CT Sytec/Sytec i Series
Functional Check /Adjustment
Copyrighte 1999 by General Electric Company
Operating Documentation
Page 2
Page 3
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
WARNING
AVERTISSEMENT
2234976
D THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
D IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
D DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
D FAILURE
PROVIDER,
OR OTHER HAZARDS.
D CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.
D SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
D NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
D LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.
T
O HEED THIS W
OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL
ARNING
MA
Y RESUL
T IN INJUR
Y T
O THE SER
VICE
WARNUNG
AVISO
D DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN
ENGLISCHER SPRACHE.
D FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
D VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST–HANDBUCH
WURDE.
D WIRD
D ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.
D SI
D NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
D LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
DIESE W
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.
ALGÚN PROVEEDOR DE SER
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIÓN.
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES
NATURALEZA.
ARNUNG NICHT BEACHTET
PROV
OCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS
NICHT
ZU RA
TE GEZOGEN UND VERST
, SO KANN
VICIOS AJENO A GEMS SOLICIT
ES ZU VERLETZUNGEN
ANDEN
A UN
IDIOMA
O DE OTRA
a
Page 4
GE MEDICAL
REV
0
ATENÇÃO
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
2234976
D ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA
DISPONÍVEL EM INGLÊS.
D SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE
D NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
DO CLIENTE
FORNECER OS SER
VIÇOS DE TRADUÇÃO.
AVVERTENZA
D O NÃO CUMPRIMENTO DESTE
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.
D IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE
SOLTANTO IN INGLESE.
D SE
UN ADDETT
MANUALE IN UNA LINGUA DIVERSA, IL
DIRETTAMENTE ALLA TRADUZIONE.
D SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
D NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER
FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.
O ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE
A
VISO PODE POR EM PERIGO A SEGURANÇA
CLIENTE È TENUT
O A PROVVEDERE
IL
b
Page 5
CT SYTEC/SYTEC I SERIES
GE MEDICAL
REV
0
SYSTEMS
FUNCTIONAL
CHECK /
ADJUSTMENT
2234976
IMPORTANT! . . . X-RAY PROTECTION
X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained
should be thoroughly read and understood by everyone who will use the equipment before you attempt to
place this equipment in operation. The General Electric Company, Medical Systems Group, will be glad to
assist
and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against
practical design of equipment can provide complete protection. Nor can
take adequate precautions to prevent the possibility of any persons carelessly exposing themselves or others to
radiation.
It
is important that everyone having anything to do with x-radiation be properly trained and fully acquainted with the
recommendations
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the
International
The
equipment is sold with the understanding that
and representatives have no responsibility for injury or damage which may result from improper use of the equipment.
Various
protective material and devices are available. It is urged that such materials or devices be used.
of
the National Council on Radiation Protection and Measurements as published in NCRP Reports
Commission on Radiation Protection, and take adequate steps to protect against injury
the General Electric Company
x-radiation other than the useful beam, no
any practical design compel the operator to
.
, Medical Systems Group, its agents,
All electrica
performed by l
b
y l
icense
b
e p
highly sophisticated
d e
erforme
l i
nstallation
icense
lectrica
d b
y q
d e
l c
ualifie
, a
ontractors
nd specia
s that are p
lectrica
l c
d GE M
edica
l e
reliminar
ontractors
. O
the
ngineering competenc
r c
onnection
l p
ersonnel
y t
. I
o p
ositionin
n a
ddition
. The p
s b
g o
, e
lectrica
etwee
roduct
e i
f the e
quipmen
l f
eed
n p
iece
s o
s i
nvolve
s required.
t a
t the site p
s into the P
f e
lectrica
l e
d (and the a
repare
owe
r D
istributio
quipment
ccompanyin
d for the e
n Unit s
, c
alibrations
g e
lectrica
quipmen
hal
l b
e p
, and t
estin
l i
nstallations
t s
hal
l b
erformed
g s
hall
) a
re
e
In performin
work
The purchaser of GE equipment shall only utiliz e qualified personnel (i.e., GE’s field engineers, personnel of thir d-party
service companies wit
on t
hes
g all e
e p
roducts wil
lectrica
h e
l work o
l c
omply wit
quivalen
n t
hes
t t
raining
e p
roducts, GE will use its own s
h the r
equirements o
, o
r l
icense
f the a
d e
lectricians) t
c
peciall
pplicabl
o p
e e
erfor
y t
raine
lectrica
m e
lectrica
d f
l c
iel
d e
odes.
l s
ngineers
ervicin
. All o
g on the e
f GE’s e
quipment.
lectrical
Page 6
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2234976
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment”
Electric representative or a hospital receiving agent. Whether noted or concealed, damage
carrier immediately
for inspection by the carrier
within
Call Traf
this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item
damaged
Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.
written on
this 14 day period.
fic and T
and extent of damage.
all
upon discovery
ransportation, Milwaukee, WI (414) 827–3449
copies of the freight or express bill
, or in any event, within 14 days after receipt, and the contents and containers held
. A transportation company will
not
before
delivery is accepted or
pay a claim for damage if an inspection is not requested
/ 8*285–3449
immediately
“signed for” by a General
MUST
be reported to the
after damage is found. At
OMISSIONS & ERRORS
GE
personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documentation.
Customers,
please contact your GE Sales or Service represenatives.
CAUTION
Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.
Devices not to be used near this equipment:
Devices
radio–controlled toy
Keep power to these devices turned of
Medical staf
which intrinsically transmit radio
, etc.
f in charge of this equipment is required to instruct technicians, patients an
2-1GENERAL2–1.
2-2FD FOR SYSTEM/OPTION SOFTWARE2–3.
2-3PRE–ATINST PROCEDURE2–5.
2-4ATINST2–11
2-5POST–ATINST PROCEDURE2–17.
2-5-1Loading Version 3.02 System Software2–25.
2-5-2Loading Version 4.00 System Software2–27.
2-5-3Loading V/R 4.01 System Software2–34.
2-5-4Loading V/R 4.02 System Software2–36.
3 – X–RA
3-1GENERAL3–1.
3-2PLANE OF ROTATION3–3.
3-3DETECTOR BEAM ON WINDOW3–6.
3-4GRAVITY SAG3–9.
3-5ISO CENTER ALIGNMENT3–14.
3-6FILTER CENTER ALIGNMENT3–18.
4 – SYSTEM CALIBRA
4-1GENERAL4–1.
4-2CALIBRATION4–2
4-3CT NUMBER ADJUSTMENT4–4.
section assumes all wiring and cabling is complete, equipment is installed, and that the following circuit
and switches are OFF.
PRIOR T
O APPL
YING POWER
CAUTION
Leave all power to the System OFF and tagged until power is called for in a specific step.
D
BREAKER IN P
D
6 BREAKERS (CB 1 – 6) in XG cabinet
D
4 REAR BREAKER SWITCHES (SW 1 – 4) and 1 BREAKER (NFB) in Gantry
.D.B (Power Distribution Box) for hospital supply
(NFB is only for System with T
able P9155AF)
2234976
breakers
D SER
D
D
Illustration 1–1
VICE SWITCH in Gantry
SYS–OFF–MNL SWITCH in Gantry
2 BREAKERS (NFB1, 2) in OC
XG Cabinet
CB6
CB1∼CB5
TM1TM2
SYSTEM1–1
Page 20
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
1-1
CHECK PRIOR T
Illustration 1–2
O APPL
Gantry
2234976
YING POWER (continued)
Servo Amp
Sub Board
TGP Board
Service Switch
(SYS/OFF/MNL)
Service Switch
DANGER!!
FATAL COLLISION HAZARD!!
WHEN ‘CONTROL’ SWITCH LOCATED ON THE GANTRY REAR BASE IS
TURNED ON (WITH THE ‘SCAN’ SWITCH ON), THE GANTRY WILL ROTATE
DUE TO INITIALIZATION.
Note
The ‘Scan’ and ‘TBL/Tilt’ switches must be turned ON before turning ‘Control’ switch ON. The
switches
turning ON sequence is as follows:
‘Scan’ ON ––> ‘TBL/T
ilt’ ON ––> ‘Control’ ON
The ‘DAS/Det’ switch ca be turned ON any time.
