Datex-Ohmeda, Inc. a General Electric Company, doing business as
GE Healthcare.
This product will perform in conformity with the description thereof
contained in this User’s Reference manual and accompanying labels
and/or inserts, when assembled, operated, maintained, and repaired
in accordance with the instruction provided. This Product must be
checked periodically. A defective Product should not be used. Parts
that are broken, missing, plainly worn, distorted, or contaminated
should be replaced immediately. Should repair or replacement
become necessary, Datex-Ohmeda recommends that a telephonic or
written request for service advice be made to the nearest DatexOhmeda Customer Service Center. This Product or any of its parts
should not be repaired other than in accordance with written
instructions provided by Datex-Ohmeda and by Datex-Ohmeda
trained personnel. The Product must not be altered without the prior
written approval of Datex-Ohmeda. The user of this Product shall
have the sole responsibility for any malfunction which results from
improper use, faulty maintenance, improper repair, damage, or
alteration by anyone other than Datex-Ohmeda.
CAUTION
U.S. Federal law restricts this device to sale by or on the
order of a licensed medical practitioner.Outside the
U.S.A., check the local laws for any restriction that may
apply.
1 Introduction
Table of Contents
Intended use .....................................1-2
Indications for use ..............................1-2
General information .............................1-2
Serial numbers .................................1-4
Trademarks ...................................1-5
Symbols used in the manual or on the equipment ........1-6
Symbols used on the equipment ...................1-6
Symbols used on the user interface ................1-9
Typeface conventions used ........................ 1-10
System information...............................1-14
System safety...................................1-16
2067226-0011-1
Aisys CS²
Intended use
The Aisys CS2 is scalable, flexible, and functionally integrated,
featuring the most advanced design, ventilation, respiratory
monitoring, and breathing system. Module bays allow for the physical
integration of legacy Datex-Ohmeda patient monitors and supports
mounting of other GE Healthcare monitors. Optionally, the open
architecture design supports mounting of non-Datex-Ohmeda patient
monitors, record keeping, and connections to the hospital information
system. The INview movable display arm helps keep the
anesthetist’s focus on the patient by offering control of all
hemodynamic, gas delivery, anesthetic agent, and ventilation
parameters.
This anesthesia system is designed for mixing and delivering
inhalation anesthetics, Air, O2, and N2O.
The anesthesia system’s small breathing system volume allows an
anesthetic agent delivery response time of less than 7 seconds.
(Agent response time defined as achieving 90% of the setting
change measured while in non-circle circuit with a 2 l/min fresh gas
flow.)
This anesthesia system uses SmartVent ventilation technology
offering Volume Control Ventilation with tidal volume compensation
and electronic PEEP. The proven SmartVent technology also
features optional Pressure Control Ventilation, Pressure Support
Ventilation with an Apnea Backup (PSVPro) that is used for
spontaneously breathing patients, Synchronized Intermittent
Mandatory Ventilation (SIMV) modes, Pressure Control VentilationVolume Guarantee (PCV-VG), Continuous Positive Airway Pressure
+ Pressure Support Ventilation (CPAP + PSV), and VCV Cardiac
Bypass. In Volume Control Ventilation, a patient can be ventilated
using a minimal tidal volume of 20 ml. In Pressure Control
Ventilation, volumes as low as 5 ml can be measured. These
advanced features allow for the ventilation of a broad patient range.
WARNING
This system is not suitable for use in an MRI
environment.
Indications for use
The system is intended to provide general inhalation anesthesia and
ventilatory support to a wide range of patients (neonate, pediatric,
and adult). The device is intended for volume or pressure control
ventilation.
General information
This anesthesia system uses the Advanced Breathing System
(ABS). This integrated breathing system is easy to remove and
disassemble and is autoclavable. Its fully integrated design
1-22067226-001
1 Introduction
enhances the system’s elegance while minimizing tube connections,
minimizing circuit volume, and increasing the work surface area.
This anesthesia system is designed for expansion and upgrades, so
it is easy to add new technologies and ventilation capabilities without
investing in a new system.
The anesthesia system is suitable for use in a patient environment.
The system must only be operated by personnel trained and
qualified in the administration of general anesthesia. The User’s
Reference manual is intended to provide training on the operation of
the system. Operate the system from the front with a clear view of
the display. It must be operated according to the instructions in this
User’s Reference manual. Make sure that all user documents are
obtained from the manufacturer.
Refer to the Technical Reference manual for service information
including: special installation instructions, installation checklist,
means of isolating the supply mains, and replacement of fuses,
supply cord, and other parts.
WARNING
Note
Explosion Hazard. Do not use this system with flammable
anesthetic agents.
Configurations available for this product depend on local market and
standards requirements. Illustrations in this manual may not
represent all configurations of the product. This manual does not
cover the operation of every accessory. Refer to the accessory
documentation for further information.
2067226-0011-3
Aisys CS²
AB.75.253
Figure 1-1 • Front view
Serial numbers
Datex-Ohmeda products have unit serial numbers with coded logic
which indicates a product group code, the year of manufacture, and
a sequential unit number for identification. The serial number can be
in one of two formats.
1-42067226-001
Trademarks
1 Introduction
AAAX11111AAAXX111111AA
The X represents an alpha character
indicating the year the product was
manufactured; H = 2004, J = 2005,
etc. I and O are not used.
Aisys, Carestation, Advanced Breathing System, ComWheel, Dfend, EZchange, Disposable Multi Absorber, Reusable Multi
Absorber, PSVPro, SmartVent, Easy-Fil, Aladin, and Aladin2 are
registered trademarks of Datex-Ohmeda, Inc.
Other brand names or product names used in this manual are
trademarks or registered trademarks of their respective holders.
The XX represents a number
indicating the year the product was
manufactured; 04 = 2004, 05 = 2005,
etc.
2067226-0011-5
Aisys CS²
Symbols used in the manual or on the equipment
Symbols replace words on the equipment, on the display, or in
product manuals.
Warnings and Cautions tell you about dangerous conditions that can
occur if you do not follow all instructions in this manual.
Warnings tell about a condition that can cause injury to the operator
or the patient.
Cautions tell about a condition that can cause damage to the
equipment. Read and follow all warnings and cautions.
Symbols used on the equipment
.
O2 Flush buttonO2 cell connection
AirAir
APL settings are approximateAnesthetic Gas Scavenging System
MaximumVacuum
Serial numberStock number
ExhaustBellows volumes are approximate
Plus, positive polarityMinus, negative polarity
Maximum mass of configured mobile
equipment
Autoclavable
On (power)Off (power)
StandbyInterference
Type BF equipmentType B equipment
Dangerous voltageFrame or chassis ground
Protective earth groundEarth ground
Direct currentAlternating current
CautionRefer to instruction manual or booklet
Caution: federal law prohibits
dispensing without prescription.
Electrical input/outputSample gas inlet to scavenging
Pneumatic inletPneumatic outlet
EquipotentialLamp, lighting, illumination
VariabilityVariability in steps
Suction bottle outletVacuum inlet
Bag position/manual ventilationMechanical ventilation
Inspiratory flowExpiratory flow
Movement in one directionMovement in two directions
LockUnlock
Open drain (remove liquid)Low pressure leak test
This way upNot autoclavable
Read to center of floatEZchange canister (CO2 bypass)
Systems with this mark agree with the
European Council Directive (93/42/
EEC) for Medical Devices when they
are used as specified in their User’s
Reference manuals. The xxxx is the
certification number of the Notified
Body used by Datex-Ohmeda’s Quality
Systems.
Date of manufactureManufacturer
Authorized representative in the
European Community
2067226-0011-7
Aisys CS²
Indicates that the waste of electrical
and electronic equipment must not be
disposed as unsorted municipal waste
and must be collected separately.
Please contact an authorized
representative of the manufacturer for
information concerning the
decommissioning of equipment.
anesthesia machine, place the display
arm and shelf in the transport position
as shown.
Single use device
Keep dryDo not stack
Temperature limitationFragile, handle with care
Protect from heat and radioactive
sources
Humidity limitationElectrical test certification
USB portNetwork
Not a USB portVGA connection
Serial connectionPinch hazard
Recyclable material
1-82067226-001
Atmospheric limitation
Symbols used on the user interface
.
1 Introduction
Lock
Indicates the touchscreen is locked.
Lock/unlock button
Button label to lock or unlock the
touchscreen.
O2% indicator on left and balance gas
indictor on right. Colors associated with
Gas indicator. Color associated with gas
settings.
gas settings.
Audio PauseSubmenu
No battery/battery failureBattery in use. Bar indicates amount of
battery power remaining.
Airway module indicatorACGO active
Drop-down menuStart/end case
PediatricAdult
Lung procedureHome screen
TimerAlarm off
PipelineCylinder
Test indicator: red for failure, yellow for
Alarm low and alarm high limit indicator
conditional outcome, and green for
pass.
Manual ventilationEnhanced temperature sensing
Agent level sensing supported. Bar
Agent level unknown
indicates amount of agent remaining.
2067226-0011-9
Aisys CS²
Typeface conventions used
Soft keys and menu items are written in bold italic typeface; for
example, System Setup.
Messages that are displayed on the screen are enclosed in single
quotes; for example, ‘Check sample gas out’.
When referring to different sections and other documents, the names
are written in italic typeface and enclosed in double quotes; for
example, "System controls and menus".
