GE Druck MCX-II User Manual

Page 1
GE
Sensing
Druck MCX-II
Multi-Calibrator
User Manual K0320
Multi-Calibrator Multikalibrator Multi-calibrateur
99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com
Page 2
2
Deutsch
Inhaltsverzeichnis ................................ 4
Français
Table des matières .............................. 6
To measure pressures (optional)
Read pressure module MCX-PM operation manual
Automatic calibration or testing (optional)
Read Linkpak-W operation manual
To communicate with ‘smart’ instrumentation (optional)
Read "communicator" operation manual
Table of Contents
Introduction
The MCX-II calibrator ............................. 8
Pressure Measurements and Calibrations
.............................................................. 8
Automatic Calibration.............................. 8
Digital Communicator ........................... 10
Standard Accessories........................... 10
Optional Accessories ............................ 10
Functionality
Parts identification ................................ 12
Screens to work with ............................ 18
Keys to work with .................................. 20
Power Sources
Power from internal batteries ............... 22
Changing from Alkaline to rechargeable
batteries ................................................. 26
Recharging............................................. 26
Operating from 115 V or 230V line
voltage (50 or 60 Hz) ............................ 28
Set-up
Date and Time Settings ........................ 30
Setting the clock to your local time.. 30 Setting the clock to your local date . 32
Temperature readings .......................... 34
Changing the temperature unit......... 34
Changing the temperature scale ...... 36
Language Setting............................... 38
Setting te language ........................... 38
Access code settings............................ 40
LCD screen viewing adjustments ........ 42
Setting the backlight automatic
shut-off mode ..................................... 42
Backlight on/off operation ................. 44
Screen contrast adjustment.............. 44
Bench or portable use ....................... 46
Keystroking ............................................ 48
Storing a key-pad operation sequence
............................................................ 48
Recalling a key-pad operation
sequence ............................................ 48
Clear key-stroking ............................. 50
To measure signals
General................................................... 52
K0320 Issue No. 1
Page 3
Millivolts ................................................. 52
Volts........................................................ 54
Milliamps/XMTR .................................. 54
Ohms ...................................................... 56
Frequency .............................................. 56
Pulse Counter Mode ............................. 58
Switch position change......................... 60
Circuit continuity tester......................... 62
To measure a thermocouple
Using compensation wires ................... 64
Using the standard test leads
(copper wires)........................................ 66
To measure a RTD
With the remote probe (optional)......... 68
Special Measurement Functions
Scaled readings..................................... 70
Setting up scaled readings (linear
relationship) ....................................... 70
Setting up scaled readings
(flow relationship) .............................. 72
Output of electrical signals
General................................................... 74
Set mode ............................................ 74
Output mode....................................... 76
Changing the output level ................. 76
Millivolts ................................................. 78
Volts........................................................ 80
Milliamps/XMTR .................................... 80
Ohms ...................................................... 82
Frequency .............................................. 82
Counter................................................... 84
To simulate a thermocouple
To start/stop ramp cycling ................ 96
Scaled readings..................................... 96
User Power supplies
Transmitter calibration
Setting up a calibration ....................... 100
To check or calibrate a transmitter ...102
Temperature transmitter simulation
Setting up a simulation ....................... 104
To simulate a transmitter.................... 106
Programmable signal converter
Setting up a signal converter ............. 108
Ce Remarks
Ce Remarks ......................................... 110
Service, Repair and Parts
Recalibration of the MCX-II ................ 112
Cold juntion compensation
calibration ......................................... 114
Lithium cell replacement..................... 118
Fault finding procedures ..................... 120
Opening and closing the calibrator
housing ............................................. 120
Fault finding procedures ..................... 122
Spare parts list................................. 128
Specifications
Protocol for RS232 System Integration
.......................................................... 142
Warranty & Liability
3
RTD Simulation
Excitation current .................................. 88
Special Output Functions
Step Mode .............................................. 92
Ramp Mode............................................ 94
Setting up the ramp mode ................ 94
To start/stop a ‘one shot’ ramp ........ 96
K0320 Issue No. 1
Page 4
Introduction
The MCX-II calibrator
The Multi-Calibrator model MCX-II has
8
been designed for testing and calibration of process instrumentation and portable test equipment. The unit provides data to comply with the ISO9000 requirements for calibration. The MCX-II can be used to measure and output analog and digital signals often used in an industrial environment. Also it can be used to simulate a wide variety of temperature sensors. Measurement and output/simulation functions can be operated and read simultaneously.
Pressure Measurements and Calibrations
A plug-in Pressure Module model MCX-PM is optionally available for pressure measurements and calibration.
K0320 Issue No. 1
Automatic Calibration
To automate calibration routines and to store calibration data the unit has been provided with a memory card slot. For the same purpose the unit has a RS232 cable connector to interface directly with a Personal Computer. To perform automatic calibration, in the field or in the workshop, the MCX­II is supported by Intecal software. The software package and Memory Cards are optionally available from your local GE Sales Office or Distributor.
Page 5
Functionallity
Parts identification
12
1 POWER switch
Switch to put the calibrator "ON" and "OFF".
2 CHG. (amber LED) charging
indicator
On when batteries are charged. The indicator should not light when non-rechargeable batteries are installed.
3 EXT. (green LED) external
power indicator
On when calibrator is powered from the optional line adaptor/charger #13603.
4 EXT. POWER INPUT
Connector for external power source.
Only suitable for optional line adaptor/charger #13603.
5 Battery compartment cover
Release screws to get access to batteries.
K0320 Issue No. 1
6 Battery selector switch
To select Alkaline or Rechargeable batteries. Remove protective sticker to get access to the switch.
IMPORTANT:
Switch position must comply with installed battery type.
Page 6
16
Functionallity
15 Pt100 PROBE
4-pole mini DIN female connector for optional sensor.
16 Control and display panel
Rotatable unit to position best for bench or portable use.
17 Auxilary Connectors
For use with pressure module MCX-PM. [Do not make any other connections to these terminals]
18 Shoulder strap (removable)
For portable use with adjustable length.
19 VIEWING ANGLE adjustment
knob
To adjust LCD contrast.
20 BACKLIGHT Push button
Switches LCD backlight ON/OFF.
21 COMMUNICATOR TERMINALS
Connections for "smart" Instrumentation. [Use is optional with communicator card]
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
K0320 Issue No. 1
FAX 781.665.0780 - TestEquipmentDepot.com
Page 7
18
Functionallity Screens to work with
The MCX-II has four types of screen display to work with:
Menu Selection Screen
Offers selection of choices. Move cursor with arrow keys to your choice.
Set up Screen
Move cursor with arrow keys to fill in the blanks.
Working Screen
Displays readings of measured and generated values. Output or simulated temperatures are set with the numerical key-pad or can be ramped up and down in different ways.
