gaymar MTA6900 Operator's Manual

®
Medi-Therm® III Hyper/Hypothermia Machine MTA6900
Manual Mode Automatic Mode Monitor Mode
www.gaymar.com P/N 11027-000 9/00
LISTED 303L
R
CLASS [CATEGORIE DE RISQUES] NO. 2G
Operator's Manual
P/N 11027-000 9/00
Before you begin . . .
Table of Contents
Section Description Page
1.0 Safety Precautions ................................. 1
2.0 Description ............................................. 2
3.0 Features ...................................................3
4.0 Operator Control Panel ...................... 4
5.0 Start-up Procedure ............................... 5
6.0 Shutdown Procedure............................ 7
7.0 Modes of Operations ........................... 8
Manual Mode.......................................... 8
Automatic Mode ................................. 10
Monitor Mode......................................12
8.0 Patient Safety........................................14
9.0 Care of Medi-Therm III......................16
10.0 Alerts/Troubleshooting ......................18
11.0 Blankets/Accessories .......................... 20
12.0 Customer Training ............................... 20
13.0 Warranty ............................................... 20
Medi-Therm® III
Important
U. S. Federal law restricts this device to sale by or on the order of a physician.
Read and understand this Medi-Therm III Operating Manual and all Safety Precautions (see page 1, section 1.0 and pp. 14–15, section 8.0) prior to using the Medi-Therm III Hyper/Hypothermia Machine.
Receiving Inspection
Refer to the Medi-Therm III Service Manual for Receiving Inspection and Check-out Procedures.
If you have any questions, contact Gaymar's Technical Service Department for assistance.
Illustrations
Figure Description Page
1 Medi-Therm III System........................ 2
2 Medi-Therm III Features ..................... 3
3 Medi-Therm III Control Panel........... 4
4 Pinch Clamps......................................... 5
5 Quick-disconnects ............................... 6
6 Clik-Tite Connector............................ 6
7 Float ........................................................ 6
8 Manual Mode......................................... 9
9 Automatic Mode ................................10
10 Monitor Mode.....................................12
11 Probe Check Well .............................. 17
Telephone: 1 800 828-7341
(716) 662-2551
Fax: (716) 662-8795
www.gaymar.com P/N 11027-000 9/00
Medi-Therm and Clik-Tite are registered trademarks of Gaymar Industries, Inc.
© 2000. Gaymar Industries, Inc. All rights reserved. www.gaymar.com
Medi-Therm® III
Safety Precautions
1.0 Safety Precautions
Read and understand this operating manual and all precautions listed below prior to using the Medi-Therm III Hyper/Hypothermia Machine.
DANGER
Do not use the Medi-Therm III machine in the presence of flammable anesthetics.
Risk of explosion can result.
WARNING
If the patient’s temperature is not responding or
does not reach the prescribed temperature in the prescribed time or deviates from the prescribed temperature range, notify the attending physician promptly. Failure to notify the physician
promptly may result in serious injury or death.
WARNING
Prevent excessive and/or prolonged tissue
pressure and shearing forces, especially over boney prominences. Skin damage may result.
Keep the area between the patient and the blanket dry. Excessive moisture may result in
skin damage.
Do not tip machine over without first draining
the water out and unplugging the power cord.
Damage to the machine or electrical shock can result.
Repairs should be performed by qualified
personnel such as certified biomedical electronics technicians or certified clinical engineers familiar with repair practices for servicing medical devices, and in accordance with the Medi-Therm III Service Manual. Do not attempt repairs to Gaymar medical equipment unless you possess these skills.
Damage to the machine or malfunction may otherwise result.
Power interruption will cause the Medi-Therm III
machine to go into a standby mode, resulting in no therapy to the patient. Follow instructions for desired mode to resume operation. Failure to
resume therapy could result in serious injury or death.
When performing the PROBE CHECK (p. 18,
10.1.2, item 4), use a protective sheath on the probe. Failure to use a sheath could result in
cross-contamination.
A physician’s order is required for setting blanket
temperature and for continued use of equipment. At least every 20 minutes or as directed by a physician, check patient’s temperature and skin condition of areas in contact with blanket; also, check blanket water temperature. Pediatric, temperature-sensitive, and operating room patients should be checked more frequently.
Failure to monitor patient may result in skin damage or inappropriate patient temperature.
Avoid placing additional heat sources between the patient and blanket. Skin damage may result.
