Gait up Physilog 5 User Manual

User manual
Thank you for purchasing Physilog®5, the Swiss quality movement sensor. This manual introduces how to use the different features of the sensor.
Table of Contents
1.Physilog®5 elements...........................................................................................................................................2
2.Charging.................................................................................................................................................................2
3.Recording with Physilog®5...............................................................................................................................3
3.1 Standalone measurement............................................................................................................................3
3.2 Live Bluetooth stream of data....................................................................................................................3
3.3 Synchronous measurement with multiple Physilog®5......................................................................3
3.4 Extension board (coming in march 2017)...............................................................................................4
4.Data Analysis........................................................................................................................................................4
5.3 How to update Date and Time of the Physilog®5................................................................................7
5.4 How to update firmware of Physilog®5 with nRF Connect..............................................................7
5.5 Sensor specifications.....................................................................................................................................8
5.6 Battery autonomy estimations.................................................................................................................9
5.7 Troubleshoot...................................................................................................................................................10
4 jan.2017 - Physilog 5 user manual.pdf - v1.1 - Check www.gaitup.com for the latest version. Copyright �2017 Gait Up SA. CH-1015 SWITZERLAND 1
1. Physilog®5 elements
2. Charging The Physilog®5 is charged via the microUSB port. Plug the microUSB connector of the cable
delivered with the Physilog®5 to the Physilog®. Plug the USB connector to a Computer or Charger (see voltage specifications below). Warning: Be careful to plug the microUSB cable in the correct orientation into the Physilog®, forcing the connector may permanently damage the Physilog® and compromise its waterproofing.
Low battery level is indicated by orange blinking during measurement or if the Physilog® blinks rapidly 3 times red when starting and doesn't start the recording (see LED indications below). A battery level estimation is available in the mobile application. When the Physilog® is discharged, plug it for charging for at least 30 minutes before use. It is recommended to avoid complete discharging of the Physilog®5, therefore check that the Physilog® sensors are off when you store them. After a long period without using the Physilog®, plug it for charging for at least 30 minutes before use. It may happen that the LED doesn't blink orange at the beginning of charging after the battery was very low, to see the charging ongoing shortly push the main button or disconnect and reconnect after some charging time.
Supply Voltage specifications The Physilog® 5 should be charged using USB at 4.2V, 100mA. For charging, the ambient temperature must lie between 0°C and 45°C.
4 jan.2017 - Physilog 5 user manual.pdf - v1.1 - Check www.gaitup.com for the latest version. Copyright �2017 Gait Up SA. CH-1015 SWITZERLAND 2
3. Recording with Physilog®5 There are two ways to start and stop a measurement with the Physilog®5:
3.1 Standalone measurement
Press on the main button (for about 2 sec, until the green light turns off) to start the measurement. During measurement the Physilog® blinks green. Hold the main button for 3 seconds, until the light turns orange to stop the measurement. When the measurement has been concluded, the LED shortly blinks green three times.
3.2 Live Bluetooth stream of data
To receive a live Bluetooth stream of Physilog® data, start the measurement from the receiving device (f.ex. Using the Gait Up app on tablet). There is no need to pair your Physilog® with the tablet prior to Bluetooth communication, just select the Physilog® from the list in the app and launch the recording from the app. During measurement the Physilog® blinks green. To stop the measurement, click on “Stop measurement” in the app or hold the main button of the sensor for 3 seconds, until the light turns orange.
If the sensor is not detected by the tablet, shortly press the main button to enter detection mode. If the sensor can still not be connected, check that it has enough battery and plug and unplug the sensor to a computer.
The Physilog®5 works in “stream and log” mode: it saves all recorded data on its SD card, also when streaming the data in real-time via Bluetooth. This ensures that the data is not lost when the sensor gets out of Bluetooth range.