K0320 Issue No. 1
NOTE:
After switching the calibrator
on, the screen will show:
Serial Serial Number
Firmware Firmware version
Hart Hart Firmware
version
Battery voltage/switch position
Days left to next calibration
Page 8
20
Functionallity
Keys to work with
The EXE key
Pressing this key will execute choices or output level adjustments.
The CE key
Press to cancel your previous key operation or to correct a typing error.
The NUMERICAL key-pad
Used to dial output levels or simulated temperatures. Also used to fill out levels and time settings for special functions.
The ARROW keys
Moves the cursor position and ramps output levels up or down.
The # key
Used to open the set-up menu.
K0320 Issue No. 1
The DOC key
Used to access the documenting functions.
The 0 key
Changes the sign when display reads all zeros.
Page 9
22
Power Sources
Power from internal batteries
Internal power is obtained from 6 x 1.5 V Alkaline batteries or 6 x 1.2 V Rechargeable batteries, Model R14, Baby or C size.
Installation of the batteries
Switch the calibrator off before you install new batteries. Release both mounting screws of the battery compartment cover. Open the compartment and place the new batteries in position as indicated in the compartment. Check for clean poles and contact plates before you replace the batteries.
Use of Alkaline batteries (supplied as standard)
The calibrator is supplied with one set of 6x Alkaline batteries. Check that the Battery Selector Switch is in the "alkaline" position before replacing the batteries. Battery switch position will be confirmed on the screen during start-up of the calibrator. Replace batteries when the screen shows the blinking battery symbol.
K0320 Issue No. 1
Page 10
24
Power Sources
IMPORTANT:
Be sure that batteries are
inserted so that they are placed correctly with respect to the (+) and (-) indications in the battery compartment. If the batteries are not inserted correctly, they may leak and damage the calibrator.
Do not mix old and new
batteries, or batteries of different types (such as carbon and alkaline.)
If the calibrator is not to be
used for a long period of time, take out the batteries.
Remove and dispose of
worn-out batteries.
Never attempt to recharge or
short-circuit the batteries.
K0320 Issue No. 1
Page 11
Power Sources
Changing from Alkaline to Rechargeable batteries
26
Remove the Alkaline batteries. Remove the switch protector sticker and slide the switch to the "Ni-Cd" position.
Place a new protector sticker to save
the new switch position. Install 6 rechargeable batteries (purchased
locally). Battery switch position will be
confirmed on the screen during start­up of the calibrator. Recharge batteries when the screen shows the blinking battery symbol.
CAUTION:
Do not fit alkaline or other non­rechargeable batteries after you have changed the switch position to "Ni-Cd".
Recharging
Use the Line Adaptor/Charger #13603 only; other equipment may cause damage to the calibrator. Check the indicated line volt switch in the CHARGE position. Connect the appropriate Adaptor/Charger plugs to the line and to the calibrator. The amber led CHG at the calibrator power panel will now be on. As adaptor and charger functions are fully independent from each other, the calibrator can be used while charging. Recharging time from complete discharge to fully charge is 14 hours.
K0320 Issue No. 1
Page 12
28
Power Sources
You may charge for periods longer
than 14 hours. Note that at lower
ambient temperatures the capacity of
Ni-Cd batteries is significantly lower. If
the batteries do not reach their normal
capacity after a 14 hours charging
period, cycle complete discharging
and charging for at least 2 times. If
batteries remain weak they should be
replaced. No particular brand of Ni-Cd
is recommended although cells rated
at 2.0 Ah have preference over
general available 1.8 Ah cells.
Operating from 115 V or 230V line voltage (50 or 60 Hz)
Use the Line Adaptor/Charger #13603
only; other equipment may cause
damage the calibrator. Check the
indicated line voltage. Connect the
appropriate Line Adaptor/Charger
plugs to the line and to the calibrator.
The green led EXT at the calibrator will
now be on.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
K0320 Issue No. 1
Page 13
30
Set-up
Date and Time Settings
The MCX-II has been equipped with an internal clock: to indicate how many days are left before last calibration expires and to record date and time on the memory card. The clock has been factory set to Paris time at the date of calibration.
Setting the clock to your local time
Leave the cursor at NONE at the INPUT and OUTPUT menu and press the # key to open the set-up menu. Select TIME. Press EXE again. Enter the local time. Press EXE to accept new time or press CE to leave the set­up menu.
K0320 Issue No. 1
Page 14
32
Set-up
Setting the clock to your local date
Leave the cursor at NONE at the INPUT and OUTPUT menu and press the # key to open the set-up menu. Select DATE. Press EXE again. Enter date. Press EXE to accept new date or press CE to leave the set-up menu.
K0320 Issue No. 1
Page 15
34
Set-up Temperature readings
Temperature simulations and measurements can be shown in either Celsius or Fahrenheit units. Factory setting is Celsius. Readings are based on either IPTS 68 or ITS 90 temperature scales. Factory setting is ITS 90.
Changing the temperature unit
Leave the cursor at NONE at the INPUT and OUTPUT menu and press the # key to open the setup menu. Select TEMP.SETTINGS and then the temperature unit. Press EXE to change.
K0320 Issue No. 1
Page 16
36
Set-up
Changing the temperature scale
Leave the cursor at NONE at the INPUT and OUTPUT menu and press the # key to open the setup menu. Select TEMP.SETTING and then the temperature scale and press EXE to change.
Select TEMP.SETTING and then the temperature scale and press EXE to change.
K0320 Issue No. 1
Page 17
38
Set-up Language Setting
The MCX-II has multilingual software. The factory setting is ENGLISH.
Setting the Language
Leave the cursor at NONE at the measure and source menu and press # key to open the set-up menu. Select SYSTEM and then LANGUAGE and press EXE. Select required language and press EXE. Press CE to leave the set-up menu.
K0320 Issue No. 1
Page 18
40
Set-up
Access code settings
The ACCESS CODE is programmable by the customer. The following menus are protected by this code;
ACCESS CODECalibration menu
(Electrical and pressure)
9410 is standard factory setting.
Leave the cursor at NONE at the measure and source menu and press # to open the set-up menu. Select SYSTEM and press EXE. Select ACCESS CODE and press EXE to confirm. You are now in mode where you can change the code. If applicable the new code can be typed in and confirmed by EXE. 0000 as code is considered as no existing ACCESS CODE.
NOTE: Registration of the
ACCESS CODE is essential.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
K0320 Issue No. 1
Page 19
42
Set-up
LCD screen viewing adjustments
Setting the backlight automatic shut off mode
Leave the cursor at NONE at the INPUT and OUTPUT menu and press the # key to open the setup menu. Select system and press EXE. Next select the BACKLIGHT MODE and press EXE. The backlight timer can be set to 30 seconds, 5 minutes or off. select appropriate choice and press EXE to acknowledge. Press CE to leave the setup menu.
NOTE:
It is recommended to switch the BACKLIGHT MODE on to save batteries if the unit is mainly used in the field.
