•This technical manual provides operating instructions for the Druck
DBC Series of Dry Well Calibrators.
Safety
•The manufacturer has designed this equipment to be safe when
operated using the procedures detailed in this manual. Do not use this
equipment for any other purpose than that stated.
•This publication contains operating and safety instructions that must be
followed to ensure safe operation and to maintain the equipment in a
safe condition. The safety instructions are either warnings or cautions
issued to protect the user and the equipment from injury or damage.
•Use suitably qualified * technicians and good engineering practice for
all procedures in this publication.
Test leads
•Only use the test leads supplied with the instrument, do not use these
test leads with any other equipment.
Power isolation
•Access to the power supply isolator switch must not be obstructed.
Toxic Materials
•There are no known toxic materials used in construction of this
equipment.
Maintenance
•The equipment must be maintained using the procedures in this
publication. Further manufacturer’s procedures should be carried out
by authorized service agents or the manufacturer’s service
departments.
Technical Advice
•For technical advice contact the manufacturer.
*A qualified technician must have the necessary technical knowledge,
documentation, special test equipment and tools to carry out the required
work on this equipment.
This equipment meets the requirements of all relevant
European safety directives. The equipment carries the CE
mark.
K400 Issue No. 1i
TRACEABILITY AND CALIBRATION REPORT
•Each DBC instrument is supplied as standard with a calibration report,
declaring traceability to a National or International standard.
WARRANTY AND LIABILITY
1Our equipment is guaranteed against defective material and
workmanship during the warranty period. Claims under warranty
can be made by returning the equipment, prepaid, to our factory. The
equipment will be replaced, repaired or adjusted at our discretion.
2The liability of GE is restricted to that given under our warranty. No
responsibility is accepted for damage, loss or other expense
incurred through the sale or use of our equipment.
3 Under no circumstances will GE be liable for any special incidental
or consequential damage.
Symbols
The following symbols are used to identify hazards on the instrument:
This symbol, on the instrument, indicates that the user
should refer to the user manual.
This symbol, on the instrument, indicates a very hot
surface that must not be touched.
K400 Issue No. 1 ii
Safety Instructions
This
instrument
During operation observe the following safety instructions:
1USE the instrument in vertical position only.
2DO NOT place any object on the protection grid of the dry well.
3DO NOT put any liquid into the dry well and/or on the inserts.
4DO NOT use fluids to clean the dry well and/or the inserts.
5BEFORE any maintenance activity, WAIT for the instrument to cool
to ambient temperature (keep the instrument switched on, to allow
the cooling fan to operate).
6DO NOT touch the protection grid until it has cooled to ambient
temperature).
7The instrument must be at ambient temperature before storing in any
transit case or packing system.
8DO NOT use this instrument near inflammable liquids or in
Hazardous areas.
9DO NOT place either your hands, or any other object, into the dry
well.
10Before use, REMOVE all traces of contamination from the device
under test.
11During normal operation, this instrument could reach high
temperatures.
12Always observe all the safety instructions.
*A qualified technician must have the necessary technical knowledge,
documentation, special test equipment and tools to carry out the required
work on this equipment.
should only be used by suitably qualified * technicians
Abbreviations
The following abbreviations are used in this manual; the
abbreviations are the same in the singular and plural.
The DBC-TC series is supplied with the following standard accessories:
•Operating manual
• Set of leads for:
2 test leads and 2 clamps for switch connection
6 test leads and 6 clamps for connection of electrical signal
• Power supply cable
•Removable insert with 3 holes (5, 6.6, 9.8 mm diameter), (3/16, 1/4,
3/8 in. diameter for USA)
• Tool for extraction of removable insert
• RS232 interface cable (1.5 metre, 5 pin/5 pin)
• Traceable calibration certificate
• 2 spare fuses (see page 4 for more information)
OPTIONAL ACCESSORIES DBC-TC series
•Option B1 for DBC 150, reference probe type PT 100, range –50°C
to +400°C
• Option B2 for DBC 650, reference probe type PT 100, range +50°C
to +650°C
• Option B3, temperature probe type PT 100, range –50°C to +650°C.
• Option D, fast cooling device
• Option E, aluminium transit case
• Option C1/C2, insert with 2 holes (5 and 13 mm diameter), (3/16,
1/2 in. diameter for USA)
• Option C3/C4, insert with 4 holes (3.4, 5, 5, 8.2 mm diameter), (1/8,
3/16, 3/16 and 5/16 in. diameter for USA)
• Option C5/C6, blank insert
• Option C7/C8, insert with special holes.
K400 Issue No. 1vii
STANDARD ACCESSORIES DBC-TS series
The DBC-TC series is supplied with the following standard accessories:
•Operating manual
• Set of leads for:
2 test leads and 2 clamps for switch connection
• Power supply cable
•Removable insert with 3 holes (5, 6.6, 9.8 mm diameter), (3/16, 1/4,
3/8 in. diameter for USA)
• Tool for extraction of removable insert
• RS232 interface cable (1.5 metre, 5 pin/5 pin)
• Traceable calibration certificate
• 2 spare fuses (see page 4 for more information)
OPTIONAL ACCESSORIES DBC-TS series
•Option B1 for DBC 150, reference probe type PT 100, range –50°C
to +400°C
• Option B2 for DBC 650, reference probe type PT 100, range +50°C
to +650°C
• Option B3, temperature probe type PT 100, range –50°C to +650°C.
