gaymar TP500 Operator's Manual

®
T/PUMP
TP500 / TP500C Heat Therapy System
®
LISTED 304L
OPERATOR'S MANUAL
P/N 10569-000 12/98
CONTENTS
Safety Precautions ...................... 1
Introduction................................. 2
OPERATOR'S MANUAL
TP500/TP500C T/PUMP
BEFORE YOU BEGIN . . .
Read and understand this T/PUMP OPERATOR'S MANUAL and all PRECAUTIONS (see page 1) prior to using the T/Pump.
Features/Specifications .............. 3
Operating Instructions ............... 5
Storage/Cleaning......................... 7
Troubleshooting.......................... 8
Warranties................................... 9
RECEIVING INSPECTION
Check the shipping carton for damage immediately upon receipt. If package damage is discovered, the device should be unpacked with the carrier's agent present. Any claims for shortage or damage must be filed with the delivering carrier by the purchaser. Do not return pumps damaged in shipment to GAYMAR without contacting our Technical Service Department for advice (see phone numbers below). If damaged goods are returned to GAYMAR without notifying the carrier, GAYMAR will assume the repairs will be made at the customer's expense.
TO RETURN PUMPS TO FACTORY FOR REPAIR OR EXCHANGE
Merchandise returned to GAYMAR must be accompanied by a Return Goods Number (RG#), issued by GAYMAR, authorizing goods to be returned. Call Customer Service or Technical Service at
(716) 662-2551 1 800 828-7341
Advise model, serial number, and nature of problem. You will be given a Return Goods Number (RG#).
www.gaymar.com P/N 10569-000 12/98
The serial number can be found on the back of the T/Pump (see figure 2, page 3).
Should you desire, repairs can be made in your facility by your own qualified personnel. Contact the GAYMAR Customer Service Department to request a T/PUMP SERVICE MANUAL. Call GAYMAR for spare parts required.
T/PUMP, T/PAD, Mul•T•Pad, Clik-Tite, and Gaymar are registered trademarks of Gaymar Industries, Inc. U. S. PATENT 4,068,870
© 1998. Gaymar Industries, Inc. All rights reserved.
OPERATOR'S MANUAL
WARNING
TP500/TP500C T/PUMP
SAFETY PRECAUTIONS
SAFETY PRECAUTIONS
DANGER
• Risk of explosion. Do not use in the presence of flammable
anesthetics.
• Risk of electric shock. Disconnect power before servicing the
T/Pump.
This device pumps warmed water through a pad. Set pad tem-
perature only as prescribed by and under the guidance of a physician. Monitor the patient's temperature and skin condition every 20 minutes or as directed by a physician. Failure to
adhere to these warnings could result in patient injury.
The following Groups/Conditions require additional surveillance:
Group/Condition at risk Potential injury
Pediatric patients Hyperthermia/
hypothermia Patients with impaired circulation Ischemia Areas of application are under pressure Ischemia In combination with topical solutions whose toxicity
may be affected by the application of heat In combination with other heat sources Thermal injury
Chemical
injury
www.gaymar.com P/N 10569-000 12/98
Only qualified medical service personnel should repair the T/Pump. Improper repair may result in death or serious
injury, equipment damage, or malfunction.
Always perform the FUNCTIONAL CHECK AND SAFETY
INSPECTION as specified in the SERVICE MANUAL after making repairs and before returning the T/Pump to patient use.
Improper repair may result in death or serious injury, equipment damage, or malfunction.
1
INTRODUCTION
OPERATOR'S MANUAL
TP500/TP500C T/PUMP
Single pad Multiple pads
INTRODUCTION
(TP500 only)
Figure 1—T/Pump Heat Therapy System
Heat therapy is effective in the dilation of blood vessels, thereby increasing the blood flow to the heated area. Heat therapy has a variety of uses, the most common being treatment of aches and pains in joints and muscles.
The GAYMAR T/Pump® Heat Therapy System provides a means of applying heat therapy by supplying temperature-controlled water through a connector hose to a Gaymar T/Pad®. The hose is terminated in easy-to-use Clik-Tite® or Colder-style connectors.
The T/Pad provides the interface for delivering the heat therapy. The unique button design allows water to flow and provides trouble-free operation when the pad is folded. This reduces the number of pads your facility must keep in inventory. The pads are applied to the part of the body requiring heat therapy, and the circulating water maintains the pad at the set point temperature. The temperature set point is key-operated to prevent tampering.
www.gaymar.com P/N 10569-000 12/98
The T/Pads can be interconnected (on model TP500 only) to provide therapy to more than one body site at a time.
2