GE CardioSys User manual

CardioSys
Operator’s Manual
Version 4.1
227 498 02 GA(e) Revision A
Contents
The product CardioSys bears the CE marking
-0366
The product fully complies with the electromagnetic immu­nity requirements of standard EN 60601-1-2 “Electroma­gnetic Compatibility – Medical Electrical Equipment”.
The device is in radio-interference protection class B in accordance with EN 55011.
The EC Declaration of Conformity covers only the accessories listed in the “Equipment” section.
This manual reflects software version 4.1.
© Marquette Hellige GmbH 1999 Postfach 600265 D-79032 Freiburg Germany
Tel. +49 761 45 43 0
Revision History
This document is subject to the Marquette Hellige change order system. The revision code, a letter that follows the document part number, changes with every update of the manual.
Part No./Revision Code Date Comment
227 498 02-A 10 Feb. 1999 Initial Release
MS-DOS and Windows are trademarks of Microsoft Corp.
2 CardioSys V4.1 227 498 02-A
Contents
1. Introduction
2. General Information on System Operation
3. Selecting and Admitting a Patient, Retrieving Stored Examinations
4.1 The Resting ECG
4.2 The Exercise ECG
4.3 The STAT ECG
4.4 The Spirometry Test
4.5 Ambulatory Blood-Pressure Measurement
4.6 Late Potential Analysis
4.7 External Programs
5. Deleting Examinations, Quitting the Program
6. System Services Menu
7. Cleaning, Disinfection and Maintenance
8. System Messages and Troubleshooting Tips
9. Technical Specifications
10. Equipment List, Order Information
Appendix
Index
227 498 02-A CardioSys V4.1 3
General Information
Contents
General information
* This manual is an integral part of the instrument and
describes its normal use. It should always be kept close to the equipment. Observance of the manual is a prerequi­site for proper instrument performance and correct operation and ensures patient and operator safety.
* The symbol m denotes: Refer to Operator’s Manual!
It serves as an indicator for important facts to be noted when operating the instrument.
* Information which refers only to certain versions of the
instrument is accompanied by the catalog number(s) of the instrument(s) concerned. The catalog number is given on the instrument nameplate.
* For detailed safety information, please refer to section
1.3 “For your safety”.
* Patient safety, specified measuring accuracy, and interfer-
ence-free operation can be guaranteed only if original Marquette Hellige devices are interconnected (e.g. basic units and plug-in modules).
* The warranty does not cover damages resulting from the
use of accessories and consumables from other manufac­turers.
* Marquette Hellige is responsible for the effects on safety,
reliability, and performance of the equipment, only – if assembly operations, extensions, readjustments,
modifications, or repairs are carried out by Marquette Hellige or by persons authorized by Marquette Hellige.
if the instrument is used in accordance with the
instructions for use.
* All publications are in conformity with the instrument
specifications and IEC publications on safety of electro­medical equipment valid at printing date. All rights are reserved for instruments, circuits, techniques, and names appearing in the manual.
* On request Marquette Hellige will provide a service
manual.
* Only use accessories which are listed in this manual and
which have been tested in combination with the device (e.g. patient cables, electrodes, transducers, sensors, consumables, etc.). If you use accessories or consumables from other manufacturers, Marquette Hellige does not guarantee safe operation or functioning of the device.
4 CardioSys V4.1 227 498 02-A
* The Marquette Hellige quality management system
complies with the DIN EN ISO 9001 and EN 46001 Standards.
