GE Druck ADTS 505 User Manual

GE
Measurement & Control
Druck ADTS 505 Air Data Test Set
User Manual K0260
© 2013 General Electric Company. All Rights Reserved. Specifications are subject to change without notice. GE is a registered trademark of General Electric Company. Other company or product names mentioned in this document may
be trademarks or registered trademarks of their respective companies, which are not affiliated with GE.
K0260 Issue No. 8
Introduction
This technical manual provides operating instructions for the Air Data Test System compatible
with the requirements of first line operation for the technician and supervisor.
Scope
This technical manual contains a brief description, operation and testing procedures for the
user of this equipment with software version V1.08.
Safety
The manufacturer has designed this equipment to be safe when operated using the
procedures detailed in this manual. Do not use this equipment for any other purpose than that stated, the protection provided by the equipment may be impaired.
This publication contains operating and safety instructions that must be followed to make sure
of safe operation and to maintain the equipment in a safe condition. The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage.
Use qualified* technicians and good engineering practice for all procedures in this publication.
PIN Protection
The ADTS 505 contains two protected menus, the operating limits (described in this manual) and the maintain calibration menu (described in the calibration manual). The factory set PIN codes are contained in an envelope addressed to the Supervisor.
IMPORTANT NOTE
Change these codes for authorised access. Unauthorised access to these two menus can make this system inaccurate and could, in control mode, cause excessive rates of pressure change.
Pressure
Do not apply pressure greater than the maximum safe working pressure to the equipment.
Toxic Materials
There are no known toxic materials used in this equipment.
Maintenance
The equipment must be maintained using the manufacturer’s procedures and should be
carried out by authorized service agents or the manufacturer’s service departments
Technical Advice
For technical advice contact the manufacturer or subsidiary.
* A qualified technician must have the necessary technical knowledge, documentation, special test
equipment and tools to carry out the required work on this equipment.
ii
K0260 Issue No. 8
Associated Publications
This lists the manuals and publications referenced in this manual.
Calibration Manual K272
Air Data Test Set ADTS 505
Quick Reference Guide K274
Air Data Test Set ADTS 505
Approved Service Agents
Markings and Symbols
This equipment meets the requirements of all relevant European safety directives. The equipment carries the CE mark.
This symbol, on the instrument, indicates that the user should refer to the user manual. This symbol, in this manual, indicates a hazardous operation.
Do not dispose of this product as household waste. Use an approved organisation that collects and/or recycles waste electrical and electronic equipment. For more information, contact one of these:
our customer service department (contact us at www.ge-mcs.com)
your local government office
Approved Service Agents
For a list of approved service agents, visit our website: www.ge-mcs.com
Internet www.ge-mcs.com
Druck ADTS 505 User Manual
iii
K0260 Issue No. 8
Table of Contents
Preliminary pages
Introduction .................................................................................................................................................................................. ............... i
Scope .................................................................................................................................................................................. ............... i
Associated Publication................................................................................................................................................................... ............... ii
Approved Service Agents.............................................................................................................................................................. ............... ii
Table of contents (this table)....................................................................................................................................................... ............... iii
Abbreviations .................................................................................................................................................................................. ............... vii
Glossary .................................................................................................................................................................................. ............... ix
Pressure units and conversion factors................................................................................................................................. ...............xi
Section Title page
1 DESCRIPTION
1.1 Introduction......................................................................................................................................................... ............... 1-1
1.2 Operating Limits............................................................................................................................................... ............... 1-3
2INSTALLATION
2.1 Packaging ........................................................................................................................................................... ...............2-1
2.2 Packaging for Storage and Transportation....................................................................................... ............... 2-1
2.3 Return Goods Procedure............................................................................................................................. ............... 2-3
2.4 Electrical Connection..................................................................................................................................... ............... 2-6
2.5 Pneumatic Pressure Connection............................................................................................................. ............... 2-7
2.6 Positioning of the ADTS 505....................................................................................................................... ...............2-8
3OPERATION
3.1 Preparation.......................................................................................................................................................... ............... 3-1
3.2 Display Functions and Units of Measure............................................................................................. ............... 3-2
3.3 Quick Reference............................................................................................................................................... ............... 3-3
3.4 First Time Operators....................................................................................................................................... ............... 3-4
3.4.1..... Operating Modes............................................................................................................................ ............... 3-6
3.5 Operation and Example Procedures..................................................................................................... ............... 3-10
3.5.1..... Operating Procedures.................................................................................................................. ............... 3-10
3.5.2..... Power-up............................................................................................................................................. ............... 3-10
3.5.3..... Control or Measure Parameter................................................................................................ ............... 3-11
3.5.4..... Leak Testing the ADTS 505......................................................................................................... ...............3-11
3.5.5..... Changing the Units of Measurement.................................................................................... ............... 3-13
3.5.6..... Setting Limits..................................................................................................................................... ............... 3-13
3.6 Tests Before Use.............................................................................................................................................. ............... 3-15
3.6.1..... Aircraft System Protection......................................................................................................... ...............3-15
3.7 Testing Aircraft Systems or UUT.............................................................................................................. ............... 3-15
3.7.1..... Testing the Aircraft Static System........................................................................................... ............... 3-16
3.7.2..... Testing the Aircraft Pitot System............................................................................................. ............... 3-17
3.7.3..... Combined Testing of the Aircraft Pitot and Static Systems........................................ ............... 3-18
