GE Druck DPI 800, Druck DPI 802 User Manual

Page 1
GE
Advanced Test Equipment Rentals
www.atecorp.com 800-404-ATEC (2832)
E
s
t
a
b
l
i
s
h
e
d
1
9
8
1
Sensing
Druck DPI 800/802
Pressure indicator and Pressure loop calibrator
User manual - K335
Page 2
A1
10c
B1
A2
10 a, b, c
1
2
3
4
A 2
A 3
9
8
7
6
5
12
DPI 802/802P
11
B1
1513
10a
10b
K335 Issue 3
14
A3
17182519 2616
20
21
22 23
24
Page 3

Table of Contents

Introduction ......................................................................... 1
Safety ..................................................................................... 2
Marks and symbols on the instrument ................................. 2
To start .................................................................................. 2
Location of items ............................................................................ 2
Items on the display ...................................................................... 3
Prepare the instrument ................................................................ 3
Power on or off ................................................................................ 3
Set up the basic operation ......................................................... 3
Select a task (Measure) ................................................................ 4
Set up the settings ......................................................................... 4
Edit functions .................................................................................... 5
Operation .............................................................................. 5
Pressure connections ................................................................... 5
Communications port connections ........................................ 6
Measure pressure ........................................................................... 6
Leak test .............................................................................................. 6
Switch test .......................................................................................... 7
Transmitter calibration ................................................................ 7
mA measurements ........................................................................ 8
UPM Pressure measurements .................................................. 8
Error indications .............................................................................. 8
Maintenance ........................................................................ 9
Clean the unit ................................................................................... 9
Replace the batteries .................................................................... 9
Calibration ............................................................................ 9
Before you start ............................................................................... 9
Procedures (Pressure P1/P2) .................................................. 10
Procedures (mA input) ............................................................... 10
Procedures (IDOS UMM) ............................................................ 10
Specification data ............................................................ 11
General ............................................................................................. 11
Pressure measurement ............................................................ 11
Electrical connectors (A2) ......................................................... 12
Customer service ............................................... Back cover

Introduction

The DPI 800 Pressure Indicator and DPI 802 Pressure Loop Calibrator are part of the Druck DPI 800 series of hand held instruments. The DPI 800 series uses Intelligent Digital Output Sensor (IDOS) technology to give instant plug and play functionality with a range of Universal Measurement Modules (UMM). Example: the Universal Pressure Module (UPM). The DPI 800/802 include these functions:
Function DPI 800 DPI 802
Measure pressure *
Measure pressure difference *** Leak Test Yes Communications port IDOS or RS232
Language selection Yes Pressure u nits 25 (Fixed ) Snapshot ** Up to 1000 displays with a date/time stamp Measure mA No 0 - 55 mA HART® resistor No Yes V dc output No 24 V Switch test No Yes Other functions Hold, Maximum/Minimum/Average, Filter,
* Refer to “Specification data”. ** Optional ite m *** High range IDOS - Low range IDOS (if applicable)
Internal IDOS: P1 and/or P2**;
External IDOS: UPM**
Difference (∆) between: P1 an d P2, or P1/P2
and a UPM
Tare, Scaled values, Backlight, Alarm
© 2005 General Electric Company. All rights reserved.
Trademarks
All product names are trademarks of their respective companies.
K335 Issue 3
[EN] English - 1
Page 4

Safety

Before you use the instrument, make sure that you read and understand all the related data. This includes: all local safety procedures, the instructions for the UMM (if applicable), and this publication.
WARNING
• Some liquid and gas mixtures are dangerous. This includes mixtures that occur because of contamination. Make sure that the instrument is safe to use with the necessary media.
• It is dangerous to ignore the specified limits for the instrument or to use the instrument when it is not in its normal condition. Use the applicable protection and obey all safety precautions.
• To prevent a dangerous release of pressure, isolate and bleed the system before you disconnect a pressure connection.
• Do not u se the instrumen t in locations w ith explosive gas, vapor or dust. There is a risk of an explosion.
• DPI 802/802P only. To prevent electrical shocks or damage to the instrument, do not connect more than 30V between the terminals, or between the terminals and the ground (earth).
Before you start an operation or procedure in this publication, make sure that you have the necessary skills (if necessary, with qualifications from an approved training establishment). Follow good engineering practice at all times.