K0320 Issue No. 1
Page 20
44
Set-up
Backlight on/off operation
Press the LIGHT key to switch the backlight ON or OFF. Each key-pad operation will postpone shut-off for another time period.
Screen contrast adjustment
Turn the knob for best screen contrast. Contrast may alter with different viewing angles.
K0320 Issue No. 1
Page 21
46
Set-up
Bench or portable use
The MCX-II calibrator can be used in the two ways as shown. Control and Display Panel can be rotated into 5 different positions for best screen viewing angle. Input and output terminals are accessible from front and back to plug-in the test leads in the most practical way.
K0320 Issue No. 1
Page 22
48
Set-up
Keystroking
Key-stroking is a solution to eliminate time consuming key-pad operations for functions you frequently use. It can store 10 different sequences of key­pad operations. Recalling a sequence will put you right back into the function as stored. You can recall a sequence from any operational situation.
Storing a key-pad operation sequence
Make all the necessary arrangements for the function you want to "key­stroke". Press # to open the set-up menu and select key-stroking. Next select the numerical key under which you want to store the sequence. To store a new function under a "used" key, repeat above procedure to overwrite the old function.
K0320 Issue No. 1
Recalling a key-pad operation sequence
Leave the cursor at NONE in the input and output menu and press the appropriate numerical key to activate the "key-stroke".
Page 23
50
Set-up
Clear keystroking
Leave the cursor at NONE at the measure and source menu and press the # key to open the set-up menu.
Select CLEAR KEYSTROKING MEMORY and press EXE. EXE will clean the KEYSTROKE memory. Press CE to leave the set-up menu.
K0320 Issue No. 1
Page 24
52
To measure signals
General
Select an input range and select DIRECT from the menu. It puts you right in the output mode. For other modes like scaled readings read the chapter " Special Output Functions". The upper part of the screen shows the window for input functions. Here, you can read the measured value.
Millivolts
Select mV and DIRECT in the input menu and press EXE to open the working screen. The upper window in the working screen indicates the measured millivolts reading. The range is 0-600 mV split into two (auto) ranges of 0 to 100 mV and
100.000 to 100.01 to 600.00 mV.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
K0320 Issue No. 1
Page 25
54
To measure signals Volts
Select VOLTS and DIRECT in the measure menu and press EXE to open the working screen. The upper window in the working screen indicates the measured Volts reading. The range is 0-60 Volts split into two (auto)ranges of 0 to 6.0000 Volts and
6.001 to 60.000 Volts.
Milliamps/XMTR
Select mA and DIRECT in the measure menu and press EXE to open the working screen. The upper window in the working screen indicates the measured milliamps reading. The percentage of 4-20 mA is indicated in small size digits to facilitate easy alarm switch adjustments. The range is 0-52.000 mA.
K0320 Issue No. 1
Page 26
56
To measure signals Ohms
Select OHMS and DIRECT in the measure menu and press EXE to open the working screen. The upper window in the working screen in the indicates the measured ohms reading. The range is 0-2000 Ohms split into two (auto)ranges of 0.01 to 400.00 Ohm and 400.0 to 2000.0 Ohm. The exitation current is 0.9 mA.
The wiring can be 2, 3, or 4 wire. The wire method is a true measurement of voltage drop across unknown resistance at a constant excitation current.
Frequency
Select FREQ in the measure menu and press EXE. Select FREQ and DIRECT in the FREQ.menu and enter a trigger level between 10 mV and 5V. Press EXE to open the working screen.
K0320 Issue No. 1
The upper window of the working screen indicates the measured frequency reading. The range is 0-20.000 Hz split into three (auto)ranges of 0-655 Hz, 655-1310 Hz and 1310-20.000 Hz.
Page 27
58
Counter Zähler Compteur
To measure signals Pulse Counter Mode
There are three different modes available:
Counts pulses up to
999.999.999
Counts the pulses received in
one minute
Counts the pulses received in
one hour
Select COUNTER in the input menu and press EXE. Enter a trigger level between 10 mV and 5.0 Volts. Select COUNTER, P/MIN or P/HOUR and press EXE to open the working screen.
Press # to set the trigger level.
K0320 Issue No. 1
NOTE:
The pulse counter starts
on the first (positive) rising edge.
Page 28
60
To measure signals Switch position change
Connect the switch as indicated. Select SWITCH in the measure menu. Generate or simulate the appropriate
source signal from the MCX-II to activate the switch. The measure (upper) reading tracks the source (lower) reading until the switch position changes. This change will immediately "freeze" the measure reading indicating the actual switch setting. Press the # button to reset. The measure reading will track the source reading again.
Make sure that NO or NC contacts are potential free.
K0320 Issue No. 1
Page 29
62
To measure signals Circuit continuity tester
To check wiring continuity use the
RTD input terminals identified with SW.
If NONE has been selected in the
output menu and SWITCH has been
selected in the input menu, press EXE
to enter the continuity check mode.
Closing the test loop will activate the
built-in buzzer top prove that the wiring loop has a resistance of 2200 Ohms or less.
To calculate the switch resistance press # to open the set-up menu and select SWITCH RESISTANCE.
NOTE:
Remote temperature
probe should not be connected in this mode.
K0320 Issue No. 1
Page 30
64
Thermocouples Thermoelementen Thermocouples
To measure a thermocouple
Go through the thermocouple input
menu and make the selections as
required. Press EXE to open the
working screen; the measurement
reading is shown in the upper window.
Resolution is 0.1 degrees for most
thermocouple types.
Using compensation wires
It is strongly recommended to use
compensation wires for thermocouple
simulations and measurements. Insert
the compensation wire lead ends into
the wire holes as indicated.
Hand-tight each terminal screw to
secure the wires and make the proper
connections at the instrument side.
Make sure that you use the right type
of wires and that polarities are not
accidentally reversed. Select
INTERNAL to compensate the cold
junction microvolts automatically.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
K0320 Issue No. 1
Page 31
To measure a thermocouple
Using the standard test leads (copper wires)
66
Air probe
Hand-tight the terminal screws before plugging the test leads in. Compensation in this case is required to offset the cold junction compensation microvolts generated by the instrument under calibration. For this we need to measure the temperature at the instrument terminals. Instrument terminal temperatures can be measured by either a hand-held thermometer or a remote temperature probe.
Hand-held thermometer
(not supplied by GE)
Read the thermometer while holding it close to the instrument terminals. Select MANUAL and open the screen to fill out the obtained reading. Press EXE to enter. Note that in this mode temperature changes at the instrument terminals can easily create calibrations errors.
Remote temperature probe
(optional)
Plug-in the Pt100 probe as shown and select EXTERNAL in the menu. Position the sensor tip as close as practically possible to the instrument terminals.
K0320 Issue No. 1
Page 32
68
Air probe
To measure a RTD
Go through the RTD input menu and
make the selections as required.
Press EXE to open the working
screen; the measurement reading is
shown in the upper window.
