Please read all precautions and instructions before
using the Sof•Matt® Low Air Loss Mattress System.
Maximum benefit to the patient at the greatest margin
of safety requires a thorough knowledge and
understanding of the correct operation and
application of the Sof•Matt Low Air Loss Mattress
System.
Users must follow the instructions and precautions
included with each mattress system prior to use.
Operate the equipment with the same care you
would use when operating precision medical
equipment.
WARRANTY
Sof•Matt Low Air Loss Mattress Systems are
warranted against defects in material and
workmanship under normal use and operation from
the date of shipment for a period of one year.
During the warranty period, Gaymar will repair or
replace at its sole option, free of charge, any
defective parts or products returned with prior
authorization prepaid to Gaymar Industries, Inc.
See Incoming Inspection and Return Policy.
Warranty does not cover products abused, misused,
or altered outside factory. There are no obligations
on the part of Gaymar for consequential damages
arising out of or in connection with the use or
performance of the product. Gaymar disclaims all
implied warranties including, but not limited to, the
implied warranties of merchantability and of fitness
for a particular purpose.
INCOMING INSPECTION AND
RETURN POLICY
Check the shipping carton for damage immediately
after receipt. If damage is discovered, the product
should be unpacked with the carrier’s agent present.
Make a claim immediately to the carrier for the damage.
Do not return damaged goods without notifying the
carrier. If products damaged during shipment are
returned to Gaymar without notifying the carrier,
Gaymar will assume that repairs will be made at the
customer’s expense. Products returned to Gaymar
must be accompanied by Returned Goods (RG)
numbers. Returned Goods authorization and numbers
may be obtained by calling Gaymar Industries'
Customer Service Department:
Toll free:1 800 828-7341
Direct:(716) 662-2551
When inquiring, please supply model and serial number,
purchase order number, whether merchandise was
bought on contract, and reason for return.
In general, deleted and outdated merchandise will not
be accepted for credit. A restocking charge of 15% will
be assessed on returns of current merchandise.
Ship returns prepaid to Gaymar Industries, Inc.,
10 Centre Drive, Orchard Park, NY 14127-2295.
Sof•Matt Low Air Loss Mattress Systems purchased by
individuals for home use must be returned to the
dealer from which they were purchased for repair.
SYMBOLS USED
Attention: consult accompanying documents
Dangerous voltage
APPLICABLE SERIAL NUMBERS
This service manual applies only to models RSM176S
(no configuration) and RSM180S configuration ‘A’
Sof•Matt Mattress Systems. The 8th character
of the serial number identifies the configuration.
For example, S/N RSM180S A B10042
identifies a configuration ‘A’ system.
23Control Panel Parts ........................................................................ 35
24Functional Test Kit ........................................................................... 36
1
PRECAUTIONS
CAUTION
WARNING
Sof•Matt® Low Air Loss Mattress System
Operator's/Service Manual
CONTRAINDICATIONS
INDICATIONS
SOF•MATT MATTRESS
PRECAUTIONS
Air support therapy is not recommended when spinal stability is a concern.
This device is intended to help treat and prevent decubitus ulcers (bed sores).
Electric shock hazard. Do not remove pump cover. Refer
servicing to qualified service personnel.
Double insulated. When servicing, use only identical
replacement parts.
• There are no operator serviceable parts within the equipment. Refer
servicing, maintenance and FUNCTIONAL CHECK AND SAFETY
INSPECTION to qualified technical personel familiar with good
practice for medical device repair.
• Always perform the FUNCTIONAL CHECK after making repairs and
before returning the Sof•Matt Mattress System to patient use.
Otherwise, personal injury, damage to the Sof•Matt Mattress
System or improper therapy may result.
PATIENT
PRECAUTIONS
2
• Some medical conditions may not respond to this type of therapy.
Patient's skin condition should be inspected regularly.
If any redness or breaks in skin occur, consult a physician.
• Disinfect the mattress between patient installations, utilizing standard
hospital protocol and FDA approved disinfectants (such as a 10%
chlorinated bleach solution).
