GE Druck DPI 841, Druck DPI 842 User Manual

Page 1
GE Infrastructure
Sensing
Druck DPI 841/842
Frequency calibrator and Frequency loop calibrator
User manual - K395
Page 2
A1
B1
10
1
A 2
9
A 3
8
2
3
4
7
6
5
11
B1
12
A2
1513
14
A3
17182619 2716
20
21
22 23
K395 Issue 1
24
25
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
Page 3

Table of Contents

Introduction ......................................................................... 1
Safety ..................................................................................... 1
Safety - Marks and symbols on the instrument ............... 2
To start .................................................................................. 2
To start - Location of items ........................................................ 2
To start - Items on the display .................................................. 2
To start - Prepare the instrument ........................................... 2
To start - Power on or off ............................................................ 3
To start - Set up the basic operation ..................................... 3
To start - Select a task (Measure and/or supply) ............. 3
To start - Set up the settings ..................................................... 4
To start - Edit functions ................................................................ 4
Operation .............................................................................. 6
Operation - Frequency connections ...................................... 6
Operation - Communications port connections .............. 6
Operation - Measure Hz or count pulses ............................. 6
Operation - Change the output values ................................. 6
Operation - Supply Hz or pulses .............................................. 7
Operation - Transmitter calibration ....................................... 7
Operation - mA measurements ............................................... 8
Operation - Switch test ................................................................ 8
Operation - UPM Pressure measurements ......................... 9
Operation - Error indications ..................................................... 9
Maintenance ..................................................................... 10
Maintenance – Clean the unit ................................................ 10
Maintenance – Replace the batteries ................................ 10
Calibration ......................................................................... 10
Calibration - Before you start ................................................. 10
Calibration - Procedures (Hz - input/output) ................... 10
Calibration - Procedures (mA input) .................................... 11
Calibration - Procedures (Amplitude output) .................. 11
Calibration - Procedures (IDOS UMM) ................................. 12
Specification data ............................................................ 12
Specification - General .............................................................. 12
Specification - Frequency ........................................................ 12
Specification - Electrical connectors (A2) ......................... 12
Customer service ............................................... Back cover
© 2005 General Electric Company. All rights reserved.
Trademarks
All product names are trademarks of their respective companies.

Introduction

The DPI 841 Frequency Calibrator and DPI 842 Frequency Loop Calibrator are part of the Druck DPI 800 series of hand held instruments. The DPI 800 series uses Intelligent Digital Output Sensor (IDOS) technology to give instant plug and play functionality with a range of Universal Measurement Modules (UMM). Example: the Universal Pressure Module (UPM). The DPI 841/842 include these functions:
Function DPI 841 DPI 842
Measure/supply a frequency or a pulse count.
Step/Ramp functions Automatic/Manual Communications port IDOS or RS232 Language selection Yes Measure pressure/Leak test ** External IDOS UPM ** Snapshot Up to 1000 displays with a date/time
Measure mA No 0 - 55 mA HART® resistor No Yes V dc output No 24 V Switch test No Yes Other functions Hold, Maximum/Minimum/Average,
Filter, Tare, Scaled values, Backlight, Alarm
* Refer to “Specification data”. ** Optional item
* Yes
stamp

Safety

Before you use the instrument, make sure that you read and understand all the related data. This includes: all local safety procedures, the instructions for the UMM (if applicable), and this publication.
WARNING
• It is dangerous to ignore the specified limits for the instrument or to use the instrument when it is not in its normal condition. Use the applicable protection and obey all safety precautions.
• Do not use the instrument in loca tions with explosive gas, vapor or dust. There is a risk of an explosion.
Continued
[EN] English - 1K395 Issue 1
Page 4
Safety (Continued)
• To prevent electrical shocks or damage to the instrument, do not connect more than 30V between the terminals, or between the terminals and the ground (earth).
• UPM only. To prevent a dangerous release of pressure, isolate and bleed the system before you disconnect a pressure connection.
Before you start an operation or procedure in this publication, make sure that you have the necessary skills (if necessary, with qualifications from an approved training establishment). Follow good engineering practice at all times.