Resolution is 0.1 or 0.01 degree
depending on selected RTD type. The
excitation current is 0.9 mA. Wiring
can be 2, 3 or 4 wire.
The 4 wire method is a true
measurement of voltage drop across
the unknown resistance at a constant
excitation current.
With the remote probe (optional)
With the optional available
temperature probe or with any other
suitable type of Pt100 sensor, you can
measure temperatures easily. Select
PROBE in the input menu and press
EXE to open the working screen; the
measurement reading is shown in the
upper window when the probe has
been plugged-in as indicated. See the
connection diagram if you use other
Pt100 sensors.
Pt100 PROBE CONNECTOR 4 POLES, MINI-DIN
K0320 Issue No. 1
Page 33
70
Special Measurement Functions
Scaled Readings
Except in temperature measure ranges all readings can be presented in a number of 5 digits and a sign. This number could represent an engineering unit like gallons/hour, revolutions/minute or any other relationship. This facility is available on both measurement and source functions. The original signal reading is still displayed in smaller sized digits.
Setting up scaled readings (linear relationship)
Select SCALE to open the set up screen. Fill out the high- and low-limits for the span (preset at 4 and 20 mA in the milliamp ranges). Fill out both scale ends for the new scale. Press EXE to show the working screen.
K0320 Issue No. 1
NOTE:
Scaling in lower ranges only
when auto ranging.
Page 34
72
Special Measurement Functions
Setting up scaled readings (flow relationship)
Only available on the mA ranges. Select SCALE and FLOW to open the set-up screen. High- and low-limits for the span are preset at 4 and 20 mA. Change span if necessary. Fill out both scale ends for the new scale. Press EXE to show the working screen.
NOTE:
Readings are flow related when simulating or reading dP-transmitters measuring across an orifice plate.
K0320 Issue No. 1
Page 35
74
Output of electrical signals
General
Select an output range and select DIRECT from the menu. It puts you right in the output mode. For other modes read section "Special Output Functions". The lower part of the screen shows the window for output functions. Here, you can read the output level labelled with either SET or OUTPUT.
Set mode
In the SET mode you change the output reading without changing the actual output at the terminals. Once you press EXE, the output will change. to the new setting.
K0320 Issue No. 1
Page 36
76
Output of electrical signals
Output mode
In the OUTPUT mode you change both the output reading and the actual output at the terminals. Entering a new value through the numerical key-pad brings you back into the SET mode.
Changing the output level
Use the
and keys to ramp the
output signal manually. Holding the key down will gradually accelerate the adjustment speed. Enter a new output value through the numerical key-pad and press EXE to activate the signal at the output terminals. Entered values outside the range will prompt the message OUTSIDE RANGE.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
K0320 Issue No. 1
Page 37
78
Output of electrical signals Millivolts
Go to millivolts in the DIRECT mode to output any signal between -10 mV and 100 mV. Adjustments are made with 1 microvolt resolution. If the adjusted output level cannot be maintained the screen will prompt the warning CHECK LOOP.
NOTE:
‘Terminal to test lead’ junctions may create e.m.f. microvolts causing an offset in the actual output signal.
Millivolts output Millivolt Geben Sortie millivolts
K0320 Issue No. 1
Page 38
80
Volts output Volt Geben Sortie volt
Transmitter (XMTR) Simulation Simulieren 2-Leiter Transmitter Simuler un transmitter 2-fils
Output of electrical signals Volts
Go to Volts in the DIRECT mode to output any signal between zero and 12 Volts. Adjustments are made with 100 microvolts. If the adjusted output level cannot be maintained the screen will prompt the warning CHECK LOOP.
Milliamps/XMTR
Use mA terminals to source a current into a resistor (active mode). Use XMTR terminals to simulate a 2-wire transmitter (passive mode). Go to milliamps in the DIRECT mode to output any signal between 0 and 24 mA. Adjustments are made with 1 microamps resolution. If the adjusted output level cannot be maintained the screen will prompt the warning CHECK LOOP. To output fixed 0, 4, 8, 12, 16 and 20 mA levels, press
key to change to fixed steps.
Use
and keys to adjust the output
current. Press
key to return to
normal output mode.
K0320 Issue No. 1
Page 39
82
Output of electrical signals Ohms
Go through the OHM menu and select either the 0-400 or the 0-2000 Ohms range and enter the DIRECT mode. Resistance is simulated across the two terminals as shown. Connecting a third or fourth wire adapts the simulation to 3 or 4 wire input. Adjustment resolution is respectively 0.01 Ohms and 0.1 Ohms.
Frequency
Pulses are symmetrical square wave and -70 mV zero based. Frequency is set by the
and keys or numerical
key-pad. Resolution is 0.1 Volt and amplitude is default set to 5.0 Volts. With frequency set at 0 Hz this function becomes available as a voltage adjustable and regulated DC supply. Frequency output can take loads up to 34 mA at 24 Volts maximum.
K0320 Issue No. 1
Frequency range 0-100 Hz
Go to pulse DIRECT mode to output any pulse signal between zero and 100 Hz. Adjustments are made with 0.01 Hz resolution.
Page 40
84
Output of electrical signals
Frequency range 0-20 kHz
Go to pulse DIRECT mode to output any pulse signal between zero and 20.000 Hz. Adjustments are made with 1 Hz resolution.
Frequency range 0-6000 P./Min.
Same as 0-100 Hz but reads in Pulses/Minute
Pulse output range
0 - 99 999 P./Hour
Same as 0-100 Hz but reads in Pulses/Hour
Counter
Go to the pulse PRESET mode and set the amplitude. Press EXE to proceed to the working screen. Use the and keys to move the cursor from upper to lower line and vice versa. Fill out the number of pulses to be transmitted as well as the Baud rate (max. 18 000 pulses/seconds). Press EXE to start the pulse transmission.
K0320 Issue No. 1
Page 41
86
To simulate a thermocouple
Go through the thermocouple menu and make the selections as required. Select the DIRECT mode to simulate any temperature within the range of the chosen thermocouple type. Adjustment resolution is 0.1 degree.
For cold junction compensation considerations and different wiring methods read section "To measure thermocouples" in this book.
K0320 Issue No. 1
Page 42
88
2 wire 2 Leiter 2 fils
3 wire 3 Leiter 3 fils
RTD Simulation
Go through the RTD menu and make the selections as required. Select DIRECT mode to simulate any temperature within the range of the chosen RTD type. Adjustment resolution is 0.1 or 0.01 degree depending on RTD type. See specifications for details.
Resistances representing temperatures are simulated across the two terminals as shown. Connecting a third or a fourth wire adapts the simulation to 3 or 4 wire inputs.
Excitation current
The excitation current to source a resistance should be between 0.175 and 3.500 mA. Excitation currents may be intermittent as with some smart temperature transmitters.