1-102067226-001
Abbreviations
1 Introduction
AbbreviationDefinition
A
AAAnesthetic agent
ABSAdvanced breathing system
ACGOAuxiliary common gas outlet
AGSSAnesthesia gas scavenging system
Alt O2Alternate O2
APLAdjustable pressure-limiting
APNApnea
ATPDAmbient temperature and pressure, dry humidity
TexpExpiratory time
TinspInspiratory time
TpauseTime where breath is paused with no flow
TVTidal volume
TVexpExpired tidal volume
TVinspInspired tidal volume
V
1-122067226-001
AbbreviationDefinition
VCVVolume control ventilation
VolVolume
1 Introduction
2067226-0011-13
Aisys CS²
System information
System classification
Device standards IEC 60601-1:2005
This system is classified as follows:
•Class I Equipment.
•Type B Equipment.
•Type BF Equipment (airway modules).
•Ordinary Equipment.
•Not for use with flammable anesthetics.
•Continuous operation.
Devices used with this anesthesia system shall comply with the
following standards where applicable:
•Breathing system and breathing system components ISO
80601-2-13.
•Anesthetic gas scavenging systems ISO 80601-2-13
•Anesthetic vapor delivery devices ISO 80601-2-13.
•Anesthetic agent monitors ISO 80601-2-55.
•Oxygen monitors ISO 80601-2-55.
•Carbon dioxide monitors ISO 80601-2-55.
•Exhaled volume monitors ISO 80601-2-13.
Device standards IEC 60601-1:1988
Devices used with this anesthesia system shall comply with the
following standards where applicable:
•Breathing system and breathing system components ISO
8835-2.
•Anesthetic gas scavenging systems ISO 8835-3
•Anesthetic vapor delivery devices ISO 8835-4.
•Anesthetic agent monitors ISO21647.
•Oxygen monitors ISO 21647.
•Carbon dioxide monitors ISO 21647.
•Exhaled volume monitors IEC 60601-2-13.
1-142067226-001
Integral system components
This anesthesia system contains the following integral components,
monitoring devices, alarm systems, and protection devices that
comply with European, international, and national standards:
•Breathing system pressure-measuring device.
•Airway pressure-limitation device.
•Exhaled-volume monitor.
•Breathing system integrity alarm.
•Breathing system continuing-pressure alarm.
•O2 monitor (optional O2 cell).
•Anesthesia ventilator.
•Breathing system.
•Anesthetic vapor delivery device.
Not integral system components
1 Introduction
System accessories
These devices are not integral to this anesthesia system:
•CO2 monitor.
•Anesthetic agent monitor.
•O2 monitor (when O2 cell is not installed).
•Suction regulator.
•EZchange canister system.
•Condenser.
When adding devices to the anesthesia system, follow the
installation instructions provided by the device manufacturer.
Whoever adds individual devices to the anesthesia system shall
provide instructions on how to enable the individual devices. For
example, a preoperative checklist.
These devices can be used as accessories on this anesthesia
system:
•Suction regulator.
•EZchange canister system.
•Condenser.
•Aladin and Aladin2 vaporizer cassettes.
2067226-0011-15
Aisys CS²
System safety
Preparing for use
WARNING
Read each component’s User’s Reference manual and
understand the following before using this system:
•All system connections.
•All warnings and cautions.
•How to use each system component.
•How to test each system component.
Before using the system:
•
•Complete all of the tests in the "Preoperative tests"
section.
•Test all other system components.
•If a test fails, do not use the equipment. Have an
authorized service representative repair the
equipment.
European, international, and national standards require
•
the following monitoring be used with this system:
•Exhaled volume monitoring.
•O2 monitoring.
•CO2 monitoring.
•Anesthetic agent monitoring be used when anesthetic
vaporizers are in use.
Single-use products are not designed or validated to be
•
reused. Reuse may cause a risk of cross-contamination,
affect the measurement accuracy, system performance,
or cause a malfunction as a result of the product being
physically damaged due to cleaning, disinfection, resterilization, or reuse.
Be aware of the risks and precautionary measures
•
related to phthalates. The following types of procedures
may increase the risk of exposure to phthalates when a
1-162067226-001
1 Introduction
device containing phthalates is used for treatment of
children or treatment of pregnant or nursing women:
•Exchange transfusion in neonates, total parenteral
nutrition in neonates, multiple procedures in sick
neonates, haemodialysis in peripuberal males, male
fetus and male infant of pregnant women, and
lactating women; and massive blood infusion into
trauma patients. Although these procedures have the
potential for increased risk of exposure, conclusive
evidence of human health risks has not been
established. As a precautionary measure, to reduce
the potential for unnecessary exposures to
phthalates, the product must be used in accordance
with the instructions for use, and practitioners should
refrain from using this product beyond the period of
time the product is medically necessary or needed.
Follow hospital procedures for the prevention and
•
treatment of malignant hyperthermia for patients sensitive
to inhalation anesthetic agents.
Risk of fire. Limit the use of supplemental oxygen
•
concentrations to less than 30 percent when using a heat
source or device that may lead to combustion. Consult
facility risk management procedures to minimize the risk
of fire if an oxygen concentration of more than 30 percent
is required for any reason.
This system is not intended for use where the
•
surrounding oxygen concentration is in excess of 25
percent. Increased oxygen concentrations can result in
an increased risk of fire.
See "Device standards IEC 60601-1 2005" and "Integral systemcomponents" for information regarding monitoring built into this
device.
Inspecting the system
Before using the system, make sure that:
•The equipment is not damaged.
•Components are correctly attached.
•The breathing circuit is correctly connected and not damaged.
•The breathing system is correctly assembled and contains
•The Aladin cassette is locked in position and contains sufficient
sufficient absorbent. Refer to the “Cleaning and Sterilization”
manual for breathing system assembly instructions.
agent.
2067226-0011-17
Aisys CS²
•Pipeline gas supplies are connected and the pressures are
correct.
•Cylinder valves are closed.
•Models with cylinder supplies have a cylinder wrench attached
to the system.
•Models with cylinder supplies have a reserve supply of O2
connected to the machine during system checkout.
•The necessary emergency equipment is available and in good
condition.
•Equipment for airway maintenance, manual ventilation, tracheal
intubation, and IV administration is available and in good
condition. In the case of system failure, the lack of immediate
access to alternative means of ventilation can result in patient
injury.
•Applicable anesthetic and emergency drugs are available.
•Check that the brake is set to prevent movement.
•The power cord is connected to an electrical outlet. The mains
indicator comes on when AC power is connected. If the indicator
is not on, the system does not have mains (electrical) power.
Use a different outlet, close the circuit breaker, or replace or
connect the power cable.
•If an optional suction regulator is present, ensure there is
adequate suction.
•If an optional O2 flowmeter is present, ensure there is adequate
flow.
Electrical safety
WARNING
Do not connect non-medical electrical equipment directly to the AC
outlet at the wall instead of an AC power source which uses a
separating transformer. Doing so may increase enclosure leakage
current above levels allowed by IEC 60601-1 in normal conditions
and under single-fault conditions. This may cause an unsafe
electrical shock to the patient or operator.
After connecting anything to these outlets, conduct a complete
system leakage current test (according to IEC 60601-1).
The system provides connections for items such as
printers, visual displays and hospital information
networks (only connect items that are intended to be part
of the system). When these items (non-medical
equipment) are combined with the system, these
precautions must be followed:
•Do not place items not approved to IEC 60601-1
closer than 1.5 m to the patient.
•All items (medical electrical equipment or nonmedical electrical equipment) connected to the
system by a signal input/signal output cable must be
1-182067226-001
1 Introduction
supplied from an AC power source which uses a
separating transformer (in accordance with IEC
60989) or be provided with an additional protective
earth conductor.
•If a portable multiple socket outlet assembly is used
as an AC power source, it must comply with IEC
60601-1-1. The assembly must not be placed on the
floor. Using more than one portable multiple socket
outlet assembly is not recommended. Using an
extension cord is not recommended.
An operator of the medical electrical system must not
•
touch non-medical electrical equipment and the patient
simultaneously. This may cause an unsafe electrical
shock to the patient.
•
Use of portable phones or other radio frequency (RF)
emitting equipment (that exceed electromagnetic
interference levels specified in IEC 60601-1-2) near the
system may cause unexpected or adverse operation.
Monitor operation when RF emitters are in the vicinity.
•
Use of other electrical equipment on or near this system
may cause interference. Verify normal operation of
equipment in the system before use on patients.
2067226-0011-19
1-202067226-001
2 System controls and menus
System controls and menus
In this section
System overview..................................2-2
Advanced breathing system components.............. 2-6
Use the optional bag support arm to hold the breathing circuit bag.
AB.82.023
1.To raise the bag support arm, squeeze the button and rotate the
arm up the top position.
2.To lower the bag support arm, squeeze the button and rotate the
arm down to the lower position.
2067226-0012-7
Aisys CS²
Aladin cassette controls
The electronically controlled vaporizer consists of the internal
electronic control unit and the Aladin agent cassette. See the
"Vaporizer" section for more information.
1
2
AB.60.038
1.Handle with release trigger
2.Lock
3.Liquid level indicator
4.Agent filling port
Figure 2-5 • Aladin2 cassette
34
2-82067226-001
Display controls
2 System controls and menus
The system uses touchscreen technology, hard keys, and a
ComWheel to access system functions, menus, and settings.
The touchscreen has numerous touch point areas that make
accessing menus and settings quick and easy. The buttons on the
right side of the screen provide direct access to commonly used
functions. The ventilation quick keys enable setup of ventilation
modes. The gas control quick keys provide a method to set up the
gas used for a case.