3.7.4..... Mach Test and Constant Mach................................................................................................ ...............3-19
3.7.5..... Leak Testing....................................................................................................................................... ............... 3-20
3.7.6..... Airspeed Switch Test..................................................................................................................... ............... 3-21
3.7.7..... Engine Pressure Ratio (EPR)....................................................................................................... ............... 3-22
3.7.8..... Go To Ground.................................................................................................................................... ............... 3-23
CONTENTS
iv
K0260 Issue No. 8
Table of Contents (contd)
Section Title page
3.8 Manual Venting of the Aircraft Pitot and Static Systems............................................................. ............... 3-24
3.9 ADTS 505 Options............................................................................................................................................ ...............3-24
3.10 Set-up Reference............................................................................................................................................. ...............3-25
4MAINTENANCE
4.1 Introduction......................................................................................................................................................... ............... 4-1
4.2 Materials and Tools......................................................................................................................................... ............... 4-2
4.3 Maintenance Tasks......................................................................................................................................... ...............4-3
4.4 Routine Maintenance..................................................................................................................................... ............... 4-4
5 TESTING AND FAULT FINDING
5.1 Introduction......................................................................................................................................................... ............... 5-1
5.2 Standard Serviceability Test...................................................................................................................... ............... 5-2
5.3 Fault Diagnosis................................................................................................................................................. ............... 5-3
5.4 Warnings and Self-test Errors.................................................................................................................. ............... 5-3
5.5 Further Testing.................................................................................................................................................. ...............5-6
5.6 Fault Finding....................................................................................................................................................... ............... 5-14
6 REFERENCE and SPECIFICATION
6.1 Introduction......................................................................................................................................................... ............... 6-1
6.2 Key-pad Selections and Functions......................................................................................................... ............... 6-1
F1 - F6................................................................................................................................................................... ............... 6-1
ALT/Ps................................................................................................................................................................... ............... 6-2
SPEED Qc.............................................................................................................................................................. ............... 6-2
MACH..................................................................................................................................................................... ............... 6-3
ROC Ps RATE........................................................................................................................................................ ............... 6-3
RATE....................................................................................................................................................................... ............... 6-4
LEAK MEASURE/CONTROL.......................................................................................................................... ............... 6-5
GROUND................................................................................................................................................................ ............... 6-5
0 to 9...................................................................................................................................................................... ............... 6-6
-000........................................................................................................................................................................ ............... 6-6
CLEAR/QUIT......................................................................................................................................................... ............... 6-6
DELETE................................................................................................................................................................... ............... 6-6
HELP....................................................................................................................................................................... ............... 6-7
ENTER.................................................................................................................................................................... ............... 6-7
CLEAR/QUIT + ENTER (ABORT)................................................................................................................... ............... 6-7
Main Menu........................................................................................................................................................... ............... 6-8
[Rate Timer]......................................................................................................................................................... ............... 6-8
[Units].................................................................................................................................................................... ............... 6-9
[EPR]....................................................................................................................................................................... ............... 6-10
[Hold]..................................................................................................................................................................... ............... 6-10
Nudge [ or ]............................................................................................................................................... ............... 6-10
6.3 SETUP .................................................................................................................................................................... ............... 6-11
6.3.1..... Set-up 1 of 2....................................................................................................................................... ............... 6-11
...............[Display],[Single, Dual, Quad, Hand Term]........................................................................... ............... 6-11
Druck ADTS 505 User Manual
v
K0260 Issue No. 8
Table of Contents (contd)
Section Title page
[Units].................................................................................................................................................................... ............... 6-12
[Limits]................................................................................................................................................................... ............... 6-13
[Alt. Corr.].............................................................................................................................................................. ............... 6-15
[Date Format]..................................................................................................................................................... ...............6-16
6.3.2..... Set-up 2 of 2....................................................................................................................................... ............... 6-17
...............[Operational Hours]....................................................................................................................... ............... 6-17
...............[Software Versions]........................................................................................................................ ............... 6-17
