Gas Measurement Instruments PS200 User Manual

PS200 Series
QUICK OPERATING INSTRUCTIONS
Refer to Essential Health and Safety Requirement (EHSR) and PS200
Series User Handbook before first use.
Switch ON
Press and hold the Right Hand button
for one second to switch the instrument ON and initiate a warm-up routine.
The display will show the serial number,
button to acknowledge.
Lastly, a sensor check is run. If sensors are
working OK and were zeroed correctly, a symbol is displayed adjacent to the gas type. If a
sensor fails the zero check, the display will be red, the audible / visual alarms activate and a symbol is displayed adjacent to the faulty gas type. T o acknowledge this fault, press the
button.
Switch ON - Manual Bump Test
Press and hold the button to switch the instrument ON and initiate a Manual Bump Test (after the warm-up routine is complete). When concluded, the instrument will operate as normal.
Normal Operation
The standard operating screen for the 4 Gas model is shown here. Other models will report the relevant gases they are capable of detecting.
Backlight / MAX & MIN
Press the button once to switch the backlight ON. (After 20 seconds it will automatically switch OFF .) Press again to view MAX readings and a further press to view the MIN readings. To reset MAX / MIN readings, press and hold the seconds when either MAX / MIN screen is displayed.
button for 2
Alarms Reset / Acknowledge
In alarm condition, the display will be red and the audible and visual alarms are activated. When the gas readings (of a Latching Alarm) are safe, press and hold the alarm. (If alarms are Non-Latching, the audible and visual alarms reset automatically when gas readings are safe.)
button to reset / acknowledge the
Switch Pump ON / OFF
Press and hold the button to toggle the pump ON / OFF (Pumped instruments only).
Manual Datalog
Press the button once to perform a manual datalog, briefly indicated on the display as ‘LOGGING’.
Switch OFF
Press and hold both and buttons, for 3 seconds, to switch the instrument OFF.
WARNINGS
LOW BATTER Y
Flashes when approximately 30 minutes operating time remains. The display flashes red, the audible alarm sounds once every two seconds and the red LED’s flash. Note: Gas alarms will continue to operate as normal.
BAT F AUL T
Flashes when approximately 3 minutes operating time remains. The display will be red, the audible alarm sounds continuously and the red LED’s remain on. After approximately 3 minutes the instrument automatically switches of f.
CHARGING
Before first use the PS200 instrument should be charged for up to 4 hours or until the ‘CHARGE COMPLETE’ message appears on the display.
The PS200 Series uses a Li-Ion (Lithium Ion) rechargeable battery. To recharge, securely connect the charging clip to the bottom of the instrument.Connect one end of the cable to the charging clip and the other to the universal charging adaptor (supplied) or any standard USB socket.
CAL EXPIRED
Displayed every 30 seconds if the calibration expiry date has been exceeded.
ZERO FAULT (and flashing
symbol)
Appears after warm-up if the instrument is switched on in gas or the instrument has been unable to zero all sensors. The display will be red, the audible alarm sounds once every 2 seconds, and the red LED’ s flash (start-up only). However , the instrument can still be used to detect and alarm on the other sensor(s) fitted.
Note:
1) If a flashing
symbol is alternating with a gas value, leave instrument on for 30 minutes then switch OFF, and then ON again in fresh air. If the fault remains, return instrument to a GMI approved Service / Repair facility.
2) If the symbol is displayed constantly, then the sensor may require replacement. Return instrument to a GMI approved Service / Repair facility.
FLOW FAULT (Pumped instruments only.)
Flashes to indicate a sample fault . The display will be red, the audible alarm sounds continuously and the red LED’s remain on. Check sample line / filters / and probe for blockage. Clear the blockage and restart pump by a press and hold of the
button.
----------------------------------------------------------------- Part Number: 64191 Issue 3 07/07/2011 -----------------------------------------------------------------
Refer to User Handbook for alternative charging options.
BUTTON OPERA TION
SINGLE PRESS PRESS & HOLD
&
LOCATING TONGUE
Backlight ON
MAX
MIN
Manual Datalog
LOCATING SLOT
CHARGING CLIP
USB CHARGING
CABLE
SWITCH ON
Pump ON / OFF
Acknowledge / Mute Alarm
Reset MAX / MIN
Switch ON - Manual Bump
Self Test
SWITCH OFF
when
(
displayed
)