Touch only one touch point at a time to ensure the correct selection
is made.
WARNING
CAUTION
Liquids on the display may degrade the performance of
the touchscreen. If liquids come in contact with the
display, lock the touchscreen and clean the display.
Unlock the touchscreen once the display has been
cleaned to resume use of the touchscreen.
Do not apply excessive force to the touchscreen as
damage may occur.
AC.22.001
6543
1.ComWheelSelects a menu item or confirms a setting. Turn clockwise or counterclockwise to scroll
through menu items or change settings.
2.Home keyRemoves all menus from the screen.
3.Screen Lock/Unlock
key
4.Start/End Case key Initiates Start or End Case function.
5.TouchscreenActivates functions when touch areas on the screen are selected.
2067226-0012-9
Locks the touchscreen. Toggles between lock and unlock functions.
1
2
Aisys CS²
6.Audio Pause keyStops audio for 120 seconds for any active, eligible high and medium priority alarms.
Prevents audio (audio off) for 90 seconds when no medium or high priority alarms are
active. Allows the operator to acknowledge any non-active medium or high priority latched
alarms.
Figure 2-6 • Display controls
Touch points
21
7
56
1.Wave fields5.Ventilator quick keys
2.Measured values6.Gas quick keys
3.Function keys7.Split screen values
4.Digit fields
Figure 2-7 • Normal/Full screen view with shaded touch point areas
3
4
AB.75.256
Measured value touch points
Touching measured values provides access to the Alarm Setup
menu and alarm limits.
2-102067226-001
1.Touch the measured value to access the Alarm Setup menu.
2.The Alarm Setup menu displays.
3.Select the alarm limit and set it to the correct value. Touch the
value on the touchscreen or push the ComWheel to confirm the
desired setting.
4.Push the Home key, touch the waveform area of the display, or
select Close to close the menu.
Active alarm touch points
When an alarm sounds the alarm message is displayed at the top of
the screen and, if applicable, the alarming numeric field and digit
field flashes. The Alarm messages at the top of the screen are
message alerts only and not active touch points.
1.Touch the flashing numeric field to access the Alarm Setup
menu and alarm limits for the active alarm.
2.The Alarm Setup menu displays with the active alarm limit
highlighted. For example: If the ‘Ppeak high’ alarm activates, the
high alarm limit setting for Ppeak displays with the highlight.
3.Select the active alarm limit and change it to the desired setting.
2 System controls and menus
2067226-0012-11
Aisys CS²
Anesthesia system display
12
21
3
56
7
8
9
11
1.Audio pause symbol and
countdown clock
2.Alarm message fieldsDisplays the active alarms.
3.Waveform fieldsDisplays the waveforms of measured values. For example: Paw, Flow, and
4.General message fields or lock
touchscreen indicator
5.Measured values fieldsDisplays the measured values. For example: Paw, Flow, and CO2.
6.ClockDisplays the current time.
7.Function keysFunctions available are: Audio Pause, Alarm Setup, Alarms On/Off, Auto
8.Digit fieldsContains information for Spirometry, Resp, Agent, and Gases.
9.Ventilation modeDisplays the selected ventilation mode. For example: Ventilator On, and
10. Ventilator quick keysDisplays Mode, associated ventilation parameters, and More Settings. For
Indicates when alarm audio is paused and the countdown clock until audio is
on.
CO2.
Displays general messages and the touchscreen lock indicator.
Limits, System Setup, Next Page, Trends, Spirometry, Procedures, Timer,
Start, and End Case.
Volume Control.
example: Mode, TV, RR, I:E, PEEP, and More Settings.
10
AB.75.258
2-122067226-001
2 System controls and menus
11. Gas quick keysDisplays O2, Total Flow, and Gas Setup.
12. Split screenContains airway pressure, gas flow values, compliance, trends, and optional
ecoFLOW information.
Figure 2-8 • Typical Normal/Full view
Digit fields
The digit field can be set to show specific information such as gas
types, gas supply, flow, agent, respiration, and spirometry loops. If
the digit field is set to show agent and no airway module is inserted,
the area is blank.
Paw, O2, and either TVexp or CO2 must show on the display during
a case. If any of these parameters are not selected to show on the
display, the right most digit field information is replaced with the
missing parameter.
See "Screen setup menu" in the "Operation" section for more
information.
Waveform fields
Split screen field
Up to three waveforms can be shown on the normal screen view.
Each waveform can be set to show specific Paw, agent, flow, or CO2
data. The corresponding numeric information shows in the measured
values field to the right of the waveform. If the waveform is set to
show the agent and no airway module is inserted, that waveform and
numeric area is blank.
When one waveform is turned off, that waveform and the
corresponding numerics information are removed from the normal
screen view. The remaining waveforms and numerics increase in
size to fill the waveform area. When two waveforms are turned off,
those waveforms and the corresponding numerics information are
removed from the normal screen view. The remaining waveform is
centered in the waveform area.
When in a case, touch the waveform field area to close the menu.
See "Screen setup menu" in the "Operation" section for more
information.
The split screen field can be set to show gas metabolics, trends,
spirometry loops, Paw gauge, airway compliance, and optional
ecoFLOW information. If None is selected, the waveforms expand to
fill the split screen area.
2067226-0012-13
Aisys CS²
Touch the spilt screen field to directly open the Screen Setup menu.
See "Screen setup menu" in the "Operation" section for more
information.
2-142067226-001
Display navigation
A
B
.
7
5
.
2
5
7
3
2
1
2 System controls and menus
Use the touchscreen and ComWheel to navigate the display.
1.MenuDisplays the title of the open menu. For example: Start Case.
2.Instructions or help informationThis shows any additional instructions or help messages.
3.Menu itemsShows Case Defaults, Volume Apnea Alarm, CO2 Alarms, Age, Ideal
Weight, and Start Case Now.
Figure 2-9 • Menu view and menu example
Using menus
Use the function keys to access the corresponding menus. When a
menu is selected, the menu field overlays the normal view and the
waveform fields start at the right edge of the menu.
1.Select the menu key to access the corresponding menu.
2067226-0012-15
Aisys CS²
Using the ComWheel
2.Select a menu item to choose the item, or turn the ComWheel
left or right to highlight a menu item and then push to confirm.
3.If the menu item selected is an adjustment, turn the ComWheel
left or right to make the setting and then push to confirm.
If the menu item has a drop-down list, select the desired value
from the list by touching the item.
4.Select Close, touch the waveform area, or push the Home key
to exit the menu.
Use the ComWheel to scroll through the quick key settings and
function keys, make selections, change settings, and confirm
settings.
•Push the ComWheel to make a selection.
•Turn the ComWheel to the right.
For menu items, the highlight moves down.
For quick keys, the highlight moves to the next key on the right.
For settings, the value changes to the next available setting.
For pull-down selections, the highlight moves to the next
available selection.
•Turn the ComWheel to the left.
For menu items, the highlight moves up.
For quick keys, the highlight moves to the next key on the left.
For settings, the value changes to the previous available setting.
For pull-down selections, the highlight moves to the previous
available selection.
•Push the ComWheel to confirm a setting.
Using quick keys
The gas settings and the main ventilator settings for each ventilation
mode can be changed using the quick keys.
1.Select a quick key to open the menu or select a parameter.
2.If Gas Setup, Mode, or More Settings is selected, a menu
displays. Select the desired value on the menu by touching the
value.
If any other quick key is selected, the value displays with a
highlight. Turn the ComWheel left or right to set the desired
value.
3.Push the ComWheel or select the quick key to confirm the
change.
2-162067226-001
3 Operation
Operation
In this section
System operation safety............................3-2
Turning on the system............................. 3-3
Start a case......................................3-4
Turning off the system............................. 3-7
Investigate all alarms that occur to help ensure adequate
patient safety.
If an alarm occurs, safeguard the patient first before
•
performing troubleshooting or doing repair procedures.
Failure to safeguard the patient could result in patient
injury.
Make sure that the patient breathing circuit is correctly
•
assembled and that the ventilator settings are clinically
appropriate before starting ventilation. Incorrect breathing
circuit assembly and incorrect ventilator settings can
injure the patient.
Make sure that the breathing circuit is correctly
•
connected and not damaged. Replace the breathing
circuit if it is damaged.
Maintain sufficient fresh gas flow when using
•
sevoflurane.
Desiccated (dehydrated) absorbent material may
•
produce dangerous chemical reactions when exposed to
inhalation anesthetics. Adequate precautions should be
taken to ensure that absorbent does not dry out. Turn off
all gases when finished using the system.
Do not leave gas cylinder valves open if the pipeline
•
supply is in use. Cylinder supplies could be depleted,
leaving insufficient reserve supply in case of pipeline
failure.
Unplug the system power cord to run the system on the
•
battery power if the integrity of the protective earth
conductor is in doubt.
The top shelf weight limit is 45 kg (100 lb).
•
Do not subject the system to excessive shock and
•
vibration. Equipment damage could occur.
Do not place excessive weight on flat surfaces or
•
drawers. Equipment damage could occur.
3-22067226-001
Turning on the system
1.Plug the power cord into an electrical outlet. Make sure the
3 Operation
system circuit breaker is on.
•The mains indicator is lit when AC power is connected.
•Battery is charging if it is not already fully charged.
1
2
CAUTION
Note
AB.91.044
1.System switch
2.Mains indicator
Figure 3-1 • Mains indicator and system switch
2.Check that the breathing system is properly connected.
