REMOTE (SLAVE)OPERATION27V-B(a) Mode SelectionV-B(b) Range SelectionV-B(c) Peak Hold28V-B(d) DisplayFormat28V-B(e)Communications Format - IEEE-48829V-B(f) Communications Format - RS-232V-B(g)ClassifiersV-B(h) Setup Save-LoadV-B(i) Probe ZeroV-B(j) Probe RelativeV-B(k) MeasureV-B(I) LockoutCOMMAND STRING ORGANIZATIONCOMMAND EXECUTION34IEEE-488 IMPLEMENTATION
TABLE OF CONTENTS
101011
131313141416171717
181919
2425
25
2728
2930303131323333
35
1
4
555888
9
SERIES
9900
9000
SERIES
GAUSSMETERS
SECTION VI COMMUNICATIONS INTERFACE38
SECTION VII DIAGNOSTIC ERROR CODES44
SECTION VIII SERIES-9900 THEORY OF OPERATION48
BLOCK DIAGRAM51SECTION IX INSTALLATION OF OPTIONS52
APPENDIX53
WARRANTYbackcover
V-F RS-232 IMPLEMENTATION36
VI-A RS-232 PORT FUNCTION DESCRIPTION38VI-B RS-232 ELECTRICAL INTERFACE/HANDSHAKING39VI-C IEEE-488 BUS FUNCTIONAL DESCRIPTION41VI-D IEEE-488 ELECTRICAL INTERFACE43
.01 X Probe can not be used for measurements above 2Gauss
lxPROBE
RANGES RESOLUTION
GAUSS TESLA GAUSS TESLA
3 G 300 uT 1 mG 0.1 uT
30 G 3 mT 1 mG 0.1 uT
300 G 30 mT 10 mG 1 uT
3 kG 300 mT 100 mG 10 uT
30 kG 3 T 1 G 100 uT
300 kG 30 T 10 G 1mT
dc ACCURACY
Ran t % of Reading t Number of Counts
30 mG to 30 G .05 14
300 G to 3 MG .05 6
AdditionalInfluences;
Temperature Coefficient: <±(0.02% of Reading, ± 3 Counts) /C
dc Calibration Reference: ± 0.1 % of Reading
1 year; 23C, ± 5C
10 XPROBE
RANGES RESOLUTION
GAUSS TESLA GAUSS TESLA
30 G 3 mT 10 mG 1 uT
300 G 30 mT 10 mG 1 uT
3 kG 300 mT 100 mG 10 uT
30 kG 3 T 1 G 100 uT
300 kG 30 T 10 G 1mT
3 MG 300 T 100 G 10mT
From OC to +50 C
SERIES
9900
9000
SERIES
GAUSSMETERS
ANALOG OUTPUT
dc ANALOG OUTPUT ACCURACY
ac ANALOG OUTPUT ACCURACY
Output Voltage: 3.0 Vrms (ac Mode) or ±3 Vdc (dc Mode) FullScaleSource Impedance: <100ohmsTermination: Standard BNCConnector
30 mGto3G30 G
300 G to 3MG
For output >10% of FullScale
1 year; 23° C, ±5° C
AdditionalInfluences:
Temperature Coefficient:<±(0.03% of Reading, +0.005% FS) I ° C
Frequency(Hz)
20-4936%50-99
100-499
500-9.9 k5%
10 k-24 k
25 k -39 k
40 k-50 k
Range ±% of Reading ±% of FullScale
.254.0.250.3.250.1
From 0° C to + 50 ° C
±% of Reading ± °/0 of Full Scale ±% of Reading ± % of FullScale
30 mG to 30 G 300 G to 3MG
17%2.017%10%2.0
7%2.07%0.25
15%2.0
22%2.022%0.25
RANGES
2.036%0.25
10%
2.05%0.25
15%
0.25
0.25
0.25
FREQUENCY RANGE
(Analog Output)
OUTPUT NOISE, ac MODE
1 Year; 23° C, ±5° C
Sinewave input, >10% of FullScale
AdditionalInfluences:
Temperature Coefficeint: <± (0.04% of Reading, +0.005% FS)/ ° C
dcMode:
acMode:
Range rmsNoise30 mGto3G 300 mV30 G 30 mV
300 G to 3 MG 7 mV
From 0° C to +50° C
dc to 400 Hz20 Hz to 50kHz
2
SERIES
9900
9000
SERIES
GAUSSMETERS
FREQUENCY RANGE(Displayed Reading)
TEMPERATURE RANGEOperating: 0°Cto+50°C
FRONT PANEL DISPLAYType: 240x 64 dotmatrix LCD with a backlight
POWERVolts: 90-125 or 180-250