(continued next column)
CAUTION
Fill with distilled water only. Failure to use distilled water may result in poor machine performance. Do not use alcohol. Alcohol may
accelerate blanket deterioration. Do not operate without water, since damage to internal components may result. Do not overfill. Overfilling may result in overflow because the water in the blanket drains back into the machine when the machine is turned off.
Place a dry absorbent sheet between the patient and the blanket when using all-vinyl blankets.
Moisture may accumulate which may result in skin damage.
www.gaymar.com P/N 11027-000 9/00
1
Description
Medi-Therm® III
Figure 1—Medi-Therm III Hyper/Hypothermia System
2.0 Description
This manual describes the operation of the Medi-Therm III Hyper/Hypothermia machine.
The Medi-Therm III Hyper/Hypothermia machine provides a means of regulating patient temperature by supplying temperature-controlled water through a connector hose to a Gaymar Hyper/Hypothermia blanket. The companion blanket provides the interface for heating or cooling the patient. A patient probe senses patient temperature.
www.gaymar.com P/N 11027-000 9/00
The Medi-Therm III machine, Hyper/Hypothermia blanket, connector hose, and patient probe comprise the Medi-Therm III system (fig. 1).
Please read and understand the precautions on page 1 and the following operating instructions before attempting to use the Medi-Therm III system. In addition, we recommend you request in-service training from your Gaymar representative.
2
Medi-Therm® III
3.0 Features
Please familiarize yourself with the location and purpose of the features (fig. 2) before reading the start-up procedures or attempting to operate the Medi-Therm III Machine.
Features
Water Fill Opening
Lift cover. Fill unit with distilled water until green band on float is fully visible (p. 6)
Operating Instructions & Precautions
Condensed instructions are located on the cover of the water fill opening and on the rear-facing side of the Operator Control Panel's sloped surface.
Operator Control Panel
Identifies displays, indicators, and controls used in Medi-Therm III operation (pp. 4-5).
Patient Probe Jack
Accepts patient probe plug connecting patient probe to machine (pp. 10, 12). Use YSI 400 series probe or equivalent.
Blanket Connections & Probe Check Well
Quick-disconnects facilitate connection to blanket water supply (p. 6). PROBE CHECK well verifies patient probe (p. 18).
ON/OFF Switch (Circuit Breaker)
Controls power to Medi-Therm III.
www.gaymar.com P/N 11027-000 9/00
Figure 2—Medi-Therm III features
3
Operator Control Panel
Medi-Therm® III
Figure 3—Medi-Therm III Control Panel
4.0 Operator Control Panel
The control panel is divided into sections according to function and use. The principal panel sections are identified in figure 3.
STATUS SECTION
ALERT SECTION
ALERT indicators light when some form of
operator action is required. In addition to the visual indicator, an audible alarm may sound. ALERT controls allow the Medi-Therm III operator to test panel lights or to silence the audible alarm.
www.gaymar.com P/N 11027-000 9/00
STATUS displays light to indicate both the machine’s flow status and how the blanket water temperature is being controlled to reach the desired temperature setting. STATUS displays function in both MANUAL and AUTO modes.
FLOW-OK indicates blanket water is flowing adequately.
HEAT indicates blanket water is being heated.
The maximum blanket temperature is 42°C (107.6°F).
IN-TEMP indicates the Medi-Therm III is controlling temperature at the desired setting.
COOL indicates blanket water is being cooled. Minimum blanket temperature is 4°C (39.2°F).
4
CHECK WATER FLOW indicates the
flow of water to the blanket has been restricted. The operator can choose whether an audible alarm accompanies this alert. (p. 7, section 5.6)
CHECK PATIENT PROBE indicates the patient probe is sensing an abnormal patient temperature (below 29°C [84.2°F] or above 45°C [113°F]). It may indicate the patient probe has become dislodged, an incorrect probe is being used, or a probe has become defective. An audible alarm accompanies this alert.
ADD WATER indicates the water reservoir is filled to less than the recommended level.
Medi-Therm® III
Start-up Procedure
REMOVE FROM USE NOW / MACHINE SHUTDOWN indicates the Medi-Therm III has
shut down due to blanket water temperature exceeding either the high or the low limit. An audible alarm accompanies this alert. Remove the Medi-Therm III
machine from use immediately.
TEST LIGHTS confirms that all indicator lights and digital displays are working, and that the audible alarm is functioning. The displays flash and the audible alarm sounds as long as the TEST LIGHTS button is pressed.
SILENCE ALARM temporarily silences the audible alarm. The alarm will reactivate in approximately 5 minutes if the condition persists.