...............[Security].............................................................................................................................................. ...............6-17
...............[Maintain Calibration]................................................................................................................... ............... 6-17
...............[Auto Leak].......................................................................................................................................... ...............6-18
6.4 Other Features................................................................................................................................................... ............... 6-19
Elapsed Time Counter.................................................................................................................................... ............... 6-19
...............Auto Zero............................................................................................................................................. ............... 6-19
...............Hand Terminal Option A.............................................................................................................. ............... 6-19
...............Hand Terminal Option B.............................................................................................................. ............... 6-19
............... ...............Template Format for Creating Test Sequence Files....................................... ............... 6-32
............... ...............Option B Advanced Hand Terminal Specification........................................... ...............6-41
6.5 Specification........................................................................................................................................................ ............... 6-42
CONTENTS
vi
K0260 Issue No. 8
Table of Illustrations
Figure Title page
1-1 ADTS 505 General View............................................................................................................................... ............... 1-2
2-1 ADTS 505 Accessories................................................................................................................................... ............... 2-4
2-2 ADTS 505 Altitude Correction On-aircraft........................................................................................... ...............2-9
2-3 ADTS 505 Altitude Correction Off-aircraft........................................................................................... ............... 2-9
2-4 ADTS 505 General View............................................................................................................................... ............... 2-10
3-1 Key-pad and Display...................................................................................................................................... ............... 3-3
4-1 Fuse Replacement........................................................................................................................................... ............... 4-5
5-1 Fault Finding Chart......................................................................................................................................... ............... 5-4
5-2 System Screen for Diagnosis.................................................................................................................... ............... 5-7
5-3 System Screen................................................................................................................................................... ............... 5-9
5-4 Key-pad Test Display..................................................................................................................................... ............... 5-10
6-1 ARINC 565 Limits Graph.............................................................................................................................. ............... 6-14
6-2 Altitude Correction On-aircraft................................................................................................................ ............... 6-15
6-3 Altitude Correction Off-aircraft................................................................................................................ ............... 6-16
6-4 Interconnection................................................................................................................................................ ............... 6-20
6-5 Power-up Displays.......................................................................................................................................... ............... 6-21
6-6 Main Menu Screen.......................................................................................................................................... ...............6-22
6-7 Main Menu Structure..................................................................................................................................... ............... 6-23
6-8 Manual Control Selections......................................................................................................................... ............... 6-24
6-9 Set-up Menu Selections............................................................................................................................... ............... 6-27
6-10 Test Sequence Selections........................................................................................................................... ............... 6-30
6-11 Customer Test Sequence File Selections............................................................................................. ............... 6-31
6-12 System Status..................................................................................................................................................... ............... 6-38
6-13 Example Error Screen.................................................................................................................................... ............... 6-39
List of Tables
Table Title page
2-1 Parts List............................................................................................................................................................... ............... 2-5
4-1 Maintenance Chart......................................................................................................................................... ............... 4-1
4-2 Materials List....................................................................................................................................................... ............... 4-2
4-3 Tool and Test Equipment Requirements.............................................................................................. ............... 4-2
5-1 Fault Finding ...................................................................................................................................................... ............... 5-5
5-2 System Screen Information....................................................................................................................... ............... 5-8
5-3 Error Messages................................................................................................................................................. ...............5-14
5-4 Warning Messages......................................................................................................................................... ............... 5-15
Druck ADTS 505 User Manual
vii
K0260 Issue No. 8
Abbreviations
The following abbreviations are used in this manual; the abbreviations are the same in the singular and plural.
AAmpere abs Absolute a.c. Alternating current ADTS Air Data Test Set ALT Altitude ARINC Air Radio Incorporated ASI Airspeed indicator CAS Calibrated airspeed COSHH Control of Substances Hazardous to Health Regulations cm Centimetre d.c. Direct current Def Define e.g. For example EPR Engine pressure ratio etc. And so on °F Degrees Fahrenheit Fig. Figure ft Foot g Gauge hHour HBC High breaking capacity Hg Mercury hm Hecto metre Hz Hertz IAS Indicated airspeed i.e. That is in Inch kg Kilogram km Kilometre kts Knots LCD Liquid crystal display mMetre mA Milliampere max Maximum mbar Millibar min Minute or minimum mm Millimetre mph Miles per hour MSDS Material safety data sheet mV Millivolts No. Number PIN Personal identification number Ps Static pressure psi Pounds per square inch Pt Total pressure (Pitot) Qc Differential pressure Ps-Pt QFE Local atmospheric pressure QNH Barometric pressure at sea level
Abbreviations
viii
K0260 Issue No. 8
Abbreviations (contd)
RGA Return Goods Authorization (Druck procedure) RH Relative humidity RMS Root mean square ROC Rate of climb Rt Rate SST Standard serviceability test VVolts +ve Positive
-ve Negative °C Degrees Celsius °F Degrees Fahrenheit
Druck ADTS 505 User Manual
ix
K0260 Issue No. 8
Glossary
Terminology
The terminology used in this manual is specific and individual interpretation must not be introduced. The terms are defined as follows:
Adjust To bring to a more satisfactory state; to manipulate controls, levers, linkages, etc. to return equipment
from an out-of-tolerance condition to an in-tolerance condition.