SIZE17.67 in. (44.9 cm) wide, 7.53 in.(19.1 cm) high including feet, 16.18 in. (41.1 cm) deep
WEIGHTNet: 35.9lbs. (16.3 kg)
(Maximum) Shipping: 43.4 lbs. (19.7 kg)
FRONT PANEL ZERO GAUSS CHAMBER
COMMUNICATION PORTSRS-232: Standard 25 pin "D" connector
WARM UP TIME1 hour to achieve rated specifications
HUMIDITY RANGE
dc Mode: dc
ac Mode: 20 Hz to 50kHz
Storage: -20° C to+70°C
Viewing Area: 5.2 in. (13.2 cm) wide, 1.6in.(4.1 cm) high
Frequency: 50-60 Hz 50-60 Hz
Current: 2.0 A 1.0 A
Cavity Size:
Attenuation:
IEEE-488: Standard 24 pin IEEE-488 connector
0°Cto+35°C 80% R H
+35°C to+ 50°C 70°/0 R H
0.344 in. (0.9cm) I.D., 2.0 in. (5.1 cm) deep80 dB in a 300 G field
3
SERIES
9900
9000
SERIES
GAUSSMETERS
1) Check the rear panel power receptacle to insure that the voltageshowingis the voltage you will be using, 115 volts ac or 230 volts ac. If it isnecessaryto change this setting, refer to Section II-A.
SIMPLIFIEDOPERATING
INSTRUCTIONS
FOR THE 9900SERIESGAUSSMETERS
2) With the power turned off, connect the required probes andprobemodules to the channels desiredvia the connectors on the front panel of the
gaussmeter. See Section II-B and II-C.
3) Connect power to the 9900 and turn on the instrument by locating thePower Switch on the front panel and depressing the side marked 1.
NOTE:At turn-on, the screen should display a large F.W. Bell logo.Beneaththe logo, various messages will indicate the progress of the initialization. If anyerror messages occur at this time, refer to Section VII, DiagnosticErrorCodes.
4)Assuming no error codes are present, the gaussmeter is now in the
MEASURE mode and will begin fieldmeasurements.
NOTE: All operating parameters such as range setting, peak/hold,classifier,display format, communication format, etc... will be initialized to the samesettingsthat were present when the gaussmeter was last turned off.
5) Prior to making accurate readings, it is necessary to zero each probe by
following the procedures below:
a) Press the ENTER/PROGRAM pushbutton. The MAIN MENU willappear.
b) Using the the up (T) and down (L) pushbuttons, advance the cursor to
PROBE ZEROfunction.
c) Press ENTER. The PROBE ZERO menu will appear.
d) Using the the up (T) and down (I) pushbuttons, advance the cursor tothe desired channel. PressENTER.
e) Install the probe tip in the ZERO GAUSS CHAMBER locateddirectlybelow the display. PressENTER.
f) The gaussmeter will proceed to zero the probe on each range setting foreither AC or DC operation, depending on the present modesetting.
g) Upon completion, the PROBE ZERO menu will reappear. Repeatstepsd-f for all remainingchannels.
h) Advance the cursor to RUN. Press ENTER to resume fieldmeasurements.
6)Consult Section IV for more information on MENU operations,or the
TABLE OF CONTENTS for references to other gaussmeterfeatures.