SYSTEM1–2
Page 21
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
1-1
CHECK PRIOR T
Illustration 1–3
O APPL
YING POWER (continued)
Operator Console
2234976
NFB1, NFB2
SYSTEM1–3
Page 22
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
1-2
GROUND CONTINUITY CHECK
V
erify that you removed the four(4) tilt lock plates and pulled out the Gantry azimuth lock pin.
1.Review illustration xxx, SYSTEM GROUNDING DIAGRAM schematics in the Installation Manual (P9018AS),
and
check that all connections are mechanically secure at both ends.
2.
Use a multimeter of equivalent device to measure the DC resistance between the following components in the
System, and verify that the resistance of each is less than
During the installation and start–up of the system you must take continuous temperature checks to make sure the
environment meets specifications listed in Section 3 of the Preinstallation manual (P9018AR). Excessive temperature,
coupled with reduced air flow from clogged filters, may overstress components and cause failures.
Filters
Check
and clean all electronic equipment dust filters. GE recommends initially doing this on a weekly basis, because
new installations
nance
interval as site environment improves.
Specifications
During installation, use the temperature/humidity recorder and thermometer to record the temperature and relative
humidity
tions.
in the scanner room. Make sure the data you collect meets the following temperature and humidity specifica
D Scanner
always have dust and dirt present at this stage. Over a period of time, you may increase the mainte
Room
–T
emperature :
20°C ∼
28
°C
2234976
-
-
Action
Contact
tions.
–
Relative Humidity :
D Operator’
the air conditioning supplier or your local GE service department when your data fails to meet any specifica
s Room and/or Equipment Room
–T
emperature :
–
Relative Humidity :
40% ∼ 70% (Non Condensing)
15°C ∼
20% ∼ 80% (Non Condensing)
30
°C
-
SYSTEM1–5
Page 24
GE MEDICAL
REV
0
1-4
POWER LINE VOL
Input Selection of 200V or 208V
V
erify that the following connections in X–Ray Generator meet the input power voltage.
SYSTEMS
FUNCTIONAL
TAGE
CT SYTEC/SYTEC I SERIES
CHECK /
ADJUSTMENT
2234976
Illustration 1–5
TM3TM4
200VC208V200VC208V
TM3TM4
200VC208V200VC208V
V
oltage Selection
For 200V System
For 208V System
SYSTEM1–6
Page 25
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
1-4
POWER LINE VOL
The
following assumes that the power source for the system was checked for mean voltage, as well as any line
disturbance
Instrument
Use
the same instrument (Dranetz, etc.) to monitor line voltage during this check that you used during the preinstalla
tion
check. The recommended line analyzer is
or Dranetz Model line monitor plug in modules. Y
mean voltage, surges, sags, short term impulses, and frequency drift.
Check Using an Instrument
Monitor the input power lines (200V
during preinstallation site checks.
D V
erify that the current line voltage equals the voltage measured during the preinstallation check.
D Verify
that mean line voltage does not fall outside of equipment specifications during the day to day opera
tion
of other site equipment during the installation period.
TAGE (continued)
AC or 208V
the Dranetz Model 606–3 (3 channel) with the 101 frequency option,
ou can use a
AC) to the System at the System. Check in particular:
similar analyzer
, as long as it can monitor and report
2234976
voltage
-
-
D Look
From the printouts, ensure that the power supply meets the following criteria :
Check Using an Oscilloscope
Monitor
between the oscilloscope’
harmonics of the third, seventh, and eleventh derivative will be the most prevalent.
Action
Investigate any condition that falls out of spec. Identify the problem source(s) and resolve the problem.
the power lines with an oscilloscope and verify that no significant distortion, noise or transients exist. Switch
for any line voltage disturbances and check for a correlation between disturbances and system mal
functions
D
Line V
D
Daily V
D Frequency :
such as “Aborts” or component failures.
oltage :
oltage V
ariation :
s long and
200V
AC (50 Hz/60 Hz) (Normal Steady State)
208V
AC (60 Hz) for U.S.A. (Normal Steady State)
+10% to –5% from nominal steady state (50Hz)
+6% to –9% from nominal steady state (60Hz)
50Hz ± 0.5Hz, 60Hz ± 0.5Hz
short
time base to monitor for spikes, as well as line drift over time. Generally
-
,
SYSTEM1–7
Page 26
GE MEDICAL
REV
0
1-5LOAD DISTRIBUTION
Before
attempting this power application and functional check,
instructions
SYSTEMS
given in ‘Checks prior to Applying Power
ELECTRIC SHOCK HAZARD !!
DANGEROUS
TO
PERSONNEL AND EQUIPMENT
VOLTAGES EXIST WITHIN THE UNIT
.
make sure that the breakers and switches match the
.’
WARNING!
, USE EXTREME CARE TO A
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
VOID INJUR
ADJUSTMENT
2234976
Y
Follow this procedure
Follow this sequence to protect you and the equipment from irreparable damage.
1.Preparation
a.
Remove the following covers:
D XG Covers (front and rear)
D OC Covers (front and rear)
D T
able Bottom Cover
D
Gantry Left side Cover
b. V
erify again that all breakers and switches are OFF according to ‘Checks prior to Applying Power
c. V
erify that wiring and cabling are correctly connected.
d.
2.
Switch ON the breaker on the power distribution box (hospital supply).
3.
Measure 200 (or 208) V
It should be
4.
Switch ON the ‘TRANS’ (CB1) breaker on the XG.
exactly
200 (or 208) ± 20V
in the order it was written. This procedure was developed with your safety in mind.
AC between terminals 1 and 2 (or 2 and 3) on TM1 on the XG.
AC.
.’
V
erify that the ‘Power
5.
Switch ON the ‘DAS’ (CB3) breaker on the XG.
6.
Measure 100V
7.
Push the ‘Power
Listen for the sound of contactor K1, K2 and K3 actuating in the XG. Check that the plunger indicators on K1,
K2
and K3 are pulled into the relay bodies.
AC between terminals 6 and 7 on TM2 on the XG. It should be
’ lamp in the front upper left corner of the XG turns ON.
’ switch on the OC.
100 ± 10V
AC.
SYSTEM1–8
Page 27
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
1-5LOAD DISTRIBUTION (continued)
8.Switch
9.Push
On the ‘OC’ (CB5), ‘TG1
voltages:
D 100 ±
D 115 ± 11.5VAC
D 200
the ‘Power
10V
AC
(or 208) ± 20V
’ switch on the OC. V
15’ (CB4) and ‘TG200/208V’ (CB2) breakers on the XG, and
.
. . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . .
AC
.
. . . . . . . . . . . . . .
FATAL COLLISION HAZARD !!
WHEN BREAKERS LOCATED ON THE GANTRY REAR BASE ARE TURNED
ON, ALWAYS TURN THE SWITCHES ON IN THE FOLLOWING ORDER.
‘SCAN’ ON → ‘TABLE/TILT’ ON → ‘DAS/DET’ ON → ‘CONTROL’ ON
(CONTROL SWITCH MUST BE TURNED ON LASTLY.)
check the following
between terminals 10 and 1
between terminals 4 and 5 on TM2 (for TG1
between terminals 1 and 2 (2 and 3) on TM2 (for TG200/208V)
erify that the K1, K2 and K3 are turned OFF
1 on TM1 (for OC)
15V).
.
DANGER!!
2234976
FAILURE TO HEED THIS WARNING MAY LEAD TO GANTRY ROTATION AND
RESULT IN PERSONAL HARM, HARM TO OTHERS OR DEATH.
10. Switch
ON all breakers and switches according to ‘Checks prior to Applying Power
FATAL COLLISION HAZARD !!
VERIFY
THA
T NO PEOPLE ARE IN THE SCAN ROOM. THE GANTR
AUTOMATICALLY IMMEDIATELY AFTER THE OC POWER SWITCH IS
TURNED
ON. WHEN THE GANTR
CAN BE CAUGHT IN THE GANTRY. FAILURE TO HEED THIS WARNING MAY
RESULT IN PERSONAL HARM, HARM TO OTHERS, OR DEATH.