Safety - Marks and symbols on the instrument

Complies with European Union directives
Read the manual Battery
Ground (Earth) P1/P2 Identifies the pressure
ON/OFF
MWP
The maximum working pressure for the connector. Example: 2 x FS (FS = Full scale pressure)
REFERENCE
The connector is a reference port for r eference pressure only. Refer to “Specification data”.
The identification on e ach pressure connector includes the type of thread (1/8 NPT, G1/8 ... ).
Warning - refer to the manual
range connector

To start

To start - Location of items

Item Description
1. On or off button.
2.
3.
4.
5.
6.
7.
8. Display. Refer to A3
9.
10. 10a. IDOS pressure connector with identification marks.
11. Connection point for some of the optional
12. Battery compartment. Refer to B1.
13., 14., 15. DPI 802/802P only. Terminals to measure current, to
Left-hand soft-key. Selects the function above it on
■ ■
the display (Item 24). Example: Edit Moves back one menu level.
ESC
Leaves a menu option. Cancels the changes to a value.
Increases or decreases a value.
▲ ▼
Highlights a different item. Holds the data on the display. To continue, press the
HOLD
HOLD button again. Shows the Select Task menu.
MENU
Selects or accepts an item or value.
OK
Selects [] or cancels [ ] a selection Right-hand soft-key. Selects the function above it on
■ ■
the display (Item 24). Example: Settings
Communications port. Use to connect a Universal
SENSOR/
PC
Measurement Module (UMM) or a RS232 cable.
Refer to “Safety”.
10b. Label to identify the pressure range for the
associated IDOS pressure connector (P1/P2): Sealed gage (sg), gage (g), absolute (a).
10c. PTFE vent filter or a reference port connector. Refer to
“Specification data”.
accessories. Refer to the datasheet.
supply 24V source, and to do switch tests.
A1 A2
2 - [EN] English
K335 Issue 3
Page 5

To start - Items on the display

Item Description
16. DPI 802/802P only. Task indication for the switch test.
17. DPI 802/802P only. The loop power supply is on.
18. The measured value satisfies one of the alarm
19. The data on the display is on hold. To continue, press
20.
21. Identifies the type of data and the measurement
22. ... 23. Identifies the settings applied to the input.
22. xxxx g The units and the type of sensor (If applicable)
23. Shows the settings applied to the measured value (If
24. A soft-key function. To select an available function,
25. The measured value or values applicable to the task
26. The Edit display to set up text labels ( 6 characters):
= switch closed
Task indication for the leak test.
Refer to: Select Task (Table 2/3)
Refer to: Select Task (Table 2/3)
conditions. Refer to: Settings (Table 4)
the HOLD button again.
Shows the battery level: 0 ... 100%.
range.
= Input = IDOS input
Refer to: Select Task (Table 2/3)
Refer to: Settings (Table 4)
OR a specified scale (x:y).
applicable).
= Filter
= Tare
press the soft-key below it. Example:
= Move left = Move right
selection.
x:y Scaling (Table 4).
OK = Accept the new text label Shift = Change the keys: 123ABC or -_+abc
= Add a space
BS = Back space (Delete character)
A3
= switch open
= Maximum = Average = Minimum

To start - Prepare the instrument

Before you use the instrument for the first time:
• Make sure that there is no damage to the instrument, and that there are no missing items.
• Remove the plastic film that protects the display. Use the tag (◗ ) in the top right-hand corner.
• Install the batteries (refer to B1). Then re-attach the cover.

To start - Power on or off

To turn the instrument on or off, press ❍ (A1 - item [1]). The instrument does a self test and then shows the applicable data. When the power is off, the last set of configuration options stays in memory. Refer to “Maintenance”.