NOTE: Greatest accuracy will be obtained after a 1-hour warm upperiod.
4
SERIES
9900
SECTION I I-A INTRODUCTION
The SERIES-9900 gaussmeter represents the latest developments in thescience of measuring magnetic flux density using the Hall effect. The unit is
GENERALDESCRIPTION
available in three configurations capable of processing a singlechannel(Model 9901), two channels (Model 9902) or three channels (Model 9903) ofmagnetic field information. Either steady-state (dc) or alternating (ac) fieldscan be measured. When matched with the appropriate F.W. Bellfourth-generation Hall generator probe, fields as low as 10 pG (0.0011.1T) or as highas 2.9999 MG (299.99 T), at frequencies up to 50 kHz, can be measured withextreme accuracy and 4-3/4 digitresolution.
The SERIES-9900 features PEAK HOLD, AUTORANGING, CLASSIFIER andRELATIVE operation, auto ZEROing, GAUSS or TESLA readout, digital andbargraph representation, diagnostics and remote operation with an IEEE-488(GPIB) instrumentation bus and an RS-232 communications port.Allinformation is displayed on an illuminated graphics liquid crystal display(LCD).
The gaussmeter employs a menu-driven format to allow the user toprogramall aspects of gaussmeter operation with ease and speed. Eachgaussmeterchannel is completely and independentlyprogrammable.
I-B FUNCTIONAL DESCRIPTION
Figures I-A and I-B depict the SERIES-9900 front and rearpanels,respectively. The gaussmeter is housed in a standard 6.87" highx 17.67"
wide x 16.18" deep cabinet featuring pop-up feet for tabletop use and
(optional) brackets for rackmounting.
9000
SERIES
GAUSSMETERS
FRONT PANEL
(1)DISPLAY
(2) POWERSWITCH
(3) ZERO GAUSSCHAMBER
(4) "T" /CHANNEL 1"RESET"
PUSHBUTTON
5
1920-pixel (240 x 64) dot matrix(graphics)LCD with backlight. Contrast is preset at thefactory.
Rocker type power switch withinternationallegends ("0"=OFF,''1"=ON).
Recessed double-wall Mu-metalchamber
used to shield the probe from straymagneticfields during the PROBE ZEROoperation.Attenuation is 80 dB in a 300 G field.
Dual-function momentary-contactpushbutton switch used to resetPEAK
readings for CHANNEL 1intheMEASUREmode or advance the cursor up in theMENUmode.
SERIES
9900
9000
SERIES
GAUSSMETERS
SECTION I continuedGeneral Description
(5) "1" /CHANNEL 2 "RESET"
PUSHBUTTON
(6) "PROGRAM/ENTER"
PUSHBUTTON
(7) MODEL-99S PROBE
INTERFACE MODULE
(7-a) PROBE CONNECTOR
(7-b) ACTIVITY INDICATOR
(8) INTERFACE MODULE
CAVITY
Dual-function momentary-contactpushbutton switch used to reset PEAK
readings for CHANNEL 2 in the MEASUREmode or advance the cursor down in theMENU mode.
NOTE: Pressing both pushbuttons (4) and(5) in the MEASUREmode will reset thePEAK reading for CHANNEL 3.
Dual-function momentary-contact
pushbutton switch used to enter the MENUmode and validate selections during MENUoperations.
Modular electronics package that acceptsfourth-generation Hall generator probe. Unitinterfaces to gaussmeterviaa 64-pin dual-rowDIN connector.
9-pin twist-lock non-magnetic connector thatmates to F.W. Bell fourth-generation Hallgenerator probes.
Red LED indicating module/probe activity.
Accepts MODEL-99S interface module (7)or CAVITY blank panel(9).Cavity containsan upper and lower card guide to ensure thatthe interface module is properly aligned withthe rear mating connector.
WARNING!If acavity is not used it must be
covered with a SERIES-9900 blank panel(9)
to ensure proper ventilation and protectionfrom contaminates.DO NOT USE
CAVITY AS A STORAGE AREA.
