GE Cardioserv User manual

CardioServ
Version 4.2
Operator’s Manual
227 446 32 Revision I
The product CardioServ bears the marking
-0459
(notified body GMED)
The authorized representative for GE Medical Systems Infor-
mation Technologies Inc. in Europe is:
Directive 93/42/EEC concerning medical devices and fulfills
the essential requirements of Annex I of this directive.
The product complies with the electromagnetic immunity
requirements of standard IEC 60601-1-2/EN 60601-1-2
Electromagnetic Compatibility – Medical Electrical Equip-
ment”.
The device is in radio-interference protection class B in
accordance with CISPR11/EN 55011.
The CE marking covers only the accessories listed in section
Order Information and Accessories.
This manual describes CardioServ with all options included
and reflects software version 4.13.
© Copyright 2005 GE Medical Systems Information Technologies GmbH
Revision History
This document is subject to the GE Medical Systems Information Technologies GmbH change order system.
The revision code, a letter that follows the document part number, changes with every update of the manual.
GE Medical Systems
Information Technologies GmbH
Munzinger Str. 3
D-79111 Freiburg, Germany
Tel. +49 (0) 7 61 45 43-0
Fax: +49 (0) 7 61 45 43-233
The country of manufacture appears on the device label.
Part No./Revision Code Date Comment
227 446 32 29 April 1997 Initial Release
227 446 32-A 14 May 1997 SW version 4.1
227 446 32-B 1 July 1997 ECO 202998
227 446 32-C 9 July 1998 ECO 060477
227 446 32-D 6 August 1999 ECO 062982
227 446 32-E 11 June 2001 ECO 066990
227 446 32-F 15 October 2003 ECO 074738
227 446 32-G 13 January 2004 ECO 075566
227 446 32-H 20 December 2004 ECO 079412
227 446 32-I 28 July 2005 ECO 080432
2 CardioServ V 4.2 227 446 32-I
Contents
1. Introduction to CardioServ
2. Controls and Indicators
3. Setup and Performance Test
4. Non-Synchronized Defibrillation
5 Cardioversion (Synchronized Defibrillation)
6. Displaying and Monitoring the ECG
7. The Memories of CardioServ
8. Recording
9. Oxygen Saturation SpO
2
10. Pacing
11. Configuring the Defibrillator Settings
12. Error Indications and Messages
13. Cleaning, Maintenance and Disposal
14. Technical Specifications
15. Order Information and Accessories
Appendix
Index
227 446 32-I CardioServ V 4.2 3
General Information
General Information
* This manual is an integral part of the instrument and
describes its normal use. It should always be kept close to
the equipment. Observance of the manual is a prerequi-
site for proper instrument performance and correct
operation and ensures patient and operator safety.
* The symbol
It serves as an indicator for important facts to be noted
when operating the instrument.
* Information which refers only to certain versions of the
instrument is accompanied by the part number(s) of the
instrument(s) concerned. The part number is given on
the instrument nameplate.
* Patient safety, specified measuring accuracy, and interfer-
ence-free operation can be guaranteed only if original
GEMS IT devices are interconnected (e.g. basic units
and plug-in modules).
* Only use accessories which are listed in this manual or
which have been tested in combination with the device
(e.g. patient cables, electrodes, transducers, sensors,
consumables, etc.). If you use accessories or consumables
from other manufacturers, GEMS IT does not guarantee
safe operation or functioning of the device.
denotes: Refer to Operator’s Manual!
* GEMS IT considers itself responsible for the effects on
safety, reliability, and performance of the equipment,
only if
assembly operations, extensions, readjustments,
modifications, or repairs are carried out by
GEMS IT or by persons authorized by GEMS IT
the instrument is used in accordance with the
instructions for use.
* All publications are in conformity with the instrument
specifications and IEC publications on safety of electro-
medical equipment valid at the time of printing. All
rights are reserved for instruments, circuits, techniques,
software programs and names appearing in the manual.
* On request GEMS IT will provide a service manual.
* The GEMS IT quality management system complies
with the standards DIN EN ISO 9001 and EN 46001.
* The warranty does not cover damage resulting from the
use of accessories and consumables from other manufac-
turers.
4 CardioServ V 4.2 227 446 32-I
Introduction to CardioServ
1. Introduction to CardioServ
This section describes
The capabilities and applications of the CardioServ
defibrillator
points to note when operating CardioServ
general points to note when handling a defibrillator
227 446 32-I CardioServ V 4.2 5
Introduction to CardioServ
1.1 General Information
CardioServ is a high-voltage electrotherapy unit
which should be handled only by specially trained
Danger
personnel. Even though the defibrillator is
equipped with various safety features, such as
internal safety discharging, its operation by
unqualified staff could be hazardous to the
patient, the operator, and any assisting personnel.