Introduction to CardioSys
1. Introduction to CardioSys
In this section you will find
an overview of the CardioSys features and capabilities
and of the available software options
a survey of the CardioSys system components
points to note during system operation
instructions on putting the system into operation
227 498 02-A CardioSys V4.1 5
1.1 Program Features
Introduction to CardioSys
CardioSys is a PC-based system for ECG recording, spiro­metry tests and ambulatory blood-pressure examinations. The following options are available to enhance the program func­tions:
Basic Software Package (RESB)
12-lead ECG acquisition (resting ECG, STAT ECG) – recording of spirometry tests – downloading of data from TONOPORT IV/IVa (am-
bulatory blood pressure)
downloading of data from electrocardiographs EK 53,
EK 56, EK 512, CardioSmart – data management for all operating modes – BDT interface
Resting ECG Measurement (RESM)
measurement of 12 ECG leads acquired simultaneously – tabular presentation of the measured values – editing of measuring marks – direct comparison of 2 resting ECGs
3-Lead Exercise ECG (ERG1)
3-lead recording and display – control of all Marquette Hellige ergometers – creation of custom protocols and phases (depending on
ergometer model)
Upgrade for 12-Lead Exercise ECG (ERG2)
optional recording of 3, 6 or 12 ECG leads – 12-lead ST measurement – automatic or manual determination of the J+X point – presentation of the ST medians – presentation of the resting ECG complex and of the
current ST beat for direct comparison
arrhythmia analysis and presentation during the test
Storage of the Full-Disclosure ECG (EGMO)
continuous storage of up to 12 ECG leads – color-coded arrhythmias – any segment of the full-disclosure ECG can be zoomed
and stored
Resting ECG Interpretation (RESI)
interpretation of the measured resting ECG with detailed
explanations – medication and the patient’s age are taken into account – reanalysis after modification of the measuring marks
6 CardioSys V4.1 227 498 02-A
Exercise Test Expert Mode (ERG3)
optional presentation of ST level and slope in the form
of a bar diagram or ST medians with reference marks – manual editing of the E, J, and J+X points – scrolling to previous stages during the exercise test – direct cross-referencing from trends to full-disclosure
ECG (final report)
Introduction to CardioSys
In-Test Reports (ESTA)
automatic or manual printout during the exercise test via
laser printer or thermal printer
report formats for laser printer: medians + 12-lead ECG
(approx. 4 seconds)
report formats for thermal printer:
10-second segments of the 12-lead ECG in different formats
comparative medians + 12-lead ECG
continuous rhythm (6 or 12 leads)
Late Potential Analysis (AVER)
recording of ventricular late potentials – recording of the standard vectors X, Y, Z (uncorrected) – manual editing of QRS onset and offset
Data Storage on File Server (NETS)
storage of the procedure data on the network file server
Monitoring of Remote Test Stations (ERGM)
during the exercise test data are continuously sent via the
network
the following data can be viewed at any of the work-
stations in the network:
station name
6-lead ECG, leads selectable
information on the exercise test and on the ergometer
HR and blood pressure
ST measurement results (resting beat, current beat)
Task List (OMAN)
patient selection on task list – task list created by host program (hospital information
system, medical reimbursement program, etc.) – supports BDT format – examination results are returned to the requesting system
ECG History (ECGH) requires hardlock
display of the median beats from up to 5 selectable
resting ECGs for visual comparison or printout – 3D presentation of the medians for serial comparison
Review Manager (REVM) requires hardlock
users can review and confirm the examination results – name of the physician who confirmed the results is
automatically entered in the patient file – display of examination procedures according to different
criteria (confirmed, unconfirmed, all)
Cardio Mailbox (CBOX) requires hardlock
transmission of multiple procedures via a modem to
CardioSys/CardioSoft storage of multiple procedures on storage medium – reception of multiple procedures via a modem (from
CardioSys/CardioSoft or CardioSmart)
227 498 02-A CardioSys V4.1 7
Introduction to CardioSys
1.2 System Configuration and Function, Setting Up the PC
CardioSys is a system cart centering around a PC and software programs. Figure 1-1 shows the CardioSys system cart.
1
F2 F3 F4 F5 F6 F7 F8 F9 F10F1
Figure 1-1. CardioSys system cart
1 Monitor 2 Supports for electrode application system or
respiration flow sensor
3 Monitor support 4 Tray for PC keyboard 5 Tray for function keyboard and mouse 6 Tray (e.g. for printer) 7 Accessory drawer 8 Power button, floppy disk drive and drive bays
(streamer, MO drive)
2 3 4
5
6 7
8
9
10
11
12 13
14
15 16
9 Shelf with power connector for defibrillator 10 Conductive castors (front castors locking) 11 Monitor retaining strip 12 Connection panel (under lid) 13 Lid lock 14 Power switch (removes power from the system) 15 Power input with instrument fuses 16 Potential equalization pin
8 CardioSys V4.1 227 498 02-A
Introduction to CardioSys
g
The lid should only be opened by Marquette
Warnin
Hellige staff or by factory-trained personnel.
Biocompatibility The parts of the product described in this operator manual, including all accessories, that come in contact with the patient during the intended use, fulfill the biocompatibility requirements of the applicable stand­ards. If you have questions in this matter, please con­tact Marquette Hellige GmbH or its representatives.
