four-digit serial number located on the label on the rear panel of the unit.
It is only possible to adjust the zero/span from the original calibration by up to ±5% of the unit's full scale range. To set span:
• Select ‘Span’ using the key and press the key
• Enter the password, press the key and choose “Continue”
• A ‘Warming up...’ message will be displayed for several seconds.
The current temperature will be displayed on the lower line.
• The following screen will then be displayed asking you to set a
tolerance.
You can now enter the value of the first pressure point that you wish to apply. After you have entered this value, apply the desired pressure and press the key. You will now see the following screen:
• Select Yes and the unit will now take fifty readings.
• The following screen is now displayed:
• Provided the readings are within tolerance and 5% of the desired pressure value the unit will move to the next point.
The sequence for the first point will now be repeated for the 2nd and 3rd presssure points.
Note: Password
3.0 Span: Note:
• If you want to change the tolerance value displayed then press the key to select the first digit.
• Using the keys, change the first value and then press the key to enter it. The next digit will then be selected. Press the key and change the digit as described above.
• Do the same for the third digit if required and then finish by pressing the key once more.
• If however the value shown does not need to be changed then press the key to continue.
You will now be asked to choose the number of calibration points (either 2 or 3). If two are chosen, by default, a zero point and a full-
scale point will be used for the subsequent zero/span. If three points are chosen then, by default, a negative pressure value, a zero point and a full-scale pressure value are used. The following example assumes that 3 points have been chosen for a gauge transducer.
• The following screen is then displayed:
MENU
MENU
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
MENU
Page 33
The “Tol_ppm” figure is the pressure stability expressed as a ppm of pressure reading
K0354, Iss. 1
Page 42
MENU
MENU
MENU
ENTER
ZERO
ENTER
ZERO
Page 34
Once all the desired pressure points have been accepted the adjustment will be performed. The message “COMPLETED” will then be displayed. Press the key to finish and return to the ‘Cal menu’.
For absolute units, the following default pressure values
will be chosen for a 2 point calibration:
1 current reading 2 full-scale
For a 3 point calibration, the following default pressure values will the chosen:
1 current reading 2 midpoint between 1st point and full-scale 3 full-scale
For the first point, you will be presented with the following screen:
The current pressure value is displayed on the upper line. Once you have set up the desired pressure point and the pressure is stable, press the key and you will see the following screen:
This displays the last pressure reading (from the previous screen) on the top line and on the second line you can set the corrected value.
Set the required value and press the key. Select ‘Yes’ or ‘No’ at the next screen and press the key. The procedure for the remainder of the pressure points on an absolute transducer is identical to that for a gauge transducer.
For both gauge and absolute transducers, the pressure point values are limited as follows:
Gauge transducers, all points:
>= -1 bar (-15 psi) and < 105% full-scale pressure
Absolute transducers, all points:
>= 0 bar (0 psi) and < 105% full-scale pressure
Additionally, pressure points must lie in the following order:
Point 3 > Point 2 > Point 1
Absolute:
Point: Pressure value:
Point: Pressure value:
Pressure point limitations:
Notes: If at any time during the procedure you wish to exit and
return to the ‘Cal menu’ then press the key. The message: 'Sure ? No Yes' appears. Select 'Yes' to exit or 'No' to continue with the zero/span calibration.
K0354, Iss. 1
Page 43
Page 35
3.1 History:
• Press the key to view the sub-menus.
• To view the overload history move the cursor to ‘overload’ and press the key. The display will show the cumulative time since the unit was last calibrated as in the example below:
• Press the key to return to the ‘cal menu’.
• As shown above, the last calibration date is displayed first.
• Pressing the or key takes you to the next screen showing the ‘last span set’ date.
• Pressing the or key again the display shows the range of the unit and the overload setting.
• Press the key to finish.
• Press the key to exit to pressure display mode.
• To view the unit’s serial number move the cursor to ‘PC6S/No’ and press the key.
• To view the serial number of the selected pressure module, move the cursor to “PM_S/No” and press the key. The accuracy of the sensor will be displayed on the lower line.