The user instructions given in this manual refer to a
CardioServ unit equipped with pacemaker and
SpO2 measuring system.
CardioServ is a light-weight, portable defibrillator with ECG
monitor and built-in recorder.
The device is designed for external and internal defibrillation.
It can be used both for semi-automatic and for manual defi-
brillation. Furthermore CardioServ is capable of monitoring
the heart rate with adjustable alarm limits.
CardioServ can be expanded with a transcutaneous pacemaker
and/or SpO2 measuring system which also monitors the
measured SpO2 values.
The device is easy and convenient to operate. Three operating
steps are sufficient to deliver the defibrillation shock. The
display can be turned 180° to allow the user to view the
information when the device is standing upright.
Defibrillation can be performed with standard defibrillation
paddles, defibrillation pads and internal electrodes.
In addition to line power operation, the defibrillator operates
on battery power from a slot-in rechargeable battery, and on
12-Volt power supplied from an emergency vehicle.
6 CardioServ V 4.2 227 446 32-I
Introduction to CardioServ
I/A
120
100
80
U
60
40
20
0
-20 0 2 4 6 8 10 12
Ra = 50 Ohm Ra = 100 Ohm Ra = 25 Ohm
c
L = 26 mH
C = 32 μF
R = 6,5 Ω
t
I
R
a
Figure 1-1. CardioServ current-discharge curve (360 J)
14
t/ms
CardioServ comes with three memories whose contents can
be documented individually:
text memory (code summary documentation)
event memory (16-s ECG)
trend memory (trend plots of HR, SpO2, 45 min,
9 hours)
The built-in recorder is initiated automatically or manually.
Extensive safety precautions have been taken to protect the
patient and the user from inadvertent delivery of defibrillation
shocks.
The current-discharge curve corresponds to an approximate
sinusoidal halfwave with aperiodic decay. Figure 1-1 shows
the discharge curve for various external resistances.
227 446 32-I CardioServ V 4.2 7
1.2 For your Safety
g
g
g
g
g
Introduction to CardioServ
The safety information given below is divided into the catego-
ries Danger, Warning and Caution.
indicates an imminently hazardous situation
which, if not avoided, will result in death or
Danger
serious injury.
indicates a potentially hazardous situation which,
Warnin
if not avoided, could result in death or serious
injury.
indicates a potentially hazardous situation which,
if not avoided, may result in minor or moderate
Caution
injury and/or damage to the equipment.
provides application tips or other useful information
to assure that you get the most from your equipment.
CardioServ is designed to comply with IEC 60601/
EN60601 requirements. It is Class I equipment and
has an internal power source.
CardioServ operates on line voltages between 95
Warnin
and 240 Volts, 49 to 65 Hz. The mains plug
must be connected to an appropriate power supply
with a non-fused earthed wire. The use of exten-
sion cords is not permitted.
Before putting the device into operation, visually
Warnin
check all connecting cables and electrodes for signs
of damage. Damaged cables and electrodes must
be replaced immediately, before use.
When disconnecting the device from the power
Warnin
line, first remove the plug from the wall outlet.
Disconnect the device from the power line and
Warnin
operate it on battery power, if the integrity of the
protective earth conductor is in doubt.
Set up the device in a location which affords
sufficient ventilation. The ventilation openings of
Caution
the device must not be obstructed. The ambient
conditions specified in the Technical Specifica-
tions section must be ensured at all times.
8 CardioServ V 4.2 227 446 32-I
Introduction to CardioServ
g
g
g
g
g
g
Warnin
Danger
Warnin
Danger
Warnin
Caution
Devices intended for emergency application must
not be exposed to low temperatures during storage
and transport to avoid moisture condensation at
the application site. Wait until all moisture has
vaporized before using the device. Avoid using the
defibrillator under conditions where prolonged
exposure to or excessive contact with moisture can
occur.
Before putting the defibrillator into operation,
make sure that the paddles and all connection
cables are dry.
CardioServ is an emergency device and must be
ready for operation at any time. For this reason,
the defibrillator battery must always be charged.
This can be achieved by leaving the defibrillator
connected to the power line when it is not needed
in an environment where only battery operation
is possible.
Possible explosion hazard if used in the presence
of concentrated oxygen.
CardioServ is suitable for application in a humid
environment provided the regulations concerning
drip-proof equipment of IEC 60601 are strictly
observed. However, avoid defibrillation in a very
moist or wet environment, unless absolutely
necessary.
Use only the original GEMS IT batteries, as
these are designed for an extended temperature
range.
Magnetic and electrical fields are capable of
Warnin
interfering with the proper performance of the
device. For this reason make sure that all external
devices operated in the vicinity of the defibrillator
comply with the relevant EMC requirements. X-
ray equipment, MRI devices and radio systems
are a possible source of interference as they may
emit higher levels of electromagnetic radiation.