11. Push
the ‘Power
V
erify that the ‘OC Power
V
erify that the 16 Fans rotate properly
’ switch on the OC.
.’
DANGER!!
Y ROT
Y COVERS HA
’ lamp turns ON, and ‘System Preparation’ is displayed on the CR
.
VE BEEN REMOVED,
ATES
PEOPLE
T monitor
.
D Six(6) Fans
D Nine(9) Fans
D One(1) Fans
.
. . . . . . . . . . . . . . . . . . . . . . . . .
.
. . . . . . . . . . . . . . . . . . . . . . . .
.
. . . . . . . . . . . . . . . . . . . . . . . .
upper rear part
bottom side of the nest chassis
behind power supplies
SYSTEM1–9
Page 28
GE MEDICAL
REV
0
1-6
EMERGENCY OFF TEST
1.
Press the emergency switch on the OC.
Verify
tector).
SYSTEMS
that contactor K1, K3 and K4 in the XG turn OFF
FUNCTIONAL
, and cut the power to the T
CT SYTEC/SYTEC I SERIES
CHECK /
able/Gantry (except DAS/De
ADJUSTMENT
2234976
-
2.Press
There are two emergency switches on the Gantry and one emergency switch on the XG. Perform this emergency
OFF
1-7FILTER
There are on Filter in the OC and two in the XG.
1.
2.
3.
4.
1-8CUST
the ‘Power
test on each on the three emergency switches.
D
OC : Horizontal. one on the bottom of OC –– housed within a gold mesh filter enclosure.
D
XG : V
Slide the filter at the bottom of the OC forward until it is removed from the OC.
Remove the front and rear covers of the XG and gentry pull the the foam out of each enclosure.
Suck up any dust on the filters with a vacuum cleaner
Slide or fit each filter back into its enclosure.
OM P
’ switch twice to reset emergency OFF
ertical, one each on the front and rear cover –– foam held in place by a white mesh enclosure.
ARAMETER
, and re–apply power
.
.
Run
the Custom Parameter program to set the correct system configuration. Y
you
change the system configuration (adding the option, etc.).
Refer to Functional Check/Adjustment manual, System, Section 2.
Note
No image data must exists on the system hard disk if you set or change the following parameters.
–
Optical Disk Option
–
Magnetic T
– Frequency
–
Machine Number
–
Hospital Name
BOOT floppy must be in the floppy disk drive to change the parameters.
ape Option
Note
ou must also
run this program any time
SYSTEM1–10
Page 29
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
1-9X–RA
Check the overheat detection function by performing the following:
1.
2.
3.
4.
5.T
Y TUBE OVERHEA
Switch OFF the ‘Scan’ switch at the Gantry rear base
Remove the both side covers from the Gantry
Disconnect the rotor cable.
If the connector is not accessible, perform the following:
a.
Set the Service switch to ON on the SUB Board.
b.
Rotate the Gantry until the X–ray tube reaches the 3 o’clock position.
c.
Set the Service switch to OFF
Switch ON the ‘Scan’ switch.
ry to perform a scan.
V
erify that a scan can not be performed and the “XG Fail” error message is reported on the status screen.
T SAFETY
.
.
2234976
V
erify that the ‘OH’ LED on the Power CONT Board in the XG is illuminated.
6.
Switch OFF the ‘Scan’ switch.
7.
Reconnect the connector
.
SYSTEM1–11
Page 30
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
1-10
SCAN TEST
This
check assumes successful completion of all functional checks for XG, Gantry T
tec
i series Operator Manual for scanning operation instructions.
DANGER!!
X–RAY EXPOSURE !!
IN
THIS SECTION, THE X–RA
THAT
Warm
Up Scan
1.
Make sure system power is ON an the ‘New Patiant’ and the ‘W
NO PEOPLE IS IN
‘W
arm Up’ should blink just after power up or system reset.
THE SCAN ROOM WHENEVER SCAN IS PERFORMED.
Y EXPOSURE IS PERFORMED. ALWAYS VERIFY
arm UP’ arear are blinking on LCD screen.
Note
2234976
able and OC. Check CT Sytec/Sy
-
2.
Make sure mothing is in the scan arear in the Gnatry room.
3.T
ouch the ‘W
Make sure all warm up scans complete without error
aum Up’ prompt to start the tube warm up program.
.
SYSTEM1–12
Page 31
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
1-10 SCAN TEST (continued)
Scoutview Scan
4.
Perform a series of scout scans to check scout scan operations:
Use the following scan parameters:
FOV
mA
Tilt
a.
Make sure that each scan is completed successfully
b. T
ilt the Gantry to any angle.
Make sure the software prohibits a scout scan with the Gantry tilted.
c.
Release the ‘Scan ’ button during a scout scan.
V
erify it aborts X–ray radiation, rotor rotation and cradle movement.
: 42 cm
: 40, 60 or 80 mA
: 0
°
2234976
.
d.
Select Auto V
Lay the water phantom horizontally on the cradle.
Perform an AP (0°) scout scan on the phantom.
Make sure that you can hear and understand each voice.
V
erify that the phantom diameter (In–Out direction) on the image is
e.
Select Auto V
Set the QA phantom on the phantom holder
Perform a L
Make sure that you can hear and understand each voice.
Release the ‘Scan’ button before the whole image is displayed and verify that the scan continues.
Push the ‘Abort’ key and verify that the cradle and scan stop.
oice S1 and Auto V
oice S2 and Auto V
T (90°) scout scan on the phantom.
oice E1.
oice E1.
.
266mm ± 3mm
.
SYSTEM1–13
Page 32
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
1-10 SCAN TEST (continued)
Localize
5.
Perform a Localize function:
a.
Display a scout image (use the image acquired at step 4).
b.
Select the Localize function.
c.
Prescribe and confirm an axial scan plan.
Make sure theat the ‘T
d.
Press the ‘T
Check that the cradle advances to the prescribed position.
Axial Scan
able Set’ button.
2234976
able Set’ button blinks.
6.
Perform an axial scan to check axial scan opreation:
Use the following scan parameters:
Interval:+10.0mm
Scan T
a.
Select each of the Auto V
V
erify that each scan completes with no error occurred.
V
erify that you can hear and understand each voice.
Recon Time
7.
Perform an axial scan using the following scan parameters:
erify that the time between the scan end and the image appearance on the CR
T monitor is within
15 sec
SYSTEM1–14
.
Page 33
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
1-10 SCAN TEST (continued)
KV and mA Meter
8.
Perform a series of axial scans (mA : 40, 60, 80, 100, 130 and 160).
V
erify that the KV meter on the XG read
Also verify that the mA meter on the XG reads the expected value ± 5%
Abort
9.
Perform an axial scan and press the ‘Abort’ button before the scan completes.
V
erify that the scan is terminated.
120KV ± 3KV
.
2234976
.
SYSTEM1–15
Page 34
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
blank
2234976
SYSTEM1–16
Page 35
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-1GENERAL
Save
existing patient data any time you perform the LFC procedure or you will lose this data.
The saving procedure is described in section 2-3 Pre–ATINST Procedure, Saving Customer
Data.
For DC–X, the system must be running on at least version 5.00 software. If the customer ’s
DC–X
system is older than version 5.00, perform FMI (FM91015 or 91016) must be performed
(version
2234976
SECTION 2 – SOFTWARE LOADING
NOTICE
NOTICE
5 upgrade) first.
This
section describes the
Disk)
to system hard disk. Perform this LFC procedure when:
D A failure or crash corrupts the hard disk so the existing software can not be used. A Load–From 0Cold
rebuilds
D CT Engineering has revised the current software to incorporate improvements or correct existing prob-
lems.
Replace the old software by performing a Load–From–Cold with the new software.
Load–From–Cold (LFC) procedure for a full system software loading from some FD (Floppy
the operating software on the disk and restores normal system operation.
SYSTEM2–1
Page 36
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-1GENERAL (continued)
Illustration 2–1
•
‘Acq Reverse’ Data
•
Communication Data
•
Ethernet Data
•
Station Data
•
Raw Data Files
•
Patient Image Data
•
Error Message Help
•
Record the Custom Parameters –––––
Load–From–Cold Standard Procedure
Start
Save Customer Data
– Start Study No.