3.1.1 Overload:
3.1.3 PC6-IDOS Serial Number:
3.1.4 Pressure Module Serial Number:
3.1.2 Span:
:
3.3.1 View:
3.3.2 Alter:
• To view the last zero calibration date move the cursor to ‘zero/span’ and press the key.
(separate PC software utility needed - contact
supplier)
To download different languages:-
• Connect PC6-IDOS to PC via the RS232 cable.
• Scroll to ‘Language’ and press the key to select.
• The PC6-IDOS will display a ‘DOWNLOADING’ message until language download is complete.
User-defined unit conversion factors and the text
associated with them:-
• Select ‘Units’ using the key and press the key.
• Select ‘View’ by pressing the key. The first user-defined unit will now be displayed with the text on the upper line and the conversion factor on the bottom line. Pressing any key will display the next user-defined unit.
(to modify the user-defined units)
• Select ‘Alter’ using the key and press the key.
• The text for the user-defined unit is modified first using the
and key.
• Press the key when finished.
• Now modify the conversion factor as above. The conversion factor is the number of user-defined units per bar.
Repeat process if required for the second user-defined unit.
To prevent confusion, it will not be possible to enter text for a user-defined unit that is identical to a factory-set unit in the PC6­IDOS. Additionally, the maximum value for the conversion factor is 100,000.
3.2 Language
3.3 Units
Note:
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
MENU
MENU
MENU
or
or
K0354, Iss. 1
Page 44
FLOWCHARTS
Page 36
K0354, Iss. 1
Page 45
ON/OFF
MENU
Unit
Max
Min
Tare
V
mA
Switch
Leak
Temp C
Date
& Time
%
File
Exit
ON/OFF
MENU MENU MENU MENU
ENTER
ZERO
Span
Units
History
Language
Exit
Menu 1
display menu
Menu 3 cal
menu
Page 37
Press all four top
keys together
ON/OFF
MENU
MENU
High
Low
V
mA
V-out
Resolution
Exit
Port
Power_Off
Event
RS232
Log
Date
& Time
Language
Zero
Display
light
Menu 2
set menu
K0354, Iss. 1
Page 46
Unit
Menu 1 sub-menu Unit & Leak
Page 38
kPa
psi
kg/cm
2
bar
mbar
mH O
2
inH O
2
mmH O
2
atm
MPa
inHg
mmHg
UNIT2
UNIT1
ENTER
ZERO
Set
time
Start
Leak
Exit
Exit
ENTER
ZERO
010
sec/min/
hr/day
MENU
MENU
ENTER
ZERO
0 01
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ENTER
ZERO
ZERO
Re-start
?
Pressure
Display
Yes
No
000
K0354, Iss. 1
Page 47
Page 39
Menu 1 V/mAsub-menu
V
mA
ENTER
ZERO
Measure
Source
Exit
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
Exit
P to V / mA
Exit
Actual Actual
Actual reading
shown in display
Actual reading
shown in display
Status must be switched
on to access source functions
(Menu 2)
P to V / mA conversion
shown in display
Converted reading
shown in display
Converted reading
shown in display
Pressure
Display
Pressure
Display
Pressure
Display
Converted Converted
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
K0354, Iss. 1
Page 48
Menu 1 sub-menu File
Page 40
Exit
View
Export
Import
Delete
ENTER
ZERO
File
Latest
File
Latest
File
First
File
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTERENTERENTER
ZEROZEROZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
MENU
MENU
MENU
Browse
File
or
Exit ?
No
Yes
Yes
Yes
Yes
Yes
No
No
No
No
Stop
Export ?
stop ?
File
Sure ?
Sure ?
All
Exit
ENTER
ZERO
ENTER ENTER
ZERO ZERO
MENU
Yes
Yes
No
No
Stop
Import ?
stop ?
*
* not necessary for file export to ‘SiCal PRO’.
