Garland EM-C180 User Manual

SERVICE MANUAL SUPPLEMENT
CAUTION
WARNING TO SERVICE
TECHNICIANS
EXCESSIVE MICROWAVE ENERGY
(a) Do not operate or allow the oven to be operated with the door open. (b) Make the following safety checks on all ovens to be serviced before activating the magnetron or other microwave source, and make repairs as necessary: (1)Interlock operation, (2) proper door closing, (3) seal and sealing surfaces (arcing, wear, and other damage), (4) damage to or loosening of hinges and latches, (5) evidence of dropping or abuse. (c) Before turning on microwave power for any service test or inspection within the microwave generating compartments, check the magnetron, wave guide or transmission line, and cavity for proper alignment, integrity, and connections. (d) Any defective or misadjusted components in the interlock, monitor, door seal, and microwave generation
and transmission systems shall be repaired, replaced, or adjusted by procedures described in this manual
before the oven is released to the owner. (e)(i) A microwave leakage check to verify compliance with the Federal performance standard should be performed on each oven prior to release to the owner. (For U.S.A) (e)(ii) A microwave leakage check to verify compliance with the Canadian Regulation, HEALTH AND WELFARE,
SOR/79-920 should be performed on each oven prior to release to the owner. (For CANADA)
See the Service Manual of EM-C180US(SM-640179)
except the items described in this Service Manual.
PRODUCT CODE NO. 437-453-00
for Model EM-C180
Microwave Oven
FORM # EM-C180 (02/05)
CAUTION
For microwave energy emission On every service call. A check for micro-
wave energy emission must be made according to the following manner.
Measurement of energy emission
In the area emitting the highest reading, switch the meter to SLOW RESPONSE and take a reading for minimum of three (3) seconds. We recommended the pattern outline shown below when the door surface is surveyed. NOTE: Periodically check to be sure that the probe tip is not worn or dirty.
Measurement must be made with the microwave oven operating at its maximum output and containing a load
of 275±15 milliliters of tap water initially at 20°±5° celsius (68±9°F) placed within the cavity at the center.
NOTE:The water container m ust be a 600 milliliter bea-
ker and made of an electrically none conduc­tive material such as glass or plastic.
The cook tray must be in place when measuring
emission.
A properly operating door and seal assembly will nor­mally register emission no greater than 4 mW/cm2 to allow for measurement uncertainty with the cooking shelf or tray in place.
All repairs must be performed in such a manner that microwave energy emission is minimal.
Follow the instructions supplied with the detector be­ing used and perform an R.F. emission test around the door front, and all edges and vent of the outer case. The cabinet (wrapper) must be in place and the oven fully assembled. When performing an emission survey, with the meter on FAST RESPONSE, the movement of the detector probe shall not exceed one (1) inch per second.
The following U.S. standard applies to microwave ov­ens:
21 CFR 1030.10, Perf ormance Standard f or Microw a v e Ovens. It requires that the power density of the microwave radiation emitted by a microwave oven shall not ex­ceed five (5) milliwatts per square centimeter at any point 5 centimeters (about 2 inches) or more from the external surface of the oven.
All microwave ovens exceeding the emission level of 4 mW/cm2 must be reported to Dept. of Service for mi­crowave ovens and the manufacturer immediately. The owner should be told not to use the microwave oven until it has been repaired completely.
If a microwave oven is found to operate with the door open, report to Dept. of Service, the manufacturer and CDRH* immediately. Also tell the owner not to use the oven. *CDRH: Center for Device and Radiological Health.
The interlock monitor switch acts as the final safety switch protecting the customer from microwave radia­tion. If the interlock monitor switch operates properly and the door interlock switch fails, the fuse will blow. If this happens, all interlock switches must be replaced. The contacts of the interlock switches may be welded together.