Do not turn on the system with the right-hand
(inspiratory) port plugged.
--
3.Turn the System switch to On.
The display shows the power-up screen.
The system does a series of automated self tests.
4.Perform a Full Test before the first case of the day.
5.Perform a preoperative checkout before each case. See the
"Preoperative checkout" section.
The system must perform a power-up self test after 12 hours of
remaining on. If the system has been on longer than 12 hours
without a power-up self test, the ‘Turn power Off and On for self
tests’ alarm occurs. Turn the power off and then back on between
cases to resolve the alarm.
--
2067226-0013-3
Aisys CS²
Start a case
Use the Start Case menu to set the case data and to start the gas
flow.
A case can be started using default settings or using custom
settings. The default settings are configured by the Super User. See
the "Super user mode" section for information on the Start Case
menu defaults.
Default Settings selection shows the first of five default case types
when the Start Case menu is accessed. Four of the default case
types are configured by the Super User. The fifth default case is Last
Case.
The Ideal Weight, Age, and Volume Apnea Alarm values are set to
the pre-selected settings defined by the Super User corresponding to
the case type.
WARNING
Make sure that the patient breathing circuit is correctly
assembled and that the ventilator settings are clinically
appropriate before starting ventilation. Incorrect breathing
circuit assembly and incorrect ventilator settings can
injure the patient.
Make sure that the preset alarm limits are appropriate for
•
the patient before starting ventilation. Incorrect alarm
settings can injure the patient.
Note
Note
Volume Apnea Alarm is not shown on the Start Case menu when
the Volume Apnea Selection is set to Disable in the Super User
settings.
The TV for Ideal Body Weight menu item from the PatientDemographics menu can only be accessed when the ventilation
mode is set to VCV, PCV-VG, SIMV VCV, and SIMV PCV-VG. Use
this setting for breath rate and tidal volume calculations based on the
set patient weight.
Minimum Alveolar Concentration
The adjusted Minimum Alveolar Concentration (MAC) is calculated
based on the patient age entered in the Start Case menu or the
Patient Demographics menu. The default patient age of the
selected case type is used if no patient age value is entered.
The MAC value is calculated from the exhaled gas concentration and
the related affects based on the age of the patient. Typically,
younger patients have better liver function and can clear a drug
faster, resulting in a higher MAC value. The MAC calculation used is
based on the Eger formula. When two agents are detected, the MAC
values of each agent are added together. The MAC value range is
0.0 to 9.9.
3-42067226-001
The adjusted MAC value shows on several areas of the screen
including in the mini-trend, agent waveform numeric information,
agent digit field, and graphical trends page.
Starting a case using default settings
Start a case using the default settings by case type defined by the
Super User.
Case Defaults contain five case type selections. Each case type has
preset values for Ideal Weight, Age, and Volume Apnea Alarm.
The first four default case types are configured and named by the
Super User. The fifth default case is Last Case.
1.Set the Bag/Vent switch to Bag.
2.Select Start Case.
The Case Defaults selection shows the first preset case type.
Ideal Weight, Age, CO2 Alarms, and Volume Apnea Alarm
show the default settings that correspond to the case type
shown.
3.Verify or change the Case Defaults selected.
4.Verify the settings are clinically appropriate.
5.Select Start Case Now. Gas flow starts.
3 Operation
Starting a case using custom settings
Ideal Weight, Age, CO2 Alarms, and Volume Apnea Alarm can be
custom set on the Start Case menu before starting a case.
Additional ventilator settings, ventilation mode, alarm settings, and
gas settings can be custom set through the Vent Mode menu and
other ventilation quick keys, Alarm Setup menu, Gas Setup menu.
1.Set the Bag/Vent switch to Bag.
2.Select Start Case.
The Case Defaults selection shows the first preset case type.
Ideal Weight, Age, CO2 Alarms, and Volume Apnea Alarm
show the default settings that correspond to the case type
shown.
3.Change Ideal Weight, Age, or Volume Apnea Alarm settings
on the menu.
The Case Defaults changes from the case name to Preset.
If the CO2 Alarms setting on the menu is changed, the Case
Defaults remains as previously selected.
2067226-0013-5
Aisys CS²
End a case
4.To change ventilation mode, select Mode. Make the change.To
change the ventilation settings, select a ventilator quick key or
More Settings. Make the change.
5.To change alarm settings, select Alarm Setup. Make the
change.
6.To change the gas settings or the circuit type, select Gas Setup.
Make the change.
7.From the Start Case menu, select Start Case Now. Gas flow
starts.
See the "Ventilator setup" section for information on the Vent Mode
menu.
See the "Gas setup" section for information on the Gas Setup menu.
See the "Alarm setup" section for information on the Alarm Setup
menu.
Use the End Case menu to stop gas flow and end the patient
alarms.
1.Set the Bag/Vent switch to Bag.
2.Select End Case.
3.Select End Case Now on the menu to put the system in standby
(stops the gas flow and patient alarms). The End Case menu
shows the gas and agent usage for the case.
3-62067226-001
Turning off the system
1.Perform the "End a case" procedure, if appropriate.
2.Turn the System switch to Standby.
3.Turn the suction switch (optional) to the off position.
4.Rotate the Auxiliary O2 knob fully clockwise to turn off the flow.
5.Disconnect or turn off any scavenging.
3 Operation
2067226-0013-7
Aisys CS²
Ventilator setup
Use the Vent Mode menu to set the ventilation mode. Use ventilator
quick keys and More Settings to change ventilator settings.
WARNING
Most anesthetic agents will cause patients to have
reduced ventilatory responses to carbon dioxide and to
hypoxemia. Therefore, triggered modes of ventilation
may not produce adequate ventilation.
The use of neuromuscular blocking agents will reduce
•
the patient’s breathing response, which will interfere with
triggering.
Important
See the "Specifications and theory of operation" section for more
information on ventilation modes.
Changing ventilator mode
1.Select the Mode quick key. The Vent Mode menu shows.
2.Select the desired ventilation mode.
3.Set and confirm the primary ventilation setting to activate the
ventilation mode.
Controls that are frequently used in the ventilation mode can be
adjusted with the ventilator quick keys and the More Settings
quick key.
Changing ventilator settings
Change the ventilator settings for the ventilation mode when a case
is running.
1.Select the ventilation setting to be adjusted. Set the desired
value.
2.Push the ComWheel to activate the change.
Optional ventilator procedures
See "Procedures" for more information on Vital Capacity and
Cycling procedures.
3-82067226-001
Auto limits
Setting auto limits
3 Operation
Use the Auto Limits menu to quickly set alarm ranges for ‘MV’, ‘TV’,
and ‘EtCO2’ during mechanical ventilation.
1.Select Auto Limits.
The menu shows the current measured values and the
proposed low and high alarm limits.
2.Check the proposed parameters.
•Select Confirm to use the proposed low and high alarm
limits.
•Select Cancel to leave the alarm limits unchanged.
•Select Case Default Limits to set the alarm limits to the
case default limits.
Note
The proposed low and high alarm limits are shown in highlighted
text. The alarm limits that are not highlighted do not change.
--
2067226-0013-9
Aisys CS²
Gas setup
Changing gas settings
Use the Gas Setup menu to adjust the Agent, O2% and total flow, to
change the balance gas, and to change the circuit type.
1.Select the Gas Setup quick key.
2.Select the setting to change from the Gas Setup menu.
3.Change the setting.
4.For Circuit, select the menu item and change using the dropdown menu. Select Confirm.
5.For Other Gas, select the menu item and change using the
drop-down menu.
6.For Agent, O2%, and Total Flow, select the setting and make
the change using the ComWheel and push to confirm the
setting.
Changing balance gas
1.Select the Gas Setup quick key.
2.Select the Other Gas menu item.
3.Select the balance gas to use with O2.
Changing circuit type
1.Select the Gas Setup quick key.
2.Select Circle or Non-Circle
3.Select Confirm.
3-102067226-001
3 Operation
Using the circle circuit
Use the circle circuit mode to combine fresh gas with recirculated
gas from the CO2 absorber. The combined gas flows out through the
inspiratory port. Patient gas is returned to the system through the
expiratory port.
Mechanical ventilation and tidal volume monitoring are available
when using the circle circuit.
O2 monitoring of fresh gas is available automatically when using the
circle circuit if the system has the airway module option or the O2
cell monitoring option. Systems with both an airway module and an
O2 cell will display the O2 values obtained from the airway module.
1.Select the Gas Setup quick key.
2.Select Circle.
Fresh gas oxygen concentration is displayed on the screen.
Fresh gas flow combines with the exhaled gas and exits out
through the inspiratory port.
3.Set the alarm limits to clinically appropriate settings.
WARNING
Using the non-circle circuit
Use the non-circle circuit mode to divert fresh gas around the
inspiratory check valve and out through the inspiratory port. This
fresh gas source may be used with circuits that do not have CO2
absorbent capability (for example, Mapleson variants). Mechanical
ventilation is not available when using the non-circle circuit. Tidal
volume monitoring is not available when using the non-circle circuit.
O2 monitoring of fresh gas is available automatically when using the
non-circle circuit if the system has the airway module option or the
O2 cell monitoring option. Systems with both an airway module and
an O2 cell will display the O2 values obtained from the airway
module.
Do not use an external ventilator when using the noncircle circuit. Do not use the non-circle circuit to drive
external ventilators or for jet ventilation.