• Option D, fast cooling device
• Option E, aluminium transit case
• Option C1/C2, insert with 2 holes (5 and 13 mm diameter), (3/16,
1/2 in. diameter for USA)
• Option C3/C4, insert with 4 holes (3.4, 5, 5, 8.2 mm diameter), (1/8,
3/16, 3/16 and 5/16 in. diameter for USA)
• Option C5/C6, blank insert
• Option C7/C8, insert with special holes.
K400 Issue No. 1 viii
1INTRODUCTION
The DBC series of temperature calibrators test and calibrate temperature devices.
The series comprises two types: a TS (Temperature Source) and a TC
(Temperature Calibrator). Multilingual firmware provides a menu-driven user
interface in English, French, German, Italian, Spanish and Portuguese. An RS232
digital interface, supplied as standard, enables remote operation of the instrument
from a compatible computer.
TS (Temperature Source)
The DBC-TS generates stable temperatures and accurately measures the
temperature of the dry well. It provides an automatic switch test facility.
There are two models of the DBC-TS, providing the same operating capabilities at
different temperature ranges:
•DBC 150-TS : -45°C below ambient temperature to +150°C.
•DBC 650-TS : +50°C to +650°C.
This series of calibrators uses the GE technology temperature controller, providing
high performance temperature stability and accuracy. Temperature generation can
be controlled either in manual mode (direct) or automatic with ramping, stepping or
preset functions. The dry well temperature can be measured either with the internal
probe or with an external certified reference probe. The key-pad, display and
electrical connections enable the user to carry out test procedures for RTD and
thermal switches with a hysteresis calculation.
TC (Temperature Calibrator)
The DBC-TC generates stable temperatures and accurately measures the
temperature of the dry well. It measures process signals from the temperature
devices under test.
There are two models of this series, providing the same operating capabilities at
different temperature ranges:
•DBC 150-TC : -45°C below ambient temperature to +150°C.
•DBC 650-TC : +50°C to +650°C.
This series of calibrators uses the GE technology temperature controller, providing
high performance temperature stability and accuracy. Temperature generation can
be controlled either in manual mode (direct) or automatic with ramping, stepping or
preset functions. The dry well temperature can be measured either with the internal
probe or with an external certified reference probe.
The key-pad, display and electrical connections enable the user to carry out test
procedures for thermocouples, RTD, switches, temperature transmitters and
converters, regulators. Both TC models contain a switch test function with a
hysteresis calculation. This Input/Measure facility enables direct connection and
reading of thermocouples, RTD, Voltages, Current and Resistance: the dual display
can simultaneously show the generated temperature (with internal or external
K400 Issue No. 1 1
reference) and a possible electrical feedback from a device under test (i. e.
temperature probes, temperature transmitters). An error calculation, with scaling
function, allows the user to check immediately the accuracy specification of the unit
under test. A 24 V auxiliary excitation provides power for temperature transmitters
under test.
1.1Automatic Calibration
An RS232 computer interface enables a Calibration Management Software to
provide automatic calibration procedures, typically with the following capabilities:
•Creation of instrument details
•Creation of test procedures and work orders
•Analysis of devices due for calibration
•Printing of calibration reports
•Export functions
Refer to section 4 SERIAL COMMUNICATION for remote control of this instrument.
1.2Parts Identification
Key to Figure 1-1
Fig no.Description
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
Dry well with insert
Protection grid
Carrying handle
4 mm socket for T/C, RTD and ohm connection, with CJ compensation
Tactile key-pad for direct selection of all the operating function (ramps, steps, preset
values, switch test, help, set-up) and to set temperature values
Cooling fan
Support feet (4)
Protection fuse (115 Vac 2x12.5 AT and 230 Vac 2x6.3 AT for DBC 650
(115 Vac 2x1.6 AT and 230 Vac 2x3.15 AT for DBC 150)
Main power selector 100 - 120 Vac and 220 - 240 Vac, 50/60 Hz
Power supply on/off switch
Main power socket: to connect the standard power cable
RS232 serial communications interface
Cursor (arrow) keys to select operating functions from the main menu and to move the
selection cursor through the menus
Digital display with Multi-language user interface
Switch test connection, suitable to connect test leads supplied as standard
4 mm socket for voltage and current input with 24V dc auxiliary excitation
Lemo connection for external reference probe: suitable for connecting PT 100 probes
with 3 or 4 wires, supplied with standard mating connector ( B1, B2 and B3 options)
2 K400 Issue No. 1
17
16
15
14
2
1
3
13
4
5
17
DBC-TS
15
12
6
7
9
11
10
8
Figure 1-1
K400 Issue No. 1 3
2INSTALLATION
2.1Preparation
General
•Remove the instrument from the carrying case and use it on a flat, stable
surface. To avoid any possible damage, either to the instrument or the unit
under test, the dry well of the instrument must be used in the vertical position.