Align
To bring into line; to line up; to bring into precise adjustment, correct relative position or coincidence.
Assemble
:To fit and secure together the several parts of; to make or form by combining parts.
Calibrate
: To determine accuracy, deviation or variation by special measurement or by comparison with a
standard.
Check:
Make a comparison of a measure of time, pressure, temperature, resistance, dimension or other quality with a known figure for that measurement.
Disconnect
:To detach the connection between; to separate keyed or matched equipment parts.
Dismantle
:To take apart to the level of the next smaller unit or down to all removable parts.
Examine
: To perform a critical visual observation or check for specific conditions; to test the condition of.
Fit
: Correctly attach one item to another.
Inspect
: Review the work carried out by Specialists to ensure it has been performed satisfactorily.
Install
: To perform operations necessary to properly fit an equipment unit into the next larger assembly or
system.
Maintain
: To hold or keep in any particular state or condition especially in a state of efficiency or validity.
Operate
: Make sure that an item or system functions correctly as far as possible without the use of test
equipment or reference to measurement.
Readjust
: To adjust again; to move back to a specified condition; to bring back to an in-tolerance condition.
Reconnect
:To rejoin or refasten that which has been separated.
Refit
: Fit an item which has previously been removed.
Glossary
x
K0260 Issue No. 8
Remove: To perform operations necessary to take an equipment unit out of the next larger assembly or
system. To take off or eliminate. To take or move away.
Repair
: To restore damaged, worn out or malfunctioning equipment to a serviceable, usable or operable
condition.
Replace
: Remove an item and fit a new or a serviced item.
Reset
: To put back into a desired position, adjustment or condition.
Service
: To perform such operations as cleaning, lubricating and replenishing to prepare for use.
Test
: Ascertain by using the appropriate test equipment that a component or system functions correctly.
Druck ADTS 505 User Manual
xi
K0260 Issue No. 8
Table of pressure units and conversion factors
Unit Conversion To convert FROM pressure VALUE 1 in pressure UNITS 1
TO pressure VALUE 2 in pressure UNITS 2, calculate as follows:
VALUE 2 = VALUE 1 x FACTOR 1
FACTOR 2
Note:
The conversion factor for pressure units referenced [1] are calculated for a water temperature of 4°C. Pressure units referenced [2] are calculated for a water temperature of 68°F these units are normally used in the USA.
tinuerusserP )slacsaP(rotcaF tinuerusserP )slacsaP(rotcaF
rab000001tf/fbl
2
3088.74
ni/fbl
2
)isp(67.4986gHni93.6833
Hm
2
O56.6089Hni
2
]1[O980.942
rabm001Htf
2
]1[O70.9892
mc/fgk
2
5.66089mta0.523101
m/fgk
2
56608.9tf/ldp
2
61884.1
gHmm223.331mc/nyd
2
1.0
gHmc22.3331rabh00000001
gHm0.223331tf/fnot
2
)KU(0.252701
H/mm
2
]1[O56608.9ni/fnot
2
)KU(00344451
H/mc
2
]1[O5660.89Hni
2
]2[)ASU(O53146.842
m/N
2
1Htf
2
]2[)ASU(O3896.3892
aPh001mm/Pk
2
0566089
aPk0001mc/Pk
2
5.66089
aPM0000001m/Pk
2
56608.9
rrot223.331
xii
K0260 Issue No. 8
Druck ADTS 505 User Manual
Description 1 - 1
K0260 Issue No. 8
1
1INTRODUCTION
1.1 Introduction
The ADTS 505 is a self-contained flight-line air data test system, enclosed in an ABS case. The
unit provides complete pressure and vacuum measuring and control for on-aircraft sense and leak testing, functional tests of air data instruments, components and systems.
The ADTS 505 displays and operates in either units of pressure measurement or aeronautical
units. In the control mode, the rate that the pressures change towards new set-points can be controlled in true aeronautical rate units.
There are two independent pneumatic channels connect to the aircraft or instrument systems,
one for static and one for pitot. They can be operated as measure only channels with leak testing facilities or each can be control channels producing true pressure conditions for altitude and airspeed. Two pneumatic outlet ports, on the front panel and identified as Ps (static) and Pt (pitot) provide connection to the aircraft system or unit under test.
To protect sensitive instruments and equipment a `ground' facility automatically and safely
controls both channels to atmospheric pressure at the previously entered rates of change and then informs the user when both channels are safely at `ground'.
Pre-defined sets of limits, stored in the system, prevent excessive pressures and rates
damaging aircraft systems and components. A further five sets of operating limits can be defined by the supervisor or quality assurance engineer. The user can select sets of limits but cannot change the values.