– Direction of Scout Image (Apex / Lateral)
– Gantry Direction from Operator
Set CPW Boot Dip Switch
Insert Boot Floppy
Pull Reset Switch
A
TINST Program
2234976
•
Data File Save –––––––––––––––––––––
•
Disk Initialize
•
Software Loading (Standard, Option)
•
File Allocation
•
Data File Loading
Starting Up System
•
Remove Floppy Disk
•
Reset CPW Dip Switch
•
Pull Rest Switch
Restore Custom Parameters
•
Including Additional Option Software Loading
•
Custom Parameter Change
(It is necessary to manually restore the parameters due to a software bug.)
– Start Study No.
– Direction of Scout Image (Apex / Lateral)
– Gantry Direction from Operator
Restore Customer Data
•
‘Acq Reverse’ Data
•
Communication Data
•
Ethernet Data
•
Station Data
•
Raw Data Files
•
Patient Image Data
•
Error Message Help
– Calibration Files
– Station No. Information
– Preset Patients Information
– Scan Parameter
– System Parameters (Custom Parameters, etc...)
End
SYSTEM2–2
Page 37
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-2
FD FOR SYSTEM/OPTION SOFTW
Prepare the following, prior to starting the LFC procedure.
Normally, option software disks and also some application software disks included in the system software (‘PG00’
disk,
‘PZ00’ disk, ....) are
if the A
TINST program (described later) failed, or when the system hard disk drive is replaced with a new one.
Functional Image
: The system includes the software.
: The system does not include the software.
not required in a version up or a re–loading procedure. However
fff
, these disks are required
SYSTEM2–4
Page 39
GE MEDICAL
REV
0
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
2234976
2-3PRE–A
Save Customer Data before proceeding. You must load Customer Data onto floppy diskette, tape or optical disk
cartridge because the LFC (Load–From Cold) procedure destroys any information on the system disk. Refer to the
Operator Manuals to complete the following steps.
1.
Record ‘Acq Reverse’ user setting parameter
a. After displaying an image on the CRT screen, write down ‘Acq Reverse’ data (ON or OFF) in ‘Set
2.
Save Communication Data.
For any system on which ELINK is installed, save Communication Data, Station No. and Ehternet Data on a
floppy
a.
TINST PROCEDURE
Note
Alphanumeric letters you must enter at the keyboard are underlined. ‘CR’ means press the ‘Enter
.
function.
Note
Usually
matically in the A
be
Press the [Mgt] key
the parameter setting data does not need to
TINST program (Data File Save Function). However
recorded due to a software bug.
disk. Otherwise, these data will be deleted during LFC.
.
be recorded, since the data will be saved auto
, only ‘Acq Reverse’ data
must
’.
Display’
-
D
Select Communication Data Load/Save → Save
D
Select Ethernet Data Handling → Save to FD
D
Select Dump Parameters → Station Number Information File
Note
When saving all information above, use on floppy disk.
SYSTEM2–5
Page 40
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-3PRE–ATINST PROCEDURE (continued)
NOTICE
If
you intend to save raw data on MOD, e sure to start with a blank cartridge, or one that does
not
need to be preserved. This procedure erase all previous data contained on
tridge.
3.Save
Raw Data Files:
Archive
wise,
Use the ‘Maintenance’ function to save the Raw Data as follows:
a.
b.
c.
d. V
Raw Data Files to a MOD, floppy
these files are destroyed during LFC.
Select
Select
Record the ID number of the Raw Data Files which you want to save.
Support
FD Dump/Load, MT Dump/Load
erify that the MOD should be set “write enable”.
from the Maintenance menu.
disks, tape, or optical disk if the customer wants to save them. Other
, or
MOD Dump/Load.
the MOD car
2234976
-
-
e.
Select
f.
Select
g.
Enter all the Raw Data File ID numbers. Then press the [End] key
4.
Save Patient Image Data:
Archive Patient Image Data to a MOD, floppy disks, tape, or optical disk if the customer wants to save them.
Otherwise,
Data. (Refer to Operator Manual.)
5.
Check if the service maintenance menu includes ‘
If
the system contains ‘Error Message Help’ function, it must be reinstalled since version–up or reloading proce
dures
Initialize
Dump (Disk –> MOD)
these files are destroyed during LFC. Use the ‘Image File Management’ function to save the Image
delete it.
to initialize the MOD.
.
.
Error Message Help
’.
-
SYSTEM2–6
Page 41
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-3PRE–ATINST PROCEDURE (continued)
6.
Record Customer Parameters.
Usually the Customer Parameter Data does not need to be recorded, since the data will be saved
automatically
rameters
D
Start Study No.
D Direction of Scout Image (Apex)
D
Direction of Scout Image (Lateral)
D
Gantry Direction from Operator
Also,
it is recommended to check which option software are currently installed
you
can confirm later that those option software have been correctly retrieved. In addition, if you nave
all
the records, they con serve as backup in the event there is trouble with the A
a.
Select
in the A
must be recorded due to a software bug.
Custom Parameters
TINST program (Data File Save Function). However
from the Maintenance menu.
Note
Note
, at least the following pa
on the system, so that
TINST program.
2234976
-
b.
Select
Custom
ters.
Custom
Therefore, enter the
Display Parameters
Parameters will be displayed
Parameters include Function Menu which noes not show in the Display Parameters mode.
.
one by one, as listed in T
Change Parameters
able 2–3. Use this table to record the parame
mode and record the parameters in T
-
able 2–4.
SYSTEM2–7
Page 42
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-3PRE–ATINST PROCEDURE (continued)
T
able 2–3
Current Model Name
Auxiliary
Optical Disk
Magnetic Optical Disk
Magnetic T
Frequency
Machine No.
Hospital Name
Start Study No.
Disk Size
ape
Custom Parameters (Display)
Unavailable . A
Unavailable . A
Unavailable . A
Unavailable . A
50 . 60
VED REFORMA
BMD
COORDINATE
CALCULATOR
DENTA
DISPLA
NETWORK
Y IN DISPLA
Custom Parameters (Display) (continued)
TION
TION
Y
Option
Unavailable . A
Unavailable . A
Unavailable . A
Unavailable . A
Unavailable . A
Unavailable . A
Unavailable . A
Unavailable . A
Unavailable . A
Unavailable . A
Unavailable . A
After terminating the A
The ATINST program will start as ‘Reloading Mode’ or ‘First Loading Mode’, depending on when the program failed.
After the ATINST program is successfully finished, restore the following data files in that order, if they were saved.
Loading
order is very important.
Calibration Files
Station Number Information
Preset Patient Information
Series Description
Scan Parameters
Option software which the customer
InSite software (if system had)
Retries the program in which the error occurred.
TINST
TINST program by selecting ‘1’ in the menu, retry A
TINST program.
TINST” to resume A
’s system contains
TINST.
TINST.
SYSTEM2–12
Page 47
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-4ATINST (continued)
Operating ATINST
It will take approximately 40 minutes for the A
TINST program to complete all its operation.
NOTICE
Save existing patient data any time you perform a LFC procedure or you will lose this data.
The
saving procedure is described in ‘Pre–A
Prepare all of the floppy diskettes listed in T
1.
Set the SW2 and SW4 of the “BOOT” switches on the CPW Board (Slot #1) to the ON option.
Illustration 2–2
Boot Switch Setting on CPW Board
ON
TINST’ procedure.
Note
able 2–1 and T
able 2–2 before starting this procedure.
2234976
1 2 3 4
OFF
2.
Insert the BOOT floppy into the disk drive.
3.
Pull the reset switch located under the right side of the Operator Console.
The system is reset and “>” will appear on the monitor
> ATINST ‘CR’
Delete IMAGE & RAW Data ? <0:OK, 1:Exit> = 0 ‘CR’
** INSTALLATION START **
Please Insert <<KEY0_[ x]>> FD and Press <ENTER>
************************************************************************
TINST program by entering A
TINST:
.