(9) CAVITY BLANK PANEL
REAR PANEL
(10) POWER RECEPTACLE/
FUSE HOLDER/LINE VOLTAGE SWITCH
(11) VENTILATION FAN Regulates internal temperature of the
6
Used to cover an unused cavity(8). OrderF.W. Bell Item #338032.
This is a multi-purpose receptacle thataccepts an international instrumentationpower line cord. The middle (ground)contact is connected to the chassis. Thisreceptacle also contains the line fuse,storage spacefor a sparefuse and a linevoltage selector.
WARNING! See SECTION II before
applying power to the gaussmeter ordamage may result!▶
gaussmeter. DO NOT COVER !
SERIES
9900
9000
SERIES
GAUSSMETERS
SECTION I continuedGeneral Description
(12) RS-232PORT
CONNECTOR
(13) IEEE-488BUS
CONNECTOR
(14) ANALOGOUTPUTS
(15) INFORMATIONLABEL
Standard 25-pin "D" type femaleconnectorfor RS-232communications.
Standard 24-pin GPIB connector forIEEE-488bus interface.
Calibrated analog outputs fromeachchannel. Standard BNCconnectors.
Label identifies the model number andserial
number along with analog output(14)
identification and line power data.
0 0 00 0
E W.BELL
IM01111111111111111111111i11 I I I I I
SERIES Q900
= 1.1 1
= if
1
06 •-'
il!,
0 0 0
FigureI-A
SERIES-9900 FrontPanel
Figure I-B
SERIES-9900 RearPanel
7
0 0
SERIES
9900
SECTION II II-A POWER RECEPTACLE
WARNING! READ THIS CAREFULLY
The SERIES-9900 gaussmeter has been factory-configured to operate on aline voltage of 115 Vac. If operation at 230 Vac is required, perform the
following steps ordamagemayresult.
9000
SERIES
GAUSSMETERS
SERIES-9900
GAUSSMETERPREPARATION
1) See figure II-A. This is the line power cord receptacle designed toaccept
an international instrumentation line cord. This assembly also contains the
line fuse and voltage selector. Remove the linecord.
2) Near the top of the assembly is a slot. Insert a narrow screwdriver andgentlyrelease the access door (the door will not open unless the line cord is
removed). The door will swing down.
3) Remove the voltage selector drum (marked "115 VAC") by pullingstraightout.
4) Remove both fuse holders (marked with a " 4") by pulling straight out. Theleft-hand holder location is the actual in-circuit fuse. The gaussmeter is
shipped with a 230 Vac, 1 A fuse in the spare location, and a 115 Vac, 2 Afuse in the active location. Reverse the positions of the holders andreinstall,
making sure that the "--," points to theright.
5) Rotate the voltage selector drum to read "230 VAC" andreinstall.
6) Close access door. The "230VAC" legend should be visible through the
access door window. Reinstall the linecord.
ACCESSSLOT
115 Vac
Rear Panel PowerReceptacle
II-B MODEL-99S INTERFACE MODULE INSTALLATION
WARNING! DO NOT INSTALL AN INTERFACE MODULE WHILETHE GAUSSMETER IS TURNED ON OR DAMAGE MAY RESULT.
1) The MODEL-99S probe interface module can occupy any one of thethree
available slots in the SERIES-9900gaussmeter.
2) Remove the blank panel by rotating both screws counter-clockwise untilthey release from the chassis. The screws will remain with the blank panel. Donot discard the panel. It must be reinstalled if the module isremoved.
8
Figure II-A
SERIES
9900
9000
SERIES
GAUSSMETERS
SECTION II continuedGaussmeter Preparation
3) Locate the upper and lower card guides in the card rack. Install themoduleso that the upper and lower card edges of the module mate with thecardguides. Slowly slide the module in until the rear connectors mate. Firmlypushthe module into the chassis. Secure with the upper and lowerscrews.
