General points to note when handling a defibrillator
Electromedical equipment such as the CardioServ
Warnin
defibrillator must only be handled by persons who
are trained in the use of such equipment and are
capable of applying it properly.
Before using the equipment, the operator must
Warnin
ascertain that it is in correct working order and
operating condition.
The defibrillator paddles must be clean and dry.
Danger
The person carrying out the defibrillation should
☞
have at least one assistant.
The operator must be trained in the use of the
Warnin
defibrillator.
All assistants must be briefed regarding the
Warnin
preparations for and execution of defibrillation.
All tasks must be assigned clearly.
Warnin
Defibrillating a patient with normal heart
rhythm may induce ventricular fibrillation.
Danger
Position the patient flat on a hard, dry surface
where the patient is electrically insulated. The
Caution
patient must not be allowed to come into contact
with metal parts, e.g., bed or litter, in order to
prevent unwanted pathways for the defibrillation
current which endanger the assistants. For the
same reason, do not place the patient on wet
ground (rain, accident in swimming pool).
Have a pacemaker at hand, if possible.
☞
Should cardiac arrest occur or be imminent during
☞
preparations for defibrillation, administer heart
massage and artificial respiration (CPR).
Do not allow the defibrillator paddles to come
into contact with other electrodes or metal parts
Danger
which are in contact with the patient.
Transducers and instruments that are not defi-
brillation-proof must be disconnected from the
Caution
patient.
Interrupt heart massage and artificial respiration
immediately before triggering the shock.
Danger
Immediately before triggering the shock discon-
nect tubes and have assistants step back.
Caution
227 446 32-ICardioServ V 4.211
Introduction to CardioServ
g
g
g
Caution
Danger
Warnin
Caution
The patient’s chest must be dry, as moisture causes
unwanted pathways for the defibrillation current.
Therefore, when using flammable skin cleansing
agents, wait until they have completely dried.
Possible explosion hazard if used in the presence
of concentrated oxygen, flammable substances
(gasoline) or anesthetic agents. Oxygenation in
the vicinity of the defibrillation paddles must be
strictly avoided; if necessary, interrupt oxygen
supply while defibrillating the patient.
To prevent sparking
– the electrodes should make full contact with
the body
– the electrodes should be pressed firmly onto
the thorax.
Do not deliver shocks into open air. High voltage
may briefly be present at the unprotected paddle
surfaces as a result and endanger the persons
present.
Warnin
Warnin
Defibrillating a patient who has an implanted
pacemaker is likely to impair the pacemaker
function or cause damage to the pacemaker.
Therefore, the following should be observed:
– Select the smallest energy level possible for the
application.
– Do not apply the defibrillation paddles in the
vicinity of the pacemaker electrodes.
– The availability of an external pacemaker is
of utmost importance in this case.
– After the defibrillation the working order of
the implanted pacemaker should be checked
immediately.
Also be aware that children require less energy for
a successful defibrillation than adults. For the
first defibrillation pulse delivered to babies and
toddlers, select an energy level of 2 joules/kg body
weight. For subsequent shocks, increase the energy
up to 4 joules/kg.
When defibrillating children it is especially
important to verify that the paddles make full
Caution
contact with the body surface. This is to be ob-
served also when using the clip-on electrodes for
children (Part No. 303 439 95).
12CardioServ V 4.2227 446 32-I
Controls and Indicators
2. Controls and Indicators
This section describes the CardioServ operating controls and
indicators and explains their function.
When operating elements in this manual are identified with
a reference number in parentheses, this number refers to
Figure 2-1 in this section.
You will also find an explanation of all signs and symbols
used on the CardioServ defibrillator.
227 446 32-ICardioServ V 4.213
Controls and Indicators
-
+
1
2
Shock
3
STERNUM
4
J (50 )
102030
7
5
2
Autosequence
2
P
a
c
e
m
a
k
e
r
S
t
i
m
u
l
a
t
e
+
P/min
+
P/min
mA
–
mA
–
Dem
Fix
50
100
300
360
u
r
6
150
200
Shock
Charge
Shock
Sync
138975
1 Select energy
Apply electrode cream to both paddles
2
Charge defibrillator
Press charge button on apex paddle
(on the device, when using adhesive
or internal electrodes)
Deliver shock3
Press buttons on both paddles simultaneously (on the device, when using
adhesive or internal electrodes)
APEX
Charge
Charge
Shock
3
2
APEX
1418 17 16 15
13
12
11
10
19
20
21
22
23
24
Option
Figure 2-1. Controls and indicators of CardioServ
14CardioServ V 4.2227 446 32-I
SpO
EKG
2
Controls and Indicators
-
+
1Defibrillator paddles
Shock
2
button to trigger the defibrillation shock – together
3
with button (8)
3 keys to adjust the LCD contrast. Press both
keys simultaneously to obtain a screen copy.