- TABLE OF CONTENTS -
Adjustment Procedures.............................................1
Specifications............................................................ 2
Power output Measurement ......................................2
Precautions and Repair Service Tips .......................2
Circuit Diagram .........................................................3
1. ADJUSTMENT PROCEDURES
TO AVOID POSSIBLE EXPOSURE TO MICROWAVE ENERGY LEAKAGE, THE FOLLOWING ADJUST­MENTS OF THE INTERLOCK SWITCHES SHOULD BE MADE ONLY BY AUTHORIZED SERVICE PER­SONNEL.
The service center should have the designated detector to measure the microwave energy leakage after the repair or adjustment.
NOTE: Detector to be used at the service center is NARDA 8100, 8200 or the equivalent.
INTERLOCK SWITCH, INTERLOCK MONITOR SWITCH AND DOOR SENSING SWITCH ADJUST­MENT
(Figure 1) (1) Loosen 2 screws securing the lever stopper.
(2) Adjust the lever stopper position so that it is pushed
up and pulled backwards until there is zero gap
between the latch lever and the switch body on the
interlock switch. At the same time there is zero gap between the latch lever and the switch body on the door sensing switch when the door latch is securely
locked. (3) Tighten the lever stopper screws securely. (4) Make sure the interlock monitor switch closes after the interlock switch opens when the door is
opened very slowly, according to “CHECKOUT
PROCEDURE FOR SWITCHES” on page 6. (5) Make sure the interlock monitor switch opens before the interlock switch closes when the door is
closed very slowly, according to CHECKOUT PRO-
CEDURE FOR SWITCHES” on page 6. (6)(i) Make sure the microwave energy leakage should
be no greater than 4 mW/cm2 to allow to measure-
ment uncertainty when measured with a detector.
(All service adjustments must be made for mini-
mum microwa ve energy leakage readings .) (For US)
(6)(ii) Make sure the microwave energy leakage is below the limit of 1 mW/cm2 (at 275cc water load), 5 mW/cm2 (at no load) and 5 mW/cm2 (at 275 cc water load without cabinet) when measured with a detector.
Test Procedures and Troubleshooting...............4•`10
Disassembly Instructions .............................. 11•`14
Exploded View and Pa rts List ........................ 15•`22
Overall Circuit Diagram.................................23•`25
(All service adjustments must be made for minimum
microwav e energy leakage readings .) (F or Canada)
NOTE: If the interlock monitor circuit operates and at
the same time the fuse blows with the door open, be sure to replace the relay circuit board , Interlock switch and monitor switch.
Lever stopper
Interlock monitor switch
Latch Lever
Interlock switch (Secondary interlock switch)
Latch Lever
Door sensing switch (Primary interlock switch)
Screws
Figure 1
- 3 -
2.SPECIFICATIONS
Microwave output ................1,800W to 180W
Frequency ...........................2,450MHz
Power supply.......................208V, 60Hz
Rated current ......................13 Amp.
Safety Device......................
Thermal protector(Magnetron) 150°C(270°F)Open (Thermostat) 80°C(144°F)Close Thermistor (Magnetron) 200°C(360°F) Open 108°C(194°F)Close
Thermistor(Duct)................120°C(216°F) Open
Fuse (Cartr idge Type) .................250V 10A
Micro switch, Relay Interlock Switch
Interlock monitor Switch Door sensing Switch and Relay RL-3 and 4
Max. input time....................Electronic Digital, up to
Manual
Overall Dimensions........422(W)x540(D)x335(H) mm
Oven cavity size ............330(W)x330(D)x230(H) mm
Effective Capacity of Oven Cavity.........19.1liters
Net weight ...........................32Kg
3. POWER OUTPUT MEASUREMENT
NOTE: The power output specification, 1800W on this model is measured with IEC measurement. The power output is measured with two(2) liters water is equiva­lent to 1800W in measurement with IEC, when mea­sured with the following power output. (1)1. Fill two beakers, one liter of tap water respec­ tively
2. Use an accurate thermometer and measure each water temperature respectively. (2) Place beakers side by side in center of the ceramic tray. (3) Close the door,set the “TIME” for two minutes. Touch the “START” key and heat the water for exactly two minutes. (4)Take the beakers out, immediately stir the water and measure the water temperatures respectively. (5) Calculate the temperature rise of water in each beaker . Then calculate the average value of the two temperature rises.( ƒ¢t) (6) The teperature rise shall be in the following range; Average Temp. Rise Minimum 23.1°C Maximum 28.3°C Po w er output is affected by the line voltage under load. (7) For correct Power output measurement, the line voltage under load must be 208±2Volts.