•
The maximum pressure at the non-circle circuit can be up
to 27 kPa (4 psi). Use a breathing circuit with a pressure
limiting device to limit the pressure at the patient
connection port, during normal and single-fault
conditions, to less than 12.5 kPa (125 cmH2O) or to the
maximum pressure required by local standards.
1.Select the Gas Setup quick key.
2.Select Non-Circle.
Fresh gas oxygen concentration is displayed on the screen.
Fresh gas flow is diverted around the inspiratory check valve
and out through the inspiratory port.
2067226-0013-11
Aisys CS²
3.Set the alarm limits to clinically appropriate settings.
3-122067226-001
System setup
3 Operation
Use System Setup to access menus and settings for Patient
Demographics, Screen Setup, Fresh Gas Usage, System Status,
Calibration, and Checkout.
Note
System Status shows the status of gas supplies, electrical supplies,
and the software version.
See the "User maintenance" section for information on the
Calibration menu item.
See the "Preoperative tests" section for information on the Checkout
menu item.
Patient demographics
Use the Patient Demographics menu to access menus and settings
for Age, Ideal Weight, TV for Ideal Body Weight, and Set Vent byWeight.
Screen setup menu
Use the Screen Setup menus to customize the screen view. Areas
of the screen can be customized to show specific information.
Screen Setup contains the Layout, Scales, Time and Date, and
More Settings submenus.
Setting waveform fields
The waveforms can be set to show agent, CO2, flow, Paw, or set to
Off. If a waveform is set to the same value as another waveform, the
previously set waveform changes to off and is removed from the
screen.
1.Select System Setup - Screen Setup.
2.Select the Layout tab.
3.Select the desired waveform button and select the value from
the drop-down menu.
4.Select Close.
Setting digit fields
The digit field can be set to show gas supply, flow, spirometry loops,
gases, respiration, or agent. If the digit field is set to show agent and
no airway module is inserted, the digit field will be blank.
2067226-0013-13
Aisys CS²
1.Select System Setup - Screen Setup.
2.Select the Layout tab.
3.Select the desired digit field button and select the value from the
drop-down menu.
4.Select Close.
Setting the split screen
Use the Split Screen setting to show metabolics, trends, spirometry
loops, Paw gauge, airway compliance, and optional ecoFLOW
information.
See the "ecoFLOW" for information on the ecoFLOW option.
Note
Note
Resistance (Raw) shows in the airway compliance split screen when
the system detects an airway module with spirometry and the
module has completed a warm-up phase.
1.Select System Setup - Screen Setup.
2.Select the Layout tab.
3.Select Split Screen and select the desired view from the dropdown menu.
4.Select Close.
Setting time and date
Use the Time and Date menu to set the time and date.
The Time and Date menu cannot be changed when a case is
running.
1.Select System Setup - Screen Setup.
2.Select the Time and Date tab.
3.Select the time or date item to change. Make the change.
The clock format factory default is 24 hours.
•When the clock format is set to 12 h, the hour selections are
in ‘1a’ format for a.m. and ‘1p’ format for p.m.
•When the clock format is set to 24 h, the hour selections are
0 to 23 in one hour increments.
4.Select Close.
Setting the data source
Use Data Source to specify the source of spirometry data.
1.Select System Setup - Screen Setup - More Settings or
Spirometry - Setup Loops.
2.Select Data Source.
3-142067226-001
3 Operation
3.Select Patient to have spirometry data sourced from the airway
module or Vent to have spirometry data sourced from the
ventilator.
4.Select Back to view changes made and access other functions
of the Spirometry menu.
Setting sweep speed
Use the Sweep Speed setting to set the waveform draw rate to fast
(6.25 mm/s) or slow (0.625 mm/s). When the sweep speed changes,
waveforms redraw at the new rate.
1.Select System Setup - Screen Setup.
2.Select More Settings.
3.Select Sweep Speed and then select Fast or Slow.
4.Select Close.
Setting display brightness
Use the brightness setting to adjust the contrast level of the display.
1.Select System Setup - Screen Setup.
2.Select More Settings.
3.Select Display Brightness.
4.Select the desired brightness level with 1 being the dimmest and
5 being the brightest.
5.Select Close.
Setting keypad brightness
Use the brightness setting to adjust the contrast level of the hard
keys on the bezel.
1.Select System Setup - Screen Setup.
2.Select More Settings.
3.Select Keypad Brightness.
4.Select the desired brightness level with 1 being the dimmest and
5 being the brightest.
Set to 0 to turn off the keypad brightness.
5.Select Close.
2067226-0013-15
Aisys CS²
Setting fresh gas controls
Use the Fresh Gas Controls selection to set the gas control style to
O2% with Total Flow or to individual gas flow.
Selecting O2% shows O2% as the first quick key and Total Flow
l/min as the second quick key.
Selecting Flow shows balance gas l/min as one of the quick keys
and O2 l/min as the other quick key.
Note
Note
Fresh gas usage
Both O2% and Flow are available if User is selected by the Super
User. If User is not selected by the Super User, either O2% or Flow
will be unavailable.
Fresh Gas Controls cannot be accessed during a case.
1.Select System Setup - Screen Setup.
2.Select More Settings.
3.Select Fresh Gas Controls.
4.Select O2% or Flow from the drop-down menu.
When set to O2%, balance gas adjusts automatically when
either the O2% or the Total Flow is changed using the gas quick
keys.
When set to Flow, balance gas and the O2% are controlled
individually using the gas quick keys.
5.Select Close.
Use Fresh Gas Usage to view the volume of O2, Air, N2O, and
agents used for the three most recent cases.
•Data only shows for gases available on the system.
•Agent data shows the three most recently used agents.
1.Select System Setup - Fresh Gas Usage.
2.Select Case Start Time to select the patient case to view.
3.Select Close.
3-162067226-001
Alarm setup
3 Operation
Use the Alarm Setup menu to set and adjust alarm limits, alarm
volume, and other alarm settings and to view alarm history. The
Alarm Setup menu contains the Primary Limits, More Limits,
Alarm History, and Configure submenus.
Setting Leak Audio to Off silences audio alarms for small leaks.
Leak Audio is automatically set to On and cannot be changed wheneither the Low MV alarm limits are off or the MV/TV Alarms is set to
Off.
Note
Setting CO2 alarms
Selecting Change to Default Limits loads the default settings as set
by the Super User or the factory defaults if no Super User settings
have been entered.
Use the CO2 Alarms setting to turn off the CO2 ‘Apnea’ alarm,
‘EtCO2 low’, ‘EtCO2 high’, and ‘FiCO2 high. Absorbent OK?’ alarms
during manual ventilation.
CO2 information is obtained from the airway module installed in the
anesthesia system module bay. The CO2 alarms setting has no
effect if there is no airway module in the anesthesia system.
1.Select Alarm Setup.
2.To turn off the CO2 alarms, set CO2 Alarms to Off.
The ‘CO2 Alarms Off’ message shows in the general message
field.
The CO2 alarm limit waveform numerics shows dashes during a
case.
The alarms remain disabled until the Bag/Vent switch is set to
Vent, the case is ended, or the CO2 Alarms is set to On.
3.To turn on the CO2 alarms, set CO2 Alarms to On.
4.Select Close.
Setting volume apnea alarm
Use the Vol Apnea Alarm setting to turn off the volume apnea alarm
during manual ventilation. The volume apnea alarm remains off until
the Bag/Vent switch is set to Vent or Vol Apnea Alarm is set to On.
Note
2067226-0013-17
Vol Apnea Alarm does not show on the Alarm Setup menu when
the Volume Apnea Selection has been disabled by the Super User.
See the "Super user mode" section for more information.
1.Select Alarm Setup.
Aisys CS²
Setting MV TV alarms
2.To turn the volume apnea alarms off, select Vol Apnea Alarm
to Off.
‘Volume Apnea Off’ shows in the general message field.
•If mechanical ventilation is started, the volume apnea
alarms are active.
•If manual ventilation is restarted, a pop-up confirmation
window appears to resume the Off setting.
3.To turn the volume apnea alarms on, set Vol Apnea Alarm to
On.
4.Select Close.
Use the MV/TV Alarms setting to turn off the MV and TV alarms.
Settings made during manual ventilation are not retained when
mechanical ventilation starts. Settings made during mechanical
ventilation are retained when manual ventilation starts.
For example, if MV/TV Alarms is set to Off during manual
ventilation, the alarms remain off until the Bag/Vent switch is set to
Vent or the MV/TV Alarms is set to On.
For example, if the MV/TV Alarms is set to Off during mechanical
ventilation, the alarms remain off when manual ventilation starts.
Setting alarm limits
WARNING
1.Select Alarm Setup.
2.To turn the volume alarms off, set MV/TV Alarms to Off.
‘MV/TV Alarms Off’ appears in the general message field.
The volume alarm limits waveform numerics shows dashes
during a case.
3.To turn the volume alarms on, set MV/TV Alarms to On.
4.Select Close.
Do not set alarm limits to extreme values. Setting limits to
extreme values can render the alarm useless.
1.Select Alarm Setup.
2.From the Primary Limits and More Limits tabs, select the
alarm limit and make the change.
3.Push the Home key, touch the waveform area of the display, or
select Close to close the menu.
3-182067226-001
Viewing alarm history
Use the Alarm History tab to view the list of the 12 most recent high
and medium priority alarms that occurred since the start of the case.
The alarm history clears at the start of a new case.
1.Select Alarm Setup.
2.Select the Alarm History tab.
3.Select Close.
Setting alarm volume
1.Select Alarm Setup.
2.Select the Configure tab.
