SYMBOLS USED WITHIN THIS
MANUAL AND ON THE PRODUCT
Warm Water Fill Line
Cold Water Fill Line
Attention, consult Operator's
Manual.
Water Levels
Water Flow
Dangerous Voltage
Type BF Applied Part
Do not penetrate with sharp
object.
Length of cycle time
Continuous
Year of manufacture
OPERATOR'S MANUAL
TP702 T/PUMP
BEFORE YOU BEGIN . . .
Read and understand this T/PUMP OPERATOR'S
MANUAL and all SAFETY PRECAUTIONS
(see page 1) prior to using the T/Pump.
Only qualied medical service personnel should repair or perform
function tests on the T/Pump. Contact your dealer or Gaymar's
Technical Service Department for assistance:
Telephone: 716 662-2551
RECEIVING INSPECTION
Check the shipping carton for damage immediately upon receipt. If
package damage is discovered, the device should be unpacked with the
carrier's agent present. Any claims for shortage or damage must be led
with the delivering carrier by the purchaser. Do not return pumps that
were damaged in shipment to Gaymar without contacting our Technical
Service Department for advice (see phone numbers below). If damaged
goods are returned to Gaymar without notifying the carrier, Gaymar will
assume the repairs will be made at the customer's expense.
TO RETURN PUMPS TO FACTORY FOR REPAIR OR
EXCHANGE
Merchandise returned to Gaymar must be accompanied by a Return
Goods Number (RG#), issued by Gaymar, authorizing goods to
be returned. Contact your local dealer or Gaymar's International
Department at:
International: +1 716-662-8636
Fax: +1 716-662-0730
Provide the model, serial number, and detailed nature of the problem.
You will be given a Return Goods Number (RG#).
The serial number can be found on the bottom of the T/Pump.
The date of manufacture of the T/Pump is encoded to the serial
number. To determine the date of manufacter, the serial number
can be broken down as follows.
A01001
Month YearDecade These numbers are
not signicant to the
date of manufacture
A January
BFebruary
CMarch
DApril
EMay
FJune
T/PUMP, T/PAD, Mul•T•Pad, Clik-Tite, and Gaymar are registered trademarks of
Gaymar Industries, Inc. U. S. PATENT 4,068,870
Incidin Extra is a registered trademark of Ecolab Healthcare Division.
• Riskofexplosion.Do not use in the presence of ammable
anesthetics.
• Riskofelectricshock.Disconnect power before servicing the
T/Pump.
• Do not attach the power cord to any moving parts of the bed
frame or any location that may be a pinch point or cause power
cord damage. If routed improperly, the power cord may become
pinched and could cause electrical shock.
WARNING
• This device pumps temperature controlled water through a
pad. Set the pad temperature only as directed by and under the
guidance of the prescriber.
• Check the skin integrity of the body surface to which therapy is
applied. Evaluate patient response to temperature application.
• Check patient's skin for adverse reactions every 30 minutes or as
directed by the prescriber.
• Failure to adhere to these warnings could result in patient injury.
• The following Groups/Conditions require additional surveillance:
Group/Condition at riskPotential Injury
Pediatric patients: The portion of an
infant’s skin surface in contact with
a pad, in relationship to their body
mass, can potentially affect their body
temperature.
Patients with impaired circulationIschemia
Areas of application under pressureIschemia
Hyperthermia/
Hypothermia
• Only qualied medical service personnel should repair the
T/Pump. Improperrepairmayresultindeathorseriousinjury,
equipmentdamage,ormalfunction.
• Use T/Pump TP702 controls with Mul•T•Pads. For catalog numbers
and descriptions, see page 3.
• Do not place additional heat sources between the patient and pad.
Skin damage may result.
CAUTION
• Federal law restricts this device to sale by or on the order of a
physician.
• Do not cover the control unit with blankets, pillows or other
insulating materials. Air ow is required to maintain system
performance.
1www.gaymar.com
INTRODUCTION
OPERATOR'S MANUAL
TP702 T/PUMP
INTRODUCTION
Heat therapy is effective in the dilation of blood vessels, thereby increasing
the blood ow to the heated area. Heat therapy has a variety of uses, the
most common being treatment of aches and pains in joints and muscles.
Cooling therapy assists in vasoconstriction, decreasing blood ow and
decreasing the metabolism in the affected area. Cooling therapy is applied
in the acute phase of injury minimizing blood loss, inammation of the
tissue, and can be effective in pain management.
The Gaymar T/Pump® Localized Temperature Therapy System provides
therapy by warming or cooling the enclosed water, and circulating it
through the Gaymar Mul•T•Pad. The pad is connected to the Gaymar
T/Pump with easy-to-use Clik-Tite® connectors.
The Mul•T•Pad provides the interface for delivering the temperature
therapy. The unique button design allows water to ow and provides
trouble-free operation when the pad is folded to form a customized t.
This reduces the number of pads your facility must keep in inventory.
The pads are applied to the part of the body requiring therapy, and the
circulating water maintains the pad at the setpoint temperature. The
setpoint temperature can be locked to prevent tampering.
Connecting the Pads
Figure 1A: Localized Temperature
Therapy System with Single Pad
with the TP702.