3.3 Synchronous measurement with multiple Physilog®5
The Physilog®5 sensors synchronize automatically at start when they are correctly configured. After starting, one of the sensors (Master) will start to blink green, the others (Slaves) shine green and don't record until they receive the synchronization signal from the Master sensor. Once they are synchronized, they blink green synchronously with the Master sensor and record. Slave sensors contain a file named “slave.txt”, the Master sensor doesn't have such a file. A sensor network can contain as many slave sensors as wished, but there must be a single Master sensor. To start Slave sensors without receiving the signal from the Master, shortly press the main button when it is shining green, the sensor will start the measurement without being synchronized to other sensors. Multiple sensor measurement start via Bluetooth command will soon be available.
4 jan.2017 - Physilog 5 user manual.pdf - v1.1 - Check www.gaitup.com for the latest version. Copyright �2017 Gait Up SA. CH-1015 SWITZERLAND 3
3.4 Extension board (coming in march 2017)
An extension board including a 3D magnetometer and a GPS sensor will be available from March 2017.
4. Data Analysis Physilog® always saves raw data on its internal memory, also when streaming the data live
via Bluetooth. Users have access to this raw data through companion software called Research ToolKit (RTK). The RTK is available for computer (PC and Mac) and Android. Functions to read raw data inside MATLAB® are also available. Please refer to the RTK user manual for more details.
USB data transfer Data saved on the SD card inside the Physilog® can be accessed using USB data transfer. Therefore plug the USB cable to the Physilog® and a computer. The Physilog®5 connects to the computer almost instantaneously and files can be accessed. Copy-paste the files to your computer for analysis. From here you can also delete the files from Physilog®'s memory. Warning: Do not remove the conf.bin file which contains the configuration information, otherwise the Physilog® will not record any data. If this happens, recreate a configuration using the Research ToolKit.
Bluetooth data transfer to Android mobile or tablet (app available 2017) With the Gait Up application installed on an Android mobile, you can stream data to your tablet or phone. Data is streamed to the device in real-time and displayed for easy check of successful data collection. When collecting data at high frequency (above 100Hz), the wireless data reception may encounter some delay and remaining data is transferred after the end of the measurement. Note: Bluetooth 4.0 or higher is required on the mobile device to be able to communicate with Physilog®5.
4 jan.2017 - Physilog 5 user manual.pdf - v1.1 - Check www.gaitup.com for the latest version. Copyright �2017 Gait Up SA. CH-1015 SWITZERLAND 4
5. Practical Considerations
5.1 Fixation
The Physilog®5 comes with a rubber clip as default attachment accessory (available from Feb
2017). Additional rubber clips and elastic straps with Velcro patches to stick onto the Physilog® can be purchased from Gait Up's online shop (shop.gaitup.com).
5.2 Handling / Maintenance
Physilog®5 should be handled carefully. In particular, it should not receive shocks, such as fall, crushing, being hit etc. Do not press too hard on the on/off button, this can damage the button. Warning: Do not use or charge the Physilog® if the sensor or its case is damaged. Contact Gait Up about what further action is needed when the case is damaged. In case the battery is damaged immediately move the device away from flammable materials and contact Gait Up's customer support.
- To benefit from the maximal battery life, fully charge Physilog® sensors before doing measurements (LED shines green when connected to computer or charger, you can also check with the app or RTK).
- Copy trial recordings to your computer: Physilog® is not the best place to keep and save important data.
- Physilog®5 can be used outside. It has the IP67 label, which means that it is waterproof up to a depth of 1 meter for 30 minutes. Do not use Physilog®5 to measure water-sport activities without additional waterproofing.
- Storage: Store in a cool (0°C – 30°C) and dry place. The provided box is the perfect place for your sensor.
- Cleaning: Sensors can be cleaned gently using alcohol wipes. Before you clean the device make sure it is turned off and disconnect all cables. Alcohol used in excess may damage the sensor’s waterproofing. Make sure to use wipes and not put liquid alcohol directly on the sensor.
- Sensors do not need yearly recalibration.
- Do not throw Physilog® sensors in the normal trash, dispose of it properly considering local laws and rules about electronic waste and batteries.
