PRECAUTIONS TO BE OBSERVED BEFORE AND DURING
SERVICING TO AVOID POSSIBLE EXPOSURE TO
EXCESSIVE MICROWAVE ENERGY
(a) Do not operate or allow the oven to be operated with the door open.
(b) Make the following safety checks on all ovens to be serviced before activating the magnetron or other
microwave source, and make repairs as necessary:
(1)Interlock operation, (2) proper door closing, (3) seal and sealing surfaces (arcing, wear, and other
damage), (4) damage to or loosening of hinges and latches, (5) evidence of dropping or abuse.
(c) Before turning on microwave power for any service test or inspection within the microwave generating
compartments, check the magnetron, wave guide or transmission line, and cavity for proper alignment,
integrity, and connections.
(d) Any defective or misadjusted components in the interlock, monitor, door seal, and microwave generation
and transmission systems shall be repaired, replaced, or adjusted by procedures described in this manual
before the oven is released to the owner.
(e)(i) A microwave leakage check to verify compliance with the Federal performance standard should be
performed on each oven prior to release to the owner. (For U.S.A)
(e)(ii) A microwave leakage check to verify compliance with the Canadian Regulation, HEALTH AND WELFARE,
SOR/79-920 should be performed on each oven prior to release to the owner. (For CANADA)
SeetheServiceManualofEM-C180US(SM-640179)
excepttheitemsdescribedinthisServiceManual.
PRODUCT CODE NO.437-453-00
for Model EM-C120
Microwave Oven
FORM # EM-C120 (02/05)
CAUTION
For microwave energy emission
On every service call. A check for micro-
wave energy emission must be made
according to the following manner.
Measurement of energy emission
Measurement must be made with the microwave oven
operating at its maximum output and containing a load
of 275±15 milliliters of tap water initially at 20°±5°
In the area emitting the highest reading, switch the
meter to SLOW RESPONSE and take a reading for
minimum of three (3) seconds. We recommended the
pattern outline shown below when the door surface is
surveyed.
NOTE: Periodically check to be sure that the probe tip
is not worn or dirty.
The following U.S. standard applies to microwave ovens:
celsius (68±9°F) placed within the cavity at the center.
NOTE:The water container m ust be a 600 milliliter bea-
ker and made of an electrically none conductive material such as glass or plastic.
The cook tray must be in place when measuring
emission.
A properly operating door and seal assembly will normally register emission no greater than 4 mW/cm2 to
allow for measurement uncertainty with the cooking
shelf or tray in place.
All repairs must be performed in such a
manner that microwave energy emission
is minimal.
Follow the instructions supplied with the detector being used and perform an R.F. emission test around the
door front, and all edges and vent of the outer case.
The cabinet (wrapper) must be in place and the oven
fully assembled.
When performing an emission survey, with the meter
on FAST RESPONCE, the movemtnt of detector probe
shall not exceed one (1) inch per second.
21 CFR 1030.10, Perf ormance Standard f or Microw a v e
Ovens.
It requires that the power density of the microwave
radiation emitted by a microwave oven shall not exceed five (5) milliwatts per square centimeter at any
point 5 centimeters (about 2 inches) or more from the
external surface of the oven.
All microwave ovens exceeding the emission level of 4
mW/cm2 must be reported to Dept. of Service for microwave ovens and the manufacturer immediately .
The owner should be told not to use the microwave
oven until it has been repaired completely.
If a microwave oven is found to operate with the door
open, report to Dept. of Service, the manufacturer and
CDRH* immediately. Also tell the owner not to use the
oven.
*CDRH: Center for Device and Radiological Health.
The interlock monitor switch acts as the final safety
switch protecting the customer from microwave radiation. If the interlock monitor switch operates properly
and the door interlocs switch fails, the fuse will blow.
If this happens, all interlock switches must be replaced.
The contacts of the interlock switches may be welded
together.
Power output Measurement ......................................1
Precautions and Repair Service Tips .......................1
Thermal protector(Magnetron) 150°C(270°F) Open
(Thermostat) 80°C(144°F)Close
Thermistor (Magnetron) 200°C(360°F) Open
108°C(194°F)Close
Thermistor(Duct)................120°C(216°F) Open
Fuse (Cartr idge Type) .................250V 10A
Micro switch, Relay
Interlock Switch
Interlock monitor Switch
Door sensing Switch and
Relay RL-3 and RL-4
Max. input time....................Electronic Digital, up to
Manual
Overall Dimensions........422(W)x540(D)x335(H) mm
Oven cavity size ............330(W)x330(D)x230(H) mm
Effective Capacity of Oven Cavity.........19.1liters
Net weight ...........................29Kg
10min./Memory 30min.