To start - Set up the basic operation

Use the Set Up menu to set up the basic operation of the instrument.
Menu:
1
Select Tas k
(Table 2) (Table 1) []/[ ]
23
▲ ▼
Menu: Set Up
45
▲ ▼
If there is additional data for a menu option, select Settings (■■) to see the values that are set up. If necessary, adjust the values.
Table 1: Menu options - Set Up
Options
(If applicable)
Descript ion
DPI 802/802P only. To add a series resistor into the mA circuit. You can then use this instrument together with a HART® communicator to set up and calibrate HA RT® devices.
To select and set up the backlight facility + timer. Additional data: Select Settings (■ ■) To select and set up the power off facility + timer. Additional data: Select Settings ( )
To show the battery level (%).
To set the display contrast (%). Increases %, decreases % To set the time + date. The calibration facility uses the date to give service and calibration messages. To set the language option.
To calibrate the instrument. Additional data: Refer to “Calibration”. To select and show the applicable status data (Software Build, Calibration Due date, Serial Number, IDOS Information).
K335 Issue 3
[EN] English - 3
Page 6

To start - Select a task (Measure)

When the instrument is set up (Table 1), use the Select Task menu to select the applicable task.
1
Menu:
Select Task
23
Display:
Task 1xxx g

To start - Set up the settings

When the task is set up (Table 2/3), use the Settings menu to adjust the measurement operation.
Display: Task
P1 + mA
Settings se lection
1
(If applicable)
23
▲ ▼
(Table 2/3) Sk2 = Settings
In Table 2/3, P1 and P2 are internal sensors, and IDOS is a Universal Measurement Module (UMM). If you attach a UMM to the communicatio ns port (A1 - item [9]), the Select Task menu shows the applicable IDOS options.
Table 2: Menu options - Select Task
Options
(If applicable)
P... An input measurement task (P1, P2*). The task
P... An input measurement task to measure the
mA DPI 802/802P only. A mA measurement task. mA(24V) DPI 802/802P only. A mA measurement task + the
IDOS UMM only. An IDOS measurement task.
Pressure options only. A leak test.
Description
(* Optional item)
includes the pressure range for the sensor.
pressure difference () between: P1 and P2*, or P1/P2* and a UPM*. P = High range IDOS - Low range IDOS
loop power supply is on. DPI 802/802P only. A switch test.
To set up the way the instrument works. Additional data: Refer to: Set Up (Table 1).
Table 3 shows all the one and two function operations that are available. If you attach a UMM, you can only use the options that include IDOS.
Table 3: Permitted 1 and 2 function operations
Function
P1 P2
(1) (1) (1) (1) mA (1) (2) (2) (2) (2) mA(24V) (1) (2) (2) (2) (2)
x (2) (2) (2) (2 ) x(2) (2) x (2)
IDOS (1) (2) (2) x x P2(1)(2)xx(2)
= DPI 802/802P only
P...
(Table 2) IDOS
■ ■
Sk2 = Settings
Menu:
Settings
(Table 4) []/[ ] Sk1 = Edit
456
■ ■
▲ ▼
Settings
x:y
If there is additional data for a menu option, select Settings (■■) to see the values that are set up. If necessary, adjust the values. Refer to “Edit functions”
Table 4: (Part of table) Menu options - Settings
Options
(If applicable)
Description
... Units To select one of the fixed units of measurement
(psi, mbar ... ). Gage sensors or sensors with differential
operation (P1, P2, IDOS). A zero correction that makes the instrument read zero at local pressure. To include maximum, minimum and average values for the measurement task. To select and set up a tare value for the measurement task (a specified value or the reading on the display). The instrument subtracts a positive tare value, and adds a negative tare value.
Additional data: Select Settings ( ) To select and set up a scale of values: One local scale for each measurement task (Maximum: 5). Additional data (Example 1/2): Select Settings (■ ■)
4 - [EN] English
K335 Issue 3
Page 7
Table 4: (Part of table) Menu options - Settings
Options
(If applicable)
Description
To select and set up the filter values to give a smoother output for the measurement task:
Band as a % of full scale (FS). The filter compares each new value with the previous value. If the new value is outside the band, it is not filtered. Low pass filter time constant in seconds. Increase the value to increase the damping factor.
Additional data: Select Settings ( ) To select and set up the alarm values for the measurement task (maximum and minimum). Additional data: Select Settings ( ) Leak Test only. To set an applicable period for the leak test (Hours:Minutes:Seconds).