The defibrillator is designed for intracardiac appli-
cation.
Devices may be connected to other devices or to
Warnin
parts of systems only when it has been made
certain that there is no danger to the patient, the
operators, or the environment as a result. In those
instances where there is any element of doubt
concerning the safety of connected devices, the
user must contact the manufacturers concerned or
other informed experts as to whether there is any
possible danger to the patient, the operator, or the
environment as a result of the proposed combina-
tion of devices. Standards IEC 60601-1-1/
EN 60601-1-1 must be complied with in all
cases.
Set up the device so that the operator has a clear,
unobstructed view of the front panel.
Liquids must not be allowed to enter the device.
Warnin
Devices into which liquids have penetrated must
be immediately cleaned and checked by a service
technician.
227 446 32-I CardioServ V 4.2 9
Introduction to CardioServ
g
g
g
Danger
Caution
Warnin
Warnin
Caution
Use only the original GEMS IT patient cables.
Do not connect other signal sources to the cables.
Patient signal inputs labelled with the
symbol are protected against damage resulting
from defibrillation and electrocautery voltages.
Nevertheless extreme care must be taken when
devices which are directly connected to the patient
remain applied during electrocautery or defibril-
lation. The distance between ECG electrodes
should be at least 15 cm. If in doubt, disconnect
the patient cable from the device while applying
the defibrillation pulse or performing
electrocautery.
For defibrillation of children use only the special
clip-on electrodes for children listed in section 15
Order Information and Accessories.
The full responsibility for the use of accessories
from other manufacturers lies with the user.
Check the device performance at regular intervals
(once a month), strictly following the instructions
given in section 3.2. Do not select high energy
levels for test discharges with defibrillation elec-
trodes shorted together.
At the end of its service life, CardioServ and its
accessories must be disposed of in compliance with
Caution
the special waste control regulations for electronic
parts. If you have any questions in this matter,
please contact GEMS IT GmbH.
Dispose of the packaging material, observing the
Warnin
applicable waste-control regulations and keeping
it out of childrens reach.
Literature
Medical Device Directive
EN 60601-1/1990 + A 1: 1993 + A2: 1995: Medical electrical
equipment. General requirements for safety
EN 60601-1-1/9.1994 + A1 12/1995: General requirements
for safety. Requirements for the safety of medical electrical
systems.
IEC Publication 513/1994: Fundamental aspects of safety
standards for medical equipment.
ROY, O.Z.: Summary of cardiac fibrillation thresholds for
60-Hz currents and voltages applied directly to the heart.
Med. & Biol. Engn. & Computing 18: 657...659 (1980).
For each defibrillation, verify that the selected
and the displayed/charged energy are identical.
Caution
CardioServ also operates on line power when the
battery is depleted or missing.
10 CardioServ V 4.2 227 446 32-I
Introduction to CardioServ
g
g
g
g
g
General points to note when handling a defibrillator
Electromedical equipment such as the CardioServ
Warnin
defibrillator must only be handled by persons who
are trained in the use of such equipment and are
capable of applying it properly.
Before using the equipment, the operator must
Warnin
ascertain that it is in correct working order and
operating condition.
The defibrillator paddles must be clean and dry.
Danger
The person carrying out the defibrillation should
have at least one assistant.
The operator must be trained in the use of the
Warnin
defibrillator.
All assistants must be briefed regarding the
Warnin
preparations for and execution of defibrillation.
All tasks must be assigned clearly.
Warnin
Defibrillating a patient with normal heart
rhythm may induce ventricular fibrillation.
Danger
Position the patient flat on a hard, dry surface
where the patient is electrically insulated. The
Caution
patient must not be allowed to come into contact
with metal parts, e.g., bed or litter, in order to
prevent unwanted pathways for the defibrillation
current which endanger the assistants. For the
same reason, do not place the patient on wet
ground (rain, accident in swimming pool).
Have a pacemaker at hand, if possible.
Should cardiac arrest occur or be imminent during
preparations for defibrillation, administer heart
massage and artificial respiration (CPR).
Do not allow the defibrillator paddles to come
into contact with other electrodes or metal parts
Danger
which are in contact with the patient.
Transducers and instruments that are not defi-
brillation-proof must be disconnected from the
Caution
patient.
Interrupt heart massage and artificial respiration
immediately before triggering the shock.
Danger
Immediately before triggering the shock discon-
nect tubes and have assistants step back.
Caution
227 446 32-I CardioServ V 4.2 11
Introduction to CardioServ
g
g
g
Caution
Danger
Warnin
Caution
The patients chest must be dry, as moisture causes
unwanted pathways for the defibrillation current.
Therefore, when using flammable skin cleansing
agents, wait until they have completely dried.