The user interface is either the key-pad and display on the front panel or one of two types of
optional hand terminal connected to the front panel. The two types of hand terminal are identified as option A and option B.
The option A hand terminal contains all the facilities of the front panel key-pad and display.
The key-pad contains fixed function keys used to select various parameters, modes and enter numeric values. The display shows various menus, each menu provides selections using six menu-defined function keys.
The option B advanced hand terminal is a computer-based device that contains all the facilities
of option A on a touch-sensitive, Windows® display, colour screen. It also up-loads and down­loads user-defined test programs and displays the resulting test data.
The integral pumps of the ADTS 505, produce pressure and vacuum supplies for the unit's
controlling requirements. The power supply connection for the unit is located on the front panel.
1 - 2 Druck ADTS 505 User Manual
K0260 Issue No. 8
FIGURE 1-1 ADTS 505 GENERAL VIEW
Option A Hand Terminal
Option B Hand Terminal
Air DataTest Set - HandTerminal
Leak Measur
e
LEAK MEASURE MODE
Mach
0.001
Mach
DO NOT
OBSTRUCT
THESE VENTS
WATER
DRAIN
Advanced Hand TerminalAdvanced Hand Terminal
g
T5.0A 250V
HBC
Description 1 - 3
K0260 Issue No. 8
1
1.2 Operating Limits
The ADTS 505 is supplied with the following pre-defined operating limits.
Civil Limits
Parameter Limit
MIN ALT -1,000 ft MAX ALT 50,000 ft MIN CAS 0.0 knots MAX CAS 450.0 knots MAX MACH 1.000 Mach MAX ROC 6,000 ft/min ARINC LIMITS OFF ALT CORRECTION 0 ft MIN Ps 115.972 mbar MAX Ps 1050.406 mbar MIN Qc 0.0 mbar MAX Qc 368.01 mbar MAX Rate Ps 109.85 mbar/min MAX Rate Qc 109.85 mbar/min
Standard Limits
Parameter Limit
MIN ALT -2,000 ft MAX ALT 60,000 ft MIN CAS 0.0 knots MAX CAS 650.0 knots MAX MACH 1.732 Mach MAX ROC 10,000 ft/min ARINC LIMITS OFF ALT CORRECTION 0 ft MIN Ps 65.00 mbar MAX Ps 1088.6 mbar MIN Qc 0.0 mbar MAX Qc 866.00 mbar MAX Rate Ps 200 mbar/min MAX Rate Qc 200 mbar/min
1 - 4 Druck ADTS 505 User Manual
K0260 Issue No. 8
Max Limits
Parameter Limit
MIN ALT -2,000 ft MAX ALT 60,000 ft MIN CAS 0.0 knots MAX CAS 650.0 knots MAX MACH 2.800 Mach MAX ROC 40,000 ft/min ARINC LIMITS OFF ALT CORRECTION 0 ft MIN Ps 10.90 mbar MAX Ps 1355.00 mbar MIN Qc -1355.00 mbar MAX Qc 2490.00 mbar MAX Rate Ps 1000 mbar/min MAX Rate Qc 1000 mbar/min
EPR Limits
Parameter Limit
MIN INLET 27.0 mbar MAX INLET 1355.0 mbar MIN OUTLET 27.0 mbar MAX OUTLET 2500.0 mbar MIN EPR 0.1 MAX EPR 10.0 MIN INLET RATE 0 MAX INLET RATE 1000 mbar/min MIN EPR RATE 0 MAX EPR RATE 60 EPR/min
Druck ADTS 505 User Manual 2 - 1
K0260 Issue No. 8
2
2 INSTALLATION
2.1 Packaging
On receipt of the ADTS 505 check the contents of the packaging against the following lists: i) ADTS 505 Air Data Test Set ii) Power supply cable - 5 m, no line plug (regional line plugs require option C) iii) Option A
comprising:
Hand terminal Hand terminal cable - 18 m (option A only) 5 m hose - red 5 m hose - blue Accessory Bag
iv) Option B
comprising:
Advanced hand terminal Hand terminal cable - 18 m (option B only) Cable, communications Cable, power supply Power pack
Accessory Bag v) 3 m hose - red vi) 3 m hose - blue vii) User Manual (this publication) viii) Quick Reference Guide K0274 ix) Kit, fuse/o-ring (also see Figure 2-1, item1)
2.2 Packaging for Storage or Transportation
To store the unit or to return the unit for calibration or repair carry out the following procedures:
1. Pack the unit as detailed in the following procedure.
2. To return the unit for calibration or repair complete the return goods procedure as
detailed in 2.3.
2 - 2 Installation
K0260 Issue No. 8
Procedure
The unit should be at zero/ambient pressure. Disconnect the hose assemblies and stow in the
lid.