NOTE: The SILENCE ALARM button will not function if
the REMOVE FROM USE NOW / MACHINE SHUTDOWN indicator is lit.
MODE SECTION
The SELECT indicator lights when a choice of operating modes is required. MODE buttons select the desired operating mode. Patient therapy is provided in either MANUAL or AUTO mode. A third mode, MONITOR, may be used to monitor patient temperature without providing therapy. After the mode button is pressed, the indicator at the top of the mode button lights. In AUTO mode, one of three additional CONTROL OPTION indicators light.
MANUAL selects the manual mode of operation. In MANUAL mode, the operator must observe patient temperature and adjust the blanket water set point temperature to obtain the desired results.
AUTO selects the automatic mode of operation. In AUTOMATIC mode, the operator sets the desired temperature and the patient temperature is automatically regulated to the operator specified set point. Automatic
cooling can be adjusted to RAPID, MODERATE, or GRADUAL rates by pressing the AUTO button again.
MONITOR selects the monitor mode of operation. In MONITOR mode, the operator may monitor patient temperature without providing therapy.
TEMPERATURE SECTION
TEMPERATURE displays identify BLANKET, PATIENT, and SET POINT temperatures.
BLANKET indicates the actual temperature of the water being delivered to the blanket. This display is illuminated when MODE is set to either MANUAL or AUTO.
PATIENT indicates actual patient temperature as sensed by the probe connected to the patient probe jack. This display is illuminated when MODE is set to MANUAL, AUTO, or MONITOR and a patient probe is plugged into the PATIENT PROBE jack.
SET POINT indicates the desired SET POINT temperature. When operating in MANUAL mode, pressing the SET POINT button (“” or “”) sets the desired BLANKET water temperature. When in AUTO mode, pressing the SET POINT button sets the desired patient temperature.
°C/°F displays the temperature in either degrees Celsius (°C) or Fahrenheit (°F). Pressing the °C/°F button will display the corresponding indicator.
5.0 Start-up Procedure
For your convenience, condensed versions of the following instructions have been permanently mounted on the cover of the water fill opening and on the rear­facing side of the Operator Control Panel's sloped surface.
www.gaymar.com P/N 11027-000 9/00
Figure 4—Pinch Clamps
Perform the following start-up procedure each time the Medi-Therm III system is placed in operation:
5.1 Close Pinch Clamps
Pinch clamps are provided on both the connector and blanket hoses to prevent water spillage when blankets are changed (fig. 4).
Close the pinch clamps on the connector hose and Hyper/Hypothermia blanket.
5
Start-up Procedure
Medi-Therm® III
5.2 Attach Blanket
Attach the Hyper/Hypothermia blanket to the Medi-Therm III machine (fig. 1, p. 2):
a. Connect a Gaymar Model DBK9 connector hose to
the Medi-Therm III machine. The hose is ten feet long. It has quick-disconnect (metal) fittings for the Medi-Therm III machine end and Clik-Tite® (plastic) connectors for the blanket end.
To connect the hose, attach one quick-disconnect fitting to the RETURN machine connection (top) and one to the SUPPLY (bottom). See figure 5. To insure a tight connection, push the retaining collar toward the cabinet and insert the quick-disconnect fitting firmly. While holding the fitting, allow the retaining collar to snap back into position. Check to make sure the fitting is fully engaged by tugging firmly on the hose fitting.
Return
5.3 Open All Pinch Clamps
Open all pinch clamps on the connector hose and the Hyper/Hypothermia blanket.
Failure to open the pinch clamps will prevent water flow and will trigger the CHECK WATER FLOW alert.
5.4 Check Water Level
Check to make sure water is at the proper level by raising the cover on the water fill opening. Fill with distilled water until the green band on the float is fully visible (fig. 7).
Figure 7—Float, showing proper water level
RETURN
SUPPLY
RETURN
SUPPLY
Supply
RETURN
SUPPLY
RETURN
SUPPLY
Figure 5—Quick-disconnects
b. Attach the two Clik-Tite connectors (fig. 6) on the
connector hose to the blanket. Insert the male fitting into the female. When fittings are fully seated, snap the locking ring into place to assure a positive lock.
CAUTION
Add distilled water only. Failure to use distilled water may result in poor machine performance.
Do not use alcohol, which may accelerate blanket deterioration.
Do not operate without water, since damage to internal components may result.
Do not overfill. Overfilling may result in overflow because the water in the blanket drains back into the machine when the machine is turned off.
www.gaymar.com P/N 11027-000 9/00
Figure 6—Clik-Tite Connector
6
Loading...
+ 16 hidden pages