4Energy selector, on/off switch
Charge
5
key to charge the unit (manual mode) and to
Shock
trigger the defibrillation shock – together with key (6).
This key assumes the function of button (8) when
internal paddles or adhesive electrodes are used.
Shock
6
key to trigger the defibrillation shock – together
with key (5). This key assumes the function of button
(2) when internal paddles or adhesive electrodes are used.
7Five selection keys F1 to F5 whose functions change
with the menu displayed. The respective key functions
are indicated by symbols or labels in the bottom line on
the LCD. From the main menu, that appears on power
up, you can access submenus which, in turn, allow the
selection of further options. The back function returns
you to the next higher menu level. The main menu
reappears automatically if you do not depress any of the
keys for about 30 seconds.
8
button to charge the unit (manual mode) and
3
2
Charge
Shock
trigger the defibrillation shock – together with button (2)
11 key to start and stop the recorder. This key assumes
the function of button (9) when internal paddles or
adhesive electrodes are used.
12 Green indicator is lit when the defibrillator operates
on line power.
13 Yellow indicator
is lit when the defibrillator battery
is being charged.
14 Connection for defibrillator paddles
(controls 15 to 18 only on models with pacemaker)
+
mA
15 Keys
16 Keys
17 Pacing mode selection key
mA
+
P/min
to adjust the pacer output
-
P/min
to adjust pacer rate
-
Dem
Fix
(demand/fixed rate)
18 Key to enable and disable the pacemaker
19 Aperture to open the paper compartment
20 Connector for power cord
21 Unassigned
22 1-Volt ECG output
23 Connector for SpO2 sensor
9 button to start and stop the recorder.
Sync
10
key to switch to the synchronized operating mode
24 Connector for patient cable (ECG signal input)
(section 5 “Cardioversion”)
227 446 32-ICardioServ V 4.215
Controls and Indicators
-
+
Explanation of the signs and symbols used on the defibrillator
Type CF equipment with highly insulated patient
connections, suitable for intracardiac application,
connections defibrillation-proof.
Type CF equipment with highly insulated patient
connections, suitable for intracardiac application,
connections not defibrillation-proof.
Standby mode (for line-power operation)
Power off
Battery charging
Recorder start
ECG signal
Contrast
Standby or preparatory state only for a part of the
equipment
On, only for a part of the equipment
Hardcopy of screen image
This symbol indicates that the waste of electrical
Refer to Operator’s Manual
and electronic equipment must not be disposed as
unsorted municipal waste and must be collected
separately. Please contact an authorized representa-
tive of the manufacturer for information concern-
ing the decommissioning of your equipment.
Direction indicator
Signal output
Audible alarm on/off
High voltage
16CardioServ V 4.2227 446 32-I
Setting Up CardioServ and Testing Its Performance
3. Setting Up CardioServ and
Testing Its Performance
In this section you will find information about
–putting CardioServ into operation
–connecting CardioServ to the 12-Volt power supply of
an ambulance vehicle
–customizing the CardioServ settings to suit your per-
sonal requirements
–testing the CardioServ performance before using it on a
patient
227 446 32-ICardioServ V 4.217
Setting Up CardioServ and Testing Its Performance
3.1 Setting Up CardioServ
The defibrillator is a high-voltage electrotherapy
device and must be handled by qualified person-
Danger
nel only. Improper use of this device can endanger
life. Do not fail to observe the information given
in this manual and only entrust the device to the
hands of trained persons.
Check the electrodes and their leads for signs of
damage every time before you use the defibrilla-
tor. In particular, make a close visual inspection
of the insulation. Replace internal electrodes or
the contact inserts when you detect signs of me-
chanical damage.
CardioServ operates on:
–line power (95 to 240 V, 49 to 65 Hz)
–battery power (rechargeable batteries), i.e., independent
of the power line
–12-Volt power supplied from the emergency vehicle
(with optional defibrillator mounting system)
*Use the power cord to connect the defibrillator to the
power line (Figure 3-1).
Figure 3-1. Power input
18CardioServ V 4.2227 446 32-I
Setting Up CardioServ and Testing Its Performance
360
x
-
+
-
+
Shock
Charge
m
a
k
e
r
S
t
i
m
u
l
a
t
e
u
mA
mA
r
+
+
P/min
m
P/min
Shock
Sync
Figure 3-2. Green indicator (indicating that defibrillator is
supplied from the power line)
Figure 3-3. Inserting the battery
*Check that the green indicator is lit (Figure 3-2).
*Check that the battery is in place (Figure 3-3).
Pull back the catch on the underside of the device to remove
the battery. When inserting it, make sure that it clicks prop-
erly into place.