Safety - Marks and symbols on the instrument

Complies with European Union directives
Read the manual Battery
Ground (Earth)
Warning - refer to the manual
ON/OFF

To start

To start - Location of items

Item Description
1. On or off button.
2.
3.
4.
5.
6.
7.
8. Display. Refer to A3
9.
10.
11. Connection point for some of the optional
12. Battery compartment. Refer to B1.
13., 14., 15. DPI 842 only. Terminals to measure current, to supply
Left-hand soft-key. Selects the function above it on
■ ■
the display (Item 25). Example: Edit Moves back one menu level.
ESC
Leaves a menu option. Cancels the changes to a value.
Increases or decreases a value.
▲ ▼
Highlights a differen t item. Holds the data on the display. To continue, press the
HOLD
HOLD button ag ain. Shows the Select Task menu.
MENU
Selects or accepts an item or value.
OK
Selects [] or cancels [ ] a selection. Right-hand soft-key. Selects the function above it on
■ ■
the display (Item 25). Example: Settings
Communications port. Use to connect a Universal
SENSOR
/ PC
Measurement Module (UMM) or a RS232 cable. Terminals to measure (IN) or supply (O UT) a
IN
OUT
frequency or a pulse rate. Refer to “Operation”.
accessories. Refer to the datasheet.
24V source, and to do switch tests.
A1 A2

To start - Items on the display

Item Description
16. DPI 842 only. Task indication for the switch test.
17. DPI 842 only. The loop power supply is on.
18. The measured value satisfies one of the alarm
19. The data on the display is on hold. To continue, press
20.
21. Identifies the type of data and the measurement
22. ... 24. The settings applied to the input or output.
22. kHz The units or a speci fied scale (x:y) - (Table 4/5)
5.0V
23.
24. = Filter
25. A soft-key function. To select an available function,
26. The measured value or values applicable to the task
27. The Edit display to set up text labels ( 6 characters):
= switch closed
UPM only. Task indication for the leak test.
Refer to: Select Task (Table 2/3)
Refer to: Select Task (Table 2/3)
conditions. Refer to: Settings (Table 4)
the HOLD button again.
Shows the battery level: 0 ... 100%.
range.
= Input = Output = IDOS input
Refer to: Select Task (Table 2/3)
, , =
...V The input trigger level (Table 4) or the output
amplitude (Table 5).
, ... ,
= Tare
press the soft-key below it. Example:
= Move left = Move right
selection.
x:y Scaling (Table 4).
OK = Accept the new text label
Shift = Change the keys: 123ABC or -_+abc
= Add a space
BS = Back space (Delete character)
A3
= switch open
Output waveform (Table 5)
= Output operation (Table 5)
= Maximum = Average = Minimum
(Table 4)

To start - Prepare the instrument

Before you use the instrument for the first time:
• Make sure that there is no damage to the instrument, and that there are no missing items.
• Remove the plastic film that protects the display. Use the tag (◗ ) in the top right-hand corner.
• Install the batteries (refer to B1). Then re-attach the cover.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
2 - [EN] English K395 Issue 1
FAX 781.665.0780 - TestEquipmentDepot.com
Page 5

To start - Power on or off

To turn the instrument on or off, press ❍ (A1 - item [1]). Th e instrument does a self test and then shows the applicable data. When the pow er is off, the last set of con figuration options stays in memory. Refer to “Maintenance”.

To start - Set up the basic operation

Use the Set Up menu to set up the basic operation of the instrument.
Menu:
1
Select Task
(Table 2) (Table 1) []/[ ]
23
▲ ▼
Menu: Set Up
45
▲ ▼
If there is additional data for a menu option, select Settings (■■
) to see the values that are set up. If
necessary, adjust the values.
Table 1: Menu options - Set Up
Options
(If applicable)
Descript ion
DPI 842 only. To add a series resistor into the mA circuit. You can then use this instrument together with a HAR T® communic ator to set up and calibrate HART® devices.
To select and set up the backlight facility + timer. Additional data: Select Settings ( ) To select and set up the power off facility + timer. Additional data: Select Settings ( )
To show the battery level (%).
To set the display contrast (%). Increases %, decreases % To set the time + date. The calibration facility uses the date to give service and calibration messages. To set the language option.
To calibrate the instrument.
Additional data: Refer to “Calibration”.
To select and show the applicable status data. (Software Build, Calibration Due date, Serial Number, IDOS Information).