Possible explosion hazard if used in the presence
of concentrated oxygen, flammable substances
(gasoline) or anesthetic agents. Oxygenation in
the vicinity of the defibrillation paddles must be
strictly avoided; if necessary, interrupt oxygen
supply while defibrillating the patient.
To prevent sparking
– the electrodes should make full contact with
the body
– the electrodes should be pressed firmly onto
the thorax.
Do not deliver shocks into open air. High voltage
may briefly be present at the unprotected paddle
surfaces as a result and endanger the persons
present.
Warnin
Warnin
Defibrillating a patient who has an implanted
pacemaker is likely to impair the pacemaker
function or cause damage to the pacemaker.
Therefore, the following should be observed:
– Select the smallest energy level possible for the
application.
– Do not apply the defibrillation paddles in the
vicinity of the pacemaker electrodes.
– The availability of an external pacemaker is
of utmost importance in this case.
– After the defibrillation the working order of
the implanted pacemaker should be checked
immediately.
Also be aware that children require less energy for
a successful defibrillation than adults. For the
first defibrillation pulse delivered to babies and
toddlers, select an energy level of 2 joules/kg body
weight. For subsequent shocks, increase the energy
up to 4 joules/kg.
When defibrillating children it is especially
important to verify that the paddles make full
Caution
contact with the body surface. This is to be ob-
served also when using the clip-on electrodes for
children (Part No. 303 439 95).
12 CardioServ V 4.2 227 446 32-I
Controls and Indicators
2. Controls and Indicators
This section describes the CardioServ operating controls and
indicators and explains their function.
When operating elements in this manual are identified with
a reference number in parentheses, this number refers to
Figure 2-1 in this section.
You will also find an explanation of all signs and symbols
used on the CardioServ defibrillator.
227 446 32-I CardioServ V 4.2 13
Controls and Indicators
-
+
1
2
Shock
3
STERNUM
4
J (50 )
102030
7
5
2
Autosequence
2
P
a
c
e
m
a
k
e
r
S
t
i
m
u
l
a
t
e
+
P/min
+
P/min
mA
mA
Dem
Fix
50
100
300
360
u
r
6
150
200
Shock
Charge
Shock
Sync
13 8 975
1 Select energy
Apply electrode cream to both paddles
2
Charge defibrillator
Press charge button on apex paddle (on the device, when using adhesive or internal electrodes)
Deliver shock3
Press buttons on both paddles simul­taneously (on the device, when using adhesive or internal electrodes)
APEX
Charge
Charge
Shock
3
2
APEX
1418 17 16 15
13
12
11
10
19
20
21
22
23
24
Option
Figure 2-1. Controls and indicators of CardioServ
14 CardioServ V 4.2 227 446 32-I
SpO
EKG
2
Controls and Indicators
-
+
1 Defibrillator paddles
Shock
2
button to trigger the defibrillation shock – together
3
with button (8)
3 keys to adjust the LCD contrast. Press both
keys simultaneously to obtain a screen copy.
4 Energy selector, on/off switch
Charge
5
key to charge the unit (manual mode) and to
Shock
trigger the defibrillation shock – together with key (6). This key assumes the function of button (8) when internal paddles or adhesive electrodes are used.
Shock
6
key to trigger the defibrillation shock – together
with key (5). This key assumes the function of button (2) when internal paddles or adhesive electrodes are used.
7 Five selection keys F1 to F5 whose functions change
with the menu displayed. The respective key functions are indicated by symbols or labels in the bottom line on the LCD. From the main menu, that appears on power up, you can access submenus which, in turn, allow the selection of further options. The back function returns you to the next higher menu level. The main menu reappears automatically if you do not depress any of the keys for about 30 seconds.
8
button to charge the unit (manual mode) and
3
2
Charge
Shock
trigger the defibrillation shock – together with button (2)
11 key to start and stop the recorder. This key assumes
the function of button (9) when internal paddles or adhesive electrodes are used.
12 Green indicator is lit when the defibrillator operates
on line power.
13 Yellow indicator
is lit when the defibrillator battery
is being charged.
14 Connection for defibrillator paddles
(controls 15 to 18 only on models with pacemaker)
+
mA
15 Keys
16 Keys
17 Pacing mode selection key
mA
+
P/min
to adjust the pacer output
-
P/min
to adjust pacer rate
-
Dem
Fix
(demand/fixed rate)
18 Key to enable and disable the pacemaker
19 Aperture to open the paper compartment
20 Connector for power cord
21 Unassigned
22 1-Volt ECG output
23 Connector for SpO2 sensor
9 button to start and stop the recorder.
Sync
10
key to switch to the synchronized operating mode
24 Connector for patient cable (ECG signal input)
(section 5 “Cardioversion”)
227 446 32-I CardioServ V 4.2 15
Controls and Indicators
-
+
Explanation of the signs and symbols used on the defibrillator
Type CF equipment with highly insulated patient
connections, suitable for intracardiac application,
connections defibrillation-proof.