3.Set Alarm Volume to the desired value.
4.Select Close.
3 Operation
The list of alarms shows in the window.
The alarm volume range is 1 to 5.
Setting apnea delay
Use the Apnea Delay setting to set the desired apnea time delay.
The apnea time delay is the amount of time that can pass without the
system detecting a measured breath before the apnea alarm occurs.
1.Select Alarm Setup.
2.Select the Configure tab.
3.Set Apnea Delay to the desired time.
•The apnea time delay range is 10 to 30 seconds.
•The apnea time settings are in 1 second increments.
4.Select Close.
Silencing leak audio alarms
Use the Leak Audio setting to silence audio alarms of small leaks.
1.Select Alarm Setup.
2.Select the Configure tab.
2067226-0013-19
Aisys CS²
3.Set Leak Audio to Off.
The audio alarms for small leaks are silenced.
4.To turn audio alarms back on, set Leak Audio to On.
5.Select Close.
Note
If the Low MV alarm limits are off or MV/TV Alarms is set to Off,
Leak Audio is automatically set to On and cannot be changed.
Setting auto MV limits
MV alarm limits can be calculated automatically for mechanical
ventilation when in VCV or PCV-VG modes and volume
compensation is enabled. Use the Auto MV Limits setting to turn on
automatic calculations of the MV alarm limits.
If the automatic calculation of the low or high minute volume alarm
limit exceeds the allowable limit, the minimum or maximum alarm
limit is used.
1.Select Alarm Setup.
2.Select the Configure tab.
3.Set Auto MV Limits to On.
4.Select Close.
--
The MV alarm limits are automatically calculated until Auto MVLimits is set to Off or until an MV limit is manually adjusted
during mechanical ventilation. The MV alarm limits are
automatically calculated based on TV and RR settings.
Setting to default limits
Use Change to Default Limits to set alarm limits to the values set
by the Super User.
1.Select Alarm Setup.
2.Select the Configure tab.
3.Select Change to Default Limits.
4.Select Close.
3-202067226-001
Alarms On Off
3 Operation
Use the Alarms On/Off menu to turn On or turn Off the CO2 AlarmsLimits and MV/TV Alarm Limits during manual ventilation. Alarm
limits are enabled at the start of mechanical ventilation.
Note
Disable alarm limits
The CO2 ‘Apnea’ alarm cannot be turned off through the Alarms
On/Off menu. Only the ‘EtCO2’ and ‘FiCO2’ alarms are turned off
using this menu.
1.Select Alarms On/Off.
2.View the alarm limit status.
3.Select Confirm to change the alarm limit settings. Select
Cancel to leave the alarm limits unchanged.
•If the CO2 Alarms Limits and MV/TV Alarm Limits are
On, the alarm limits change to Off.
•If the CO2 Alarms Limits or MV/TV Alarm Limits are On,
the alarm limits change to Off.
•If the CO2 Alarms Limits and MV/TV Alarm Limits are
Off, the alarm limits change to On.
2067226-0013-21
Aisys CS²
Next page
Select Next Page to change the screen view. A default view and four
configurable screen views are available. A general message displays
identifying the page number of the screen view.
See the "Super user mode" section for information on setting the
preset screen views.
3-222067226-001
Trends
Setting trends
3 Operation
Use the Trends menu to view patient trends and set the time scale.
There are three views for patient trends: measured (numerical),
settings, and graphical. Trend information is saved every 15 seconds
for the most recent 24 hours.
1.Select Trends.
2.Select the desired view.
3.Select Scroll to move through the current trend view.
4.Select Time Scale to select the desired scale from the dropdown menu.
5.Select Next Page to view additional parameters.
6.Select Close.
2067226-0013-23
Aisys CS²
Spirometry
Use the Spirometry menu to:
•Set the loop type.
•Adjust the loop scaling.
•Save a loop to memory.
•Access the Setup Loops menu.
•View a saved loop.
•Delete a saved loop.
There are three types of spirometry loops: Pressure-Volume (Paw-Vol), Flow-Volume (Flow-Vol), and Pressure-Flow (Paw-Flow). The
spirometry loops show in the spirometry window and can be set to
show alongside the waveforms as the split screen.
1
4
3
Setting loop type
2
AB.98.039
1.Volume axis
2.Pressure axis
3.Real-time loop
4.Reference loop (appears on display in gray)
Figure 3-2 • Example of a Paw-Vol loop
1.Select Spirometry.
2.Select Loop Type and select the loop from the drop-down list.
3.Select Close.
3-242067226-001
Setting loop graph scaling
Use Spirometry Scaling to set the scales of the spirometry loop
graph. The available settings for the volume, Paw, and flow graph
axes are dependent on the set patient type of adult or pediatric.
•Auto automatically adjusts the volume, Paw, and flow axes of
the loop based on the minimum and maximum breath reading
shown in the waveform.
•Linked links the adjustment of the volume, Paw, and flow axes
of the loop graph together. Change one of the scales and the
remaining two scales automatically change based on the one
set scale.
•Indep. allows the axes of the loop graph to be changed
separately for the volume, Paw, and flow axes.
1.Select Spirometry - Setup Loops.
2.Select Spirometry Scaling and set the scale type from the
drop-down list.
3.Select Back to view changes made and access other functions
of the Spirometry menu.
3 Operation
Setting patient and sensor type
Patient and sensor type refer to the style of airway adapter used with
the airway module. If spirometry data is obtained from the airway
module, make sure that the sensor type matches the type of airway
adapter used. Adult or pediatric patient types are available.
WARNING
Make sure that the set sensor type corresponds to the
type of airway adapter in use. If the sensor type is not set
correctly, the information displayed may not be accurate.
1.Select Spirometry - Setup Loops.
2.Select Patient and Sensor and then select Adult or Pedi
depending on the sensor used.
3.Select Back to view changes made and access other functions
of the Spirometry menu.
Setting the data source
Use Data Source to specify the source of spirometry data.
2067226-0013-25
Aisys CS²
Setting spirometry volume type
1.Select System Setup - Screen Setup - More Settings or
Spirometry - Setup Loops.
2.Select Data Source.
3.Select Patient to have spirometry data sourced from the airway
module or Vent to have spirometry data sourced from the
ventilator.
4.Select Back to view changes made and access other functions
of the Spirometry menu.
The volume shown on the spirometry split screen can be set to
minute volume or tidal volume.
1.Select Spirometry - Setup Loops.
2.Select Show MV or TV and select MV or TV from the dropdown list.
Set to TV to show TVinsp and TVexp on the spirometry split
screen.
Set to MV to show MVexp and TVexp on the spirometry split
screen.
3.Select Back to view changes made and access other functions
of the Spirometry menu.
Saving, viewing, and deleting spirometry loops
Spirometry loops can be saved, viewed, and deleted through the
Spirometry menu.
1.Select Spirometry.
2.To store a loop to memory, select Save Loop. Up to six loops
can be saved.
3.To view a saved loop, set Show Ref. Loop to the time at which
it was saved.
4.To delete a saved loop, set Delete Ref. Loop to the time at
which it was saved.
5.Select Close.
3-262067226-001
Procedures
3 Operation
Use the Procedures menu to pause the gas flow, start cardiac
bypass, perform or change settings for a vital capacity procedure, or
perform or change the settings for a cycling procedure.
Note
Note
Pause gas flow
Vital Capacity shows in the menu if it is set to Yes by the Super
User. Vital Capacity is only selectable during mechanical ventilation.
Cycling shows in the menu if it is set to Yes by the Super User.
Cycling is only selectable during mechanical ventilation.
Use Pause Gas Flow to temporarily suspend the flow of gas during
a case. Using Pause Gas Flow while the breathing circuit is
disconnected prevents the flow of gas into the room. Pause GasFlow is available during both mechanical ventilation and manual
ventilation.
1.Select Procedures.
2.Select Pause Gas Flow.
The amount of time remaining in the gas flow pause shows in
the window.
Gas flow stops for 1 minute and automatically resumes after 1
minute.
If mechanical ventilation is on, mechanical ventilation stops for 1
minute and then automatically resumes after 1 minute.
3.Resume the flow of gas at any time during the pause by
selecting Restart Gas Flow.
Cardiac bypass
There are two types of cardiac bypass. Manual ventilation cardiac
bypass is standard. VCV cardiac bypass is optional.
Manual ventilation cardiac bypass suspends alarms for patients on
cardiac bypass when the ventilator is not mechanically ventilating.
The volume, apnea, low agent, CO2, and respiratory rate alarms are
suspended. The alarms are enabled when cardiac bypass is turned
off or mechanical ventilation is started.
Systems with the VCV cardiac bypass option enabled can
mechanically ventilate while in VCV mode. The VCV mode is the
only ventilation mode available while using VCV cardiac bypass. The
volume, apnea, low agent, CO2, low Paw, and respiratory rate
alarms are suspended. The alarms are enabled when VCV cardiac
bypass is turned off or mechanical ventilation is stopped.
2067226-0013-27
Aisys CS²
WARNING
Manual ventilation cardiac bypass and VCV cardiac
bypass modes should only be used when the patient is
receiving extra-corporeal oxygenation by means of a
heart-lung machine. These modes of ventilation are not
intended to provide metabolic levels of ventilation to the
patient.
Using manual ventilation cardiac bypass
1.Set the Bag/Vent switch to Bag.
2.Select Procedures.
3.Select Start Cardiac Bypass.
The ‘Cardiac Bypass’ message shows in the waveforms and in
the general message field when manual ventilation cardiac
bypass is active.