II-C PROBE INSTALLATION
1) Install the probe connector so that the molded keys in the connector body
line up with similar keyways in the interface module's receptacle (Figure II-B).Push the connector in until the threaded sleeve on the connectormakescontact with the receptacle. Rotate the sleeve clockwise to secure theconnector to thereceptacle.
0 0}
PROBECONNECTOR
MODULERECEPTACLE
Probe Connector andReceptacle
II-D POWER-UP
Locate the power switch on the front panel and depress the side marked"1".The display's backlight should turn on and the "F.W. Bell" logo will appear.Beneath the logo various messages will appear to indicate the progress ofthe power-up initialization. Internal diagnostics are performed first, followedby retrieval of calibration data from each of the probe interface modules and
their probes. Finally, all display screens are constructed and variousoperating
parametersinitialized.
If any of the internal diagnostic tests fail, an error message will appear and thegaussmeter will halt the power-up procedure. See Section VII if thisoccurs.
If no aberations are detected, the MEASURE mode is entered and fieldmeasurements will begin. The activity indicator light on each probeinterfacemodule will light if a probe is connected. If no modules are presentor noprobes connected, an error message will appear.
All operating parameters, such as range settings, peakhold,classifiers,display format, communications, etc. will be initialized to the samecondition
they were in when the gaussmeter was last turned off. (See Section III-I.)
When the gaussmeter is turned off again it will save the present setup fornext time. The probe's zero and relative values are notsaved.
9
Figure II-B
SERIES
9900
SECTION III III-A DISPLAY FORMAT
In the MEASURE mode of operation the format of the display varies with the
number of probes present. The user can also modify the display format with
FIELDMEASUREMENTOPERATIONS
the DISPLAY FORMAT menu. (MENU mode is discussed in Section IV.)
For single-channel operation the entire display area is used as seen inFigureIII-A. For dual-channel operation, Figure 111-B, the display is split in half witheach half dedicated to one channel. For three-channel operation thedisplayis split into three sections, one for each channel, as seen in figure III-C.CHANNEL 1 is always the left-most channel in the chassis, CHANNEL 2 themiddle channel and CHANNEL 3 the right-most.
In all cases the display fields contain the same information.Refer to Figure III-D for the followingdiscussions.
III-B DIGITAL READING
This area 0 contains the reading's polarity, five digits ofinformationincluding a decimal point, and the scale (range) of thereading.
In the dc mode of operation the polarity will be "+" or "—" (or blank if thereading is exactly zero). In the ac mode a sinusoidal symbol will appear (-).
The mode of operation (dc or ac) is selected from the MODESELECTION
menu.
9000
SERIES
GAUSSMETERS
In the ac mode the reading represents thetrue rmsvalue of the field
waveform.
The user can remove the digital reading from the display using theDISPLAY
FORMAT menu. If this is done the digits and scale will be removed, but the
polarity symbol will remain to remind the user that a dc or ac field is being
measured.
If the CLASSIFIER function has been turned on, the digital field reading willbe replaced by, or alternate with, the phrase "LOW," "HIGH"or "ACCEPT."See Section IV-K for moreinformation.
11111111 I in, I I F JnuTin-RNGE:EDDNG D.67B9 G
PERK iii3E.GE G. 123.4SWG
1.1.112 12H
One- Channel DisplayScreen
18361 G113. 12KG
Two-Channel DisplayScreen
Figure111-A
AUTOPERK
AUTOPERK
Figure III-B
I I L
FAGE:3=4G 13.67139 cM+136.85 G. 123.4514c
RNGE:31:1014G 0.67Eg GM*136.8S G. 123.4514G
LIFZIMONIIIMMON
WIN
10
SERIES
9900
9000
SERIES
GAUSSMETERS
SECTION III continuedField Measurement Operations
Ill-C BARGRAPH
1 1r %.11LJa11 L, a if W40'1U If! trilMOMMOMFMMIO
AUTO—I:INGE:30014Si 0.6789 GPERK p 136.85 G. 123.45KG
IL now lui ffivalumt ul IN g meliffINMOREIIINJ
AUTO RNSE:300KG4_PERK p 136.85 651
motimuktnuliffivuNISMINMERWMMO
AUTO I:INGE:3001MT. 0.6789 GPERK p 136.85 Gia 123.46KG
Three-Channel DisplayScreen
S
Figure III-C
WIZEN LEE • IFEN
0.6789 G
123.45KG
..M71119
00
Figure III-D
Information Fields
The bargraph provides an "analog" representation of the present field
reading. Each bar on the graph represents 1/150th (0.667%) of full scale of
the presently-selected range. For instance, if the range is 3 kG (3000G),
each bar represents 20 G (3000/150=20). On the 300 µ-1- range, each bar
represents 2 µT.