Type CF equipment with highly insulated patient
connections, suitable for intracardiac application,
connections not defibrillation-proof.
Standby mode (for line-power operation)
Power off
Battery charging
Recorder start
ECG signal
Contrast
Standby or preparatory state only for a part of the
equipment
On, only for a part of the equipment
Hardcopy of screen image
This symbol indicates that the waste of electrical
Refer to Operators Manual
and electronic equipment must not be disposed as
unsorted municipal waste and must be collected
separately. Please contact an authorized representa-
tive of the manufacturer for information concern-
ing the decommissioning of your equipment.
Direction indicator
Signal output
Audible alarm on/off
High voltage
16 CardioServ V 4.2 227 446 32-I
Setting Up CardioServ and Testing Its Performance
3. Setting Up CardioServ and Testing Its Performance
In this section you will find information about
putting CardioServ into operation
connecting CardioServ to the 12-Volt power supply of
an ambulance vehicle
customizing the CardioServ settings to suit your per-
sonal requirements
testing the CardioServ performance before using it on a
patient
227 446 32-I CardioServ V 4.2 17
Setting Up CardioServ and Testing Its Performance
3.1 Setting Up CardioServ
The defibrillator is a high-voltage electrotherapy
device and must be handled by qualified person-
Danger
nel only. Improper use of this device can endanger
life. Do not fail to observe the information given
in this manual and only entrust the device to the
hands of trained persons.
Check the electrodes and their leads for signs of
damage every time before you use the defibrilla-
tor. In particular, make a close visual inspection
of the insulation. Replace internal electrodes or
the contact inserts when you detect signs of me-
chanical damage.
CardioServ operates on:
line power (95 to 240 V, 49 to 65 Hz)
battery power (rechargeable batteries), i.e., independent
of the power line
12-Volt power supplied from the emergency vehicle
(with optional defibrillator mounting system)
* Use the power cord to connect the defibrillator to the
power line (Figure 3-1).
Figure 3-1. Power input
18 CardioServ V 4.2 227 446 32-I
Setting Up CardioServ and Testing Its Performance
360
x
-
+
-
+
Shock
Charge
m
a
k
e
r
S
t
i
m
u
l
a
t
e
u
mA
mA
r
+
+
P/min
m
P/min
Shock
Sync
Figure 3-2. Green indicator (indicating that defibrillator is
supplied from the power line)
Figure 3-3. Inserting the battery
* Check that the green indicator is lit (Figure 3-2).
* Check that the battery is in place (Figure 3-3).
Pull back the catch on the underside of the device to remove
the battery. When inserting it, make sure that it clicks prop-
erly into place.
When a battery is inserted, the yellow indicator
starts
flashing as soon as the defibrillator is connected to the power
line (to indicate that the battery is charging). The battery is
fully charged after 16 hours and the indicator is continuously
lit.
* Check that the supply of chart paper is sufficient (Figure
3-4). A stripe marks the last 3 meters of the roll.
Figure 3-4. Recording strip
227 446 32-I CardioServ V 4.2 19
Setting Up CardioServ and Testing Its Performance
g
-
+
-
+
-
+
Power Supply From Emergency Vehicles
A qualified technician can be called in to connect the
CardioServ to the 12-Volt supply of an emergency vehicle.
The following points must be noted:
The negative terminal of the ambulance power supply
system must be connected to the ambulance chassis for
grounding.
The positive lead of the ambulance power supply system
intended to supply the current must be protected with a
10-A fuse.
T
e
>
s
9
t
5
>
%
8
0
%
<
8
0
%
T
e
S
s
ta
t
rt
Use only the defibrillator mounting system listed in
section 15 Order Information and Accessories or the
external charging unit, if your CardioServ has been
modified accordingly.
Check that the contacts for power supply from the
Figure 3-5. Accu Service Unit
defibrillator mounting system on the underside of
CardioServ (next to battery) are clean. Do not damage
We recommend our Accu Service Unit for optimal
them in any way.
care of the batteries. It prolongs the batteries service
life and guarantees their operational readiness at all
Important Information on Battery-Power Operation
times.
Rechargeable batteries require special maintenance and con-
A NiCd battery should not be charged while
Warnin
located in direct sunlight, over a radiator, in cold
storage, or in other temperature extremes (not
below 5 °C). When the instrument is charging,
ambient temperatures exceeding 40 °C may
adversely affect battery capacity and life.
20 CardioServ V 4.2 227 446 32-I
tinued checks to assure they function in emergency situations.
It is normal for batteries of this type to selfdischarge when not
in use.
The battery charges automatically when CardioServ is con-
nected to the power line (yellow indicator (13) flashes).