4 wire 4 Leiter 4 fils
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
K0320 Issue No. 1
Page 43
90
Special Output Functions
Step Mode
In the step mode the MCX-II can generate preset output levels in three different ways;
INPUT
OHM FREQ. COUNT ER
SWI TCH
BATT ERY
OUT P UT
mA/ XMT
mV VOLT T/ C RT D
K0320 Issue No. 1
Page 44
92
Special Output Functions
Free programmable, Select PROGR
Number of steps (2 to 10) and levels are free programmable.
STEP
PROGR.
10 % 20% 30%
10% Divisions, Select 10%
Free programmable span is automatically divided in 10 steps of 10% each.
20% Divisions, Select 20%
Free programmable span is automatically divided in 5 steps of 20% each.
K0320 Issue No. 1
25% Divisions, Select 25%
Free programmable span is automatically divided in 4 steps of 25% each.
Stepping through the fixed levels can either be performed manually or automatically.
Page 45
94
Special Output Functions
Ramp Mode
The MCX-II can generate automatically a time linear increasing or decreasing output signal. Signal limits are free to set within any of the available output ranges. Time scales can be set between 1 and 9999 seconds.
NOTE:
The automatic ramp mode cannot be used together with the frequency measurement mode and the pulse count mode.
INPUT
OHM FREQ. COUN TER
SWI TCH
BAT TERY
OUT P UT
mA/ XMT
mV VOLT T/ C RT D
Setting up the ramp mode
Select RAMP and fill out the set-up screen. The high-end value sets the maximum output. The low-end value sets the minimum output. Fill out the required travel and dwell time in seconds and press EXE.
K0320 Issue No. 1
Page 46
96
Special Output Functions
To start/stop a ‘one shot’ ramp
Press to start a one shot ramp for signal increase and press
to start
this ramp for signal decrease. Press
and simultaneously to
‘freeze’ the ramp action. To restart press
or . Press CE to
cancel the ramping operation.
To start/stop ramp cycling
Press
and direct after each other
to start a continuous up and down ramp cycling with dwelling at the high and low ends. The first arrow key hit determines the take-off direction. Press
and simultaneously to
‘freeze’ the ramp action. To restart press
or . Press CE to cancel the
ramping operation.
K0320 Issue No. 1
Scaled readings
Dial directly in scaled readings to SET a related output signal as set-up. Press EXE to actually output the signal. The and key functions are available to fine adjust the output.
Page 47
98
User Power supplies
Power supplies available for the user are designed to power transmitters, control and transducer wheatstone bridges. The following power supplies are available at the terminals as indicated:
24 VDC fixed, combined with the mA input terminals, current limited at 32 mA.
24 VDC fixed, combined with the mA output terminals, current limited at 32 mA.
0-24 VDC adjustable, combined with the frequency output terminals. Resolution is 0.1 Volt. With frequency set at 0 Hz, output becomes available as a voltage adjustable and regulated power supply. Current limited at 34 mA.
K0320 Issue No. 1
Page 48
100
Transmitter calibration
In this function transmitter source and measure readings are both scaled in engineering units for quick comparison. Transmitter may be a linearized, non­linearized, 4-20 or 0-20 mA model for thermocouples or RTD’s.
Setting up a calibration
Leave the cursor on NONE in the input menu and select the required output function.
Go through the menu and make the selections as required. In the mode screen, select CAL to calibrate a transmitter. Proceed to select all transmitter characteristics.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
K0320 Issue No. 1
Page 49
102
Transmitter calibration
To check or calibrate a transmitter
All normal key-pad functions are available to change the output. Use the
key to change to fixed steps.
Use and keys to step through 0%, 25%, 50%, 75% and 100% of the span as set-up. Press
to
return to the normal key-pad functions. An ideal transmitter will show equal readings. The difference of both readings is expressed in % of the transmitter span or reading and is shown in the upper window in small size digits.
K0320 Issue No. 1
Page 50
104
0-20 mA 4-20 mA
mA out
Control system Regelsystem Système de régulation
4-20 mA
XMTR
Control system Regelsystem Système de régulation
Temperature transmitter simulation
User can dial a temperature in degrees to output a 4-20 mA or 0-20 mA signal. This signal can be used to simulate transmitters at control system input terminals. The simulated transmitter may be linearized or non­linearized. Function is available for all type of temperature sensors supplied as standard in the MCX-II.
Setting up a simulation
Leave the cursor on NONE in the input menu and select the required RTD or T/C function (ignore the CJ compensation menu). In the mode screen, select SIM to simulate a transmitter and proceed to select all transmitter characteristics. Fill-out the transmitter span and press EXE to open the working screen. The lower windows shows the MCX-II output milliamps in small digits.
K0320 Issue No. 1
Page 51
106
Temperature transmitter simulation
To simulate a transmitter
All normal key-pad functions are available to change the (mA) output. Use the
key to move to fixed steps.
Use and keys to step through 0%, 25%, 50%, 75% and 100% of the span as set-up. Press
to
return to the normal key-pad functions. Next to the simulated temperature the lower window shows the MCX-II output milliamps in small size digits.
K0320 Issue No. 1
Page 52
108
Programmable signal converter
With this unique feature any input signal (including the probe) can be converted into any on the MCX-II available output signals with full galvanic isolation. Zero and span settings for both input and output are fully programmable.
Setting up a signal converter
Make all input and output selections as you would do when you use input and output functions separately in the DIRECT output mode. When the normal working screen is displayed press # to open the set-up screen and select convert. Fill out the input- and output span and press EXE again to proceed to the signal converter working screen. Actual input and output signals are now displayed.
K0320 Issue No. 1
NOTE:
The signal converter function can not be used in pulse counting ranges.
Page 53
110
Ce Remarks
Ce Remarks
The MCX-II prompts EMI (Electro­Magnetic Interference) whenever the functionality is influenced by electromagnetic disturbances. The results obtained during this state are unreliable.
The MCX-II is protected against failures by fast transient/burst. It restarts itself whenever this situation occurs and returns to the main screen.
K0320 Issue No. 1
In this way a save state of operation can be assumed.
Page 54
112
Service, repair and parts
Recalibration of the MCX-II
Leave the cursor at NONE in the INPUT and OUTPUT menu and press the # key to open the settings menu. Select CALIBRATION and press EXE. Fill out the access code and press EXE to open the calibration menu.
The access code is fixed at 9410 and works as an barrier to prevent accidental access to MCX-II calibration ranges.
The calibration menu shows all relevant ranges to calibrate. Each range can be recalibrated independent of any other range. Only the cold junction compensationsystem is influenced by millivolts recalibration.