Switch OFF and disconnect from the electrical power supply. Disconnect the power
supply cable and the hand terminal cable. Disconnect the hand terminal cable from the hand terminal.
Fit the lid to the unit.
The power supply cable, hand terminal cable and the hand terminal should be placed
in the original packing material.
If available, use the original packing material. When using packing materials other than the
original, proceed as follows.
Wrap unit in polyethylene sheeting.
Select a double-wall cardboard container. Inside dimensions must be at least 15 cm greater than the equipment. The carton must meet test strength requirements of >
125 kg.
Protect all sides with shock-absorbing material to prevent equipment movement within
the container.
Seal carton with approved sealing tape.
Mark carton “FRAGILE” on all sides, top, and bottom of shipping container.
To return the unit for calibration or repair complete the return goods procedure as detailed in 2.3.
Environment
The following conditions apply for both shipping and storage:
Temperature Range........... ............-20 to +70°C (-4 to +158°F)
Altitude......... ............Up to 15,000 feet (4,570 metres)
Druck ADTS 505 User Manual 2 - 3
K0260 Issue No. 8
2
2.3 Returned Goods Procedure
Should the unit require calibration or become unserviceable it can be returned to the Druck Service Department.
Please contact our Service Department, either by 'phone, fax or E-mail, to obtain a Returned Goods Authorization (RGA) number or (Return Material Authorization [RMA] in USA), providing the following information:
Product (i.e. ADTS 505) Serial number Details of defect/work to be undertaken Calibration traceability requirements Operating conditions
Safety Precautions
You must also tell us if the product has been in contact with anything hazardous or toxic and, the relevant COSHH (MSDS in USA) references and precautions to be taken when handling.
IMPORTANT NOTICE
Service or calibration by unauthorized sources will affect the warranty and may not guarantee further performance
2 - 4 Installation
K0260 Issue No. 8
FIGURE 2-1 ADTS 505 ACCESSORIES
Air Data Test S
et - Hand T
erminal
Leak Measure
LMEAK EASURE MODE
M
ach
0
.0
0
1
Mach
Advanced Hand TerminalAdvanced Hand Terminal
g
191-198
2
1
5
18m
3A
6
9
12
13
11
Note: Item 13 supplied for the country of use
10
18m
4
Druck ADTS 505 User Manual 2 - 5
K0260 Issue No. 8
2
# not illustrated † alternative
TABLE 2-1 PARTS LIST
No Part Number Description
Qty per
assy
1
AS505-18-3124M0
Kit, fuse/o-ring comprising: o-ring AN4 fuse, T5.0A/250V HBC
1
2
AS505-53-3124M0
comprising: Hose, red, ST/OPN, AN4, 3m
1
Hose, blue, ST/OPN, AN4, 3m
1
3A
AS505-40-3124M0 Cable, AC Power, 5M (open end)
1
3B#
AS505-40-3124M0 Cable, AC Power, 5M (250V UK plug)
1
3C#
AS505-41-3124M0 AC Power, 5M (115V US plug)
1
3D#
AS505-42-3124M0 AC Power, 5M (250V European plug)
1
4
AS505-56-3124M0
Option A Hand terminal Assembly
1
4.1#
-
Option A and B Hose, red, 5M Hose, blue, 5M
1
5
AS505-54-3124M0
Option A Cable, Hand terminal 18M (option)
1
6
AS505-57-3124M0
Option A and B Bag, Accessory, Hand Terminal
1
7#
AS505-60-3124M0 Handbook, User Manual, K0260
1
8#
AS505-61-3124M0 Handbook, Calibration Manual, K0272
1
9
AS505-56-3124M1
Option B Hand terminal, Advanced
1
10
AS505-54-3124M1
Option B Cable, Hand terminal 18M, Safety marking (option)
1
11
AS204-06-3435M0 Cable, Communications PC
1
12
AS204-07-3435M0 Power pack
1
13
- part of item 12 -Adaptor, power pack (comprising: 4)
1
14
AS204-10-3435M0 Stylus, pack (comprising: 3)
1
2 - 6 Installation
K0260 Issue No. 8
2.4 Electrical Connection
WARNING: VOLTAGES IN EXCESS OF 30 VOLTS (RMS) AC OR 50 VOLTS DC, IN CERTAIN
CIRCUMSTANCES, CAN BE LETHAL. CARE MUST BE TAKEN WHEN WORKING ON LIVE, EXPOSED CONDUCTORS.
Power Supply Connection
The unit must be connected to the correct electrical power supply as stated, adjacent to the power connector.
Open end cables Only a qualified technician (see page i) must fit a connector to a cable supplied by GE (see Table 2-1, item 3).
CAUTIONS:
1T
HE SUPPLY MUST PROVIDE CONNECTION TO A PROTECTIVE GROUND TERMINAL. THE UNIT MUST, AT
ALL TIMES, BE CONNECTED TO THE SUPPLY EARTH (GROUND).