Setting Up CardioServ and Testing Its Performance
In order to ensure its functioning as an emergency device, the
defibrillator should not be disconnected from the power line
for more than 48 hours.
A fully charged battery supplies power for 35 defibrillation
shocks of 360 joules or 2 hours of monitoring (or 1.2 hours of
monitoring if the CardioServ unit has pacemaker and SpO
option). It takes 16 hours to charge a depleted battery.
Proceed as follows to test the battery charge level:
Disconnect CardioServ from the power line.
Trigger a test discharge of 360 joules (see
section 3.2 Testing the Defibrillator Perform-
ance).
If you are not prompted to charge the battery,
the charge level should be sufficient for at least
5 more 360-joule shocks.
Monthly battery maintenance and checks:
1. Disconnect CardioServ from the power line and
discharge fully charged battery in the monitoring
mode. To do so, set energy selector switch to ,
(SpO2 sensor not connected) and wait until device
switches off.
2
Inserting CardioServ In Its Softcase
* Open both zips on the CardioServ softcase.
* Undo the two Velcro strips on the front of the softcase.
* Open the Velcro flap located in front of the CardioServ
printer.
* Place the softcase on the small base, so that the softcase is
positioned as though you were carrying it.
* Disconnect the paddles on the CardioServ and remove
them.
* Hold all the opened up parts of the softcase to the side
and put the CardioServ into the softcase.
* Pull the softcase by the two Velcro strips on the black
base into position, ensuring that the feet of the
CardioServ are in the cut-out openings provided.
* Close the two front Velcro strips as well as the flap in
front of the printer at the side.
2. Check how long it takes before battery is depleted.
If the time is less than 1.8 hours, the battery is too
old or improperly maintained and must be
replaced.
3. Recharge the battery. This will take 16 hours.
For easy, convenient care and maintenance of the
batteries, use our Accu Service Unit”.
227 446 32-I CardioServ V 4.2 21
* Close the zips.
* If the individual cut-out openings for the external con-
nections, paddles or feet are not correctly positioned,
pull the material at this point into the right shape.
* Reconnect the paddles and put them back into the
CardioServ recesses.
Customizing the defibrillator settings
Setting Up CardioServ and Testing Its Performance
Further steps that can be taken while setting up the defibrilla-
tor include customizing the device functions. This allows you
to select defibrillator default settings which suit your personal
preferences. CardioServ saves these configured settings and
reactivates them automatically every time you switch the
Refer to section 11 The Defaults Menu for a detailed
explanation of how to customize the CardioServ, including
the language selection (available languages are English,
French, German, Italian, Portuguese, Russian, Spanish and
Swedish).
defibrillator on. The following chart shows the factory settings
of all parameters and the optional adjustments.
retemaraP noitpircseD gnitteSyrotcaF egnartnemtsujdA sgnittesdezimotsuC
stimiLRH stimilmralaRH 061/04 ffo003ot51ffo
ytivitisneS ezislangisGCE 1 Vm/mc2,1,5.
daeL 1 selddap+sdaeldradnatslla
.qesotuA slevelygrenegnicneuqesotua J063,J002,J002 ,J063,J003,J002,J051
OpS
stimiL stimilmrala ffo/09 ffo001ot51ffo
2
kcoL-C OpSdezinorhcnys-GCE
OpS
2
resU )sretcarahc04(emanrotxeteerf
emiT.tnI OpS
mralAliaFdaeL ffo s03/ffo
peeBSRQ ffo ffo/no
enoTmralA ffo ffo/no
tuotnirPmralA tratsredrocer.motua ffo ffo/no
tuotnirPkcohS tratsredrocer.motua no ffo/no
tuotnirP.tnoC tratslaunamnopupotsredrocer ffo ffo/no
rekamecaP etareslup MPB06 nim/P051ot51
yalpsiD lamron desrever/lamron
yalpsiD seerged0 seerged081/seerged0
emuloV slangisoiduallafoemulov duol wol/duol
retliFCA zH05 ffo/zH06/zH05
retliFelcsuM tcafitranoitomfonoitanimile no ffo/no
emiT/etaD
tamroFetaD YYYY:MM:DD YYYY:DD:MM/YYYY:MM:DD
egaugnaL namreG ,namreG,hcnerF,hsilgnE
tluafedyrotcaF sgnittesyrotcafhsilbatseer ffo ffo/no
2
2
emitnoitargetni s8 )dednemmocerton(s21,s8,s4
tnemerusaem ffo ffo/no
,naissuR,eseugutroP,nailatI
hsidewS,hsinapS
Table 1
22 CardioServ V 4.2 227 446 32-I
Setting Up CardioServ and Testing Its Performance
m
J (50 Ω)
20
10
7
5
2
Autosequence
P
a
c
e
m
a
k
e
r
S
t
i
Figure 3-6. Switching on CardioServ
a b
Paddle
c
0
160 BPM
140 ECG
HR Alarm
d
3.2 Testing the Defibrillator
30
50
100
150
200
300
360
Charge
Shock
Shock
Sync
e
f
Performance
On power up CardioServ runs an automatic selftest. Any
malfunctions identified during this test result in an error
message displayed on the LCD (refer to section 12 “Error
Indications and Messages). As a further performance test a
trial defibrillation can be triggered.