To start - Edit functions

Example 1) Set up a label for x:y Scaling = %.
Settings
x:y
Sk1 = Edit A3 - item [26]
1 Edit 2 Edit 3
■ ■
Edit ... Edit 45
■ ■
▲ ▼
Shift

Operation

This section gives examples of how to connect and use the instrument. Before you start:
• Read and understand the “Safety” section.
• Do not use a damaged instrument.

Operation - Pressure connections

CAUTION: To prevent damage, do not apply torque to the body of the instrument. If available, use the flat faces on the pressure connector to hold the instrument in position.
Use an applicable method to seal the pressure connections, and then tighten to the applicable torque (Figure 1 and Table 5).
Note: If the instrument uses G1/8 connections, and the pressure is ≥1500 psi (100 bar), connection method (b) is mandatory.
a) 1/8 NPT
connections
12 1
3
c) G1/8 connections:
Pressure < 1500 psi (100 bar)
12
b) G1/8 connections:
Pressure ≥ 1500 psi (100 bar)
2
4
d) M5 or 10-32 UNF
connections:
Refer to “Table 10”
2
5
%_
A3 - item [26]
OK
■ ■
Example 2) Set up values for x:y Scaling = 0 to 100%.
Settings
x:y
Sk1 = Edit
Settings
1
▲ ▼
= Flow scaling (mA, pressure only)
2 Edit 34
x:y
■ ■
■ ■
K335 Issue 3
4
Figure 1: Connection methods
Table 5: Key to figure 1
Item Des cription
1 Process connector. Max imum torque:
1/8 NPT: 26 lbf.ft (35 Nm)
▲ ▼
G1/8: 18.4 lbf.ft (25 Nm) 2 Applicable instrument pressure connector (Table 10) 3 (1/8 NPT only) Thread with an applicable sealant 4 (G1/8 only) Applicable bonded seal 5 Reference port connector: M5 or 10-32 UNF (Table 10)
Maximum torque: 1.5 lbf.ft (2 Nm)
[EN] English - 5
Page 8

Operation - Communications port connections

Use the communications port (A1 - item [9]) to attach an IDOS Universal Measurement Module (UMM). When you attach the cable from a UMM (Figure 8), the instrument automatically changes the menus to give you all the applicable options (Table 2/3).

Operation - Measure pressure

To measure the pressure or the pressure difference (if there are two IDOS pressure ranges).
1. Connect the instrument (Figure 2) and, if necessary, adjust the Set Up (Table 1).
2. Select an applicable pressure task from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 4).
3. If necessary, do a zero correction (Table 4).
4. Before you disconnect the instrument, isolate and bleed the system.
p1
p1
p2
p2
P2P1
Settin

Operation - Leak test

To do a leak test on a pressure system:
1. Connect the instrument (Figure 3) and, if necessary, adjust the Set Up (Table 1).
2. Select an applicable leak test from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 4).
3. Set the period for the leak test (Table 4).
4. If necessary, do a zero correction (Table 4).
5. To start the leak test, select Start (
). When the test is
finished, the instrument calculates the leak rate in the applicable units/minute.
6. Before you disconnect the instrument, isolate and bleed the system.
Start Settings
Figure 3: Example configuration - Leak test
Settings Settings
a) P1 and/
or P2
b) P1 - P2
Figure 2: Example configuration - Measure pressure
The examples in Figure 2 show:
• how to measure one or two pressures in a system
• how to measure pressure difference
Note:
P = High range IDOS - Low range IDOS. If
Pressure (p2) > Pressure (p1), the result is negative.
6 - [EN] English
K335 Issue 3
Page 9

Operation - Switch test

DPI 802/802P only. To do tests on a pressure switch:
1. Connect the instrument (Figure 4) and, if necessary, adjust the Set Up (Table 1).
2. Select the applicable switch test from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 4). The display shows the switch condition (open or closed) in the top right-hand corner.
3. If necessary, do a zero correction (Table 4).
4. Slowly apply pressure to the system until the switch changes condition (open or closed).
5. Slowly release the pressure until the switch changes condition again. The display shows the pressure values to open and close the switch.
6. To do the test again, press ESC to reset the values.
7. Before you disconnect the instrument, isolate and bleed the system.
4. For each calibration point, apply an applicable pressure and wait for the pressure system to stabilize.
5. Before you disconnect the instrument, isolate and bleed the system.
Maximum: 30V
Settings
Figure 5: Example configuration - Transmitter calibration
with external loop power
Settings
Figure 4: Example configuration - Switch test