To start - Select a task (Measure and/or supply)

When the instrument is set up (Table 1), use the Select Task menu to select the applicable task.
1
Menu:
Select Tas k
(Table 2/3)
23
▲ ▼
Display:
Hz output
Sk1 = Edit Sk2 = Settings
In Table 2/3, IDOS is a Universal Measurement Module (UMM). If you attach a UMM to the communications port (A1 - item [9]), the Select Task menu shows the applicable IDOS options.
Table 2: Menu options - Select Task
Options
(If applicable)
IDOS UMM only. An IDOS measurement task.
UPM only. A leak test.
Description
Hz or
An input measurement task:
Pulses
Hz - Measure the frequency Pulses - Count the number of pulses
Hz or
An output task:
Pulses
Hz - Supply an output frequency
Pulses - Generate a specified number of pulses mA DPI 842 only. A mA measurement task. mA(24V) DPI 842 only. A mA measurement task + the loop
power supply is on .
DPI 842 only. A switch test.
To set up the way the instrument works.
Additional data: Refer to: Set Up (Table 1) .
Table 3 shows all the one and two function operations that are available. If you attach a UMM, you can only use the options that include IDOS.
Table 3: Permitted 1 and 2 function operations
Function
mA (1) x (2) x x (2) mA(24V) (1) x (2) x x (2)
IDOS (1) (2) (2) (2) x x
Hz Hz Pulses Pulses IDOS (1) (1) (1) (1) (1)
x x (2) x x (2) xxxxx(2)
= DPI 842 only
[EN] English - 3K395 Issue 1
Page 6

To start - Set up the settings

When the task is set up (Table 2/3), use the Settings menu to adjust the input and/or output operation.
Display: Task
Hz + mA
Sk1 = Start/Stop Sk2 = Settings
Menu:
Settings
(Table 4) []/[ ] Sk1 = Edit
Menu:
Settings
(Table 5) []/ [ ]
Settings selection
1
■ ■
4A 5A 6A
▲ ▼
4B 5B
▲ ▼
(If applicable)
■ ■
23
▲ ▼
Settings
x:y
If there is additional data for a menu option, select Settings (■■
) to see the values that are set up. If
necessary, adjust the values. Refer to “Edit functions”.
Table 4: (Part of table) Menu options - Settings (Input)
Options
(If applicable)
Trigger level To set the amplitude at which the instrument
Description
... Units Frequency Units (Hz input only). To select one of
these units: Hz: Range < 1000Hz kHz: Range 1 ... 50kHz Pulses/minute (cpm) Pulses/hour (cph)
... Units UPM only = “Pressure Units” if you select an IDOS
task (Table 2/3). Select one of the fixed units of measurement (psi, mbar ... ).
senses a frequency signal. Default = 5V. Auto Detect []/[ ]: Set this option to make the
instrument calculate the value from the available signal. To include maximum, minimum and average values for the measurement task.
Table 4: (Part of table) Menu options - Settings (Input)
Options
(If applicable)
Description
To select and set up a tare value for the measurement tas k (a specified value or the reading on the display). The instrument subtracts a positive tare value, and adds a negative tare value.
Additional data: Select Settings ( ) To select and set up a scale of values: One local scale for each measurement task (Maximum: 5). Additional data (Example 1/2): Select Settings ( ) To select and set up the filter values to give a smoother output for the measurement task:
Band as a % of full scale (FS). The filter
compares each new value with the previous value. If the new value is outside the band, it is not filtered.
Low pass filter time constant in seconds.
Increase the value to increase damping factor. Additional data: Select Settings ( ) To select and set up the alarm values for the measurement task (maximum and minimum).
Additional data: Select Settings ( ) UPM only. Gage sensors or sensors with differential
operation. A zero correction that makes the instrument read zero at local pressure. Leak Test only. To set an applicable period for the leak test (Hours:Minutes:Seconds).

