Gaymar SPR Plus III Service Manual

SPR•Plus® IIISPR•Plus® III
P/N 100695000 Rev A 10/10
Service Manual
SPR•Plus® III Control Unit
• CL360
LISTED UL 2601-1
LISTED CAN/CSA C22.2 NO. 601.1
®
®
SPR•PLUS® III PRODUCTS ARE MADE IN THE USA
GAYMAR
®
GAYMAR INDUSTRIES, INC.
10 Centre Drive
Orchard Park, NY
14127-2295
Phone:
1.800.828.7341
716.662.2551
FAX:
1.800.993.7890
716.662.0730
SPR•Plus® IIISPR•Plus® III
24
This page is left blank intentionall y
SPR•Plus® IIISPR•Plus® III
Important
Use the SPR•Plus® III Control Unit only under the direction of a physician.
Before using the SPR•Plus III Control Unit, read and understand the SPR•Plus III Operator's Manual and the SAFETY PRECAUTIONS.
Only qualied medical service personnel should repair the SPR•Plus III Control Unit.
In the event of any questions, contact Gaymar's Technical Service Department for assistance.
USA:
Phone 1.800.828.7341
(716) 662-2551
Fax (716) 662-8795
International:
Phone (716) 662-8636
Fax (716) 662-0730
SPR•Plus is a registered trademark of Gaymar Industries, Inc. © 2000. Gaymar Industries, Inc. All rights reserved. www.gaymar.com
Table of Contents
Section Description Page
1.0 Indications/Contraindications/Precautions . 1
2.0 Receiving Inspection ............................................2
3.0 Repair Policy ............................................................2
4.0 Specications ..........................................................3
5.0 Care and Maintenance .........................................3
6.0 Theory ........................................................................4
7.0 Control Panel ...........................................................6
8.0 Functional Check and Safety Inspection .......7
9.0 Troubleshooting .................................................. 10
10.0 Repair Procedures ............................................... 12
Illustrations
Figure Description Page
1 SPR•Plus III System ................................................. 4
2 SPR•Plus III Control Unit ......................................5
3 Control Panel ........................................................... 6
4 Cushion Simulator .................................................8
5 Calibration Check ................................................11
6 Compressor Pump .............................................13
7 PC Board .................................................................14
8a Pump Assembly ................................................... 15
8b Pump Assembly ................................................... 17
9 Compressor and Support Assembly ............ 19
10 Wiring Diagram ................................................... 21
Tables
Table Description Page
1 Func tional Check /
Safety Inspection Form ........................................9
2 Troubleshooting Chart...................................... 10
23
This page is left blank intentionall y
SPR•Plus® IIISPR•Plus® III
1
Contraindications/Safety Precautions
• If a large bulge in the top surface of the mattress is noticed, call your Dealer or Gaymar Industries' Customer Service Depar tment. A large bulge may cause the patient to roll over when not desired, resulting in potential injury or discomfort.
• U. S. Federal law restricts this device to sale by or on the order of a physician.
• For grounding reliability, plug only into a grounded outlet labeled “Hospital Grade.”
• Use minimal layers of sheeting and incontinence pads. Too many layers between the patient's skin and the SPR•Plus III Cushion will reduce the pressure-redistibution capabilities of the therapy system.
• Do not pull linens tightly over cushion, but leave loose. Tight sheets may cause “hammocking” and reduce eectiveness of therapy.
• Some medical conditions may not respond to therapy of this type. Patient's skin should be inspected regularly for redness or breaks. If redness or breaks in the skin occurs, consult the attending physician.
• Disconnec t power before servicing the SPR•Plus III Control Unit.
• Repairs should be per formed only by qualied personnel familiar with repair practices for servicing medical devices.
1.0 Indications for Use
This device is intended to assist in prevention and treatment of decubitus ulcers (bed sores).
1.1 Contraindications
Air support therapy is not recommended when spinal stability is a concern.
1.2 Safety Precautions
Review the following SAFETY PRECAUTIONS prior to using the SPR•Plus III System.
• Explosion hazard. Do not use in the presence of ammable anesthetics.
• Risk of electric shock. Refer servicing to qualied service personnel.
• Be sure the cushion is securely installed on mattress. Be sure straps are securely positioned over corners of mattress. Failure to secure cushion may permit cushion shifting and may cause improper therapy or patient injury.