The Mul•T•Pads can be interconnected using Clik-Tite® connectors to
provide therapy to more than one body site at a time.
Figure 1B: Localized
Temperature Therapy
System with Multiple Pads.
2www.gaymar.com
OPERATOR'S MANUAL
TP702 T/PUMP
INTRODUCTION
Catalog Descriptions
Catalog #Product Name
TP702Professional Control Unit
TP12E Mul•T•Pad: 13"w x 18"l (33cm x 45cm)
Nonwoven fabric on one side, pliable polymer on the
other side. 20 per carton
TP22E Mul•T•Pad: 15"w x 22"l (38cm x 55cm)
Nonwoven fabric on one side, pliable polymer on the
other side. 20 per carton
TP22G Mul•T•Pad: 15"w x 22"l (38cm x 55cm)
Heavy Polymer, Reuseable. 10 per carton
TP26E Mul•T•Pad: 18"w x 26"l (45cm x 66cm)
Nonwoven fabric on one side, pliable polymer on the
other side. 10 per carton
To order any of these products contact your dealer or Gaymar's
International Department at:
International: +1 716-662-8636
Fax: +1 716-662-0730
Or, visit our website at www.gaymar.com
3www.gaymar.com
INDICATIONS AND CONTRAINDICATIONS
OPERATOR'S MANUAL
TP702 T/PUMP
INDICATIONS FOR USE
Localized temperature therapy is recommended in treating the following
applications:
Orthopedic Conditions such as:
acute injuries, chronic pain, lower back pain, muscle spasm and
strains
Skin Trauma such as:
abscesses, boils, bruises, burns and contusions
Other conditions such as:
chronic arthritis, neuritis, phlebitis, tendonitis and I.V. inltration,
infection and localized pain.
Other applications only as prescribed.
OK for Use With:
• Non-acute traumatized tissue.
CLINICAL
CONTRAINDICATIONS
• Impaired mental status.
• Insensate body surface.
• O2 therapy. However, if oxygen tent is in use, do not use pump inside
tent.
Heating contraindications are:
• Application to a body surface with compromised blood ow
(Ischemia, area under pressure, arterial insufciency).
• Application to a patient with an increased tendency to bleeding
(aggravates potential for hemorrhage).
• Application to a body surface with possibility of malignancy (tissue
metabolism is increased and therefore, the growth potential of the
malignant tissues).
• Treatment of hematoma within rst 24-48 hours (potential for
re-bleeding and hemorrhage).
• Recent sprain or fracture (acute inammatory response).
• In combination with topical solutions whose toxicity may be
affected by the application of heat.
• In combination with other heat sources.
Cooling contraindications are:
• Application to a body surface with compromised blood ow or
hypersensitivity to cold. (Ischemia, Reynauds Syndrome).
4www.gaymar.com
OPERATOR'S MANUAL
TP702 T/PUMP
FEATURES
PUMP FEATURES
Easy to Use
Keypad
Attached hose10 ft (305 cm) dual hose. Connectors allow pads to be
Flow indicatorIndicates no ow. Turns off heater if pump is tipped.
Warm/Cool
Delivery
Therapy Cycles Choose from 20-minute, 30-minute, or Continuous
On/Standby
Button
Over Temp Safety
Thermostat
Self CheckAutomatic system check at startup.
Hose/cord
Management
Comfortable
Handle Design
Dual Micro
Processor
Tethered EasyOpen Cap
Handle VentsThe vents in the handle allow air ow to keep the
See KEYPAD FEATURES on page 6.
connected to the pump (Figures 1A and 1B, page 2).
Four temperature setpoints on the TP702.
cycles.
Indicates power is supplied to the unit.
Limit thermostat shuts off heater if the high temperature
limit is exceeded.
Convenient and easy storage areas for hose and cord.
Designed for a more comfortable grip when moving the
pump.
Two electronic circuits, one over temperature sensing
circuit.
Prevents misplacing the cap. Only 1/4 of a turn is needed
to remove or secure the cap.
motor and heater inside the unit cool.
Easy to Use Keypad
Flow Indicator
Temperature Setpoints for
warm or cool delivery
Therapy Time Cycles)
Vents in Handle
Figure 2: T/Pump Features
On/Standby Button
Comfortable Handle
Tethered Easy-Open Cap
Cord storage
Hose storage
Hose Connections
5www.gaymar.com
FEATURES
KEYPAD FEATURES
TP702
Indicator/
Warning
Light
OPERATOR'S MANUAL
TP702 T/PUMP
Figure 3: TP702 Keypad
See the Troubleshooting section on page 11.
Water Flow
Check hoses or clamps for kinks or occlusions.
Water Level
Check water level.
SetpointsPress the button at the bottom of the setpoint indicator
to toggle through the four setpoints. Temperatures are
identied in °C.
Setpoint
Lock
Therapy
Cycles
Prevents tampering.
Press and hold for 2 seconds to lock or unlock the
setpoint.
Continuous cycle,
30-minute cycle
20-minute cycle.
Green indicates the unit is on.
Yellow indicates power is supplied to the unit but the
unit is not on.
6www.gaymar.com