SYSTEM2–13
Page 48
CT SYTEC/SYTEC I SERIES
GE MEDICAL
REV
0
2-4ATINST (continued)
5.
Remove the BOOT floppy from the floppy drive and insert KEY0 floppy supplied with the system.
SYSTEMS
FUNCTIONAL
CHECK /
ADJUSTMENT
2234976
Press the <Enter> key
******************************* Displayed *******************************
Please Insert FD for STORE and Press <ENTR>
************************************************************************
Always
prompted
The following operations are carried out automatically in the order described:
D
D
D
D
D
If a non–initialize floppy diskette is inserted, it will be initialized automatically before saving data.
use either blank floppies to store Data Files, or answer “Y
during ST
Saving Data files
Disk Initialize
Software Loading (Standard, Option)
File Allocation
Data File Loading
.
ORE.
Note
Note
es” to “Initialize OK
?” question when
6.
Save Data Files:
Remove
the scratch
the
7.
Initialize Hard Disk:
Remove the floppy diskette that data files are saved on and insert “BOOT” floppy diskette.
Press the <Enter> key
the KEY0 diskette, then insert the scratch floppy diskette with No.1 written on it. Y
floppy
disks in order according to screen messages to save data onto these disks. Be sure to number
floppies in the order used (1–4) for later reloading.
Note
If a non–initialize floppy diskette is inserted, it will be initialized automatically before saving data.
. This starts VINIT program so that the hard disk will be initialized automatically
** DFRRV End **
************************************************************************
.
.
SYSTEM2–15
Page 50
CT SYTEC/SYTEC I SERIES
GE MEDICAL
REV
0
2-4ATINST (continued)
10.
Data File Loading
Remove the “BOOT” floppy diskette and inset No.1 floppy diskette that includes the parameters saved during
“Saving
11.
Replace the No.1 floppy diskette with No.2 and press the <Enter> key according to screen messages. Repeat
this procedure until all parameters are loaded.
12. After all FDs have been loaded, press the <End> key to terminate the Data File Loading. Then the following
screen
SYSTEMS
Data files” procedures at the first stage of A
will appears.
******************************* Displayed *******************************
** INSTALLATION END **
*************************************************
* We will complete all works for installation *
* Put back the Dip Switch (1–4:OFF)*
* and pull the Rest Button.*
* Don’t Forget other operations as follows:*
* (*) Retrieve Image & Raw Data*
* (*) Change CUSTOM PARAMETER if necessary *
*************************************************
the system reset switch located under the right side of the Operator Console.
.
.
SYSTEM2–17
Page 52
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-5POST–ATINST PROCEDURE (continued)
Restoring Custom Parameter
Use the CHANGE P
to change the custom parameters;
to load additional option software;
to change the number of raw data files.
The
custom parameter data will be restored automatically in the ATINST program (Data File Loading
Function). However, regarding the following, it is necessary to manually restore due to a software
bug:
1.
Insert the BOOT floppy into the disk drive.
ARAMETERS function of the CUST
– Start
–
–
–
Study No.
Direction of Scout Image (Apex)
Direction of Scout Image (Lateral)
Gantry direction from Operator
OM P
ARAMETERS program for the following purposes:
Note
2234976
2.T
ouch the [Maint] key and enter $$
3.
Select
Custom Parameters
4.
Select
Change Parameters
If a parameter is not required to be changed, press the [END] key to proceed to the next item.
V
erify or change the custom parameters which are displayed one by one as listed T
When installing addition option software, the CUST
can
not be loaded using TCOPY
5.
Remove the BOOT floppy diskette.
6.
Pull the system reset switch located under the right side of the Operator Console.
7.
Check that the system start up normally
8.Using
‘Display Parameter
are
displayed as “Installed”.
’ of the CUST
to display the Maintenance menu.
from Maintenance menu.
Note
OM PARAMETER menu must be used. Note that software
.
.
OM P
ARAMETERS menu, check if all the optional software you installed
able 2–5.
SYSTEM2–18
Page 53
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-5POST–ATINST PROCEDURE (continued)
T
able 2–5
Set Model Name
Optical Disk
Magnetic Optical Disk
Magnetic T
Frequency
Machine No.
Hospital Name
Start Study No.
Scout Image Orientation
Gantry Direction from Operator
Axial Image Orientation (from)
ape
Custom Parameters (Change Mode)
This parameter does not appear if the hard disk contains images.
This parameter does not appear if the hard disk contains images.
see Note1
see Note2
see Note3
Press [END] to proceed to next item.
2234976
Function Menu
Thickness 2/3 mm
Option
InSite Password
No. of Raw Data Files
Note1When
creased
Note2
rewritten
has
Note3When
be rewritten against the original data. The indication visible in the display might be the same as the original one, but
data has been rewritten. So, enter the original data recorded during ‘Pre–A
rameter will not appear if ‘Gantry’ is selected as the kind of BASE IN SCOUT IMAGE ORIENTATION (Lateral).
by 1. So, enter the original No.
against the original data. The indication visible
been rewritten. So, enter the original data recorded during ‘Pre–A
‘Start Study No.’ is automatically saved and loaded by A
When ‘Scout Image Orientation’ is saved and loaded automatically
‘Gantry Direction from Operator
see Note4
see Note5
see Note6
Password for InSite can be changed.
see Note7 This parameter does not appear if the hard disk contains images.
TINST
, the original
. Using A
in the display might be the same as the original one, but data
TINST Procedure’ to New Data.
’ is automatically saved and loaded by A
TINST Procedure’ to New Data. This pa
study number will be in
TINST
the default data will be
TINST
, the default data will
-
-
SYSTEM2–19
Page 54
GE MEDICAL
REV
0
2-5POST–ATINST PROCEDURE (continued)
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
2234976
Note4Function
******************************* Displayed *******************************
** FUNCTION MENU
the number of vacant images on the system HD is decreased to below the number
(4. MIN V
3. AUT
(and not the retrieved ones) are automatically deleted from the HD to reserve the specified image space (4.
CANT
study
to be reserved.
4. MIN V
on the HD becomes equal to this specified number, the
is displayed on the Status screen.
ACANT IMAGE SP
O DELETE OPTION
IMAGE SP
No. on the HD through to the study which will allow the specified space (4.MIN V
ACANT IMAGE SP
The following menu with the current settings is displayed:
CURRENT
an OD and MOD are equipped in the
system)
ORE OPTIONAUTO:
ACE FOR SCAN).
ON:
When the AUT
ACE FOR SCAN). Delete operation
ACE FOR SCAN
Images are automatically stored to the OD (AUT
O ST
ORE function in enabled,
is performed on files starting from the study having the smallest
During
an image retrieve operation. If the number of vacant images
retrieve operation is terminated and a “DK FULL” message
or AUT
or MOD; This appears only if both
O)
or ORIGINAL)
or ORIGINAL)
or ORIGINAL)
studies which are already stored
ACANT IMAGE FOR SCAN)
O ST
ORE DEVICE),
MIN VA-
Note5Thickness
is
adjusted to 2mm, “2” must be selected.
Note6Enter
ware
options you want to install are loaded. The software options loaded will be indicated by “Installed”.
Note7If
appropriate
are
many bad sectors on the disk, a “T
number
ate.
the number of raw data files needs to be changed, select “0” and
value in response to “Program Area”. It will take several minutes to complete
of Raw Data to be entered appears on
can be changed only for CT Sytec, 3000i, Plus, and 4000i. The default setting is “3”. If the system
the appropriate No. of the option software you want to install. Repeat this procedure until all the soft
press the [ENTER] key
data file reservation. If there
oo Many Bad Sector” message and the new appropriate (maximum possible)
the monitor
. Enter the new number again if it is considered as appropri
-
. Then enter the
-
SYSTEM2–20
Page 55
GE MEDICAL
REV
0
2-5POST–ATINST PROCEDURE (continued)
Restoring Customer Data
1.
Set ‘Acq Reverse’ user setting parameter
SYSTEMS
.
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
2234976
a. After
2.
Restore ‘Communication Data’.