To start - Edit functions

Example 1) Set up a label for x:y Scaling = %.
Settings
x:y
Sk1 = Edit A3 - item [27]
1 Edit 2 Edit 3
■ ■
Edit ... Edit 45
A3 - item [27]
Shift
■ ■
%_
OK
■ ■
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
4 - [EN] English K395 Issue 1
FAX 781.665.0780 - TestEquipmentDepot.com
Page 7
Example 2) Set up values for x:y Scaling = 0 to 100%.
Settings
x:y
Sk1 = Edit
Settings
1
▲ ▼
= Flow scaling (mA, pressure only)
2 Edit 34
x:y
■ ■
▲ ▼
■ ■
Table 5: (Part of table) Menu options - Settings (Output)
Options
Amplitude To set the amplitude and the mode of the output
Waveform To set the waveform for the output signal:
Descript ion
Hz
... Units Frequency Units. To select one of these units:
Hz: Range < 1000Hz kHz: Range 1 ... 50kHz Pulses/minute (cpm ) Pulses/hour (cph)
signal. Amplitude = 5V (Default). Bipolar mode []/[ ]: Set this option to make the signal pass thru zero.
= Unipolar = Bipolar
,,
To select and set up a value f or the “Nudge” output. Example: 0.010 kHz incremen ts. Additional data: Select Settings ( )
To select and set up values for the “Span Check”
output. Example output cycle:
Additional data (Table 6): Select Settings ( ) To select and set up values for the “% Step” output.
% Step
Example output cycle:
Additional data (Table 6): Select Settings ( )
= Sine, square, or triangle
This cycle repeats automatically.
Auto Repeat - Opt ional
Table 5: (Part of table) Menu options - Settings (Output)
Options
To select and set up values for the “Ramp” output.
Options
Description
Hz
To select and set up values for the “Defined Step”
... Step
output. Example output cycle:
Additional data (Table 6): Select Settings ( )
Example output cyc le:
Additional data (Table 6): Select Settings ( )
Description
Pulses
Counts: To set the total number of output pulses. Rate: To set the frequency of the output. Amplitude: To set the amplitude of the output
signal. Amplitu de = 5V (Default). Bipolar mode []/[ ]: Set this option to make the signal pass thru zero.
= Unipolar = Bipolar
Auto Repeat - Optional
Auto Repeat - Optional
Table 6: Additional data for Settings (Output):
Item Value Span Check
Low (0%) Set th e 0% value. High (100%) Set th e 100% value. Dwell (d) Set the per iod (Hours:Minutes:Seco nds) between
% Step Low (0%), High (100%), Dwell (d): As above. Step Size (s) ... %Set the change in value for each step as a
Defined Step Low (0%), High (100%), Dwell (d): As above. Step Size (s) Set the change in value for each step as a
Ramp Low (0%), High (100%), Dwell (d): As above. Travel (t) Set the period (Hours:Minutes:Secon ds) to go from
Auto Repeat If applicable, select this ite m to repeat a cycle
each change in value.
percentage of the full-scale range (High - Low).
frequency value.
the Low (0%) value to the High (100%) value.
continuously.
[EN] English - 5K395 Issue 1
Page 8

Operation

This section gives examples of how to connect and use the instrument. Before you start:
• Read and understand the “Safety” section.
• Do not use a damaged instrument.

Operation - Frequency connections

To prevent instrument errors, make sure that the frequency connections (A1-item [10]) are correct.

Operation - Communications port connections

Use the communications port (A1 - item [9]) to attach an IDOS Universal Measurement Module (UMM). When you attach the cable from a UMM (Figure 7/8), the instrument automatically changes the menus to give you all the applicable options (Table 2/3).

Operation - Measure Hz or count pulses

To measure Hz or count pulses:
1. Connect the instrument (Figure 1) and, if necessary, adjust the Set Up (Table 1).
2. Select a Hz or Pulses input task from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 4).
3. For Pulses, use Start/Stop ( count. The display shows the time (hh:mm:ss) since you started the count.
) to start and stop the

Operation - Change the output values

When the output operation is set up (Table 5), use one of these procedures to change the output values:
Table 7: Procedures to change the output
Output Procedure
,
,
Pulses
Select Edit (■ ■) and/or use the ▲ ▼ buttons. See the example below. Select Start/Stop (■ ■) or use the ▲ ▼ buttons to make the step changes manually. Select Start/Stop (■ ■).
Example procedure (“Nudge” output):
Display:
Hz output
Sk1 = Edit Sk2 = Settings
Display:
Hz output
1 Edit 23Edit 4
■ ■ ■ ■
56
▲ ▼
1.010
Increment = 0.010
(Table 5)
1.020
+01.000
Pulses
SettingsStartSettings
a) Hz b) Pulses
Figure 1: Example configuration - To measure Hz
or count Pulses
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
6 - [EN] English K395 Issue 1
FAX 781.665.0780 - TestEquipmentDepot.com
Page 9