Failure to disinfect may risk cross-contamination and
infection.
• Keep bedding dry. Use incontinence pads if necessary.
Failure to keep the patient's skin clean and dry may lead to
maceration.
Sof•Matt® Low Air Loss Mattress System
Operator's/Service Manual
INTRODUCTION
Gaymar Sof•Matt Low Air Loss Mattress Systems are designed to aid in the
prevention and treatment of pressure ulcers. The Sof•Matt Low Air Loss Mattress
System (fig. 1) has an internal air pump which maintains adequate pressure in its
upper and lower cells.
Please read and understand the precautions and instructions within this manual
before attempting to use the mattress system. Carefully follow the instructions
included with each mattress. In addition, we recommend that you request
in-service training from your Gaymar representative.
INTRODUCTION
Figure 1—Sof•Matt Mattress System
PREPARING THE BED
The mattress is designed for use on bed frames which provide firm, solid
support. Use otherwise may cause the mattress to be ineffective.
NOTE: • When the mattress is used on a link-spring bed frame, it may
cause the RSM mattress to sag. Additional support can be
provided by using three appropriately sized support boards. The
RSM mattress contains straps inside the bottom of its cover to
accommodate these boards and hold them in place. See figure 2.
• Support boards (Gaymar P/N 10623-000) are available from
your dealer. Boards may also be made as follows: Use 1/2" thick
interior grade plywood board, good both sides. Cut each board
to 33-1/2 inches (85 cm) long by 9 inches (23 cm) wide. Radius
all corners. Sand all surfaces and edges smooth. Make sure all
edges are rounded and smoothed to prevent wood slivers from
causing punctures, which are not covered by warranty.
Figure 2—Support Boards
3
OPERATING INSTRUCTIONS
Sof•Matt® Low Air Loss Mattress System
Operator's/Service Manual
OPERATING
INSTRUCTIONS
NOTE: This product is not known to be susceptible to interference from
outside sources in its normal environment, non does it produce radiated
or line conducted emission that will interfere with other equipment in its
vicinity.
1.Place the Sof•Matt
at the proper end of the bed, facing up (fig. 1, p. 3).
2.Plug the power cord into a polarized electrical outlet. The mattress will
immediately begin to inflate. The power cord is the only means to
disconnect this product from the supply (MAINS).
3.Position the Firmness Control (located behind the control panel flap) initially
as shown in figure 3.
Mattress on the bed frame with the FOOT END markings
Figure 3—Firmness Control
4.The patient may be placed on the mattress after 45 minutes. The heel
section will completely inflate within 90 minutes. For rapid inflation, insert
a portable inflator (e.g., Gaymar model SC505) into the CPR valve and
inflate for one minute. Replace the CPR plug quickly to prevent air from
escaping.
5.Apply linens, leaving the red CPR strap exposed.
SECURING LINENS (OPTIONAL)
To secure linens to the Sof•Matt Mattress, thread the four corners
through the D rings attached to the mattress (fig. 4). To ensure proper
therapy, do not pull linens taut. The linens should remain loose and wrinkly
on surface of mattress.
Figure 4—Securing Linens with D Rings
4
Sof•Matt® Low Air Loss Mattress System
CAUTION
Operator's/Service Manual
6.Place the patient on the mattress. If necessary, the Firmness Control setting
may now be adjusted for patient comfort.
NOTE: When elevating the head of the bed above 30°,
the Firmness Control setting may need to be increased.
OPERATING INSTRUCTIONS
MAINTAINING
PATIENT COMFORT
AND PROTECTION
1.If the patient appears to be bottomed when the head of the bed has been
raised to a sitting position (30° or more above horizontal), adjust the
firmness as described in step 6, page 4. When the bed is returned to a flat
sleeping position (0°), adjust as necessary for patient comfort.
Too Soft
Too Firm
Figure 5—Visual Firmness Check
2.To determine if the Sof•Matt Mattress is functioning properly, visually
check twice each day that the mattress is correctly inflated (fig. 5).