For any system on which ELINK is installed, restore Communication Data, Station No. and Ehternet Data from
the floppy disk.
a. Press [Mgt] key.
b.
displaying an image on the CR
Select
The data has been saved using different menu in ‘Saving Customer Data’ procedure. However, in
version
Communication Data Load/Save
3 system, you have only to restore them using ‘Communication Data Load/Save’ menu.
T screen, set ‘Acq Reverse’ data (ON or OFF) in ‘Set Display’ function.
.
Note
SYSTEM2–21
Page 56
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-5POST–ATINST PROCEDURE (continued)
3.
Restore Raw Data.
If
you saved raw data files prior to performing the A
function
a.
b. T
c.
d.
to retrieve raw data.
Insert the BOOT floppy into the disk drive.
ouch the [Maint] key and enter $$
Select
Select
(Example: Lading raw data files from MOD)
******************************* Displayed *******************************
** RAW DATA DUMP/LOAD (MOD) **
1.DUMP (DIDK –> MOD)
2.LOAD (MOD –> DISK)
3.DIRECTORY
4.INITIALIZE
Select No. = 2 ‘CR’
SUPPORT
FD DUMP/LOAD, MT DUMP/LOAD
from the Maintenance menu.
to display the Maintenance menu.
2234976
TINST procedure, restore them to the disk. Use Maintenance
, or
MOD DUMP/LOAD
.
*** LOAD ( FROM MOD TO DISK ) ***
Volume Name = ORAW
Enter MOD START FILE No. = 0 ‘CR’(Enter
*** RAW DATA FILE SELECT ***
** FILE No. = 0
Enter SERIES No. = 1 ‘CR’
** FILE No. = 1
Enter SERIES No. = 2 ‘CR’
************************************************************************
The
MOD raw data files will have the following ID No. on the disk.
0:0.XX.1;1 (ST
The file No. is updated and the system waits for enter of another file Series No.
Restore Patient Image Data.
If you archived patient image data prior to performing the ATINST procedure, restore it to the disk. Use Image
File Management function to retrieve it (refer to Operator manual).
5.
Reinstall Error Help software (Class–C):
If the system contains ‘error Message Help’, reinstall the software since this version–up procedures delete it.
a.
b.
c.
SYSTEMS
Note
The new EH00 floppy disk must be prepared. Old software must not be reloaded.
Enter the Management menu by pressing [Mgt] key
Insert the BOOT floppy disk.
Enter $$
and “>” prompt will be displayed.
FUNCTIONAL
.
CHECK /
ADJUSTMENT
2234976
d.
Remove the BOOT floppy diskette and insert the EH00 floppy diskette.
Press the [END] key to skip a menu and repeat this until ‘FD CONTENTS’ is displayed.
f.
Remove the BOOT floppy diskette, and insert the CD00 floppy diskette.
ersion 3.02 System Software (continued)
TE software. if possible:
in ‘Restoring Custom Parameter
in this step.
to display the Maintenance menu.
Custom Parameters
Change Parameters
from Maintenance menu.
Note
’, the software above is installed, but they
2234976
MUST be installed
g.
Press the [END] key to skip a menu and repeat this until the Custom Parameter Change Mode is finished.
h.
Remove the floppy diskette.
3.
Pull the system reset switch located under the right side of the Operator Console.
4.
Check that the system starts up normally and the “VER. 03.02” and logo appear on the CR
T monitor
.
SYSTEM2–26
Page 61
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-5POST–ATINST PROCEDURE (continued)
2-5-2Loading V
This software version–up is quite different from normal version–up. Do not pull ‘System
Rest’
ware.
1.Verify
2.
that the system should be running on version 3.00 or later system software and
net
option software, if not, perform the upgrade to version 3.00 and install the Ethernet option software.
Save System Parameter:
a.
Insert the BOOT floppy into the disk drive.
b. T
ouch the [Maint] key and enter $$
c.
Select
ersion 4.00 System Software
until completion of the ‘Custom Parameter Change’ after copying the version 4.00
Otherwise, the system will no longer starts up.
to display the Maintenance menu.
System Parameter Dump/Load
NOTICE
from the Maintenance menu.
2234976
soft
-
the system has the Ether
-
d.
Select
e.
Remove the BOOT floppy diskette and insert the first scratch floppy diskette.
After
Parameter”)
3.
Save Communication Data:
a.
Before saving Communication Data, change ‘Interval T
i.
ii.
iii.
iv.
v.
b.
Insert a second scratch floppy diskette into the floppy disk drive.
c.
Select
System Parameter Dump
completion
and press the [Cancel] key to terminate the Maintenance menu.
Press [Mgt] key to display the Management menu.
Select Transmit Image
Select
Select
Press the [END] key to skip a menu and repeat this until Management menu is displayed.
Auto Transmission
Semiauto T
Communication Data Load/Save
.
of ‘System Parameter Dump’, remove the scratch floppy diskette (label this floppy “System
from the Management menu.
, and change the value of ‘Interval Time’ to 20sec.
ransmission
, and change the value of ‘Interval Time’ to 20sec.
from the Management menu.
ime’ of the ‘Auto T
ransmission’:
d.
Select
After
munication Data”) and press the [END] key to skip a menu and repeat this until Management menu is displayed.
Communication Data Save
completion of ‘Communication Data Save’, remove the scratch floppy diskette (label this floppy
.
SYSTEM2–27
“Com
-
Page 62
GE MEDICAL
REV
0
2-5POST–ATINST PROCEDURE (continued)
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
2234976
2-5-2Loading V
4.W
rite down the installed option software:
a.
Insert the BOOT floppy into the disk drive.
b. T
ouch the [Maint] key and enter $$
c.
Select
d.
Select
e.
Press the [END] key to skip a menu and repeat this until ‘FD CONTENTS’ is displayed.
f. Write
eters.
Table
Custom Parameters
Display Parameters
down the installed option software. Use the T
2–6
Option SoftwareV
PagingPG00
3D Imaging
Real–time Reformation
ersion 4.00 System Software (continued)
to display the Maintenance menu.
from the Maintenance menu.
.
able 2–6 to check and write down the site’
List of Option Software to Reinstall
olume Name
DIM3
RTRF
s Custom Param
Check Item
Installed . Not Installed
Installed . Not Installed
Installed . Not Installed
g. After completion of loading version 4.00 standard software: 2131960 floppy diskette 1/5∼5/5, remove the
V400–5/5
h.
Insert the following floppy diskette into the floppy disk drive.
i. T
ype the same command as in step f. to copy all files on the floppy diskette to the system.
floppy diskette.
FD Number
2131961
2131962
2131963
Model Name
Sytec 2000i, S
Sytec, Sytec 3000i
Sytec 4000i, Plus
SYSTEM2–30
Page 65
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-5POST–ATINST PROCEDURE (continued)
2-5-2Loading V
j. After
nate
k.
Remove the floppy diskette.
Do
not pull ‘SYSTEM RESET’.
perform the procedure outlined
Software and Set the Multi language Function’.
i.
ii.
iii.
iv.
ersion 4.00 System Software (continued)
completion of TCOPY
the Management menu.
Insert the ‘BOOT–[0] V/R 3.xx’ floppy diskette into the floppy disk drive.
Set the SW2 and SW4 of the BOOT switches on the CPW Board to the ON position.
Pull the reset switch located under the right side of the Operator console.
Enter the following at the “>” prompt:
, enter RT to return of the Management menu, and press the [END] key to termi
Otherwise, the system will no longer starts up. IN THIS CASE
below
2234976
-
NOTICE
,
, and then, follows to next procedure ‘Reinstall Option
> FN SYS:0.&.TCOPY.LO ‘CR’
> RU ‘CR’
SELECT No. = 2S ‘CR’
WHAT FILE NAME ?
SOURCE : SYS:0.ELNKFMP. ‘CR’
OBJECT : SYS:0.FMP. ‘CR’
DO YOU WANT SELECT ? N ‘CR’
IF FILE EXISTS, OVERWRITE OK ? Y ‘CR’
v.Remove
vi.
Set all bits on the BOOT SELECT switch on the CPW Board to the OFF position.
vii.
Pull the system reset switch.
the ‘BOOT–[0] V/R 3.xx’ floppy diskette.