The bargraph has a scale marker at every 5th bar. The 50th bar is marked with
a digit "1," the 100th with"2" and the 150thwith"3."
The user can remove the bargraph from the display using theDISPLAY
FORMATmenu.
III-D ACTIVE CHANNEL INDICATOR
The active-channel indicator ® appears only on the one-channel and two-
channel display. The indicator is a pictorial diagram of the SERIES-9900
gaussmeter with one of the channel locations blacked out. This reminds the
users that the information contained on the display is for the channelshown.
On the three-channel display, the top field is for CHANNEL 1(the left-mostchannel), the middle field is for CHANNEL 2 (middle channel) and the bottom
field for CHANNEL 3 (right-mostchannel).
11
SERIES
9900
SECTION III continued III-E INFORMATION BLOCKField Measurement Operations
The information block ® contains various annunciators andinformationabout the present state of the channelas follows:
9000
SERIES
GAUSSMETERS
"AUTO"ANNUNCIATOR
"PEAK"ANNUNCIATOR
RANGE SETTING The present range setting for thechannel.
RELATIVEOFFSET
CLASSIFIERS
III-F CHANNEL DEACTIVATION
In some cases the user may have several channels installed but may beinterested in observing only one or two. A channel can be deactivated byturning offboththe digital readingandthe bargraph via theDISPLAY
FORMAT menu, or by disconnecting the probe from the interfacemodule.This places the channel in a standby position and removes it from thedisplay.
For instance if all three channels are present and CHANNEL 2 isdeactivated,the gaussmeter will automatically reformat for a two-channeldisplay;CHANNEL 1 on the top half and CHANNEL 3 on the bottom half of thedisplay.
If the autoranging function has been turned on(RANGE SELECT menu), the"AUTO"annunciator willbepresent.
If the PEAK HOLD function has been turned on(PEAK HOLD menu), the "PEAK"annunciatorwill be present. The "PEAK" annunciatoris alsoused to indicate a particular overrangeconditionas described in Section li I-G.
This field contains the RELATIVE fieldreadingthat was present when the PROBERELATIVEoperation began (See PROBERELATIVEmenu). The reading is preceded by the"R"annunciator. If the RELATIVE functionis turnedoff this field will beblank.
These fields contain the user-definedCLASSIFIER settings preceded by the C/L andC/H annunciators. If the CLASSIFIERfunctionhas been turned off these fields will beblank.
In the REMOTE mode of operation, field measurements are notpossibleunless the deactivated channel is reactivated. This can be accomplishedfromthe remote device (see SectionV).
III-G OVERRANGE CONDITIONS
If the present field density exceeds the present range of thegaussmeterboth the digital reading and the bargraph will flash. Overrange occurs whenthe magnitude of the reading exceeds 29999 (2.9999 kG, 299.99T,etc.).Field readings will continue up to about 9% higher than this magnitude, or
32767.
In some instances the magnetic field being measured may contain spikes orripples that cause the present range to be exceeded even though theaveragereading is within limits. When this occurs the "PEAK" annunciator willflash to indicate that the displayed reading may be inaccurate and a higherrange shouldbeused.
12
SERIES
9900
9000
SERIES
GAUSSMETERS
SECTION III continuedField Measurement Operations
III-H PEAK HOLD OPERATION
When the PEAK HOLD function is engaged (via the PEAK HOLD menu) thelargest absolute field reading will be held on the display. For instance a+200.00 G reading will replace a +100.00 G reading and be held, and a
-250.00 G reading will replace the previously-held +200.00 Greading.