Operation - Supply Hz or pulses

To supply Hz or pulses:
1. Connect the instrument (Figure 2) and, if necessary, adjust the Set Up (Table 1).
2. Select a Hz or Pulses output task from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 5).
For Pulses, the display shows the time (hh:mm:ss) to complete the count at the specified rate.
3. Supply the output values to the system (Table 7).
Pulses
Maximum: 30V
Edit Settings
Pulses
Edit Settings Start Settings
a) Hz b) Pulses
Figure 2: Example configuration - To supply Hz or Pulses

Operation - Transmitter calibration

DPI 842 only. To calibrate a transmitter:
1. Connect the instrument (Figure 3/4) and, if necessary, adjust the Set Up (Table 1).
2. Select the applicable calibration task from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 4/5).
3. Supply the output values to the system (Table 7).
Figure 3: Example configuration - Transmitter calibration
with external loop power
Edit Settings
Figure 4: Example configuration - Transmitter calibration
with internal loop power
[EN] English - 7K395 Issue 1
Page 10

Operation - mA measurements

DPI 842 only. To measure a current:
1. Connect the instrument (Figure 5) and, if necessary, adjust the Set Up (Table 1).
2. Select the applicable mA input task from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 4).
Maximum: 30V
Settings Settings
a) with external loop
power
b) with internal loop
power
Figure 5: Example configuration - mA measurement

Operation - Switch test

DPI 842 only. To do tests on a frequency operated switch:
1. Connect the instrument (Figure 6) and, if necessary, adjust the Set Up (Table 1).
2. Select the applicable switch test from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 5). The display shows the switch condition (open or closed) in the top right-hand corner.
3. Supply the output values to the system (Table 7).
• Example - “Nudge” output.
a. Use Edit (
) to set a value less than the switch
value.
b. Use the ▲ ▼ buttons to change the value in small
increments.
• Example - “Ramp” output. a. Set “High” and “Low” values that are applicable to
the switch value (Table 6). Then, to get an accurate switch value, set a long “Travel” period.
b. Use Start/Stop (
) to start and stop the “Ramp”
cycle.
4. If necessary, supply the output values in the opposite direction until the switch changes condition again.
The display shows the applicable values to open and close the switch.
5. To do the test again, press ESC to reset the values.
Edit Settings
Figure 6: Example configuration - Switch test
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
8 - [EN] English K395 Issue 1
FAX 781.665.0780 - TestEquipmentDepot.com
Page 11

Operation - UPM Pressure measurements

Read all the instructions supplied with the UPM and then use the specified procedures to connect it (Figure 7/8).
Settings
a) Pressur e b) Leak test
Start Settings
Figure 7: Example configuration - Pressure measurement
with a UPM
When the connections are complete, make the necessary IDOS selections (Table 2/3). If you re-attach a UPM, the instrument uses the same measurement units that you used before. The instrument keeps a record for the last 10 modules.

UPM - Measure the pressure

To measure the pressure (Figure 7):
1. Select the applicable pressure task from Select Task (Table 2/3) and, if necessary, adjust the Set Up (Table 1), and the Settings (Table 4/5).
2. If necessary, do a zero correction (Table 4).
To measure pressure with another operation (Figure 8), use the same procedure.
Settings
Figure 8: Example configuration - To measure pressure
and frequency

UPM - Leak test

To do a leak test on a pressure system (Figure 7):
1. Select an applicable leak test from Select Task (Table 2/3) and, if necessary, adjust the Set Up (Table 1), and the Settings (Table 4).
2. Set the period for the leak test (Table 4).
3. If necessary, do a zero correction (Table 4).
4. To start the leak test, select Start (
). When the test is
finished, the instrument calculates the leak rate in the applicable units/minute.

Operation - Error indications

If the display shows <<<< or >>>> :
• Make sure that the range is correct.
• Make sure that all the related equipment and connections are serviceable.
[EN] English - 9K395 Issue 1
Page 12

Maintenance

This section gives procedures to maintain the unit in a good condition. Return the instrument to the supplier for all repairs.

Maintenance – Clean the unit

Clean the case with a moist, lint-free cloth and a weak detergent. Do not use solvents or abrasive materials.

Maintenance – Replace the batteries

B1
To replace the batteries, refer to B1. Then re-attach the cover. Make sure that the time and date are correct. The calibration facility uses the date to give service and calibration messages. All the other configuration options stay in memory.