2T
HE POWER SUPPLY CABLE AND CONNECTOR MUST BE CORRECTLY RATED FOR THE POWER SUPPLY.
Note: The ADTS 505 is normally supplied with an approved power supply cable for use in the country
o f d e li ve ry . T hi s c an li mi t t he ma xi mu m s up pl y v ol ta ge th at ca n b e s af el y u se d. e.g. a NEMA 5-15P terminated cable, for use in the U.S.A., is approved for a maximum of 125 V ac; it must be replaced for a higher supply voltage.
Make sure that the power supply is off before connecting the power cable. If required, connect the hand terminal to the connector on the front panel.
Note: Connecting the hand terminal to the unit disables the front panel key-pad.
Fuses
The two fuses, located in the holders and mounted on the front panel, protect the unit. The fuses are connected in the live and neutral supply circuit and are T5.0A/250V HBC type.
Note: Only replace the fuses with the correct type, see Figure 2-1 item1, to replace see section 4-5.
External earth/ground connection
An external earth (ground) cable may be connected to the stud on the front panel of the unit
providing integrity of the earth (ground) connection.
F
IGURE 2-2 POWER SUPPLY LABEL
Pin
(ADTS)
European
Colour
U.S.
Color
Function
1
Brown Black Live
4
Blue White Neutral
Centre
Green/Yellow Green Protective Earth (Ground)
100 / 120 / 230V ~ 50 / 60Hz 200VA
115V ~ 400Hz 200VA
Druck ADTS 505 User Manual 2 - 7
K0260 Issue No. 8
2
Connection
The ADTS 505 provides a power supply to the advanced hand terminal when connected to the hand terminal connector on the front panel.
The power supplies must be isolated when connecting the advanced hand terminal in the hazardous area.
Using the advanced hand terminal when not connected to the ADTS 505
Using the advanced hand terminal with a pc to create test scripts.
The unit must be connected to the correct electrical power supply, see paragraph 2.4.
Before use, make sure the SELV power adaptor supplied with the instrument is correct for the power supply voltage. The Safety Extra Low Voltage (SELV) power adaptor complies with EN61010 (including safety requirements for laboratory instruments).
Note: The instrument can be powered from other DC power supplies of the correct voltage range. It
is the user’s responsibility to make sure the power supply is safe.
Make sure that the power supply is off before connecting the power cable.
Communications Connection
Communications cable - parts list item 11
2.5 Pneumatic Pressure Connections
Ps (static)- AN4 (MS33656-4) Pt (pitot)- AN4 (MS33656-4)
When not in use, a blanking cap must be fitted.
Note: When carrying out a leak test, a leak of this blanking cap affects the performance of the ADTS 505.
Pin No. ADTS
1
2
BLK
BLK-W
Function
Connector
Cable
Colour
0V
+ VIN
Jack plug outer
Jack plug inner
3
5
4
RED
RS232 Tx
PIN 2
PIN 3
PIN 5
BLK
GRN
RS232 0V
RS232 Rx
2 - 8 Installation
K0260 Issue No. 8
2.6 Positioning of the ADTS 505
WARNING: IN AN ENCLOSED AREA WITH FUEL VAPOUR PRESENT THIS EQUIPMENT MUST BE
PLACED AT LEAST 0.5 METRES ABOVE FLOOR LEVEL. THIS EQUIPMENT CONTAINS A D.C. MOTOR WITH BRUSHES THAT COULD CAUSE A SPARK.
C
AUTION:TO OPERATE, PLACE THE UNIT ON A HORIZONTAL SURFACE (FRONT PANEL UPPERMOST) OR VERTICALLY
(
POWER SUPPLY CONNECTOR UPPERMOST) THIS ALLOWS THE WATER IN THE WATER FILTER TO VENT.
W
ATER COULD CONTAMINATE THE CONTROLLER MANIFOLD AND AFFECT CONTROLLER PERFORMANCE.
Note: In control mode, the water drain, located at one end of the unit near the carrying handle, produces
a flow of air and some water. The amount of water depends on the humidity and the operating time in control mode.
C
AUTION:BEFORE USE, CHECK THE WATER DRAIN PIPE, IT MUST BE FREE OF OBSTRUCTION. WHEN IN CONTROL
MODE CHECK THAT A SMALL FLOW OF AIR COMES OUT OF THE DRAIN PIPE.
Notes:
1 When checking for a small flow of air do not block the drain pipe completely this causes a back-
pressure in the pipe and controller instability.
2 If no air flows from the drain pipe, when in control mode, switch off and start again. If no air
flows after a re-start, switch off and do not use the ADTS 505, return the unit to the repair depot.
C
AUTION: THE SIDE VENTS MUST NOT BE OBSTRUCTED, THIS UNIT REQUIRES AN AIRFLOW FOR THE INTERNAL COOLING FANS.