4.PRECATIONS AND REPAIR SERVICE TIPS PRELIMINARY
10min./Memory 30min.
A. SINCE NEARLY 4,000 V OLTS EXISTS IN SOME
CIRCUITS OF THIS MICROWAVE OVEN, RE­PAIRS SHOULD BE CARRIED OUT WITH GREAT CARE.
B. TO AVOID POSSIBLE EXPOSURE TO MICRO-
WAVE ENERGY LEAKAGE, THE FOLLOWING PRECA TIONS MUST BE TAKEN BEFORE SER­VICING.
(1) Before the power is applied.
(a) Open and close door several times to make
sure the interlock switch, door sensing switch and interlock monitor switch operate properly. (Listen for the clicking sound from switches.)
Make sure the interlock monitor switch is
closed after the interlock switch and door sens-
ing are open when the door is opened. (See pages 1 and 6) (b) Make sure the perf orated screen and the choke
dielectric of the door are correctly mounted.
(2) After the power is applied.
(a) Open and close the door to see if the interlock
mechanism operates properly. (b) Check microwave energy leakage with a leak-
age detector and confirm the energy leakage
is below 5mW/cm
2
(3) Do not operate the unit until it is completely repaired of any of the following conditions. (a) Door is not closed firmly against the cavity front.
(b) The hinge is broken. (c) The choke dielectric or the door seal is dam-
aged. (d) The door is bent or warped, or there is any
other visible damage to the oven that may
cause microwave energy leakage.
Note: Always keep the seal clean.
(e) Make sure that there are no defective parts in
the interlock mechanism. (f) Make sure that there are no defective par ts in
the microwave generating and transmission
assembly. (especially wave guide). (4) The following items should be checked after
the unit is repaired. (a) The interlock monitor switch is connected cor-
rectly and firmly. (b) The magnetron gasket on the magnetron is
properly positioned. (c) Waveguide and oven cavity are intact. (No
leakage of microwave energy). (d) The door can be properly closed and the safety
switches work properly. (e) The oven must be stopped when the door is
opened or the time is up.
The ov en must not be operated with any of the abo v e components removed or bypassed.
- 2 -
6. TEST PROCEDURES AND TROUBLESHOO TING
CAUTION
-DISCONNECT THE POWER SUPPLY CORD FROM THE WALL OUTLET WHENEVER REMO VING THE CABINET FROM THE UNIT. PROCEED WITH TESTS ONLY AFTER DISCHARGING THE HIGH VOLTAGE CAPACITORS AND REMOVING THE LEAD WIRES ON THE PRIMARY WINDING OF THE HIGH VOLTAGE TRANSFORM­ERS FOR LOWER AND UPPER MA GNETRONS.
(SEE FIGURE 3) A. TEST PROCEDURES
COMPONENT
MAGNETRON
CHECKOUT PROCEDURE RESULT
1) Check for resistance: Across the filament ter­minal of the magnetron with an ohm - meter on Rx1 scale.
Figure 4
2) Check for resistance: Between each filament terminal of the magnetron and the chassis ground with an ohm-meter on highest scale.
PRIMARY WINDINGS
Secondary Windings
Filament Windings
Figure 3
Normal reading: Less than 1 ohm.
Normal reading: Infinite ohms.
HIGH-VOLTAGE TRANSFORMER
1) Measure the resistance:
With an ohm-meter on R x1 scale. a. Primary winding; b. Filament winding; c. Secondary winding;
2) Measure the resistance:
with an ohm-meter on highest scale. a. Primary winding to ground; b. Filament winding to ground;
Figure 5
Figure 6
Normal reading: Approximately 1.0 ohms
Less than 1 ohm. Approximately 60 ohms Normal reading:
Infinite ohms. Infinite ohms.