- Make sure to know which Physilog® file corresponds to which trial for your analysis – Last digits of file name are automatically incremented (exceeding 999 and firmware update
4 jan.2017 - Physilog 5 user manual.pdf - v1.1 - Check www.gaitup.com for the latest version. Copyright �2017 Gait Up SA. CH-1015 SWITZERLAND 5
resets the increment back to 1). Note that date of file generation (recorded between one start and stop of the Physilog®) is visible in the XLS or CSV file and the Matlab structure “header”. If you wish to adapt date and time, read section below.
- Do not remove the “conf.bin” file from the Physilog, otherwise it will create empty files (no sensor activated). If this happens, recreate a configuration in the Research ToolKit software.
Safety information
Physilog®5 includes a lithium battery. This battery may only be charged over a limited temperature
range. Never attempt to dock or charge your Physilog® when the temperature is outside of the range of 0 to 45°C.
Physilog®5 should be charged through USB connection with a computer. An external charger may be
used; note that no charger is provided by Gait Up. Gait Up declines any responsibility due to charger usage.
Supply voltage should be as follows: DC, 4.2V. Current consumption is 100mA. All external circuits
connected to the Physilog®5 should be «Safety Extra Low Voltage» and «limited Power Sources» circuits as described in the following standards: IEC60950-1:2005+/A1:2010 & EN60950­1:2006+/A11:2009+/A1:2010+/A12:2011
Warnhinweise:
Physilog®5 enthält eine Lithium Batterie. Die Batterie soll nur in einer definierten Temperaturspanne
aufgeladen werden. Versuchen Sie nie Ihren Physilog® an den Computer anzuschliessen oder aufzuladen, wenn die Aussentemperatur nicht zwischen 0°C und 45°C ist.
Sensoren sollen durch eine USB Verbindung mit einem Computer aufgeladen werden. Ein externes
Ladegerät kann benutzt werden; Gait Up empfiehlt und liefert aber kein externes Ladegerät. Gait Up übernimmt keine Haftung falls ein Ladegerät gebraucht wird.
Die erlaubte Netzspannung ist: DC, 4.2V. Stromaufnahme beträgt 100mA. Alle an den Physilog®5
angeschlossenen externe Stromkreise sollen die «Sicherheitskleinspannung” und “mit begrenzter Leistung” Regeln erfüllen, die in den folgenden Normen beschrieben sind: IEC60950-1:2005+/A1:2010 und EN60950-1:2006+/A11:2009+/A1:2010+/A12:2011
Informations relatives à la sécurité:
Physilog®5 contient une batterie au lithium. La batterie ne doit être chargée que si la température
extérieure est dans les limites définies. N’essayez jamais de connecter ou charger votre Physilog®5 si la température ambiante est en dehors des limites de 0 à 45°C.
Physilog®5 doit être chargé par une connexion USB avec un ordinateur. Un chargeur externe peut être
utilisé; mais Gait Up ne livre et ne recommande pas de chargeur. Gait Up refuse toute responsabilité liée à l’utilisation d’un chargeur externe.
La tension de réseau devrait être: DC, 4.2V. La consommation de courant est de 100mA. Tout circuits
connectés à Physilog®5 doivent être «très basse tension de protection» et «source à tension limité» comme décrit dans les normes suivantes: IEC60950-1:2005+/A1:2010 et EN60950­1:2006+/A11:2009+/A1:2010+/A12:2011
4 jan.2017 - Physilog 5 user manual.pdf - v1.1 - Check www.gaitup.com for the latest version. Copyright �2017 Gait Up SA. CH-1015 SWITZERLAND 6
5.3 How to update Date and Time of the Physilog®5
- Create a new text file on the Desktop of your computer or directly on the Physilog® sensor (right click, new>text file)
- Leave it empty and rename it to TIME.txt
- Copy-paste the file to the Physilog® if you created it on the Desktop
- The Physilog® date and time are updated to the hour when the file has been created
Create a new text file each time you update the date (TIME.txt file previously created will update to the date it was created, so in the past), and do not rename the existing TIME.txt. The TIME.txt file will disappear from Physilog's memory when disconnecting the Physilog® from the computer, therefore it is normal to not find it the next time you open the Physilog® on the computer.