The system comes pre-installed and is ready for operation. All connections for system components and peripheral devices (including the patient cable) are located under the lid (12) which can be opened with a tool only.
The program is a Windows application. Its scope can be expanded by activation of individual optional programs. This manual describes the program with all options included. If you have not purchased one or the other option, simply skip the corresponding sections.
Being familiar with PC programs may be a help, but in this manual you will find everything you need to know about using a mouse and a PC keyboard.
Whenever you are unsure about a program function or oper­ating step, just click on the Help button to obtain online information.
Access to the program is restricted to registered users. At the time a user is registered, the password and the user’s rights are assigned (user, physician, with or without the right to make system changes).
227 498 02-A CardioSys V4.1 9
1.3 For Your Safety
g
g
Introduction to CardioSys
The safety information given below is divided into the catego­ries “Danger”, “Warning” and “Caution”.
indicates an imminently hazardous situation which, if not avoided, will result in death or
Danger
serious injury.
indicates a potentially hazardous situation which, if not avoided, could result in death or
Warnin
serious injury.
indicates a potentially hazardous situation which, if not avoided, may result in minor or
Caution
moderate injury and/or damage to the equip­ment.
provides application tips or other useful informa-
tion to assure that you get the most from your equipment.
CardioSys is a class I system.
Before connecting the device to the power line, check that the voltage and frequency ratings of
Caution
the power line are the same as those indicated on the system’s nameplate. If this is not the case, do not connect the system to the power line.
Do not use extension cables with multiple
Warnin
power outlets to connect the system to the power line.
Before putting the system into operation visu­ally inspect all connecting cables for signs of
Danger
damage. Damaged cables and connectors must be replaced immediately.
When disconnecting the system from the power line, remove the plug from the wall outlet first.
Danger
Then you may disconnect the power cord from the device. If you do not observe this sequence, there is a risk of coming into contact with line voltage by inserting metal objects, such as the pins of electrode leads, into the sockets of the power cord by mistake.
10 CardioSys V4.1 227 498 02-A
Introduction to CardioSys
g
g
g
g
g
Caution
Warnin
Warnin
CardioSys can be operated on an uninter­ruptible power supply (UPS). The UPS should not be disconnected from the power line for prolonged periods of time to prevent deep discharging of the battery. For this reason, switch the CardioSys system off and on with the system’s own power switch and do not use the switch of the UPS or the main power switch of the office. As a general rule, the UPS must first be con­nected to the power line and switched on; then you may switch on CardioSys. Do not switch on the CardioSys system during a power outage. Only the CardioSys system may be connected to the UPS. The UPS must be installed within the medi­cally used room, but outside the patient envi­ronment. Observe the information given in the UPS user manual.
The power socket of the CardioSys system cart is reserved for a defibrillator for recharging of the defibrillator battery (max. power 30 VA).
All devices of a system must be connected to the same power supply circuit. Devices which are not connected to the same circuit must be electrically isolated when operated (electrically isolated RS232 interface).
Set up the device in a location which affords sufficient ventilation. The ventilation openings
Caution
of the device must not be obstructed. The ambient conditions specified in the “Technical Specifications” section must be ensured at all times.
The system should be set up to provide the opera-
tor an unobstructed view of the control panel.
Devices intended for emergency application must not be stored or transported at tempera-
Danger
tures which cause moisture condensation at the application site. Wait until all moisture con­densation has evaporated before using the device.
If the system is used as an emergency device, a
Warnin
second ECG recorder must be available.
The device is not intended for intracardiac
Warnin
application.
The device is not suitable for operation in areas
Warnin
where an explosion hazard may occur. Explo­sion hazards may be caused by the use of flam­mable anesthetics, skin-cleansing agents and disinfectants. Great care must be exercised when the monitor is used in an oxygen-enriched atmosphere. The atmosphere is considered to be oxygen-enriched when the room air contains more than 25% of oxygen or nitrous oxide.
227 498 02-A CardioSys V4.1 11
Introduction to CardioSys
g
g
g
g
g
g
Warnin
Warnin
Warnin
Warnin
The patient signal input is labelled with the
symbol, indicating that it is protected against damage resulting from defibrillation and electrocautery. Nevertheless extreme care must be taken when devices which are directly connected to the patient remain applied during defibrillation. The distance between ECG elec­trodes and defibrillation paddles should be at least 6 in./15 cm. If in doubt, disconnect the patient cable from the device while applying the defibrillation pulse.