Operation - Transmitter calibration

DPI 802/802P only. To calibrate a transmitter:
1. Connect the instrument (Figure 5 or Figure 6) and, if necessary, adjust the Set Up (Table 1).
2. Select the applicable calibration task from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 4).
3. If necessary, do a zero correction (Table 4).
K335 Issue 3
Settings
Figure 6: Example configuration - Transmitter calibration
with internal loop power
[EN] English - 7
Page 10

Operation - mA measurements

DPI 802/802P only. To measure a current:
1. Connect the instrument (Figure 7) and, if necessary, adjust the Set Up (Table 1).
2. Select the applicable mA task from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 4).
Maximum: 30V

Operation - UPM Pressure measurements

Read all the instructions supplied with the UPM and then use the specified procedures to connect it (Figure 8).
Settings Settings
a) with external loop
power
b) with internal loop
power
Figure 7: Example configuration - mA measurement
Settings
a) Pressur e b) Leak test
Start Settings
Figure 8: Example configuration - Pressure measurement
with a UPM
When the connections are complete, make the necessary IDOS selections (Table 2/3). Then use the same procedures as specified for the P1/P2 connections. If you re-attach a UPM, the instrument uses the same measurement units that you used before. The instrument keeps a record for the last 10 modules.

Operation - Error indications

If the display shows <<<< or >>>> :
• Make sure that the range is correct.
• Make sure that all the related equipment and connections are serviceable.
8 - [EN] English
K335 Issue 3
Page 11

Maintenance

This section gives procedures to maintain the unit in a good condition. You must return the unit to the supplier for all repairs.

Maintenance - Clean the unit

Clean the case with a moist, lint-free cloth and a weak detergent. Do not use solvents or abrasive materials.

Maintenance - Replace the batteries

B1
To replace the batteries, refer to B1. Then re-attach the cover. Make sure that the time and date are correct. The calibration facility uses the date to give service and calibration messages. All the other configuration options stay in memory.

Calibration

Note: GE can provide a calibration service that is traceable to international standards.
We recommend that you return the instrument to the manufacturer or an approved service agent for calibration. If you use an alternative calibration facility, make sure that it uses these standards.

Calibration - Before you start

To do an accurate calibration, you must have:
• the calibration equipment specified in Table 6.
• a stable temperature environment: 70 ± 2°F (21 ± 1°C)
Table 6: Calibration equipment
Function Calibration equipment
Pressure An applicable pressure standard
mA mA calibrator.
Before you start the calibration, make sure that the time and date on the instrument are correct (Table 1).
(primary or secondary) with a total uncertainty of 0.01% reading or better.
Accuracy: Refer to Table 9.
Selection sequence:
Select Task (Table 2) ➤ Set Up (Table 1) ➤ Calibration ➤
Display: Enter
1
Calibration PIN
Factory PIN = 4321 (Table 7)
23
▲ ▼
+
Menu:
Select Channel
45
▲ ▼
Table 7: Calibration options
Options
(If applicable)
Description
(* Optional item)
P... To calibrate the specified IDOS pressure input
(P1 or P2*).
IDOS ... UMM only. To calibrate the specified IDOS UMM.
Refer to the user manual for the IDOS UMM.
mA DPI 802/802P only. To calibrate the mA input.
Calibration Due: To set the date of the next calibration for the instrument. After the specified calibration date, there is a warning message. There is a selection box to stop the warning.
To change the c alibration PIN (Personal Identification Number).
When you select a channel, the display shows the applicable instructions to complete the calibration. When the calibration is complete, select Calibration Due and set the new calibration date for the instrument.
K335 Issue 3
[EN] English - 9
Page 12

Calibration - Procedures (Pressure P1/P2)