4.Select Close.
Using VCV cardiac bypass
1.Start mechanical ventilation in VCV mode.
2.Select Procedures.
3.Select Start Cardiac Bypass.
PEEP is set to 5 cmH2O.
TV settings of less than 170 ml prior to starting cardiac bypass
remain at the set TV.
TV settings of more than 170 ml prior to starting cardiac bypass
change to 170 ml.
The ‘VCV Cardiac Bypass’ message shows in the waveforms
and in the general message field when VCV cardiac bypass is
active.
Note
PEEP and TV settings can be changed after entering cardiac bypass
mode.
--
4.Select Close.
Vital capacity
Use the Vital Capacity procedure to deliver a pressure breath for a
set time. The Vital Capacity procedure provides a simple way to
deliver one pressure breath during mechanical ventilation without
making multiple ventilator setting changes. The PEEP on Exit
setting provides a way to change the ventilation PEEP setting
automatically at the end of the Vital Capacity procedure.
3-282067226-001
3 Operation
The Pressure Hold, Hold Time, and PEEP on Exit settings can be
preset by the Super User. These settings can be changed by the
user before starting the procedure.
Note
Note
PEEP on Exit shows if it is set to Yes by the Super User.
Using vital capacity
1.Select Procedures.
2.Select Start Vital Capacity.
One pressure breath is given at the set pressure.
The pressure is held for the set time.
PEEP is set to the PEEP on Exit setting.
3.Select Stop Vital Capacity at any time to stop the procedure.
4.Select Close.
If the procedure is stopped before completion, the PEEP on Exit
setting is not used.
Changing vital capacity settings
1.Select Procedures.
2.Select the setting to change and make the change.
•Set Pressure Hold to between 20 and 60 cmH2O.
•Set Hold Time to between 10 and 40 seconds.
•Set PEEP on Exit to Off or between 4 and 30 cmH2O.
Cycling
Note
3.Select Start Vital Capacity.
4.Select Close.
Use the Cycling procedure to deliver pressure breaths through a
series of ventilation steps. The Cycling procedure provides a flexible
way to deliver pressure breaths during ventilation without making
multiple ventilator setting changes. Up to seven preset steps with
multiple breaths are available.
Each procedure defaults steps and ventilation settings which can be
preset by the Super User. The ventilation settings of each step can
be changed by the user before starting a procedure.
There is a limited amount of gas in the bellows. No additional gas
enters the bellows during the cycling procedure. Increase fresh gas
flow to avoid bellows collapse.
Using cycling
1.Select Procedures.
2067226-0013-29
Aisys CS²
2.Select Cycling.
3.Select a Procedure to perform.
4.Select Start Cycling.
The procedure begins.
Procedure progress shows in the procedure window.
5.Stop the procedure anytime by selecting Stop Cycling.
6.Select Close.
Changing cycling settings
1.Select Procedures.
2.Select Cycling.
3.Select a Procedure to perform.
4.Select Adjust Settings.
The first setting of Step 1 in the procedure window is selected.
5.Push the ComWheel to enter the adjustment window.
6.Use the ComWheel to navigate the adjustment window and
change a value.
7.Select Start Cycling.
8.Select Close.
3-302067226-001
Timer function
Using the timer
3 Operation
Use Start as a timer function. When selected a clock will display
counting up from zero.
1.Select Start to start the clock.
2.Select Stop to pause or stop the clock.
If Start is selected, the clock will resume operation.
If Start is selected and held for longer than 1 second, the clock
will reset to zero.
2067226-0013-31
Aisys CS²
ecoFLOW
This feature provides a split screen view that shows the approximate
minimum O2 flow to maintain a preset inspired O2 concentration.
Also shown is the approximate agent used per hour and the cost.
40
20
60
10
9
8
0
-20
15
9
3
2
1
AgentTotal Flow
21.91
80
100
1
2
100
10
7
6
5
O2 Total
1.30
Fi25 O2 flow
0.60
Iso
4
3
041
22.
.
C
A
4
1.AgentShows Agent cost and flow information.
2.Agent costThe cost of the current agent flow. This value is determined by the agent flow multiplied
by the agent cost set in Super user mode.
3.Agent flowThe measured value of the liquid agent flow from the vaporizer. The agent flow may have
a delayed response. For example: Iso.
4.FiO2 flow markerThe graphical representation on the flow tube of the FiO2 flow value. This marker can be
removed by disabling it in Super user mode.
5.FiO2 flowThe minimum O2 flow needed to maintain the set inspired O2 flow. This item can be
disabled in Super user mode. For example: Fi25 O2 flow.
6.O2 total (numeric)The numeric representation of the total O2 flow. If N2O is the balance gas, this equals the
set O2 flow. If Air is the balance gas, this is the set O2 flow plus 21% of the Air flow. For
example: O2 Total.
7.O2 total (graphical)The graphical representation of the total O2 flow. If N2O is the balance gas, this equals
the set O2 flow. If Air is the balance gas, this is the set O2 flow plus 21% of the Air flow.
8.Remaining gas flowIf N2O is the balance gas, this equals the set N2O flow. If Air is the balance gas, this N2
is 79% of the Air flow.
9.Flow bobbinThe height of this represents the total fresh gas flow delivered to the patient.
10. Total flowShows Total Flow information.
Figure 3-3 • ecoFLOW feature
3-322067226-001
Using ecoFLOW
3 Operation
1.Select System Setup - Screen Setup.
2.Select the Layout tab.
3.Select Split Screen and select ecoFLOW from the drop-down
menu.
4.Select Close.
2067226-0013-33
Aisys CS²
Alternate O2 control
WARNING
The Alternate O2 control is not an auxiliary source of O2.
When Alternate O2 control is enabled, flow from the
•
electronic mixer is stopped and the agent concentration
is set to off. O2 is flowing through the Alternate O2
control to the breathing system. To activate anesthetic
agent flow to the breathing system, set the agent to the
desired concentration.
•
Agent delivery cannot be activated in the case of certain
electronic mixer or agent delivery failures. The Agent
quick key will be blank if agent delivery is unavailable.
Use the Alternate O2 control to deliver O2 through an independent
pneumatic path to the vaporizer and patient circuit. Alternate O2 is
connected to the system O2 supply. Alternate O2 control activates
automatically in the case of certain failures or errors. It can also be
activated manually.
The Alternate O2 control is available approximately 20 seconds after
the system is turned on. The Alternate O2 flow is adjustable to
approximately 0.5 l/min to 10 l/min as indicated on the flow tube.
1
3
1.ON/OFF button
2.Flow control
3.Flow tube
Figure 3-4 • Alternate O2 control
Using Alternate O2 control
1.Push the Alternate O2 ON/OFF button.
The O2 flow is indicated on the flow tube.
2
AB.91.077
3-342067226-001
3 Operation
2.Use the flow control to adjust the O2 flow.
3.Set the agent to the desired concentration.
4.To end Alternate O2 control, push the Alternate O2 ON/OFF
button.
2067226-0013-35
Aisys CS²
EZchange canister mode
Use the optional EZchange canister mode for continued ventilation of
the patient while changing the absorber canister. The EZchange
canister mode seals the breathing circuit when the canister holder is
down. While the absorber canister is out of the breathing circuit, the
patient re-breaths exhaled gases without any gas passing through
the absorbent.
Systems with EZchange canister mode have a label on the canister
holder. An EZchange canister has side rails that fit in the EZchange
canister holder.
Figure 3-5 • EZchange canister holder label
AB.82.042
1.Side rails
Figure 3-6 • EZchange canister
1
AC.22p.010
3-362067226-001
Using EZchange canister mode
3 Operation
Note
Make sure that the absorber canister has side rails. If the canister
does not have side rails, it will not work in the EZchange canister
holder.
1.Push the absorber canister release to activate the EZchange
canister mode.
The canister swings down to the EZchange position.
‘CO2 Absorber Out of Circuit’ shows in the waveform area on
the anesthesia display.
2.Remove the canister from the holder.
3.Install a canister with fresh absorbent into the holder.
4.Push the canister back up and snap it into absorber position.
The exhaled gas flows through the absorber, removing CO2.
2067226-0013-37
Aisys CS²
Condenser
Use the optional condenser to remove water in the system that is
produced from the reaction of CO2 gas with the absorbent. The
condenser is connected between the outlet of the absorber canister
and the inlet of the circuit module. Moisture in the gas is condensed
into water droplets, which run into the condenser’s reservoir.
3
1.Drain button
2.Reservoir
3.Condenser
Figure 3-7 • Condenser
Draining the condenser
AB.82.043
12
1.Visually check the condenser reservoir daily. Drain the reservoir
daily.
2.Place a container under the reservoir.
3-382067226-001
3 Operation
3.Push the drain button to empty any water in the condenser.
2067226-0013-39
Aisys CS²
Auxiliary Common Gas Outlet
Use the optional Auxiliary Common Gas Outlet (ACGO) switch to
direct the fresh gas flow through the ACGO port on the front of the
system. The ACGO may be used to provide fresh gas to an auxiliary
manual breathing circuit.
Mechanical ventilation is not available when using an auxiliary
manual breathing circuit with fresh gas from the ACGO. The Bag/
Vent switch, APL valve, and CO2 absorber are not part of the
external circuit. Volume and pressure monitoring are not available.
O2 monitoring of fresh gas is available when the ACGO is selected if
the system has the airway module option or the O2 cell monitoring
option. Systems with both an airway module and an O2 cell will
display the circuit O2 value obtained from the airway module.