•The fan system at the base of the instrument must be free from any obstruction.
• The dry well must be completely clean.
•Carefully read the SAFETY section of this manual.
AFTER USE, THE PROBES WILL BE VERY HOT. HANDLE WITH
CARE AND WAIT FOR THE PROBES TO COOL COMPLETELY.
•Check the probe diameter and operating range and use reduction inserts with
the correct sized holes (the hole diameter should comply with EA-10/13
guidelines). A large temperature difference between probe and dry well causes
large measuring errors. This difference increases as the difference between
the diameters increases.
•Always clean the probes, before use.
•Do not force the probes into the dry well, insert the probe down to the bottom of
the dry well.
4 K400 Issue No. 1
•When testing a probe with a certified reference probe, place both probes close
together and at the same depth.
insert
NO
NO
Reduction
YES
*
YES
160mm
50
mm
Internal reference probe
zone
Calibration
NOYES
* dimension should comply with EA-10/13 guidelines - for temperature
ranges -80° C to +660°C maximum 0.5mm
IMPORTANT NOTE
Reduction inserts with special holes could be necessary when carrying out a
calibration procedure on temperature device with specific mechanical
dimensions. Refer to GE for details of specific application solutions.
K400 Issue No. 1 5
2.4Electrical Connections
•Connection of a Thermocouple sensor
V IN
24V Out
35mA max
Ref RTD
0V
X MTR
CAT I
max 30V
35mA
max
mA IN
•Use of screw connections for T/C wires
-
+
1
2
3
4
6 K400 Issue No. 1
•Connection of a 3-wire RTD.
X MTR
24V Out
35mA max
0V
CAT I
max 30V
35mA
max
V IN
mA IN
•Connection of a 4-wire RTD
X MTR
24V Out
35mA max
0V
CAT I
max 30V
35mA
max
V IN
mA IN
Ref RTD
Ref RTD
•Connection of a 2-wire T/I transmitter
V IN
24V Out
35mA max
(external 24 V dc excitation)
Ref RTD
0V
X MTR
CAT I
max 30V
35mA
max
mA IN
0V user supply (SELV)
I
T
-
K400 Issue No. 1 7
+
24V user supply floating (SELV)
• Connection of a 2-wire T/I transmitter
(with 24 V dc auxiliary excitation)
X MTR
24V Out
35mA max
0V
Ref RTD
CAT I
max 30V
35mA
max
V IN
mA IN
I
T
•General
The temperature probes to be tested fit into an appropriate size reduction insert
located in the 30 mm diameter hole of the dry well. The dry well can be heated by
either cartridge heaters (DBC 650) or Peltier cells (DBC 150). A fan system fitted
on the bottom of the instrument reduces the temperature during cooling operations.
Operating procedure
The DBC Series can be used to carry out comparison test using a variety of
methods:
Reading of temperature calibration with the internal regulation probe
The generated temperature
is considered as reference
value and is shown on the
DBC display: this
temperature is measured
with the standard internal
probe.
Unit under Test
Dry Well
Internal Temperature
Probe
8 K400 Issue No. 1
Calibration with external reference probe
l
The temperature value to be considered as standard is measured with an external
and certified reference probe. This measurement reading can be directly shown
either on the DBC display (connecting the reference probe to the REF RTD socket
on the top of the instrument) or with an external, certified electronic calibrator.
Unit under Test
External Certified
Reference Probe
Dry Wel
Internal Temperature
Probe
K400 Issue No. 1 9
Power, reading and calibration of devices
under test (
The DBC-TC series has the input capability to
directly measure the electrical signal of unit
under test (T/C, RTD, mA, V, Ohms): using an
error calculation function (with scaling
capability), calibration procedures can be
carried out and the UUT accuracy immediately
measured.
A 24 V auxiliary excitation supply is also
available to power the loop of the device due for
calibration. The generated temperature can be
measured with the internal or the external
reference as described above.
DBC-TC only)
•External probe connections
A
1 Wire 1
2 Wire 1a
3 Wire 2
4 Wire 2a
A
10 K400 Issue No. 1
3OPERATI0N
3.1Key-pad Functions
7
4
1
0
infodirect
help
8
5
2
#
ce
9
6
3
+/-
setup
menu
ent
• ent (Enter) key
1.Pressing this key to confirms the selected choices or the temperature set-
point.
•ce (clear entry) key
2.Pressing this key cancels the previous key operation or corrects a typing
error.
•menu/setup key
3.Pressing this key recalls the Menu Selection Screen. Pressing this key after
the # key (2nd function), selects the SETUP menu display.
•# key
4.Pressing this key selects DIRECT temperature generation mode. Pressing
this key with other keys recalls second functions (# + set-up, # + , # +
etc.).
•help/Info key
5Pressing this key recalls a related display containing more information, either
on the selected function or on the operation to be performed. Pressing this
key after the # key, changes the display containing information relating to the
selected temperature reference or to the selected test function (for example
switch test).
K400 Issue No. 1 11
Loading...
+ 47 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.