• Do not operate system with bed side rails down. Patients should never be left unattended with lowered side rails, since patient injury could occur.
• Do not reuse the SPR•Plus III Cushion. Reuse of cushion may risk cross-contamination and infection.
• Disinfect the control unit and hose between patient installations, utilizing standard hospital protocol and disinfectants. Failure to disinfect may risk cross-contamination and infection.
• Deate mattress before performing CPR or CPR will be ineective.
• Keep bedding dr y. Use incontinence pads if necessary. Failure to keep the patient's skin clean and dry may lead to maceration.
(continued next column)
Wiring Diagram
22
8
SPR•Plus® IIISPR•Plus® III
2
Receiving Inspection/Repair Policy
2.0 Receiving Inspection
Upon receipt, unpack the SPR•Plus III Control Unit and inspect for concealed damage. Save all pack ing material. If any damage is found, notify the carrier at once and ask for a written inspection. Prepare a written description of any damage. Photograph any damage.
Failure to take the above action within 15 days of receipt may result in loss of claim.
Do not return the SPR•Plus III Control Unit to GAYMAR. Contact Gaymar's Technical Service Department for instructions.
USA only: 1 800 828-7341 (716) 662-2551 Outside USA: (716) 662-8636
3.0 Repair Policy
The SPR•Plus III Control Unit (CL360) is warranted free of defects in material and workmanship for a period of one (1) year.
The SPR•Plus III Low Air Loss Cushion (CL212/CL272) is warranted free of defects in material and workmanship for a period of thirty (30) days.
The Control Unit and Low Air Loss Cushion are warranted under the terms and conditions of the Gaymar warranty in place at the time of purchase. A copy of the warranty is available upon request. Gaymar disclaims all implied warranties including, but not limited to, the implied warranties of merchantability and of tness for a particular purpose.
Control Units may be returned to the factory for servicing (see 3.3, Return Authorization).
For customers who choose to repair Gaymar SPR•Plus III Control Units at their location, this manual contains information to allow a qualied biomedical technician to make necessary repairs. For technical support, contact Gaymar's Technical Service Department.
3.1 In-Warranty Repairs
All in-warranty eld repairs must be authorized by Gaymar's Technical Service Department before proceeding.
3.2 Out-of-Warranty Repairs
The following repair options are available when servicing SPR•Plus III Control Units:
1. Defective Components - replacement parts may be ordered by specifying the Gaymar part number as shown in the parts lists (pp. 13 – 23).
2. SPR•Plus III Control Unit Repairs - If the SPR•Plus III Control Unit becomes inoperative and the cause cannot be determined, the complete control unit may be returned to the factory for servicing at the purchaser's expense (see 3.3, Return Authorization).
3.3 Return Authorization
Please be sure to obtain a return goods (RG) authorization number from GAYMAR's Customer Service Department before returning the SPR•Plus III Control Unit or any component parts to GAYMAR.
USA only: 1 800 828-7341 (716) 662-2551
Outside USA: (716) 662-8636
21
Wiring Diagram
Figure 9—Wiring Diagram
ATTENTION
Static-Sensitive Devices
NOTES:
1. ALL RESISTANCES ARE IN OHMS.
2. ALL RESISTORS ARE 5% CARBON UNLESS NOTED.
3. ALL CAPACITORS ARE IN MICROFARADS UNLESS NOTED.
4. CAUTION: THE ACTUAL ASSEMBLIES ARE SUSCEPTIBLE TO DAMAGE FROM ELECTROSTATIC DISCHARGE. HANDLING OF THE ASSEMBLIES SHALL BE ACCOMPLISHED WITH ADEQUATE GROUNDING PR OVISIONS TO PREVENT ELECTROSTATIC DISCHARGE DAMAGE.
5. THIS DRAWING CORRELATES TO DRAWING (D) 09055 REVISION: D.
6 FOR CLP250 & CL360 P/N 09055-000 F3 AND F4 ARE 3/10A. FOR CL750 P/N 09055-001 F3 AND F4 ARE .25A. 7 FOR CLP250 & CL360 P/N 09055-000 JUMPERS ARE IN 120V LOCATIONS. FOR CL750 P/N 09055-001 JUMPERS ARE IN 240V LOCATIONS. 8 FOR CL750 P/N 09055-001 C12 IS 0.01-800V.