Completes
Import
K0354, Iss. 1
Page 49
10.5000
High
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
or
Low
ENTER
ZERO
ENTER
ZERO
-10.5000
or
MENU
Once finished editing, press the Key to go back to Pressure Display
*
*
*
Menu 2
High & Low
Edit next digit
Edit next digit
Page 41
For setting negative values
For setting negative values
K0354, Iss. 1
Page 50
Menu 2 sub-menu V/mA (Measure)
Page 42
ENTER
ZERO
ENTER
ZERO
Measure
Source
Exit
Exit
Root
Pressure
Display
FS
Text
Zero
Finished Editing ?
Finished Editing ?
Finished Editing ?
Finished Editing ?
Finished
Editing ?
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
mA
V
Menu Menu
Menu Menu
Yes Yes
Yes Yes
Yes
No No
No No
No
Menu
Measure Z. Conv
00.000 000.0000
Measure FS Conv.
50.000 050.0000 Measure Text:
XXXXXX
Measure FS Conv.
050.000 00.0001
Measure Z. Conv
025.010 00.0000
Off
On
ENTER
ZERO
ENTER
ZERO
MENU
K0354, Iss. 1
Page 51
Menu 2 sub-menu V/mA (Source - Range)
Page 43
ENTER
ZERO
Measure
Source
Exit
Range
Zero
FS
Exit
Convert P to V/mA Setup Exit
Pressure
Display
mA
V
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
S.Range Z.
00.0000
S.Range FS.
10.0000
Change value as required
Change value as required
Change
Value?
Change
Value?
Yes
Yes
No
No
MENU
MENU
K0354, Iss. 1
Page 52
Menu 2 sub-menu V/mA (Source - Convert)
ENTER
ZERO
Measure
Source
Exit
Range
Text
Convert P to V/mA Setup Exit
Pressure
Display
mA
V
ENTER
ZERO
ENTER
ZERO
Zero
FS
Exit
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
S.Con. Z. Conv.
00.000 000.0000
S.Conv. FS Conv.
10.000 010.0000
S.Convert Text:
XXX
Change values as required
Change values as required
Change text as required
Change
Value?
Change
Value?
Enter
text?
Yes
Yes
Yes
No
No
No
MENU
MENU
MENU
Page 44
K0354, Iss. 1
Page 53
Zero
FS
Exit
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
S.P V. Z. Conv.
00.000 000.0000
S.P V FS Conv.
10.000 010.0000
Change values as required
Change values as required
Change
Value?
Change
Value?
Yes
Yes
No
No
MENU
MENU
Menu 2 sub-menu V/mA (Source - P to V / mA)
ENTER
ZERO
Measure
Source
Exit
Range Convert P to V/mA Setup Exit
Pressure
Display
mA
V
ENTER
ZERO
ENTER
ZERO
Page 45
K0354, Iss. 1
Page 54
Menu 2 sub-menu V/mA (Source - Setup)
ENTER
ZERO
Measure
Source
Exit
P to V/mA Setup Exit
Exit
Status
Root
Step
Ramp
Pressure
Display
mA
V
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
Page 46
On
On
Off
Off
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
Step:
25 (%FS)0
Page 47
ENTER
ZERO
ENTER
ZERO
Change value as required
Change
Value?
Yes
No
MENU
Range Convert
K0354, Iss. 1
Page 55
Menu 2 sub-menu V/mA (Source - Setup - Ramp)
Page 47
ENTER
ZERO
Status Dwell_time
Rate
Exit
ENTER
ZERO
Off
On
ENTER
ZERO
ENTER
ZERO
MENU
Finished Editing ?
Finished Editing ?
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
Menu Menu
Yes Yes
No No
Ramp: Rate:
1.00 V/sec0
Dwell_time:
0 Sec1
Ramp
K0354, Iss. 1
Page 56
Menu 2 sub-menu V-out, Resolution, Event & RS232.Display light,
Normal High
Low
ENTER
ZERO
Resolution
Exit
Timed
Key
ENTER
ZERO
Event
ENTER
ZERO
111
sec/min/
hr/day
MENU
ENTER
ZERO
ENTER
ZERO
100
ENTER
ZERO
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
Exit
Status
ENTER
ZERO
RS232
Stop bits
Baud-rate
ENTER
ZERO
On
Off
1
2
ENTER
ZERO
or
2400, 4800
9600, 19200
ENTER
ZERO
1 01
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
38400
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
Pressure
Display
Pressure
Display
Pressure
Display
Pressure
Display
Pressure
Display
ENTER
ZERO
ENTER
ZERO
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
MENU
1200
ENTER
ZERO
ENTER
ZERO
Off
On
ENTER
ZERO
V-out
Display
light
Pressure
Display
Page 48
K0354, Iss. 1
Page 57
Menu 2 sub-menu Log
Exit
Log
File Exists ?