Select the ranges to be calibrated and press EXE. Follow the instructions on the next screens. Fill out the calibration readings and press EXE to confirm or to return to the calibration menu. Select an other range to continue recalibration or press CE to leave the calibration menu.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
K0320 Issue No. 1
FAX 781.665.0780 - TestEquipmentDepot.com
Page 55
114
Service, repair and parts
Cold Junction Compensation Calibration
To recalibrate or check the cold junction measurement system for both input and output, use an electronic thermometer with a maximum error of ±0.07°C (0.13°F). Remove the compensation wire fixation screws from the T/C terminals. Insert the thermometer probe in the appropriate T/C terminal hole and make sure a good thermal contact is made down at the bottom of the terminal. Take measurements samples of both terminals and calculate the average temperature. Enter this temperature to recalibrate. Do not use thermal conductivity gel as this can not easily be removed from the terminal.
K0320 Issue No. 1
Page 56
116
Service, repair and parts
NOTES:
1. Calibration standards should
have a valid calibration certification and should be at least 3 times more accurate as the published accuracies for the MCX-II for each range.
2. As derived from electrical parameters, temperature sensor ranges do not require any individual recalibration.
3. Be sure that you use the test leads, as standard supplied with the MCX-II or other low e.m.f. test leads.
4. The MCX-II should be powered from the on board batteries during recalibration.
5. Recalibration must be carried out at reference ambient temperature and humidity. MCX-II must be exposed to this condition for at least 2 hours before the actual recalibration starts.
6. Recalibrate mV ranges before starting recalibration of the Cold Junction Compensation system.
7. When accessing the calibration menu the temperature unit is set to °C and the temperature scale is set to IPTS 68.
8. Calibration is recommended between 12 and 14 months.
K0320 Issue No. 1
Page 57
118
Processor board p/n 25100 Processorkarte p/n 25100 Carte processeur p/n 25100
Service, repair and parts
Replacing the lithium battery
Clock problems are in most cases caused by a weak lithium cell. To replace the cell, open the unit as described under chapter “fault finding procedures”. Use a small screwdriver to lift the cell from its socket. Remove the cell by sliding it away. Replace by 3 Volts cell type CR2025 only.
K0320 Issue No. 1
Page 58
Service, repair and parts
120
Fault finding procedures
Opening and closing the calibrator housing
Remove the batteries and disconnect the unit from other equipment. Take the four T/C wire screws out and release all visible Phillips screws at the back and inside the battery compartment. Lift gently the housing upper half off and put it away. Disconnect the power cable from the board and the cable to the rotatable unit. The electronic board assembly and the rotatable unit can now be taken out of the lower half of the housing.
To open the rotatable unit release the two screws at the same half on each side.
To close the calibrator housing in reverse order the following must be considered;
1. Place the rotatable unit back in the click mechanism slots as found before.
2. Make sure that the cable between base unit and rotatable unit is correctly routed as indicated.
3. Put the three printed circuit board spacers in place. Left-hand and right­hand spacers have brass wire guide tubes; make sure they fit into the mV terminals correctly.
4. Check the "battery type" selector switch position before turning the unit on.
K0320 Issue No. 1
Page 59
122
Service, repair and parts
Fault finding procedures
When the MCX-II doesn’t work properly or doesn’t work at all, it is recommended to return the unit to one of the GE Service Centres or to your local distributor after you have checked the following;
Most of the problems with portable
instruments are related to power
supply defects, blown fuses or
displaced connectors.
The unit doesn’t work at all.
1. When normally line powered, change to battery operation to check the line adaptor.
2. When normally battery powered, check the contact springs for corrosion and spring force.
K0320 Issue No. 1
3. Check total battery voltage
with a voltmeter. Alkaline should read around 9 Volts and NiCd’s should read around 7.2 Volts when new or fully charged.
4. Open the MCX-II as described in the previous paragraph and check whether the PCB interconnecting plugs are still in position.
Page 60
124
Input board p/n 25102 Eingangskarte p/n 25102 Platine d'entrée p/n 25102
Power board p/n 25101 Versorgungskarte p/n 25101 Platine alimentation p/n 25101
Service, repair and parts
5. The unit doesn’t measure milliamps.
Check the fuse accessible from the front. Replace fuse only by p/n 13119 or Littelfuse No. 217.400.
6. The unit doesn’t measure Ohms or RTD’s
Open the MCX-II and check the fuse on the input board. Replace fuse only by p/n 13119 or Littelfuse No. 217.400.
7. The Optional “Communicator” doesn’t work.
Open the MCX-II and check the fuse on the power board. Replace fuse only by p/n 13119 or Littelfuse No.217.400.
Most other fault symptoms are not logically related to just one possible defect. When you have spare printed circuit boards available, you could replace board by board to find the faulty board. Note that a board replacement requires a recalibration. Note that this procedure may also create a problem in a new healthy board by specific defects in a original board.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
K0320 Issue No. 1
Page 61
Kundendienst, Reparatur und Ersatzteile Maintenance, répartions et pièces détachées
5. Das Gerät mißt keine Milliampere
Prüfen Sie die von der Vorderseite erreichbare Sicherung. Ersetzen Sie sie gegebenenfalls nur durch eine p/n 13119 oder Littelfuse Nr.
217.400.
6. Das Gerät mißt keine Ohm oder Widerstandsthermometer
Öffnen Sie den MCX-II und prüfen Sie die Sicherung auf der Eingangskarte. Ersetzen
5. L'appareil ne mesure pas le millampères
Vérifier le fusible accessible à l'avant. Le remplacer uniquement par un fusible réf. 13119 ou un fusible "Littelfuse" N
0
217.400.
6. L'appareil ne mesure pas les ohms ou les RTD.
Ouvrir le MCX-II et vérifier le fusible sur la carte d'entrée. Le remplacer uniquement par un fusible réf. 13119 ou on fusible "Littelfuse" N
0
217.400. Sie sie gegebenenfalls nur durch eine p/n 13119 oder Littelfuse Nr. 217.400.
7. Die Option "Communicator" funktioniert nicht
Öffnen Sie den MCX-II und prüfen Sie die Sicherung auf der Versorgungskarte. Ersetzen Sie sie gegeben enfalls nur durch eine p/n 13119 oder Littelfuse Nr. 217.400.
7. La carte de communication optionelle ne fonctionne pas.
Ouvrir le MCX-II précédemment et vérifier le fusible sur la carte alimentation. Le remplacer uniquement par un fusible réf. 13119 ou un fusible "Littelfuse" N
0
217.400.
125
Die meisten anderen Fehlersymptome stehen nicht in einer logischen Beziehung zu einem einzigen möglichen Fehler. Wenn Sie Ersatzkarten zur Verfügung haben, können Sie Karte für Karte ersetzen und so die Fehlerhafte finden. Bedenken Sie, daß der Austausch einer Karte eine Neukalibrierung erfordert. Bedenken Sie ferner, daß durch diese vorgehensweise ein Fehler in einer neuen intakten Karte entstehen kann, verursacht durch einen bestimmten Fehler in einer Originalkarte.
La plupart des autres symptômes peuvent avoir plusieurs causes. Si des cartes de rechange sont disponibles, remplacer les cartes une à une afin d'identifier la carte défectueuse. A noter qu'en cas de remplacement d'une carte, il faut réétalonner l'appareil. Il se peut aussi que cette opération crée un problème dans une carte neuve saine du fait de défauts propres à la carte d'orgine.