1. Connect the instrument to the pressure standard (Figure 2).
2. Let the equipment get to a stable temperature (minimum: 30 minutes since the last power on).
3. Use the calibration menu (Table 7) to do a two-point calibration (Zero and +FS) or a three-point calibration (-FS, Zero and +FS). Refer to Table 8. The display shows the applicable instructions to complete the calibration.
Table 8: Calibration pressures
Ranges: g/d Nominal applied pressure
10.0 psi (700 mbar) -FS 0 +FS > 10.0 psi (700 mbar) -13.1 (-900) 0 +FS
† For a three-point calibration, do not apply more than -90% of
the specified FS for the unit.
Ranges: a Nominal applied pressure
5.00 psi (350 mbar) < 0.02 (1.0) +FS
30.0 psi (2 bar) < 0.07 (5.0) +FS
100.0 psi ( 7 bar) < 0.29 (20.0) +FS
300.0 psi (20 bar) < 0.73 (50.0) +FS
Ranges: sg Nominal applied pressure
5000 psi (350 bar) Use atmospheric
psi (mbar)
-FS † Zero +FS
psi (mbar)
Zero +FS
Zero +FS
+FS
pressure as zero.
4. To make sure that the calibration is correct, select the applicable pressure task (Table 2) and apply these pressure values:
• Ranges g/d or sg: 0, 20, 40, 60, 80, 100 (%FS) Then: Go back to 0 in the same steps. Then (three-point calibration only):
-20, -40, -60, -80, -100 (%FS) Then: Go back to 0 in the same steps.
• Ranges a: 0, 20, 40, 60, 80, 100 (%FS) Then: Go back to 0 in the same steps.
5. Make sure that the error is in the specified limits:
• Standard accuracy: The specified accuracy (refer to “Specification data”) includes an allowance for temperature changes, reading stability for one year, and the uncertainty of the standard used for calibration. In step 4, make sure that the error between the applied pressure and the reading on the unit is not more than
0.015% FS.
• Premier precision: The specified precision (refer to “Specification data”) includes an allowance for temperature changes and the uncertainty of the standard used for calibration. In step 4, make sure that the error between the applied pressure and the reading on the unit is not more than the specified value for Premier precision.

Calibration - Procedures (mA input)

1. DPI 802/802P only. Connect the instrument to the calibration equipment (Figure 7).
2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 7) to do a three-point calibration (-FS, Zero and +FS). The display shows the applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the applicable mA task (Table 2) and apply these values:
• mA: -55, -40, -24, -18, -12, -6, 0 (short circuit) Then: 0, 6, 12, 18, 24, 40, 55.
5. Make sure that the error is in the specified limits (Table 9).
Table 9: mA input error limits
Applied mA
±55 0.0022 0.005 ±40 0.0018 0.004 ±24 0.0014 0.003 ±18 0.0004 0.003 ±12 0.0003 0.002 ±6 0.0002 0.002 0 (Short circuit) - 0.001
Calibrator error (mA)
Permitted DPI 800/802 error (mA)

Calibration - Procedures (IDOS UMM)

Refer to the user manual for the IDOS UMM. When the calibration is complete, the instrument automatically sets a new calibration date in the UMM.
10 - [EN] English
K335 Issue 3
Page 13

Specification data

All accuracy statements are for one year.

Specification - General

Languages English [De fault] Operating temperature Storage temperature Humidity 0 to 90% without condensation
Shock/Vibration BS EN 61010:2001; Def Stan 66-31, 8.4 cat III
14 … 122°F (-10 … 50°C)
-4 … 158°F (-20 … 70°C)
(Def Stan 66-31, 8.6 cat III)