When a battery is inserted, the yellow indicator
starts
flashing as soon as the defibrillator is connected to the power
line (to indicate that the battery is charging). The battery is
fully charged after 16 hours and the indicator is continuously
lit.
*Check that the supply of chart paper is sufficient (Figure
3-4). A stripe marks the last 3 meters of the roll.
Figure 3-4. Recording strip
227 446 32-ICardioServ V 4.219
Setting Up CardioServ and Testing Its Performance
g
-
+
-
+
-
+
Power Supply From Emergency Vehicles
A qualified technician can be called in to connect the
CardioServ to the 12-Volt supply of an emergency vehicle.
The following points must be noted:
–The negative terminal of the ambulance power supply
system must be connected to the ambulance chassis for
grounding.
–The positive lead of the ambulance power supply system
intended to supply the current must be protected with a
10-A fuse.
T
e
>
s
9
t
5
>
%
8
0
%
<
8
0
%
T
e
S
s
ta
t
rt
–Use only the defibrillator mounting system listed in
section 15 “Order Information and Accessories” or the
external charging unit, if your CardioServ has been
modified accordingly.
Check that the contacts for power supply from the
Figure 3-5. Accu Service Unit
defibrillator mounting system on the underside of
CardioServ (next to battery) are clean. Do not damage
We recommend our Accu Service Unit for optimal
☞
them in any way.
care of the batteries. It prolongs the batteries’ service
life and guarantees their operational readiness at all
Important Information on Battery-Power Operation
times.
Rechargeable batteries require special maintenance and con-
A NiCd battery should not be charged while
Warnin
located in direct sunlight, over a radiator, in cold
storage, or in other temperature extremes (not
below 5 °C). When the instrument is charging,
ambient temperatures exceeding 40 °C may
adversely affect battery capacity and life.
20CardioServ V 4.2227 446 32-I
tinued checks to assure they function in emergency situations.
It is normal for batteries of this type to selfdischarge when not
in use.
The battery charges automatically when CardioServ is con-
nected to the power line (yellow indicator (13) flashes).
Setting Up CardioServ and Testing Its Performance
In order to ensure its functioning as an emergency device, the
defibrillator should not be disconnected from the power line
for more than 48 hours.
A fully charged battery supplies power for 35 defibrillation
shocks of 360 joules or 2 hours of monitoring (or 1.2 hours of
monitoring if the CardioServ unit has pacemaker and SpO
option). It takes 16 hours to charge a depleted battery.
Proceed as follows to test the battery charge level:
☞
–Disconnect CardioServ from the power line.
–Trigger a test discharge of 360 joules (see
section 3.2 “Testing the Defibrillator Perform-
ance”).
–If you are not prompted to charge the battery,
the charge level should be sufficient for at least
5 more 360-joule shocks.
Monthly battery maintenance and checks:
☞
1. Disconnect CardioServ from the power line and
discharge fully charged battery in the monitoring
mode. To do so, set energy selector switch to ,
(SpO2 sensor not connected) and wait until device
switches off.
2
Inserting CardioServ In Its Softcase
*Open both zips on the CardioServ softcase.
*Undo the two Velcro strips on the front of the softcase.
*Open the Velcro flap located in front of the CardioServ
printer.
*Place the softcase on the small base, so that the softcase is
positioned as though you were carrying it.
*Disconnect the paddles on the CardioServ and remove
them.
*Hold all the opened up parts of the softcase to the side
and put the CardioServ into the softcase.
*Pull the softcase by the two Velcro strips on the black
base into position, ensuring that the feet of the
CardioServ are in the cut-out openings provided.
*Close the two front Velcro strips as well as the flap in
front of the printer at the side.
2. Check how long it takes before battery is depleted.
If the time is less than 1.8 hours, the battery is too
old or improperly maintained and must be
replaced.
3. Recharge the battery. This will take 16 hours.
For easy, convenient care and maintenance of the
batteries, use our “Accu Service Unit”.
227 446 32-ICardioServ V 4.221
*Close the zips.
*If the individual cut-out openings for the external con-
nections, paddles or feet are not correctly positioned,
pull the material at this point into the right shape.
*Reconnect the paddles and put them back into the
CardioServ recesses.
Customizing the defibrillator settings
Setting Up CardioServ and Testing Its Performance
Further steps that can be taken while setting up the defibrilla-
tor include customizing the device functions. This allows you
to select defibrillator default settings which suit your personal
preferences. CardioServ saves these configured settings and
reactivates them automatically every time you switch the
Refer to section 11 “The Defaults Menu” for a detailed
explanation of how to customize the CardioServ, including
the language selection (available languages are English,
French, German, Italian, Portuguese, Russian, Spanish and
Swedish).
defibrillator on. The following chart shows the factory settings
On power up CardioServ runs an automatic selftest. Any
malfunctions identified during this test result in an error
message displayed on the LCD (refer to section 12 “Error
Indications and Messages”). As a further performance test a
trial defibrillation can be triggered.