The energy selector is used to switch CardioServ on and off.
Once you have become familiar with CardioServ you can thus
switch on the defibrillator and select the required energy in
one single operation.
* Set the energy selector to the position. No energy
will be stored in this position of the switch.
The defibrillator beeps and displays a checkered pattern
(LCD performance test). Next the main display appears
(Figure 3-7).
g
The Paddle message (b, Figure 3-7) indicates that no pa-
tient cable is connected and that the ECG signal is acquired
via the defib paddles. Upon connection of the patient cable
FreezeSpO2ECG
Next
Menu
the selected ECG lead is displayed here. CardioServ is now
ready for operation.
Figure 3-7. CardioServ main display
a heart/pulse-rate reading with alarm limits b ECG signal source c 1-mV calibration pulse d alarm message, QRS blip e signal trace, channel 1 f signal trace, channel 2 g menu
227 446 32-I CardioServ V 4.2 23
Setting Up CardioServ and Testing Its Performance
Do not trigger more than 5 consecutive test
discharges (or internal safety discharges) within
Caution
15 minutes at max. energy setting.
Paddle
0
160 BPM
140 ECG
HR Alarm
Figure 3-8. Display of selected energy
360 J
Next
MenuFreezeSpO2ECG
Test Discharge
A test discharge can be triggered to check the defibrillator
discharge circuit. For this test the stored energy is discharged
into the device via two contacts in the paddle compartments.
* Set the energy selector to 360 joules (50 joules, if inter-
nal electrodes are connected). The display first shows the
selected energy (Figure 3-8).
2
Charge
* Press the
(Press the
Shock
button on the paddle to charge the unit.
3
Charge
key on the defibrillator, when using
Shock
internal electrodes.)
You may now watch the defibrillator charging.
* When the selected energy level has been reached,
CardioServ beeps and the stored energy is displayed
(Figure 3-9).
Do not deliver shocks into open air. High voltage
may briefly be present at the unprotected paddle
Caution
surfaces as a result and endanger the persons
onds. To do so, simultaneously press the buttons on
Charge
both paddles (or press the two
Shock
Shock
keys on the
defibrillator, when using internal electrodes).
present.
If more than 5% of the available energy are lost before the
defibrillation pulse is triggered, CardioServ recharges until the
required energy level is reached.
24 CardioServ V 4.2 227 446 32-I
* Trigger the defibrillation pulse within the next 30 sec-
Setting Up CardioServ and Testing Its Performance
Paddle
0
160 BPM
140 ECG
HR Alarm
Energy available
Figure 3-9. Display of available energy
If you do not trigger the defibrillation pulse within
30 seconds, an internal safety discharge is initiated
automatically.
360 J
Next
MenuFreezeSpO2ECG
After defibrillation, the beeping sound stops, and the energy
actually delivered into a 50-ohm resistance is displayed for
10 seconds in place of the stored energy (Figure 3-9). The
delivered energy must not deviate more than ±15% or
±4 joules (whichever is greater) from the selected value. A
recording is initiated at the same time (16-second strip).
Should the discharge circuit be interrupted (paddles not
properly placed on contacts in the compartments, defective
lead), an internal safety discharge is initiated 200 ms after the
defibrillation shock has been triggered. In this case the deliv-
ered energy is “0”.
If the defibrillator cannot store the selected energy so that
selected and stored energy values differ, the LCD shows the
message Energy high or Energy low. The defibrillation
pulse can be triggeed all the same.
* Switch off CardioServ (set energy selector to ).
The message Energy high or Energy low
indicates that CardioServ needs to be repaired.
Caution
If, in spite of this energy storage problem, the
device has to be employed, it will display the
message Self-test failed. Charge Energy Error
Testing the Pacemaker Performance
The performance of the pacemaker can be tested with a
commercially available pacemaker tester (e.g. CS300 Simula-
tor from GEMS IT, part no. 417 983-001).
upon power up. In this situation press one of the
function keys and proceed as usual.
Performance Test
Caution
Test the defibrillator performance once a week:
test 1 – defibrillator connected to mains, battery
removed,
test 2 – defibrillator disconnected from mains,
battery inserted
227 446 32-I CardioServ V 4.2 25
For your notes
Setting Up CardioServ and Testing Its Performance
26 CardioServ V 4.2 227 446 32-I
Non-Synchronized Defibrillation
4. Non-Synchronized Defibrillation
This section describes first how to perform a non-synchro-
nized defibrillation, using the standard defibrillation
paddles. The subsequent explanations refer to the use of
internal and disposable adhesive electrodes.