Specification - Pressure measurement

The % full scale (FS) statements for Standard accuracy and Premier precision are only applicable if there is a regular zero correction by the IDOS instrument.
Ranges: Gage and differential operation (g/d), sealed gage (sg), absolute (a)
+/- psi: 0.36 (+/- mbar: 25) g/d 0.1 0.03 1/2 +/- psi: 1, 3, 5, 10 (+/- mbar: 70, 200, 350, 700) g/d 0.075 0.03 1/2 psi: -15 to [15 or 30] (bar: -1 to [1 or 2]) psi: -15 to [50, 100, 150, or 300] (bar: -1 to [3.5, 7, 10, or 20]) psi: 500, 1000, 1500, 2000, 3000 (bar: 35, 70, 100, 135, 200) psi: 5 (mbar : 350) a 0.1 - 2 psi: 30 (bar: 2) a 0.075 - 2 psi: 100, 300 (bar: 7, 20) a 0.075 - 3 psi: 5000, 10000 (bar: 350, 700) sg 0.05 - 3
* Standard accuracy is for 32 … 122°F (0 … 50°C);
Stability: 1 year
† Premier precision is for 65 … 82°F (18 … 28°C);
Stability: ≤ 10 psi (700 mbar) = 0.02% of reading/Year Stability: > 10 psi (700 mbar) = 0.01% of reading/Year Premier precision for 41 … 113°F (5 … 45°C):
10 psi (700 mbar): 0.075% FS
>
10 psi (700 mbar): 0.014% FS
Type Standard Accuracy *
g/d 0.05 0.01 1/2
g/d 0.05 0.01 1/3
g/d 0.05 0.01 1/3
EMC BS EN 61326-1:1998 + A2:2001 Safety Electrical - BS EN 61010:2001;
Size (L: W: H) Ma ximum: 7.5 x 3.3 x 2.0 in
Weight 19 oz (530 g) Power supply 3 x A A alkaline batteries Duration
(Measure)
% FS
Notes:
1. Reference port media: Non-corrosive, dry gas
2. + port media: Non-corrosive, non-conductive liquid or Non-cor rosive, dr y gas
3. + port media: Media applicable to stainless steel
Pressure Equipment Directive - Class: Sound Engineering Practice (SEP); CE Marked
(190 x 85 x 50 mm)
P1: ≈ 50 hours P1 + mA: 5 hours (24 V Source at 12 mA)
Premier Precision †
% FS
Notes
K335 Issue 3
[EN] English - 11
Page 14
Table 10: DPI 800 series pressure connections
Ranges Pressure connections
g/d: 30 psi g (2 bar g) Standard or Premier
g/d: > 30 psi g (2 bar g) Standard only g/d: > 30 psi g (2 bar g) Premier only
sg or a: all ranges G1/8 female (+ port) OR
(One IDOS range) 1/8 NPT female (+ port) + 1/8 NPT female reference port OR G1/8 female (+ port) + G1/8 female reference port 1/8 NPT female (+ port) OR G1/8 female (+ port) 1/8 NPT female (+ port) + 10-32 UNF reference port OR G1/8 female (+ port) + M5 reference port
1/8 NPT female (+ port)
Table 11: Maximum pressure (+ port)
Ranges: g/d, sg, a MWP Maximum transient /
5 psi (350 mbar) 2 x FS 4 x FS > 5 psi (350 mbar) 1.2 x FS 2 x FS
intermittent pressure
Table 12: Maximum pressure (Reference port)
Ranges: g/d only MWP
5 psi (350 mbar) 2 x FS 10 to 15 psi
(700 mbar to 1 bar) 30 psi (2 bar) 30 psi (2 bar)
1.2 x FS

Specification - Electrical connectors (A2)

Range (Measure) 0 to ±55 mA Accuracy 0.02% of reading + 3 counts Temperature coe fficient 14 … 50°F, 86 … 122°F (-10 … 10 °C, 30 … 50°C) Switch detection Open and closed. 2 mA current. Loop power output 24 V ± 10%
HART® resistor Connectors (A2) Three 0.16 in (4 mm) sockets
0.0011% FS / °F (0.002% FS / °C)
250 (menu selec tion)
Pressure connections
(Two IDOS ranges) Two 1/8 NPT female (+ port) + 10-32 UNF reference port OR Two G1/8 female (+ port) + M5 reference port Two 1/8 NPT female (+ port) OR Two G1/8 female (+ port) Two 1/8 NPT female (+ port) + 10-32 UNF reference port OR Two G1/8 female (+ port) + M5 reference port Two 1/8 NPT female (+ port) OR Two G1/8 female (+ port)
12 - [EN] English
K335 Issue 3
Page 15

Customer Service

Visit our web site: www.gesensing.com
Loading...