5.4 How to update firmware of Physilog®5 with nRF Connect
The firmware is updated via Bluetooth. Therefore install the Android application “nRF Connect for Mobile” on a mobile or tablet and follow the steps below:
- Download the latest firmware version from the email to the Android device
- Open the nRF Connect app and enable all necessary permissions (Bluetooth and Position)
- Go to Scanner tab and scan for surrounding Physilogs (shortly push the main button of the Physilog®5 to enter detection mode)
- Select connect of one Physilog®5 sensor in the list
- In the top rigth corner of the app click on the DFU button (on the right of “Disconnect”) and select “Distribution Packet (ZIP)” from the proposed options
- Get the new firmware file from where it was downloaded (Some device may only allow to select from Dropbox)
- Wait until the graph in the app has shown that 100% of the firmware was transferred
- Disconnect the Physilog® by clicking on “Disconnect” and close the tab of the Physilog®5
- Plug the Physilog® to the computer and update the date and time (see above)
4 jan.2017 - Physilog 5 user manual.pdf - v1.1 - Check www.gaitup.com for the latest version. Copyright �2017 Gait Up SA. CH-1015 SWITZERLAND 7
5.5 Sensor specifications
Component Manufacturer Remarks Inertial Sensors STMicroelectronics 3D Accelerometer up to 16g
3D Gyroscope up to 2000°/s
Ambient Sensor STMicroelectronics Barometric altitude from 260 to 1260 hPa
Temperature sensor accuracy of +/-1.5°C
Radio Chip Nordic Semiconductors Multi-standard: Bluetooth Low Energy (BLE), Ant+,
and Near field communication (NFC) Internal Memory Apacer Class 10 microSD Card, 8Gb Microcontroller Nordic Semiconductors ARM® Cortex® M4 with floating-point for on-board
processing Micro-usb interface Amphenol FCI & Microchip Waterproof IP67, with dedicated chip for fast data
transfer. High-speed USB 2.0 Battery Renata Lithium Ion Polymer Accumulators 3.7V 140mAh Plastic Enclosure ABS Polylac® PA-757 Biocompatible with bi-color LED and 8mm button
4 jan.2017 - Physilog 5 user manual.pdf - v1.1 - Check www.gaitup.com for the latest version. Copyright �2017 Gait Up SA. CH-1015 SWITZERLAND 8
Operating Temperature
From-20° to 45°C
Assembly Locatis Electronics SA (CH) Weight 11 grams
Dimensions 26,5 x 47,5 x 10 mm Extension Board
(optional)
Ublox Melexis
Low-power GPS/GNSS module
High-quality 3D Magnetometer
Component Manufacturer Remarks
3D orientation 3D orientation of the sensor is calculated on-board during the measurement. A 3D model of
the Physilog®5 which indicates its orientation is shown in the app for illustration purposes. The Matlab function to read the .BIN file gives the quaternions and Euler angles at the output.