It is important that the position of the ADTS 505 in relation to the aircraft altitude sensors is
known. An altitude correction must be made to allow for the difference in height between the reference level and the aircraft's altitude sensors (Figures 2-2 and 2-3). The Reference section contains details of altitude correction.
WARNING: OBSERVE THE APPROPRIATE SAFETY INSTRUCTIONS AND TESTING PROCEDURES
DETAILED IN THE AIRCRAFT MAINTENANCE MANUALS AND COMPONENT MAINTENANCE MANUALS.
Druck ADTS 505 User Manual 2 - 9
K0260 Issue No. 8
2
FIGURE 2-2 ADTS 505 ALTITUDE CORRECTION ON-AIRCRAFT
FIGURE 2-3 ALTITUDE CORRECTION OFF-AIRCRAFT
2 - 10 Installation
K0260 Issue No. 8
Power supply
fuses
Power supply
switch
Hand terminal
connector (option)
Ps (static)
connector
Pt (pitot)
connector
Power supply
connector
G
()
stud
round
Earth
Key-pad
and display
Air
vent
Water
drain
DO
NOT
OBSTRUCT
THESE
VENTS
WAT
E
R
D
R
A
IN
T5.0A
250V
HBC
FIGURE 2-4 ADTS 505 GENERAL VIEW
Druck ADTS 505 User Manual 3 - 1
K0260 Issue No. 8
3
3OPERATION
3.1 Preparation
WARNING:
OBSERVE SAFETY PRECAUTIONS STATED IN LOCAL ORDERS AND THE AIRCRAFT OR EQUIPMENT SERVICING PROCEDURES.
Make sure the electrical and pneumatic connectors, electrical cables and pipes and positioning of
the ADTS 505 comply with the instructions and requirements in Section 2 Installation.
Carry out the following before use:
If necessary, carry out the maintenance task detailed in Section 4.
Make sure the air data test system power supply switch on the front panel is set to
OFF. Connect the air data test system to the electrical supply, make sure the supply includes a connection to a protective earth.
Inspect the pneumatic hoses for damage, ingress of dirt and moisture. Make sure the
aircraft adaptors are serviceable and the pipe connections are air-tight.
Note: Do not connect the air data test system to a contaminated aircraft system. Inspect the static
vents and pitot probes for dirt and debris before connecting. If necessary, check the pitot-static system water drain traps.
Connect, to the air data test system, the hoses necessary for the test procedures to be carried out:
STATIC output (Ps), PITOT output (Pt). Temporarily seal the free ends of the hoses.
Note: When connected, take care not to kink or stand on the hoses.
Before use, the ADTS 505 should be tested, for first time users see section 3.4, for users requiring
more operating detail see section 3.5.
This section contains a quick reference chart detailing all the functions of the key-pad. Further
quick reference charts, at the end of this section, detail the set-up menu.
Review and become familiar with the whole procedure before starting the test process on an
aircraft or component.
3 - 2 Operation
K0260 Issue No. 8
3.2 Display Functions and Units of Measure
When operating in either pressure measuring or pressure controlling modes, the ADTS 505 can display the following information:
Aeronautical Functions
Display Abbreviation Displayed Units
(if applicable)
Altitude ALT ft, m Calibrated Airspeed CAS kts, km/h, mph Mach MACH ­Rate of Climb ROC ft/min, m/min, m/s, hm/min Rate of Airspeed Rt CAS kts/min, km/h/min, mph/min
Pressure Functions Display Abbreviation Displayed Units
(if applicable)
Static (Absolute) Ps [P] Pitot (Absolute) Pt [P] Dynamic or Impact (Differential) Qc [P] Engine Pressure Ratio EPR ­Rate of Ps Rt Ps [P] /min Rate of Pt Rt Pt [P] /min Rate of Qc Rt Qc [P] /min Rate of EPR Rt EPR EPR/min
Where [P] is the currently selected pressure units from the following list:
mbar, inHg, mmHg, inH2O (4°C), inH2O (20°C), inH2O (60°F), psi, hPa, kPa.
Operating Range and Performance
The ADTS 505 is supplied with a full-scale range of 650 knots for measurement and control of the pitot pressure channel. The unit can measure altitude up to 105,000 ft; altitude control depends on the performance of the pump, the integral pump is capable of achieving -2000 to 60,000 ft. Refer to section 6 for more details of performance and specification.
Sets of factory-defined limits known as ARINC565, Standard, Military, Max and EPR can be selected through the SETUP menu (see Reference section 6). Operators may also define up to five sets of additional limits for different aircraft types. When configuring the display to aeronautical or pressure units operators should be aware that when units of pressure are selected, wider full-scale pressure limits will be enabled for some parameters.
Loading...
+ 94 hidden pages