Select File
Last File
Edited
Filename ?
Yes
Yes
Yes
Yes Yes
On
Yes
No
No
No
No No
Off
No
Edit
Record count
Finished ?
ENTER
ZERO
Log
Delete
View
Status
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
MENU
MENU
MENU
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
Overwrite ?
Sure ? Sure ?
Status
All
Exit
File
ENTER
ZERO
Pressure
Display
ENTER
ZERO
ENTER
ZERO
Browse
File
or
Exit File ?
No
Yes
Yes
Export
Import
Pressure
display
Select
File
ENTER
ZERO
ENTER
ZERO
ENTER
ENTER ENTER
ZERO
ZERO ZERO
ENTER
ZERO
MENU
Yes
Yes
No
No
Stop
Export ?
stop ?
MENU
Yes
Yes
No
No
Stop
Import ?
stop ?
Page 49
Exit Log
View?
No
Deletes
selected
File.
Delete: All: Please Wait...
Completes
Export.
Completes
Import.
Completes
Export.
ENTER
ZERO
* not necessary for file export to ‘SiCal PRO’.
*
K0354, Iss. 1
Page 58
Page 50
Menu 2
sub-menu Date&Time , Zero , Port, Power_OffLanguage,
Pressure
Display
Set: nglish German Exit
E
Pressure
Display
P2P1
Port
ZERO
ENTER
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
Power_Off Time:
20 Min1
Timed
Pressure
Display
Power_Off
ENTER
ZERO
Exit
ENTER
ZERO
ENTER
ZERO
Key
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
Exit
View
Set
ENTER
ENTER
ZERO
ZERO
Date&Time
Language
ENTER
ZERO
ENTER
ENTER
ZERO
ZERO
145625 151003
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
Menu
Menu
Menu
No
No
Yes
Yes
ENTER
ZERO
Edit next
digit
Finished
editing
Time ?
Change
Language?
Select
language
ZERO
ENTER
ZERO
Yes
ZERO RESET
Press Enter
Zero
ENTER
Menu
No
Pressure
Display
Pressure
Display
‘ xit’ orE
e.g.
PC6-IDOS resets
K0354, Iss. 1
Page 59
Menu 3 cal Span-menu
Warming up . . .
Cal Span
History
Language
Exit
Units
ZERO
ENTER
Enter Password:
ENTER
ZERO
Stop?
Yes
Pressure Display
ENTER
ZERO
ENTER
ZERO
No
Yes
Yes
Yes
Within
Tolerance ?
** COMPLETED **
Press ENTER
50
counts
done
ENTER
ZERO
NoNo
No
ENTER
ZERO
Set 20.00 bar Press ENTER to cont..
20.00 bar
0.0002
0
Set:
Sure ?: No Set:
Yes
0.0002
Maximum pressure of transducer
Within
50 ppm ?
Within
50 ppm ?
50
counts
done
Yes
No No
Calibrating
Settling
*
For 3 calibration points, you will also be asked to enter a negative pressure point at the start.
For absolute units, only 2 points are used; barometric and full-scale pressure.
Example
MENU
ENTER
ZERO
Change tol_ppm.
500
0 05
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
050
050
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
MENU
Release Pressure
Press ENTER to cont..
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
Continue
Exit
Enter Cal Points
32
A
A
A
A
B
B
B
B
MENU
ENTER
ZERO
YesYes
Within
Tolerance ?