SYSTEM2–31
Page 66
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-5POST–ATINST PROCEDURE (continued)
2-5-2Loading V
6.
Reinstall Option Software and Set the Multi–Language Function:
Although the ‘FD CONTENTS’ screen may show the options of T
reinstall those
a.
Insert the ‘BOOT–[0] V/R 3.xx’ floppy into the disk drive.
b. T
ouch the [Maint] key and enter $$
c.
Select
d.
Select
e.
Press the [END] key to skip a menu and repeat this until ‘FD CONTENTS’ is displayed.
f.
Remove the BOOT floppy diskette.
ersion 4.00 System Software (continued)
Installed
Custom Parameters
Change Parameters
options to enable the new features contained in version 4.00 software.
to display the Maintenance menu.
from the Maintenance menu.
.
Note
able 2–6 as
2234976
Installed, you MUST
g.
Enter one option No. which was checked as ‘
h.
Load the other potions, which were checked as ‘
i.
Press the [END] key to skip a menu and repeat this until ‘SET LANGUAGE TYPE’ is displayed.
j.
Enter the language number
On the Japanese system ‘Current Language = Japanese’ is displayed and this parameter can not
be
changed. Y
changed
Plus” in ‘SET MODE NAME’ of change mode of the ‘CUSTOM PARAMETERS’. In the same way,
when changing English or French system to Japanese system, select “PreSage” after deleting all
images
k.
Press the [END] key to skip a menu and repeat this until the following message is displayed.
** CUSTOM PARAMETER END
*********************************
* PULL THE SYSTEM RESET BUTTON *
*********************************
l. Remove
in the hard disk.
ou have only to press [END] key to skip this setting. If the Japanese system must be
to English or French, delete all image in the hard disk first. The select “Sytec 4000i or Sytec
the floppy diskette, and pull the system reset switch located under the right side of Operator console.
.
Installed
Installed
Note
’ in T
able 2–6, and load the option software.
’ in T
able 2–6, by repeating previous step.
m. Check
that the system starts up normally
, and version “4.00” and system logo appear on the CR
T monitor
SYSTEM2–32
.
Page 67
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
2-5POST–ATINST PROCEDURE (continued)
2-5-2Loading V
7.
Restore Communication Data:
a.
Press [Mgt] key to display the Management menu.
b.
Insert the ‘BOOT–[0] V/R 3.xx’ floppy into the disk drive.
c.
Select
d. Remove the BOOT floppy diskette, and insert the second scratch floppy diskette saved ‘Communication
Data’.
e. Select Load from FD.
After
8.Restore
ersion 4.00 System Software (continued)
Communication Data Load/Save
completion of ‘Communication Data
System Parameters:
from the Management menu.
Load’, press [Cancel] key to terminate the Management program.
2234976
NOTICE
It
takes approximately 5 minutes to load the FD. Do not misunderstand that
UP’
or failed.
a.
Insert the ‘BOOT–[0] V/R 3.xx’ floppy into the disk drive.
b. T
ouch the [Maint] key and enter $$
c.
Select
d.
Select
e.
Remove the BOOT floppy diskette, and insert the first scratch floppy diskette saved ‘System Parameters’.
After completion of ‘System Parameter Load’, press [Cancel] key to terminate the Maintenance menu.
f. Remove
g.
Check that the system starts up normally
tor.
System Parameter Dump/Load
Parameter Load from FD
the floppy diskette, and pull the system reset switch located under the right side of Operator console.
to display the Maintenance menu.
from the Maintenance menu.
.
, and version “VER. 4.00” and system logo appear on the CR
status is ‘HUNG
T moni
-
SYSTEM2–33
Page 68
GE MEDICAL
REV
0
2-5POST–ATINST PROCEDURE (continued)
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
2234976
2-5-3
Loading V/R 4.01 System Software
NOTICE
The CT Sytec / Sytec i series system must running on V/R 4.00.
Copy
one(1) version 4.01 floppy diskette (version 4.01 standard software: 2144661 floppy diskette 1/1) using TCOPY
SELECT No.1 2 3 4
1: COPY2:COPY AND VERIFY
3:VERIFY4:VERIFY WITH COMMENT
** <OPTION> **
N:NON–WRITE TO MT HEADER(ONLY FOR MT I/O)
S:STOP ON VERIFY ERRORS(EXCEPT SELECT NO.1)
T:TERMINATE ON VERIFY ERROR(EXCEPT SELECT N0.1)
EXAMPLE: SELECT NO.= 2NS
SELECT NO. = 2S ‘CR’
WHAT FILE NAME
SOURCE : V402 ‘CR’
OBJECT : SYS ‘CR’
DO YOU WANT SELECT ? N ‘CR’
IF FILE EXISTS, OVERWRITE OK ? Y ‘CR’
.
.
.
TCOPY:TERMINATED
>
using the following commands:
7.After
8.
9.Check
completion of
of
the Management menu, and press the [END] key to terminate the Management menu.
Remove the floppy diskette, and pull the system reset switch located under the right side of Operator console.
that the system starts up normally
loading
version 4.02 standard software: 2236432 floppy diskette 1/1
, and version “VER. 4.02” and
system logo appear on the CR
, enter RT to return
T monitor
.
SYSTEM2–37
Page 72
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
blank
2234976
SYSTEM2–38
Page 73
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
ROTATION HAZARD !
IN THIS ‘X–RAY ALIGNMENT’, GANTRY COMPONENTS ARE ADJUSTED AND SCANS ARE
PERFORMED. ALWAYS SWITCH OFF THE ‘SCAN’ SWITCH BEFORE SERVICING OR
ADJUSTING THE GANTRY. ALSO VERIFY THAT NO PEOPLE ARE IN THE SCAN ROOM
WHENEVER
3-1GENERAL
2234976
SECTION 3 – X–RAY ALIGNMENT
WARNING!
A SCAN IS PERFORMED.
This ‘X–Ray Alignment’ consists of the following check/adjustment procedures:
Plane of Rotation
Detector Beam–On–Window
Gravity SAG
Iso Center Alignment
Filter Center Alignment
Perform these procedures in the following order:
Plane of Rotation → Detector Beam–On–Window
→ Filter Center Alignment
The ‘Gravity SAG’ procedure is a check procedure for the Gantry rotating mechanism (bend, vibrations, etc.).
Perform this check if considered necessary
.
→ (Gravity SAG) →
Iso Center Alignment
SYSTEM3–1
Page 74
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
3-1GENERAL (continued)
Requirements
The critical components for X–Ray Alignment are all mounted directly to the rotating Gantry; they are:
D
X–Ray T
D
Collimator (includes a bow–tie filter and an adjustable beam defining exit slit)
D Detector
T
ools and materials required for X–Ray alignment:
D
Polaroid T
D
Polaroid T
D
Film Holder (Z9807NA and Z9807NB) for Detector Beam–On–Window alignment
D
G–SAG Pin (Z9807NH)
ube
ype 52 Film Packs
ype 45 Film Holder (developer for type 52 film)
2234976
Gantry Geometry
Theta :Tangent to Gantry rotation.
clockwise (+)
counterclockwise (–)
Z–Axis :
Note:
D
D
D
Polaroid Film Use
Use a Polaroid Film pack, a Polaroid film holder or squeeze roller for development as follows:
Parallel to the axis of Gantry rotation.
toward the T
away from the T
The tube unit can be adjusted in both the theta and
the Z direction.
The bow–tie filter in the collimator can be adjusted
only in the theta direction.
The detector can be adjusted only in the Z direction.
able (–)
able (+)
(–)
(–)
(+)
(+)
1.
Develop a Polaroid film pack with the film holder
2.
Break the chemical pod over a table edge, or use a squeeze roller ad follows:
a.
Apply heavy pressure to break the pod, and ease up slightly to spread the developer evenly ever the film.
.
SYSTEM3–2
Page 75
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
3-2PLANE OF ROTATION
This procedure checks and adjusts the plane of rotation of the X–Ray fan beam until it is perpendicular to the axis
of
Gantry rotation to within ± 0.75 mm across the largest Gantry aperture (63 cm).
1.
Switch OFF the ‘Scan’ and ‘TBL/T
2.
