Gaymar TP702 Operator's Manual

T/PUMP
®
TP702 Localized Temperature Therapy System
Operator's Manual [EN]
0086
P/N 101516000 Rev A 7/11www.gaymar.com
BEFORE YOU BEGIN . . .
CONTENTS
SAFETY PRECAUTIONS .........................1
INTRODUCTION .....................................2
INDICATIONS FOR USE ........................4
CLINICAL CONTRAINDICATIONS ...4
PUMP FEATURES ....................................... 5
KEYPAD FEATURES TP702 .....................6
CLIK-TITE® CONNECTORS ..................7
START-UP PROCEDURE ..........................8
SHUTDOWN PROCEDURE .................. 8
THERAPY CYCLES,
PROFESSIONAL ONLY. .........................9
TROUBLESHOOTING.. ...........................10
STORAGE / CLEANING ....................... 13
PREVENTIVE MAINTENANCE/
SERVICE....................................................15
SPECIFICATIONS ................................... 16
WARRANTIES ......................................... 17
SYMBOLS USED WITHIN THIS MANUAL AND ON THE PRODUCT
Warm Water Fill Line
Cold Water Fill Line
Attention, consult Operator's Manual.
Water Levels
Water Flow
Dangerous Voltage
Type BF Applied Part
Do not penetrate with sharp object.
Length of cycle time
Continuous
Year of manufacture
OPERATOR'S MANUAL
BEFORE YOU BEGIN . . . Read and understand this T/PUMP OPERATOR'S
MANUAL and all SAFETY PRECAUTIONS (see page 1) prior to using the T/Pump.
Only qualied medical service personnel should repair or perform function tests on the T/Pump. Contact your dealer or Gaymar's Technical Service Department for assistance:
Telephone: 716 662-2551
RECEIVING INSPECTION
Check the shipping carton for damage immediately upon receipt. If package damage is discovered, the device should be unpacked with the carrier's agent present. Any claims for shortage or damage must be led with the delivering carrier by the purchaser. Do not return pumps that were damaged in shipment to Gaymar without contacting our Technical Service Department for advice (see phone numbers below). If damaged goods are returned to Gaymar without notifying the carrier, Gaymar will assume the repairs will be made at the customer's expense.
TO RETURN PUMPS TO FACTORY FOR REPAIR OR EXCHANGE
Merchandise returned to Gaymar must be accompanied by a Return Goods Number (RG#), issued by Gaymar, authorizing goods to be returned. Contact your local dealer or Gaymar's International Department at:
International: +1 716-662-8636
Fax: +1 716-662-0730
Provide the model, serial number, and detailed nature of the problem. You will be given a Return Goods Number (RG#).
The serial number can be found on the bottom of the T/Pump.
The date of manufacture of the T/Pump is encoded to the serial number. To determine the date of manufacter, the serial number can be broken down as follows.
A 0 1 0 0 1
Month Year Decade These numbers are
not signicant to the date of manufacture
A January B February C March
D April
E May F June
T/PUMP, T/PAD, Mul•T•Pad, Clik-Tite, and Gaymar are registered trademarks of
Gaymar Industries, Inc. U. S. PATENT 4,068,870
Incidin Extra is a registered trademark of Ecolab Healthcare Division.
© 2009. Gaymar Industries, Inc. All rights reserved.
G July H August
I September J October
K November
L December
OPERATOR'S MANUAL TP702 T/PUMP
SAFETY PRECAUTIONS
SAFETY PRECAUTIONS
DANGER
• Riskofexplosion.Do not use in the presence of ammable anesthetics.
• Riskofelectricshock.Disconnect power before servicing the T/Pump.
• Do not attach the power cord to any moving parts of the bed frame or any location that may be a pinch point or cause power cord damage. If routed improperly, the power cord may become pinched and could cause electrical shock.
WARNING
• This device pumps temperature controlled water through a pad. Set the pad temperature only as directed by and under the guidance of the prescriber.
• Check the skin integrity of the body surface to which therapy is applied. Evaluate patient response to temperature application.
• Check patient's skin for adverse reactions every 30 minutes or as directed by the prescriber.
• Failure to adhere to these warnings could result in patient injury.
• The following Groups/Conditions require additional surveillance:
Group/Condition at risk Potential Injury
Pediatric patients: The portion of an infant’s skin surface in contact with a pad, in relationship to their body mass, can potentially affect their body temperature.
Patients with impaired circulation Ischemia Areas of application under pressure Ischemia
Hyperthermia/ Hypothermia
• Only qualied medical service personnel should repair the T/Pump. Improperrepairmayresultindeathorseriousinjury,
equipmentdamage,ormalfunction.
• Use T/Pump TP702 controls with Mul•T•Pads. For catalog numbers and descriptions, see page 3.
• Do not place additional heat sources between the patient and pad. Skin damage may result.
CAUTION
• Federal law restricts this device to sale by or on the order of a physician.
• Do not cover the control unit with blankets, pillows or other insulating materials. Air ow is required to maintain system performance.
1www.gaymar.com
INTRODUCTION
OPERATOR'S MANUAL
INTRODUCTION
Heat therapy is effective in the dilation of blood vessels, thereby increasing the blood ow to the heated area. Heat therapy has a variety of uses, the most common being treatment of aches and pains in joints and muscles.
Cooling therapy assists in vasoconstriction, decreasing blood ow and decreasing the metabolism in the affected area. Cooling therapy is applied in the acute phase of injury minimizing blood loss, inammation of the tissue, and can be effective in pain management.