Turn the pressure control knob clockwise to increase firmness.
If the inflation is still incorrect at the maximum setting, remove from
service and contact your dealer.
3.To provide greater comfort, follow the procedure in step 6, page 4.
Check patient after changing setting to confirm that the mattress surface
is correctly inflated.
Some medical conditions may not respond to therapy of this type.
Check patient's skin frequently. Consult physician if any
redness or skin break occurs.
Serious injury could result if the patient's skin condition is
left unchecked.
5
OPERATING INSTRUCTIONS
Sof•Matt® Low Air Loss Mattress System
Operator's/Service Manual
POWER OUTAGE/
PATIENT TRANSPORT
FEATURE
DEFLATING THE
MATTRESS
REINFLATING THE
MATTRESS
When power is removed from the mattress, therapy will be continued;
however, the height of the mattress will gradually begin to decrease.
To maintain therapy, power should be restored within 4 hours.
When transporting, unplug the power cord and bundle it on top of the
mattress to keep it from underfoot.
The lower cell will maintain internal cell pressure for approximately four hours.
The upper cell will partially deflate and the foot end will completely deflate.
The Sof•Matt Mattress can be deflated
for storage by pulling firmly on the red
CPR strap.
To reinflate, make certain to reinstall the CPR pin below the flap and to
resecure the retaining snap. Plug the system into a polarized electrical outlet.
The mattress will automatically inflate in approximately 90 minutes.
For rapid inflation, a portable inflator can be used by inserting it into the CPR
valve.
Figure 6—CPR Plug (pull strap to deflate)
6
Sof•Matt® Low Air Loss Mattress System
Operator's/Service Manual
CLEANING AND CARE
CLEANING AND CARE
1.Unplug the power cord from the electrical outlet.
2.To clean, use soap and water and a clean cloth to wipe down the power
cord and mattress. Do not use abrasive cleaners. Wipe dry with a clean
cloth. Do not autoclave.
NOTE: Blood and other body fluids must be thoroughly cleaned
from all surfaces before applying the disinfectant.
3.Apply an FDA approved disinfectant or 10% chlorinated bleach solution
(chlorinated bleach with 5.25% sodium hypochlorite) to the external
surfaces of the mattress.
Allow to completely dry. The solution contact time is what makes
disinfection effective.
4.Wipe down the mattress with a clean, dry cloth to remove any excess
disinfectant.
NOTE:• Mattress cover must be dry prior to storage or
application of linens.
• Failing to remove excess disinfectant could
eventually degrade the polyurethane coating of the
mattress cover.
STORAGE AND CARE,
SOF•MATT MATTRESS
AND COVER
Individual states have established rules and regulations pertaining to
the cleaning and sterilization of mattresses that are intended to be
resold, rented, or otherwise transferred from one user to another.
Some states prohibit reselling mattresses and require them to be
disposed of properly. Gaymar recommends contacting your state
bedding department to obtain information on your state's rules and
regulations if reusing, renting, or transferring mattresses.
Clean in accordance with the Cleaning and Care instructions above.
Plastic material should not be exposed to high temperature. Do not allow heat
lamps or heating pads to come close to the mattress cover. Do not autoclave.
Do not place sharp objects in contact with the mattress. Holes in the mattress
will degrade system performance.