The energy selector is used to switch CardioServ on and off.
Once you have become familiar with CardioServ you can thus
switch on the defibrillator and select the required energy in
one single operation.
*Set the energy selector to the position. No energy
will be stored in this position of the switch.
The defibrillator beeps and displays a checkered pattern
(LCD performance test). Next the main display appears
(Figure 3-7).
g
The “Paddle” message (b, Figure 3-7) indicates that no pa-
tient cable is connected and that the ECG signal is acquired
via the defib paddles. Upon connection of the patient cable
FreezeSpO2ECG
Next
Menu
the selected ECG lead is displayed here. CardioServ is now
ready for operation.
Figure 3-7. CardioServ main display
a heart/pulse-rate reading with alarm limits
b ECG signal source
c 1-mV calibration pulse
d alarm message, QRS blip
e signal trace, channel 1
f signal trace, channel 2
g menu
227 446 32-ICardioServ V 4.223
Setting Up CardioServ and Testing Its Performance
Do not trigger more than 5 consecutive test
discharges (or internal safety discharges) within
Caution
15 minutes at max. energy setting.
Paddle
0
160 BPM
140 ECG
HR Alarm
Figure 3-8. Display of selected energy
360 J
Next
MenuFreezeSpO2ECG
Test Discharge
A test discharge can be triggered to check the defibrillator
discharge circuit. For this test the stored energy is discharged
into the device via two contacts in the paddle compartments.
*Set the energy selector to 360 joules (50 joules, if inter-
nal electrodes are connected). The display first shows the
selected energy (Figure 3-8).
2
Charge
*Press the
(Press the
Shock
button on the paddle to charge the unit.
3
Charge
key on the defibrillator, when using
Shock
internal electrodes.)
You may now watch the defibrillator charging.
*When the selected energy level has been reached,
CardioServ beeps and the stored energy is displayed
(Figure 3-9).
Do not deliver shocks into open air. High voltage
may briefly be present at the unprotected paddle
Caution
surfaces as a result and endanger the persons
onds. To do so, simultaneously press the buttons on
Charge
both paddles (or press the two
Shock
Shock
keys on the
defibrillator, when using internal electrodes).
present.
If more than 5% of the available energy are lost before the
defibrillation pulse is triggered, CardioServ recharges until the
required energy level is reached.
24CardioServ V 4.2227 446 32-I
*Trigger the defibrillation pulse within the next 30 sec-
Setting Up CardioServ and Testing Its Performance
Paddle
0
160 BPM
140 ECG
HR Alarm
Energy
available
Figure 3-9. Display of available energy
If you do not trigger the defibrillation pulse within
☞
30 seconds, an internal safety discharge is initiated
automatically.
360 J
Next
MenuFreezeSpO2ECG
After defibrillation, the beeping sound stops, and the energy
actually delivered into a 50-ohm resistance is displayed for
10 seconds in place of the stored energy (Figure 3-9). The
delivered energy must not deviate more than ±15% or
±4 joules (whichever is greater) from the selected value. A
recording is initiated at the same time (16-second strip).
Should the discharge circuit be interrupted (paddles not
properly placed on contacts in the compartments, defective
lead), an internal safety discharge is initiated 200 ms after the
defibrillation shock has been triggered. In this case the “deliv-
ered energy” is “0”.
If the defibrillator cannot store the selected energy so that
selected and stored energy values differ, the LCD shows the
message “Energy high” or “Energy low”. The defibrillation
pulse can be triggeed all the same.
*Switch off CardioServ (set energy selector to ).
The message “Energy high” or “Energy low”
indicates that CardioServ needs to be repaired.
Caution
If, in spite of this energy storage problem, the
device has to be employed, it will display the
message “Self-test failed. Charge Energy Error”
Testing the Pacemaker Performance
The performance of the pacemaker can be tested with a
commercially available pacemaker tester (e.g. CS300 Simula-
tor from GEMS IT, part no. 417 983-001).
upon power up. In this situation press one of the
function keys and proceed as usual.
Performance Test
Caution
Test the defibrillator performance once a week:
test 1 – defibrillator connected to mains, battery
removed,
test 2 – defibrillator disconnected from mains,
battery inserted
227 446 32-ICardioServ V 4.225
For your notes
Setting Up CardioServ and Testing Its Performance
26CardioServ V 4.2227 446 32-I
Non-Synchronized Defibrillation
4. Non-Synchronized Defibrillation
This section describes first how to perform a non-synchro-
nized defibrillation, using the standard defibrillation
paddles. The subsequent explanations refer to the use of
internal and disposable adhesive electrodes.
At the end of this section you will find a summary of all