Magnetic and electrical fields are capable of interfering with the proper performance of the device. For this reason do not operate the system in the vicinity of equipment emitting strong electromagnetic radiation (X-ray equip­ment, diathermy systems or MRI devices).
Only use the original Marquette Hellige pa­tient cable. Do not connect other signal sources to the cable.
With all peripheral devices connected, CardioSys represents a medical system. It fulfills the requirements of the system standard IEC 60601-1-1. The entire system may be operated in the patient environment.
Warnin
Danger
Warnin
If the system cart is installed in the patient envi­ronment, only the special printer modified by Marquette Hellige and equipped with an addi­tional protective earth connection (potential equalization) may be operated. Observe the as­sembly instructions supplied with the printers.
For stress-test examinations, a defibrillator and a pacemaker, regularly checked for proper functioning, must be readily accessible.
Extreme care must be exercised when using medical electrical equipment. Many parts of the man/machine circuit are conductive, such as the patient, connectors, electrodes, trans­ducers. It is very important that these conduc­tive parts do not come into contact with other grounded, conductive parts when connected to the isolated patient input of the device. Such contact would bridge the patient’s isolation and cancel the protection provided by the isolated input. In particular, there must be no contact of the neutral electrode and ground.
12 CardioSys V4.1 227 498 02-A
Introduction to CardioSys
g
g
g
g
g
Warnin
Danger
Devices may only be interconnected with each other or to parts of systems when it has been made certain that there is no danger to the patient, the operators, or the environment as a result. In those instances where there is any element of doubt concerning the safety of connected devices, the user must contact the manufacturers concerned or other informed experts as to whether there is any possible danger to the patient, the operator, or the environment as a result of the proposed combi­nation of devices. Standards IEC 60601-1-1/ EN 60601-1-1 must be complied with in all cases.
Liquids must not be allowed to enter the de­vices. Devices into which liquids have pen­etrated must be checked by a service technician before being used again.
Medical technical equipment such as the
CardioSys system must only be used by persons who have sufficient knowledge, who have received adequate training in the use of such equipment and who are capable of applying it properly.
Before using the system, the operator is obliged
Warnin
to verify that it is in correct working order and operating condition.
The operator must be familiar with the opera­tion of the system.
Caution
Verify the performance of the system once a month.
Caution
Dispose of the packaging material, observing
Warnin
the applicable waste control regulations and keeping it out of children’s reach.
Chemicals required for the application or maintenance of the devices must under all
Danger
circumstances be stored, prepared and kept at hand in the containers provided by the manu­facturer. Failure to observe this instruction
At the end of its service life, CardioSys and its accessories must be disposed of in compliance
Caution
with the applicable waste control regulations. If you have any questions in this matter, please contact Marquette Hellige GmbH.
may result in severe consequences for the pa-
Warnin
tient.
Warnin
CardioSys is not intended for use as a vital signs physiological monitor.
Only use bicycle ergometers and treadmills that fulfill the applicable IEC 60601-1 require­ments. CardioSys and the associated bicycle ergometer or treadmill must be connected to the same power supply circuit via separate wall outlets.
227 498 02-A CardioSys V4.1 13
Introduction to CardioSys
g
Warnin
Only modems complying with the requirements of IEC 60950 or UL1950 may be connected to the system. Furthermore, the regulations appli­cable in the respective countries must be ob­served. The modem must be set up inside the medically used room, but outside the patient environ­ment.
Literature
Medical Device Directive
EN60601-1/1990 + A1:1993 + A2:1995: Medical electrical equipment. General safety standards
EN60601-1-1/9.1994 and A1 12.95: General safety stand­ards. Standards for the safety of medical electrical systems.
IEC Publication 513/1994: Fundamental aspects of safety standards for medical equipment.
ROY, O.Z.: Summary of cardiac fibrillation thresholds for 60-Hz currents and voltages applied directly to the heart. Med. & Biol. Engn. & Computing 18: 657...659 (1980).
14 CardioSys V4.1 227 498 02-A
Introduction to CardioSys
1.4 Putting the System Into Operation, Testing the Performance
* Before using the system for the first time, please read the
information given in section 1.3 “For your Safety”.
* Turn on power with the main power switch (14).
* Switch on the system with the power button on the front
panel (press for approx. 1 second).