The Gaymar T/Pump® Localized Temperature Therapy System provides therapy by warming or cooling the enclosed water, and circulating it through the Gaymar Mul•T•Pad. The pad is connected to the Gaymar T/Pump with easy-to-use Clik-Tite® connectors.
The Mul•T•Pad provides the interface for delivering the temperature therapy. The unique button design allows water to ow and provides trouble-free operation when the pad is folded to form a customized t. This reduces the number of pads your facility must keep in inventory. The pads are applied to the part of the body requiring therapy, and the circulating water maintains the pad at the setpoint temperature. The setpoint temperature can be locked to prevent tampering.
Connecting the Pads
Figure 1A: Localized Temperature Therapy System with Single Pad with the TP702.
The Mul•T•Pads can be interconnected using Clik-Tite® connectors to provide therapy to more than one body site at a time.
Figure 1B: Localized Temperature Therapy System with Multiple Pads.
2 www.gaymar.com
OPERATOR'S MANUAL TP702 T/PUMP
INTRODUCTION
Catalog Descriptions
Catalog # Product Name
TP702 Professional Control Unit
TP12E Mul•T•Pad: 13"w x 18"l (33cm x 45cm)
Nonwoven fabric on one side, pliable polymer on the other side. 20 per carton
TP22E Mul•T•Pad: 15"w x 22"l (38cm x 55cm)
Nonwoven fabric on one side, pliable polymer on the other side. 20 per carton
TP22G Mul•T•Pad: 15"w x 22"l (38cm x 55cm)
Heavy Polymer, Reuseable. 10 per carton
TP26E Mul•T•Pad: 18"w x 26"l (45cm x 66cm)
Nonwoven fabric on one side, pliable polymer on the other side. 10 per carton
To order any of these products contact your dealer or Gaymar's International Department at:
International: +1 716-662-8636
Fax: +1 716-662-0730
Or, visit our website at www.gaymar.com
3www.gaymar.com
INDICATIONS AND CONTRAINDICATIONS
OPERATOR'S MANUAL
INDICATIONS FOR USE
Localized temperature therapy is recommended in treating the following applications:
Orthopedic Conditions such as: acute injuries, chronic pain, lower back pain, muscle spasm and
strains
Skin Trauma such as: abscesses, boils, bruises, burns and contusions
Cold Indications muscle spasm contusions tendonitis pain management
Other conditions such as: chronic arthritis, neuritis, phlebitis, tendonitis and I.V. inltration,
infection and localized pain.
Other applications only as prescribed.
OK for Use With:
• Non-acute traumatized tissue.
CLINICAL CONTRAINDICATIONS
• Impaired mental status.
• Insensate body surface.
O2 therapy. However, if oxygen tent is in use, do not use pump inside tent.
Heating contraindications are:
• Application to a body surface with compromised blood ow (Ischemia, area under pressure, arterial insufciency).
• Application to a patient with an increased tendency to bleeding (aggravates potential for hemorrhage).
• Application to a body surface with possibility of malignancy (tissue metabolism is increased and therefore, the growth potential of the malignant tissues).
• Treatment of hematoma within rst 24-48 hours (potential for re-bleeding and hemorrhage).
• Recent sprain or fracture (acute inammatory response).
• In combination with topical solutions whose toxicity may be affected by the application of heat.
• In combination with other heat sources.
Cooling contraindications are:
• Application to a body surface with compromised blood ow or hypersensitivity to cold. (Ischemia, Reynauds Syndrome).
4 www.gaymar.com
OPERATOR'S MANUAL TP702 T/PUMP
FEATURES
PUMP FEATURES
Easy to Use Keypad
Attached hose 10 ft (305 cm) dual hose. Connectors allow pads to be
Flow indicator Indicates no ow. Turns off heater if pump is tipped.
Warm/Cool Delivery
Therapy Cycles Choose from 20-minute, 30-minute, or Continuous
On/Standby Button
Over Temp Safety Thermostat
Self Check Automatic system check at startup.
Hose/cord Management
Comfortable Handle Design
Dual Micro Processor
Tethered Easy­Open Cap
Handle Vents The vents in the handle allow air ow to keep the
See KEYPAD FEATURES on page 6.
connected to the pump (Figures 1A and 1B, page 2).
Four temperature setpoints on the TP702.
cycles.
Indicates power is supplied to the unit.
Limit thermostat shuts off heater if the high temperature limit is exceeded.
Convenient and easy storage areas for hose and cord.
Designed for a more comfortable grip when moving the pump.
Two electronic circuits, one over temperature sensing circuit.
Prevents misplacing the cap. Only 1/4 of a turn is needed to remove or secure the cap.
motor and heater inside the unit cool.
Easy to Use Keypad
Flow Indicator
Temperature Setpoints for
warm or cool delivery
Therapy Time Cycles)
Vents in Handle
Figure 2: T/Pump Features
On/Standby Button
Comfortable Handle
Tethered Easy-Open Cap
Cord storage
Hose storage
Hose Connections
5www.gaymar.com
FEATURES
KEYPAD FEATURES TP702
Indicator/ Warning Light
OPERATOR'S MANUAL
Figure 3: TP702 Keypad
See the Troubleshooting section on page 11.
Water Flow
Check hoses or clamps for kinks or occlusions.
Water Level
Check water level.
Setpoints Press the button at the bottom of the setpoint indicator
to toggle through the four setpoints. Temperatures are identied in °C.
Setpoint Lock
Therapy Cycles
Prevents tampering.
Press and hold for 2 seconds to lock or unlock the setpoint.
Continuous cycle,
30-minute cycle
20-minute cycle.
Green indicates the unit is on. Yellow indicates power is supplied to the unit but the unit is not on.
6 www.gaymar.com