5.6 Battery autonomy estimations
Measuring channels Approx. Autonomy *
3D Accelerometer 3D Gyroscope Barometric pressure Hours
256Hz 256Hz 64Hz 20h 256Hz - - 24h
*Autonomy measured for firmware version 0.0.9 at room temperature
4 jan.2017 - Physilog 5 user manual.pdf - v1.1 - Check www.gaitup.com for the latest version. Copyright �2017 Gait Up SA. CH-1015 SWITZERLAND 9
5.7 Troubleshoot
LED indications
Situation Description Meaning Connected via USB Green continuous Fully charged
Orange blinking Charging Multicolor blinking Data transfer Red continuous Error 1
During measurement
Green blinking Measuring Green continuous Waiting for synchronization Orange blinking Low battery
Error indications 2 rapid red blinks, pause – repeated 3 times in total Error 2
3 rapid red blinks, pause – repeated 3 times in total Error 3 4 rapid red blinks, pause – repeated 3 times in total Error 4 5 rapid red blinks, pause – repeated 3 times in total Error 5 6 rapid red blinks, pause – repeated 3 times in total Error 6
Troubleshoot for error indications
Error Type* Troubleshoot Error 1 Disconnect and replug the Physilog® Error 2 Contact Gait Up Error 3 Charge the Physilog® Error 4 Contact Gait Up Error 5 Contact Gait Up Error 6 Empty the memory card
*see LED indication above
Soft reset The soft reset is the first manipulation which can be done if the Physilog® is not working
properly. To do a soft reset, hold the main button for 15 seconds. The LED shines orange and then stops. If the problem persists, fill out the contact form on the support page (www.gaitup.com/support2017) or write an email to: contact@gaitup.com. Please indicate Physilog® serial number and describe the problem in as much detail as possible.
4 jan.2017 - Physilog 5 user manual.pdf - v1.1 - Check www.gaitup.com for the latest version. Copyright �2017 Gait Up SA. CH-1015 SWITZERLAND 10
6.Limited Warranty & Support Policy
Warranty: Gait Up offers 12 months parts and labour on Physilog® starting from the date of delivery. If within one year from the date of delivery to the customer the equipment does not comply with the foregoing Limited warranty, Gait Up will at Gait Up’s option, repair, replace or refund the purchase price of the defective equipment free of charge to the customer. Customers requesting repair, replacement or refund are required to ship, the Physilog® to Gait Up. As a condition of this warranty, customers must contact Gait Up’s customer service for instructions on and approval of shipment prior to returning any defective Physilog®. The warranty shall not apply to any product or component thereof which has been repaired or altered by anyone other than Gait Up in any manner so as, in Gait Up’s judgement, to affect its service ability, or any product been subject to alteration, accident, misuse, abuse, neglect or abnormal wear. Gait up warrants solely to the original purchaser (customer). Only the terms expressed in this warranty shall apply and no distributor, corporation or individual is authorized to amend, modify or extend this warranty in any way. Gait Up shall have no liability for any consequential, incidental or special damages by reason of any act or omission or arising out of or in connection with the equipment or its rental, delivery, installation, maintenance, operation, performance or use, including without limitation any loss of use, lost revenue, lost profits or a cost associated with downtime. The obligations contained in this paragraph continue beyond the term of this limited warranty. Physilog® and Gait Analysis Software are not considered as proper Medical Devices, since they do not support directly diagnosis, but they provide data which have to be analysed and approved by medical doctors for them to make their diagnosis. Reclamations regarding medical devices will not be considered.
Support Policy: Support does not include:
support for 3rd party hardware, software, mailing lists or web content writing or debugging customer applications and deployments detailed explanations of the engineering principles behind our software and hardware support for systemic problems beyond the scope of the actual Physilog® and existing systems software
(IT issues, computer operation)
support for customers whose conduct fails to meet professional standards
Occasionally we engage in more interactive support or consulting. Interactive support is a customer courtesy and is provided at Gait Up’s discretion. It does not invalidate the support policy described above. There is no guarantee of performance, timeliness, or establishment of a continuous support relationship. Consulting is subject to acceptance of a formal statement of work.
4 jan.2017 - Physilog 5 user manual.pdf - v1.1 - Check www.gaitup.com for the latest version. Copyright �2017 Gait Up SA. CH-1015 SWITZERLAND 11
Contact
At Gait Up, we welcome your feedback and questions.
Please contact us at:
EPFL Innov’ Park - C
CH-1015 Lausanne tel: +41 21 633 7527
mail: contact@gaitup.com
web: www.gaitup.com
4 jan.2017 - Physilog 5 user manual.pdf - v1.1 - Check www.gaitup.com for the latest version. Copyright �2017 Gait Up SA. CH-1015 SWITZERLAND 12