Upon power up the PC runs automatic self tests. When the program detects any malfunctions, an error message is dis­played. If no problem exists, you will see the usual Windows screen.
The ECG program too runs a self test when a signal acquisi­tion screen is called up. If you do not see any error messages, the program and the computer are intact. Error messages may also appear during program operation.
Do not run more than one other Windows application
at the same time as CardioSys. – Do not activate a Windows screen saver. – Run a daily backup of your data (Appendix, sec-
tions 3.1). – Quit the program in the prescribed manner (Chapter 5.
“Deleting Examinations, Quitting the Program”) before
switching off the system with the power button (8).
Turn off power with the main power switch (14) only
when the CardioSys cart will not be used for a prolonged
period of time
227 498 02-A CardioSys V4.1 15
Introduction to CardioSys
1.5 Putting the Thermal Printer AR 200 Into Operation
Controls and indicators of the thermal printer are shown in Figure 1-2.
2
1
3
5
7
6
Figure 1-2. Controls and indicators of the thermal printer
1 Green indicator, is lit when the printer is
connected to the power line Yellow indicator, is lit to indicate a problem
2 Paper guide 3 Paper tray door latch (lift latch to open door) 4 Paper tray door (the printer stops when the door
is opened)
When the door is open, take care to
Caution
insure no small objects enter the printer.
910
8
4
5 Specification labels 6 Instrument fuses 7 Power button (printer is turned on with the
system’s main power switch)
8 Power input 9 Potential equalization pin 10 Parallel port
Top View
Rear Panel View
16 CardioSys V4.1 227 498 02-A
Introduction to CardioSys
Caution
Explanation of Symbols as Used on the Printer
printer problem
interface
T3, 15A/250V
fuse
Chart-Paper Loading
Before loading chart paper, make sure that the printer is switched off (button 7, Figure 1-2).
Caution
The indicators will go off after a few seconds.
* Pulling upward on the paper tray door latch (3, Figure
1-2), open the paper tray door.
* Remove the cardboard backing of the old pad.
* When inserting the new pad, be aware of the following
points: – the last sheets of the paper pad have a red stripe, so
this side of the pad should be at the bottom of the tray
the black squares should be facing down and lo-
cated on the side of the printer nearest the paper tray door latch
place the pad, including the cardboard backing, on
the tray and remove the cardboard on the top.
* Pull the first sheet of the pad straight out over the paper
guide, at the same time closing the paper tray door.
* Turn on the printer.
Do not attempt to operate the printer without thermal paper. Do not pull on the sheets to remove them from the printer. After loading a new pad, press keys
)
+
to ensure that the next printout
9 ]
Ctrl
+
starts at the top of the page.
The thermal printer AR 200 does not require an additional protective earth connection
Caution
(device in compliance with IEC 60601).
227 498 02-A CardioSys V4.1 17
Introduction to CardioSys
Printer Problems (Thermal Printer)
melborP ydemeR
neergdnatnirptonseoD
delebalDEL rewoP tonseod
thgil
.no
neergdnatnirptonseoD
delebalDEL rewoP tilsi
.repapfo
tubrepapsevomrepaP
tnirptonseodredrocer
Reinstalling the Thermal Printer Driver
Usually the printer driver is installed. If it needs to be rein­stalled, proceed as follows:
sielbacrewoptahtkcehC
.detcennocylreporp
sySoidraCehtfikcehC
.nosirewopmetsys
neewtebselbackcehC dnametsyssySoidraC
-erotdeenyaM(.retnirp
).metsysehttoob
tuosiretnirpehtfikcehC
ehtrednuthguacsparcs
.daehtnirp
sahrepaptahtkcehC
.kramnoitanigaptcerroc
* Exit the application to Windows.
sihctiwsrewoptahtkcehC
* From the Program Manager screen, call up
Main Group
Control Panel
Printers.
* Select “Add Printer”.
* Select “Install unlisted or updated driver”.
* Select “Install...”.
repapynaeraerehtfikcehC
* Place the disk with the printer driver into the diskette
.dedaolyltcerrocnisirepaP
drive and click on “OK”.
* Check the “Default printer” box.
* Select “Close”.
* Restart the application.