The user can reset a held reading at any time by pressingandreleasing thatchannel's RESET pushbutton on the front panel.As the pushbutton ispressed the ''PEAK" annunciator will flash to indicate that the resetcommandhas been recognized but the actual reset operation will not occur until thepushbuttonis released.
One pushbutton is dedicated to resetting CHANNEL 1, another forCHANNEL 2 and both for CHANNEL 3. In the case of a CHANNEL3reset,
the user might press one pushbutton before the other,say CHANNEL 1before CHANNEL 2. In this case the CHANNEL 1 "PEAK" annunciator willflash. As soon as the CHANNEL 2 pushbutton is pressed the CHANNEL 1annunciator will return to normal and the CHANNEL 3 annunciator willflash.This will not cause a problem since the actual resetting operation will notoccur untilbothpushbuttonsare released, properly resetting CHANNEL 3'speak reading only.
III-I POWER UP INITIALIZATION
The gaussmeter permanently stores each channel's MEASURE modesetup.When the gaussmeter is powered off and on again, the previous settingsarerestored and the gaussmeter is reinitialized to those settings. The followinginformationis saved:
• MODE (ac/dc ANDGAUSS/TESLA)
• RANGE SETTING (INCLUDINGAUTORANGE)
•PEAK HOLD ON/OFF (LAST PEAK READING WILL NOT BESAVED)
• CLASSIFIERSETTINGS
• CLASSIFIER ON/OFFSTATUS
• DIGITS ON/OFF
• BARGRAPH ON/OFF
• BACKLIGHT ON/OFF
• IEEE-488 PRIMARYADDRESS
• RS-232 PARITY, STOP BITS, CHARACTER LENGTH AND BAUD RATE
The RELATIVE mode will be turned off and the relative offset will be reset tozero.
III-J PROBE ORIENTATION VERSUS POLARITY
In the dc mode of operation the polarity of the reading versus theorientation
of the probe is depicted in Figure III-E. (Note the"F.W. BELL" logo is in anupright, right-reading orientation.)As shown, the magnetic flux linestravelingin the direction indicated by a "B" will result in a positive (+)polarity.
B
=Cr BELL
Figure III-E
Probe Orientation Versus ReadoutPolarity
13
C
SERIES
9900
SECTION III continued III-K ANALOG OUTPUTSField Measurement Operations
On the rear panel of the gaussmeter are connections that, whenconnected
to an oscilloscope,allow the user to observe the actual magnetic field
waveforms.
In the ac mode, the waveform is the actual field waveform calibrated to 3volts
rms full scale. For instance, with the gaussmeter programmed for a 3 kGrange
setting (3000 G) a 2.0000 Vrms signal correlates to a 2.0000 kG fielddensity.
In the dc mode, the output is instantaneously proprotional to the field in
magnitude and polarity from dc to 400Hz.
NOTE: The gaussmeter digitally corrects the magnetic field density signals for
errors due to probe offset, amplifier offset, frequency-related attenuation,
temperature-related effects, etc. before the final reading is displayed.The
signals available at the ANALOG OUTPUT connectors ARE NOT
CORRECTED for these errors.
III-L REMOTE TERMINAL ("MASTER") OPERATION
There are two modes of remote operation available in the SERIES-9900
gaussmeter. In the "SLAVE" mode, available from either the RS-232serial
port or the IEEE-488 bus, a remote device (typically a computer system or
industrial controller) controls all gaussmeter operations. This is discussed in
Section V.
9000
SERIES
GAUSSMETERS
In the "MASTER" mode, available only from the RS-232 serial port, the
gaussmeter controls a remote terminal or printer device. Thisis particularly
useful if the the user and gaussmeter must be separated by somedistance,
or if the user wishes to have a hardcopy of all gaussmeteractivity.
