Gamry Interface 1000, Interface 5000 Quick Start Manual

Interface 1000
QUICK-START GUIDE
USB Potentiostat Installation
Guide #2
When possible, connect the Chassis Ground on the back of your potentiostat to a known, good earth ground.
Connect the cell cable to the Calibration side of the UDC4 dummy cell included with your instrument.
Place the dummy cell inside the Calibration Shield, close the lid, and connect the black oating-ground lead of your cell cable to the Shield’s grounding post.
Open Gamry Framework™. Select Experiment > Utilities > Calibrate Instrument
Choose your potentiostat. Select the radio button for calibration type Both, and click the OK button.
(Even though AC calibration for the Interface™ 1000 is performed at the Factory, be sure to click the Both radio button.)
After you conrm several instructional messages, the calibration proceeds automatically, and you are notied if the calibration is successful.
Did you receive a
CALIBRATION WARNING?
TROUBLESHOOTING A FAILED CALIBRATION
Calibration is used to check the potentiostat’s health, and to “zero” many of the measurement circuits to your laboratory environment. A warning does not necessarily indicate a critical failure, and Gamry can use calibration information to determine the source of the warning.
Double-check the following:
• Regular connections are to Calibration side of the UDC4.
• Floating-ground cable is connected to calibration shield.
• If possible, the Chassis Ground is attached to a known, good earth ground.
Click the Retry button, and the rest of the calibration restarts. Click the Ignore button for any other calibration warnings that may appear, and continue to Step 3.
After a failed calibration attempt:
Find Calibration Results PC6-#####.txt in your My Gamry Data folder.*
E-mail the le, along with complete contact information, to techsupport@gamry.com
* ##### is the serial number of your potentiostat.
WHAT DOES GAMRY SOFTWARE DO?
Gamry Framework
Potentiostat control for exible data aquisition. Select from standardized experiments grouped by research type, or use the Sequence Wizard to build complex automated experiments.
Echem Analyst
Quick and easy data analysis. Open data les with Echem Analyst for specialized analysis algorithms and high-quality plots. Customize, overlay, and scale plots, or export data.
My Gamry Data
The default data-folder location for Gamry Framework, with a shortcut on your desktop after installation. Change the folder location within Gamry Framework via Options > Path.
Virtual Front Panel
Software-based front panel for quick access to Gamry potentiostats’ functions, like a front panel of an early analog potentiostat; and to perform simple electrochemical experiments.
Electrochemical Signal Analyzer
TM
TM
TM
TM
TM
Designed specically for the acquisition and analysis of time-dependent electrochemical noise signals.
Resonator
Data-acquisition and -control software for the Gamry eQCM™. Contains a full suite of physical electrochemistry techniques.
Electrochemistry Toolkit
A sophisticated package for complete access to the capabilities of Gamry potentiostats in the software environment of your choice.
TM
TM
Calibration QSG: USB Potentiostat 1000 - 988-00024 rev. 2 - Gamry Instruments © 2016