18 CardioSys V4.1 227 498 02-A
General Information on System Operation
2. General Information on System
Operation
This section describes some basic facts on program opera­tion. You will learn
how to use the program with a mouse
how to enter data via the PC keyboard
how to start the program
how to select the language
227 498 02-A CardioSys V4.1 19
General Information on System Operation
2.1 Operating the Program with a
Mouse
All operating steps required to run the program can be per­formed with a mouse. The screen layout permits swift pro­gram operation, and text entries via the keyboard have been reduced to a minimum. The system allows you to create libraries for various situations, which you can copy from the list to the respective text boxes in a dialog window, using the mouse.
If you are already familiar with the functions and handling of a PC mouse, you can skip this paragraph.
Some General Remarks on the PC Mouse
When entering text via the keyboard, use the
key to move from text box to text box. This will save time as your hands can stay on the keyboard all the time.
Figure 2-1. Mouse buttons
20 CardioSys V4.1 227 498 02-A
An arrow or mouse pointer appears on all screens. This mouse pointer moves around on the screen as you move the mouse on the table. Sliding the mouse to the right, for in­stance, shifts the pointer to the right.
The program can be used with a two or a three-button mouse (Figure 2-1). The buttons are operated with the index finger and the middle finger. The left button (index finger) controls almost all functions, while the right button (middle finger) is seldom used.
General Information on System Operation
Patient
Procedure System Services Help
Baker, Dave, 41yrs REST ECG/ Signal Acquisition
Test Information
Setup
aVR
aVL
aVF
X 35 HZ X 50 HZ
Hellige
Freeze ECG Options
10 mm/mV
Figure 2-2. Screen with operating controls
Select Procedure
Patient
Procedure all
Status
Location
Date (from - to)
Show new procedure list
12/04/1996 03:38:17pm 0 Exercise Test
12/04/1996 03:29:24pm x 0 Exercise Test 12/04/1996 03:10:13pm 0 Exercise Test 12/04/1996 03:00:11pm * 0 Exercise Test 11/20/1996 10:53:39am 0 Exercise Test
Baker, Dave, 41yrs
all
-
1 3 2
Figure 2-3. Excerpt from the procedure list
1 moves the list down (line by line) 2 moves the list up (line by line) 3 allows you to scroll to a general location in the list 4 clears the window without saving entries
»
25 mm/s
V1
V2
V3
V4
V5
V6
Combined Report…
HR (BPM)
SYS/DIA (mmHg)
R
Lead Sequence
STANDARD
Display
Confirm
Send
Delete
Cancel
Help
1 2
Spline
ON
3
85
5
4
4
The icons displayed on the monitor screen resemble the operating controls on a conventional ECG recorder (Figure 2-2).
To select a function you place the mouse pointer on the respective icon and press the left mouse button. You will see that the key on the screen is “pressed” at the same time.
6
L
Pressing a mouse button is also called “clicking”. When, later
FN
on in this manual, you are asked to “double-click” this means that you should press the left mouse button twice in rapid succession.
On many screens you will see windows offering various selections (Figure 2-3).
To select an option, move the mouse pointer to the item and press the left mouse button. This selects the item, and the window disappears.
When a list of selections is too long to fit in the window, the window has a scroll bar at its right edge. Clicking the arrow at the upper end of the scroll bar moves the list down one by one (1, Figure 2-3), clicking the lower arrow moves the list up (2, Figure 2-3).
A so-called scroll box between the two arrows allows you to move quickly through the list by clicking the scroll box, holding the mouse button depressed, and dragging the box up or down (3, Figure 2-3).
227 498 02-A CardioSys V4.1 21
General Information on System Operation
Select Procedure
Patient
Procedure all
Status
Location
Date (from - to)
Show new procedure list
12/04/1996 03:38:17pm 0 Exercise Test
12/04/1996 03:29:24pm x 0 Exercise Test 12/04/1996 03:10:13pm 0 Exercise Test 12/04/1996 03:00:11pm * 0 Exercise Test 11/20/1996 10:53:39am 0 Exercise Test
Baker, Dave, 41yrs
all
-
1
3 2
Figure 2-3. Excerpt from the procedure list
1 moves the list down (line by line) 2 moves the list up (line by line) 3 allows you to scroll to a general location in the list 4 clears the window without saving entries
Display
Confirm
Send
Combined Report…
Delete
Cancel
Help
The scroll box also indicates which section of the list is cur­rently displayed. If the box is just below the upper arrow you see the top portion of the list, if the box is in the middle of the scroll bar, the middle portion of the list is on display, etc. When you see an horizontal scroll bar below a picture, the two arrows and the scroll box allow you to shift the picture left or right.