WARNING
Using the ACGO
The Bag/Vent switch and the APL valve do not control
the ACGO or any breathing circuit connected to the
AGCO. Do not use these controls when using a breathing
circuit with fresh gas from the ACGO. Patient injury may
occur.
•
Volume and pressure monitoring are not available
through the system when using the ACGO port. Monitor
the patient using other methods.
•
Do not use an external ventilator on the ACGO. Do not
use the ACGO to drive external ventilators or for jet
ventilation.
•
The maximum pressure at the ACGO can be up to 55
kPa (8 psi). Use a breathing circuit with a pressure
limiting device to limit the pressure at the patient
connection port, during normal and single-fault
conditions, to less than 12.5 kPa (125 cmH2O) or to the
maximum pressure required by local standards.
1.Set the ACGO switch to the ACGO position.
Fresh gas flows through the ACGO port.
Fresh gas oxygen concentration is displayed on the screen.
2.Set the alarm limits to clinically appropriate settings.
3.To stop fresh gas flow through the ACGO port, set the ACGO
switch to the circle circuit position.
Note
3-402067226-001
Displayed fresh gas oxygen concentration may not reflect FiO2
during spontaneous breathing or in rebreathing circuits. Use an
external O2 monitor if using a rebreathing circuit on ACGO.
--
3 Operation
Note
A sample of the fresh gas is diverted to the O2 cell in the breathing
system. The sample flow to the O2 cell is dependent on the pressure
in the external circuit. The sample flow reduces the fresh gas flow
rate to the auxiliary breathing circuit equal to the amount diverted to
the O2 cell.
--
2067226-0013-41
3-422067226-001
4 Preoperative checkout
Preoperative checkout
In this section
Every day before your first patient....................4-2
Before every patient...............................4-3
2067226-0014-1
Aisys CS²
Every day before your first patient
Check that necessary emergency equipment is available and
□
in good condition.
Check that the equipment is not damaged and that
□
components are correctly attached.
Check that the pipeline gas supplies are connected. If
□
equipped with cylinders, check that there is sufficient reserve
capacity and that the cylinder valve is closed.
Connect scavenging and verify operation.
□
Check that the amount of liquid agent in the cassette is
□
adequate. Install the agent cassette into the active bay. The
cassette is properly inserted when the agent is identified on
the display.
Check that the breathing circuit is correctly connected, not
□
damaged, and the breathing system contains sufficient
absorbent in the canister.
Turn the System switch On.
□
Perform a Full Test from the Checkout menu.
□
Do an Agent Delivery check in the Checkout menu for each
□
agent cassette that will be used during the day.
Check that an adequate reserve O2 supply is available.
□
Check that the ventilator functions correctly:
□
•Connect a test lung to the patient breathing circuit
connection.
•Set the ventilator to VCV mode and the settings to TV to
400 ml, RR to 12, I:E to 1:2, Tpause to Off, and Pmax to
40 cmH2O.
•Set the gas flow to the minimum settings.
•Start a case.
•Set the Bag/Vent switch to Vent.
•Fill the bellows using O2 flush.
•Check that mechanical ventilation starts. Check that the
bellows inflate and deflate. Check that the display shows
the correct ventilator data. Check that there are no
inappropriate alarms.
Turn the system circuit breaker off and check that mechanical
□
ventilation continues while the system is running on battery
power. After completing the check, turn the system circuit
breaker on. The mains indicator is lit when AC power is
connected.
Set the appropriate controls and alarm limits for the case.
□
4-22067226-001
Before every patient
4 Preoperative checkout
Note
This check does not need to be done before the first case of the day
if the "Every day before your first patient" checklist was done.
Check that the necessary emergency equipment is available
□
and in good condition.
Select Checkout and perform a Low P Leak test.
□
If the cassette was changed since the last case:
□
•Check that the amount of liquid agent in the cassette is
adequate.
•Check that the cassette is properly inserted. The cassette
is properly inserted when the agent is identified on the
display.
Check that an adequate reserve O2 supply is available.
□
Check that the breathing circuit is correctly connected, not
□
damaged, and the breathing system contains sufficient
absorbent in the canister.
Select Checkout and perform a Circuit Leak test.
□
Check that the ventilator functions correctly:
□
•Connect a test lung to the patient breathing circuit
connection.
•Set the ventilator to VCV mode and the settings for TV to
400 ml, RR to 12, I:E to 1:2, Tpause to Off, and Pmax to
40 cmH2O.
•Set the gas flow to the minimum settings.
•Start a case.
•Set the Bag/Vent switch to Vent.
•Fill the bellows using O2 flush.
•Check that mechanical ventilation starts. Check that the
bellows inflate and deflate. Check that the display shows
the correct ventilator data. Check that there are no
inappropriate alarms.
•Set the appropriate controls and alarm limits for the case.
Make sure that the alarms function. See the "Alarm tests".
Set the appropriate controls and alarm limits for the case.
Positive low pressure leak test (ACGO systems only)....5-10
2067226-0015-1
Aisys CS²
Aladin cassette installation
1.Using the liquid level indicator, check that the cassette is filled to
the appropriate level.
2.If using Aladin2, unlock the cassette handle before installing it
into the active bay.
3.Insert the cassette into the active bay until a click is heard. An
audible click indicates that the cassette is in position.
4.Turn the lock on the handle to the horizontal position (Aladin2
cassettes only).
5.The cassette is properly inserted when the agent is identified on
the display. Make sure that the displayed agent matches the
cassette.
Note
Store the cassettes in the cassette storage bay when they are not in
use. See the "Vaporizer cassettes" section for additional cassette
information.
5-22067226-001
Flow and pressure calibration
5 Preoperative tests
Important
Room temperature fluctuations of more than 5°C may affect sensor
measurements. Recalibrate the flow sensors if the room temperature
changes by more than 5°C.
Calibrate the flow sensors by removing the flow sensor module from
the system. On-screen instructions are available through SystemSetup - Calibration - Flow and Pressure.
1.Set the Bag/Vent switch to Bag.
2.Remove the flow sensor module.
3.Wait for ‘No insp flow sensor’ and ‘No exp flow sensor’ alarms to
occur.
4.Reinsert the flow sensor module. Make sure the flow sensors
are securely latched when they are reinserted. Wait for alarms
to clear.
5.Start mechanical ventilation when ready.
2067226-0015-3
Aisys CS²
Circuit compliance compensation
The ventilator adjusts gas delivery and monitoring to compensate for
the compliance of the patient circuit if:
•The system has an airway module installed.
•The Circuit Compliance is set to On from Super User System Config. - Ventilator Settings.
•The Checkout is completed after the system is turned on.
Circuit compliance is determined during the Checkout - Full Test or
when the Vent and Gas check is done as an individual test. Circuit
compliance must be set by the Super User for the Circuit
Compliance compensation function to be active. See the “Super user
mode” section for information on the circuit compliance setting.
In volume modes, circuit compliance compensation increases the
volume delivered at the inspiratory port by taking into consideration
the circuit compliance value. In all modes, circuit compliance
compensation adjusts the volume measurements. Circuit compliance
compensation provides consistent ventilator accuracy at the patient
circuit.
WARNING
Perform a Vent and Gas test after changing the patient
tube type. Changing the patient breathing circuit after
completing a Vent and Gas test affects the volume
measurements in all modes.
Breathing circuits and breathing circuit components are
•
available in many different configurations from multiple
suppliers. Attributes of the breathing circuits such as
materials, tube length, tube diameter, and configuration
of components within the breathing circuit, may result in
hazards to the patient from increased leakage, added
resistance, or changed circuit compliance.
5-42067226-001
Checkout menu
5 Preoperative tests
The Checkout menu shows on the display after turning on the
system. To access the Checkout menu between cases, select
Checkout. Step-by-step instructions show in the Checkout menu
during the tests. Use the Checkout menu to:
•Perform a Full Test.
•Perform any of the individual tests.
•View the Test Log.
•Start a case.
2067226-0015-5
Aisys CS²
Full test
The Full Test or the individual tests must be performed at least once
within every 24-hour period.
Perform the Full Test at the start of each day. The full test runs
automatically and beeps to indicate when it is finished or if
interaction is required.
The Full Test does the following tests: Vent and Gas, Circuit Leak,
and Circuit O2 Cell (if circuit O2 cell is present). When one of the
tests is completed, the next test begins.
1.From the Checkout menu, select Full Test and follow the
instructions.
2.If a test fails, follow the instructions to perform a retest or accept
the results.
3.When the Full Test is completed, start a case.
Vent and gas
Circuit leak
Note
Note
In case of a patient emergency, the Full Test may be bypassed by
selecting Start Case. The general message ‘Please Do Checkout’ is
displayed if a Full Test or all of the individual tests are not completed
with passing results within 24 hours.
The Vent and Gas test checks the agent delivery, airway module,
Bag/Vent switch, proper gas supply pressures, ventilator operation
and leak, battery and electrical power, circuit compliance, and flow
control operation. This is a two-step test.
To run this test, follow the on-screen instructions. When the test
passes, the next test starts.
Any cassette may be used during the test. Use a non-desflurane
cassette during the test to check the full functionality of the internal
electronic control unit.
The Circuit Leak test checks the Bag/Vent switch, proper gas
supply pressures, airway pressure measurement transducer, APL
valve, and manual circuit leak.
To run this test, follow the on-screen instructions. When the test
passes, the next test starts.
5-62067226-001
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.