At the end of this section you will find a summary of all
necessary operating steps (brief operating instructions).
227 446 32-I CardioServ V 4.2 27
Non-Synchronized Defibrillation
g
4.1 Defibrillation with Standard
Electrodes
The information given in section 1.2 must be
Warnin
observed without fail to ensure safe and reliable
application of the device.
Always switch off CardioServ before exchanging
the defibrillation paddles.
Danger
Paddle
0
160 BPM
140 ECG
HR Alarm
200 J
SpO2ECG
Figure 4-1. Main display, indication of ECG lead, manual
operation and selected energy
Next
MenuFreeze
The energy selector can be set to the autosequence position,
where the defibrillator automatically sequences the preset
energy levels. The preset factory default setting is the sequence
recommended by AHA/ERC (200 J, 200 J, 360 J). The
factory default values can be changed from the defaults menu.
You can choose among 150 J, 200 J, 300 J and 360 J.
* Set the energy selector (4) to “Autosequence or to the
required energy value (this turns on CardioServ).
The defibrillator beeps and displays a checkered pattern (LCD
performance test). Next the main display appears (Figure
4-1). When a patient cable is connected the selected ECG
lead is displayed (selectable).
* Check that the energy selector locks in on the correct
position and that the display shows the selected energy
(Figure 4-1).
The energy depends on the defibrillation mode, on the
patients age and constitution. In external application the
thickness of the tissue is also a factor which influences the
amount of energy required.
The energy necessary for successful ventricular defibrillation
without damaging the myocardium has for many years been a
matter of scientific controversy. The manufacturer is therefore
not able to give any recommendations.
28 CardioServ V 4.2 227 446 32-I
Figure 4-2. Removing the paddles
g
Do not apply the paddles over
Warnin
sternum or clavicle
nipples
implanted pacemaker or defibrillator.
Non-Synchronized Defibrillation
In emergency situations AHA recommends for adult patients
1. defibrillation with 200 joules; if unsuccessful, repeat
2. defibrillation with 200 joules; if unsuccessful, repeat
3. defibrillation with max. energy setting (360 joules).
Please note that children require less energy for successful
ventricular defibrillation than adults. For the first defibrilla-
tion pulse delivered to babies and small children, select an
energy level of 2 joules/kg body weight. For subsequent
shocks, the energy may be increased to 4 joules/kg.
In compliance with IEC requirements the energy adjusted on
this defibrillator is not the stored energy, but the energy
released into an external resistance of 50 ohms (patient resist-
ance + electrode-to-skin contact resistance). The energy
selector is labelled accordingly.
* Remove the paddles from their compartments (as shown
in Figure 4-2). Carefully dry the electrodes, if they are
wet. The handles, in particular, must be completely dry.
Apply an ample amount of electrode cream to each
paddle.
3
STERNUM
* Apply the electrodes to the patients thorax so that the
greatest possible amount of energy flows through the
myocardium (the imaginary connecting line between the
2
Charge
3
APEX
two electrode centers should be identical with the cardiac
median line; Figure 4-3).
* Press paddles firmly down onto the patients thorax..
Figure 4-3. Paddle application points
227 446 32-I CardioServ V 4.2 29
Non-Synchronized Defibrillation
2
Charge
Shock
3
STERNUM
2
3
Charge
Shock
8
APEX
Figure 4-4. Buttons to initiate energy storage and to trigger the
defibrillation shock
Paddle
0
160 BPM
140 ECG
HR Alarm
The ECG signal trace appears on the monitor.
* Do not touch the patient any more and warn all those
present.
2
Charge
* Press the
Shock
button (8) on the apex paddle to initiate
3
charging (Figure 4-4). When using internal electrodes or
defibrillation pads, press
Charge
on the device.
Shock
When the selected energy level has been reached (message
Energy available), CardioServ beeps and the stored energy is
displayed (Figure 4-5).
* Trigger the defibrillation shock within the next 30 sec-
onds by simultaneously pressing the buttons on both
paddles (Figure 4-4). When using internal electrodes or
defibrillation pads, press
Charge
Shock
and
Shock
on the device.
If more than 5% of the available energy are lost before the
defibrillation shock is triggered, CardioServ recharges until
the required energy level is reached.
Energy available
FreezeSpO2ECG
200 J
Next
Menu
Figure 4-5. Display of available energy
When the defibrillator is already charged you can
increase the energy level simply by turning the energy
selector to the new setting. To decrease the energy
level, set the selector to the lower value and initiate
charging again.
30 CardioServ V 4.2 227 446 32-I
Loading...
+ 96 hidden pages