Calibration

Note: GE can provide a calibration service that is traceable to international standards.
We recommend that you return the instrument to the manufacturer or an approved service agent for calibration. If you use an alternative calibration facility, make sure that it uses these standards.

Calibration - Before you start

To do an accurate calibration, you must have:
• the calibration equipment specified in Table 8.
• a stable temperature environment: 70 ± 2°F (21 ± 1°C)
Function Calibration equipment
Hz 1) Frequency meter
Pressure UPM only. Refer to the user manual for the IDOS
mA mA calibrator.
Amplitude (V) 1) Frequency meter
Before you start the calibration, make sure that the time and date on the instrument are correct (Table 1).
Table 8: Calibration equipment
(ppm = parts per million)
Total error: 7 ppm or better Resolution: 8 digits (minimum)
2) Signal generator
UPM.
Accuracy: Refer to Table 11.
Total error: 7 ppm or better Resolution: 8 digits (minimum)
2) Digital volt meter (DVM)
Selection sequence:
Select Task (Table 2) ➤ Set Up (Table 1) ➤ Calibration ➤.
Display: Enter
1
Calibration PIN
Factory PIN = 4321 (Table 9)
23
▲ ▼
+
Menu:
Select Channel
45
▲ ▼
Table 9: Calibration options
Options Description
Hz To calibrate the input or output frequency. ... Amplitu de. To select and calibrate the output
amplitude for the Square Wave and then the Sine/Triangular Wave.
IDOS ... UMM only. To calibrate the specified IDOS UMM.
Refer to the user manual for the IDOS UMM.
mA DPI 842 only. To calibrate the mA input.
Calibration Due: To set the date of the next calibration for the instrument. After the specified calibration date, there is a warning message. There is a selection box to stop the warning. To change the calibration PI N (Personal Identification Number).
When you select a channel, the display shows the applicable instructions to complete the calibration. When the calibration is complete, select Calibration Due and set the new calibration date for the instrument.

Calibration - Procedures (Hz - input/output)

1. Connect the instrument to the calibration equipment (Figure 1).
2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on).
3. Set up the equipment with these conditions:
Frequency meter: Gate time = one second Signal generator: Output = 10V, unipolar,
square wave Frequency = 990 Hz
DPI 841/842: Input units = Hz (Table 4)
Input trigger level = 5V (Table 4)
4. Use the calibration menu (Table 9) to do the calibration. The display shows the applicable instructions to complete the calibration.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
10 - [EN] English K395 Issue 1
FAX 781.665.0780 - TestEquipmentDepot.com
Page 13
5. To make sure that the calibration is correct, set up the equipment to do one of these calibration checks:
• Hz input calibration check (Figure 1):
Frequency meter: Gate time = one second Signal generator: Output = 10V, unipolar,
square wave
DPI 841/842: Input trigger level = 5V (Table 4)
Units (Table 4): Hz or kHz as specified in Table 10.
• Hz output calibration check (Figure 2):
Frequency meter: Gate time = one second DPI 841/842: Units (Table 5): Hz or kHz as
specified in Table 10.
6. Measure or supply the specified values (Table 10): Hz then kHz. Make sure that the error is in the specified limits.
Table 10: Hz error limits (Measure/Supply)
Measure/ Supply
Hz kHz
25 0.002 0.0014 2.5000 0.0002 0.000042 100 0.002 0.0021 10.0000 0.0002 0.000112 250 0.004 0.0035 20.0000 0.0003 0.000205 500 0.006 0.0058 30.0000 0.0004 0.000298 990 0.011 0.0104 50.0000 0.0006 0.000483
Permitted DPI 841/842 error (Hz)
Measure/ Supply
Permitted DPI 841/842 error (kHz)

Calibration - Procedures (mA input)

1. DPI 842 only. Connect the instrument to the calibration equipment (Figure 5).
2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a three-point calibration (-FS, Zero and +FS). The display shows the applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the applicable mA input task (Table 2) and apply these values:
• mA: -55, -40, -24, -18, -12, -6, 0 (short circuit) Then mA: 0, 6, 12, 18, 24, 40, 55.
5. Make sure that the error is in the specified limits (Table 11).
Table 11: mA input error limits
Applied mA
±55 0.0022 0.005 ±40 0.0018 0.004 ±24 0.0014 0.003 ±18 0.0004 0.003 ±12 0.0003 0.002 ±6 0.0002 0.002 0 (Short circuit) - 0.001
Calibrator error (mA)
Permitted DPI 841/842 error (mA)