2. PO WER OUTPUT MEASUREMENT
NOTE: The power output specification, 1200W on this
model is measured with IEC measurement. The power
output is measured with two(2) liters water is equivalent to 1200W in measurement with IEC, when measured with the following power output.
(1)1. Fill two beakers, one liter of tap water respec tively
2. Use an accurate thermometer and measure each
water temperature respectively.
(2) Place beakers side by side in center of the
ceramic tray.
(3) Close the door,set the “TIME” for two minutes.
Touch the “START” key and heat the water for
exactly two minutes.
(4)Take the beakers out, immediately stir the water
and measure the water temperatures respectively.
(5) Calculate the temperatures rise of water in each
beaker . Then calculate the average value of the two
temperature rises.(Ģt)
(6) The teperature rise shall be in the following range;
Average Temp. Rise
Minimum 15.4°C
Maximum 18.8°C
Po w er output is affected by the line voltage under load.
(7) For correct Power output measurement, the line
voltage under load must be 120±2 Volts.
- 1 -
4. TEST PROCEDURES AND TROUBLESHOO TING
CAUTION
-DISCONNECT THE POWER SUPPLY CORD FROM THE WALL
OUTLET WHENEVER REMO VING THE CABINET FROM THE UNIT.
PROCEED WITH TESTS ONLY AFTER DISCHARGING THE HIGH
VOLTAGE CAPACITORS AND REMOVING THE LEAD WIRES ON
THE PRIMARY WINDING OF THE HIGH VOLTAGE TRANSFORMERS FOR LOWER AND UPPER MAGNETRONS.
(SEE FIGURE 3)
A. TEST PROCEDURES
COMPONENT
CHECKOUT PROCEDURERESULT
PRIMARY
WINDINGS
Secondary
Windings
Filament
Windings
Figure 2
1) Measure the resistance:
HIGH-VOLTAGE
TRANSFORMER
5. DISASSEMBY INSTRUCTIONS
F. CHANGING POWER SUPPLY CORD
(See exploded view on page 5)
(1) Unfasten 1 screw for ground and pull out the
2 wires of the power cord from the terminal plate.
(2) Remove 1 screw for the bottom bracket of the
cord bushing.
(3) Install the new power supply cord with the
reverse procedure of above (1) to (2).
With an ohm-meter on R x1 scale.
a. Primary winding;
b. Filament winding;
c. Secondary winding;
2) Measure the resistance:
with an ohm-meter on highest scale.
a. Primary winding to ground;
b. Filament winding to ground;
Normal reading:
Approximately 1.0 ohms
Less than 1 ohm.
Approximately 83 ohms
Normal reading:
Infinite ohms.
Infinite ohms.
Note: Remove varnish of measured point.
Figure 3
WARNING:
For changing the power supply cord, never use
other than the following.
Maintenance:
The microwave ovens are designed, manufactured, and tested for years of dependable operation. However, the
oven may require service from time to time if the consumable components listed below are not replaced at the
appropriate time. For protection from unexpected service calls and undue inconvenience, we recommend that
the user has the listed parts replaced at the inter vals below, (at customer cost).
This will avoid the trouble of repeated service calls after the expiration of the warranty period.
Consumable components:
When more than 1,250 hours of accumulative cooking time or more than 200,000 cycles of door opening/closing
is observed by key operations, the following consumable components will be replaced.
(Maintenance light in window display indicates when accumulative cooking time reaches 1,250 hours.)
Light
1. Magnetron Tube, Part No. 415 002 7702
2. Printed Circuit Board-Relay, Part No. 617 137 3844
3. Switch base Assembly, Part No. 617 205 1208
4. Door Latch, part No. 617 068 1087
When more than 2,000 hours of accumulative cooking time is observed, the following consumable compo-
nents should be replaced.
5. Blower motor, Part No. 617 140 1585
When slow rotating of blower motor is observed after removing dust from blower motor, blower motor must
be replaced.
6. Door hinge, Part No. 617 120 3028
When a worn door hinge is observed and proper door adjustments can not be made, the door hinge must
be replaced.
7. Door Assembly, Part No. 617 178 0734
When a worn door pin is observed and proper door adjustments can not be made, the door assembly must
be replaced.
- 4 -
Loading...
+ 14 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.