No
*
Page 51
K0354, Iss. 1
Page 60
Page 52
Menu 3 cal History
-menu
Overload Hi story
0.0 secs
Last Calibr ated
00/00/00
Last Span Set
00/00/00
Range 0000 Overload 00 00
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
MENU
MENU
MENU
MENU MENU
Overload
PC6S/No PM_S/No
Span
Exit
ENTER
ZERO
Cal
ENTER
ZERO
MENU
or or
or
or
Span
History
Language
Exit
Units
IPM178918 3
Accuracy 0. 025%
Ver2.XXx 22/11/01 PC6-XXXX-X XXX
K0354, Iss. 1
Page 61
Page 53
Menu 3 cal Language-menu
ENTER
MENU
ZERO
No
Yes
Cal
CAL: Langua ge Downloadi ng...
Stop? es NoY
Special software utility is needed for exporting languages to PC6-IDOS (Contact Supplier)
Completes import
Completes import
Stop
Import ?
ENTER
ZERO
ENTER
ZERO
Yes
No
Span
History
Language
Exit
Units
K0354, Iss. 1
Page 62
54Page
Menu 3 cal Units
-menu
Span
History
Language
Exit
Units
Cal
ENTER
ENTER
ZERO
ZERO
Enter Password:
ENTER
ZERO
ENTER
ZERO
Exit
Alter
View
ENTER
ZERO
ENTER
ZERO
NB There are 2 user-defined units
K0354, Iss. 1
Page 63
APPENDIX 1:
External Pressure Module
(EPM)
55Page
K0354, Iss. 1
Page 64
The EPM has been developed to work in conjunction with the
The illustration on Page 57 shows a typical application.
The EPM is an intelligent module that contains a pressure transducer, analogue to digital converter, and the necessary calibration data stored during EPM calibration.
During operation, the EPM sends pressure and temperature related information digitally to the PC6-IDOS, which acts as an intelligent terminal.
The EPM or the PC6-IDOS internal pressure sensor can be selected during calibration using the calibrator menu system.
PC6-IDOS Pressure Calibrator to provide additional pressure ranges
without incurring the full cost of extra calibrators.
For the safe connection and operation of the EPM, refer to the EPM user manual (document K380).
For a full specification and a list of the available pressure ranges, refer to the
datasheet.
EXTERNAL PRESSURE MODULE (EPM)
56Page
K0354, Iss. 1
Page 65
57Page
ITEM UNDER TEST
EPM
H
I
G
H
P
R
I
M
E
EXTERNAL PRESSURE SOURCE
K0354, Iss. 1
Page 66
APPENDIX 2:
Calibration Procedures
58Page
K0354, Iss. 1
Page 67
59Page
A.2.0 Calibration Procedures
Calibration Procedures are created with SiCal PRO. Please read this section in conjunction with the SiCal PRO literature.
Once the file is opened you will see the set text followed by the calibration text scrolling on the lower line. To stop the
scrolling (and return to normal pressure display
mode) press the key.
Use the key to log each point. If the current calibration point is outside the accuracy statement that you have set up in the calibration procedure, you will receive a “FAIL” message along with the FS % accuracy figure. If you wish to log this point, press the key again. Otherwise press the key and you can adjust the point and log it again.
Select 'View' from the “File” menu (section 1.13.1, page 18) Choose either “data” or “Procedure” with the keys
and press the key.
Once you have logged a calibration procedure and you are happy with the data, the file can be uploaded back to SiCal PRO. This file, now complete, can be used for certificates, etc. Please see section 1.13.2, page 19 and the SiCal PRO literature for details.
A.2.1 File Import.
A.2.1 File Log.
Calibration procedures are downloaded to PC6-IDOS by choosing Import from the File menu (section 1.13.3, page
19).
To use a Calibration Procedure, open the file by choosing “Log” from the “Log” menu (section 2.6.1, page 27). Select the desired calibration file by using the keys and press the key. Now press the key twice.
lower line
If you choose “data” you will see the following: procedure serial number, description, pressure range,
transmitter range (if applicable), accuracy, set text, and the temperature at which the procedure was carried out.