The terminal/printer may be connected directly to the gaussmeter, or
indirectly through a modem following the electrical connections shown in
Section VI. The RS-232 characteristics of both devices must match. The
gaussmeters' parameters can be modified via theCOMMUNICATIONS
FORMAT menu.All MEASURE mode and MENU mode operationsare
accessible from the terminal. See Section IV-M for furtherinformation.
Whenever field measurements are taking place the gaussmeter will transmit a
continuous listing of field measurements and channel status inaformatted
ASCII character stream as shown in Figure Ill-F. Note that CHANNEL 1 is
programmed for the RELATIVE mode, CHANNEL 2 has beendeactivated
and CHANNEL 3 has been programmed for CLASSIFIERoperation.
CHANNEL 1 has a 1X bulk Hall generator probe attached, CHANNEL 3 a 10X
thin filmdevice.
Initially, a FORMFEED (FF) control character is transmitted. If a printer isconnected this should cause the paper feeder to advance to the top of thenext sheet of paper. If a terminal is connected this usually clears thescreen.The FF is followed by CARRIAGE RETURN (CR) and LINEFEED (LF)controlcharacter and then a three-line header containing 79 printable characters perline. Each line is followed by a CR and LF.
Each line of measurement data contains 79 printable ASCIIcharactersfollowed by a CR and LF. The third (CHANNEL 3) information line is followedby one more LF to create a blank line between each groupofreadings.
14
SERIES
9900
SECTION III continued
Field Measurement Operations
F.W. BELL *** SERIES-9900 GAUSSMETER *** VERSION 1.0
9000
SERIES
GAUSSMETERS
Chan
Reading RelativeRangeClass-LOClass-HiClassifyAuto
1-1.6676 kG -55.338 G3kGOFF
INACTIVE
23+22.445 G30 G-5.1000 G+20.000 G
1-1.7288 kG -55.338 G3kG2INACTIVE3
+22.998 G30 G-5.1000G
1-1.7299 kG -55.338 G3kG2INACTIVE
+10.900 G
3
1
-1.7299 kG -55.338 GINACTIVE
2
+01.222 G
3
1-1.7299 kG -55.338 G3kG2INACTIVE3
-70.339 G
•••
30 G-5.1000G
3kGOFFOFF
30 G
30kG-5.1000 G+20.000 GLOWONOFFTHIN 10X
NOTE: Once 64 lines (including header and blank lines) have been
transmitted,theprocess starts over with the FFcharacter.
-5.1000G
+20.000 GHIGH
+20.000 GACCEPTONOFF
+20.000 GACCEPT
OFF
HIGHON
OFF
OFFOFF
OFFOFF
OFFONBULK 1X
ONOFFTHIN 10X
ONOFFTHIN 10X
PeakProbeType
ONBULK 1X
OFFTHIN 10X
ONBULK 1X
THIN 10X
ONBULK 1X
ONBULK 1X
F.W. BELL *** SERIES-9900 GAUSSMETER *** VERSION 1.0
Chan Reading Relative Range Class-LO Class-Hi Classify Auto Peak Probe Type
1 -1.7299 kG -55.338 G 3 kG OFF OFF ON BULK 1X2 INACTIVE3 +22.445 G 30 G -5.1000 G +20.000 G HIGH ON OFF THIN 10X
NOTE: The user should be aware that the serial transmission rateis muchslower than the rate at which the display is updated with new readings. Inother wordsthegaussmeter does not wait for a transmissiontocompletebefore taking another reading. Readings taken during transmissions willnever appear in a transmission, as the following scenarioillustrates:
Occasionally, the gaussmeter will inititate an internal calibration cycleinorderto maintain optimum performance. During this time, thephrase"GAUSSMETER BEING RECALIBRATED" will appear on the display and fieldmeasurement activity will be suspended. A recalibration willoccur:
a) Immediately after the power-up initializationsequence.b) Any time a new probeis connected.c) Whenever the internal temperature has changed more than ±5°Csince
the last recalibrationcycle.
16
Loading...
+ 46 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.