Remove the front and both side Gantry covers.
3.V
erify that the bow–tie filter is in the collimator
4.Attach
a Polaroid film at the Gantry 0 deg. position using adhesive tape (the side of
TOWARD
LEDS” should face the Gantry rear cover).
The
position of the film within the scan circle is critical for accurate results during this check/adjust
ment procedure.
If front and rear Gantry covers are removed, use the 42 cm PP phantom as follows
a. Raise the T
b. Set the 42cm PP phantom to the phantom holder
c. Attach the edge of the ‘Type 52’ film onto top part of the circumference of the phantom.
d. Move the cradle until the center to the film reaches under the collimator slit.
able to its highest position.
ilt’ switches at the Gantry rear base.
.
the film marked “THIS SIDE
Note
.
2234976
-
Mark a ‘T’ on the film to identify the T
5.
Switch ON the ‘Scan’ and ‘TBL/T
6.
Perform an axial scan with the following parameters:
Scan T
Thickness:
KV:
mA:
Remove the film pack.
Develop the film. W
ime
:3.6 sec
:4.5 sec
10 mm
120 KV
40 mA
ait the required film development time described on the film pack.
if the narrow exposure lies within the wider exposure, and the center lines of the two beams coincide within
≤
0.75 mm
> 0.75
WhereXF
Measure
–––> proceed to step 9.
mm –––> proceed to the next step 8.
: the front width not covered by the narrow beam
XR
: the rear width not covered by the narrow beam
Narrow and W
XF–XR
2
ide Exposure
TABLEGANTRY
XF
XR
Narrow Beam
(T
ube at the HOME position)
Wide Beam
(T
ube at 180 deg. position)
SYSTEM3–4
Page 77
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
3-2PLANE OF ROTATION (continued)
8.
Adjust the X–Ray tube Z–axis position:
a.
Rotate the Gantry until the X–Ray tube reaches the HOME position.
b.
Switch OFF the ‘Scan’ and ‘TBL/Tilt’ switches.
c.
Loosen the 4 mounting bolts.
d.
Locate the Adjustment Nuts for Z–Axis; note that there are two Z–Axis nuts for adjustment.
e.
Loosen the nut that faces the direction in which you want to shift the tube.
f. Turn
g. T
h. T
the other nut in the direction in which you want to shift the
D
One turn of the nut results in approximately 1mm on tube movement.
D
Use the micro dial gauge to help you adjusting the tube.
ighten the nut previously loosened in step e.
ighten the 4 mounting bolts.
Note
tube (the required tube shift is
1
7
2234976
XF–XR
).
i.
Switch ON the ‘Scan’ and ‘TBL/T
j.
Return to step 4.
9.
Proceed to Section 3-3 ‘Detector Beam–On–Window’.
Illustration 3–2
TableGantry
X–Ray T
Mounting Plate
ilt’
ube Positioning
Cathode
Mounting Bolts
(Socket Head)
Anode
Adjustment Nut
for Z–Axis
Mounting Bolts
(Socket Head)
Adjustment Nut
for Theta
SYSTEM3–5
Page 78
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
3-3DETECT
To
ensure the entire width of the X–Ray beam falls within the detector window
tion
check/adjustment of the detector
1.Preparation:
Y
ou do not have to perform the following, if you have just completed the ‘Plane of Rotation’ procedure.
a.
b.
c.
d.
2.
Set the film holders with Polaroid film to the Detector:
a.
b. Insert
OR BEAM ON WINDOW
, follow this procedure for a Z–Axis posi
.
Switch OFF the ‘Scan’ and ‘TBL/T
Remove the front and both side Gantry covers.
Remove the bow–tie filter from the collimator (refer to the Component Replacement manual).
Switch ON the ‘Scan’ and ‘TBL/T
Switch OFF the ‘Scan’ switch.
a Polaroid film to the two film holders with
and
the “INSER
T THIS END” forward the Gantry when placed on the Detector
ile’ switches at Gantry rear base.
ilt’ switches.
the side marked “THIS SIDE TOWARD LENS” facing down
.
2234976
-
Make
sure that the films are fully inserted into the holders, otherwise you may lose part of the image. One film
holder
is for the right side of the Detector
Each film holder has a side tab to position it to the window slot of the Detector, and two end tabs
to align it to the end of the Detector housing. A cutout in the front of the holder provides reference
marks
on the exposed film relating the center line of the Detector window and the X–Ray beam Z–
Axis position.
c.
Mount the two film holders with film onto both ends of the Detector
d.
Switch ON the ‘Scan’ switch.
Illustration 3–3
Film Holder Placement
, and the other is for the left side.
Note
.
Film
Film
SYSTEM3–6
Page 79
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
3-3DETECTOR BEAM ON WINDOW (continued)
3.
Perform a stationary scan on the films:
a.
Insert the BOOT floppy into the disk drive.
b. T
ouch the [Maint] key and enter $$
c.
Select
d.
Select
e.
In this menu, perform the following
SUPPORT
OFF LINE SCAN
Azimuth:
Scan T
Thickness:
KV:
mA:
(Others):(defaults)
from the Maintenance menu.
from the Support menu.
Present Degree
ime:3.6 sec
4.5 sec
10 mm
120 KV
80 mA
to display the Maintenance menu.
stationary scan:
(CT Sytec 2000i, S)
(CT Sytec 3000i, 4000i, Plus)
2234976
f.
Switch OFF the ‘Scan’ switch.
g. Remove
h. Mark
i.
Switch ON the ‘Scan’ switch.
4.
Examine the films for axial centering of the X–Ray beam to the center line of the Detector window:
See
illustration 3–4. Y
Illustration
Measure if the x–ray beam centering falls within the following values:
the films, and develop
‘ch1’ or ‘ch635’ and ‘T
ou should see both the top and bottom edges of x–ray
3–4
Beam–On–W
LEFTRIGHT
the films. (W
able’ or ‘Gantry’ on the film as appropriate.
indow Exposure
ait the required film development time described on the film pace.
H635
H1
Xw
. The edges should be well defined.
Xw
– H1
2
If these are not met, perform step 5. If these are met, proceed to step 6.
≤ 1 mm
Xw
2
– H635
≤ 1 mm
H1 – H635
≤ 1 mm
SYSTEM3–7
Page 80
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
3-3DETECTOR BEAM ON WINDOW (continued)
5.
Adjust the Detector Z–Axis position:
a.
Switch OFF the ‘Scan’ switch.
b.
Locate the Detector mounting/adjustment points (see illustration ).
Each mounting location has an adjustment nut for shifting the Detector along the Z–Axis.
c.
Loosen the (Allen cap head) locking screw
d. Turn the Z–Axis adjustment nut in the direction you want to shift the Detector (i.e., toward Table or toward
Gantry) and over the distance required to meet specification using vernier calipers to measure Z–Axis sift
measured
One T
e. T
ighten the locking screw to lock the Detector into position.
f. Make
same.
in step 4.
urn : approximately 1mm shift.
this adjustment at each of the 4 mounting locations. The measurement for all 4 locations should be
.
2234976
the
g. Switch
h.
6.
Proceed to the ‘Gravity SAG’ or ‘Iso Center Alignment’ procedure.
ON the ‘Scan’ switch.
Return to step 2.
Illustration 3–5
Beam–On–Window Adjustment
Detector Support
Plate
Use V
ernier Calipers
Shock
Mount
Detector
Z–Axis
Adjustment Nut
Locking Screw
(Allen Head)
SYSTEM3–8
Page 81
GE MEDICAL
SYSTEMS
CT SYTEC/SYTEC I SERIES
FUNCTIONAL
CHECK /
ADJUSTMENT
REV 0
3-4GRA
1.Preparation:
2.
VITY SAG
Y
ou do not have to perform the following, if you have just completed the ‘Beam–On–Window’ procedure.
a.
Switch OFF the ‘Scan’ and ‘TBL/T
b.
Remove the front and both side Gantry covers.
c.
Remove the bow–tie filter from the collimator (refer to the Component Replacement manual).
d.
Switch ON the ‘Scan’ and ‘TBL/T
Perform an axial scan:
a.