K0320 Issue No. 1
Page 62
Service, repair and parts
The following boards are accommodated in the MCX-II:
126
Power supply board p/n 25101
Processor board p/n 25100
Input board p/n 25102
Output board p/n 25103
Digital communicator board
(Optional) p/n 25107
LCD/Key-pad processor board (in rotatable panel) p/n 25105
Key-pad board (in rotatable panel) p/ n 25106
For GE Sales and Service Points see back cover.
K0320 Issue No. 1
Page 63
128
Service, repair and parts
Spare parts list
Part No.
Power board Assy 152 Processor board Assy 149 Input board Assy 153/II Output board Assy 154 LCD display 22500 LCD display board Assy 156/II Key-pad board Assy 157 Board spacers (3X) 25207 T/C terminal screws (2X) 25301 6x Test leads, 6x clips Assy 162 2x Switch protectors Operating Manual 25500 Line adaptor/charger 115VAC 13603/115 Line adaptor/charger 230VAC 13603/230 Pt100 hand-held probe Assy 170 Switch protector (3X) 24003 Battery compartment cover 25204 Set of battery contact springs 13013/16 LCD display window 25125 Set of fuses (10X) 13119 Fuse holder 13118 Carrying case 24005 Shoulder strap 24007 Lithium cell 13183
K0320 Issue No. 1
Page 64
Electrical functions
Measure millivolts
auto.ranging
Output millivolts
130
Measure Volts
auto.ranging Output Volts Measure milli-ampére Output milli-ampére Simulate 2-wire XMT Measure Ohms
auto.ranging Simulate Ohms Simulate Ohms Measure frequency
auto.ranging
auto.ranging Counter Counts/minute Counts/hour Output frequency
Output pulses preset
Temperature functions
Pt1000 Measure
Pt500 Measure
Pt200 Measure
Pt100 Measure
Pt50 Measure
D-100 Measure
Ni100 Measure
Ni120 Measure
Cu10 Measure
Measure/Simulation TC - J Measure/Simulation TC - L Measure/Simulation TC - K
Measure/Simulation TC - T
Measure/Simulation TC - U
Measure/Simulation TC - B
Measure/Simulation TC - R
Measure/Simulation TC - S
Measure/Simulation TC - E
Measure/Simulation TC - N
Measure/Simulation TC - C
Measure/Simulation TC - D
Simulation
Simulation
Simulation
Simulation
Simulation
Simulation
Simulation
Simulation
Simulation
Specifications
Range
+ 0...100mV +100...600mV
-10...100mV + 0...6V + 6...60V
0...12V + 0...52mA
0...24mA
4...24mA
0...400Ω
400...2000Ω
0...400Ω
0...2000Ω
0...655Hz
655...1310Hz
1310...10.000Hz
0...99999999 600000 9999999
0...100Hz 6000 p/min
0...10000Hz 99999 p/hour
0...99999999
Range
-200/400°C -328/752°F
-200/400°C -328/752°F
-200/850°C -328/1562°F
-200/850°C -328/1562°F
-200/850°C -328/1562°F
-200/850°C -328/1562°F
-200/850°C -328/1562°F
-200/850°C -328/1562°F
-200/850°C -328/1562°F
-200/850°C -328/1562°F
-200/649°C -328/1200°F
-200/649°C -328/1200°F
-60/250°C -76/482°F
-60/250°C -76/482°F
-80/260°C -112/500°F
-80/260°C -112/500°F
-200/260°C -328/500°F
-200/260°C -328/500°F
-210/1200°C -346/2192°F
-200/900°C -328/1652°F
-200/1372°C -328/2502°F
-270/-200°C -454/-328°F
-75/400°C -103/752°F
-180/-75°C -292/-103°F
-270/-180°C -454/-292°F
-100/600°C -184/1112°F
-200/-100°C -328/-184°F
1200/1820°C 2192/3308°F
500/1200°C 932/2192°F
250/500°C 482/932°F
50/250°C 122/482°F
300/1769°C 572/3216°F
-50/300°C -58/572°F
400/1769°C 752/3216°F
100/400°C 212/752°F
-50/100°C -58/212°F
-150/1000°C -238/1832°F
-270/-150°C -454/-238°F
-20/1300°C -4/2372°F
-200/-20°C -328/-4°F
-270/-200°C -454/-328°F 0/1500°C 32/2732°F
1500/2000°C 2732/3632°F 2000/2320°C 3632/4208°F
0/1700°C 32/3092°F
1700/2200°C 3092/3992°F 2200/2495°C 3992/4523°F
Resolution
0.001
0.01
0.001
0.0001
0.001
0.0001
0.001
0.001
0.001
0.01
0.1
0.01
0.1
0.01
0.1 1 1 1 1 0,01 1 1 1 1
Resolution
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.03°C 0.06°F
0.03°C 0.06°F
0.06°C 0.11°F
0.06°C 0.11°F
0.03°C 0.06°F
0.03°C 0.06°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.3°C 0.6°F
0.3°C 0.6°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.2°C 0.4°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.2°C 0.4°F
0.3°C 0.6°F
0.5°C 1.0°F
0.1°C 0.2°F
0.2°C 0.4°F
0.1°C 0.2°F
0.2°C 0.4°F
0.2°C 0.4°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.2°C 0.4°F
0.1°C 0.2°F
0.1°C 0.2°F
0.2°C 0.4°F
0.1°C 0.2°F
0.1°C 0.2°F
0.2°C 0.4°F
K0320 Issue No. 1
Page 65
1 year accuracy
0.004% Rdg. + 0.004% f.s. +1 lsd
0.005% Rdg. + 0.005% f.s. +1 lsd
0.003% Rdg. + 0.004% f.s. +1 lsd
0.009% Rdg. + 0.003% f.s. +1 lsd
0.009% Rdg. + 0.003% f.s. +1 lsd
0.004% Rdg. + 0.002% f.s. +1 lsd
0.010% Rdg. + 0.003% f.s. +1 lsd
0.010% Rdg. + 0.005% f.s. +1 lsd
0.010% Rdg. + 0.005% f.s. +1 lsd
0.005% Rdg. + 0.008% f.s. +1 lsd
0.01Hz
0.1Hz 1 Hz infinite 1 c/min 1 c/hour
0.01Hz 1 p/min 1Hz 36 p/hour infinite
0.012% f.s. +1 lsd
0.012% f.s. +1 lsd
0.010% f.s. +1 lsd
1 year accuracy + (abs. + 1lsd)
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.2°C 0.4°F
0.3°C 0.5°F
0.15°C 0.27°F
0.12°C 0.22°F
0.25°C 0.45°F
0.20°C 0.36°F
0.15°C 0.27°F
0.12°C 0.22°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
1.0°C 1.8°F
1.5°C 2.7°F
0.1°C 0.2°F
0.1°C 0.2°F
0.1°C 0.2°F
0.3°C 0.6°F
0.1°C 0.2°F
0.2°C 0.4°F
0.5°C 0.9°F
0.1°C 0.2°F
0.2°C 0.4°F
0.4°C 0.7°F
0.7°C 1.3°F
1.5°C 2.7°F
2.0°C 3.6°F
0.5°C 0.9°F
1.0°C 1.8°F
0.5°C 0.9°F
0.8°C 1.5°F
1.0°C 1.8°F
0.1°C 0.2°F
0.3°C 0.6°F
0.1°C 0.2°F
0.4°C 0.7°F
1.0°C 1.8°F
0.2°C 0.4°F
0.3°C 0.7°F
0.6°C 1.1°F
0.2°C 0.4°F
0.4°C 0.7°F
1.1°C 2.0°F
Remarks
R - input > 20MΩ R - input > 20MΩ R - output < 0.2Ω R - input > 1MΩ R - input > 1MΩ R - output < 0.2Ω R - input 2,5Ωfused R - max. 900Ω V-max. 56 Volts at 0.9mA excitation at 0.9mA excitation at 1.0mA excitation at 1.0mA excitation R - input > 300kΩ R - input > 300kΩ R - input > 300kΩ R - input > 300kΩ R - input > 300kΩ R - input > 300kΩ 0 - 24V / 34mA max. 0 - 24V / 34mA max. 0 - 24V / 34mA max. 0 - 24V / 34mA max. 0 - 24V / 34mA max.