4
Some screens which offer various selections can be cleared by two methods: either with
OK
or with
Cancel
(Figure 2-3).
When you select tions made, with
OK
the program accepts the selec-
Cancel
, the selections will be ignored.
If you wish to highlight text click the first letter of the text block, hold the left mouse button down and drag the pointer to the last letter. Then release the mouse button. The marked text is displayed in reverse video.
You clear the HELP window by clicking “File”
and “Exit”.
To obtain information on program operation, click the
Help
button or .
22 CardioSys V4.1 227 498 02-A
General Information on System Operation
2.2 The Most Important Keyboard Controls
Use the keyboard to enter text and numbers. The following is a description of functions which you do not know from a conventional typewriter. If you are already familiar with a PC keyboard, you may skip this section.
* The Backspace key (1, Figure 2-4) clears the characters
from left to right, one by one.
* The cursor keys (5, Figure 2-4) are used to move the
cursor up, down, right and left to any desired position in the text, for instance, to insert text (Ins key)
F6F5 F7 F8F2F1 F3 F4 F10F9 F11 F12Esc
* or to delete text (Del key) (3, Figure 2-4).
* The Home and End keys (2, Figure 2-4) move the
cursor to the first and last position in the text.
* With Return (4) and on Windows screens with TAB (6)
you move the cursor to the beginning of a new, blank line.
Scroll
Lock
2
Num Lock
Caps Lock
Scroll Lock
1
Print Pause
¬
!
1"2£3$4
`
Caps Lock
|
\
%
5
&
^
6
7*89)0_-
MNBVCXZ
(
<,>
LKJHGFDSA
.
+
=
}
{
POIUYTREWQ
: ;
]
[
~
@
#
'
?
/
CtrlCtrl Alt GrAlt
4
Insert
Delete
3
Home
End
Page
Page
Down
÷
Num
Up
Lock
Home
O
Ins
x
987
PgUp
654
321
PgDn
,
Del
356
_
+
Enter
Figure 2-4. The keyboard
227 498 02-A CardioSys V4.1 23
Start program
Start program
marquette HELLIGE CardioSoft (Supervisor)
120/80
General Information on System Operation
2.3 Starting the Program, Explanation of the Function Keys and Selecting the Language
marquette HELLIGE CardioSoft
4.1e
Figure 2-5. Title screen
When the password function is enabled (section
6.1 “System Configuration –> General”) enter the user name and password first.
The program may be configured to default to an
operating mode, such as resting ECG. In this case you will see the patient selection screen instead of the title screen. Chapter 3 “Selecting and Admit­ting a Patient” explains how to retrieve a patient from the database.
Starting the Program
* Turn on the system and start the program by double-
clicking on . The title screen (Figure 2-5) appears. When the password function is enabled in the System Configuration (tab “General”), you will see the user registration window.
Enter the name and password and click on the
button.
24 CardioSys V4.1 227 498 02-A
General Information on System Operation
Explanation of the Function Keys
The function keys are used to call up the different program functions:
Opens a window for patient selection or for admission of a new patient (section 3.1, 3.2)
Recording of a resting ECG (section 4.1)
Recording of an exercise ECG (section 4.2)
Recording of a STAT ECG (section 4.3)
Initiates a pulmonary function test (section 4.4)
Initiates the ambulatory BP program for program-
120/80
ming of the BP monitor and data evaluation (section 4.5)
Initiates the late potential analysis program (sec­tion 4.6)
Initiates a bronchodilation test (section 4.7)
Calls up further examination procedures
Opens a window for selection of stored procedures
Opens the printer menu and allows you to print the displayed procedure
Allows you to compare the current procedure with another one
Allows you to delete the displayed procedure
Calls up system services (chapter 6)
Calls up the online help function
Allows you to quit the program
227 498 02-A CardioSys V4.1 25
General Information on System Operation
Patient
(Unbenannt)
System
Configuration
Procedure
Procedure
Copy Patient
Records
Hellige
System Services Help
PPG Hellige PPG Hellige System Services Menu
Archiving
Compression
HELLIGE CARDIOSOFT
T1.0a
Voice
Recorder
from ECG
System/ECG
Recorder
Figure 2-6. System services selection window
System Configuration
General Modem MUSE
Institution Name and Address Name Street City
Remote Cart Name
Remote Cart Name
Physician Technician
Default Procedure Trace Background Screen Diagonal Remote Test Screen
Test Patient (temporary) Anti-Aliasing of ECG Waveforms Patient Selection prior to STAT ECG Enable Test Information Window
System Configuration
Save Load
Print
Option Code
black-yellow (antial.) 15 in NETBIOS-Protocol
Country Settings
For Service only
Archive Patient
Records
ID ID
Windows control menu (Restart program for changes to become effective)
Menu labels
Enable Password Function
Setup
Load
Save
HelpCancel
Database…
User List…
Modify password…
Icon Bar…
OK
Selecting the Language
The program offers different languages for the user interfaces and for the printed reports.