Calibration - Procedures (Amplitude output)

1. Connect the instrument to the calibration equipment (Figure 2).
2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on).
3. Set up the DPI 841/842 with these conditions:
Output Hz = 0 (For direct current output) Output amplitude: Set to Unipolar (Table 5) Output waveform = Square Wave (Table 5)
4. Use the calibration menu (Table 9) to do the Square Wave calibration. The display shows the applicable instructions to complete the calibration.
Calibration values: Low = 0.1V, High = 20V
5. Change the output waveform to Sine Wave (Table 5).
6. Use the calibration menu (Table 9) to do the Sine/Triangular Wave calibration. The display shows the applicable instructions to complete the calibration.
Calibration values: Low = 0.1V, High = 20V
[EN] English - 11K395 Issue 1
Page 14
7. To make sure that the calibration is correct, set up the DPI 841/842 with these conditions:
Output Hz = 0 (For direct current output) Output amplitude: Set to Unipolar (Table 5)
8. Supply the specified values (Table 12): Square Wave then Sine Wave. Make sure that the error is in the specified limits.
Table 12: Amplitude output error limits
Amplitude Volts (V)
0.1 0.05
1.0 0.10
10.0 0.10
15.0 0.15
24.0 0.24
Permitted DPI 841/842 error (V)

Calibration - Procedures (IDOS UMM)

Refer to the user manual for the IDOS UMM. When the calibration is complete, the instrument automatically sets a new calibration date in the UMM.

Specification data

All accuracy statements are for one year.

Specification - General

Languages English [Default] Operating temperature Storage temperature Humidity 0 to 90% without condensation
Shock/Vibration BS EN 61010:2001; Def Stan 66-31, 8.4 cat III EMC BS EN 61326-1:1998 + A2:2001 Safety Electrical - BS EN 61010:2001; CE Marked Size (L: W: H) 7.1 x 3.3 x 2.0 in
Weight 14 oz (400 g) Power supply 3 x AA alkaline batteries Duration (Measure) Duration (Supply)
14 … 122°F (-10 … 50°C)
-4 … 158°F (-20 … 70°C)
(Def Stan 66-31, 8.6 cat III)
(180 x 85 x 50 mm)
Hz, pulses: ≈ 60 hours mA: 35 hours Hz, pulses: ≈ 20 hours mA: 10 hours (24 V Sour ce at 12 mA)

Specification - Frequency

cpm = Pulses/minute, cph = Pulses/hour
Range (Measure): Accuracy:
0 ... 999.999 Hz 1 ... 50.0000 kHz cpm: 0 ... 99999 cph: 0 ... 99999
Range (Supply): Accuracy:
0 ... 999.99 Hz 1 ... 50.000 kHz cpm: 0 ... 99999 cph: 0 ... 99999
Temperature coefficient 14 … 50°F, 86 … 122°F (-10 … 10°C, 30 … 50°C)
Waveform (Sine, square, triangle)
Voltage input 0 ... 30 V Trigger level 0 ... 24 V, Resolution: 0.1 V Output amplitude 0 ... 24 V dc ± 1%
Connectors (A1 - Item 10) Four 0.16 in (4 mm) sockets
For all the ranges:
0.003% of reading + 2 counts
0.003% of reading + 0.0023 Hz
0.003% of reading + 0.0336 Hz
0.003% of reading + 0.138 cpm
0.003% of reading + 0.5 cph
0.0011% FS / °F (0.002% FS / °C)
Unipolar Bipolar
0 ... 24 V ac ± 5% (Current ≤ 20 mA)

Specification - Electrical connectors (A2)

Range (Measure) 0 to ±55 mA Accuracy 0.02% of reading + 3 counts Temperature coefficient 14 … 50°F, 86 … 122°F (-10 … 10°C, 30 … 50°C) Switch detection Open a nd closed. 2 mA current. Loop power output 24 V ± 10%
HART® resistor Connectors (A2) Three 0.16 in (4 mm) sockets
0.0011% FS / °F (0.002% FS / °C)
250 (menu selec tion)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
12 - [EN] English K395 Issue 1
FAX 781.665.0780 - TestEquipmentDepot.com
Loading...