Following this header data is the data that you have logged into the file. Note that if you have only just downloaded the calibration procedure from SiCal PRO
then the data area will be blank.
If you choose “Procedure” you will see the same header information as described above followed by the calibration instructions and pressure points for each point.
NB if you try to modify the number of records or filename then the file will revert to a normal file (”N” rather than “C” will appear at the end of the filename display).
A.2.2 File View.
A.2.2 File Export.
MENU
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
PRINT
LOG
PRINT
LOG
K0354, Iss. 1
Page 68
APPENDIX 3:
Panel Mount Dimensions
60Page
K0354, Iss. 1
Page 69
61Page
PANEL MOUNT DIMENSIONS
3U x 42HP
39 HP
See page vi for ordering information.
+10V
V
+24V
mA
GND
SWITCH
Druck PC6-IDOS
K0354, Iss. 1
Page 70
INDEX
62Page
A
B
C
D
E
F
H
I
K
L
M
N
O
P
Q
R
S
T
U
V
W
Z
absolute ...................... v, 34
accuracy....................... v
activate a function........ iv
alarm. ........................ 21
arrangement................ 4
auto switch off.............. 12
back-light.................... 26
battery......................... iii, vi
baud-rate.................... v, 29
cal menu..................... 8, 32
calibration procedure .. 32, 58
charger....................... 11
connection panel........ 3, 4
connections................ 3, 10
connection fault ......... 13
convert ...................... 23
coupling...................... 3, 4
current input................ 10
current overscale ....... 13
data entry.................... iv
date & time.................. 30, 50
display (LCD).............. 3
display light................. 26, 48
display hold................. 12
downloading ............... 19
display symbols ......... 13
dwell-time ................... 24, 47
enter............................. iv
EPM............................. v, 3, 10, 55
event............................ 26, 48
export .......................... 19, 49
file ............................... 18, 40
filename ...................... 27
flowcharts.................... 36
functions...................... 5
FS (full scale) ............. v, 22
high.............................. 21, 41
history ......................... 35, 52
import .......................... 19, 49
indicator....................... ii
input............................. 3
keypad......................... 2
keys ............................ iv
language .. ................. 30, 35, 50, 53
LCD............................ 3
leak. (test) ................. 17, 38
log............................... 27, 49
low............................... 21, 41
maximum (Max)........... 15
measure ...................... 15, 22
memory errors ........... 12
menu selection............ iv
messages & parameters..12
microprocessor............ vi
minimum (Min) ........... 15
mA ............................... 15, 22
nudge ......................... 25
overload....................... v, 12
overscale.................... 13
P to V / mA ................. 23
password .................... 33
PM_S/No ................... 8, 35
Port ........................ 7, 31, 50
Power_Off .................. 7, 31, 50
pressure connections .. vi, 10
pressure media ........... vi
pressure range............ v
print.............................. 12
quick reference............ iv
quick connect coupling.. vi, 3, 10 (Quick-Fit)
range .......................... 23
ramp ........................... 24, 25, 47
rate ............................. 24, 47
resolution..................... v, 26, 48
root ............................. 22, 24
RS232 ......................... 11, 29, 48
safety warning............. iii
serial number ............. 8, 33, 52
set menu..................... 20
short detected ........... 13
software....................... vi
span ............................ 32, 51
specifications............... v, vi
status .......................... 24, 29
step ............................. 24
stop bits....................... 29
storage......................... 19
source ......................... 15, 23
switch test ................... 11
tare............................... iv, 15
temperature.................. 15
text .............................. 22
timed ........................... 11, 26
unit .............................. 15, 35
units ............................ 35, 54
v, mA ........................... 15, 22, 39, 42
voltage input................ 10
voltage output.............. 11
voltage overscale
v-out ........................... 26, 48
warning........................ iii
zero........................ 22
maintenance................ vi
panel mount ............... vi, 60
..........13
....... , 30, 50
K0354, Iss. 1
Page 71
Customer service
Druck has joined other GE high-technology sensing businesses under a new name GE Sensing.
Visit our Website:
www.gesensing.com
Page 72
Loading...