To select one of the languages, proceed as follows:
Click the button.
The window showing the system services opens (Figure 2-6).
Select .
You will see Figure 2-7. (This paragraph describes language selection only. Other settings to be made in this window are discussed in section 6.1)
Select the “Country Settings” tab (arrow, Figure 2-7).
Help
Cancel
Figure 2-7. System configuration menu – “General” tab
26 CardioSys V4.1 227 498 02-A
General Information on System Operation
System Configuration
General Modem MUSE Option Code Country Settings
Language Time Format Date Format Unit of Measure Speed Unit ST Level Line Frequency Lead Label
Enable Swedish Patient ID
Enable Card Reader
English 24-Hour Format DD.MM.YYYY cm, kg km/h mV 50 Hz IEC
Help
Cancel
OK
Figure 2-8. System configuration menu – “Country Settings” tab
A screen similar to Figure 2-8 appears.
Select your language from the “Language” list box and click
OK
to quit the menu.
227 498 02-A CardioSys V4.1 27
For your notes
General Information on System Operation
28 CardioSys V4.1 227 498 02-A
Selecting and Admitting a Patient, Retrieving Stored Examinations
3. Selecting and Admitting a Patient, Retrieving Stored Examinations
In this section you will find information on
how to select a patient from the database
how to admit a new patient to the system
how to retrieve examinations stored for a specific
patient
227 498 02-A CardioSys V4.1 29
Selecting and Admitting a Patient, Retrieving Stored Examinations
1 2 3 4
Select patient (local database)
Search for … Last Name Patient ID
Last Name
Master
Miller Richards Roberts Shields Stevenson Thompson next page…
New Patient… Last Name
Patient ID
Gender Race
First Name
Jim
Jane Chantal Jennifer Stuart Robert Anthony
Master 8738999
Male
Date of Birth
10/05/1949
06/04/1943 03/12/1977 07/27/1977 11/11/1912 11/22/1966 09/19/1932
First Name Date of Birth
Height Weight
Patient ID
8738999
654987 7766699 543876 765321 333999 665544
Jim 10/05/1949
176 66/0
MM/DD/YYYY
in lb
Pacemaker
12 1113
Figure 3-1. Patient selection screen
1 Displays more patients 2 Bar cursor, highlighting selected patient 3 Text box for patient’s last name 4 Click to select highlighted patient and go to the
title screen 5 Click to enter new patient data 6 Clears the patient data 7 “Unnamed” button (the test will not be assigned
to a patient) 8 Click to display the online help screen 9 Clears the menu (entries are not saved) 10 Displays the title screen 11 Click to save the data to the database 12 Pacemaker patient: yes X, no 13 Patient data window
Depending on the system configuration you will
be working with the local database or with the database of the MUSE CV system.
Select
New Patient
Delete
Unnamed
Help
Cancel
Close
Accept
3.1 Retrieving a Patient From the Database
An examination procedure can be assigned to a patient only when this patient has been retrieved from the database or
5
when the patient is admitted to the system first. Similarly, you
6
have to select a patient from the database before recalling one
7
of the procedures stored for this patient (“Unnamed” allows
8
you to run a procedure without assigning it to a patient).
9
10
Click on to display the patient selection screen.
Local Database (Figure 3-1)
Type the last name or ID number of the patient whose record you wish to retrieve (it is sufficient to enter the first letters or numbers).
Click at (1) to display a new page of patient records.
Check the patient data. If it needs to be changed: – type over the old text, and – click
Accept
.
To select the highlighted patient record – click
Select
, or
– press .
As an alternative, you can select a patient record by double-clicking on the name.
30 CardioSys V4.1 227 498 02-A
Loading...
+ 242 hidden pages