Note: Remove varnish of measured point.
- 6 -
COMPONENT
HIGH-VOLTAGE CAP A CITOR
Including internal bleeder resistor
CHECKOUT PROCEDURE RESULT
Measure the resistance: Across two terminals with an ohm-meter on highest scale.
Figure 7
Normal reading:
Momentarily indicates several ohms, and gradually to 10 meg-ohms.
Abnormal reading:
Indicates continuity or 10 meg-ohms from the beginning.
HIGH-VOLTAGE DIODE
Measure the resistance: Across two terminals with an ohm-meter on highest scale.
Figure 8
Normal reading:
Indicate about middle position in one direction (forward) and infinite ohms in the reverse direction, using ohm meter with a 9V battery.
NOTE
- Some digital meter may show more than 0 ohms or infinite ohms even in a forward direction because the low mea­suring voltage of the meter does not allow the meter to pass through the high voltage diode. Use an ohm meter with a 9V battery.
Abnormal reading:
Indicates continuity or infinite ohms in both directions.
- 7 -
COMPONENT
TOUCH KEY BOARD
CHECKOUT PROCEDURE RESUL T
Measure the resistance between terminals of FPC connector after removing it from S101.(Figure 10)
NOTE
- When reconnecting the FPC connector, make sure the holes on the connector are properly inserted in hook of the plastic fastener in S101.
MATRIX CIRCUIT FOR TOUCH KEY BOARD FPC CONNECTOR
When When not Resistance touched touched Value Less than More than 1 K ohms 1 meg ohms
When checking the POWER key , connect the ohm-meter as illustrated below.
TERMINAL OF FPC CONNECTOR FIGURE 10
CHECKOUT PROCEDURE FOR SWITCHES
Disconnect the lead wires from the switches and check for the continuity of the switches, connecting an ohm-meter to its terminals.
SWITCHES
(SEE Figure 1 on page 1)
INTERLOCK SWITCH
DOOR SENSING SWITCH
INTERLOCK MONITOR SWITCH
CHECKOUT PROCEDURES DOOR OPEN DOOR CLOSED
Terminals "COM" and "NO"
Ter minals "COM" and "NC"
CAUTION: After checking the switches, make sure that the interlock monitor switch is properly connected according
to the CIRCUIT DIAGRAM on page 3.
- 8 -
WARNING: When removing the cabinet, you m ust disconnect
the power supply cord from the wall oulet f or your safety. Only the chec kout procedure below needs the power supply on. TAKE GREAT CARE to avoid possible electrical shock.
For your safety, proceed with the test only after removing the wire leads from the pri­mary winding of the high voltage transformer
Primary Winding
Primary Winding
Lower
Upper
COMPONENT
POWER P.C.B
CHECKOUT PROCEDURE RESULT
Check voltage at S104 and S105 after remov­ing each connector (female) from power circuit board. Pin No.3 (Ground) and 4,5, 1,2 at S105. Pin No.1 and 2 at S104.
CAUTION:
For your safety, proceed with the test only after removing the wire leads from the pri-
mary winding of high voltage transformer. Test procedures: a) Make sure that the power supply cord is not
plugged in. b) Remove the connector S104 and S105 from
the power circuit board. c) Plug the power supply cord : back in. d) And then, measure each voltage.
Normal reading: Connection
Pin No., Voltage(V)
S105
#3 to #4 DC 12 #3 to #5 DC 16 #3 to #1 DC 30 #3 to #2 DC 35
S104
#1 to #2 AC 2.4
CONTROL P.C.B
Measure the voltage: Between test points TP-1, TP-2 ,TP-3 and ground (See figure 16 on page
23)
Note
- Proceed with the check of the control P.C.B to see if any one of the measured values is differ­ent from the specified values.
- 9 -
Test point
TP, Voltage(V) TP-1 DC - 5
TP-2 DC - 12 TP-3 DC - 16 TP-4 DC - 35
Loading...
+ 19 hidden pages