7
SPECIFICATIONS
Sof•Matt® Low Air Loss Mattress System
Operator's/Service Manual
SPECIFICATIONS
LACISYHP
snoisnemiDmc02xmc391xmc98
thgieWgk2.21
epO
LACIRTCELE
noitacifissalC,straptneitapdeilppaonhtiwFBepyT
()"8x"67x"53
ro
mc02xmc302xmc98
)"8x"08x"53(
()bl72
tneibmAgnitar
egnaRerutarepmeT
tropsnarTdnaegarotS
egnaRerutarepmeT
tF°02
()F°021o
C°2.23otC°6.51
)F°09otF°06(
C°9.84otC°7.6-
elbatropsnart,detalusnielbuod,IIssalC
nitnocrofelbatiustnempiuqe
suou
noitcetorprofdeifissalctoN.noitarepo
toN.diuqilfossergnilufmrahtsniaga
serpehtniesurofdeifissalc
.scitehtsenaelbammalf
foecne
tupnIAV9,zH06,V021~
droCrewoPdrocGWA81,)tf51(m6.4
edisnidetacol(sesuF
elbisseccaton,pmup
)rotarepoehtot
egakaeLtnerruC.xamserepmaorcim001
SLAVORPPAEDOC1-1062LU
alophtiw
erepma4/1
gulpdezir
,gnitcatsafyrev,V052,erutainimbuS
1.106.oN2.22ASC/NAC
8
Sof•Matt® Low Air Loss Mattress System
CAUTION
Operator's/Service Manual
FUNCTIONAL CHECK
•The FUNCTIONAL CHECK must be performed by qualified technical
personnel familiar with good practice for medical device repair.
Otherwise, damage to the Sof•Matt System and
improper therapy may result.
•Testing will verify proper operation of the Sof•Matt System.
Always perform the FUNCTIONAL CHECK after making repairs.
Record each maintenance check on the INSPECTION FORM
(fig. 8, p. 11).
Operation of the Sof•Matt System at other than specified
operating pressures can result in poor or reduced patient
support. Improper repair can result in patient injury.
TOOLS REQUIRED
FUNCTIONAL CHECK
1. OPERATING
PRESSURE TEST
• Functional Test Kit (P/N 78281-000—see fig. 24, p. 35)
• Manometer, 0-50 mm Hg (± 1 mm Hg accuracy)
Operate the Sof•Matt Mattress for at least two hours at nominal
120VAC line voltage prior to performing the Operating Pressure Test.
NOTE: Perform steps 1-3 prior to the two-hour warm-up.
1.See figure 7 (p. 10). Use the Functional Test Kit (fig. 24, p. 35). Replace the
CPR plug with the test pin. Connect the female CPC connector and hose
to the manometer.
2.• To measure lower cell pressure—Connect the shortest tube (L) from
the test pin to the connector on the manometer.
• To measure foot cell pressure—Connect the medium length tube (F)
from the test pin to the connector on the manometer.
• To measure upper cell pressure—Connect the longest tube (U) from
the test pin to the connector on the manometer.
3.Turn the Firmness Control knob to the lowest position (Softer), at
approximately 8 o'clock. After a two-hour warm-up, the cell pressures
should be:
• lower cell pressure:25.0 ± 2.0 mm Hg;
• foot cell pressure:5.5 ± 2.0 mm Hg;
• upper cell pressure:12.0 ± 2.0 mm Hg.
9
FUNCTIONAL CHECK
1. OPERATING
PRESSURE TEST
(continued)
Sof•Matt® Low Air Loss Mattress System
Operator's/Service Manual
Figure 7—Test pin
2. CURRENT
LEAKAGE
4.Turn the Firmness Control knob to the highest position (Firmer), at
approximately 5 o'clock. After the system stabilizes, the cell pressures
should be:
• lower cell pressure:25.0 ± 2.0 mm Hg;
• foot cell pressure:5.5 ± 2.0 mm Hg;
• upper cell pressure: 32.0 ± 3.0 mm Hg.
5.If the pressure(s) in steps 3 or 4 are not met, see Troubleshooting, page 12.
NOTE: When testing has been completed, remove the manometer
and test pin. Replace the original CPR dump plug.
The internal pump contained within the mattress system has double
insulated construction. It is not necessary to measure current leakage.
However, if you choose to do so, refer to UL 2601-1, Clause 19.
3. GROUND
RESISTANCE
10
Measuring ground resistance is not applicable to the internal pump.
Double insulated appliances are permitted to have two-conductor
cords (refer to NFPA 99--Standard for Health Care Facilities, paragraph
9-2.1.2.2).
Loading...
+ 28 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.