Remarks
IEC 751 IEC 751 IEC 751 IEC 751 IEC 751 IEC 751 IEC 751 IEC 751 IEC 751 IEC 751 JIS C1604 1989 JIS C1604 1989 DIN 43760 Aug. 1985 DIN 43760 Aug. 1985 Minco 7 Minco 7 Minco 16 - 9 Minco 16 - 9 IEC 584-1 DIN 43710 Dec. 1985 IEC 584-1 IEC 584-1 IEC 584-1 IEC 584-1 IEC 584-1 DIN 43710 Dec. 1985 DIN 43710 Dec. 1985 IEC 584-1 IEC 584-1 IEC 584-1 IEC 584-1 IEC 584-1 IEC 584-1 IEC 584-1 IEC 584-1 IEC 584-1 IEC 584-1 IEC 584-1 IEC 584-1 IEC 584-1 IEC 584-1 ASTM E988-96 ASTM E988-96 ASTM E988-96 ASTM E988-96 ASTM E988-96 ASTM E988-96
131
K0320 Issue No. 1
Page 66
Special functions
132
Specifications
1 Steps; 10 programmable, 10%, 20%,
25% div. stepping by key or adj. timer
2 Ramp; programmable travel time for
up/down and dwell
3 Scaling; in 5 digits and sign on all
electrical ranges
4 Temp. XMT cal.; both input and output
readings in temp. units. Cal.function extended for all output functions.
5 Temp. XMT sim.; mA output reads in
temperature units
6 Signal converter; any input into any
output, fully isolated
Note: Thermocouple accuracies do not include cold junction compensation errors
Reference 22°C ±1, RH 45% ±15 Calibration Traceable to National Standards Accuracies For 17°C to 27°C; outside these
Cold Junction error ±0.4°C (0.8 °F) max. in transient
Long term drift after 1 year 0.005% / year or 1000 hours operating Cold Junction modes Automatic, Manual, Remote sensor Temperature scale Selectable IPTS68 or ITS90 Warm up time 2 minutes for rated accuracies Re-calibration From key-pad (no inside trimmers)
Operating temp. -10°C to 50°C (14°F to 122°F) Storage temp. -20°C to 70°C (-4°F to 158°F) Relative humidity 0 - 90% non-condensing In-/output isolation 500VDC continuous Electrical protection 50 Volts, 30 sec. max.
Output mismatch Displays " check loop"
limits 0.0005% / °C typical on mV
±0.2°C (0.4°F) max. when steady
Recommended for each year
Ohms out to ground; 5 Volts max.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
K0320 Issue No. 1
Page 67
Overrange Displays ">>>>> " Underrange Displays "<<<<< " Readings Temp. °C or °F Batteries (LR14 or C) 6x 1.5V alkaline or 1.2V NiCd
Battery life Alkaline; 22 hours at 20°C (68°F)
Battery life 20 mA out Alkaline; 12 hours at 20°C (68°F)
Low battery warning Flashes a battery symbol Line power With optional adaptor 115 - or 230VAC,
Recharge NiCd's With optional adaptor 115 - or 230VAC,
User terminals Gold plated, standard 4mm plugs T/C comp.wire terminals Screw fixable, 2mm diameter max. Protection IP 53 Housing Textured high-impact ABS plastic Control panel Rotatable, 5 clicks of 30° Size 265 x 160 x 56/80mm without case
Carrying case Black Cordura Weight 2.2kg (5.5lbs) with batteries Keystroking 10 operation sequences storage Contact change detect. Reading freezes on open or close Date and time Recorded when PCMCIA card is used PCMCIA station PCMCIA card type 1 or 2 Detect. level freq.in Adjustable; 10mV to 5.0 Volts Pulse output level Adjustable; 0 - 24 Volt, 0.1V resolution,
Pulse output form Symmetrical square, zero based - 70mV Pulse output speed Adjustable in Hz or Baud rate Remote sensor input Mini DIN Suitable for Pt100 (385 curve)
LCD display Full graphic, with contrast control and
Ohms simulation 0.18 to 3.5mA excitation current
Ohms input terminals 2, 3 or 4 wire automatic mode select.
Auxiliary connectors For use with optional pressure module Dig. communicator board Slot supplied as standard for electronic
20 mA out NiCd; 7 hours at 20°C (68°F)
NiCd; 14 hours at 20°C (68°F)
green LED indicator
amber LED indicator
(10.6 x 6.4 x 2.2 /3.2) inch
2% accuracy
full range, 3 or 4 wire
backlight timer
either polarity
True 4 - wire system
board
133
K0320 Issue No. 1
Page 68
Specifications
Protocol for RS232 System Integration
142
Serial port settings
Bauds: 9600 Parity: none Bits: 8 Stop: 1 Hardware handshake: none
K0320 Issue No. 1
Page 69
144
Warranty & Liability
Our equipment is warranted against defective material and workmanship for a period of 18 months from date of shipment. Claims under warranty can be made by returning the equipment prepaid to our factory. The equipment will be replaced, repaired or adjusted at our option. The liability of GE is restricted to that given under our warranty. No responsability is accepted for damage, loss or other expense incurred through sale or use of our equipment. Under no condition shall GE be liable for any special, incidental or consequential damage.
NOTE: Units with defects caused by
battery electrolyte from leaking batteries are definitely excluded for repair or replacement under warranty.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
K0320 Issue No. 1
Loading...